Agilent Technologies Part No.
86100- 90002
Printed in USA
February 2000
Agilent Technologies, Inc.
Lightwave Division
1400 Fountaingrove Park way
Santa Rosa, CA 9 5403-1799,
USA
(707) 577-1 400
Notice.
The information contained in
this document is subject to
change without notice. Companies, names, and data used
in examples herein are fictitious unless otherwise noted.
Agilent Technologies makes
no warranty of any kind with
regard to this material, including but not limited to, the
implied warranties of merchantability and fitness for a
particular purpose. Agilent
Technologies shall not be liable for er rors contained herein
or for incidental or consequential damages in connection with the furnishing,
performance, or use of this
material.
Restricted Rights Legend.
Use, duplication, or disclosure by the U.S. Government
is subject to restrictions as set
forth in subparagraph (c) (1)
(ii) of the Rights in Technical
Data and Computer Software
clause at DFARS 252.2 27-7013
for DOD agencies, and subparagraphs (c) (1) and (c) (2 )
of the Commercial Computer
Software Restricted Rights
claus e a t FAR 52 .22 7-19 for
other agencies.
Warrant y.
This Agilent Technologies
instrument product is warranted against defects in
material and workmanship for
a period of one year from date
of shipment. During the warranty period, Agilent Technologies Company will, at its
option, either repa ir or
replace products which prove
to be defective. For warranty
service or repair, this product
must be returned to a service
facility designated by Agilent
Technologies. Buyer shall prepay shipping charges to Agilent Technologies and Agilent
Technologies shall pay shipping charges to return the
product to Buyer. However,
Buyer shall pay all shipping
charges, duties, and taxes for
products returned to Agilent
Technologies from another
country.
Agilent Technologies warrants that its software and
firmware designated by Agilent Technologies for use with
an instrument will execute its
programming instructions
when properly installed o n
that instrument. Agilent Technologies does not warrant that
the operation of the instrument, or sof tware, or firmware
will be uninterrupted or errorfree.
Limitation of Warranty.
The foregoing warranty shall
not apply to defects resulting
from improper or inadequate
maintenance by Buyer, Buyersupplied software or interfacing, unauthorized modification or misuse, operation
outside of the environmental
specifications for the product,
or improper site preparation
or maintenance.
No other warranty is
expressed or implied. Agilent
Technologies specifi cally disclaims the implied warranties
of merchantability and fitness
for a particular purpose.
Exclusive Remedies.
The remedies provided herein
are buyer's sole and exclusive
remedies. Agilent Technologies shall not be li able for any
direct, indirect, special, incidental, or consequential damages, whether based on
contract, tort, or any o ther
legal theory.
Safety Symbols.
CAUTION
caution
The
sign de notes a
hazard. It calls attention to a
procedure which, if not correctly performed or adhered
to, could result in damage to
or destruction of the product.
Do not proceed beyond a caution sign until the indicated
conditions are fully understood and met.
WARNING
warning
The
sign denotes a
hazard. It calls attention to a
procedure which, if not correctly performed or adhered
to, could result in injury or
loss of life. Do not proceed
beyond a warning sign until
the indicated conditions are
fully understood and met.
The instruction manual symbol. The product is marked with this
warning symbol when
it is necessary for the
user to refer to the
instructions in the
manual.
The laser radiation
symbol. This warning
symbol is marked on
products which have a
laser output.
The AC symbol is used
to indicate the
required nature o f the
line module input
power.
|The ON symbols are
used to mark the positions of the instrument
power line switch.
❍The OFF symbols
are used to mark the
positions of the instrument power line
switch.
The CE mark is a registe r ed tra demark of
the European Community.
The CSA mark is a registe r ed tra demark of
the Canadian Standards Association.
The C-Tick mark is a
registered trademark
of the Australian Spectrum Management
Agency.
