Agilent technologies Xcalibur Gemini User Manual

e
Xcalibur Gemini
Single Crystal Diffractometers
User Manual
Version 2.4, September 2011
Agilent Technologies XRD Products
10 Mead Road, Yarnton, Oxfordshire. OX5 1QU, UK Tel: +44(0)1865 291600 Fax: +44 (0)1865 291601
http://www.agilent.com/chem
Notices
CAUTION
WARNING
© Agilent Technologies, Inc. 2011
No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Agilent Technologies, Inc. as governed by United States and international copyright laws.
For Assistance and Support
http://www.agilent.com/find/assist
Limitation of Warranty
The foregoing warranty shall not apply to defects resulting from improper or inadequate maintenance by Buyer, Buyer­supplied software or interfacing, unauthorized modification or misuse, operation outside of the environmental specifications for the product, or improper site preparation or maintenance. No other warranty is expressed or implied. Agilent Technologies specifically disclaims the implied warranties of Merchantability and Fitness for a Particular Purpose.
Warranty
The material contained in this document
is provided “as is,” and is subject to
being changed, without notice, in future editions. Further, to the maximum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a particular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or performance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the separate agreement shall control.
Technology Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.
Restricted Rights Legend
If software is for use in the performance of a U.S. Government prime contract or subcontract, Software is delivered and
licensed as “Commercial computer software” as defined in DFAR 252.227­7014 (June 1995), or as a “commercial item” as defined in FAR 2.101(a) or as “Restricted computer software” as
defined in FAR 52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclosure of Software is subject to
Agilent Technologies‟ standard
commercial license terms, and non-DOD Departments and Agencies of the U.S. Government will receive no greater than Restricted Rights as defined in FAR
52.227-19(c)(1-2) (June 1987). U.S. Government users will receive no greater than Limited Rights as defined in FAR
52.227-14 (June 1987) or DFAR 252.227­7015 (b)(2) (November 1995), as applicable in any technical data.
Safety Notices
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.
Safety Summary
General Safety Precautions
The following general safety precautions must be observed during all phases of operation of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design, manufacture, and intended use of the instrument.
Agilent Technologies Inc. assumes no liability for the customer's failure to comply with these requirements.
Before operation, review the instrument and manual for safety markings and instructions. You must follow these to ensure safe operation and to maintain the instrument in safe condition.
General
This product is a Safety Class 1 instrument (provided with a protective earth terminal). The protective features of this product may be impaired if it is used in a manner not specified in the operation instructions.
Environment Conditions
This instrument is intended for indoor use in an installation category II, pollution degree 2 environment.
Refer to the specifications tables for the ac mains voltage requirements and ambient operating temperature range.
Before Applying Power
Verify that all safety precautions are taken. The power cable inlet of the instrument serves as a device to disconnect from the mains in case of hazard. The instrument must be positioned so that the operator can easily access the power cable inlet. When the instrument is rack mounted the rack must be provided with an easily accessible mains switch.
Ground the Instrument
To minimize shock hazard, the instrument chassis and cover must be connected to an electrical protective earth ground. The instrument must be connected to the ac power mains through a grounded power cable, with the ground wire firmly connected to an electrical ground (safety ground) at the power outlet. Any interruption of the protective (grounding) conductor or disconnection of the protective earth terminal will cause a potential shock hazard that could result in personal injury.
Do Not Operate in an Explosive Atmosphere
Do not operate the instrument in the presence of flammable gases or fumes.
Do Not Remove the Instrument Cover
Operating personnel must not remove instrument covers. Component replacement and internal adjustments must be made only by qualified personnel.
Instruments that appear damaged or defective should be made inoperative and secured against unintended operation until they can be repaired by qualified service personnel.
This product complies with the WEEE Directive (2002/96/EC) marketing requirements. The affixed label indicates that you must not discard this electrical/electronic product in domestic household waste.
Do not dispose in domestic household waste.
To return unwanted products, contact your local Agilent office, or see
www.agilent.com/environment/product/
for more information.
Environmental Information
Important Information
Product:
Xcalibur or Gemini
Model Configuration:
X-ray source: ENHANCE and/or ULTRA Detector: Eos, Atlas or Titan
Electrical Ratings:
1/N AC 230 V 50/60 Hz 4200 Watts
This user manual applies to the Xcalibur systems manufactured in Poland by Agilent Technologies.
Before attempting to operate the system, PLEASE READ THE INSTRUCTIONS.
This product should only be used by persons legally permitted to do so.
If the equipment is used in a manner not specified in the User Manual, the protection provided by the equipment may be impaired.
Important Health and Safety Notice
When returning components for service or repair it is essential that the item is shipped together with a signed declaration that the product has not been exposed to any hazardous contamination or that appropriate decontamination procedures have been carried out so that the product is safe to handle.
Care has been taken to ensure the information in this manual is accurate and at an appropriate level. Please inform Agilent Technologies if you have any suggestions for corrections or improvements to this manual.
Xcalibur service and support is available for technical and operational issues as indicated below.
E-mail: XRDSupport@agilent.com Phone: +44 (0)1865 291600 between 8 a.m. and 4.30 p.m. (UK time), Monday to Friday Fax: +44 (0)1865 291601
Agilent Technologies acknowledges all trademarks and registrations.
Copyright 2011 Agilent Technologies Limited. All rights reserved. No part of this document may be reproduced or distributed in any form, or by any means, or stored in a database or retrieval system, without prior written permission of Agilent Technologies.
Xcalibur/Gemini Version 2.4
USER MANUAL Page v
Contents
1. Health and Safety Information .............................................................................. 1
1.1 General ......................................................................................................................................................... 1
1.2 Electrical Safety .......................................................................................................................................... 2
1.2.1. Potential Electrical Hazards ........................................................................................................... 2
1.2.2. Recommended Precautions ........................................................................................................... 2
1.2.3. First Aid ............................................................................................................................................. 3
1.3 Mechanical Handling Safety ..................................................................................................................... 3
1.4 Safe Mechanical Practice ......................................................................................................................... 3
1.5 Moving Parts ............................................................................................................................................... 4
1.6 Extreme Temperatures .............................................................................................................................. 4
1.7 X-ray Radiation ............................................................................................................................................ 4
1.8 Vacuum ........................................................................................................................................................ 5
1.9 High Pressures ............................................................................................................................................ 5
1.10 Hazardous or Toxic Materials ................................................................................................................ 6
1.11 Maintenance ............................................................................................................................................. 6
2. Introduction ............................................................................................................. 7
2.1 Scope ............................................................................................................................................................ 7
2.2 How To Use This Manual .......................................................................................................................... 7
2.3 System Overview ........................................................................................................................................ 7
3. Specifications ......................................................................................................... 8
3.1 Environmental Requirements ................................................................................................................... 8
3.2 Services ........................................................................................................................................................ 8
3.2.1. Electrical Supply .............................................................................................................................. 8
3.2.2. Water Cooling .................................................................................................................................. 9
3.2.3. Helium Gas Supply For Ultra .......................................................................................................... 9
3.3 Performance Data ..................................................................................................................................... 10
3.3.1. X-ray Tube (Typical Operating Conditions) ............................................................................... 10
3.3.2. CCD Detector .................................................................................................................................. 10
3.3.2.1 General .......................................................................................................................................... 10
3.3.2.2 Eos CCD Detector ........................................................................................................................ 11
3.3.2.3 Atlas CCD Detector ..................................................................................................................... 11
3.3.2.4 Titan CCD Detector ..................................................................................................................... 11
3.3.2.5 PC CCD Interface ......................................................................................................................... 11
3.3.2.6. CCD Detector Maximum Theta and Resolution .................................................................... 12
3.3.2.7. CCD Detector Maximum Distance ........................................................................................... 13
3.3.3. Four-circle Kappa Geometry X-ray Goniometer ........................................................................ 14
3.4 Electrical Data ........................................................................................................................................... 14
4. Technical Description .......................................................................................... 15
4.1 Overview of Xcalibur/Gemini ................................................................................................................. 15
4.2 KMW200CCD Chiller and KMW3000C Chiller...................................................................................... 18
4.3 Low Temperature Option ......................................................................................................................... 18
4.4 Safety Features ......................................................................................................................................... 18
5. Handling, Installation, Storage and Transit Information ................................. 20
Xcalibur/Gemini Version 2.4
USER MANUAL Page vi
5.1 Reception and Handling .......................................................................................................................... 20
5.1.1. Delivery ............................................................................................................................................ 20
5.1.2. Unpacking ....................................................................................................................................... 20
5.1.3. Mechanical Handling .................................................................................................................... 21
5.2 Installation and Setting to Work ............................................................................................................ 22
5.2.1. Preparation of Site and Services ................................................................................................ 22
5.2.2. Setting to Work .............................................................................................................................. 25
5.3 Storage ....................................................................................................................................................... 25
6. Operation ............................................................................................................... 26
6.1 Controls and Indicators ........................................................................................................................... 26
6.2 System Start-Up ........................................................................................................................................ 28
6.3 X-ray Tube Warm-up Procedure ............................................................................................................ 29
6.4 System Standby and Normal Shutdown Procedure ........................................................................... 29
6.5 Emergency Shutdown .............................................................................................................................. 30
6.5.1. Emergency Shutdown Procedure ............................................................................................... 30
7. Maintenance Schedules ...................................................................................... 31
7.1 Introduction ............................................................................................................................................... 31
7.2 Weekly Maintenance Schedule ............................................................................................................. 31
7.3 Monthly Maintenance Schedule ............................................................................................................ 31
7.4 Six Monthly Maintenance Schedule ..................................................................................................... 32
7.5 Yearly Maintenance Schedule ................................................................................................................ 32
7.6 10,000 Hours Maintenance Schedule ................................................................................................... 32
8. Maintenance Instructions ................................................................................... 34
8.1 Refining the Machine Parameter File ................................................................................................... 34
8.2 Changing the X-ray Tube of Ultra (Horizontal X-ray tubes) ............................................................... 35
8.3 Changing the X-ray Tube of Enhance (Vertical X-ray Tubes)............................................................ 37
8.4 Changing the Collimator of Enhance ..................................................................................................... 38
8.5 Aligning the X-ray Collimator of Enhance ............................................................................................ 39
8.6 Aligning the Video Microscope .............................................................................................................. 39
8.7 Checking the Door Safety Interlocks ..................................................................................................... 41
8.8 Checking the Emergency stop ................................................................................................................ 41
8.9 Checking the X-ray Radiation Levels ..................................................................................................... 42
8.10 CCD Detector – Pumping Out Vacuum ............................................................................................... 42
9. Trouble Shooting .................................................................................................. 45
10. Spares .................................................................................................................. 52
10.1 Fuses......................................................................................................................................................... 52
10.2 Bulbs ......................................................................................................................................................... 52
11. Disposal Instructions ......................................................................................... 53
11.1 X-ray Tube and CCD Detector .............................................................................................................. 53
11.2 Third Party Equipment ........................................................................................................................... 53
12. Additional Information ....................................................................................... 54
12.1 Third Party Information.......................................................................................................................... 54
12.2 Drawings .................................................................................................................................................. 55
12.2.1. Mechanical Drawings ................................................................................................................. 55
Xcalibur/Gemini Version 2.4
USER MANUAL Page vii
12.2.2. Electrical Drawings ..................................................................................................................... 57
Table of Figures
Figure 4.1.1 Components of a typical Xcalibur/Gemini system ..................................................................... 15
Figure 4.1.2 View of Xcalibur Ultra diffractometer ........................................................................................... 16
Figure 4.1.3 View of Gemini diffractometer ....................................................................................................... 17
Figure 4.1.4 Goniometer phi, kappa, omega and theta axes ........................................................................... 18
Figure 6.1.1 Location of switches ........................................................................................................................ 27
Figure 8.6.1 Goniometer Head Adjustments ...................................................................................................... 39
Figure 8.6.2 Video Microscope Adjustments ..................................................................................................... 39
Figure 8.7.1 View of the access plate and pump out port of the Eos CCD detector .................................. 43
Figure 8.7.2 View of the pump out port of the Atlas/Titan CCD detector.................................................... 43
Figure 12.2.1 Xcalibur / Gemini Suggested Layout .......................................................................................... 55
Figure 12.2.2 System and Component Dimensions .......................................................................................... 56
Xcalibur/Gemini Version 2.4
USER MANUAL Page viii
HEALTH AND SAFETY INFORMATION
WARNING
Indicates a potential hazard which may result in injury or death
CAUTION
Indicates a potential hazard which may result in damage to equipment
Protective conductor terminal
Earth (ground) terminal
CAUTION Risk of electric shock
CAUTION Refer to accompanying documents
WARNING Radiation Hazard
CAUTION Risk of finger squeeze between moving parts
CAUTION
Do not take risks. You have a responsibility to ensure the safe condition and safe operation of equipment.
