Agilent 86100A Quick Start Guide

Quick Start Guide
© Copyright Agilent Technol­ogies, Inc. 2000 All Rights Reserved. Repro­duction, adaptat ion, or trans­lation without prior written permission is prohibited, except as allowed under copy­right laws.
Agilent Technologies Part No. 86100- 90002 Printed in USA February 2000
Agilent Technologies, Inc. Lightwave Division 1400 Fountaingrove Park way Santa Rosa, CA 9 5403-1799, USA (707) 577-1 400
Notice.
The information contained in this document is subject to change without notice. Com­panies, names, and data used in examples herein are ficti­tious unless otherwise noted. Agilent Technologies makes no warranty of any kind with regard to this material, includ­ing but not limited to, the implied warranties of mer­chantability and fitness for a particular purpose. Agilent Technologies shall not be lia­ble for er rors contained herein or for incidental or conse­quential damages in connec­tion with the furnishing, performance, or use of this material.
Restricted Rights Legend.
Use, duplication, or disclo­sure by the U.S. Government is subject to restrictions as set forth in subparagraph (c) (1) (ii) of the Rights in Technical Data and Computer Software clause at DFARS 252.2 27-7013 for DOD agencies, and sub­paragraphs (c) (1) and (c) (2 ) of the Commercial Computer
Software Restricted Rights claus e a t FAR 52 .22 7-19 for other agencies.
Warrant y.
This Agilent Technologies instrument product is war­ranted against defects in material and workmanship for a period of one year from date of shipment. During the war­ranty period, Agilent Technol­ogies Company will, at its option, either repa ir or replace products which prove to be defective. For warranty service or repair, this product must be returned to a service facility designated by Agilent Technologies. Buyer shall pre­pay shipping charges to Agi­lent Technologies and Agilent Technologies shall pay ship­ping charges to return the product to Buyer. However, Buyer shall pay all shipping charges, duties, and taxes for products returned to Agilent Technologies from another country.
Agilent Technologies war­rants that its software and firmware designated by Agi­lent Technologies for use with an instrument will execute its programming instructions when properly installed o n that instrument. Agilent Tech­nologies does not warrant that the operation of the instru­ment, or sof tware, or firmware will be uninterrupted or error­free.
Limitation of Warranty.
The foregoing warranty shall not apply to defects resulting from improper or inadequate maintenance by Buyer, Buyer­supplied software or interfac­ing, unauthorized modifica­tion or misuse, operation outside of the environmental
specifications for the product, or improper site preparation or maintenance.
No other warranty is expressed or implied. Agilent Technologies specifi cally dis­claims the implied warranties of merchantability and fitness for a particular purpose.
Exclusive Remedies.
The remedies provided herein are buyer's sole and exclusive remedies. Agilent Technolo­gies shall not be li able for any direct, indirect, special, inci­dental, or consequential dam­ages, whether based on contract, tort, or any o ther legal theory.
Safety Symbols.
CAUTION
caution
The
sign de notes a hazard. It calls attention to a procedure which, if not cor­rectly performed or adhered to, could result in damage to or destruction of the product. Do not proceed beyond a cau­tion sign until the indicated conditions are fully under­stood and met.
WARNING
warning
The
sign denotes a hazard. It calls attention to a procedure which, if not cor­rectly performed or adhered to, could result in injury or loss of life. Do not proceed beyond a warning sign until the indicated conditions are fully understood and met.
The instruction man­ual symbol. The prod­uct is marked with this warning symbol when it is necessary for the user to refer to the instructions in the manual.
The laser radiation symbol. This warning symbol is marked on products which have a laser output.
The AC symbol is used to indicate the required nature o f the line module input power.
|The ON symbols are used to mark the posi­tions of the instrument power line switch.
❍The OFF symbols are used to mark the positions of the instru­ment power line switch.
The CE mark is a reg­iste r ed tra demark of the European Commu­nity.
The CSA mark is a reg­iste r ed tra demark of the Canadian Stan­dards Association.
The C-Tick mark is a registered trademark of the Australian Spec­trum Management Agency.
This text denotes the
ISM1-A
instrument is an Industrial Scientific and Medical Group 1 Class A product.
ii

