This Agilent Technologies instrument product is warranted against defects in material and
workmanship for a period of one year from date of shipment. During the warranty period,
Agilent Technologies will at its option, either repair or replace products which prove to be
defective.
For warranty service or repair, this product must be returned to a service facility
designated by Agilent Technologies. Buyer shall prepay shipping charges to Agilent
Technologies and Agilent Technologies shall pay shipping charges, duties, and taxes for
products returned to Agilent Technologies from another country.
Agilent Technologies warrants that its software and firmware designated by Agilent
Technologies for use with an instrument will execute its programming instructions when
properly installed on that instrument. Agilent Technologies does not warrant that the
operation of the instrument, or software, or firmware will be uninterrupted or error free.
Limitation of Warranty
The foregoing warranty shall not apply to defects resulting from improper or inadequate
maintenance by Buyer, Buyer-supplied software or interfacing, unauthorized modification
or misuse, operation outside of the environmental specifications for the product, or
improper site preparation or maintenance.
NO OTHER WARRANTY IS EXPRESSED OR IMPLIED. AGILENT
TECHNOLOGIES SPECIFICALLY DISCLAIMS THE IMPLIED WARRANTIES OF
MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
Limitation of Remedies and Liability
THE REMEDIES PROVIDED HEREIN ARE BUYER’S SOLE AND EXCLUSIVE
REMEDIES. AGILENT TECHNOLOGIES SHALL NOT BE LIABLE FOR ANY
DIRECT, INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES,
WHETHER BASED ON CONTRACT, TORT, OR ANY OTHER LEGAL THEORY.
i
Responsibilities of the Customer
Responsibilities of the Customer
The customer shall provide;
1Access to the products during the specified periods of coverage to perform mainte-
nance.
2Adequate working space around the products for servicing by Agilent Technologies
personnel.
3Access to and use of all information and facilities determined necessary by Agilent
Technologies to service and/or maintain the products. (In so far as these items may
contain proprietary or classified information, the customer shall assume full
responsibility for safeguarding and protection from wrongful use.)
4Routine operator maintenance and cleaning as specified in the Agilent Technologies
Operating and Service Manuals.
5Consumables such as paper, disks, magnetic tapes, ribbons, inks, pens, gases, solvents,
lamps, filters, fuses, seals, etc.
Certification
Agilent Technologies certifies that this product met its published specifications at the time
of shipment from the factory. Agilent Technologies further certifies that its calibration
measurements are traceable to the United States National Bureau of Standards and
Technology, to the extent allowed by the Bureau’s calibration facility, and to the
calibration facilities of other International Standards Organization members.
Assistance
Product maintenance agreements and other customer assistance agreements are available
for Agilent Technologies products.
For any assistance, contact your local Agilent Sales and Service Office. For a list of
contact information, see “Sales and Service Offices” on page ix.
ii
Notices
Notices
The material contained in this document is subject to change without notice. AGILENT
TECHNOLOGIES MAKES NO WARRANTY OF ANY KIND WITH REGARD TO
THIS MATERIAL, INCLUDING, BUT NOT LIMITED TO, THE IMPLIED
WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR
PURPOSE. Agilent Technologies inc. shall not be liable for errors contained herein or for
incidental or consequential damages in connection with the furnishing, performance or use
of this material.
Agilent Technologies assumes no responsibility for the use or reliability of its software on
equipment that is not furnished by Agilent Technologies.
Restricted Rights Legend
If Software is for use in the performance of a U.S. Government prime contract or
subcontract, Software is delivered and licensed as "Commercial computer software" as
defined in DFAR 252.227-7014 (June 1995), or as a "commercial item" as defined in FAR
2.101(a) or as "Restricted computer software" as defined in FAR 52.227-19 (June 1987) or
any equivalent agency regulation or contract clause. Use, duplication or disclosure of
Software is subject to Agilent Technologies’ standard commercial licenseterms, and nonDOD Departments and Agencies of the U.S. Government will receive no greater than
Restricted Rights as defined in FAR 52.227- 19(c)(1-2) (June 1987). U.S. Government
users will receive no greater than Limited Rights as defined in FAR52.227-14 (June 1987)
or DFAR 252.227-7015 (b)(2) (November 1995), as applicable in any technical data.
Copyright 200X Agilent Technologies Inc. All Rights Reserved.
Statement of Compliance
This product conforms to EN61010-1(1993) / IEC 1010-1(1990) +A1(1992) +A2(1994) /
CSA C22.2 No. 1010.1(1993) Safety requirements for Electrical Equipment for
Measurement, Control and Laboratory Use, and has been supplied in a safe condition. The
instruction documentation contains information and warnings which must be followed by
the user to ensure safe operation and to maintain the instrument in a safe condition.
iii
Electromagnetic Compatibility (EMC) Information
Electromagnetic Compatibility (EMC) Information
This product has been designed to meet the protection requirements of the European
Communities Electromagnetic Compatibility (EMC) directive:
EN55011:1991 (Group 1, Class A)
EN50082-1:1992
- IEC 1000-4-2 (1995) ESD
- IEC 1000-4-3 (1995) Radiated Susceptibility
- IEC 1000-4-4 (1995) EFT
In order to preserve the EMC performance of this product, any cable which becomes worn
or damaged, must be replaced with the same type and specification.
Sound Emission
Manufacturer’s Declaration
This statement is provided to comply with the requirements of the German Sound
Emission Directive, from 18 January 1991.
This product has a sound pressure emission (at the operator position) < 70 dB(A).
❒ Sound Pressure Lp < 70 dB(A).
❒ At Operator Position.
❒ Normal Operation.
❒ According to ISO 7779:1988/EN 27779:1991 (Type Test).
Herstellerbescheinigung
Diese Information steht im Zusammenhang mit den Anforderungen der
Maschinenlärminformationsverordnung vom 18 Januar 1991.
❒ Schalldruckpegel Lp < 70 dB(A).
❒ Am Arbeitsplatz.
❒ Normaler Betrieb.
❒ Nach ISO 7779:1988/EN 27779:1991 (Typprfung).
iv
Declaration of Conformity
according to ISO/IEC Guide 22 and EN45014
Manufacturer’s Name:Agilent Technologies.
Declaration of Conformity
Manufacturer’s Address:
Agilent Technologies
South Queensferry
West Lothian, EH30 9TG
Scotland, United Kingdom
Declares that the product
Product Name
:
Model Numbers:
Product Options:
Conforms with the protection requirements of European Council Directive 89/336/EEC on the approximation
of the laws of the member states relating to electromagnetic compatibility.
Against EMC test specifications EN 55011:1991 (Group 1, Class A) and EN 50082-1:1992
As Detailed in:
Assessed by:
Technical Report Number:6893/2200/CBR, dated 23 September 1997
GSM MS Test Set
Agilent Technologies 8922M and 8922S
This declaration covers all options of the above products as detailed in
TCF A-5951-9852-02
Electromagnetic Compatibility (EMC)
Technical Construction File (TCF) No. A-5951-9852-02
Dti Appointed Competent Body
EMC Test Centre, GEC-Marconi Avionics Ltd.,
Maxwell Building, Donibristle Industrial Park,
KY11 5LB
Scotland, United Kingdom
Supplementary Information:
The product conforms to the following
safety standards:
The product herewith complies with the requirements of the Low Voltage Directive 73/23/EEC, and carries the CEmarking accordingly.
South Queensferry, Scotland17 November 1997
LocationDateR.M. Evans / Quality Manager
EN 61010-1(1993) / IEC 1010-1(1990) +A1(1992) +A2(1994)
CSA-C22.2 No. 1010.1-93
EN 60825-1(1994) / IEC 825-1(1993)
v
Safety Information
Safety Information
The following general safety precautions must be observed during all phases of operation
of this instrument. Failure to comply with these precautions or with specific warnings
elsewhere in this manual violates safety standards of design, manufacture, and intended
use of the instrument. Agilent Technologies Inc. assumes no liability for the customer’s
failure to comply with these requirements.
GENERAL
This product is a Safety Class 1 instrument (provided with a protective earth terminal).
The protective features of this product may be impaired if it is used in a manner not
specified in the operation instructions. All Light Emitting Diodes (LEDs) used in this
product are Class 1 LEDs as per IEC 60825-1.
ENVIRONMENTAL CONDITIONS
This instrument is intended for indoor use in an installation category II, pollution degree 2
environment. It is designed to operate at a maximum relative humidity of 95% and at
altitudes of up to 2000 meters. Refer to the specifications tables for the ac mains voltage
requirements and ambient operating temperature range.
BEFORE APPLYING POWER
Verify that the product is set to match the available line voltage, the correct fuse is
installed, and all safety precautions are taken. Note the instrument’s external markings
described under "Safety Symbols".
GROUND THE INSTRUMENT
To minimize shock hazard, the instrument chassis and cover must be connected to an
electrical protective earth ground. The instrument must be connected to the ac power
mains through a grounded power cable, with the ground wire firmly connected to an
electrical ground (safety ground) at the power outlet. Any interruption of the protective
(grounding) conductor or disconnection of the protective earth terminal will cause a
potential shock hazard that could result in personal injury.
FUSES
Only fuses with the required rated current, voltage, and specified type (normal blow, time
delay,etc.) should be used. Do not use repaired fuses or short-circuited fuse holders. To do
so could cause a shock or fire hazard.
vi
Safety Information
DO NOT OPERATE IN AN EXPLOSIVE ATMOSPHERE
Do not operate the instrument in the presence of flammable gases or fumes.
DO NOT REMOVE THE INSTRUMENT COVER
Operating personnel must not remove instrument covers. Component replacement and
internal adjustments must be made only by qualified service personnel.
Instruments that appear damaged or defective should be made inoperative and secured
against unintended operation until they can be repaired by qualified service personnel.
WARNING The WARNING sign denotes a hazard. It calls attention to a procedure,
practice, or the like, which, if not correctly performed or adhered to, could result in
personal injury. Do not proceed beyond a WARNING sign until the indicated conditions
are fully understood and met.
Caution The CAUTION sign denotes a hazard. It calls attention to an operating
procedure, or the like, which, if not correctly performed or adhered to, could result in
damage to or destruction of part or all of the product. Do not proceed beyond a CAUTION
sign until the indicated conditions are fully understood and met.
vii
Safety Symbols
Safety Symbols
The following symbols on the instrument and in the manual indicate precautions which
must be taken to maintain safe operation of the instrument
Safety Symbols
The Instruction Documentation Symbol. The product is marked with this symbol when
it is necessary for the user to refer to the instructions in the supplied documentation.
