Agilent 8991A Data Sheet

Page 1
NAVAIR 17-20GP-48
GP-48
THIS PUBLICATION SUPERSEDES NAVAIR 17-20GP-48
TECHNICAL MANUAL
INSTRUMENT CALIBRATION PROCEDURE
PEAK POWER ANALYZER
HEWLETT-PACKARD
8991AOPT003
DATED 1 MARCH 1998
DISTRIBUTION STATEMENT C. DISTRIBUTION AUTHORIZED TO U.S. GOVERNMENT AGENCIES AND THEIR CONTRACTORS TO PROTECT PUBLICATIONS REQUIRED FOR OFFICIAL USE OR FOR ADMINISTRATIVE OR OPERATIONAL PURPOSES ONLY. OTHER REQUESTS FOR THIS DOCUMENT SHALL BE REFERRED TO NAVY MEASUREMENT SCIENCE DIRECTORATE, NAVAL WARFARE ASSESSMENT STATION, P.O. BOX 5000, CORONA, CA 92878-5000.
DESTRUCTION NOTICE - FOR UNCLASSIFIED, LIMITED DOCUMENTS, DESTROY BY ANY METHOD THAT WILL PREVENT DISCLOSURE OF CONTENTS OR “RECONSTRUCTION OF THE DOCUMENT”.
1 MAY 2000
Page 2
NAVAIR 17-20GP-48
LIST OF EFFECTIVE PAGES
The purpose of this list is to identify the pages in this document affected by any technical content changes made since the previous release of the document.
NOTE: On a revised page, with the exception of the Title, the A, and the i pages, the technical changes are indicated by a vertical line in the outer margin of the page.
Page
No. ** Revision*
Title 0 A 0 i 0 1 to 14 0
Page
No. ** Revision*
Page
No. ** Revision*
*Zero in this column indicates an original page.
“R” in this column indicates a revised page.
** All pages not listed are blank pages.
A
Page 3
NAVAIR 17-20GP-48
TABLE OF CONTENTS
Section Page
1 Introduction and Description.............................................................................................................. 1
2 Equipment Requirements................................................................................................................... 2
3 Preliminary Operations...................................................................................................................... 3
4 Calibration Process ............................................................................................................................ 5
4.1 Vertical Accuracy Tests.................................................................................................................. 5
4.2 Horizontal Accuracy Tests.............................................................................................................. 6
4.3 Rise Time Tests (Channels 2 and 3)................................................................................................ 7
4.4 Rise Time Tests (Channels 1 and 4)................................................................................................ 8
4.5 Trigger Sensitivity Tests (Channels 1 and 4)................................................................................... 9
4.6 Instrumentation Uncertainty Tests................................................................................................... 10
4.7 Check Source Tests........................................................................................................................ 10
Navy Calibration Checklist................................................................................................................ 11
TABLES
Table Page
1 Calibration Description...................................................................................................................... 1
2 Equipment Requirements................................................................................................................... 2
i
Page 4
NAVAIR 17-20GP-48
SECTION 1
INTRODUCTION AND DESCRIPTION
1.1 This procedure describes the calibration of the Hewlett-Packard 8991AOPT003 Peak Power Analyzer. The instrument
being calibrated is referred to herein as the TI (Test Instrument).
NOTE
This procedure does not calibrate the Hewlett-Packard 84812A or 84815A peak power sensors commonly used with the TI. To calibrate these peak power sensors, use NAVAIR 17-20GD-24.
1.2 This procedure was derived from a draft prepared by Naval Air Warfare Center, Aircraft Division (542800A), Patuxent
River, MD. All comments concerning this procedure should be directed to Measurement Science Directorate, Naval Warfare Assessment Station, P.O. Box 5000, Corona, CA 92878-5000.
1.3 This procedure includes tests of essential performance parameters only. Any malfunction noticed during calibration,
whether specifically tested for or not, should be corrected.
Table 1. Calibration Description
TI
Characteristics
Type-N connector
dimension
In accordance with MIL-C-39012
composite: 0.187 to 0.207 in.
Performance
Specifications
for female connector
Vertical accuracy
(Ch. 2 and 3)
Gain tolerance: +1.5% of display range Offset tolerance: +(2% of offset value + 20% of V/div setting)
Horizontal accuracy Timebase range: 2 ns to 5 s/div.
Tolerance: +(0.005% iv + 2% of sec/div setting + 1 ns)
Rise time
Tolerance: <5 ns Tested by applying a fast rise
(Ch. 2 and 3)
Rise time
(Ch. 1 and 4)
Tolerance
<10 ns at high bandwidth; <1 µs at low bandwidth; <100 µs at CW; 84815A tolerance at High BW: <45 ns
Test
Method
Measured with a coaxial gage.
