Agilent Technologies 86100b User Manual

Agilent 86100B Wide-Bandwidth Oscilloscope Getting Started

Notices
© Agilent Technologies, Inc. 2000-2002
No part of this manual may be repro­duced in any form or by any means (including electronic storage and retrieval or translation into a foreign lan­guage) without prior agreement and writ­ten consent from Agilent Technologies, Inc. as governed by United States and international copyright lays.
Manual Part Number
86100-90054
Edition
First edition, October 2002
Printed in USA
Agilent Technologies, Inc. Lightwave Division 3910 Brickway Boulevard Santa Rosa, CA 95403, USA
Warranty
The material contained in this document is provided “as is,” and is subject to being changed, without notice, in future edi­tions. Further, to the maximum extent permitted by applicable law, Agilent dis­claims all warranties, either express or implied, with regard to this manual and any information contained herein, includ­ing but not limited to the implied warran­ties of merchantability and fitness for a particular purpose. Agilent shall not be liable for errors or for incidental or conse­quential damages in connection with the furnishing, use, or performance of this document or of any information con­tained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the mate­rial in this document that conflict with these terms, the warranty terms in the separate agreement shall control.
Technology Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.
Safety Notices
CAUTION
Caution denotes a hazard. It calls attention to a procedure which, if not correctly per­formed or adhered to, could result in damage to or destruction of the product. Do not proceed beyond a caution sign until the indicated conditions are fully understood and met.
WARNING
Warning denotes a hazard. It calls attention to a procedure which, if not correctly per­formed or adhered to, could result in injury or loss of life. Do not proceed beyond a warning sign until the indicated conditions are fully understood and met.
Restricted Rights Legend
If software is for use in the performance of a U.S. Government prime contract or subcontract, Software is delivered and licensed as “Commercial computer soft­ware” as defined in DFAR 252.227-7014 (June 1995), or as a “commercial item” as defined in FAR 2.101(a) or as “Restricted computer software” as defined in FAR
52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclosure of Software is subject to Agilent Technologies’ standard commercial license terms, and non-DOD Departments and Agencies of the U.S. Government will receive no greater than Restricted Rights as defined in FAR
52.227-19(c)(1-2) (June 1987). U.S. Gov­ernment users will receive no greater than Limited Rights as defined in FAR
52.227-14 (June 1987) or DFAR 252.227­7015 (b)(2) (November 1995), as appli­cable in any technical data.
ii

Contents

1 Setting Up the Infiniium DCA
Step 1. Inspect the Shipment 1–5 Step 2. Install any Modules 1–9 Step 3. Consider Environmental Specifications 1–10 Step 4. Connect the Keyboard and Mouse (Optional) 1–11 Step 5. Connect the Line Cord 1–12 Step 6. Turn on the line power 1–15 Step 7. Set the time and date 1–17 Step 8. Connect a Printer 1–18 Step 9. Calibrate the Touch Screen 1–22 Step 10. Configure the Instrument on a LAN 1–23 Step 11. Avoid Costly Repairs 1–24 Step 12. Perform a Module Calibration 1–26 Step 13. For More Information 1–29 Cleaning Connections for Accurate Measurements 1–30 Returning the Instrument for Service 1–39 Hard Drive Recovery 1–42 Agilent Technologies Service Offices 1–43
2 Working in Comfort
3 Using the Infiniium DCA
Eye/Mask Mode 3–7 Oscilloscope Mode 3–8 TDR/TDT Mode 3–9 Using Plug-in Modules 3–10 Using the Touch Screen 3–12 Using the Quick Measure Feature 3–13 If the Display Goes Blank 3–14 Hiding and Moving Dialog Boxes 3–15 Menus 3–16
4 Using the Built-In Information System
The Infiniium DCA Help Contents 4–3 Description of the Contents Topics 4–4 Getting Help While Changing Instrument Settings 4–6 Learning About Measurements Results 4–7
Contents-1
Contents
Hiding the Help 4–8 Printing the Contents of a Topic 4–9 Selecting the Help Language 4–10
5 Regulatory Information
Compliance with German Noise Requirements 5–2 Compliance with Canadian EMC Requirements 5–2 Declaration of Conformity - Mainframe 5–3 Declaration of Conformity - Plug-in Modules 5–4
Contents-2
1
Step 1. Inspect the Shipment 1-5 Step 2. Install any Modules 1-9 Step 3. Consider Environmental Specifications 1-10 Step 4. Connect the Keyboard and Mouse (Optional) 1-11 Step 5. Connect the Line Cord 1-12 Step 6. Turn on the line power 1-15 Step 7. Set the time and date 1-17 Step 8. Connect a Printer 1-18 Step 9. Calibrate the Touch Screen 1-22 Step 10. Configure the Instrument on a LAN 1-23 Step 11. Avoid Costly Repairs 1-24 Step 12. Perform a Module Calibration 1-26 Step 13. For More Information 1-29 Cleaning Connections for Accurate Measurements 1-30 Returning the Instrument for Service 1-39 Hard Drive Recovery 1-42 Agilent Technologies Service Offices 1-43

Setting Up the Infiniium DCA

Setting Up the Infiniium DCA
Setting Up the Infiniium DCA
Setting Up the Infiniium DCA
This chapter shows you how to properly set up your Infiniium DCA. Follow the setup steps in the order presented. In addition, you will learn how to use proper optical connection cleaning techniques to avoid costly repairs. Lastly, this chapter explains how to return your instrument for service.

General Safety Considerations

This product has been designed and tested in accordance with the standards listed in the Manufacturers Declaration of Conformity, and has been supplied in a safe condition. The documentation contains information and warnings which must be followed by the user to ensure safe operation and to maintain the product in a safe condition.
Install the instrument according to the enclosure protection provided. This instrument does not protect against the ingress of water. This instrument pro­tects against finger access to hazardous parts within the enclosure.
1-2
Setting Up the Infiniium DCA
General Safety Considerations
WARNING If this product is not used as specified, the protection provided by the
equipment could be impaired. This product must be used in a normal condition (in which all means for protection are intact) only.
WARNING No operator serviceable parts inside. Refer servicing to qualified
service personnel. To prevent electrical shock do not remove covers.
WARNING To prevent electrical shock, disconnect the instrument from mains
before cleaning. Use a dry cloth or one slightly dampened with water to clean the external case parts. Do not attempt to clean internally.
CAUTION Do not use too much liquid in cleaning the instrument. Water can enter the
front-panel keyboard, damaging sensitive electronic components.
WARNING This is a Safety Class 1 Product (provided with a protective earthing
ground incorporated in the power cord). The mains plug shall only be inserted in a socket outlet provided with a protective earth contact. Any interruption of the protective conductor inside or outside of the instrument is likely to make the instrument dangerous. Intentional interruption is prohibited.
CAUTION VENTILATION REQUIREMENTS: When installing the product in a cabinet, the
convection into and out of the product must not be restricted. The ambient temperature (outside the cabinet) must be less than the maximum operating temperature of the product by 4° C for every 100 watts dissipated in the cabinet. If the total power dissipated in the cabinet is greater than 800 watts, then forced convection must be used.
CAUTION This product is designed for use in Installation Category II and Pollution
Degree 2 per IEC 1010 and 664 respectively.
Installation Category: Installation Categories (overvoltage categories) are determined by the transient overvoltage levels that may be expected. CAT I: Mains isolated. CAT II: Line voltage in appliance and to wall outlet. CAT III: Line voltage behind wall outlet to next level of distribution.
1-3
Setting Up the Infiniium DCA

Instrument Markings

Instrument Markings
The instruction manual symbol. The product is marked with this warn­ing symbol when it is necessary for the user to refer to the instructions in the manual.
The laser radiation symbol. This warning symbol is marked on products which have a laser output.
The AC symbol is used to indicate the required nature of the line mod­ule input power.
The Standby symbol is used to mark the position of the instrument power line switch.
The CE mark is a registered trademark of the European Community.
The CSA mark is a registered trademark of the Canadian Standards Association.
The C-Tick mark is a registered trademark of the Australian Spectrum Management Agency.
This text denotes the instrument is an Industrial Scientific and Medical
ISM1-A
Group 1 Class A product.
1-4

Step 1. Inspect the Shipment

Setting Up the Infiniium DCA
Step 1. Inspect the Shipment
Inspect the shipping container for damage.
Inspect the instrument.
Verify that you received the options and accessories you ordered.
1-5
Setting Up the Infiniium DCA

