Agilent technologies N2790A User Manual

N2790A Differential Voltage Probe
User’s Guide
For Safety, Regulatory, and publishing information, see the pages at the back of this book.
© Copyright Agilent Technologies 2010
All Rights Reserved.

Contents
Inspecting the Probe 3 Compatibility 4 Probe Parts Supplied 5 Characteristics and Specifications 6 Cleaning the Probe 8 Using the N2790A Probe 9 Service Strategy 10 Characterization Plots 11 Performance Verification Procedures 13

Inspecting the Probe

Inspecting the Probe
Inspect the shipping container for damage.
Keep a damaged shipping container or cushioning material until the contents of the shipment have been checked for completeness and the probe has been checked mechanically and electrically.
Check the accessories.
Any accessories that were supplied with the probe are listed in the “Probe Parts Supplied” section of this manual.
• If the contents are incomplete or damaged, notify your Agilent Technologies Sales Office.
Inspect the instrument.
• If there is mechanical damage or defect, or if the probe does not operate properly or pass calibration tests, notify your Agilent Technologies Sales Office.
• If the shipping container is damaged, or the cushioning materials show signs of stress, notify the carrier as well as your Agilent Technologies Sales Office. Keep the shipping materials for the carrier’s inspection. The Agilent Technologies office will arrange for repair or replacement at Agilent Technologies’ option without waiting for claim settlement.
Inspecting the Probe
N2790A Differential Voltage Probe
The N2790A differential probe allows you to safely measure high-voltage floating circuits with the oscilloscope grounded. It is ideal for many applications such as motor speed controls, power supply designs, and electronic high-power converters. The N2790A offers sufficient dynamic range and bandwidth for your application to make the floating measurements found in high-speed power electronic circuits safely and accurately.
With 100 MHz bandwidth, switchable attenuation of 50:1 and 500:1, and a maximum differential voltage input of 1400V (DC + peak AC), it provides the versatility for a broad range of applications including high-voltage circuits. Each probe comes with sharp browser probe tips for use on small components and in tight places, alligator clips for connecting to larger cables, and pincher clips.
The probe is powered by the autoprobe interface.

Compatibility

The N2790A differential probe is compatible with the following Agilent oscilloscopes:
• InfiniiVision Series 5000, 6000 (except 100 MHz), and 7000 with version 5.2 software or later
• Infiniium 9000 (with version 2.00 software or later), 8000 (with version 5.61 software or later), and 54830 Series (with version 5.61 software or later)
Inspecting the Probe
2
3
4

Probe Parts Supplied

The following diagram and table show the parts supplied with the N2790A Probe.
The probe tips you receive may look slightly different than the ones shown in this picture.
Parts for N2790A Probe Item Description Part Number
1 Differential voltage probe N2790-68701 2 Probe tips - browser tip (x2)
3 Probe tips - pincher clip (x2)
4 Probe tips - alligator plunger clip (x2)
5 Plastic pouch 6 Offset adjustment tool
To order the Probe Tip Accessory Kit (consisting of 2 browser tips, 2 pincher clips, and 2 alligator plunger tips), order part number N2790-68700. Probe tips are only for use with the N2790A differential voltage probe.
*Probe tips are not orderable separately.
(black and red pair)
(black and red pair)
(black and red pair)
A054*

Characteristics and Specifications

Characteristics and Specifications
Characteristics and specifications for the N2790A Differential Voltage Probe are shown below.
Electrical Characteristics
Attenuation ratios 50:1 and 500:1 (Selectable via switch on probe.) Bandwidth (-3 dB)* 100 MHz Gain Accuracy* \4% @ 0-20 oC, \2% @ 20-30 oC, \4% @ 30-
DC CMRR -70 dB @ 500VDC AC CMRR -80 dB @ 50/60 Hz
Input impedance Between inputs: 8 M , 3.5 pF Noise referenced to input < 300 mVrms @ 500:1
Propagation Delay 14.0 ns @ 50:1, 12.8 ns @ 500:1 Switch positions 500:1, 50:1 Internal power Autoprobe interface
*Denotes warranted specification after 20 minute warm-up
o
55
C
-50 dB @ 100 kHz
-50 dB @ 1 MHz
< 50 mVrms @ 50:1
Input Characteristics
Cable length 170 cm (66.9 inches) Length of input leads 40 cm (15.7 inches) Maximum common mode
operating voltage (rms or DC + peak AC)
1
Maximum common mode operating voltage derating frequency
2
Maximum differential mode operating voltage
(DC + peak AC)
1
Refer to the warnings on page 8 for further details on the input voltage safety rating
2
Refer to the derating plot on page 12
600V CAT III 1000V CAT II
2 MHz
1400 V @ 500:1 140 V @ 50:1
Characteristics and Specifications
Output Characteristics
Output cable Safety designed BNC cable Maximum output voltage range \2.8 V into 1 M at 500:1
Environmental Specifications
Temperature Operating: -10 °C to +50 °C
Nonoperating: -51 °C to +71 °C
Altitude Operating: 2,000 m (6,561 ft)
Nonoperating: 15,300 m (50,200 ft)
Humidity Operating: 80% room humidity @ 40 °C
Nonoperating: 90% room humidity @ 65 °C
Pollution Degree Pollution Degree 2
Safety Specifications
UL Listed: File No. E326557 Compliant with: UL 61010-031, CSA 22.2 No. 61010-031-07

