Agilent 8920B Users Guide

HP 8920B RF Communications Test Set
User’s Guide
Firmware Version B.06.00 and above
POWE
OF O
!
MAX POWER
SCREEN CONTROL
USER
k1’
INCR
k1
k2’ k2 k3’ k3
ASSIG
k4
RELEA
k5
SHIFT
CANCE
MIC/
!
ANT IN
MAX POWER 200
DUPLEX OUTRF IN/OUT
HP Part No. 08920-90221
Printed in U. S. A.
October 1999
Rev. D
CONFIHELPMS SG HOLD PRINT DUPLETXRX PREV TESTS
DATA FUNCTIONS
REF
METER
AVG
INCR
INCR
LO HI
CURSOR CON-
PUSH TO
ON/OFF
INSTRUMENT STATE
ADRS
SAVE
LOCAL
RECAL
DATA
789 456
123
_
+
0
NO
YES
AUDIO SQUELCVOL-
MAX
!
MEAS
ENTER
dB
GHz
%
MHz
s
kHz
Hzms% Ωppm
AUDIO IN
!
PRESE
MEMO
LOHI
MAX
1
Copyright © Hewlett-Packard Company 1995
Notice Information contained in this document is subject to change without notice.
All Rights Reserved. Reproduct ion, adaptation, or translati on without prior written permission is prohibited, except as allowed under the copyright laws.
This material may be reproduced by or for the U.S. Government pursuant to the Copyright License under the clause at DFARS 52.227-7013 (APR 1988).
Hewlett-Packard Company Learning Products Department 24001 E. Mission Liberty Lake, WA 99019-9599 U.S.A.
2
Manufacturer’s Declaration
This statement is provi ded to c omply with the req uir ements of the Ger man So und Emission Directive, from 18 January 1991.
This product has a sound pressure emission (at the operator position) < 70 dB(A).
Sound Pressure Lp < 70 dB(A).
At Operator Position.
Normal Operation.
According to ISO 7779:1988/EN 27779:1991 (Type Test).
Herstellerbescheinigung
Diese Information steht im Zusammenhang mit den Anforderungen der
Maschinenlärminformationsverordnung vom 18 Januar 1991.
Schalldruckpegel Lp < 70 dB(A).
Am Arbeitsplatz.
Normaler Betrieb.
Nach ISO 7779:1988/EN 27779: 19 91 (T ypp rüfung).
3
Safety Considerations
GENERAL This product and related doc umentation must be re viewed for famil iarizatio n with
safety markings and instructions before operation. This product has been designed and tested in accordance with IEC Publication
1010, "Safety Requirements for Electronic Measuring Apparatus," and has been supplied in a s af e c ondi ti on. This instructi on d ocumentation contains i nformation and warnings which must be followed by the user to ensure safe operation and to maintain the product in a safe condition.
SAFETY EARTH GROUND A uninterruptible safety earth ground must be provided from the main power
source to the product input wiring terminals, power cord, or supplied power cord set.
CHASSIS GROUND TERMINAL To prevent a potential shoc k hazard, always co nnect the rear -panel chassi s ground
terminal to earth ground when operating this instrument from a dc power source. SAFETY SYMBOLS Indicates instrument damage ca n occur if ind icate d opera ting l imits are exc eeded.
!
Refer to the instructions in this guide. Indicates hazardous voltages. Indicates earth (ground) terminal
WARNING A WARNING note denotes a hazard. It calls attention to a procedure, practice, or the
like, which, if not correctly performed or adhered to, could result in personal injury. Do not proceed beyond a WARNING sign until the indicated conditions are fully un­derstood and met.
CAUTION A CAUTION note denotes a hazard. It calls attention to an operation procedure, practice,
or the like, which, if not correctly performed or adhered to, could resu lt in damage to or destruction of part or all of the product. Do not proceed beyond an CAUTION note until the indicated conditions are fully understood and met.
4
Safety Considerations for this Instrument
WARNING This product is a Safety Class I instrument (provided with a protective
earthing ground incorporated in the power cord). The mains plug shall only be inserted in a socket outlet provided with a protective earth contact. Any interruption of the protective conductor inside or outside of the product is likely to make the product dangerous. Intentional interruption is prohibited..
Whenever it is likely that the protection has been impaired, the instrument must be made inoperative and be secured against any unintended operation.
If this instrument is to be energized via an autotransformer (for voltage reduction), make sure the common terminal is connected to the earth terminal of the power source.
If this product is not used as specified, the protection provided by the equipment could be impaired. This product must be used in a normal condition (in which all means for protection are intact) only.
No operator serviceable parts in this product. Refer servicing to qualified personnel. To prevent electrical shock, do not remove covers.
Servicing instructions are for use by qualified personnel only. To avoid electrical shock, do not perform any servicing unless you are qualified to do so.
The opening of covers or removal of parts is likely to expose dangerous voltages. Disconnect the product from all voltage sources while it is being opened.
Adjustments described in the manual are performed with power supplied to the instrument while protective covers are removed. Energy available at many points may, if contacted, result in personal injury.
The power cord is connected to internal capacitors that my remain live for 5 seconds after disconnecting the plug from its power supply.
For Continued protection against fire hazard, replace the line fuse(s) only with 250 V fuse(s) or the same current rating and type (for example, normal blow or time delay). Do not use repaired fuses or short circuited fuseholders.
5
WARNING: Always use the three-prong ac power cord supplied with this product. Failure to
ensure adequate earth grounding by not using this cord may cause product damage.
This product is designed for use in Installa t ion Category II and Pollution Degree 2 per IEC 1010 and IEC 664 respectively. For indoor use only.
This product has autoranging line voltage input, be sure the supply voltage is within the specified range.
To prevent electrical shock, disconnect instrument from mains (line) before cleaning. Use a dry cloth or one slightly dampened with water to clean the external case parts. Do not attempt to clean internally.
Ventilation Requirements: When installing t he pro duct in a cabinet, the convection into and out of the product must not be restricted. The ambient temperature (outside the cabinet) must be less than the maximum operating
temperature of the product by 4° C for every 100 watts dissipated in the cabinet. If the total power dissipated in the cabinet is greater than 800 watts, then forced convection must be used.
Product Markings CE - the CE mark is a regist ered trademark of the European Community. A CE
mark accompanied by a year indicated the year the design was proven. CSA - the CSA m ark is a regis tered trade mark of the Canadian Standards Associ-
ation.
CERTIFICATION Hewlett-Packard Company certifies that this product met its published specifica-
tions at the time of shipment from the factory. Hewlett-Packard further certifies that its calibration measurements are traceable to the United States National In-
stitute of Standards and Technology, to the extent allowed by the Institute’s cali­bration facility, and to the calibration facilities of other International Standards Organization members
6
Hewett-Packard Warranty Statement for Commercial Products
HP 8920B RF Communications Test Set
Duration of Warranty: 1 year
1. HP warrants HP hardware, accessories and supplies against defects in materials and
workmanship for the period specified above. If HP receives notice of such defects during the warranty period, HP will, at its option, either repair or replace products which prove to be defective. Replacement products may be either new or like-new.
2 HP warrants that HP software will not fail to execute its programming instructions, for
the period specified above, due to defects in material and workmanshi p when properly installed and used. If HP receives notice of such defects during the warranty period, HP will replace software media which does not execute its programming instructions due to such defects.
3. HP does not warrant that the operation of HP products will be uninterrupted or error
free. If HP is unable, within a reasonable time, to repair or replace any product to a condition as warranted, customer will be entitled to a refund of the purchase price upon prompt return of the product.
4 HP products may contain remanuf actured parts equival ent to new in performance or
may have been subject to incidental use.
5. The warranty period begins on the date of deliv ery or on the date of insta llation if
installed by HP. If customer schedules or delays HP installation more than 30 days after delivery, warranty begins on the 31st day from delivery.
6 Warranty does not apply to defects resulting from (a) improper or inadequate mai nte-
nance or calibration, (b) software, interfacing, parts or supplies not supplied by HP, (c) unauthorized modification or misuse, (d) operation outside of the published environ­mental specifications for the product, or (e) improper site preparation or maintenance.
7 TO THE EXTENT ALLOWED BY LOCAL LAW, THE ABOVE WARRANTIES
ARE EXCLUSIVE AND NO OTHER WARRANTYOR CONDITION, WHETHER WRITTEN OR ORAL IS EXPRESSED OR IMPLIED AND HP SPECIFICALLY DISCLAIMS ANY IMPLIED WARRANTIES OR CONDITIONS OR MERCHANT­ABILITY, SATISFACTORY QUALITY, AND FITNESS FOR A PARTICULAR PURPOSE.
8 HP will be liable for damage to tangible property per incident up to the greater of
$300,000 or the actual amount paid f or the product th at is the subject of the claim, and for damages for bodily injury or death, to the extent that all such damages are deter­mined by a court of competent jurisdiction to have b een directly caused b y a defective HP product.
7
9. TO THE EXTENT ALLOWED BY LOCAL LAW, THE REMEDIES IN THIS
WARRANTY STATEMENT ARE CUSTOMER’S SOLE AND EXCLUSIVE REMEDIES. EXCEPT AS INDICATED ABOVE, IN NO EVENT WILL HP OR ITS SUPPLIERS BE LIABLE FOR LOSS OF DATA OR FOR DIRECT, SPECIAL, INCIDENTAL, CONSEQUENTIAL (INCLUDING LOST PROFIT OR DATA), OR OTHER DAMAGE, WHETHER BASED IN CONTRACT, TORT, OR OTHERWISE.
FOR CONSUMER TRANSACTIONS IN AUSTRALIA AND NEW ZEALAND: THE WARRANTY TERMS CONTAINED IN THIS STATEMENT, EXCEPT TO THE EXTENT LAWFULLY PERMITTED, DO NOT EXCLUDE RESTRICT OR MODIFY AND ARE IN ADDITION TO THE MANDATORY STATUTORY RIGHTS APPLICABLE TO THE SALE OF THIS PRODUCT TO YOU.
ASSISTANCE Product maintenance agreements and other customer assistance agreements are
available for Hewle tt-Packard products . For any assistance , contact your near est Hewlett-Packard Sales and Service Office.
8
DECLARATION OF CONFORMITY
Manufacturer’s Name:
according to ISO/IEC Guide 22 and EN 45014
Hewlett-Packard Co.
Manufacturer’s Address:
Spokane Division 24001 E. Mission Avenue Liberty Lake, Washington 99019-9599 USA
declares that the product
Product Name: Model Number: Product Options:
RF Communications Test Set / Cell Site Test Set HP 8920A, HP 8920B, and HP 8921A This declaration covers all options of the above
product.
conforms to the following Product specifications:
Safety: IEC 348:1978 / HD 401S1:1981
1)
EMC: CISPR 11:1990 / EN 55011:1991 Group 1, Class A
EN 50 08 2-1 : 1992 IEC 801-2:1991 - 4 kV CD, 8 kV AD IEC 801-3:1984 - 3V/m IEC 801-4:1988 - 0.5 kV Sig. Lines, 1 kV Power Lines
Supplementary Information:
This product herewith complies with the requirements of the Low Voltage Directive 73/23/EEC and the EMC Directive 89/336/EEC and carries the CD-marking accordingly
1) The product was introduced prior to 12/1993 Spokane, Washington USA Date October 17, 1996 Vince Roland/Quality Manager
European Contact: Your local Hewlett-Packard Sales and Service Office or Hewlett-Packard GmbH Department ZQ/Standards Europe, Herrenberger Strasse 130, D-71034 B
9
ö
blinger, Germany (FAX+49-7031-14-3143)
.
Table 1 Regional Sales and Service Offices
United States of America U.S. Instrument Suppor t Center For Test & Measurement Equipment Repair & Calibration. Hewlett-Packard Company Englewood, Colorado 80112
Telephone: (800) 403-0801 Fax: (888) 857-8161
South Eastern Europe Sales and Service Hewlett-Packard Ges. m.b.h. Liebigasse 1 P.O. Box 72 A-1222 Vienna, Austria
Telephone: 43 222 2500 0 Telex: 13 4425
Asia Sales and Service Hewlett-Packard Asia Ltd. 22-30/F Pere gr i ne T ow e r Lippo Center 89 Queensway, Central Hong Kong G.P.O. Box 863 Hong Kong
United States of America Customer Information Center For Assistance On All HP Prod­ucts. Hewlett-Packard Company Tel: (800) 752-0900
6:00 am to 5:00 pm Pacific Time Parts Direct: 1-800-227-8164
European Multicountry Region Sales and Service Hewlett-Packard S.A. P.O. Box 95 150, Route dv Nant_dl_AVRIL CH-1217 Meyrin 2 Geneva, Switzerland
Telephone: (41/22) 780-8111 Fax: (41/22) 780-8542
Japan Hewlett-Packard Japan, Ltd. Measurement Assistance Center 9-1, Takakura-Cho, Hachioji-Shi Tokyo 192-8510, Japan
Telephone:(81)-426-56-7832 Fax: (81)-426-56-7840
United Kingdom Sales and Service Hewlett-Packard Ltd. Cain Road Amen Corner Bracknell, Berkshire RG12 1HN United Kingdom
Telephone: 44 344 360000 Fax: 44 344 363344
Northern Europe Sales and Service Hewlett-Packard Nederland B.V. Startbaan 16 1187 XR Amstelveen, The Netherlands P.O. Box 667
Telephone: 31/20 5476911 X 6631 Fax: 31-20-6471825NL
International Sales Branch Head­quarters Sales and Service Hewlett-Packard S.A. 39 Rue Veyrot P.O. Box 365 1217 Meyrin 1 Geneva, Switzerland
Telephone: 852-848-7777 Fax: 852-868-4997
Australia, New Zealand Sales and Service Hewlett-Packard Ltd. P.O. Box 221 31-41 Joseph Street Blackburn, Victoria 3130
Telephone: (61/3) 895-2895 Fax: (61/3) 898-9257
10
Canada Sales and Service Hewlett-Packard (Canada) Ltd. 5150 Spectrum Way Mississauga, Ontario L4W 5G1 Canada
Telephone: (416) 206-4725 Fax: (416) 206-4739
Telephone: 41-22-780-4111 Fax: 41-22-780-4770
Canada Service Center Hewlett-Packard Company 17500 Transcanada Highway S. Serv Road Kirkland, Quebec H9J 2X8 Canada
Telephone: (416) 206-3295
Table 1 Regional Sales and Service Offices (Continued)
Canada Service Center Hewlett-Packard Ltd. 11120 178 Street Edmonton, Alberta T5S 1P2 Canada
Telephone: (403) 486-6666 Fax: (403) 489-8764
Latin America Hewlett-Packard Company LAHQ Mexico City Col. Lomas de Virreyes 11000 Mexico D.F. Mexico
Telephone: (52/5) 326-4000 Fax: (52/5) 202 7718
United Kingdom Sales and Service Hewlett-Packard Ltd. Cain Road Amen Corner Bracknell, Berkshire RG12 1HN United Kingdom
Telephone: 44 344 360000 Fax: 44 344 363344
11
In this Book The HP 8920B is referred to in this docum ent as "Test Set."
Chapter 1, Get Started
This chapter describes the basic operation of the Test Set. It also provides a quick check that verifies that the Test Set is operating properly.
Chapter 2, Configuring Your Test Set
This chapter describes various instrument configuration settings that affect the general operation of the instrument.
Chapter 3, Operating Overview
This chapter contains detailed op erating inst ructions and examples for u sing several instrument features.
Chapters 4 through 23, Screen and Field Descriptions
These chapters contains reference informa tion fo r each screen an d its fields . Man y of
the descriptions contain signal flow diagrams that relate the screen’s f ields to the func­tions they perform. The screens are arranged in alphabetical order by title at the top of the screen; Signaling Encoder and Signaling Decoder are alphabetized by the names Encoder and Decoder.
Chapter 24, Connector, Key, and Knob Descriptions
This chapter describes the purpose and use of each connector and control.
Chapter 25, Modifications, Accessories, Manuals, Support
This chapter describes retrofit kits, accessories, manuals, and customer support avail­able for your Test Set.
Error Messages
This section discusses error and operating messages.
12

