Agilent 86205A Specifications Sheet

Network Analyzers
294 Accessories
HP 8753E
Series
HP 86205A/86207A
HP 86205A/86207A RF Bridges
The HP 86205A/86207A high directivity RF bridges offer unparalleled performance in a variety of general-purpose applications. They are ideal for accurate reflection measurements and signal leveling applications.
Specifications Summary
HP 86205A HP 86207A
Impedance 50 75 Freq. Range 300 kHz to 6 GHz 300 kHz to 3 GHz
5
Directivity 30 dB, 0.3 MHz to 5 MHz 30 dB, 0.3 MHz to 5 MHz
Coupling Factor (< 3 GHz) 16.0 dB, + 0.15 dB/GHz
Insertion Loss 1.5 dB, + 0.1 dB/GHz Maximum Input 25 dBm RF Connectors 50 Type-N (female) 75 Type-N (female)
40 dB, 5 MHz to 2 GHz 40 dB, 5 MHz to 1.3 GHz 30 dB, 2 GHz to 3 GHz 35 dB, 1.3 GHz to 2 GHz 20 dB, 3 GHz to 5 GHz (typ.) 30 dB, 2 GHz to 3 GHz 16 dB, 5 GHz to 6 GHz (typ.)
(> 3 GHz) 16.5 dB, – 0.20 dB/GHz
Physical Characteristics
Size: 93 mm H x 160 mm W x 23 mm D (3.7 in x 6.3 in x 1 in) Weight: Net, 0.57 kg (1.3 lb); shipping, 1.8 kg (4 lb)
HP 11850C/D Three-Way Power Splitters Specifications Summary
HP 11850C HP 11850D
Impedance 50 75 Frequency Range DC to 3 GHz DC to 2 GHz Tracking ±0.25 dB, ±3˚ ±0.2 dB, ± 2.5˚ Equivalent Source Match 30 dB at 1.3 GHz 30 dB at 1.3 GHz
(ratio or leveling) 20 dB at 3 GHz 20 dB at 3 GHz
Nominal Insertion Loss 9.5 dB + 1 dB/GHz 7.8 dB Input Port Match
DC to 1.3 GHz 20 dB 20 dB
1.3 GHz to F
RF Connectors
RF Input: (female) 50 ΩType-N 50 ΩType-N All Others: (female) 50 ΩType-N 75 ΩType-N
1
max
F
1
max
is the upper frequency limit of the associated power splitter.
10 dB 10 dB
HP 11851B RF Cable Kit
This kit includes three 610-mm (24-in) 50 cables phase matched to 4˚ at 1.3 GHz and one cable 860 mm (34-in). Connectors are type-N (male).
Maximum Input Power: 250 mW (+24 dBm) RF Connectors: 50 type-N (f) and 75 type-N (m) standard,
50 type-N (m) and 75 type-N (f) Option 004
Type-N Accessory Kits
Each kit contains a type-N (female) short, a type-N (male) short, two type-N (male) barrels, two type-N (female) barrels, and a storage case.
HP 11853A 50 Type-N Accessory Kit
The HP 11853A accessory kit furnishes the RF components required for measurement of devices with 50 type-N connectors using the HP 11850C, 85044A, 85046A, or 85047A.
HP 11855A 75 Type-N Accessory Kit
The HP 11855A accessory kit furnishes the RF components required for measurement of devices with 75 type-N connectors using the HP 11850D, 85044B, or 85046B. This kit also contains a 75 Ω type-N (male) termination.
BNC Accessory Kits
The BNC accessory kit contains two type-N (male) to BNC(female) adapters, two type-N (male) to BNC (male) adapters, two type-N (female) to BNC (female) adapters, two type-N (female) to BNC (male) adapters, a BNC (male) short, and a storage case.
HP 11854A 50 BNC Accessory Kit
The HP 11854A accessory kit furnishes the RF components required for measurement of devices with 50 BNC connectors using the HP 11850C, 85044A, 85046A, or 85047A.
HP 11856A 75 BNC Accessory Kit
The HP 11856A furnishes RF components required for measurement of devices with 75 BNC connectors using the HP 11850D, 85044B, or 85046B. This kit also contains a 75 BNC (male) termination.
Test Port Cables
HP 11857D 50 APC-7 Test Port Cables
The HP 11857D includes two precision 61-cm (24-in) cables, phase matched to 2˚ at 1.3 GHz for use with the HP 8753D/E, 85046A or 85047A S-parameter test sets. Connectors are 50 APC-7.
HP 11857B 75 Type-N Test Port Cables
The HP 11857B includes two precision 61-cm (24-in) cables, phase matched to 2˚ at 1.3 GHz for use with the HP 8753D/E Option 075 or HP 85046B S-parameter test set. One cable has 75 type-N (male) con­nectors on both ends; the other has one type-N (male) and one type-N (female) connector.
Transit Cases
HP offers a complete line of sturdy transit cases that protect your instrument from shock, vibration, moisture, impact, and contamination, providing a secure enclosure for shipping. Model 9211-2657 fits the HP 8753E and model 9211-2656 fits the HP 8752C.
Calibration Kits
The calibration kits in the HP 8753 family contain precision standards used in accuracy enhancement procedures to characterize the system­atic errors of an HP 8753 measurement system.
HP 85031B 7-mm Calibration Kit
The HP 85031B calibration kit contains a set of precision 7-mm fixed terminations, and a one-piece open/short circuit used to calibrate the HP 8753 and its 50 test sets for measurement of devices with preci­sion 7-mm connectors. This kit is specified 300 kHz to 6 GHz.
HP 11852B 50 Ω/75 Ω Minimum Loss Pad
The HP 11852B is a low SWR minimum loss pad required for measure­ments on 75 devices with the HP 8753.
Frequency Range: DC to 3.0 GHz Insertion Loss: 5.7 dB Return Loss: 75 : typically 30 dB, 50 : typically 26 dB