Tested by having the TI
measure DC voltage from an oscilloscope calibrator. Vertical accuracy = gain accuracy when offset = 0 and (gain accuracy + offset Accuracy) otherwise.
The TI measures known
horizontal markers from an oscilloscope calibrator.
pulse from an oscilloscope calibrator.
For the high-bandwidth
(<10 ns) test, a Pulse-modulated signal is Characterized with a crystal Detector and oscilloscope and then applied to TI. For all other tests, the signal itself has a rise time fast enough to be used as a standard.
Trigger sensitivity
(Ch. 1 and 4)
Minimum level:
-15 dBm high bandwidth;
-25 dBm low bandwidth
Verified with a signal
generator.
1
Page 5
NAVAIR 17-20GP-48
Characteristics
Instrumentation
uncertainty (Ch. 1 and 4)
TI
Tolerance: Verified with an automated
Performance
Specifications
 
0.07
 
dB at high bandwidth;
 
0.07
 
dB at low bandwidth, CW
IEEE-488 (HP-IB)
data bus
Conforms to IEEE-STD 488-1978
for talker and listener
Check source Frequency: 1.05 Hz
Output level: 10 dBm Tolerance: +0.5 dB
SECTION 2
Test
Method
procedure.
1
1
 
++±26dBminsignal
 
++±33dBminsignal
Verified indirectly during the
instrumentation uncertainty tests.
Verified with the TI after its
instrumentation uncertainty has been verified.
EQUIPMENT REQUIREMENTS
NOTE
Minimum use specifications are the principal parameters required for performance of the calibration, and are included to assist in the selection of alternate equipment, which may be used at the discretion of the using laboratory. Satisfactory performance of alternate items shall be verified prior to use. All applicable equipment must bear evidence of current calibration.
The instruments utilized in this procedure were selected from those known to be available at Navy calibration facilities, and the listing by make or model number carries no implication of preference, recommendation, or approval for use by other agencies. It is recognized that equivalent equipment produced by other manufacturers may be capable of equally satisfactory performance in this procedure.
Table 2. Equipment Requirements
Item Minimum Use
Specifications
Calibration Equipment
2.1 Coaxial gage Uncertainty: +0.001 in. Maury Microwave A007A
2.2 Special BNC cable Delay: 5 ns +10%
Hewlett-Packard 10503A
Length: Approx. 1.5 meters
2.3 Peak power sensors
2.3.1 Peak power
Compatible with TI Hewlett-Packard 84812A
sensor #1
2.3.2 Peak power
Compatible with TI Hewlett-Packard 84815A
sensor #2
2
Page 6
Table 2. Equipment Requirements
NAVAIR 17-20GP-48
Item Minimum Use
Specifications
2.4 Oscilloscope Bandwidth: >200 MHz Tektronix TDS620A
2.5 Oscilloscope calibrator
2.6 Feedthrough termination
2.7 Signal generator Frequency range: 1 to 5 Hz
2.8 Crystal detector Frequency range: 5 GHz Hewlett-Packard
2.9 Calibration software No substitute NAVAIR
2.10 Instrument controller No substitute Fluke 1722AAP
2.11 IEEE bus cable Meets IEEE STD 488-1978 and
Voltage: 0 to 20 V dc Uncertainty: +0.375% iv Marker spacing: 10 ns to 10 ms Uncertainty: +0.01% iv Fast Edge mode rise time: <2.5 ns
Impedance: 50
Power range: 0 to -25 dBm Pulse modulation capability with rise time <30 ns
FCC regulation 79-555-14686
Tektronix CG5011 with
015-0611-00 Programmable Pulse Head
Tektronix 011-0049-00
Hewlett-Packard
83630AOPT001, 008, H53
8470BOPT012
17-20AXG-11-056D, CHG: 00
Local supply
Calibration Equipment
2.12 Calibration cable No substitute NAWCAD Patuxent River
6209
2.13 Digital multimeter Used in Section 4.6 Hewlett-Packard 3458A
2.14 Cables, adapters, tees, etc.
Throughout this procedure, TI “hardkeys” are those on the front panel, while “softkeys” are those blank buttons on the side of the TI display for which the TI display will indicate the “softkey” name according to the current menu selection.
3.1 Ensure that all power switches are set to off, and set all auxiliary equipment controls as necessary to avoid damage to the
equipment and so that dangerous voltages will not be present on output terminals when the power switches are turned on.