Indicates the field wiring terminal that must be connected to earth ground before operating the equipment - protects against electrical shock in case of fault.
Frame or chassis ground terminal - typically connects to the equipment's metal frame.
Alternating current (AC)
Direct current (DC)
Warning, risk of electric shock.
N
L
ISM 1-AThis is a symbol of an Industrial, Scientific, and Medical Group 1 Class A product.
Terminal for Neutral conductor on permanently installed equipment.
Terminal for Line conductor on permanently installed equipment..
The CE mark shows that the product complies with all relevant European Legal Directives.
The CSA mark is a registered trademark of the Canadian Standards Association, and
indicates compliance to the standards defined by them.
Indicates that a laser is fitted. The user must refer to the manual for specific Warning or
Caution information to avoid personal injury or damage to the product.
viii
Sales and Service Offices
Any adjustment, maintenance, or repair of this product must
be performed by qualified personnel. Contact your customer
engineer through your local Agilent Technologies Service
Center. You can find a list of local service service
representatives on the web at:
http://www.agilent-tech.com/services/English/index.html
You can also contact one of the following centers and ask for
Agilent Technologies
Test & Measurement
European Marketing Organisation
P.O. Box 999
1180 AZ Amstelveen
The Netherlands
(tel) (31 20) 547 9999
Latin America:
Agilent Technologies
Latin American Region Headquarters
5200 Blue Lagoon Drive, Suite #950
Miami, Florida 33126
U.S.A.
(tel) (305) 267 4245
(fax) (305) 267 4286
United States:
Agilent Technologies
Test and Measurement Call Center
P.O. Box 4026
Englewood, CO 80155-4026
(tel) 1 800 452 488
In any correspondence or telephone conversations, refer to
the power sensor by its model number and full serial number.
With this information, the Agilent Technologies
representativecan quickly determine whether your unit is still
within its warranty period.
This guide gives a brief description on how to make each of the measurements required to
test a GSM mobile phone. More detailed descriptions are given in the Agilent 8922M/S
GSM Test Set User’s Guide.
Agilent 8922M/S GSM Test Set User’s Guide.
This guide contains information on how to set up the Agilent 8922M/S for making
measurements and verifying performance. It also contains more detailed information on
each of the screens, keys, and connectors, and how to use the IBASIC facilities available
on the Agilent 8922M/S.
Agilent 8922M/S GSM Test Set Performance Test Software.
This 3.5 inch floppy disk allows you to verify the performance of the Agilent 8922.
Instructions on how to use this are detailed in the Agilent 8922M/S GSM Test Set User’s
Guide.
Agilent 8922M/S GSM Test Set Programming Reference Guide.
This guide describes, in detail, each of the GPIB command sets for the Agilent 8922M/S.
Agilent 8922 Multi-Band Test System User’s Guide1.
This is a supplementary user’s guide that describes the additional features found with the
Multi-Band test system. The test system is used for testing and making measurements of
dual band mobiles.
1.The user’s guide is only available with the HP 8922M/S Option 010.
xi
Typeface Conventions
Typeface Conventions
Italics
DisplayDisplay text is used to show examples, fields, and prompts that are
PRESET
Soft keysDisplay text is used to show examples, fields, and prompts that are
Soft keysAll software listings in this manual can be identified with this font.
Italic type is used for emphasis.
displayed on the Agilent 8922M/S screen.
Keycaps on the Agilent 8922M/S keyboard are enclosed in boxes.
displayed on the Agilent 8922M/S screen.
xii
1
Installing Your Agilent 8922M/S
1-1
Installing Your Agilent 8922M/S
Using this Chapter
Using this Chapter
Use the following procedure to get the Agilent Technologies 8922M/S powered-up
correctly. After completing this procedure, refer to the Quick Start Guide for an
introduction to operating the Agilent Technologies 8922M/S and Chapter 2,
“Making Measurements” for more extensive information on using the Agilent
Technologies 8922M/S.
Equipment Supplied •Fuse Envelope and Fuse
•REF OUT/REF IN cable (Option 001 only)
•Power Cord
CAUTIONTo avoid potential injury, ensure that two people are employed in lifting the Agilent
8922M/S out of the box and for any other instrument moves.
NOTE:If you have the Agilent 8922M/S Option 010 Multi-Band Test System, refer to the
appropriate Agilent 8922 Multi-Band User’s Guide for more information on
connection and operating differences.
1-2
Installing Your Agilent 8922M/S
Fuses and Power Cords
Fuses and Power Cords
CAUTIONBefore plugging this instrument into the Mains (line) voltage, be sure the correct
voltage on the line voltage selection card has been selected.
Line Voltage and
Fuse Selection
Figure 1-1 Voltage Selection Card and Fuse Installation
Other FusingNon user replaceable fusing on the secondary of this instrument power supply are;
Verify that the line voltage selection card is matched to the power source
(see Figure 1-1 on page 1-3). Order fuse Agilent part 2110-0083 (2.5 A 250 V,
normal blow) for replacement.
1Connect a 1, 2, 5, 10, or 13 MHz signal to the REF IN. If you are using option
001, connect as shown, see Figure 1-2 on page 1-5, with the supplied cable (OPT
001 REF OUT to REF IN).
2Connect the supplied power cord to the Agilent 8922M/S and power up the
instrument.
1
Power Source
Figure 1-2 Rear View Connections
1-5
Installing Your Agilent 8922M/S
Installation Overview
3Access the CONFIG screen to customize the set-up of your Agilent 8922M/S.
This is done by:
•Moving to the CONFIG field in the bottom right-hand corner of the Cell
Control screen, (the first screen that appears after power-up or after selecting
PRESET
). Rotate the cursor control knob (refer to “a”, see Figure 1-3 on
page 1-6) until you are on the CONFIG field.
•Selecting the CONFIG field by pushing the cursor control knob (diagram “b”,
see Figure 1-3 on page 1-6).
Figure 1-3 Access CONFIG Screen
1-6
Installing Your Agilent 8922M/S
Installation Overview
4To guarantee the correct operation of the Agilent 8922M/S with all mobile
phones, you should use a high-stability timebase.
This step details how to configure the Agilent 8922M/S for use with either the
option 001, high-stability timebase, or an external reference signal.
A If you have option 001 installed and wish to use this as your reference, ensure
that the OPT 001 REF OUT field in the Configure screen is set to On.
Access the Reference field in the Configure screen and select 10 MHz.
B If you wish to use an external reference signal, set the Reference field to
1, 2, 5, 10, or 13 MHz according to the frequency reference you intend to use.
A
Figure 1-4 High-stability timebase set-up
To customize your Agilent 8922M/S further, refer to Chapter 4, Screens.
B
1-7
Installing Your Agilent 8922M/S
General Information
General Information
Operation and
Storage
Environment
Instrument OptionsRefer to Agilent 8922M/S Specifications in Chapter 3, Performance Verification for
SpecificationsRefer to Agilent 8922M/S Specifications in Chapter 3, Performance Verification for
Refer to General Specifications in the Agilent 8922M/S Specifications section of
Chapter 3, Performance Verification for information about the operation or storage
environment.
information about instrument options.
instrument specifications.
1-8
2
Making Measurements
2-1
Making Measurements
Using This Chapter
Using This Chapter
Use this chapter to obtain an overview of how to operate the Agilent 8922M/S GSM
Test Sets.
This chapter is divided into the following sections:
•Making Measurements
•Agilent 8922M/S Operating Modes - explains how to configure the Agilent
8922M/S so that measurements can be made.
•Measurements - details how to perform the many different measurements
available on the Agilent 8922M/S.
•Advanced Features - advanced information for experienced users.
To help you understand this chapter, you may wish to consult the following sources
of information:
•Quick Start Guide - This briefly explains the basic operating principles of the
Agilent 8922M/S including making a call and simple measurements.
•Typeface Conventions - Found at the front matter of this manual. This explains
how to interpret the different typefaces used in this manual.
•Field Types - Refer to Chapter 4, Screens. Explains in detail the different kinds
of “fields” or areas which are displayed on the screens. Fields are mainly used for
measurement results, data entry and access to other screens.
•Screens -Chapter 4, provides more detailed information on each of the fields
contained within all of the screens available on the Agilent 8922M/S.
NOTE:If you have the Agilent 8922M/S Option 010 Multi-Band Test System, refer to the
appropriate Agilent 8922 Multi-Band User’s Guide for more information on making
dual band measurements.
2-2
Making Measurements
Agilent Technologies 8922M/S Operating Modes
Agilent Technologies 8922M/S Operating Modes
This section details the procedures necessary to control the Agilent 8922M/S and
GSM mobile phone in each of the main operating modes. It is recommended that
you read this section before attempting the “Measurements” section.
The Agilent 8922M/S GSM TestSets have three main modes of operation, these are:
•ACTIVE CELL
•TEST MODE
•CW GENERATOR
Active CellThe first mode is the default mode and configures the Agilent 8922M/S as a GSM
Base Station emulator. This allows calls to be made from the Agilent 8922M/S to
the mobile phone and vice versa. Measurements can then be made to verify the
mobile phone’s performance.
Test ModeThis mode is used when measurements need to be made on the mobile phone
without a call being set up.
CW GeneratorThis mode configures the Agilent 8922M/S as a standard Continuous Wave (CW)
Signal Generator. This is used for test applications that require an unmodulated RF
carrier.
2-3
Making Measurements
Agilent Technologies 8922M/S Operating Modes
To configure the Agilent 8922M/S to the desired operating mode, carry out the
following instructions, referring to Figure 2-1 on page 4 for the position of the
fields.
•Ensure (1) is set to mobile phone type you require (GSM900, E-GSM, DCS1800,
or PCS1900).
NOTEFor use with DCS1800 or PCS1900, refer to the HP/Agilent 83220A/E Users Guide
which explains how to configure the Agilent 8922M/S to test other mobile phone
formats.
•Use the knob to move the cursor to the lower field below the Operating Mode (2).
•Push the knob and select the desired Operating Mode (ACTIVE CELL, TEST
MODE, or CW GENERATOR) from the menu which appears at the bottom
right-hand side of the screen.
1
2
Figure 2-1 Agilent 8922M/S Operating Modes
In addition to the three main operating modes, there are three other modes available.
These are:
•ACTIVE CELL+
•TEST MODE+
•CW GENERATOR+
These “+” modes have all the features of the other modes but include extra
diagnostic information on the screen. This information is designed to help advanced
users in fault-finding mobile phones.
