Agilent 8753D Operator Manual

Installation Note
11884C 6 GHz Operation Upgrade Kit for the 8753D Option 011 Network Analyzer
Part Number 08753-90266 Supersedes August 1998 Printed in USA August 2001
Notice.
The information contained in this document is subject to change without notice.
Agilent T echnologies makes no warranty of any kind with re gard to this material, including but not limited to, the implied warranties of merchantability and fitness for a particular purpose. Agilent Technologies shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.
© Copyright 1995, 1998, 2001 Agilent Technologies, Inc

11884C 6 GHz Operation Upgrade Kit

Product Affected 8753D Option 011 Network Analyzer To Be Performed by: [X] Agilent Technologies Qualified
Personnel

Purpose

The 11884C 6 GHz upgrade kit provides the 8753D Option 011 Network Analyzer with optional 6 GHz operation. The upgrade kit is keyed to the serial number of an individual network analyzer.

Items Included in the Kit

Table 1 describes the parts included in the 11884C upgrade kit. Check the contents of the kit against this table.
Table 1 11884C Upgrade Kit Contents
Quantity Description
1 6 GHz source assembly 1 Customer letter 1 6 GHz registration number label 1 Keyword label 1 Installation note 1 Option installation label 1 6 GHz Nameplate
Installation Note 08753-90266 3

Equipment and Tools Required

Table 2 Required Epuipment and Tools
Item Part or Model Number
Power meter 436A/437B/438A Power sensor 8482A Power sensor 8481A Power Splitter, 2 way 50Ω (2) 8481A Frequency counter 5350B Cable Set, RF 24-inch 11500B Cable Set, RF 50Type-N 11851B Cable, BNC-to-BNC 8120-1840 Adapter, Type-N (f) to BNC (m) 1250-0077 Adapter, Type-N (m) to Type-N (m) 1250-1475 Adapter, Type-N (f) to Type-N (f) 1250-1472 Adapter, Type-N (m) to SMA (f) 1250-1250 Adapter, Type-N (f) to SMA (m) 1250-1562 Adapter, BNC Alligator clip 8120-1292 GPIB cable assembly 10833A Low-pass filter 9135-0198 Attenuator, 10 dB (Option 002 only) 8491A Option 010 Attenuator, 20 dB (2) 8491A Option 020 Attenuator, 30 dB (Option 002 only) 8491A Option 030
3.5-inch floppy disk 92192A (box of 10) Small needle-nose pliers T-10 TORX screwdriver T-15 TORX screwdriver 5/16 inch open-end wrench 5/16 inch torque wrench (10 in-lbs) Electrostatic discharge (ESD) grounding wrist strap and mat
4 Installation Note 087533-90266

Safety Considerations

WARNING Before you disassemble the instrument, turn the power switch OFF and
unplug the instrument. Failure to unplug the instrument can result in personal injury.
CAUTION Electrostatic discharge (ESD) can damage or destroy electronic components.
Perform these procedures only at a static-safe workstation and wear a grounding strap. Refer to the documentation that pertains to your instrument for information about static-safe workstations and ordering static-safe accessories.
Installation Note 08753-90266 5

Installation Procedure for the 8753D Option 011

Verify the Serial Number
Refer to the keyword label in the box below. Verify that the analyzer’s serial number matc hes the serial number below. This keyword will not work on any other analyzer. If the serial numbers do not match, contact your nearest Agilent Technologies sales or service office for assistance.
Keyword Label
Remove the Feet and Covers
Refer to Figure 1.
1. Disconnect the power cord.
2. Remove the top cover: a. Remove both of the upper rear feet (item 1) by loosening the attaching screws (item 2).
b. Loosen the top cover screw (item 3). c. Slide the cover towards the back of the instrument.
6 Installation Note 087533-90266
Figure 1 Cover and Feet Removal
Installation Note 08753-90266 7

Remove the 3 GHz Source Assembly

Refer to Figure 2
3. Remove the source bracket (item 4).
4. Disconnect the two semi-rigid cables (W1 and W24) and one flexible cable (W26) from the source assembly (A3).
5. Lift the two retention clips (item 5) at the front and rear of the source module to an upright position.
6. The source is seated in a motherboard edge connector. Hold the two loose semi-rigid cables (W1 and W24) to the right and gently pull up on the source bracket handle (item 6) to lift the source assembly out of the instrument.