3.2 Verify that the N-type connector of the TI sensor check source conforms to the MIL-C-39012 composite specification
(0.187” to 0.207”) as follows:
As required Local supply
SECTION 3
PRELIMINARY OPERATIONS
NOTE
3.2.1 If the coaxial gage has been standardized with a MIL-C-39012 master gage block, verify that the gage dial indicates
between -0.020 and 0.000 inches.
3
Page 7
NAVAIR 17-20GP-48
3.2.2 If the coaxial gage has been standardized with a MIL-C-71B master gage block, verify that the gage dial indicates
between -0.010 and +0.010 inches.
3.3 Connect the TI and auxiliary equipment to appropriate power sources. Turn the auxiliary equipment power switches to
on. Then, while holding any TI front panel key down, turn the TI power on. (This ensures that all TI parameters are reset.)
3.4 Allow sufficient time for the TI and auxiliary equipment to warm up (the TI requires one hour warm up time).
3.5 Press the TI UTIL hardkey, and then the ‘more’ (if necessary) and ‘selftest menu’ softkeys. Disconnect everything from the TI inputs, if necessary, and then press the ‘test all’ softkey and ensure that the TI eventually displays “PASSED” for all 11 self tests.
3.6 Press the TI ‘exit menu’ and ‘instr cal menu’ softkeys. Set the CALIBRATOR switch on the TI rear panel to UNPROTECTED, removing any label covering the switch if necessary.
3.7 Press the TI ‘cal select’ softkey as necessary to select VERTICAL CAL. Press the TI ‘channel’ softkey as necessary to select ALL. Press the TI ‘start cal’ softkey and follow the instructions on the screen, using BNC cables and a BNC tee connector as necessary.
3.8 A full calibration lasts about 10 minutes. Verify that when the calibration is completed, the TI ‘status:’ at the top of the display indicates “passed cal”.
3.9 Press the TI ‘cal select’ softkey to select ‘delay cal’. Press the TI ‘channel’ softkey as necessary to select channel 1. Press the TI ‘start cal’ softkey and follow the instructions on the screen, making connections with the special BNC cable (item
2.2) and peak power sensor #1 (item 2.3.1 ) as necessary.
3.10 After completion of the self tests, verify that the TI ‘status:’ at the top of the display indicates “passed cal”.
3.11 Repeat steps 3.9 and 3.10 for TI channels 2, 3, and 4, except press the TI ‘channel’ softkey as necessary to select the appropriate channel.
3.12 Disconnect the special BNC cable and peak power sensor #1 from the TI. Press the TI ‘exit menu’ softkey and set the TI CALIBRATOR switch (rear panel) to PROTECTED. Push the TI DISPLAY hardkey.
3.13 Set each of the the TI softkey settings to the following:
DISPLAY to avg # of avg to 16 off/frame/axes/grid to grid connect dots to on
3.14 Press the TI DEFINE MEAS hardkey and then press the TI ‘continuous’ softkey as necessary to select ‘on’.
3.15 Press the oscilloscope SETUP, Recall Factory Setup (on the bottom menu), and OK Confirm Factory Init keys. When the oscilloscope is ready, press the VERTICAL MENU key and set the impedance to 50Ω.
4
Page 8
NAVAIR 17-20GP-48
.1
1
100
88 to 112
.2
"
200
176 to 224
.5
"
500
440 to 560
"2"
940 to 1060
"3"
1.44 to 1.56 V
SECTION 4
CALIBRATION PROCESS
NOTE
Unless otherwise specified, verify the results of each test and take corrective action whenever the test requirement is not met, before proceeding.
4.1 VERTICAL ACCURACY TESTS
4.1.1 Set the oscilloscope calibrator to output +DC voltage. Connect the oscilloscope calibrator pulse head output to the TI CH 2 input.
4.1.2 Press the TI CHAN/VERT hardkey, press the TI CHANNEL softkey as necessary to select CHANNEL 2, and then set CHANNEL 2 to ‘on’. Ensure that the TI ‘offset’ is set to 0 V.
4.1.3 Press the TI Blue, AVG and 2 hardkeys to display the channel 2 average value. Set the oscilloscope calibrator output to ON.
4.1.4 Set the oscilloscope calibrator to output each of the following positive dc voltages/division and the corresponding number of divisions. For each value, set the TI ‘scale’ values as indicated. Wait until the TI displays a flat line and the numerical value has settled. Then, verify that the displayed TI value is within the tolerance limits listed.
NOTES
The TI ‘scale’ value is just above the TI ‘offset’ value on the display.
Pressing the TI CLEAR DISPLAY hardkey reduces settling time.