Contents

1 Get Started
Before Connecting a Radio 40
Accessing the Test Set’s Screens 41
Changing A Field’s Setting 43
How do I Verify that the Test Set is Operating Properly? 46
Instrument Functional Diagram 47
13
Contents
2 Configuring Your Test Set
General Oper ating Information 50
14
Contents
3 Operating Overview
Interaction Between Screens 54
Displaying Measurements 58
Entering and Changing Numbers 64
Printing A Screen 67
Using Measurement Limit Indicators 68
Averaging Measurements 70
Setting A Measurement Reference 71
Using Memory Cards 72
Saving and Recalling Instrument Setups 77
Using USER Keys 81
Setting an RF Generator/Analyzer Offset 85
Using Remote Control 86
15
Contents
4 Adjacent Channel Power Screen
How the Test Set Measures Adjacent Channel Power (ACP) 92
Field Descriptions 93
16
Contents
5 AF Analyzer Screen
Block Diagram 102
17
Contents
6 Call Processing Subsystem
Description of the Call Processing Subsystem 116
Using the Call Processing Subsystem 120
The CALL CONTROL Screen 124
Using the CALL CONTROL Screen to Test Call Processing Functions 142
Using the CALL CONTROL Screen to test AMPS Authentication 153
AUTHENTICATION Screen 164
CALL DATA Screen 167
CALL DATA Screen Message Field Descriptions 172
Using the CALL DATA Screen 188
CALL BIT Screen 196
CALL BIT Screen Message Field Descriptions 203
Using the CALL BIT Screen 241
ANALOG MEAS Screen 246
Using the ANALOG MEAS Screen 249
CALL CONFIGURE Screen 252
18
Contents
7 Configure Screen
Field Descriptions 256
19
Contents
8 Signaling Decoder Screen
Field Descriptions for Decoder Modes 270
AMPS-TACS, NAMPS-NTACS Decoder 271
Using the AMPS/TACS, NAMPS/NTACS Decoder 276
Continuous Digital Controlled Squelch System Decoder 280
Using the CDCSS Decoder 284
Digital Paging Decoder 285
Dual-Tone Multi-Frequency (DTMF) Decoder 289
Using the DTMF Decoder 294
EDACS Decoder 295
Using the EDACS Decoder 298
Function Generator Decoder 301
Using the Function Generator Decoder 304
LTR Decoder 305
Using the LTR Decoder 309
MPT 1327 Decoder 311
NMT Decoder 316
Using the NMT Decoder/Encoder 321
Creating NMT Tests 326
Tone Sequence Decoder 335
20
Contents
9 Duplex Test Screen
Block Diagram 340
Field Descriptions 341
21
Contents
10 Signaling Encoder (AF Generator 2)
Field Descriptions for Encoder Modes 354
AMPS-TACS NAMPS-NTACS Encoder 355
Using the AMPS-TACS, NAMPS-NTACS Encoder 367
CDCSS Encoder 368
Using the CDCSS Encoder 372
Digital Paging Encoder 373
DTMF Sequence Encoder 378
EDACS Encoder 382
Using the EDACS Encoder 388
Function Generator Encoder 393
LTR Encoder 396
Using the LTR Encoder 400
MPT 1327 Encoder 404
Using the MPT 1327 Encoder 417
Nordic Mobile Telephone (NMT) Encoder 420
Tone Sequence Encoder 432
22
Contents
11 Help Screen
Field Descriptions 436
23
Contents
12 I/O Configure Screen
Field Descriptions 438
24
Contents
13 Message Screen
Field Descriptions 444
25
Contents
14 Oscilloscope Screen
Field Descriptions 446
Using the Oscilloscope 453
26
Contents
15 Print Configure Screen
Field Descriptions 456
27
Contents
16 Radio Interface Screen
Radio Interface Functional Description 460
Field Descriptions 462
Using the Radio Interface (Manual Operation) 465
Using The Radio Interface (Remote Operation) 471
28
Contents
17 RF Analyzer Screen
Block Diagram 474
Field Descriptions 475
29
Contents
18 RF Generator Screen
Block Diagram 486
Field Descriptions 487
30
Contents
19 RX Test Screen
Block Diagram 496
Field Descriptions 497
31
Contents
20 Service Screen
Field Descriptions 504
32
Contents
21 Spectrum Analyzer Screen
Field Descriptions 508
Using the Spectrum Analyzer 520
33
Contents
22 Tests Screen
Description of the Tests Subsystem 522
TESTS (Main Menu) 524
TESTS (Channel Information) 527
TESTS (Test Parameters) 529
TESTS (Order of Tests) 530
TESTS (Pass/Fail Limits) 532
TESTS (Save/Delete Procedure) 534
TESTS (Execution Conditions) 537
TESTS (External Devices) 540
TESTS (Printer Setup) 543
TESTS (IBASIC Controller) 546
ROM Programs 548
34
Contents
23 TX Test Screen
Block Diagram 552
Field Descriptions 553
35
Contents
24 Connector, Key, and Knob Descriptions
Connector Descriptions 566
Key Descriptions 584
Knob Descriptions 588
36
Contents
25 Modifications, Accessories, Manuals, Support
Modifications 590
Accessories 593
HP Support for Your Instrument 605
37
Contents
Index 621
38
1

