Agilent 8560E Data Sheet

Calibration Guide
Agilent Technologies 8560 E-Series and EC-Series
Spectrum Analyzers
Manufacturing Part Number: 08560-90160
Supersedes December 1999
Printed in USA November 2000
© Copyright 1991-2000 Agilent Technologies, Inc.
Notice
The information contained in this document is subject to change without notice. Agilent Technologies makes no warranty of any kind with regard to this material,
including but not limited to, the implied warranties of merchantability and fitness for a particular purpose. Agilent Technologies shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.
Certification
Agilent Technologies certifies that this product met its published specifications at the time of shipment from the factory. Agilent Technologies further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology, to the extent allowed by the Institute's calibration facility, and to the calibration facilities of other International Standards Organization members.
Regulatory Information
The specifications and characteristics chapters and General Regulatory and Safety Information chapter in this manual contain regulatory information.
Warranty
This Agilent Technologies instrument product is warranted against defects in material and workmanship for a period of one year from date of shipment. During the warranty period, Agilent Technologies will, at its option, either repair or replace products which prove to be defective.
For warranty service or repair, this product must be returned to a service facility designated by Agilent Technologies. Buyer shall prepay shipping charges to Agilent Technologies and Agilent Technologies shall pay shipping charges to return the product to Buyer. However, Buyer shall pay all shipping charges, duties, and taxes for products returned to Agilent Technologies from another country.
Agilent Technologies warrants that its software and firmware designated by Agilent Technologies for use with an instrument will execute its programming instructions when properly installed on that instrument. Agilent Technologies does not warrant that the operation of the instrument, or software, or firmware will be uninterrupted or error-free.
L
IMITATION OF WARRANTY
The foregoing warranty shall not apply to defects resulting from improper or inadequate maintenance by Buyer, Buyer-supplied software or interfacing, unauthorized modification or misuse, operation outside of the environmental specifications for the product, or improper site preparation or maintenance.
NO OTHER WARRANTYISEXPRESSEDORIMPLIED. AGILENT TECHNOLOGIES SPECIFICALLY DISCLAIMS THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
E
XCLUSIVE REMEDIES
THE REMEDIES PROVIDED HEREIN ARE BUYER'S SOLE AND EXCLUSIVE REMEDIES. AGILENT TECHNOLOGIES SHALL NOT BE LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES, WHETHER BASED ON CONTRACT, TORT, OR ANY OTHER LEGAL THEORY.
Assistance
Product maintenance agreements and other customer assistance agreements are available for Agilent Technologies products.
For any assistance, contact your nearest Agilent Technologies Sales and Service Office.
Cleaning
The instrument front and rear panels should be cleaned using a soft cloth with water or a mild soap and water mixture.
General Safety Considerations
WARNING Before this instrument is switched on, make sure it has been properly
grounded through the protective conductor of the ac power cable to a socket outlet provided with protective earth contact.
Any interruption of the protective (grounding) conductor, inside or outside the instrument, or disconnection of the protective earth terminal can result in personal injury.
WARNING There are many points in the instrument which can, if contacted,
cause personal injury. Be extremely careful. Any adjustments or service procedures that require operation of the
instrument with protective covers removed should be performed only by trained service personnel.
CAUTION Before this instrument is switched on, make sure its primary power circuitry
has been adapted to the voltage of the ac power source. Failure to set the ac power input to the correct voltage could cause damage to
the instrument when the ac power cable is plugged in.
Differences between Agilent Technologies 8560 E-Series and EC-Series Spectrum Analyzers
1. EC-series analyzers use a color LCD display which is not adjustable. E-series instruments use a monochrome CRT display which can be adjusted for focus and intensity. E-series analyzers use a high-voltage module (HVM), located in the power supply, to supply power to the CRT. EC-series analyzers do not use an HVM.
2. EC-series analyzers offer fast analog-to-digital-conversion (FADC) digitized sweep times as a standard feature. EC-series FADC circuitry is integrated into the A2 controller board. E-series instruments offer FADC feature as an option (Option 007) which is implemented through the use of the A16 FADC board. FADC enables you to view displays with fast sweep times, between 30 ms and 50 µs.
3. EC-series instruments offer a rear-panel VGA connection as a standard feature. E-series instruments do not have a rear-panel VGA connector. The VGA connection is always active and does not require user activation.
In all other operational respects the 8560 EC-series and E-series are identical.
Contents
1. Using Operation Verification Software Volume I
What You'll Find in This Chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Getting Started . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Using Operation Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Operation Verification Menu Softkeys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Test Descriptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
10 MHz Reference Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
1ST LO OUTPUT Amplitude Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Calibrator Amplitude Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Displayed Average Noise Level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Fast Sweep Time Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Frequency Readout/Frequency Counter Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Frequency Response . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Frequency Span Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .45
IF Gain Uncertainty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Image and Multiple Responses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Input Attenuator Switching Uncertainty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Noise Sidebands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
RES BW Accuracy and Selectivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
RES BW Switching and IF Alignment Uncertainty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Residual FM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Scale Fidelity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Second Harmonic Distortion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Operation Verification Error Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
2. 8560E/EC Specifications and Characteristics
Specifications and Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Calibration Cycle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Frequency Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Amplitude Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .73
Tracking Generator Specifications (Option 002) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
Inputs and Outputs Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
General Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
Frequency Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .87
Amplitude Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .91
Option 002 Tracking Generator Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
Inputs and Outputs Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
Regulatory Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
3. 8561E/EC Specifications and Characteristics
Specifications and Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102
Calibration Cycle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
Frequency Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
Amplitude Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 112
Inputs and Outputs Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119
General Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120
Frequency Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 124
Amplitude Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 128
Contents
Inputs and Outputs Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .130
Regulatory Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .134
4. 8562E/EC Specifications and Characteristics
Specifications and Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .138
Calibration Cycle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .139
Frequency Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .140
Amplitude Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .149
Inputs and Outputs Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .157
General Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .158
Frequency Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .162
Amplitude Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .166
Inputs and Outputs Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .169
Regulatory Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .173
5. 8563E/EC Specifications and Characteristics
Specifications and Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .178
Calibration Cycle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .179
Frequency Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .180
Amplitude Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .189
Inputs and Outputs Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .197
General Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .198
Frequency Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .202
Amplitude Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .206
Inputs and Outputs Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .209
Regulatory Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .213
6. 8564E/EC Specifications and Characteristics
Specifications and Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .218
Calibration Cycle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .219
Frequency Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .220
Amplitude Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .229
Inputs and Outputs Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .237
General Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .238
Frequency Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .242
Amplitude Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .246
Inputs and Outputs Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .250
Regulatory Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .254
7. 8565E/EC Specifications and Characteristics
Specifications and Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .258
Calibration Cycle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .259
Frequency Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .260
Amplitude Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .269
Inputs and Outputs Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .277
General Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .278
Frequency Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .282
Amplitude Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .286
Contents
Inputs and Outputs Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 290
Regulatory Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 294
8. Using Performance Tests: 3335A Source not Available
What You'll Find in This Chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 298
Before You Start . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 299
Required Test Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 301
11a. Resolution Bandwidth Switching and IF Alignment Uncertainty . . . . . . . . . . . . . . . . 308
12a. Resolution Bandwidth Accuracy and Selectivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 312
13a.Input AttenuatorSwitchingUncertainty:8560E/EC,8561E/EC,8562E/EC,and8563E/EC 321
14a. Input Attenuator Switching Uncertainty: 8564E/EC and 8565E/EC . . . . . . . . . . . . . . 329
15a. IF Gain Uncertainty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 338
16a. Scale Fidelity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 346
33a. Second Harmonic Distortion: 8560E/EC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 359
34a. Second Harmonic Distortion: Agilent 8561E/EC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 361
36a. Frequency Response: 8560E/EC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 366
37a. Frequency Response: 8561E/EC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 378
38a. Frequency Response: 8562E/EC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 394
39a. Frequency Response: Agilent 8563E/EC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 417
40a. Frequency Response: Agilent 8564E/EC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 436
41a. Frequency Response: Agilent 8565E/EC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 458
43a. Third Order Intermodulation Distortion: 8560E/EC . . . . . . . . . . . . . . . . . . . . . . . . . . . 480
44a. Third Order Intermodulation Distortion: Agilent 8561E/EC . . . . . . . . . . . . . . . . . . . . 484
45a. Third Order Intermodulation Distortion: Agilent 8562E/EC, Agilent 8563E/EC . . . . 491
46a. Third Order Intermodulation Distortion: Agilent 8564E/EC, Agilent 8565E/EC . . . . 500
9. General Regulatory and Safety Information
Safety Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 510
General Safety Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 511
10. Using Performance Tests Volume II
What You'll Find in This Chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 514
Performance Tests versus Operation Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 515
Before You Start . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 516
85629B Functional Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 521
Running the Functional Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 523
1. 10 MHz Reference Output Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 534
2. 10 MHz Reference Output Accuracy (Option 103) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 538
3. Fast Sweep Time Accuracy (EC-Series and E-Series with Option 007) . . . . . . . . . . . . . . 540
4. Calibrator Amplitude Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 542
5. Displayed Average Noise Level: 8560E/EC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 544
6. Displayed Average Noise Level: 8561E/EC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 550
7. Displayed Average Noise Level: 8562E/EC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 556
8. Displayed Average Noise Level: 8563E/EC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 563
9. Displayed Average Noise Level: 8564E/EC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 571
10. Displayed Average Noise Level: 8565E/EC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 580
Contents
11. Resolution Bandwidth Switching and IF Alignment Uncertainty . . . . . . . . . . . . . . . . .590
12. Resolution Bandwidth Accuracy and Selectivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .593
13. Input Attenuator Switching Uncertainty: 8560E/EC, 8561E/EC, 8562E/EC/, 8563E/EC. . . 601
14. Input Attenuator Switching Uncertainty: 8564E/EC, 8565E/EC. . . . . . . . . . . . . . . . . . .608
15. IF Gain Uncertainty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .615
16. Scale Fidelity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .620
17. Residual FM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .630
18. Noise Sidebands: 8560E, 8561E, and 8563E . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .633
19. Noise Sidebands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .636
20. Image, Multiple, and Out-of-Range Responses: 8560E/EC . . . . . . . . . . . . . . . . . . . . . . .640
21. Image, Multiple, Out-of-Band, and Out-of-Range Responses: 8561E/EC . . . . . . . . . . . .643
22. Image, Multiple, and Out-of-Band Responses: 8562E/EC . . . . . . . . . . . . . . . . . . . . . . . .647
23. Image, Multiple, and Out-of-Band Responses: 8563E/EC . . . . . . . . . . . . . . . . . . . . . . . .651
24. Image, Multiple, and Out-of-Band Responses: 8564E/EC . . . . . . . . . . . . . . . . . . . . . . . .657
25. Image, Multiple, and Out-of-Band Responses: 8565E/EC . . . . . . . . . . . . . . . . . . . . . . . .664
26. Frequency Readout/Frequency Count Marker Accuracy: 8560E/EC . . . . . . . . . . . . . . . .672
27. Frequency Readout/Frequency Count Marker Accuracy: 8561E/EC . . . . . . . . . . . . . . . .675
28. Frequency Readout Accuracy/Frequency Count Marker Accuracy: 8562E/EC . . . . . . . .678
29. Frequency Readout Accuracy/Frequency Count Marker Accuracy: 8563E/EC . . . . . . . .681
30. Frequency Readout Accuracy/Frequency Count Marker Accuracy: 8564E/EC . . . . . . . .685
31. Frequency Readout Accuracy/Frequency Count& Marker Accuracy: 8565E/EC. . . . . . .689
32. Pulse Digitization Uncertainty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .694
33. Second Harmonic Distortion: 8560E/EC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .698
34. Second Harmonic Distortion: 8561E/EC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .700
35. Second Harmonic Distortion: 8562E/EC, 8563E/EC, 8564E/EC, 8565E/EC . . . . . . . . . .705
36. Frequency Response: 8560E/EC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .712
37. Frequency Response: 8561E/EC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .723
38. Frequency Response: 8562E/EC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .738
39. Frequency Response: 8563E/EC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .758
40. Frequency Response: 8564E/EC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .774
41. Frequency Response: 8565E/EC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .793
42. Frequency Span Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .813
43. Third Order Intermodulation Distortion: 8560E/EC . . . . . . . . . . . . . . . . . . . . . . . . . . . .816
44. Third Order Intermodulation Distortion: 8561E/EC . . . . . . . . . . . . . . . . . . . . . . . . . . . .820
45. Third Order Intermodulation Distortion: 8562E/EC, 8563E/EC . . . . . . . . . . . . . . . . . . .826
46. Third Order Intermodulation Distortion: 8564E/EC, 8565E/EC . . . . . . . . . . . . . . . . . . .834
47. Gain Compression: 8560E/EC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .842
48. Gain Compression: 8561E/EC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .845
49. Gain Compression: 8562E/EC, 8563E/EC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .849
50. Gain Compression: 8564E/EC, 8565E/EC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .854
51. 1ST LO OUTPUT Amplitude: 8560E/EC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .859
52. 1ST LO OUTPUT Amplitude: 8561E/EC, 8562E/EC, 8563E/EC, 8564E/EC,8565E/EC 863
53. Sweep Time Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .866
54. Residual Responses: 8560E/EC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .872
55. Residual Responses: 8561E/EC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .875
56. Residual Responses: 8562E/EC, 8563E/EC, 8564E/EC, 8565E/EC . . . . . . . . . . . . . . . . .878
57. IF INPUT Amplitude Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .881
58. Gate Delay Accuracy and Gate Length Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .884
Contents
59. Delayed Sweep Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 890
60. Tracking Generator Level Flatness . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 893
61. Absolute Amplitude and Vernier Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 897
62. Maximum Leveled Output Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 900
63. Power Sweep Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 903
64. RF-Power-Off Residuals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 906
65. Harmonic Spurious Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 909
66. Non-Harmonic Spurious Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 913
67. LO Feedthrough Amplitude . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 919
68. Tracking Generator Feedthrough . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 922
69. Frequency Tracking Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 926
70. Tracking Generator Frequency Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 929
11. 8560E/EC Performance Test Record
Test Record. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 934
12. 8561E/EC Performance Test Record
Test Record. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 958
13. 8562E/EC Performance Test Record
Test Record. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 980
14. 8563E/EC Performance Test Record
Test Record. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1002
15. 8564E/EC Performance Test Record
Test Record. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1028
16. 8565E/EC Performance Test Record
Test Record. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1058
Contents
10
1 Using Operation Verification
Software Volume I
11
Using Operation Verification Software Volume I
What You'll Find in This Chapter
What You'll Find in This Chapter
This automated test software is designed to give a high confidence level of spectrum analyzer operation in a reasonable amount of time. Instructions for using the software, brief descriptions of the tests, and a list of the software error messages are included in this chapter. Refer to
Table 1-3, on page 20, for the names of manual performance tests that
are not included in the automated operation verification software.