Options and Accessories

NOTE Keep the shipping container and cushioning material until you have inspected the
contents of the shipment for completeness and have checked the instrument mechanically and electrically.
If anything is missing or defective, contact your nearest Agilent Technologies Sales Office. Refer to Agilent Technologies Service Offices on page 1-43. If the shipment was damaged, contact the carrier, then contact the nearest Agi­lent Sales Office. Keep the shipping materials for the carrier’s inspection. The Agilent Sales Office will arrange for repair or replacement at Agilent’s option without waiting for claim settlement.
Options and Accessories
The following table lists the accessories that may be on the Infiniium DCA shipping list. The information on your actual shipping list is more accurate and should supersede the information in this table.
Table 1-1. Standard Accessories
Accessory Part Number
Keyboard 1150-7809
Mouse C3751-60201
Getting Started (this book) 86100-90050
Programmer’s Guide 86100-90049
Filler panel 86101-60005
APC 3.5 (f-f) Adapter/connector saver 5061-5311
The following table lists the available product options. Make sure that you received all of the options that you ordered.
Table 1-2. Instrument Options
Option Description
001 12 GHz Trigger Bandwidth
AX4 Rack mount flange kit
AXE Rack mount flange kit with handles
UK6 Commercial calibration certificate with test data
1-6
Table 1-3. Optional Accessories
Accessory Description
10086A ECL terminator
11667B Power splitter, dc to 26.5 GHz, APC 3.5 mm
11667C Power splitter, dc to 50 GHz, 2.4 mm
11898A Plug-in module remote/extender
11982A High-speed lightwave receiver
54006A 6 GHz Resistive divider probe kit
54008B 24 ns delay line
54701A 2.5 GHz active probe
83430A 2.5 Gb/s Lightwave transmitter (1550 nm)
83433A 10 Gb/s Lightwave transmitter
83434A 10 Gb/s Lightwave receiver
83440B/C/D Optical-to-Electrical Converters (6/20/34 GHz)
83446A Lightwave clock and data receiver
83480A-K16 Switch matrix
83480A-K17 RIMM fixture
N1020A 6 GHz TDR probe kit (passive)
N1025A 1 GHz Active differential probe
Setting Up the Infiniium DCA
Options and Accessories
Table 1-4. Additional Adapters for Electrical Channels
Adapter Agilent Part Number
SMA (f-f) 1250-1158
APC 3.5 (f-f) 5061-5311
1-7
Setting Up the Infiniium DCA
Options and Accessories
Table 1-5. Fiber-Optic Adapters for Optical Channels
Front Panel Fiber-Optic Adapter Description Agilent Part Number
Diamond HMS-10 81000AI
FC/PC 81000FI
D4 81000GI
SC 81000KI
DIN 81000SI
1-8
ST 81000VI
Biconic 81000WI
Setting Up the Infiniium DCA

Step 2. Install any Modules

Step 2. Install any Modules
Up to two modules can be inserted into the Infiniium DCA. All documentation (including specifications) for plug-in modules is located in the Infiniium DCA’s Help system. To access the Help, touch or click Contents on the Help menu.
1 Slide the module into an available front-panel slot.
You can remove or install a module while the instrument is operating.
2 Finger tighten the two knurled screws on the module’s front panel.
1-9
Setting Up the Infiniium DCA

Step 3. Consider Environmental Specifications

Step 3. Consider Environmental Specifications
Review the following specifications to ensure that your operating or storage environment is suitable for the instrument.
Table 1-6. Environmental Specifications
Use indoor Temperature
Operating 10 °C to +40 °C (50 °F to + 104 °F) Non-operating -40 °C to +70 °C (-40 °F to +158 °F)
Humidity
Operating Up to 90% humidity (non-condensing) at +40 °C (+104 °F) Non-operating Up to 95% relative humidity at + 65 °C (+149 °F)
Altitude
Operating Up to 4,600 meters (15,000 ft) Non-operating Up to 15,300 meters (50,000 ft)
Vibration
Operating Random vibration 5-500 Hz, 10 minutes per axis, 2.41 g(rms) Non-operating Random vibration 5-500 Hz, 10 minutes per axis, 0.3 g(rms);
Resonant search, 5-500Hz swept sine, 1 octave/minute sweep rate, 0.75g, 5 minute resonant dwell at 4 resonances per axis.
Power requirements
Voltage 90 to 132 or 198 to 264 Vac, 48-66 Hz CAT II Power (including modules) 545 VA; 375 W
Weight
Mainframe 15.5 kg (34 lb) Modules 1.2 kg (2.6 lb)
Dimensions
Height 215.1 mm (8.47 in) Width 425.5 mm (16.75 in) Depth 566 mm (22.2 in) Depth with front connectors
and rear feet
629 mm (24.8 in)
1-10
Setting Up the Infiniium DCA

Step 4. Connect the Keyboard and Mouse (Optional)

Step 4. Connect the Keyboard and Mouse (Optional)
Connect the standard PC-compatible mouse and keyboard that is supplied with the instrument.
NOTE Do not stack other objects on the keyboard; this will cause self-test failures on power-on.
1-11
Setting Up the Infiniium DCA

Step 5. Connect the Line Cord

Step 5. Connect the Line Cord
Connect the line cord as shown in the figure below.
The Infiniium DCA automatically adjusts for line input voltages in the range of 100 to 240 VAC. There is no manual selection switch. The line cord provided is matched by Agilent to the country in which the order originates. Refer to
Table 1-8, Available Line Cords, on page 1-13 for a list of available line power
cords.
Install the instrument so that the detachable power cord is readily identifiable and is easily reached by the operator. The detachable power cord is the instrument disconnecting device. It disconnects the mains circuits from the mains supply before other parts of the instrument. The front panel switch is
1-12
Setting Up the Infiniium DCA
Step 5. Connect the Line Cord
only a standby switch and is not a LINE switch. Alternatively, an externally installed switch or circuit breaker (which is readily identifiable and is easily reached by the operator) may be used as a disconnecting device.
CAUTION Always use the three-prong AC power cord supplied with this instrument.
Failure to ensure adequate earth grounding by not using this cord may cause instrument damage.
CAUTION Do not connect ac power until you have verified the line voltage is correct as
described in Table 1-7. Damage to the equipment could result.
CAUTION This instrument has autoranging line voltage input. Be sure the supply voltage
is within the specified range.
Table 1-7. Line Power Requirements
Power 115 VAC: 110 VA MAX. / 60 WATTS MAX. / 1.1 A MAX.
230 VAC: 150 VA MAX. / 70 WATTS MAX. / 0.6 A MAX.
Voltage nominal: 115 VAC / 230 VAC CAT II
range 115 VAC: 90–132 V range 230 VAC: 198–254 V
Frequency nominals: 50 Hz / 60 Hz
range: 47–63 Hz
Table 1-8. Available Line Cords
Plug Type Cable Part No. Plug Description
250V 8120-1351
8120-1703
250V 8120-1369
8120-0696
Straight *BS1363A
90°
Straight *NZSS198/ASC
90°
Length (in/cm)
90/228
90/228
79/200
87/221
Color Country
Gray
Mint Gray
Gray
Mint Gray
United Kingdom, Cyprus, Nigeria, Zimbabwe, Singapore
Australia, New Zealand
1-13
Setting Up the Infiniium DCA
Step 5. Connect the Line Cord
Table 1-8. Available Line Cords
Plug Type Cable Part No. Plug Description
250V 8120-1689
8120-1692
8120-2857p
125V 8120-1378
8120-1521
8120-1992
250V 8120-2104
8120-2296
Straight *CEE7-Y11
90°
Straight (Shielded)
Straight *NEMA5-15P
90°
Straight (Medical) UL544
Straight *SEV1011
1959-24507
Type 12 90°
220V 8120-2956
8120-2957
250V 8120-4211
8120-4600
Straight *DHCK107
90°
Straight SABS164
90°
Length (in/cm)
79/200
79/200
79/200
90/228
90/228
96/244
79/200
79/200
79/200
79/200
79/200
79/200
Color Country
Mint Gray
Mint Gray
Coco Brown
Jade Gray
Jade Gray
East and West Europe, Saudi Arabia, So. Africa, India (unpolarized in many nations)
United States, Canada, Mexico, Philippines, Taiwan
Black
Mint Gray
Switzerland
Mint Gray
Mint Gray
Denmark
Mint Gray
Jade Gray Republic of South Africa
India
100V 8120-4753
8120-4754
* Part number shown for plug is the industry identifier for the plug only. Number shown for
cable is the Agilent Technologies part number for the complete cable including the plug.
1-14
Straight MITI
90°
90/230
90/230
Dark Gray Japan