2-4
ACTIVE CELL
Making Measurements
ACTIVE CELL
Making a Call From
the Mobile Phone to
the Agilent 8922M/S
This is the default mode after cycling the power or selecting.
A functional mobile phone will lock on (camp on) to a signal which is produced by
the Agilent 8922M/S. The characteristic of this signal appears in the BASESTATION fields, (1), see Figure 2-2 on page 2-6. These fields show the Channel
number and the Amplitude of the signal. Once the mobile phone has camped on to
the signal, it is possible to make a call between the mobile phone and the Agilent
8922M/S.
•Cycle the power on the Agilent 8922M/S or select.
•Insert a Test SIM card into the mobile phone. [The Test SIM (Subscriber Identity
Module) holds the user’s customized information. Agilent supply both micro and
standard Test SIMs.]
•Connect the mobile phone to the RF IN/OUT connector on the front panel of the
Agilent 8922M/S.
•Switch on the mobile phone and wait for it to camp to the Agilent 8922M/S (most
mobile phones display 001-01 when the mobile phone has camped).
•Dial any number on the mobile phone and press send.
The call set-up should now proceed. The CALL STATUS field on the screen should
display CONNECTED, (2), see Figure 2-2 on page 2-6. Once the call has been
established, the following information is displayed:
•Peak Power transmitted by the mobile phone is displayed in the center of the
screen.
•In the CELL STATUS area, (3), see Figure 2-2 on page 2-6, the mobile phone
reports its:
•Transmit power level (TX Lev)
•Received level (RX Lev)
•Received signal quality (RX Qual)
•Confirmation of the Traffic Channel and Timeslot is displayed to the left of the
mobile phone reports.
PRESET
PRESET
If the call does not proceed, there may be a problem with the mobile phone. Refer to
“Test Mode” in this chapter which explains how to configure the Agilent 8922M/S
to troubleshoot the mobile phone. When a call is connected, any speech received
from the mobile phone is echoed back by the Agilent 8922M/S, into the mobile
phone with a 0.5 second delay. This allows provisional checking of the mobile
phone’s audio sections by speaking into the mobile phone and listening for the echo
of your voice.
2-5
Making Measurements
ACTIVE CELL
2
3
3
4
6
5
Figure 2-2 Active Cell Mode
Making a Call From
the Agilent 8922M/S
to the Mobile Phone
NOTEThe Network number cannot be used to make a call. The Network Telephone number
To call (or page) the mobile phone from the Agilent 8922M/S, it is necessary for the
Agilent 8922M/S to know the number of the SIM in the mobile phone. [This number
is the International Mobile Subscriber Identity (IMSI). This and other information is
stored on the SIM card.]
is allocated to the SIM and is used in a real network to page the IMSI.
There are two ways the Agilent 8922M/S can receive this information:
•Make a call from the mobile phone to the Agilent 8922M/S.
Once the call is originated from the mobile phone, the Agilent 8922M/S
automatically reads the IMSI on the SIM card. If the previous call on the Agilent
8922M/S was made with the SIM card inserted in the mobile phone, and the
mobile phone is camped to the Agilent 8922M/S, the mobile can be called by
pressing.
•Enter the MS Information screen and enter the IMSI into the Agilent 8922M/S
manually.
2-6
ORG CALL
Making Measurements
ACTIVE CELL
Press the following keys to do this:
SHIFTCELL CONFIG
•, (MS INFO)
•Move the cursor to the Paging IMSI field and enter the IMSI using the
numeric data entry keypad.
CELL CNTLORG CALL
•,
Performing a “location update” from the MS INFO screen allows the Agilent
8922M/S to update the IMSI. This can be done by either:
•changing the Current location parameters and waiting for the mobile phone to
re-camp.
•settingIMSI Attach/Detach to On before poweringonthe phone. When the
phone camps its IMSI is set on the MS INFO screen and a call can be made from
the Agilent 8922M/S.
Changing Channel, Timeslot, and the Transmit Level
You can alter the parameters of the call before the call has been set up, or during a
call. They are displayed on the right-hand side of the screen under MOBILEPHONE, (4), see Figure 2-2 on page 2-6. The parameters are:
•Channel
•Transmit Level (TX Level)
•Timeslot
Channel To change channel, highlight the field and enter a new channel number
from the keypad. There is no interruption of communication between the Agilent
8922M/S and the mobile phone. For additional information on valid Absolute RF
Channel Numbers (ARFCNs), consult the specifications in Chapter 3.
TX Level This is a coded number used by the Agilent 8922M/S to command the
mobile phone to transmit at a particular power. When the TX Level is changed, two
things happen:
•The mobile phone should change its transmitted power.
•The Amplitude field in the Expected Input area of the screen, (5),
see Figure 2-2 on page 2-6, automatically adjusts to the nominal value defined by
the TX Level field. This allows the RF analyzer in the Agilent 8922M/S to
align itself with the mobile phone’sexpected output amplitude. If thesignal is not
within 3 dB of theexpected amplitude, it isnecessary to modify the Amplitude
field so that it is within 3 dB. Directly entering a value in the Amplitude field
does not change the transmitted level code transmitted to the mobile phone.
2-7
Making Measurements
ACTIVE CELL
There may be many reasons for the measured level not being close to the expected
level. The two most likely are that, either, the mobile phone is not operating
correctly, or, there is some power loss between the Agilent 8922M/S RF IN/OUT
connector and the mobile phone. If you suspect it is the second case, you can
compensate the Agilent 8922M/S generator settings and measurement results for
external losses or gains. The compensation is carried out in the CONFIGURE screen
which is accessed using the CONFIG field on the bottom right-hand side of the
screen, (6), see Figure 2-2 on page 2-6. Refer to Chapter 4, Screens for further
information.
Timeslot To change the timeslot, highlight the field and enter a new timeslot value
from the keypad. The timeslots can vary from 2 to 6. (Timeslots 0, 1, and 7 are
reserved for maintaining communication between the Agilent 8922M/S and the
mobile phone.)
NOTEWhile the Active Cell is selected, the three MOBILE PHONE parameters are also
available on the bottom right-hand side of all measurement screens. This provides
control of the mobile phone during measurements.
2-8
Making Measurements
TEST MODE
TEST MODE
To enter this mode, select TEST MODE as described in the section titled “Agilent
8922M/S Operating Modes”.
In TEST MODE, the Agilent 8922M/S no longer controls the mobile phone. The
TEST MODE is used when it is not desired, or not possible, to set up a call between
the Agilent 8922M/S and the mobile phone.
The MOBILE PHONE area available in Active Cell controls three functions
simultaneously:
•The traffic channel transmitted by the Agilent 8922M/S.
•The traffic channel transmitted by the mobile phone.
•The corresponding channel used by the Agilent 8922M/S to measure the mobile
phone.
In Test Mode, these functions are now separated. The traffic channel transmitted by
the Agilent 8922M/S is now controlled by the Traffic Chan parameters, (1),
see Figure 2-3 on page 2-10. The mobile phone’s transmission must be controlled
manually using the mobile phone manufacturer’s service and test modes. The
Agilent 8922M/S measurements are controlled by the fields below ExpectedInput, (2), see Figure 2-3 on page 2-10.
These fields are:
•Channel
•TX Level
•Burst Type
2-9
Making Measurements
TEST MODE
NOTEWhile the Test Mode is selected, the three MEASURE ON parameters are also
available on the bottom right-hand side of all measurement screens. This provides
control of the Expected Input parameters during measurements.
4
2
3
Figure 2-3 Test Mode
Mobile Phone
Transmitter Testing
Using Test Mode
To measure an incoming signal from the mobile phone’s transmitter, perform the
following steps:
•Select TEST MODE.
•In the Expected Input area, (2), see Figure 2-3 onpage 2-10, enter the GSM
channel number of the signal to be measured. This will automatically update the
input Frequency field below it, (3), see Figure 2-3 on page 2-10. If a
frequency needs to be measured which does not correspond to a standard GSM
channel, enter the non-standard value into the Frequency field. This value can
be anywhere from 10 MHz to 1000 MHz. A direct entry in the Frequency field
over-rides the frequency defined by the Channel field above.
•Enter the expected transmit level of the mobile phone in the TX Level field.
This will automatically update the Amplitude field below with the equivalent
TX level represented in dBm. If a non-standard input level is expected, enter the
value directly into the Amplitude field, this overrides the TX Level field
above.
•Entertheexpected Colour Code of the input signal. The Colour Codeisafunction
of the central “midamble” of the transmitted burst and is needed so that
measurements are correctly synchronized to the received burst. If the Colour
1
2-10
Code is not known, it can be determined and corrected from measurements
described later. Refer to the “Advanced Features” section mentioned later in this
chapter.
NOTEIn the Active Cell mode, the Colour Code is automatically set.
Once these have been selected, the Agilent 8922M/S is ready to measure incoming
signals of the type specified. The mobile phone should be set up to generate a
corresponding test signal to the one expected. This will require access to the mobile
phone manufacturer’s servicing or test modes.
Making Measurements
TEST MODE
Mobile Phone
Receiver Testing
Using Test Mode
It is possible to analyze the mobile phone’s response to the Agilent 8922M/S
Broadcast Channel (BCH), by varying the BCH number and amplitude, (4),
see Figure 2-3 on page 2-10. TEST MODE provides the Agilent 8922M/S with a
“forced” traffic channel (TCH) generator which can be turned on and off without the
need for any signaling or the presence of a mobile phone. A forced TCH can be
generated on any channel in the GSM Base Station range and is enabled by the On/Off toggle field (1) below the Traffic Chan or the and
ORG CALLEND CALL
keys.
The presence of both the BCH and forced TCH signals from the Agilent 8922M/S
allows the mobile phone’s receiver to be stimulated with signals identical to those
used on a real call. These signals can be used in conjunction with the mobile phone
manufacturer’s service and test modes to help measure and troubleshoot the mobile
phone. With these features, it is possible to make bit error rate measurements and
test the mobile phone’s receiver sensitivity when there is no call set up.
2-11
Making Measurements
CW GENERATOR
CW GENERATOR
To enter this mode, select CW GENERATOR as described in the section titled
“Agilent 8922M/S Operating Modes”. The CW Generator mode has the same
measurement capabilities as the Test Mode but replaces the GSM BCH and forced
TCH signals with a single unmodulated RF carrier. The frequency and amplitude of
the Continuous Wave (CW) signal is controlled by the Channel, Amplitude,
and the Frequency fields (1). When a GSM channel number is entered in the
Channel field, this automatically updates the generator Frequency field below it.