What Is Operation Verification?
Operation Verification automates many of the performance tests. It is recommended as a check of instrument operation for incoming inspection or after a repair. It performs 80 to 85 percent of the manual performance tests in less than 60 minutes (75 minutes for the Agilent 8564E/EC and Agilent 8565E/EC). It is designed to test an instrument operating within a 20 °Cto30°C temperature range using a minimum set of test equipment. Refer to Table 1-1, on page 16 for a list of tests performed.
If a test does not pass, the related manual performance test needs to be run. Related manual performance tests are listed at the top of each operation verification test description and the test procedures are in “Using Performance Tests” Chapter 10, of Volume II of the calibration guide.
Operation Verification versus Performance Tests
The performance tests verify that the analyzer performance meets all specifications. Performance verification consists of executing all of the manual performance tests. It is time-consuming and requires extensive test equipment.
12 Chapter1
Using Operation Verification Software Volume I
Getting Started
Getting Started
First, make sure you have a compatible controller (computer), the proper test equipment, and a printer for recording test results. The following paragraphs describe requirements for controllers, test equipment, and printers. Once the proper equipment is identified, proceed to “Equipment Connections.”
Spectrum Analyzers
The 8560 E-Series and EC-Series Operation and Verification software tests the following spectrum analyzers:
Spectrum Analyzer Software Revision
8560E/EC all
Agilent 8561E/EC Agilent 8562E/EC Agilent 8563E/EC Agilent 8564E/EC Agilent 8565E/EC
all D.00.00 and later all C.00.00 and later C.00.00 and later
Errors will occur if this software is used to test spectrum analyzers other than those listed above. Other versions of the operation verification software are available for testing other 8560 Series spectrum analyzers.
Controller (Computer)
Operation Verification software requires using any combination of one of the following controllers and the HP BASIC operating system:
Controller
9000 model 216 9000 model 236 9000 model 310
Chapter 1 13
Using Operation Verification Software Volume I
Getting Started
Operating System
HP BASIC 2.0 with extensions 2.1 HP BASIC 3.0 and required BIN files HP BASIC 4.0 and required BIN files
Other 9000 Series 300 controllers are compatible with the Operation Verification software; however, the graph printouts might not be full width. This is especially true with medium- and high-resolution displays.
The Operation Verification program requires at least 500K of free memory. The computer can have either single or dual GPIB ports. Refer to “Dual-Bus Operation,” on page 28, under “Using Operation
Verification,” on page 22, for information on using the program with
dual GPIB ports.
Test Equipment
Table 1-1, on page 16, lists the operation verification tests and the test
equipment required for each test. You do not need all the test equipment connected to perform operation verification. You need only connect the equipment specified in each test to run that test.
NOTE The Agilent 3335A Synthesizer Level Generator, which is used in
several operation verification tests, has been obsoleted. If an Agilent 3335A is not available, alternate test equipment can be used to perform comparable tests manually. Chapter 8 in this volume contains the information on alternate test equipment and manuals test procedures you will need to perform these tests.
Table 1-2, on page 19, summarizes the equipment required to run the
Operation Verification tests. Some tests, like 10 MHz Reference Accuracy, can use various model numbers of a particular equipment type. Information about selecting the equipment model number you want to use is provided in “Setting GPIB Addresses,” on page 25, in this chapter under “Using Operation Verification,” on page 22.
NOTE The validity of Operation Verification program measurements depends
in part on required test equipment measurement accuracy. Verify proper calibration of test equipment before testing the analyzer with the software.
14 Chapter1
Using Operation Verification Software Volume I
Getting Started
Printers
All test results are sent to a GPIB printer. The program does not run without being connected to a GPIB printer. Virtually any GPIB graphics workstation printer can work. These tests have been run using the HP ThinkJet, 2671G, 82906A, and 9876G printers.
Warmup Time Test Equipment Warmup Allow sufficient warmup time for test
equipment. Refer to their individual operating and service manuals for warmup specifications.
Spectrum Analyzer Warmup Warm the spectrumanalyzer up for at least five minutes before performing the first test.
NOTE Since the preselector in the Agilent 8561E/EC, Agilent 8562E/EC,
Agilent 8563E/EC, Agilent 8564E/EC and Agilent 8565E/EC spectrum analyzers is not peaked during the Frequency Response test, allow a 30 minute warmup at 20 °Cto30°C before beginning this test. If you start running All Testsafter five-minute warmup, the spectrum analyzer will have warmed up for 30 minutes by the time the Frequency Response test begins.
Chapter 1 15
Using Operation Verification Software Volume I
Getting Started
Table 1-1 Equipment Required for Each Operation Verification Test
Test Name Equipment Required
10 MHz Reference Accuracy
1ST LO OUTPUT Amplitude Accuracy
Calibrator Amplitude Accuracy
Displayed Average Noise Level
Fast Sweep Time Accuracy
Agilent
8560E/EC
5342A, 5343A, or 5350B*
8902A or 436A or 438A
8485A or 8481A
8902A, 436A or 438A
8481A, or 8482A
909D 909D 909D 909D 85138B 85138B
5342A, 5343A, or 5350B**
Agilent
8561E/EC
5342A, 5343A, or 5350B*
8902A or 436A or 438A
8485A or 8481A
8902A, 436A or 438A
8481A, or 8482A
5342A, 5343A, or 5350B**
Agilent
8562E/EC
5342A, 5343A, or 5350B*
8902A or 436A or 438A
8485A or 8481A
8902A, 436A or 438A
8481A, or 8482A
5342A, 5343A, or 5350B**
Agilent
8563E/EC
5342A, 5343A, or 5350B*
8902A or 436A or 438A
8485A or 8481A
8902A, 436A or 438A
8481A, or 8482A
5342A, 5343A, or 5350B**
Agilent
8564E/EC
5342A, 5343A, or 5350B*
8902A or 436A or 438A
8485A or 8481A
8902A, 436A or 438A
8481A, or 8482A
5342A, 5343A, or 5350B**
Agilent
8565E/EC
5342A, 5343A, or 5350B*
8902A or 436A or 438A
8485A or 8481A
8902A, 436A or 438A
8481A, or 8482A
5342A, 5343A, or 5350B**
Frequency Readout/ Counter Accuracy
Frequency Span Accuracy
8340A/B or 83640A or 83650A
8120-4921
8340A/B or 83640A or 83650A
11667A/B
8120-4921
8340A/B or 83640A or 83650A
8120-4921
8340A/B or 83640A or 83650A
11667A/B
8120-4921
8340A/B or 83640A or 83650A
8120-4921
8340A/B or 83640A or 83650A
11667A/B
8120-4921
8340A/B or 83640A or 83650A
8120-4921
8340A/B or 83640A or 83650A
11667B
8120-4921
83640A or 83650A
8120-6164
83640A or 83650A
11667C
8120-6164
83650A
8120-6164
83650A
11667C
8120-6164
16 Chapter1
Using Operation Verification Software Volume I
Table 1-1 Equipment Required for Each Operation Verification Test
Test Name Equipment Required
Getting Started
Frequency* Response
IF Gain* Uncertainty
Image and Multiple Responses
Agilent
8560E/EC
8340A/B or 83640A or 83650A
3335A 8902A or
436A or 438A
8482A 11667A/B
8120-4921
3335A 3335A 3335A 3335A 3335A 3335A
8340A/B or 83640A or 83650A
8902A or 436A or 438A
8481A 11667A/B
8120-4921
Agilent
8561E/EC
8340A/B or 83640A or 83650A
3335A 8902A or
436A or 438A
8481A 11667A/B
8120-4921
8340A/B or 83640A or 83650A
8902A or 436A or 438A
8481A 11667A/B
8120-4921
Agilent
8562E/EC
8340A/B or 83640A or 83650A
3335A 8902A or
436A or 438A
8481A 11667A/B
8120-4921
8340A/B or 83640A or 83650A
8902A or 436A or 438A
8481A 11667A/B
8120-4921
Agilent
8563E/EC
8340A/B or 83640A or 83650A
3335A 8902A or
436A or 438A
8485A 11667B
8120-4921
8340A/B or 83640A or 83650A
8902A or 436A or 438A
8485A 11667B
8120-4921
Agilent
8564E/EC
83640A or 83650A
3335A 8902A or
436A or 438A
8487A 11667C
8120-6164
83640A or 83650A
8902A or 436A or 438A
8487A 11667C
8120-6164
Agilent
8565E/EC
83650A
3335A 8902A or
436A or 438A
8487A 11667C
8120-6164
83650A
8902A or 436A or 438A
8487A 11667C
8120-6164
Input* Attenuator Switching Uncertainty
Noise Sidebands
RES BW* Accuracy/ Selectivity
3335A 3335A 3335A 3335A 3335A 3335A
8662A or
8663A or CAL OUTPUT signal
3335A 3335A 3335A 3335A 3335A 3335A
8662A or
8663A or CAL OUTPUT signal
Chapter 1 17
8662A or
8663A or CAL OUTPUT signal
8662A or
8663A or CAL OUTPUT signal
8662A or
8663A or CAL OUTPUT signal
8662A or
8663A or CAL OUTPUT signal
Using Operation Verification Software Volume I
Getting Started
Table 1-1 Equipment Required for Each Operation Verification Test
Test Name Equipment Required
RES BW* Switching Uncertainty
Residual FM
Scale* Fidelity
Second Harmonic Distortion
Agilent
8560E/EC
Agilent
8561E/EC
Agilent
8562E/EC
Agilent
8563E/EC
Agilent
8564E/EC
Agilent
8565E/EC
3335A 3335A 3335A 3335A 3335A 3335A
8662A or 8663A
8662A or 8663A
8662A or 8663A
8662A or 8663A
8662A or 8663A
8662A or 8663A
3335A 3335A 3335A 3335A 3335A 3335A
8340A/B or 83640A or 83650A
8902A or 436A or 438A
8485A or 8481A
11667A/B
0955-0306 8120-4921
8340A/B or 83640A or 83650A
8902A or 436A or 438A
8485A or 8481A
11667A/B 11689A (2
required)
0955-0306 8120-4921
8340A/B or 83640A or 83650A
8902A or 436A or 438A
8485A or 8481A
11667A/B 11689A (2
required)
0955-0306 8120-4921
8340A/B or 83640A or 83650A
8902A or 436A or 438A
8485A or 8481A
11667B 11689A (2
required)
0955-0306 8120-4921
83640A or 83650A
8902A or 436A or 438A
8485A or 8481A
11667C 11689A (2
required)
0955-0306 8120-6164
83650A
8902A or 436A or 438A
8485A or 8481A
11667C 11689A (2
required)
0955-0306 8120-6164
* The 3335A Synthesizer Level Generator, which is used in several operation verification tests, has been obsoleted. If an 3335A is not available, alternate test equipment can be used to perform comparable tests manually. Chapter 8 in this volume contains the information on alternate test equipment and manuals test procedures you will need to perform these tests.
* *The Agilent 5350B is supported only in software revision D.00.00 and later.
18 Chapter1
Using Operation Verification Software Volume I
Table 1-2 Required Test Equipment Summary
Type of Equipment Model Number
Controller* Series 200 9000 model 216 ( 9816)
9000 model 236 ( 9836)
or 9000 model 310
Synthesizer/level generator 3335A
Getting Started
Synthesized sweeper (10 MHz to 26.5 GHz)
(10 MHz to 40 GHz) (10 MHz to 50 GHz)
Synthesized signal generator 8662A/8663A Measuring receiver 8902A Power meter 436A or 438A (alternate) Microwave frequency counter 5343A Microwave frequency counter 5342A (alternate) Microwave frequency counter 5350B (alternate) † Power sensor (100 kHz to 4.2 GHz) 8482A Power sensor (50 MHz to 26.5 GHz)
(50 MHz to 50 GHz)
Power sensor (10 MHz to 18 GHz) 8481A (alternate for 8560E/EC, 8561E/EC, and
Power splitter (dc to 18 GHz)
(dc to 26.5 GHz)
(dc to 50 GHz)
8340A/B ( 8560E/EC, 8561E/EC, 8562E/EC, and 8563E/EC)
85640A ( 8564E/EC) 85650A ( 8565E/EC)
8485A ( 8560E/EC, 8561E/EC, 8562E/EC, and 8563E/EC)
8487A ( 8564E/EC and 8565E/EC)
8562E/EC) 11667A ( 8560E/EC, 8561E/EC, and 8562E/EC)
11667B ( 8560E/EC, 8561E/EC, 8562E/EC, and 8563E/EC)
11667C ( 8564E/EC and 8565E/EC)
4.4 GHz low-pass filter (two required) 11689A ( 8561E/EC, 8562E/EC, 8563E/EC, 8564E/EC, and 8565E/EC)
50 termination (dc to 26.5 GHz)
(dc to 50 GHz) 50 MHz low-pass filter 0955-0306 Miscellaneous cables and adapters As per test setup GPIB printer See “Printers”
*
500 kilobytes of free memory is required for the test program.
Th 5350B is supported only in software revision D.00.00 and later.
909D ( 8560E/EC, 8561E/EC, 8562E/EC, and 8563E/EC) 85138B ( 8564E/EC and 8565E/EC)
Chapter 1 19
Using Operation Verification Software Volume I
Getting Started
Table 1-3 Manual Performance Tests That Are Not Automated
Pulse Digitization Uncertainty Third Order Intermodulation Distortion Gain Compression Sweep Time Accuracy Residual Responses IF Input Amplitude Accuracy Gate Delay Accuracy and Gate Length Accuracy Delayed Sweep Accuracy
These tests apply only to 8560E/EC Option 002:
Tracking Generator Level Flatness Absolute Amplitude and Vernier Accuracy Maximum Leveled Output Power Power Sweep Range RF-Power-Off Residuals Harmonic Outputs Non-Harmonic Outputs Tracking Generator Feedthrough LO Feedthrough Amplitude Frequency Tracking Range Tracking Generator Frequency Accuracy
Equipment Connections
Computer (Controller) Setup
For 9000 model 216 or model 236 computers, setup instructions are provided in Chapter 1, “Computer Installation,” of the BASIC Operating Manual. For 9000 model 310 computers, setup information is provided in Configuration Reference Manual for Series 300 computers.