In This Book

In This Book
This book gives you the information you need to begin using the Infiniium DCA. It contains five chapters:
Chapter 1, “
Chapter 1, “Setting Up the Infiniium DCA
Chapter 1, “Chapter 1, “
ments, air flow, and setup information.
Chapter 2, “
Chapter 2, “Working in Comfort
Chapter 2, “Chapter 2, “
fortably and safely while operating the Infiniium DCA.
Chapter 3, “
Chapter 3, “Using the Infiniiu m DCA
Chapter 3, “Chapter 3, “
the graphical user interface, and tells you how to perform basic operations with the Infiniium DCA.
Chapter 4, “
Chapter 4, “Using the Built-In Information System
Chapter 4, “Chapter 4, “
information system contents and navigation. The built-in information system contains all of the information that is generally found in a user’s guide.
Chapter 5, “
Chapter 5, “Regulatory Information
Chapter 5, “Chapter 5, “
tion.
Setting Up the Infiniium DCA”””” contains inspection, power require-
Setting Up the Infiniium DCASetting Up the Infiniium DCA
Working in Comfort”””” contains recommendations for working com-
Working in ComfortWorking in Comfort
Using the Infiniium DCA”””” gives an overview of the front panel and
Using the Infiniium DCAUsing the Infiniium DCA
Using the Built-In Information System”””” describes the built-in
Using the Built-In Information SystemUsing the Built-In Information System
Regulatory Information”””” contains important regulatory informa-
Regulatory InformationRegulatory Information
CAUTION
• For detailed information on how the Infiniium DCA makes measurements and how to use the instrument, see the built-in information system in the instru­ment.
• For information on programming the Infiniium DCA using a computer with a GPIB interface card, refer to the
The Infiniium DCA uses a specially designed Windows 98 application program (Windows ® is a U.S. registered trademark of Microsoft Corp.) All Infiniium DCA functionality is directly available from within the Infiniium DCA application. Windows 98 configuration changes made outside of the Infiniium DCA application may not work correctly and could cause the instrument to become inoperable. Do not try to access or make changes to the Windows Operating system. Repairs caused by the improper use of the Infiniium DCA will not be covered under warranty.
Infiniium DCA Programmer’s Guide
.
iii
General Safety Considerations
General Safety Considerations
This product has been designed and tested in accordance with IEC Publica­tion 1010, Safety Requirements for Electronic Measuring Apparatus, and has been supplied in a safe condition. The instruction documentation contains information and warnings which must be followed by the user to ensure safe operation and to maintain the product in a safe condition.
Install the instrument according to the enclosure protection provided. This instrument does not protect against the ingress of water. This instrument pro­tects against finger access to hazardous parts within the enclosure.
WARN ING
WARN ING
WARN ING
CAUTION
If this product is not used as specified, the protection provided by the equipment could be impaired. This product must be used in a normal condition (in which all means for protection are intact) only.
No operator serviceable parts inside. Refer servicing to qualified service personnel. To prevent electrical shock do not remove covers.
To prevent electrical shock, disconnect the instrument from mains before cleaning. Use a dry cloth or one slightly dampened with water to clean the external case parts. Do not attempt to clean internally.
Do not use too much liquid in cleaning the instrument. Water can enter the front-panel keyboard, damaging sensitive electronic components.
iv

