Agilent 8922M Users Guide

User’s Guide
Agilent Technologies 8922M/S GSM Test Set
Agilent Part No. 08922-90211
Printed in UK
January 1998
© Copyright 1998, Agilent Technologies. All rights reserved. No part of this manual may be reproduced in any form or by any means(including electronic storage and retrieval or translation into a foreign language) without prior agree­ment and written consent from Agilent Technologies Inc. as governed by United States and international copyright laws.
Contents
1 Installing Your Agilent 8922M/S
Declaration of Conformity.................................................................... v
Sales and Service Offices...................................................................viii
Agilent Technologies 8922M/S Documentation Description............... x
Typeface Conventions..........................................................................xi
Using this Chapter .............................................................................1-2
Fuses and Power Cords .....................................................................1-3
Installation Overview ........................................................................1-5
General Information...........................................................................1-8
2 Making Measurements
Using This Chapter............................................................................2-2
Agilent Technologies 8922M/S Operating Modes ............................2-3
ACTIVE CELL..................................................................................2-5
TEST MODE.....................................................................................2-9
CW GENERATOR..........................................................................2-12
Measurements..................................................................................2-13
If You Have Problems with a Measurement ...................................2-30
Advanced Features...........................................................................2-39
3 Verifying Performance
About This Chapter............................................................................3-2
Setting up the Tests............................................................................3-3
Getting the Right Equipment ............................................................3-4
Installing and Operating the Software ..............................................3-5
Understanding the Tests.....................................................................3-7
Understanding Test Failures ............................................................3-13
Agilent Technologies 8922M/S Specifications................................3-14
Contents-1
Contents
4 Screens
Field Types........................................................................................ 4-2
Audio ................................................................................................4-4
Bit Error ............................................................................................ 4-8
Bit Error 2 ....................................................................................... 4-11
Cell Configuration - GSM 900 ....................................................... 4-16
Cell Configuration - E-GSM, DCS 1800, PCS 1900 ..................... 4-22
Cell Control - Active Cell ............................................................... 4-27
Cell Control - Active Cell + ........................................................... 4-31
Cell Control - Test Mode ................................................................ 4-33
Cell Control - CW Generator ......................................................... 4-35
Cell Control 2 ................................................................................. 4-37
Configure ........................................................................................4-49
CW Measurement ........................................................................... 4-54
Fast Bit Error ..................................................................................4-56
I/O Configuration ........................................................................... 4-59
Logging .......................................................................................... 4-63
Measurement Sync ......................................................................... 4-64
Message ..........................................................................................4-69
MS Information / Signaling ............................................................ 4-70
Oscilloscope, Main Controls .......................................................... 4-75
Oscilloscope, Trigger Controls ....................................................... 4-77
Oscilloscope, Marker Controls ....................................................... 4-80
Output RF Spectrum, Main View (Option 006 Only) ....................4-82
Output RF Spectrum, Trace View (Option 006 Only) .................... 4-84
Phase and Frequency Error - Multiburst OFF................................. 4-86
Phase and Frequency Error - Multi-burst ON................................. 4-88
Phase/Freq, Phase Err ..................................................................... 4-91
Phase/Freq, Data Bits ..................................................................... 4-93
Pwr Ramp: Rise Edge .................................................................... 4-95
Pwr Ramp, Top 2 dB ...................................................................... 4-97
Contents-2
Contents
Pwr Ramp, Fall Edge ......................................................................4-99
Pwr Ramp: Summary ....................................................................4-101
Pwr Ramp: Pulse (Option 006 Only) ............................................ 4-104
Pwr Ramp: Pulse Rise (Option 006 Only) ....................................4-107
Pwr Ramp: Pulse Fall (Option 006 Only) .....................................4-109
RF Generator / RF Analyzer (AF Gen) ........................................4-111
RF Generator / RF Analyzer (RF Analyzer) .................................4-112
RF Generator / RF Analyzer (RF Gen) .........................................4-115
Service ..........................................................................................4-118
SMS Cell Broadcast ......................................................................4-119
Spectrum Analyzer, Main Controls (Option 006 Only) ...............4-122
Spectrum Analyzer, RF Gen Controls (Option 006 Only) ...........4-124
Spectrum Analyzer, Marker Controls (Option 006 Only) ............4-125
Spectrum Analyzer, Auxiliary Controls .......................................4-127
Tests...............................................................................................4-129
5 Keys
Key Map ............................................................................................5-2
Function Keys....................................................................................5-3
Local Keys ......................................................................................5-10
Global Keys ....................................................................................5-11
Units Keys .......................................................................................5-12
Contents-3
Contents
6 Connectors
7 Messages
Front-Panel Connectors of the Agilent Technologies 8922M/S ....... 6-2
Rear-Panel Connectors of the Agilent Technologies 8922M/S....... 6-10
Signal Descriptions for SYSTEM BUS ......................................... 6-18
Timing Diagrams............................................................................. 6-27
Communication Failures .................................................................. 7-2
Firmware Error ................................................................................. 7-3
Sync Status ....................................................................................... 7-4
Protocol Error Messages .................................................................. 7-5
Timers................................................................................................ 7-6
Disconnects ....................................................................................... 7-8
Protocol Log Examples Of Typical Calls........................................ 7-11
Monitoring For Protocol Failure And Recovery During Test......... 7-12
Contents-4
8 Instrument BASIC
Agilent Technologies 8922M/S Instrument BASIC Overview......... 8-2
Configuration and Instrument Control .............................................. 8-4
Loading, Storing, and Running ....................................................... 8-10
Entering and Editing Programs ...................................................... 8-16
Memory Cards................................................................................. 8-20
Programming and Using the TESTS Subsystem............................. 8-26
Contents
A APPENDIX A
B Glossary
Purpose ............................................................................................. A-2
Equipment Required......................................................................... A-3
Connecting the Agilent 8922M to the HP/Agilent 37900D .............A-4
Setting Up the Agilent Technologies 8922M ................................... A-5
Setting Up the HP/Agilent 37900D.................................................. A-6
How to Obtain a Protocol Log.......................................................... A-8
Additional Information................................................................... A-11
Protocol Log of a Typical Call........................................................A-13
.......................................................................................................... B-2
Index 1
Contents-5
Contents
Contents-6
Warranty
Warranty
This Agilent Technologies instrument product is warranted against defects in material and workmanship for a period of one year from date of shipment. During the warranty period, Agilent Technologies will at its option, either repair or replace products which prove to be defective.
For warranty service or repair, this product must be returned to a service facility designated by Agilent Technologies. Buyer shall prepay shipping charges to Agilent Technologies and Agilent Technologies shall pay shipping charges, duties, and taxes for products returned to Agilent Technologies from another country.
Agilent Technologies warrants that its software and firmware designated by Agilent Technologies for use with an instrument will execute its programming instructions when properly installed on that instrument. Agilent Technologies does not warrant that the operation of the instrument, or software, or firmware will be uninterrupted or error free.
Limitation of Warranty
The foregoing warranty shall not apply to defects resulting from improper or inadequate maintenance by Buyer, Buyer-supplied software or interfacing, unauthorized modification or misuse, operation outside of the environmental specifications for the product, or improper site preparation or maintenance.
NO OTHER WARRANTY IS EXPRESSED OR IMPLIED. AGILENT TECHNOLOGIES SPECIFICALLY DISCLAIMS THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
Limitation of Remedies and Liability
THE REMEDIES PROVIDED HEREIN ARE BUYER’S SOLE AND EXCLUSIVE REMEDIES. AGILENT TECHNOLOGIES SHALL NOT BE LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES, WHETHER BASED ON CONTRACT, TORT, OR ANY OTHER LEGAL THEORY.
i
Responsibilities of the Customer
Responsibilities of the Customer
The customer shall provide; 1 Access to the products during the specified periods of coverage to perform mainte-
nance.
2 Adequate working space around the products for servicing by Agilent Technologies
personnel.
3 Access to and use of all information and facilities determined necessary by Agilent
Technologies to service and/or maintain the products. (In so far as these items may contain proprietary or classified information, the customer shall assume full responsibility for safeguarding and protection from wrongful use.)
4 Routine operator maintenance and cleaning as specified in the Agilent Technologies
Operating and Service Manuals.
5 Consumables such as paper, disks, magnetic tapes, ribbons, inks, pens, gases, solvents,
lamps, filters, fuses, seals, etc.
Certification
Agilent Technologies certifies that this product met its published specifications at the time of shipment from the factory. Agilent Technologies further certifies that its calibration measurements are traceable to the United States National Bureau of Standards and Technology, to the extent allowed by the Bureau’s calibration facility, and to the calibration facilities of other International Standards Organization members.
Assistance
Product maintenance agreements and other customer assistance agreements are available for Agilent Technologies products.
For any assistance, contact your local Agilent Sales and Service Office. For a list of contact information, see “Sales and Service Offices” on page ix.
ii
Notices
Notices
The material contained in this document is subject to change without notice. AGILENT TECHNOLOGIES MAKES NO WARRANTY OF ANY KIND WITH REGARD TO THIS MATERIAL, INCLUDING, BUT NOT LIMITED TO, THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. Agilent Technologies inc. shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance or use of this material.
Agilent Technologies assumes no responsibility for the use or reliability of its software on equipment that is not furnished by Agilent Technologies.
Restricted Rights Legend
If Software is for use in the performance of a U.S. Government prime contract or subcontract, Software is delivered and licensed as "Commercial computer software" as defined in DFAR 252.227-7014 (June 1995), or as a "commercial item" as defined in FAR
2.101(a) or as "Restricted computer software" as defined in FAR 52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclosure of Software is subject to Agilent Technologies’ standard commercial licenseterms, and non­DOD Departments and Agencies of the U.S. Government will receive no greater than Restricted Rights as defined in FAR 52.227- 19(c)(1-2) (June 1987). U.S. Government users will receive no greater than Limited Rights as defined in FAR52.227-14 (June 1987) or DFAR 252.227-7015 (b)(2) (November 1995), as applicable in any technical data. Copyright 200X Agilent Technologies Inc. All Rights Reserved.
Statement of Compliance
This product conforms to EN61010-1(1993) / IEC 1010-1(1990) +A1(1992) +A2(1994) / CSA C22.2 No. 1010.1(1993) Safety requirements for Electrical Equipment for Measurement, Control and Laboratory Use, and has been supplied in a safe condition. The instruction documentation contains information and warnings which must be followed by the user to ensure safe operation and to maintain the instrument in a safe condition.
iii
Electromagnetic Compatibility (EMC) Information
Electromagnetic Compatibility (EMC) Information
This product has been designed to meet the protection requirements of the European Communities Electromagnetic Compatibility (EMC) directive:
EN55011:1991 (Group 1, Class A) EN50082-1:1992
- IEC 1000-4-2 (1995) ESD
- IEC 1000-4-3 (1995) Radiated Susceptibility
- IEC 1000-4-4 (1995) EFT In order to preserve the EMC performance of this product, any cable which becomes worn
or damaged, must be replaced with the same type and specification.
Sound Emission
Manufacturer’s Declaration
This statement is provided to comply with the requirements of the German Sound Emission Directive, from 18 January 1991.
This product has a sound pressure emission (at the operator position) < 70 dB(A).
Sound Pressure Lp < 70 dB(A).At Operator Position.Normal Operation.According to ISO 7779:1988/EN 27779:1991 (Type Test).
Herstellerbescheinigung
Diese Information steht im Zusammenhang mit den Anforderungen der Maschinenlärminformationsverordnung vom 18 Januar 1991.
Schalldruckpegel Lp < 70 dB(A).Am Arbeitsplatz.Normaler Betrieb.Nach ISO 7779:1988/EN 27779:1991 (Typprfung).
iv
Declaration of Conformity
according to ISO/IEC Guide 22 and EN45014
Manufacturer’s Name: Agilent Technologies.
Declaration of Conformity
Manufacturer’s Address:
Agilent Technologies South Queensferry West Lothian, EH30 9TG Scotland, United Kingdom
Declares that the product
Product Name
:
Model Numbers: Product Options:
Conforms with the protection requirements of European Council Directive 89/336/EEC on the approximation of the laws of the member states relating to electromagnetic compatibility.
Against EMC test specifications EN 55011:1991 (Group 1, Class A) and EN 50082-1:1992
As Detailed in:
Assessed by:
Technical Report Number:6893/2200/CBR, dated 23 September 1997
GSM MS Test Set
Agilent Technologies 8922M and 8922S
This declaration covers all options of the above products as detailed in TCF A-5951-9852-02
Electromagnetic Compatibility (EMC) Technical Construction File (TCF) No. A-5951-9852-02
Dti Appointed Competent Body EMC Test Centre, GEC-Marconi Avionics Ltd., Maxwell Building, Donibristle Industrial Park, KY11 5LB Scotland, United Kingdom
Supplementary Information:
The product conforms to the following safety standards:
The product herewith complies with the requirements of the Low Voltage Directive 73/23/EEC, and carries the CE­marking accordingly.