Step 6. Turn on the line power

Setting Up the Infiniium DCA
Step 6. Turn on the line power
Press the power switch at the lower left-hand corner of the front panel.
After about one minute, the Infiniium DCA will be ready to use. The display will look similar to the following figure.
1-15
Setting Up the Infiniium DCA

Interacting With the Display

NOTE The detachable power cord is the instrument disconnecting device. It disconnects the
mains circuits from the mains power supply before other parts of the instrument. The front panel switch is only a standby switch and is not a LINE switch disconnecting device. Install the instrument so that the detachable power cord is readily identifiable and is easily reached by the operator.
Interacting With the Display
Although you can operate all instrument functions using a keyboard and mouse, the touch screen feature makes your Infiniium DCA even easier to use. You can explore the instruments functions and settings by touching elements on the display (such as menus, buttons, and other controls) with your finger or the supplied stylus.
Although this book uses the term “touch” to encourage you to experience using the touch screen, you can always click using a mouse or PC-compatible pointing device instead.
NOTE The display will always function as a touch screen, even when a mouse and keyboard are
connected.
CAUTION Avoid touching the screen with a sharp object, as this could result in damage to
the display. Use your finger or the supplied stylus.
1-16
Setting Up the Infiniium DCA

Step 7. Set the time and date

Step 7. Set the time and date
1 Touch the current time and date shown near the top of the display. The Set
Tim e and Date dialog box opens.
2 Enter the current time and date in one of the following ways:
Click the time and date text boxes, and enter the desired number.
Click the up or down arrows next to the time and date text boxes until the
desired number is displayed.
3 Touch Set Time & Date Now to apply the changes and close the dialog box.
1-17
Setting Up the Infiniium DCA

Step 8. Connect a Printer

Step 8. Connect a Printer
Use of a printer with the Infiniium DCA is optional. Table 1 -9 lists the PCL-lan­guage printers that are currently supported for use with the instrument. To add a local printer, you must use the Add Printer Wizard to select the printer driver for the printer as outlined in this procedure; the Add Printer Wizard does not run automatically. For more information on connecting and using a printer, refer to the built-in information system. To add a network printer, con­tact your network administrator.
Unsupported printers
If your printer is not in the list of supported printers, you will need the printer driver disk that came with your printer. To install the printer driver from this disk, click Have Disk on the Add Printer Wizard dialog box. The instrument will then prompt you to put your printer driver disk into the front-panel disk drive (drive A) or rear-panel CD-Rom drive. Be sure to install the Windows 98 printer driver for your printer. If a printer driver disk was not provided or you cannot find the disk, you can contact the printers manufacturer to obtain the appropriate Windows 98 printer driver.
Table 1-9. Supported Printers
HP DeskJet 850 HP DeskJet 1600C
HP DeskJet 890C HP LaserJet 4P
HP DeskJet 970 HP LaserJet 4000N
HP DeskJet 1200C
To connect a printer
1 Turn on the power to the Infiniium DCA.
2 On the File menu, select Print Setup and then Add Printer.
An online Help window opens with instructions for connecting a printer. You do not need to read the displayed instructions at this time.
1-18
Setting Up the Infiniium DCA
Step 8. Connect a Printer
3 Turn the power off the Infiniium DCA. Do not exit the online Help before
turning the instrument off.
4 Connect the printer to the instruments rear-panel Parallel port. Refer to
Rear-panel features on page 3-6 to locate the Parallel port. Use the cable
that came with your printer.
5 Turn the printer power on.
6 Wait five seconds and then turn the power on the Infiniium DCA again.
7 The following message appears. Touch No.
8 The following message appears. Touch OK.
9 The instrument
automatically recognizes the printer connected on the printer port and displays the Add Printer Wizard. Touch Next.
1-19
Setting Up the Infiniium DCA
Step 8. Connect a Printer
10 Select Local printer and
then touch Next.
11 Select the printer
manufacturer and then printer. Then, touch Next. If your printer is not listed, touch Have Disk. The instrument will then prompt you to put your printer driver disk into the front-panel disk drive (drive A) or CD-Rom drive. Be sure to install the Windows 98 printer driver for your printer.
12 Select LPT1 for the
printer port. Touch Next.
1-20
13 Touch Yes and then
touch Next.
14 Touch Yes to print a test
page. Then, touch Next.
15 After confirming that the
test page printed correctly, touch Yes.
Setting Up the Infiniium DCA
Step 8. Connect a Printer
16 Turn the Infiniium DCA’s power off and then on again. The printer is now
installed and ready to use.
1-21
Setting Up the Infiniium DCA

Step 9. Calibrate the Touch Screen

Step 9. Calibrate the Touch Screen
The touch screen configuration utility ensures that the touch screen is both aligned and orientated with the display. Therefore, when you touch an ele­ment on the display screen, the instrument can detect the task you want to perform.
1 On the Utilities menu, choose Touch Screen Calibration. The touch screen
configuration utility opens.
2 Touch the center of the X. The X will move to each corner of the display screen.
The final X will appear in the center of the display screen. When the display is calibrated, the touch screen configuration utility closes.
The calibration will time out within 10 seconds if you do not touch the center of the X. You can also press Esc on an optional keyboard to exit the touch screen calibration.
1-22
Setting Up the Infiniium DCA

Step 10. Configure the Instrument on a LAN

Step 10. Configure the Instrument on a LAN
You can configure this instrument on a LAN for file sharing and using network printers. Contact your companys network administrator to set up your instru­ment with the appropriate client, protocol, and configuration for your local area network (LAN). The network setup is different for every company. If you plan on sharing your instrument hard disk drive with other computers on your LAN, you should have the network administrator enable file sharing as part of the network setup. For more information on setting up a network drive, refer to the built-in information system.
1-23
Setting Up the Infiniium DCA

Step 11. Avoid Costly Repairs

Step 11. Avoid Costly Repairs
CAUTION Electrical channel input circuits and the trigger input circuit can be damaged
by electrostatic discharge (ESD). Therefore, avoid applying static discharges to the front-panel input connectors. Prior to connecting any coaxial cable to the connectors, momentarily short the center and outer conductors of the cable together. Avoid touching the front-panel input connectors without first touching the frame of the instrument. Be sure that the instrument is properly earth-grounded to prevent buildup of static charge. Wear a wrist-strap or heel­strap.
CAUTION Optical channel fiber-optic connectors are easily damaged when connected to
dirty or damaged cables and accessories. When you use improper cleaning and handling techniques, you risk expensive instrument repairs, damaged cables,
1-24
Setting Up the Infiniium DCA
Step 11. Avoid Costly Repairs
and compromised measurements. Before you connect any fiber-optic cable to the Infiniium DCA, refer to Cleaning Connections for Accurate
Measurements on page 1-30.
1-25
Setting Up the Infiniium DCA

Step 12. Perform a Module Calibration

Step 12. Perform a Module Calibration
A module calibration establishes calibration factors for all installed modules. The calibration factors compensate for imperfections in the measurement sys­tem (for example, variations due to ambient temperature), resulting in the best instrument precision. It is recommended that you perform this calibration routinely for best measurement accuracy.
The module calibration is recommended whenever the following occurs:
Instrument power has been cycled
A module has been removed and then reinserted
A change in the temperature of the mainframe exceeds 5°C compared to the temperature of the last vertical (amplitude) calibration (T > 5°C)
The time since the last calibration has exceeded 10 hours
Refer to the Help for more information about instrument calibrations.
Warm-up time
Allow the instrument, with modules installed, to warm up for a period of at least one hour before performing a vertical calibration. This ensures that the module reaches its equilibrium temperature.
CAUTION Electrical channel input circuits and the trigger input circuit can be damaged
by electrostatic discharge (ESD). Therefore, avoid applying static discharges to the front-panel input connectors. Prior to connecting any coaxial cable to the connectors, momentarily short the center and outer conductors of the cable together. Avoid touching the front-panel input connectors without first touching the frame of the instrument. Be sure that the instrument is properly earth-grounded to prevent buildup of static charge. Wear a wrist-strap or heel­strap.
1-26
Setting Up the Infiniium DCA
Step 12. Perform a Module Calibration
To perform a module calibration
1 On the Calibrate menu, choose All Calibrations.
The All Calibrations dialog box opens. This dialog box allows you to view the calibration status of modules and initiate the calibration.
2 Touch the Modules tab.
The Modules tab opens and allows you to perform a module calibration on either the left or right module.
3 Touch the Calibrate Module button (located at the bottom of
the tab) for the module you need to calibrate. This starts the Module calibration.
4 Remove all external devices and signals from the module.
The module calibration uses known signal levels in the instrument. Introduc­ing an outside signal source will interfere with the calibration factors and decrease the accuracy of the calibration.
5 Touch Continue. A progress indicator appears.
6 After the calibration completes, choose All Calibrations on the Calibration
menu.
7 The Module tab shows the status of the calibration. The tab will look similar to
the following figure. Notice that the tab indicates that the Current user
1-27
Setting Up the Infiniium DCA
Step 12. Perform a Module Calibration
module calibration is valid.
1-28
Setting Up the Infiniium DCA