20 Chapter1
Using Operation Verification Software Volume I
Getting Started
GPIB Cables
All test equipment controlled by GPIB should be connected to the internal GPIB of the controller (select code 7). If the controller has only one GPIB connector, connect the spectrum analyzer to it as well. If the controller has dual GPIB connectors, connect the spectrum analyzer under test to the second GPIB (typically, select code 8).
10 MHz Reference
The 10 MHz REF IN/OUT on the spectrum analyzer under test should be connected to the synthesized sweeper external frequency reference. That is, it should be connected to the FREQUENCY STANDARD EXT of the 8340A/B or the 10MHz REF INPUT of the 83640A or 83650A.
Connect the synthesized sweeper 10 MHz REF OUTPUT to the 40/N MHz REF INPUT of the 3335A. Do not connect the spectrum analyzer 10 MHz REF IN/OUT to the external frequency reference input of the 8663A; doing so invalidates the Noise Sidebands test results.
NOTE Terminate the 3335A 10 MHz REF OVEN OUTPUT in 50 Ω. Do not
connect the 10 MHz output to the external frequency reference input of any other test equipment.
Test Setups
Test setups for each test are included with the test. These are in the “Test Descriptions” section of this chapter. The program prompts the operator to make appropriate equipment connections if the correct equipment setup is not detected.
Chapter 1 21
Using Operation Verification Software Volume I
Using Operation Verification
Using Operation Verification
Loading the Program
Load BASIC into the computer. BASIC choices are:
• BASIC 2.0 and extensions 2.1
• BASIC 3.0 or 4.0, which must include the following binaries: MAT
IO GRAPH GRAPHX PDEV IB MS CLOCK CS80 DISC KBD
For configuration instructions, refer to the BASIC Operating Manual. Next, insert the Operation Verification software disk into the disk drive, then type:
LOAD “VERIFY_6XE”,1
NOTE A double-sided disk drive must be used. The Operation Verification
software will not fit on a single-sided formatted disk. Press
Series 300 computers, to load the software and start the program running.
EXECUTE on 9000 Series 200 computers, or RETURN on 9000
22 Chapter1
Using Operation Verification Software Volume I
Using Operation Verification
Program Operation
Operation Verification consists of three menus. They are the Conditions Menu, the Test Menu, and the Sensor Utilities Menu, which are accessed from the Conditions Menu. Program operation is controlled through a combination of softkeys and user prompts. Some prompts, primarily in the Conditions and Sensor Utilities Menus, require computer keyboard entries. Terminate keyboard entries with the
RETURN or ENTER key. Most prompts, however, tell the user what to do
next or provide informational messages. If the message (any key) follows a prompt, pressing any key on the
keyboard continues the program. If the message (any key or `Q' to quit) follows a prompt, pressing any key except Q continues the program. Pressing Q terminates the current procedure at the next, most logical point in the program.
Conditions Menu
The first menu screen displayed is the Conditions Menu. The pointer displayed along the left edge of the screen may be moved with the knob (if one is present) or the up () and down () arrow keys. Notice that the menu has two pages. Moving the pointer below the last entry on the page brings up the next page. Similarly, moving the pointer above the first entry on a subsequent page brings up the preceding page. The two pages of the Conditions Menu have a four-line overlap. The last four lines of page 1 appear as the first four lines on page 2.
Chapter 1 23
Using Operation Verification Software Volume I
Using Operation Verification
Test Record Header Information
The information in the first six entries of this menu is printed out as part of the operation verification test record. The spectrum analyzer model number and serial number are stored in the analyzer memory. Software revisions before C.00.00 will not display the complete list of options if there are more than two. The program queries these numbers via GPIB and displays them. If the spectrum analyzer under test does not respond at the address listed under GPIB Addresses, or no address is listed, a message appears where the model and serial numbers are normally displayed.
The program also queries the time and date in the computer. If an 9000 Series 200 computer is used, it might be necessary to reset the time and date; 9000 Series 300 computers have built-in real-time clocks.
Entries for Operator, Test Conditions, and Other Comments are optional. Blank spaces are provided on the test record if no entry is made. To make or to change an entry, move the pointer to the line where the entry is to be made or changed. Press
Change Entry and type
in your new entry. Entries for Operator, Test Conditions, and Other Comments can be up to 37 characters long, but only the first 25 characters of the Operator entry are printed on the test record.
System Mass Storage File Location
Calibration factor data for different power sensors and a customized set of conditions may be stored on disk. The mass storage unit specifier (msus) for the disk containing this information should be entered as the system mass storage file location. Refer to the BASIC Operating Techniques Manual for information on the syntax of the msus. Software revision C.00.00 and later allows a system mass storage file location with more than 37 characters though only the first and last 17 characters will be displayed.
The Operation Verification program disk comes write-protected from the factory. If you want to use this disk for storing your power sensor and conditions data files, it is necessary to disable the write-protect mechanism.
NOTE A double-sided disk drive must be used. The Operation Verification
software will not fit on a single-sided formatted disk.
24 Chapter1
Using Operation Verification Software Volume I
Using Operation Verification
Power Sensors
The Operation Verification program supports four models of power sensors, but only two models are necessary to run all the tests. The Agilent 8481A may be substituted for the Agilent 8482A. The Agilent 8485A is required for the Agilent 8562E/EC and Agilent 8563E/EC, and the Agilent 8487A is required for the Agilent 8564E/EC and Agilent 8565E/EC. Refer to “Sensor Utilities
Menu,” on page 28, for more information regarding storing, viewing,
editing, and purging cal factor data for power sensors. To select a particular sensor of a certain model number, move the
pointer to the desired model number and press
Change Entry. Enter the
last five digits of the power sensor serial number (that is, the serial number suffix). The program checks to see that a data file containing the cal factor data for that particular sensor exists.
To create, edit, view, or purge power sensor cal factor data files, press
Sensor Utils to bring up the Sensor Utilities Menu. Refer to “Sensor
Utilities Menu,” on page 28, in this chapter for more information.
A WARNING message appears if the program does not find a data file for the sensor. If this occurs, check that the system mass storage file location specifies the disk where the power sensor data resides. If the system mass storage file location is correct, the cal factor data for that particular sensor has not been stored.
NOTE Power sensor data files created using the VERIFY_62 Operation
Verification software (for 8560A/61A/61B/62A/62B/63A spectrum analyzers) are not compatible with the VERIFY_6XE Operation Verification software.
Refer to “Sensor Utilities Menu,” on page 28, in this chapter for additional information.
Setting GPIB Addresses
The last 11 lines (last 12 lines, for revision D.00.00 and later) of the Conditions Menu are for selecting the GPIB addresses of test equipment used for the Operation Verification program. It is not necessary to use all the test equipment listed. Some model numbers listed are “alternates.” Table 1-1, on page 16, lists the test equipment required for each test and Table 1-2, on page 19, lists model numbers allowed for a particular type of test equipment.
NOTE For software revisions B.01.00 and earlier, the Conditions Menu has
two entries for Agilent 8340A/B synthesized sweepers. Only Agilent 8340A/B #1 is used.
Software revisions D.00.00 and later have an entry for an Agilent 5350 Series microwave frequency counter.
Chapter 1 25
Using Operation Verification Software Volume I
Using Operation Verification
Entering zero as the test equipment address results in that model number being unavailable in the program (NA is displayed in the address field). To minimize possible confusion later, enter a zero for the address of each piece of test equipment that is not available.
Enter the address for each piece of test equipment that is available, including the spectrum analyzer under test. Addresses must contain the select code of the bus to which the equipment is connected followed by the equipment address on that bus. For example, if the 8902A is at address 14 on a bus with a select code of 7, enter an address of 714. If the 8902A were on a bus with a select code of 12, you should enter an address of 1214.
A question mark (?) next to a GPIB address indicates the address has not been checked to verify a response. An asterisk (*) next to a GPIB address indicates the address was checked and that an instrument responds at that address. If there is neither an asterisk nor a question mark next to an address, the address has been checked and no response was detected.
Storing and Loading the Conditions File
The information in the Conditions Menu may be stored for future use by pressing
Store Conds. A file named CONDITIONS is created on the
disk specified by the system mass storage file location.
NOTE CONDITIONS files for software versions B.01.00 and earlier are not
compatible with software versions C.00.00 and C.01.00. Versions C.00.00 and later CONDITIONS files include entries for the 83640A and 83650A sweepers and the 8487A power sensor.
CONDITIONS files for software versions C.01.00 and earlier are not compatible with software versions D.00.00 and later. Versions D.00.00 and later CONDITIONS files include entries for 5350 Series microwave frequency counters.
When running the Operation Verification program in the future, set the system mass storage file location to read the disk where the CONDITIONS file is located and press
Load Conds. If the
CONDITIONS file resides on the default system mass storage file location, the CONDITIONS file is loaded automatically the next time the program is run.
The default system mass storage file location is :,700,1. For the 9000 Model 236 ( 9836), it is necessary to use an external disk drive. The disk drive must support double-sided format.
26 Chapter1
Using Operation Verification Software Volume I
Using Operation Verification
Getting to the Test Menu
Once all necessary items in the Conditions Menu are selected, you can run a test by pressing Test Menu. Before the Test Menu is displayed, these things occur:
• Appropriate power sensor data files are loaded.
• GPIB is checked for a response at each address.
• Serial and model number of the spectrum analyzer under test are queried.
• Reference level calibration is performed.
NOTE Power sensor data files created using the VERIFY_62 Operation
Verification software (for 8560A/61A/61B/62A/62B/63A spectrum analyzers) are not compatible with the VERIFY_6XE Operation Verification software.
Refer to “Test Menu,” on page 30, “Test Menu” in this chapter for more details on running the tests. If a printer is unavailable, the Conditions Menu is displayed again rather than the Test Menu. All test results must be sent to the printer.
NOTE Without a printer, Operation Verification tests do not run.
Verifying the GPIB
To see which test equipment responds on GPIB, press
Verify Bus. This
check only verifies that there is a response at the address listed; it cannot tell that a particular piece of equipment is at a particular address. This is useful for verifying GPIB connections without entering the Test Menu.
Querying the Spectrum Analyzer Serial Number
The Operation Verification program automatically queries the spectrum analyzer serial and model number on three occasions: at program initiation, when loading the CONDITIONS file, and when entering the Test Menu. To query the analyzer serial and model numbers at any other time, press
Query DUT S/N. This is helpful for
testing multiple spectrum analyzers; you do not have to reload the CONDITIONS file or restart the program.
Exiting Operation Verification
Press
Exit Program to exit the Operation Verification program.
Chapter 1 27
Using Operation Verification Software Volume I
Using Operation Verification
Dual-Bus Operation
The Operation Verification program may be used on dual GPIB systems, such as the microwave test set. In these systems, all the test equipment is connected to GPIB at select code 7, and the device under test (for example, the spectrum analyzer) is connected to GPIB at select code 8.
To run this program in a dual-bus configuration, enter equipment addresses as described in “Setting GPIB Addresses,” on page 25, making sure that each address properly identifies the bus select code to which it is connected. Program operation is the same for dual-bus and single-bus configuration.
Sensor Utilities Menu
Operation Verification needs to know the cal factors of each power sensor being used. Create, edit, view, and delete data files containing cal factors for each power sensor in the Sensor Utilities Menu. Power sensor data filenames include the last five digits of the power sensor serial number. For example,
for 8481A power sensors the filename is SEN81NNNNN for 8482A power sensors the filename is SEN82NNNNN for 8485A power sensors the filename is SEN85NNNNN for 8487A power sensors the filename is SEN87NNNNN
where NNNNN represents the last five digits of the power sensor serial number (the serial number suffix). Note that the first two digits in the filename correspond to the last two digits of the power sensor model number.
All power sensor data files available on the system mass storage location file are listed upon entering the Sensor Utilities Menu.
NOTE Power sensor data files created using the VERIFY_62 Operation
Verification software (for 8560A/61A/61B/62A/62B/63A Spectrum Analyzers) are not compatible with the VERIFY_6XE Operation Verification software.
28 Chapter1
Adding a Power Sensor Data File
Using Operation Verification Software Volume I
Using Operation Verification
Toadd a new power sensor data file, press
Add File, and enter the power
sensor model number as requested. An error message is displayed if a disk is not found at the current system mass storage file location.
When prompted for the power sensor serial number, enter only the last five digits (the serial number suffix). You are then prompted for a cal factor frequency and for the cal factor. These frequency/cal-factor pairs need not be entered in order of increasing frequency; the program inserts the pairs in their proper place. All frequencies should be entered in MHz.
A 50 MHz Cal-Factor must be entered in order to calibrate the power sensor. Some power sensors do not include a 50 MHz Cal-Factor on their chart or calibration record; it is listed as part of the Calibration Procedure on the case of the power sensor.
If a mistake is made entering a cal factor, enter the frequency of the erroneous cal factor at the next frequency prompt. Enter the correct cal factor at the next prompt. If an error was made entering the frequency value, enter the erroneous frequency at the next frequency prompt and a zero for the cal factor to delete that frequency point.
Once all cal factor data for a power sensor is entered, enter an S at the next frequency prompt. The power sensor data is then stored on disk.
Viewing and Editing a Power Sensor Data File
Press
View/Edit to view or edit a power sensor data file. Only data files
listed on the screen can be viewed or edited. If a file is created but data is not stored, the filename is listed, but no data is viewed and it cannot be edited.
To change the cal factor at a particular frequency, enter that frequency at the frequency prompt, then enter the new cal factor.
To delete a frequency/cal factor pair, enter the frequency of the pair to be deleted and a cal factor of zero. Add a frequency/cal factor pair by entering the new frequency at the frequency prompt and the new cal factor.
Deleting a File
Press
Delete File to remove a listed file. At the prompt, enter the
filename exactly as it appears on-screen. You are asked for confirmation to delete the file.
Chapter 1 29
Using Operation Verification Software Volume I
Using Operation Verification
Changing the System Mass Storage File Location
To add, edit, or view power sensor data on a disk other than the one currently specified by the system mass storage file location, press
System File. Enter the msus of the new system mass storage file
location. All power sensor files residing on that disk are listed. Upon returning to the Conditions Menu, the system mass storage file location is the one determined in the Sensor Utilities Menu.
Listing Available Power Sensor Data Files
Press
List Files to list all power sensor data files on the currently
specified system mass storage location file.
Returning to the Conditions Menu
Press
Cond Menu to return to the Conditions Menu.
