Agilent Technologies N3280A User Manual

USER’S GUIDE
Agilent Technologies
Model N3280A
Component Test DC Source
5
Agilent Part No. 5964-8248
Microfiche No. 5964-8249
June, 2001

Warranty Information

CERTIFICATION
Agilent Technologies certifies that this product met its published specifications at time of shipment from the factory. Agilent Technologies further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology, to the extent allowed by the Bureau's calibration facility, and to the calibration facilities of other International Standards Organization members.
WARRANTY
This Agilent Technologies hardware product is warranted against defects in material and workmanship for a period of three years from date of delivery. Agilent Technologies software and firmware products, which are designated by Agilent Technologies for use with a hardware product and when properly installed on that hardware product, are warranted not to fail to execute their programming instructions due to defects in material and workmanship for a period of 90 days from date of delivery. During the warranty period Agilent Technologies will, at its option, either repair or replace products which prove to be defective. Agilent does not warrant that the operation for the software firmware, or hardware shall be uninterrupted or error free.
For warranty service, with the exception of warranty options, this product must be returned to a service facility designated by Agilent Technologies. Customer shall prepay shipping charges by (and shall pay all duty and taxes) for products returned to Agilent Technologies for warranty service. Except for products returned to Customer from another country, Agilent Technologies shall pay for return of products to Customer.
Warranty services outside the country of initial purchase are included in Agilent Technologies' product price, only if Customer pays Agilent Technologies international prices (defined as destination local currency price, or U.S. or Geneva Export price).
If Agilent is unable, within a reasonable time to repair or replace any product to condition as warranted, the Customer shall be entitled to a refund of the purchase price upon return of the product to Agilent Technologies.
LIMITATION OF WARRANTY
The foregoing warranty shall not apply to defects resulting from improper or inadequate maintenance by the Customer, Customer-supplied software or interfacing, unauthorized modification or misuse, operation outside of the environmental specifications for the product, or improper site preparation and maintenance. NO OTHER WARRANTY IS EXPRESSED OR IMPLIED. AGILENT TECHNOLOGIES SPECIFICALLY DISCLAIMS THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
EXCLUSIVE REMEDIES
THE REMEDIES PROVIDED HEREIN ARE THE CUSTOMER'S SOLE AND EXCLUSIVE REMEDIES. AGILENT TECHNOLOGIES SHALL NOT BE LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES, WHETHER BASED ON CONTRACT, TORT, OR ANY OTHER LEGAL THEORY.
ASSISTANCE
The above statements apply only to the standard product warranty. Warranty options, extended support contacts, product maintenance agreements and customer assistance agreements are also available. Contact your nearest Agilent Technologies Sales and Service office for further information on Agilent Technologies' full line of Support Programs.

Safety Summary

F
y
s
y
The following general safety precautions must be observed during all phases of operation of this instrument.
ailure to comply with these precautions or with specific warnings elsewhere in this manual violates safet
tandards of design, manufacture, and intended use of the instrument. Agilent Technologies assumes no liabilit
for the customer's failure to comply with these requirements.
GENERAL
This product is a Safety Class 1 instrument (provided with a protective earth terminal). The protective features of this product may be impaired if it is used in a manner not specified in the operation instructions. Any LEDs used in this product are Class 1 LEDs as per IEC 825-1.
ENVIRONMENTAL CONDITIONS
This instrument is intended for indoor use in an installation category II, pollution degree 2 environment. It is designed to operate at a maximum relative humidity of 95% and at altitudes of up to 4500 meters. Refer to the specifications tables for the ac mains voltage requirements and ambient operating temperature range.
BEFORE APPLYING POWER
Verify that the product is set to match the available line voltage, the correct fuse is installed, and all safety precautions are taken. Note the instrument's external markings described under "Safety Symbols".
GROUND THE INSTRUMENT
To minimize shock hazard, the instrument chassis and cover must be connected to an electrical ground. The instrument must be connected to the ac power mains through a grounded power cable, with the ground wire firmly connected to an electrical ground (safety ground) at the power outlet. Any interruption of the protective (grounding) conductor or disconnection of the protective earth terminal will cause a potential shock hazard that could result in personal injury.
ATTENTION: Un circuit de terre continu est essentiel en vue du fonctionnement sécuritaire de l'appareil. Ne jamais mettre l'appareil en marche lorsque le conducteur de mise … la terre est d‚branch‚.
FUSES
Only fuses with the required rated current, voltage, and specified type (normal blow, time delay, etc.) should be used. Do not use repaired fuses or short-circuited fuseholders. To do so could cause a shock or fire hazard.
Vous devrez impérativement utiliser des fusibles calibrés aux spécifications de courant, tension et type (coupure, délai de coupure, etc ...). N'utilisez jamais de fusibles réparés et ne court-circuitez pas les supports de fusibles. Sinon, vous risquez de provoquer un choc électrique ou un incendie.
DO NOT OPERATE IN AN EXPLOSIVE ATMOSPHERE
Do not operate the instrument in the presence of flammable gases or fumes.
DO NOT REMOVE THE INSTRUMENT COVER
Operating personnel must not remove instrument covers. Component replacement and internal adjustments must be made only by qualified service personnel.
Instruments that appear damaged or defective should be made inoperative and secured against unintended operation until they can be repaired by qualified service personnel.
SAFETY SYMBOLS
Direct current
Alternating current
Both direct and alternating current
Three-phase alternating current
Earth (ground) terminal
Protective earth (ground) terminal
Frame or chassis terminal
Terminal is at earth potential. Used for measurement and control circuits designed to be operated with one terminal at earth potential.
Terminal for Neutral conductor on permanently installed equipment
Terminal for Line conductor on permanently installed equipment
WARNING
Caution
On (supply)
Off (supply)
Standby (supply). Units with this symbol are not completely disconnected from ac mains when this switch is off. To completely disconnect the unit from ac mains, either disconnect the power cord or have a qualified electrician install an external switch.
In position of a bi-stable push control
Out position of a bi-stable push control
Caution, risk of electric shock
Caution, hot surface
Caution (refer to accompanying documents)
The WARNING sign denotes a hazard. It calls attention to a procedure, practice, or the like, which, if not correctly performed or adhered to, could result in personal injury. Do not proceed beyond a WARNING sign until the indicated conditions are fully understood and met.
The CAUTION sign denotes a hazard. It calls attention to an operating procedure, or the like, which, if not correctly performed or adhered to, could result in damage to or destruction of part or all of the product. Do not proceed beyond a CAUTION sign until the indicated conditions are fully understood and met.