1. Health and Safety Information
1.1 General
In normal operation the system is designed to operate safely. All users of Xcalibur should be aware of potential hazards which exist in and around equipment of this type and the ways of avoiding possible injury and equipment damage which may result from inappropriate ways of working. A description of such potential hazards and how to avoid them is given in this section.
This manual adopts the following convention:
Warning symbols on the equipment are:
See original manufacturers' manuals for further safety data on third party equipment supplied with the system. A list of these is given in this manual.
Xcalibur/Gemini Version 2.4
USER MANUAL Page 1
HEALTH AND SAFETY INFORMATION
WARNING
The equipment should only be operated and maintained by authorised operators of the system. An authorised operator is a person who has undergone specialist radiation training and has been trained in the use of Xcalibur by Agilent Technologies personnel.
WARNING
All of the electrical equipment supplied as part of the system should be provided with a protective ground. Do not remove protective grounds as this may give rise to an electrical safety hazard. It is vitally important that the system is properly grounded at all times.
Follow local and national electrical regulations and procedures.
Do not defeat interlocks, remove connectors, disconnect equipment, open safety covers, dismantle or modify equipment unless you are qualified and authorised to do so and you are fully conversant with its operation and potential hazards, or have total assurance through your local electrical permit to work system that the equipment has been made safe.
Ensure that the mains supply is fused at an appropriate rating, or fitted with a circuit breaker, and that it can be isolated locally via a clearly labelled, clearly visible and easily accessible isolating switch. Isolate the supply before carrying out any maintenance work.
Never switch off the CCD detector’s power supply when the KMW200CCD is
operational. To switch off the CCD detector use the KMW200CCD’s front panel power switch.
1.2 Electrical Safety
In normal use the user is protected from the dangers associated with the voltage, current and power levels used by the equipment. Only personnel qualified to work with the voltages and currents used by this equipment should attempt to disconnect, dismantle or modify the equipment.
1.2.1. Potential Electrical Hazards
The following list is not intended as a complete guide to all the electrical hazards on the system, but serves to illustrate the range of potential hazards that exist:
electric shock electric burn fire of electrical origin electric arcing
1.2.2. Recommended Precautions
Xcalibur/Gemini Version 2.4
USER MANUAL Page 2
HEALTH AND SAFETY INFORMATION
WARNING
Do not touch any unshielded wires or connectors while mains power is supplied to the system.
Do not allow water or any other foreign objects to come into contact with the equipment’s electrical components.
WARNING
Do not attempt to administer first aid to someone who may have suffered electric shock until the source of the shock has been isolated. Mains voltages are present in the system. High voltages are used by the X-ray tube and power supply. These can cause serious injury or death. Only personnel qualified to work with high voltages and currents should perform service or maintenance work on such equipment.
WARNING
Lifting points are provided for safe handling of components and safe handling practice must be observed to comply with local regulations. Check that lifting points are used only for the job intended. The system itself and some components are heavy and require careful handling. Use safe lifting procedures for heavy items to prevent possible strain injury.
1.2.3. First Aid
A course in first aid to include methods of artificial respiration is recommended for those whose work involves equipment that may produce a high voltage.
1.3 Mechanical Handling Safety
1.4 Safe Mechanical Practice
In normal use personnel are not required to undertake mechanical work. However, servicing or repair may necessitate access to any part of the system. Only personnel who have been trained by Agilent Technologies to carry out service work on this equipment should attempt to dismantle, modify or repair the equipment.
Water connections should be made and tested in accordance with any local and national safety regulations.
Xcalibur/Gemini Version 2.4
USER MANUAL Page 3
HEALTH AND SAFETY INFORMATION
WARNING
Injury could result if clothing or body parts become caught in moving mechanisms. Keep clothing, hands and body parts away from moving mechanisms.
WARNING
Between the moving parts of the goniometer there are a number of places at which a potential finger squeeze or shearing hazard exists. The warning triangle shown to the left is visible on the goniometer close to these places. Keep hands away from the goniometer when parts are moving.
WARNING
1. Systems fitted with the low temperature option use liquid nitrogen and/or
liquid helium as a coolant. Liquid nitrogen and liquid helium are cryogenic liquids and can cause cold burns. Wear gloves when handling cryogenic liquids and use eye protection. Refer to the information supplied with the equipment for more information.
2. During operation both the X-ray tube and the heat sink of the CCD power
supply become hot. In normal use they are located inside a cabinet and hot parts are not accessible. During maintenance periods, however, it may be necessary to open the cabinet so that adjustments can be made. Therefore great care must be taken to avoid touching either the X-ray tube or the heat sink of the CCD power supply when they are operating and for a period of 20 minutes after operation.
WARNING
This equipment contains an X-ray tube. Ensure that safe working practices relating to radiation are employed. Follow any local, national or international rules and guidelines. Intentional or reckless misuse of the X-ray generator or its safety devices including safety interlocks and cabinet shielding can result in serious injury or even death.
1.5 Moving Parts
There are a number of moving parts in the system which are powered by electric motors.
1.6 Extreme Temperatures
1.7 X-ray Radiation
During operation, there is an acceptable level of X-ray radiation as based on the recommendations on risk published by the International Commission of Radiological Protection (ICRP) and endorsed by the Health Protection Agency (HPA) in the UK. For use in the UK, the Ionising Radiations‟ Regulations
Xcalibur/Gemini Version 2.4
USER MANUAL Page 4
HEALTH AND SAFETY INFORMATION
WARNING
To prevent injury to personnel and possible damage to the equipment, please note the following guidelines:
1. Only authorised personnel who have received appropriate instruction and are
aware of the laboratory rules that govern the use of this type of system should operate the system.
2. Never dismount the beam stop when the system is operational.
3. Do not operate the system without the collimator, unless performing the beam
alignment procedure.
4. Use appropriate X-ray detection equipment to perform regular radiation
checks as per any laboratory rules
5. Use only genuine hardware X-ray tubes, X-ray generators, monochromators,
goniometer heads and collimators, as recommended by Agilent Technologies. Use of other products may compromise the performance of the shielding and safety system, and may invalidate your warranty.
WARNING
When handling and using X-ray tubes and the CCD detector, particular care should be taken to avoid injury caused by possible implosion of the vacuum tube. Wear eye protection.
WARNING
Know the law about high pressure gas cylinders and follow it. High pressure cylinders are often used to store gases (typically at pressures up to 200 bar). Most countries have laws about using them.
Chain cylinders to a fixed object or keep them in specially designed trolleys
Only use approved and tested high pressure fittings
1999 should be adhered to. For countries outside the UK the appropriate laws apply such as registration and inspection.
Customers should be aware of their duty of safety to their employees and visitors.
1.8 Vacuum
1.9 High Pressures
Xcalibur/Gemini Version 2.4
USER MANUAL Page 5
HEALTH AND SAFETY INFORMATION
WARNING
If Beryllium is exposed to fire, it may oxidise to highly toxic beryllium oxide powder. Do not attempt to clear up the remains of any fire, but contact the relevant local agency stating that there is an incident involving possible beryllium or beryllium oxide contamination.
1.10 Hazardous or Toxic Materials
Beryllium and beryllium oxide are toxic materials. Follow appropriate handling, shipping, use, storage and disposal procedures and regulations. Refer to BrushWellman Material Safety Data Sheet No. M10 for further information.
1.11 Maintenance
The manufacturer will not be held responsible for the safety, reliability or performance of the equipment unless assembly operations, extensions, re-adjustments, modifications and repairs are carried out only by persons authorised by the manufacturer. Interchangeable parts which are subject to deterioration during operation must be serviced or interchanged during the intervals given.
Xcalibur/Gemini Version 2.4
USER MANUAL Page 6
INTRODUCTION
2. Introduction
2.1 Scope
This manual applies to the Xcalibur system designed and manufactured by Agilent Technologies.
2.2 How To Use This Manual
This manual is aimed at operators and maintenance personnel of the Xcalibur system. Operators of the system should be computer literate, familiar with X-ray diffraction techniques, have had training in the use of the Xcalibur system by Agilent Technologies staff, and have had training about radiation safety.
This manual is intended to provide operators with a practical guide to the system and its operation. This is intended to familiarise the user with how the system works and provide a better understanding of the system operation.
All personnel who are likely to operate the system or who are likely to come into contact with any of the system components should read the HEALTH AND SAFETY INFORMATION section of the manual. This provides basic information aimed at highlighting the safety hazards associated with the equipment.
More detailed information and instructions for component parts of the system are given in the third party manuals supplied with the system, which are listed in this manual. These manuals should also be read and understood before operating the system.
The purpose of this manual is to:
explain how to operate the equipment explain how to interface to the equipment list performance characteristics of the equipment describe how the equipment operates assist with simple fault finding and maintenance
2.3 System Overview
The Xcalibur series are single crystal diffractometers that use the property of X-ray diffraction to determine the crystal structure of materials. They are intended for use with single crystals of chemical substances (inorganic, organic or organo-metallic), mineralogical and biological samples. The Xcalibur may also be used in the analysis of powder samples. Intended samples should have a maximum unit cell dimension of 100 Ångstrom in any direction for small molecule diffractometers and 500 Ångstrom for macromolecular systems.
The Xcalibur may be used with crystal conditioning devices. Specifically, low and high temperature attachments and high pressure cells. Some minor modifications may be required by Agilent Technologies to enable use of these devices.
Xcalibur/Gemini Version 2.4
USER MANUAL Page 7
3. Specifications
Ambient temperature during operation
18 – 28 C
Stability of ambient temperature during operation
1 C
Storage temperature
>10 C <40 C
Relative humidity
20 - 80 % non – condensing
Location
Clean, dust free environment >2m from air conditioning or heating units
Floor covering
Conductive or, if carpeted, covered with electrostatic mats
Floor strength
Able to bear system weight of ~500 kg over an area of
0.64 m2
Total power requirements
230 V 10 % (or 208V for USA) for X-ray generators and diffractometer
240, 220 or 110V (depending on region) for temperature attachments and computer monitor.
