Agilent Technologies 1141A
Differential Probe and 1142A
Probe Control and Power
Module
Agilent Technologies 1141A Differential Probe and
1142A Probe Control and Power Module
This manual contains information for use and service of the differential
probe system, the 1141A Differential Probe and 1142A Probe Control
and Power Module. In this document, the two models will be treated as
a system.
Each of the two instrument models that make up the differential probe
system has a serial number sticker. The sticker for the 1141A
Differential Probe is inside the probe, in the bottom cover. (See
chapter 3 for disassembly procedure.)
The 1141A/1142A probe system allows measurement of small
differential signals in the presence of much larger common-mode
signals. It has the following major features:
• 200 MHz bandwidth
• Variable offset
• dc reject
• ac coupling
• Remote operating capability
The variable offset capability can be used to measure small ac signals in
the presence of much larger dc levels. Remote operation of key
features allows the use of the probe system in automatic test situations.
The probe can be used with an oscilloscope, spectrum analyzer, or any
instrument where differential probing is required and a compatible
50 Ω input is available.
2
Contents
1 Operating the Probe
Accessories Supplied 6
Accessories Available 8
To inspect the probe 9
Using the probe with other instruments 9
Recommended Test Equipment 9
2 Calibration Tests and Adjustment
Equipment Required 26
The Test Board 26
Calibration Tests27
dc Gain Accuracy 28
Bandwidth 34
CMRR Test 36
Calibration Test Record 39
Differential Probe 60
Control and Power Module 60
Attenuator Adapters 61
Test Board 62
Service Policy63
Troubleshooting64
Probe Troubleshooting 64
Probe Control and Power Module Troubleshooting 64
Removing and Replacing Assemblies66
Differential Probe 66
Probe Adapters 68
Probe Control and Power Module 70
Replaceable Parts71
Parts List 71
Ordering Information 71
Direct Mail Order System 71
Manufacturers’ Codes 72
Exploded View 73
4
1
Operating the Probe
5
Operating the Probe
Accessories Supplied
Introduction
This chapter shows you how to connect and operate the 1141A
Differential Probe and 1142A Probe Control and Power Module as a
differential probe system.
Accessories Supplied
The following items are supplied as part of the 1141A/1142A probe
system. Item numbers refer to the numbers in Figure 1-1 on page -7
and Figure 1-2 on page -8. Those without item numbers are supplied
but not shown in figures. See the Replaceable Parts List for parts not
listed below.
ItemDescriptionQtyPart Number
1Differential Probe1
210x Attenuator Adapter15063-2144
3100x Attenuator15063-2145
4ac Coupling Adapter15063-2146
5Two-inch Extension Leads (package 5)15959-9334
6Mini Grabbers21400-1422
7Five-inch Ground Lead15061-6162
8Shielded Signal Lead101141-68702
9Test Board101141-66504
10Flat-blade Alignment Tool18710-1961
11Circuit Connection Posts (strip of 20)11251-5943
Probe Control and Power Module11142A
Power Cord1see parts list
Carrying case
User and Service Manual1
6
Figure 1-1
Operating the Probe
Accessories Supplied
1141A Differential Probe and Accessories
7
Figure 1-2
Figure 1-3
Operating the Probe
Accessories Available
1141A Miscellaneous Accessories
1142A Probe Control and Power Module
Accessories Available
The following accessories can be ordered.
• 5959-9335 Long Extension Lead (5.5 inch/14 cm), package of 5
• 5090-4833 Mini grabber for SMT, package of 20
8
Operating the Probe
To inspect the probe
To inspect the probe
Inspect the shipping container for damage. If the shipping container or cushioning
material is damaged, it should be kept until the contents of the shipment have been
checked for completeness and the instrument had been checked mechanically and
electrically. Accessories supplied with the instrument are listed in see “Accessories
Supplied” on page 6 of this manual.
If the contents are incomplete, if there is mechanical damage or defect, or if the
instrument does not pass calibration tests, notify the nearest Agilent Technologies office.
If the shipping container is damaged, or the cushioning materials show sign of stress,
notify the carrier as well as the nearest Agilent Technologies office. Keep the shipping
materials for the carrier’s inspection. The office will arrange for repair or replacement
at Agilent Technologies’ option without waiting for a claim settlement.