Oscilloscope
Calibrator
(V/D)
4.1.5 Set the oscilloscope calibrator to output each of the following positive dc voltages/division and the corresponding number of divisions. For each value, set the TI ‘offset’ and ‘scale’ values as indicated. Wait until the TI displays a flat line and the numerical value has settled, and then verify that the displayed TI value is within the tolerance limits listed.
Oscilloscope
Calibrator
(DIV)
TI
Scale
(mV/div.)
TI
Tolerance Limits
(mV dc)
5
Page 9
NAVAIR 17-20GP-48
110.7
100
0.954 to 1.046
""1.0
"
0.948 to 1.052
""1.3
"
0.942 to 1.058
5"4.7
"
4.87 to 5.13
""5.0
"
4.87 to 5.13
""5.3
200
4.86 to 5.14
10"9.4
"
9.75 to 10.25
""10.0
"
9.74 to 10.26
""10.6
"
9.72 to 10.28
20"18.5
500
19.5 to 20.5
""20.0
"
19.4 to 20.6
Oscilloscope
Calibrator
(V/D)
4.1.6 Press the TI CHAN/VERT hardkey, and then set CHANNEL 2 to ‘off’. Press the TI Blue and CLR MEAS hardkeys.
4.1.7 Set the oscilloscope calibrator OUTPUT to off.
4.1.8 Connect the oscilloscope calibrator pulse head output to the TI CH 3 input. Press the TI CHAN/VERT hardkey, press the TI CHANNEL softkey as necessary to select CHANNEL 3, and then set CHANNEL 3 to ‘on’. Ensure that the TI ‘offset’ is set to 0 V.
Oscilloscope
Calibrator
(DIV)
TI
Offset
(V)
TI
Scale
(mV/div.)
TI
Tolerance Limits
(mV dc)
4.1.9 Press the TI Blue, AVG and 3 hardkeys to display the channel 3 average value. Set the oscilloscope calibrator output to ON.
4.1.10 Repeat steps 4.1.4 through 4.1.7 for TI channel 3.
4.1.11 Disconnect the equipment from the TI.
4.2 HORIZONTAL ACCURACY TESTS
4.2.1 Connect the oscilloscope calibrator pulse head output to the TI CH 2 input. Press the oscilloscope calibrator MARKERS and OUTPUT ON buttons to output horizontal markers.
4.2.2 Set the oscilloscope calibrator to 10 ns/div and press the TI AUTOSCALE hardkey. Adjust the TI ‘scale’ setting (CHAN/VERT hardkey menu) as necessary to display the entire signal on the screen.
4.2.3 Press the TI Blue, PRI (1/PRF) and 2 hardkeys to display the channel 2 period.
4.2.4 Set the oscilloscope calibrator to output each of the following marker values. For each value, press the TI TIMEBASE hardkey and set the TI TIMEBASE setting to the given value. Adjust the TI ‘delay’ setting if necessary so that the TI displays only two markers in the middle of the screen. Wait until the numerical value has settled and then verify that the TI value is within the tolerance limits listed.
6
Page 10
NAVAIR 17-20GP-48
10 ns
2 ns
8.9600 to 11.0400 ns
.1 µs
20 ns
98.5950 to 101.4050 ns
1 µs
200 ns
994.9500 to 1005.0500 ns
10 µs
2 µs
9.9585 to 10.0415 µs
.1 ms
20 µs
99.5940 to 100.4060 µs
1 ms
200 µs
995.9490 to 1004.0510 µs
10 ms
2 ms
9.9595 to 10.0405 ms
Oscilloscope Calibrator
Marker Setting
(Per Division)
4.2.5 Set the oscilloscope calibrator OUTPUT to off.
4.2.6 Connect the oscilloscope calibrator pulse head output to the TI CH 3 input. Press the oscilloscope calibrator OUTPUT ON button to output horizontal markers.
4.2.7 Set the oscilloscope calibrator to 10 ns/div and press the TI AUTOSCALE hardkey. Adjust the TI ‘scale’ setting (CHAN/VERT hardkey menu) as necessary to display the entire signal on the screen.
4.2.8 Press the TI Blue, PRI (1/PRF) and 3 hardkeys to display the channel 3 period.
4.2.9 Repeat steps 4.2.4 and 4.2.5 for TI channel 3.
Timebase Setting
TI
TI
Tolerance Limits
4.2.10 Set the oscilloscope calibrator output to off and then disconnect the pulse head from the TI.
4.3 RISE TIME TESTS (CHANNELS 2 AND 3)
4.3.1 Connect the oscilloscope calibrator pulse head output through a feedthrough termination to the TI CH 2 input. Press the oscilloscope calibrator FAST EDGE and OUTPUT ON buttons and set it for positive polarity at a frequency of 100 kHz.