Declaration Page

Manufacturer’s Name: Manufacturer’s Address:
declares that the product:
Product Name:
DECLARATION OF CONFORMITY
According to ISO/IEC Guide 22 and CEN/CENELEC EN 45014
Responsible Party Alternate Manufacturing Site Agilent Technologies, Inc. Agilent Technologies Power Products PGU 140 Green Pond Road Rockaway, New Jersey 07866 U.S.A
Component Test dc Source
South Queensferry West Lothian EH30 9TG United Kingdom
Model Number:
Product Options:
Conforms with the following European Directives:
The product herewith complies with the requirements of the Low Voltage Directive 73/23/EEC and the EMC Directive 89/336/EEC (including 93/68/EEC) and carries the CE Marking accordingly
EMC information:
The product herewith complies with the requirements of the EMC Directive 89/336/EEC (including 93/68/EEC) and carries the CE Marking accordingly (European Union).
As detailed in
Assessed by:
Safety information:
The product herewith complies with the requirements of the Low Voltage Directive 73/23/EEC and carries the CE-marking accordingly
N3280A
This declaration covers all options of the above product(s).
Electromagnetic Compatibility (EMC) Certificate of Conformance No.TCF CC/TCF/01/016 based on Technical Construction File (TCF) No. ANJ13, dated 8/03/2001
Celestica Ltd, Appointed Competent Body Westfields House, West Avenue Kidsgrove, Stoke-on-Trent Straffordshire, ST7 1TL United Kingdom
Supplemental information
The product conforms to the following safety standards:
IEC 1010-1:1990+A1+A2 / EN 61010-1:1993 +A2 UL 3111-1:1994 CSA C22.2 No. 1010.1:1993
March 19, 2001 Date Hank Kowalla / Quality Manager at PPPGU
For further information, please contact your local Agilent Technologies sales office, agent or distributor. Authorized EU-representative: Agilent Technologies Deutschland GmbH, Herrenberger Stra Böblingen, Germany
βe 130, D71034

Acoustic Noise Information

Herstellerbescheinigung
Diese Information steht im Zusammenhang mit den Anforderungen der Maschinenläminformationsverordnung vom 18 Januar 1991.
* Schalldruckpegel Lp <70 dB(A) * Am Arbeitsplatz * Normaler Betrieb * Nach EN 27779 (Typprüfung).
Manufacturer's Declaration
This statement is provided to comply with the requirements of the German Sound Emission Directive, from 18 January 1991.
* Sound Pressure Lp <70 dB(A) * At Operator Position * Normal Operation * According to EN 27779 (Type Test).
Printing History
The edition and current revision of this manual are indicated below. Reprints of this manual containing minor corrections and updates may have the same printing date. Revised editions are identified by a new printing date. A revised edition incorporates all new or corrected material since the previous printing date.
Changes to the manual occurring between revisions are covered by change sheets shipped with the manual. In some cases, the manual change applies only to specific instruments. Instructions provided on the change sheet will indicate if a particular change applies only to certain instruments.
This document contains proprietary information protected by copyright. All rights are reserved. No part of this document may be photocopied, reproduced, or translated into another language without the prior consent of Agilent Technologies. The information contained in this document is subject to change without notice.
Copyright 2001 Agilent Technologies, Inc. Edition 1 March, 2001 Update 1 June, 2001