Step 13. For More Information

Step 13. For More Information
There are several ways to learn more about your Infiniium DCA.
Continue reading this book. Chapter 3, Using the Infiniium DCA will help you get started using this instrument.
Refer to the Help. The Help contains the information that would normally be in the users guide.
Visit our website at http://www.agilent.com. Use the keywords “Infiniium DCA or 86100B in your search
1-29
Setting Up the Infiniium DCA

Cleaning Connections for Accurate Measurements

Cleaning Connections for Accurate Measurements
Today, advances in measurement capabilities make connectors and connec­tion techniques more important than ever. Damage to the connectors on cali­bration and verification devices, test ports, cables, and other devices can degrade measurement accuracy and damage instruments. Replacing a dam­aged connector can cost thousands of dollars, not to mention lost time! This expense can be avoided by observing the simple precautions presented in this book. This book also contains a brief list of tips for caring for electrical connec­tors.

Choosing the Right Connector

A critical but often overlooked factor in making a good lightwave measure­ment is the selection of the fiber-optic connector. The differences in connec­tor types are mainly in the mechanical assembly that holds the ferrule in position against another identical ferrule. Connectors also vary in the polish, curve, and concentricity of the core within the cladding. Mating one style of cable to another requires an adapter. Agilent Technologies offers adapters for most instruments to allow testing with many different cables. Figure 1-1 on
page 1-31 shows the basic components of a typical connectors.
The system tolerance for reflection and insertion loss must be known when selecting a connector from the wide variety of currently available connectors. Some items to consider when selecting a connector are:
How much insertion loss can be allowed?
Will the connector need to make multiple connections? Some connectors are
better than others, and some are very poor for making repeated connections.
What is the reflection tolerance? Can the system take reflection degradation?
Is an instrument-grade connector with a precision core alignment required?
Is repeatability tolerance for reflection and loss important? Do your specifica-
1-30
Setting Up the Infiniium DCA
Choosing the Right Connector
tions take repeatability uncertainty into account?
Will a connector degrade the return loss too much, or will a fusion splice be re­quired? For example, many DFB lasers cannot operate with reflections from connectors. Often as much as 90 dB isolation is needed.
Figure 1-1. Basic components of a connector.
Over the last few years, the FC/PC style connector has emerged as the most popular connector for fiber-optic applications. While not the highest perform­ing connector, it represents a good compromise between performance, reli­ability, and cost. If properly maintained and cleaned, this connector can withstand many repeated connections.
However, many instrument specifications require tighter tolerances than most connectors, including the FC/PC style, can deliver. These instruments cannot tolerate connectors with the large non-concentricities of the fiber common with ceramic style ferrules. When tighter alignment is required, Agilent Technologies instruments typically use a connector such as the Diamond HMS-10, which has concentric tolerances within a few tenths of a micron. Agi­lent Technologies then uses a special universal adapter, which allows other cable types to mate with this precision connector. See Figure 1-2.
1-31
Setting Up the Infiniium DCA
Choosing the Right Connector
Figure 1-2. Universal adapters to Diamond HMS-10.
The HMS-10 encases the fiber within a soft nickel silver (Cu/Ni/Zn) center which is surrounded by a tough tungsten carbide casing, as shown in
Figure 1-3.
Figure 1-3. Cross-section of the Diamond HMS-10 connector.
The nickel silver allows an active centering process that permits the glass fiber to be moved to the desired position. This process first stakes the soft nickel silver to fix the fiber in a near-center location, then uses a post-active staking to shift the fiber into the desired position within 0.2 µm. This process, plus the keyed axis, allows very precise core-to-core alignments. This connector is found on most Agilent Technologies lightwave instruments.
The soft core, while allowing precise centering, is also the chief liability of the connector. The soft material is easily damaged. Care must be taken to mini­mize excessive scratching and wear. While minor wear is not a problem if the
1-32
Setting Up the Infiniium DCA

Inspecting Connectors

glass face is not affected, scratches or grit can cause the glass fiber to move out of alignment. Also, if unkeyed connectors are used, the nickel silver can be pushed onto the glass surface. Scratches, fiber movement, or glass contamina­tion will cause loss of signal and increased reflections, resulting in poor return loss.
Inspecting Connectors
Because fiber-optic connectors are susceptible to damage that is not immedi­ately obvious to the naked eye, bad measurements can be made without the user even being aware of a connector problem. Although microscopic exami­nation and return loss measurements are the best way to ensure good connec­tions, they are not always practical. An awareness of potential problems, along with good cleaning practices, can ensure that optimum connector perfor­mance is maintained. With glass-to-glass interfaces, it is clear that any degra­dation of a ferrule or the end of the fiber, any stray particles, or finger oil can have a significant effect on connector performance.
Figure 1-4 shows the end of a clean fiber-optic cable. The dark circle in the
center of the micrograph is the fibers 125 µm core and cladding which carries the light. The surrounding area is the soft nickel-silver ferrule. Figure 1-5 shows a dirty fiber end from neglect or perhaps improper cleaning. Material is smeared and ground into the end of the fiber causing light scattering and poor reflection. Not only is the precision polish lost, but this action can grind off the glass face and destroy the connector.
Figure 1-6 shows physical damage to the glass fiber end caused by either
repeated connections made without removing loose particles or using improper cleaning tools. When severe, the damage on one connector end can be transferred to another good connector that comes in contact with it.
The cure for these problems is disciplined connector care as described in the following list and in “Cleaning Connectors” on page 1-36.
Use the following guidelines to achieve the best possible performance when making measurements on a fiber-optic system:
Never use metal or sharp objects to clean a connector and never scrape the connector.
Avoid matching gel and oils.
1-33
Setting Up the Infiniium DCA
Inspecting Connectors
Figure 1-4. Clean, problem-free fiber end and ferrule.
Figure 1-5. Dirty fiber end and ferrule from poor cleaning.
Figure 1-6. Damage from improper cleaning.
While these often work well on first insertion, they are great dirt magnets. The oil or gel grabs and holds grit that is then ground into the end of the fiber. Also, some early gels were designed for use with the FC, non-contacting con-
1-34
Setting Up the Infiniium DCA
Inspecting Connectors
nectors, using small glass spheres. When used with contacting connectors, these glass balls can scratch and pit the fiber. If an index matching gel or oil must be used, apply it to a freshly cleaned connector, make the measurement, and then immediately clean it off. Never use a gel for longer-term connections and never use it to improve a damaged connector. The gel can mask the extent of damage and continued use of a damaged fiber can transfer damage to the instrument.
When inserting a fiber-optic cable into a connector, gently insert it in as straight a line as possible. Tipping and inserting at an angle can scrape material off the inside of the connector or even break the inside sleeve of connectors made with ceramic material.
When inserting a fiber-optic connector into a connector, make sure that the fi­ber end does not touch the outside of the mating connector or adapter.
Avoid over tightening connections.
Unlike common electrical connections, tighter is not better. The purpose of the connector is to bring two fiber ends together. Once they touch, tightening only causes a greater force to be applied to the delicate fibers. With connec­tors that have a convex fiber end, the end can be pushed off-axis resulting in misalignment and excessive return loss. Many measurements are actually improved by backing off the connector pressure. Also, if a piece of grit does happen to get by the cleaning procedure, the tighter connection is more likely to damage the glass. Tighten the connectors just until the two fibers touch.
Keep connectors covered when not in use.
Use fusion splices on the more permanent critical nodes. Choose the best con-
nector possible. Replace connecting cables regularly. Frequently measure the return loss of the connector to check for degradation, and clean every connec­tor, every time.
All connectors should be treated like the high-quality lens of a good camera. The weak link in instrument and system reliability is often the inappropriate use and care of the connector. Because current connectors are so easy to use, there tends to be reduced vigilance in connector care and cleaning. It takes only one missed cleaning for a piece of grit to permanently damage the glass and ruin the connector.
Measuring insertion loss and return loss
Consistent measurements with your lightwave equipment are a good indica­tion that you have good connections. Since return loss and insertion loss are key factors in determining optical connector performance they can be used to determine connector degradation. A smooth, polished fiber end should pro-
1-35
Setting Up the Infiniium DCA

