THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED “AS IS,” AND IS
SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER,
TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, AGILENT DISCLAIMS
ALL WARRANTIES, EITHER EXPRESS OR IMPLIED WITH REGARD TO THIS MANUAL
AND ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE
IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR
PURPOSE. AGILENT SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL
OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR
PERFORMANCE OF THIS DOCUMENT OR ANY INFORMATION CONTAINED HEREIN.
SHOULD AGILENT AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT WITH
WARRANTY TERMS COVERING THE MATERIAL IN THIS DOCUMENT THAT CONFLICT
WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT WILL
CONTROL.
DFARS/Restricted Rights Notice
If software is for use in the performance of a U.S. Government prime contract or
subcontract, Software is delivered and licensed as “Commercial computer software” as
defined in DFAR 252.227-7014 (June 1995), or as a “commercial item” as defined in
FAR 2.101(a) or as “Restricted computer software” as defined in FAR 52.227-19 (June
1987) or any equivalent agency regulation or contract clause. Use, duplication or
disclosure of Software is subject to Agilent Technologies’ standard commercial license
terms, and non-DOD Departments and Agencies of the U.S. Government will receive no
greater than Restricted Rights as defined in FAR 52.227-19(c)(1-2) (June 1987). U.S.
Government users will receive no greater than Limited Rights as defined in FAR
52.227-14 (June 1987) or DFAR 252.227-7015 (b)(2) (November 1995), as applicable in
any technical data.
ii
Safety Notes
The following safety notes are used throughout this document. Familiarize yourself
with each of these notes and its meaning before performing any of the procedures in
this document.
WARNINGWarning denotes a hazard. It calls attention to a procedure
which, if not correctly performed or adhered to, could result in
injury or loss of life. Do not proceed beyond a warning note
until the indicated conditions are fully understood and met.
CAUTIONCaution denotes a hazard. It calls attention to a procedure that, if not
correctly performed or adhered to, could result in damage to or
destruction of the instrument. Do not proceed beyond a caution sign
until the indicated conditions are fully understood and met.
Statement of Compliance
This instrument has been designed and tested in accordance with IEC Publication 1010, Safety
Requirements for Electronic Measuring Apparatus, and has been supplied in a safe condition. The
instruction documentation contains information and warnings which must be followed by the user to
ensure safe operation and to maintain the instrument in a safe condition.
Definitions
• Specifications describe the performance of parameters covered by the product warranty
(temperature –0 to 55 °C, unless otherwise noted.)
• Typical describes additional product performance information that is not covered by the
product warranty. It is performance beyond specification that 80% of the units exhibit
with a 95% confidence level over the temperature range 20 to 30 °C. Typical
performance does not include measurement uncertainty.
• Nominal values indicate expected performance or describe product performance that is
useful in the application of the product, but is not covered by the product warranty.
The Agilent Z5623A Option H48 Multiport Test Set is designed for use
with 50 Ω Network Analyzers such as the Agilent PNA Series (Models
E8356A, E8357A, and E8358A). Figure 1-1 shows a typical equipment
setup.
The test set reduces the time required to test multiport devices having
up to eight ports (distribution amplifiers, taps, switches, couplers, etc.).
It does this by reducing the number of device reconnects the operator
must perform. The test set can connect each of its test ports to any of
the following:
•Reflection Port
•Transmission Port
•50 Ω termination internal to the test set.
Switching is performed with mechanical switches.
The test set is controlled by means of its GPIB interface. The control
can be performed either by a PNA Series network analyzer or by an
external computer.
NOTE
This User's and Service Guide documents the use of the test set with an
Agilent E8358A network analyzer.
Figure 1-1 Typical Equipment Setup
Network Analyzer
Port 1
RF Cables
Reflection
Port
1
Z5623A Option H48
Test Ports
2
3
4
56
Port 2
Transmission
8
7
Port
1-2 Agilent Technologies Z5623A Option H48
Instrument Description
Specifications
Specifications
Agilent provides two different types of specifications for the test set:
• Guaranteed performance specs
• Typical performance specs
Typical performance specs have been benchmarked during product
development, but are not tested by the factory and are not guaranteed.