If it is necessary to generate a non GSM channel frequency, enter a value directly
into the frequency field. This range varies from 10 MHz through 1000 MHz. A
direct frequency entry over-rides the Channel field above. The unmodulated RF
signal can be used for any general purpose application and can be particularly useful
for checking some aspects of the mobile phone’s receiver where a CW signal is
needed.
Figure 2-4 CW Generator
2-12
1
Making Measurements
Measurements
Measurements
The measurements available on the Agilent 8922M/S can all be accessed from the
cell control screen by selecting a measurement field under MEASUREMENTS,
(see Figure 2-5 on page 2-13) and pushing the knob.
GSM Specific
Measurements
Ancillary
Measurements
The measurements available are:
•Peak Carrier Power
•Phase and Frequency Error
•Power Ramp Mask
•Bit Error Rate
•Output RF Spectrum (modulation or ramping, Option 006 only)
You can also use the toolkit capabilities of the Agilent 8922M/S. These additional
measurements are:
•Spectrum Analyzer (Option 006 only)
•Scope
•Audio Measurements
•CW Measurements
GSM Specific Measurements
Figure 2-5 Agilent 8922M/S Measurements
Ancillary Measurements
2-13
Making Measurements
Measurements
After making one type of measurement, another can be made by simply pressing
CELL CNTL
and, using the knob, selecting the next measurement of your choice.
Peak Carrier Power
Measurement
NOTESelect the Pwr Zero field, disconnect external source and lower BCH power, (3),
The peak transmitter carrier power averages the transmitter carrier power for a
single burst. This average is calculated over the time that the data information bits
are transmitted.
see Figure 2-6 on page 2-14, to zero the power meter before any signal is applied.
Method
The peak transmitter carrier power value is displayed on the cell control screen.
After you have set up a call (Active Cell) or manually aligned the Agilent
8922M/S to the mobile phone (Test Mode/CW Generator), the Peak Power
reading is displayed in the center of the screen (1).
Below the CALL STATUS area of the cell control screen are reports from the
mobile phone. These appear only when using Active Cell. They indicate the TX
Level which the phone is transmitting and the RX Level and RX Quality of the
received signal, (2), see Figure 2-6 on page 2-14.
2
1
3
Figure 2-6 Peak Power Measurements
2-14
Making Measurements
Measurements
Phase and
Frequency Error
Measurement
NOTEThe test is run automatically when the screen is selected.
Phase error and frequency error are measures of the modulation and noise
performance of the radio’s transmitter path.
Method
Select the PHASE FRQ field on the cell control screen to access the phase and
frequency measurements.
The measurement fields in this screen are RMS Phase Error, Peak Phase Error and
Frequency Error (1), see Figure 2-7 on page 2-15.
The Agilent 8922M/S input sensitivity (2) can be varied, if necessary, to ensure that
the input signal level matches the Agilent 8922M/S RF Analyzer. Valid
measurements are only made when the signal is within 3 dB of the RF Analyzer
setting.
The Agilent 8922M has the option of using multi-burst measurements (4). This
gives the ability to make several phase and frequency measurements using more
than one burst. For more information refer to Chapter 4, Screens, Phase and
Frequency Error screen.
1
2
4
3
Figure 2-7 Phase and Frequency Error Measurements
Two additional screens in the phase and frequency measurement section can be
revealed if you select (3).
2-15
Making Measurements
Measurements
These are:
•PHASE ERR - this displays the phase error graphically. The phase error trace is
displayed using an autoscaling phase error axis versus data bits (numbered 0
through 147).
•DATA BITS - this screen allows you display a screen which details the values
of the 148 bits in the timeslot (including midamble). If a known test signal is
being used, the reception of these bits can be verified.
Power Ramp Mask
Measurements
To avoid unwanted interference and to ensure successful reception at the Base
Station, the mobile phone’stransmitted signal must conform to GSM standards. The
purpose of the power ramp is to display the pulsed signal and verify that it conforms
to these standards.
Method
Selecting PWR RAMP on the cell control screen gives you access to the power ramp
measurement screens.
The marker (1), see Figure 2-8 on page 2-16 can be used to make a measurement at
a discrete point in time along the signal trace. The amplitude at this point is
displayed in the top left-hand corner of the screen.
2
Figure 2-8 Power Ramp Measurements
2-16
1
3
Making Measurements
Measurements
The power ramp measurements are divided into three screens where you can view
different parts of the signal and one screen which displays a series of amplitude
values at various times during the burst. These screens can be revealed by
highlighting and selecting from the View field (2):
•Rise Edge - displays the top 30 dB of the rising section of the waveform.
•Top 2 dB - displays the signal during the middle part of the burst allowing
analysis of the ripple of the signal.
•Fall Edge - displays the signal during the falling edge of the burst allowing
analysis of the fall time of the signal.
•Summary - details the amplitude measurements made at the times selected in the
12 time fields. You can choose your own time-positions, or use the default
settings.
The sensitivity of the Agilent 8922M/S receiver can be varied (3) allowing you to
verify whether the input signal level matches the Agilent 8922M/S RF Analyzer.
Valid measurements are only made when the signal is within 3 dB of the RF
Analyzer setting.
Measurement
Summary
The Measurement Summary field on the DSP Analyzer Ampl Main screen displays
whether HI/LO limits set for the measurement display fields, (Ampl1-12, pk+
flatness, or pk- flatness) were exceeded in the last measurement. The possible
Measurement Summary displays are:
Failed
One or more measurement limit was exceeded.
Passed
No measurement limits were exceeded.
- - - -
No measurement limits are set, or, all of the Ampl and Pk measurement displays are
turned off.
A blank field
The blank field is displayed when the measurement is armed. It will remain blank
until the measurement is complete.
Refer to “Pulse Measurements” within this section for a description of Pulse
Measurements. These measurements are available with option 006 only.
2-17
Making Measurements
Measurements
Pulse Measurements
(Available if Option
006, Spectrum
Analyzer, is Fitted)
If you have option 006 (spectrum analyzer) installed, you can make measurements
on the lower portion of the pulse. These measurements can be accessed from the
Power Ramp screens.
•The main measurements, Pulse On/Off Rise and Pulse On/Off Fall are displayed
at the top of the screen (1), see Figure 2-9 on page 2-18
•The sensitivity of the Agilent 8922M/S receiver can be varied (2) allowing you
to verify whether the signal still falls within the boundaries of the GSM power
mask.
1
32
Figure 2-9 Pulse Measurements
The Agilent 8922M/S provides three pulse measurements which can be selected
using View (3). These are:
•Pulse - pulse on/off ratio measures the ratio of the average transmitter- power
(pulse on) to a specified time-position when power is reduced (pulse off). The
default settings are 28µs before bit 0 and 28µS after bit 147.
•Puls Rise - this screen displays the signal during the initial rise of the pulse.
•Puls Fall - this screen displays the signal during the final fall of the pulse.
2-18
Making Measurements
Measurements
Bit Error Rate
Measurement
NOTEThere are two sets of results displayed.
The Bit Error Rate measurement allows you to test the sensitivity of the mobile
phone’s receiver. By reducing the signal transmitted by the Agilent 8922M/S, you
can test the ability of the receiver to accurately decode its incoming signal. Data bits
that are decoded are sent back to the Agilent 8922M/S. The Agilent 8922M/S
compares them to original signal that was sent out and the differences are derived
from this. The audio echo function that works in the Cell Control screen is turned off
when making a BER measurement.
Method
To run a bit error rate test, select the BIT ERROR field in the cell control screen.
This reveals the bit error measurement screen.
This test will run automatically if you set the Single/Cont field to CONT.To run
a test set the Run/Stop field to RUN. This toggle field start/stops the BER test (1)
ratio, see Figure 2-10 on page 2-19.
•Intermediate Results - this is a running total of the bit errors as the bits are being
tested (2), (this is useful where a large number of bits are being tested).
•Final Results - this shows the completed BER (3). This is displayed after all the
bits have been tested.
3
2
1
Figure 2-10 Bit Error Rate Measurement
4
2-19
Making Measurements
Measurements
By varying the Base station Amplitude field (4), you can test the mobile phone’s
receiver sensitivity. Actual results can be compared with the values reported by the
radio.
NOTEIf the Base Station amplitude is lowered too much, the radio will lose the call. This
will need to be re-established by increasing the base station amplitude, returning to
the cell control screen, and originating the call before you can continue
measurements.
NOTENot all phone types support burst-by-burst BER measurements.
OutputRFSpectrum
Measurements
(Available if Option
006, Spectrum
Analyzer, is Fitted)
The Output RF spectrum measurement shows the spectral power (due to the effects
of ramping or modulation) at a specified frequency offset. These are relative
measurements that compare the result at the offset frequency with the value at the
center frequency. You can also view the output RF spectrum trace from this screen.
Method
Select OUT RF SP in the cell control screen.
To make a measurement, do the following:
•Ensure the Freq Offset is set to zero (1), see Figure 2-11 on page 2-21.
•Select Ramp Ref or Mod Ref (2) (depending on whether you want to make a
ramping measurement or a modulation measurement).
2-20
Making Measurements
Measurements
This sets the reference level to which the offset frequency values are compared.
•Select Ramping or Modulation (3).
•Set your Freq Offset value (4).
5
2,3
Figure 2-11 Output RF Spectrum Measurements
•The trace of the output RF spectrum can be viewed if you highlight View and
select Trace (5).
OutputRFSpectrum
MeasurementsUsing
a 3-Pole
Synchronously
TunedMeasurement
Filter
GSM Recommendation 11.10 and 11.20 set the requirements for the out-of-channel
performance of base station and mobile station transmitters with the Output RF
Spectrum specification. The specification calls for the measurement of transmitted
energy at several offsets from the carrier frequency. Two types of measurements are
required:
•Check the interference due to modulation.
•Check the interference due to ramping of the signal amplitude (switching
transients).
Limits are specified for each type of measurement at specified frequency offsets.
The GSM specified limits are based on a 5-pole synchronously tuned measurement
filter.
1,4
2-21
Making Measurements
Measurements
In the Agilent 8922, Output RF Spectrum measurements are made using a 3-pole
synchronously tuned measurement filter. The measurement results will differ from
measurements using a 5-pole filter.
The difference between measurements of the “ideal” signals using a 3-pole filter and
a 5-pole filter are shown in Table 1 and Table2. These values are added to the GSM
specifications to obtain equivalent limits for Output RF Spectrum measurements
using 3-pole synchronously tuned measurement filter. The adjusted Output RF
Spectrum limits are based on a 3-pole filter are shown in Table 3 and Table 4.