Test Menu
The Test Menu displays all tests that can be performed by the Operation Verification program. Notice that the menu has two pages. Tests may be run in any of five modes. These are listed below.
• All Tests runs all 17 tests in the sequence shown on screen.
• Single Sequence runs a user-defined sequence of tests once.
• Single Test runs one test once.
• Repeat Sequence runs a user-defined sequence of tests until testing is aborted.
• Repeat Test runs a single test until the testing is aborted.
If GPIB controlled equipment for a given test does not respond over GPIB, that test is flagged MISSING ETE (missing electronic test equipment). These tests cannot be run and, if they are included as part of a sequence (All Tests,Single Sequence, or Repeat Sequence), they are ignored. See “List Equipment,” below.
If a test does not apply to the spectrum analyzer,the test will be flagged TEST NOT APPLICABLE. For example, if the spectrum analyzer is an E-Series instrument which does not have Option 007, the Fast Sweep Time Accuracy test will have the message TEST NOT APPLICABLE next to it.
Equipment connection prompts are displayed on the computer screen. Most tests check equipment connections and only prompt the operator if a misconnection is detected.
30 Chapter1
Using Operation Verification Software Volume I
Using Operation Verification
If more than one power meter (or the measuring receiver and one of the power meters) is present, the program asks which model to use as the power meter. Enter the model number without the alphabetic character (for example, enter 8902 for an 8902A). Similarly, if more than one frequency counter is present, the program asks which counter to use. Again, enter the model number without the alphabetic character.
The test currently being run and its test number are indicated in the screen title block of the spectrum analyzer under test.
All Tests
To run all 17 tests in the sequence shown, press
All Tests. The pointer
moves to each test as it is being run. All Tests can be run in approximately 60 minutes (75 minutes for an Agilent 8564E/EC or Agilent 8565E/EC).
Three softkeys are displayed when running All Tests. Press
ABORT TEST to abort the current test and continue to the next test.
Press
ABORT SEQUENCE to abort the All Test mode. Pressing Restart
aborts and restarts the current test. If the spectrum analyzer is in the middle of a sweep, no action is taken until the sweep is completed.
Single Sequence
Use this mode to perform a subset of the tests, to run a particular test a specified number of times, or to run all 16 tests in a sequence different from the All Tests sequence. After pressing
Single Sequence, you are
prompted for a test number. The sequence is displayed after each prompt. Up to 25 test numbers may be entered (test number duplication is permitted). Enter a zero to terminate building the sequence and begin testing.
If an error is made in entering the sequence, enter a zero at the next prompt, then press
ABORT SEQUENCE. Now press Single Sequence to
reenter the correct sequence. The ABORT TEST, ABORT SEQUENCE, and Restart softkeys have the
same function as in the All Test mode.
Single Test
Press is running, press
Chapter 1 31
Single Test to run the test indicated by the pointer. Once the test
Restart to abort and restart the test.
Using Operation Verification Software Volume I
Using Operation Verification
Repeat Sequence
The Repeat Sequence mode performs a user-defined set of tests repeatedly until the sequence is aborted. For example, if the desired sequence is test numbers 6,7,8,6,7,8,6,7,8,…, press
Repeat Sequence
and enter the sequence of 6,7,8. When the last test of this sequence is completed, the sequence is repeated.
The ABORT TEST, ABORT SEQUENCE, and Restart softkeys have the same function as in the All Test mode.
Repeat Test
Use the Repeat Test mode to run a single test indefinitely. Move the pointer to the test to be repeated. Testing can be stopped by pressing
ABORT REPEAT. Press ABORT TEST to abort and restart the test.
Calibrate Power Sensor
The Operation Verification program keeps track of which power sensor is being used and the elapsed time since it was last calibrated.
The program prompts the user to recalibrate the power sensor if more than 2 hours elapses since the last calibration. Also, if the power sensor is changed, the new power sensor must be calibrated.
If there is a significant change in ambient temperature, or improved power meter accuracy is desired, it is advisable to recalibrate the sensor more often than the program requires. Press
Cal Sensor and follow the
instructions on the computer screen to recalibrate the power sensor.
List Equipment
To obtain a list of required test equipment for running a test, move the pointer to the test, press
List Equip. All GPIB controlled equipment and
passive devices, other than required cables and adapters, are listed. If a test is flagged MISSING ETE but all test equipment appears present, press Menu and verify that the equipment is present. Press
List Equip to see what is needed, then return to the Conditions
Cond Menu to
return to the Conditions Menu from the Test Menu.
32 Chapter1
Using Operation Verification Software Volume I
Using Operation Verification
Test Results
At the end of each test, a PASS, SHORT PASS, or MEASUREMENT IS OUT OF TOLERANCE message is printed on the test record and displayed next
to the test on the computer screen. PASS indicates that the test is fully completed and all measurements
are within specification limits. SHORT PASS indicates that the test was abbreviated, usually due to
equipment limitations, but the measurements made were within specification limits. Not all tests can be abbreviated. The meaning of Short Pass varies between each test; refer to “Test Descriptions” in this section for more information.
NOTE A Short Pass is sufficient for passing Operation Verification alone. If
Operation Verification is used as part of performance verification, all tests must yield a PASS result.
MEASUREMENT IS OUT OF TOLERANCE indicates that one or more of the measurements made during the test did not meet specification limits. If the data is shown in tabular form, the symbol <<<< is placed next to the out-of-tolerance data. In the event of a measurement-out-of-tolerance condition, it is recommended that any related manual performance test be performed to verify out-of-tolerance conditions. The related performance test for each operation verification test is listed at the beginning of each test description in this chapter.
NOTE Because test results are expected to change over a period of time,
Agilent Technologies warrants only the specification range and not the repeatability of data for any given specification.
Chapter 1 33
Using Operation Verification Software Volume I
Operation Verification Menu Softkeys
Operation Verification Menu Softkeys
This section provides a brief description of each menu of softkeys. More detailed information is provided in “Program Operation” in this chapter. The softkey order shown below may vary with what appears on the computer display; the order differs depending on whether an 9000 Series 200 or Series 300 computer is used.
Conditions Menu
Test Menu displays the Test Menu, which allows you to run tests
once, repeatedly, or in a user-defined sequence. Refer to “Test Menu.”
Load Conds loads the CONDITIONS file from the disk specified by
the system mass storage file location.
Sensor Utils displays the Sensor Utilities Menu, which allows
viewing, editing, and adding power sensor data files. Refer to “Sensor Utilities Menu.”
Change Entry permits changing a Conditions Menu entry, indicated
by the pointer along the left edge of the computer display. Press the
RETURN or ENTER keys to terminate
an entry.
Verify Bus checks each listed GPIB address for response.
Verify Bus does not verify that a particular piece of
equipment is at a specified address.
Query DUT S/N queries the GPIB for the serial number and model
number of the spectrum analyzer under test.
Store Conds stores the current conditions in the CONDITIONS file
on the specified system mass storage file location.
Exit Program exits the Operation Verification program.
34 Chapter1
Using Operation Verification Software Volume I
Operation Verification Menu Softkeys
Sensor Utilities Menu
View/Edit allows user to view and edit power sensor data files. Add File creates a new power sensor data file. Delete File deletes a power sensor data file. User is asked for
confirmation before deletion takes place.
List Files lists all power sensor data files on the disk currently
specified by the system mass storage file location.
System File allows user to change the currently specified system
mass storage file location.
Cond Menu returns you to the Conditions Menu.
Test Menu
All Tests runs all 16 tests in the order listed by the program. Single Sequence allows entry of a test sequence that is run once. Single Test runs the test indicated by the pointer, once. Repeat Sequence allows entry of a test sequence that runs repeatedly
until you abort testing.
Repeat Test runs the test indicated by the pointer repeatedly until
you abort the testing.
Cal Sensor allows you to recalibrate the current power sensor and
resets the internal “time-since-last-calibration” timer.
List Equip lists the required equipment for the test indicated by
the pointer.
Cond Menu returns to the Conditions Menu.
Chapter 1 35
Using Operation Verification Software Volume I
Test Descriptions
Test Descriptions
Each of the following test descriptions include the related specification, related performance test, and a test setup illustration used in Operation Verification. Operation Verification is designed to test a spectrum analyzer operating within a temperature range of 20 °C to 30 °C.
36 Chapter1
Using Operation Verification Software Volume I
10 MHz Reference Accuracy
10 MHz Reference Accuracy
Related Specification
Frequency Reference Accuracy (Option 103)
Related Performance Test
2. 10 MHz Reference Output Accuracy (Option 103) in Chapter 10 of Volume II of the calibration guide.
Test Description
The frequency of the CAL OUTPUT of the spectrum analyzer is counted by the microwave frequency counter and is compared to the specification. Measuring the CAL OUTPUT signal yields higher resolution than measuring the 10 MHz reference directly. This test applies only to Option 103 spectrum analyzers (the temperature-compensated crystal oscillator option).
NOTE A Short Pass will be indicated if the analyzer is a non-Option 103 but
passes the specification for Option 103. Earlier revisions of the operation verification software measure the
10 MHz REF IN/OUT frequency directly.
Figure 1-1 10 MHz Reference Accuracy Test Setup
Chapter 1 37
Using Operation Verification Software Volume I
1ST LO OUTPUT Amplitude Accuracy
1ST LO OUTPUT Amplitude Accuracy
Related Specification
1ST LO OUTPUT Amplitude
Related Performance Test
1ST LO OUTPUT Amplitude
Test Description
The 1ST LO OUTPUT power is measured with a power meter. For spectrum analyzers without Option 002 (Tracking Generator), the spectrum analyzer is placed in external mixing mode and harmonic-locked to N=6. This allows for the maximum tuning range of the 1ST LO. For spectrum analyzers with Option 002 (8560E/ECs only), internal mixing mode is used. This limits the lowest 1st LO frequency that can be set.
Figure 1-2 1ST LO OUTPUT Amplitude Accuracy Test Setup
38 Chapter1
Using Operation Verification Software Volume I
Calibrator Amplitude Accuracy
Related Specification
CAL OUTPUT Amplitude
Related Performance Test
Calibrator Amplitude and Frequency Accuracy
Test Description
The amplitude of the CAL OUTPUT signal is measured using a power sensor and either the measuring receiver or the power meter. The measured amplitude is compared to the specification.
Figure 1-3 Calibrator Amplitude Accuracy Test Setup
Calibrator Amplitude Accuracy
Chapter 1 39
Using Operation Verification Software Volume I
Displayed Average Noise Level
Displayed Average Noise Level
Related Specification
Displayed Average Noise Level
Related Performance Test
Displayed Average Noise Level
Test Description
The spectrum analyzer INPUT 50 is terminated in 50 Ω. The resolution bandwidth, video bandwidth, and input attenuation are set according to the spectrum analyzer specifications. The displayed average noise level is measured at several points in each band and the results are compared with the specification.
Figure 1-4 Displayed Average Noise Level Test Setup
40 Chapter1
Using Operation Verification Software Volume I
Fast Sweep Time Accuracy
Fast Sweep Time Accuracy
Related Specification
Sweep Time Accuracy (all EC-Series instruments and E-Series instruments with Option 007)
Related Performance Test
Fast Sweep Time Accuracy (all EC-Series instruments and E-Series instruments with Option 007)
Test Description
The sweeptime accuracy for sweeptimes <30 ms, in EC-Series spectrum analyzers and E-Series spectrum analyzers with Option 007, is dependent on the 10 MHz frequency reference absolute accuracy. The test measures the 300 MHz CAL OUTPUT frequency since it is referenced to the 10 MHz reference. Measuring the CAL OUTPUT signal yields higher resolution than measuring the 10 MHz reference directly.
Figure 1-5 Fast Sweep Time Accuracy Test Setup
Chapter 1 41
Using Operation Verification Software Volume I
Frequency Readout/Frequency Counter Accuracy
Frequency Readout/Frequency Counter Accuracy
Related Specifications
Frequency Readout Accuracy Frequency Count Marker Accuracy
Related Performance Test
Frequency Readout Accuracy/ Frequency Count Marker Accuracy
Test Description
The frequency of the synthesized sweeper signal is measured using both the normal marker and the frequency count marker. Both the “frequency readout × frequency reference accuracy” and “marker frequency × frequency reference accuracy” terms of the specification are zero, since the spectrum analyzer provides the frequency reference for the synthesized sweeper. The marker frequencies are compared to the specification.
Figure 1-6 Frequency Readout/Counter Accuracy Test Setup
42 Chapter1
Using Operation Verification Software Volume I
Frequency Response
Frequency Response
Related Specification
Frequency Response
Related Performance Test
Frequency Response
Test Description
The spectrum analyzer frequency response is tested with two setups: the first, using the synthesizer/level generator, for frequencies between the spectrum analyzer minimum frequency and 50 MHz; and the second, using the synthesized sweeper and a measuring receiver or power meter, for frequencies above 50 MHz. If the synthesizer/level generator is not available, the frequency response above 50 MHz can still be tested. If the synthesizer/level generator is available, but you do not wish to perform the test below 50 MHz, enter a Q when prompted to connect the Agilent 3335A output to the spectrum analyzer input.
In both parts of this test, a signal of known amplitude is applied to the input of the spectrum analyzer and the analyzer marker amplitude is read. The frequency response relative to the calibrator frequency (300 MHz), within a given frequency band, is calculated and compared to specification. The band-switching uncertainty specification is verified by calculating the band-to-band frequency response. The band-to-band frequency response specification is equivalent to the sum of the in-band frequency response specifications of the two bands in question and the band-switching uncertainty specification.
While the >50 MHz part of the test is running, a graph of frequency response relative to the CAL OUTPUT signal will be plotted on the computer display. This graph will be dumped to the printer when the test has been completed. If one of the band-to-band frequency response entries is out-of-tolerance, the <<<< symbol will be placed to the right of the row where the out-of-tolerance condition was detected. It will not necessarily be placed directly to the right of the out-of-tolerance entry. Check each entry in that row against the specification (listed in parentheses) to find the entry that is out of tolerance. A Short Pass will occur if the >50 MHz part of the test is within specification but the <50 MHz part of the test was not performed.
Chapter 1 43
Using Operation Verification Software Volume I
Frequency Response
Figure 1-7 Frequency Response Test Setup (<50 MHz)
Figure 1-8 Frequency Response Test Setup (>50 MHz)
44 Chapter1
Using Operation Verification Software Volume I
Frequency Span Accuracy
Frequency Span Accuracy
Related Specification
Frequency Span Accuracy
Related Performance Test
Frequency Span Accuracy
Test Description
Two sources provide two signals of precise frequency separation. One source is a synthesized sweeper, and the second source is the CAL OUTPUT signal. The frequency separation is measured using the spectrum analyzer delta-marker function and compared to the specification. The frequency reference for the synthesized sweeper is provided by the spectrum analyzer.
