11884C 6 GHz Operation Upgrade Kit
for the 8753D Option 011 Network Analyzer
Part Number 08753-90266 Supersedes August 1998
Printed in USA August 2001
Notice.
The information contained in this document is subject to change without notice.
Agilent T echnologies makes no warranty of any kind with re gard to this material, including but not limited
to, the implied warranties of merchantability and fitness for a particular purpose. Agilent Technologies
shall not be liable for errors contained herein or for incidental or consequential damages in connection with
the furnishing, performance, or use of this material.
Product Affected8753D Option 011 Network Analyzer
To Be Performed by:[X] Agilent Technologies Qualified
Personnel
Purpose
The 11884C 6 GHz upgrade kit provides the 8753D Option 011 Network Analyzer with optional 6 GHz
operation. The upgrade kit is keyed to the serial number of an individual network analyzer.
Items Included in the Kit
Table 1 describes the parts included in the 11884C upgrade kit. Check the contents of the kit against
this table.
Power meter436A/437B/438A
Power sensor8482A
Power sensor8481A
Power Splitter, 2 way 50Ω (2)8481A
Frequency counter5350B
Cable Set, RF 24-inch11500B
Cable Set, RF 50Ω Type-N11851B
Cable, BNC-to-BNC8120-1840
Adapter, Type-N (f) to BNC (m)1250-0077
Adapter, Type-N (m) to Type-N (m)1250-1475
Adapter, Type-N (f) to Type-N (f)1250-1472
Adapter, Type-N (m) to SMA (f)1250-1250
Adapter, Type-N (f) to SMA (m)1250-1562
Adapter, BNC Alligator clip8120-1292
GPIB cable assembly10833A
Low-pass filter9135-0198
Attenuator, 10 dB (Option 002 only)8491A Option 010
Attenuator, 20 dB (2)8491A Option 020
Attenuator, 30 dB (Option 002 only)8491A Option 030
3.5-inch floppy disk92192A (box of 10)
Small needle-nose pliers
T-10 TORX screwdriver
T-15 TORX screwdriver
5/16 inch open-end wrench
5/16 inch torque wrench (10 in-lbs)
Electrostatic discharge (ESD) grounding wrist strap and mat
4 Installation Note 087533-90266
Safety Considerations
WARNINGBefore you disassemble the instrument, turn the power switch OFF and
unplug the instrument. Failure to unplug the instrument can result in
personal injury.
CAUTIONElectrostatic discharge (ESD) can damage or destroy electronic components.
Perform these procedures only at a static-safe workstation and wear a grounding
strap. Refer to the documentation that pertains to your instrument for
information about static-safe workstations and ordering static-safe accessories.
Installation Note 08753-90266 5
Installation Procedure for the 8753D Option 011
Verify the Serial Number
Refer to the keyword label in the box below. Verify that the analyzer’s serial number matc hes the serial
number below. This keyword will not work on any other analyzer. If the serial numbers do not match,
contact your nearest Agilent Technologies sales or service office for assistance.
Keyword Label
Remove the Feet and Covers
Refer to Figure 1.
1. Disconnect the power cord.
2. Remove the top cover:
a. Remove both of the upper rear feet (item 1) by loosening the attaching screws (item 2).
b. Loosen the top cover screw (item 3).
c. Slide the cover towards the back of the instrument.
6 Installation Note 087533-90266
Figure 1Cover and Feet Removal
Installation Note 08753-90266 7
Remove the 3 GHz Source Assembly
Refer to Figure 2
3. Remove the source bracket (item 4).
4. Disconnect the two semi-rigid cables (W1 and W24) and one flexible cable (W26) from the source
assembly (A3).
5. Lift the two retention clips (item 5) at the front and rear of the source module to an upright position.
6. The source is seated in a motherboard edge connector. Hold the two loose semi-rigid cables
(W1 and W24) to the right and gently pull up on the source bracket handle (item 6) to lift the source
assembly out of the instrument.