Table 13-Pole Specification Difference Output RF Spectrum due to Modulation
Offset from Carrier (kHZ)
100200250400600 to
1800
Difference (dB)02.51.09.00
Table 23-Pole Specification Difference Output RF Spectrum due to Switching
Offset from Carrier (kHZ)
40060012001800
Difference (dB)6420
NOTEIt is important to note that these values are estimates based on simulation and
measurements. They are intended to allow the user to estimate their transmitters
conformance to GSM Output RF Spectrum requirements.
2-22
Making Measurements
Measurements
Table 3Adjusted Limits Based on 3-Pole Filter Output RF Spectrum Due to
Spectrum Analyzer
(Available if Option
006, Spectrum
Analyzer, is Fitted)
NOTEThe spectrum analyzer can detect very low power signals where Active Cell and
NOTEIf both the input and output are set to RF IN/OUT, large errors in measured signal
The spectrum analyzer allows you to view the mobile phone’s signal over a wide
dynamic range. It also allows you to view any adjacent interference which may
exist.
even Test Mode cannot operate. Broken cable or connectors can be found using
this function.
Method
Select SPEC ANL in the cell control screen.
Selecting (1) provides access to other functions of the spectrum analyzer. These are:
•Main - the default screen has basic spectrum analyzer functions,
see Figure 2-12 on page 2-25.
•RF Gen - this controls the RF generator in the Agilent 8922M/S. The generator
signal can be fed back into the spectrum analyzer for signal confirmation. The
Aux RF OUT port should be connected to the RF In/Out, and the RF Output set
to “Aux RF OUT”.
level will be seen.
•Marker - this screen allows you to control the marker for the input signal.
•Auxiliary - the auxiliary screen controls theinputs and the attenuator settings.
2-24
Making Measurements
Measurements
NOTEThe RF Generator and spectrum analyzer can be tuned to different frequencies
allowing the inspection of the IF signals inside the mobile phone.
1
2
Figure 2-12 Spectrum Analyzer Measurement
Use the MeasReset (2) to reset the trace and measurements in the Agilent 8922M/
S. This is useful if you are using the Max Hold field (3).
The resolution bandwidth of the spectrum analyzer is automatically coupled to the
frequency span. Reducing the span to less than 200 kHz can be useful when
checking the mobile phone’s transmitter.
3
2-25
Making Measurements
Measurements
ScopeThe oscilloscope function of the Agilent 8922M/S allows you to view the
demodulated signal from the mobile phone. This can be used for fault-finding in the
audio path.
Select SCOPE in the cell control screen to gain access to the oscilloscope function of
the Agilent 8922M/S.
Selecting (1), see Figure 2-13 on page 2-26, gives access to other functions of the
oscilloscope. These are:
•Main - gives you the basic oscilloscope functions for viewing and measuring the
trace.
•Trigger - allows you to trigger the input signal from a variety of sources.
•Marker - offers you the capability to measure the voltage at discrete points in
time along the trace. The result is shown in the top right-hand corner of the
screen. You may find it helpful to use the,, (HOLD) feature
when using the marker.
SHIFT
PREV
1
Figure 2-13 Oscilloscope Measurements
Use the MeasReset (2) to reset the trace and measurements in the Agilent 8922M/
S. This is useful for single triggered measurements.
Set AF Anl In (3) to change the source of the signal being directed to the Audio
Analyzer Input. Refer to Screens, Chapter 4 for further information.
2-26
2
3
Making Measurements
Measurements
AudioThe audio function measures the audio frequency and voltage of the demodulated
signal from the mobile phone or from a number of other sources selectable using AFAnl In (5), see Figure 2-14 on page 2-27.
Select AUDIO in the cell control screen to gain access to the audio function of the
Agilent 8922M/S.
The measurements of the audio voltage and frequency commence within a short
time of the screen being accessed.
The ac voltage (1) and the audio frequency (2) are the two main measurements made
in this screen. Additional measurements, such as DC Level, can be made by
selecting the AF Freq field. (2). The audio frequency generator (3) is used to
generate an audio signal from dc up to 25 kHz. The AF analyzer fields sets the
conditions for measuring the audio signal (4).
1
3
Figure 2-14 Audio Measurements
2
4
5
2-27
Making Measurements
Measurements
CW MeasurementThe CW Measurement screen displays the carrier frequency and power of a
continuous (non-pulsed) signal. The CW Power measurement offers a greater
dynamic range than is available when making pulsed measurements.
CW Power is a broadband measurement. The CW Frequency measurement is
obtained using a tuned, selective input. The RF analyzer should be set to within 500
kHz of the expected signal frequency.
In the GSM band, the value that is entered should be set to the nearest 100 kHz.
Select CW MEAS in the cell control screen to gain access to the power measurements
of the Agilent 8922M/S.
The CW measurements are made automatically when you access the screen.
The CW Frequency field (1), see Figure 2-15 on page 2-28, displays the carrier
frequency of the signal and CW Power (2) shows the power of the carrier signal. It is
also possible to display the difference between the measured frequency and the
value selected by choosing CWFreqErr from field 1.
1
Figure 2-15 CW Measurements
These measurements can be reset using MeasReset. Select Pwr Zero before
power is applied to zero the power meter in the Agilent 8922M/S.
2
2-28
Making Measurements
Measurements
NOTEAlthough CW Measurements is a broadband measurement, it uses calibration data
that relies on the expected input frequency being set correctly. The Power Detector
is connected so that it will only make measurements on signals present at the RF In/
Out port.
2-29
Making Measurements
If You Have Problems with a Measurement
If You Have Problems with a Measurement
This section tells you what to do if either of the following screen display events
occurs:
•Message Line Messages (on the top of the screen).
•Is a Message Line displayed at the top of the screen.
•Possible Solutions to Message Line Errors.
•Sync Status Messages
•Is an Error Message Displayed in the Sync Status Field.
Refer to Chapter 7, “Messages”, for more information.
Is a Message Line
displayed at the top
of the screen?
At position 1, see Figure 2-16 on page 2-31, a message line appears if:
•Attempt To Arm or Query an Inactive Measurement - check that a measurement
has been selected from the To Screen.
•Trigger too late, decrease trigger delay - See Trigger Timing B
•Trigger too early, increase trigger delay - See Trigger Timing B
•Measurement Armed, awaiting trigger - See Trigger Timing A
2-30
Making Measurements
If You Have Problems with a Measurement
Figure 2-16
2-31
Making Measurements
If You Have Problems with a Measurement
Possible Solutions to
MessageLine Errors
The signal processing hardware that is used to generate measurement results has to
be told when to take data samples. If the trigger to do this is misaligned with the
signal, there may be errors in the results that are displayed.
Trigger Timing A
Check if the trigger is being received.
On the MEAS SYNC screen (pressto gain access), check if the
correct burst type has been defined.
Trigger Timing B
Check if the trigger is being received at the correct time.
The First Bit field (A) , see Figure 2-17 on page 2-32, on the Phase Freq: Data
Bits screen displays the time difference between when a trigger is being received
and when the first bit of a burst occurred. The time difference is only valid if
FMErrCount is 0. The timing difference can be corrected by altering the trigger
delay (B).
MEAS SYNC
A
Figure 2-17
B
2-32
Trigger Range For Pulsed RF
Making Measurements
If You Have Problems with a Measurement
2-33
Making Measurements
If You Have Problems with a Measurement
Is an Error Message
Displayed in the
Sync Status field?
The sync status field displays an error message for the following errors:
•Bad Sync - demodulation error, perform a Phase/Frequency error measurement
to identify which of the Sync Status error listed below may be the possible
problem.
•FM Errors - see solutions 1, 2, 3, 4
•Short Burst - see solutions 1, 4
•Level Late - see solutions 1, 2, 3, 4
•Level Short - see solutions 1, 2, 3, 4
•Low Level- see solution 3
•RF Ovrload - see solution 3
2-34
Making Measurements
If You Have Problems with a Measurement
Possible Solutions to
Sync Status Errors
Solution 1 - Trigger Timing
Check if the trigger is being received.
On the MEAS SYNC screen (pressto gain access), check if the
correct burst type has been defined. On the Data Bits screen:
A The First Bit field on the Phase Freq:Data Bits screen displays the time
difference between when a trigger is being received and when the first bit of
a burst occurred (A). The time difference is only valid if FMErrCount is 0.
B Check the delay in the Trig Delay field (B).
MEAS SYNC
A
Figure 2-18
B
2-35
Making Measurements
If You Have Problems with a Measurement
Solution 2 - Midamble Sync
On the MEAS SYNC screen (pressto gain access),
•Check the definition of the signal’s burst type (A).
•Check the Burst Sel field (B).
MEAS SYNC
A
B
Check the bit pattern of your measurement
Perform a Data Bits measurement (D). An “M” will display under the
bits that are identified as the midamble bits.
D
2-36
Making Measurements
If You Have Problems with a Measurement
Solution 3 - Level
Check the following fields on the RF Generator/RF Analyzer screen (press
SHIFTCELL CNTL
3Pulse field (a), (if signal is pulsed) Ext or Hop Trig should be selected.
4Amplitude field (b), for the expected amplitude
5Frequency field (c), for the correct frequency.
6RF Input field (d), for the correct connector choice.
7AGC Mode field (e). If either Open or Auto is the selected mode, check the
Check the actual amplitude of the input signal.
Perform a Peak Carrier Power measurement.
,, (RFG/RFA) to gain access).
value in the Open Loop DAC (f) Value field. If Closed is the selected mode,
check that the burst is repetitive (at least one timeslot every two frames).
a
b
c
d
e
f
2-37
Making Measurements
If You Have Problems with a Measurement
Solution 4 - Amplitude Envelope
•Check if the Pulse Amplitude is ± 1 dB of the expected value during the useful
part of the burst.
If FM Errors:
Perform a Phase Freq:Data Bits measurement. Dashes (a) will display
under the bits where the power is too low.
If No FM Errors:
Perform pulse demodulation measurements. Connect PULSE
(DEMODULATION OUT) connector to SCOPE IN (MEASURE)
connector on the front panel.
a
---------
2-38
Making Measurements
Advanced Features
Advanced Features
Other ScreensThe screens mentioned in this section are not necessary for simple measurement of
mobile phones as the work is done by the functions mentioned previously. However,
the advanced user may find it worthwhile to know what these screens are and what
they are capable of doing.