63 A
All single phase supplies should be connected to the same phase.
Number of outlets required
1 single-phase outlet (16 A) for diffractometer system (comprising goniometer, interface, water chillers and computer etc.)
1 single-phase outlet (32 A) for each X-ray generator
4 single-phase outlets (15 A total) for temperature attachments and computer monitor.
3.1 Environmental Requirements
SPECIFICATIONS
3.2 Services
3.2.1. Electrical Supply
Xcalibur/Gemini Version 2.4
USER MANUAL Page 8
Voltage fluctuation
< 10 % (with line voltage regulator fitted if necessary)
Location of outlets
On wall behind system (except computer monitor supply which should be near intended computer table location)
Protection
Circuit breaker to be fitted to all electrical outlets
3.2.2. Water Cooling
Maximum flow rate
1.8 l/min
Pressure
0.5 – 3 bar gauge
Temperature stability
5 C
Temperature range
5 – 18 C
Composition
Refer to third party manual provided by X-ray tube manufacturer.
Purity
99.99%
Regulator
low pressure regulator
Minimum pressure
0.5 bar gauge
Flow rate
30 - 40 cm3/min @ ATP
SPECIFICATIONS
3.2.3. Helium Gas Supply For Ultra
Xcalibur/Gemini Version 2.4
USER MANUAL Page 9
3.3 Performance Data
Tube
Voltage (kV) setting
Current (mA) setting
Resulting power (kW)
Cu 2.2kW tube
50
40
2.0
Mo 3kW tube
55
40
2.2
Maximum radiation dose due to scattering at outside surface of the enclosure (door closed, shutter open)
<0.5 µSv/h (with Mo 3 kW tube operating at 2.2 kW) Maximum radiation dose due to scattering (door open, with shutter closed)
<0.5 µSv/h (with Mo 3 kW tube operating at 2.2 kW) CCD chip
Kodak KAF4320-E
Scintillator material
Gadox
Peltier cooling
-40ºC (two stage cooler)
Temperature stability
± 0.05ºC(micro-processor PID)
Analogue-to-digital resolution
True 18 bit
System noise (so-called read noise)
<10 e- RMS
Dark current
< 0.05 e-/pix.s
Communication
Gigabit Ethernet
Pixel Sampling rate
5.6 MHz
Readout time (complete duty cycle including chip readout, CDS, analogue-to-digital conversion, transfer detector-PC, disk storage)
0.22s (4 x 4 binning)
0.46s (2 x 2 binning)
1.59s (1 x 1 binning)
3.3.1. X-ray Tube (Typical Operating Conditions)
SPECIFICATIONS
3.3.2. CCD Detector
3.3.2.1 General
Xcalibur/Gemini Version 2.4
USER MANUAL Page 10
3.3.2.2 Eos CCD Detector
Active area
90 – 92 mm diagonal
Overall dimensions (L x W x H)
218 X 178 X 234 MM
Pixel size on scintillator
31 μm
Scintillator material
Gadox
Fibre optic reduction
1.3:1; low distortion reduction taper
Active area
135 mm diagonal
Overall dimensions (L x W x H)
243 X 178 X 234 MM
Weight
16 kg
Pixel size on scintillator
48 μm
Fibre optic reduction
2:1; low distortion reduction taper
Active area
165 mm diagonal
Overall dimensions (L x W x H)
262 X 192 X 241 MM
Weight
20 kg
Pixel size on scintillator
60 μm
Fibre optic reduction
2.5:1; low distortion reduction taper
Communication
Gigabit Ethernet
Drivers
WinXP
Typical host computer
Intel Core 2 Duo class PC >2 GHz
1.0 Gb RAM
20” colour display
3.3.2.3 Atlas CCD Detector
SPECIFICATIONS
3.3.2.4 Titan CCD Detector
3.3.2.5 PC CCD Interface
Xcalibur/Gemini Version 2.4
USER MANUAL Page 11
3.3.2.6. CCD Detector Maximum Theta and Resolution
Mo
Cu
Maximum resolution in 2 theta
155 °
155 °
Maximum resolution in d-value
0.37Å
0.79Å
Mo
Cu
Maximum resolution in 2 theta
133 °
155 °
Maximum resolution in d-value
0.39Å
0.79Å
Mo
Cu
Maximum resolution in 2 theta
155 °
133 °
Maximum resolution in d-value
0.37 Å
0.84 Å
Mo
Cu
Maximum resolution in 2 theta
165 °
163 °
Maximum resolution in d-value
0.36Å
0.78Å
Mo
Cu
Maximum resolution in 2 theta
141 °
163 °
Maximum resolution in d-value
0.38Å
0.78Å
Eos on Xcalibur
Eos on Gemini
SPECIFICATIONS
Eos on Gemini Ultra
Atlas on Xcalibur
Atlas on Gemini
Xcalibur/Gemini Version 2.4
USER MANUAL Page 12
Atlas on Gemini Ultra
Mo
Cu
Maximum resolution in 2 theta
165 °
141 °
Maximum resolution in d-value
0.36 Å
0.82 Å
Mo
Cu
Maximum resolution in 2 theta
157 °
156 °
Maximum resolution in d-value
0.37Å
0.79Å
Mo
Cu
Maximum resolution in 2 theta
139 °
156 °
Maximum resolution in d-value
0.38Å
0.79Å
Mo
Cu
Maximum resolution in 2 theta
157 °
138 °
Maximum resolution in d-value
0.37 Å
0.83 Å
CCD detector to sample distance
42 to 150 mm (Eos detector)
52 to 150 mm (Atlas detector)
62 to 150 mm (Titan detector)
Titan on Xcalibur
Titan on Gemini
SPECIFICATIONS
Titan on Gemini Ultra
3.3.2.7. CCD Detector Maximum Distance
Xcalibur/Gemini Version 2.4
USER MANUAL Page 13
Type
Four–circle Kappa geometry X-ray goniometer
Sphere of omega, kappa, phi axes coincidence
10 m Maximum load Phi axis
2 kg
Reproducibility
0.00125 deg for Omega and Theta
0.0025 deg for Kappa
0.005 deg for Phi
Scanning speed range
0.005 to 3.0 deg/sec
X-ray shutter response time
3 ms
X-ray Generator
System
Power connection
1/N AC 230 V ± 10% 50/60 Hz
1/N AC 230 V ± 10% 50/60 Hz
Maximum mains current at 230V
19 A
4 A
Power outlet fuse
32 A
16A
3.3.3. Four-circle Kappa Geometry X-ray Goniometer
SPECIFICATIONS
3.4 Electrical Data
X-ray Generator and System
Xcalibur/Gemini Version 2.4
USER MANUAL Page 14
4. Technical Description
Note:
The numbers on this diagram refer to the numbered list above
1 2 3
4
5
6
4.1 Overview of Xcalibur/Gemini
The Xcalibur/Gemini system consists of:
1. A kappa geometry, 4-circle diffractometer
2. A CCD area detector
3. An instrument cabinet with electronics rack
4. A stand for optional equipment like high or low temperature attachments
5. System software installed on PC workstation
6. A water chiller for the X-ray tube (type KMW3000C)
TECHNICAL DESCRIPTION
Figure 4.1.1 Components of a typical Xcalibur/Gemini system
The diffractometer and CCD area detector are mounted inside a cabinet. The cabinet experiment area is mounted on top of the electronics rack. Water chillers for the CCD detector and the X-ray generator are positioned to the side of the instrument. The PC workstation is located close to the instrument to allow a clear view of the diffractometer and convenient access to it.
Xcalibur/Gemini Version 2.4
USER MANUAL Page 15
TECHNICAL DESCRIPTION
Key
1. X-ray tube
2. 4-circle Kappa goniometer
3. X-ray Shutter
4. Collimator
5. Beamstop
6. Beryllium Window
7. Video microscope
1 3 4 5 6
2 7
The diffractometer consists of an X-ray tube, a 4-axis Kappa goniometer (omega, kappa, phi and theta axis) for sample orientation with a detector arm, which has a universal mount capable of supporting any Agilent Technologies CCD area detector or scintillation point detector. The CCD area detector and the point detector are used to measure the X-quanta diffracted from the sample.
Figure 4.1.2 View of Xcalibur Ultra diffractometer
The X-rays are generated by a sealed tube, which is mounted on the goniometer and powered by the high voltage X-ray generator. The X-ray optics consist of a high speed shutter located next to the tube shield, a monochromator for selecting a specified bandwidth from tube spectrum and a collimator for limiting beam divergence. The sample can be viewed with the video microscope, which is attached to the stand doming the instrument. The video image is made available through CrysAlis the PC monitor just outside the cabinet.
Xcalibur/Gemini Version 2.4
USER MANUAL Page 16
Pro
software on
TECHNICAL DESCRIPTION
Figure 4.1.3 View of Gemini diffractometer
The CCD area detector works according to the following principle: The X-rays enter the detector through a Beryllium window to the vacuum-sealed detector unit. A scintillation screen absorbs the X­rays and emits visible light, which is conducted via a fibre optic reduction taper towards the scientific grade CCD chip. The CCD signal is digitised to 18-bit resolution by a correlated double sampling circuit with analog-to-digital converter located in the detector head. The data is then transferred via a gigabit Ethernet data link to the PC workstation. The control program stores it for further analysis to the hard disk.
The four-axis goniometer is driven by microprocessor-controlled stepping motors with 12,800 microsteps per revolution.
The sample is aligned using a video microscope. The sample picture is displayed on the control computer by using the built in frame grabber. The cabinet halogen lighting provides brilliant, high contrast illumination of the sample at all goniometer positions. Optionally, a fibre optic halogen flood light system can be ordered to provide additional lighting focussed onto the sample.
The X-ray generator is located in the electronics rack in the instrument cabinet.
The generator power trips out if the X-ray tube overheats and warning lights on the front panel indicate the operational status of the generator.
The generator is controlled remotely using Crysalis
Pro
software.
Gemini systems contain two generators.
Xcalibur/Gemini Version 2.4
USER MANUAL Page 17
TECHNICAL DESCRIPTION
Figure 4.1.4 Goniometer phi, kappa, omega and theta axes
The programs controlling the measurement procedures of the system are WIN32 applications, which run under Windows XP or Windows 7. The data acquisition and reduction are run concurrently with the program Crysalis
Pro
.
4.2 KMW200CCD Chiller and KMW3000C Chiller
Information about the Chillers is in separate manuals supplied with this equipment.
4.3 Low Temperature Option
If a cryogenic cooler is to be fitted to the diffractometer system, then the correct adapter must be used to mount the cooler on the cryo-support stand. Further details can be obtained from Agilent Technologies.
4.4 Safety Features
In normal operation the X-rays are generated and projected in a totally enclosed cabinet constructed of lead-doped Perspex called Kyowaglas (equivalent to 0.3mm of lead sheet).
Xcalibur has magnetic switches mounted at the base of the front doors, and at the top of the side and rear panels of the protective enclosure. During normal operation (i.e. interlocks applied) the X-ray shutter will not open unless all the cabinet doors are closed. If the doors are then opened whilst the X-ray shutter is open the X-ray shutter is immediately caused to close.