South Queensferry, Scotland 17 November 1997
Location Date R.M. Evans / Quality Manager
EN 61010-1(1993) / IEC 1010-1(1990) +A1(1992) +A2(1994) CSA-C22.2 No. 1010.1-93 EN 60825-1(1994) / IEC 825-1(1993)
v
Safety Information
Safety Information
The following general safety precautions must be observed during all phases of operation of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design, manufacture, and intended use of the instrument. Agilent Technologies Inc. assumes no liability for the customer’s failure to comply with these requirements.
GENERAL
This product is a Safety Class 1 instrument (provided with a protective earth terminal). The protective features of this product may be impaired if it is used in a manner not specified in the operation instructions. All Light Emitting Diodes (LEDs) used in this product are Class 1 LEDs as per IEC 60825-1.
ENVIRONMENTAL CONDITIONS
This instrument is intended for indoor use in an installation category II, pollution degree 2 environment. It is designed to operate at a maximum relative humidity of 95% and at altitudes of up to 2000 meters. Refer to the specifications tables for the ac mains voltage requirements and ambient operating temperature range.
BEFORE APPLYING POWER
Verify that the product is set to match the available line voltage, the correct fuse is installed, and all safety precautions are taken. Note the instrument’s external markings described under "Safety Symbols".
GROUND THE INSTRUMENT
To minimize shock hazard, the instrument chassis and cover must be connected to an electrical protective earth ground. The instrument must be connected to the ac power mains through a grounded power cable, with the ground wire firmly connected to an electrical ground (safety ground) at the power outlet. Any interruption of the protective (grounding) conductor or disconnection of the protective earth terminal will cause a potential shock hazard that could result in personal injury.
FUSES
Only fuses with the required rated current, voltage, and specified type (normal blow, time delay,etc.) should be used. Do not use repaired fuses or short-circuited fuse holders. To do so could cause a shock or fire hazard.
vi
Safety Information
DO NOT OPERATE IN AN EXPLOSIVE ATMOSPHERE
Do not operate the instrument in the presence of flammable gases or fumes.
DO NOT REMOVE THE INSTRUMENT COVER
Operating personnel must not remove instrument covers. Component replacement and internal adjustments must be made only by qualified service personnel.
Instruments that appear damaged or defective should be made inoperative and secured against unintended operation until they can be repaired by qualified service personnel.
WARNING The WARNING sign denotes a hazard. It calls attention to a procedure, practice, or the like, which, if not correctly performed or adhered to, could result in personal injury. Do not proceed beyond a WARNING sign until the indicated conditions are fully understood and met.
Caution The CAUTION sign denotes a hazard. It calls attention to an operating procedure, or the like, which, if not correctly performed or adhered to, could result in damage to or destruction of part or all of the product. Do not proceed beyond a CAUTION sign until the indicated conditions are fully understood and met.
vii
Safety Symbols
Safety Symbols
The following symbols on the instrument and in the manual indicate precautions which must be taken to maintain safe operation of the instrument
Safety Symbols
The Instruction Documentation Symbol. The product is marked with this symbol when it is necessary for the user to refer to the instructions in the supplied documentation.
Indicates the field wiring terminal that must be connected to earth ground before operat­ing the equipment - protects against electrical shock in case of fault.
Frame or chassis ground terminal - typically connects to the equipment's metal frame.
Alternating current (AC)
Direct current (DC) Warning, risk of electric shock.
N L
ISM 1-A This is a symbol of an Industrial, Scientific, and Medical Group 1 Class A product.
Terminal for Neutral conductor on permanently installed equipment.
Terminal for Line conductor on permanently installed equipment..
The CE mark shows that the product complies with all relevant European Legal Direc­tives.
The CSA mark is a registered trademark of the Canadian Standards Association, and indicates compliance to the standards defined by them.
Indicates that a laser is fitted. The user must refer to the manual for specific Warning or Caution information to avoid personal injury or damage to the product.
viii
Sales and Service Offices
Any adjustment, maintenance, or repair of this product must be performed by qualified personnel. Contact your customer engineer through your local Agilent Technologies Service Center. You can find a list of local service service representatives on the web at:
http://www.agilent-tech.com/services/English/index.html You can also contact one of the following centers and ask for
a test and measurement sales representative.
Asia Pacific:
Agilent Technologies 19/F, Cityplaza One, 1111 King’s Road, Taikoo Shing, Hong Kong, SAR (tel) (852) 2599 7889 (fax) (852) 2506 9233
Sales and Service Offices
Japan:
Agilent Technologies Japan Ltd. Measurement Assistance Center 9-1, Takakura-Cho, Hachioji-Shi Yokyo, 192-8510 (tel) (81) 426 56 7832 (fax) (81) 426 56 7840
Australia/New Zealand:
Agilent Technologies Australia Pty Ltd 347 Burwood Highway Forest Hill, Victoria 3131 (tel) 1-800 629 485 (Australia) (fax) (61 3) 9272 0749 (tel) 0 800 738 378 (New Zealand) (fax) (64 4) 802 6881
ix
Sales and Service Offices
Canada
Agilent Technologies Canada Inc. 5150 Spectrum Way, Mississauga, Ontario L4W 5G1 (tel) 1 877 894 4414
Europe:
Agilent Technologies Test & Measurement European Marketing Organisation P.O. Box 999 1180 AZ Amstelveen The Netherlands (tel) (31 20) 547 9999
Latin America:
Agilent Technologies Latin American Region Headquarters 5200 Blue Lagoon Drive, Suite #950 Miami, Florida 33126 U.S.A. (tel) (305) 267 4245 (fax) (305) 267 4286
United States:
Agilent Technologies Test and Measurement Call Center P.O. Box 4026 Englewood, CO 80155-4026 (tel) 1 800 452 488
In any correspondence or telephone conversations, refer to the power sensor by its model number and full serial number. With this information, the Agilent Technologies representativecan quickly determine whether your unit is still within its warranty period.
x
Agilent Technologies 8922M/S Documentation Description
Agilent Technologies 8922M/S Documentation Description
Documentation Shipped with Your GSM Test Set
Agilent 8922M/S GSM Test Set Quick Start Guide.
This guide gives a brief description on how to make each of the measurements required to test a GSM mobile phone. More detailed descriptions are given in the Agilent 8922M/S
GSM Test Set User’s Guide.
Agilent 8922M/S GSM Test Set User’s Guide.
This guide contains information on how to set up the Agilent 8922M/S for making measurements and verifying performance. It also contains more detailed information on each of the screens, keys, and connectors, and how to use the IBASIC facilities available on the Agilent 8922M/S.
Agilent 8922M/S GSM Test Set Performance Test Software.
This 3.5 inch floppy disk allows you to verify the performance of the Agilent 8922. Instructions on how to use this are detailed in the Agilent 8922M/S GSM Test Set User’s
Guide.
Agilent 8922M/S GSM Test Set Programming Reference Guide.
This guide describes, in detail, each of the GPIB command sets for the Agilent 8922M/S.
Agilent 8922 Multi-Band Test System User’s Guide1.
This is a supplementary user’s guide that describes the additional features found with the Multi-Band test system. The test system is used for testing and making measurements of dual band mobiles.
1. The user’s guide is only available with the HP 8922M/S Option 010.
xi
Typeface Conventions
Typeface Conventions
Italics
Display Display text is used to show examples, fields, and prompts that are
PRESET
Soft keys Display text is used to show examples, fields, and prompts that are
Soft keys All software listings in this manual can be identified with this font.
Italic type is used for emphasis.
displayed on the Agilent 8922M/S screen. Keycaps on the Agilent 8922M/S keyboard are enclosed in boxes.
displayed on the Agilent 8922M/S screen.
xii
1
Installing Your Agilent 8922M/S
1-1
Installing Your Agilent 8922M/S
Using this Chapter
Using this Chapter
Use the following procedure to get the Agilent Technologies 8922M/S powered-up correctly. After completing this procedure, refer to the Quick Start Guide for an introduction to operating the Agilent Technologies 8922M/S and Chapter 2, “Making Measurements” for more extensive information on using the Agilent Technologies 8922M/S.
Equipment Supplied Fuse Envelope and Fuse
REF OUT/REF IN cable (Option 001 only)
Power Cord
CAUTION To avoid potential injury, ensure that two people are employed in lifting the Agilent
8922M/S out of the box and for any other instrument moves.
NOTE: If you have the Agilent 8922M/S Option 010 Multi-Band Test System, refer to the
appropriate Agilent 8922 Multi-Band User’s Guide for more information on connection and operating differences.
1-2
Installing Your Agilent 8922M/S
Fuses and Power Cords
Fuses and Power Cords
CAUTION Before plugging this instrument into the Mains (line) voltage, be sure the correct
voltage on the line voltage selection card has been selected.
Line Voltage and Fuse Selection
Figure 1-1 Voltage Selection Card and Fuse Installation
Other Fusing Non user replaceable fusing on the secondary of this instrument power supply are;
Verify that the line voltage selection card is matched to the power source (see Figure 1-1 on page 1-3). Order fuse Agilent part 2110-0083 (2.5 A 250 V, normal blow) for replacement.
Fuse Current Type Volatge
F1 5A F 5.0A H 250V F2 3A F 3.0A H 250V F3 10A F 10A 32V F4 3A F 3.0A H 250V F5 0.5A F 0.5A H 250V F6 3A F 3.0A H 250V
1-3
Power Cords
Installing Your Agilent 8922M/S
Fuses and Power Cords
Agilent Part
1-4
Agilent
Installing Your Agilent 8922M/S
Installation Overview
Installation Overview
1 Connect a 1, 2, 5, 10, or 13 MHz signal to the REF IN. If you are using option
001, connect as shown, see Figure 1-2 on page 1-5, with the supplied cable (OPT 001 REF OUT to REF IN).
2 Connect the supplied power cord to the Agilent 8922M/S and power up the
instrument.
1
Power Source
Figure 1-2 Rear View Connections
1-5
Installing Your Agilent 8922M/S
Installation Overview
3 Access the CONFIG screen to customize the set-up of your Agilent 8922M/S.
This is done by:
Moving to the CONFIG field in the bottom right-hand corner of the Cell Control screen, (the first screen that appears after power-up or after selecting
PRESET
). Rotate the cursor control knob (refer to “a”, see Figure 1-3 on
page 1-6) until you are on the CONFIG field.
Selecting the CONFIG field by pushing the cursor control knob (diagram “b”, see Figure 1-3 on page 1-6).
Figure 1-3 Access CONFIG Screen
1-6
Installing Your Agilent 8922M/S
Installation Overview
4 To guarantee the correct operation of the Agilent 8922M/S with all mobile
phones, you should use a high-stability timebase. This step details how to configure the Agilent 8922M/S for use with either the
option 001, high-stability timebase, or an external reference signal. A If you have option 001 installed and wish to use this as your reference, ensure
that the OPT 001 REF OUT field in the Configure screen is set to On. Access the Reference field in the Configure screen and select 10 MHz.
B If you wish to use an external reference signal, set the Reference field to
1, 2, 5, 10, or 13 MHz according to the frequency reference you intend to use.
A
Figure 1-4 High-stability timebase set-up
To customize your Agilent 8922M/S further, refer to Chapter 4, Screens.
B
1-7
Installing Your Agilent 8922M/S
General Information
General Information
Operation and Storage Environment
Instrument Options Refer to Agilent 8922M/S Specifications in Chapter 3, Performance Verification for
Specifications Refer to Agilent 8922M/S Specifications in Chapter 3, Performance Verification for
Refer to General Specifications in the Agilent 8922M/S Specifications section of Chapter 3, Performance Verification for information about the operation or storage environment.
information about instrument options.
instrument specifications.
1-8
2
Making Measurements
2-1
Making Measurements
Using This Chapter
Using This Chapter
Use this chapter to obtain an overview of how to operate the Agilent 8922M/S GSM Test Sets.
This chapter is divided into the following sections:
Making Measurements
Agilent 8922M/S Operating Modes - explains how to configure the Agilent 8922M/S so that measurements can be made.
Measurements - details how to perform the many different measurements available on the Agilent 8922M/S.
Advanced Features - advanced information for experienced users. To help you understand this chapter, you may wish to consult the following sources
of information:
Quick Start Guide - This briefly explains the basic operating principles of the
Agilent 8922M/S including making a call and simple measurements.
Typeface Conventions - Found at the front matter of this manual. This explains
how to interpret the different typefaces used in this manual.
Field Types - Refer to Chapter 4, Screens. Explains in detail the different kinds
of “fields” or areas which are displayed on the screens. Fields are mainly used for measurement results, data entry and access to other screens.
Screens -Chapter 4, provides more detailed information on each of the fields
contained within all of the screens available on the Agilent 8922M/S.