•RF Generator/RF Analyzer - this screen controls the Agilent 8922M/S RF
generator and RF analyzer. Thisscreen can be accessed bypressing the following
keys:
SHIFTCELL CNTL
•,, (RFG/RFA)
•MS Information - the MS INFO (Mobile Subscriber Information) screen gives
you access to the information about the mobile phone itself. A Location Update
can be verified from this screen. This screen can be accessed by pressing:
SHIFTCELL CONFIG
•,, (MS INFO)
•Cell Configuration - the CELL CONFIG screen gives you access to the control
parameters for the configuration of the cell. This screen can be accessed by
pressing:
CELL CONFIG
•
•CellControl2 - gives you the capabilityto control the additional cell parameters.
This screen is relevant in the set up of a call. This screen can be accessed by
highlighting the More field in any main screen and selecting CELL CNTL2.
•Hopping - the hopping screen allows you manual control of the hopping
frequencies. This feature is only available in the Agilent 8922G compatibility
mode.
•Logging - the logging screen allows control over the Protocol Interface port on
the rear panel (option 003). To access this screen, highlight More and select
LOGGING.
•Configuration - the configuration screen is a full instrument control screen
which allows you to set up the Agilent 8922M/S. Screen set up, compatibility
mode, radio type as well as other miscellaneous pieces of information are
contained on this screen. To access this screen, highlight More and select
CONFIG.
•I/OConfiguration - the I/Oconfiguration screen allows you to set up the general
interface of the instrument, such as printer options. This screen can be accessed
by highlighting the More field in any main screen and selecting I/O CONFIG.
2-39
Making Measurements
Advanced Features
•SMS Cell Broadcast - the SMS CB screen allows you to test whether the mobile
phone is capable of correctlyreceiving and displaying a message sent from a base
station using the Short Message Service (SMS) Cell Broadcast protocol. This
screen can be accessed by highlighting the More field in any main screen and
selecting SMS CB.
Using RF Rise
Triggering
Effect of the RF
Generator on RF
Analyzer
Measurements
When using RF Rise triggering of measurements, the Trigger Delay should be
adjusted so that the wanted data is captured during the sampling process. The actual
measurement is made at the Trigger Delay time after the RF Rise. When changing
from RF Rise triggering to another trigger mode, the Trigger Delay should be
adjusted to the appropriate value as the default Trigger Delay is not automatically
restored.
Example:In the Agilent 8922M/S, when using the RF Rise, a Trig-
ger Delay of 0 is recommended. When changing from RF
Rise to Ext Demod trigger mode, the default value of
473.4T should be re-entered for Trigger Delay.
The value 473.4T is derived from the 468.75 bits of delay
between downlink and uplink plus the 4.7 bits of modulation delay inside the Agilent 8922M/S. All timing is referenced to the beginning of bit zero on the downlink TCH
timeslot, (DCH or BCCH timeslot if TCH is not active.)
When the RF Generator is set to a high level relative to the analyzer RF input, some
error may be introduced in the analyzer’s power detector. This is due to the isolation
limits between generator and analyzer on the RF IN/OUT port. This only occurs
when the RF Generator and RF Analyzer share the RF IN/OUT port. If the
RF Generator uses the AUX RF OUT port, no error occurs.
Error10LOG 10
RFin 20–
--------------------------
10
RFgen 14+
-------------------------------
+
10
10
RFin 20–()–dBm=
RFin is the amplitude in dBm of the signal into the RF IN/OUT port. Normally the
setting in dBm of the RF Analyzer Amplitude. RFgen is the setting in dBm of the
RF Generator.
Because the two components are not always in phase, the actual error will often be
slightly less.
2-40
Making Measurements
Advanced Features
Power Ramp Setting
Time
Peak Carrier Power
Measurement,
Assumptions And
Algorithm
The mobile transmitter output ramp should settle so that it complies with the power
mask (GSM Rec 11.10) by the time output level calibration is done in the
manufacturing process. The power mask is displayed in the Power Ramp screens.
Peak Carrier Power measurement method is not made to GSM Rec. 11.10. If you
need to make a measurement to this specification, go to the Power Ramp Summary
screen. The field in the top right-hand corner displays the Peak Carrier Power to
GSM Rec. 11.10. Refer to PWR RAMP: SUMMARY screen (item 5) for a
description.
The algorithm for the Peak Carrier Power measurement is based on the use of a fast
sampler taking continuous asynchronous measurements of the power detected at the
RF IN/OUT port.
The sampler accumulates 75µS sample bins containing the maximum power
observed during the period and the minimum power observed during the period.
Sampling rate is 10 MHz. The maximum and minimum values observed are based
on 750 individual samples within the period.
A trigger is provided at the middle of bit 0 of the burst. (When using RF Rise
triggering, the actual trigger may occur several bits earlier due to the ramp up of the
burst.) In response to this trigger, the following 6 sample bins (0 - 5) from the
sampler are saved and reported to the measurement processor.
Timeslot
0µS
577µS
Mobile Burst
Amplitude
Trigger
Sample
Bins
-2-10123456789
375µS Window
75µS/sample_bin
2-41
Making Measurements
Advanced Features
The beginning of the 0 sample bin will be between 0 and 75µS (bit 0 - 20) after the
trigger as the sampler runs asynchronously to the trigger. The last sample bin will
complete sampling between 450µS and 525µS (bit 121 - 142) into the burst.
The measurement processor discards the 0 sample bin. This effectively removes the
ramp-up overshoot. Each measurement is also RF Power Qualified to ensure only
valid bursts are measured. Measurements with a result below the qualification level
are discarded.
The remaining sample bins 1-5 are averaged for the displayed measurement result.
Each measurement is therefore based on a 375µS (102 bit) sampling of the burst.
A mobile which does not meet the amplitude mask and has erratic amplitude across
the burst could give erroneous results with this measurement algorithm. It is
therefore essential that mobiles first comply with the amplitude mask in
GSM Rec. 11.10.
Making
Synchronized
External
MeasurementsUsing
Triggers From The
Agilent 8922M/S
There are two ways of making a synchronized external measurement using triggers
from the Agilent 8922M/S:
1Use the G_EXT_TRIG on the rear panel SYSTEM_BUS connector. See
Chapter 6, Connectors for a pin out diagram.
This is a separate trigger directly from the Protocol Processor. It is only
selectable from the SERVICE screen. Once set it will provide continuous
triggers on each frame. Caution must be used as this trigger can have up to a
0.75 bit error due to the 156/157 bit variance in timeslots.
(GSM Recommendations allow use of 156/157 bit timeslots instead of the
156.25 bit length.)
To use the G_EXT_TRIG signal:
Go to the CONFIG screen.
Select the SERVICE field on the right hand side of the screen.
Select the LATCH field and find the selection labeled:
‘g_ext_trig_enable’ (SERV:LATCH:SEL ‘g_ext_trig_enable’)
Enter a ‘1’ in the VALUE field. (SERV:LATCH:VAL 1)
Select the LATCH field and find the selection labeled:
‘g_ext_trig’ (SERV:LATCH:SEL ‘g_ext_trig’)
Enter the TIMESLOT and BITPOSITION on which you want the
trigger to occur.
Example:432 hex is TIMESLOT=4 BITPOSITION=32 hex. The
maximum TIMESLOT=7 and the maximum BITPOSITION=9C hex (Bit 0 - 157).
2-42
Making Measurements
Advanced Features
NOTEThe GPIB requires the use of base 10 values, 432h = 1074d, for example:
SERV:LATCH:VAL 1074.
The value entered is relative to the beginning of the zero bit of the zero
slot on the downlink baseband.
A 4.7 bit modulator delay occurs between the baseband and the RF
domain which must also be taken into account. Since the value may
only be integer, round to the nearest whole value.
The following timing error is introduced due to 156/157 bit timing.
Timeslot=0,4 No Error
Timeslot=1,5 +.75 bit error (early trigger)
Timeslot=2,6 +.50 bit error
Timeslot=3,7 +.25 bit error
2Use MEAS_TRIG_OUT on the rear panel SYSTEM_BUS connector.
This is the delayed trigger from the Measurement System. In the normal
ACTIVATED operation, the Protocol Processor sends triggers to the
Measurement System on the zero bit of each downlink TCH burst. The
Measurement System applies the ‘Meas Trig’ ‘Trigger Delay’ to this input and
then, after the elapsed time, begins making its measurement. After this elapsed
time, the Measurement System also sends a trigger transition to the
MEAS_TRIG_OUT signal on the rear panel SYSTEM_BUS connector.
The trigger is only generated when a measurement is being made. This signal
may be used to trigger an externally connected measurement device such as the
HP/Agilent 71150/250 GSM high performance spectrum analyzers.
2-43
Making Measurements
Advanced Features
Unexpected
Operations
These are descriptions of operations which may be unclear to the user as to how they
occurred. These are NOT defects.
1The Agilent 8922M/S occasionally sends ABORTS to the attached protocol mon-
itor. These ABORTS are normal and logging data is not lost.
2In the FA and SD/4+FA control channel configurations, the Agilent 8922M/S
sends speech frames rather than fill frames on the FACCH when there is nothing
else to send.
3The DRX ON mode corresponds to the NORMAL paging mode. The
Agilent 8922M/S sends continuous paging mode NORMAL on the idle PCH
blocks. The DRX OFF mode corresponds to the REORGANIZE paging mode.
The Agilent 8922M/S sends continuous paging mode REORGANIZE on the idle
PCH blocks. When in the REORGANIZE paging mode, the Agilent 8922M/S is
guaranteed to page on a random PCH block which is not the block expected in
the NORMAL paging mode. The active PCH block will page the MS with the
paging mode set to SAME. The SAME mode means “the same paging mode as
last indicated.”
4TMSI reallocation in the Agilent 8922M/S uses the two (2) octet TMSI size.
5The Agilent 8922M/S does not turn off the transmitter during DTX when the
voice activity detector (VAD) determines that there is no speech input present.
Instead, the Agilent 8922M/S sends a silence descriptor (SID) on every speech
frame.
6USE MEM is not intended to be used with Pulse (Pulse ON/OFF Ratio)
measurements - do not select USE MEM after making a pulse measurement.
7During a Pulse (Pulse ON/OFF Ratio) measurement the following message can
occur if the device under test is generating a larger signal than is expected: ‘RF
Overpower occurred during Pulse On/Off Ratio measurement’. This message
cannot be queried via GPIB. Use the Hardware 1 Status Register Bit 5 to detect
this condition when doing pulse measurements.
NOTEOption 006 (spectrum analyzer) must be installed to make pulse measurements.