b. Band 1 (300 kHz to 1.3 GHz)
c. Band 2 (1.3 GHz to 3.0 GHz)
d. Band 3 (3.0 GHz to 6.0 GHz)
e. Band 4 (6.0 GHz to 9.0 GHz)
f.When using an external termination on the Reflection or
Transmission port
g.When using one of the 50 Ω internal terminations
h.From any test set port to the Reflection or Transmission port
i. Set by the power handling capability of the 50 Ω internal
a. Power measured at test ports 1-8 on Z5623A Option
H48 when connected to the E8358A using the
8120-4782 RF Jumper cables supplied.
b. Band 1 (300 kHz to 1.3 GHz)
c. Band 2 (1.3 GHz to 3.0 GHz)
d. Band 3 (3.0 GHz to 6.0 GHz)
e. Band 4 (6.0 GHz to 9.0 GHz)
f. IF Bandwidth set to 10 Hz
g. Conditions: 2 Windows, 4 Channels, 8 Traces,
2-port cal, no band crossings, 201points.
b
c
d
e
Agilent Technologies Z5623A Option H481-5
Instrument Description
Electrical Requirements
Electrical Requirements
The alternating-current (AC) power that is supplied to the test set must
meet the following requirements:
Voltage: 90 to 250 Vac
Frequency: 48 to 66 Hz
Available power: 40 watts minimum
If the available AC line voltage is outside the 90 to 250 Vac range, an
autotransformer that provides third wire continuity to earth ground
may be used.
Environmental Requirements
CAUTION
Operating Environment
Indoor use only
Operating temperature: 0 to 55 °C
Maximum relative humidity: 80 percent for temperatures up to 31 °C
decreasing linearly to 50 percent relative humidity at 40 °C
Altitude: up to 15,000 feet (4,572 meters)
Enclosure protection: IP 20, according to IEC 529
This product is designed for use in INSTALLATION CATEGORY II,
and POLLUTION DEGREE 2, per IEC 101 and 664 respectively.
Non-Operating Storage Conditions
Temperature: –40 °C to +70 °C
Humidity: 0 to 90 percent relative at +65 °C (non-condensing)
Altitude: 0 to 15,240 meters (50,000 feet)
1-6 Agilent Technologies Z5623A Option H48
Instrument Description
General Characteristics
General Characteristics
Weight
Net: Approximately 9 kg
Shipping: Approximately 20 kg
Cabinet Dimensions
These dimensions exclude front and rear panel protrusions.
89 mm H by 425 mm W by 500 mm D (3.5 in by 16.75 in by 19.7 in)
Miscellaneous Characteristics
RF connectors: Type N female
Switch type: Mechanical
I/O Control: GPIB
Available Options
UK6
Option UK6 provides a commercial calibration certificate including
actual test data. Data includes test results including reflection,
transmission, and crosstalk for all test ports.
Rack Ear Mounts
Option 908, part number 5062-3974, provides rack mounts that make it
quick and easy to install or remove the test set from a mainframe.
For further information on these options, please contact Agilent
Technologies. Refer to Chapter 10 , “Contacting Agilent.”
Agilent Technologies Z5623A Option H481-7
Instrument Description
Cleaning and Shipping Instructions
Cleaning and Shipping Instructions
Cleaning
Clean the instrument cabinet using a damp cloth only.
Shipping
Always transport or ship the instrument using the original packaging if
possible. If not, comparable packaging must be used.
1-8 Agilent Technologies Z5623A Option H48
2Installation
This chapter contains the following sections:
• Checking the Shipment
• Recommended Additional Equipment
• Familiarization with Safety Requirements
• Electrical Preparations
• Environmental Preparations
• Test Set Familiarization
• Setting the GPIB Address of the Test Set
• Connecting and Turning on the Test Set
Agilent Technologies Z5623A Option H482 -1
Installation
Checking the Shipment
Checking the Shipment
After the test set has been unpacked, keep the original packaging
materials so they can be used if you need to transport the instrument.
Check the items received against Table 2-1 to make sure you have
received everything.
Inspect the test set and all accessories for any signs of damage that may
have occurred during shipment. If your test set or any accessories
appear to be damaged or missing, call Agilent Technologies. Refer to
Chapter 10 , “Contacting Agilent.”
