Hewlett-Packard Company
Lightwave Operations
1400 Fountaingrove Parkway
Santa Rosa, CA 95403-1799,
USA
(707) 577-1400
Notice.
The information contained in
this document is subject to
change without notice. Companies, names, and data used
in examples herein are fictitious unless otherwise noted.
Hewlett-Packard makes no
warranty of any kind with
regard to this material, including but not limited to, the
implied warranties of merchantability and fitness for a
particular purpose. HewlettPackard shall not be liable for
errors contained herein or for
incidental or consequential
damages in connection with
the furnishing, performance,
or use of this material.
Restricted Rights Legend.
Use, duplication, or disclosure by the U.S. Government
is subject to restrictions as set
forth in subparagraph (c) (1)
(ii) of the Rights in Technical
Data and Computer Software
clause at DFARS 252.227-7013
for DOD agencies, and subparagraphs (c) (1) and (c) (2)
of the Commercial Computer
Software Restricted Rights
clause at FAR 52.227-19 for
other agencies.
Warrant y.
This Hewlett-Packard instrument product is warranted
against defects in material and
workmanship for a period of
one year from date of shipment. During the warranty
period, Hewlett-Packard Company will, at its option, either
repair or replace products
which prove to be defective.
For warranty service or repair,
this product must be returned
to a service facility designated by Hewlett-Packard.
Buyer shall prepay shipping
charges to Hewlett-Packard
and Hewlett-Packard shall pay
shipping charges to return the
product to Buyer. However,
Buyer shall pay all shipping
charges, duties, and taxes for
products returned to HewlettPackard from another country.
Hewlett-Packard warrants
that its software and firmware
designated by Hewlett-Packard for use with an instrument
will execute its programming
instructions when properly
installed on that instrument.
Hewlett-Packard does not
warrant that the operation of
the instrument, or software,
or firmware will be uninterrupted or error-free.
Limitation of Warranty.
The foregoing warranty shall
not apply to defects resulting
from improper or inadequate
maintenance by Buyer, Buyersupplied software or interfacing, unauthorized modification or misuse, operation
outside of the environmental
specifications for the product,
or improper site preparation
or maintenance.
No other warranty is
expressed or implied.
Hewlett-Packard specifically
disclaims the implied warranties of merchantability and fitness for a particular purpose.
Exclusive Remedies.
The remedies provided herein
are buyer's sole and exclusive
remedies. Hewlett-Packard
shall not be liable for any
direct, indirect, special, incidental, or consequential damages, whether based on
contract, tort, or any other
legal theory.
Safety Symbols.
CAUTION
The
hazard. It calls attention to a
procedure which, if not correctly performed or adhered
to, could result in damage to
or destruction of the product.
Do not proceed beyond a caution sign until the indicated
conditions are fully understood and met.
WAR N IN G
The
hazard. It calls attention to a
procedure which, if not correctly performed or adhered
to, could result in injury or
loss of life. Do not proceed
beyond a warning sign until
the indicated conditions are
fully understood and met.
sign denotes a
caution
warning
sign denotes a
The instruction manual symbol. The product is marked with this
warning symbol when
it is necessary for the
user to refer to the
instructions in the
manual.
The laser radiation
symbol. This warning
symbol is marked on
products which have a
laser output.
The AC symbol is used
to indicate the
required nature of the
line module input
power.
The ON symbols are
|
used to mark the positions of the instrument
power line switch.
The OFF symbols
❍
are used to mark the
positions of the instrument power line
switch.
The CE mark is a registered trademark of
the European Community.
The CSA mark is a registered trademark of
the Canadian Standards Association.
The C-Tick mark is a
registered trademark
of the Australian Spectrum Management
Agency.
This text denotes the
ISM1-A
instrument is an
Industrial Scientific
and Medical Group 1
Class A product.
ii
General Safety Considerations
General Safety Considerations
This product has been designed and tested in accordance with IEC Publication 1010, Safety Requirements for Electronic Measuring Apparatus, and has
been supplied in a safe condition. The instruction documentation contains
information and warnings which must be followed by the user to ensure safe
operation and to maintain the product in a safe condition.
WARNING
WARNING
WARNING
WARNING
WARNING
This is a Safety Class I product (provided with a protective earthing
ground incorporated in the power cord). The mains plug shall only be
inserted in a socket outlet provided with a protective earth contact.
Any interruption of the protective conductor inside or outside of the
instrument is likely to make the instrument dangerous. Intentional
interruption is prohibited.
No operator serviceable parts inside. Refer servicing to qualified
personnel. To prevent electrical shock, do not remove covers.
If this product is not used as specified, the protection provided by the
equipment could be impaired. This product must be used in a normal
condition (in which all means for protection are intact) only.
For continued protection against fire hazard, replace line fuse only
with same type and rating (type 2A/250V). The use of other fuses or
materials is prohibited.
On Option 002 dual input instruments, any light on an
unselected “A” channel will likely be output on one of the
unselected “B” channels.
On Option 002 dual input instruments, any light on an
unselected “B” channel will likely be output on an unselected
“A” channel.
To avoid exposure to light energy, always cover all unused channels.
iii
General Safety Considerations
CAUTION
CAUTION
CAUTION
CAUTION
Always use the three-prong ac power cord supplied with this product. Failure
to ensure adequate earth grounding by not using this cord may cause product
damage.
This instrument has autoranging line voltage input. Be sure the supply voltage
is within the specified range.
CAUTION, VENTILATION REQUIREMENTS: When installing the product in a
cabinet, the convection into and out of the product must not be restricted. The
ambient temperature (outside the cabinet) must be less than the maximum
operating temperature of the product by 4°C for every 100 watts dissipated in
the cabinet. If the total power dissipated in the cabinet is greater than 800
watts, then forced convection must be used.
This product is designed for use in INSTALLATION CATEGORY II and
POLLUTION DEGREE 2, per IEC 1010 and 664 respectively.
iv
Contents
1 General Information
Channels, Options, and Accessories 1-3
Specifications and Regulatory Information 1-7
Regulatory Information 1-11
Care of Fiber-Optic Connectors 1-13
Inspecting connectors 1-13
Cleaning optical connectors 1-17
Measuring insertion loss and return loss 1-19
Returning the Instrument for Service 1-20
Preparing the instrument for shipping 1-21
Hewlett-Packard Sales and Service Offices 1-23
2 Installing
Step 1. Inspect the shipment 2-3
Step 2. Check the fuse 2-4
Step 3. Connect the line-power cable 2-5
Step 4. Turn on the lightwave switch 2-7
Step 5. Performing an operational check 2-8
If The Operational Check Fails 2-10
3 Using the Switch
Front-panel features 3-3
Rear-panel features 3-5
Changing Switch Position 3-6
To set single port A switches 3-7
To set dual port A switches 3-8
Adjusting Display Contrast 3-8
Saving Switch States 3-8
To save a state 3-9
To recall a state 3-10
4 Programming
General Information 4-3
Setting the switches 4-4
Returning the switch to manual control 4-5
Response generation 4-6
Programming over HP-IB 4-7
Contents-1
Contents
Programming over RS-232 4-9
Common Commands 4-12
Standard SCPI Commands 4-23
Instrument Specific Commands 4-27
Error Messages 4-31
Programming Examples 4-32
Example 1: Switch position using the *WAI command 4-33
Example 2: Switch position using the Status Byte Register 4-35
Example 3: Switch position using the *OPC command 4-37
Example 4: Input/output multiplexers 4-39
5Servicing
Spare Channel Replacement Procedure 5-4
Electrostatic Discharge Information 5-7
Contents-2
1
Channels, Options, and Accessories1-3
Specifications and Regulatory Information1-7
Care of Fiber-Optic Connectors1-13
Returning the Instrument for Service1-20
Hewlett-Packard Sales and Service Offices1-23
General Information
General Information
General Information
General Information
The HP 86060C-series lightwave switches cover a broad range of switching
capacity and provide for accurate and repeatable measurements. Configuring
the switch is easy because the signal routing is shown graphically on the display. You can easily integrate the switch into an automated test system using
SCPI-compatible commands via HP-IB or RS-232 interfaces. The HP 86060Cseries lightwave switches are temperature stabilized.
• The HP 86060C is a compact switch with 4 to 8 output channels and 1 or 2 inputs.
• The HP 86061C is a mid-size switch in a half-width chassis, with 1 or 2 input
channels. It can accommodate from 4 to 12 output channels on the front panel,
and up to 18 outputs on the rear panel.
WARNING
CAUTION
CAUTION
• The HP 86062C is a full-width switch with 20 to 100 output channels.
On Option 002 dual input instruments, any light on an
unselected “A” channel will likely be output on one of the
unselected “B” channels.
On Option 002 dual input instruments, any light on an
unselected “B” channel will likely be output on an unselected
“A” channel.
To avoid exposure to light energy, always cover all unused channels.
Improper connector care, cleaning, or use of mismatched cable connectors can
invalidate the published specifications and damage connectors. Clean all cables
before applying to any connector. Repair of damaged connectors due to
improper use is not covered under warranty. Refer to “Care of Fiber-Optic
Connectors” on page 1-13 for proper cleaning procedures.
OPTION 3XX INSTRUMENTS: To avoid damage, handle the pigtail fiber with
care. Use only an appropriate fiber cleaver tool for cutting the fiber. Do not pull
the bare fiber out of its jacket, crush it, kink it, or bend it past its minimum bend
radius.
1-2
General Information
Channels, Options, and Accessories
Channels, Options, and Accessories
Fiber-optic cables
The lightwave switch use one of three types of fiber-optic cables. To determine which fiber-optic cable type your lightwave switch uses, refer to the rearpanel serial number label. This label indicates the installed options which are
defined in the following list:
The lightwave switches are based on a moving fiber technology where an input
fiber is aligned with any one of “N” fixed output fibers. The input fiber is positioned by means of a precision stepper motor. Lightwave switches with two
input fibers allow the user to position either input A1 or A2 to a specific output channel. The non-selected input
put channel.
Lightwave switches with three or more “B” channels have an additional position called channel O or OFF.
may
may not
or
align with another out-
Special ordered instruments
Normal lightwave switches have only one layer installed. (A switch layer is a
switch matrix of “A” ports and “B” ports.) However, special ordered instruments may have multiple switch layers installed. If the rear panel shows more
than one set of “A” ports and “B” ports, the instrument has multiple switch
layers. Other switch configurations, such as non-blocking matrices are available as special orders.
Serial numbers
Hewlett-Packard makes frequent improvements to its products to enhance
their performance, usability, or reliability, and to control costs. HP service personnel have access to complete records of design changes to each type of
equipment, based on the equipment’s serial number. Whenever you contact
1-3
General Information
Channels, Options, and Accessories
Hewlett-Packard about your lightwave switch, have the complete serial number available to ensure obtaining the most complete and accurate information
possible.
A serial-number label is attached to the rear of the lightwave switch. It contains the serial number and the options installed in the lightwave switch.
Whenever you specify the serial number or refer to it in obtaining information
about your lightwave switch, be sure to use the complete number, including
the full prefix and suffix.
