No part of this manual may be reproduced in
any form or by any means (including
electronic storage and retrieval or translation
into a foreign language) without prior
agreement and written consent from Agilent
Technologies, Inc. as governed by United
States and international copyright laws.
Manual Part Number
G2790-90012
Edition
First edition, January 2019
Printed in USA
Agilent Technologies, Inc.
2850 Centerville Road
Wilmington, DE 19808-1610 USA
The material contained in this document is
provided “as is,” and is subject to being
changed, without notice, in future editions.
Further, to the maximum extent permitted
by applicable law, Agilent disclaims all
warranties, either express or implied, with
regard to this manual and any information
contained herein, including but not limited
to the implied warranties of merchantability
and fitness for a particular purpose. Agilent
shall not be liable for errors or for incidental
or consequential damages in connection
with the furnishing, use, or performance of
this document or of any information
contained herein. Should Agilent and the
user have a separate written agreement
with warranty terms covering the material in
this document that conflict with these
terms, the warranty terms in the separate
agreement shall control.
Technology Licenses
The hardware and/or software described in
this document are furnished under a license
and may be used or copied only in accordance
with the terms of such license.
Restricted Rights Legend
U.S. Government Restricted Rights. Software
and technical data rights granted to the
federal government include only those rights
customarily provided to end user customers.
Agilent provides this customary commercial
license in Software and technical data
pursuant to FAR 12.211 (Technical Data) and
12.212 (Computer Software) and, for the
Department of Defense, DFARS 252.227-7015
(Technical Data -Commercial Items) and
DFARS 227.7202-3 (Rights in Commercial
Computer Software or Computer Software
Documentation).
Safety Notices
A CAUTION notice denotes a
hazard. It calls attention to an
operating procedure, practice, or
the like that, if not correctly
performed or adhered to, could
result in damage to the product or
loss of important data. Do not
proceed beyond a CAUTION notice
until the indicated conditions are
fully understood and met.
A WARNING notice denotes a
hazard. It calls attention to an
operating procedure, practice, or
the like that, if not correctly
performed or adhered to, could
result in personal injury or death.
Do not proceed beyond a WARNING
notice until the indicated conditions
are fully understood and met.
Site Preparation Guide
Contents
1Agilent 8860 GC Site Preparation
Site Preparation Checklist 6
Bench Preparation 7
Maximum Length of Cables and Hoses 10
2GC Installation Kits
Installation Kits 14
3Dimensions and Weights
Dimensions and Weight 18
Foreline Pump Requirements for Systems Including an MSD 20
ALS Dimensions and Weight 21
4Environmental Conditions
Environmental Conditions 24
Heat Dissipation 25
ALS Environmental Conditions 26
5Exhaust Venting
Exhaust Venting 28
Venting hot air 28
Venting other gases 29
Exhaust vent fittings 30
6GC System Power Requirements
Power Requirements 32
USA fast heating oven, 240 V 34
Canadian installation 34
Common instrument power cord plugs 34
ALS Power Requirements 38
7Gas Selection and Plumbing
Gas and Reagent Selection 40
Hydrogen Carrier Gas 42
Gas and Reagent Purity 42
Site Preparation Guide3
Gas Supplies 43
GC/MS Gas Requirements 45
Performance verification 48
Gas Plumbing 49
Supply tubing for most carrier and detector gases 50
Supply tubing for hydrogen gas 50
Two-stage pressure regulators 51
Pressure regulator-gas supply tubing connections 51
Filters and traps 52
ALAN Requirements
Site LAN Network 56
4Site Preparation Guide
1Agilent 8860 GC Site Preparation
Site Preparation Checklist 6
Bench Preparation 7
Maximum Length of Cables and Hoses 10
This guide outlines the site requirements for GC, GC/MS, and automatic liquid sampler (ALS)
installation. Site requirements include the necessary space, electrical supplies, gas supplies,
operating supplies and consumables required to successfully install the GC and related
instruments and systems.
The site must meet the requirements specified in this guide before beginning installation.
Refer to the Agilent Web site at www.agilent.com for the most up-to-date listing of GC, GC/MS,
and ALS supplies and consumables.
Site Preparation Guide5
1Agilent 8860 GC Site Preparation
Site Preparation Checklist
Site Preparation Checklist
For typical system requirements for system installation, see the diagrams on page 8 through
page 9.
Use the following checklist to ensure that the site is properly prepared for GC system
installation.
1Ensure that the appropriate installation hardware has been acquired.
See “Installation Kits” on page 14.
2Ensure that the location in which the GC system is being installed meets the
requirements for environmental conditions. See “Environmental Conditions” on
page 24. Also see “Heat Dissipation” on page 25.
3Prepare bench space for the GC system. Ensure that the bench has the size and
weight capacity to accommodate the GC and associated components. See “Bench
Preparation” on page 7. Also see “Dimensions and Weight” on page 18.
4Ensure that system components are oriented so that they can be connected
properly. See “Maximum Length of Cables and Hoses” on page 10.
5If the system being installed includes an MSD, ensure that the bench allows for
proper installation and connection of the foreline pump. See “Foreline Pump
Requirements for Systems Including an MSD” on page 20.
6Ensure that appropriate venting is provided for the GC system. See “Exhaust
Venting” on page 28.
7Ensure that a dedicated power circuit is available for each device in the system.
See “Power Requirements” on page 32.
8Ensure that appropriate gas and reagent supplies are provided for the GC system.
See “Gas and Reagent Selection” on page 40.
9Ensure that appropriate gas plumbing is provided for the GC system. See “Gas
Plumbing” on page 49.
10 If the GC system being installed includes a data system, ensure that the PC meets
the requirements necessary to properly support the GC system. For more
information, see the site prep guide for your data system.
11 If the GC being installed is to be connected to a site LAN, ensure that the appropriate
cabling is available. See “Site LAN Network” on page 56.
6Site Preparation Guide
1Agilent 8860 GC Site Preparation
Bench Preparation
Bench Preparation
When planning a bench layout:
•Consider component dimensions, weights, and space requirements. See “Dimensions and
Weight” on page 18.
•Consider the lengths of cables and hoses for connection of components. See “Maximum
Length of Cables and Hoses” on page 10.
•For systems that include an MS, consider foreline pump requirements. See “Foreline
Pump Requirements for Systems Including an MSD” on page 20.
•Allow space for operational access.
•Note that the 7200 Q-TOF requires 48 cm (1.6 ft) of space in front to allow for the RIS
probe extraction tool handle when installed.
•Note that some repairs to the GC/MS, or to the GC itself, will require access to the back of
the instrument(s).
Examples are provided here for systems including a GC with an ALS, computer and printer.
Most examples also include an MS.
See several example layouts below.
Site Preparation Guide7
1Agilent 8860 GC Site Preparation
Typical GC System - 8860 GC with computer and printer.
Total weight: ~84 kg (186 lb)
Maximum power consumption: ~3,250 VA (11,090 btu/hr)
78 cm (31-in.)
without heat
deflector
65 cm (26-in.)
with heat
deflector
Leave 30 cm (12-in.)
open space for
operational access
Use isolated circuits with
dedicated grounds for GC
and MSD.
Application
Gas
*
* Use 1/8-in. Swagelok gas connections
Purity
Supply
Pressure
(psi)
†
† 1 psi = 6.89 kPa
CarrierHelium
Hydrogen
Nitrogen
99.9995
99.9995
99.9995
50 - 80
50 - 80
50 - 80
Detectors
TCDHelium99.999550 - 80
FID, NPD,
FPD, TCD
Hydrogen99.999550 - 80
ECD, FID,
FPD, NPD,
TCD
Nitrogen99.999550 - 80
FID, NPD,
FPD
AirZero
grade
50 - 80
80 cm
(32-in.)
(without
ALS)
100 cm (40-in.)
(with ALS)
~2.35 m (7.7 ft.)
~86 cm (34-in.)~56 cm (22-in.)~56 cm (22-in.)~44 cm (17.3-in.)