This text denotes the
ISM1-A
instrument is an
Industrial Scientific
and Medical Group 1
Class A product.
ii
In This Book
In This Book
This book gives you the information you need to begin using the Infiniium
DCA. It contains five chapters:
Chapter 1, “
Chapter 1, “Setting Up the Infiniium DCA
Chapter 1, “Chapter 1, “
ments, air flow, and setup information.
Chapter 2, “
Chapter 2, “Working in Comfort
Chapter 2, “Chapter 2, “
fortably and safely while operating the Infiniium DCA.
Chapter 3, “
Chapter 3, “Using the Infiniiu m DCA
Chapter 3, “Chapter 3, “
the graphical user interface, and tells you how to perform basic operations
with the Infiniium DCA.
Chapter 4, “
Chapter 4, “Using the Built-In Information System
Chapter 4, “Chapter 4, “
information system contents and navigation. The built-in information system
contains all of the information that is generally found in a user’s guide.
Chapter 5, “
Chapter 5, “Regulatory Information
Chapter 5, “Chapter 5, “
tion.
Setting Up the Infiniium DCA”””” contains inspection, power require-
Setting Up the Infiniium DCASetting Up the Infiniium DCA
Working in Comfort”””” contains recommendations for working com-
Working in ComfortWorking in Comfort
Using the Infiniium DCA”””” gives an overview of the front panel and
Using the Infiniium DCAUsing the Infiniium DCA
Using the Built-In Information System”””” describes the built-in
Using the Built-In Information SystemUsing the Built-In Information System
Regulatory Information”””” contains important regulatory informa-
Regulatory InformationRegulatory Information
CAUTION
• For detailed information on how the Infiniium DCA makes measurements and
how to use the instrument, see the built-in information system in the instrument.
• For information on programming the Infiniium DCA using a computer with a
GPIB interface card, refer to the
The Infiniium DCA uses a specially designed Windows 98 application program
(Windows ® is a U.S. registered trademark of Microsoft Corp.) All Infiniium
DCA functionality is directly available from within the Infiniium DCA
application. Windows 98 configuration changes made outside of the Infiniium
DCA application may not work correctly and could cause the instrument to
become inoperable. Do not try to access or make changes to the Windows
Operating system. Repairs caused by the improper use of the Infiniium DCA
will not be covered under warranty.
Infiniium DCA Programmer’s Guide
.
iii
General Safety Considerations
General Safety Considerations
This product has been designed and tested in accordance with IEC Publication 1010, Safety Requirements for Electronic Measuring Apparatus, and has
been supplied in a safe condition. The instruction documentation contains
information and warnings which must be followed by the user to ensure safe
operation and to maintain the product in a safe condition.
Install the instrument according to the enclosure protection provided. This
instrument does not protect against the ingress of water. This instrument protects against finger access to hazardous parts within the enclosure.
WARN ING
WARN ING
WARN ING
CAUTION
If this product is not used as specified, the protection provided by the
equipment could be impaired. This product must be used in a normal
condition (in which all means for protection are intact) only.
No operator serviceable parts inside. Refer servicing to qualified
service personnel. To prevent electrical shock do not remove covers.
To prevent electrical shock, disconnect the instrument from mains
before cleaning. Use a dry cloth or one slightly dampened with water
to clean the external case parts. Do not attempt to clean internally.
Do not use too much liquid in cleaning the instrument. Water can enter the
front-panel keyboard, damaging sensitive electronic components.