Cleaning the Probe

!!!!
WARNING SHOCK HAZARD!
These probes are designed for use with oscilloscopes that have a common terminal at GROUND POTENTIAL (in accordance with OSHA requirements and the National Electric Code). Exposed metallic surfaces of the probe and the oscilloscope MUST BE GROUNDED. Failure to ground the common terminal during certain applications, such as those requiring the oscilloscope to be powered from an external battery, might expose the operator to an electrical shock hazard that could be lethal (depending on voltage and current conditions.)
The probe inputs are safety rated to a maximum of 1000 Vrms CAT II and 600 Vrms CAT III between either input and earth ground. These maximum ratings apply regardless of the attenuation setting of the probe. Do not apply voltages greater than 1000 Vrms CAT II or 600 Vrms CAT III between either input and ground.
The probe tips that are provided are rated for 1000 Vrms CAT II and 600 Vrms CAT III usage. Other probe tips may be used if they have the appropriate ratings and the same style connectors.
Cleaning the Probe
Disconnect the probe from all power sources and clean it with a soft cloth dampened with a mild soap and water solution. Be careful not to get water in the attenuation switch. Make sure the probe is completely dry before reconnecting it to a power source.

Using the N2790A Probe

Using the N2790A Probe

Powering the N2790A

The N2790A probe is powered by the autoprobe interface found on Agilent InfiniiVision and Infiniium Se ries oscilloscopes.

Selecting the Correct Attenuation on the Probe and the Oscilloscope

The probe has two modes of operation, 50:1 and 500:1 attenuation. Select the desired attenuation on the probe (via the switch on the probe) and the correct probe attenuation setting on the oscilloscope. This will ensure that accurate voltage readings can be made directly from the oscilloscope.

Adjusting the Probe Offset Level

The probe offset level is adjusted using the tool provided with the N2790A probe. Adjust the offset level by turning the potentiometer adjustment located on the autoprobe pod.

Measurement Tips

• Use the 50:1 range on the differential voltage probe for smaller signals, such as ripple on a high voltage reference lead.
• Connect the red probe cable to a more positive or more negative voltage level than the black probe cable.
• There is an overrange LED on the probe to indicate when you have exceeded the maximum voltage of the probe.

Service Strat egy

Service Strategy
For repair, calibration, and to ensure the N2790A performs to its warranted specifications, send the probe to an Agilent Service Center for calibration testing procedures. The probe should be tested once a year or as required by other standards. If repair is needed and the N2790A is under warranty, normal warranty services will apply. If the N2790A is not under warranty, repair costs will be applied.

To return the Probe to Agilent Technologies for Service

Call (877) 477-7278 for further details and the location of your nearest Agilent Technologies Service Office.
1 Write the following information on a tag and attach it to the probe.
• Name and address of the owner
• Probe model number
• Description of service required or failure indication
2 Retain all accessories. 3 Return the probe in its case or pack the probe in foam or other
shock-absorbing material and place it in a strong shipping container. You can use the original shipping materials or order materials from an Agilent Technologies Sales Office. If neither are available, place 3 to 4 inches of shock-absorbing material around the instrument and place it in a box that does not allow movement during shipping.
4 Seal the shipment container securely. 5 Mark the shipping container as FRAGILE. In all correspondence, refer to the
instrument by model number and full serial number.
10

Characterization Plots

10
6
10
7
10
8
-9
-6
-3
0
3
6
Vout/Vin
Vin
Vout
BW(-3dB) = 147
10
7
10
8
-80
-70
-60
-50
-40
-30
Characterization Plots
Graph of dB(Vin), dB(Vout), and dB(Vout/Vin) frequency response
Graph of dB (Vout/Vin) frequency response when inputs driven in common (common mode rejection)
11
Characterization Plots
Derating plot - above 2 MHz, the maximum input operating voltage decreases as the frequency increases.
               