Bench Preparation
8Site Preparation Guide
1Agilent 8860 GC Site Preparation
Typical GC/MS System - 8860 GC, 5977 MSD, with computer and
printer.
Total weight: ~123 kg (271 lb)
Maximum power consumption: ~4,350 VA (14,843 btu/hr)
78 cm (31-in.)
without heat
deflector
65 cm (26-in.)
with heat
deflector
If bench backs up to
wall, drill 1.5-in. hole
for foreline pump
hose.
Leave 30 cm (12-in.) open
space for operational
access
Use isolated circuits with
dedicated grounds for GC
and MSD.
Place foreline pump on floor or
on vibration reducing bench.
Application
Gas
*
* Use 1/8-in. Swagelok gas connections
Purity
Supply
Pressure
(psi)
†
† 1 psi = 6.89 kPa
CarrierHelium
Hydrogen
Nitrogen
99.9995
99.9995
99.9995
50 - 80
50 - 80
50 - 80
Detectors
TCDHelium99.999550 - 80
FID, NPD,
FPD, TCD
Hydrogen99.999550 - 80
ECD, FID,
FPD, NPD,
TCD
Nitrogen99.999550 - 80
FID, NPD,
FPD
AirZero
grade
50 - 80
80 cm
(32-in.)
(without
ALS)
100 cm (40-in.)
(with ALS)
~2.35 m (7.7 ft.)
~86 cm (34-in.)~56 cm (22-in.)~56 cm (22-in.)~44 cm (17.3-in.)
Bench Preparation
Site Preparation Guide9
1Agilent 8860 GC Site Preparation
CAUTION
Maximum Length of Cables and Hoses
Maximum Length of Cables and Hoses
The distance between system modules may be limited by some of the cabling and the vent or
vacuum hoses.
Table 1Cable and hose lengths
ItemLength
Remote cable2 m (6.6 ft)
LAN cable10 m (32.8 ft)
Power cords2 m (6.6 ft)
Vacuum hose1.3 m (4.24 ft)
Foreline pump power cord2 m (6.6 ft)
7697A Headspace Sampler transfer line99 cm (39-in.)
G1888 Headspace Sampler transfer line80 cm (31.5-in.)
The supporting surface for the 7200/7250 Q-TOF GC/MS system should be kept relatively
vibration free. Do not put the rough pump on your laboratory bench with the 7200/7250
Q-TOF GC/MS due to the vibration that the pump creates. Vibration can lead to a loss of
mass accuracy and resolution.
10Site Preparation Guide
2GC Installation Kits
Installation Kits 14
This section provides details for available installation hardware.
Refer to the Agilent Web site at www.agilent.com for the most up-to-date listing of GC, GC/MS,
and ALS supplies and consumables.
Site Preparation Guide13
2GC Installation Kits
NOTE
Installation Kits
Installation Kits
Installation kits are not supplied with the GC. If you did not order the GC with the factory
plumbing option (305), Agilent highly recommends the following kits in Table 2.
•Agilent recommends purchasing the installation kit(s) that provides parts useful for GC
installation. (Tab le 2 lists the appropriate installation kits.)
In addition to these installation kits, fittings and reducers are required to convert gas cylinder
regulator fittings (for example, 1/4-inch male NPT) to the 1/8-inch female Swagelok fittings
needed to connect gases to the instrument. These fittings are not included with the GC or with
the installation kits. See “Gas Plumbing” on page 49 for details.
Table 2Installation kits
KitPart numberContents
Recom mended for GCs wit h FID, NPD, F PD:
GC Supply Gas Installation
Kit with Gas Purifiers
See Figure 1.
Recommended for GCs with TCD/ECD, MS, and MSD:
19199NIncludes Gas Clean Filter system kit
CP736530 (with 1 oxygen, 1 moisture,
and 2 charcoal filters), 1/8-inch brass
nuts and ferrules, copper tubing,
1/8-inch brass tees, tubing cutter,
1/8-inch brass caps, universal external
split vent trap with replacement
cartridges, and 1/8-inch ball valve.
GC Supply Gas Installation
Kit
See Figure 2.
Gas Clean carrier gas filter
kit, 1/8-inch
See Figure 3.
Figure 1.GC Supply Gas Installation Kit with Gas Purifiers 19199N
Figure 2.GC Supply Gas Installation Kit 19199M
Figure 3.Gas Clean carrier gas filter kit, 1/8-inchCP17974
Site Preparation Guide15
2GC Installation Kits
Installation Kits
16Site Preparation Guide
3Dimensions and Weights
Dimensions and Weight 18
Foreline Pump Requirements for Systems Including an MSD 20
ALS Dimensions and Weight 21
This section lists the dimensions of the GC, GC/MS, and automatic liquid samplers (ALS).
Site Preparation Guide17
3Dimensions and Weights
Dimensions and Weight
Dimensions and Weight
1Ensure that you can accommodate the shipping pallets when you take delivery.
See Table 3.
2Prepare laboratory bench space before the system arrives. Make sure the prepared area is
clean, clear, and level. Pay special attention to the total height requirements. Avoid bench
space with overhanging shelves. See Table 4 .
Table 3Pal let d i mens ions and weights
ProductHeightWidthDepthWeight
GC
8860 Series GC shipping pallet
With third side detector (side mount)
MS
7200/7250 Q-TOF MS96 cm (38-inch)130 cm (51-inch)91 cm (36-inch)175 kg (385 lb)
7200 Flight Tube66 cm (26-inch)66 cm (26-inch)147 cm (58-inch)36.4 kg (80 lb)
7250 Flight Tube66 cm (26-inch)206 cm (81-inch)81 cm (32-inch)87 kg (191 lb)
76 cm (30-inch)
76 cm (30-inch)
86 cm (34-inch)
87 cm (34-inch)
103 cm (40.5-inch)
108 cm (42.5-inch)
N/A
Table 4Instrument dimensions, weights, and required clearances
ProductHeightWidthDepthWeight
GC
8860 Series GC
With third side detector (side mount)
• GC operational /oven accessRequires
• GC rear ventilation / maintenance
clearance
MSD
5975 Series MSD
• Diffusion pump 41 cm (16-inch)30 cm (12-inch)54 cm (22-inch)39 kg (85 lb)
• Standard turbo pump41 cm (16-inch)30 cm (12-inch)54 cm (22-inch)39 kg (85 lb)
• Foreline pump
Standard
Oil-free
• GC/MS operational and maintenance
access
5977 Series MSD
• Diffusion pump 41 cm (16-inch)30 cm (12-inch)54 cm (22-inch)39 kg (85 lb)
50 cm (19.2-inch)
50 cm (19.2-inch)
21 cm (8-inch)
19 cm (7.5-inch)
59 cm (23-inch)
68 cm (27-inch)
54 cm (21-inch)
54 cm (21-inch)
50 kg (112 lb)
57 kg (125.4 lb)
≥ 30 cm (12-inch) open space above GC.
Requires
Requires
wall to dissipate hot air and allow for routine maintenance.
13 cm (5-inch)
32 cm (13-inch)
Requires 30 cm (1 ft) to the left of the unit.
≥ 27 cm (10.7-inch) open space in front of GC.