Table 2-1 Accessories Supplied with Z5623A Option H48
DescriptionAgilent Part NumberQuantity
Power CordSee Figure 2-2 on page 2-61
Front Handle Kit5063-92261
Rack Mount Kit5063-92321
RF Cable8120-47822
GPIB Cable .5MHP 10833D1
User’s and Service GuideZ5623-900161
2 -2 Agilent Technologies Z5623A Option H48
Recommended Additional Equipment
Recommended Additional Equipment
The equipment listed in Table 2-2 is not shipped with the test set, but is
recommended for use with it. The remainder of this User’s and Service
Guide assumes that the user has the listed equipment (or its
equivalent) available.
Table 2-2 Recommended Equipment Not Supplied With Test Set
DescriptionModel Or Part NumberQuantity
Installation
PNA Series Network
Analyzer
Electronic Calibration
Module
Keyboard with USB
cable
Mouse with USB cableHewlett-Packard Model No.
USB Quad HubPeracom Model No.
USB Cable
RF Cable (24”)
9/16” WrenchAgilent Part No. 8710-17701
3/4” Torque WrenchAgilent Part No. 8710-17661
Type N 50 Ohm Male
Short
Type N 50 Ohm Female
Short
Agilent Model No.
E8356A, E8357A, or E8358A
Agilent Part No. 85072-600071
Hewlett-Packard Model No.
SK-2502U
M-UB48
UH4000A
Peracom Model No. CA2000A
Agilent Part No. 8120-4781
Agilent Part No. 85032-600162
Agilent Part No. 85032-600152
a
a
a,b
a,b
a
1
1
1
1
3
Type N 50 Ohm Female
Termination
Type N 50 Ohm
Female-Female Adapter
Agilent Part No. 85032-600182
Agilent Part No. 85032-600211
a. Or equivalent
b. Available from Peracom Networks, Inc., Cary, North Carolina
(www.peracom.com).
Agilent Technologies Z5623A Option H482 -3
Installation
Familiarization with Safety Requirements
Familiarization with Safety Requirements
This document contains two types of safety notices: Warnings and
Cautions.
A Warning denotes a hazard that may endanger the operator.
A Caution denotes a hazard that may endanger the instrument.
Before proceeding to the Electrical Preparations described in the next
section, turn to Chapter 9, “Safety and Regulatory Information,” and
review the Warnings, Cautions, and safety markings that apply to this
instrument.
Electrical Preparations
1. Ensure that the “Electrical Requirements” on page 1-6 are met.
2. Verify that the power cable is not damaged and that the power
source outlet provides a protective earth ground contact. Note that
Figure 2-1 on page 2-5 depicts only one type of power source outlet.
Refer to Figure 2-2 on page 2-6 to see the different types of power
cord plugs that can be used with your test set.
Cables are available in different lengths. For descriptions and part
numbers of cables other than those described in Figure 2-2, call
Agilent Technologies. Refer to Chapter 10 , “Contacting Agilent.”
3. If this product is to be powered by autotransformer, make sure the
common terminal is connected to the neutral (grounded) side of the
ac power supply.
2 -4 Agilent Technologies Z5623A Option H48
Figure 2-1 Protective Earth Ground
Installation
Electrical Preparations
WARNING
This is a Safety Class I product (provided with a protective
earthing ground incorporated in the power cord). The mains
plug shall only be inserted into a socket outlet provided with a
protective earth contact. Any interruption of the protective
conductor, inside or outside the instrument, is likely to make
the instrument dangerous. Intentional interruption of the
protective conductor is prohibited.
Agilent Technologies Z5623A Option H482 -5
Installation
Electrical Preparations
Figure 2-2 Power Cables
Plug Type
250V
250V
125V
125V
250V
230V
220V
E
LN
E
LN
EE
NNLLEENNLL
EE
NNLLEENNLL
E
E
L
N
NL
E
LN
a
Number
8120-8705
8120-8709
8120-1369
8120-0696
8120-1378
8120-1521
8120-4753
8120-4754
8120-1689
8120-1692
8120-2104
8120-2296
8120-2956
Cable
Part
b
Plug
Description
Straight
BS 1363A
90
Straight
AS 3112
90
Straight
NEMA 5-15P
90
Straight
NEMA 5-15P
90
Straight
CEE 7/VII
90
Straight
SEV Type 12
90
Straight
SR 107-2-D
Length
cm (in.)