Table of Contents

Warranty Information 2 Safety Summary 3 Declaration Page 5 Acoustic Noise Information 6 Printing History 6 Table of Contents 7
GENERAL INFORMATION 13
Document Orientation 13 Safety Considerations 13 Options and Accessories 14 Description 14
Remote Programming 14
Output Characteristics 15
Voltage Priority Operation 15 Current Priority Operation 16
Measurement Characteristics 17
Start of a Measurement 18
INSTALLATION 19
Inspection 19
Damage 19 Packaging Material 19 Additional Items 19 Cleaning 19
Location 20
Bench Operation 20 Rack Mounting 20
Power Connections 21
Connect the Power Cord 21
Output Connections 22
Outputs 1 - 4 22 Current Ratings 23 Voltage Drops and Lead Resistance 23 Coaxial Guard Connections 24 Maintaining Stability 25 OVP Considerations 25
External Trigger Connections 25 Computer Connections 25
GPIB Interface 25 GPIB Address 26
TURN-ON CHECKOUT 27
Front Panel Description 27 Checkout Procedure 28 In Case of Trouble 29
Selftest Error Messages 29 Runtime Error Messages 29 Line Fuse 29
INTRODUCTION TO PROGRAMMING 31
External References 31
GPIB References 31 SCPI References 31
GPIB Capabilities of the DC Source 31 Introduction to SCPI 32
Conventions Used in This Guide 32
Types of SCPI Commands 32
Multiple Commands in a Message 33 Moving Among Subsystems 33 Including Common Commands 33 Using Queries 33
Types of SCPI Messages 34
The Message Unit 34 Channel List Parameter 34 Headers 35 Query Indicator 35 Message Unit Separator 35 Root Specifier 35 Message Terminator 35
SCPI Data Formats 35
Numerical Data Formats 35 Suffixes and Multipliers 36 Response Data Types 36
SCPI Command Completion 36 Using Device Clear 37
PROGRAMMING THE DC SOURCE 39
Introduction 39 Programming the Output 39
Power-on Initialization 39 Enabling the Output 39 Output Voltage 39 Overvoltage Protection 40 Output Current 40 Output Mode 40 Oscillation Protection 41
Triggering Output Changes 41
Output Trigger Model 41 Setting the Voltage and Current Trigger Levels 42 Enabling the Output Trigger System 42 Selecting the Output Trigger Source 42 Generating Output Triggers 43
Making Measurements 43
Average Measurements 43 Power Line Cycles 44 Measurement Samples and Time Interval 44 Current Ranges 45 Window Functions 45 Returning All Measurement Data From the Data Buffer 45
Triggered Measurements 45
Measurement Trigger Model 46 Enabling the Measurement Trigger System 46 Selecting the Measurement Trigger Source 46 Selecting the Sensing Function 47 Output Settling Delay 47 Generating Measurement Triggers 47
Pre-trigger and Post-trigger Data Acquisition 48
Programming the Status Registers 49
Operation Status Group 50 Questionable Status Group 51 Standard Event Status Group 51 Status Byte Register 51 Determining the Cause of a Service Interrupt 52 Servicing Operation Status and Questionable Status Events 52
LANGUAGE DICTIONARY 53
Introduction 53
Subsystem Commands 53 Common Commands 53 Programming Parameters 53
SCPI Programming Commands - At a Glance 54 Calibration Commands 57
CALibrate:CURRent 57 CALibrate:CURRent:LIMit[:POSitive] CALibrate:CURRent:LIMit:NEGative 57 CALibrate:CURRent:MEASure 57 CALibrate:DATA 58 CALibrate:DATE 58 CALibrate:LEVel 58 CALibrate:PASSword 58 CALibrate:SAVE 59 CALibrate:STATe 59 CALibrate:VOLTage 59
Measurement Commands 60
FETCh:ARRay:CURRent? FETCh:ARRay:VOLTage? 60 FETCh:CURRent? FETCh:VOLTage? 60 MEASure:ARRay:CURRent? MEASure:ARRay:VOLTage? 61 MEASure:CURRent? MEASure:VOLTage? 61 SENSe:CURRent:RANGe 61 SENSe:FUNCtion 62 SENSe:SWEep:NPLCycles 62 SENSe:SWEep:OFFSet:POINts 62 SENSe:SWEep:POINts 63 SENSe:SWEep:TINTerval 63 SENSe:WINDow 63
Output Commands 64
OUTPut 64 OUTPut:OSCProtect 64 OUTPut:PROTection:CLEar 64 [SOURce:]CURRent[:IMMediate] [SOURce:]CURRent:TRIGgered 65 [SOURce:]CURRent:LIMit[:IMMediate] [SOURce:]CURRent:LIMit:TRIGgered 65 [SOURce:]CURRent:LIMit:BWIDth 65 [SOURce:]CURRent:MODE [SOURce:]CURRent:LIMit:MODE 66 [SOURce:]DELay 66 [SOURce:]DELay:MODE 66 [SOURce:]FUNCtion:MODE 67 [SOURce:]VOLTage:ALC:BWIDth 67 [SOURce:]VOLTage[:IMMediate] [SOURce:]VOLTage:TRIGgered 67 [SOURce:]VOLTage:MODE 68 [SOURce:]VOLTage:PROTection:STATe 68
Status Commands 69
STATus:OPERation[:EVENt]? 69 STATus:OPERation:CONDition? 69 STATus:OPERation:ENABle 69
STATus:OPERation:NTR STATus:OPERation:PTR 70 STATus:PRESet 70 STATus:QUEStionable[:EVENt]? 70 STATus:QUEStionable:CONDition? 71 STATus:QUEStionable:ENABle 71 STATus:QUEStionable:NTR STATus:QUEStionable:PTR 71
System Commands 72
SYSTem:ERRor? 72 SYSTem:VERSion? 72
Trigger Commands 73
ABORt 73 INITiate:NAME 73 TRIGger:ACQuire 73 TRIGger:ACQuire:SOURce 74 TRIGger[:TRANsient]:SOURce 74 TRIGger[:TRANsient] 74
Common Commands 75
*CLS 75 *ESE 75 *ESR? 75 *IDN? 76 *OPC 76 *OPT? 76 *RST 77 *SRE 77 *STB? 77 *TRG 78 *TST? 78 *WAI 78
SPECIFICATIONS 79
Introduction 79
PERFORMANCE TESTS AND CALIBRATION 83
Introduction 83 Equipment Required 83 Performance & Verification Tests 84
Measurement Techniques 84 Electronic Load 85 Programming 85 Test Setup 85
Voltage Priority Tests 86
Voltage Programming and Readback Accuracy 86 Positive Current Limit (+CL) 86 Negative Current Limit (-CL) 87
Current Priority Tests 88
Current Programming and Readback Accuracy 88
Load Effect Tests 89
Voltage Priority, Constant Voltage Load Effect 89 Voltage Priority, +Current Limit Load Effect 90 Voltage Priority, -Current Limit Load Effect Test 90 Current Priority Constant Current Test 91
Source Effect Tests 91
Voltage Priority, Constant Voltage Source Effect 91 Voltage Priority, +Current Limit Source Effect 92 Voltage Priority, -Current Limit Source Effect 92
10
Current Priority, Constant Current Source Effect 93
Ripple and Noise Tests 94
Voltage Priority Ripple and Noise 94 Current Priority Ripple and Noise 95
Transient Response Tests 95
Voltage Priority, Transient Recovery Time 95 Current Priority Transient Recovery Time 96
Performance Test Equipment Form 97 Performance Test Record Form 98 Performing the Calibration Procedure 99
Enable Calibration Mode 99 Voltage Priority Mode Programming and Measurement Calibration 99 Negative Current Limit Calibration 100 Positive Current Limit Calibration 100
0.5A Range Current Measurement Calibration 100 15mA Range Current Measurement Calibration 101 Current Priority Mode Programming and 0.5mA Range Measurement Calibration 101 Saving the Calibration Constants 101 Changing the Calibration Password 102 Calibration Error Messages 102
ERROR MESSAGES 103
Error Number List 103
LINE VOLTAGE SELECTION 107
EARLIER VERSION OUTPUT CONNECTORS 109
Mating Connector Part Numbers 109 Rear Panel Pinout Assignments 109
INDEX 111
11
1