Figure 1-9 Frequency Span Accuracy Test Setup
Chapter 1 45
Using Operation Verification Software Volume I
IF Gain Uncertainty
IF Gain Uncertainty
Related Specification
IF Gain Uncertainty
Related Performance Test
IF Gain Uncertainty
Test Description
A signal source of known amplitude is connected to the spectrum analyzer and an amplitude reference is set. The signal source amplitude is stepped down as the spectrum analyzer is stepped down, and the signal amplitude is measured at each point. The amplitude variation with respect to the reference is compared to the specification. The test is performed in 1 dB steps from 0 dBm to 12 dBm reference levels, and in 10 dB steps from 0 dBm to 80 dBm reference levels. The 10 dB steps are tested in both log and linear scale factors.
Figure 1-10 IF Gain Uncertainty Test Setup
46 Chapter1
Using Operation Verification Software Volume I
Image and Multiple Responses
Image and Multiple Responses
Related Specifications
Image and Multiple Responses (8560E/EC) Image, Multiple, and Out-of-Band Responses (Agilent 8561E/EC,Agilent 8562E/EC, Agilent 8563E/EC, Agilent 8564E/EC, and Agilent 8565E/EC) Out-of-Range Responses (8560E/EC and Agilent 8561E/EC)
Related Performance Test
Image, Multiple, and Out-of-Range Responses (8560E/EC) Image, Multiple, Out-of-Band, and Out-of-Range Responses (Agilent 8561E/EC) Image, Multiple, and Out-of-Band Responses (Agilent 8562E/EC,Agilent 8563E/EC,Agilent 8564E/EC, and Agilent 8565E/EC)
Test Description
Image, multiple, out-of-band, and out-of-range responses are tested, as applicable, in each frequency band. A signal is applied to the signal analyzer input, then a reference amplitude measurement is made. The signal source is then tuned to a frequency that causes either an image, multiple, out-of-band, or out-of-range response. The amplitude displayed on the spectrum analyzer is measured and the difference between this measurement and the reference amplitude measurement is calculated.
Figure 1-11 Image, Multiple, Out-of-Band, and Out-of-Range Responses Test
Setup
Chapter 1 47
Using Operation Verification Software Volume I
Input Attenuator Switching Uncertainty
Input Attenuator Switching Uncertainty
Related Specification
Input Attenuator Switching Uncertainty
Related Performance Test
Input Attenuator Accuracy
Description
The output of theAgilent 3335A is applied to the input of the spectrum analyzer, and an amplitude reference is set. The spectrum analyzer IF gain uncertainty is characterized using the Agilent 3335A as the reference. The 3335A is then reset to a fixed amplitude and the input attenuator is stepped from 10 dB to 70 dB (10 to 60 dB for the Agilent 8564E/EC and Agilent 8565E/EC). At each step, the amplitude deviation from the reference is measured using the marker functions. The input attenuator accuracy is calculated from the marker value and the characterized IF gain uncertainty. The input attenuator accuracy then is compared to the specification.
Figure 1-12 Input Attenuator Accuracy Test Setup
48 Chapter1
Using Operation Verification Software Volume I
Noise Sidebands
Noise Sidebands
Related Specification
Noise Sidebands
Related Performance Test
Noise Sidebands
Test Description
A clean signal source is applied to the input of the spectrum analyzer and the noise level at frequency offsets above and below the carrier are measured. These sideband levels are compared to the specification.
Noise sidebands measured at offsets of 1 kHz and less using the CAL OUTPUT signal will be up to 6 dB better than when measured using the synthesized signal generator. This difference is due to phase-coherency of the CAL OUTPUT signal and the internal local oscillators. Noise sidebands measured at offsets of 10 kHz and greater will be the same if the CAL OUTPUT signal or the synthesized signal generator is used (the signal generator frequency is set to 300 MHz).
NOTE Test results will be invalid if the source and the spectrum analyzer use
the same frequency reference.
Figure 1-13 Noise Sidebands Test Setup
Chapter 1 49
Using Operation Verification Software Volume I
RES BW Accuracy and Selectivity
RES BW Accuracy and Selectivity
Related Specifications
Resolution Bandwidth Accuracy Resolution Bandwidth Selectivity
Related Performance Test
Resolution Bandwidth Accuracy and Selectivity
Description
The output of a synthesizer/level-generator is connected to the input of the spectrum analyzer. The spectrum analyzer is set to a span approximately twice the resolution bandwidth setting (for measuring the 3 dB bandwidth).
The synthesizer/level-generator output is then reduced in amplitude by 3 dB. A marker reference is set and the synthesizer/level-generator output is increased 3 dB to its previous level. A sweep is taken, then the markers are used to measure the 3 dB bandwidth.
The 60 dB bandwidths are measured in a similar manner, with the span set about 15 to 20 times the resolution bandwidth setting. The ratio between the 60 dB and 3 dB bandwidths are calculated and stored.
RES BW settings less than 300 Hz are not measured. These bandwidths are digitally derived; therefore, their accuracy and shape factors are guaranteed by design.
Figure 1-14 RES BW Accuracy and Selectivity Test Setup
50 Chapter1
Using Operation Verification Software Volume I
RES BW Switching and IF Alignment Uncertainty
RES BW Switching and IF Alignment Uncertainty
Related Specifications
Resolution Bandwidth Switching Uncertainty IF Alignment Uncertainty
Related Performance Test
Resolution Bandwidth Switching and IF Alignment Uncertainty
Description
A signal is applied to the input of the spectrum analyzer and the signal amplitude is measured in each resolution bandwidth setting. The amplitude variation with respect to the 300 kHz resolution bandwidth is calculated and compared to the specifications.
Figure 1-15 RES BW Switching Uncertainty Test Setup
Chapter 1 51
Using Operation Verification Software Volume I
Residual FM
Residual FM
Related Specification
Residual FM
Related Performance Test
Residual FM
Test Description
A clean signal source is connected to the spectrum analyzer, and the analyzer resolution bandwidth is set to 300 Hz. The slope of the signal is measured for use in calculating the residual FM. The source is tuned to the middle of the slope just measured with the analyzer in zero span. The trace is read into the controller and a 10 Hz RES BW is simulated by executing a 50 ms moving average on the trace data. Only 20 ms worth of averaged data is examined to simulate the 20 ms measurement window.
Figure 1-16 Residual FM Test Setup
52 Chapter1
Using Operation Verification Software Volume I
Scale Fidelity
Scale Fidelity
Related Specification
Scale Fidelity
Related Performance Test
Scale Fidelity
Test Description
A signal source of known amplitude is connected to the spectrum analyzer and the source amplitude is adjusted for a top-screen reference. The source amplitude is stepped down and the displayed amplitude is measured at each step. The scale fidelity is tested in 2 dB steps in 2 dB/division and linear, and in 10 dB steps in 10 dB/division.
The amplitude variation with respect to the reference is measured and compared to the specification. In log mode, the amplitude difference between adjacent steps is calculated and compared to the specification.
Figure 1-17 Scale Fidelity Test Setup
Chapter 1 53
Using Operation Verification Software Volume I
Second Harmonic Distortion
Second Harmonic Distortion
Related Specification
Second Harmonic Distortion
Related Performance Test
Second Harmonic Distortion
Test Description
This test consists of two parts: a low-band distortion test and a high-band distortion test. The high-band distortion test is only applicable to the Agilent 8561E/EC, Agilent 8562E/EC, Agilent 8563E/EC, Agilent 8564E/EC, and Agilent 8565E/EC. The low-band distortion test can be performed using either the frequency synthesizer or the synthesized sweeper. The high-band distortion test can only be performed using a synthesized sweeper. After the low-band distortion test has been completed, if a synthesized sweeper is available, the operator may choose whether or not to perform the high-band distortion test.
Before making the second harmonic distortion measurement, the filters are checked for sufficient rejection at the second harmonic. A warning message will be displayed if the filter has insufficient rejection. If the filter is acceptable, the test will proceed. The test is performed at 50 MHz for low band and at 2.95 GHz for high band (these are the fundamental frequencies).
Before checking the second harmonic distortion in high band, a frequency response check is made to reduce the measurement uncertainty due to the spectrum analyzer frequency response. Two filters are necessary for the high-band distortion test to ensure sufficient rejection at the second harmonic.
For an Agilent 8561E/EC, Agilent 8562E/EC,Agilent 8563E/EC, Agilent 8564E/EC, or Agilent 8565E/EC a Short Pass occurs if the low-band distortion test is within specification, and the high-band test is not performed. Agilent 8562E/EC, Agilent 8563E/EC, Agilent 8564E/EC, or Agilent 8565E/EC analyzers having software revisions D.00.00 and later will report a short pass even if both the low-band and high-band distortion tests are within specification. These analyzers specify distortion in three frequency ranges, but the software tests only two of these ranges.
54 Chapter1
Using Operation Verification Software Volume I
Second Harmonic Distortion
Figure 1-18 Low-Band Second Harmonic Distortion Test Setup
Figure 1-19 High-Band Second Harmonic Distortion Test Setup
Chapter 1 55
Using Operation Verification Software Volume I
Operation Verification Error Messages
Operation Verification Error Messages
Operation Verification displays prompts and error messages on the computer display. Error messages are preceded with ERROR:. For more information on prompts, refer to “Program Operation” in this chapter.
The error messages listed below are in three groups: messages beginning with alphabetic characters, those beginning with numeric characters, and others beginning with variables such as filenames or instrument model numbers. If an error message is not found in either of the first two groups, it probably begins with a variable. Refer to the third group that begins with variables. The error message descriptions include recommended corrective action.
Error Messages Beginning with Alphabetic Characters
Address must be from 0 to 30 inclusive
GPIB addresses must be in the range from 0 to 30. Press any key and enter new address in this range.
Cal Factor outside of 0 to 150% range entered
Cal Factorentries must be within this range. Press any key and enter the frequency again. Then enter a Cal Factor in the proper range.
CONDITIONS file from system file location is not compatible! WARNING
CONDITIONS files from versions prior to C.00.00 did not include an entry for the 8487A power sensor. Similarly, CONDITIONS files from versions prior to D.00.00 did not include an entry for the 5350 Series microwave frequency counter. Default CONDITIONS data will be used. Modify the CONDITIONS data as necessary and press
Conditions Menu DUT ID disagrees with responding DUT ID
The model and/or serial number of the spectrum analyzer under test (DUT) listed in the “Conditions
Menu,”on page 23, does not agree with that of the DUT
which is responding over GPIB. Press any key and follow the instructions in the next three prompts.
STORE CONDS.
56 Chapter1
Using Operation Verification Software Volume I
Operation Verification Error Messages
Could not set calibrator signal to 10.00 or 10.17 dBm
While attempting the noise sidebands test using the CAL OUTPUT signal, the amplitude of the CAL OUTPUT signal could not be set to one of the values indicated. Check the CAL OUTPUT amplitude and the range of the reference level calibration adjustment.
Counter reads <frequency value> Hz. Check counter setup
The microwave frequency counter read a frequency far exceeding the specification of the 10 MHz reference. Check the test setup and press any key. The counter will read the frequency once more and assume that the value is correct.
Data not accepted, check entry format
The data just entered was not valid. Press any key and try again, checking for the proper entry format.
DUT doesn't respond at address listed
Program attempted to address the spectrum analyzer under test at the address listed, but the spectrum analyzer did not respond. Check the GPIB connections and the address listed.
File <filename> not found
The filename of the power sensor data file entered could not be found on the currently specified system mass storage file location. Check the filename and the system mass storage file location.
Insufficient equip. to do test <test number> : <test name>
The required GPIB-controlled test equipment for the test indicated is not available. Press any key and choose another test.
Low-pass filter(s) don't have at least <value> dB rejection
The low-pass filters are checked for rejection at the second harmonic in the Second Harmonic Distortion test. If insufficient rejection is detected, the part of the test using the tested filter cannot be run.
Maximum cal amplitude <10 dBm
Program checked that CAL OUTPUT was connected to INPUT 50 and found that the REF LVL CAL adjustment could not be set for a marker amplitude of ≥−10 dBm. Check CAL OUTPUT amplitude and REF LVL CAL adjustment range.
Chapter 1 57
Using Operation Verification Software Volume I
Operation Verification Error Messages
No GPIB address listed for DUT
Program attempted to address the spectrum analyzer (DUT), but no GPIB address was listed for it. Press any key and enter an address for the spectrum analyzer.
No more tests may be linked; enter 0 at next prompt
When entering a sequence of tests, the sequence string (including commas) cannot exceed 78 characters. Press any key and then a 0 at the next prompt. The testing sequence will begin.
Non-numeric entry other than S entered, or frequency <=0
When entering a frequency of a frequency/Cal-Factor pair to be added, edited, or deleted, the entry must either be a number greater than 0, or S to store the current data.
No sensor file found for <sensor model> S/N <sensor serial #>
A power sensor data file for the indicated power sensor could not be found on the currently specified system mass storage file location specifier. Check the sensor model and serial numbers and the system mass storage file location.
No 8662/63 reference oscillator. Check INT-EXT switches
No 10 MHz reference oscillator for the 8662A/63A was detected. Check the INTrnal/EXTernal frequency reference switches on its rear panel. The 8662A/63A should be using its internal reference oscillator.
Power meter reads <value> dBm
The power meter has read a value far exceeding the specification of the CAL OUTPUT amplitude; check that power sensor is connected to CAL OUTPUT and press any key. The power will be read once more and assumed to be valid.
Printer not available; cannot perform tests
All test results are sent to the printer. If a printer is not available, tests cannot be performed.
REF LVL CAL adjustment range <5 dB
In checking that the CAL OUTPUT was connected to INPUT 50 Ω, the REF LVL CAL adjustment was found to have insufficient range. Check REF LVL CAL range manually.
Select code <value> does not currently support GPIB
58 Chapter1
Using Operation Verification Software Volume I
Operation Verification Error Messages
operations
The address just entered specified a select code which is not a GPIB interface. Check the address entered and the select code of the appropriate interface.
Sensor serial number must be from 1 to 99999
The power sensor serial number entered was not in the range indicated. Enter the serial number correctly.
System mass storage file location catalog cannot be read
Program attempted to read the catalog of the system mass storage file location. Check the msus of the system mass storage file location.