Using the probe with other instruments
The 1141A/1142A probe system can be used with other instruments as well as
oscilloscopes. You can use it with a spectrum analyzer or frequency counter, or any
instrument with an input that can be terminated with 50 Ω.
If you are going to use the probe system with an instrument other than an oscilloscope,
you may need to set up the probe with an oscilloscope first. This will allow you to select
coupling and reject modes, and set offset, so the output of the probe is compatible with
signal requirements of the other instrument.
Recommended Test Equipment
The following table is a list of the test equipment required to test calibration, make
adjustments, and troubleshoot this instrument. The table indicates the critical
specifications of the test equipment and for which procedure the equipment is necessary.
Equipment other than the recommended model may be used if it satisfies the critical
specifications listed in the table.
The following paragraphs cover system preparation and initial adjustments.
Power Requirements
The 1141A/1142A probe system (specifically the 1142A) requires a power source of
either 90 to 132/198 to 264 Vac, 47 to 440 Hz, 25 VA maximum.
CAUTIONBefore connecting power to this instrument, be sure the line voltage switch on the rear
panel of the instrument is set properly.
Line Voltage Selection
Before applying power, verify the setting of the LINE SELECT switch on the rear panel
of the 1142A. The slide switch can be set to either 115 or 230 V.
10
Operating the Probe
Recommended Test Equipment
WARNINGBefore connecting this instrument, the protective earth terminal of the instrument must
be connected to the protective conductor of the (Mains) power cord. The Mains plug
must be inserted in a socket outlet provided with a protective each contact. The
protective action must be negated by the use on an extension cord (power cable)
without a protective conductor (grounding). Grounding one conductor of a twoconductor outlet does not provide an instrument ground.
This instrument is provided with a three-wire power cable. When connected to an
appropriate ac power outlet, this cable grounds the instrument cabinet. The type of
power cable plug shipped with the instrument depends on the country of destination.
The 1142A Power Control and Power Module does not have a power switch. A power
switch is not required because of the low mains power requirement.
Figure 1-4
1142A Rear Panel
Procedure
Use the power cord to connect the 1142A to the ac mains.
1
2 Connect the 1141A probe cable power connector to the PROBE connector on
the rear panel of the 1142A power module.
3 Connect the output of the probe to the input of the oscilloscope.
4 Set the input impedance of the oscilloscope to 50 Ω.
If the oscilloscope does not have a selectable 50 Ω input impedance, connect a 50 Ω BNC
feedthrough termination between the probe output and the input of the oscilloscope.
5 If making an initial equipment setup, continue with the initial adjustment in
the following section.
Initial Adjustment
For a given combination of 1141A Differential Probe and 1142A Probe Control and Power
Module, you may want to adjust the Offset Null and DC Reject Gain. Typically, you need
to make these adjustments only once, before the probe is first used. You can make them
any time to optimize the system. These adjustments do not affect the specifications of
the probe system.
11
Figure 1-5
Operating the Probe
Recommended Test Equipment
• Offset null zeroes the dc level at the output of the probe. The range of adjustment is
about ±4 mV.
• DC Reject Gain adjusts the gain of the dc reject circuit to accurately null the dc
component of an input signal. The range of adjustment is about ±0.5%.
1142A Front Panel
Equipment Needed
The following equipment is necessary for initial adjustment.
• 5 V power supply
• DVM that can measure 25 µV
•50 Ω BNC feedthrough terminator
• 01141-66504 test board
Equipment Setup
Use the following procedure to setup the differential probe system for initial adjustment.
CAUTIONDo not exceed ±7 V when using the test board for this procedure. If the voltage is too
high, it will cause excessive power dissipation in the 50 Ω termination on the test board.
1 Use the probe setup procedure to set up the probe system.
2 Connect a 50 Ω BNC feedthrough terminator to the output of the probe.
3 Disconnect all accessories from the input of the probe.
4 Connect the DVM to measure the dc output of the probe at the 50 Ω load.
12
Figure 1-6
Operating the Probe
Recommended Test Equipment
5 Set up the 1142A:
a Set the Local/Remote push button to Local.
b Under DC Couple, press Zero offset.
6 Set the power supply output to 5 V.
7 Arrange a connection between the power supply and the test board. The
negative terminal of the supply should connect to the shield of the test board
BNC.