Cleaning Connectors

duce a good return-loss measurement. The quality of the polish establishes the difference between the “PC” (physical contact) and the “Super PC con­nectors. Most connectors today are physical contact which make glass-to-glass connections, therefore it is critical that the area around the glass core be clean and free of scratches. Although the major area of a connector, excluding the glass, may show scratches and wear, if the glass has maintained its polished smoothness, the connector can still provide a good low level return loss con­nection.
If you test your cables and accessories for insertion loss and return loss upon receipt, and retain the measured data for comparison, you will be able to tell in the future if any degradation has occurred. Typical values are less than 0.5 dB of loss, and sometimes as little as 0.1 dB of loss with high performance con­nectors. Return loss is a measure of reflection: the less reflection the better (the larger the return loss, the smaller the reflection). The best physically contacting connectors have return losses better than 50 dB, although 30 to 40 dB is more common.
Visual inspection of fiber ends
Visual inspection of fiber ends can be helpful. Contamination or imperfections on the cable end face can be detected as well as cracks or chips in the fiber itself. Use a microscope (100X to 200X magnification) to inspect the entire end face for contamination, raised metal, or dents in the metal as well as any other imperfections. Inspect the fiber for cracks and chips. Visible imperfec­tions not touching the fiber core may not affect performance (unless the imperfections keep the fibers from contacting).
Cleaning Connectors
The procedures in this section provide the proper steps for cleaning fiber­optic cables and Agilent Technologies universal adapters. The initial cleaning, using the alcohol as a solvent, gently removes any grit and oil. If a caked-on layer of material is still present, (this can happen if the beryllium-copper sides of the ferrule retainer get scraped and deposited on the end of the fiber during insertion of the cable), a second cleaning should be performed. It is not uncommon for a cable or connector to require more than one cleaning.
CAUTION Agilent Technologies strongly recommends that index matching compounds
not be applied to their instruments and accessories. Some compounds, such as gels, may be difficult to remove and can contain damaging particulates. If you think the use of such compounds is necessary, refer to the compound manufacturer for information on application and cleaning procedures.
1-36
Setting Up the Infiniium DCA
Table 1-10. Cleaning Accessories
Item Agilent Part Number
Isopropyl alcohol 8500-5344
Cotton swabs 8520-0023
Small foam swabs 9300-1223
Compressed dust remover (non-residue) 8500-5262
Table 1-11. Dust Caps Provided with Lightwave Instruments
Item Agilent Part Number
Laser shutter cap 08145-64521
FC/PC dust cap 08154-44102
Biconic dust cap 08154-44105
DIN dust cap 5040-9364
HMS10/dust cap 5040-9361
ST dust cap 5040-9366
Cleaning Connectors
To clean a non-lensed connector
CAUTION Do not use any type of foam swab to clean optical fiber ends. Foam swabs can
leave filmy deposits on fiber ends that can degrade performance.
1 Apply pure isopropyl alcohol to a clean lint-free cotton swab or lens paper.
Cotton swabs can be used as long as no cotton fibers remain on the fiber end after cleaning.
2 Clean the ferrules and other parts of the connector while avoiding the end of
the fiber.
3 Apply isopropyl alcohol to a new clean lint-free cotton swab or lens paper.
4 Clean the fiber end with the swab or lens paper.
Do not scrub during this initial cleaning because grit can be caught in the swab and become a gouging element.
5 Immediately dry the fiber end with a clean, dry, lint-free cotton swab or lens
paper.
6 Blow across the connector end face from a distance of 6 to 8 inches using
1-37
Setting Up the Infiniium DCA
Cleaning Connectors
filtered, dry, compressed air. Aim the compressed air at a shallow angle to the fiber end face.
Nitrogen gas or compressed dust remover can also be used.
CAUTION Do not shake, tip, or invert compressed air canisters, because this releases
particles in the can into the air. Refer to instructions provided on the compressed air canister.
7 As soon as the connector is dry, connect or cover it for later use.
If the performance, after the initial cleaning, seems poor try cleaning the con­nector again. Often a second cleaning will restore proper performance. The second cleaning should be more arduous with a scrubbing action.
To clean an adapter
The fiber-optic input and output connectors on many Agilent Technologies instruments employ a universal adapter such as those shown in the following picture. These adapters allow you to connect the instrument to different types of fiber-optic cables.
Figure 1-7. Universal adapters.
1 Apply isopropyl alcohol to a clean foam swab.
Cotton swabs can be used as long as no cotton fibers remain after cleaning. The foam swabs listed in this sections introduction are small enough to fit into adapters.
Although foam swabs can leave filmy deposits, these deposits are very thin, and the risk of other contamination buildup on the inside of adapters greatly out­weighs the risk of contamination by foam swabs.
2 Clean the adapter with the foam swab.
3 Dry the inside of the adapter with a clean, dry, foam swab.
4 Blow through the adapter using filtered, dry, compressed air.
Nitrogen gas or compressed dust remover can also be used. Do not shake, tip, or invert compressed air canisters, because this releases particles in the can into the air. Refer to instructions provided on the compressed air canister.
1-38
Setting Up the Infiniium DCA

Returning the Instrument for Service

Returning the Instrument for Service
The instructions in this section show you how to properly package the instru­ment for return to a Agilent Technologies service office. For a list of offices,
refer to Agilent Technologies Service Offices” on page 1-43.
If the instrument is still under warranty or is covered by an Agilent mainte­nance contract, it will be repaired under the terms of the warranty or contract (the warranty is at the front of this manual). If the instrument is no longer under warranty or is not covered by an Agilent maintenance plan, Agilent will notify you of the cost of the repair after examining the unit.
When an instrument is returned to a Agilent service office for servicing, it must be adequately packaged and have a complete description of the failure symptoms attached.
When describing the failure, please be as specific as possible about the nature of the problem. Include copies of additional failure information (such as the instrument failure settings, data related to instrument failure, and error mes­sages) along with the original cal data disks and the instrument being returned.
Please notify the service office before returning your instrument for service. Any special arrangements for the instrument can be discussed at this time. This will help the Agilent service office repair and return your instrument as quickly as possible.

Preparing the instrument for shipping

1 Write a complete description of the failure and attach it to the instrument.
Include any specific performance details related to the problem. The following
1-39
Setting Up the Infiniium DCA
Preparing the instrument for shipping
information should be returned with the instrument.
Type of service required.
Date instrument was returned for repair.
Description of the problem:
Whether problem is constant or intermittent.
Whether instrument is temperature-sensitive.
Whether instrument is vibration-sensitive.
Instrument settings required to reproduce the problem.
Performance data.
Company name and return address.
Name and phone number of technical contact person.
Model number of returned instrument.
Full serial number of returned instrument.
List of any accessories returned with instrument.
The original cal data disks.
2 Cover all front or rear-panel connectors that were originally covered when you
first received the instrument.
CAUTION Cover electrical connectors to protect sensitive components from electrostatic
damage. Cover optical connectors to protect them from damage due to physical contact or dust.
CAUTION Instrument damage can result from using packaging materials other than the
original materials. Never use styrene pellets as packaging material. They do not adequately cushion the instrument or prevent it from shifting in the carton. They may also cause instrument damage by generating static electricity.
3 Pack the instrument in the original shipping containers. Original materials are
available through any Agilent office. Or, use the following guidelines:
Wrap the instrument in antistatic plastic to reduce the possibility of damage caused by electrostatic discharge.
For instruments weighing less than 54 kg (120 lb), use a double-walled, cor­rugated cardboard carton of 159 kg (350 lb) test strength.
The carton must be large enough to allow approximately 7 cm (3 inches) on all sides of the instrument for packing material, and strong enough to accom­modate the weight of the instrument.
Surround the equipment with approximately 7 cm (3 inches) of packing ma­terial, to protect the instrument and prevent it from moving in the carton. If packing foam is not available, the best alternative is S.D-240 Air Cap from
1-40
Setting Up the Infiniium DCA
Preparing the instrument for shipping
Sealed Air Corporation (Commerce, California 90001). Air Cap looks like a plastic sheet filled with air bubbles. Use the pink (antistatic) Air Cap to reduce static electricity. Wrapping the instrument several times in this ma­terial will protect the instrument and prevent it from moving in the carton.
4 Seal the carton with strong nylon adhesive tape.
5 Mark the carton FRAGILE, HANDLE WITH CARE”.
6 Retain copies of all shipping papers.
1-41
Setting Up the Infiniium DCA