iv
Contents
In This Book iii
1 Setting Up the Infiniium DCA
Installing the Infiniium DCA 1-3
Cleaning Connections for Accurate Measurements 1-26
Returning the Instrument for Service 1-35
Agilent Technologies Service Offices 1-38
2 Wo r k i ng in Co mfort
3 Using the Infiniium DCA
Eye/Mask Mode 3-7
Oscilloscope Mode 3-8
TDR/TDT Mode 3-9
Using Plug-in Modules 3-10
Using the Touch Screen 3-12
Using the Quick Measure Feature 3-14
If the Display Goes Blank 3-15
Hiding and Moving Dialog Boxes 3-16
Menus 3-17
4 Using the Built-In Information System
The Contents Topic 4-3
Using Setup Guides 4-4
Getting Help While Changing Instrument Settings 4-6
Learning About Measurements Results 4-7
Hiding the Built-in Information System 4-8
Printing the Contents of a Topic 4-9
5 Regulatory Informa tion
Contents-1
Contents
Contents -2
1
Installing the Infiniium DCA1-3
Step 1. Inspect the shipment1-3
Step 2. Install any modules1-7
Step 3. Connect the keyboard and mouse1-8
Step 4. Connect the line cord1-9
Step 5. Turn on the line power1-12
Step 6. Set the time and date1-13
Step 7. Connect a printer1-14
Step 8. Configuring the touch screen1-19
Step 9. Avoid costly repairs1-21
Step 10. Perform a vertical calibration1-22
Step 11. For more information1-25
Cleaning Connections for Accurate Measurements1-26
Returning the Instrument for Service1-35
Agilent Technologies Service Offices1-38
Setting Up the Infiniium DCA
Setting Up the Infiniium DCA
Setting Up the Infiniium DCA
Setting Up the Infiniium DCA
This chapter shows you how to set up your Infiniium DCA and connect the
power and the accessories. It also shows you how to use proper optical connection cleaning techniques to avoid
costly
repairs.
WARN ING
CAUTION
CAUTION
This is a Safety Class 1 Product (provided with a protective earthing
ground incorporated in the power cord). The ma ins plug shall only be
inserted in a socket outlet provided with a protective earth contact.
Any interruption of the protective conductor inside or outside of the
instrument is likely to make the instrument dangerous. Intentional
interruption is prohibited.
VENTILATION REQUIREMENTS: When installi ng the product in a cabinet, the
convection into and out of the product must not be restricted. The ambient
temperature (outside the cabinet) must be less than the maximum operating
temperature of the product by 4° C for every 100 watts dissipated in the
cabinet. If the total power dissipated in the cabinet is greater than 800 watts,
then forced convection must be used.
This product is designed for use in Installation Category II and Pollution
Degree 2 per IEC 1010 and 664 respectively.
Installation Category: Installation Categories (overvoltage categories) are
determined by the transient overvoltage levels that may be expected. CAT I:
Mains isolated. CAT II: Line voltage in appliance and to wall outlet. CAT III:
Line voltage behind wall outlet to next level of distribution.
1-2
Installing the Infiniium DCA
Step 1. Inspect the shipment
Setting Up the Infiniium DCA
Installing the Infiniium DCA
❒
Inspect the shipping container for damage.
❒
Inspect the instrument.
❒
Verify that you received the options and accessories you ordered.
1-3
Setting Up the Infiniium DCA
Step 1. Inspect the shipment
Keep the shipping container and cushioning material until you have inspected
the contents of the shipment for completeness and have checked the instrument mechanically and electrically.
If anything is missing or defective, contact your nearest Agilent Technologies
Sales Office. Refer to “Agilent Technologies Service Offices” on page 1-38. If
the shipment was damaged, contact the carrier, then contact the nearest Agilent Sales Office. Keep the shipping materials for the carrier’s inspection. The
Agilent Sales Office will arrange for repair or replacement at Agilent’s option
without waiting for claim settlement.