Test number must be between 0 and 17
Valid test numbers are in the range 0 and 17 for entering a sequence. Entering a 0 will terminate sequence entry and begin testing sequence. Press any key and enter a valid test number at the next prompt.
This program does not support the current DUT
The operation verification software supports only the 8560 E-Series and EC-Series. Refer to the Spectrum Analyzers section in Getting Started.
Unable to load CONDITIONS file from listed system file location
Program attempted to load the CONDITIONS file from the listed system mass storage file location. Check the msus of the System mass storage file location and the presence of the CONDITIONS file.
Unable to load data from <sensor filename>
Program found power sensor data file, but could not read the data from the file. Use the Sensor Utilities to delete the file and enter new data.
Unable to obtain catalog from <system mass storage file location>
The system could not verify that the system mass storage file location entered was available. Check the msus of the system mass storage file location.
Unable to reach power level of <value> dBm
Program was unable to set the source amplitude for a desired power meter reading. Check the test setup.
WARNING: CONDITIONS file from system file location is not
Chapter 1 59
Using Operation Verification Software Volume I
Operation Verification Error Messages
compatible!
CONDITIONS files from versions prior to C.00.00 did not include an entry for the 8487A power sensor. Similarly, CONDITIONS files from versions prior to D.00.00 did not include an entry for the 5350 Series microwave frequency counter. Default CONDITIONS data will be used. Modify the CONDITIONS data as necessary and press
STORE CONDS.
Error Messages Beginning with Numeric Characters
8481A Sensor cal data minimum frequency not <= 50 MHz
Program requires the 8481A power sensor to have a Cal Factor at or below 50 MHz. Use Sensor Utilities to add a Cal Factor at or below 50 MHz.
8481A Sensor cal data maximum frequency not >= 300 MHz
Program requires the 8481A power sensor to have a Cal Factorat or above 300 MHz. Use Sensor Utilities to add a Cal Factor at or above 300 MHz.
8482A Sensor cal data minimum frequency not <= 50 MHz
Program requires the 8482A power sensor to have a Cal Factor at or below 50 MHz. Use Sensor Utilities to add a Cal Factor at or below 50 MHz.
8482A Sensor cal data maximum frequency not >= 300 MHz
Program requires the 8482A power sensor to have a Cal Factorat or above 300 MHz. Use Sensor Utilities to add a Cal Factor at or above 300 MHz.
8485A Sensor cal data minimum frequency not = 50 MHz
Program requires the 8485A power sensor to have a Cal Factorat 50 MHz. Use Sensor Utilities to add a Cal Factor at 50 MHz.
8485A Sensor cal data maximum frequency not >= 26.5 GHz
Program requires the 8485A power sensor to have a Cal Factor at or above 26.5 GHz. Use Sensor Utilities to add a Cal Factor at or above 26.5 GHz.
8487A Sensor cal data minimum frequency not = 50 MHz
Program requires the 8487A power sensor to have a Cal Factorat 50 MHz. Use Sensor Utilities to add a Cal Factor at 50 MHz.
60 Chapter1
Using Operation Verification Software Volume I
Operation Verification Error Messages
8487A Sensor cal data maximum frequency not >= 50 GHz
Program requires the 8487A power sensor to have a Cal Factor at or above 50 GHz. Use Sensor Utilities to add a Cal Factor at or above 50 GHz.
8662/63 Error # <error number>
The 8662A/63A generated the error listed. Consult the
8662A or 8663A manual.
8662/63 Frequency reference out of tolerance
The 8662A/63A frequency reference is out of tolerance. Consult the 8662A or 8663A manual.
8662/63 Malfunction. Origin unknown
The 8662A/63A has detected a malfunction. Consult the 8662A or 8663A manual.
8662/63 Oven not yet warmed up
The 8662A/63A 10 MHz oven oscillator is cold. Allow the oven to warm up.
8662/63 Should be on INTernal reference
The 8662A/63A is in EXTernal frequency reference mode. Set the 8662A/63A to INTernal frequency reference.
Error Messages Beginning with Variables
<filename> file not found
The file indicated could not be found at the listed system mass storage file location. Check the filename and the system mass storage file location.
<keyboard entry> is a non-numeric entry
The program expected a numeric entry but did not receive one. Enter a numeric entry.
<number of instruments> instruments have GPIB addresses of <GPIB address>
The indicated number of instruments have all been set to the same GPIB address. Review the addresses and eliminate the duplication.
Chapter 1 61
Using Operation Verification Software Volume I
Operation Verification Error Messages
<power meter model number> doesn't read signal to be in 1 +/5 dBm range
The power meter (or measuring receiver) does not read a power level within the range indicated. Check for loose connections.
<source model number> signal not in +5 +/5 dBm range
The source indicated was set for +5 dBm output, but the spectrum analyzer measured the amplitude to be outside the ±5 dB range. Check test setup.
<source model number> signal not in 1 +/5 dBm range
The source indicated was set for + 5 dBm output and the source output is fed through a power splitter to the spectrum analyzer under test. The spectrum analyzer should measure the amplitude to be within 5 dB of
1 dBm (6 dB loss through power splitter). Check test setup.
<source model number> +10 dBm signal not in +10 +/8 dBm range
The source indicated was set for a +10 dBm output and the source output is fed through a low-pass filter(s) to the spectrum analyzer under test. The spectrum analyzer should measure the amplitude to be within 8 dB of +10 dBm (the filters have some insertion loss). Check test setup.
<source model number> has a cold oven
The 10 MHz reference oven oscillator has not warmed up yet. Allow the oven to warm up.
<source model number> is unlevelled
The source indicated has been programmed for an amplitude which results in an unlevelled condition. Check the test setup for loose connections.
62 Chapter1
2 8560E/EC Specifications and
Characteristics
63
8560E/EC Specifications and Characteristics
Specifications and Characteristics
Specifications and Characteristics
The following tables list the spectrum analyzer specifications. Unless stated otherwise, all specifications describe the analyzer warranted performance under the following conditions:
• Five-minute warmup in ambient conditions
• Auto-coupled controls
• Digital trace display
• IF ADJ ON
• REF LVL CAL adjusted
• 1ST LO OUTPUT terminated in 50 ohms
• 2ND IF OUTPUT (Option 001 analyzers) terminated in 50 ohms
• Two-year calibration cycle
• Environmental requirements met
NOTE REF LVL CAL uses the CAL OUTPUT signal to calibrate the reference
level. Internal temperature changes determine how often this adjustment should be performed. Amplitude temperature drift is a nominal 1 dB/10 °C. The nominal temperature variation within the instrument is 10 °C, most of which occurs during the first 30 minutes after power-on. Internal temperature equilibrium is reached after 2 hours of operation at a stable ambient temperature.
Characteristics provide useful information in the form of typical, nominal, or approximate values for analyzer performance. Tables of spectrum analyzer characteristics follow the specifications.
Calibration Cycle
The performance tests in Chapter 2 should be used to check the analyzer against its specifications every two years. Specifications are listed in this chapter.
The frequency reference must be adjusted and checked at the same time. Refer to the “10 MHz Frequency Reference Adjustment” in the 8560E and 8560EC Spectrum Analyzer Service Guide.
64 Chapter2
Frequency Specifications
Frequency Range
Internal Mixing AC Coupled 100 kHz to 2.9 GHz DC Coupled 30 Hz to 2.9 GHz
External Mixing 18 GHz to 325 GHz
Non-Option 002 and Non-Option 327
External Mixing Bands
Frequency Band Frequency Range Harmonic Mixing Mode (N*)
8560E/EC Specifications and Characteristics
Frequency Specifications
Preselected Unpreselected
K 18.0 to 26.5 n/a 6 A 26.5 to 40.0 8+ 8 Q 33.0 to 50.0 10+ 10 U 40.0 to 60.0 10+ 10 V 50.0 to 75.0 14+ 14 E 60.0 to 90.0 n/a 16
W 75.0 to 110.0 18+ 18
F 90.0 to 140.0 n/a 24 D 110.0 to 170.0 n/a 30 G 140.0 to 220.0 n/a 36 Y 170.0 to 260.0 n/a 44
J 220.0 to 325.0 n/a 54
* N is the harmonic mixing mode. Fornegative mixing modes (as indicated by the “”), the desired 1st LO harmonic is higher than the tuned frequency by the 1st IF (3.9107 GHz for the 30 Hz to
2.9 GHz band, 310.7 MHz for all other bands). For positive mixing modes, the desired 1st LO harmonic is lower than the tuned frequency by 310.7 MHz.
Chapter 2 65
8560E/EC Specifications and Characteristics
Frequency Specifications
Frequency Readout Accuracy
Accuracy of START, CENTER, STOP or MARKER frequency
SPAN > 2 MHz × N*
<±(frequency readout × frequency reference accuracy†+5% of frequency span + 15% of resolution bandwidth + 10 Hz)
SPAN 2 MHz × N*
<±(frequency readout × frequency reference accuracy†+1% of frequency span + 15% of resolution bandwidth + 10 Hz)
* N is the harmonic mixing mode.
frequency reference accuracy = (aging × period of time since adjustment) + initial achievable
accuracy + temperature stability.
Frequency Count Marker
Frequency Count Marker Resolution Selectable from 1 Hz to 1 MHz
Frequency Count Marker Accuracy (for signal-to-noise ratio 25 dB)
Delta Frequency Count Accuracy (for signal-to-noise ratio 25 dB)
<±(marker frequency × frequency reference accuracy‡ + 2 Hz × N† + 1 LSD)
<±(delta frequency × frequency reference accuracy‡ + 4 Hz × N† + 2 LSD)
† N is the harmonic mixing mode. ‡ Frequency Reference Accuracy = (aging × period of time since adjustment + initial achievable
accuracy + temperature stability).
Frequency Reference Accuracy
Non-Option 103
Aging
<±0.5 × 10−9/day (after 7 day warmup) <±1 × 10−7/year
Temperature Stability
<±1 × 10−8, 10 °C to +55 °C, referenced to 25 °C
Option 103
Aging Settability Temperature Stability
<±2 × 10−6/year <±1 × 10
6
<±1 × 10−6, 10 °C to +55 °C, referenced to 25 °C
66 Chapter2
Stability
Residual FM
(zero span, 10 Hz RES BW)
8560E/EC Specifications and Characteristics
Frequency Specifications
Non-Option 103
< 1.0 Hz × N* peak-to-peak in 20 ms
Option 103 < 10 Hz × N* peak-to-peak in 20 ms
Noise Sidebands
For Frequencies 1 GHz
(Refer to the characteristics section for frequencies > 2.9 GHz)
Offset Non-Option 103 Option 103
100 Hz
serial number prefix <3424A serial number prefix 3424A
1 kHz
10 kHz 30 kHz
††
††
<80 dBc/Hz <88 dBc/Hz
<97 dBc/Hz <113 dBc/Hz <113 dBc/Hz
§
#
<70 dBc/Hz <70 dBc/Hz
<90 dBc/Hz <113 dBc/Hz <113 dBc/Hz
§
#
100 kHz
‡‡
serial number prefix <3424A
<113 dBc/Hz
§
<113 dBc/Hz
§
serial number prefix 3424A <117 dBc/Hz** <117 dBc/Hz**
* N is the harmonic mixing mode.
See Resolution Bandwidth Usability in “Frequency Characteristics” for further information.
Add 5.2 dB × (f/1 GHz) 1) for f > 1 GHz and f 2.9 GHz.
§
Add 2.5 dB × (f/1 GHz) 1) for f > 1 GHz and f 2.9 GHz.
#
Add 3.0 dB × (f/1 GHz) 1) for f > 1 GHz and f 2.9 GHz.
**Add 2.0 dB for frequencies > 1 GHz and 2.9 GHz
††
For resolution bandwidth 1 kHz or frequency span 745 kHz.
‡‡
For resolution bandwidth 3 kHz or frequency span > 745 kHz.
Chapter 2 67
8560E/EC Specifications and Characteristics
Frequency Specifications
Frequency Span
Range
Internal Mixing 0 Hz, 100 Hz to 2.9 GHz over the 10-division display horizontal
axis, variable in approximately 1% increments, or in a 1, 2, 5 sequence.
External Mixing
Minimum span = 100 Hz × N*
Accuracy
SPAN > 2 MHz × N* <±5% SPAN 2 MHz × N* <±1%
* N is the harmonic mixing mode. † Resolution bandwidths 100 Hz are not available in external mixing. External mixing is not
available for Option 002 or Option 327.
68 Chapter2
8560E/EC Specifications and Characteristics
Frequency Specifications
Resolution Bandwidths (3 dB)
Range*
Non-Option 103 1 Hz to 1 MHz (selectable in a 1, 3, 10 sequence) and
2 MHz (3 MHz at 6 dB)
Option 103 10 Hz to 1 MHz (selectable in a 1, 3, 10 sequence) and
2 MHz (3 MHz at 6 dB)
Accuracy
1 Hz to 300 kHz RES BW <±10% 1 MHz RES BW <±25% 2 MHz RES BW <+50%, 25%
Selectivity (60 dB/3 dB bandwidth
ratio)
RES BW 300 Hz <15:1 RES BW 100 Hz <5:1
Bandwidth Shape
1 and 2 MHz RES BW Approximately Gaussian 300 Hz to 300 kHz RES BW Synchronously tuned, 4-pole filters 1 Hz to 100 Hz RES BW Digital, approximately Gaussian
* Resolution bandwidths 100 Hz are not available in external mixing.
Video Bandwidth
(Post-detection low-pass filter averages displayed
noise for a smooth trace.)*
Range
1 Hz to 3 MHz† in a 1, 3, 10 sequence
* Video bandwidth filtering is not available in resolution bandwidths 100 Hz when SPAN 0 Hz
with firmware revisions 930809 and earlier.
The video bandwidth upper limit is 450 kHz in sample detection mode.
Chapter 2 69
8560E/EC Specifications and Characteristics
Frequency Specifications
Sweep
Sweep Time
Range
Span = 0
Non-Option 007 (E-Series)
Analog display 50 µs to <30 ms Digital display 30 ms to 6,000 s*
Option 007 E-Series and all EC-Series instruments
Digital display 50 µs to 6,000 s
Span 100 Hz
RES BW 300 Hz
50 ms to 2,000 s
RES BW 100 Hz 50 ms to 100,000 s (span-dependent)
Accuracy (Span = 0 Hz)
Non-Option 007 (E-Series)
Sweep time 30 ms to 6,000 s* <±1% Sweep time <30 ms <±10%
Option 007 E-Series and all EC-Series instruments
Sweep time 30 ms to 6,000 s* <±1% Sweep time <30 ms <±0.1%
Sweep Trigger
Delayed, Free Run, Single, Line, External, or Video
* 30 ms to 100 s for analyzers with serial prefix <3310A.