Access to the interior of the cabinet via the front doors is required when changing samples. When the doors are opened during sample change and alignment, the X-ray generator remains on for optimum performance.
Indicator lights mounted on the outside of the enclosure show when the X-ray generator power is on (orange light) and whether the X-ray shutter is open (red light) or closed (green). If these lights are
Xcalibur/Gemini Version 2.4
USER MANUAL Page 18
TECHNICAL DESCRIPTION
defective then the X-ray generator will not operate (in the case of the orange light) and the X-ray shutter will not open in the case of the red and green shutter lights.
The door of the cabinet should remain locked when the system is unattended to prevent unauthorised access to the system.
Xcalibur/Gemini Version 2.4
USER MANUAL Page 19
HANDLING, INSTALLATION, STORAGE AND TRANSIT INFORMATION
WARNING The packing crates are heavy and could cause serious injury and damage to the equipment if not handled correctly. Use suitable lifting equipment and procedures. Only lift the packing cases from the bottom.
CAUTION Do not remove the equipment from the packing crates until they have been moved to their designated installation site. The equipment has been carefully packed to protect the equipment from damage in transit. Removal of the packing equipment could make the equipment vulnerable to damage during transit. Do not unpack the equipment until Agilent Technologies personnel are on site.
5. Handling, Installation, Storage and Transit Information
5.1 Reception and Handling
5.1.1. Delivery
Carry out the following steps on delivery of the Xcalibur system:
1. When the system arrives, check that there is no visible damage, with the delivery driver present.
If damage has occurred contact the carrier and Agilent Technologies immediately.
2. Check that shock-watch and tilt indicators fitted to the outside of the packing cases have not
been activated. If the indicators have been activated notify Agilent Technologies immediately.
3. Check the number of delivered boxes/crates against the packing list. If any items are missing
contact Agilent Technologies within 3 days.
4. Always lift packing cases from the bottom using suitable lifting equipment (refer to list of
component weights in the following section.
5. Move packing cases into the designated installation site.
6. Contact Agilent Technologies to notify them that the equipment is awaiting installation by a
factory trained service representative.
5.1.2. Unpacking
1. Retain all packing material until installation of the system is completed.
2. Ensure that special tools are stored safely for use during maintenance periods.
Xcalibur/Gemini Version 2.4
USER MANUAL Page 20
HANDLING, INSTALLATION, STORAGE AND TRANSIT INFORMATION
Description
NET Weight kg
Dimensions (width x height x depth) cm
Centre of gravity
Lifting points
Kappa goniometer
106
43 x 64 x 47
Offset from centre of unit towards side of X-ray tube mount
At four corners (DO NOT lift from below)
X-ray generator
40
48 x 22 x 69
Centre of unit
From the sides and below
Enclosure
90
115 x 100 x 100
N/A
N/A
Electronics rack
60
80 x 85 x 80
Centre of unit
By hand from top four corners, otherwise from below
KMW200CCD
19
60 x 10 x 37
Right hand side of unit
Underneath at the four corners
KMW3000C
56
38 x 54 x 96
Centre of unit
At the four corners and from below
Helijet
10
30 x 25 x 10 (head only)
Centre of component parts
From below with transfer tube removed
Cryojet
15
15 x 35 x 15 (head only)
Centre of component parts
From below or using handles whilst supporting transfer tube.
Autofill controller (ILM210)
5
44 x 10 x 30
Centre of unit
By hand from below
Helijet controller (ITC502)
5
44 x 10 x 30
Centre of unit
By hand from below
Helijet gas flow controller (GFC1)
12
45 x 26 x 38
Centre of unit
Front panel handles
5.1.3. Mechanical Handling
5.1.3.1 Weights, Dimensions and Lifting Points
Xcalibur/Gemini Version 2.4
USER MANUAL Page 21
HANDLING, INSTALLATION, STORAGE AND TRANSIT INFORMATION
Box
(No.)
Item
Length (cm)
Width (cm)
Height (cm)
Weight (kgs)
1
Goniometer
96
61
79
160 2 Protection cabinet
96
116
59
200 3 Electronics rack
91
89
105
170 4 Helijet (option)
100
67
100
110 5 KMW3000C Chiller
105
47
79
100 6 Eos CCD detector
98
85
99
105
Atlas CCD detector
116
94
112
130 Titan CCD detector
116
94
112
145 7 160L LN2 dewar (option)
70
72
155
119 8 21LCD TFT Monitor
45
28
58
11 9 Cryojet (option)
70
79
189
143
10
X-ray generator
76
60
34
38
11
X-ray generator (for Gemini)
76
60
34
38
5.1.3.2 Boxed Weights, Dimensions and Lifting Points on Delivery
The weights and dimension above are an estimate and should only act as an indication of the lifting requirements when the system is delivered. All boxes are fitted with the facility to use forks to unload. There is 15 cm clearance from floor to the base of each box.
It is recommended that a fork lift truck is available to unload the delivery vehicle with a pallet truck to move the packing cases into the systems final location.
5.2 Installation and Setting to Work
5.2.1. Preparation of Site and Services
5.2.1.1. Environmental Requirements
It is the customer‟s responsibility to ensure that all local building and safety regulations are met.
Ensure that the environmental conditions of the installation site conform to the requirements stated in the SPECIFICATIONS section of this manual.
USER MANUAL Page 22
Xcalibur/Gemini Version 2.4
HANDLING, INSTALLATION, STORAGE AND TRANSIT INFORMATION
5.2.1.2. System Layout
Adequate space is required around the system for servicing. The minimum clearance from the walls and a suggested system layout are shown in drawing number OD-01-00-15.
When the low temperature option is fitted an extra 100 cm space on the left-hand side of the system is required.
Unpacked, the largest subassembly will fit through a door aperture of 85 cm. Check the door aperture to ensure the system can be assembled in its designated area.
5.2.1.3. Electrical Services
A 3-phase 63 A supply, with one or two 32 A outlets (for X-ray generator – one for Xcalibur or two for Gemini) and one 16 A outlet is required. Additionally four single-phase outlets are required as described in drawing OD-01-00-015-C.
Use only the power cables supplied.
Do not connect the electrical power supply circuit to any other devices. Limit the electrical noise in the system by attaching the earth cable exclusively to a system earth terminal with a resistance of less than 0.5 ohms.
Fit a line voltage regulator if the power supply voltage fluctuates more than ±10%.
Locate the mains outlet on the wall behind the system. The mains outlet should be of the circuit breaker type. (Outlet and connecting plugs are not supplied). The mains plug should be readily accessible by the operator when the equipment has been installed.
In areas where the mains power supply is unreliable an „uninterruptible power supply‟ (UPS) is recommended. The UPS should have specifications of 10 kVA with 3-phase input and single phase output. For example the following unit from APC will run Xcalibur/Gemini from a 400V 3-phase input for about 100 minutes:
APC Smart-UPS VT 10kVA 400V w/3 Batt Mod Exp to 4, Start-Up 5X8, Int Maint Bypass, Parallel Capable + (1)SYBT4 Battery Unit
Xcalibur/Gemini Version 2.4
USER MANUAL Page 23
HANDLING, INSTALLATION, STORAGE AND TRANSIT INFORMATION
Description
Voltage (V)
Frequency (Hz)
Maximum nominal mains current (A)
X-ray generator(s)
1/N AC 230 ±10%
50/60
24
CCD detector
90-130 / 180-260
50/60
2.5
KMW200CCD cooler
100 - 240
50/60
2.5
Goniometer Interface
90-130 / 180-260
50/60
2.5
Cabinet lights, sample illumination
1/N AC 230 ±10%
50/60
2.5
Computer and peripherals
100 - 240
50/60
2.5
Cryojet controller (option)
100 - 240
50/60
2
Autofill controller (option)
100 - 230
50/60
0.6
KMW3000C Cooler
1/N AC 230 ±10%
50/60
1.5
Thermo chiller (option)
208 -240
50/60
5.2.1.4. Water Supply
CCD detector Cooling
The CCD detector is water cooled by the KMW200CCD Cooler. Ten metre long water pipes are supplied with the system to connect the cooler to an external water supply.
X-ray Tube Cooling
A cooling system is required to dissipate the heat produced by the X-ray tube. A closed circuit cooling system should be installed to minimise the effects of particles, low pressure and water temperature fluctuation on the performance of the system from local tap water.
The KMW3000C Chiller is a closed circuit cooling system suitable for this purpose. It is supplied with two hoses, 10 metres long with 9 mm inner diameter and 15 mm outer diameter. The maximum distance between the KMW3000C chiller and the system is 10 metres. The distance between the water supply and the chiller is not limited but the supply must deliver 0.5 – 5 bar gauge pressure with a minimum 5 litres/min flow and a gravity drain with an elevation that does not exceed 1metre.
The supply should have a wall mounted shut valve.
An additional chiller, such as the ThermoNESLAB, is required when the location of the diffractometer does not have any running water or lab-cooling system, or if a continuous flow of water cannot be put down the drain. The KMW3000C is a water-to-water heat exchanger and therefore the heat from the X-ray tube is removed by the KMW3000C closed water circuit into the main water bath but this heat load has to be extracted by an additional cold water supply. Refer to the third party ThermoNESLAB manual for information about this equipment, if supplied.
Xcalibur/Gemini Version 2.4
USER MANUAL Page 24
HANDLING, INSTALLATION, STORAGE AND TRANSIT INFORMATION
5.2.1.5. Low Temperature Option
A suitable high vacuum pump, ideally 70 Litres/sec turbo is required to periodically evacuate the Cryojet head and the heater leg. A KF16 fitting is required for this procedure.
To demonstrate the operation of the Cryojet, 100 litres of liquid nitrogen are required.
The customer should supply a suitable table/rack for both the ILM and the Cryojet controller.
5.2.1.6. CCD Camera Pumping
A high vacuum pump is required to periodically evacuate the CCD camera.
5.2.1.7. Helijet Option
The customer should provide a minimum of 50 litres of liquid helium and a minimum of 1 full helium gas cylinder of at least 99.99% grade helium gas in order for the operation of the Helijet to be demonstrated.
5.2.2. Setting to Work
5.2.2.1. Equipment Required
Table for computer (for integration with system) 1 set of Allen keys Phillips (+) screw drivers (assorted sizes) Flat headed screw drivers (assorted sizes)
5.2.2.2. Personnel Required for Installation
5 persons for lifting of heavy components
5.2.2.3. Setting up Procedures
Agilent Technologies personnel normally perform installation. The duration of the installation is typically 4 working days, to include assembly, testing, maintenance and software training.
5.3 Storage
Before installation commences, or when the system is not being used for extended periods, store the diffractometer in accordance with the environmental conditions for temperature and humidity stated in the SPECIFICATIONS section of this manual.
Always store Xcalibur in a secure room
Xcalibur/Gemini Version 2.4
USER MANUAL Page 25
OPERATION
WARNING
Local rules and regulations may apply to the use of the diffractometer. If these exist, refer to these local rules before operating the system.
WARNING
Do not operate the diffractometer unless ALL X-ray tubes are fitted. If the X-ray tube is not present then a high voltage will appear on a bare connector and water will flow out of the tube shield. When a tube reaches the end of its useful life it must be left fitted until a new tube is available for exchange.