4.3.2 Press the TI AUTOSCALE hardkey, followed by the Blue, RISETIME and 2 hardkeys to measure the channel 2 rise time.
4.3.3 Press the TI TIMEBASE hardkey, and set the TI TIMEBASE setting to 2 ns/div.
4.3.4 Ensure that a single rising-edge pulse can be seen on the TI display. Press the TI ‘delay’ softkey and then adjust the ‘delay’ setting so that the center of the rising edge crosses the third vertical graticule from the left.
4.3.5 Allow the TI to settle for 30 seconds or until the < sign with the rise time value disappears. Verify that the rise time value is <5 ns.
4.3.6 Set the oscilloscope calibrator OUTPUT to off.
NOTE
In some instances, the < sign may not disappear, indicating that the rise time is too short for the timebase setting. If the < has not disappeared after 30 seconds, ensure that the most stable displayed value is <5 ns.
4.3.7 Connect the oscilloscope calibrator pulse head through the feedthrough termination to the TI CH 3 input. Press the oscilloscope calibrator OUTPUT ON.
7
Page 11
NAVAIR 17-20GP-48
4.3.8 Press the TI AUTOSCALE hardkey, followed by the Blue, RISETIME and 3 hardkeys to measure the channel 3 rise time.
4.3.9 Repeat steps 4.3.3 through 4.3.6 for TI channel 3.
4.3.10 Set the oscilloscope calibrator output to off and then disconnect the pulse head from the TI.
4.4 RISE TIME TESTS (CHANNELS 1 AND 4)
4.4.1 Set the signal generator controls for a 0 dBm output at 5 GHz. Set the signal generator controls for internal pulse modulation, setting the internal pulse rate to 1 kHz and pulse width to 500 µs.
4.4.2 Connect one end of the crystal detector to the signal generator RF output and the other end of the detector through a BNC cable to the oscilloscope CH 1 input. Ensure that the signal generator output is set to on. Press the oscilloscope AUTOSET key.
4.4.3 Press the oscilloscope SHIFT and ACQUIRE MENU keys, select ‘Mode’ and ‘Average’ and then set the average to 16. Press the oscilloscope TRIGGER MENU key and set the edge slope to negative.
4.4.4 Press the oscilloscope CH 1, MEASURE, and ‘Select Measrment’ keys; then press ‘more’ (as necessary) and select FALL TIME.
NOTE
The crystal detector inverts the signal. Thus, the rise time of the signal will be measured as fall time on the oscilloscope.
4.4.5 Adjust the oscilloscope vertical scale and position as necessary to display full square waves on the screen. Ensure that the oscilloscope “C1 Fall” measurement does not display the words “Low Signal Amplitude.”
4.4.6 The oscilloscope will still display the words “Low Resolution.” Adjust the oscilloscope horizontal timebase to 50 ns/div and ensure that “Low Resolution” disappears from the oscilloscope screen. Note the “C1 Fall” value on the oscilloscope. Round this value to the nearest 0.5 ns and record it as value T (in ns).
4.4.7 Calculate and record the maximum TI rise time from the formula:
MAXIMUM TI RISE TIME (ns) = 1 + T2 + 100
where, ‘T’ (in ns) is the value recorded in step 4.4.6
4.4.8 Disconnect the crystal detector and BNC cable from between the oscilloscope and signal generator.
4.4.9 Connect peak power sensor #1 to the TI CH 1 input and to the signal generator RF output. Press the TI AUTOSCALE hardkey.
4.4.10 Press the TI CARRIER FREQ hardkey, and then set the TI CARRIER FREQ softkey selection to CH 1=CH 4. Set the TI frequency to 5 GHz via the keypad.
4.4.11 Press the TI CHAN/VERT hardkey. Adjust the ‘ref. Level’ softkey selection until the peak of the TI waveform is between the top two horizontal lines. Press the TI Blue, RISETIME and 1 hardkeys to measure the channel 1 rise time.
4.4.12 Push the TI TIMEBASE hardkey, and then decrease the TI TIMEBASE setting until the < sign with the rise time value disappears. Press the TI ‘delay’ softkey and then adjust the ‘delay’ setting as necessary to keep the leading edge of the pulse centered on the display.
8
Page 12
NAVAIR 17-20GP-48
4.4.13 Adjust the TI ‘delay’ setting to put the center of the rising edge on the third vertical graticule from the left. Wait for the measurement to settle. Verify that the rise time measurement value displayed on the TI screen is less than or equal to the maximum TI rise time calculated in step 4.4.7.