Get Started

39
Chapter 1, Get Started

Before Connecting a Radio

Before Connecting a Radio
The RF IN/OUT port should be used for all transmitter tests when the radio is connected directly to the Test Set. (All UUT transmitter power measurements are made through this port). Off-the-air measurements can be made using the ANT IN port.
CAUTION: Overpower Damage — Refer to the Test Set’s front panel for max imum input power level.
Exceeding this level can cause permanent instrument damage.
Speaker or Audio OutMic Audio InMic KeyAntenna
Radio Under Test
Figure 1 Connecting a Radio to the Test Set
40
S:\HP892XX\USRGUIDE\manual\getstart.fb

Accessing the Test Set’s Screens

List of Screens

The following table lists all the screens that could be provided by the Test Set.
Table 2
Chapter 1, Get Started
Accessing the Test Set’s Screens
Analog Measurement Screens
Adjacent Channel Power Configure AF Analyzer I/O Configure Decoder Print Configure Duplex Encoder Help Oscilloscope Message Radio Interface RF Analyzer Service RF Generator RX Test Call Control Spectrum Analyzer Call Data TX Test Call Bit
Software Control Screens
Instrument Configuration
User Assistance Screens
Service Assistance Screen
Call Processing Screens
Call Configure
Screens
Tests Analog Measure Tests (IBASIC Controller) Authentication
41
Chapter 1, Get Started
Accessing the Test Set’s Screens

Accessing Screens

Test Set’s screens can be accessed through
Front-panel keys
The front-panel Cursor Control knob (using the To Screen menu, see item 5 in fig-
ure 3 on page 43)
HP-IB, using the DISPlay subsystem.
Knob
Access to
Additional
Screens
Knob Access to
Screens
Front-Panel Key
Access to Screens
Figure 2 Accessing the Screens
42
S:\HP892XX\USRGUIDE\manual\getstart.fb

Changing A Field’s Setting

There are several types of CRT display fields in the Test Set. This section describes some of the different types of fields.
Chapter 1, Get Started
Changing A Field’s Setting
3
Figure 3 Different Types of Fields
1
rxscrn.wmf
intro4.wmf
542
43
Chapter 1, Get Started
Changing A Field’s Setting

Unit-of-Measure Field

Unit-of-measure can be changed to display measurements in different values or magnitudes. See item 1 in
To change a unit-of-measure
1. Position the cursor at the unit field on the display.
2. Press a key labeled with a different unit-of-measure (such as W).
If the new units are valid, the measurement value is displayed in the unit.

Underlined Immediate-Action Field

Underlined immediate ac tion fiel ds pro vide a choi ce of tw o set tings. Se e it em 2 in
figure 3 to see an example of an underlined immediate-action field.
To change an underlined entry
1. Position the cursor at the field.
2. Push the CURSOR CONTRO L knob or th e EN TER key to move the underline under
the desired choice.
figure 3 to see an example of a units-of-measure field.
The underlined setting is immediately activated when selected.
44
S:\HP892XX\USRGUIDE\manual\getstart.fb

One-of-Many Field

Numeric-Entry Field

Chapter 1, Get Started
Changing A Field’s Setting
One-of-many fields displ ay a lis t of choi ces when selected. See item 3 in figure 3 to see an example of a one-of many field.
To make a one-of-many choice
1. Position the cursor at the field.
2. Push the Cursor Control knob or the ENTER key to display the choices.
3. Move the cursor through the choices by turning the knob.
4. Push the Cursor Control knob or the ENTER key to make the choice.
The choice is immediately activated when selected.
The To Screen menu (see item 5 in figure 3 ) is a variation of the one-of-many field.
Numeric-entry field s contain val ues for sett ings like Ext ernal Load Resi stance and RF Generator Frequency.See item 4 in
figure 3 to see an example of a numeric-
entry field.
To change a value
1. Position the cursor at the field.
2. Key in the desired number using the DATA keys.
3. Press ENTER to select the choice.
OR
1. Position the cursor at the field.
2. Push the Cursor Control knob to highlight the desired choice.
3. Turn the knob to increment or decrement the value.
4. Push the Cursor Control knob or the ENTER key to select the choice.
45
Chapter 1, Get Started
How do I Verify that the Test Set is Operating Properly?
How do I Verify that the Test Set is Operating Properly?
If your Test Set powers-up and displays the RX TEST screen, but you suspect an instrument problem, use the Instrument Quick Check to verify operation of the basic instrument functions.
If no failure is indicated by this test, but you still suspect a problem, refer to the
“Performance Tests” information in the Assembly Level Repair Manual.

Instrument Quick Check

1. Set up the quick check: a. Connect a cable between the DUPLEX OUT and ANT IN ports.
b. Turn instrument power on (if it is not already on). c. Press PRESET. d. Press DUPLEX to access the DUPLEX TEST screen. e. Set the Tune Mode field to Manual f. Set the Tune Freq field to 825 MHz. g. Set the Input Port field to Ant. h. Set the RF Gen Freq field to 825 MHz. i. Set the Amplitude field to -10 dBm. j. Set the Output Port field to Dupl k. Verify that AFGen1 Freq is set to 1.0000 kHz, and that AFGen1 To is set to
FM and 3.00 kHz.
l. Set the AF Anl In field to FM Demod. m. Set the Filter 1 field to 300Hz HPF. n. Set the Filter 2 field to 3kHz LPF. o. Verify that De-Emphasis is Off p. Set the Detector field to Pk+-/2. q. Turn the VOLUME knob clockwise until you hear a tone (1 kHz default for
AFGen1 Freq).
.
.
.
2. Check the following readings:
SINAD should be >35 dB. FM Deviation should be about 3.0 kHz.
3 Access the OSCILLOSCOPE screen using the To Screen menu. With the default
Vert/div setting of 2 kHz and a default Time/div setting of 200 µs
see two complete sinewaves across the screen.
1
4 Access the SPECTRUM ANALYZER
an 850 MHz FM carrier.
1. O ptional on some Test Set models.
46
S:\HP892XX\USRGUIDE\manual\getstart.fb
using the To Screen menu. You shou ld see
, you should

Instrument Functional Diagram

Chapter 1, Get Started
Instrument Functional Diagram
Figure 4 Instrument Functional Diagram (1 of 2)
47
Chapter 1, Get Started
Instrument Functional Diagram
intr-bd2.wmf
Figure 5 Instrument Functional Diagram (2 of 2)
48
S:\HP892XX\USRGUIDE\manual\getstart.fb
2

Configuring Your Test Set

The CONFIGURE and I/O CONFIGURE screens contain a number of settings used to alter instrument operation and hardware communication settings. The HP-IB address, screen intensity, serial communication parameters , and sever al other settings , are changed in these screens.
Most CONFIGURE and I/O CONFIGURE screen entries are saved when the instrument is turned off.
49
Chapter 2, Configuring Your Test Set

General Operating Information

General Operating Information
The following configuration information discusses general operating information for some of the fields in these screens.

To Set Screen Intensity

1. Access the CONFIGURE screen.
2. Select the Intensity field.
3. Rotate the knob to change the setting (1=dim, 8=bright).

To Set RF Voltage Interpretation (50 /emf)

1. Access the CONFIGURE screen.
2. Position the cursor in front of the RFGen Volts field.
3. Press the Cursor Control knob or press ENTER to select 50 ohm or emf.
Voltage settings can control either:
the voltage across a 50-ohm load, or
the open circuit voltage (emf).
This setting affects the RF Generator’s and the Tracking Generator’s amplitudes.