Table 1-1. Output Channels
HP 86060C Compact
Lightwave Switch
Number of Output Channels
04
06
08
HP 86061C Mid-Size
Lightwave Switch
04
08
12
16
HP 86062C Full-Size
Lightwave Switch
20
24
28
32
40
48
56
64
72
80
00 (100 output channels)
1-4
General Information
Channels, Options, and Accessories
Table 1-2. Options
Option Description
Number of Input Channels (select one):
Option 001 Single input channel
Option 002 Two input channels
Wavelength and Fiber Type (select one):
Option 109
Option 163
Option H51
Port Type (select one):
Option 050 Connectors on front panel. (Only available on an HP 86060C or
Option 051 Connectors on rear panel. (For connectorized outputs only.)
Option 052 3 meter fiber out of the rear panel. (For connectorized outputs, the
Option 2XX Where XX is the number of connectorized output channels.
(Note: Option 200 is 100 connectorized output channels.)
Option 3XXWhere XX is the number of non-connectorized output channels.
(Note: Option 300 is 100 non-connectorized output channels.)
Connector Type (for connectorized ports or fibers only):
Option 012 FC/PC connectors
Option 014 ST connectors
Option 017 SC connectors
FC/APC or SC/APC connectors (special order)
1-5
General Information
Channels, Options, and Accessories
Table 1-2. Options
Option Description
Optional Accessories
Option ABJUser’s manual in Japanese
Option UK6Commercial calibration certificate with test data
Option 1CMRack-mount flange kit
Option 1CNFront handle kit
Option 1CPRack mount flange kit with handles
(2 of 2)
Table 1-3. Accessories
HP Part Number Description
5062-3957 Rack mount adapter kit for a single half-width instrument.
5062-3977 Rack mount adapter kit for two adjacent half-width instruments.
5062-4079 Lock link kit for the HP 5062-3977.
5952-4079 Fiber Optics Handbook, an introduction to, and a reference for, fiber-optic
measurements.
1-6
General Information
Specifications and Regulatory Information
Specifications and Regulatory Information
This section lists specifications and regulatory information of the HP 86060Cseries lightwave switches. Specifications apply over the temperature range
+0°C to +55°C (unless otherwise noted). All specifications apply after the
instrument’s temperature has been stabilized after 120 minutes of continuous
operation.
Table 1-4 on page 1-8 lists specification,
The distinction between these terms is described as follows:
characteristics
, and nominal values.
Specifications
CharacteristicsCharacteristics
Nominal values
Calibration cycle
CAUTION
Specifications describe warranted performance.
ranted, performance parameters.
Nominal value indicates the expected, but not warranted, value of the parameter.
HP warrants instrument specifications over the recommended calibration
interval. To maintain specifications, periodic recalibrations are necessary. We
recommend that the HP 86060C-series switches be calibrated at an HP service
facility every 24 months.
Improper connector care, cleaning, or use of mismatched cable connectors can
invalidate the published specifications and damage connectors. Clean all cables
before applying to any connector. Repair of damaged connectors due to
improper use is not covered under warranty. Refer to “Care of Fiber-Optic
Connectors” on page 1-13 for proper cleaning procedures.
provide useful information by giving functional, but nonwar-
Characteristics are printed in italics.
1-7
General Information
Specifications and Regulatory Information
Table 1-4. Optical Interface Specifications and Characteristics
Insertion Loss
Single-mode switches
Multi-mode switches
Insertion Loss Stability
Repeatability
Sequential switching
Random switching
Optical Return Loss
Single-mode
Multimode
Polarization Dependent Loss
Isolation
Typical Switching Life
a
1.0 dB
(0.7 dB)
0.8 dB
(0.6 dB)
b
c
d
e
±0.03 dB
±0.008 dB
±0.025 dB
58 dB
(62 dB)
20 dB
(25 dB)
0.05 dB
(0.02 dB)
80 dB
–
(–100 dB)
(±0.025)
(±0.005)
(±0.01)
10 million cycles, minimum
Switching Time
Between adjacent channels
Each additional channel
a. Insertion loss does not include connectors. Include an additional 0.5 dB (
connector.
b. Drift of any channel relative to one assigned reference channel at ±3°C deviation of ambient temperature
over 7 day period.
c. Repeatability measured after four (4) hours warm-up and with an eight (8) second pause between
movements.
d. Excludes external pigtail backscatter and connector reflections.
e. Polarization dependent loss only applies to single-mode switches and is measured at 1550 nm.
330 msec
50 msec
0.25 dB characteristic
) for each
1-8
Table 1-5. Switching Time Sample (msec)
General Information
Specifications and Regulatory Information
Switch Sizea
Switch
1×4HP 86060C,
Starting Channel to
Adjacent Channels
Plus Additional
Time/Channel
Maximum
Switching Time
29040370
b
HP 86061C
1×8HP 86060C,
29040530
HP 86061C
1×56HP 86062C2587.5663
1×100HP 86062C2587.5993
a. Note that the switch mechanism used for channel count greater than 48 is different, hence switching time.
b. Switching time = (switching between starting and adjacent channel) + (additional time/channel) × remaining channel increments to reach
last channel.
Table 1-6. General Specifications (1 of 2)
OPTICAL CONNECTORS
Option 012FC/PC connectors
Option 014ST connectors
Option 017SC connectors
GENERAL SPECIFICATIONS
Temperature Range
Operating
Storage
Humidity
Operating
Storage
a,b,c
+0°C to +55°C
–40°C to +70°C
Maximum relative humidity 95% for temperatures up to 40°C (non-condensing)
Maximum relative humidity less than 90% at 65
C
°
AltitudeAltitude up to 15,000 feet (4,572 meters).
EMI CompatibilityConducted and radiated emissions meet the requirements of CISPR Publication 11 and
EN 55011 Group 1, Class A.
Power Requirements100/115/230/240 V (range 90 to 254 Vac),
50/60 Hz (range 47 to 63 Hz)
Power ConsumptionUp to 80 VA
Installation CategoryCategory II per I.E.C. 1010
Pollution DegreeDegree 2 per I.E.C. 664
1-9
General Information
Specifications and Regulatory Information
Table 1-6. General Specifications (2 of 2)
UsageFor indoor use.
Enclosure ProtectionIP 2 0, according to IEC 529
Weight
(dependent on # of channels)
HP 86060C
HP 86061C
HP 86062C
Dimensions (H × W × D)
d
HP 86060C
HP 86061C
HP 86062C
a. All HP 86060C-series lightwave switches must specify one of the following options, except when specifying Option 3xx.
b. Unlike most HP lightwave instruments, connector types are not interchangeable.
c. Other connector types are available upon request.
d. Feet add 12.5 mm to the height of the instrument.
This instrument is in conformance with the German Regulation on Noise Declaration for Machines (Laermangabe nach der Maschinenlaermrerordnung
–3.GSGV Deutschland).
Notice for Germany: Noise Declaration
Acoustic Noise EmissionGeraeuschemission
LpA < 70 dBLpA < 70 dB
Operator positionam Arbeitsplatz
Normal positionnormaler Betrieb
per ISO 7779nach DIN 45635 t.19
1-11
General Information
Specifications and Regulatory Information
1-12
General Information
Care of Fiber-Optic Connectors
Care of Fiber-Optic Connectors
Advances in measurement capabilities make connectors and connection techniques more important than ever. Damage to the connectors on calibration
and verification devices, test ports, cables, and other devices can degrade
measurement accuracy and damage instruments. Replacing a damaged connector can be very expensive, not to mention the lost time. This expense can
be avoided by observing the simple precautions presented in this section.
CAUTION
The connectors on the switch are
to the the lightwave switch it is
procedures in “Cleaning optical connectors” on page 1-17.
not
easily cleaned. Before connecting cables
very
important they are cleaned following the
Inspecting connectors
Because fiber-optic connectors are susceptible to damage that is not immediately obvious to the naked eye, bad measurements can be made without the
user even being aware of a connector problem. Although microscopic examination and return loss measurements are the best way to ensure good connections, they are not always practical. An awareness of potential problems, along
with good cleaning practices, can ensure that optimum connector performance is maintained. With glass-to-glass interfaces, it is clear that any degradation of a ferrule or fiber endface, any stray particles, or finger oil on the
endface, can have a significant effect on connector performance. Many times
an instrument must be serviced to replace a damaged connector when thousands of dollars and lost time could have been avoided if better care were
given to the fiber-optic connector.
Figure 1-1 is a close-up micrograph of a clean cable endface. In contrast, Fig-
ure 1-2 shows a connector endface that was either not cleaned, or not prop-
erly cleaned. Material is smeared and ground into the endface causing light
scattering and poor reflection. Not only is the precision polish lost, but this
action can grind off the glass face and destroy the connector.
1-13
General Information
Care of Fiber-Optic Connectors
Figure 1-1. A clean and problem-free connector
Figure 1-2. A dirty endface from poor cleaning
Repeated connections made without removing loose particles, or using
improper cleaning tools can lead to physical damage of the glass fiber endface,
as shown in Figure 1-2. When severe, the damage on one connector end can
be transferred to another good connector that comes in contact with it.
1-14
General Information
Care of Fiber-Optic Connectors
Figure 1-3. A damaged fiber end from using an improper cleaning procedure
The cure for these problems is disciplined connector care. as described in the
following list and in “Cleaning optical connectors” on page 1-17.
Guidelines
Use the following guidelines to achieve the best possible performance when
making measurements on a fiber-optic system:
• Keep connectors covered when not in use.
• Use dry connections whenever possible.
• Never use metal or sharp objects to clean the connector and never scrape the
connector.
• Avoid matching gel and oils.
While these often work well on first insertion, they are great dirt magnets. The
oil or gel grabs and holds grit that is then ground into the fiber endface. Also,
some early gels were designed for use with the FC, non-contacting connectors, using small glass spheres. When used with contacting connectors, these
glass balls can scratch and pit the fiber. If an index matching gel or oil must be
used, apply it to a freshly cleaned connector, make the measurement, and
then immediately clean it off. Never use a gel for longer-term connections and
never use it to improve a damaged connector. The gel can mask the extent of
damage and continued use of a damaged fiber can transfer damage to the
instrument.
• When inserting a fiber-optic cable into a connector or adapter, gently insert it
in as straight a line as possible. Make sure the fiber end does not touch the outside of the mating connector or adapter. Tipping and inserting at an angle can
1-15
General Information
Care of Fiber-Optic Connectors
scrape material off the inside of the connector or even break the inside sleeve
of connectors made with ceramic material.
• Avoid over tightening connections.
Unlike common electrical connections, tighter is
not
better. The purpose of
the connector is to bring the endfaces of two fibers together. Once they touch,
tightening only causes a greater force to be applied to the delicate endfaces.
With some connectors, the end can set itself off-axis with a tight connection,
due to the curved face, resulting in misalignment and excessive return loss.
Many measurements are actually improved by backing off the connector pressure. Also, if a piece of grit does happen to get by the cleaning procedure, the
tighter connection is more likely to damage the glass. Tighten the connectors
just until the two fibers touch.
• Keep connections covered when not in use.
• Use fusion splices on the more permanent critical nodes. Choose the best connector possible. Replace connecting cables regularly. Frequently measure the
return loss of the connector to check for degradation, and clean
every
tor,
time.
every
connec-
All connectors should be treated like the high-quality lens of a good camera.
