Agilent 86207A Data Sheet

Agilent 86205A & 86207A 50 & 75 Ω RF Bridges
Technical Data
300 kHz to 6 GHz (50 ) 300 kHz to 3 GHz (75 )
40 dB directivity
Excellent directivity allows you to measure high return loss devices and good port match lets you measure low return loss devices. This is especially important when making reflection measurements with scalar network analyzers, vector network analyzers and spectrum analyzers.
Wide frequency range
The bridges have an exceptionally wide RF frequency range: 300 kHz to 6 GHz for the 50 Ω 86205A and 300 kHz to 3 GHz for the 75 86207A. They are ideal acces­sories when used with the 8753 family of RF vector net­work analyzers.
Low insertion loss
These bridges offer low insertion loss, (nominally 1.5 dB), which is significantly less than the typical 6 to 8 dB nor­mally associated with RF bridges. Low insertion loss means more power to the device under test. This is required for the measurement of high-power solid state amplifiers and traveling wave tube amplifiers.
Flat coupling factor
The frequency response of the coupled arm for these bridges is very flat, within ±0.2 dB of the nominal value of 16 dB. This is important in applications such as external power leveling where a power meter or diode detector is used to remotely level the output power from an RF source and when measuring power level dependent devices such as the LO input to a mixer or the RF input to an amplifier during compression testing.
The 86205A/86207A high directivity RF bridges offer unparal­leled performance in a variety of general purpose applica­tions. They are ideal for accurate reflection measurements and signal-leveling applications. They combine the directivity and broad-band frequency range of directional bridges and the low insertion loss and flat coupling factor of directional couplers.
Specifications
Specifications describe the instrument’s warranted per­formance over the temperature range of 0 °C to 55 °C. Supplemental characteristics are intended to provide information useful in applying the instrument by giving supplemental, but not warranted performance parame­ters. These are denoted as “typical.”
86205A
Frequency range 300 kHz to 6 GHz Impedance 50 Ohms (nominal) Directivity (25 °C, ±5 °C) 30 dB, 0.3 MHz to 5 MHz
40 dB, 5 MHz to 2 GHz 30 dB, 2 GHz to 3 GHz (typical) 20 dB, 3 GHz to 5 GHz (typical) 16 dB, 5 GHz to 6 GHz
Port match 23 dB, 0.3 MHz to 2.0 GHz (1.15 SWR)
20 dB, 2.0 GHz to 3 GHz (1.22 SWR) (typical) 18 dB, 3 GHz to 5 GHz (1.29 SWR) (typical) 16 dB, 5 GHz to 6 GHz (1.38 SWR)
Supplemental characteristics
Insertion loss 1.5 dB, +0.1 dB/GHz Insertion loss deviation ±0.2 dB Coupling factor (< 3 GHz) 16.0 dB, +0.15 dB/GHz
(> 3 GHz) 16.5 dB, –0.20 dB/GHz Coupling factor deviation (< 3 GHz) ±0.2 dB, (> 3 GHz) ±0.4 dB Maximum input power 25 dBm Maximum DC, volts (through arm) 30 VDC Maximum DC, volts (coupled arm) 0 VDC Maximum DC, amps (through arm) 1 amp Below 1 MHz directivity and port match will be slightly degraded above 200 ma bias current. Connectors Type-N (female) Dimensions 160 W x 93 H x 23 mm D
(6.3 W x 3.7 H x 1 in D) Weight Net 0.57 kg (1.3 lbs),
shipping 1.80 kg (4.0 lbs)
Models
86205A RF bridge (50 Ω) 86207A RF bridge (75 Ω)
806207A
Frequency range 300 kHz to 3 GHz Impedance 75 Ohms (nominal) Directivity (25 °C, ±5 °C) 30 dB, 0.3 MHz to 5 MHz
40 dB, 5 MHz to 1.3 GHz 35 dB, 1.3 GHz to 2 GHz (typical) 30 dB, 2 GHz to 3 GHz
Port match 20 dB, 0.3 GHz to 1.3 GHz (1.22 SWR)
(typical) 18 dB, 1.3 GHz to 2 GHz (1.29 SWR) (typical) 18 dB, 2 GHz to 3 GHz (1.29 SWR)
www.agilent.com/find/emailupdates
Get the latest information on the products and applications you select.
For more information on Agilent Technologies’ products, applications or services, please contact your local Agilent office. The complete list is available at:
www.agilent.com/find/contactus
Product specifications and descriptions in this document subject to change without notice.
© Agilent Technologies, Inc. 2005, 2001 Printed in USA, February 8, 2005 5091-3117E
–10
–20
–30
dB
–40
–50
–60
0.3 MHz 10 MHz 100 MHz 1 GHz 6 GHz
Directivity
Frequency
–10
–20
–30
dB
–40
–50
–60
0.3 MHz 1 MHz 10 MHz 1 GHz 3 GHz
Directivity
100 MHz
Frequency
–1.1
–1.3
–1.5
–1.7
dB
–1.9
–2.1 –2.3 –2.5
0.3 MHz 3 GHz 6 GHz
–15.6
–15.8
–16.0
–16.2
dB
–16.4
–16.6 –16.8 –17.0
0.3 MHz 3 GHz 6 GHz
Insertion Loss
Frequency
Coupling Factor
Frequency