To Set the Date and Time

1. Access the CONFIGURE screen.
2. Select the Date field and use the DATA keys to enter the d ate in the fo rmat shown be-
low the field.
3. Select the Time field and use the DATA keys to enter the time in the format shown below the field.
The Test Set has a built-in clock that keeps track of the date and time. It is powered by an internal battery to keep it operating when the instrument is off.
50
S:\HP892XX\USRGUIDE\manual\configts.fb

To Change the Beeper Volume

1. Access the CONFIGURE screen.
2. Select the Beeper field to display the volume choices.
3. Select the desired choice.
The beeper alerts you to important operating and measurement conditions. It beeps any time a message is displayed at the top of the screen. These messages warn you of conditio ns such as ex ceeding th e RF input level or t rying to s et a field to an unacceptab le value. Therefore, it is recommende d that y ou do not disable the beeper.

To Verify or Change the Low-Battery Setting

1. Access the CONFIGURE screen.
Chapter 2, Configuring Your Test Set
General Operating Information
2. The current time setting is shown under the Low Battery field.
3. Select that field to display a list of setting choices.
Select the desired time, or
Select Disable to eliminate the low-battery warning.
The low-battery warning system is used to alert you when you have not used any front-panel controls within a specified amount of time. This setting is only used with DC power. It does not actually monitor the DC supply voltage. Since batteries are most often used for a DC supply, this function helps you conserve power by reminding you that the Test Set is still turned on.
When the specified time has elapsed between front-panel entries, the Beeper sounds and a message appears at the top of the screen alerting you to the condition.
This setting is saved when the instrument is turned off.
51
Chapter 2, Configuring Your Test Set
General Operating Information
52
S:\HP892XX\USRGUIDE\manual\configts.fb
3

Operating Overview

The information in this section discusses some fre que ntl y used operating features of the Test Set.
From reading
What “fields” and “screens” are.
How to use the Cursor Control knob to select different fields and screens.
chapter 1, "Get Started" you should understand:
53
Chapter 3, Operating Overview

Interaction Between Screens

Interaction Between Screens
Most fields operate globally; changing the setting in any screen automatically changes that setting in all screens where it is available. example of this field type.
AFGen1 Freq is an
Figure 6 Example of How Global Fields Work
Priority fields give the RX TEST and TX TEST screens priority control of their settings. No matter wh at these fields were set to in other screens, if the RX TEST or TX TEST screen is accessed, the field changes to whatever it was last set to in
these scree ns. The RF Generator’s
Amplitude field is an example of this field
type. These fields and their preset values are listed in
54
S:\HP892XX\USRGUIDE\manual\opoverv.fb
table 3.
Chapter 3, Operating Overview
Interaction Between Screens
Table 3 Priority RX TEST and TX TEST Fields
Priority Field RX TEST TX TEST
RF Gen Amplitude Presets to 80 dBm (changeable) Always Off AFGen1 To Presets to FM (changeable) Always Audio Out AF Anl In Always Audio In Presets to FM Demod (changeable) Detector Always RMS Presets to Pk + Max (changeable) De-emphasis Always Off
AF Anl Measurement Presets to SINAD (changeable) Presets to Audio Freq (changeable)
Presets to 750 (changeable)
µs
Using your Test Set, duplicate the steps in figure 7 to demonstrate how the Priority fields oper ate .
55
Chapter 3, Operating Overview
Interaction Between Screens
fig2-2.wmf
Figure 7 Example of How Priority Fields Work
56
S:\HP892XX\USRGUIDE\manual\opoverv.fb

Priority CALL CONFIGURE Field

The Input Att field located on the CALL CONFIGURE screen is another priority field. Similar fields labeled ADJACENT CHANNEL POWER screen, RF ANALYZER screen, and the SPECTRUM ANALYZER screen.
Chapter 3, Operating Overview
Interaction Between Screens
Input Atten are located on the
Setting the
Input Atten fields found on the other screens. Setting the Input Atten field
updates all other
Input Att field. Accessing the CALL CONFIGURE screen changes the input
attenuator to the current value of the
Atten
Input Att field from the CALL CONFIGURE screen updates all
Input Atten fields but does not change the setting of the
Input Att field and updates all Input
fields.
57
Chapter 3, Operating Overview

Displaying Measurements

Displaying Measurements
Figure 8 Where To Access Measurements
58
S:\HP892XX\USRGUIDE\manual\opoverv.fb
scntxrx.wmf

Displaying RF Measurements

Transmitter Frequency
Chapter 3, Operating Overview
Displaying Measurements
TX Frequency
figure 8 on page 58.)
in
Transmitter Frequency Error
TX Freq Error
figure 8 on page 58.)
(1) in
Transmitter Power
TX Power RF In (Refer to item (2) in figure 8 on page 58). If Ant (antenna) is selected, the
is displayed when Tune Mode is set to Auto. (Refer to item (1)
is displayed when Tune Mode is set to Manual. (Refer to item
is only measured and displayed here when the Input Port is set to
measurement is replaced by four dashes (- - - -). You can measure low power levels on the ANT IN port using the Spectrum
Analyzer.
1
Refer to the TX Power field description, on page 561 and the TX Pwr Zero field
description, on page 563
CAUTION: Connecting a signal of >200 mW to the ANT IN (antenna) port can cause instrument damage
(although internal protection circuits can typically withstand a short-duration signal of 1 or 2 Watts). If the overpower circuit is triggered, remove the signal from the ANT IN port and turn the Test Set off and on to reset it.
for more information on measuring transmitter power.
1. O ptional on some Test Set models.
59
Chapter 3, Operating Overview
Displaying Measurements

Displaying AF Measurements

FM Deviation, AM Depth, AC Level
The AF Anl In setti ng determine s the AF Anal yzer’s inp ut and the measureme nt displayed in the top-right corner of the measurement area (see measurements are available in the TX TEST, DUPLEX TEST, RF GENERATOR, RF ANALYZER, and AF ANALYZER screens. (Refer to item
figure 8 on page 58.)
(3) in
Table 4 AF Measurements Selected by AF Analyzer Input Setting
Measurement AF Anl In Setting
FM Deviation FM Demod, FM Mod
table 4). These
AM Depth AM Demod, AM Mod AM Depth
a. AC Level is also measured in the RX TEST screen, but
a
always uses the AUDIO IN connector as the input. (Refer to item (5) in figure 8 on page 58.)
SSB Demod, AudioIn, Radio Int, Ext Mod, Mic Mod, Audio Out
60
S:\HP892XX\USRGUIDE\manual\opoverv.fb
Chapter 3, Operating Overview
Displaying Measurements
SINAD, Distortion, SNR, AF Frequency, DC Level, DC Current
Selecting the curr ently-displaye d measurement c auses the To Sc reen menu to be replaced by a list of measuremen t choices. Select the new choice to repla ce the old measurement. These measurements are available in the RX TEST, TX TEST, DUPLEX TEST, RF GENERATOR, RF ANALYZER, and AF ANALYZER screens. (Refer to item (4) in
figure 8 on page 58)
The Distortion measurement is only for a 1 kHz tone. The SINAD measurement is normally shown using an analog-type meter and
small digits, but can be changed to display in large digits only. (See
Analog METER Format" on page 63
.)
"To Use the
DC Current can only be measured using the rear-panel DC CURRENT MEASUREMENT connections.
Selecting SNR (Signal/Noise Ratio) turns off the other audio measurement. For more information on making this measurement, see the
description, on page 501
.
RF Gen Freq field
AF Power
AF Power is measured in the RX TEST screen by specifying the external load resistance,
Ext Load R, and changing the unit of measure for the AC Level
measurement to W (W att s), m W, or dBm . (Th e mi lliwa tt ( mW) unit is sel ected by pressing SHIFT, ENTER) Refer to item (5)
figure 8 on pa ge 58.
61
Chapter 3, Operating Overview
Displaying Measurements

To Change the Measurement’s Unit-of-Measure

1. Position the cursor in front of the present unit-o f-measurement.
2. Press the key labeled with the desired unit.
All measurements allow you to change the associated unit-of- measure. For instance; the changed to display in mW, dBm, V, mV, or dB
Select mW by pressing SHIFT, ENTER. For example; to display transmitter power in units of dBm instead of Watts:
1. Move the cursor in front of the unit-of-measure for the TX Power measurement (W).
2. Press the dBm key. The measurement value is changed immediately to display in dB m.
TX Power measurement is usually displayed in Watts, but can be
µV.
62
S:\HP892XX\USRGUIDE\manual\opoverv.fb

To Use the Analog METER Format

To display measurement results using the analog meter format, use the following procedure.
1. Position the cursor in front of the unit-of-measure for the measurement you want to dis-
play.
2. Press and release the SHIFT key, then the INCR SET key to display the Meters menu
in the lower-right corner of the screen.
3. Select On/Off to display the meter.
4. Repeat steps 1 and 2 to enter each meter end point and the meter intervals.
5. Repeat steps 1, 2, and 3 to cancel the METER function.
The METER function displays an equivalent analog display. (This is the SINAD
measurement’s default state when the instrument is turned on or preset). As the measurement is displayed graphically on the meter, the value is also displayed in small digits below the m eter.
Chapter 3, Operating Overview
Displaying Measurements
You can specify the high and low end points and display interval, or you can use the default m eter settings .
This function is only available for measurements displayed using the large digits, such as the measurements displayed in the RX TEST and TX TEST screens.