Contents

In This Book iii
1 Setting Up the Infiniium DCA
Installing the Infiniium DCA 1-3 Cleaning Connections for Accurate Measurements 1-26 Returning the Instrument for Service 1-35 Agilent Technologies Service Offices 1-38
2 Wo r k i ng in Co mfort
3 Using the Infiniium DCA
Eye/Mask Mode 3-7 Oscilloscope Mode 3-8 TDR/TDT Mode 3-9 Using Plug-in Modules 3-10 Using the Touch Screen 3-12 Using the Quick Measure Feature 3-14 If the Display Goes Blank 3-15 Hiding and Moving Dialog Boxes 3-16 Menus 3-17
4 Using the Built-In Information System
The Contents Topic 4-3 Using Setup Guides 4-4 Getting Help While Changing Instrument Settings 4-6 Learning About Measurements Results 4-7 Hiding the Built-in Information System 4-8 Printing the Contents of a Topic 4-9
5 Regulatory Informa tion
Contents-1
Contents
Contents -2
1
Installing the Infiniium DCA 1-3
Step 1. Inspect the shipment 1-3 Step 2. Install any modules 1-7 Step 3. Connect the keyboard and mouse 1-8 Step 4. Connect the line cord 1-9 Step 5. Turn on the line power 1-12 Step 6. Set the time and date 1-13 Step 7. Connect a printer 1-14 Step 8. Configuring the touch screen 1-19 Step 9. Avoid costly repairs 1-21 Step 10. Perform a vertical calibration 1-22
Step 11. For more information 1-25 Cleaning Connections for Accurate Measurements 1-26 Returning the Instrument for Service 1-35 Agilent Technologies Service Offices 1-38

Setting Up the Infiniium DCA

Setting Up the Infiniium DCA
Setting Up the Infiniium DCA
Setting Up the Infiniium DCA
This chapter shows you how to set up your Infiniium DCA and connect the power and the accessories. It also shows you how to use proper optical con­nection cleaning techniques to avoid
costly
repairs.
WARN ING
CAUTION
CAUTION
This is a Safety Class 1 Product (provided with a protective earthing ground incorporated in the power cord). The ma ins plug shall only be inserted in a socket outlet provided with a protective earth contact. Any interruption of the protective conductor inside or outside of the instrument is likely to make the instrument dangerous. Intentional interruption is prohibited.
VENTILATION REQUIREMENTS: When installi ng the product in a cabinet, the convection into and out of the product must not be restricted. The ambient temperature (outside the cabinet) must be less than the maximum operating temperature of the product by 4° C for every 100 watts dissipated in the cabinet. If the total power dissipated in the cabinet is greater than 800 watts, then forced convection must be used.
This product is designed for use in Installation Category II and Pollution Degree 2 per IEC 1010 and 664 respectively.
Installation Category: Installation Categories (overvoltage categories) are determined by the transient overvoltage levels that may be expected. CAT I: Mains isolated. CAT II: Line voltage in appliance and to wall outlet. CAT III: Line voltage behind wall outlet to next level of distribution.
1-2