The weak link in instrument and system reliability is often the inappropriate
use and care of the connector. Because current connectors are so easy to use,
there tends to be reduced vigilance in connector care and cleaning. It takes
only one missed cleaning for a piece of grit to permanently damage the glass
and ruin the connector.
Visual inspection of fiber ends
Visual inspection of fiber ends can be helpful. Contamination or imperfections
on the cable end can be detected, as well as cracks or chips in the fiber itself.
Use a microscope with 100× to 200× magnification to inspect the entire end
face for contamination, raised metal, or dents in the metal as well as any other
imperfections. Inspect the fiber for cracks and chips. Visible imperfections not
touching the fiber core may not affect performance, unless the imperfections
keep the fibers from contacting.
1-16
Cleaning optical connectors
General Information
Care of Fiber-Optic Connectors
CAUTION
CAUTION
The connectors on the switch are
to the the lightwave switch it is
not
easily cleaned. Before connecting cables
very
important they are cleaned following the
procedures in this section.
The procedures in this section provide the proper steps for cleaning fiberoptic cables. The initial cleaning, using alcohol as a solvent, gently removes
any grit and oil. If a caked-on layer of material is still present, this can happen
if the beryllium-copper sides of the ferrule retainer get scraped and deposited
on the end of the fiber during insertion of the cable, a second cleaning should
be performed. It is no uncommon for a cable or connector to require more
than one cleaning.
Hewlett-Packard strongly recommends that index matching compounds
not
be
applied to their instruments and accessories. Some compounds, such as gels,
may be difficult to remove and can contain damaging particulates. If you think
the use of such compounds is necessary, refer to the compound manufacturer
for information on application and cleaning procedures.
Table 1-7. Cleaning Accessories
ItemHP Part Number
Isopropyl alcohol8500-5344
Cotton swabs8520-0023
Small foam swabs9300-1223
Compressed dust remover (non-residue)8500-5262
1-17
General Information
Care of Fiber-Optic Connectors
Table 1-8. Dust Caps Provided with Lightwave Instruments
ItemHP Part Number
Laser shutter cap08145-64521
FC/PC dust cap08154-44102
Biconic dust cap08154-44105
DIN dust cap5040-9364
HMS10/HP dust cap5040-9361
ST dust cap5040-9366
Cleaning a non-lensed connector
CAUTION
CAUTION
Do not use any type of foam swab to clean optical fiber ends. Foam swabs can
leave filmy deposits on fiber ends that can degrade performance.
1
Apply pure isopropyl alcohol to a clean lint-free cotton swab or lens paper.
2
Clean the ferrules and other parts of the connector while avoiding the end of
the fiber.
3
Apply isopropyl alcohol to a new clean lint-free cotton swab or lens paper.
4
Clean the fiber end with the swab or lens paper.
not
Do
scrub during this initial cleaning because grit can be caught in the
swab and become a gouging element.
5
Immediately dry the fiber end with a clean, dry, lint-free cotton swab or lens
paper.
6
Blow across the connector end face from a distance of 6 to 8 inches using
filtered, dry, compressed air. Aim the compressed air at a shallow angle to the
fiber end face.
Nitrogen gas or compressed dust remover can also be used.
Do not shake, tip, or invert compressed air canisters. This releases particles
from the can into the air. Refer to the instructions on the compressed air
canister.
7
As soon as the connector is dry, connect or cover it for later use.
1-18
General Information
Care of Fiber-Optic Connectors
If the performance, after the initial cleaning, seems poor try cleaning the connector again. Often a second cleaning will restore proper performance. The
second cleaning should be more arduous with a scrubbing action.
Measuring insertion loss and return loss
Consistent measurements with your lightwave equipment are a good indication that you have good connections. Since return loss and insertion loss are
key factors in determining optical performance they can be used to determine
connector degradation A smooth, polished fiber end should produce a good
return loss measurement. The quality of the polish establishes the difference
between the “PC” (physical contact) and the “Super PC” connectors. Most
connectors today are physical contact which make glass-to-glass connections,
therefore it is critical that the area around the glass core be clean and free of
scratches. Although the major area of a connector, excluding the glass, may
show scratches and wear, of the glass has maintained its polished smoothness,
the connector can still provide a good low level return loss connection.
If you test your cables and accessories for insertion loss and return loss upon
receipt, and retain the measured data for comparison, in the future you will be
able to tell if any degradation has occurred. Typical values are less than 0.5 dB
of loss, and sometimes as little as 0.1 dB of loss with high performance connectors. Return loss is a measure of reflection: the less reflections the better.
The larger the return loss, the smaller the reflection. The most physically contacting connectors have return losses better than 50 dB, although 30 to 40 dB
is more common.
1-19
General Information
Returning the Instrument for Service
Returning the Instrument for Service
The instructions in this section show you how to properly return the instrument for repair or calibration. Always call the HP Instrument Support Center
first to initiate service
This ensures that the repair (or calibration) can be properly tracked and that
your instrument will be returned to you as quickly as possible. Call this number regardless of where you are located. Refer to “Hewlett-Packard Sales and
Service Offices” on page 1-23 for a list of service offices.
If the instrument is still under warranty or is covered by an HP maintenance
contract, it will be repaired under the terms of the warranty or contract (the
warranty is at the front of this manual). If the instrument is no longer under
warranty or is not covered by an HP maintenance plan, Hewlett-Packard will
notify you of the cost of the repair after examining the unit.
When an instrument is returned to a Hewlett-Packard service office for servicing, it must be adequately packaged and have a complete description of the
failure symptoms attached. When describing the failure, please be as specific
as possible about the nature of the problem. Include copies of additional failure information (such as the instrument failure settings, data related to instrument failure, and error messages) along with the instrument being returned.
before
returning your instrument to a service office.
1-20
General Information
Returning the Instrument for Service
Preparing the instrument for shipping
1
Write a complete description of the failure and attach it to the instrument.
Include any specific performance details related to the problem. The following
information should be returned with the instrument.
• Type of service required.
• Date instrument was returned for repair.
• Description of the problem:
• Whether problem is constant or intermittent.
• Whether instrument is temperature-sensitive.
• Whether instrument is vibration-sensitive.
• Instrument settings required to reproduce the problem.
• Performance data.
• Company name and return address.
• Name and phone number of technical contact person.
• Model number of returned instrument.
• Full serial number of returned instrument.
• List of any accessories returned with instrument.
CAUTION
CAUTION
2
Cover all front or rear-panel connectors that were originally covered when you
first received the instrument.
Cover electrical connectors to protect sensitive components from electrostatic
damage. Cover optical connectors to protect them from damage due to physical
contact or dust.
Use original packaging or equivalent. Instrument damage can result from using
packaging materials other than the original materials. Never use styrene pellets
as packaging material. They do not adequately cushion the instrument or
prevent it from shifting in the carton. They may also cause instrument damage
by generating static electricity.
3
Pack the instrument in the original shipping containers. Original materials are
available through any Hewlett-Packard office. Or, use the following guidelines:
• Wrap the instrument in antistatic plastic to reduce the possibility of damage
caused by electrostatic discharge.
• For instruments weighing less than 54 kg (120 lb), use a double-walled, cor-
rugated cardboard carton of 159 kg (350 lb) test strength.
• The carton must be large enough to allow approximately 7 cm (3 inches) on
1-21
General Information
Returning the Instrument for Service
all sides of the instrument for packing material, and strong enough to accommodate the weight of the instrument.
• Surround the equipment with approximately 7 cm (3 inches)
1
of packing
material, to protect the instrument and prevent it from moving in the carton.
If packing foam is not available, the best alternative is S.D-240 Air Cap™
from Sealed Air Corporation (Commerce, California 90001). Air Cap looks
like a plastic sheet filled with air bubbles. Use the pink (antistatic) Air Cap™
to reduce static electricity. Wrapping the instrument several times in this
material will protect the instrument and prevent it from moving in the carton.
Step 1. Inspect the shipment2-3
Step 2. Check the fuse2-4
Step 3. Connect the line-power cable2-5
Step 4. Turn on the lightwave switch2-7
Step 5. Performing an operational check 2-8
If The Operational Check Fails2-10
Installing
Installing
Installing
Installing
WARNING
CAUTION
Before installing the lightwave switch, refer to “General Safety
Considerations” on page -iii of this manual.
OPTION 3XX INSTRUMENTS: To avoid damage, handle the pigtail fiber with
care. Use only an appropriate fiber cleaver tool for cutting the fiber. Do not pull
the bare fiber out of its jacket, crush it, kink it, or bend it past its minimum bend
radius.
Install the instrument so that the front panel ON/OFF switch is readily identifiable and is easily reached by the operator. The ON/OFF switch, or the
detachable power cord, is the instrument disconnecting device. It disconnects
the mains circuit from the mains supply after the EMC filters and before other
parts of the instrument. Alternatively, an external installed switch or circuit
breaker (which is easily identifiable and is easily reached by the operator)
may be used as a disconnecting device.
Install the instrument according to the enclosure protection provided. This
instrument does
protect against finger access to hazardous parts within the enclosure.
not
protect against the ingress of water. This instrument does
2-2
Step 1. Inspect the shipment
Step 1. Inspect the shipment
1
Verify that all components ordered have arrived by comparing the shipping
forms to the original purchase order. Inspect all shipping containers.
If your shipment is damaged or incomplete, save the packing materials and
notify both the shipping carrier and the nearest Hewlett-Packard sales and
service office. HP will arrange for repair or replacement of damaged or
incomplete shipments without waiting for a settlement from the
transportation company. Notify the HP customer engineer of any problems.
2
Make sure that the serial number and options listed on the instrument’s rearpanel label match the serial number and options listed on the shipping
document.
Installing
2-3
Installing
Step 2. Check the fuse
Step 2. Check the fuse
1
Locate the line-input connector on the instrument’s rear panel.
2
Disconnect the line-power cable if it is connected.
3
Use a small flat-blade screwdriver to open the pull-out line fuse drawer.
WARNING
For continued protection against fire hazard, replace line fuse only
with same type and ratings (2A/250V). The use of other fuses or
materials is prohibited.
Figure 2-1. Changing the fuse
4
Verify that the value of the line fuse in the pull-out drawer is correct.
115V operation, 5×20 mm, 2A, 250 V, fast acting UL/CSA fuse: . . 2110-0702
230V operation, 5×20 mm, 2A, 250 V, fast acting IEC fuse:. . . . . . 2110-0702
2-4
Step 3. Connect the line-power cable
Step 3. Connect the line-power cable
1
Verify that the line power meets the requirements shown in the following table.
Table 2-1. HP 86060C-Series Power Requirements
CharacteristicRequirement
Input Voltagewithin range 90 to 254 Vac
Frequencywithin range 47 to 63 Hz
Power80 VA (maximum)
Installing
2
Connect the line-power cord to the instrument’s rear-panel connector.
3
Connect the other end of the line-power cord to the power receptacle.
The lightwave switch is equipped with a three-wire power cable, in accordance with international safety standards. When connected to an appropriate
power line outlet, this cable grounds the instrument cabinet.