To Make Beat Frequency Measurements

1. Select the DUPLEX TEST screen to set up for beat frequency oscillator measurements.
2. Set the AF Anl In field to SSB Demod.
3. Manually adjust the Tune Freq field to the desired carrier frequency.
63
Chapter 3, Operating Overview

Entering and Changing Numbers

Entering and Changing Numbers
Values for numeric entry fields can be entered and changed using various methods, depending on y our testing needs. The unit-of-measure fo r some of these
fields can also be changed (such as changing the RF Generator’s units from dBm to

To Enter Numbers

1. Position the cursor in front of the numeric entry field to be changed.
2. Use one of the following methods:
a. enter the number and unit-of-measure directly using the keypad,
or
µV).
Amplitude
b. press the Cursor Control knob or ENTER to highlight the field, and use the knob,
or c. use the down-arrow or the up-arrow key s to increment or decrement the present
value.
Decimal Values
Decimal values are used for most numeric en try fi elds, suc h as th e RF Gen Freq setting. The acceptabl e e ntr ies for decimal values a re 0 thr ough 9, ., +/-, and EEX.
The +/- key is used for entering negative numbers. For example; when entering the RF Generator
47 dBm: +/- 4 7 dBm.
Amplitude you can enter this sequence to set the value to
The EEX key can be used when entering exponential notation. For example; to
3
× 10
enter 1.25
kHz you could use the sequence: 1 . 2 5 EEX 3 kHz.
64
S:\HP892XX\USRGUIDE\manual\opoverv.fb
Hexadecimal Values
Hexadecimal (Hex) values are used for entering some signaling parameters in the ENCODER, such as AMPS communications parameters, such as the RADIO INTERFACE field. The acceptable entries for decimal values are 0 through 9 and A through F. No unit-of-measure is associated with these values.
Hexadecimal values are either entered from the keypad (A through F are shifted functions), or by using the selected (such as the AMPS

To Enter and Change the Unit-of-Measure

Entering the Unit-of-Measure for Settings
When a number is entered, the unit-of-measure is either specified or implied.
Chapter 3, Operating Overview
Entering and Changing Numbers
Filler data field, and for specifying remote
Output Data
Choices menu displayed when certain fields are
Filler field).
When the unit is implied, the current unit is used. For example; if the present RF frequency is 250 MHz, and you want to change it to 225 MHz, you would enter this sequence: 2 2 5 ENTER.
When the unit is specif ied, the units change to whatever you specify. For ex ample; if the present
RF Gen Freq setting is 250 MHz, a nd you want to cha nge it to 455
kHz, you would enter this sequence: 4 5 5 kHz.
Changing the Unit-of-Measure for Settings
To change the present unit-of- measure , position the cursor in front of the field and press the key labeled with the desired unit. For example, position the cursor in front of the
RF Gen Freq field and push GHz or kHz to display the setting in
either of these units.
65
Chapter 3, Operating Overview
Entering and Changing Numbers

To Change the Increment or Decrement Setting

Using the Pre-Defined Increment/Decrement Keys
The INCR ×10] and INCR ÷10] keys change the increment/decrement value by a factor of 10.
For example; if the the knob or push of the down-arrow or up-arrow keys, pushing INCR
Tune Freq presently changes by 10 MHz for every click of
×10] once
changes the increment value to 100 MHz.
Specifying An Increment Value
The INCR SET key is used to assign a specific increment value. The increment value may use differen t units than the field you are incrementing/decrementing. For instance; if the RF Generator
Amplitude setting is displayed in dBµV, you
could increment in units of dB or mV. To change the increment value;
1. Move the cursor to the numeric entry field to be changed.
2. Press INCR SET, and enter the desired value and unit-of-measure using the DATA
keys.
3. Use the down-arrow and up-arrow keys or CURSOR CONTROL knob to change the
field’s value by the increment value you set.
Example of Setting an Increment Value
This example changes the Tune Freq in increments of 15 MHz.
1. Access the TX TEST screen and position the cursor in front of the Tune Freq field.
2. Press 1 0 0 MHz to set the frequency at 100 MHz.
3. Press INCR SET 1 5 MHz.
4. Turn the Cursor Control knob. The field’s value changes by 15 MHz for each knob
click.
66
S:\HP892XX\USRGUIDE\manual\opoverv.fb

Printing A Screen

To Print A Screen’s Contents

1. Connect a printer to the appropriate rear-panel connector.
2. Access the PRINT CONFIGURE screen from the More menu and set the Printer
Port field to the appropriate type of printer connection.
If HP-IB is selected, enter the HP-IB Printer Address of the printer.
3. Select the type of printer you are using in the Model field. If your printer is not lis ted,
configure your printer to emulate one that is listed.
4. Enter a Print Title using the knob, if desired. This text will appear at the top of
your printout.
Chapter 3, Operating Overview
Printing A Screen
5. Display the screen you want to print and press and release the SHIFT key, then the
TESTS key to access the PRINT function.
To interrupt prin ti ng, s el ect the Abort Print field on the PRINT CONFIGURE screen.
67
Chapter 3, Operating Overview

Using Measurement Limit Indicators

Using Measurement Limit Indicators
The LO LIMIT and HI LIMIT functions are used to define a measurement
“window” to alert you to measurements that are outside these limits. When limits are assigned, Lo and/or Hi appear by the measurement.
A measurement that goes a bove or be low the defined limits causes three things to happen:
1. A message appears at the top of the screen indicating a limit was exceeded.
2. The Lo or Hi indicator by the measurement flashes.
3. The Beeper beeps if it is has been enabled in the CONFIGURE screen.
Limits are helpful when you can’t watch the Test Set’s display while you are making an adjustment on the eq uipment you are test ing or r epairing . They ar e also a convenient way of alerting you to long-term measurement drift without having to observe the screen.

To Set A HI and/or LO LIMIT

1. Position the cursor in front of the unit-of-measur e for the measuremen t that you are set-
ting limits for.
2. Press and release the SHIFT key, then the down-arrow key to access the LO LIMIT
function, and enter the measurement’s low-limit value and its unit-o f -measu re.
3. Press and release the SHIFT key, then the up-arrow key to access the HI LIMIT func­tion, and enter the measurement’s high-limit value and its unit-of-measure.
1. The fundamental unit for the LIMITs does not have to be the same as the measure-
ment’s units. For instance; when measuring AC Level in Volts, you can set HI and LO LIMITs in units of dBm.
1
1
68
S:\HP892XX\USRGUIDE\manual\opoverv.fb

To Reset or Remove Limits

To reset a limit that has been exceeded
1. Position the cursor in front of the measurement’s unit-of-measure.
2. Press and release the SHIFT key, then the down-arrow (or up-arrow key) to access the
LO LIMIT (or HI LIMIT) function, then press ENTER or MEAS RESET.
To remove a limit
1. Position the cursor in front of the unit-of-measure for the assigned limit.
2. Press and release the SHIFT key, then the down-arrow (or up-arrow key) to access the
LO LIMIT (or HI LIMIT) function, then press ON/OFF.
Example of Setting HI and LO LIMITs
This example sets limits for the TX Freq Error measurement. Limits are being set to indicate if a 100 MHz carrier varies more than
Chapter 3, Operating Overview
Using Measurement Limit Indicators
± 10 kHz.
1. Position the cursor in front of the unit-of-measure for the TX FREQ ERROR measure-
ment (the default is kHz).
2. Press and release the SHIFT key, then the down-arrow to access the LO LIMIT func-
tion, then enter 1 0 kHz.
3. Press and release the SHIFT key, then the up-arrow to access the HI LIMIT function,
then enter 1 0 kHz.
69
Chapter 3, Operating Overview

Averaging Measurements

Averaging Measurements
The AVG (average) fun cti on al lows you to display the ave rag e va lue of a number of measurements. You enter the number of measurement samples used to calculate and display the measurement average. This dampens the effects of rapidly changing measurements, providing a more usable measurement display.

To Use Measurement Averaging

1. Position the cursor in front of the measurement’s unit-of-measure.
2. Press and release the SHIFT key, then the INCR ×10 key to access the AVG function.
The default number of average samples is displayed below the measurement.
Enter the desired number of measurement samples to be used for calculating the av-
erage, or
Press ON/OFF to use the currently-displayed number of samples.
3. To turn averaging off, position the cursor in front of the unit-of-measure and press and release the SHIFT key, then the INCR ×10 key to access the AVG function, then press the ON/OFF key.
When the averaging function is first enabled, a numeric average is calculated and displayed each time a measurement is made. This continues until the specified number of samples is reached. From that point on, the averaging function performs an exponential filtering operation that mimics an RC filter.
Because of the exponential response, any large measurement changes result in a displayed value that ramps up or down to the actual measured value.
Pressing MEAS RESET clears the measur ement hi story for all measure ments and starts the averaging process over.
For more information on the theory of this filtering technique, refer to the April 1986 issue of the HP Journal, page 24.
Example of Using Measurement Averaging
This example enables the SINAD measurement to be averaged using 25 samples.
1. Press PRESET and wait for the instrument to display the RX TEST screen.
2. Position the cursor in front of the unit-of-measure for the SINAD measurement (default
is dB).
3. Press and release the SHIFT key, then the INCR ×10 key to access the AVG function, enter 2 5, then press the ENTER key. Avg appears below the displayed measurement value to indicate that averaging is being used.
70
S:\HP892XX\USRGUIDE\manual\opoverv.fb

Setting A Measurement Reference

The REF SET function establishes a measurement reference point. This allows you to make a direct comparison between two measurement results, or between a measurement standard and the actual measurement results.
Referenced measurements are displayed in one of two ways, depending on the type of measurement:
Displayed value = Measurement Reference. The difference between the measured value and the reference value is displayed in the same unit-of-measure.
or Displayed value = Measurement ÷ Reference. A ratio of the measured value to the
reference value is displayed in dB.
Chapter 3, Operating Overview
Setting A Measurement Reference

To Use the Present Value as a Reference

Position the cursor in front of the unit- of-measure for the measurement y ou want to set the reference for.
1. Press and release the SHIFT key, then the INCR ÷10 key to access the REF SET func-
tion; then press enter ENTER.
2. Ref appears below the measurement.
The measurement displayed is now referenced to the measurement value present when the reference was s et.