Control
Type
Location
Effect
Emergency Stop
Red button
Front, left corner of electronics rack
Shutdown X-ray generator power and stop goniometer movement
Remote Control
Key pad
Inside protective enclosure
Control and drive goniometer; Control must be released from computer
Cabinet Illumination
Switch
Front of electronics rack
Illuminate inside of enclosure
Sample illumination
Switch
Front of electronics rack
Illuminate crystal sample (optional extra)
Video Microscope
Switch
Front of electronics rack
Switch on / off video microscope
Flow gauge
(only on systems fitted with Ultra)
Turn knob
Inside front right of electronics rack
Control of Helium flow rate through Ultra optic housing (should be 30 – 40 cm/min)
6. Operation
Xcalibur is a computer-controlled system. All functions are controlled from the computer terminal except when control is released from the computer terminal to the remote control unit. Power is switched on and off via manual switches located on the system.
The KMW200 and KMW3000C chillers must be connected to Xcalibur, before operating the system.
6.1 Controls and Indicators
Control
Xcalibur/Gemini Version 2.4
USER MANUAL Page 26
Power
Control
Type
Location
Effect
Diffractometer goniometer interface
Red switch
Interface front panel, left; Inside front door of electronics rack
Power on / off to goniometer interface
CCD power supply
Switch
Inside enclosure underneath power supply box
Power on / off to CCD detector
Water Cooler (for CCD)
Red Switch
Cooler front panel, left; Inside front door of electronics rack
Power on / off to Water Cooler (also to CCD detector)
Mains power supply
Switch
Inside left panel, electronics rack
Power on/off to system
Safety Override Key (for Agilent service use only)
Cabinet Illumination Switch
Video Microscope Switch
Sample Illumination Switch (optional)
Flow Gauge (only on systems fitted with Ultra)
OPERATION
Figure 6.1.1 Location of switches
Xcalibur/Gemini Version 2.4
USER MANUAL Page 27
OPERATION
Indicator
Location
Meaning
Orange light (X-ray Mo)
Outside, top right of the protective enclosure
X-ray Mo generator power is on
Orange light (X-ray Cu)
Outside, top right of the protective enclosure
X-ray Cu generator power is on
Red light (Shutter open)
Outside, top right of the protective enclosure
X-ray shutter open
Green Light (Shutter closed)
Outside, top right of the protective enclosure
X-ray shutter closed
Red light (only on systems fitted with Ultra)
Back light of flow gauge inside front right of electronics rack
He gas flow out of range through gauge. X-ray interlock is open.
Green light (only on systems fitted with Ultra)
Back light of flow gauge inside front right of electronics rack
He gas flow is within range through gauge. X-ray interlock is closed.
Red light (Shutter open)
Inside top protective cabinet on beamstop support near X-ray shutter)
X-ray shutter is open
6.2 System Start-Up
Safety devices in the diffractometer protect against damage to the system and the operator during start up. The following initial switch on procedure should be followed.
1. Plug in all mains power cords to their wall mounted sockets (system, computer monitor).
2. Switch on KMW200CCD chiller using the power switch on the front panel (open the front
door of the electrical rack to access).
3. Wait for the chiller to start-up. The error state lights on the front panel will flash three times
when the chiller is ready. Then check that the detector power supply box is switched on (red lamp lights on the top).
4. Allow 60-90 minutes for the CCD detector to cool the Peltier and chip to -40ºC.
5. Turn on the interface using the red switch located on the left hand side of the interface front
panel. On the MGC model: ensure 4 green LEDs are illuminated on the right hand side of the interface panel. (These lights are labeled Power, EXT I/O, Ready and Enable). On the DC08 model: wait 15 seconds for the LEDs to self-test and then ensure 2 green LEDs are illuminated (these are labeled: Power and Ext Supply)
6. Turn on the computer monitor.
7. Switch on the PC and launch the CrysAlis
initialise. If you have the interface DC08 model then the front panel green LED „Goniometer Ready‟ should light.
Xcalibur/Gemini Version 2.4
USER MANUAL Page 28
Pro
programme. The goniometer should auto-
OPERATION
WARNING
When handling and using X-ray tubes particular care should be taken to avoid injury caused by exposure to the X-rays, exposure to high voltages, possible implosion of the vacuum tube and contact with Beryllium and Beryllium oxide.
8. Switch on the KMW3000C chiller using the key provided. (Ensure that the external circuit
water cooling is also turned on).
9. Turn on the X-ray generator (switch on the generator back panel).
10. To generate X-rays open the X-ray device server in the Crysalis
28 kV, 1 mA then press OK. The generator should now ramp up to these settings. If the X-ray tube has not been running for >1 hour follow the warm-up procedure suggested in the manufacturers manual.
11. After closing all protection cabinet panels the system will enable the X-ray shutter interlock
(to allow it to be opened after a command from the software). If you have the interface DC08 model then the front panel green LED „Shutter Enable‟ should light.
Pro
GUI. Click on start and enter
6.3 X-ray Tube Warm-up Procedure
If the X-ray tube is stored for a lengthy period, the vacuum may become impaired. We recommend carrying out the following warm-up procedure for new tubes and tubes that have not been used for more than 100 hours.
Use Crysalis the tube manufacturer‟s manual.
If, after completion of this procedure, a flashover occurs in the next 30 minutes, repeat the whole procedure.
Pro
X-ray device server software control to carry out the burn in procedure as detailed in
6.4 System Standby and Normal Shutdown Procedure
Depending on the period of diffractometer inactivity, follow one of the 3 procedures below.
6.4.1 Shutdown Procedure (Diffractometer not expected to be used for 2-3 days)0.
1. Ramp down the X-ray generator to 10 mA, 30 kV (improves lifetime of the X-ray tube).
2. Park the goniometer in its home position simply by exiting the Crysalis
3. Turn off the video microscope using the front panel mounted switch.
4. Turn off the cabinet & sample illumination using the front panel mounted switches.
5. Warm up/turn off sample cooling device (following manufacturers guidelines) – where
applicable.
6. Logoff the computer and shut down.
Pro
program.
7. Turn off the computer Monitor („O‟ on the screen).
Xcalibur/Gemini Version 2.4
USER MANUAL Page 29
OPERATION
WARNING
The emergency stop button will only shut down high voltage from the X-ray generator. Other equipment on the system will still be powered on.
6.4.2 Shutdown Procedure (Diffractometer not expected to be used for 1 week or more)0.
1. Turn off X-ray generator using the X-ray device server in the Crysalis
flow for 15 minutes to cool tube.
2. After 15 minutes turn off the KMW3000C X-ray tube chiller using the key provided. You may
wish to turn off the external water supplying circuit, although this is not necessary since a thermostatic valve will cease flow when the bath reaches a pre-determined temperature.
3. Park the goniometer in its home position simply by exiting the Crysalis
4. Turn off the video microscope using the front panel mounted switch.
5. Turn off the cabinet & sample illumination using the front panel mounted switches.
6. Warm up/turn off sample cooling device (following manufacturers guidelines) – where
applicable.
7. Logoff the computer and shut down.
8. Turn off the computer Monitor („O‟ on the screen).
9. Turn off the diffractometer interface (red illuminated switch located on the left side of the
front face).
Pro
GUI). Allow water to
Pro
program.
6.5 Emergency Shutdown
The emergency shutdown procedure should be used:
If there is a fire or any other emergency requiring the evacuation of personnel from the area
AND
The procedure can be performed without endangering any persons‟ safety.
6.5.1. Emergency Shutdown Procedure
0.
1. Press the red emergency stop button located on the front left of the system.
2. Switch off all power at the mains electrical supply.
3. Turn off mains water supply.
0.
Xcalibur/Gemini Version 2.4
USER MANUAL Page 30
MAINTENANCE SCHEDULES
CAUTION
1. Maintenance tasks must only be carried out by authorised operators who have undergone specialist radiation training. Refer to local rules for further details.
2. Failure to perform scheduled maintenance tasks properly and at the correct intervals can affect the safety and performance of this system.
3. Before performing any maintenance task ensure that you have read and understood the HEALTH AND SAFETY INFORMATION at the beginning of this manual and any local rules governing the use of the diffractometer system
Action
Personnel
Estimated task duration
Estimated elapsed time
1. Check door and cabinet window
safety interlocks
Authorised Operator
2 minutes
2 minutes
Action
Personnel
Estimated task duration
Estimated elapsed time
1. Check Emergency Stop
Authorised Operator
10 minutes
10 minutes
2. Check X-ray radiation levels
Authorised Operator
20 minutes
20 minutes
7. Maintenance Schedules
7.1 Introduction
Various maintenance tasks must be performed to ensure that the diffractometer system continues to operate safely and reliably. These tasks are detailed in the maintenance schedules given below.
Planned maintenance that can be performed by the user is limited to replacing consumable items, alignment procedures, pump-down procedures, and checking radiation levels and safety features. The user‟s authorised service representative should carry out other tasks. If in any doubt about the performance of the diffractometer, contact Agilent Technologies.
7.2 Weekly Maintenance Schedule
Tools and Materials:
None
7.3 Monthly Maintenance Schedule
Tools and Materials:
Radiation meter
Xcalibur/Gemini Version 2.4
USER MANUAL Page 31
MAINTENANCE SCHEDULES
Action
Personnel
Estimated task duration
Estimated elapsed time
1. Refining machine parameter file
Authorised Operator
4 – 5 hours
4 – 5 hours
2. Check alignment of video
microscope and cryogenic cooling device (if fitted)
Authorised Operator
1 hour
1 hour
Action
Personnel
Estimated task duration
Estimated elapsed time
1. Refine machine parameter file
Authorised Operator
4 - 5 hours
4 – 5 hours
2. CCD detector – pump out
vacuum
Authorised Operator
30 minutes
3- 6 hours (depending on detector)
7.4 Six Monthly Maintenance Schedule
This maintenance schedule should also be completed after adjustment of the X-ray beam, collimator and CCD detector position.
Tools: Ylid test crystal - (C11H10SO2)
7.5 Yearly Maintenance Schedule
Tools and Materials:
Ylid test crystal - (C11H10SO2) Special „T‟ tool Phillips screw driver Rotary vacuum pump capable of obtaining a pressure of <0.04 mbar Vacuum tubing and adaptor
7.6 10,000 Hours Maintenance Schedule
Tools and Materials:
Set of Allen keys Set of flat-headed screwdrivers Water receptacle (e.g. Bucket) Special tools –special tool 1 and special tool 2 (see section 8 for identification)
Xcalibur/Gemini Version 2.4
USER MANUAL Page 32
MAINTENANCE SCHEDULES
Action
Personnel
Estimated task duration
Estimated elapsed time
1. Changing the X-ray tube
Authorised Operator
30 minutes
30 minutes
2. Alignment of X-ray optics
Authorised Operator
2 hours
2 hours
Xcalibur/Gemini Version 2.4
USER MANUAL Page 33
MAINTENANCE INSTRUCTIONS
WARNING
1. Read and understand the Health and Safety Information section of this manual before performing any maintenance procedures.
2. Follow any local, national or international rules and guidelines that apply to this equipment when performing maintenance tasks.
3. Maintenance tasks must only be performed by authorised operators of the diffactometer.
Task Time:
4 – 5 hours
When:
After adjustment of beam, collimator, CCD detector position etc. Otherwise twice a year.