Table 1-1. Standard Accessories
AccessoryPart Number
KeyboardE2610-68700
MouseC3751-60201
Quick Start Guide (this book)86100-90002
Programmer’s Guide86100-90003
Table 1-2. Product Options
OptionDescription
00112 GHz Trigger Bandwidth
AX4Rack mount flange kit
AXERack mount flange kit with handles
UK6Commercial calibration certificate with test data
1-4
Table 1-3. Optional Accessories
AccessoryDescription
83446ALightwave clock and data receiver
11982AHigh-speed lightwave receiver
83440B/C/DHigh-speed lightwave receiver
54118A500 MHz to 18 GHz trigger
N1020ATDR probe kit
54006A6 GHz hand-held probe
547012.5 GHz active probe
83480A-K17Switch matrix
83480A-K17RIMM fixture
10086AECL terminator
83430ALightwave digital source
11898APlug-in module remote/extender
54008B24 ns delay line
Setting Up the Infiniium DCA
Step 1. Inspect the shipment
Table 1-4. Additional Adapters for Electrical Channels
AdapterAgilent Part Number
SMA (f-f)1250-1158
APC 3.5 (f-f)1250-1749
1-5
Setting Up the Infiniium DCA
Step 1. Inspect the shipment
Table 1-5. Fiber-Optic Adapters for Optical Channels
Front Panel
Fiber-Optic
AdapterDescriptionAgilent Part Number
Diamond HMS-1081000AI
FC/PC81000FI
D481000GI
SC81000KI
DIN81000SI
1-6
ST81000VI
Biconic81000WI
Setting Up the Infiniium DCA
Step 2. Install any modules
Step 2. Install any modules
Up to two modules can be inserted into the Infiniium DCA. All documentation
(including specifications) for plug-in modules is located in the Infiniium DCA’s
built-in information system. To access the built-in information system, touch
or click Contents on the Help menu .
1
Slide the module into an available front-panel slot.
You can remove or install a module while the instrument is operating.
2
Finger tighten the two knurled screws on the module’s front panel.
1-7
Setting Up the Infiniium DCA
Step 3. Connect the keyboard and mouse
Step 3. Connect the keyboard and mouse
• Connect the standard PC-compatible mouse and keyboard.
Although you can operate all instrument functions using a keyboard and
mouse, the touch screen feature makes your Infiniium DCA even easier to use.
The touch screen feature is always available, even when you connect the
mouse. For more information, refer to Chapter 4, “Using the Built-In Informa-
tion System”, or use the built-in information system.
not
Do
stack other objects on the keyboard; this will cause self-test failures on
power-on.
1-8
Step 4. Connect the line cord
Setting Up the Infiniium DCA
Step 4. Connect the line cord
CAUTION
CAUTION
CAUTION
The Infiniium DCA automatically adjusts for line input voltages in the range of
100 to 240 VAC. There is no manual selection switch. The line cord provided is
matched by Agilent to the country in which the order originates. Refer to
Table 1-8, “Available Line Cords,” on page 1-11 for a list of available line power
cords.
Always use the three-prong AC power cord supplied with this instrument.
Failure to ensure adequate earth grounding by not using this cord may cause
instrument damage.
Do not
connect ac power until you have verified the line voltage is correct as
described in Table 1-6. Damage to the equipment could result.
This instrument has autoranging line voltage input. Be sure the supply voltage
is within the specified range.
1-9
Setting Up the Infiniium DCA
Step 4. Connect the line cord
Table 1-6. Line Power Requirements
Power115 VAC: 110 VA MAX. / 60 WATTS MAX. / 1.1 A MAX.
230 VAC: 150 VA MAX. / 70 WATTS MAX. / 0.6 A MAX.
Voltagenominal: 115 VAC / 230 VAC CAT II
range 115 VAC: 90–132 V
range 230 VAC: 198–254 V
Frequencynominals: 50 Hz / 60 Hz
range: 47–63 Hz
Table 1-7. Environmental Specifications
Useindoor
Temperature
Operating10 °C to +40 °C (50 °F to + 104 °F)
Non-operating-40 °C to +70 °C (-40 °F to +158 °F)
Humidity
OperatingUp to 90% humidity (non-condensing) at +40 °C (+104 °F)
Non-operatingUp to 95% relative humidity at + 65 °C (+149 °F)
Altitude
OperatingUp to 4,600 meters (15,000 ft)
Non-operatingUp to 15,300 meters (50,000 ft)
Vibration
OperatingRandom vibration 5-500 Hz, 10 minutes per axis, 2.41 g(rms)
Non-operatingRandom vibration 5-500 Hz, 10 minutes per axis, 0.3 g(rms);
Resonant search, 5-500Hz swept sine, 1 octave/minute sweep
rate, 0.75g, 5 minute resonant dwell at 4 resonances per axis.