General Information

Document Orientation

This manual describes the operation of the Agilent Model N3280A Component Test DC Source. Unless otherwise noted, the unit will be referred to by the description "dc source" throughout this manual.
The following Getting Started Map is a general guide to the location of information in this manual. Refer to the table of contents or index for a complete list of information.
Getting Started Map
Task Where to find information
General information
Capabilities and characteristics
Installing the unit
Line connections Load connections Computer connections
Checking out the unit
Verifying proper operation
Using the programming interface
GPIB interface
Programming the unit using SCPI commands
SCPI commands SCPI programming examples SCPI language dictionary
Specifications
Verifying and Calibrating the Unit
Chapter 1
Chapter 2
Chapter 3
Chapter 4
Chapters 5 and 6
Appendix A
Appendix B

Safety Considerations

This dc source is a Safety Class 1 instrument, which means it has a protective earth terminal. That terminal must be connected to earth ground through a power source equipped with a ground receptacle. Refer to the Safety Summary page at the beginning of this guide for general safety information. Before installation or operation, check the dc source and review this guide for safety warnings and instructions. Safety warnings for specific procedures are located at appropriate places in the guide.
13
1 - General Information

Options and Accessories

Table 1-1. Options
Option Description
100 220 230
8ZL Add instrument feet - for bench mounting (p/n 5041-9167)
1
AXS
1
1CM
1
Support rails are required when rack mounting units. Use E3663A support rails for Agilent rack cabinets. If you are
using non-Agilent rack cabinets, contact the rack manufacturer to obtain support rails for your cabinet.
Item Part Number
GPIB cables 1.0 meter (3.3 ft) Agilent 10833A
Rack mount with slide - for two side-by-side units Order 5063-9255 and 1494-0015 Rack mount with slide - for one unit Order 5063-9255, 1494-0015, and 5002-3999
87106 Vac, 4763 Hz 191233 Vac, 4763 Hz 207253 Vac, 4763 Hz
Rack mount kit for two side-by-side N3280A units. Consists of: Lock-link kit (p/n 5061-9694), Flange kit (p/n 5063-9212), Tie bracket (p/n 5002-1587) Rack mount kit for one unit (p/n 5063-9240)
Table 1-2. Accessories
2.0 meters (6.6 ft) Agilent 10833B
4.0 meters (13.2 ft) Agilent 10833C
0.5 meters (1.6 ft) Agilent 10833D

Description

The Agilent Model N3280A Component Test DC Source is a quad output dc power supply designed to simplify the testing of integrated circuits. It has the following key features and performance capabilities:
High density – four isolated outputs in a 2U half-rack package Four quadrant bipolar output High programming and measurement accuracy (refer to Appendix A) Active guard available for accurate current measurements Solid-state output and sense terminal disconnect relays High GPIB throughput
Additional features include:
Positive and negative overvoltage protection shutdown Over-temperature and oscillation protection Programmable current limit in voltage priority mode

Remote Programming

NOTE: With the exception of the power switch, there are no front panel controls for the Agilent
N3280A dc source. The N3280A can be controlled only with SCPI programming commands.
The dc source may be remotely programmed via the GPIB bus. GPIB programming is with SCPI commands (Standard Commands for Programmable Instruments), which make dc source programs compatible with those of other GPIB instruments. Dc source status registers allow remote monitoring of a wide variety of operating conditions. Refer to chapters 5 and 6 for more information.
14
General Information - 1