Hard Drive Recovery

Hard Drive Recovery
The Infiniium DCA hard drive recovery system offers a way to recover from situations when the instrument hard drive configuration becomes corrupt due to inadvertent user changes to the instrument or to a virus infection.
The Infiniium DCA hard disk recovery system consists one floppy disk, one CD-ROM, and instructions. The recovery process will re-image the internal disk drive on the Agilent 86100B. After the recovery process is complete, the Infiniium DCA firmware is set to the latest revision as indicated on the recov­ery disks.
NOTE Please be advised that all user files and the user network configuration settings will be
lost during the recovery process.
To order a free hard drive recovery package, visit
http://www.agilent.com/comms/dca
1-42
Setting Up the Infiniium DCA

Agilent Technologies Service Offices

Agilent Technologies Service Offices

Call Center

For technical assistance, you can contact your local Agilent Call Center.
In the Americas, call 1 (800) 452-4844
In other regions, visit http://www.agilent.com and click Contact Us.

Service Center

Before returning an instrument for service, you must first call the Agilent Technologies Instrument Support Center.
In all regions, call (800) 403-0801
1-43
Setting Up the Infiniium DCA
Service Center
1-44
2

Working in Comfort

Working in Comfort
Working in Comfort
Working in Comfort
To optimize your comfort and productivity, it is important that you set up your work area correctly and use your Agilent product properly. With that in mind, we have developed some setup and use recommendations for you to follow based on established ergonomic principles.
Improper and prolonged use of keyboards and input devices are among those tasks that have been associated with repetitive strain injury (RSI) to soft tis­sues in the hands and arms. If you experience discomfort or pain while using your Agilent Technologies product, discontinue use immediately and consult your physician as soon as possible.
Please study the recommendations offered here in this chapter. You may also wish to consult your employers human resources department or other rele­vant departments for guidance specific to your company.

About Repetitive Strain Injury

Because your comfort and safety are our primary concern, we strongly recom­mend that you use the infiniium DCA in accordance with established ergo­nomic principles and recommendations. Scientific literature suggests that there may be a relationship between injury to soft tissues - especially in the hands and arms - and prolonged improper use of keyboards or other equip­ment requiring repeated motions of the hands and forearms. This literature also suggests that there are many other risk factor that may increase the chance of such injury, commonly called Repetitive Strain Injury.
What is RSI? Repetitive Strain Injury (RSIalso known as cumulative trauma disorder or
repetitive motion injury) is a type of injury where soft tissues in the body, such as muscles, nerves, or tendons, become irritated or inflamed. RSI has been a reported problem for those who perform repetitive tasks such as assembly line work, meatpacking, sewing, playing musical instruments, and computer work.
2-2
Working in Comfort

Mice and Other Input Devices

RSI has also been observed in those who frequently engage in activities such as carpentry, knitting, housework, gardening, tennis, windsurfing and lifting children.
What causes RSI? The specific causes of RSI have not been established. Nevertheless, the inci-
dence of RSI has been associated with a variety of risk factors, including:
Too many uninterrupted repetitions of an activity or motion.
Performing an activity in an awkward or unnatural posture.
Maintaining a static position for prolonged periods.
Failing to take frequent short breaks.
Other environmental and social factors.
In addition, there have been reports associating the occurrence of RSI with the use of keyboards, mice, and other input devices. Also, certain medical con­ditions, such as rheumatoid arthritis, obesity and diabetes, may predispose people to this type of injury.
What if I experience discomfort?
If you are experiencing any discomfort, seek professional medical advice immediately. Typically, the earlier a problem is diagnosed and treated, the eas­ier it is to resolve.
Mice and Other Input Devices
Various aspects of using mice and other input devices may increase your risk of discomfort or injury. Observing the following recommendations may reduce that risk.
Try to keep your hand, wrist, and forearm in a neutral position while using your mouse or other input device.
If you use your thumb to rotate the ball on a trackball or spaceball, keep it in a relaxed, natural shape, and maintain a neutral posture in your hand, wrist, and forearm.
Hold the mouse gently by draping your fingers over it. Keep your hand relaxed and fingers loose. Do not grip the mouse tightly.
It takes very little pressure or force from your fingers to activate the buttons or scroll wheel on your mouse, scrolling mouse, trackball, or other input device. Using too much force can place unnecessary stress on the tendons and muscles
2-3
Working in Comfort
Mice and Other Input Devices
in your hands, wrists, and forearms.
If you are using a scrolling mouse, be sure to keep your fingers and hand in a relaxed, neutral position when activating the scroll wheel. Also, this type of mouse features software that can minimize the number of mouse movements or button clicks.
When using a mouse, trackball, or other input device, position it as close to the keyboard as possible, and keep it at the same level as you do not have to stretch while using it.
Use a good quality mouse pad to enable the mouse to work most effectively and reduce unnecessary hand and wrist movements.
Be sure to keep your mouse and trackball clean. Regular removal of accumulat­ed dust and dirt helps ensure proper tracking and reduces unnecessary hand and wrist motions.
2-4
3
Introduction 3-2 Oscilloscope Mode 3-8 Eye/Mask Mode 3-7 TDR/TDT Mode 3-9 Using Plug-in Modules 3-10 Using the Touch Screen 3-12 Using the Quick Measure Feature 3-13 If the Display Goes Blank 3-14 Hiding and Moving Dialog Boxes 3-15 Menus 3-16

Using the Infiniium DCA

Using the Infiniium DCA
Introduction
Introduction
The Infiniium DCA is designed to measure a variety of high speed digital com­munication waveforms. You can operate it in any of the following three instru­ment modes:
Eye/Mask Mode (100 Mb/s to 10 Gb/s) for eye diagram analysis and standards testing (for example, SONET/SDH, Gigabit Ethernet, and Fiber Channel).
TDR/TDT Mode for time domain reflectometry measurements
Oscilloscope Mode (DC to 20 GHz or 50 GHz)
3-2
Using the Infiniium DCA
Introduction
Use the touch screen
You can explore the instruments menus and change its settings by touching the display. Notice that the measurement toolbar changes depending on the selected instrument mode. Touch the channel, time and delay, and trigger level buttons located at the bottom of the display to change the horizontal and vertical scales. Or, you can change these settings using front-panel knobs and buttons.
3-3
Using the Infiniium DCA
Introduction
Refer to the Help System
To learn how to use your Infiniium DCA and to locate information on specifica­tions (including plug-in modules), instrument features, and measurement configuration refer to Chapter 4, Using the Built-In Information System in this book.
You must provide a trigger input signal
Because high-speed oscilloscopes are not capable of triggering directly on a test signal, you must first connect an external timing reference signal to the front-panel trigger input connector. This input allows the Infiniium DCA to synchonize to the input signal. You can select free-run triggering when an external trigger is unavailable. However, you will only be able to view the waveform for amplitude information only without any timing information. This is an easy way to examine the amplitude of a signal.
3-4
Using the Infiniium DCA
Introduction
Front-panel features
The following figure shows the input connectors, controls, power switch, and other front-panel features available on the Infiniium DCA.
3-5
Using the Infiniium DCA
Introduction
Rear-panel features
The following figure shows the rear-panel input connections available on the Infiniium DCA.
3-6
Using the Infiniium DCA

