Agilent 8991A Data Sheet

NAVAIR 17-20GP-48
GP-48
THIS PUBLICATION SUPERSEDES NAVAIR 17-20GP-48
TECHNICAL MANUAL
INSTRUMENT CALIBRATION PROCEDURE
PEAK POWER ANALYZER
HEWLETT-PACKARD
8991AOPT003
DATED 1 MARCH 1998
DISTRIBUTION STATEMENT C. DISTRIBUTION AUTHORIZED TO U.S. GOVERNMENT AGENCIES AND THEIR CONTRACTORS TO PROTECT PUBLICATIONS REQUIRED FOR OFFICIAL USE OR FOR ADMINISTRATIVE OR OPERATIONAL PURPOSES ONLY. OTHER REQUESTS FOR THIS DOCUMENT SHALL BE REFERRED TO NAVY MEASUREMENT SCIENCE DIRECTORATE, NAVAL WARFARE ASSESSMENT STATION, P.O. BOX 5000, CORONA, CA 92878-5000.
DESTRUCTION NOTICE - FOR UNCLASSIFIED, LIMITED DOCUMENTS, DESTROY BY ANY METHOD THAT WILL PREVENT DISCLOSURE OF CONTENTS OR “RECONSTRUCTION OF THE DOCUMENT”.
1 MAY 2000
NAVAIR 17-20GP-48
LIST OF EFFECTIVE PAGES
The purpose of this list is to identify the pages in this document affected by any technical content changes made since the previous release of the document.
NOTE: On a revised page, with the exception of the Title, the A, and the i pages, the technical changes are indicated by a vertical line in the outer margin of the page.
Page
No. ** Revision*
Title 0 A 0 i 0 1 to 14 0
Page
No. ** Revision*
Page
No. ** Revision*
*Zero in this column indicates an original page.
“R” in this column indicates a revised page.
** All pages not listed are blank pages.
A
NAVAIR 17-20GP-48
TABLE OF CONTENTS
Section Page
1 Introduction and Description.............................................................................................................. 1
2 Equipment Requirements................................................................................................................... 2
3 Preliminary Operations...................................................................................................................... 3
4 Calibration Process ............................................................................................................................ 5
4.1 Vertical Accuracy Tests.................................................................................................................. 5
4.2 Horizontal Accuracy Tests.............................................................................................................. 6
4.3 Rise Time Tests (Channels 2 and 3)................................................................................................ 7
4.4 Rise Time Tests (Channels 1 and 4)................................................................................................ 8
4.5 Trigger Sensitivity Tests (Channels 1 and 4)................................................................................... 9
4.6 Instrumentation Uncertainty Tests................................................................................................... 10
4.7 Check Source Tests........................................................................................................................ 10
Navy Calibration Checklist................................................................................................................ 11
TABLES
Table Page
1 Calibration Description...................................................................................................................... 1
2 Equipment Requirements................................................................................................................... 2
i
NAVAIR 17-20GP-48
SECTION 1
INTRODUCTION AND DESCRIPTION
1.1 This procedure describes the calibration of the Hewlett-Packard 8991AOPT003 Peak Power Analyzer. The instrument
being calibrated is referred to herein as the TI (Test Instrument).
NOTE
This procedure does not calibrate the Hewlett-Packard 84812A or 84815A peak power sensors commonly used with the TI. To calibrate these peak power sensors, use NAVAIR 17-20GD-24.
1.2 This procedure was derived from a draft prepared by Naval Air Warfare Center, Aircraft Division (542800A), Patuxent
River, MD. All comments concerning this procedure should be directed to Measurement Science Directorate, Naval Warfare Assessment Station, P.O. Box 5000, Corona, CA 92878-5000.
1.3 This procedure includes tests of essential performance parameters only. Any malfunction noticed during calibration,
whether specifically tested for or not, should be corrected.
Table 1. Calibration Description
TI
Characteristics
Type-N connector
dimension
In accordance with MIL-C-39012
composite: 0.187 to 0.207 in.
Performance
Specifications
for female connector
Vertical accuracy
(Ch. 2 and 3)
Gain tolerance: +1.5% of display range Offset tolerance: +(2% of offset value + 20% of V/div setting)
Horizontal accuracy Timebase range: 2 ns to 5 s/div.
Tolerance: +(0.005% iv + 2% of sec/div setting + 1 ns)
Rise time
Tolerance: <5 ns Tested by applying a fast rise
(Ch. 2 and 3)
Rise time
(Ch. 1 and 4)
Tolerance
<10 ns at high bandwidth; <1 µs at low bandwidth; <100 µs at CW; 84815A tolerance at High BW: <45 ns
Test
Method
Measured with a coaxial gage.
Tested by having the TI
measure DC voltage from an oscilloscope calibrator. Vertical accuracy = gain accuracy when offset = 0 and (gain accuracy + offset Accuracy) otherwise.
The TI measures known
horizontal markers from an oscilloscope calibrator.
pulse from an oscilloscope calibrator.
For the high-bandwidth
(<10 ns) test, a Pulse-modulated signal is Characterized with a crystal Detector and oscilloscope and then applied to TI. For all other tests, the signal itself has a rise time fast enough to be used as a standard.
Trigger sensitivity
(Ch. 1 and 4)
Minimum level:
-15 dBm high bandwidth;
-25 dBm low bandwidth
Verified with a signal
generator.
1
NAVAIR 17-20GP-48
Characteristics
Instrumentation
uncertainty (Ch. 1 and 4)
TI
Tolerance: Verified with an automated
Performance
Specifications
 