NOTE: If you have the Agilent 8922M/S Option 010 Multi-Band Test System, refer to the
appropriate Agilent 8922 Multi-Band User’s Guide for more information on making dual band measurements.
2-2
Making Measurements
Agilent Technologies 8922M/S Operating Modes
Agilent Technologies 8922M/S Operating Modes
This section details the procedures necessary to control the Agilent 8922M/S and GSM mobile phone in each of the main operating modes. It is recommended that you read this section before attempting the “Measurements” section.
The Agilent 8922M/S GSM TestSets have three main modes of operation, these are:
ACTIVE CELL
TEST MODE
CW GENERATOR
Active Cell The first mode is the default mode and configures the Agilent 8922M/S as a GSM
Base Station emulator. This allows calls to be made from the Agilent 8922M/S to the mobile phone and vice versa. Measurements can then be made to verify the mobile phone’s performance.
Test Mode This mode is used when measurements need to be made on the mobile phone
without a call being set up.
CW Generator This mode configures the Agilent 8922M/S as a standard Continuous Wave (CW)
Signal Generator. This is used for test applications that require an unmodulated RF carrier.
2-3
Making Measurements
Agilent Technologies 8922M/S Operating Modes
To configure the Agilent 8922M/S to the desired operating mode, carry out the following instructions, referring to Figure 2-1 on page 4 for the position of the fields.
Ensure (1) is set to mobile phone type you require (GSM900, E-GSM, DCS1800,
or PCS1900).
NOTE For use with DCS1800 or PCS1900, refer to the HP/Agilent 83220A/E Users Guide
which explains how to configure the Agilent 8922M/S to test other mobile phone formats.
Use the knob to move the cursor to the lower field below the Operating Mode (2).
Push the knob and select the desired Operating Mode (ACTIVE CELL, TEST MODE, or CW GENERATOR) from the menu which appears at the bottom right-hand side of the screen.
1
2
Figure 2-1 Agilent 8922M/S Operating Modes
In addition to the three main operating modes, there are three other modes available. These are:
ACTIVE CELL+
TEST MODE+
CW GENERATOR+
These “+” modes have all the features of the other modes but include extra diagnostic information on the screen. This information is designed to help advanced users in fault-finding mobile phones.
2-4
ACTIVE CELL
Making Measurements
ACTIVE CELL
Making a Call From the Mobile Phone to the Agilent 8922M/S
This is the default mode after cycling the power or selecting . A functional mobile phone will lock on (camp on) to a signal which is produced by the Agilent 8922M/S. The characteristic of this signal appears in the BASE STATION fields, (1), see Figure 2-2 on page 2-6. These fields show the Channel number and the Amplitude of the signal. Once the mobile phone has camped on to the signal, it is possible to make a call between the mobile phone and the Agilent 8922M/S.
Cycle the power on the Agilent 8922M/S or select .
Insert a Test SIM card into the mobile phone. [The Test SIM (Subscriber Identity Module) holds the user’s customized information. Agilent supply both micro and standard Test SIMs.]
Connect the mobile phone to the RF IN/OUT connector on the front panel of the Agilent 8922M/S.
Switch on the mobile phone and wait for it to camp to the Agilent 8922M/S (most mobile phones display 001-01 when the mobile phone has camped).
Dial any number on the mobile phone and press send.
The call set-up should now proceed. The CALL STATUS field on the screen should display CONNECTED, (2), see Figure 2-2 on page 2-6. Once the call has been established, the following information is displayed:
Peak Power transmitted by the mobile phone is displayed in the center of the screen.
In the CELL STATUS area, (3), see Figure 2-2 on page 2-6, the mobile phone reports its:
Transmit power level (TX Lev)
Received level (RX Lev)
Received signal quality (RX Qual)
Confirmation of the Traffic Channel and Timeslot is displayed to the left of the mobile phone reports.
PRESET
PRESET
If the call does not proceed, there may be a problem with the mobile phone. Refer to “Test Mode” in this chapter which explains how to configure the Agilent 8922M/S to troubleshoot the mobile phone. When a call is connected, any speech received from the mobile phone is echoed back by the Agilent 8922M/S, into the mobile phone with a 0.5 second delay. This allows provisional checking of the mobile phone’s audio sections by speaking into the mobile phone and listening for the echo of your voice.
2-5
Making Measurements
ACTIVE CELL
2
3
3
4
6
5
Figure 2-2 Active Cell Mode
Making a Call From the Agilent 8922M/S to the Mobile Phone
NOTE The Network number cannot be used to make a call. The Network Telephone number
To call (or page) the mobile phone from the Agilent 8922M/S, it is necessary for the Agilent 8922M/S to know the number of the SIM in the mobile phone. [This number is the International Mobile Subscriber Identity (IMSI). This and other information is stored on the SIM card.]
is allocated to the SIM and is used in a real network to page the IMSI.
There are two ways the Agilent 8922M/S can receive this information:
Make a call from the mobile phone to the Agilent 8922M/S. Once the call is originated from the mobile phone, the Agilent 8922M/S
automatically reads the IMSI on the SIM card. If the previous call on the Agilent 8922M/S was made with the SIM card inserted in the mobile phone, and the mobile phone is camped to the Agilent 8922M/S, the mobile can be called by
pressing .
Enter the MS Information screen and enter the IMSI into the Agilent 8922M/S manually.
2-6
ORG CALL
Making Measurements
ACTIVE CELL
Press the following keys to do this:
SHIFT CELL CONFIG
, (MS INFO)
Move the cursor to the Paging IMSI field and enter the IMSI using the numeric data entry keypad.
CELL CNTL ORG CALL
•,
Performing a “location update” from the MS INFO screen allows the Agilent 8922M/S to update the IMSI. This can be done by either:
changing the Current location parameters and waiting for the mobile phone to re-camp.
settingIMSI Attach/Detach to On before poweringonthe phone. When the phone camps its IMSI is set on the MS INFO screen and a call can be made from the Agilent 8922M/S.
Changing Channel, Timeslot, and the Transmit Level
You can alter the parameters of the call before the call has been set up, or during a call. They are displayed on the right-hand side of the screen under MOBILE PHONE, (4), see Figure 2-2 on page 2-6. The parameters are:
Channel
Transmit Level (TX Level)
Timeslot
Channel To change channel, highlight the field and enter a new channel number from the keypad. There is no interruption of communication between the Agilent 8922M/S and the mobile phone. For additional information on valid Absolute RF Channel Numbers (ARFCNs), consult the specifications in Chapter 3.
TX Level This is a coded number used by the Agilent 8922M/S to command the mobile phone to transmit at a particular power. When the TX Level is changed, two things happen:
The mobile phone should change its transmitted power.
The Amplitude field in the Expected Input area of the screen, (5), see Figure 2-2 on page 2-6, automatically adjusts to the nominal value defined by the TX Level field. This allows the RF analyzer in the Agilent 8922M/S to align itself with the mobile phone’sexpected output amplitude. If thesignal is not within 3 dB of theexpected amplitude, it isnecessary to modify the Amplitude field so that it is within 3 dB. Directly entering a value in the Amplitude field does not change the transmitted level code transmitted to the mobile phone.
2-7
Making Measurements
ACTIVE CELL
There may be many reasons for the measured level not being close to the expected level. The two most likely are that, either, the mobile phone is not operating correctly, or, there is some power loss between the Agilent 8922M/S RF IN/OUT connector and the mobile phone. If you suspect it is the second case, you can compensate the Agilent 8922M/S generator settings and measurement results for external losses or gains. The compensation is carried out in the CONFIGURE screen which is accessed using the CONFIG field on the bottom right-hand side of the screen, (6), see Figure 2-2 on page 2-6. Refer to Chapter 4, Screens for further information.
Timeslot To change the timeslot, highlight the field and enter a new timeslot value from the keypad. The timeslots can vary from 2 to 6. (Timeslots 0, 1, and 7 are reserved for maintaining communication between the Agilent 8922M/S and the mobile phone.)
NOTE While the Active Cell is selected, the three MOBILE PHONE parameters are also
available on the bottom right-hand side of all measurement screens. This provides control of the mobile phone during measurements.
2-8
Making Measurements
TEST MODE
TEST MODE
To enter this mode, select TEST MODE as described in the section titled “Agilent 8922M/S Operating Modes”.
In TEST MODE, the Agilent 8922M/S no longer controls the mobile phone. The TEST MODE is used when it is not desired, or not possible, to set up a call between the Agilent 8922M/S and the mobile phone.
The MOBILE PHONE area available in Active Cell controls three functions simultaneously:
The traffic channel transmitted by the Agilent 8922M/S.
The traffic channel transmitted by the mobile phone.
The corresponding channel used by the Agilent 8922M/S to measure the mobile phone.
In Test Mode, these functions are now separated. The traffic channel transmitted by the Agilent 8922M/S is now controlled by the Traffic Chan parameters, (1), see Figure 2-3 on page 2-10. The mobile phone’s transmission must be controlled manually using the mobile phone manufacturer’s service and test modes. The Agilent 8922M/S measurements are controlled by the fields below Expected Input, (2), see Figure 2-3 on page 2-10.
These fields are:
Channel
TX Level
Burst Type
2-9
Making Measurements
TEST MODE
NOTE While the Test Mode is selected, the three MEASURE ON parameters are also
available on the bottom right-hand side of all measurement screens. This provides control of the Expected Input parameters during measurements.
4
2
3
Figure 2-3 Test Mode
Mobile Phone Transmitter Testing Using Test Mode
To measure an incoming signal from the mobile phone’s transmitter, perform the following steps:
Select TEST MODE.
In the Expected Input area, (2), see Figure 2-3 onpage 2-10, enter the GSM
channel number of the signal to be measured. This will automatically update the input Frequency field below it, (3), see Figure 2-3 on page 2-10. If a frequency needs to be measured which does not correspond to a standard GSM channel, enter the non-standard value into the Frequency field. This value can be anywhere from 10 MHz to 1000 MHz. A direct entry in the Frequency field over-rides the frequency defined by the Channel field above.
Enter the expected transmit level of the mobile phone in the TX Level field. This will automatically update the Amplitude field below with the equivalent TX level represented in dBm. If a non-standard input level is expected, enter the value directly into the Amplitude field, this overrides the TX Level field above.
Entertheexpected Colour Code of the input signal. The Colour Codeisafunction of the central “midamble” of the transmitted burst and is needed so that measurements are correctly synchronized to the received burst. If the Colour
1
2-10
Code is not known, it can be determined and corrected from measurements described later. Refer to the “Advanced Features” section mentioned later in this chapter.
NOTE In the Active Cell mode, the Colour Code is automatically set.
Once these have been selected, the Agilent 8922M/S is ready to measure incoming signals of the type specified. The mobile phone should be set up to generate a corresponding test signal to the one expected. This will require access to the mobile phone manufacturer’s servicing or test modes.
Making Measurements
TEST MODE
Mobile Phone Receiver Testing Using Test Mode
It is possible to analyze the mobile phone’s response to the Agilent 8922M/S Broadcast Channel (BCH), by varying the BCH number and amplitude, (4), see Figure 2-3 on page 2-10. TEST MODE provides the Agilent 8922M/S with a “forced” traffic channel (TCH) generator which can be turned on and off without the need for any signaling or the presence of a mobile phone. A forced TCH can be generated on any channel in the GSM Base Station range and is enabled by the On/ Off toggle field (1) below the Traffic Chan or the and
ORG CALL END CALL
keys. The presence of both the BCH and forced TCH signals from the Agilent 8922M/S
allows the mobile phone’s receiver to be stimulated with signals identical to those used on a real call. These signals can be used in conjunction with the mobile phone manufacturer’s service and test modes to help measure and troubleshoot the mobile phone. With these features, it is possible to make bit error rate measurements and test the mobile phone’s receiver sensitivity when there is no call set up.
2-11
Making Measurements
CW GENERATOR
CW GENERATOR
To enter this mode, select CW GENERATOR as described in the section titled “Agilent 8922M/S Operating Modes”. The CW Generator mode has the same measurement capabilities as the Test Mode but replaces the GSM BCH and forced TCH signals with a single unmodulated RF carrier. The frequency and amplitude of the Continuous Wave (CW) signal is controlled by the Channel, Amplitude, and the Frequency fields (1). When a GSM channel number is entered in the Channel field, this automatically updates the generator Frequency field below it. If it is necessary to generate a non GSM channel frequency, enter a value directly into the frequency field. This range varies from 10 MHz through 1000 MHz. A direct frequency entry over-rides the Channel field above. The unmodulated RF signal can be used for any general purpose application and can be particularly useful for checking some aspects of the mobile phone’s receiver where a CW signal is needed.
Figure 2-4 CW Generator
2-12
1
Making Measurements
Measurements
Measurements
The measurements available on the Agilent 8922M/S can all be accessed from the cell control screen by selecting a measurement field under MEASUREMENTS, (see Figure 2-5 on page 2-13) and pushing the knob.