Eye/Mask Mode

Eye/Mask Mode
The Eye/Mask mode allows you to per­form NRZ (Non-Return to Zero), or RZ (Return to Zero) eye diagram measure­ments and eye mask tests. The eye dia­gram is typically produced by triggering the instrument with a synchronous clock signal.
Eye/Mask mode measurements only work when an eye diagram, and not a pulse, is present on the screen. Mea­surements made on a pulse waveform while in Eye/Mask mode will fail.
Once you are in Eye/Mask mode, press the Autoscale button. This ensures that an optimum eye diagram is displayed on the graticule. An optimum eye diagram consists of a full display of the eye in addition to portions of the wave­form preceding and following the eye.
When you select the Eye/Mask mode, the measurement toolbars shown here appear. These measurement toolbars include mea­surements such as rise time, fall time, jitter, and extinction ratio, as well as the mask test start and stop controls.
Eye measurement and mask test results appear below the display graticule. A maxi­mum of four eye measurement test results are displayed and include any failures or instrument messages that may occur.
To learn more about the Eye/Mask mode and switching from NRZ to RZ measurement mode, consult the Help.
3-7
Using the Infiniium DCA

Oscilloscope Mode

Oscilloscope Mode
Use the Oscilloscope mode for pulse type waveforms that are triggered by an external signal. Oscilloscope mode measurements only work when a single valued waveform, and not an eye dia­gram, is present on the screen.
When you select the Oscilloscope mode, the measurement toolbars shown here appear. These measurement toolbars include measurements such as rise time, fall time, overshoot, and Vp-p. All measure­ments, with the exception of jitter, can be performed on any of the following sources: channels, waveform memories, and math functions. One of the above sources must be turned on in order to perform an oscillo­scope measurement.
If more than one source is turned on and you initiate a measurement, the Enter Mea­surement Info dialog box opens. This dialog box allows you to select the source you want to use for the measurement.
Measurement results appear below the dis­play graticule. A maximum of four pulse measurement test results are displayed and include any failures or instrument messages that may occur.
To learn more about the Oscilloscope mode, consult the Help.
3-8

TDR/TDT Mode

The TDR/TDT mode allows you use time domain reflectometry (TDR) to verify the proper functioning of the physical components of your device under test using a time-delayed electri­cal pulse. You can also use time domain transmission (TDT) to measure both the attenuation and propagation delay of your device under test.
TDR/TDT mode measurements when using a single-ended or differential TDR/TDT 5475x series plug-in module.
When you select the TDR/TDT mode, the measure­ment toolbar shown here appears. This measurement toolbar includes measurements such as Quick TDR, normalization, rise time, and fall time. Quick TDR allows you to view an immediate display of the charac­teristics of your device under test. You can also use this toolbar to perform tasks such as setting up the stimu­lus and destination for your measurement.
You can use markers to to verify the impedance, volt­age, and percent of reflection of a selected point on the waveform trace. TDR/TDT markers are initiated by turning on the X-axis marker(s) only. The Y-axis marker(s) will then track the waveform at the position determined by the X-axis marker. You can view the position of the marker values in the marker values tab located below the display graticule.
To learn more about the TDR/TDT mode, consult the Help.
Using the Infiniium DCA
TDR/TDT Mode
3-9
Using the Infiniium DCA

Using Plug-in Modules

Using Plug-in Modules
The Infiniium DCA holds up to two plug-in modules, which can provide up to four measurement channels. Each input channel has its own vertical controls. The amplitude scale knob adjusts the amplitude scale used for the input chan­nel. Use the amplitude offset knob to position the displayed signal. You can also set the amplitude scale and offset to specific values by touching of chan­nel buttons shown to at the bottom of the display.
To install a module
Slide the module into an available slot and finger tighten the two front-panel locking screws.
3-10
Using the Infiniium DCA
Using Plug-in Modules
You can use 8348x and 5475x series plug-in modules
In addition to a complete line of modules designed specifically for this instru­ment, modules designed for use with HP/Agilent 83480A and 54750A main­frames can also be installed into the Infiniium DCA. The 8348x and 5475x series plug-in modules are equipped with a trigger input on the individual module front panels. Repeatedly press the Infiniium DCAs front-panel Trigger Source button until the label for the desired trigger source is lit.
CAUTION Electrical channel input circuits and the trigger input circuit can be damaged
by electrostatic discharge (ESD). Avoid applying static discharges to the front­panel input connectors. Prior to connecting any coaxial cable to the connectors, momentarily short the center and outer conductors of the cable together. Avoid touching the front-panel input connectors without first touching the frame of the instrument. Be sure that the instrument is properly earth-grounded to prevent buildup of static charge. Wear a wrist-strap or heel­strap.
CAUTION Optical channel fiber-optic connectors are easily damaged when connected to
dirty or damaged cables and accessories. When you use improper cleaning and handling techniques, you risk expensive instrument repairs, damaged cables, and compromised measurements. Before you connect any fiber-optic cable to the digital communications analyzer, refer to Cleaning Connections for
Accurate Measurements on page 1-30.
3-11
Using the Infiniium DCA

Using the Touch Screen

Using the Touch Screen
There may be times when you would like to disable the touch screen feature. For example, when pointing out a measurement result without initiating a change to the instrument. You can temporarily disable the touch screen for this purpose.
A touch screen configuration utility can be used to ensure that the touch screen is both aligned and orientated with the display. It also allows you to adjust the touch screen speed to simulate a double-click and to turn the touch screen sound on or off.
To temporarily disable the touch screen
Touch the disable button which is located at the top of the display. Touch the button a second time to reenable the touch screen.
3-12
Using the Infiniium DCA

Using the Quick Measure Feature

Using the Quick Measure Feature
Press Quick Measure and the Infiniium DCA automatically performs four measurements on a displayed signal. You can configure Quick Measure to include any four measurements located on the measurement tool bar. Quick Measure has a default set of measurements for each instrument mode. For example, in Oscilloscope mode the default measurements are: rise time, fall time, period, and V amplitude. In Eye/Mask mode you can also configure Quick Measure to perform a mask test.
To configure Quick Measure
On the Measure menu, choose Quick Measure Config.
3-13
Using the Infiniium DCA

If the Display Goes Blank

If the Display Goes Blank
The display automatically turns off if the Infiniium DCA is powered over an eight hour period without being used. The purpose of this feature is to extend the life of the display. You can adjust the amount of time that must elapse before the screen is turned off by the backlight saver utility (for example, 2, 4, 6, or 8 hours). This feature is especially useful if you are using remote mea­surements or leaving the instrument unattended for long periods of time. Although not recommended, you can disable this control so that the backlight is never turned off during inactivity.
To reactivate the display
Touch the display screen
Move the mouse (if applicable)
Touch any button on the instrument front panel
Turn any knob on the instrument front panel
To configure the backlight saver
1 On the Setup menu, choose Display.
2 Touch Enabled to enable or disable the backlight control.
3-14
Using the Infiniium DCA

Hiding and Moving Dialog Boxes

Hiding and Moving Dialog Boxes
Sometimes, dialog boxes may cover portions of a displayed signal and grati­cule that you wish to see. You can either move the dialog box or make it trans­parent.
To move a dialog box
Touch the top of the dialog box, and drag your finger across of display to move the dialog box to a more convenient location.
To make a dialog box transparent
1 Select Dialog Boxes from the Setup menu.
2 Touch Opaque or Transparent.
3-15
The File menu
Using the Infiniium DCA

Menus

Menus
The instrument menus provide access to all functions. This section shows the layout of each menu. For detailed information on each of these menus, refer to the Help.
3-16
The Control menu
Using the Infiniium DCA
Menus
3-17
The Setup menu
Using the Infiniium DCA
Menus
3-18
The Measure menu
Using the Infiniium DCA
Menus
The Calibrate menu
3-19
The Utilities menu
The Help menu
Using the Infiniium DCA
Menus
3-20
4
The Infiniium DCA Help Contents 4-3 Description of the Contents Topics 4-4 Getting Help While Changing Instrument Settings 4-6 Learning About Measurements Results 4-7 Hiding the Help 4-8 Printing the Contents of a Topic 4-9 Selecting the Help Language 4-10

Using the Built-In Information System

Using the Built-In Information System
Using the Built-In Information System
Using the Built-In Information System
Where is the operating manual for your infiniium DCA? It is built into your instrument! To access the built-in information system, referred to as Help, simply touch Contents on the Help menu. This will open the contents topic that is shown in the figure on the following page. In this chapter, youll learn features that are unique to the Infiniium DCAs Help system as well as tips that will make the system more useful to you.
Exploring the Help should be familiar to you because it is similar to other Win­dows applications. You can, of course, use the touch screen or the mouse to explore the Help.
4-2
Using the Built-In Information System