If your power supply has standard binding posts, you can connect a banana-to-BNC
adapter to the supply and connect a BNC cable between the supply and the test board.
Adjustment
Warm up the 1141A for 30 minutes before making adjustments.
1
With the 1141A probe inputs unconnected, adjust Offset Null on the 1142A for
a minimum reading on the DVM.
The voltage swing of the adjustment is approximately ±4 mV.
On the 1142A, under DC Reject, press 5.0 Hz.
2
3 Read and record the reading on the DVM, _________ mV.
4 Connect the probe to the test board in the position shown below
(signal to + input).
Signal to + input
5 After the DVM reading stabilizes, adjust DC Reject Gain to the reading recorded
in step 3.
With a 5 V supply, the voltage swing is approximately ±12.5 mV. With a lower supply, the
voltage swings proportionally less.
Using the Accessories
The 1141A Differential Probe and accessories are designed to provide a variety of ways
to connect to circuitry and make measurements. In the descriptions, any method used
to connect to the probe signal inputs also applied to the adapters. The figure below
shows, in a general way, the use of accessories.
13
Figure 1-7
Operating the Probe
Recommended Test Equipment
Basic Accessory Connections
Probe Tips
Probe tips fit into the receptacles in the probe and are held in place with probe tip caps.
If necessary, you can solder the probe tips into a circuit or wires can be soldered to the
tips. If you solder to the probe tips, be careful not to melt the plastic probe tip caps.
NoteBecause of the close tolerances between the probe tip caps and probe tips, it will be
difficult to separate the probe tips and caps once the probe tips have been soldered.
Ground Leads
The circular end of the ground lead fits over the screw on the top side of the probe.
Extension Leads
The extension leads provide a flexible connection between circuitry and the probe.
• To provide a male connection to other circuitry, connect the extension lead over the
probe tips.
• To provide a female connection, remove the probe tip caps and probe tips and connect
the extension leads to the probe.
14
Operating the Probe
Recommended Test Equipment
NoteUse extension leads and similar connection accessories carefully. Extension leads
compromise the high-frequency specifications of the probe. CMRR is particularly
sensitive to unbalanced input parameters.
To prevent pickup of stray fields when you use extension lead, either the ones supplied
with the 1141A or others, dress them carefully as follows:
• Connect the leads at right angles to the circuitry under test.
• Keep the leads as parallel as possible before they connect to the probe.
Mini-Grabbers
Mini-grabbers can be attached to the probe or adapter through the extension leads.
1
Remove the probe tip caps and tips.
2 Attach the extension leads to the probe or adapter.
3 Attach the mini-grabbers to the extension leads.
Circuit Connector Posts
These 0.025-inch square posts can be used to connect either directly to the probe or to
the extension leads.
Solder the posts directly into your circuitry or use them to make extension leads that
plug into the inputs of the probe or adapters.
Shielded Signal Leads
The shielded signal leads allow connection to points in a circuit that are up to 10 inches
apart. The leads are shielded to within approximately 1/2 inch of the end of the lead so
they minimize pick-up due to stray fields from adjacent circuitry.
1
Connect the end with the ground connector to the probe pins and ground of
the differential probe or adapter.
2 Connect the free ends of the leads to 0.025-inch square or 0.030-inch round pins
in your circuitry or to the mini-grabbers.
NoteEach lead has an input capacitance of approximately 15 pF. This capacitance may limit
the bandwidth of your measurement (depending on the impedance of the circuit). Also,
CMRR may be affected because of slight differences between the input capacitance of
the two leads. CMRR is also affected by differences in impedance between the two
measurements points.
15
Figure 1-8
Operating the Probe
Recommended Test Equipment
Adapters
There are three adapters for use with the differential probe. Two adapters are
attenuators, a 10x and a100x. The other is an ac adapter for blocking dc from the probe
input.
The adapters are installed on the probe after the probe tip caps and probe tips have been
removed. The adapter fastens to the probe using a thumb wheel located on the underside
of the adapter. the figure below shows a good way to hold the probe while attaching the
adapter.
1
Remove the probe tip caps and probe tips, from the probe.
2 Fit the adapter over the end of the probe and rotate the thumb wheel with your
finger until the adapter fits snugly.
A snug fit is important because the ground is maintained through the thumb wheel screw.