≥ 25 cm (10-inch) clearance between back of instrument and
31 cm (12-inch)
28 cm (11-inch)
11 kg (23.1 lb)
16 kg (35.2 lb)
• Performance turbo pump41 cm (16-inch)30 cm (12-inch)54 cm (22-inch)41 kg (90 lb)
18Site Preparation Guide
3Dimensions and Weights
Dimensions and Weight
Table 4Instrument dimensions, weights, and required clearances (continued)
ProductHeightWidthDepthWeight
• Foreline pump
Standard
Oil-free (MVP-070)
Oil-free (IDP3)
21 cm (8-inch)
19 cm (7.5-inch)
18 cm (7-inch)
13 cm (5-inch)
32 cm (13-inch)
35 cm (14-incn)
31 cm (12-inch)
28 cm (11-inch)
14 cm (6-inch)
11 kg (23.1 lb)
16 kg (35.2 lb)
10 kg (21 lb)
• GC/MS operational and maintenance
access
MS
7000 and 7010 Triple Quad MS
• EI Mainframe47 cm (18.5-inch)35 cm (14-inch)86 cm (34-inch)59 kg (130 lb)
• EI/CI Mainframe47 cm (18.5-inch)35 cm (14-inch)86 cm (34-inch)63.5 kg (140 lb)
• Foreline pump 28 cm (11-inch)18 cm (7-inch)35 cm (14-inch)21.5 kg (47.3 lb)
• GC/MS operational and maintenance
access
7200 Q-TOF MS
• Mainframe133 cm (52.5-inch)90 cm (34.5-inch)100 cm (39.5-inch)138 kg (305 lb)
• Foreline pump28 cm (11-inch)18 cm (7-inch)35 cm (14-inch)21.5 kg (47.3 lb)
7250 Q-TOF MS
• Mainframe190 cm (74.8-inch)90 cm (34.5-inch)100 cm (39.5-inch)138 kg (350 lb)
• Foreline pump DS20228 cm (11-inch)18 cm (7-inch)35 cm (14-inch)21.5 kg (47.3 lb)
• Foreline pump IDP-1536.4 cm (14.3-inch)33.3 cm (13.1-inch)48.5 cm (19.1-inch)45.5 kg (100 lb)
• GC/Q-TOF operational and maintenance
access
Headspace sampler (HS)
7697A Headspace sampler
111 vial model
12 vial model
• GC with 7697A Headspace samplerRequires 69 cm (27-inch) to the right of the GC (G4557A), or 64 cm
G1888 Headspace sampler56 cm (22-inch)46 cm (18.1-inch)64 cm (25-inch)46.3 kg (102 lb)
ALS
80 cm (32-inch)
61 cm (24-inch)
Requires 30 cm (1 ft) to the left of the unit.
Requires 30 cm (1 ft) to the left of the unit.
Requires 40 cm (16-inch) clearance on both sides of the unit.
Requires 30 cm (12-inch) clearance behind the unit.
69 cm (27-inch)
64 cm (25-inch)
(25-inch) to the right of the GC (G4556A).
70 cm (27.5-inch)
69 cm (27-inch)
46 kg (101 lb)
38.2 kg (84 lb)
• GC with 7693A ALS injectorRequires 50 cm (19.5-inch) above the GC3.9 kg (8.6 lb) each
• GC with 7693A ALS trayRequires 43 cm (16.8-inch) left of the GC
Requires 4.2 cm (1.7-inch) in front of GC
• GC with 7650A ALS injectorRequires 50 cm (19.5-inch) above the GC3.9 kg (8.6 lb) each
• GC with CTC PAL AutosamplerRequires 76.6 cm (30.2-inch) above the GC and
65 to 98 cm (25.6 to 38.6-inch) to the right of the
GC, depending on configuration
6.8 kg (15 lb) each
Site Preparation Guide19
3Dimensions and Weights
CAUTION
Foreline Pump Requirements for Systems Including an MSD
Foreline Pump Requirements for Systems
Including an MSD
1If using a 7200 or 7250 Q-TOF MS, the length of the quadrupole vacuum hose is 130 cm
(4 ft 3-inch) from the high vacuum pump to the foreline pump, and the length of the foreline
pump power cord is 2 m (6 ft 6-inch).
2If your bench abuts a wall, drill 4 cm (1.5-inch) diameter holes through the rear of the bench
for the vacuum hose and power cord.
Make sure the 7200/7250 Q-TOF GC/MS foreline pump is located where it is not likely to be
touched by operators.
20Site Preparation Guide
3Dimensions and Weights
ALS Dimensions and Weight
ALS Dimensions and Weight
Select the laboratory bench space before the system arrives. Pay special attention to the total
height requirements. Avoid bench space with overhanging shelves. See Table 5.
The instrument needs space for proper convection of heat and ventilation. Allow at least
20 cm clearance between the back of the instrument and wall to dissipate hot air.
Table 5Required height, width, depth, and weight
ProductHeight (cm)Width (cm)Depth (cm)Weight (kg)
G4513A Injector5116.516.53.9
G4514A Tray
7650A Injector5122244.5
Additional space requirements
• GC with 7693A ALS injector50 cm (19.5-inch) above the GC
• GC with 7693A ALS tray45 cm (17.5-inch) left of the GC
• GC with 7650 ALS injector50 cm (19.5-inch) above the GC
*
2944436.8
9 cm (3.6-inch) in front of the GC
3 cm (1.2-inch) to the left of the GC
* The G4520A Tray with a bar code reader is available with a G4514A Tray and G4515A bar code reader.
Site Preparation Guide21
3Dimensions and Weights
ALS Dimensions and Weight
22Site Preparation Guide
4Environmental Conditions
Environmental Conditions 24
Heat Dissipation 25
ALS Environmental Conditions 26
This section outlines the environmental requirements for use or storage of the GC, GC/MS, and
automatic liquid sampler (ALS). Heat dissipation information is also provided.
Site Preparation Guide23
4Environmental Conditions
NOTE
NOTE
Environmental Conditions
Environmental Conditions
Ensure that the instrument will be operated or stored within the recommended environmental
ranges. This optimizes instrument performance and lifetime. The specified conditions assume
a non-condensing, non-corrosive atmosphere. See Tab le 6 .
Also, see “Heat Dissipation” on page 25.
Performance can be affected by sources of heat and cold from heating, air conditioning
systems, or drafts.
Table 6Environmental conditions for operation and storage
ProductCondition Temperature rangeHumidity rangeMaximum altitude
8860 GCStandard oven ramp5 to 45 °C5 to 90%3,100 m
Storage–20 to 65 °C0 to 90%
MSD
5975 Series MSDOperation15 to 35 °C
(59 to 95 °F)
Storage–20 to 70 °C
(–4 to 158 °F)
5977 Series MSDOperation15 to 35 °C
(59 to 95 °F)
Storage–20 to 70 °C
(–4 to 158 °F)
MS
7010 or 7000 Triple Quad MSOperation15 to 35 °C
(59 to 95 °F)
Storage–20 to 70 °C
(–4 to 158 °F)
7200 or 7250 Q-TOF MSOperation15 to 35 °C
(59 to 95 °F)
Storage–20 to 70 °C
(–4 to 158 °F)
*
*
†
†
20 to 80%4,615 m
0 to 95%
20 to 80%4,615 m
0 to 95%
40 to 80%5,000 m
0 to 95%
20 to 80%2,500 m
0 to 95%
‡
* Operation requires constant temperature (variations < 2 oC/hour)
† Operation requires constant temperature (variations < 2 oC/hour)
‡ An altitude of 3,700 meters (12,000 feet) is supported if the ambient temperature is less than 30 °C
Air pressure 75 kPa to 106 kPa. No hoar-frost, dew, water, rain, or percolating
24Site Preparation Guide
4Environmental Conditions
Heat Dissipation
Heat Dissipation
•Use Table 7 to estimate the additional heat dissipated from the equipment. Maximums
represent the heat given off when heated zones are set for maximum temperatures.
Table 7Heat dissipation
Instrument
8860 GC7681 BTU/hour maximum
5975 Series MSD3000 BTU/hour (3165 kJ/h)
5977 Series MSD3000 BTU/hour (3165 kJ/h)
7010 or 7000 Triple Quad MS3700 BTU/hour (3904 kJ/h)
7200 or 7250 Q-TOF MS6200 BTU/hour (6541 kJ/h)
If using the optional G4522A Cooling Accessory, you will need to supply:
(8103 kJ/h)
5120 BTU/hour maximum (100 V
power option) (5402 kJ/h)
•A water chiller.
•Tubing and 1/8-inch Swagelok fittings to connect the chilled water and return water to the
chiller.