GSM Specific Measurements
Ancillary Measurements
The measurements available are:
Peak Carrier Power
Phase and Frequency Error
Power Ramp Mask
Bit Error Rate
Output RF Spectrum (modulation or ramping, Option 006 only)
You can also use the toolkit capabilities of the Agilent 8922M/S. These additional measurements are:
Spectrum Analyzer (Option 006 only)
Scope
Audio Measurements
CW Measurements
GSM Specific Measurements
Figure 2-5 Agilent 8922M/S Measurements
Ancillary Measurements
2-13
Making Measurements
Measurements
After making one type of measurement, another can be made by simply pressing
CELL CNTL
and, using the knob, selecting the next measurement of your choice.
Peak Carrier Power Measurement
NOTE Select the Pwr Zero field, disconnect external source and lower BCH power, (3),
The peak transmitter carrier power averages the transmitter carrier power for a single burst. This average is calculated over the time that the data information bits are transmitted.
see Figure 2-6 on page 2-14, to zero the power meter before any signal is applied.
Method
The peak transmitter carrier power value is displayed on the cell control screen. After you have set up a call (Active Cell) or manually aligned the Agilent 8922M/S to the mobile phone (Test Mode/CW Generator), the Peak Power reading is displayed in the center of the screen (1).
Below the CALL STATUS area of the cell control screen are reports from the mobile phone. These appear only when using Active Cell. They indicate the TX Level which the phone is transmitting and the RX Level and RX Quality of the received signal, (2), see Figure 2-6 on page 2-14.
2
1
3
Figure 2-6 Peak Power Measurements
2-14
Making Measurements
Measurements
Phase and Frequency Error Measurement
NOTE The test is run automatically when the screen is selected.
Phase error and frequency error are measures of the modulation and noise performance of the radio’s transmitter path.
Method
Select the PHASE FRQ field on the cell control screen to access the phase and frequency measurements.
The measurement fields in this screen are RMS Phase Error, Peak Phase Error and Frequency Error (1), see Figure 2-7 on page 2-15.
The Agilent 8922M/S input sensitivity (2) can be varied, if necessary, to ensure that the input signal level matches the Agilent 8922M/S RF Analyzer. Valid measurements are only made when the signal is within 3 dB of the RF Analyzer setting.
The Agilent 8922M has the option of using multi-burst measurements (4). This gives the ability to make several phase and frequency measurements using more than one burst. For more information refer to Chapter 4, Screens, Phase and Frequency Error screen.
1
2
4
3
Figure 2-7 Phase and Frequency Error Measurements
Two additional screens in the phase and frequency measurement section can be revealed if you select (3).
2-15
Making Measurements
Measurements
These are:
PHASE ERR - this displays the phase error graphically. The phase error trace is displayed using an autoscaling phase error axis versus data bits (numbered 0 through 147).
DATA BITS - this screen allows you display a screen which details the values of the 148 bits in the timeslot (including midamble). If a known test signal is being used, the reception of these bits can be verified.
Power Ramp Mask Measurements
To avoid unwanted interference and to ensure successful reception at the Base Station, the mobile phone’stransmitted signal must conform to GSM standards. The purpose of the power ramp is to display the pulsed signal and verify that it conforms to these standards.
Method
Selecting PWR RAMP on the cell control screen gives you access to the power ramp measurement screens.
The marker (1), see Figure 2-8 on page 2-16 can be used to make a measurement at a discrete point in time along the signal trace. The amplitude at this point is displayed in the top left-hand corner of the screen.
2
Figure 2-8 Power Ramp Measurements
2-16
1
3
Making Measurements
Measurements
The power ramp measurements are divided into three screens where you can view different parts of the signal and one screen which displays a series of amplitude values at various times during the burst. These screens can be revealed by highlighting and selecting from the View field (2):
Rise Edge - displays the top 30 dB of the rising section of the waveform.
Top 2 dB - displays the signal during the middle part of the burst allowing analysis of the ripple of the signal.
Fall Edge - displays the signal during the falling edge of the burst allowing analysis of the fall time of the signal.
Summary - details the amplitude measurements made at the times selected in the 12 time fields. You can choose your own time-positions, or use the default settings.
The sensitivity of the Agilent 8922M/S receiver can be varied (3) allowing you to verify whether the input signal level matches the Agilent 8922M/S RF Analyzer. Valid measurements are only made when the signal is within 3 dB of the RF Analyzer setting.
Measurement Summary
The Measurement Summary field on the DSP Analyzer Ampl Main screen displays whether HI/LO limits set for the measurement display fields, (Ampl1-12, pk+ flatness, or pk- flatness) were exceeded in the last measurement. The possible Measurement Summary displays are:
Failed
One or more measurement limit was exceeded.
Passed
No measurement limits were exceeded.
- - - -
No measurement limits are set, or, all of the Ampl and Pk measurement displays are turned off.
A blank field
The blank field is displayed when the measurement is armed. It will remain blank until the measurement is complete.
Refer to “Pulse Measurements” within this section for a description of Pulse Measurements. These measurements are available with option 006 only.
2-17
Making Measurements
Measurements
Pulse Measurements (Available if Option 006, Spectrum Analyzer, is Fitted)
If you have option 006 (spectrum analyzer) installed, you can make measurements on the lower portion of the pulse. These measurements can be accessed from the Power Ramp screens.
The main measurements, Pulse On/Off Rise and Pulse On/Off Fall are displayed
at the top of the screen (1), see Figure 2-9 on page 2-18
The sensitivity of the Agilent 8922M/S receiver can be varied (2) allowing you
to verify whether the signal still falls within the boundaries of the GSM power mask.
1
32
Figure 2-9 Pulse Measurements
The Agilent 8922M/S provides three pulse measurements which can be selected using View (3). These are:
Pulse - pulse on/off ratio measures the ratio of the average transmitter- power (pulse on) to a specified time-position when power is reduced (pulse off). The default settings are 28µs before bit 0 and 28µS after bit 147.
Puls Rise - this screen displays the signal during the initial rise of the pulse.
Puls Fall - this screen displays the signal during the final fall of the pulse.
2-18
Making Measurements
Measurements
Bit Error Rate Measurement
NOTE There are two sets of results displayed.
The Bit Error Rate measurement allows you to test the sensitivity of the mobile phone’s receiver. By reducing the signal transmitted by the Agilent 8922M/S, you can test the ability of the receiver to accurately decode its incoming signal. Data bits that are decoded are sent back to the Agilent 8922M/S. The Agilent 8922M/S compares them to original signal that was sent out and the differences are derived from this. The audio echo function that works in the Cell Control screen is turned off when making a BER measurement.
Method
To run a bit error rate test, select the BIT ERROR field in the cell control screen. This reveals the bit error measurement screen.
This test will run automatically if you set the Single/Cont field to CONT.To run a test set the Run/Stop field to RUN. This toggle field start/stops the BER test (1) ratio, see Figure 2-10 on page 2-19.
Intermediate Results - this is a running total of the bit errors as the bits are being
tested (2), (this is useful where a large number of bits are being tested).
Final Results - this shows the completed BER (3). This is displayed after all the
bits have been tested.
3
2
1
Figure 2-10 Bit Error Rate Measurement
4
2-19
Making Measurements
Measurements
By varying the Base station Amplitude field (4), you can test the mobile phone’s receiver sensitivity. Actual results can be compared with the values reported by the radio.
NOTE If the Base Station amplitude is lowered too much, the radio will lose the call. This
will need to be re-established by increasing the base station amplitude, returning to the cell control screen, and originating the call before you can continue measurements.
NOTE Not all phone types support burst-by-burst BER measurements.
OutputRFSpectrum Measurements (Available if Option 006, Spectrum Analyzer, is Fitted)
The Output RF spectrum measurement shows the spectral power (due to the effects of ramping or modulation) at a specified frequency offset. These are relative measurements that compare the result at the offset frequency with the value at the center frequency. You can also view the output RF spectrum trace from this screen.
Method
Select OUT RF SP in the cell control screen. To make a measurement, do the following:
Ensure the Freq Offset is set to zero (1), see Figure 2-11 on page 2-21.
Select Ramp Ref or Mod Ref (2) (depending on whether you want to make a ramping measurement or a modulation measurement).
2-20
Making Measurements
Measurements
This sets the reference level to which the offset frequency values are compared.
Select Ramping or Modulation (3).
Set your Freq Offset value (4).
5
2,3
Figure 2-11 Output RF Spectrum Measurements
The trace of the output RF spectrum can be viewed if you highlight View and
select Trace (5).
OutputRFSpectrum MeasurementsUsing a 3-Pole Synchronously TunedMeasurement Filter
GSM Recommendation 11.10 and 11.20 set the requirements for the out-of-channel performance of base station and mobile station transmitters with the Output RF Spectrum specification. The specification calls for the measurement of transmitted energy at several offsets from the carrier frequency. Two types of measurements are required:
Check the interference due to modulation.
Check the interference due to ramping of the signal amplitude (switching transients).
Limits are specified for each type of measurement at specified frequency offsets. The GSM specified limits are based on a 5-pole synchronously tuned measurement filter.
1,4
2-21
Making Measurements
Measurements
In the Agilent 8922, Output RF Spectrum measurements are made using a 3-pole synchronously tuned measurement filter. The measurement results will differ from measurements using a 5-pole filter.
The difference between measurements of the “ideal” signals using a 3-pole filter and a 5-pole filter are shown in Table 1 and Table2. These values are added to the GSM specifications to obtain equivalent limits for Output RF Spectrum measurements using 3-pole synchronously tuned measurement filter. The adjusted Output RF Spectrum limits are based on a 3-pole filter are shown in Table 3 and Table 4.
Table 1 3-Pole Specification Difference Output RF Spectrum due to Modulation
Offset from Carrier (kHZ)
100 200 250 400 600 to
1800
Difference (dB) 0 2.5 1.0 9.0 0
Table 2 3-Pole Specification Difference Output RF Spectrum due to Switching
Offset from Carrier (kHZ)
400 600 1200 1800
Difference (dB) 6420
NOTE It is important to note that these values are estimates based on simulation and
measurements. They are intended to allow the user to estimate their transmitters conformance to GSM Output RF Spectrum requirements.
2-22
Making Measurements
Measurements
Table 3 Adjusted Limits Based on 3-Pole Filter Output RF Spectrum Due to
Modulation
Power Control Level
0 (43 dBm) 0 dB 0.5 dB -27.5 dB -32 dB -51 dB -70 dB 0 (39 dBm) 0 0.5 -27.5 -32 -51 -66 0 (37 dBm) 0 0.5 -27.5 -32 -51 -64 5
(33 dBm)
Table 4 Adjusted Limits Based on 3-Pole Filter Output RF Spectrum Due to Switching
PowerLevel(dBm) Offset from Carrier (kHZ)
43 -3 dB -17 dB -19 dB -24 dB 41 -5 -17 -19 -24 39 -7 -17 -19 -24 37 -9 -17 -19 -24
0 100 200 250 400 600 to
0 0.5 -27.5 -32 -51 -60
400 600 1200 1800
Offset from Carrier (kHZ)
1800
35 -11 -17 -19 -24 33 -13 -17 -19 -24 31 -15 -19 -21 -26 29 -17 -21 -23 -28 27 -17 -22 -25 -30 25 -17 -22 -27 -32 23 -17 -22 -29 -34 21 -17 -22 -30 -36
2-23
Making Measurements
Measurements
Spectrum Analyzer (Available if Option 006, Spectrum Analyzer, is Fitted)
NOTE The spectrum analyzer can detect very low power signals where Active Cell and
NOTE If both the input and output are set to RF IN/OUT, large errors in measured signal
The spectrum analyzer allows you to view the mobile phone’s signal over a wide dynamic range. It also allows you to view any adjacent interference which may exist.
even Test Mode cannot operate. Broken cable or connectors can be found using this function.
Method
Select SPEC ANL in the cell control screen. Selecting (1) provides access to other functions of the spectrum analyzer. These are:
Main - the default screen has basic spectrum analyzer functions, see Figure 2-12 on page 2-25.
RF Gen - this controls the RF generator in the Agilent 8922M/S. The generator signal can be fed back into the spectrum analyzer for signal confirmation. The Aux RF OUT port should be connected to the RF In/Out, and the RF Output set to “Aux RF OUT”.
level will be seen.
Marker - this screen allows you to control the marker for the input signal.
Auxiliary - the auxiliary screen controls theinputs and the attenuator settings.
2-24
Making Measurements
Measurements
NOTE The RF Generator and spectrum analyzer can be tuned to different frequencies
allowing the inspection of the IF signals inside the mobile phone.
1
2
Figure 2-12 Spectrum Analyzer Measurement
Use the MeasReset (2) to reset the trace and measurements in the Agilent 8922M/ S. This is useful if you are using the Max Hold field (3).