A loosely attached adapter compromises the mechanical and electrical integrity of the
combination.
Attaching the Adapters
16
Figure 1-9
Operating the Probe
Recommended Test Equipment
Adapter Combinations
The figure below shows the allowed adapter and probe connections. There are two
specific combinations that should not be used.
• Do not attach the ac adapter between an attenuator adapter and the probe.
An attenuator adapter must be terminated by the input resistance of the probe. The
ac adapter isolates the probe input resistance.
• Do not cascade two attenuator adapters.
The attenuator adapters are designed to be terminated by the 1 MΩ resistance of the
probe. The input resistance of the attenuator adapter is 9 MΩ for the 10x adapter and
10 MΩ for the 100x adapter.
Allowed Adapter Connections
17
Figure 1-10
Operating the Probe
Recommended Test Equipment
Connector Compatibility
The following are general connector characteristics for the probe, adapters, and
accessories.
• The female connectors on the probe, adapters, and other accessories are designed to
mate with 0.030-inch round or 0.0250-inch square pins.
• The probe, adapter, and extension lead pins are 0.030-inch round.
• The strip of circuit connection posts provided as an accessory has 0.025-inch square
pins.
• The mini-grabber has a 0.25-inch square pin.
• The ground connection at the end of the probe and adapters (where the adapters
fasten) accepts an M3 metric screw.
Test Board
The primary use of the test board is to apply test and calibration signals to the input of
the probe or adapters. Specific use of the test board is covered wherever it applies.
Grounding
Grounding is very important when probing circuitry. Improper grounding can increase
the common mode signal level. This reduces the effectiveness of the differential probe.
The mechanical connections at the input of the probe are ground for probe signals. The
screw where the ground lead attaches (see figure 1-7) fastens to this ground. Also, the
attenuator and ac adapter fasten to this ground through the screw connection and the
ground is carried through each adapter to its front.
Probe Grounds
Coupling Functions
There are three methods for blocking or compensating for the dc component of a signal.
Each has specific advantages.
18
Probe System Coupling Functions
Operating the Probe
Recommended Test Equipment
dc offsetdc Rejectac Coupling
Adapter
dc Blocked
Probe alone
Probe with 10x
Probe with 100x
Set-up needed
± 20 V
± 200 V
± 500 V
Adjust offset to put
signal on screen
Remote Control?Yes
CMRR
NoNoYes
1
± 20 V
± 200 V
± 500 V
Select DC Reject
low-frequency
corner
1
Yes
± 20 V
± 500 V
± 500 V
Attach ac Coupling
adapter to
differential probe
No
degradation?
Low-frequency
NoYesYes
degradation?
1
Isolated external dc reference and control signals are needed
dc Reject
dc Reject is the best method of eliminating the dc component of a signal when dc is not
a factor in the measurement.
The key characteristics are:
• The low-frequency component (from dc to the selected corner frequency) is
automatically nulled by the dc reject circuitry.
• Probe CMRR specifications are not compromised as happens when the ac coupling
adapter is uses.
• There is a selectable low frequency corner with -3dB points at 0.05 Hz, 0.5 Hz,
or 5.0 Hz.
• The voltage reject range is ±20 V with the probe alone, ±200 V with the 10x attenuator,
and ±500 V with the 100x attenuator.
19
Operating the Probe
Recommended Test Equipment
To use dc reject:
Remove the ac adapter if it is installed.
1
2 On the front panel of the 1142A, press Local.
3 Under DC Reject on the front panel, press 5.0 Hz or 0.5 Hz individually, or
5.0 Hz and 0.5 Hz simultaneously to get 0.05 Hz.
Within the frequency and voltage characteristics noted elsewhere in this manual, low
frequencies are nulled from the input signal.
Offset
Offset is the best method to use when the low-frequency corners associated with dc
reject and the ac adapter interfere with the measurement.
The key characteristics are:
• The user manually null the dc component with the offset adjustment.
• Offset is dc coupled so there is no low frequency roll-off.
• Probe CMRR specifications are not compromised as happens when the ac coupling
adapter is used.
• The voltage offset range is ±20 V with the probe alone, ±200 V with the 10x attenuator,
and ±500 V with the 100x attenuator. (With the 100x attenuator, the offset range is
restricted by the maximum input voltage rating rather than the operating range of
the offset).