Various power cables are available to connect the lightwave switch to the
types of ac power outlets unique to specific geographic areas. The cable
appropriate for the area to which the lightwave switch is originally shipped is
included with the unit. You can order additional ac power cables for use in different areas. Table 2-2 on page 2-6 lists the available ac power cables, illustrates the plug configurations, and identifies the geographic area in which
each cable is appropriate.
2-5
Table 2-2. Power Cables
Installing
Step 3. Connect the line-power cable
Plug Type
Cable Part
No.
250V8120-1351
8120-1703
250V8120-1369
8120-0696
250V8120-1689
8120-1692
8120-2857p
125V8120-1378
8120-1521
8120-1992
250V8120-2104
8120-2296
220V8120-2956
8120-2957
Plug Description
Straight *BS1363A
90°
Straight *NZSS198/
ASC
90°
Straight *CEE7-Y11
90°
Straight (Shielded)
Straight *NEMA5-15P
90°
Straight (Medical)
UL544
Straight *SEV1011
1959-24507
Type 12 90°
Straight *DHCK107
90°
Length
(in/cm)
90/228
90/228
79/200
87/221
79/200
79/200
79/200
90/228
90/228
96/244
79/200
79/200
79/200
79/200
ColorCountry
Gray
Mint Gray
United Kingdom,
Cyprus, Nigeria,
Zimbabwe,
Singapore
Gray
Mint Gray
Mint Gray
Mint Gray
Coco Brown
Australia, New
Zealand
East and West
Europe, Saudi
Arabia, So. Africa,
India (unpolarized in
many nations)
Jade Gray
Jade Gray
United States,
Canada, Mexico,
Philippines, Taiwan
Black
Mint Gray
Switzerland
Mint Gray
Mint Gray
Denmark
Mint Gray
250V8120-4211
8120-4600
100V8120-4753
8120-4754
* Part number shown for plug is the industry identifier for the plug only. Number shown for
cable is the HP part number for the complete cable including the plug.
2-6
Straight SABS164
90°
Straight MITI
90°
79/200
79/200
90/230
90/230
Jade GrayRepublic of South
Africa
India
Dark GrayJapan
Step 4. Turn on the lightwave switch
Step 4. Turn on the lightwave switch
1
Turn the lightwave switch on by pressing the line switch. The liquid-crystal
display (LCD) displays the message:
Installing
Initializing
Screen Saver
A screen-saver has been built in to the switch to prolong the lifetime of the backlit LCD.
The screen-saver turns off the LCD backlighting after 10 minutes elapses without a
front-panel key being pressed. The time interval is not adjustable. To resume operation,
press any key.
When the switch is turned on, it automatically resets to channel 0 (reset, optical off position).
2-7
Installing
Step 5. Performing an operational check
Step 5. Performing an operational check
CAUTION
Return loss
Improper connector care, cleaning, or use of mismatched cable connectors can
invalidate the published specifications and damage connectors. Clean all cables
before applying to any connector. Repair of damaged connectors due to
improper use is not covered under warranty. Refer to “Care of Fiber-Optic
Connectors” on page 1-13 for proper cleaning procedures.
Return loss can be tested using a number of different test equipment configurations. Some of these are:
• an HP 8703A lightwave component analyzer
• an HP 8702B lightwave component analyzer with the appropriate source, receiver and lightwave coupler
• an HP 8504B precision reflectometer
• an HP 8153A lightwave multimeter and HP 81534A return loss module
Many other possibilities exist. The basic requirements are an appropriate
lightwave source, a compatible lightwave receiver, and a compatible lightwave
coupler.
Refer to the manuals provided with your lightwave test equipment for information on how to perform a return loss test.
Typical return loss is better than 40 dB. For actual specifications on your particular cable or accessory, refer to the manufacturer.
Insertion loss
Insertion loss can be tested using a number of different test equipment configurations. Some of these are:
• an HP 8702B or HP 8703A lightwave component analyzer system with a lightwave source and receivers
• an HP 83420 lightwave test set with an HP 8510 network analyzer
• an HP 8153A lightwave multimeter with a source and a power sensor module
2-8
Installing
Step 5. Performing an operational check
Many other possibilities exist. The basic requirements are an appropriate
lightwave source and a compatible lightwave receiver. Refer to the manuals
provided with your lightwave test equipment for information on how to perform an insertion loss test.
Typical insertion loss for cables is less than 1 dB, and can be as little as 0.1 dB.
For actual specifications on your particular cable or accessory, refer to the
manufacturer.
2-9
Installing
If The Operational Check Fails
If The Operational Check Fails
If the HP 86060C does not pass the operational check, you should review the
procedure being performed when the problem occurred. A few minutes spent
performing some simple checks may save waiting for your instrument to be
repaired. Before calling Hewlett-Packard or returning the unit for service,
please make the following checks:
1
Is the line fuse good?
2
Does the line socket have power?
3
Is the unit plugged in to the proper ac power source?
4
Is the unit turned on?
5
If other equipment, cables, and connectors are being used with the lightwave
switch, are they connected properly and operating correctly?
6
Review the procedure for the test being performed when the problem
appeared. Are all the settings correct?
7
Are the connectors clean? Refer to “Care of Fiber-Optic Connectors” on page
1-13 for more information about cleaning the connectors.
Refer to “Spare Channel Replacement Procedure” on page 5-4 for more information.
If the HP 86060C lightwave switch still fails, return it to Hewlett-Packard for
repair; if the lightwave switch is still under warranty or is covered by an
HP maintenance contract, it will be repaired under the terms of the warranty
or contract (the warranty is at the front of this manual). If the lightwave
switch is no longer under warranty or is not covered by an HP maintenance
plan, Hewlett-Packard will notify you of the cost of the repair after examining
the unit. Refer to “Returning the Instrument for Service” on page 1-20 for
more information.
2-10
3
Front-panel features3-3
Rear-panel features 3-5
Changing Switch Position 3-6
Adjusting Display Contrast 3-8
Saving Switch States3-8
Using the Switch
Using the Switch
Using the Switch
Using the Switch
This chapter describes the front and rear-panel features. It also provides stepby-step procedures for configuring the lightwave switch. Position the lightwave switch according to the enclosure protection provide. This instrument
does not protect against the ingress of water. This instrument protects against
finger access to hazardous parts within the enclosure.
WARNING
WARNING
To prevent electrical shock, disconnect the HP 86060C-series switch
from the mains before cleaning. Use a dry cloth or one slightly
dampened with water to clean the external case parts. Do not attempt
to clean internally.
On Option 002 dual input instruments, any light on an
unselected “A” channel will likely be output on one of the
unselected “B” channels.
On Option 002 dual input instruments, any light on an
unselected “B” channel will likely be output on an unselected
“A” channel.
To avoid exposure to light energy, always cover all unused channels.
3-2
Front-panel features
Figure 3-1. The HP 86060C-series front-panel functional area
Screen Saver
Using the Switch
Front-panel features
LINE
Switch
Display
key
HELP
key
A screen-saver has been built in to the switch to prolong the lifetime of the backlit LCD.
The screen-saver turns off the LCD backlighting after 10 minutes elapses without a
front-panel key being pressed. The time interval is not adjustable. To resume operation,
press any key.
Turns the lightwave switch on and off. The front-panel
the mains circuits from the mains supply after the EMC filters and before
other parts of the instrument.
Graphically shows current signal path of the switc
the RMT, LSN, TLK, and SRQ lines.
Use this key to adjust the 0contrast of the display.
Press to use built-in Help. Then press any of the front-panel keys. A short
explanation of that key’s function will be displayed.
h and the current HP-IB status of
LINE
switch disconnects
3-3
LOCAL
key
Using the Switch
Front-panel features
Press this key to display the HP-IB address of the lightwave switch. You can
also change the address using the numeric keypad. If a computer has placed
the instrument in remote control, is this key to reenable front-panel control.
SAVE
&
RECALL
SWITCH PORT
keys
key
Use these keys to save and recall switch configurations. Ten internal memory
registers, selected using the numeric keypad, are available.
Repeatedly pressing this key activates either “A” or “B” channels. Once activated, use the arrow keys to select the active switch port.
3-4
Rear-panel features
Using the Switch
Rear-panel features
Optical
connector(s)
HP-IB connector
RS-232 connector
Figure 3-2. The HP 86060C-series rear-panel functional area
The number of optical connectors depends on the HP 86060-series switch.
The connectors are grouped as Port A and Port B.
Provides for remote control of the lightwave switch via the HP-IB interface
bus. Refer to “Programming over HP-IB” on page 4-7.
Provides for remote control of the lightwave switch via RS-232. Refer to “Programming over RS-232” on page 4-9.
3-5
Using the Switch
Changing Switch Position
Changing Switch Position
3-6
Using the Switch
To set single port A switches
To set single port A switches
The 1 × N switch has a single Port A channel and multiple Port B channels.
1
To select a Port B channel, press:
The Port B channels are shown in inverse video and the prompt,
active
2
Use the arrow keys to change the Port B channel.
You can also use the numeric keys to enter the desired Port B channel. For
example: 4 followed by
The new connection is displayed on the front-panel display.
, appears at the bottom of the display.
ENTER
SWITCH PORT
Port B
.
3-7
Using the Switch
To set dual port A switches
To set dual port A switches
The 2 × N switch has two Port A channels and multiple Port B channels.
1
To select the Port A channel, press:
The Port A channels are shown in inverse video and the prompt,
active
2
To select a Port B channel, again press:
The Port B channels are shown in inverse video and the prompt,
active
3
Use the arrow keys to change the channel.
You can also use the numeric keys to enter the desired channel. For example:
4
followed by
The new connection is displayed on the front-panel display.
, appears at the bottom of the display.
, appears at the bottom of the display.
ENTER
.
SWITCH PORT
Port A
SWITCH PORT
Port B
Adjusting Display Contrast
To adjust the contrast of the display, press the key. Use the arrow keys to
select the desired contrast, then press
ENTER
.
Saving Switch States
3-8
Using the Switch
To save a state
To save a state
To save the currently displayed switch state in one of the ten internal storage
registers, press
SAVE
, and then press one of the numeric keys (0–9).
3-9
Using the Switch
To recall a state
To recall a state
To recall a previously saved switch state from one of the ten internal storage
registers, press
RECALL
, and then press one of the numeric keys (0–9).
3-10
4
General Information4-3
Programming over HP-IB 4-7
Programming over RS-232 4-9
Common Commands 4-12
Standard SCPI Commands4-23
Instrument Specific Commands4-27
Error Messages 4-31
Programming Examples4-32
Programming
Programming
Programming
Programming
The programming instructions in this manual conform to the IEEE 488.2 Standard Digital Interface for Programmable Instrumentation and to the Standard
Commands for Programmable Instruments (SCPI). The programming instructions provide the means of remote control.
Where to begin . . .
• To program the HP 86060-series lightwave switch, it is necessary to add either
an HP-IB or RS-232 interface to the rear panel of the switch.
• The programming examples for individual commands in this manual are written in HP BASIC 6.0 for an HP 9000 Series 200/300 Controller.
• For more information regarding the HP-IB, the IEEE 488.2 standard, or the
SCPI standard, refer to the following books:
Hewlett-Packard Company.
Interface Bus,
Hewlett-Packard Company.