8The Output RF Spectrum measurement and Pulse On/Off Rise and Fall
measurement results shown on the Main view screens are not accessible via
GPIB or IBASIC. These measurement results are available on the Trace view
screens. All GPIB and IBASIC actions with these fields refer to those fields
shown on the Trace view screens.
2-44
Making Measurements
Advanced Features
9The RF Generator may become uncalibrated in a hopping situation when the last
CW frequency setting was outside the GSM or DCS band. If you are using the RF
Generator outside the normal GSM or DCS bands, make sure you set a valid
GSM or DCS frequency in the CW frequency before entering a hopping mode.
This includes the ACTIVE CELL (ACTIVATED state on the CELL CONFIG 2
screen) on the Agilent 8922M/S.
10 When using RF Rise triggering it is recommended to set MS_Parms:Timing
Advance to MANUAL with the Timing Advance set to zero. If AUTO timing
advance is used with RF Rise triggering, the Agilent 8922M/S, operating as a
base station, will progressively command the mobile to advance timing until it is
at the maximum value of 64. It is difficult to recover from this state without
ending the call with the mobile. Changing back to EXT_DEMOD triggering will
cause the call to drop. This is because the Agilent 8922M/S is listening for the
mobile 64 bits later than the RF Rise triggering left it.
11 When making measurements over the GPIB, it is often necessary to place time-
out constraints in software to account for mobile failure. When a time-out is
encountered while a measurement is armed but not complete, it is necessary to
perform a “CLEAR” operation on the GPIB. A “RESET” operation is
insufficient.
Example
ASSIGN @gpib TO 714
OUTPUT @gpib;”TRIGger:ASTate ’ARM’”
ON TIMEOUT 7,5 GOTO Flag1
OUTPUT @gpib;”MEASure:DSP:PHASe:FREQuency?”
ENTER @gpib;The_answer
PRINT ‘FREQUENCY:’; The_answer
OFF TIMEOUT 7
GOTO Flag2
Flag1:OFF TIMEOUT 7
12 Some measurement results may be displayed as ---- due, for example, to
mobile failure. The measurement has been armed but has not completed. In these
circumstances a GPIB query of the measurement will not return a result. This
event should be catered for by adding a time-out to the query and a “CLEAR”
operation as described in paragraph 11, above.
13 When changing between screens, the DSP results may briefly show a very large
number. This is expected behavior.
When controlling the instrument via GPIB, some phase and frequency results
may consistently return a very large number. This is most likely caused by
attempting to read result from another DSP screen when measurement triggering
is set to Continuous. The alternative solutions are either to change to the
appropriate DSP screen before reading the result, or to use single-shot
measurement triggering.
2-46
Making Measurements
Advanced Features
2-47
3
Verifying Performance
3-1
Verifying Performance
About This Chapter
About This Chapter
The tests in this chapter verify the electrical performance of the Agilent 8922M/S
GSM Test Set using the Agilent 8922 Performance TestSoftware provided with the
product. If the instrument passes this verification, its operation and specifications are
assured within the measurement uncertainties provided in the performance test print
out. Because of the specialized nature of the Agilent 8922M/S and the equipment
required to support it, it is recommended that calibration and repair be performed
only by specially equipped Agilent Technologies service centers.
A list of specifications is found at the end of this chapter.
3-2
Setting up the Tests
This chapter contains the following information:
Verifying Performance
Setting up the Tests
Getting the Right
Test Equipment
Installing and
Operating the
Software
Required Test Equipment lists the test equipment needed for the performance
tests. This is the only equipment supported by the Agilent 8922 Performance Test
Software and is required to verify instrument operation. Equipment substitutions or
manual performance tests are not recommended or supported by Agilent
Technologies.
Performance Test Software describes how to install and operate the Agilent 8922
Performance Tests software.
3-3
Verifying Performance
Getting the Right Equipment
Getting the Right Equipment
The following equipment is required to do all of the performance tests. The test
descriptions have an equipment list that specifies the equipment used for each
particular test.
BASIC and DS Disk Drive
GPIB/RS-232 printerHP 2225A
HP9000 Series 200/300
3-4
Verifying Performance
Installing and Operating the Software
Installing and Operating the Software
PerformanceTest Software is supplied on a 3.5-inch, double-sided floppy disk and
is written to run with BASIC 5.0 and later. Modifications to the program should be
limited to changing the default addresses and storing copies for back-up purposes.
Understanding the
Tests
To Load the
Program.
Using the
Compatibility
Switch
Test Descriptions contains a description of each test that is performed by the
Performance Test software. This description is intended to help locate problems if
the software fails to execute properly or to help users understand the test
methodology that is used in each performance test. The descriptions are not step by
step procedures for manual performance tests.
To verify the performance of the Agilent 8922M, you need to convert the instrument
back from an Agilent 8922G, or convertthe Agilent 8922S to an Agilent 8922E. To
do this refer to the following section titled “Using the Compatibility Switch - Back
Conversion”.
You are now ready to run the Performance Test Software.
1Put the disk in the disk drive.
2Type LOAD “PT_8922”, press.
After you have completed the Performance Tests, return the instrument back to the
Agilent 8922M or Agilent 8922S by following the instructions in the following
section titled “Using the Compatibility Switch - Forward Conversion”.
Back Conversion
To turn the instrument from the Agilent 8922M/S back to an Agilent 8922G or an
Agilent 8922E, select the following keys:
•CONFIG (this is accessible from the Cell Control screen in the bottom right-hand
corner).
•Compatible, select 8922G or 8922E
•HP-IB Adrs (22)
•
PRESET
ENTER
The instrument is now set up as an Agilent 8922G or Agilent 8922E and ready for
Performance Verification testing.
3-5
Verifying Performance
Installing and Operating the Software
Forward Conversion
To return the instrument from an Agilent 8922G back to an Agilent 8922M or an
Agilent 8922E to an Agilent 8922S, select the following keys:
•More (this is accessible from the Cell Control screen in the bottom right-hand
corner). Scroll down the list and select CONFIG.
•Compatible, select 8922M or 8922S
•HP-IB Adrs (14)
•
PRESET
The instrument is returned to an Agilent 8922M or Agilent 8922S.
To Configure the
GPIB Addresses
1With the program loaded, type EDIT DEFAULT_ADDRESS, press.
2Modify each line to indicate the proper instrument address (700-730).
It is now possible to re-store the program as “PT_8922” or store it under a
different name.
To Run the Program 1Type RUN, press.
ENTER
2Follow the directions as they appear on the screen.
Notes on Running the Program.
The first screen which appears is the GPIB status of each piece of test equipment
that is supported. It is only necessary to have the instruments responding that will be
used in each particular test. Make certain that each instrument you will be using is
responding at the proper address. Duplicate addresses may make an instrument
appear to be responding but this is not allowed. Press “I” (for Ignore) to continue
past this screen.
The second screen prompts you for the instrument model. If you have disk
08922-10006, select HP 8922G (for Agilent 8922M performance testing) or HP
8922E (for Agilent 8922S performance testing). The third screen which will appear
is the main Performance Tests selection menu. Three options are available on this
screen:
•Select the performance test to run, remember the test instruments and UUT must
be responding over GPIB.
ENTER
•Turn the printer function ON or OFF. If the printer function is turned on it must
be responding over GPIB or the program will lock up.
•Exit from the program.
Press the key corresponding to the option that you would like to perform. The other
screens that appear are connection instructions, error messages and output results.
3-6
Verifying Performance
Understanding the Tests
Understanding the Tests
This section describes the theory of each performance test, lists the equipment
needed for the test, and provides some problem solving information.
The UUT is set to generate CW signals at various levels and frequencies. The HP/
Agilent 8902A/11722A is used as a power meter to measure high level signals and
the HP/Agilent 8902A tuned RF level function is used to measure low level signals.
Things To Check In Case Of Problems
•The HP/Agilent 8902A/11722A must be calibrated and zeroed for accurate
measurementresults.The HP/Agilent 11722A calibration factors must be entered
into the HP/Agilent 8902A.
•Low level signals from the Agilent 8922M/S must be very stable to be measured
by the tuned RF level function of the HP/Agilent 8902A. The instrument
timebases should be locked together. Also, if the Agilent 8922 source has
unusually high residual FM it may cause the low level measurements to fail
because the HP/Agilent 8902A cannot lock onto the signal.
•Mismatch uncertainty causes a high level of uncertainty in this measurement.
3-7
Verifying Performance
Understanding the Tests
Test 02: Signal
Generator Spectral
Purity
Test 03: Signal Pulse
Modulation
Equipment Required
HP/Agilent 8566B Spectrum Analyzer
Theory of the Test
The UUT is set to generate a CW signal at various levels and frequencies. The HP/
Agilent 8566B is used to measure the signal level and then the level of the
harmonics or spurious signals.
Things To Check In Case Of Problems
•The instruments should have their timebases locked together to assure that the
frequency offsets for the spurious measurements are accurate.
Equipment Required (excluding Agilent 8922M/S)
HP/Agilent 8116APulse Generator
Theory of the Test
The HP/Agilent 8116A pulse generator drives the pulse input of the UUT and
causes it to generate pulsed RF signals. These signals are detected with the negative
diode detector and analyzed with the oscilloscope. The UUT is set to generate two
different on/off ratios: 30 dB and 80 to 90 dB. Rise time and fall time for each of
these is measured.
When the Agilent 8922S is the UUT, the RF Generator is set to generate a pulsed
GMSK signal. This signal is then analyzed with the UUT detector and oscilloscope.
Things To Check In Case Of Problems
•The oscilloscope is used to automatically measure rise and fall times. If the RF
signal has high video feedthrough, it may cause the oscilloscope to make a faulty
reading by triggering on the video feedthrough instead of the actual pulse.
•A diode detector with a greatly different sensitivity may cause the demodulated
waveform to overrange on the oscilloscope or have reduced accuracy because of
the resolution of the oscilloscope.
The HP/Agilent 8904A is used to generate a 270.833 kHz clock and random data.
This drives the UUT RF generator to generate the 0.3 GMSK modulated signals
with random data. This signal is then analyzed by the UUT to assure that the
frequency and phase errors are correct.
If the UUT is an Agilent 8922S, clock and data signals are generated internally,and
the HP/Agilent 8904A is not used.
Things To Check In Case Of Problems
•This test should only be performed if the UUT analyzer is known to be accurate.
Run test #9 first if there is any question about the accuracy of the UUT analyzer.
•Even when the accuracy of the UUT analyzer is known, its specification is such
that this test still has a high degree of measurement uncertainty.