To use offset:
1
Remove the ac adapter if it is installed.
2 On the front panel of the 1142A, press Local and Variable offset.
3 Adjust the Coarse and Fine Variable Offset until the signal is displayed on the
screen of the oscilloscope.
ac Adapter
The ac adapter must be used when the dc component of the signal exceeds the operating
range of the dc reject or offset methods. The ac adapter block the dc and low frequency
component of the input by forming a high pass filter with the input impedance of the
probe or adapter.
The key characteristics are:
• The ac adapter safely blocks ±200 Vdc when attached directly to the probe or
±500 Vdc when attached to a 10x or 100x adapter.
• The probe and adapters have different input impedances, so they have different low
frequency corners with the ac adapter. When the ac adapter is directly on the probe
the -3dB corner is 15 Hz. When the ac adapter is on an attenuator the corner is 1.5 Hz.
• The low-frequency CMRR when using the ac adapter is not as good as when using the
probe alone or the probe with a 10x or 100x adapter.
CAUTIONIf you measure a node having a high dc potential, the blocking capacitors in the ac
adapter will charge to that potential. After making such measurements, discharge the
capacitors by grounding both inputs of the ac adapter. This will prevent damage by a
high voltage discharge into sensitive circuitry when the next measurement is made.
20
Operating the Probe
Recommended Test Equipment
To use ac coupling:
Attach the ac coupling adapter to the input of the probe or the input of the
1
attenuator adapter.
2 On the 1142A, press Local and Zero offset.
Remote operation
For automatic test applications, the coupling and offset functions provided by the 1142A
Probe Control and Power Module can be remotely controlled through a connector on the
rear panel of the module. The connection is through a standard 9-pin female
D-subminiature connector. This style is the same as that used on some personal
computer monitor cables, which provides an economical way to connect the 1142A to
the controller interface on an automatic test system.
The following table gives the connections.
Remote Input Connections
PinFunctionPinFunctionConnector
1Function Select 1 (A1R)6Function Select 0 (A0R)
2Digital common7N.C.
3N.C.8N.C.
4External offset common9External offset
5Shield
NOTETo minimize dc offset errors and potential noise coupling, electrically isolate all
connections between the Remote Input connector and the controlling system.
21
Operating the Probe
Recommended Test Equipment
Function Select
The easiest way to control the function select lines is contact closures between the lines
and Digital common (pin 2) of the remote input connector. (TTL compatible control
signals can be used; but to avoid problems with ground loops, they must be electrically
isolated.) The following truth table shows the functions provided by the function select
lines. For the Remote Inputs, “0” represents a closure and “1” represents an open circuit.
Remote Functions Select Truth Table
Remote InputFunction
A1R
(Pin 1)
000.05 Hz Reject
010.5 Hz Reject
105 Hz Reject
11DC Couple
Variable Offset
The remote variable offset can be used when the dc couple function is remotely selected.
The offset voltage must be referenced to the External offset common (pin 4) of the
remote input connector. It must be electrically isolated from the controlling system. The
following table shows the offset range and remote offset requirements for probe and
adapter combinations.
Remote Offset Input Requirements
Probe alone± 20 V± 10 V
Probe with 10x adapter± 200 V± 10 V
Probe with 100x adapter± 500 V± 2.5 V
A0R
(Pin 6)
Offset RangeRemote
Requirements
Cleaning Requirements
If the instrument requires cleaning: (1) Remove power from the instrument. (2) Clean
the external surfaces of the instrument with a soft cloth dampened with a mixture of
mild detergent and water. (3) Make sure that the instrument is completely dry before
reconnecting it to a power source.
22
Operating the Probe
Recommended Test Equipment
Differential Amplifiers and CMRR
The 1141A Differential Probe is a high-impedance differential amplifier. A characteristic
of differential amplifiers is the ability to reject signals that are common to the two inputs.
The common mode rejection ratio (CMRR) is the measurement of this ability. It is
expressed as the ratio between the amplitudes of the common mode and differential
signals which product equal outputs. For example, if a common mode signal of 1 V and
differential signal of 1 mV both produce outputs of 1 mV, the CMRR is 1000:1.