To Set a Specific Reference

1. Position the cursor in front of the unit-of-measure for the measurement you want to set
the reference for.
2. Press and release the SHIFT key, then the INCR ÷10 key to access the REF SET func-
tion.
3. Enter a reference value.
4. Ref appears below the measurement value to indicate a reference has been set.
The measurement displayed is now referenced to the value you entered.
71
Chapter 3, Operating Overview

Using Memory Cards

Using Memory Cards
OTP (One Time Programmable) cards provide removable read-only storage. File editing and erasure are not possible. These cards cannot be programmed by the Test Set; they require a special memory card programmer to save files.
SRAM cards provide removable read/write memory for your files, similar to a flexible disk. Data can be stored, re-stored, read, or erased as needed.
SRAM memory cards require a battery to maintain stored information.
Table 5 Memory Card Part Numbers
Memory Type Part Number
64 kilobytes SRAM HP 83230A 1 Mbyte SRAM HP 83231A
72
S:\HP892XX\USRGUIDE\manual\opoverv.fb

Inserting and Removing Memory Cards

Figure 9 illustrates how to insert a memory card into the Test Set’s front panel. To
remove a memory card, simply pull it out. The Test Set’s memory-card label is marked with an arrow that must be inserted
on the left side (when you are facing the Test Set) of the front-panel slot.
Memory cards may be inserted and removed with the Test Set powered on or off.
Chapter 3, Operating Overview
Using Memory Cards
Figure 9 Inserting a Memory Card
73
Chapter 3, Operating Overview
Using Memory Cards

Setting the Write-Protect Switch

The SRAM memory card’s write-protect switch lets the user secure its contents from being overwritten or erased. The switch has two positions (see
Read-write – The memory-card contents can be changed or erased, and new files may written on the card.
Read-only – The memory-card contents can be read by the Test Set, but cannot be changed or erased.
figure 10 ):
Figure 10 Setting the SRAM Write-Protect Switch
74
S:\HP892XX\USRGUIDE\manual\opoverv.fb

The Memory Card Battery

SRAM memory cards use a lithium battery to power the card. Listed below are the
batteries for the Test Set’s SRAM cards. SRAM cards typically retain data for over 1 year at 25
SRAM Card Battery Part Numbers - CR2025 or HP 1420-0509Replacing the Battery
Turn the Test Set on and insert the memory card. An inserted memory card takes power
from the Test Set, preventing the card’s contents from being lost.
1. Hold the memory card in the slot with one hand and pull the battery holder out with your other hand. (See figure 11 .)
The Test Set’s SRAM cards have a Battery Holder Lock switch immediately above the Write-Protect switch. If the switch is in the locked position the battery cannot be removed. Ensure that the Battery Holder Lock switch is in the unlocked position before trying to remove the battery.
Chapter 3, Operating Overview
Using Memory Cards
° C. To retain data, the battery should be replaced annually.
2. Install the battery with the side marked “+” on the same side marked “+” on the battery holder. Avoid touching the flat sides of the battery, finger o ils may contaminate battery contacts in the memory-card.
3. Re-insert the battery holder into the memory card.
The Test Set’s SRAM cards have a Battery Holder Lock switch immediately above the Write-Protect switch. Ensure that the Battery Holder Lock switch is in the locked position after installing the new battery.
4. Remove the memory card from the Test Set.
75
Chapter 3, Operating Overview
Using Memory Cards
Figure 11 Replacing the Memory Card’s Battery
WARNING: Do not mutilate, puncture, or dispose of batteries in fire. The batteries can burst or explode,
releasing hazardous chemicals. Discard unused batteries according to the manufacturer’s instructions.

Memory Card Initialization

All new SRAM cards must be initialized before they can be used to store information. The RAM_MANAGER procedure stored on the internal ROM Disk can be used to quickly initialize any SRAM memory card.
SRAM Memory Cards can al so be initialize d f ro m th e TESTS screen by insert ing the memory card i nto the front -panel slot an d selec ting t he Save /Delet e fie ld, then selecting on screen instructions to complete the process.
If the error message the memory card has not been properly initialized. Check the SRAM battery to
ensure that it’s charged and inserted correctly in the battery holder.
Init Card or pressing the k3 USER k ey to i nitia lize a card. Follow t he
ERROR 85 Medium uninitialized appears on the screen
76
S:\HP892XX\USRGUIDE\manual\opoverv.fb

Saving and Recalling Instrument Setups

The SAVE and RECALL functions al lo w you t o store different instr ument se tups and retrieve them later, eliminating the task of r e-configuring the Test Set.
The number of available sa ve regi ster s depends on how many cha nges were made to the base instrument setup for e ach save. (See smaller the number of changes, the greater t he number of save regi sters that can be used (typically over 200).
Save/Recall register settings can be saved to several types of mass storage. This
allows you to “back up the settings in case you need to clear them from memory
"Memory Considerations" on page 80) for running large programs, or when a
(see firmware upgrade is performed (see
Chapter 3, Operating Overview
Saving and Recalling Instrument Setups
"BASE Settings" on page 80.) The
"Save/Recall" on page 441).

To Save an Instrument Setup

Use the More menu to access the I/O CONFIGURE screen. )
1. Select the storage media using the Save/Recall field. (The default is internal mem- ory.
2. Make any changes to the instrument that you want to save in a register.
3. Press and release the SHIFT key then the RECALL key to access the SAVE function.
4. Use the DATA keys or the Save menu at the bottom right of the screen to enter the
register’s name.

To Recall an Instrument Setup

Use the More menu to access the I/O CONFIGURE screen.
1. Select the media to recall settings from using the Save/Recall field. The default is internal memory.
2. Press RECALL.
3. Use the knob to select the desired setup to be recalled from the Recall menu at the
bottom-right of the screen.
77
Chapter 3, Operating Overview
Saving and Recalling Instrument Setups
Example of Saving and Recalling an Instrument Setup
This example SAVES changes made to t he RX TEST screen, a nd then RECALLS them. The register is s aved to wh ere ver th e memory - unless you have changed it).
1. Access the RX TEST screen and set the RF Gen Freq to 500 MHz.
2. Set Amplitude to -35 dBm.
3. Press and release the SHIFT key then the RECALL key to access the SAVE function.
A prompt appears at the top of the screen asking you to enter a name.
4. Using the DATA keys, press 1 2 3 ENTER to assign a name to these changes.
5. Press PRESET and wait for the instrument to return to normal operation.
6. If not already displayed, access the RX TEST screen. Notice that the RF Gen Freq
and Amplitude settings are reset to their preset values.
7. Press RECALL 1 2 3 ENTER. The RF Gen Freq and Amplitude are changed to
the settings you saved in register 123 (500 MHz and -35 dBm).
Save/Recall field is set (internal

To Remove (Clear) an Individual Save Register

Specify where the register is stored using the Save/Recall field on the I/O CONFIG­URE screen.
1. Press RECALL.
2. Use the knob to position the cursor in front of the register to be removed from the Re-
call menu at the bottom-right of the screen. The register name and percentage of memory occupied by that register are indicated at the top of the screen.
3. Press ON/OFF. A prompt appears, asking if you want to delete the save register.
4. Press YES.
78
S:\HP892XX\USRGUIDE\manual\opoverv.fb

To Clear All Save Registers

1. Press RECALL.
2. Use the knob to position the cursor in front of the *Clr All* entry in the Recall
menu at the bottom-right of the screen.
3. Press the knob or press ENTER. A prompt appears at the top of the screen to verify that you want to clear all registers.
4. Press YES.

Register Names

You can use any number, letter, or combination of numbers and letters as a name for storing instrument settings. For instance; if you want to save a setup for testing
a “Vulcan7” radio, you can save the setting as “VULCAN7”. Two register names are reserved for special purposes: POWERON and BASE.
Chapter 3, Operating Overview
Saving and Recalling Instrument Setups

POWERON Settings

When the Test Set is turned on, it uses a set of instrument setup parameters specified at the time of manufacture. You can have the instrument power up in a different state by making the desired changes to the original settings, and then saving them using the name POWERON.
The next time the instrument is turned on, the instrument returns to the state present when you saved the POWERON setting. For instance; if the OSCILLOSCOPE screen was displayed when POWERON was saved, it is the screen that is displayed when you turn the instrument on.
79
Chapter 3, Operating Overview
Saving and Recalling Instrument Setups

BASE Settings

The BASE register contains any field settings the user has saved that are different from the instrument preset state. It establishes a reference point for all future saves. If a base is not saved, the preset state is used as the reference.
When you save an instrument setup, the new setup is compared to the base settings, and any differences are stored under the register name you supply. Because only differences are stored, a much larger number of instrument setups can be saved than if the contents of every field was saved.
When you recall an ins trument se tting, ev ery fiel d is reset to the base sett ings. The saved settings are then used to re-establish the desired instrument setup.
CAUTION: Since each save/recall register only contains the differences between th e setup bein g saved an d
the present base register setti ngs, changing the base settings causes all other saved setu ps to be erased from memory (including the POWERON setting if one has been saved). Unless you consistently change the same fields to the same value each time you use the instrument, you should avoid creating your own BASE settings.