Programmable Instruments,
International Institute of Electrical and Electronics Engineers.
Standard 488.1-1987, IEEE Standard Digital Interface for
Programmable Instrumentation.
International Institute of Electrical and Electronics Engineers.
Standard 488.2-1987, IEEE Standard Codes, Formats, Protocols and
Common commands For Use with ANSI/IEEE Std 488.1-1987.
NY, 1987.
4-2
1987.
Tutorial Description of Hewlett-Packard
SCPI—Standard Commands for
1991.
IEEE
New York, NY, 1987.
IEEE
New York,
General Information
This instrument has three types of commands:
• Common commands
• Standard SCPI commands
• Instrument specific commands
Programming
General Information
Common
commands
Standard SCPI
commands
Instrument
specific commands
The common commands are the commands defined by IEEE 488.2. These
commands control some functions that are common to all IEEE 488.2 instruments. Common command headers consist of only a single mnemonic preceded by an asterisk.
Example:
The standard SCPI commands are the STATUS subsystem commands required
for compatibility with SCPI. In most instruments, the STATUS subsystem commands are used to report device-dependent errors. In the lightwave switch,
these commands have no function but are included for SCPI compatibility.
Standard SCPI command headers are compound headers consisting of two or
more mnemonics.
Example:
Instrument specific commands are those commands which are specific to the
control of the switch. These commands control switch movements and report
the configuration of the switch. Instrument specific commands are compound
headers consisting of two or more mnemonics.
Example:
*RST
:STATUS:OPERATION:ENABLE
:ROUTE:LAYER:CHANNEL
4-3
Programming
Setting the switches
Setting the switches
Use the [:ROUTe]:[LAYer]:CHANnel command to set the channel connections
for a particular switch layer. A layer is a particular switch matrix that creates a
path from an “A” port to a “B” port. When the lightwave switch has multiple
switch layers, the switch layers are referred to by the word LAYER in the
ROUTE:LAYER:CHANNEL command. The numeric value at the end of the
mnemonic LAYER selects the switch block to which the
ROUTE:LAYER:CHANNEL command should be applied. For example, in the
following program statement, the command is applied to switch layer 2 of the
instrument.
OUTPUT 711;":ROUTE:LAYER2:CHANNEL A2,B4"
The following HP BASIC statement command moves the switch on switch
layer 1 to channel 1 of port A and channel 1 of port B:
OUTPUT 711;":ROUTE:LAYER1:CHANNEL A1,B1"
This next example sets port A to channel 2 and port B to channel 5:
OUTPUT 711;":ROUTE:LAYER1:CHANNEL A2,B5”
To learn more about this command, refer to “[:ROUTe]:[LAYer]:CHANnel” on
page 4-28.
The current switch setting can be queried. The query
:ROUTE:LAYER1:CHANNEL? places the current channel setting on layer 1 in
the output queue. In HP BASIC, the controller input statement:
ENTER <device address>;Setting$
passes the value across the bus to the controller and places it in the variable
Setting$.
4-4
Returning the switch to manual control
Returning the switch to manual control
Programming
To return the switch to manual control after remote operation, press
LOCAL
.
4-5
Programming
Response generation
Response generation
As defined by IEEE 488.2, query responses may be buffered for the following
conditions:
• When the query is parsed by the instrument.
• When the controller addresses the instrument to talk so that it may read the
response.
The responses to a query are buffered when the query is parsed.
Command headers immediately followed by a question mark (?) are queries.
Query commands are used to find out information regarding the instrument’s
current state. After receiving a query, the instrument interrogates the
requested function and places the answer in its output queue. The answer
remains in the output queue until it is read or another command is issued.
When read, the answer is transmitted across the bus to the designated listener
(typically a controller).
The output queue must be read before the next program message is sent. For
example, when you send the query :SYSTEM:CONFIG? you must follow that
query with an input statement. In HP BASIC, this is usually done with an
ENTER statement immediately followed by a variable name. This statement
reads the result of the query and places the result in a specified variable.
4-6
Programming
Programming over HP-IB
Programming over HP-IB
This section describes the HP-IB interface functions and some general concepts. In general, these functions are defined by IEEE 488.2. They deal with
general interface management issues, as well as messages which can be sent
as interface commands.
Default address
Command and
data concepts
Addressing
The HP-IB address is factory preset to 711. To change the HP-IB address,
LOCAL
press
enter a different address, press the two numeric keys for that address. For
example: 14 for address 714, then press
The interface has two modes of operation:
• command mode
•data mode
The bus is in the data mode when the ATN line is false. The data mode is used
to convey device-dependent messages across the bus.
The address is used to determine which instrument on the interface bus with
which the controller is communicating.
• Each device on the HP-IB resides at a particular address, 0–30.
• The active controller specifies which devices talk and which listen.
• An instrument may be talk addressed, listen addressed, or unaddressed by the
controller.
If the controller addresses the instrument to talk, the instrument remains configured to talk until it receives an interface clear message (IFC), another instrument’s talk address (OTA), its own listen address (MLA), or a universal untalk
command (UNT).
. The last two digits of the current HP-IB address are displayed. To
ENTER
.
If the controller addresses the instrument to listen, the instrument remains
configured to listen until it receives an interface clear message (IFC), its own
talk address (MTA), or a universal unlisten command (UNL).
4-7
Programming
Programming over HP-IB
Interface select
code (selects
interface)
Instrument
address (selects
instrument)
Lockout
Each interface card has a unique interface select code. This code is used by
the controller to direct commands and communications to the proper interface. The default is typically "7" for HP-IB controllers.
Each instrument on an HP-IB must have a unique instrument address between
decimal 0 and 30. The device address passed with the program message must
include not only the correct instrument address, but also the correct interface
select code.
For example, if the instrument address for the instrument is 4 and the interface
select code is 7, when the program message is passed, the routine performs its
function on the instrument at device address 704.
For this instrument, the address is typically set to "11" at the factory. This
address can be changed by pressing the
LOCAL
key on the front panel.
NOTE
The examples in this manual assume the instrument is at device address 711.
With HP-IB, the instrument is placed in the lockout mode by sending the local
lockout command (LLO). The instrument can be returned to local by sending
the go-to-local command (GTL) to the instrument.
Bus commands
NOTE
Cycling the power also restores front panel control.
The following commands are IEEE 488.1 bus commands (ATN true). IEEE
488.2 defines many of the actions which are taken when these commands are
received by the instrument.
The device clear (DCL) or selected device clear (SDC) commands clear the
input and output buffers, reset the parser, and clear any pending commands.
The interface clear (IFC) command halts all bus activity. This includes unaddressing all listeners and the talker, disabling serial poll on all devices, and
returning control to the system controller.
4-8
Programming
Programming over RS-232
Programming over RS-232
This section describes the interface functions and some general concepts of
the RS-232 interface. The RS-232 interface on this instrument is HewlettPackard’s implementation of EIA Recommended Standard RS-232, "Interface
Between Data Terminal Equipment and Data Communications Equipment
Employing Serial Binary Data Interchange." With this interface, data is sent
one bit at a time and characters are not synchronized with preceding or subsequent data characters. Each character is sent as a complete entity without
relationship to other events.
NOTE
IEEE 488.2 is designed to work with IEEE 488.1 as the physical interface. When RS-232
is used as the physical interface, as much of IEEE 488.2 is retained as the hardware differences will allow. No IEEE 488.1 messages such as DCL, GET, and END are available.
Interface
operation
Cables
The switch can be programmed with a controller over RS-232 using an interface cable that is appropriate for your application. The operation and exact
connections for this interface are described in more detail in the following sections. When you are using a controller to program a switch over RS-232, you
are normally operating directly between two DTE (Data Terminal Equipment)
devices as compared to operating between a DTE device and a DCE (Data
Communications Device) device.
The type of RS-232 cable you use to connect the controller to the switch will
depend on your application. The following paragraphs describe which lines of
the switch are used to control the operation of the RS-232 bus relative to the
switch. To locate the proper cable for your application, refer to the reference
manual for your controller.
4-9
Programming
Programming over RS-232
3-wire interface
Interface settings
The switch uses a 3-wire RS-232 interface. It provides a simple connection
between devices because you can ignore hardware handshake requirements.
The switch uses the following connections on its RS-232 interface for 3-wire
communication:
Switch Computer
SGND (Signal Ground) Pin 5Pin 5
TD (Transmit Data from switch) Pin 2Pin 2
RD (Receive Data into switch) Pin 3Pin 3
The TD (Transmit Data) line from the switch must connect to the RD
(Receive Data) line on the controller. Likewise, the RD line from the switch
must connect to the TD line on the controller. The RS-232 interface on the
switch ignores all signals on the DCD, DSR, RTS, and CTS lines.
NOTE
If the instrument is used with QBASIC programming language, pin 6, 7, and 8 of the DTE
connector should be shorted together.
The baud rate, stop bits, parity, protocol, and data bits must be configured
exactly the same for both the controller and the switch to properly communicate over the RS-232 interface. The RS-232 interface capabilities of the lightwave switch are listed below:
The baud rate is factory set to 9600 baud. To change the baud rate, press the
LOCAL
key twice. The current baud rate is displayed. Use the arrow keys to
change the baud rate to 1200 or 9600. When the desired rate is displayed,
ENTER
press
.
Data bits are the number of bits sent and received per character that represent the binary code of that character.
4-10
Programming
Programming over RS-232
Information is stored in bytes (8 bits at a time) in the switch. Data can be sent
and received just as it is stored, without the need to convert the data.
Communicating
over the RS-232
interface
RS-232 commands
Each RS-232 interface card has its own interface select code. This code is
used by the controller to direct commands and communications to the proper
interface. Unlike HP-IB, which allows multiple devices to be connected
through a single interface card, RS-232 is only connected between two devices
at a time through the same interface card. Because of this, only the interface
code is required for the device address.
Generally, the interface select code can be any decimal value between 0 and
31, except for those interface codes which are reserved by the controller for
internal peripherals and other internal interfaces. This value can be selected
through switches on the interface card. For more information, refer to the reference manual for your interface card or controller.
RS-232 control of the switch is initiated by sending the OPEN RS232 COM
command over the interface. This places the switch in the remote mode and
locks out the front panel. Pressing the
LOCAL
key will bring the instrument
back to local mode.
Many of the commands used for controlling the switch and for retrieving data
from the switch are the same as for the HP-IB interface. Refer to the individual
command descriptions to determine if the command applies to both HP-IB
and RS-232.
To end communications with the switch over the RS-232 interface, send the
CLOSE RS232 COM command. This command returns control to the front
panel of the instrument.
4-11
Programming
Common Commands
Common Commands
The following commands are required by the IEEE 488.2–1987 standard.
*CLS (Clear Status)
The *CLS (clear status) common command clears all the event registers summarized in the Status Byte register. With the exception of the output queue,
all queues that are summarized in the Status Byte Register are emptied. The
error queue is also emptied. Neither the Standard Event Status Enable Register, nor the Service Request Enable Register are affected by this command.
After the *CLS command, the instrument is left in the idle state. The command does not alter the instrument setting. *OPC/*OPC? actions are cancelled.