The ability to reject common mode signals is dependent on the balance designed into
the differential amplifier. At higher frequencies it becomes harder to balance circuit
parasitics and parameters of devices so CMRR degrades as frequency increases. Also,
stray coupling increases with frequency and coupling may vary between the two
differential paths.
The CMRR of the 1141A Differential Probe is specified at the input of the probe and
cannot be affected expect by adjustments in the probe. However, the way the probe is
connected into the circuitry being tested can have a big influence in the overall result of
the measurement, especially at high frequencies.
The following things can affect the effective CMRR of a test setup:
• The connection to the circuit under test.
The method used to connect the probe is important because it involves the symmetry
of the differential input circuitry. For example, using different lengths of wire to
connect the circuit to the two probe inputs unbalances the inductance and
capacitance at the inputs. The effective CMRR will be reduced, especially at high
frequencies. Additionally, coupling from adjacent circuitry will be less balanced.
• The impedance of the source.
This is another instance where the symmetry of the differential circuit is important.
The impedance of the source forms a network with the input impedance of the source
forms a network with the input impedance of the connections and the probe. This
network determines the frequency response for the measurement. If each side of the
differential source has a different impedance, the frequency response of each side
will be different and the unbalance is reflected in a reduced CMRR. Of course, lower
source impedances have less effect on the frequency response of the measurement.
• The ground connection.
A poorly located ground connection allows ground loops to add to the common mode
signal.
• Frequency.
Frequency is the most important factor in CMRR only because all of the factors
mentioned above are frequency dependent. The unbalances of capacitance and
inductance are more important as frequency increases. Therefore, good highfrequency practice is important when using a high impedance differential probe.
On the other hand, if the differential probe is ac coupled to the circuit under test (the
ac adapter is being used) the CMRR will be degraded below a certain frequency; the
lower the frequency the worse the CMRR. This is because unbalance in the series
capacitances of the ac coupler becomes more significant the lower the frequency.
23
Operating the Probe
Recommended Test Equipment
24
2
Calibration Tests and Adjustment
25
Calibration Tests and Adjustment
Equipment Required
Introduction
This chapter is divided into two sections. The first section gives
calibration tests and the second adjustment procedures for the 1141A
Differential Probe and 1142A Probe Control and Power Module.
Equipment Required
A complete list of equipment required for the calibration tests and adjustments is listed
in “Recommended Test Equipment” on page 9. Equipment required for individual
procedures is listed at the procedure. Any equipment satisfying the critical specifications
listed may be substituted for the recommended model.
The Test Board
The test board is a supplied accessory for use during calibration tests and adjustments
to connect signals to the differential probe (with or without adapters). A BNC connector
connects the test board to a cable from the signal generator. The board includes a 50 Ω
termination (two 100 Ω resistors).
CAUTIONThe power rating of the 50 Ω termination is 1.0 W. Keep the signal input below 7 Vdc
or rms to avoid degrading the termination.
Once the probe tip caps and probe tips have been removed, the probe can be connected
to the test board in one of three ways, as shown in the figure below. At each position of
the probe, probe inputs are connected to a different combination of signal and ground.
A separate terminal on the test board connects the ground of the probe to the signal
ground.
Figure 2-1
Test Board Showing Probe Positions
26
Calibration Tests and Adjustment
The Test Board
Calibration Tests
These procedures test the probe’s electrical performance using applicable specifications
given in “Performance Specifications and Characteristics” on page 55 as performance
standards. Specifications applicable to individual tests are noted at the test for reference.
Testing Interval
The calibration testing procedures may be performed for incoming inspection of the
instrument and should be performed periodically thereafter to ensure and maintain peak
performance. The recommended test interval is yearly or every 2,000 hours of operation.
Amount of use, environmental conditions, and the user’s experience concerning need
for testing will contribute to verification requirements.
Calibration Test Record
The results of the calibration tests may be tabulated in the Calibration Test Record
provided at the end of the calibration tests. The Calibration Test Record listed the
calibration tests and provides an area to mark test results. The results recorded in the
Calibration Test Record during initial inspection may be used for later comparisons of
the tests during periodic maintenance, troubleshooting, and after repairs or adjustments.
Calibration Test Procedures
Procedures may be done individually or in any order.
NOTEAllow the instrument to warm up for at least 30 minutes prior to beginning calibration
tests.
27
Loading...
+ 63 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.