Replace the Source
Refer to Figure 2.
7. Slide the edges of the sheet metal partition (item 7) of the 6 GHz source assembly into the guides at
the front and back of the source compartment.
8. Press down on the assembly to ensure that it is well seated in the motherboard connector.
9. Push the retention clips down.
10.Reconnect the two semi-rigid cables (W1 and W24) and one flexible cable (W26) to the source
assembly. Torque these connections to 10 in-lbs.
11.Replace the source bracket.
8 Installation Note 087533-90266
Figure 2A3 Source Assembly Removal
Installation Note 08753-90266 9
Configure the 8753D Option 011 for 6 GHz Operation
Refer to Figure 3.
12.Remove the instrument bottom cover.
13.Move the A9 CC Jumper to the ALT position. Refer to Figure 3.
Figure 3 A9 CC Jumper Location
14.Switch on the power for the 8753D Option 011.
15.Press , , , , , to erase the logo.
16.Enter the keyword from the label located on page 6. You may use an external keyboard or use the
analyzer’s front panel knob to position the arrow below each character of the keyword. Press
SELECT LETTERDONE
17.Press , , , , . At the prompt, press
EXECUTE TEST YES
The analyzer will displayOption Cor DONE when the test is complete.
18.Press , , , . At the prompt, press , .
The analyzer will display*Source Def DONE when the test is complete.
19.Press , , , . At the prompt, press , .
The analyzer will display*Pretune Def DONE when the test is complete.
20.Press , , , . At the prompt, press , .
The analyzer will displayABUS Cor DONE when the test is complete.
21.Press , , , . At the prompt, press , .
The analyzer will displayPretune Cor DONE when the test is complete.
PRESETDISPLAY
to enter each character. When all characters are entered, press .
SYSTEM
RETURN TESTS
RETURN TESTS
RETURN TESTS
RETURN TESTS
SERVICE MENU TESTS
,.
MORE TITLE ERASE TITLE
56x1
44x1
45x1
46x1
48x1
EXECUTE TEST YES
EXECUTE TEST YES
EXECUTE TEST YES
EXECUTE TEST YES
10 Installation Note 087533-90266
22.Refer to "Adjustments" in the8753D Network Analyzer Option 011 Service Guide and perform the
following adjustments:
•RF Output Power Correction Constants
•Sampler Magnitude and Phase Correction Constants
•Cavity Oscillator Frequency Correction Constants
•Source Spur Avoidance Tracking Adjustment
•EEPROM Backup Disk Procedure
23.Refer to "System Verification and Performance Tests" in the8753D Network AnalyzerOption 011 Service Guide and perform the following system tests:
•Source Frequency Range and Accuracy
•Source Power Range, Linearity, and Accuracy
•Source and Receiver Harmonics (necessary only for analyzers with both Option 002 and 006)
24.Switch off the analyzer and disconnect the power cord. Move the A9 CC Jumper back to the NRM
position.
Refer to Figure 3.
Install the Keyword Label
Refer to Figure 4.
25.Place the keyword label inside the instrument at the location shown.
Figure 4 Option Keyword Label Location for the 8753D Option 011
Installation Note 08753-90266 11
Reassemble the Instrument
26.Replace the instrument top and bottom covers and all of the rubber feet.
27.Place the 11884C installation label on the rear panel of the network analyzer. Write the date of the
installation in the space provided.
Replace the Analyzer's Nameplate
28.Use a sharp knife to slowly peel away the old nameplate.
29.Adhere the new nameplate, which is included in the upgrade kit, onto the analyzer's front panel.
This completes the 11884C 6GHz operation upgrade kit installation for the 8753D Option 011.
Contacting Agilent
By internet, phone or fax, get assistance with all your test and measurement needs.
Online assistance: www.agilent.com/find/assist
12 Installation Note 087533-90266