Usage:
Command Syntax: *CLS
Example: OUTPUT 711;"*CLS"
HP-IB only
*ESE (Event Status Enable)
The *ESE command sets the bits in the Standard Event Status Enable Register and enables the corresponding bits in the Standard Event Status Register.
The Standard Event Status Enable Register contains a mask value for the bits
to be enabled in the Standard Event Status Register. A bit set to one in the
Standard Event Status Enable Register enables the corresponding bit in the
Standard Event Status Register. A zero disables the bit. Refer to Table 4-1 for
information about the Standard Event Status Enable Register bits, bit weights,
and what each bit masks.
4-12
Programming
Common Commands
The Standard Event Status Enable Register is cleared at power-on. The *RST
and *CLS commands do not change the register.
The *ESE query returns the value of the Standard Event Status Enable Register.
Usage:
Command Syntax:
Where:
HP-IB only
*ESE <mask>
<mask> ::= 0 to 255
Example: OUTPUT 711;"*ESE 64"
In this example, the *ESE 64 command enables URQ (user request) bit 6 of
the Standard Event Status Enable Register. Therefore, when a front-panel key
is pressed, the ESB (event summary bit) in the Status Byte Register is also
set.
Query Syntax:
Returned Format:
Where:
*ESE?
<mask><NL>
<mask> ::= 0 to 255 (integer–NR1 format)
Table 4-1. Standard Event Status Enable Register
(High–Enables the ERS bit)
BitBit WeightEnables
7128PON – Power On
664URQ – User Request
532CME – Command Error
416EXE – Execution Error
38NOT USED
24QYE – Query Error
4-13
Programming
Common Commands
Table 4-1. Standard Event Status Enable Register
(High–Enables the ERS bit)
BitBit WeightEnables
12NOT USED
01OPC – Operation Complete
*ESR (Event Status Register)
The *ESR query returns the value of the Standard Event Status Register.
When you read the Event Status Register, the value returned is the total of the
bit weights of all of the bits that are set to one at the time you read the byte.
Table 4-2 shows each bit in the Event Status Register and its bit weight.
Reading the register clears the Event Status Register.
Usage:
Query Syntax:
Returned Format:
Where:
HP-IB only
*ESR?
<status><NL>
<status> ::= 0 to 255 (integer–NR1 format)
Example: OUTPUT 711;"*ESR?"
ENTER 711;Event
PRINT Event
Table 4-2. Standard Event Status Register
BitBit WeightCondition
7128PON – Power On
664URQ – User Request
532CME – Command Error
4-14
Programming
Common Commands
Table 4-2. Standard Event Status Register
BitBit WeightCondition
416EXE – Execution Error
38NOT USED
24QYE – Query Error
12NOT USED
01OPC – Operation Complete
*IDN (Identification Number)
The *IDN query returns a string value which identifies the instrument type
and firmware version.
An *IDN query must be the last query in a program message. Any queries after
the *IDN query in a program message are ignored.
Usage:
Query Syntax:
Returned Format:
Where:
HP-IB and RS-232
*IDN?
"HEWLETT-PACKARD 8606XC, 0, VERSION <X.X>"
<X.X> = firmware revision number
8606XC is the model number and can be 86060C, 86061C, or 86062C.
Example: DIM Id$[50]
OUTPUT 711;"*IDN?"
ENTER 711;Id$
PRINT Id$
4-15
Programming
Common Commands
*OPC (Operation Complete)
The *OPC command sets the operation complete bit in the Standard Event
Status Register when all pending device operations have finished.
The *OPC query places an ASCII "1" in the output queue when all pending
device operations have finished.
NOTE
The *OPC command can be used to ensure all switch movement operations have completed before continuing the program. By following a ROUTE:LAYER:CHANNEL command
with an *OPC query and an ENTER statement, the program will pause until the response
(ASCII "1") is returned by the instrument.
Usage:
Command Syntax:
Example: OUTPUT 711;"*OPC"
Query Syntax:
Returned Format:
Example: OUTPUT 711;"*OPC?"
HP-IB only
*OPC
*OPC?
1<NL>
ENTER 711;Op$
*RCL (Recall)
The *RCL command recalls the state of the instrument from the specified
instrument state register. If the instrument state register has not been previously stored, the *RCL command will restore the instrument to its power-on
state.
Usage:
HP-IB and RS-232
4-16
Programming
Common Commands
Command Syntax:
Where:
Example: OUTPUT 711;"*RCL 3"
*RCL <value>
<value> ::= 0 to 9 (integer–NR1 format)
*RST (Reset)
The *RST command returns the switch to its power-up condition. For all layers, each port is set to its OFF position or channel 1.
Usage:
Command Syntax:
Example: OUTPUT 711;"*RST"
HP-IB and RS-232
*RST
*SAV (Save)
The *SAV command saves the current state of the instrument to the specified
instrument state register.
Usage:
Command Syntax:
Where:
Example: OUTPUT 711;"*SAV 3"
HP-IB and RS-232
*SAV <value>
<value> ::= 0 to 9 (integer–NR1 format)
4-17
Programming
Common Commands
*SRE (Service Request Enable)
The *SRE command sets the bits in the Service Request Enable Register. The
Service Request Enable Register contains a mask value for the bits to be
enabled in the Status Byte Register. A bit set to one (1) in the Service Request
Enable Register enables the corresponding bit in the Status Byte Register. A
zero (0) disables the bit. Table 4-3 lists the bits in the Service Request Enable
Register and what they mask.
The Service Request Enable Register is cleared at power-on. The *RST and
*CLS commands do not change the register.
The *SRE query returns the value of the Service Request Enable Register.
Usage:
Command Syntax:
Where:
Example: OUTPUT 711;"*SRE 32"
Query Syntax:
Returned Format:
Where:
HP-IB only
*SRE <mask>
<mask> ::= 0 to 255
In this example, the *SRE 32 command enables ESB (event summary) bit 5 of
the Status Byte Register, the MSS (master summary status) bit 6 in the Status
Byte Register is also set.
*SRE?
<mask><NL>
<value> ::= 0 to 255 (integer–NR1 format)
4-18
Table 4-3. Service Request Enable Register
Service Request Enable Register
(High–Enables the SRE bit)
BitBit WeightEnables
7128Not Used
664MSS – Master Summary Status
532ESB – Event Status Bit
416MAV – Message Available
38Not Used
24Not Used
12Not Used
01OPP – Operation Pending
Programming
Common Commands
4-19
Programming
Common Commands
*STB (Status Byte)
The *STB query returns the current value of the instrument’s status byte. The
MSS (Master Summary Status) bit 6 indicates whether or not the device has at
least one reason for requesting service.
When you read the Status Byte Register, the value returned is the total of the
bit weights of all the bits set to one (1) at the time you read the byte. Table 4-4
shows each bit in the Status Byte Register and its bit weight.
The *STB query does not affect the contents of the Status Byte Register.
NOTE
To read the instrument’s status byte with RQS reported on bit 6, use the interface Serial
Poll.
NOTE
The *STB query can be used to determine when the switch has settled to a new position. After sending a :ROUTE:LAYER:CHANNEL command, bit 0 of the Status Byte Register will be set to one while the switch is moving and return to zero when the switch has
settled.
Usage:
Query Syntax:
Returned Format:
Where:
Example: OUTPUT 711;"*STB?"
HP-IB only
*STB?
<value><NL>
<value> ::= 0 to 255 (integer – NR1 format)
ENTER 711;Value
PRINT Value
4-20
Table 4-4. Status Byte Register
BitBit WeightCondition
7128Not Used
664MSS – Master Summary Status
532ESB – Event Status Bit
416MAV – Message Available
38Not Used
24Not Used
12Not Used
01OPP – Operation Pending
Programming
Common Commands
*TST (Test)
The *TST query performs a self-test on the instrument. The result of the test
is placed in the output queue. A zero indicates the test passed and a non-zero
value indicates the test failed. If a test fails, refer to “Step 5. Performing an
operational check” on page 2-8.
Usage:
Query Syntax:
Returned Format:
Where:
Example: OUTPUT 711;"*TST?"
HP-IB and RS-232
*TST?
<result><NL>
<result> ::= 0 or non-zero value
0 indicates the test passed.
non-zero indicates the test failed.
ENTER 711;Result
PRINT Result
4-21
Programming
Common Commands
*WAI (Wait)
The *WAI command prevents the instrument from executing any further commands until the current command has finished executing. All pending operations are completed during the wait period.
NOTE
The *WAI command can be used to ensure all switch movement operations have completed before continuing the program. Following a ROUTE:LAYER:CHANNEL command
with a *WAI command followed by a query, will ensure the query is not answered until
the switch has settled to its new position.
Usage:
Command Syntax:
Example: OUTPUT 711;":ROUTE:LAYER1:CHANNEL A2,B4"
HP-IB only
*WAI
OUTPUT 711;"*WAI"
OUTPUT 711;"SYSTEM:CONFIG?"
ENTER 711;DUMMY$
4-22
Programming
Standard SCPI Commands
Standard SCPI Commands
:STATus:<node>:CONDition
The :STATus:<node>:CONDition query returns the value for the condition register for the node. Condition registers have no function in this instrument, but
the query is included for compatability with the SCPI standard. This query
always returns the value 0.
The :STATus:<node>:ENABle command sets the enable register for the node.
Enable registers have no function in this instrument, but the command is
included for compatability with the SCPI standard.
The :STATus:<node>:ENABle query returns the value of the enable register
for the node.
The :STATus:<node>[:EVENT] query returns the value of the event register
for the node. Event registers have no function in this instrument, but the
query is included for compatability with the SCPI standard. This query always
returns the value 0.
The :STATus:PRESet command presets the enable registers for all status
nodes. Enable registers have no function in this instrument, but the command
is included for compatability with the SCPI standard. Table 4-5 shows the
value of each enable register.
Usage:
Query Syntax:
Example: OUTPUT 711;":STATUS:PRESET"
HP-IB only
:S TATus:P RESe t
Table 4-5. Values of the Enable Registers
Status NodePreset Value
Operation 0
Questionable 0
:SYSTem:ERRor
The :SYSTem:ERRor query returns the next error number and error description in the error queue over the interface. This instrument has an error queue
100 errors deep and operates on a first-in, first-out basis. Repeatedly sending
the query :SYSTEM:ERROR? returns the error numbers and descriptions in
the order in which they occur until the queue is empty. Any further queries
returns "+0,No errors" until another error occurs. Refer to Table 4-6 for the
error numbers and descriptions.
Usage:
Query Syntax:
Returned Format:
HP-IB and RS-232
:SYSTem:ERRor?
<value>, <string>
4-25
Programming
Standard SCPI Commands
Where:
<value> = an integer error code (NR1)
<string> = text of error message
Example: DIM Error$[50]
OUTPUT 711;":SYSTEM:ERROR?"
ENTER 711;Error$
PRINT Error$
4-26
Programming
Instrument Specific Commands
Instrument Specific Commands
The following commands are specific to remote operation of the HP 86060Cseries lightwave switches.
CLOSE RS232 COM
The CLOSE RS232 COM command disables remote operation of the instrument over the RS-232 interface and enables the front-panel keyboard. This
command is the same as pressing the
the RS-232 interface.