Replace the Source

Refer to Figure 2.
7. Slide the edges of the sheet metal partition (item 7) of the 6 GHz source assembly into the guides at the front and back of the source compartment.
8. Press down on the assembly to ensure that it is well seated in the motherboard connector.
9. Push the retention clips down.
10.Reconnect the two semi-rigid cables (W1 and W24) and one flexible cable (W26) to the source assembly. Torque these connections to 10 in-lbs.
11.Replace the source bracket.
8 Installation Note 087533-90266
Figure 2 A3 Source Assembly Removal
Installation Note 08753-90266 9
Configure the 8753D Option 011 for 6 GHz Operation
Refer to Figure 3.
12.Remove the instrument bottom cover.
13.Move the A9 CC Jumper to the ALT position. Refer to Figure 3.
Figure 3 A9 CC Jumper Location
14.Switch on the power for the 8753D Option 011.
15.Press , , , , , to erase the logo.
16.Enter the keyword from the label located on page 6. You may use an external keyboard or use the analyzer’s front panel knob to position the arrow below each character of the keyword. Press
SELECT LETTER DONE
17.Press , , , , . At the prompt, press
EXECUTE TEST YES
The analyzer will displayOption Cor DONE when the test is complete.
18.Press , , , . At the prompt, press , . The analyzer will display*Source Def DONE when the test is complete.
19.Press , , , . At the prompt, press , . The analyzer will display*Pretune Def DONE when the test is complete.
20.Press , , , . At the prompt, press , . The analyzer will displayABUS Cor DONE when the test is complete.
21.Press , , , . At the prompt, press , . The analyzer will displayPretune Cor DONE when the test is complete.
PRESET DISPLAY
to enter each character. When all characters are entered, press .
SYSTEM
RETURN TESTS
RETURN TESTS
RETURN TESTS
RETURN TESTS
SERVICE MENU TESTS
,.
MORE TITLE ERASE TITLE
56 x1
44 x1
45 x1
46 x1
48 x1
EXECUTE TEST YES
EXECUTE TEST YES
EXECUTE TEST YES
EXECUTE TEST YES
10 Installation Note 087533-90266
22.Refer to "Adjustments" in the8753D Network Analyzer Option 011 Service Guide and perform the following adjustments:
•RF Output Power Correction Constants
•Sampler Magnitude and Phase Correction Constants
•Cavity Oscillator Frequency Correction Constants
•Source Spur Avoidance Tracking Adjustment
•EEPROM Backup Disk Procedure
23.Refer to "System Verification and Performance Tests" in the8753D Network Analyzer Option 011 Service Guide and perform the following system tests:
•Source Frequency Range and Accuracy
•Source Power Range, Linearity, and Accuracy
•Source and Receiver Harmonics (necessary only for analyzers with both Option 002 and 006)
24.Switch off the analyzer and disconnect the power cord. Move the A9 CC Jumper back to the NRM position.
Refer to Figure 3.
Install the Keyword Label
Refer to Figure 4.
25.Place the keyword label inside the instrument at the location shown.
Figure 4 Option Keyword Label Location for the 8753D Option 011
Installation Note 08753-90266 11

Reassemble the Instrument

26.Replace the instrument top and bottom covers and all of the rubber feet.
27.Place the 11884C installation label on the rear panel of the network analyzer. Write the date of the installation in the space provided.

Replace the Analyzer's Nameplate

28.Use a sharp knife to slowly peel away the old nameplate.
29.Adhere the new nameplate, which is included in the upgrade kit, onto the analyzer's front panel.
This completes the 11884C 6GHz operation upgrade kit installation for the 8753D Option 011.