229 (90)
229 (90)
210 (79)
200 (78)
203 (80)
203 (80)
229 (90)
229 (90)
200 (78)
200 (78)
200 (78)
200 (78)
200 (78)
Cable
Color
Mint Gray
Mint Gray
Gray
Gray
Jade Gray
Jade Gray
Gray
Gray
Mint Gra y
Mint Gray
Gray
Gray
Gray
For Use
in Country
Option 900
United Kingdom, Hong
Kong, Cyprus, Nigeria,
Singapore, Zimbabwe
Option 901
Argentina, Australia,
New Zealand, Mainland
China
Option 903
United States, Canada,
Brazil, Colombia,
Mexico,Philippines,
Saudi Arabia, Taiwan
Option 918
Japan
Option 902
Continental Europe,
Central African Republic,
United Arab Republic
Option 906
Switzerland
Option 912
Denmark
8120-2957
250V
E
L
8120-4211
N
8120-4600
250V
E
N
8120-5182
L
8120-5181
a. E =earth ground, L = line, and N = neutral.
b. Plug identifier numbers describe the plug only. The Agilent Technologies part number is for the complete cable assembly.
90
Straight
IEC 83-B1
90
Straight
SI 32
90
200 (78)
200 (78)
200 (78)
200 (78)
200 (78)
Gray
Mint Gray
Mint Gray
Jade Gray
Jade Gray
Option 917
South Africa, India
Option 919
Israel
2 -6 Agilent Technologies Z5623A Option H48
Environmental Preparations
1. Ensure that the “Environmental Requirements” on page 1-6 are met.
2. If you are installing the test set into a cabinet, ensure there are at
least two inches of clearance around the sides and back of the test
set and the system cabinet. See Figure 2-3. The convection into and
out of the test set must not be restricted. The ambient temperature
(outside the cabinet) must be less than the maximum operating
temperature of the test set by 4 °C for every 100 watts dissipated in
the cabinet.
Figure 2-3 Ventilation Clearance Requirements
Installation
Environmental Preparations
CAUTION
If the total power dissipated in the cabinet is greater than 800 watts,
forced convection must be used.
Agilent Technologies Z5623A Option H482 -7
Installation
Environmental Preparations
3. Set up a static safe workstation such as the one shown in Figure 2-4.
Electrostatic discharge (ESD) can damage or destroy electronic
components.
Figure 2-4 Example of an Antistatic Workstation
2 -8 Agilent Technologies Z5623A Option H48
Test Set Familiarization
This section familiarizes the user with various front and rear panel
features of the test set.
Front Panel
Figure 2-5 Front Panel Features
Installation
Test Set Familiarization
REFLECTION
Port
POWER ON
Switch
POWER ON Switch
The POWER ON Switch turns the AC power to the test set on and off.
The switch is located at the bottom left corner of the front panel. See
Figure 2-5.
The switch disconnects the mains circuits from the mains supply after
the EMC filters and before other parts of the instrument.
PORT
CONNECTION
Status LCD
Test Ports 1 through 8
TRANSMISSION
Port
CONTROL
LINES
Connector
GROUND
Connector
CAUTION
REFLECTION Port, TRANSMISSION Port, and Test Ports 1–8
The REFLECTION Port, the TRANSMISSION Port, and Test Ports 1
through 8 are 50 Ω connectors that are used to connect to the device
under test.
Do not input more than 1 Watt maximum RF+DC to these ports or
damage to the internal RF switches or the analyzer may occur.
The GROUND Connector
The GROUND Connector provides a convenient front panel ground
connection for a standard banana plug.
Agilent Technologies Z5623A Option H482 -9
Installation
Test Set Familiarization
The PORT CONNECTION Status LCD
The PORT CONNECTION Status LCD displays the following:
• Which test port is connected to the REFLECTION port
• Which test port is connected to the TRANSMISSION port
• The status (O = “Open”, L = “Low”) of each of the five control lines
(C5 through C1)
All test ports not displayed on the Status LCD are internally
terminated in 50 Ω.
The CONTROL LINES Connector
The 9-pin CONTROL LINES Connector provides the user with a way of
controlling either the Device Under Test (DUT) or other equipment. The
test set cannot source any current from this connector. Therefore, the
user must provide an external power supply and some additional
circuitry if he wishes to use the connector. Refer to“Using the Control
Lines Connector” on page 7-6 for further details.