The resolution bandwidth of the spectrum analyzer is automatically coupled to the frequency span. Reducing the span to less than 200 kHz can be useful when checking the mobile phone’s transmitter.
3
2-25
Making Measurements
Measurements
Scope The oscilloscope function of the Agilent 8922M/S allows you to view the
demodulated signal from the mobile phone. This can be used for fault-finding in the audio path.
Select SCOPE in the cell control screen to gain access to the oscilloscope function of the Agilent 8922M/S.
Selecting (1), see Figure 2-13 on page 2-26, gives access to other functions of the oscilloscope. These are:
Main - gives you the basic oscilloscope functions for viewing and measuring the trace.
Trigger - allows you to trigger the input signal from a variety of sources.
Marker - offers you the capability to measure the voltage at discrete points in time along the trace. The result is shown in the top right-hand corner of the screen. You may find it helpful to use the , , (HOLD) feature when using the marker.
SHIFT
PREV
1
Figure 2-13 Oscilloscope Measurements
Use the MeasReset (2) to reset the trace and measurements in the Agilent 8922M/ S. This is useful for single triggered measurements.
Set AF Anl In (3) to change the source of the signal being directed to the Audio Analyzer Input. Refer to Screens, Chapter 4 for further information.
2-26
2
3
Making Measurements
Measurements
Audio The audio function measures the audio frequency and voltage of the demodulated
signal from the mobile phone or from a number of other sources selectable using AF Anl In (5), see Figure 2-14 on page 2-27.
Select AUDIO in the cell control screen to gain access to the audio function of the Agilent 8922M/S.
The measurements of the audio voltage and frequency commence within a short time of the screen being accessed.
The ac voltage (1) and the audio frequency (2) are the two main measurements made in this screen. Additional measurements, such as DC Level, can be made by selecting the AF Freq field. (2). The audio frequency generator (3) is used to generate an audio signal from dc up to 25 kHz. The AF analyzer fields sets the conditions for measuring the audio signal (4).
1
3
Figure 2-14 Audio Measurements
2
4
5
2-27
Making Measurements
Measurements
CW Measurement The CW Measurement screen displays the carrier frequency and power of a
continuous (non-pulsed) signal. The CW Power measurement offers a greater dynamic range than is available when making pulsed measurements.
CW Power is a broadband measurement. The CW Frequency measurement is obtained using a tuned, selective input. The RF analyzer should be set to within 500 kHz of the expected signal frequency.
In the GSM band, the value that is entered should be set to the nearest 100 kHz. Select CW MEAS in the cell control screen to gain access to the power measurements of the Agilent 8922M/S.
The CW measurements are made automatically when you access the screen. The CW Frequency field (1), see Figure 2-15 on page 2-28, displays the carrier
frequency of the signal and CW Power (2) shows the power of the carrier signal. It is also possible to display the difference between the measured frequency and the value selected by choosing CWFreqErr from field 1.
1
Figure 2-15 CW Measurements
These measurements can be reset using MeasReset. Select Pwr Zero before power is applied to zero the power meter in the Agilent 8922M/S.
2
2-28
Making Measurements
Measurements
NOTE Although CW Measurements is a broadband measurement, it uses calibration data
that relies on the expected input frequency being set correctly. The Power Detector is connected so that it will only make measurements on signals present at the RF In/ Out port.
2-29
Making Measurements
If You Have Problems with a Measurement
If You Have Problems with a Measurement
This section tells you what to do if either of the following screen display events occurs:
Message Line Messages (on the top of the screen).
Is a Message Line displayed at the top of the screen.
Possible Solutions to Message Line Errors.
Sync Status Messages
Is an Error Message Displayed in the Sync Status Field.
Refer to Chapter 7, “Messages”, for more information.
Is a Message Line displayed at the top of the screen?
At position 1, see Figure 2-16 on page 2-31, a message line appears if:
Attempt To Arm or Query an Inactive Measurement - check that a measurement has been selected from the To Screen.
Trigger too late, decrease trigger delay - See Trigger Timing B
Trigger too early, increase trigger delay - See Trigger Timing B
Measurement Armed, awaiting trigger - See Trigger Timing A
2-30
Making Measurements
If You Have Problems with a Measurement
Figure 2-16
2-31
Making Measurements
If You Have Problems with a Measurement
Possible Solutions to MessageLine Errors
The signal processing hardware that is used to generate measurement results has to be told when to take data samples. If the trigger to do this is misaligned with the signal, there may be errors in the results that are displayed.
Trigger Timing A
Check if the trigger is being received.
On the MEAS SYNC screen (press to gain access), check if the correct burst type has been defined.
Trigger Timing B
Check if the trigger is being received at the correct time.
The First Bit field (A) , see Figure 2-17 on page 2-32, on the Phase Freq: Data Bits screen displays the time difference between when a trigger is being received and when the first bit of a burst occurred. The time difference is only valid if FMErrCount is 0. The timing difference can be corrected by altering the trigger delay (B).
MEAS SYNC
A
Figure 2-17
B
2-32
Trigger Range For Pulsed RF
Making Measurements
If You Have Problems with a Measurement
2-33
Making Measurements
If You Have Problems with a Measurement
Is an Error Message Displayed in the Sync Status field?
The sync status field displays an error message for the following errors:
Bad Sync - demodulation error, perform a Phase/Frequency error measurement to identify which of the Sync Status error listed below may be the possible problem.
FM Errors - see solutions 1, 2, 3, 4
Short Burst - see solutions 1, 4
Level Late - see solutions 1, 2, 3, 4
Level Short - see solutions 1, 2, 3, 4
Low Level- see solution 3
RF Ovrload - see solution 3
2-34
Making Measurements
If You Have Problems with a Measurement
Possible Solutions to Sync Status Errors
Solution 1 - Trigger Timing
Check if the trigger is being received. On the MEAS SYNC screen (press to gain access), check if the correct burst type has been defined. On the Data Bits screen:
A The First Bit field on the Phase Freq:Data Bits screen displays the time
difference between when a trigger is being received and when the first bit of a burst occurred (A). The time difference is only valid if FMErrCount is 0.
B Check the delay in the Trig Delay field (B).
MEAS SYNC
A
Figure 2-18
B
2-35
Making Measurements
If You Have Problems with a Measurement
Solution 2 - Midamble Sync
On the MEAS SYNC screen (press to gain access),
Check the definition of the signal’s burst type (A).
Check the Burst Sel field (B).
MEAS SYNC
A
B
Check the bit pattern of your measurement
Perform a Data Bits measurement (D). An “M” will display under the bits that are identified as the midamble bits.
D
2-36
Making Measurements
If You Have Problems with a Measurement
Solution 3 - Level
Check the following fields on the RF Generator/RF Analyzer screen (press
SHIFT CELL CNTL
3 Pulse field (a), (if signal is pulsed) Ext or Hop Trig should be selected. 4 Amplitude field (b), for the expected amplitude 5 Frequency field (c), for the correct frequency. 6 RF Input field (d), for the correct connector choice. 7 AGC Mode field (e). If either Open or Auto is the selected mode, check the
Check the actual amplitude of the input signal. Perform a Peak Carrier Power measurement.
, , (RFG/RFA) to gain access).
value in the Open Loop DAC (f) Value field. If Closed is the selected mode, check that the burst is repetitive (at least one timeslot every two frames).
a
b
c
d
e
f
2-37
Making Measurements
If You Have Problems with a Measurement
Solution 4 - Amplitude Envelope
Check if the Pulse Amplitude is ± 1 dB of the expected value during the useful part of the burst.
If FM Errors:
Perform a Phase Freq:Data Bits measurement. Dashes (a) will display under the bits where the power is too low.
If No FM Errors:
Perform pulse demodulation measurements. Connect PULSE (DEMODULATION OUT) connector to SCOPE IN (MEASURE) connector on the front panel.
a
---------
2-38
Making Measurements
Advanced Features
Advanced Features
Other Screens The screens mentioned in this section are not necessary for simple measurement of
mobile phones as the work is done by the functions mentioned previously. However, the advanced user may find it worthwhile to know what these screens are and what they are capable of doing.
RF Generator/RF Analyzer - this screen controls the Agilent 8922M/S RF
generator and RF analyzer. Thisscreen can be accessed bypressing the following keys:
SHIFT CELL CNTL
, , (RFG/RFA)
MS Information - the MS INFO (Mobile Subscriber Information) screen gives
you access to the information about the mobile phone itself. A Location Update can be verified from this screen. This screen can be accessed by pressing:
SHIFT CELL CONFIG
, , (MS INFO)
Cell Configuration - the CELL CONFIG screen gives you access to the control
parameters for the configuration of the cell. This screen can be accessed by pressing:
CELL CONFIG
CellControl2 - gives you the capabilityto control the additional cell parameters.
This screen is relevant in the set up of a call. This screen can be accessed by highlighting the More field in any main screen and selecting CELL CNTL2.
Hopping - the hopping screen allows you manual control of the hopping
frequencies. This feature is only available in the Agilent 8922G compatibility mode.
Logging - the logging screen allows control over the Protocol Interface port on
the rear panel (option 003). To access this screen, highlight More and select LOGGING.
Configuration - the configuration screen is a full instrument control screen
which allows you to set up the Agilent 8922M/S. Screen set up, compatibility mode, radio type as well as other miscellaneous pieces of information are contained on this screen. To access this screen, highlight More and select CONFIG.
I/OConfiguration - the I/Oconfiguration screen allows you to set up the general
interface of the instrument, such as printer options. This screen can be accessed by highlighting the More field in any main screen and selecting I/O CONFIG.
2-39
Making Measurements
Advanced Features
SMS Cell Broadcast - the SMS CB screen allows you to test whether the mobile phone is capable of correctlyreceiving and displaying a message sent from a base station using the Short Message Service (SMS) Cell Broadcast protocol. This screen can be accessed by highlighting the More field in any main screen and selecting SMS CB.
Using RF Rise Triggering
Effect of the RF Generator on RF Analyzer Measurements
When using RF Rise triggering of measurements, the Trigger Delay should be adjusted so that the wanted data is captured during the sampling process. The actual measurement is made at the Trigger Delay time after the RF Rise. When changing from RF Rise triggering to another trigger mode, the Trigger Delay should be adjusted to the appropriate value as the default Trigger Delay is not automatically restored.
Example: In the Agilent 8922M/S, when using the RF Rise, a Trig-
ger Delay of 0 is recommended. When changing from RF Rise to Ext Demod trigger mode, the default value of
473.4T should be re-entered for Trigger Delay. The value 473.4T is derived from the 468.75 bits of delay
between downlink and uplink plus the 4.7 bits of modula­tion delay inside the Agilent 8922M/S. All timing is refer­enced to the beginning of bit zero on the downlink TCH timeslot, (DCH or BCCH timeslot if TCH is not active.)
When the RF Generator is set to a high level relative to the analyzer RF input, some error may be introduced in the analyzer’s power detector. This is due to the isolation limits between generator and analyzer on the RF IN/OUT port. This only occurs when the RF Generator and RF Analyzer share the RF IN/OUT port. If the RF Generator uses the AUX RF OUT port, no error occurs.

Error 10LOG 10
   
RFin 20

--------------------------

10
RFgen 14+

-------------------------------

+
10
10
RFin 20() dBm=
RFin is the amplitude in dBm of the signal into the RF IN/OUT port. Normally the setting in dBm of the RF Analyzer Amplitude. RFgen is the setting in dBm of the RF Generator.
Because the two components are not always in phase, the actual error will often be slightly less.
2-40
Making Measurements
Advanced Features
Power Ramp Setting Time
Peak Carrier Power Measurement, Assumptions And Algorithm
The mobile transmitter output ramp should settle so that it complies with the power mask (GSM Rec 11.10) by the time output level calibration is done in the manufacturing process. The power mask is displayed in the Power Ramp screens.
Peak Carrier Power measurement method is not made to GSM Rec. 11.10. If you need to make a measurement to this specification, go to the Power Ramp Summary screen. The field in the top right-hand corner displays the Peak Carrier Power to GSM Rec. 11.10. Refer to PWR RAMP: SUMMARY screen (item 5) for a description.
The algorithm for the Peak Carrier Power measurement is based on the use of a fast sampler taking continuous asynchronous measurements of the power detected at the RF IN/OUT port.
The sampler accumulates 75µS sample bins containing the maximum power observed during the period and the minimum power observed during the period. Sampling rate is 10 MHz. The maximum and minimum values observed are based on 750 individual samples within the period.
A trigger is provided at the middle of bit 0 of the burst. (When using RF Rise triggering, the actual trigger may occur several bits earlier due to the ramp up of the burst.) In response to this trigger, the following 6 sample bins (0 - 5) from the sampler are saved and reported to the measurement processor.