Tools:
Ylid test crystal - (C11H10SO2)
8. Maintenance Instructions
Some procedures in this section require that commands are entered in the command line. To open the command line:
1. Launch Crysalis
2. Click on the “command line“ icon on the left hand side tool bar. 0.
Pro
8.1 Refining the Machine Parameter File
Procedure:
See section „Refining of Machine Parameter File‟, located in the CrysAlis
Pro
software manual.
Xcalibur/Gemini Version 2.4
USER MANUAL Page 34
MAINTENANCE INSTRUCTIONS
Task Time:
30 mins
When:
Sealed source X-ray tubes have a limited useful lifetime. Operation of a 2.2 kW tube at typical output of 1.5 kW will normally provide 10,000 hours of use (actual values may vary, see third party manual).
Note: X-ray tubes also have a limited shelf life.
Who:
Agilent authorised operator, trained on radiation safety
Tools:
Set of Allen keys Set of flat-headed screwdrivers Water receptacle (e.g. Bucket)
WARNING
Do not touch the Beryllium windows in the X-ray tube. Beryllium is a potentially toxic material.
CAUTION
The customer may change the X-ray tube of Enhance Ultra but the focussing optic inside this X-ray source must be precisely aligned to each new X-ray tube and this may only be performed by a trained Agilent Technologies service engineer. Missing this alignment could result in worse performance of this X-ray source.
CAUTION
Use only ceramic X-ray tubes not glass X-ray tubes
Remove all packaging material and red plugs from water openings before fitting
X-ray tube must be inserted in the correct orientation with the Agilent Technologies logo upside down.
CAUTION
Have a receptacle ready to receive water from the pipes and X-ray tube shield.
8.2 Changing the X-ray Tube of Ultra (Horizontal X-ray tubes)
Procedure
1. Switch off the power to the X-ray tube to be changed.
2. Turn off the KMW3000C chiller to stop water circulating.
3. Leave for at least 20 minutes to ensure that the X-ray tube is not hot (test temperature before
proceeding).
Xcalibur/Gemini Version 2.4
USER MANUAL Page 35
MAINTENANCE INSTRUCTIONS
CAUTION
Failure to fit these o-rings will result in water leakage which could damage both the X-ray tube and the instrument.
WARNING
Ensure that the X-ray tube is correctly mounted. Incorrect mounting may result in exposure of personnel to X-ray radiation.
4. Disconnect the top water pipe from back of Ultra X-ray tube shield (This is attached with a quick
release connector).
5. Remove the Ultra front face panel. To do this locate the metal block on the left side of the X-ray
tube housing, which is behind the front face panel, find the single screw at the rear of this block which is holding the front face panel and use a 5mm Allen key to completely undo the screw. Lift off the front face panel.
6. Unscrew the temperature sensor from the front plate (metal base, with label) of the X-ray tube.
7. Release the two screws holding the X-ray tube in place and withdraw the X-ray tube from the X-
ray shield. Do this by holding the front plate (metal base, with label) and pulling towards the front of the Ultra. Ensure that the X-ray tube is held level and withdraw parallel to the X-ray shield.
8. For disposal of used X-ray tubes see the DISPOSAL chapter of this manual.
9. Make sure the two o-rings remain in their grooves at the front of the tube shield.
10. Replace the X-ray tube with an identical specification Cu X-ray tube. Holding by its metal base,
remove the new X-ray tube from its protective box
11. From the front of the ULTRA insert the X-ray tube into the tube shield, in the same orientation as
the previous one, and secure with two screws.
12. Re-mount the thermocouple to the base of the X-ray tube and re-mount the ULTRA front face
panel. Secure in place using a 5mm Allen key and the single screw described above.
13. Re-connect the water supply to the back of the ULTRA X-ray tube shield.
14. Switch on power to the generator and KMW3000C chiller unit.
Xcalibur/Gemini Version 2.4
USER MANUAL Page 36
MAINTENANCE INSTRUCTIONS
Task Time:
30 minutes
When:
Sealed source X-ray tubes have a limited useful lifetime. Operation of a 3 kW Mo tube at typical output of 2 kW, and a 2.2kW Cu tube at 1.5 kW, will normally provide 10,000 hours of use (actual values may vary, see third party manual).
Note: X-ray tubes also have a limited shelf life.
Who:
Agilent authorised operator, trained on radiation safety
Tools:
Set of Allen keys Set of flat-headed screwdrivers Water receptacle (e.g. Bucket)
WARNING
Do not touch the Beryllium windows in the X-ray tube. Beryllium is a potentially toxic material.
CAUTION
Use only ceramic X-ray tubes not glass X-ray tubes
Remove all packaging material and red plugs from water openings before fitting
X-ray tube must be inserted in the correct orientation with the Agilent Technologies logo upside down.
8.3 Changing the X-ray Tube of Enhance (Vertical X-ray Tubes)
Procedure
1. Switch off the power to the X-ray tube to be changed.
2. Turn off the KMW3000C chiller to stop water circulating.
3. Leave for at least 20 minutes to ensure that the X-ray tube is not hot (test temperature before
proceeding).
4. Switch off and remove any cryogenic device mounted on the diffractometer.
5. Remove the plate for mounting cryogenic devices (3x M8 screws to be removed from top of
attachments tower).
6. Unscrew (one screw at the front and the back) the end cap from the top of the Enhance and
remove (this may also need the loosening of the conical cover at the base end of the collimator for the Cu source) and lay to one side along with the integrally mounted thermal sensor.
7. Unscrew and slide out the X-ray tube in an upwards direction. A little water inside the end of
the X-ray tube plate will drip out and should be cleaned up, but otherwise all water in the system will be retained inside the pipes.
8. Make sure the two o-rings remain in their grooves at the top of the tube shield.
Xcalibur/Gemini Version 2.4
USER MANUAL Page 37
MAINTENANCE INSTRUCTIONS
CAUTION
Failure to fit these o-rings will result in water leakage which could damage both the X-ray tube and the instrument.
Task Time:
<1 min
When:
Different sizes of collimator may be necessary for different experiments Three standard collimators of sizes 0.5, 0.8 and 1.0 mm are provided with the diffractometer system. A 0.3 mm standard length collimator and a short 0.3 and 0.5 mm collimator (for use in high pressure experiments) are available by special order. Different lengths of collimator are used for Mo and Cu sources, the longer collimators are for the Mo source.
Tools:
None
WARNING
Ensure that the collimator is correctly mounted. Incorrect mounting may result in exposure of personnel to X-ray radiation.
WARNING
Ensure that the X-ray tube is correctly mounted. Incorrect mounting may result in exposure of personnel to X-ray radiation.
9. Fitting the new tube is the reverse procedure (maintaining the same orientation of the tube).
10. For disposal of used X-ray tubes see the DISPOSAL chapter of this manual.
11. Holding by its metal base, remove the new X-ray tube from its protective box.
12. Slide the X-ray tube into position on the diffractometer and secure with two screws.
13. Fit the end cap (with integral temperature sensor) onto the top of Enhance and secure using
the two fixing screws.
14. Re-fit the plate for mounting cryogenic devices.
15. Re-fit any cryogenic device.
16. Switch on KMW3000C water chiller.
8.4 Changing the Collimator of Enhance
Procedure
1. Ensure that the X-ray shutter is closed and all interlocks are in place.
2. No sample should be mounted on the goniometer.
3. Pull out the collimator from its housing, pulling from the right hand side. Be careful not to touch
the beamstop or CCD detector.
4. Select the required size of collimator and push into position on the diffractometer such that it is
flush to the housing surface. The collimator should be inserted such that the flat edge with the collimator size label is uppermost.0.
Xcalibur/Gemini Version 2.4
USER MANUAL Page 38
MAINTENANCE INSTRUCTIONS
Task Time:
< 30 minutes
When:
When it becomes impossible to centre a sample correctly on the video microscope image.
Tools:
Pre-mounted 0.3 mm steel ball mounted on a goniometer head.
Figure 8.6.1 Goniometer Head Adjustments
Figure 8.6.2 Video Microscope Adjustments
Left / right Adjusters Height Adjuster
Equivalent to Height Adjustment Left / right Adjustment
Focus
8.5 Aligning the X-ray Collimator of Enhance
The X-ray collimator is factory aligned via the collimator housing. This means that re-alignment is not needed when changing between the collimators supplied by Agilent Technologies.
If collimator re-alignment is required please contact Agilent Technologies for assistance.
8.6 Aligning the Video Microscope
When first assembled the Goniometer sample centre and the video microscope focal point should be close to each other, however the video image only displays a field of view approximately 1.2mm x
0.8mm and its depth of field is similarly limited. Adjustments need to be made so that the sample centre and microscope focal point coincide exactly. Both the Goniometer head and the video microscope have adjustments to move their centres in three planes. Agilent Technologies supply a pre-mounted 0.3mm steel ball to aid this procedure.
The three adjustments for the microscope can be seen in the figure above (Video Microscope Adjustments).
Xcalibur/Gemini Version 2.4
USER MANUAL Page 39
MAINTENANCE INSTRUCTIONS
NOTES
An indication that the video microscope is not properly centred is if the sample cannot be centred to the middle of the video image, then small adjustments of the microscope may be needed during the sample centring process. The focus may also need adjusting occasionally if the Crystal size varies widely from one experiment to another.
For a correctly centred sample, and video microscope, the variation in position of the sample (Sphere of Confusion) as it rotates to any of the positions on the remote control (except home) should be one small division on the image (10μm). In the unlikely event that correct centring cannot be achieved contact Agilent Technologies for advice.
To centre the video microscope and therefore the steel ball on the centre of the video image use the following procedure:
Procedure
1. Within the CrysAlis
goniometer to the remote control pad.
2. Mount the 0.3mm steel ball plus goniomenter head onto the goniometer.
3. Press the „Lower‟ then ‟0‟ keys and wait for the goniometer to stop moving.
4. Adjust the goniometer head such that the steel ball is visible on the video image.
5. Press „180‟ and watch the steel ball as it is rotated, it will probably move off image. Use the
adjustments on both the Head and Microscope equally to move the ball back to the centre, press „0‟. The ball should end up close to the centre of the image, adjust again to make it central, repeat as necessary until the movement of the ball is minimised.
6. Press „90‟ and move the ball to the centre using the head only.
7. Press „270‟ and watch the ball as it rotates. If the microscope is correctly centred left to right it
will be possible to centre the ball using only the Head adjustments, if not small adjustments of the microscope may be required, repeat until the ball appears to spin on the spot for all four angle positions.
8. Press „0‟, then „Upper‟, watching the ball as it moves. Use the height adjustments on both the
Head and the microscope equally to re-centre the ball, press „Lower‟ and repeat step 8 until there is no movement of the ball as the goniometer moves between the two positions.
Pro
software, press „F12‟ on the computer keyboard to release control of the
9. Repeat steps 5 to 7 at the „Upper‟ position.
10. Adjust the focus on the microscope to bring the ball into sharp focus. The video microscope is
now aligned. The video microscope should not need to be adjusted again, unless it is knocked or extra weight is added to, or removed from the support arm, for example, a low temperature device is added or removed.