A container or drain to dispose of condensate from the tray.
Site Preparation Guide25
4Environmental Conditions
ALS Environmental Conditions
ALS Environmental Conditions
Operating the instrument within the recommended ranges optimizes instrument performance
and lifetime. The sampler system operates in the same environment as its parent GC.
See “Environmental Conditions” on page 24.
The conditions assume a non-condensing, noncorrosive atmosphere.
Table 8Environmental conditions for operation and storage
Product
G4513A Injector
G4514A Tray
G4515A Bar Code Reader*
7650 InjectorOperation0 to 40 °C5–95%4,300 m
G4517A ControllerOperation–5 to 45 °CMaximum relative humidity of 80% for
* The G4520A Tray with a bar code reader is available with a G4514A Tray and G4515A bar code reader.
*
Conditions
Operation0 to 40 °C5–95%4,300 m
Storage
Operating temp
range
Operating humidity range
temperatures up to 31 °C decreasing
linearly to 50% relative humidity at 40 °C
Maximum altitude
2,000 m
26Site Preparation Guide
5Exhaust Venting
Exhaust Venting 28
Venting hot air 28
Venting other gases 29
Exhaust vent fittings 30
This section outlines the exhaust venting requirements for GC, GC/MS, and automatic liquid
sampler (ALS) installation.
Site Preparation Guide27
5Exhaust Venting
WARNING
Exhaust Venting
Exhaust Venting
During normal operation, the GC exhausts hot oven air. Depending on the installed inlet and
detector types, the GC can also exhaust (or vent) uncombusted carrier gas and sample. Proper
venting of these exhausts is required for operation and safety.
Venting hot air
Do not place temperature-sensitive items (for example, gas cylinders, chemicals, regulators,
and plastic tubing) in the path of the heated exhaust. These items will be damaged and
plastic tubing will melt. Be careful when working behind the instrument during cool-down
cycles to avoid burns from the hot exhaust.
1Hot air (up to 450 °C) from the oven exits through a vent in the rear of the instrument. Allow
at least 25 cm (10-inch) clearance behind the instrument, or 30 cm (12-inch) behind a
Q-TOF GC/MS to dissipate this hot air. See Figure 4.
Figure 4.Exhaust outlet.
2For most applications, an optional oven exhaust deflector is available. The exhaust
deflector allows for less bench depth than a GC without the exhaust deflector installed.
• The GC exhaust deflector is included if GC option 306 is ordered. The exhaust deflector
requires 14 cm (5.5 inches) behind the instrument. For GCs with the exhaust deflector
option installed, the exhaust rate is about 65 ft
deflector, the exhaust rate is about 99 ft
diameter is 10 cm (4 in). See Figure 5.
• For part numbers for the exhaust deflectors, see Tab l e 9.
28Site Preparation Guide
3
3
/min (1.84 m3/min). Without the
/min (2.8 m3/min). The deflector outlet
5Exhaust Venting
NOTE
Venting other gases
Table 9Exhaust deflector part numbers
InstrumentPart number
GCG1530-80650
7200/7250 Q-TOF GC/MS, GC Q-TOFG3850-80650
Figure 5.Exhaust deflector G1530-80650
Venting other gases
During normal operation of the GC with many detector and inlet types, some of the carrier gas
and sample vents outside the instrument through the split vent, septum purge vent, and
detector exhaust. If any sample components are toxic or noxious, or if hydrogen is used as the
carrier gas or detector fuel gas, these exhausts must be vented to a fume hood.
Exhaust venting must comply with all local environmental and safety codes. Contact your
Environmental Health & Safety (EHS) specialist.
1Place the GC in the hood or attach a large diameter venting tube to the relevant outlet for
proper ventilation. See “Exhaust vent fittings” on page 30.
2To further prevent contamination from noxious gases, attach a chemical trap to the
vent(s).
3If using an ECD, always connect the ECD exhaust vent to a fume hood or vent it to the
outside. See the latest revision of 10 CFR Part 20 (including Appendix B), or the applicable
state regulation. For countries other than the United States, consult with the appropriate
agency for equivalent requirements. Agilent recommends a vent line internal diameter of
6 mm (1/4-inch) or greater. With a line of this diameter, the length is not critical.
Site Preparation Guide29
5Exhaust Venting
Exhaust vent fittings
4Vent the GC/MS system externally to the building via an ambient-pressure vent system,
within 460 cm (15 ft) of both the GC split vent and GC/MS foreline pump, or vent to a fume
hood.
Exhaust vent fittings
The various inlet and detector vents terminate in the following fittings:
•TCD, ECD: The detector exhaust terminates in a 1/8-inch od tube.
•SSL: The split vent terminates in a 1/8-inch Swagelok female fitting.
•All inlets: The septum purge vent terminates in 1/8-inch od tubing.
30Site Preparation Guide
6GC System Power Requirements
Power Requirements 32
USA fast heating oven, 240 V 34
Canadian installation 34
Common instrument power cord plugs 34
ALS Power Requirements 38
This section details the power requirements for GC, GC/MS, and automatic liquid sampler
(ALS) installation.
Site Preparation Guide31
6GC System Power Requirements
WARNING
WARNING
CAUTION
Power Requirements
Power Requirements
Power consumption and requirements depend on the country to which the unit ships.
The number and type of electrical outlets depend on the size and complexity of the system.
To protect users, the metal instrument panels and cabinet are grounded through the
three-conductor power line cord in accordance with International Electrotechnical
Commission (IEC) requirements.
A proper earth ground is required for GC operations. Any interruption of the grounding
conductor or disconnection of the power cord could cause a shock that could result in
personal injury.
Be sure to verify proper receptacle grounding.
Do not use extension cords with Agilent instruments. Extension cords normally are not rated
to carry enough power and can be a safety hazard.
The length of the power cord is 2 meters (6.6 feet).
Do not use power line conditioners with Agilent instruments. Doing so may cause damage to
the equipment.
1Ensure that each instrument in your GC system can be connected to a dedicated circuit
with an isolated ground. (Note that ALS instruments receive their power from the GC.)
2Power requirements are printed near the power cord attachment on the rear panel of each
instrument. Although your GC should arrive ready for operation in your country, compare
its power requirements with those listed in Ta ble 10. If the voltage option you ordered is
not suitable for your installation, contact Agilent Technologies.
Table 10 Power requirements
Product
8860 GCStandard100
8860 GCStandard120
8860 GCStandard200/220/230/240
Oven type
Line voltage (VAC)
single phase
(–10% / +10%)
single phase
(–10% / +10%)
single phase
(–10% / +10%)
Frequency
(Hz)
48–63150012.515 Amp
48–63225018.820 Amp
48–6322509.6/9.3/
Maximum
continuous power
consumption (VA)
Current
rating
(amps)
9.3/9.2
Power outlet
current rating
Dedicated
Dedicated
10 Amp
Dedicated
32Site Preparation Guide
6GC System Power Requirements
NOTE
Power Requirements
Table 10 Power requirements (continued)
Product
MSD
5975 Series MSD120
5975 Series MSD220–240
5975 Series MSD200
5977 Series MSD120
5977 Series MSD220–240
5977 Series MSD200
MS
7010 or 7000 Triple Quad MS120
7010 or 7000 Triple Quad MS220–240
7010 or 7000 Triple Quad MS200
Oven type
Line voltage (VAC)
(–10% / +5%)
(–10% / +5%)
(–10% / +5%)
(–10% / +5%)
(–10% / +5%)
(–10% / +5%)
(–10% / +5%)
(–10% / +5%)
(–10% / +5%)
Frequency
(Hz)
50/60 ± 5%1100 (400 for foreline
50/60 ± 5%1100 (400 for foreline
50/60 ± 5%1100 (400 for foreline
50/60 ± 5%1100 (400 for foreline
50/60 ± 5%1100 (400 for foreline
50/60 ± 5%1100 (400 for foreline
50/60 ± 5%16001515 Amp
50/60 ± 5%16001515 Amp
50/60 ± 5%16001515 Amp
Maximum
continuous power
consumption (VA)
pump only)
pump only)
pump only)
pump only)
pump only)
pump only)
Current
rating
(amps)
810 Amp
810 Amp
810 Amp
810 Amp
810 Amp
810 Amp
Power outlet
current rating
Dedicated
Dedicated
Dedicated
Dedicated
Dedicated
Dedicated
Dedicated
Dedicated
Dedicated
7200 or 7250 Q-TOF MS200–240
HS
7697A HeadspaceAmericas: 120
7697A Headspace200/220/230/240
All
Data system PC
(monitor, CPU, printer)
Data system PC
(monitor, CPU, printer)
(–10% / +5%)
single phase
(–10% / +10%)
single/split phase
(–10% / +10%)
100/120
(–10% / +5%)
200-240
(–10% / +5%)
The GC and related equipment meet the following International Electrotechnical Commission
(IEC) classifications: Equipment Class I, Laboratory Equipment, Installation Category II, and
Pollution Degree 2.