Memory Considerations

When the Save/Recall field of the I/O CONFIGURE screen is set to
Internal, programs are saved to the same non-volatile RAM used to create
RAM Disk(s) and run I BASIC p rog ram s . By saving a large number of instrument setups, you reduce the amount of RAM available to run programs. If you get a
“memory overflow” message while trying to load a program, you must clear one or more save/recall registers to fr ee RAM space.
Instrument Hardware Changes
Recalling a saved register that uses a hardware option that has been removed (such as an audio filter) results in unspecified operation. Re-install the needed option before attempting to recall the associated register(s).
80
S:\HP892XX\USRGUIDE\manual\opoverv.fb

Using USER Keys

Chapter 3, Operating Overview
Using USER Keys
User keys instantly access instrument settings without using the knob. You can use user keys to move quickly between fields on the same screen, and to access field settings that are not normally available on the screen you are using.
Local user keys are used to move bet ween sett in gs on the scre en that is displa yed . When the user key is pressed, the cursor instantly moves to, and selects, the assigned field; eliminating the need to turn and push the knob. Five local user keys are available for each screen: k1, k2, k3, k4, and k5.
Five factory-assi gned local user k eys ar e available in e ach screen; however, us ing these keys removes any other local user keys you may have already set up.
Global user keys are used to access settings that are not available on the current screen. Three global user keys are available: k1’, k2’, and k3’. (These are shifted functions of the local user keys.)
When defining user keys, the ASSIGN function is used to create key definitions; the RELEASE function removes the definitions. Re-assigning a user key to a different field setting automatically releases it from the setting it was previously associated with.
81
Chapter 3, Operating Overview
Using USER Keys

To Use the Pre-Assigned Local USER Keys

1. Press and release the SHIFT key, then the k4 key to access the ASSIGN function; then
press the ENTER key. The numbers 1 thr oug h 5 app ear in front of various fields. (See
figure 12 .)
2. Press the different local user keys (k1 to k5) and notice how the cursor immediately
moves to the corresponding field.
3. To stop using the default local user keys, press and release the SHIFT key, then the k5
key to access the RELEASE function; then press the ENTER key.
Figure 12 An Example of Pre-Assigned Local User Keys
82
S:\HP892XX\USRGUIDE\manual\opoverv.fb
scnusr.wmf

To Assign Local USER Keys

1. Move the cursor to the field you want to assign a local user key to.
2. Press and release the SHIFT key, then the k4 key to access the ASSIGN function. Then
press a local USER key (k1-k5). The user key number appears in front of the field you assigned it to.
Example of Assigning a Local USER Key
Use this example to assign local USER key k1 to the Filter 1 field in the RX TEST screen.
1. Access the RX TEST screen and position the cursor in front of the Filter 1 field.
2. Press and release the SHIFT key, then the k4 key to access the ASSIGN function; then
press k1. A small 1 appears next to the field indicating that USER key k1 has been as­signed to it.
3. Move the cursor to any other field on the s creen and press k1. The cursor immediately returns to the Filter 1 field. The field is also highlighted to change the entry using the CURSOR CONTROL knob or arrow keys.
Chapter 3, Operating Overview
Using USER Keys

To Release Local USER Keys

1. Display the screen containing the user key assignment to be removed.
2. Press and release the SHIFT key, then the k5 key to access the RELEASE function; then
press the USER key (k1-k5).
83
Chapter 3, Operating Overview
Using USER Keys

To Assign Global USER Keys

1. Move the cursor to the field you want to assign a global user key to.
2. Press and release the SHIFT key, then the k4 key to access the ASSIGN function. Then
press SHIFT and a global USER key (k1’ - k3’). Unlike a local user key, the user key number does not appear at this field; instead, a prompt appears at the top of the screen confirming the key assignment.
Example of Assigning a Global USER Key
Use this example to assi gn glob al USER k ey k1’ t o the AF Anl In field, and then access this field from the OSCILLOSCOPE screen.
1. Access the AF ANALYZER screen and position the cursor in front of the AF Anl In
field.
2. Press and release the SHIFT key, then the k4 key to access the ASSIGN function.
3. Press SHIFT, k1’. Notice the prompt Global User key 1 assigned. at the top
of the screen.
4. Access the OSCILLOSCOPE screen.
5. Press SHIFT, k1’.
AF Anl Input
present input is set to FM Demod). To change the input, use the arrow keys (down-arrow or up-arrow), or press ENTER to access the
A field that is accessed using a global user key is only displayed at the top of the screen while it is being accessed. Moving the cursor to any other field in the screen causes the user key field to disappear until it is accessed again.

To Release Global USER Keys

1. Move the cursor to the field with the global user key assigned to it.
2. Press and release the SHIFT key, then the k5 key to access the RELEASE function.
Then press SHIFT and the USER key to be released (k1’-k3’).
, FM Demod is displayed at the top of the screen (assuming the
Choices menu.
84
S:\HP892XX\USRGUIDE\manual\opoverv.fb

Setting an RF Generator/Analyzer Offset

You can set a fixed frequency offset between the RF Generator and the RF Analyzer. This feature is convenient for testing radios with a fixed transmit/receive frequency offset.

To Set an RF Offset

1. Access the CONFIGURE screen.
2. Position the cursor in front of the RF Offset field, and press the Cursor Control knob,
or press ENTER to turn the offset On or Off.
3. Select the (Gen)-(Anl) field and enter the frequency offset value.
Chapter 3, Operating Overview
Setting an RF Generator/Analyzer Offset
Example of Setting an RF Offset
1. Access the CONFIGURE screen.
2. Set the RF Offset to On.
3. Enter an offset frequency ((Gen)-(Anl)) of 10 MHz.
4. Access the DUPLEX screen.
5. Set the Tune Mode to Manual.
6. Select the RF Gen Freq field, and rotate the Cursor Control knob to vary the RF Gen-
erator’s frequency.
7. Notice that the Tune Freq value changes to maintain the 10 MHz difference between the generator and the analyzer.
1
1. Manual tuning is used in this example to prevent possible unexpected Tune
Frequency changes during the procedure.
85
Chapter 3, Operating Overview

Using Remote Control

Using Remote Control
The Test Set can be remotely controlled several ways:
Using HP-IB control from a computer/controller.
Using IBASIC programs on memory cards.
Using an ASCII terminal connected to the serial port.

Using HP-IB Control

The Programmer’s Guide contains information on writing HP-IB control programs for the Test Set. Programming examples and a syntax listing provide general HP-IB operation guidelines.

Running IBASIC Programs from Memory Cards

The documentation shipped with HP 11807 software packages explains how to run those programs from memory cards. Refer to the Programmer’s Guide for detailed information on using memory cards with your own IBASIC programs.
86
S:\HP892XX\USRGUIDE\manual\opoverv.fb

Using an ASCII Terminal

Connecting an ASCII terminal to the serial port allows you to remotely operate the Test Set by entering characters that represent each front-panel control.
Before you can use this feature, you must first set the required serial port settings in the I/O CONFIGURE screen, and make any hardware connections.
Chapter 3, Operating Overview
Using Remote Control
The Serial Port connections are described in
Knob Descriptions"
To Configure for Serial Port Operation
1. Access the I/O CONFIGURE screen.
2. Set the Serial In field to Inst.
3. Set the IBASIC Echo field to On.
4. Set the Inst Echo field to On.
5. Set the remaining serial communications fields according to your terminal/computer’s
serial communication requirements. These fields include:
Serial Baud
Parity
Data Length
•Stop Length
Rcv Pace
Xmt Pace
6. The Test Set now responds to the equivalent characters sent to it by the terminal/com ­puter.
Equivalent Front-Panel Control Characters
Table 6 on page 88
.
lists the terminal/computer keystrokes that equate to front-
chapter 24, "Connector, Key, and
panel controls. Each sequence must be preceded by the Escape key. For example, to remotely access th e CONFIGURE scr een, pre ss the Esc key , then
press the C key on your te rminal/com puter. Be sure to use uppe r -case C for this example.
87
Chapter 3, Operating Overview
Using Remote Control
Table 6 Equivalent Front-Panel Control Characters
Equivalent
Function
ESC
Function
Character
CANCEL ! A X PERCENT MHZ_V ( EEX Z S_KHZ_MV ) YES_ON_OFF [ BACKSPACE - NO_PPM_W ] ENTER . RX a RELEASE 0 TX b K1 1 DUPLEX c K2 2 PREV d K3 3 TESTS_MAIN e K4 4 LOCAL f K5 5 RECALL g K1_PRIME 6 MEAS_RESET h K2_PRIME 7 PRESET i K3_PRIME 8 INCR_DIV_10 j ASSIGN 9 INCR_SET k KNOB_TURN_CCW < INCR_TIMES_10 l KNOB_TURN_CW > DOWN m MSSG A UP n HELP B SEVEN o CONFIG C EIGHT p HOLD D NINE q PRINT E FOUR r ADRS F FIVE s SAVE G SIX t REF_SET J ONE u METER K TWO v AVG L THREE w LO_LIMIT M ZERO x HI_LIMIT N POINT y
Equivalent
ESC
Character
88
S:\HP892XX\USRGUIDE\manual\opoverv.fb
Chapter 3, Operating Overview
Table 6 Equivalent Front-Panel Co ntrol Characters (Continued)
Using Remote Control
Equivalent
Function
E R PLUS_MINUS z F S OHM_PCT_DEL_DBUV B U DB_GHZ_DBM | CVMS_HZ_UV DW
ESC
Character
Function
Equivalent
Character
Alternate sequences for 5 commonly-used functions are also available. Hold down the Cntl (control) key and select the corresponding key for the desired function. Example: Cntl H moves the cursor to the left one space.
ENTER - J or M CANCEL - C BACKSPACE - H KNOB_TURN_CW - R KNOB_TURN_CCW - L
ESC
89
Chapter 3, Operating Overview
Using Remote Control
90
S:\HP892XX\USRGUIDE\manual\opoverv.fb
4