11. When fully centred in the Upper and Lower positions the goniometer should be sent back to
„Home‟ once this is complete the ball (or Crystal) will appear to be displaced towards the left of
the video image. This is an optical illusion caused by the angle of the microscope with respect to
the Crystal at „Home‟ and can be ignored. .
Xcalibur/Gemini Version 2.4
USER MANUAL Page 40
MAINTENANCE INSTRUCTIONS
Task Time:
2 minutes
When:
Once a week
Tools:
None
Task Time:
10 minutes
When:
Once a month
Tools:
None
8.7 Checking the Door Safety Interlocks
Procedure
1. With the diffractometer in normal operation and at normal X-ray generator settings, open the
shutter using the shutter open button in the Service Power Tool of the CrysAlis
2. Open the front doors of the enclosure cabinet. The X-ray shutter should immediately close.
3. Close the enclosure doors.
4. Repeat step 1 and open the side panels of the cabinet. The X-ray shutter should immediately close.
5. Close the side panels.
6. Repeat step 1 and open the back panels of the cabinet. X-ray shutter should immediately close.
7. Close the back panels.
Pro
software.
8. Using the weekly maintenance sheet (see appendix). Record the date, persons testing and sign off
the outcome.
8.8 Checking the Emergency stop
Procedure
1. With the diffractometer running at minimum settings (for example 28 kV, 1 mA), press the red
Emergency Stop button.
2. The X-ray generator should switch off immediately, release the emergency stop button by twisting
anti-clockwise, restart the X-ray generator. Return the generator / X-ray tube to required settings.
3. Switch off the interface, wait 5 seconds, restart the interface and then in the Service Power Tool
of CrysAlis
Pro
click on the radio button gon reinit and click Start.
4. Using the monthly maintenance schedule (see appendix) record details of the test, stipulating the
date, person testing, outcome and signature. .
Xcalibur/Gemini Version 2.4
USER MANUAL Page 41
MAINTENANCE INSTRUCTIONS
Task Time:
20 minutes
When:
Once a month
Tools:
Appropriate Radiation meter for X-ray energy levels (Mo14 keV, Cu 8 keV)
Task Time:
3 hours
When:
Once a year
Tools:
Special „T‟ tool (carried by Agilent Technologies service engineer) 1 set of Allen keys Rotary vacuum pump capable of obtaining a pressure of 1x10-1 mbar Vacuum tubing and adaptor
8.9 Checking the X-ray Radiation Levels
Procedure
1. With the X-ray generator on (at normal operational settings) and the shutter closed, use the
radiation meter to sweep the area inside the cabinet and around the X-ray tube housing, fast shutter and collimator for any radiation leak.
2. Close the protective enclosure doors and open the X-ray shutter. In the Service Power Tool of
CrysAlis
3. Sweep the outside of the enclosure using the radiation meter, paying particular attention to the
plane of the X-ray tube, the area of the secondary beam stop (mounted on the right hand side enclosure window) and the door seals.
Pro
click on Shutter Open.
4. Close the X-ray shutter by clicking on the Shutter Close button in the Service Power Tool.
5. Using the monthly maintenance schedule log, record details of the test, stipulating the date,
person testing, outcome and signature.
8.10 CCD Detector – Pumping Out Vacuum
Procedure 0.
1. Switch off the cooling water for the CCD detector at the cooler/ chiller unit.
2. Allow one hour for the detector to warm to room temperature.
3. For the Eos detector unscrew front four corners of the collision sensor of the CCD detector. This
is spring loaded so be ready to catch it as you release the screws. Remove the front of the CCD detector.
Xcalibur/Gemini Version 2.4
USER MANUAL Page 42
MAINTENANCE INSTRUCTIONS
Pump out port (with „T‟ tool
attached)
Pump out port (with cover unscrewed)
CAUTION
Only use new, clean tubing. Apply vacuum before removing the CCD detector vacuum seal to prevent oil etc. diffusing into the detector and causing damage.
CAUTION
Do not turn off the vacuum pump until the CCD detector has been sealed using the relevant stopper.
Figure 8.7.1 View of the access plate and pump out port of the Eos CCD detector
4. For the Atlas or Titan detector unscrew the cap from the pump out port
Figure 8.7.2 View of the pump out port of the Atlas/Titan CCD detector
5. Screw the „T‟ tool into place on the pump out port.
6. Attach vacuum pump to the side arm of „T‟ tool using a suitable adaptor and vacuum tubing.
7. Switch on vacuum pump and evacuate tubing / pumping line to 1x10
-1
bar pressure.
8. Depress „T‟ tool plunger and screw into place. This screws into the o-ring sealed metal stopper
that secures the detector vacuum.
9. Slowly withdraw plunger to remove stopper. Secure plunger in extended position.
10. Leave under vacuum for 20 minutes.
11. Slowly depress „T‟ tool plunger to replace stopper and unscrew plunger
Xcalibur/Gemini Version 2.4
USER MANUAL Page 43
MAINTENANCE INSTRUCTIONS
12. Switch off vacuum pump and disconnect pipes.
13. Unscrew „T‟ tool and replace the cap on the port of Atlas or Titan. For Eos replace the front
access plate of CCD detector, leaving 1mm gap equally all around the edge of the collision sensor.
14. Switch on the cooling water for the CCD detector at the cooler/chiller unit and monitor the cool
down using the CCD device server until the set temperature is obtained.
Xcalibur/Gemini Version 2.4
USER MANUAL Page 44
9. Trouble Shooting
Symptom
Fault
Solution
X-ray shutter will not open (resulting in blank images in the software)
1. Shutter is not enabled
If you have interface model DC08: open front door of electrical rack and look for Shutter Enable green LED illuminated on front panel. If it is not then proceed with the checks below.
2. Protective cabinet doors are not
closed properly
Close doors properly. If the interface is model DC08 then the front panel
will show the green LED „Shutter Enable‟. Otherwise if the safety
circuit is closed a green lamp is illuminated within the control electronics. Access is via the left side panel of the electronics cabinet.
3. Shutter-open and /or shutter
closed lights defective
Check light connections and LED light display.
To check if LED arrays are working;
turn off the interface, wait 5 seconds and turn back on the interface. For a short period all shutter indicators should be illuminated (i.e. open/closed on the exterior of the cabinet and the Shutter Open (red) lamp inside the cabinet on the beamstop mounting)
4. Protective cabinet doors were
opened whilst the shutter was open and safety circuit was activated
In CrysAlis
Pro
re-start experiment.
5. X-ray generator is turned off
Turn on X-ray generator and start high voltage.
6. Shutter signal cable
Ensure that the shutter cable
(Fischer connection) is secured to the optics/shutter housing.
TROUBLE SHOOTING
Xcalibur/Gemini Version 2.4
USER MANUAL Page 45
TROUBLE SHOOTING
Symptom
Fault
Solution
7. He flow is out of normal range
(ULTRA)
Adjust He flow rate to normal range so that lamp behind flow gauge shows green colour, not red.
8. Beam stop is in the „up‟ position
(ULTRA)
Move beam stop to the „down‟
position.
9. Emergency stop has been
pressed.
Ensure emergency stop has not
been pressed (generator will also not function). If yes, twist to release and turn off/on the interface.
Generator will not generate X-rays
1. No power reaching the generator
1. Ensure fuse or circuit breaker of
mains supply is OK.
2. Ensure switch on the unit (rear
panel) is in the correct (1) position.
2. Computer RS232 communication
is not connected
Check COM cable from generator is
connected at the rear of the computer. Check COM port number has been detected by CrysAlis
Pro
.
3. Emergency stop has been pressed
and not released.
1. Twist emergency stop button
clockwise to release.
2. Restart high voltage on generator.
3. Switch off system interface at the
front panel, wait 3 s and then restart. Type gon reinit on Crysalis
Pro
command line
4. If Mo X-rays are required then
change over radiation type (Gemini models only) using software button.
Xcalibur/Gemini Version 2.4
USER MANUAL Page 46
TROUBLE SHOOTING
Symptom
Fault
Solution
4. No cooling water is flowing
1. Check KMW3000C water cooler is
switched on.
2. Check water cooler has no error
state
3. Check all necessary water valves
are open
4. Check external cooling water
supply is flowing and has an acceptable temperature.
5. X-ray tube temperature is too high
There is a temperature switch on
the top of the X-ray tube. If it is hot then wait for it to cool down. Investigate the reason for overheating of the X-ray tube before proceeding.
6. X-ray lamp on cabinet is defective
Check light connections and LED
light display.
7. Additional X-ray lamp is defective
Some laboratory installations have
an additional X-ray lamp in the interlock circuit. This might be wall-mounted near to the system or outside the laboratory. Check the lamp connections and the replace the bulb with the same type, if necessary.
Problem reading kV and mA from generator
1. RS232 communication cable not
connected properly
Check the RS232 cable is connected to back of PC (serial COM port) and to back of the generator
2. Check PC serial port is working
CrysAlis
Pro
should search through all
available serial ports
Xcalibur/Gemini Version 2.4
USER MANUAL Page 47
TROUBLE SHOOTING
Symptom
Fault
Solution
Software cannot control goniometer
No communication with interface
1. Check interface is switched
on.
2. For interface model MGC:
Check RS232 serial cable is connected between rear of PC and the interface. For interface model DC08: Check Ethernet cable is connected between PC and system and to rear panel of interface.
3. For interface model MGC:
check emergency stop on the front of the interface has not been pressed. If yes, twist it to release, turn off power on front panel, wait 5s and then restart.
Goniometer axis will not move
1. Check the goniometer is
connected to the interface with two cables. (Always turn off the interface before plugging/unplugging these cables)
2. Check yellow/green
grounding cable is connected to the goniometer.
3. Contact Agilent
Technologies for further support.
Xcalibur/Gemini Version 2.4
USER MANUAL Page 48
TROUBLE SHOOTING
Symptom
Fault
Solution
Goniometer / Detector collision
1. Collision of the goniometer with
the collision sensor on the front of Eos CCD detector (interface front panel will show a red LED for the collision error state). Other detector types do not have a collision sensor.
Turn off the KMW200CCD using the switch in the front of the electrical rack (this will disable the collision sensors; great care should be taken while the machine is in this state as considerable damage could be caused).
In CrysAlis
Pro
CCD open the service
power tool and select MGCUTIL and
press „start‟.
Click on „parameters‟ and the two „Edit deg‟ buttons. Set both to
1.0000 and click on „slow‟. Click
„enter‟
Select „Detector‟ and „drive relative +‟. Click on „Start‟. This will move
the detector away from the goniometer axis in 1 mm steps. This should be repeated several times until there is a clear gap between the detector and the goniometer axis.
Turn on the KMW200CCD chiller.
Follow steps 2 – 9 from “Power loss
to goniometer” below
Goniometer lost angle positions
1. Collision of the goniometer
See above
Xcalibur/Gemini Version 2.4
USER MANUAL Page 49
Symptom
Fault
Solution
2. Power loss to goniometer
1. Restore power to system, restart the
PC and CrysAlis
Pro
program.
2. Switch off the interface; wait 5
seconds and then switch on.
3. In CrysAlis
Pro
open the service power
tool and select MGCUTIL and press
„start‟.