Installing the Infiniium DCA

Step 1. Inspect the shipment

Setting Up the Infiniium DCA
Installing the Infiniium DCA
Inspect the shipping container for damage.
Inspect the instrument.
Verify that you received the options and accessories you ordered.
1-3
Setting Up the Infiniium DCA
Step 1. Inspect the shipment
Keep the shipping container and cushioning material until you have inspected the contents of the shipment for completeness and have checked the instru­ment mechanically and electrically.
If anything is missing or defective, contact your nearest Agilent Technologies Sales Office. Refer to Agilent Technologies Service Offices on page 1-38. If the shipment was damaged, contact the carrier, then contact the nearest Agi­lent Sales Office. Keep the shipping materials for the carriers inspection. The Agilent Sales Office will arrange for repair or replacement at Agilents option without waiting for claim settlement.
Table 1-1. Standard Accessories
Accessory Part Number
Keyboard E2610-68700
Mouse C3751-60201
Quick Start Guide (this book) 86100-90002
Programmer’s Guide 86100-90003
Table 1-2. Product Options
Option Description
001 12 GHz Trigger Bandwidth
AX4 Rack mount flange kit
AXE Rack mount flange kit with handles
UK6 Commercial calibration certificate with test data
1-4
Table 1-3. Optional Accessories
Accessory Description
83446A Lightwave clock and data receiver
11982A High-speed lightwave receiver
83440B/C/D High-speed lightwave receiver
54118A 500 MHz to 18 GHz trigger
N1020A TDR probe kit
54006A 6 GHz hand-held probe
54701 2.5 GHz active probe
83480A-K17 Switch matrix
83480A-K17 RIMM fixture
10086A ECL terminator
83430A Lightwave digital source
11898A Plug-in module remote/extender
54008B 24 ns delay line
Setting Up the Infiniium DCA
Step 1. Inspect the shipment
Table 1-4. Additional Adapters for Electrical Channels
Adapter Agilent Part Number
SMA (f-f) 1250-1158
APC 3.5 (f-f) 1250-1749
1-5
Setting Up the Infiniium DCA
Step 1. Inspect the shipment
Table 1-5. Fiber-Optic Adapters for Optical Channels
Front Panel Fiber-Optic Adapter Description Agilent Part Number
Diamond HMS-10 81000AI
FC/PC 81000FI
D4 81000GI
SC 81000KI
DIN 81000SI
1-6
ST 81000VI
Biconic 81000WI
Setting Up the Infiniium DCA

Step 2. Install any modules

Step 2. Install any modules
Up to two modules can be inserted into the Infiniium DCA. All documentation (including specifications) for plug-in modules is located in the Infiniium DCA’s built-in information system. To access the built-in information system, touch or click Contents on the Help menu .
1
Slide the module into an available front-panel slot.
You can remove or install a module while the instrument is operating.
2
Finger tighten the two knurled screws on the modules front panel.
1-7
Setting Up the Infiniium DCA

Step 3. Connect the keyboard and mouse

Step 3. Connect the keyboard and mouse
Connect the standard PC-compatible mouse and keyboard.
Although you can operate all instrument functions using a keyboard and mouse, the touch screen feature makes your Infiniium DCA even easier to use. The touch screen feature is always available, even when you connect the mouse. For more information, refer to Chapter 4, “Using the Built-In Informa-
tion System, or use the built-in information system.
not
Do
stack other objects on the keyboard; this will cause self-test failures on
power-on.
1-8