12

Performance Verification Procedures

Performance Verification Procedures
The following procedure can be used to test and verify the DC differential gain accuracy and bandwidth of the N2790A high voltage differential probe.
Required Test Equipment
Description Critical Specifications Recommended Model
Digitizing Oscilloscope Bandwidth: >100 MHz
Signal Generator Precision DC voltage source
BNC Adapter BNC (f) to Dual Banana (m)
50 BNC Feed Through Adapter
1 M/50 selectable input Amplitude accuracy: greater
than or equal to 0.25% 1 M/50 selectable load
Sine wave greater than or equal to 25 MHz
Adapter
50 precision feed through Agilent 11048C Termination between
Part Number
Agilent MSO9254A Display probe output
Fluke 9500B or Fluke 5520A for DC source.
Fluke 9500B or Agilent E8257D +1EU option for AC source
Agilent 1251-2277 Interconnection
DC Differential Gain Accuracy
WARNING: These procedures require the application of high voltage to the inputs of the N2790A probe. Because the adapter used in this procedure has exposed metal surfaces, only qualified personnel should perform any testing with voltage levels exceeding 30 V rms. All pertinent safety rules and guidelines for elevated voltage measurements should be followed and adhered to.
WARNING: Generator produces hazardous voltages. To avoid risk of shock, do not touch exposed metal parts after the generator output is enabled.
1 Set the volts/division on channel 1 of the oscilloscope to 20 mV/div. Set the seconds/
div to 200 s and the acquisition mode to average 32.
2 Set the volts/division on channel 2 of the oscilloscope to 50 mV/div. Trigger on channel
2, select 50 impedance.
3 Set the calibrator 9500B’s channel 2 as the trigger channel (50  load). Connect the
calibrator’s channel 2 active head to channel 2 of the oscilloscope.
4 Connect the active head channel 1 from the Fluke 9500B to channel 1 of the
oscilloscope. Set the calibrator channel 1 to 0.1V peak-to-peak amplitude and 1 KHz (square wave, 1 M load), enable the output.
Functions
Signal source for DC gain and bandwidth
between probe and generator
probe and calibrator for bandwidth verification
13
Performance Verification Procedures
5 Select the amplitude measurement on the oscilloscope and record the DC amplitude
(approximately 100 mV) of the square wave. This measurement is only the oscilloscope.
6 Disable the Fluke calibrator output, disconnect the channel 1 active head from channel
1 of the oscilloscope.
7 Connect the output of the N2790A probe to channel 1 of the oscilloscope. 8 Attach the BNC adapter to the Fluke channel 1 active head. 9 Attach the differential probe input leads by clipping the alligator clamp to the BNC
adapter banana post.
10 Set the probe to 50:1. Set the Fluke calibrator to 10 V and 1 KHz standard amplitude
output (channel 1, square wave, 1 M load).
11 Enable the output of the calibrator. 12 Record the DC amplitude of the square wave and divide 50 into just the amplitude of
the oscilloscope. Verify that the probe gain accuracy is +/-2% + sc ope gain accuracy.
13 Set the calibrator to 100 V output and set the attenuation button on the probe to 500:1. 14 Record the DC amplitude of the square wave and divide 500 into just the amplitude of
the oscilloscope. Verify that the probe gain accuracy is +/-2% + scope gain accuracy.
15 Disable the calibrator output and leave the setup connected for the next procedure.

Bandwidth

1 Connect the 50 BNC feed through adapter to the calibrator’s active head channel 1. 2 Attach the BNC (f)-to-banana post adapter to the BNC feed through adapter. 3 Attach the differential probe input leads by clipping the alligator clamp to the BNC
adapter banana post.
4 Set the probe to 50:1 attenuation. 5 Set the volts/division on channel 1 of the oscilloscope to 10 mV/div and seconds/
division to 20 ns/div. Set the trace to the center of the oscilloscope. Input impedance shoudl be set to 1 . Disable average mode.
6 Set the calibrator to sine wave, 5 V peak-to-peak amplitude. Set the frequency to 100
MHz, 50 load.
7 Write down the peak-to-peak amplitude measured by the oscilloscope. It should be
greater than or equal to 71 mV + scope error.
8 Disable the generator output and disconnect the probe input.
14