4.4.14 Press the TI CHAN/VERT hardkey and then press the ‘bandwidth’ softkey as necessary to set the bandwidth to ‘low’. Press the TI TIMEBASE hardkey, and then set the TI TIMEBASE setting to 500 ns/div. Verify that the TI rise time measurement value displayed is <1000 ns.
4.4.15 Press the TI CHAN/VERT hardkey and then press the ‘bandwidth’ softkey as necessary to set the bandwidth to CW. Press the TI TIMEBASE hardkey, and then set the TI TIMEBASE setting to 50 µs/div. Verify that the TI rise time measurement value displayed is <100 µs.
4.4.16 Connect peak power sensor #1 to the TI CH 4 input and to the signal generator RF output. Press the TI AUTOSCALE hardkey.
4.4.17 Press the TI CHAN/VERT hardkey. Adjust the REF LEVEL softkey selection until the peak of the TI waveform is between the top two horizontal lines. Press the TI Blue, RISETIME and 4 hardkeys to measure the channel 4 rise time.
4.4.18 Repeat steps 4.4.12 through 4.4.15 for TI channel 4, using peak power sensor #1.
4.4.19 Connect peak power sensor #2 to the TI CH 1 input and to the signal generator RF output. Press the TI AUTOSCALE hardkey.
4.4.20 Press the TI CHAN/VERT hardkey. Adjust the REF LEVEL softkey selection until the peak of the TI waveform is between the top two horizontal lines. Press the TI Blue, RISE TIME and 1 hardkeys to measure the channel 1 rise time.
4.4.21 Repeat steps 4.4.12 and 4.4.13 for TI channel 1, using peak power sensor #2, except verify that the rise time is <45 ns, and then repeat steps 4.4.14 and 4.4.15.
4.4.22 Disconnect the peak power sensors (#1 and #2) from the TI and the signal generator, as applicable.
4.5 TRIGGER SENSITIVITY TESTS (CHANNELS 1 AND 4)
4.5.1 Set the signal generator controls for a -14 dBm output at 1 GHz. Set the signal generator controls for internal pulse modulation, setting the internal pulse rate to 1 kHz and the width to 500 µsec.
4.5.2 Connect peak power sensor #1 to the TI CH 1 input and to the signal generator RF output.
4.5.3 Press the TI AUTOSCALE hardkey, and then press the CARRIER FREQ hardkey. Set the TI frequency to 1 GHz via the keypad.
4.5.4 Press the TI Blue, TOP and 1 hardkeys to measure the channel 1 top value. Slowly adjust the signal generator output level (if necessary) until the TI indicates <-15 dBm, allowing time for settling. Verify that the waveform can still be seen on the TI.
4.5.5 Press the TI CHAN/VERT hardkey and set the bandwidth to LOW.
4.5.6 Set the signal generator controls for a -24 dBm output. Adjust the TI trigger to about -27 dBm and adjust the reference level until the display is visible and the peak is near the top of the screen.
4.5.7 Slowly adjust the signal generator output level until the TI indicates <-25 dBm, allowing time for settling. Verify that the waveform can still be seen on the TI.
4.5.8 Connect peak power sensor #1 to the TI CH 4 input and to the signal generator RF output. Set the signal generator output level to -14 dBm.
9
Page 13
NAVAIR 17-20GP-48
4.5.9 Press the TI AUTOSCALE hardkey, and then press the CARRIER FREQ hardkey. Set the TI frequency to 1 GHz via the keypad.
4.5.10 Press the TI Blue, TOP and 4 hardkeys to measure the channel 4 top value. Slowly adjust the signal generator output level (if necessary) until the TI indicates <-15 dBm, allowing time for settling. Verify that the waveform can still be seen on the TI.
4.5.11 Press the TI CHAN/VERT hardkey and set the bandwidth to LOW.
4.5.12 Set the signal generator controls for a -24 dBm output. Adjust the TI trigger to about -27 dBm and adjust the reference level until the display is visible and the peak is near the top of the screen.
4.5.13 Slowly adjust the signal generator output level until the TI indicates <-25 dBm, allowing time for settling. Verify that the waveform can still be seen on the TI.
4.5.14 Disconnect the equipment from the TI.
4.6 INSTRUMENTATION UNCERTAINTY TESTS
4.6.1 Insert the program disk into the instrument controller. Set the instrument controller as appropriate to run the disk.
4.6.2 Use the instructions provided with the calibration software to connect the TI to the instrument controller with an IEEE bus cable.
4.6.3 Follow the instructions in the calibration software to perform the indicated tests, using the calibration cable and the digital multimeter. Verify that all software tests pass.