Output Characteristics

Voltage Priority Operation

Each Agilent N3280A output is a four-quadrant bipolar dc source that can be operated in either voltage or current priority mode. In voltage priority mode the output is controlled by a bi-polar constant voltage feedback loop, which maintains the output voltage at its positive or negative programmed setting. The output voltage will remain at its programmed setting as long as the load current remains within the positive or negative current limit. A single positive value programs both the positive and negative current limit.
Figure 1-1 shows the voltage priority operating characteristics of the dc source. The area in quadrants 1 and 3 shows the characteristics of the output when it is being operated as a source (sourcing power). The area in quadrants 2 and 4 shows the characteristics of the output when it is being operated as a load (sinking power).
+ 10.25V
-
- 10.25V
- 512.5mA
+ OV
-I limit
Output Voltage
+
V setting
1+
324
+I limit
-
- OV
+ 512.5mA
Key
Sinking power
Sourcing power
Programmable
Output Current
Figure 1-1. Output Characteristic (Voltage Priority)
The heavy line illustrates the locus of possible operating points as a function of the output load, which may be purely resistive, or possibly include external voltage or current sources. In voltage priority mode, the constant voltage loop will regulate the output voltage as the load changes, unless the output current attempts to exceed the current limit setting.
If this occurs, either the negative or the positive current limit loop will regulate the output current at the programmed value. Either a CV (constant voltage), CL+ (positive current limit), or CL (negative current limit) status flag is set to indicate which loop is presently controlling the output.
15
1 - General Information
If the output voltage exceeds either the positive or negative overvoltage set point, the output will shut down and be disabled, automatically opening the output and sense relays. This leaves the output in a high-impedance state.
The full ±512.5 milliampere output current is available only in voltage priority mode. In this mode, the output voltage should be programmed to the desired positive or negative value. A positive current limit value should also be programmed. Note that the negative current limit tracks the positive current limit set point. The output will regulate at the desired voltage level, provided that the current limit has been set higher that the actual output current requirement of the external load. Note that if the current limit is set to a value between zero and 75 µA, the actual current limit will be ±75 µA. Thus, it is not possible to program current limit values less than 75 µA in voltage priority mode. (This limitation does not apply in current priority mode.)

Current Priority Operation

Each Agilent N3280A output is a four-quadrant bipolar dc source that can be operated in either voltage or current priority mode. In current priority mode the output is controlled by a bi-polar constant current feedback loop, which maintains the output current (source or sink) at its programmed setting. The output current will remain at its programmed setting as long as the load voltage remains within the positive and negative voltage limits. The voltage limits are not programmable and vary somewhat with the output current. When the output current is zero, the voltage limits are typically ±10.75 V.
Figure 1-2 shows the current priority operating characteristics of the dc source. The area in quadrants 1 and 3 shows the characteristics of the unit when it is being operated as a source (sourcing power). The area in quadrants 2 and 4 shows the characteristics of the unit when it is being operated as a load (sinking power).
+ 12V
-
+V limit
Output Voltage
+
+ 10.75V
+VL status set
I setting
132
4
Key
Sinking power
Sourcing power
Programmable
+ 9.5V
+
Output Current
16
- 9.5V
- 0.5125mA
-VL status set
- 10.75V
-
-V limit
+ 0.5125mA
- 12V
Figure 1-2. Output Characteristic (Current Priority)
General Information - 1
The heavy line illustrates the locus of possible operating points as a function of the output load, which may be purely resistive, or possibly include external voltage or current sources. In current priority mode, the constant current loop will regulate the output current as the load changes, until the positive or negative voltage limit is reached. A CC (constant current) status flag indicates when the current loop is controlling the output.
If the output voltage reaches either the positive or negative voltage limit, the unit no longer operates in constant current mode and the output current is no longer held constant. Instead, the output current is limited at either the positive or negative voltage limit line. When the unit is sinking power, the output voltage will continue to increase in the positive or negative direction as more current is forced into the unit. Note that a VL+ (positive voltage limit) or VL (negative voltage limit) status bit will be set to register a voltage limit at about 0.8 V before the positive or negative voltage line is reached.
The maximum current available in current priority mode is about 0.5 mA, which is ideal for testing sensitive devices such as input diodes. In this mode, the output current must be programmed to the desired positive or negative value. However, the positive and negative voltage limits are not programmable, and vary with the actual output current as shown in the figure. The typical positive voltage limit ranges from about 10.75V at no load to about 9.5V at full load. The typical negative voltage limit ranges from about –10.75V to about –9.5V.
NOTE: Overvoltage protection is not functional in current priority mode.

Measurement Characteristics

The N3280A uses a digitizing measurement system with a single timebase for all output channels. The number of measurement samples and the sampling interval of the timebase can be explicitly programmed. These values will apply to measurements taken on all outputs. For example, if simultaneous measurements are made on four output channels and one of the three channels is set to one power line cycle (PLC), then all three channels will be set to one power line cycle per measurement.
Conversely, each output channel of the N3280A has its own measurement buffer. This means that each output can be configured to measure a different parameter (either voltage or current), and a different current range. However, the number of measurement samples and sampling interval for each type of measurement is the same for all channels.
There is one voltage measurement range and three current measurement ranges. The current range must be selected explicitly. If a measured value exceeds the presently selected range, an error message is returned. Voltage measurements and current measurements using the 0.5A or 15mA range can be made to full accuracy using the default measurement sample (5 data points @30.4µs intervals = 152 µs). To achieve full accuracy on the 0.5mA current range, a longer sampling interval of one power line cycle (PLC) is required to filter out line noise. Thus, a full accuracy measurement on the 0.5mA current range will typically take between 18 and 21.3 ms, depending on the line frequency.
Note that faster measurements using lower PLC values (<1) are only appropriate for loads that do not draw currents with a significant noise component. If the load current is noisy, it may be necessary to increase the sampling interval to provide additional filtering.
All voltage and current measurements return the average value of the samples taken. Measurements can be made using either a Rectangular or Hanning window. The default Rectangular window is used on all
17
1 - General Information
measurement ranges to make fast measurements. The Hanning window can be used to reduce errors caused by other periodic noise sources, provided that the sample period is long enough to capture three or more noise waveform cycles. Using a Hanning window will result in slower measurement speed.

