89410A-10
SERVICE NOTE
SUPERSEDES: None
89410A Vector Signal Analyzer
Serial Numbers: 0000A00000/US40514 521
Modifications to A20 ADC assembly require new repair strategy.
Situation:
To improve the ma rg in of uni ts failing the ma nufacturing amp litude linearity test, a modif ica tion has been im plemented on the A20 ADC asse mbly with date cod e 41 03 or greater. If the
instrume nt ha s option AY7, a second input channel, th en this modificati on m ust be the performed on both A20 ADC asse mbl ie s. This modification also requires firmware revision
A.08.23 or greater to be inst a ll ed .
Solution:
1. When replacing an A20 ADC assembl y upgr a de the firmwa re version to A.08.23 or
greater. Firmware upgrades are available on the web using the URL
http://www.agilen t.c om/find/89400 or upgrade kits can be purchased usin g part number 89400-84401 . To verify the firmware version that is inst alle d pr e ss S ystem Utility, more, fi rmwa re ve rsion.
2. When replacing a n A20 ADC assembly in an instr ument with opti on AY7, se co nd input channel, verify that the ADC revision levels are compatible. Cycle the power on
the 89410A while holding down the ’.’ and ’3 ’ keys on the front panel . Continue to
press the keys until Additional Hardwar e is di spl ay ed . The te st log will now show the
revision levels of Channel 1 ADC and Channel 2 ADC. A rev 5 (date code 4103)
ADC or greater cannot be used with rev 1 t hrough 4 ADC. Order exchange A20 assemblies for both channels if needed.
ADMINISTRATIVE INFORMATION
SERVICE NOTE CLASSIFICATION:
INFORMATION ONLY
AUTHOR:
DWH
© 2001 AGILENT TECHNOLOGIES
PRINTED IN U.S.A.
ENTITY:
A100
DATE: June 2001
ADDITIONAL INFORMATION:
PC0 #A 100-10 101431