The Infiniium DCA Help Contents

The Infiniium DCA Help Contents
The following figure shows the 86100B Help contents. This is the starting point for learning how to use your Infiniium DCA.
In addition to the 11 topics presented in the contents, be sure to use the index buttons located along the bottom of the topic window. These can be very help­ful in locating a topic of interest.
4-3
Using the Built-In Information System

Description of the Contents Topics

Description of the Contents Topics
The following list describes the information you can find in each of the con­tents topics.
This topic contains links to useful information you may want to read first.
Getting Started
This topic contains links to utilizing the instrument functionality.
Using the Instrument
Setup Guide
Messages and Errors
Modules and Specifications
4-4
This topic provides step-by-step procedures for many configuration and measurement tasks.
This topic contains information about the status of the instrument or an action that has been performed.
This topic contains descriptions and specifications of plug-in modules used with the Infiniium DCA.
Connector Care
Managing Files and Networks
Instrument Specifications
Using the Built-In Information System
Description of the Contents Topics
This topic provides information about optical and elec­trical connector care.
This topic provides information about file management, using your network, and printing.
This topic contains descriptions of all instrument main­frame specifications.
Glossary
Technical Support
New Features
This topic provides a list of terms and their definitions.
This topic provides technical support information.
This topic contains the links to both new and enhanced features in this version of the Infiniium DCA software.
4-5
Using the Built-In Information System

Getting Help While Changing Instrument Settings

Getting Help While Changing Instrument Settings
Help is available directly from dialog boxes in the graphical user interface.
To see overview information in an open dialog box, touch the Book icon. A top-
ic window opens and describes the dialog box and its controls.
To get information on a specific item in the dialog box, touch the What’s this?
icon. Then touch the desired control on the dialog box. A window appears with a description of the selected control, as shown in the following figure.
Book icon
What’s this? icon
4-6
Using the Built-In Information System

Learning About Measurements Results

Learning About Measurements Results
After a measurement is performed, the displays shows the Measure tab which includes the measurement results. The fastest way to get help on measure­ment configuration or algorithms is to touch the Setup & Info button. The following figure shows the location of this button.
4-7
Using the Built-In Information System

Hiding the Help

Hiding the Help
When using the Help, you can temporarily hide it so that you can see the entire display. Then, you can re-display the Help with the same topic still showing.
To hide the Help, touch the Minimize button located at the top of the Help
window.
To re-display the Help, touch the Restore button that is shown on the display.
4-8
Using the Built-In Information System

Printing the Contents of a Topic

Printing the Contents of a Topic
You can print the entire contents of a Help topic or a pop-up window. To print the entire contents of a Help topic, simply touch the Print button located at the top of the Help window.
To print the entire contents of a pop-up Help window, by following the steps below.
1 Connect a printer and a mouse pointing device to the Infiniium DCA.
2 Place the pointer over the topic, and click the right mouse button. Then click
Print Topic.
4-9
Using the Built-In Information System

Selecting the Help Language

Selecting the Help Language
Help is available in more than one language. On the Help menu, touch Select Help Language, and then touch the language you want to view. Please
ensure that the Help file is closed before changing the language selection. You can only view one version of help at a time.
Help Version
Refer to the English version of help to view the most current information about instru­ment features and specifications.
4-10
5

Regulatory Information

Regulatory Information

Compliance with German Noise Requirements

Compliance with German Noise Requirements
This is to declare that this instrument is in conformance with the German Reg­ulation on Noise Declaration for Machines (Laermangabe nach der Maschinen­laermrerordnung -3.GSGV Deutschland).
Notice for Germany: Noise Declaration
Acoustic Noise Emission Geraeuschemission
LpA < 70 dB LpA < 70 dB
Operator position am Arbeitsplatz
Normal position normaler Betrieb
per ISO 7779 nach DIN 45635 t.19

Compliance with Canadian EMC Requirements

This ISM device complies with Canadian ICES-001.
Cet appareil ISM est conforme a la norme NMB du Canada.
5-2
Regulatory Information

Declaration of Conformity - Mainframe

Declaration of Conformity - Mainframe
5-3
Regulatory Information

Declaration of Conformity - Plug-in Modules

Declaration of Conformity - Plug-in Modules
5-4

Index

Numerics
12 GHz trigger bandwidth, 1–6 5475x series plug-in modules, 3–11 8348x series plug-in modules, 3–11
A
accessories
connecting, 1–11
adapters
APC 3.5 (f-f), 1–7
SMA (f-f), 1–7 Add Printer Wizard, 1–18 Agilent
Web site, 1–29 Agilent offices, 1–43 algorithms, 4–7 amplitude offset, 3–10 amplitude scale, 3–10 Applications menu, 3–18 Autoscale button, 3–7
B
blank screen, 3–14 Book icon, 4–6 built-in information system, 4–2
temporarily hiding, 4–8
C
calibration
module, 1–26 calibration certificate, 1–6 care of fiber optics, 1–24, 3–11 CE mark, 1–4 cleaning, 1–3
adapters, 1–38
fiber-optic connections, 1–30, 1–37
non-lensed connectors, 1–37 clock signal, 3–7 compressed dust remover, 1–37 connecting
accessories, 1–11
printer, 1–12 connector
care, 1–30 Contents topic, 4–2
contents topic, 4–3 cotton swabs, 1–37 CSA mark, 1–4
D
dialog box
making transparent, 3–15
moving, 3–15 dialog boxes, 4–6 dimensions, 1–10 dust caps, 1–37
E
electrical channel input circuits, 1–26 electrostatic discharge, 1–26 Eye/Mask Mode, 3–2, 3–7
F
fiber optics
care of, 1–24, 3–11
cleaning connections, 1–30
connectors, covering, 1–40 fiber-optic adapters, 1–8 File menu, 3–16 file sharing, 1–23 foam swabs, 1–37 front panel
power switch, 1–24
H
hard drive recovery, 1–42 help, 4–2 help language, selecting, 4–10 help topic, printing, 4–9
I
input
connector, 1–30 instrument
checking the shipment, 1–5
positioning, 1–11
powering up, 1–24
returning for service, 1–39
Index-1
Index
instrument modes, 3–2 ISM1-A, 1–4 isopropyl alcohol, 1–37
K
keyboard, 1–6, 1–11
L
LAN, 1–23 language version, 4–10 line cords, 1–13 line power
requirements, 1–15
M
Measure menu, 3–17 measurement channels, 3–10 measurement results, 4–7 measurement toolbar, 3–3 measurement toolbars, 3–7, 3–8, 3–9 menu
bar, 3–16 Minimize button, 4–8 module calibration, 1–26 module, installing, 3–10 modules, installing, 1–9 mouse, 1–6, 1–11
N
network printers, 1–23 network setup, 1–23 noise declaration, 5–2
O
online help, 4–2 Options menu, 3–19 Oscilloscope Mode, 3–2, 3–8
P
package contents, 1–5 packaging for shipment, 1–40 plug-in modules, 3–10
positioning the instrument, 1–11 power
requirements, 1–15
switch, 1–24 power switch, 1–15 Print Setup, 1–18 printer
connecting, 1–12 printer driver, 1–18, 1–20 procedures, step-by-step, 4–4
Q
Quick Measure button, 3–13
R
rack mount, 1–6 recovery
hard drive, 1–42 Restore button, 4–8
S
Safety Class 1 Product, 1–3 sales and service offices, 1–43 service, 1–39
sales and service offices, 1–43 Setup & Info button, 4–7 shipment, checking package content, 1–5 shipping procedure, 1–39 supported printers, 1–18, 1–20 swabs, 1–37 synchronous clock signal, 3–7
T
TDR/TDT Mode, 3–2, 3–9 time and date, 1–17 time-domain reflectometry, 3–2 timing reference signal, 3–4 touch calibration, 1–22 touch screen, 1–16, 3–3
configuration, 1–22
disabling, 3–12 Touch Screen Config, 1–22 trigger input signal, 3–4 Trigger Source button, 3–11
Index-2
U
Utilities menu, 1–22
W
warm-up time, 1–26 Web site, Agilent, 1–29 weight, 1–10 What’s this? icon, 4–6 wire bail, positioning, 1–11
Index
Index-3
Index
Index-4
Loading...