Agilent Technologies N3970A User Manual

N3970A OPTICAL POWER METER
Operating and
Maintenance Guide
N3970A OPTICAL POWER METER SOURCE QUICK REFERENCE GUIDE
To remove interchangeable connector, move interface to
mid position, and pull off adaptor.
To defeat auto power-off, hold POWER for 3 seconds at
turn on until ON and perm are displayed.
Select POWER, Source, then:
To set reference, push Abs/Rel and hold Set Ref for
3 seconds (3 buzzer sounds).
To display reference, hold Abs/Rel, so Ref flashes.
To stop / start display update, push Hold.
If a test tone is detected, meter will display Hz.
AUTOTEST OPERATION
No user action is required for meter to enter Autotest
mode.
Autotest:
seconds.
Can also push Autotest on meter for faster lock-on.
To display one wavelength only, on the power meter
push the left side of -/+ and again for the other wavelength. Push right side of -/+ (twice) to exit.
Autotest remains for >8 seconds after signal loss, to allow
quick connection change.
To e x i tautotest, remove test lead or turn off source for >8 seconds.
Push Autotest on source and wait a few
MANUAL OPERATION
To turn on and select wavelength, push: POWER, Power
Meter, -/+.
To display linear mode, push dBm/W.
To display max/min values, push MAX/MIN repeatedly.
Hold MAX/MIN to reset (3 buzzer sounds). .
MEMORY
Enter Power Meter or Autotest modes first
While in Autotest mode, it is only possible to STORE.
To clear all memory, push both RECALL and CANCEL
together for 3 seconds. clr will display.
To store in next location, push STORE.
To change next store location, push SELECT -/+, SET.
To recall, push RECALL -/+ to scroll through data.
To exit memory display, push CANCEL.
ADDITIONAL AUTOTEST MEMORY OPERATION
To read or cancel memory, first exit Autotest mode, and
enter Power Meter mode.
Autotest memory: use -/+ to scroll wavelengths.
INTRODUCTION AND A P P L I C AT I O N S
N3970A
OPTICAL POWER METER
Congratulations on your purchase of this instrument, which has been engineered to provide the best possible reliability, convenience and performance. To get the best use from your equipment and ensure its safe operation, please spend a few minutes to read this manual.
1
CO N T E N T S
Service and Support 3 Introduction & applications 4-5 Safety Summary 6-7 Battery and external power 8 Optical connector 9-10 How to use: Getting started and turning on 11-12 Autotest operation 13-14 Power meter operation 15 Storing & recalling readings 16-17 RS232 interface 18
Care of your instrument 19 Accuracy considerations 20-21 Calibration, maintenance & instrument re t u rn s 22-24 Specifications - Definition of terms 25 Specifications and ordering information 26-28 Performance test 29-38 Cleaning information 39-40 Cleaning procedures 41 Disclaimer & warranty 42 Notes 43-44
BEFORE READING FURTHER, OPEN THE CONTROL PANEL PICTURE AT THE BACK OF THE MANUAL
2
S E RVICE AND SUPPORT
Any adjustment, maintenance, or repair of this product must be p e rf o rmed by qualified personnel. Contact your customer engineer through your local Agilent Technologies Serv i c e Center. You can find a list of local service representatives on the Web at: http://www.agilent.com/find/assist
If you do not have access to the Internet, one of these centers can direct you to your nearest representative:
United States 1 800 452 4844 Canada 1 877 894 4414
(905) 206 4120 (Fax)
Europe (31 20) 547 2323
(31 20) 547 2390 (Fax)
Japan (81) 426 56 7832
(81) 426 56 7840 (Fax)
Latin America (305) 269 7500
(305) 269 7599 (Fax)
N3970A UM-1
Australia 1 800 629 485
(613) 9272 0749 (Fax)
New Zealand 0800 738 378
64 4 495 8950 (Fax)
Asia-Pacific (852) 3197 7777
(852) 2506 9284 (Fax)
3
IN T R O D U C T I O N A N D AP P L I C AT I O N S
The N3970A Optical Power Meter is used for field or laboratory testing of fiber optic systems at typically 850, 1300, 1310 and 1550 nm. It features superb measurement confidence, ease of use and reduced cost of ownership, and is for use by installers, technicians and engineers.
Superior measurement confidence is achieved with a unique Total Uncertainty specification over the full temperature , measurement and connector range, and a NATA traceable calibration certificate. The display hold and back light facilities help acquire readings, and the flexible memory helps record your results
Autotest mode enables automatic λ (wavelength) detection and attenuation measurement (in combination with an Autotest source). A single or dual λsource and meter in Autotest mode automatically measure single or dual λ attenuation in one direction.