Start of a Measurement

The dc source delays the start of a measurement until a previous output voltage or current change has settled. When voltage or current settings are changed in either voltage priority or in current priority mode, an internal timer is started that delays any subsequent measurements. At power-on or after *RST this delay allows the output to settle to better than 0.1% of its final value. In voltage priority mode, the final value is based on a 20 ohm load. In current priority mode, the final value is based on a short-circuit load.
The settling delay can also be explicitly programmed. This may be required, for example, if the load requires more or less delay than the representative load or if the measurement requires less accuracy.
18
2

Installation

Inspection

Damage

When you receive your dc source, inspect it for any obvious damage that may have occurred during shipment. If there is damage, notify the shipping carrier and the nearest Agilent Sales and Support Office immediately. The list of Agilent Sales and Support Offices is at the back of this guide. Warranty information is printed in the front of this guide.

Packaging Material

Until you have checked out the dc source, save the shipping carton and packing materials in case the unit has to be returned. If you return the dc source for service, attach a tag identifying the owner's name and address, the model number, and a brief description of the problem.

Additional Items

Table 2-1. Items Supplied
Item Part Number Description
Power Cord contact the nearest Agilent
Sales and Support Office
4 - Output connectors
Trigger connector
Line Fuse 2110-0638
User's Guide 5964-8248 This manual.
1253-4893 A 6-terminal connector plug for connecting the output,
1252-8670 3-terminal digital plug for connecting the trigger input
2110-0773
A power cord appropriate for your location.
sense, ground, and guard. The connector installs in the back of the unit.
signal. The connector installs in the back of the unit.
3.15 AT (time delay) for 100/120 Vac operation
1.6 AT (time delay) for 220/230 Vac operation

Cleaning

Use a dry cloth or one slightly dampened with water to clean the external case. Do NOT open the unit.
WARNING: To prevent electric shock, unplug the unit before cleaning.
19
2 - Installation

Location

Figure 2-1 gives the dimensions of your dc source. The dc source must be installed in a location that allows enough space at the sides and back of the unit for adequate air circulation (see Bench Operation).
NOTE: This dc source generates magnetic fields that may affect the operation of other
instruments. If your instrument is susceptible to operating magnetic fields, do not locate it in the immediate vicinity of the dc source. Typically, at 5 millimeters from the dc source, the electromagnetic field is less than 5 gauss. Many CRT’s, such as those used in computer displays, are susceptible to magnetic fields much lower than 5 gauss. Check susceptibility before mounting any display near the dc source.

Bench Operation

Do not block the fan exhaust at the rear of the unit.
A fan cools the dc source by drawing air in through the sides and exhausting it out the back. Minimum clearances for bench operation are 1 inch (25 mm) along the sides.

Rack Mounting

The dc source can be mounted in a standard 19-inch rack panel or cabinet. Table 1-1 documents the part numbers for the various rack mounting options that are available for the dc source. Installation instructions are included with each rack mount option.
NOTE: Support rails or an instrument shelf is required when rack mounting units.
20
Figure 2-1. Outline Diagram
Installation - 2

Power Connections

Connect the Power Cord

Connect the power cord to the IEC 320 connector on the rear of the unit. If the wrong power cord was shipped with your unit, contact your nearest Agilent Sales and Support Office to obtain the correct cord (refer to the list at the back of this guide).
Check the line voltage rating label on the back of the unit to make sure that it agrees with your ac mains voltage. Refer to appendix E if the voltage at your site is different from the voltage indicated on the unit.
Figure 2-2 identifies all rear panel connections on the dc source.
1 42 3 5
QQQQ GPIB
connector
RRRR Trigger Connector
SSSS Address Switch
TTTT Output
Connectors (4)
UUUU Line
Figure 2-2. Rear Panel Connectors and Switches
GPIB connector for computer connection.
A 3-terminal trigger input connector. Only the center and left-most terminals are used.
Switch to select GPIB address. Refer to the end of this chapter.
Pin 1 = Active guard Pin 2 = High sense Pin 3 = High output Pin 4 = Low output Pin 5 = Low sense Pin 6 = chassis ground connection
AC line cord is installed here. Also used to set the ac line voltage see Appendix E.
21
2 - Installation

Output Connections

Turn the unit off before connecting any wires.

Outputs 1 - 4

Disconnect the mating plug from the unit by pulling it straight back.
The output connectors (outputs 1-4) have a termination for the Hi and Lo output terminals, the Hi and Lo sense terminals, a guard terminal, and an earth ground terminal (see figure 2-3). For proper operation of the dc source, you must connect the Hi sense and Lo sense terminals to their respective high and low monitoring points. Install the connector plug with its sense terminals connected before applying power to the unit.
CAUTION: Connect the sense leads carefully so that they do not become open-circuited. If the sense
leads are left unconnected or become open during operation, the dc source will revert to a local sense mode using internal sense protect resistors. This will result in an incorrect voltage being applied at the load terminals.
The 6-pin connector is removable and accepts wires sizes from AWG 28 to AWG 16. Insert the wire into the wire terminal, then use a small, flat-bladed screwdriver to tighten the wire terminal. Agilent Technologies does not recommend using wire sizes smaller than AWG 24. After you insert the mating plug into the output connector, tighten the two locking screws to secure the connection.
OUTPUT 1 MATING PLUG
TIGHTEN SCREWS
LOCKING SCREW
INSERT WIRES
TWIST PAIR
SHOWN
Hsen Hi Lo Lsen
TWIST LEADS
_
LOAD
+
Figure 2-3. Remote Sense Connections
Figure 2-4 shows how to connect remote sense and load leads when using a removable test fixture. For best transient response and load regulation, keep the resistance and inductance as low as possible, as illustrated in the figure. The addition of a low-leakage RC network may help improve output transient response when the unit is operating in voltage priority mode.
22
TIGHTEN SCREWS
Installation - 2
OUTPUT 1 MATING PLUG SHOWN
LOCKING SCREW
INSERT WIRES