50/60 ± 5%1800 (1200 for
foreline pump)
50/60 ± 5%8506.215 Amp
50/60 ± 5%8503.8/3.4/
50/60 ± 5%100010/8.315 Amp
50/60 ± 5%10004.1-510 Amp
1515 Amp
3.3/3.1
Dedicated
Dedicated
10 Amp
Dedicated
Dedicated
Dedicated
Site Preparation Guide33
6GC System Power Requirements
USA fast heating oven, 240 V
USA fast heating oven, 240 V
The 240 V fast heating oven requires 240 V/15 A power. Do not use 208 V power. Lower
voltage causes slow oven ramps and prevents proper temperature control. The power cord
supplied with your GC is rated for 250 V/15 A, and is a two-pole, three-wire cord with grounding
(type L6-15R/L6-15P).
Canadian installation
When installing a GC in Canada, make sure your GC power supply circuit meets the following
additional requirements:
•The circuit breaker for the branch circuit, which is dedicated to the instrument, is rated for
continuous operation.
•The service box branch circuit is marked as a “Dedicated Circuit.”
Common instrument power cord plugs
Table 11 shows common Agilent power cord plugs.
Table 11 Power cord terminations
Part number
8121-0675Argentina240164.5C19AS 3112
8120-1369Australia,
8120-8619Australia240162.5C19AS 3112
8121-1787Brazil240162.5C19IEC 60906-1
Country
New Zealand
Voltage Amps
240102.5C13AS 3112
Cable
length (m)
GC connector
type
Termination type Plug
8121-1809Brazil240102.5C13IEC 60906-1
34Site Preparation Guide
6GC System Power Requirements
Common instrument power cord plugs
Table 11 Power cord terminations
Part number
8120-6978Chile240102.5C13CEI 23-16
8121-0070China220162.5C19GB 1002
8121-0723China220102.5C13GB 1002
8120-3997Denmark,
Country
Greenland
Voltage Amps
230102.5C13AFSNIT 107-2-01
Cable
length (m)
GC connector
type
Termination type Plug
8120-8622Denmark,
Switzerland
8120-8621Europe220 / 230
8121-1222Korea220 / 230
8121-1226Korea220 / 230
230162.5C19Swiss/Denmark
/ 240
/ 240
/ 240
1302
162.5C19CEE/7/V11
162.5C19CEE/7/V11
102.5C13CEE/7/V11
Site Preparation Guide35
6GC System Power Requirements
Common instrument power cord plugs
Table 11 Power cord terminations
Part number
8121-0710India, South
8120-5182Israel230102.5C13Israeli SI32
8120-0161Israel23016, 16
8120-6903Japan200204.5C19NEMA L6-20P
Country
Africa
Voltage Amps
240152.5C19AS 3112
AWG
Cable
length (m)
2.5C19Israeli SI32
GC connector
type
Termination type Plug
8120-8620United Kingdom,
Hong Kong,
Singapore,
Malaysia
8120-8705United Kingdom,
Hong Kong,
Singapore,
Malaysia
8120-6894United States120202.5C19NEMA 5-20P
8120-1992United States120132.5C13NEMA 5-20P
240132.5C19BS1363/A
240102.3C13BS1363/A
36Site Preparation Guide
6GC System Power Requirements
Common instrument power cord plugs
Table 11 Power cord terminations
Part number
8121-0075United States240152.5C19NEMA L6-15P
8120-6360Taiwan, South
8121-1301Thailand220151.8C19
Country
America
Voltage Amps
120202.5C19NEMA 5-20P
Cable
length (m)
GC connector
type
Termination type Plug
G
Site Preparation Guide37
6GC System Power Requirements
WARNING
ALS Power Requirements
ALS Power Requirements
The ALS components draw power from the GC. No other power source is required.
The G4517A controller, used with the 8860 Series GC, requires one electrical outlet with a
dedicated ground. The controller can be set for either 100-120 V or 200-240 V.
Do not use extension cords with Agilent instruments. Extension cords normally are not rated
to carry enough power and can be a safety hazard.
The length of the power cord is 2 meters (6.6 feet).
38Site Preparation Guide
7Gas Selection and Plumbing
Gas and Reagent Selection 40
Hydrogen Carrier Gas 42
Gas and Reagent Purity 42
Gas Supplies 43
GC/MS Gas Requirements 45
Gas Plumbing 49
Supply tubing for most carrier and detector gases 50
Supply tubing for hydrogen gas 50
Two-stage pressure regulators 51
Pressure regulator-gas supply tubing connections 51
Filters and traps 52
This section outlines the requirements for gas selection and plumbing.
Refer to the Agilent Web site at www.agilent.com for the most up-to-date listing of GC, GC/MS,
and ALS supplies and consumables.
Site Preparation Guide39
7Gas Selection and Plumbing
WARNING
NOTE
Gas and Reagent Selection
Gas and Reagent Selection
Table 12 lists gases usable with Agilent GCs and capillary columns. When used with capillary
columns, GC detectors require a separate makeup gas for optimum sensitivity. The MS and
MSD use GC carrier gas.
If you are using any MS system, use of hydrogen as the carrier gas may require hardware
modifications for the best performance. Contact your Agilent service representative. Hydrogen
is not supported as a carrier gas with the 7200/7250 GC/Q-TOF system.
When using hydrogen (H2) as the carrier gas or fuel gas, be aware that hydrogen gas can flow
into the GC and create an explosion hazard. Therefore, be sure that the supply is turned off
until all connections are made and ensure the inlet and detector column fittings are either
connected to a column or capped at all times when hydrogen gas is supplied to the
instrument.
Hydrogen is flammable. Leaks, when confined in an enclosed space, may create a fire or
explosion hazard. In any application using hydrogen, periodically leak test all connections,
lines, and valves before operating the instrument or after maintenance. Always turn off the
hydrogen supply at its source before working on the instrument.
Please refer to the Hydrogen Safety Guide shipped with your instrument.
Hydrogen use is specifically prohibited with the 7200 or 7250 GC/Q-TOF system.
Nitrogen and Argon/Methane are generally not suitable for GC/MS carrier gas.
Table 12 Gases usable with Agilent GCs and capillary columns
Detector type
Electron capture (ECD)Hydrogen
Flame ionization (FID)Hydrogen
Flame photometric (FPD)Hydrogen
Nitrogen-Phosphorus (NPD)Helium
Carrier
Helium
Nitrogen
Argon/Methane (5%)
Helium
Nitrogen
Helium
Nitrogen
Argon
Nitrogen
Preferred makeup
NitrogenNitrogen
Nitrogen
Nitrogen
Nitrogen
Nitrogen
Nitrogen
Nitrogen
Nitrogen
Nitrogen
Nitrogen
Alternate choice
Nitrogen
Argon/Methane (5%)
Nitrogen
Helium
Helium
Helium
*
Helium
Helium
Detector, anode
purge, or reference
Anode purge must be
same as makeup
Hydrogen and air for
detector
Hydrogen and air for
detector
Hydrogen and air for
detector
40Site Preparation Guide
7Gas Selection and Plumbing
Gas and Reagent Selection
Table 12 Gases usable with Agilent GCs and capillary columns (continued)
Detector type
Thermal conductivity (TCD)Hydrogen
* Depending on bead type, higher makeup gas flow rates (> 5 mL/min) may introduce cooling effects or shorten bead life.