Step 4. Connect the line cord

Setting Up the Infiniium DCA
Step 4. Connect the line cord
CAUTION
CAUTION
CAUTION
The Infiniium DCA automatically adjusts for line input voltages in the range of 100 to 240 VAC. There is no manual selection switch. The line cord provided is matched by Agilent to the country in which the order originates. Refer to
Table 1-8, Available Line Cords, on page 1-11 for a list of available line power
cords.
Always use the three-prong AC power cord supplied with this instrument. Failure to ensure adequate earth grounding by not using this cord may cause instrument damage.
Do not
connect ac power until you have verified the line voltage is correct as
described in Table 1-6. Damage to the equipment could result.
This instrument has autoranging line voltage input. Be sure the supply voltage is within the specified range.
1-9
Setting Up the Infiniium DCA
Step 4. Connect the line cord
Table 1-6. Line Power Requirements
Power 115 VAC: 110 VA MAX. / 60 WATTS MAX. / 1.1 A MAX.
230 VAC: 150 VA MAX. / 70 WATTS MAX. / 0.6 A MAX.
Voltage nominal: 115 VAC / 230 VAC CAT II
range 115 VAC: 90–132 V range 230 VAC: 198–254 V
Frequency nominals: 50 Hz / 60 Hz
range: 47–63 Hz
Table 1-7. Environmental Specifications
Use indoor Temperature
Operating 10 °C to +40 °C (50 °F to + 104 °F) Non-operating -40 °C to +70 °C (-40 °F to +158 °F)
Humidity
Operating Up to 90% humidity (non-condensing) at +40 °C (+104 °F) Non-operating Up to 95% relative humidity at + 65 °C (+149 °F)
Altitude
Operating Up to 4,600 meters (15,000 ft) Non-operating Up to 15,300 meters (50,000 ft)
Vibration
Operating Random vibration 5-500 Hz, 10 minutes per axis, 2.41 g(rms) Non-operating Random vibration 5-500 Hz, 10 minutes per axis, 0.3 g(rms);
Resonant search, 5-500Hz swept sine, 1 octave/minute sweep rate, 0.75g, 5 minute resonant dwell at 4 resonances per axis.
Power requirements
Voltage 90 to 132 or 198 to 264 Vac, 48-66 Hz CAT II Power (including modules) 545 VA; 375 W
Weight
Mainframe 12.7 kg (28 lb) Modules 1.2 kg (2.6 lb)
Dimensions
Height 215.1 mm (8.47 in) Width 425.5 mm (16.75 in) Depth 521 mm (20.5 in) Depth with front connectors
and rear feet
584 mm (23.0 in)
1-10
Table 1-8. Available Line Cords
Setting Up the Infiniium DCA
Step 4. Connect the line cord
Plug Type Cable Part No. Plug Description
250V 8120-1351
8120-1703
250V 8120-1369
Straight *BS1363A 90°
Straight *NZSS198/ASC 90°
8120-0696
250V 8120-1689
8120-1692 8120-2857p
125V 8120-1378
8120-1521 8120-1992
250V 8120-2104
8120-2296
Straight *CEE7-Y11 90° Straight (Shielded)
Straight *NEMA5-15P 90° Straight (Medical) UL544
Straight *SEV1011 1959-24507 Type 12 90°
Length (in/cm)
90/228 90/228
79/200
87/221
79/200 79/200 79/200
90/228 90/228 96/244
79/200 79/200
Color Country
Gray Mint Gray
United Kingdom, Cyprus, Nigeria, Zimbabwe, Singapore
Gray
Australia, New Zealand
Mint Gray
Mint Gray Mint Gray Coco Brown
East and West Europe, Saudi Arabia, So. Africa, India (unpolarized in many nations)
Jade Gray Jade Gray Black
United States, Canada, Mexico, Philippines, Taiwan
Mint Gray
Switzerland
Mint Gray
220V 8120-2956
8120-2957
250V 8120-4211
8120-4600
100V 8120-4753
8120-4754
* Part number sho wn for plug is the i ndustry identifier for the plug only. Number sho wn for cable is the Agilent
Technologies part number for the complete cable including the plug.
Straight *DHCK107 90°
Straight SABS164 90°
Straight MITI 90°
79/200 79/200
79/200 79/200
90/230 90/230
Mint Gray
Denmark
Mint Gray
Jade Gray Republic of South
Africa India
Dark Gray Japan
1-11
Setting Up the Infiniium DCA

Step 5. Turn on the line power

Step 5. Turn on the line power
Press the power switch at the lower left-hand corner of the front panel.
After about one minute, the display will look similar to the following figure. The Infiniium DCA is ready to use.
1-12
Setting Up the Infiniium DCA

Step 6. Set the time and date

Step 6. Set the time and date
1
Touch the current time and date shown near the top of the display. Notice how the instrument responds to your touch by displaying the current time and date settings. You can touch the screen with your finger, ballpoint pen, or a stylus.
You can also click the current time and date with a mouse. Although this book uses the term “touch” to encourage you to experience using the touch screen, you can always click using the mouse instead.
CAUTION
Avoid touching the screen with a sharp object as this could result in damage to the display.
2
Touch the various displayed settings to set the correct time for your location. After setting the correct time, touch or click Set Time & Date Now to change the time.
1-13
Setting Up the Infiniium DCA