Current Ratings

Hsen Hi Lo Lsen
TWIST LEADS
TWIST PAIR
ADDITION OF LOW-LEAKAGE
RC NETWORK MAY IMPROVE
TRANSIENT RESPONSE IN
VOLTAGE PRIORITY MODE.
KEEP RESISTNCE AND INDUCTANCE LOW. USE TWISTED PAIR OR SANDWICHED PCB TRACKS.
_
+
FIXTURE
CONNECTIONS
Figure 2-4. Remote Sense Connections with Test Fixture
LOAD
The following table lists the characteristics of AWG (American Wire Gauge) copper wire for some common wire sizes that can be accommodated in the output connectors.
Table 2-2. Ampacity and Resistance of Stranded Copper Conductors
AWG No. Maximum Ampacity (in
free air)
Resistance (at 20 deg. C)
ΩΩΩΩ/m ΩΩΩΩ/ft
24 3.52 0.0843 0.0257 22 5.0 0.0531 0.0162 20 8.33 0.0331 0.0101 18 15.4 0.0210 0.00639 16 19.4 0.0132 0.00402

Voltage Drops and Lead Resistance

To optimize the performance and transient response in your test system, please observe the following guidelines:
Twist the load leads together and keep them short. The shorter the leads, the better the performance.
Twist the sense leads together, but do not bundle the sense leads with the load leads.
For best performance, keep the total cable length to the load to about 5 meters (15 ft) or less.
The load wires must also be of a diameter large enough to avoid excessive voltage drops due to the impedance of the wires. In general, if the wires are heavy enough to carry the maximum short circuit current without overheating, excessive voltage drops will not be a problem.
23
2 - Installation
NOTE: Any voltage drop in the load leads must be subtracted from the full-scale voltage
available at the output terminals.

Coaxial Guard Connections

An active guard connection is available at the output connector. When the guard connection is extended to a test fixture for example, it can be used to eliminate the effects of leakage current that can exist between the Hi and Lo output terminals when testing high-impedance devices. In particular, the Hi output terminal and the Hi sense terminal may benefit from guarding. In this way, any leakage current that is not load current will be collected by the circuit and not be included in the output current measurement.
The guard connection is always enabled and provides a buffered voltage that is at approximately the same potential as the Hi output terminal. The output impedance of the guard is approximately 2.1K ohms.
If you are using tri-axial cables to extend the guard connection to the test fixture, use the center connector for the Hi connection, the inner shield for the guard connection, and the outer shield as the Lo connection (see figure 2-5).
OUTPUT 1 MATING PLUG
TIGHTEN SCREWS
SHOWN
LOCKING SCREW
INSERT WIRES
Hsen Hi Lo Lsen
TRIAXIAL CABLE
TEST FIXTURE
_
+
GUARD SHIELD
Figure 2-5. Guard Connections for Test Fixtures
24
Installation - 2

Maintaining Stability

In voltage priority mode, the constant voltage loop has the following three compensation bandwidths:
30 kHz, 20 kHz; and 10 kHz
In current limit operation, only two compensation bandwidths are available:
30 kHz and 10 kHz
If the output of your unit is being shut down by the oscillation protection circuit because of long load wires or a high Q load impedance, you can reprogram the output compensation bandwidth to try and eliminate the oscillation. As shipped from the factory, the compensation bandwidth is set to 30 kHz.

OVP Considerations

CAUTION: Disabling the OVP protection circuit may cause excessive output voltages, such as can
occur if remote sense leads are shorted, to damage the equipment under test.
The dc source is shipped from the factory with its overvoltage protection circuit enabled. You can disable the OVP circuit using the VOLTage:PROTection:STATe command as explained in chapter 6. The overvoltage circuit automatically turns the output off and opens the output relays if the output voltage exceeds +11.5V (±0.3V) or −11.5V (±0.3V)

External Trigger Connections

This rear panel connector has an external trigger input.
The trigger input pin is normally at a TTL high level. To generate a trigger, you can provide a negative­going TTL signal to the trigger input, or momentarily connect a short (contact closure) from the trigger input pin to the chassis ground pin on the trigger connector. In any case, the device that you use to implement the trigger must be able to sink approximately 1mA.
The external trigger input can trigger both output voltage/current changes and output measurements.

Computer Connections

The dc source can be controlled through a GPIB interface.