50 ms to 100 s for analyzers with serial prefix <3424A.
# Video trigger is not available in RES BW settings 100 Hz.
#
70 Chapter2
8560E/EC Specifications and Characteristics
Delayed Sweep
Trigger Modes Free Run, Line, External, Video* Range
Span = 0
Non-Option 007 (E-Series) +2 µs to +65.535 ms Option 007 E-Series and all EC-Series
instruments
Sweep time 30 ms +2 µs to +65.535 ms
Frequency Specifications
Sweep time < 30 ms
9.9 ms to +65.535 ms
††
Span 100 Hz +2 µs to +65.535 ms
Resolution 1 µs Accuracy
#
Serial prefix 3310A and above ±1 µs Serial prefix <3310A
20 °C to 30 °C ±(1 µs + (0.05% × DELAY SWEEP setting)
10 °C to +55 °C ±(1 µs + (0.12% × DELAY SWEEP setting)
* Video trigger is not available in RES BW settings 100 Hz.
††
Negative delayed sweep (pre-trigger) is also limited to 50 × sweep time.
#
In E-Series instruments without Option 007, the delay time will be subject to up to ±0.5 µs of jitter due to synchronization of the input edge trigger to the internal 1 MHz timebase. In E-series instruments with Option 007 and in EC-instruments using sweeptimes <30 ms, the delay time will experience up to ±83 ns of jitter due to synchronization of the input edge trigger to the internal 12 MHz timebase.
Chapter 2 71
8560E/EC Specifications and Characteristics
Frequency Specifications
Time-Gated Spectrum Analysis
Gate Delay* Edge Mode Level Mode
Range 3 µs to 65.535 ms 0.5 µs Resolution 1 µs
Accuracy
Serial prefix 3310A and above <±1 µs Serial prefix <3310A
20 °C to 30 °C ± (1 µs + (0.05% × GATE DELAY setting)
10 °C to 55 °C ± (1 µs + (0.12% × GATE DELAY setting)
* Time from GATE TRIGGER INPUT to positive edge of GATE OUTPUT.
The gate delay time will experience up to ±0.5 µs of jitter due to synchronization of the input
edge trigger to the internal 1 MHz timebase.
Gate Length*
Range 1 µs to 65.535 ms Resolution 1 µs Accuracy
Serial prefix 3310A and above <±1 µs Serial prefix <3310A
20 °C to 30 °C ± (0.2 µs + (0.05% × GATE LENGTH setting)
10 °C to 55 °C ± (0.2 µs + (0.12% × GATE LENGTH setting)
* Time from positive edge to negative edge of GATE OUTPUT.
Marker Frequency Resolution SPAN/600 to a minimum of 1 Hz
Marker Time Resolution Sweep time/600
72 Chapter2
Amplitude Specifications
Measurement Range
Maximum Safe Input Power
8560E/EC Specifications and Characteristics
Amplitude Specifications
Average Continuous Power
(input attenuation 10 dB)
Peak Pulse Power
(input attenuation 30 dB)
DC Voltage
AC Coupled <±50 V DC Coupled <±0.2 V
Gain Compression
10 MHz to 2.9 GHz <1.0 dB (≤−5 dBm at input mixer‡)
Mixer level = input level input attenuation.
+30 dBm (1 W)
+50 dBm (100 W) for pulse widths 10 µs and <1% duty cycle.
Chapter 2 73
8560E/EC Specifications and Characteristics
Amplitude Specifications
Displayed Average Noise Level
With no signal at input, 1 Hz video bandwidth, and 0 dB input attenuation, tracking generator off.
Frequency Range 10 Hz RES BW 1 Hz RES BW
(Option 103) (Non-Option 103)
30 Hz <80 dBm <90 dBm 100 Hz <80 dBm <90 dBm 1 kHz <95 dBm <105 dBm 10 kHz <110 dBm <120 dBm 100 kHz <110 dBm <120 dBm 1 MHz to 10 MHz <130 dBm <140 dBm 10 MHz to 2.9 GHz
serial number prefix <3632A <135 dBm <145 dBm serial number prefix 3632A <141 dBm <151 dBm
Option H13 <141 dBm <151 dBm
74 Chapter2
Spurious Responses
8560E/EC Specifications and Characteristics
Amplitude Specifications
All input-related spurious responses, except as noted
Mixer Level* Distortion
below.
10 MHz to 2.9 GHz 40 dBm <75 dBc
Second Harmonic Distortion
(40 dBm mixer level*)
Applied Signal Frequency Range
1 MHz to 1.45 GHz (serial number prefix <3632A) 40 dBm <72 dBc 20 MHz to 1.45 GHz (serial number prefix
40 dBm <79 dBc
3632A)
20 MHz to 1.45 GHz (Option H13) 40 dBm <79 dBc
Third Order Intermodulation Distortion
(with two signals at input mixer, spaced 1 kHz apart)
Frequency Range
1 MHz to 2.9 GHz (serial number prefix <3632A) 30 dBm each <78 dBc 20 MHz to 2.9 GHz (serial number prefix 3632A) 30 dBm each <82 dBc 20 MHz to 2.9 GHz (Option H13) 30 dBm each <82 dBc
Image and Multiple Responses
Frequency Range
10 MHz to 2.9 GHz 10 dBm <80 dBc
Out of Range Responses
(for input signals 2.9 GHz to 12 GHz)
Frequency Range
10 MHz to 2.9 GHz 10 dBm <80 dBc
* Mixer Level = input level input attenuation
Residual Responses
>200 kHz with no signal at input, 0 dB input attenuation, N† 1
N harmonic mixing number
<90 dBm
Chapter 2 75
8560E/EC Specifications and Characteristics
Amplitude Specifications
Display Range
Amplitude Scale 10 vertical display divisions, with the reference level (0 dB) at
the top graticule line.
Calibration
LOG 10 dB/DIV for 100 dB display from reference level.*
5 dB/DIV for 50 dB display expanded from reference level. 2 dB/DIV for 20 dB display expanded from reference level.
1 dB/DIV for 10 dB display expanded from reference level.
LINEAR 10% of reference level per division over the top nine divisions
(all 10 divisions for RES BW 100 Hz) when calibrated in voltage.
* 10 dB/DIV for 70 dB display from reference level for RES BW 100 Hz when SPAN = 0 Hz.
In E-Series instruments these scales are not available for sweep times < 30 ms without Option
007.
Accuracy
Reference Level Range
LOG, adjustable in 0.1 dB steps 120 dBm to +30 dBm LINEAR, settable in 1% steps 2.2 µV to 7.07 V
76 Chapter2
8560E/EC Specifications and Characteristics
Amplitude Specifications
Reference Level Uncertainty
Frequency Response
(with 10 dB input attenuation) Typical
Relative (referenced to midpoint between highest and
lowest peak excursions)
DC Coupled, 30 Hz to 2.9 GHz <±1.0 dB <±0.8 dB
DC Coupled, 100 MHz to 2.9 GHz (serial number prefix
3632A)
AC Coupled, 100 kHz to 2.9 GHz <±1.4 dB <±0.9 dB
Absolute (referenced to 300 MHz CAL OUTPUT)
DC Coupled, 30 Hz to 2.9 GHz <±1.5 dB <±1.0 dB AC Coupled, 100 kHz to 2.9 GHz <±1.7 dB <±1.1 dB
Calibrator Uncertainty
10 dBm, 300 MHz <±0.3 dB
Input Attenuator Switching Uncertainty
(20 to 70 dB settings, referenced to 10 dB input attenuation)
30 Hz to 2.9 GHz <±0.6 dB/10 dB step, 1.8 dB max.
<±0.7 dB <±0.7 dB
(20 °C to 30 °C)
IF Gain Uncertainty
(0 dBm to 80 dBm reference levels with 10 dB input attenuation) <±1.0 dB
Resolution Bandwidth Switching Uncertainty
(Referenced to 300 kHz resolution bandwidth at the reference level.)* <±0.5 dB * Scale fidelity is not the same for RES BW 100 Hz as for RES BW 300 Hz. Therefore, signals
not at the reference level will experience an additional amplitude difference when switching between these two sets of RES BW settings, due to differences in scale fidelity.
Chapter 2 77
8560E/EC Specifications and Characteristics
Amplitude Specifications
Pulse Digitization Uncertainty
(Pulse response mode, PRF >720/sweep time)
LOG
Resolution Bandwidth 1 MHz <1.25 dB peak-to-peak Resolution Bandwidth = 2 MHz <3 dB peak-to-peak
LINEAR
Resolution Bandwidth 1 MHz <4% of reference level peak-to-peak Resolution Bandwidth = 2 MHz <12% of reference level peak-to-peak
IF Alignment Uncertainty
(additional uncertainty when using 300 Hz RES BW only) <±0.5 dB
78 Chapter2
Scale Fidelity*
LOG
Incremental
8560E/EC Specifications and Characteristics
Amplitude Specifications
0 to 90 dB range
RES BW 300 Hz <±0.1 dB/dB RES BW 100 Hz <±0.2 dB/2 dB
Cumulative
0 to 90 dB range
RES BW 300 Hz <±0.1 dB/dB from the reference level to a maximum of ±0.85 dB RES BW 100 Hz <±0.2 dB/2 dB from the reference level to a maximum of ±0.85 dB
0 to 100 dB range
RES BW 300 Hz ±2.5 dB characteristic RES BW 100 Hz maximum of ±1.5 dB
LINEAR <±3% of reference level
* Scale fidelity is not the same for RES BW 100 Hz as for RES BW 300 Hz. Therefore, signals
not at the reference level will experience an additional amplitude difference when switching between these two sets of RES BW settings due to the differences in scale fidelity.
0 to 70 dB range for RES BW 100 Hz when SPAN = 0 Hz.
Marker Amplitude Resolution*
(Sweep time 30 ms)
Scale: LOG 10 dB/DIV (1/6) dB
LOG 5 dB/DIV (1/12) dB LOG 2 dB/DIV (1/30) dB LOG 1 dB/DIV (1/60) dB LINEAR Reference Level/600
* In E-Series instruments markers are not available for sweep times < 30 ms with RES BW
300 Hz without Option 007. For Option 007, see the characteristics section.
Chapter 2 79
8560E/EC Specifications and Characteristics
Tracking Generator Specifications (Option 002)
Tracking Generator Specifications (Option 002)
Frequency
Frequency Range 300 kHz to 2.9 GHz Frequency Readout
Accuracy
Minimum Resolution
Bandwidth
* Frequency Reference Accuracy = (aging × period of time since adjustment + initial achievable
accuracy + temperature stability).
±(frequency reference accuracy* × frequency + 5% × span + 295 Hz)
300 Hz
Amplitude
Output Level Range
Range 10 dBm to +1 dBm
10 dBm to +2.8 dBm (typical)
Resolution 0.10 dB Maximum Leveled Output Power +1 dBm minimum Power Sweep Range 10 dB, 0.1 dB resolution
Output Level Accuracy
Absolute Amplitude Accuracy
(0 dBm at 300 MHz, 25 °C ± 10 °C) ±0.75 dB
Vernier Accuracy
(refer to 0 dBm at 300 MHz, 25 °C ± 10 °C) ±0.20 dB/dB, ±0.5 dB max.
Level Flatness
(0 dBm relative to 300 MHz) ±2.0 dB
Total Absolute Accuracy (25 °C ± 10 °C) ±3.25 dB
80 Chapter2
Spurious Outputs (at +1 dBm output power)
8560E/EC Specifications and Characteristics
Tracking Generator Specifications (Option 002)
Harmonic Spurious (300 kHz to 2.9 GHz)
<25 dBc
Non-Harmonic Spurious
300 kHz to 2.0 GHz
2.0 GHz to 2.9 GHz
<27 dBc <23 dBc
LO Feedthrough (3.9 GHz to 6.8 GHz) <16 dBm
Fundamental Frequency.
Residuals (RF power off)
300 kHz to 2.9 GHz <78 dBm
Dynamic Range
TG Feedthrough*
300 kHz to 1 MHz <95 dBm 1 MHz to 2.0 GHz <115 dBm
2.0 GHz to 2.9 GHz <110 dBm
* Leakage measured with maximum levelled output power into 50 and with 50 on INPUT
50Ω.
Power Sweep
Range 10 dB Resolution 0.1 dB
Chapter 2 81
8560E/EC Specifications and Characteristics
Inputs and Outputs Specifications
Inputs and Outputs Specifications
CAUTION Any electrostatic discharge according to IEC 801-2/1991 to the center
pins of any of the connectors may cause damage to the associated circuitry.
IF INPUT (Deleted on Option 002 and Option 327)
Connector SMA female, front panel Input level for full-screen deflections 30 dBm ± 1.5 dB
(external mixing mode, 0 dBm reference level, 30 dB conversion
loss)
GPIB
Connector IEEE-488 bus connector Interface Functions SH1, AH1, T6, TE0, L4, LE0, SR1, RL1, PP1, DC1, DT1, C1, C28 Direct Plotter Output Supports HP 7225A, HP7440A, HP 7470A, HP 7475A,HP 7550A Direct Printer Output Supports HP 3630A PaintJet, HP 2225A ThinkJet
CAL OUTPUT
Connector BNC female, front panel Frequency
300 MHz ±(300 MHz × frequency reference accuracy†)
Amplitude 10 dBm ±0.3 dB
Frequency Reference Accuracy = (aging × period of time since adjustment + initial achievable accuracy +
temperature stability).
1ST LO OUTPUT
Connector SMA female, front panel Amplitude
Non-Option 002 +16.5 dBm ±2.0 dB Option 002 +14.5 dBm ±3.0 dB
10 MHz REF IN/OUT
Connector BNC female, rear panel Output Frequency
Frequency Reference Accuracy = (aging × period of time since adjustment + initial achievable accuracy +
temperature stability).