Carrier
Helium
Nitrogen
Preferred makeup
Must be same as carrier
and reference
Alternate choice
Must be same as carrier
and reference
Table 13 lists gas recommendations for packed column use. In general, makeup gases are not
required with packed columns.
Table 13 Gases usable with Agilent GCs and packed columns
Detector type
Electron capture (ECD)Nitrogen
Flame ionization (FID)Nitrogen
Flame photometric (FPD)Hydrogen
Nitrogen-Phosphorus (NPD)Helium
Carrier gas
Argon/methane
Helium
Helium
Nitrogen
Argon
Comments
Maximum sensitivity
Maximum dynamic range
Maximum sensitivity
Acceptable alternative
Optimum performance
Detector, anode purge, or
reference
Nitrogen
Argon/Methane
Hydrogen and air for detector.
Hydrogen and air for detector.
Hydrogen and air for detector.
Detector, anode
purge, or reference
Reference must be
same as carrier and
makeup
Nitrogen
Thermal conductivity (TCD)Helium
Hydrogen
Nitrogen
Argon
* Slightly greater sensitivity than helium. Incompatible with some compounds.
† For analysis of hydrogen or helium. Greatly reduces sensitivity for other compounds.
Acceptable alternative
General use
Maximum sensitivity
Hydrogen detection
Maximum hydrogen sensitivity
For installation checkout, Agilent requires the gas types shown in Tab le 14.
Table 14 Gases and reagents required for checkout
DetectorGases required
FIDCarrier: helium
Makeup: nitrogen
Fuel: hydrogen
Aux gas: Air
TCDCarrier and reference: helium
Reference must be same as
*
†
carrier and makeup.
*
Site Preparation Guide41
7Gas Selection and Plumbing
Hydrogen Carrier Gas
Table 14 Gases and reagents required for checkout (continued)
DetectorGases required
NPDCarrier: helium
Makeup: nitrogen
Fuel: hydrogen
Aux gas: Air
ECDCarrier: helium
FPDCarrier: helium
CI MS (external)Reagent gas: methane
CI MS (internal)Reagent: methanol
Anode purge and makeup: nitrogen
Makeup: nitrogen
Fuel: hydrogen
Aux gas: Air
MS and MSD systems purchased with a Self-Cleaning Ion Source also require a source of
hydrogen gas in addition to helium carrier gas. This source can be shared but must meet
carrier gas purity requirements.
Hydrogen Carrier Gas
Refer to the Agilent 8860 GC Safety Manual for important safety information about hydrogen
gas.
If hydrogen is being used as a carrier gas, or for the JetClean ion source system, special
considerations apply due to hydrogen’s flammability and chromatographic properties.
•Agilent highly recommends the G3388B Leak Detector to safely check for leaks.
•Hydrogen carrier gas requires special considerations for supply tubing. See “Gas
Plumbing” on page 49.
•In addition to the supply pressure requirements listed in “Gas Supplies” on page 43, Agilent
also recommends users of hydrogen gas consider gas source and purification needs. See
the additional recommendations in “Requirements for hydrogen as a carrier gas or for
use in JetClean systems” on page 45.
•When using hydrogen carrier gas with a ECD, TCD, or any other detector that vents
uncombusted gases, plan to vent the detector output to a fume hood or similar location.
Uncombusted hydrogen can present a safety hazard. See “Exhaust Venting” on page 28.
•When using hydrogen carrier gas, also plan to safely vent inlet split vent flows and purge
vent flows. See “Exhaust Venting” on page 28.
Gas and Reagent Purity
Agilent recommends that carrier and detector gases be 99.9995% pure. See Table 1 5 . Air
needs to be zero grade or better. Agilent also recommends using high quality traps to remove
hydrocarbons, water, and oxygen.
42Site Preparation Guide
7Gas Selection and Plumbing
NOTE
Swagelok nut and ferrules
Female Swagelok fittings on GC
Gas Supplies
Table 15 Carrier, collision and reagent gas purity
Carrier, collision and reagent gas
requirements
Helium (carrier)99.9995% Hydrocarbon free
Hydrogen (carrier) (carrier and
self-cleaning ion source)
Nitrogen (carrier)99.9995%
Nitrogen (drying gas, nebulizer
pressure)
Methanol
* Purity specification is the minimum acceptable purity. Major contaminants can be water,
† Required reagent for performance verification in internal CI mode only. Evaporation residue
*
†
oxygen, or air. Drying gas and nebulizer pressure gas can be supplied by a nitrogen gas
generator, house nitrogen system, or liquid nitrogen dewar.
<.0001%.
Purity
99.9995% SFC grade
99.999%Research grade
99.9%Reagent grade. Purge and
Notes
trap grade
recommended.
Gas Supplies
General requirements
Supply instrument gases using tanks, an internal distribution system, or gas generators. If
used, tanks require two-stage pressure regulators with packless, stainless steel diaphragms.
The instrument requires 1/8-inch Swagelok connections to its gas supply fittings.
See Figure 6.
Site Preparation Guide43
Plumb the gas supply tubing/regulators so that one 1/8-inch Swagelok female connector is
available for each gas needed at the instrument.
Figure 6.Example Swagelok connector and hardware
7Gas Selection and Plumbing
Gas Supplies
Table 16 lists available Agilent two-stage tank regulators. All Agilent regulators are supplied
The minimum supply pressure for Auxiliary EPC and PCM modules is 138 kPa (20 psi) greater
than the pressure used in your method. For example, if you need a pressure of 138 kPa (20 psi)
for the method, the supply pressure must be at least 276 kPa (40 psi). Table 1 9 lists the
maximum carrier pressure for Auxiliary EPC and PCM modules.
*
310–690 kPa
(45–100 psig)
690–827 kPa
(100–120 sig)
380–690 kPa
(55–100 psig)
380–690 kPa
(55–100 psig)
44Site Preparation Guide
7Gas Selection and Plumbing
GC/MS Gas Requirements
Table 19 Delivery pressures for Auxiliary EPC and PCM modules, in kPa (psig)
Aux EPCPCM 1PCM 2 or PCM Aux
Carrier (max)827 (120)827 (120)827 (120) with Forward pressure control
345 (50) with Backpressure control
Requirements for hydrogen as a carrier gas or for use in JetClean systems
Not all systems can use hydrogen as a carrier gas. See Gas Selection.
Hydrogen can be supplied from a generator or from a cylinder.
Agilent recommends use of a high-quality hydrogen gas generator. A high-quality generator
can consistently produce purity > 99.9999%, and the generator can include built-in safety
features such as limited storage, limited flow rates, and auto-shutdown. Select a hydrogen
generator that provides low (good) specifications for water and oxygen content.
If using a hydrogen gas cylinder, Agilent recommends use of Gas Clean Filters to purify the
gas. Consider additional safety equipment as recommended by your company safety
personnel.
GC/MS Gas Requirements
Table 20 lists typical flows resulting from selected carrier gas source pressures.
Table 20 5977 and 5975 Series MSD carrier gases
Carrier gas requirementsTypical pressure rangeTypical flow (mL/min)
Helium (required)
(column and split flow)
*
Hydrogen (optional)
(column and split flow)
Methane reagent gas
(required for CI operation)
Isobutane reagent gas (optional)103 to 172 kPa
Ammonia reagent gas (optional)34 to 55 kPa
Carbon dioxide reagent gas
(optional)
* Hydrogen gas can be used for the carrier gas but specifications are based on helium as the
carrier gas. Please observe all hydrogen gas safety cautions.