Timeslot
0µS
577µS
Mobile Burst
Amplitude
Trigger
Sample Bins
-2 -1 0 1 2 3 4 5 6 7 8 9 375µS Window
75µS/sample_bin
2-41
Making Measurements
Advanced Features
The beginning of the 0 sample bin will be between 0 and 75µS (bit 0 - 20) after the trigger as the sampler runs asynchronously to the trigger. The last sample bin will complete sampling between 450µS and 525µS (bit 121 - 142) into the burst.
The measurement processor discards the 0 sample bin. This effectively removes the ramp-up overshoot. Each measurement is also RF Power Qualified to ensure only valid bursts are measured. Measurements with a result below the qualification level are discarded.
The remaining sample bins 1-5 are averaged for the displayed measurement result. Each measurement is therefore based on a 375µS (102 bit) sampling of the burst.
A mobile which does not meet the amplitude mask and has erratic amplitude across the burst could give erroneous results with this measurement algorithm. It is therefore essential that mobiles first comply with the amplitude mask in GSM Rec. 11.10.
Making Synchronized External MeasurementsUsing Triggers From The Agilent 8922M/S
There are two ways of making a synchronized external measurement using triggers from the Agilent 8922M/S:
1 Use the G_EXT_TRIG on the rear panel SYSTEM_BUS connector. See
Chapter 6, Connectors for a pin out diagram. This is a separate trigger directly from the Protocol Processor. It is only
selectable from the SERVICE screen. Once set it will provide continuous triggers on each frame. Caution must be used as this trigger can have up to a
0.75 bit error due to the 156/157 bit variance in timeslots. (GSM Recommendations allow use of 156/157 bit timeslots instead of the
156.25 bit length.) To use the G_EXT_TRIG signal:
Go to the CONFIG screen. Select the SERVICE field on the right hand side of the screen. Select the LATCH field and find the selection labeled: ‘g_ext_trig_enable’ (SERV:LATCH:SEL ‘g_ext_trig_enable’) Enter a ‘1’ in the VALUE field. (SERV:LATCH:VAL 1) Select the LATCH field and find the selection labeled: ‘g_ext_trig’ (SERV:LATCH:SEL ‘g_ext_trig’) Enter the TIMESLOT and BITPOSITION on which you want the trigger to occur.
Example: 432 hex is TIMESLOT=4 BITPOSITION=32 hex. The
maximum TIMESLOT=7 and the maximum BITPOSI­TION=9C hex (Bit 0 - 157).
2-42
Making Measurements
Advanced Features
NOTE The GPIB requires the use of base 10 values, 432h = 1074d, for example:
SERV:LATCH:VAL 1074.
The value entered is relative to the beginning of the zero bit of the zero slot on the downlink baseband. A 4.7 bit modulator delay occurs between the baseband and the RF domain which must also be taken into account. Since the value may only be integer, round to the nearest whole value. The following timing error is introduced due to 156/157 bit timing.
Timeslot=0,4 No Error Timeslot=1,5 +.75 bit error (early trigger) Timeslot=2,6 +.50 bit error Timeslot=3,7 +.25 bit error
2 Use MEAS_TRIG_OUT on the rear panel SYSTEM_BUS connector.
This is the delayed trigger from the Measurement System. In the normal ACTIVATED operation, the Protocol Processor sends triggers to the Measurement System on the zero bit of each downlink TCH burst. The Measurement System applies the ‘Meas Trig’ ‘Trigger Delay’ to this input and then, after the elapsed time, begins making its measurement. After this elapsed time, the Measurement System also sends a trigger transition to the MEAS_TRIG_OUT signal on the rear panel SYSTEM_BUS connector.
The trigger is only generated when a measurement is being made. This signal may be used to trigger an externally connected measurement device such as the HP/Agilent 71150/250 GSM high performance spectrum analyzers.
2-43
Making Measurements
Advanced Features
Unexpected Operations
These are descriptions of operations which may be unclear to the user as to how they occurred. These are NOT defects.
1 The Agilent 8922M/S occasionally sends ABORTS to the attached protocol mon-
itor. These ABORTS are normal and logging data is not lost.
2 In the FA and SD/4+FA control channel configurations, the Agilent 8922M/S
sends speech frames rather than fill frames on the FACCH when there is nothing else to send.
3 The DRX ON mode corresponds to the NORMAL paging mode. The
Agilent 8922M/S sends continuous paging mode NORMAL on the idle PCH blocks. The DRX OFF mode corresponds to the REORGANIZE paging mode. The Agilent 8922M/S sends continuous paging mode REORGANIZE on the idle PCH blocks. When in the REORGANIZE paging mode, the Agilent 8922M/S is guaranteed to page on a random PCH block which is not the block expected in the NORMAL paging mode. The active PCH block will page the MS with the paging mode set to SAME. The SAME mode means “the same paging mode as last indicated.”
4 TMSI reallocation in the Agilent 8922M/S uses the two (2) octet TMSI size. 5 The Agilent 8922M/S does not turn off the transmitter during DTX when the
voice activity detector (VAD) determines that there is no speech input present. Instead, the Agilent 8922M/S sends a silence descriptor (SID) on every speech frame.
6 USE MEM is not intended to be used with Pulse (Pulse ON/OFF Ratio)
measurements - do not select USE MEM after making a pulse measurement.
7 During a Pulse (Pulse ON/OFF Ratio) measurement the following message can
occur if the device under test is generating a larger signal than is expected: ‘RF Overpower occurred during Pulse On/Off Ratio measurement’. This message cannot be queried via GPIB. Use the Hardware 1 Status Register Bit 5 to detect this condition when doing pulse measurements.
NOTE Option 006 (spectrum analyzer) must be installed to make pulse measurements.
8 The Output RF Spectrum measurement and Pulse On/Off Rise and Fall
measurement results shown on the Main view screens are not accessible via GPIB or IBASIC. These measurement results are available on the Trace view screens. All GPIB and IBASIC actions with these fields refer to those fields shown on the Trace view screens.
2-44
Making Measurements
Advanced Features
9 The RF Generator may become uncalibrated in a hopping situation when the last
CW frequency setting was outside the GSM or DCS band. If you are using the RF Generator outside the normal GSM or DCS bands, make sure you set a valid GSM or DCS frequency in the CW frequency before entering a hopping mode. This includes the ACTIVE CELL (ACTIVATED state on the CELL CONFIG 2 screen) on the Agilent 8922M/S.
10 When using RF Rise triggering it is recommended to set MS_Parms:Timing
Advance to MANUAL with the Timing Advance set to zero. If AUTO timing advance is used with RF Rise triggering, the Agilent 8922M/S, operating as a base station, will progressively command the mobile to advance timing until it is at the maximum value of 64. It is difficult to recover from this state without ending the call with the mobile. Changing back to EXT_DEMOD triggering will cause the call to drop. This is because the Agilent 8922M/S is listening for the mobile 64 bits later than the RF Rise triggering left it.
11 When making measurements over the GPIB, it is often necessary to place time-
out constraints in software to account for mobile failure. When a time-out is encountered while a measurement is armed but not complete, it is necessary to perform a “CLEAR” operation on the GPIB. A “RESET” operation is insufficient.
Example
ASSIGN @gpib TO 714 OUTPUT @gpib;”TRIGger:ASTate ’ARM’” ON TIMEOUT 7,5 GOTO Flag1 OUTPUT @gpib;”MEASure:DSP:PHASe:FREQuency?” ENTER @gpib;The_answer PRINT ‘FREQUENCY:’; The_answer OFF TIMEOUT 7 GOTO Flag2 Flag1:OFF TIMEOUT 7
CLEAR @gpib PRINT “Timeout Occurred” OUTPUT @gpib;”TRIGger:ASTate ‘DISARM’”
Flag2:END
2-45
Making Measurements
Advanced Features
12 Some measurement results may be displayed as ---- due, for example, to
mobile failure. The measurement has been armed but has not completed. In these circumstances a GPIB query of the measurement will not return a result. This event should be catered for by adding a time-out to the query and a “CLEAR” operation as described in paragraph 11, above.
13 When changing between screens, the DSP results may briefly show a very large
number. This is expected behavior. When controlling the instrument via GPIB, some phase and frequency results
may consistently return a very large number. This is most likely caused by attempting to read result from another DSP screen when measurement triggering is set to Continuous. The alternative solutions are either to change to the appropriate DSP screen before reading the result, or to use single-shot measurement triggering.
2-46
Making Measurements
Advanced Features
2-47
3
Verifying Performance
3-1
Verifying Performance
About This Chapter
About This Chapter
The tests in this chapter verify the electrical performance of the Agilent 8922M/S GSM Test Set using the Agilent 8922 Performance TestSoftware provided with the product. If the instrument passes this verification, its operation and specifications are assured within the measurement uncertainties provided in the performance test print out. Because of the specialized nature of the Agilent 8922M/S and the equipment required to support it, it is recommended that calibration and repair be performed only by specially equipped Agilent Technologies service centers.
A list of specifications is found at the end of this chapter.
3-2
Setting up the Tests
This chapter contains the following information:
Verifying Performance
Setting up the Tests
Getting the Right Test Equipment
Installing and Operating the Software
Required Test Equipment lists the test equipment needed for the performance
tests. This is the only equipment supported by the Agilent 8922 Performance Test Software and is required to verify instrument operation. Equipment substitutions or manual performance tests are not recommended or supported by Agilent Technologies.
Performance Test Software describes how to install and operate the Agilent 8922 Performance Tests software.
3-3
Verifying Performance
Getting the Right Equipment
Getting the Right Equipment
The following equipment is required to do all of the performance tests. The test descriptions have an equipment list that specifies the equipment used for each particular test.
Equipment HP/Agilent Model Number
Measuring Receiver 8902A Sensor Module 11722A Audio Analyzer 8903B Voltmeter 3456A Signal Generator 8657A/B Opt. 022 Multifunction Synthesizer 8904A Opt. 001/002 Pulse Generator 8116A Spectrum Analyzer 8566B Technical Computer with Rocky Mountain
BASIC and DS Disk Drive GPIB/RS-232 printer HP 2225A
HP9000 Series 200/300
3-4
Verifying Performance
Installing and Operating the Software
Installing and Operating the Software
PerformanceTest Software is supplied on a 3.5-inch, double-sided floppy disk and
is written to run with BASIC 5.0 and later. Modifications to the program should be limited to changing the default addresses and storing copies for back-up purposes.
Understanding the Tests
To Load the Program.
Using the Compatibility Switch
Test Descriptions contains a description of each test that is performed by the
Performance Test software. This description is intended to help locate problems if the software fails to execute properly or to help users understand the test methodology that is used in each performance test. The descriptions are not step by step procedures for manual performance tests.
To verify the performance of the Agilent 8922M, you need to convert the instrument back from an Agilent 8922G, or convertthe Agilent 8922S to an Agilent 8922E. To do this refer to the following section titled “Using the Compatibility Switch - Back Conversion”.
You are now ready to run the Performance Test Software.
1 Put the disk in the disk drive. 2 Type LOAD “PT_8922”, press .
After you have completed the Performance Tests, return the instrument back to the Agilent 8922M or Agilent 8922S by following the instructions in the following section titled “Using the Compatibility Switch - Forward Conversion”.
Back Conversion
To turn the instrument from the Agilent 8922M/S back to an Agilent 8922G or an Agilent 8922E, select the following keys:
CONFIG (this is accessible from the Cell Control screen in the bottom right-hand corner).
Compatible, select 8922G or 8922E
HP-IB Adrs (22)
PRESET
ENTER
The instrument is now set up as an Agilent 8922G or Agilent 8922E and ready for Performance Verification testing.
3-5
Verifying Performance
Installing and Operating the Software
Forward Conversion
To return the instrument from an Agilent 8922G back to an Agilent 8922M or an Agilent 8922E to an Agilent 8922S, select the following keys:
More (this is accessible from the Cell Control screen in the bottom right-hand corner). Scroll down the list and select CONFIG.
Compatible, select 8922M or 8922S
HP-IB Adrs (14)
PRESET
The instrument is returned to an Agilent 8922M or Agilent 8922S.
To Configure the GPIB Addresses
1 With the program loaded, type EDIT DEFAULT_ADDRESS, press . 2 Modify each line to indicate the proper instrument address (700-730).
It is now possible to re-store the program as “PT_8922” or store it under a different name.
To Run the Program 1 Type RUN, press .
ENTER
2 Follow the directions as they appear on the screen.
Notes on Running the Program.
The first screen which appears is the GPIB status of each piece of test equipment that is supported. It is only necessary to have the instruments responding that will be used in each particular test. Make certain that each instrument you will be using is responding at the proper address. Duplicate addresses may make an instrument appear to be responding but this is not allowed. Press “I” (for Ignore) to continue past this screen.
The second screen prompts you for the instrument model. If you have disk 08922-10006, select HP 8922G (for Agilent 8922M performance testing) or HP 8922E (for Agilent 8922S performance testing). The third screen which will appear is the main Performance Tests selection menu. Three options are available on this screen:
Select the performance test to run, remember the test instruments and UUT must be responding over GPIB.