Adjacent Channel Power Screen

NOTE: This screen is displayed on the Test Set with Option 102, Spectrum Analyzer.
This screen is used to measure Adj acent Cha nnel Power . This is a measur ement of the power of signals at a specific channel spacing above and below the RF
Analyzer’s center frequency. This screen is accessed by selecting from the
To Screen menu.
AD CH PWR
91
Chapter 4, Adjacent Channel Power Screen
How the Test Set Measures Adjacent Channel Power (ACP)
How the Test Set Measures Adjacent Channel Power (ACP)
When you access t his scr een, the Te st Set au tomatical ly st arts a multi-st ep proces s for measurin g ACP:
1. AF Generator 1 is turned off if the Carrier Ref field is set to Unmod.
2. The amplitude of the center frequency (Tune Freq) is measured to establish a
reference.
3. AF Generator 1 is turned back on if it was previously turned off.
4. The power in each of the adjacent channels is analyzed.
5. Adjacent Channel Power is calculated and displayed. This value can be displayed as an
absolute power level or as a ratio referenced to the center frequency’s level.
Which Input Port to Use. The TX Power measurement is used to calculate absolute Adjacent Channel Power. Since TX Power can only be measured using the RF IN/OUT port, you must use this port to measure ACP Level. ACP Ratio can be measured using either the RF IN/OUT or the ANT IN port. Measuring ACP on AM Transmitters. When measur ing AM signals, the reference level must be measured on an unmodulated carrier; so the Carrier Ref field must be set to Unmod. After the reference is measured, the power in the adjacent channels must be mea­sured with modulatio n . This req uires th e m odulating signal to be turned off and on repeat­edly as measurements are being calculated and displayed. Since the Test Set automatically turns AFGen1 on and off w hen the Carrier Ref field is set to Unmod, you must use AFGen1 and the AUDIO OUT port as the modulation source for making AM ACP mea­surements.
92
S:\HP892XX\USRGUIDE\manual\acpscrn.fb

Field Desc ri p tions

Chapter 4, Adjacent Channel Power Screen
Field Descriptions
Figure 13 The Adjacent Channel Power Screen

ACP Meas

This field selects the format for displaying upper and lower adjacent channel power levels.
Ratio displays the power levels relative to the power around the center frequency (Tune Freq). Levels can be displayed in dB or as a percentage (%).
Level displays the absolute power levels in mW, W, dBm, V, mV, and dBmV.
Operating Considerations
TX Power
Four dashes are displaye d for these mea surements when the
Ant (ANT IN). ACP Ratio can be measured on either the ANT IN port or RF IN/OUT port.
For more information, refer to
Power (ACP)" on page 92
and ACP Level can only be measured through the RF IN/OUT port.
.
adchpwr.wmf
Input Port is set to
"How the Test Set Measures Adjacent Channel
93
Chapter 4, Adjacent Channel Power Screen
Field Descriptions

AFGen1 Freq

This field sets the frequency for the first audio frequency sinewave generator.

AFGen1 To

This field sets two values:
The upper field determines whether the AFGen1 signal modulates the RF Generator or is output through the AUDIO OUT connector.
The lower field sets the depth of modulation for FM and AM in kilohertz and amplitude (including Off) for audio out. The AUDIO OUT level is always in volts rms.
Operating Considerations
This is a priority control field. Accessing the RX TEST or TX TEST screen overrides any changes made to this field in other screens.

Carrier Ref

Channel BW

See Also
"Interaction Between Screens" on page 54
Use the carrier reference field to indicate whether the carrier (Tune Freq) being measured should be unmodulated or modulated when making the ACP reference measurement. (For more information, refer to
Adjacent Channel Power (ACP)" on page 92
Operating Considerations
FM transmitters can be measured with the carrier modulated or unmodulated.
For AM transmitters, the carrier must be measured while unmodulated. AFGen1 and the AUDIO OUT port must be used as the modulation source whenever Unmod is selected.
"How the Test Set Measures
.)
Use the channel bandwidth field to specify the bandwidth of the carrier and adjacent channels to be measured. (See
figure 14 .)
94
S:\HP892XX\USRGUIDE\manual\acpscrn.fb
Chapter 4, Adjacent Channel Power Screen
Field Descriptions
adchpwr1.wmf
Figure 14 Relationship Between Tune Freq, Ch Offset, and Channel BW Fields
95
Chapter 4, Adjacent Channel Power Screen
Field Descriptions

Ch Offset

Use the channel offset field to enter the frequency difference between the Tune
setting and the center of the adjacent channels you want to measur e. Thi s is
Freq
an absolute value; only positive values can be entered. See

Ext TX key

This field controls a switch at the MIC/ACC connector. Use it to “key” an external transmitter.
See Also
"MIC/ACC" on page 572

Input Atten

figure 14 .
Input Attenuation sets the amount of input attenuation for the RF IN/OUT and ANT IN connectors. This function controls two settings:
The upper field determines if you want the instrument to set the attenuation automati­cally (Auto), or if you want to set the value manually (Hold).
The lower field displays the present attenuation value, and is used to set the desired at­tenuation level when the upper area is set to Hold.
Operating Considerations
Input Attenuator autor angi ng can interfere with oscill oscope or signaling decoder operation under certain conditions.
See "Input Atten," in chapter 17, on page 477
for additional information.
96
S:\HP892XX\USRGUIDE\manual\acpscrn.fb

Input Port

Chapter 4, Adjacent Channel Power Screen
Field Descriptions
This field selects the RF IN/OUT or ANT IN port for making RF measurements. The RF IN/OUT port must be used for making TX Power or ACP Level measurements on this screen.
Operating Considerations
Power levels for each port are printed on the Test Set’s front panel. If the RF power at the RF IN/OUT port exceeds allowable limits, a loud warning signal sounds and a message appears at the top of the screen. If this occurs, disconnect the RF power, press the MEAS RESET key, and allow the Test Set to cool off for approximately two minutes before making any other measurements on this port.
The ANT IN (antenna input) connector provides a highly-sensitive input for very low level signals (such as “off the air” measurements). You cannot measure TX (RF) Power or ACP Level on this screen using the ANT IN port.
CAUTION: Connecting a signal of >200 mW to the ANT IN port can cause instrument damage
(although internal protection circuits can typically withstand a short-duration signal of 1 or 2 Watts).
If the overpower circuit is triggered (signified by a warning message at the top of the screen), remove the signal from the ANT IN port, and press the MEAS RESET key or turn the Test Set off and on to reset it.

Lower and Upper ACP [Ratio:Level]

These two measurement s display th e amount of power in signals abo ve and below
Tune Freq signal. The level is displayed as a ratio (referenced to the power
the around the
See Also
ACP Meas
Tune Freq) or as an absolute value.
field description, on page 93
97
Chapter 4, Adjacent Channel Power Screen
Field Descriptions

Res BW

The resolution bandwidth field selects the IF filter used when measuring the power of the carrier and the adjacent channels.
Operating Considerations
Using a narrower bandwidth filter (300 Hz) slows the measurement, but rejects carrier leakage and out of channel spurs. Using a wider bandwidth filter (1 kHz) speeds measurements, but may allow unwanted spurs and carrier leakage to be integrated into th e measureme nt when measuri ng at the edges of the select ed channel bandwidth.

Tune Freq

This field sets the center frequency for the RF signal to be analyzed.

Tune Mode

See Also
Tune Mode
field description (this page)
This field selects Automatic or Manual tuning of the RF Analyzer.
Auto tuning causes the RF Analyz er to f ind th e sign al with th e great est a mplitu de
>-36 dBm, and to set the Tune Frequency for that signal.
Manual tuning requires the operator to set the Tune Frequency for the RF signal
to be analyzed.
Operating Considerations
Changing the Tune Mode also changes the RF frequency display. Automatic tuning enables the
TX Freq Error measurement.
TX Frequency measurement. Manual tuning enables the
98
S:\HP892XX\USRGUIDE\manual\acpscrn.fb

TX Freq Error/TX Frequency

This measurement displays Transmitter Frequency Error or absolute Transmitter Frequency.
See Also
Chapter 4, Adjacent Channel Power Screen
Field Descriptions

TX Power

TX Pwr Meas

Tune Mode
field description (this page)
Transmitter Power measures RF power at the RF IN/OUT port.
Operating Considerations
Only the RF IN/OUT port can be used for measuring TX Power on this screen. When the
Input Port is set to Ant, four dashe s (- - - -) appear in place of di git s
for this measurement.
1
Use the Spectrum Analyzer
to measure low-level RF power (200 mW) at the
ANT IN port.
The transmitter power measurement field specifies how transmitter power measurements are made:
Peak can be used to measure AM, FM, and un-modulated (CW) signals.
Sample can be used to measure FM or CW signals with increased measurement speed. This method can only be used with signals having no amplitude modulation component.

TX Pwr Zero

Operating Considerations
If you change this field, zero the power measurement (using the TX Pwr Zero field) before measuring power.
The transmitter power zero function establishes a 0.0000 W reference for measuring RF power at the RF IN/OUT port.
1. O ptional on some Test Set models.
99
Chapter 4, Adjacent Channel Power Screen
Field Descriptions
CAUTION: RF power must not be applied while zeroing. Set the RF Generator screen’s Amplitude
field to off to prevent internal cross-coupling into the power detector while zeroing.
Operating Considerations
When power is applied to the RF IN/OUT connector, the temperature of the internal circuit ry incr eases. This can cause chan ges in t he TX Power mea surement when low power levels are measured immediately following high power measurements.
When alternately making high and low power measurements, always zero the power meter immediately before making the low power measurements; this provides the best measurement accuracy.
100
S:\HP892XX\USRGUIDE\manual\acpscrn.fb
Loading...