345 to 552 kPa
(50 to 80 psi)
345 to 552 kPa
(50 to 80 psi)
103 to 172 kPa
(15 to 25 psi)
(15 to 25 psi)
(5 to 8 psi)
103 to 138 kPa
(15 to 0 psi)
20 to 50
20 to 50
1 to 2
1 to 2
1 to 2
1 to 2
7010 and 7000 Series MS
Table 21 lists typical flows resulting from selected carrier gas source pressures.
Site Preparation Guide45
7Gas Selection and Plumbing
GC/MS Gas Requirements
Table 21 7010 and 7000 Triple Quad MS carrier gases
Carrier gas requirementsTypical pressure rangeTypical flow (mL/min)
Helium (required)
(column and split flow)
*
Hydrogen (optional)
(column and split flow)
Methane reagent gas
(required for CI operation)
Ammonia reagent gas (optional)34 to 55 kPa
Isobutane reagent gas (optional)
Carbon dioxide reagent gas (optional)
Nitrogen for collision cell (nitrogen source is supplied to EPC
module in GC.)
* Hydrogen gas can be used for the carrier gas but specifications are based on helium as the carrier gas. Please observe all hydrogen gas safety
cautions.
† Reagent available with manual tune only.
†
†
345 to 552 kPa
(50 to 80 psi)
345 to 552 kPa
(50 to 80 psi)
103 to 172 kPa
(15 to 25 psi)
(5 to 8 psi)
103 to 172 kPa
(15 to 25 psi)
103 to 138 kPa
(15 to 20 psi)
1.03 to 1.72 bar
(104 to 172 kPa, or 15 to 25 psi)
20 to 50
20 to 50
1 to 2
1 to 2
1 to 2
1 to 2
1 to 2 (mL/min)
7200 and 7250 Series Q-TOF MS
Table 22 lists the limits on total gas flow into the 7200/7250 Q-TOF GC/MS.
Table 22 7200/7250 Q-TOF GC/MS total gas flow limitations
Feature72007250
High vacuum pump 1Split-flow turboSplit-flow turbo
High vacuum pump 2Split-flow turboTurbo
High vacuum pump 3TurboTurbo
Carrier gas optimal gas flow, mL/min
Carrier gas maximum recommended gas flow, mL/min2.02.0
Carrier gas maximum gas flow, mL/min
Reagent gas flow (EI/CI – CI application)1.0 to 2.0NA
Collision cell gas flow rate, mL/min (Nitrogen)1.51.0
Collision cell gas flow rate, mL/min (Helium)4.0
Maximum column id0.32 mm (30 m long)0.32 mm (30 m long)
* Total gas flow into the MS = column flow + reagent gas flow (if applicable) + collision cell gas flow.
† Expect degradation of spectral performance and sensitivity.
*
†
1.0 to 1.51.0 to 1.5
2.42.4
Table 23 lists typical flows resulting from selected carrier and reagent gas source pressures.
46Site Preparation Guide
7Gas Selection and Plumbing
WARNING
GC/MS Gas Requirements
Table 23 7200/7250 Q-TOF GC/MS carrier and reagent gas flows
Carrier and reagent gas requirementsQ-TOFTypical pressure rangeTypical flow
Helium (required for carrier and IRM)7200173 to 207 kPa
Nitrogen for RIS transfer line actuator72006.1 to 6.8 bar
Nitrogen for collision cell (nitrogen source is supplied to
EPC module in GC.)
Helium for collision cell (helium source is supplied to
EPC module in GC.)
7200/72500.7 to 2.0 ar
72500.7 to 2.0 bar
(25 to 30 psi)
(612 to 690 kPa, or
(90 to 100 psi)
(70 to 207 kPa, or
10 to 30 psi)
(70 to 207 kPa, or
10 to 30 psi)
The use of hydrogen is specifically prohibited with the 7200/7250 GC/Q-TOF.
1.0 to 2.0 (mL/min)
Up to 30 L/min
1 to 2 (mL/min)
4 (mL/min)
Site Preparation Guide47
7Gas Selection and Plumbing
Performance verification
GC/MS systems with a JetClean ion source system installed use helium as the GC carrier gas
and an additional supply of hydrogen gas to the MS analyzer. Table 2 4 shows typical supply
pressures needed for operation. These values reflect the pressures supplied to the
instruments, not setpoints.
Table 24 JetClean ion source system gas supply pressures
Gas supplyPressure delivered at the GC
Helium690 kPa (100 psi)
Hydrogen
* Any delivery pressure ≤ 621 kPa (90 psi) is acceptable as long as it is 69 kPa (10 psi) higher
than the maximum hydrogen pressure needed during operation.
Performance verification
Performance verification requires the following:
•Helium carrier gas.
•For MS systems using chemical ionization or methane reagent gas.
≤ 621 kPa (90 psi)
*
48Site Preparation Guide
7Gas Selection and Plumbing
WARNING
Two-stage regulator
Main supply
on/off valve
Main gas
supply
On/off valve
Gas Clean filter
system. See “Filters
and traps”
on page 52
for more information.
Shutoff valve
Gas Clean filter configuration will vary depending on the
application.
Gas Plumbing
Gas Plumbing
All compressed gas cylinders should be securely fastened to an immovable structure or
permanent wall. Compressed gases should be stored and handled in accordance with the
relevant safety codes.
Gas cylinders should not be located in the path of heated oven exhaust.
To avoid possible eye injury, wear eye protection when using compressed gas.
Site Preparation Guide49
Figure 7.Recommended filters and plumbing configuration from a carrier gas cylinder
•If you have not requested option 305 (pre-plumbed tubing), you must supply pre-cleaned,
1/8-inch copper tubing and a variety of 1/8-inch Swagelok fittings to connect the GC to
inlet and detector gas supplies. See the Installation Kits for recommended parts.
•Agilent strongly recommends two-stage regulators to eliminate pressure surges.
High-quality, stainless-steel diaphragm-type regulators are especially recommended.
•On/off valves mounted on the outlet fitting of the two-stage regulator are not essential but
are very useful. Be sure the valves have stainless-steel, packless diaphragms.
•Agilent strongly recommends installation of shut-off valves at each GC inlet supply fitting
to allow the GC to be isolated for maintenance and troubleshooting. Order part number
0100-2144. (Note that some optional installation kits include one shut-off valve. See
“Installation Kits” on page 14for more information.)
•If you purchased automated valving, the valve actuation requires a separate pressurized,
dry air supply at 380 kPa (55 psig). This air supply must end in a male fitting compatible
with a 1/4-inch id plastic tube at the GC.
7Gas Selection and Plumbing
CAUTION
CAUTION
Supply tubing for most carrier and detector gases
•FID, FPD, and NPD detectors require a dedicated air supply. Operation may be affected by
pressure pulses in air lines shared with other devices.
•Flow- and pressure-controlling devices require at least
10 psi (138 kPa) pressure differential across them to operate properly. Set source
pressures and capacities high enough to ensure this.
•Situate auxiliary pressure regulators close to the GC inlet fittings. This ensures that the
supply pressure is measured at the instrument (rather than at the source); pressure at the
source may be different if the gas supply lines are long or narrow.
•Never use liquid thread sealer to connect fittings.
•Never use chlorinated solvents to clean tubing or fittings.
See “Installation Kits” on page 14 for more information.
Supply tubing for most carrier and detector gases
Use only preconditioned copper tubing (part number 5180-4196) to supply gases to the
instrument. Do not use ordinary copper tubing—it contains oils and contaminants.
Do not use methylene chloride or other halogenated solvent to clean tubing that will be used
with an electron capture detector. They will cause elevated baselines and detector noise until
they are completely flushed out of the system.
Do not use plastic tubing for suppling detector and inlet gases to the GC. It is permeable to
oxygen and other contaminants that can damage columns and detectors.
Plastic tubing can melt if near hot exhaust or components.