•A possibility to verify an instrument that fails this test may be to use a second
Agilent 8922M/S to analyze the signal or use the HP/Agilent 11836A 0.3 GMSK
measurement software. This software (and its associated hardware) has a lower
measurement uncertainty but is not currently supported with the Agilent 8922
Performance Tests software.
The UUT is set to generate audio signals at various levels and frequencies. The
voltmeter measures AC and DC levelaccuracy, the audio analyzer measures residual
distortion, and the measuring receiver measures frequency accuracy.
Things To Check In Case Of Problems
•The measuring receiver and the UUT should have their timebases locked to make
accurate frequency measurements.
The HP/Agilent 8904A is used to generate accurate test signals which are analyzed
by the UUT.The voltmeter is used to reduce measurement uncertainty by accurately
characterizing the test signal level to predict the correct response of the UUT. The
voltmeter is used in the AC level, DC level, SINAD and distortion measurements.
Distortion test signals are generated by summing two sinewaves in the HP/Agilent
8904A, one as the desired signal, the other as a distortion product.
Things To Check In Case Of Problems
•Forfrequency accuracy measurements the HP/Agilent 8904A timebase should be
locked to the UUT.
•The voltmeter reading is used as a reference to normalize the reading from the
UUT. The output “measured” result is the actual reading from the UUT
normalized by the actual reading from the voltmeter. If the voltmeter makes a
faulty reading, it will normalize the actual UUT response and cause it to appear
faulty also. If the front panel reading from the UUT is significantly different from
the printed “measured” response, the normalization may be the cause.
The audio analyzer is used to generate CW reference signals which are
characterized by the voltmeter and fed into the UUT. The voltmeter reading is used
to normalize the reading from the UUT oscilloscope. The oscilloscope
measurements are taken using the UUT marker function which is set to the peak of
the input sinewave.
Things To Check In Case Of Problems
•The voltmeter reading is used as a reference to normalize the reading from the
UUT. The output “measured” result is the actual reading from the UUT
normalized by the actual reading from the voltmeter. If the voltmeter makes a
faulty reading, it will normalize the actual UUT response and cause it to appear
3-10
Verifying Performance
Understanding the Tests
faulty also. If the front panel reading from the UUT is significantly different from
the printed “measured” response, the normalization may be the cause. If the
oscilloscope triggers incorrectly, the peak search may not find the actual peak
response on the display.
The signal generator is used to generate CW signals at various frequencies and
power levels. These are accurately characterized by the HP/Agilent 8902A/11722A
and used as a reference. The CW signals are then used to test the UUT which is
compared to the reference reading taken by the HP/Agilent 8902A/11722A.
Things To Check In Case Of Problems
•Mismatch uncertainties cause a high level of measurement uncertainty in this
measurement. If the HP/Agilent 11722A or HP/Agilent 8657A/B has a high
SWR, the accuracy of the results may be reduced.
The HP/Agilent 8904A frequency modulates the GMSK signal generator with a
known amount of phase error. This test signal is used to verify the UUT’s ability to
measure this known signal. The actual measurements are compared to the predicted
to determine the accuracy of the UUT. The HP/Agilent 8904A is also used to create
the 270.833 kHz clock signal and the pulse generator is used to create a random data
pattern. Polarity of the data signal from the HP/Agilent 8657A/B Option 022 is not
important in this measurement.
Things To Check In Case Of Problems
•Make certain the option 022 part of the 8657A/B is locked up to the 270.833 kHz
3-11
Verifying Performance
Understanding the Tests
clocksignal. The option does not have GPIBcontrol so the program cannot verify
that the option is operational.
Test A: RF Analyzer
Pulse Demodulation
Test B: Spectrum
Analyzer (Option
006 only)
Equipment Required
HP/Agilent 8657A/B Signal Generator
HP/Agilent 8116A Pulse Generator
Theory of the Test
The Pulse generator drives the signal generator to generate Pulsed RF signals at
various frequencies. The UUT demodulates these signals and the internal
oscilloscope measures the rise and fall time of the demodulated waveform. This test
verifies the demodulator rise and fall time specifications.
Things To Check In Case Of Problems
•Theoscilloscopemeasurement functions are used to measure the 10% to 90% rise
and fall time. If the demodulated waveform has significant video feedthrough or
the oscilloscope cannot trigger the measurements properly, an incorrect
measurement may result.
Equipment Required
HP/Agilent 8657A/B Signal Generator
Theory of the Test
The signal generator is used to stimulate the UUT spectrum analyzer at various
frequencies. At each frequency, the spectrum analyzer is tuned to measure its own
image and spurious responses. The level of the signal generator is set to various
levels to determine the spectrum analyzer’s log-linearity.
Things To Check In Case Of Problems
•The absolute accuracy of the signal generator causes a significant measurement
uncertainty in the log linearity measurements. A more accurate measurement is
to reduce the signal level using a precision attenuator. This is an acceptable
secondary manual test if the UUT fails the automated test.
3-12
Verifying Performance
Understanding Test Failures
Understanding Test Failures
This section is intended to be used in conjunction with the Agilent 8922 Series
Assembly Level Repair Guide for assembly level repair and troubleshooting. If a
performance test fails and hardware is suspected, the following table is a guideline
to help identify the hardware assemblies most likely to cause each failure. Before
replacing an assembly, the memory card diagnostics (supplied with the Assembly
Level Repair Guide) should also be run to help verify which assembly is defective.
ASSEMBLIES TESTED (P=Primary S=Secondary)
Performance Test
Sig Gen LevelSSPSPSSSS
Sig Gen Spectral PuritySSSSSSSSS
Sig Gen Pulse ModSPSPSSSSSS
Sig Gen .3 GMSKPSSSSSSSSP
AF GeneratorSS
AF AnalyzerP PS
OscilloscopeSS
RF Analyzer PowerSP
RF Analyzer GMSK ModPSPSS
RF Analyzer Pulse DemodSPSS
Spectrum AnalyzerSSSPS
NOTE:Ifyou have the Agilent 8922M/S Option 010 Multi-Band Test System, refer to the
appropriate Agilent 8922 Multi-Band User’s Guide for more information on
specifications relevant to dual band testing.
GSM900 and
E-GSM900
Functionality
Bit/Frame Error Rate Measurements:Class 1a, 1b, and Class II bits in both raw and residual form.
Also Burst-by-Burst Bit Error Measurement
MS Power Output Level Control:0 to 19 with RF analyzer auto adjust
Broadcast Channel Capability:BCCH + CCCH or BCCH + CCCH + SDCCH/4
Control Channels (SDCCH, FACCH,
SACCH)
Call Control CapabilitiesBS originated call (FS/EFS), MS originated call (FS/EFS),
Traffic ChannelsTCH (FS/EFS)
HSCSD2x1 and 2x2, 9.6 kbs and 14.4 kbs
TimingAuto, manual, uplink-downlink, offset measurement
HoppingTwo independent, user definable MA tables with offsets
Speech Encoding/DecodingFull rate speech
Speech Echo ModeUser selectable delay 0 to 5 sec on Agilent 8922M, fixed
Measurement CoordinationFlexible control of burst type, ARFCN, and timeslot.
SACCH MEAS ResultRXLEV, RXQUAL, and timing advance.
SMS Cell BroadcastTwo predefined messages with user definable attributes
1. GSM frequency bands are 880 to 915 MHz and 925 to 960 Mhz.
2. Level accuracy degrades 0.2 dB when using the RF in/out connector for both RF generator and RF analyzer. In 30dB
pulse mode, level accuracy specifications are typical.
200 milliwatts
−4 dBm
3-15
Verifying Performance
Agilent Technologies 8922M/S Specifications
Spectral Purity
Spurious Signals (for ≤+1 dBm output level at Aux RF Out or ≤−19 dBm output level at RF In/Out.
Harmonics:<
Non-harmonics:<
−25 dBc
−50 dBc, >5 kHz offset from carrier
0.3 GMSK Modulation
After one timeslot, 577µs, from an isolated RF Generator Trigger in the GSM
Frequency bands.
±10 kHz
±3 dB
±0.6 dB ± noise effects (+4 to +41 dBm) (0.2 mW)
±0.6 dB ± noise effects (-5 to +4 dBm) (0.2 mW)
0.2 dB
Power Measurement Accuracy
Accuracy (dB)
0.5
0.4
Verifying Performance
Agilent Technologies 8922M/S Specifications
0.3
0.2
0.1
0
Pulse ON/OFF Ratio Measurement (Requires Option 006)
ON power is averaged over the useful part of the burst. OFF is averaged over a one
bit interval centered at a user specified time. Non-hopped mode only.
Input Frequency Setting Error
Input Level Setting Error
Timing Accuracy
-505101520253035
Power (dBm)
±10 kHz
±3 dB
±1.7 µs (± 1.1 µs typical)
3-19
Verifying Performance
Agilent Technologies 8922M/S Specifications
Accuracy (ON/OFF 40 dB, RF In/Out only)
OFF Power (dBm)ON/OFF Ratio Accuracy
−30 to −1±2.4 dB±1.1 typically
−37 to −30±2.9 dB±1.3 typically
−42 to −37±3.7 dB±1.7 typically
−47 to −42±4.2 dB±2.1 typically
Amplitude Envelope Measurement
After one timeslot, 577µs, from an isolated Receiver Hop Trigger in the GSM
Frequency bands
Measurement Range
RF In/Out
Aux RF In
Input Frequency Setting Error
−6 to +41 dBm
−36 to 20 dBm
±10 kHz
Inaccuracy due to Noise (for overshoots ≤ 1 dB):
0 dB
Relative
Level
±1 dB±3 dB±3 dB w/5 averages
<
±0.15 dB pk<±0.2 dB pk<±0.2 dB pk
Input Level Setting Error
−6 dB<±0.2 dB<±0.3 dB<±0.3 dB
−30 dB
<+3.0 dB<+4.2 dB<+2.2 dB
−3.8 dB−7.5 dB−2.6 dB
3-20
Verifying Performance
Agilent Technologies 8922M/S Specifications
Phase and Frequency Measurements
After one timeslot, 577µs, from an isolated Receiver Hop Trigger in the GSM
Frequency bands.
Range
RF In/Out
Aux RF In
Input Frequency Setting Error
Input Level Setting Error
RMS phase error accuracy
Peak phase error accuracy
Frequency error accuracy
Supplemental Characteristics
After three timeslots, 1.73 ms, from an isolated Receiver Hop Triggering the GSM frequency band.