ENTER
Turn the printer function ON or OFF. If the printer function is turned on it must be responding over GPIB or the program will lock up.
Exit from the program.
Press the key corresponding to the option that you would like to perform. The other screens that appear are connection instructions, error messages and output results.
3-6
Verifying Performance
Understanding the Tests
Understanding the Tests
This section describes the theory of each performance test, lists the equipment needed for the test, and provides some problem solving information.
Test 01: Signal Generator Level
Equipment Required
Measuring Receiver HP/Agilent 8902A Sensor Module HP/Agilent 11722A
Theory of the Test
The UUT is set to generate CW signals at various levels and frequencies. The HP/ Agilent 8902A/11722A is used as a power meter to measure high level signals and the HP/Agilent 8902A tuned RF level function is used to measure low level signals.
Things To Check In Case Of Problems
The HP/Agilent 8902A/11722A must be calibrated and zeroed for accurate measurementresults.The HP/Agilent 11722A calibration factors must be entered into the HP/Agilent 8902A.
Low level signals from the Agilent 8922M/S must be very stable to be measured by the tuned RF level function of the HP/Agilent 8902A. The instrument timebases should be locked together. Also, if the Agilent 8922 source has unusually high residual FM it may cause the low level measurements to fail because the HP/Agilent 8902A cannot lock onto the signal.
Mismatch uncertainty causes a high level of uncertainty in this measurement.
3-7
Verifying Performance
Understanding the Tests
Test 02: Signal Generator Spectral Purity
Test 03: Signal Pulse Modulation
Equipment Required
HP/Agilent 8566B Spectrum Analyzer
Theory of the Test
The UUT is set to generate a CW signal at various levels and frequencies. The HP/ Agilent 8566B is used to measure the signal level and then the level of the harmonics or spurious signals.
Things To Check In Case Of Problems
The instruments should have their timebases locked together to assure that the frequency offsets for the spurious measurements are accurate.
Equipment Required (excluding Agilent 8922M/S)
HP/Agilent 8116APulse Generator
Theory of the Test
The HP/Agilent 8116A pulse generator drives the pulse input of the UUT and causes it to generate pulsed RF signals. These signals are detected with the negative diode detector and analyzed with the oscilloscope. The UUT is set to generate two different on/off ratios: 30 dB and 80 to 90 dB. Rise time and fall time for each of these is measured.
When the Agilent 8922S is the UUT, the RF Generator is set to generate a pulsed GMSK signal. This signal is then analyzed with the UUT detector and oscilloscope.
Things To Check In Case Of Problems
The oscilloscope is used to automatically measure rise and fall times. If the RF signal has high video feedthrough, it may cause the oscilloscope to make a faulty reading by triggering on the video feedthrough instead of the actual pulse.
A diode detector with a greatly different sensitivity may cause the demodulated waveform to overrange on the oscilloscope or have reduced accuracy because of the resolution of the oscilloscope.
3-8
Verifying Performance
Understanding the Tests
Test 04: Signal Generator 0.3 GMSK Modulation
Equipment Required (excluding Agilent 8922S)
HP/Agilent 8904A Option 001/002 Multifunction Synthesizer
Theory of the Test
The HP/Agilent 8904A is used to generate a 270.833 kHz clock and random data. This drives the UUT RF generator to generate the 0.3 GMSK modulated signals with random data. This signal is then analyzed by the UUT to assure that the frequency and phase errors are correct.
If the UUT is an Agilent 8922S, clock and data signals are generated internally,and the HP/Agilent 8904A is not used.
Things To Check In Case Of Problems
This test should only be performed if the UUT analyzer is known to be accurate. Run test #9 first if there is any question about the accuracy of the UUT analyzer.
Even when the accuracy of the UUT analyzer is known, its specification is such that this test still has a high degree of measurement uncertainty.
A possibility to verify an instrument that fails this test may be to use a second Agilent 8922M/S to analyze the signal or use the HP/Agilent 11836A 0.3 GMSK measurement software. This software (and its associated hardware) has a lower measurement uncertainty but is not currently supported with the Agilent 8922 Performance Tests software.
Test 05: Audio Frequency Generator
Equipment Required
HP/Agilent 3456AVoltmeter HP/Agilent 8903BAudio Analyzer HP/Agilent 8902AMeasuring Receiver
Theory of the Test
The UUT is set to generate audio signals at various levels and frequencies. The voltmeter measures AC and DC levelaccuracy, the audio analyzer measures residual distortion, and the measuring receiver measures frequency accuracy.
Things To Check In Case Of Problems
The measuring receiver and the UUT should have their timebases locked to make accurate frequency measurements.
3-9
Verifying Performance
Understanding the Tests
Test 06: Audio Frequency Analyzer
Equipment Required
HP/Agilent 8904A Option 001/002Multifunction Synthesizer HP/Agilent 3456AVoltmeter
Theory of the Test
The HP/Agilent 8904A is used to generate accurate test signals which are analyzed by the UUT.The voltmeter is used to reduce measurement uncertainty by accurately characterizing the test signal level to predict the correct response of the UUT. The voltmeter is used in the AC level, DC level, SINAD and distortion measurements. Distortion test signals are generated by summing two sinewaves in the HP/Agilent 8904A, one as the desired signal, the other as a distortion product.
Things To Check In Case Of Problems
Forfrequency accuracy measurements the HP/Agilent 8904A timebase should be locked to the UUT.
The voltmeter reading is used as a reference to normalize the reading from the UUT. The output “measured” result is the actual reading from the UUT normalized by the actual reading from the voltmeter. If the voltmeter makes a faulty reading, it will normalize the actual UUT response and cause it to appear faulty also. If the front panel reading from the UUT is significantly different from the printed “measured” response, the normalization may be the cause.
Test07: Oscilloscope Equipment Required
HP/Agilent 8903BAudio Analyzer HP/Agilent 3456AVoltmeter
Theory of the Test
The audio analyzer is used to generate CW reference signals which are characterized by the voltmeter and fed into the UUT. The voltmeter reading is used to normalize the reading from the UUT oscilloscope. The oscilloscope measurements are taken using the UUT marker function which is set to the peak of the input sinewave.
Things To Check In Case Of Problems
The voltmeter reading is used as a reference to normalize the reading from the UUT. The output “measured” result is the actual reading from the UUT normalized by the actual reading from the voltmeter. If the voltmeter makes a faulty reading, it will normalize the actual UUT response and cause it to appear
3-10
Verifying Performance
Understanding the Tests
faulty also. If the front panel reading from the UUT is significantly different from the printed “measured” response, the normalization may be the cause. If the oscilloscope triggers incorrectly, the peak search may not find the actual peak response on the display.
Test08:RF Analyzer Level
Test09:RF Analyzer GMSK Modulation
Equipment Required
HP/Agilent 8657A/B Signal Generator HP/Agilent 8902A Measuring Receiver HP/Agilent 11722A Sensor Module
Theory of the Test
The signal generator is used to generate CW signals at various frequencies and power levels. These are accurately characterized by the HP/Agilent 8902A/11722A and used as a reference. The CW signals are then used to test the UUT which is compared to the reference reading taken by the HP/Agilent 8902A/11722A.
Things To Check In Case Of Problems
Mismatch uncertainties cause a high level of measurement uncertainty in this measurement. If the HP/Agilent 11722A or HP/Agilent 8657A/B has a high SWR, the accuracy of the results may be reduced.
Equipment Required
HP/Agilent 8116APulse Generator HP/Agilent 8904A Opt. 001/002Multifunction Synthesizer HP/Agilent 8657A/B Option 022.Signal Generator
Theory of the Test
The HP/Agilent 8904A frequency modulates the GMSK signal generator with a known amount of phase error. This test signal is used to verify the UUT’s ability to measure this known signal. The actual measurements are compared to the predicted to determine the accuracy of the UUT. The HP/Agilent 8904A is also used to create the 270.833 kHz clock signal and the pulse generator is used to create a random data pattern. Polarity of the data signal from the HP/Agilent 8657A/B Option 022 is not important in this measurement.
Things To Check In Case Of Problems
Make certain the option 022 part of the 8657A/B is locked up to the 270.833 kHz
3-11
Verifying Performance
Understanding the Tests
clocksignal. The option does not have GPIBcontrol so the program cannot verify that the option is operational.
Test A: RF Analyzer Pulse Demodulation
Test B: Spectrum Analyzer (Option 006 only)
Equipment Required
HP/Agilent 8657A/B Signal Generator HP/Agilent 8116A Pulse Generator
Theory of the Test
The Pulse generator drives the signal generator to generate Pulsed RF signals at various frequencies. The UUT demodulates these signals and the internal oscilloscope measures the rise and fall time of the demodulated waveform. This test verifies the demodulator rise and fall time specifications.
Things To Check In Case Of Problems
Theoscilloscopemeasurement functions are used to measure the 10% to 90% rise and fall time. If the demodulated waveform has significant video feedthrough or the oscilloscope cannot trigger the measurements properly, an incorrect measurement may result.
Equipment Required
HP/Agilent 8657A/B Signal Generator
Theory of the Test
The signal generator is used to stimulate the UUT spectrum analyzer at various frequencies. At each frequency, the spectrum analyzer is tuned to measure its own image and spurious responses. The level of the signal generator is set to various levels to determine the spectrum analyzer’s log-linearity.
Things To Check In Case Of Problems
The absolute accuracy of the signal generator causes a significant measurement uncertainty in the log linearity measurements. A more accurate measurement is to reduce the signal level using a precision attenuator. This is an acceptable secondary manual test if the UUT fails the automated test.
3-12
Verifying Performance
Understanding Test Failures
Understanding Test Failures
This section is intended to be used in conjunction with the Agilent 8922 Series Assembly Level Repair Guide for assembly level repair and troubleshooting. If a
performance test fails and hardware is suspected, the following table is a guideline to help identify the hardware assemblies most likely to cause each failure. Before replacing an assembly, the memory card diagnostics (supplied with the Assembly Level Repair Guide) should also be run to help verify which assembly is defective.
ASSEMBLIES TESTED (P=Primary S=Secondary)
Performance Test
Sig Gen Level S S P S PSSSS Sig Gen Spectral Purity S S S S SSSSS Sig Gen Pulse Mod S PSPS SSSSS Sig Gen .3 GMSK P SSSS SSSSP AF Generator S S AF Analyzer P P S Oscilloscope S S RF Analyzer Power S P RF Analyzer GMSK Mod P S P S S RF Analyzer Pulse Demod S P S S Spectrum Analyzer S S S P S
A2 A3 A4 A5 A6 A9 A11A12 A13 A14 A15 A16 A17 A18 A19 A23 A24 A25 A26 A27
3-13
Verifying Performance
Agilent Technologies 8922M/S Specifications
Agilent Technologies 8922M/S Specifications
NOTE: Ifyou have the Agilent 8922M/S Option 010 Multi-Band Test System, refer to the
appropriate Agilent 8922 Multi-Band User’s Guide for more information on specifications relevant to dual band testing.
GSM900 and E-GSM900 Functionality
Bit/Frame Error Rate Measurements: Class 1a, 1b, and Class II bits in both raw and residual form.
Also Burst-by-Burst Bit Error Measurement
MS Power Output Level Control: 0 to 19 with RF analyzer auto adjust Broadcast Channel Capability: BCCH + CCCH or BCCH + CCCH + SDCCH/4 Control Channels (SDCCH, FACCH,
SACCH) Call Control Capabilities BS originated call (FS/EFS), MS originated call (FS/EFS),
Traffic Channels TCH (FS/EFS) HSCSD 2x1 and 2x2, 9.6 kbs and 14.4 kbs Timing Auto, manual, uplink-downlink, offset measurement Hopping Two independent, user definable MA tables with offsets Speech Encoding/Decoding Full rate speech Speech Echo Mode User selectable delay 0 to 5 sec on Agilent 8922M, fixed
Measurement Coordination Flexible control of burst type, ARFCN, and timeslot. SACCH MEAS Result RXLEV, RXQUAL, and timing advance. SMS Cell Broadcast Two predefined messages with user definable attributes
BCCH + CCCH, BCCH + CCCH + SDCCH/4, SDCCH/8 (non-hopped), SACCH/FACCH
MS camp on, BS call disconnect, MS call disconnect
delay on Agilent 8922S
(code, update number, language and identifier).
RF Generator
Frequency
Specifications
Range 10 MHz to 1 GHz Resolution 1 Hz Accuracy
Stability Same as reference
Reference accuracy
3-14
± 0.5 Hz
Verifying Performance
Agilent Technologies 8922M/S Specifications
Supplemental Characteristics:
Frequency Overrange Switching Speed
To 1015 MHz with uncalibrated output and modulation 577 µs over the GSM frequency bands in hop mode (refer to 0.3
GMSK modulation specs
Output
RF In/Out Connector
Level Range Level Resolution
Level Accuracy GSM Bands
50 MHz to 1 GHz
1
2
14 to 127 dBm
0.1 dB
±1.0 dB, levels ≥−127dBm,
maximum over range > typically for levels
12 dB ±1.0 dB,
≥−127 dBm while hopping.