Performance Verification Test Record

Performance Verification Test Record
Agilent Technologies
Recommended Test Interval: 1 Year Recommended Date of Next Certification:_________ Certification Temperature:_____________________
N2790A 100 MHz Differential Probe Serial No.:_______________________ Certification Date:_________________ Tested By:_______________________
_______________________________ _______________________________
Test Probe Settings Limits Results
Gain 50:1 196 mV to 204 mV Gain 500:1 196 mV to 204 mV Bandwidth 50:1 Greater than or equal to 71 mV
15
Performance Verification Test Record
16
Safety
!
Notices
This apparatus has been designed and tested in accordance with IEC Publi cation 1010, Safety Require­ments for Measuring Apparatus, and has been supplied in a safe condi tion. This is a Safety Class I instrument (provided with terminal for protective earthing). Before applying power, verify that the cor rect safety precautions are taken (see the following warnings). In addition, note the external markings on the instrument that are described under "Safety Symbols."
Warnings
• Before turning on the instrument, you must con nect the protective earth ter­minal of the instrument to the protective conductor of the (mains) power cord. The mains plug shall only be inserted in a socket outlet provided with a protective earth contact. You must not negate the protective action by using an extension cord (power cable) without a pro tective conductor (ground­ing). Grounding one conductor of a two-conduc­tor outlet is not sufficient protection.
• Only fuses with the required rated current, volt age, and specified type (nor­mal blow, time delay, etc.) should be used. Do not use repaired fuses or short­circuited fuseholders. To do so could cause a shock or fire hazard.
• If you energize this instru­ment by an auto trans­former (for voltage reduction or mains isola tion), the common terminal must be connected to the earth terminal of the power source.
• Whenever it is likely that the ground protection is impaired, you must make the instrument inoperative and secure it against any unintended operation.
• Service instructions are for trained service personnel. To avoid dangerous electric
-
-
-
-
shock, do not perform any service unless qualified to do so. Do not attempt internal service or adjustment unless another person, capable of rendering first aid and resuscitation, is present.
• Do not install substitute
-
parts or perform any unau thorized modification to the instrument.
• Capacitors inside the instrument may retain a charge even if the instru ment is disconnected from its source of supply.
• Do not operate the instru­ment in the presence of flammable gasses or fumes. Operation of any electrical instrument in such an envi ronment constitutes a defi­nite safety hazard.
• Do not use the instrument in a manner not specified by the manufacturer.
To clean the instrument If the instrument requires cleaning: (1) Remove power
from the instrument. (2) Clean the external surfaces of the instrument with a soft cloth dampened with a mix ture of mild detergent and water. (3) Make sure that the instrument is com pletely dry before recon-
­necting it to a power source.
-
Safety Symbols
Instruction manual sym­bol: the product is marked with this symbol when it is necessary for you to refer to
-
the instruction manual in order to protect against damage to the product..
-
Hazardous voltage symbol.
Earth terminal symbol:
­Used to indicate a circuit common connected to grounded chassis.
-
-
Agilent Technologies P.O. Box 2197 1900 Garden of the Gods Road Colorado Springs, CO 80901
Notices
© Agilent Technologies, Inc. 2010 No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and writ ten consent from Agilent Technologies, Inc. as gov­erned by United States and international copyright laws.
Manual Part Number N2790-97002, December 2010
Print History N2790-97002, Dec 2010 N2790-97001, August 2010 N2790-97000, January 2009
Agilent Technologies, Inc. 1900 Garden of the Gods Road
Colorado Springs, CO 80907 USA
Restricted Rights Legend If software is for use in the
performance of a U.S. Gov­ernment prime contract or subcontract, Software is delivered and licensed as “Commercial computer software” as defined in DFAR 252.227-7014 (June
1995), or as a “commercial item” as defined in FAR
2.101(a) or as “Restricted computer software” as defined in FAR 52.227-19 (June 1987) or any equiva lent agency regulation or contract clause. Use, dupli cation or disclosure of Soft­ware is subject to Agilent Technologies’ standard commercial license terms, and non-DOD Departments and Agencies of the U.S. Government will receive no greater than Restricted Rights as defined in FAR
52.227-19(c)(1-2) (June
1987). U.S. Government users will receive no greater than Limited Rights as defined in FAR 52.227-14 (June 1987) or DFAR
252.227-7015 (b)(2) (Novem ber 1995), as applicable in any technical data.
-
ted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, includ ing but not limited to the implied warranties of merchantability and fit ness for a particular purpose. Agilent shall not be liable for errors or for incidental or con sequential damages in connection with the fur nishing, use, or perfor­mance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with war ranty terms covering the material in this docu ment that conflict with these terms, the war­ranty terms in the sepa­rate agreement shall control.
Technology Licenses The hardware and/or soft-
ware described in this docu­ment are furnished under a license and may be used or copied only in accordance with the terms of such license.
-
-
-
WARNING A WARNING notice
denotes a hazard. It calls attention to an operating procedure,
-
practice, or the like that, if not correctly performed or adhered
-
to, could result in personal injury or death. Do not proceed
-
beyond a WARNING notice until the
-
indicated conditions are fully understood and met.
CAUTION
-
A CAUTION notice
-
denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
Document Warranty
The material contained in this document is pro vided “as is,” and is sub­ject to being changed, without notice, in future editions. Further, to the maximum extent permit
-
-

Agilent Technologies Printed in the Japan
Manual Part Number N2790-97002

*N2790-97002*

Loading...