GPIB Interface

Follow the GPIB card manufacturer's directions for card installation and software driver setup. Dc sources may be connected to the GPIB interface in series configuration, star configuration, or a combination of the two, provided the following rules are observed:
The total number of devices including the GPIB interface card is no more than 15.
The total length of all cables used is no more than 2 meters times the number of devices connected
together, up to a maximum of 20 meters. (Refer to table 1-2 for a list of available GPIB cables.)
Do not stack more than three connector blocks together on any GPIB connector.
Make sure all connectors are fully seated and the lock screws are firmly finger-tightened.
25
2 - Installation

GPIB Address

Each dc source has its own GPIB bus address, which can be set using the rear panel Address switch. The dc source is shipped with its GPIB address set to 5. Refer to the following table for additional address switch positions.
4 3 2 1 0
1
0
Address = 5
Handle
Table 2-3. Settings for Power Module Configuration Switch
GPIB Switch Setting GPIB Switch Setting
Address 4 3 2 1 0 Address 4 3 2 1 0
000000801000 100001901001 2 0 001 0 10 01 0 10 3 0 001 1 11 01 0 11 4 0 010 0 12 01 1 00
5
00101 13 01I01 6 0 011 0 14 01 1 10 7 0 011 1 15 01 1 11
26

Turn-On Checkout

Front Panel Description

N3280A 10V, 0.5A
Component Test DC Source
3
1 2 3
QQQQ Line
Switch
RRRR Unit
Indicators
SSSS Channel
Indicators
Figure 3-1. Front Panel, Overall View
AC mains power switch.
Unit indicators light to indicate the following operating conditions:
Power The dc source is turned on. Active The dc source is addressed to talk or listen. Error There is a message in the SCPI error queue.
Channel indicators light to indicate the following channel conditions:
On The specified output channel is enabled. Prot The specified output channel has entered protection mode due to:
Overtemperature, Overvoltage, Oscillation protect, or Power clear. Query the status registers of the affected channel to determine which protection feature is tripped.
27
3 – Turn-On Checkout

Checkout Procedure

Successful tests in this chapter provide a high degree of confidence that your unit is operating properly. Complete performance tests are given in Appendix B.
NOTE: To perform the checkout procedure, you will need a computer with a GPIB interface.
You will also need a digital multimeter for making voltage and current measurements.
If you have not already done so, connect your unit to the computer's GPIB interface. Also connect the power cord to the unit and plug it in.
Procedure Explanation
1. Connect the Hi sense terminal to the Hi terminal. Connect the Lo sense terminal to the Lo terminal. Connect the voltage inputs of the voltmeter across the Hi and Lo sense terminals of output 1.
2. Turn the unit on. The unit undergoes a self­test when you first turn it on.
3. Check that the fan is on. You should be able to hear the fan and feel air coming from
4.
Program
5.
Program "
6. Create a variable for a measurement. Program Read the variable value.
7.
Program
8.
Program Read the variable value.
9.
Program
"Output On, (@1)"
Voltage 10, (@1)"
"Measure:Voltage? (@1)"
"Voltage -10, (@1)"
"Measure:Voltage? (@1)"
"Output Off, (@1)"
The external voltmeter is used to verify the output.
During selftest, all indicators light simultaneously and then light individually in a clockwise manner to test the functionality of the display
the back of the unit.
Turn the output on.
Check the voltmeter display to verify the voltage programming.
Reads the voltage of output 1. This should agree with the value displayed on the voltmeter.
Check the voltmeter display to verify the voltage programming.
Reads the voltage of output 1. This should agree with the value displayed on the voltmeter.
Turn the output off.
10. Connect the current measurement inputs of the ammeter across Hi and Lo output terminals of output 1. Observe polarity.
11.
Program
12.
Program
13.
Program
14. Create a variable for a measurement. Program Read the variable value.
15.
Program Disconnect the multimeter.
16. Repeat steps 3 through 15 for outputs 2, 3, and 4.
"Output On, (@1)"
"Function:Mode CURR, (@1)"
"Current 0.0005, (@1)"
"Measure:Current? (@1)"
"Output Off, (@1)"
28
Use the ammeter to short the output of the unit and verify the output current.
Turn the output on.
Program the unit for current priority mode.
Reads the current of output 1. This should agree with the value displayed on the ammeter.
Turn the output off.
Substitute the channel that you are programming after the @ symbol. For example, if you are programming channel 2, program "(@2)" in all commands.
Turn-On Checkout - 3

In Case of Trouble

Dc source failure may occur during power-on selftest or during operation. Either the Error or the Prot indicator on the front panel may be lit to indicate that a failure has occurred. If this occurs, turn the power off and then back on to see if the error persists. If the error persists, the dc source requires service.

Selftest Error Messages

Error numbers and messages are read back with the SYSTem:ERRor? query. SYSTem:ERRor? returns an NR1 and a string error message.
Table 3-1. Power-On Selftest Errors
Error No. Failed Test
Error 0 No error
Error 1
Error 2
Error 3
Error 4
Error 5
Error 10
Output 1 non-volatile RAM CAL section checksum failed
Output 2 non-volatile RAM CAL section checksum failed
Output 3 non-volatile RAM CAL section checksum failed
Output 4 non-volatile RAM CAL section checksum failed
Non-volatile RAM CONFIG section checksum failed
RAM selftest

Runtime Error Messages

Appendix C lists other error messages that may appear at runtime.

Line Fuse

If the dc source appears "dead" with the Power LCD off and the fan is not running, check your ac mains to be certain line voltage is being supplied to the dc source. Also check that the line module on the rear of the unit is set to the correct voltage. If the ac mains is normal, the internal line fuse may be defective.
Refer to Appendix E and follow the procedure described in the appendix for accessing and replacing the line fuse located inside the unit. Unless the line voltage setting is incorrect, do not change the line voltage setting.
NOTE: If the dc source has a defective fuse, replace it only once. If it fails again, the dc source
requires service.
29
Loading...
+ 86 hidden pages