Step 7. Connect a printer

Step 7. Connect a printer
Use of a printer with the Infiniium DCA is optional. Table 1-9 lists the PCL-lan­guage printers that are currently supported for use with the instrument. You must use the Add Printer Wizard to select the printer driver for the print er as outlined in this procedure; the Add Print er Wizard does not run automatically. For more information on connecting and using a printer, refer to the built-in information system.
Unsupported printers
If your printer is not in the list of supported printers, you will need the printer driver disk that came with your printer. To install the printer driver from this disk, click Have Disk on the Add Printer Wizard dialog box. The instrument will then prompt you to put your printer driver disk into the front-panel disk drive (drive A). Be sure to install the Windows 98 printer driver for your printer. If a printer driver disk was not provided or you cannot find the disk, you can contact the printer’s manufacturer to obtain the appropriate Windows 98 printer driver.
Table 1-9. Supported Printers
HP DeskJet 850 HP DeskJet 1600C
HP DeskJet 890C HP LaserJet 4P
HP DeskJet 970 HP LaserJet 4000N
HP DeskJet 1200C
To connect a printer
1
Connect the printer to the instruments rear-panel
Parallel
port. Use the cable
that came with your printer.
2
On the File menu, select Print Setup and then Add Printer.
3
Select Printer that is Plug and Play.
An online Help window opens with instructions for connecting a printer. You do not need to read the displayed instructions at this time.
4
Turn the Infiniium DCAs power off. Do not exit the online Help before turning the instrument off.
5
Turn the printers power on.
1-14
Setting Up the Infiniium DCA
Step 7. Connect a printer
.
6
Wait five seconds and then turn the Infiniium DCAs power on again.
7
When the following message appears, select “No”.
8
When the following message appears select “OK”.
1-15
Setting Up the Infiniium DCA
Step 7. Connect a printer
9
The instrument automatically recognizes the printer connected on the printer port and displays the Add Printer Wizard. Touch Next.
10
Select Local printer and then touch Next.
11
Select the printer manufacturer and then printer. Then, touch Next. If your printer is not listed, click Have Disk. The in strument will then prompt you to put your printer driver disk into the front-panel disk drive (drive A). Be sure to install the Windows 98 printer driver for your printer.
1-16
12
Select LPT1 for the printer port. Touch Next.
13
Select Yes and then touch Next.
Setting Up the Infiniium DCA
Step 7. Connect a printer
14
Select Yes to print a test page. Then, touch Next.
1-17
Setting Up the Infiniium DCA
Step 7. Connect a printer
15
After confirming that the test page printed correctly, touch Yes.
16
Turn the Infiniium DCA’s power off and then on again. The printer is now installed and ready to us e.
1-18
Setting Up the Infiniium DCA

Step 8. Configuring the touch screen

Step 8. Configuring the touch screen
The touch screen configuration utility ensures that the touch screen is both aligned and orientated with the display. It also allows you to adjust the touch screen speed to simulate a double-click and to turn the touch screen sound on or off.
1
On the Utilities menu, select Touch Screen Config. The touch screen configuration utility is displayed.
2
Touch Calibration.
3
Touch the Calibrate button and follow the displayed instructions. You do not need to select Calibrate TouchSurround; this feature is not used in configuring the touch screen.
4
After returning to the Configuration Utilities dialog box shown in Step 1, touch Properties.
1-19
Setting Up the Infiniium DCA
Step 8. Configuring the touch screen
5
The Display Area Configuration dialog box shown here is displayed. Although the Beep on Touch and the Drag and Drop settings are bes t left selected, you can turn them off if desired. This figure shows the default settings. Pressing the front-panel Default Setup button does not change these settings.
The Double Touch Feature
Since the double touch feature is rarely used in the Infiniium DCA, youll probably never need to adjust Double Touch Speed or Timed Double Touch. One example of using double touch is when you open a file. You can double touch the file name instead of pressing the Open button.
Timed Double Touch allows a single touch to be interpreted as a double touch if you con­tinue the touch for a set amount of time. If this setting is set to None (the default set­ting), two touches are required. You then set the Double Touch Speed for the time needed between two touches. The Test Box button is used to test your double touch set­tings.
1-20