4. Click on „parameters‟ and the two
„Edit deg‟ buttons. Set both to 1.0000 and click on „slow‟. Click „enter‟
5. Select offending angle from the radio
boxes. Click on “Set angles” and „Start‟. Input actual position of each
axis, as read from the mechanical goniometer. Pay special attention to
„+‟ or „-„
6. Select “Rsync all” and click „Start‟.
Note that at the end of the Rsnync
operation the “Rsync” flags should
read 1. This is located on the text line below the parameter button.
7. If the detector distance is not at
home: select „Detector‟ and „drive relative -‟. Click on „Start‟. This will move the detector towards home in 1 mm steps. Look at the HOME flag number for detector distance. It might initially be 1 if the detector is far from home. Just before home this flag will display 0 for about 2mm and when it changes back to 1 you should stop since this is home.
8. Select “Hsync all” and click „Start‟.
Note that at the end of the Hsync operation the “Hsync” flags should read 1. This is located on the text line below the parameter button. Click on „OK‟
9. Select “gon reinit” and „start‟ in the
service power tool. This will reinitialise all axes and leave the goniometer in its home position ready for operation.
TROUBLE SHOOTING
Xcalibur/Gemini Version 2.4
USER MANUAL Page 50
TROUBLE SHOOTING
Symptom
Fault
Solution
3. Emergency stop button pressed
Release the emergency stop button by twisting clockwise and follow the instructions on the PC screen:
Power off interface; wait 5 seconds and then power on.
Open service power tool, select “gon reinit” and „start‟ in CrysAlis
Pro
. This will reinitialise all drives and return the goniometer to the position it was at when the emergency stop button was pressed
Eos, Atlas, Titan: Strange artefacts on CCD images (lines, blocks, chequer board)
1. Communication errors on
Ethernet link between CCD and PC.
1. Reset the CCD detector by
switching off the Water Cooler power on the front panel, waiting 10s, and restarting.
2. Check that the Ethernet link
restarts shortly afterwards by checking Windows icon bar.
3. Close and restart Crysalis
Pro
.
Xcalibur/Gemini Version 2.4
USER MANUAL Page 51
10. Spares
Device
Location
Value
Package
Designators
Power distribution
Electronics rack ((left hand side rear panel)
10 A / 250 V fast
5x20 mm
F1
Fan panel
Electronics rack (rear panel)
315 mA / 250 V fast
5x20 mm
F3
Cabinet illumination
Electronics rack (rear panel)
6.3 A / 250 V fast
5x20 mm
F4
Shutter driver
Shutter driver PCB
0.8 A / 250 V fast
5x20 mm
F1
0.8 A / 250 V fast
5x20 mm
F2
Interface (MGC model)
Interface front panel
2 A / 250 V
5x20 mm
F1
System control board
System control board (left hand side of rack)
800 mA T800 mA / 250 V
5x20 mm glass "slow"
F1 F2
Device
Location
Description
Fast shutter
Front panel
MultiLED T5, 5 RED 12 V / 25 mA (Swisstac)
CCD
Top panel of CCD camera
MultiLED T5, 5 RED 12 V / 25 mA (Swisstac)
MultiLED T5, 5 GREEN 12 V / 25 mA (Swisstac)
Protective enclosure illumination
Inside (top) protective enclosure
Halogen 12 V AC / 35 W (Osram) type: 44865WFL (x3)
Fibre-optic lights
Inside light source box
Halogen 12 V 75 W (Osram) type: 64617
10.1 Fuses
SPARES
10.2 Bulbs
Xcalibur/Gemini Version 2.4
USER MANUAL Page 52
DISPOSAL INSTRUCTIONS
11. Disposal Instructions
11.1 X-ray Tube and CCD Detector
The X-ray tube and CCD detectors have beryllium windows. Dispose of Beryllium in accordance with local government regulations.
11.2 Third Party Equipment
Refer to third party manuals for information about disposing of third party equipment.
Xcalibur/Gemini Version 2.4
USER MANUAL Page 53
ADDITIONAL INFORMATION
Title
Supplier
Material Safety Data Sheet M10 Beryllium
Brushwellman
KMW200CCD chiller manual
Agilent Technologies
KMW3000C chiller manual
Agilent Technologies
Cryojet User Manual
Oxford Instruments
optional
Cryostream User Manual
Oxford Cryostreams
optional
Cobra User Manual
Oxford Cryostreams
optional
Helijet User Manual
Agilent Technologies
optional
12. Additional Information
12.1 Third Party Information
Information marked optional is only supplied when that option is fitted to the system.
Xcalibur/Gemini Version 2.4
USER MANUAL Page 54
12.2 Drawings
12.2.1. Mechanical Drawings
ADDITIONAL INFORMATION
Figure 12.2.1 Xcalibur / Gemini Suggested Layout
Xcalibur/Gemini Version 2.4
USER MANUAL Page 55
Xcalibur / Gemini Dimensions
115
200
116
80
100
80
54
37 75
96
ADDITIONAL INFORMATION
KMW3000C Dimensions
Figure 12.2.2 System and Component Dimensions
Xcalibur/Gemini Version 2.4
USER MANUAL Page 56
12.2.2. Electrical Drawings
Drawing no
Title
Number of pages
EW-02-00-001-A
KMW150CCD / KMW200CCD ELECTRICAL BLOCK SCHEME
1
EC-02-01-001-A
XCALIBUR / GEMINI STEPPING MOTORS CABLE
2
EC-24-01-001-A
XCALIBUR / GEMINI ENCODERS CABLE
2
EC-01-02-001-A
CCD POWER SUPPLY 230VAC SUPPLY CABLE
2
EW-03-02-001-A
KMW3000C ELECTRICAL CONNECTION SCHEME
1
EC-04-02-001-A
XCALIBUR / GEMINI CCD POWER SUPPLY TO PC (RS232)
2
EC-02-05-001-B
LCD MONITOR AND VIDEO MICROSCOPE POWER SUPPLY CABLE
2
EC-02-05-001-B
LCD MONITOR AND VIDEO MICROSCOPE POWER SUPPLY CABLE
2
EC-24-08-001-A
XCALIBUR / GEMINI REMOTE CONTROL CABLE
2
EC-04-09-001-A
XCALIBUR / GEMINI INTERFACE RS232 CABLE
2
EC-04-02-002-A
XCALIBUR / GEMINI FIP60 TO INTERFACE CABLE
2
EC-01-09-002-A
PC 230VAC SUPPLY CABLE
2
EC-04-02-003-B
CCD DETECTOR TEMPERATURE AND COLLISION
2
EF-00-01-001-A
XCALIBUR / GEMINI FAN PANEL- ELECTRICAL CONNECTION
1
EG-01-02-000-A
ALL SYSTEMS INSTALLATION
OD-1-04-01
CCD DETECTOR CONNECTIONS
1
OD-1-07-01
VIDEO MICROSCOPE KIT – BLOCK DIAGRAM
1
OD-1-07-02
EC-05-05-002-B VIDEO MICROSCOPE TO BNC „T‟ ADAPTOR CABLE
1
OD-1-07-03
EC-05-05-001-B BNC „T‟ ADAPTOR TO FRAME GRABBER CABLE
1
OD-1-07-04
EC-05-05-003-A BNC „T‟ ADAPTOR TO MONITOR CABLE
1
BCK-005-1
DETECTOR HEAD SUPPLY CABLE
1
BCK-006-1
DETECTOR HEAD COOLER CABLE
1
ADDITIONAL INFORMATION
Xcalibur/Gemini Version 2.4
USER MANUAL Page 57
EG-01-00-013-A
GEMINI ELECTRICAL PANEL
1
EC-24-07-000-A
GEMINI FRONT ELECTRICAL PANEL
1
EC-24-07-010-A
GEMINI FRONT PANEL
1
EG-01-00-001-A
GEMINI PCB CONNECTIONS
1
EG-01-00-002-A
GEMINI PCB CONNECTIONS
1
EG-01-00-003-A
GEMINI PCB CONNECTIONS
1
EG-01-00-004-A
GEMINI PCB CONNECTIONS
1
EG-01-00-006-A
GEMINI PCB CONNECTIONS
1
EG-01-00-007-A
GEMINI PCB CONNECTIONS
1
EG-01-00-009-A
GEMINI PCB CONNECTIONS
1
EG-01-00-010-A
GEMINI PCB CONNECTIONS
1
EG-01-00-011-A
GEMINI PCB CONNECTIONS
1
EC-24-11-001-B
GEMINI PROTECTION CABINET SENSORS
1
EG-01-00-005-A
GEMINI WARNING LAMPS
1
EG-01-00-012-A
GEMINI WATER PANEL
1
Gemini
ADDITIONAL INFORMATION
Xcalibur/Gemini Version 2.4
USER MANUAL Page 58
Appendices
Anode
Cu
1.540598(2)
1.544426(2)
1.39225(1)
1.00248
Mo
0.7093187(4)
0.713609(6)
0.632305(9)
1.00605
Compound name:
Ylid (2-Dimethylsulfuranylidene-1,3-indandione)
Compound formula:
(C11H10SO2)
Cell constants:
a= 5.947 Å
b= 9.026 Å
c= 18.399 Å
= = =90
To convert Fahrenheit to Celsius:
subtract 32 from F then multiply by 5/9
C=5/9(F-32) To convert Celsius to Fahrenheit:
multiply by 9/5 then add 32
F=(9/5C)+32
To convert Celsius to Kelvin:
add 273.15
K=C+273.15
To convert Fahrenheit to Kelvin
first convert F to C then add 273.15
Appendix 1 X-ray Tubes Wavelengths
Appendix 2 Standard Crystal Parameters
ADDITIONAL INFORMATION
Appendix 3 Temperature Scales Conversion
Xcalibur/Gemini Version 2.4
USER MANUAL Page 59
ADDITIONAL INFORMATION
Weekly Maintenance Record Sheet
Week No.
Date of maintenance
Name of person performing test
Signature
Comments
1
2
3 4 5 6 7 8
9
10
11
12
13
14
15
16
17
18
19
Appendix 4 Maintenance Records
The attached record sheets are provided as examples of the type of records of maintenance checks that should be completed for Xcalibur/Gemini. These records are to show only that maintenance has been completed as directed in this manual. Other maintenance records should be kept as required by any local, national or international regulations.
Xcalibur/Gemini Version 2.4
USER MANUAL Page 60
ADDITIONAL INFORMATION
Weekly Maintenance Record Sheet
Week No.
Date of maintenance
Name of person performing test
Signature
Comments
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
Xcalibur/Gemini Version 2.4
USER MANUAL Page 61
ADDITIONAL INFORMATION
Weekly Maintenance Record Sheet
Week No.
Date of maintenance
Name of person performing test
Signature
Comments
43
44
45
46
47
48
49
50
51
52 Monthly Maintenance Record Sheet
Month
Date of maintenance
Name of person performing test
Signature
Emergency Stop Check OK? Y/N
X-ray radiation check OK? Y/N
Comments
1
2
3
4
5
6
USER MANUAL Page 62
Xcalibur/Gemini Version 2.4
ADDITIONAL INFORMATION
7
8
9
10
11
12
Six Monthly Maintenance Record Sheet
Date of maintenance
Name of person performing test
Signature
Comments
Yearly Maintenance Record Sheet
Date of maintenance
Name of person performing test
Signature
Machine parameter file refined? Y/N
CCD detector vacuum pumped? Y/N
Comments
Xcalibur/Gemini Version 2.4
USER MANUAL Page 63
Loading...