LOCAL
key while in remote operation over
Usage:
Command Syntax:
Example: com_port=9
RS-232 only
CLOSE RS232 COM
OUTPUT com_port; "CLOSE RS232 COM"
OPEN RS232 COM
The OPEN RS232 COM command enables remote operation of the instrument
over the RS-232 interface and locks out the front-panel keyboard. This command must be sent before sending any other commands over the RS-232
interface. Press the
interface.
Usage:
Command Syntax:
RS-232 only
OPEN RS232 COM
LOCAL
key to return to local mode and lock out the RS-232
4-27
Programming
Instrument Specific Commands
Example: com_port=9
OUTPUT com_port; "OPEN RS232 COM"
[:ROUTe]:[LAYer]:CHANnel
The [:ROUTe]:[LAYer]:CHANnel command configures the channel connections. In the command syntax, the keyword “
switch layer is a switch matrix of “A” ports and “B” ports. Normal lightwave
switches have only one layer installed. Special ordered instruments may have
multiple switch layers installed. If the front and rear panels show more than
one set of “A” ports and “B” ports, the instrument has multiple switch layers.
Standard lightwave switches do not require that the <layer> argument be
specified as shown in the command syntax. If no layer is specified, it defaults
to layer 1. The minimum layer number is always 1.
The minimum channel number is either 0 or 1, depending on whether the port
has an "OFF" position. The maximum number of layers and channels is dependent on the switch configuration. If the command parameters are outside the
permitted range for the switch configuration, the switch position is not
changed and an error is generated.
The [:ROUTe]:[LAYer]:CHANnel query returns the current port settings for the
specified layer. If no layer is specified, the default value is layer 1.
LAYer
” refers to a switch layer. A
WARNING
Usage:
Command Syntax:
Where:
On Option 002 dual input instruments, any light on an
unselected “A” channel will likely be output on one of the
unselected “B” channels.
On Option 002 dual input instruments, any light on an
unselected “B” channel will likely be output on an unselected
“A” channel.
To avoid exposure to light energy, always cover all unused channels.
HP-IB and RS-232
[:ROUTe]:[LAYer<layer>]:CHANnel <channel_list>
<layer> a positive integer (NR1). Set to 1 if the instrument does not have
4-28
Instrument Specific Commands
multiple layers.
<channel list> = A<channel>|B<channel>|A<channel>,B<channel>
<channel> = a non-negative integer (NR1) | OFF
A<channel>,B<channel>
<channel> = a non-negative integer (NR1)
OUTPUT 711;":ROUTE:LAYER3:CHANNEL?"
ENTER 711;Setting$
PRINT Setting$
:SYSTem:CONFig
The :SYSTem:CONFig query returns the switch configuration of the instrument. For each layer, the minimum and maximum channel numbers for each
port are given.
<i> = number of layers on switch
<j1> = minimum available channel on port A, layer 1
<k1> = maximum available channel on port A, layer 1
<l1>= minimum available channel on port B, layer 1
<m1> = maximum available channel on port B, layer 1
<j2> = minimum available channel on port A, layer 2
<k2> = maximum available channel on port A, layer 2
<l2> = minimum available channel on port B, layer 2
<m2> = maximum available channel on port B, layer 2
Example: DIM Config$
OUTPUT 711;":SYSTem:CONFIG?"
ENTER 711;Config$
PRINT Config$
4-30
Error Messages
Table 4-6. Error Messages
Error NumberDescription
–105GET not allowed
–110Command Header error
–120Numeric Data error
–140Character Data error
–150String Data error
–170Expression error
–220Parameter error
–350Too many errors
–410Query INTERRUPTED
–420Query UNTERMINATED
–430Query DEADLOCKED
+300Frame error
+310Invalid command byte
+320Invalid switch module byte
+330Invalid data byte
+340Packet checksum error
+4xxSwitch module xx not responding
+5xxSwitch module xx motor failed
Programming
Error Messages
4-31
Programming
Programming Examples
Programming Examples
This section includes a number of programming examples to illustrate the use
of remote commands in actual programs. These programming examples do not
cover the full command set for the instrument. They are intended only as an
introduction to the method of programming the instrument
The example programs in this chapter are as follows:
Example 1: This simple program uses the ROUTE:LAYER:CHANNEL
command to move the switch to a new position. The
program shows how to use the *WAI command to ensure
that the switch has settled to its new position.
Example 2: This program is similar to the first example program.
Instead of using the *WAI command, the Status Byte
Register is read repeatedly using the *STB query. When
bit 0 of the Status Byte Register returns to zero, the switch
has settled to its new position.
.
Example 3: Repeating the same program as the first two examples, the
*OPC command and *ESR query are now used to
determine that the switch has settled to its new position.
The *OPC command is sent before the
:ROUTE:LAYER:CHANNEL command, and then the
Standard Event Status Register is continuously read until
bit 0 is set to one.
Example 4: This example illustrates the use of two switches in an
automated system to periodically monitor a number of
devices under test (DUTs). The test system includes an
HP 8153A optical multimeter with an HP 81554SM laser
source and an HP 81532A optical power sensor. This
program measures the optical power through each device
under test every 5 minutes and displays a message if the
power drops below 1 microwatt.
4-32
Programming
Example 1: Switch position using the *WAI command
Example 1: Switch position using the *WAI command
This program prompts the operator for the desired switch position and then
moves the switch to this position. The switch error queue is then read and
printed. The program shows how to use the *WAI command to ensure that the
switch has settled to its new position.
(VAL$(B_position))
80 Command$="ROUTE:LAYER1:CHANNEL "&Channel$
90 OUTPUT Switch_addr;Command$
100GOSUB Wait_to_settle
110 REPEAT
120 OUTPUT Switch_addr;"SYSTEM:ERROR?"
130 ENTER Switch_addr;Error_return$
140 PRINT Error_return$
150 UNTIL (VAL(Error_return$)=0)
170 GOTO Exit_prog
180 !
190 Wait_to_settle: ! wait for switch to settle
200 OUTPUT Switch_addr;"*WAI"
210 OUTPUT Switch_addr;"SYSTEM CONFIG?"
220 ENTER Switch_addr;Dummy$
230 RETURN
240 !
250 Exit_prog:!
260 END
Line No.
1
10 to 20 Declare some variables for use in the program.
30 Set the HP 8606X Optical Switch address variable,
Switch_addr
40 to 60 Clear the screen and prompt the operator for the desired
switch position. Store the positions in variables
A_position and B_position
70 Set
Channel$
, to 711 (factory default).
.
to represent the switch channel positions in
4-33
Programming
Example 1: Switch position using the *WAI command
the form appropriate for the
HP-IB command. For example, if
B_position=3
80 Set
Command$
, then
to represent the full HP-IB command to set
the desired switch position, appending
example
Channel$
"ROUTE:LAYER1:CHANNEL A1,B3"
90 Send
Command$
to the HP 8606X Optical Switch via the
given above,
ROUTE:LAYER:CHANNEL
A_position=1
Channel$
would equal
Channel$
Command$
.
and
"A1,B3"
. For the
would equal
HP-IB interface.
100 Call the subroutine
110 to 150 These lines implement a
Wait_to_settle
REPEAT-UNTIL
.
loop that
continuously queries the HP 8606X for error status. The
returned error message(s) are printed to the screen. This
loop exits when the numeric value of the error string equals
0. This will occur when the error message
error"
is returned.
"+0, no
170 Go to the end of the program.
190 to 230 The
wait_for_settle
subroutine.
200 Output the *WAI command to the HP 8606X Optical
Switch. This command will prevent the switch from
executing any further commands until the previous
command (that is, the switch setting command) has
completed. When the program continues after the
completion of the command, the switches are guaranteed
to have settled.
.
210 Output system configuration query to switch.
220 Read back switch configuration. Since this query was
preceded by the *WAI command, the switch movement
must be settled before the query is responded to.
230 Return execution to the line after call to subroutine
(line 110).
4-34
Programming
Example 2: Switch position using the Status Byte Register
Example 2: Switch position using the Status Byte Register
This program is identical in functionality to the first sample program except a
different method is used for determining when the switch has settled. The settling routine used here reads the Status Byte Register repeatedly until bit
0 returns to zero.
Status_byte
20 DIM Command$[80],Channel$[80],Error_return$[80]
30 Switch_addr=711
40 CLEAR SCREEN
50 INPUT "Enter A-port position : ",A_position
60 INPUT "Enter B-port position : ",B_position
70 Channel$="A"&TRIM$(VAL$(A_position))&",B"&TRIM$
(VAL$(B_position))
80 Command$="ROUTE:LAYER1:CHANNEL "&Channel$
90 OUTPUT Switch_addr;Command$
100 GOSUB Wait_to_settle
110 REPEAT
120 OUTPUT Switch_addr;"SYSTEM:ERROR?"
130 ENTER Switch_addr;Error_return$
140 PRINT Error_return$
150 UNTIL (VAL(Error_return$)=0)
170 GOTO Exit_prog
180 !
190 Wait_to_settle: ! wait for switch to settle
200 REPEAT
210 OUTPUT Switch_addr;"*STB?"
220 ENTER Switch_addr;Status_byte
230 UNTIL NOT BIT(Status_byte,0)
240 RETURN
250 !
260 Exit_prog:!
270 END
4-35
Programming
Example 2: Switch position using the Status Byte Register
Description Line No.
10 to 170 Same as in Example 1 except for declaration of
Status_byte
.
190 to 240 The new
200 Start
Wait_to_settle
REPEAT
loop.
subroutine.
210 Send the *STB? command to the switch. This queries the
switch to return the value of the status byte.
220 Read the status byte.
230 If the LSB (Least Significant Bit) of the status byte is
0 (that is, the switch is settled), exit the loop. If the LSB is
1 (that is, the switch is moving), then loop back to line 200.
240 Return from subroutine.
4-36
Programming
Example 3: Switch position using the *OPC command
Example 3: Switch position using the *OPC command
This program is identical in functionality to the first two example programs,
except that it uses yet another method for determining when the switch is settled. This settling method sends the *OPC command before the
ROUTE:LAYER:CHANNEL commands and then reads the Standard Event Status Register repeatedly until bit 0 is set to one.
(VAL$(B_position))
80 Command$="ROUTE:LAYER1:CHANNEL "&Channel$
81 OUTPUT Switch_addr;"*OPC"
90 OUTPUT Switch_addr;Command$
100 GOSUB Wait_to_settle
110 REPEAT
120 OUTPUT Switch_addr;"SYSTEM:ERROR?"
130 ENTER Switch_addr;Error_return$
140 PRINT Error_return$
150 UNTIL (VAL(Error_return$)=0)
170 GOTO Exit_prog
180 !
190 Wait_to_settle: ! wait for switch to settle
200 REPEAT
210 OUTPUT Switch_addr;"*ESR?"
220 ENTER Switch_addr;Esr_byte
230 UNTIL BIT(Esr_byte,0)
240 RETURN
250 !
260 Exit_prog:!
270 END
Description
Line No.
10 to 170 Same as in Example 1 except for declaration of
190 to 240 The new
Wait_to_settle
subroutine.
Esr_byte
4-37
.
Programming
Example 3: Switch position using the *OPC command
200 Start
REPEAT
loop.