4
High operational reliability & availability is achieved with 250 hour battery life, 3 year recommended re-calibration cycle, negligible warm up and no range-changing delays. Dark current offset is not required. An external power input is also provided.
The combination of features enables reduced equipment inventory, simpler training, shorter procedures, lower skill levels and direct labour savings in many situations, combining to give lower cost of ownership.
The innovative optical connector port provides drop protection during use, and is dust protected by a snap on cover. The interchangeable optical connector system is very versatile, and easily disassembled for cleaning.
The simple user interface guides the operator through functional sequences, and a hidden keypad is used to access advanced features such as log/linear display, and memory functions.
The tone detector is very sensitive, and is useful for fiber identification, and calibrating the modulation frequency of test tone sources.
The instrument has shock absorbent corners, and a tough polycarbonate housing which has passed extensive drop testing.
Typical applications
Singlemode testing, with versatile interc h a n g e a b l e connectors.
Telecom construction, maintenance, CATV, LAN and R&D applications.
Measurements where traceability and documentation are required.
Quality assurance and acceptance tests.
Automated testing and re p o rting using optional Excel software .
Test tone measurement & fiber identification.
N3970A UM-1
IN T R O D U C T I O N A N D AP P L I C AT I O N S
5
SAFETY SUMMARY
The following general safety precautions must be observed during all phases of operation, service, and repair of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design, manufacture, and intended use of the instrument. Agilent Technologies Inc. assumes no liability for the customer’s failure to comply with these requirements.
Before operation, review the instrument and manual, for safety markings and instructions. You must follow these to ensure safe operation and to maintain the instrument in safe condition.
WARNING
The WARNING sign denotes a hazard. It calls attention to a procedure, practice or the like, which, if not correctly performed or adhered to, could result in injury or loss of life. Do not proceed beyond a WARNING sign until the indicated conditions are fully understood and met.
6
CAUTION
The CAUTION sign denotes a hazard. It calls attention to an operating pro c e d u re, or the like, which, if not corre c t l y p e rf o rmed or adhered to, could result in damage to or destruction of part or all of the product. Do not proceed beyond a CAUTION sign until the indicated conditions are fully understood and met.
CAUTION
m
The apparatus will be marked with this symbol when it is necessary for the user to refer to the instruction manual in order to protect the apparatus against damage.
SAFETY SUMMARY
Initial Inspection
Inspect the shipping container for damage. If there is damage to the container or Cushioning, keep them until you have checked the contents of the shipment for completeness and verified the instrument both mechanically and electrically.
The Perf o rmance Tests give pro c e d u res for checking the operation of the instrument. If the contents are incomplete, mechanical damage or defect is apparent, or if an instrument does not pass the operator’s checks, notify the nearest Agilent Technologies Sales/Service Office (see page 3).
WARNING
To avoid any hazard, do not perform electrical tests when there are signs of shipping damage to any portion of the outer enclosure (covers, panels, and so on).
WARNING
You MUST return instruments with malfunctions to an Agilent Technologies Service Center for repair and calibration (see page 3).
N3970A UM-1
Line Power Requirements
The Agilent N3970A Power Meter operates at line power when applied to the optional external power supply N3979A 9V DC
Operating Environment
The Agilent N3970A Power Meter can be operated at temperatures between - 10°C and + 55°C and at relative
humidity of <95%.
Storage and Shipment
The Agilent N3970A Power Meter can be stored or shipped at temperatures between - 25°C and + 70°C and at relative humidity of less than 95%. Protect the module fro m temperature extremes that may cause condensation within it.
REQUIREMENT COMPLIANCE
N 10149
7
BATTERY AND EXTERNAL POWER
The instrument is powered by two 1.5V dry batteries (Alkaline C)for an operating period of typically 250 hours or two 1.2V rechargeable batteries for an operating period of approximately 125 hours.
Do not use lithium batteries or other batteries with a nominal voltage greater than 1.8 V. The instrument will be damaged.
When the batteries are low, the low-battery indicator is shown on the display. At this stage, there is approximately enough energy for another 30 minutes of use.
For saving of energy the instrument automatically turns off after 10 minutes without operation.
For changing the battery, open the cover of the battery compartment at the base of the instrument, remove the batteries, insert new batteries (check polarity) and close the cover again.
Rechargeable batteries must be charged by an external charger.
8
For operation of the instrument by external power use Agilent AC adaptor N3979A (9V DC, 1.33 A, ).