10 MHz ±(10 MHz × frequency reference accuracy†)
82 Chapter2
8560E/EC Specifications and Characteristics
General Specifications
General Specifications
Environmental Specifications
Military Specification per MIL-T-28800, Type III, Class 3, Style B (EC)/Style C (E), as follows:
Calibration Interval 2 years Warmup 5 minutes from ambient conditions* Temperature
Operating 10 °C to +55 °C (E) / 0 °C to +55 °C (EC)
Non-operating 51 °C to +71 °C Humidity 95% at 40 °C for 5 days Altitude
Operating 15,000 feet
Non-operating 50,000 feet Rain resistance Drip-proof at 16 liters/hour/square foot Vibration
5 to 15 Hz 0.060 inch peak-to-peak excursion
16 to 25 Hz 0.040 inch peak-to-peak excursion
26 to 55 Hz 0.020 inch peak-to-peak excursion Pulse Shock
Half Sine 30 g for 11 ms duration Transit Drop 8 inch drop on 6 faces and 8 corners Power Main Voltage fluctuations within the range specified in the
spectrum analyzer “Power Requirements.”
Power Main Operating environment within the limits of installation
category II according to IEC 1010.
Pollution Operating environment within the limits of pollution degree
2 according to IEC 664.
* Two hours for conditions of internal condensation, 30 minutes to meet frequency response
specifications without preselector peaking. When operating outside the 20 °Cto30°C ambient temperature range, preselector peaking is always required to meet frequency response specifications.
Chapter 2 83
8560E/EC Specifications and Characteristics
General Specifications
Electromagnetic Compatibility
Conducted and radiated interference is in compliance with CISPR, Publication 11 (1990).
Military Specification Meets the requirements of MIL-STD-461C, Part 2, with the
exceptions shown below:
Conducted Emissions
CE01 (Narrowband) 1 kHz to 15 kHz only CE03 (Narrowband) Full limits CE03 (Broadband) 20 dB relaxation from 15 kHz to 100 kHz
Conducted Susceptibility
CS01 Full Limits CS02 Full Limits with 40 dB relaxation at IF frequencies. Refer
also to “Radiated Immunity” in Amplitude Characteristics.
CS06 Full Limits
Radiated Emissions
RE01 Test probe at 15 cm, front and rear panel search excluded. RE02 Full limits to 1 GHz
Radiated Susceptibility
RS03 Limited to 3 V/m from 14 kHz to 1 GHz with 40 dB
relaxation at IF frequencies. Refer also to “Radiated Immunity” in Amplitude Characteristics.
84 Chapter2
8560E/EC Specifications and Characteristics
General Specifications
Power Requirements
115 Vac Operation
Voltage 90 V to 140 V rms
Current 3.2 A rms maximum
Frequency 47 Hz to 440 Hz 230 Vac Operation
Voltage 180 V to 250 V rms
Current 1.8 A rms maximum
Frequency 47 Hz to 66 Hz Maximum Power Dissipation 180 W
Audible Noise <5.0 Bels, 20 °C to 30 °C (ISO DP7779)
Weight 20 kg (44 lb)
Chapter 2 85
8560E/EC Specifications and Characteristics
General Specifications
Dimensions
With Handle and Front Cover: Without Handle and Front Cover: (A) 202 mm (7-15/16 in) high (B) 187 mm (7-3/8 in) high (C) 366 mm (14-7/16 in) wide (D) 337 mm (13-1/4 in) wide (F) 503 mm (19-13/16 in) deep (E) 461 mm (18-1/8 in) deep
86 Chapter2
Frequency Characteristics
These are not specifications. Characteristics provide useful information about instrument performance.
Frequency Reference Accuracy
Non-Option 103
8560E/EC Specifications and Characteristics
Frequency Characteristics
Initial Achievable Accuracy
(includes gravitational sensitivity,
retrace, and settability)
Daily Aging (average over 7 days after being powered on for 7 days)
Warmup
(Internal frequency reference
selected)
After 5 minutes
After 15 minutes
Option 103
<±2.2 × 10
<±5 × 10
<±1 × 10−7 of final frequency* (0 °C to +55 °C) <±1 × 10−6 of final frequency* (10 °C)
<±1 × 10−8 of final frequency* (10 °C to +55 °C)
8
10
Initial Achievable Accuracy
* Final frequency is defined as frequency 60 minutes after power-on with analyzer set to internal
frequency reference.
Bandwidth Selectivity
RES BW 100 Hz <4.5:1 RES BW = 1 MHz <8:1 RES BW = 2 MHz <5.5:1
Chapter 2 87
<±1 × 10
6
8560E/EC Specifications and Characteristics
Frequency Characteristics
Impulse Bandwidth
RES BW 2 MHz 2.93 MHz ±10% RES BW 1 MHz 1.60 MHz ±7% RES BW 300 kHz 491 kHz ±7% 300 Hz RES BW 100 kHz 1.62 × RES BW ± 10%
Stability
Noise Sidebands For frequencies 1 GHz, 100 kHz offset from carrier, and
frequency span > 2 MHz
121 dBc/Hz
88 Chapter2
8560E/EC Specifications and Characteristics
Frequency Characteristics
Figure 2-1 Noise Sidebands Normalized to 1 Hz BW versus Offset from
Carrier
Figure 2-2 Typical On-Screen Dynamic Range vs. Offset from 1 GHz Center
Freq. for all RBW's
Chapter 2 89
8560E/EC Specifications and Characteristics
Frequency Characteristics
Sweep
Sweep Time Accuracy
Span 100 Hz <±15%
Resolution Bandwidth Usability* (Non-Option 103)
RES BW Maximum Usable Frequency
Video Average
Video Average ON, 10 Video Averages
OFF
Source/Spectrum Analyzer Frequency
References
Locked
Independent
30 Hz >2.9 GHz >2.9 GHz >2.9 GHz
10 Hz >2.9 GHz >2.9 GHz >2.9 GHz
3 Hz >2.9 GHz >2.9 GHz 2 GHz 1 Hz >2.9 GHz >2.9 GHz 200 MHz
* Resolution Bandwidth Usability is the maximum usable frequency for a given resolution
bandwidth. The maximum usable frequency is limited by signal instability resulting from spectrum analyzer residual FM during the measurement interval. Measurements at frequencies less than the maximum usable frequency will have a typical amplitude uncertainty of less than 1 dB. These characteristics apply after a 30 minute warmup.
Source and spectrum analyzer share the same frequency reference.
Source and spectrum analyzer do not share the same frequency reference.
90 Chapter2
Amplitude Characteristics
Dynamic Range
Figure 2-3 Nominal Dynamic Range
8560E/EC Specifications and Characteristics
Amplitude Characteristics
Amplitude Measurement Range
Spurious Responses Frequency Range Distortion
(all input related spurious responses <1 kHz from the carrier)
10 MHz to 2.9 GHz <55 dBc
Amplitude Accuracy
Input Attenuator Repeatability <±0.2 dB
Pulse Digitization Uncertainty
(Pulse response mode, PRF >720/sweep time)
Standard Deviation <0.2 dB
Chapter 2 91
8560E/EC Specifications and Characteristics
Amplitude Characteristics
Marker Amplitude Resolution
(EC-Series instruments and E-Series instruments with
Option 007, sweep time < 30 ms, zero span)
Scale: LOG 10 dB/DIV (1/2) dB
LOG 5 dB/DIV (1/4) dB LOG 2 dB/DIV (1/10) dB LOG 1 dB/DIV (1/20) dB LINEAR Reference Level/200
Demodulation
Spectrum Demodulation
Modulation Type AM and FM (5 kHz peak deviation) Audio Output Internal speaker and phone jack with volume control. Pause Time at Marker Frequency 100 ms to 60 s
Radiated Immunity
Radiated Immunity
When tested at 3 V/m according to IEC 801-3/1984 the displayed average noise level will be within specifications over the full immunity test frequency range of 27 to 500 MHz except that at the immunity test frequency of 310.7 MHz ± selected resolution bandwidth the displayed average noise level may be up to 80 dBm. When the analyzer tuned frequency is identical to the immunity test signal frequency there may be signals of up to 90 dBm displayed on the screen.
92 Chapter2
8560E/EC Specifications and Characteristics
Option 002 Tracking Generator Characteristics
Option 002 Tracking Generator Characteristics
Tracking Drift Usable in a 1 kHz RES BW after 5 minutes with SRC PWR set to ON*.
Usable in a 300 Hz RES BW after 30 minutes with SRC PWR set to ON*.
* Setting the LINE switch ON does not automatically set SRC PWR to ON.
Effective Source Match <1.92:1
Dynamic Range
300 kHz to 1 MHz 96 dB 1 MHz to 2.0 GHz 116 dB
2.0 GHz to 2.9 GHz 111 dB
Chapter 2 93
8560E/EC Specifications and Characteristics
Inputs and Outputs Characteristics
Inputs and Outputs Characteristics
CAUTION Any electrostatic discharge according to IEC 801-2/1991 to the center
pins of any of the connectors may cause damage to the associated circuitry.
INPUT 50
Connector Precision Type N female, front panel Impedance 50 VSWR (at tuned frequency)
10 dB Input Attenuation <1.5:1 0 dB Input Attenuation <3.0:1
LO Emission Level* (average)
10 dB Input Attenuation <70 dBm
* Level of 1st LO, 3.0 to 6.8 GHz, present at INPUT 50 connector.
IF INPUT (Deleted on Option 002 and Option 327)
(2nd IF input for use with external mixers)
Connector SMA female, front panel Impedance (dc coupled) 50 Frequency 310.7 MHz Noise Figure 7 dB 1 dB Gain Compression Level 23 dBm Full Screen Level 30 dBm
(Gain Compression and Full Screen Levels apply with 30 dB
conversion loss setting and 0 dBm reference level.)
1ST LO OUTPUT
Connector SMA female, front panel Impedance 50 Frequency Range
3.8107 GHz to 6.8107 GHz for analyzers equipped with Option 002.
94 Chapter2
3.0000 GHz to 6.8107 GHz
8560E/EC Specifications and Characteristics
Inputs and Outputs Characteristics
CAL OUTPUT
Connector BNC female, front panel Impedance 50
10 MHz REF IN/OUT
Connector BNC female, rear panel Impedance 50 Output Amplitude 0 dBm Input Frequency
Input Amplitude 2 to +10 dBm External Reference Phase Noise
Analyzer noise sideband performance will not be
degraded if the external reference phase noise is within the limits given below.
Non-Option 103 <135 dBc/Hz at 100 Hz offset Option 103 <110 dBc/Hz at 100 Hz offset
VIDEO OUTPUT* (Deleted on Option 327)
Connector BNC female, rear panel Impedance (dc coupled) 50 Amplitude (RES BW 300 Hz) 0 to +1 V full scale Scaling
RES BW 300 Hz linear or log 100 dB/V
RES BW 100 Hz 4.8 kHz, auto-ranged level with dc offset
10 MHz × (1±2×10−5)
* The VIDEO OUTPUT is a video signal for RES BW 300 Hz with switching transients and IF
ADJ signals between sweeps. For RES BW 100 Hz the output is an IF signal with transients and IF ADJ signals between and during sweeps.
Chapter 2 95
8560E/EC Specifications and Characteristics
Inputs and Outputs Characteristics
LO SWP|FAV OUTPUT and LO SWP|0.5 V/GHz OUTPUT*
Connector BNC female, rear panel Impedance (dc coupled) 120 LO SWP OUTPUT (no load) 0 to + 10 V
0.5 V/GHz OUTPUT Internal Mixer Mode 0.5 V/GHz of tuned frequency (no load) External Mixer Mode ([(1.5 V/GHz) × LO frequency)] 0.2054 V) ±50 mV
0.25 V/GHz OUTPUT
0.25 V/GHz of tuned frequency (no load)
* This connector is labeled LO SWP|0.5 V/GHz OUTPUT on older spectrum analyzers and
LO SWP|FAV OUTPUT on newer spectrum analyzers.
The 0.25 V/GHz output is available only in the 8564E and 8565E.
BLKG/GATE OUTPUT
Connector BNC female, rear panel Impedance 50 Blanking Mode
Amplitude during sweep Low TTL Level Amplitude during retrace High TTL Level
Gate Mode
Gate On High TTL level Gate Off Low TTL level
EXT/GATE TRIG INPUT
Connector BNC female, rear panel Impedance 10 k Trigger Level Settable to high TTL, or low TTL, or edge
triggered TTL
PROBE POWER (front panel)
Voltage +15 V dc, 12.6 V dc Current 150 mA maximum, each
96 Chapter2
8560E/EC Specifications and Characteristics
Inputs and Outputs Characteristics
EARPHONE
Connector 1/8 inch miniature monophonic jack, rear panel Power Output 0.2 W into 4
2ND IF OUT
(Option 001 instruments only)
Connector SMA female, rear panel Impedance 50 Frequency 310.7 MHz
Serial Number Prefix 3 dB BW Noise Figure Conversion
Gain
<3632A >28 MHz 25 dB 15 dB3632A >28 MHz 20 dB 8dB
RF OUTPUT 50
(Option 002 analyzers only)
Connector Type N female Impedance 50 Maximum Safe Reverse Level +30 dBm, 30 Vdc Output VSWR (0 dB attenuation) 1.92:1
EXT ALC INPUT
(Option 002 analyzers only)
Connector BNC female Impedance >10 k Polarity Use with negative detector
ALT SWEEP OUT
(Option 005 analyzers only)
Connector BNC female, rear panel Impedance 100
Chapter 2 97
8560E/EC Specifications and Characteristics
Regulatory Information
Regulatory Information
The following information applies to the 8560EC spectrum analyzer.
98 Chapter2
8560E/EC Specifications and Characteristics
Regulatory Information
The following information applies to the 8560E spectrum analyzer.
Chapter 2 99
8560E/EC Specifications and Characteristics
Regulatory Information
Notice for Germany: Noise Declaration
LpA < 70 dB am Arbeitsplatz (operator position) normaler Betrieb (normal position) nach DIN 45635 T. 19 (per ISO 7779)
CE: European Community
The product design was approved for the European Community.
ISM1-A: Industrial Scientific and Medical Group 1, Class A
The instrument meets the requirements of CISPER 11, Clause 4.
CSA: Canadian Standards Association
The CSA mark is a registered trademark of the Canadian Standards Association. The product meets the requirements.
IEC: International Electrotechnical Commission
This instrument has been designed and tested in accordance with IEC Publication 1010, Safety Requirements for Electronic Measuring Apparatus, and has been supplied in a safe condition. The instruction documentation contains information and warnings which must be followed by the user to ensure safe operation and to maintain the instrument in a safe condition.
This product conforms to Enclosure Protection IP 2 0 according to IEC-529. The enclosure protects against finger access to hazardous parts within the enclosure; the enclosure does not protect against the entrance of water.
100 Chapter2
Loading...