The tubing diameter depends on the distance between the supply gas and the GC and the total
flow rate for the particular gas. Tubing of 1/8-in diameter is adequate when the supply line is
less than 15 feet (4.6 m) long.
Use larger diameter tubing (1/4-inch) for distances greater then 15 feet (4.6 m) or when
multiple instruments are connected to the same source. Use larger diameter tubing if high
demand is anticipated (for example, air for an FID).
Be generous when cutting tubing for local supply lines—a coil of flexible tubing between the
supply and the instrument lets you move the GC without moving the gas supply. Take this
extra length into account when choosing the tubing diameter.
Supply tubing for hydrogen gas
Agilent recommends using new chromatographic quality stainless steel tubing and fittings
when using hydrogen.
•Do not re-use old tubing when installing or switching to hydrogen supply lines for carrier
gas or the JetClean ion source system. Hydrogen gas tends to remove contaminants left
on old tubing by previous gases (by helium, for example). These contaminants can appear
in output as high background noise or hydrocarbon contamination for several weeks.
•Especially do not use old copper tubing, which can become brittle.
50Site Preparation Guide
7Gas Selection and Plumbing
WARNING
Two-stage regulator
Adapter to 1/8-inch female
Swagelok fitting
Two-stage pressure regulators
Do not use old copper tubing with hydrogen gas. Old copper tubing can become brittle and
create a safety hazard.
Two-stage pressure regulators
To eliminate pressure surges, use a two-stage regulator with each gas tank. Stainless steel,
diaphragm-type regulators are recommended.
The type of regulator you use depends on the gas type and supplier. The Agilent catalog for
consumables and supplies contains information to help you identify the correct regulator, as
determined by the Compressed Gas Association (CGA). Agilent Technologies offers
pressure-regulator kits that contain all the materials needed to install regulators properly.
Pressure regulator-gas supply tubing connections
Use PTFE tape to seal the pipe-thread connection between the pressure regulator outlet and
Site Preparation Guide51
the fitting to which you connect the gas tubing. Instrument grade PTFE tape (part number
0460-1266), from which volatiles have been removed, is recommended for all fittings. Do not use pipe dope to seal the threads; it contains volatile materials that will contaminate the
tubing.
Pressure regulators typically end in fittings that must be adapted to the correct style or size.
Table 25 lists parts needed to adapt a standard 1/4-inch male NPT fitting to a 1/8-inch or
1/4-inch Swagelok fitting.
7Gas Selection and Plumbing
Filters and traps
Table 25 Parts for adapting NPT fittings
DescriptionPart number
Swagelok 1/8-inch to female 1/4-inch NPT, brass 0100-0118
Swagelok 1/4-inch to female 1/4-inch NPT, brass 0100-0119
Reducing union, 1/4-inch to 1/8-inch, brass, 2/pk5180-4131
Filters and traps
Using chromatographic-grade gases ensures that the gas in your system is pure. However, for
optimum sensitivity, install high-quality filters or traps to remove traces of water or other
contaminants. After installing a filter, check the gas supply lines for leaks.
Agilent recommends the Gas Clean Filter system. The Gas Clean Filter system delivers high
purity gases to your analytical instruments, reducing the risk of column damage, sensitivity
loss, and instrument downtime. The filters are designed for use with the GC, GC/MS, ICP-OES,
ICP-MS, LC/MS, and any other analysis instrument using carrier gas. Six filters are available,
including CO2, oxygen, moisture, and organics trap (charcoal).
Filter types
Each Gas Clean Filter type is designed to filter out a specific impurity that may exist in the gas
supply. The following filter types are available:
•Oxygen - Prevents oxidation of the GC column, septum, liner, and glass wool.
•Moisture - Delivers fast stabilization times for increased GC productivity, and prevents
hydrolization damage to the stationary phase, column, liner, glass wool, or septum in the
GC.
•Process Moisture - Prevents oxidation of GC components and is safe to use with acetylene
in process GC applications.
•Charcoal - Removes organic compounds and ensures correct performance of FID
detectors in the GC.
•GC/MS - Delivers fast stabilization times for increased GC productivity, removes oxygen,
moisture, and hydrocarbons from the carrier gas for MS applications, and provides
ultimate GC column protection.
Table 26 lists the most common Gas Clean Filter system kits. See the Agilent online store or
contact your local Agilent sales representative for additional filters, parts, and accessories
applicable to your instrument configuration.
52Site Preparation Guide
7Gas Selection and Plumbing
Filters and traps
Table 26 Recommended Gas Clean Filter kits
DescriptionPart numberUse
Gas Clean Filter kit (connecting unit for one filter, including one carrier gas filter, 1/8-inch
connections, a smart sensor, and mounting bracket for the GC)
Gas Clean Filter kit (connecting unit for four filters, including four filters, 1/4-inch
connections)
Gas Clean Filter kit (connecting unit for four filters, including four filters, 1/8-inch
connections)
GC/MS Gas Clean Filter kit (includes one connecting unit and two GC/MS filters, 1/8-inch
connections)
GC/MS Gas Clean Filter kit (includes one connecting unit and two GC/MS filters, 1/4-inch
connections)
GC/MS Gas Clean Filter installation kit (includes CP17976, 1 m copper tubing, and two
1/8-inch nuts and ferrules)
TCD filter kit (with oxygen and moisture filters)CP738408TCD
CP179880Carrier gas only
CP7995FID, FPD, NPD
CP736530FID, FPD, NPD
CP17976ECD, GC/MS
CP17977ECD, GC/MS
CP17978ECD, GC/MS
Each separate gas supply requires its own filters.
See also “Installation Kits” on page 14.
Site Preparation Guide53
7Gas Selection and Plumbing
Filters and traps
54Site Preparation Guide
ALAN Requirements
Site LAN Network 56
This section outlines the site LAN requirements for GC, GC/MS, and automatic liquid sampler
(ALS) installation.
Site Preparation Guide55
ALAN Requirements
NOTE
NOTE
NOTE
NOTE
NOTE
Site LAN Network
Site LAN Network
Agilent Technologies is not responsible for connecting to or establishing communication
with your site LAN network. The Agilent representative will test the system’s ability to
communicate on a mini-hub or LAN switch only.
If you intend to connect your system to your site’s LAN network, you must have an additional
shielded twisted pair network cable (8121-0940).
The IP addresses assigned to the instrument(s) must be fixed (permanently assigned)
addresses. If you intend to connect your system to your site’s network, each piece of
equipment must have a unique, fixed (static) IP address assigned to it.
For a Single Quad GC/MS system, Agilent recommends, sells, and supports the use of a PC
with one (1) network interface card (NIC) and a network switch to isolate the GC/MS system
from the site LAN. The network switch supplied with Agilent systems prevents
instrument-to-PC network traffic from entering the site LAN and keeps site LAN network
traffic from interfering with instrument-to-PC communications. Agilent develops and tests all
Single Quad GC/MS hardware and software using the single NIC configuration and has no
known network configuration issues. Alternate network configurations can be configured
and managed by the end user at their own risk and expense.
For Triple Quad and 7200 Q-TOF GC/MS systems, Agilent recommends, sells, and supports
the use of a PC with two network interface cards (NIC) to provide both a site LAN connection
and an isolated GC/MS system connection. Agilent develops and tests all Triple Quad and
Q-TOF GC/MS hardware and software using the dual NIC configuration, and has no known
network configuration issues. Alternate network configurations can be configured and
managed by the end user at their own risk and expense.
For the 7250 Q-TOF GC/MS systems, Agilent recommends, sells, and supports the use of a
PC with three network interface cards (NIC) to provide a site LAN connection, MS only
connection, and an isolated GC/MS system connection. Agilent develops and tests Q-TOF
GC/MS hardware and software using the triple NIC configuration, and has no known network
configuration issues. Alternate network configurations can be configured and managed by
the end user at their own risk and expense.
56Site Preparation Guide
This page intentionally left blank.
57Site Preparation Guide
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.