Use of the external power supply disconnects the batteries. Ensure that the mains supply has the correct voltage rating and
safety compliance as relevant. Do not use non-compliant or doubtful mains equipment.
OPTICAL CONNECTOR
The optical port is mounted on a swivel, which allows the connector to be angled outwards for accessibility, and then pushed back and covered with the snap cover to provide dirt and drop protection.
The other port can be used to store a spare interchangeable optical connector.
To install the connector adaptor, align the locating slot on the side of the adaptor with that on the instrument connector, and push it on.
To remove the connector adaptor, move the connector port to it’s mid-way point (see dotted line on diagram at back of manual), and then pull off the adaptor. It is not possible to remove the adaptor in the ‘open’ or ‘shut’ positions. With some styles of connector, it is easier to pull off the adaptor with a test lead in place, since this gives better grip.
Different styles of connector adaptor ( ST, SC, FC, ) can be easily fitted by the user.
N3970A UM-1
Optical connectors are precision components, and require care in use:
Always clean the mating connector tip and ferrule before
mating, using approved materials.
This instrument can be used with both PC and APC
connector styles (unlike sources, which are factory configured for PC only ).
The soft carry case enables the instrument to be stored
without removing the test leads. This practice saves time, and is to be recommended.
When not in use, keep all test ports and connectors covered
and away from dust.
Do not touch connector tips with your fingers,
since body oils and dirt can impair connector performance.
Do not use damaged or incompatible connectors.
9
OPTICAL CONNECTOR
Cleaning
The glass power meter interface does not make contact with the inserted connector: there is a slight air gap. Therefore it should not wear out, and only needs occasional cleaning.
To clean the glass power meter interface: remove the interchangeable adaptor to access the glass interface, then use a sticky material such as ‘ Blu tac ‘ to dab the glass and remove any dirt. A soft brush, alcohol, or air can are also appropriate. Do not use anything hard that could scratch the glass.
In this manual, refer to the sections on Cleaning Information and Procedures for further information.
Bare fiber adaptors
Bare fiber adaptors used with this power meter must achieve fiber eccentricity of +_ 100 microns and end tolerance of
+
_ 0.3 mm, relative to the ferrule end. Be careful not to scratch the detector interface when using adaptors, since permanent
10
meter inaccuracy may result. Preferred bare fiber adaptors consist of a connector with fiber retention device. If you work regularly with bare fibers, to avoid potential damage to the meter, it is generally preferable to use an alternative a rrangement such as a multimode pigtail with a v-groove or mechanical splice.
GETTING STA RTED AND TURNING ON
This and following sections show you how to use your instrument:
Check that the correct accessories have been supplied, and an optical connector adaptor has been fitted. If you have any queries, call your supplier to resolve the situation.
Put in the batteries, or plug external power into the socket at the top of the instrument (see Battery and External Power section).
Push the green POWER button. The display will come on. If the battery is low, this will be indicated on the display. Pushing POWER again will turn the instrument off.
To stop the instrument turning off 10 minutes after the last key press, hold POWER down for 3 seconds during turn-on. PERM on the display indicates that the unit will stay on permanently.
To back light the display temporarily, select LIGHT. To back light continuously, hold down the button for 3 seconds. Use of the back light will drain the batteries faster.
N3970A UM-1
This instrument uses a menu driven keypad. The function of each button is shown on the display.
To access the optical connector, grasp a top corner of the instrument, and pull off the cover. The connector can be swivelled to improve access.
After instrument turn on, the instrument will perform a self­calibration sequence, set the dark current adjustment, and then display absolute power in dBm at the previously set λ. If HI or LO are displayed, the input is out of range.
The hidden keypad is accessed by pulling up the hinged display cover. The hidden keypad can operate in two modes. The second mode is accessed by SHIFT. Access to the second mode can be denied / permitted by typing the button sequence : POWER (turn on instrument), SHIFT, CANCEL, SET, SELECT, RECALL, STORE.
11
GETTING STA RTED AND TURNING ON
To disable or enable the buzzer, hold down SHIFT for 3 seconds. To add the carry strap: slip the end of the strap through the slit
on the curved section at the back of the instrument, and secure the buckle.
This instrument requires no warm up, and no user adjustment of dark current to achieve it’s specified performance.
To learn how to use the instrument features, go to the next section.
Should the instrument fail to turn on, the microprocessor may need re-booting. To do this, remove batteries and external power for at least 40 seconds. This will not lose stored data.
12
Loading...
+ 32 hidden pages