±1.5 dB, levels ≥−107 dBm ±2.0 dB, levels ≥−127 dBm
10 MHz to 50 MHz
±2.0 dB, levels ≥−107 dBm ±2.5 dB, levels ≥−127 dBm
Reverse Power
SWR
Aux RF Out Connector
Level Range +4 to Level Resolution
Level Accuracy GSM Bands
2
1
15 Watts continuous 100 Watts for 10 seconds/minute
1.5:1
127 dBm
0.1 dB
±1.0 dB, levels ≥−107 dBm ±1.0 dB, typically for levels≥−107 dBm while hopping.
50 MHz to 1 GHz
±1.5 dB, levels ≥−107 dBm ±2.0 dB, levels ≥−127 dBm
10 MHz to 50 MHz
±2.0 dB, levels ≥−107 dBm ±2.5 dB, levels ≥−127 dBm
Reverse Power SWR 2.0:1, level <
1. GSM frequency bands are 880 to 915 MHz and 925 to 960 Mhz.
2. Level accuracy degrades 0.2 dB when using the RF in/out connector for both RF generator and RF analyzer. In 30dB pulse mode, level accuracy specifications are typical.
200 milliwatts
4 dBm
3-15
Verifying Performance
Agilent Technologies 8922M/S Specifications
Spectral Purity
Spurious Signals (for +1 dBm output level at Aux RF Out or ≤−19 dBm output level at RF In/Out.
Harmonics: < Non-harmonics: <
25 dBc
50 dBc, >5 kHz offset from carrier
0.3 GMSK Modulation
After one timeslot, 577µs, from an isolated RF Generator Trigger in the GSM Frequency bands.
Phase Error Peak Phase Error Frequency Error
Amplitude Flatness Clock Input (Agilent 8922M only)
Frequency 270.833 kHz Level
Data Input (Agilent 8922M only)
Format Level
Supplemental Characteristics:
After three timeslots, 1.73 ms, from an isolated RF Generator Trigger in the GSM frequency bands
Phase Error Peak Phase Error Frequency Error
1° rms4° peak ± [0.02 ppm (18 Hz) + reference accuracy], for normal bursts
Typically
±[0.03 ppm (27 Hz) + reference accuracy], for RACHs.
±0.25 dB peak.
± 2 Hz (relative to reference).
TTL
Non differentially encoded input. TTL
0.5° rms2.0° peak ±[0.01ppm (9 Hz) + reference accuracy] for normal bursts.
±[0.02ppm (18 Hz) + reference accuracy] for RACH bursts.
Pulse Modulation
Input Levels (Agilent 8922M only) TTL
Rise/Fall Time (10% to 90%)
Supplemental Characteristics:
On/Off Ratio
5 µs
>80 dB
3-16
30 dB Pulse Modulation (Agilent 8922M only)
All timeslots 30 dB higher than desired/active timeslot, to test adjacent timeslot rejection.
Supplemental Characteristics: Input Levels TTL Rise/Fall Time (10 to 90%)
5 µs
AM for Level Control (Agilent 8922M Only)
For output levels +1 dBm at Aux RF Out or ≤−19 dBm at RF In/Out
Supplemental Characteristics: Input
Range Impedance Sensitivity Calibration
Rise/Fall Time (10 to 90%)
1.0V to +0.6V
600 nominal, DC Coupled 100% AM per volt, nominal 0 VDC input produces calibrated output from the RF Generator
10 µs
Verifying Performance
Agilent Technologies 8922M/S Specifications
3-17
Verifying Performance
Agilent Technologies 8922M/S Specifications
RF Analyzer
Frequency
Specifications
Range 10 MHz to 1 GHz Resolution 1 Hz Hop Mode
Resolution Offset Frequency
Offset Resolution RF In/Out SWR <1.5:1 Supplemental Characteristics:
Frequency Overrange
Offset Resolution
100 kHz
50 kHz
1 Hz
To 1015 MHz 500 Hz for FM demodulation out
CW RF Frequency Measurement
Range 10 MHz to 1 GHz Level Range
RF In/Out
Aux RF In Input Frequency Setting Error Accuracy Supplemental Characteristics:
Minimum Resolution
6 to +41 dBm
36 to +20 dBm ±500 kHz ±(1 Hz + reference accuracy)
1 Hz
CW RF Power Measurements (RF In/Out Only)
Range -5 to +41 dBm Input Frequency Setting Error
Accuracy Supplemental Characteristics:
Accuracy
Minimum Resolution
±500 kHz ±0.5 dB ± noise effects (+4 to +41 dBm) (0.2 mW)
±0.5 dB ± noise effects (-5 to +4 dBm) (0.2 mW)
0.01 dB
3-18
Peak/Transmitter Carrier Power Measurement
RF In/Out only. After one timeslot, 577 µs, from an isolated Receiver Hop Trigger in the GSM bands.
Range -5 to +41 dBm Input Frequency Setting Error
Input Level Setting Error Accuracy Supplemental Characteristics:
Accuracy Minimum Resolution
±10 kHz ±3 dB ±0.6 dB ± noise effects (+4 to +41 dBm) (0.2 mW)
±0.6 dB ± noise effects (-5 to +4 dBm) (0.2 mW)
0.2 dB
Power Measurement Accuracy
Accuracy (dB)
0.5
0.4
Verifying Performance
Agilent Technologies 8922M/S Specifications
0.3
0.2
0.1
0
Pulse ON/OFF Ratio Measurement (Requires Option 006)
ON power is averaged over the useful part of the burst. OFF is averaged over a one bit interval centered at a user specified time. Non-hopped mode only.
Input Frequency Setting Error Input Level Setting Error Timing Accuracy
-5 0 5 10 15 20 25 30 35
Power (dBm)
±10 kHz ±3 dB ±1.7 µs (± 1.1 µs typical)
3-19
Verifying Performance
Agilent Technologies 8922M/S Specifications
Accuracy (ON/OFF 40 dB, RF In/Out only)
OFF Power (dBm) ON/OFF Ratio Accuracy
30 to 1 ±2.4 dB ±1.1 typically
37 to 30 ±2.9 dB ±1.3 typically
42 to 37 ±3.7 dB ±1.7 typically
47 to 42 ±4.2 dB ±2.1 typically
Amplitude Envelope Measurement
After one timeslot, 577µs, from an isolated Receiver Hop Trigger in the GSM Frequency bands
Measurement Range
RF In/Out
Aux RF In Input Frequency Setting Error
6 to +41 dBm
36 to 20 dBm ±10 kHz
Inaccuracy due to Noise (for overshoots 1 dB):
0 dB
Relative
Level
±1 dB ±3 dB ±3 dB w/5 averages
<
±0.15 dB pk <±0.2 dB pk <±0.2 dB pk
Input Level Setting Error
6 dB <±0.2 dB <±0.3 dB <±0.3 dB
30 dB
<+3.0 dB <+4.2 dB <+2.2 dB
3.8 dB 7.5 dB 2.6 dB
3-20
Verifying Performance
Agilent Technologies 8922M/S Specifications
Phase and Frequency Measurements
After one timeslot, 577µs, from an isolated Receiver Hop Trigger in the GSM Frequency bands.
Range
RF In/Out
Aux RF In Input Frequency Setting Error Input Level Setting Error RMS phase error accuracy Peak phase error accuracy Frequency error accuracy
Supplemental Characteristics
After three timeslots, 1.73 ms, from an isolated Receiver Hop Triggering the GSM frequency band.
RMS phase error accuracy Peak phase error accuracy Frequency error accuracy
6 to +41 dBm
36 to +20 dBm ±10 kHz ±3 dB1° rms4° peak ±[0.02 ppm (18 Hz) + reference accuracy], for normal bursts.
Typically
±[0.03ppm (27 Hz) + reference accuracy] for RACH bursts.
0.5° rms2° peak ±[0.01 ppm (9 Hz) + reference accuracy], for normal bursts.
±[0.02 ppm (18 Hz) + reference accuracy] for RACH bursts.
0.3 GMSK Data Recovery (Agilent 8922M only)
After one timeslot, 577µs, from an isolated Receiver Hop Trigger in the GSM Frequency bands
Range
RF In/Out
Aux RF In Input Frequency Setting Error Required Input Phase Accuracy
Demodulation Duty Cycle 1 timeslot per frame Outputs Data, Clock and Data Valid Data Output Clock Clocked at 1 MHz rate Delay, data
Output Level TTL
6 to +41 dBm
36 to +20 dBm ±100 Hz5°
, 20° peak
rms
1 frame (4.62 ms).
3-21
Verifying Performance
Agilent Technologies 8922M/S Specifications
FM Demodulation Output (Agilent 8922M only)
Range
RF In/Out
Aux RF In Sensitivity Input Frequency Setting Error Input Level Setting Error Supplemental Characteristics
3 dB Bandwidth
Output Impedance
DC Offset
Pulse Demodulation Output (Agilent 8922M only)
Range
RF In/Out
Aux RF In Input Frequency Setting Error Input Level Setting Error Rise Time (10 to 90%) Fall Time (90 to 10%) Supplemental Characteristics:
Output Impedance
Output Level
6 to +41 dBm
36 to +20 dBm
20µV/Hz
±5% (into an open circuit) ±50 kHz, with 100 kHz pk deviation ±3 dB
DC to 270 kHz 600
5 mV
6 to +41 dBm
36 to +20 dBm ±50 kHz ±3 dB2.5 µs2.5 µs
600 , DC Coupled 2 Vpk into an open circuit
3-22
Output RF Spectrum Measurement (Requires Option 006)
After one timeslot, 577µs, from an isolated Receiver Hop Trigger in the GSM Frequency bands.
Range
RF In/Out Aux RF In
Input Levels for Optimum Dynamic Range
RF In/Out
Aux RF In Input Frequency Setting Error Input Level Setting Error Supplemental Characteristics:
Log Linearity
Amplitude Flatness
Amplitude Resolution
Verifying Performance
Agilent Technologies 8922M/S Specifications
6 to +41 dBm
36 to +20 dBm
+7, +17, +27, +37 dBm
23, 13, 3, +7 dBm ±10 kHz, ±3 dB
±0.4 dB ±1.0 dB
0.4 dB
Dynamic Range (dB)
Offset (kHz)
100 200 300 400 600 800 to 1800
Range (dB) 24 42 53 60 63 64
NOTE: When using Output RF Spectrum due to the Ramping measurement, the dynamic
range is decreased by 12 dB (due to Peak Hold).
3-23
Verifying Performance
Agilent Technologies 8922M/S Specifications
Spectrum Analyzer Specifications (Option 006)
Frequency Range 10 MHz to 1 GHz Frequency Span/Resolution
Bandwidth (coupled)
Span Bandwidth
<50 kHz 300 Hz <200 kHz 1 kHz <1.5 MHz 3 kHz
4 MHz
Display Log, 10 dB/div Display Range 80 dB Log Linearity
Reference Level
RF In/Out +44 to Aux RF In +23 to
Non-harmonic Spurious
Responses Residual Responses Image Rejection >50 dB Supplemental Characteristics:
Level Accuracy
Frequency Overrange
Displayed Average Noise Level <
±1.1 dB
50 dBc max, for inputs 30 dBm
<
±2.5 dB
To 1015 MHz
30 kHz
24 dBm
55 dBm
70 dBm (no input signal, 0 dB attenuation)
116 dBm (0 dB attenuation, <50 kHz spans).
3-24
Verifying Performance
Agilent Technologies 8922M/S Specifications
Audio Source
Frequency
Specifications
Range DC to 25 kHz Accuracy 0.025% of setting Supplemental Characteristics
Minimum Resolution
Output Level
Range 0.1 mV to 4 V Maximum Output Current 20 mA peak
Output Impedance <1 Accuracy
Residual Distortion (THD + noise, amplitude >200 mV
Supplemental Characteristics
Minimum Resolution Level
DC Coupled Offset
rms
)
0.1 Hz
rms
±(2% of setting + resolution)
0.1%, 20 Hz to 25 kHz in 80 kHz BW.
0.01V: 50µV
Level
0.1V: 0.5 mV
Level
1V: 5 mV
Level >1V: 50 mV <50 mV
Audio Analyzer
Frequency Measurement
Specifications
Range 20 Hz to 400 kHz Accuracy
External Input 20 mV Supplemental Characteristics
Minimum Resolution
±(0.02% + 1 count + reference accuracy)
to 30 V
rms
f < 10kHz: 0.01 Hz f < 100kHz: 0.1 Hz
100kHz: 1 Hz
f
rms
3-25
Loading...