Power requirements
Voltage90 to 132 or 198 to 264 Vac, 48-66 Hz CAT II
Power (including modules)545 VA; 375 W
Weight
Mainframe12.7 kg (28 lb)
Modules1.2 kg (2.6 lb)
Dimensions
Height215.1 mm (8.47 in)
Width425.5 mm (16.75 in)
Depth521 mm (20.5 in)
Depth with front connectors
and rear feet
584 mm (23.0 in)
1-10
Table 1-8. Available Line Cords
Setting Up the Infiniium DCA
Step 4. Connect the line cord
Plug TypeCable Part No.Plug Description
250V8120-1351
8120-1703
250V8120-1369
Straight *BS1363A
90°
Straight *NZSS198/ASC
90°
8120-0696
250V8120-1689
8120-1692
8120-2857p
125V8120-1378
8120-1521
8120-1992
250V8120-2104
8120-2296
Straight *CEE7-Y11
90°
Straight (Shielded)
Straight *NEMA5-15P
90°
Straight (Medical) UL544
Straight *SEV1011
1959-24507
Type 12 90°
Length
(in/cm)
90/228
90/228
79/200
87/221
79/200
79/200
79/200
90/228
90/228
96/244
79/200
79/200
ColorCountry
Gray
Mint Gray
United Kingdom,
Cyprus, Nigeria,
Zimbabwe,
Singapore
Gray
Australia, New
Zealand
Mint Gray
Mint Gray
Mint Gray
Coco Brown
East and West
Europe, Saudi
Arabia, So. Africa,
India (unpolarized in
many nations)
Jade Gray
Jade Gray
Black
United States,
Canada, Mexico,
Philippines,
Taiwan
Mint Gray
Switzerland
Mint Gray
220V8120-2956
8120-2957
250V8120-4211
8120-4600
100V8120-4753
8120-4754
* Part number sho wn for plug is the i ndustry identifier for the plug only. Number sho wn for cable is the Agilent
Technologies part number for the complete cable including the plug.
Straight *DHCK107
90°
Straight SABS164
90°
Straight MITI
90°
79/200
79/200
79/200
79/200
90/230
90/230
Mint Gray
Denmark
Mint Gray
Jade GrayRepublic of South
Africa
India
Dark GrayJapan
1-11
Setting Up the Infiniium DCA
Step 5. Turn on the line power
Step 5. Turn on the line power
• Press the power switch at the lower left-hand corner of the front panel.
After about one minute, the display will look similar to the following figure.
The Infiniium DCA is ready to use.
1-12
Setting Up the Infiniium DCA
Step 6. Set the time and date
Step 6. Set the time and date
1
Touch the current time and date shown near the top of the display. Notice how
the instrument responds to your touch by displaying the current time and date
settings. You can touch the screen with your finger, ballpoint pen, or a stylus.
You can also click the current time and date with a mouse. Although this book
uses the term “touch” to encourage you to experience using the touch screen,
you can always click using the mouse instead.
CAUTION
Avoid touching the screen with a sharp object as this could result in damage to
the display.
2
Touch the various displayed settings to set the correct time for your location.
After setting the correct time, touch or click Set Time & Date Now to change
the time.
1-13
Setting Up the Infiniium DCA
Step 7. Connect a printer
Step 7. Connect a printer
Use of a printer with the Infiniium DCA is optional. Table 1-9 lists the PCL-language printers that are currently supported for use with the instrument. You
must use the Add Printer Wizard to select the printer driver for the print er as
outlined in this procedure; the Add Print er Wizard does not run automatically.