210 Send the *ESR? command to the switch. This queries the
switch for the Event Status Register value.
220 Read the value of the Event Status Register.
230 If the LSB of the ESR is 1 (that is, the switch is settled),
exit the loop. If the LSB is 0 (that is, switch is moving),
then loop back to line 200.
240 Return from subroutine.
4-38
Programming
Example 4: Input/output multiplexers
Example 4: Input/output multiplexers
This program illustrates how two HP 86060C-series switches may be used to
function as input and output multiplexers in an automated test system. For
this example, two 1 × 8 switches are used to test 8 optical devices under test.
The block diagram for this test system is shown below.
Block diagram of the test system
This example test system uses an HP 8153A optical multimeter equipped with
an HP 81554SM laser source and an HP 81532A optical sensor. This program
periodically (every 5 minutes) measures the optical power through each
device under test and displays an error message if any measured power drops
below 1 microwatt.
Opt_meter_addr
20 INTEGER Meas_count,Current_dut
30 REAL Watts_read,Min_power
40 DIM In_switch$[80],Out_switch$[80],Dummy$[50]
50 !
60 CLEAR SCREEN
70 PRINT "THIS PROGRAM MEASURES ALL DUT POWERS EVERY 5
MINUTES."
4-39
Programming
Example 4: Input/output multiplexers
80 PRINT "IT WILL STOP MEASURING AFTER 50 MEASUREMENT
CYCLES."
90 PRINT "TO HALT PROGRAM EARLIER, PRESS F8."
100 !
110 GOSUB Init_system
120 !
130 ON TIME 300 GOSUB Measure_duts
140 ON KEY 8 LABEL "QUIT" GOTO End_prog
150 !
160 Meas_count=0
170 REPEAT
180 UNTIL Meas_count=50
190 GOTO End_prog
200 !
210 Init_system:! Initialize HPIB instruments
220 CLEAR (7) ! clear HPIB interface
230 ! set HPIB instrument addresses
240 In_switch_addr=711
250 Out_switch_addr=712
260 Opt_meter_addr=722
270 ! set minimum power allowed to 1 microwatt
280 Min_power=1.E-6
290 ! Turn on autoranging
300 OUTPUT @Opt_meter_addr;"SENSE2:POWER:RANGE:AUTO
ON"
310 ! Select Watts as output units
320 OUTPUT @Opt_meter_addr;"SENSE2:POWER:UNIT WATT"
330 ! Select 1550 nm wavelength from source
340 OUTPUT @Opt_meter_addr;"SOURCE1:POWER:WAVELENGTH
UPPER"
350 RETURN
360 !
370 !
380 Measure_duts:! measure all eight duts
390 ! Turn on laser
400 OUTPUT @Opt_meter_addr;"SOURCE1:POWER:STATE ON"
410 ! step through and measure all DUTs
420 FOR Current_dut=1 TO 8
430 ! build hpib commands to send to switches
440 In_switch$="ROUTE:LAYER1:CHANNEL
B"&TRIM$(VAL$(Current_dut))
450 Out_switch$="ROUTE:LAYER1:CHANNEL
B"&TRIM$(VAL$(Current_dut))
460 ! send command to switches
470 OUTPUT @In_switch_addr;In_switch$
480 OUTPUT @Out_switch_addr;Out_switch$
490 ! wait for switches to settle
500 OUTPUT @In_switch_addr;"*WAI"
510 OUTPUT @In_switch_addr;"SYSTEM:CONFIG?"
520 ENTER @In_switch_addr;Dummy$
530 OUTPUT @Out_switch_addr;"*WAI"
4-40
Programming
Example 4: Input/output multiplexers
540 OUTPUT @Out_switch_addr;"SYSTEM:CONFIG?"
550 ENTER @Out_switch_addr;Dummy$
560 ! measure optical power
570 OUTPUT @Opt_meter_addr;"READ2:POW?"
580 ENTER @Opt_meter_addr;Watts_read
590 IF (Watts_read<Min_power) THEN
600 PRINT USING """DUT #"",K,"" Power
Low!""";Current_dut
610 END IF
620 NEXT Current_dut
630 Meas_count=Meas_count+1
640 RETURN
650 !
660 End_prog:! quit program
670 ! turn off time initiated branching
680 OFF TIME
690 END
Description Line No.
10 to 40 Declare some variables to use in program.
60 to 90 Clear screen and print heading.
110 Call
130 Set up time initiated branching. Every 300 seconds,
140 Set up function key initiated branching. When f8 is pressed,
160 Initialize measurement counter to zero.
170 to 180 Loop, waiting for events, until measurement count gets to
190 Exit program.
210 to 350 The
220 Clear the HP-IB bus.
240 to 260 Define HP-IB addresses for input and output switches,
280 Define
300 to 340 Configure optical multimeter to enable autoranging, set
Init_system
Measure_duts
subroutine.
subroutine will be called.
program will end.
50.
Init_system
subroutine.
optical multimeter.
Min_power
to be 1 microwatt.
power units to Watts, and select the upper wavelength
(1550 nm) on the source.
4-41
Programming
Example 4: Input/output multiplexers
350 Return to calling line.
380 to 600 The
Measure_duts
subroutine.
400 Turn laser source on.
420 to 580
FOR-NEXT
loop to step through and measure each of the
8 devices-under-test.
440 to 450 Create the HP-IB commands to select the input and output
switch positions to measure the current device-under-test.
For example, if
Out_switch$
B3"
. Note that since the switches are 1x8, it is not
Current_dut=3
would equal
"ROUTE:LAYER1:CHANNEL
In_switch$
then
and
necessary to specify the position of the A ports.
500 to 550 Use the *WAI command to ensure both switches have
settled.
570 to 610 Read the optical power from the optical multimeter. If this
power is less than
Min_power
then print appropriate error
message to screen.
630 Increment
Meas_count
.
640 Return to calling line.
660 to 690
End_prog
subroutine. Turn off time-initiated branching
and end program.
4-42
5
Spare Channel Replacement Procedure5-4
Electrostatic Discharge Information 5-7
Servicing
Servicing
Servicing
Servicing
This chapter provides a procedure to replace an internal fiber-optic cable with
a spare cable. This procedure may be needed if a fiber-optic cable becomes
damaged through improper connections. Spare fiber-optic cables are provided
inside the lightwave switch.
Before servicing this lightwave switch, familiarize yourself with the safety
markings on the instrument and the safety instructions in this manual. This
instrument has been manufactured and tested according to international
safety standards. To ensure safe operation of the instrument and the personal
safety of the user and service personnel, the cautions and warnings in this
manual must be heeded.
WARNING
WARNING
WARNING
WARNING
CAUTION
Refer to the summary of safety considerations at the front of this
manual.
These servicing instructions are for use by qualified personnel only.
To avoid electrical shock, do not perform any servicing unless you are
qualified to do so.
Failure to ground the lightwave switch properly can result in personal
injury, as well as instrument damage.
The opening of covers or removal of parts is likely to expose
dangerous voltages. Disconnect the instrument from all voltage
sources while it is being opened.
Electrostatic discharge (ESD) can damage or destroy electronic components.
All work on electronic assemblies should be performed at a static-safe work
station. Refer to “Electrostatic Discharge Information” on page 5-7 for more
information on preventing ESD.
5-2
Servicing
Servicing
Required service tools
To enable extra fiber/connector the following tools are required:
This procedure is included so the instrument can be repaired quickly in the
field. Performing this procedure violates the calibration seal. When spare fibers
have replaced bad port fibers, the unit should be sent to an HP Service Center,
at your earliest convenience, to have new port fiber/connectors spliced into the
instrument and to have the instrument calibrated.
1
Disconnect the power cord from the instrument.
2
Remove the top cover using a #15 or #20 TORX driver.
3
Locate out the spare cable/connector. Refer to Figure 5-1 on page 5-5 and
Figure 5-2 on page 5-6. The spare cable/connector is on the bottom of the
switch assembly, secured with a tie-wrap. Remove the tie-wrap.
Switches with 50 channels or less have one spare fiber. Switches with more
than 50 channels have two spare fibers, labeled S1 and S2.
4
Replace the defective port cable/connector assembly with the spare assembly.
Note the label on the spare cable/connector and record the identification
number (S1 or S2) for use later on in this procedure.
Connect the power cord to the instrument. Turn the instrument on.
7
Help
Press
keypad: 86060, 86061, or 86062.
, then enter the model number of the lightwave switch from the
8
Press the
9
Perform this step
the lightwave switch has multiple switch layers installed. (A switch layer is a
switch matrix of “A” ports and “B” ports. Normal lightwave switches have only
one layer installed.) If the rear panel shows more than one set of “A” ports and
“B” ports, the instrument has multiple switch layers. The display will prompt
5-4
4 key
.
only
if the lightwave switch has multiple layers. Determine if
Spare Channel Replacement Procedure
you for the switch layer on which to install the spare fibers:
Press 1 for layer 1
Press 2 for layer 2
10
The display will prompt you to enter 1 for spare fiber 1, or 2 for spare fiber 2.
You recorded this identification number in Step 4.
11
The display will prompt you to enter the number of the front or rear panel
channel to be replaced. Enter the number of the channel and press
The display will show the intended cable changes.
Enter
Servicing
.
12
Press
.
Enter
to confirm the changes.
Figure 5-1. Top view of the HP 86060C or 86061C lightwave switch
5-5
Servicing
Spare Channel Replacement Procedure
Figure 5-2. Top view of the HP 86062C lightwave switch
5-6
Electrostatic Discharge Information
Electrostatic Discharge Information
Electrostatic discharge (ESD) can damage or destroy electronic components.
All work on electronic assemblies should be performed at a static-safe work
station. The following figure shows an example of a static-safe work station
using two types of ESD protection:
• Conductive table-mat and wrist-strap combination.
• Conductive floor-mat and heel-strap combination.
Servicing
Both types, when used together, provide a significant level of ESD protection.
Of the two, only the table-mat and wrist-strap combination provides adequate
ESD protection when used alone.
5-7
Servicing
Electrostatic Discharge Information
To ensure user safety, the static-safe accessories must provide at least 1 MΩ of
isolation from ground. Refer to Table 3 on page 5-8 for information on ordering
static-safe accessories.
WARNING
Reducing ESD
Damage
These techniques for a static-safe work station should not be used
when working on circuitry with a voltage potential greater than
500 volts.
The following suggestions may help reduce ESD damage that occurs during
testing and servicing operations.
• Personnel should be grounded with a resistor-isolated wrist strap before removing any assembly from the unit.
• Be sure all instruments are properly earth-grounded to prevent a buildup of
static charge.
Table 3. Static-Safe Accessories
HP Part Number Description
9300-0797
9300-0980Wrist-strap cord 1.5 m (5 ft)
9300-1383Wrist-strap, color black, stainless steel, without cord, has four adjustable
Set includes: 3M static control mat 0.6 m
(15 ft) ground wire. (The wrist-strap and wrist-strap cord are not included.
They must be ordered separately.)
links and a 7 mm post-type connection.
× 1.2 m (2 ft× 4 ft) and 4.6 cm
9300-1169ESD heel-strap (reusable 6 to 12 months).
5-8
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.