4.6.4 Disconnect all equipment from the TI.
4.7 CHECK SOURCE TESTS
4.7.1 Connect the TI SENSOR CHECK input to the TI CH 1 input, using peak power sensor #1.
4.7.2 Press the TI UTIL hardkey and ‘more’ softkey, and set the check source to ‘pulse’. Press the TI CARRIER FREQ hardkey and enter a frequency of 1.05 GHz.
4.7.3 Press the TI AUTOSCALE hardkey and ensure that a square wave appears on the TI display. Press the TI Blue, TOP and 1 hardkeys to measure the channel 1 top value. Verify that the top value is between 9.5 and 10.5 dBm.
4.7.4 Press the TI UTIL hardkey and set the check source to CW. Press the TI AUTOSCALE hardkey, followed by the TI Blue, AVG and 1 hardkeys to measure the channel 1 average value. Verify that the TI displayed average value is between 9.5 and 10.5 dBm.
4.7.5 Unless other measurements are to be performed, turn all power switches to off or standby and disconnect the equipment from the TI.
10
Page 14
NAVAIR 17-20GP-48
CALIBRATION CHECKLIST TEST INST (S) Hewlett-Packard 8991AOPT003 Peak Power Analyzer
PROC. NO. NA 17-20GP-48 MFR. MODEL SER. NO.
PROCEDURE MEASURED VALUES OUT
STEP FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN OF CALIBRATION TOLERANCES
NO. TOL
(1) (2) (3) (4) (5) (6) (7)
3.2 (Connector Dimensions)
3.2.1 Check Source: C-39012 - - ck( ) -0.020" to 0.000"
3.2.2 Check Source: C-71B - - ck( ) -0.010" to +0.010"
3.8 Self test - - ck( ) ‘passed cal’
3.10 CH 1 self test - - ck( ) ‘passed cal’
3.11 CH 2 self test - - ck( ) ‘passed cal’ " CH 3 self test - - ck( ) ‘passed cal’ " CH 4 self test - - ck( ) ‘passed cal’
4.1 Vertical Accuracy Tests
(Channel 2)
4.1.4 100 mV/div scale 100 mV 88 to 112 mV " 200 " " 200 mV 176 to 224 mV " 500 " " 500 mV 440 to 560 mV " " " " 1000 mV 940 to 1060 mV " " " " 1.5 V 1.44 to 1.56 V
4.1.5 1 V @ Offset: 0.7 V 1.0 V 0.954 to 1.046 V " 1 V @ Offset: 1.0 V 1.0 V 0.948 to 1.052 V " 1 V @ Offset: 1.3 V 1.0 V 0.942 to 1.058 V " 5 V @ Offset: 4.7 V 5.0 V 4.87 to 5.13 V " 5 V @ Offset: 5.0 V 5.0 V 4.87 to 5.13 V " 5 V @ Offset: 5.3 V 5.0 V 4.86 to 5.14 V " 10 V @ Offset: 9.4 V 10.0 V 9.75 to 10.25 V " 10 V @ Offset: 10.0 V 10.0 V 9.74 to 10.26 V " 10 V @ Offset: 10.6 V 10.0 V 9.72 to 10.28 V " 20 V @ Offset: 18.5 V 20.0 V 19.5 to 20.5 V " 20 V @ Offset: 20.0 V 20.0 V 19.4 to 20.6 V
Page 1 of 4
11
Page 15
NAVAIR 17-20GP-48
CALIBRATION CHECKLIST TEST INST (S) Hewlett-Packard 8991AOPT003 Peak Power Analyzer
PROC. NO. NA 17-20GP-48 MFR. MODEL SER. NO.