For more information on connecting and using a printer, refer to the built-in
information system.
Unsupported printers
If your printer is not in the list of supported printers, you will need the printer driver disk
that came with your printer. To install the printer driver from this disk, click Have Disk on
the Add Printer Wizard dialog box. The instrument will then prompt you to put your
printer driver disk into the front-panel disk drive (drive A). Be sure to install the
Windows 98 printer driver for your printer. If a printer driver disk was not provided or you
cannot find the disk, you can contact the printer’s manufacturer to obtain the appropriate
Windows 98 printer driver.
Table 1-9. Supported Printers
HP DeskJet 850HP DeskJet 1600C
HP DeskJet 890CHP LaserJet 4P
HP DeskJet 970HP LaserJet 4000N
HP DeskJet 1200C
To connect a printer
1
Connect the printer to the instrument’s rear-panel
Parallel
port. Use the cable
that came with your printer.
2
On the File menu, select Print Setup and then Add Printer.
3
Select Printer that is Plug and Play.
An online Help window opens with instructions for connecting a printer. You
do not need to read the displayed instructions at this time.
4
Turn the Infiniium DCA’s power off. Do not exit the online Help before turning
the instrument off.
5
Turn the printer’s power on.
1-14
Setting Up the Infiniium DCA
Step 7. Connect a printer
.
6
Wait five seconds and then turn the Infiniium DCA’s power on again.
7
When the following message appears, select “No”.
8
When the following message appears select “OK”.
1-15
Setting Up the Infiniium DCA
Step 7. Connect a printer
9
The instrument
automatically recognizes
the printer connected on
the printer port and
displays the Add Printer
Wizard. Touch Next.
10
Select Local printer and
then touch Next.
11
Select the printer
manufacturer and then
printer. Then, touch
Next. If your printer is
not listed, click Have
Disk. The in strument will
then prompt you to put
your printer driver disk
into the front-panel disk
drive (drive A). Be sure
to install the Windows 98
printer driver for your
printer.
1-16
12
Select LPT1 for the
printer port. Touch Next.
13
Select Yes and then
touch Next.
Setting Up the Infiniium DCA
Step 7. Connect a printer
14
Select Yes to print a test
page. Then, touch Next.
1-17
Setting Up the Infiniium DCA
Step 7. Connect a printer
15
After confirming that the test page printed correctly, touch Yes.
16
Turn the Infiniium DCA’s power off and then on again. The printer is now
installed and ready to us e.
1-18
Setting Up the Infiniium DCA
Step 8. Configuring the touch screen
Step 8. Configuring the touch screen
The touch screen configuration utility ensures that the touch screen is both
aligned and orientated with the display. It also allows you to adjust the touch
screen speed to simulate a double-click and to turn the touch screen sound on
or off.
1
On the Utilities menu, select
Touch Screen Config. The touch
screen configuration utility is
displayed.
2
Touch Calibration.
3
Touch the Calibrate button and
follow the displayed instructions.
You do not need to select
Calibrate TouchSurround; this
feature is not used in configuring
the touch screen.
4
After returning to the
Configuration Utilities dialog box
shown in Step 1, touch
Properties.
1-19
Setting Up the Infiniium DCA
Step 8. Configuring the touch screen
5
The Display Area Configuration
dialog box shown here is
displayed. Although the Beep on
Touch and the Drag and Drop
settings are bes t left selected, you
can turn them off if desired. This
figure shows the default settings.
Pressing the front-panel Default
Setup button does not change
these settings.
The Double Touch Feature
Since the double touch feature is rarely used in the Infiniium DCA, you’ll probably never
need to adjust Double Touch Speed or Timed Double Touch. One example of using double
touch is when you open a file. You can double touch the file name instead of pressing
the Open button.
Timed Double Touch allows a single touch to be interpreted as a double touch if you continue the touch for a set amount of time. If this setting is set to None (the default setting), two touches are required. You then set the Double Touch Speed for the time