CAUTION
CAUTION
Do not apply more than +22 Vdc to pins 1 through 5 of the CONTROL
LINES Connector.
Do not employ external circuitry that forces the test set to sink more
than 250 mA on any control line (pins 1 through 5 of the CONTROL
LINES Connector) when that line is commanded to the “Low” state
(+0.63 Vdc).
2 -10 Agilent Technologies Z5623A Option H48
Rear Panel
Figure 2-6 Rear Panel Features
Installation
Test Set Familiarization
Not Used
GPIB
Connector
GPIB
Address
Line Module
Switches
GPIB Connector
This connector allows the test set to be controlled either by the network
analyzer or by an external controller.
GPIB Address Switches
The GPIB Address Switches set the GPIB address of the test set. See
“Setting the GPIB Address of the Test Set” on page 2-13 for
information.
Line Module
The line module contains the power cable receptacle and the line fuse.
Power Cables
The line power cable is supplied in one of several configurations,
depending on the destination of the original shipment.
Each instrument is equipped with a three-wire power cable. When
connected to an appropriate ac power receptacle, this cable grounds the
instrument chassis. The type of power cable shipped with each
instrument depends on the country of destination. See Figure 2-2,
“Power Cables,” on page 2-6 for the part numbers of these power cables.
WARNING
This is a Safety Class I product (provided with a protective
earthing ground incorporated in the power cord). The mains
plug shall only be inserted in a socket outlet provided with a
protective earth contact. Any interruption of the protective
conductor, inside or outside the instrument, is likely to make
the instrument dangerous. Intentional interruption is
prohibited.
Agilent Technologies Z5623A Option H482 -11
Installation
Test Set Familiarization
The Line Fuse
The line fuse (F 3 A/250 V, part number 2110-0780) and a spare reside
within the line module. Figure 2-7 illustrates where the fuses are and
how to access them.
Figure 2-7 Location of Line Fuses
2 -12 Agilent Technologies Z5623A Option H48
Setting the GPIB Address of the Test Set
A bank of five switches is used to set the GPIB address of the test set.
The switch bank is located on the rear panel of the test set and is shown
in Figure 2-6 on page 2-11. A diagram of the switch bank is presented
in Figure 2-8.
The switch bank sets the GPIB address in binary (base 2) format. Each
switch controls one bit of the address. To set a bit to 1, move the
corresponding switch to the up position. To set a bit to 0, move the
corresponding switch to the down position. The binary address shown
in Figure 2-8 is 01100.
Figure 2-8 GPIB Address Switches
ADDRESS:
HP-IB
168421
Installation
Setting the GPIB Address of the Test Set
Equation 2-1
1
0
12
PARALLEL
To convert the binary address to decimal (base 10) format, just multiply
each bit by the decimal value of that bit, then add the results. The
decimal value of each bit is shown above the corresponding switch.
Applying this conversion method to Figure 2-8 we obtain
016×0=
18×8=
14×4=
02×0=
02×0=
Sum12=
Therefore, the binary GPIB address (01100) shown in Figure 2-8 is
equal to 12 in decimal (base 10) format.
If all bits are set to 1, the GPIB address is 11111 in binary format, or 31
in decimal format. (16 + 8 + 4 + 2 + 1 = 31.) Therefore, the GPIB address
can have any value from 0 to 31 inclusive.
Agilent Technologies Z5623A Option H482 -13
Installation
Connecting and Turning on the Test Set
NOTE
The test set is shipped from the factory with the GPIB address set to 12.
Connecting and Turning on the Test Set
Make the connections shown in Figure 2-9, “Detailed Equipment
Connection Diagram.” The figure shows two RF cables connecting the
network analyzer to the test set. If you intend to verify the performance
of the test set, use the long RF cables listed in Table 2-3. Otherwise, use
the short ones.
Table 2-3 RF Cables Connecting Network Analyzer to Test Set
LengthPart No.Purpose
Long8120-4781Verification of test set performance
Short8120-4782Measurement of multiport devices
Plug power cords into the test set. Turn both instruments on.
NOTE
Allow the test set to warm up for at least 2 hours. For the most stable
and accurate measurements, leave the test set turned on at all times.
2 -14 Agilent Technologies Z5623A Option H48
Loading...
+ 108 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.