PROCEDURE MEASURED VALUES OUT
STEP FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN OF CALIBRATION TOLERANCES
NO. TOL
(1) (2) (3) (4) (5) (6) (7)
(Channel 3)
4.1.10 100 mV/div scale 100 mV 88 to 112 mV " 200 " " 200 mV 176 to 224 mV " 500 " " 500 mV 440 to 560 mV " " " " 1000 mV 940 to 1060 mV " " " " 1.5 V 1.44 to 1.56 V " 1 V @ Offset: 0.7V 1.0 V 0.954 to 1.046 V " 1 V @ Offset: 1.0V 1.0 V 0.948 to 1.052 V " 1 V @ Offset: 1.3V 1.0 V 0.942 to 1.058 V " 5 V @ Offset: 4.7V 5.0 V 4.87 to 5.13 V " 5 V @ Offset: 5.0V 5.0 V 4.87 to 5.13 V " 5 V @ Offset: 5.3V 5.0 V 4.86 to 5.14 V " 10 V @ Offset: 9.4V 10.0 V 9.75 to 10.25 V " 10 V @ Offset: 10.0 V 10.0 V 9.74 to 10.26 V " 10 V @ Offset: 10.6 V 10.0 V 9.72 to 10.28 V " 20 V @ Offset: 18.5 V 20.0 V 19.5 to 20.5 V " 20 V @ Offset: 20.0 V 20.0 V 19.4 to 20.6 V
4.2 Horizontal Accuracy Tests
(Channel 2) (ns) (ns)
4.2.4 10 ns mkrs: 2 ns/div 10.0 8.9600 to 11.0400 " " " " "
.1 µs mkrs: 20 ns/div
1 µs mkrs: 20 ns/div
10 µs mkrs: 2 µs/div
.1 ms mkrs: 20 µs/div
1 ms mkrs: 20 µs/div
100.0 98.5950 to 101.4050
1000.0 994.9500 to 1005.0500
10.0 9.9585 to 10.0415
100.0 99.5940 to 100.4060
1000.0 995.9490 to 1004.0510
" 10 ms mkrs: 2 ms/div 10.0 9.9595 to 10.0405
Page 2 of 4
12
Page 16
NAVAIR 17-20GP-48
CALIBRATION CHECKLIST TEST INST (S) Hewlett-Packard 8991AOPT003 Peak Power Analyzer
PROC. NO. NA 17-20GP-48 MFR. MODEL SER. NO.
PROCEDURE MEASURED VALUES OUT
STEP FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN OF CALIBRATION TOLERANCES
NO. TOL
(1) (2) (3) (4) (5) (6) (7)
(Channel 3) (ns) (ns)
4.2.9 10 ns mkrs: 2 ns/div 10.0 8.9600 to 11.0400 " " " " "
.1 µs mkrs: 20ns/div
1 µs mkrs: 200 s/div
10 µs mkrs: 2 µs/div
.1 ms mkrs: 20µs/div
1 ms mkrs: 200µs/div
100.0 98.5950 to 101.4050
1000.0 994.9500 to 1005.0500
10.0 9.9585 to 10.0415
100.0 99.5940 to 100.4060
1000.0 995.9490 to 1004.0510
" 10 ms mkrs: 2 ms/div 10.0 9.9595 to 10.0405
4.3 Rise Time Tests (Channels 2 and 3)
4.3.5 Channel 2 - - ck ( ) <5 ns
4.3.9 Channel 3 - - ck ( ) <5 ns
4.4 Rise Time Tests (Channels 1 and 4)
(Sensor #1)
4.4.6 C1 Fall (T) ck ( ) NA Record fall time meas.
4.4.7 Max. TI Rise Time (ns) = " ck ( ) NA Calculate and record
2
100T1
++
ck ( )
4.4.13 Ch 1 rise time: BW-high - - ck ( ) < (value of step 4.4.7)
4.4.14 Ch 1 rise time: BW-low - - ck ( ) <1000 ns
4.4.15 Ch 1 rise time: BW-CW - - ck ( )
<100 µs
4.4.18 Ch 4 rise time: BW-high - - ck ( ) < (value of step 4.4.7) " Ch 4 rise time: BW-low - - ck ( ) <1000 ns " Ch 4 rise time: BW-CW - - ck ( )
<100 µs
(Sensor #2)
4.4.21 Ch 1 rise time: BW-high - - ck ( ) <45 ns " Ch 1 rise time: BW-low - - ck ( ) <1000 ns " Ch 1 rise time: BW-CW - - ck ( )
<100 µs
Page 3 of 4
13
Page 17
NAVAIR 17-20GP-48
CALIBRATION CHECKLIST TEST INST (S) Hewlett-Packard 8991AOPT003 Peak Power Analyzer
PROC. NO. NA 17-20GP-48 MFR. MODEL SER. NO.
PROCEDURE MEASURED VALUES OUT
STEP FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN OF CALIBRATION TOLERANCES
NO. TOL
(1) (2) (3) (4) (5) (6) (7)
4.5 Trigger Sensitivity Tests
4.5.4 Ch 1 @ -15 dBm: BW high - - ck( ) Waveform visible
4.5.7 Ch 1 @ -25 dBm: BW low - - ck( ) "
4.5.10 Ch 4 @ -15 dBm: BW high - - ck( ) "
4.5.13 Ch 4 @ -25 dBm: BW low - - ck( ) "
4.6 Instrumentation Uncertainty Tests
4.6.3 Software calibration - - ck( ) All tests pass
4.7 Check Source Tests (dBm) (dBm)
4.7.3 Pulse: TOP 10 9.5 to 10.5
4.7.4 CW: AVG " "
14
Page 4 of 4
Loading...