0.07
 
dB at high bandwidth;
 
0.07
 
dB at low bandwidth, CW
IEEE-488 (HP-IB)
data bus
Conforms to IEEE-STD 488-1978
for talker and listener
Check source Frequency: 1.05 Hz
Output level: 10 dBm Tolerance: +0.5 dB
SECTION 2
Test
Method
procedure.
1
1
 
++±26dBminsignal
 
++±33dBminsignal
Verified indirectly during the
instrumentation uncertainty tests.
Verified with the TI after its
instrumentation uncertainty has been verified.
EQUIPMENT REQUIREMENTS
NOTE
Minimum use specifications are the principal parameters required for performance of the calibration, and are included to assist in the selection of alternate equipment, which may be used at the discretion of the using laboratory. Satisfactory performance of alternate items shall be verified prior to use. All applicable equipment must bear evidence of current calibration.
The instruments utilized in this procedure were selected from those known to be available at Navy calibration facilities, and the listing by make or model number carries no implication of preference, recommendation, or approval for use by other agencies. It is recognized that equivalent equipment produced by other manufacturers may be capable of equally satisfactory performance in this procedure.
Table 2. Equipment Requirements
Item Minimum Use
Specifications
Calibration Equipment
2.1 Coaxial gage Uncertainty: +0.001 in. Maury Microwave A007A
2.2 Special BNC cable Delay: 5 ns +10%
Hewlett-Packard 10503A
Length: Approx. 1.5 meters
2.3 Peak power sensors
2.3.1 Peak power
Compatible with TI Hewlett-Packard 84812A
sensor #1
2.3.2 Peak power
Compatible with TI Hewlett-Packard 84815A
sensor #2
2
Table 2. Equipment Requirements
NAVAIR 17-20GP-48
Item Minimum Use
Specifications
2.4 Oscilloscope Bandwidth: >200 MHz Tektronix TDS620A
2.5 Oscilloscope calibrator
2.6 Feedthrough termination
2.7 Signal generator Frequency range: 1 to 5 Hz
2.8 Crystal detector Frequency range: 5 GHz Hewlett-Packard
2.9 Calibration software No substitute NAVAIR
2.10 Instrument controller No substitute Fluke 1722AAP
2.11 IEEE bus cable Meets IEEE STD 488-1978 and
Voltage: 0 to 20 V dc Uncertainty: +0.375% iv Marker spacing: 10 ns to 10 ms Uncertainty: +0.01% iv Fast Edge mode rise time: <2.5 ns
Impedance: 50
Power range: 0 to -25 dBm Pulse modulation capability with rise time <30 ns
FCC regulation 79-555-14686
Tektronix CG5011 with
015-0611-00 Programmable Pulse Head
Tektronix 011-0049-00
Hewlett-Packard
83630AOPT001, 008, H53
8470BOPT012
17-20AXG-11-056D, CHG: 00
Local supply
Calibration Equipment
2.12 Calibration cable No substitute NAWCAD Patuxent River
6209
2.13 Digital multimeter Used in Section 4.6 Hewlett-Packard 3458A
2.14 Cables, adapters, tees, etc.
Throughout this procedure, TI “hardkeys” are those on the front panel, while “softkeys” are those blank buttons on the side of the TI display for which the TI display will indicate the “softkey” name according to the current menu selection.
3.1 Ensure that all power switches are set to off, and set all auxiliary equipment controls as necessary to avoid damage to the
equipment and so that dangerous voltages will not be present on output terminals when the power switches are turned on.
3.2 Verify that the N-type connector of the TI sensor check source conforms to the MIL-C-39012 composite specification
(0.187” to 0.207”) as follows:
As required Local supply
SECTION 3
PRELIMINARY OPERATIONS
NOTE
3.2.1 If the coaxial gage has been standardized with a MIL-C-39012 master gage block, verify that the gage dial indicates
between -0.020 and 0.000 inches.
3
Loading...
+ 11 hidden pages