Agilent 8510XF Service Quick Reference Guide

Agilent Technologies 8510XF Network Analyzer System
Service Quick Reference Guide
Agilent Part Number: E7340-90013 Printed in USA Print Date: May 2002
Supersedes: June 2001

Notice

Restricted Rights Legend
If software is for use in the performance of a U.S. Government prime contract or subcontract, Software is delivered and licensed as "Commercial computer sofware" as defined in DFAR 252.227-7014 (June 1995), or as a "commercial item" as defined in FAR 2.101(a) or as "Restricted computer software" as defined in FAR52.227-19(June1987) or any equivalentagency regulation or contract clause. Use, duplication or disclosure of Software is subject Agilent Technologies’ standard commercial license terms, and non­DOD Departments and Agencies of the U.S. Government will receive no greater than Restricted Rights as defined FAR 52.227-19(c)(1-2) (June
1987). U.S. Government users will receiveno greater than Limited Rights as defined in FAR 52.227-14 (June 1987) or DFAR 252.227-7015 (b)(2) (November 1995), as applicable in any technical data.
Hewlett-Packard to Agilent Technologies Transition
This documentation supports a product that previously shipped under the Hewlett-Packard company brand name. The brand name has now been changed to Agilent Technologies. The two products are functionally identical, only our name has changed. The document still includes references to Hewlett-Packard products, some of which have been transitioned to Agilent Technologies.
Windows ® is a registered copyright of Microsoft corportation.
© Copyright 1998, 2001–2002 Agilent Technologies, Inc
8510XF Service Quick Reference Guide

What You’ll Find in This Manual…

Chapter 1 Installation “Preflight” Checkout
Chapter 2 Upgrading a Source for 8510XF
Chapter 3 Making 8510XF Adjustments
Vertical Alignment Adjustment
Degaussing the Display
8360 Series Sources Full User CAL
Power Level Calibration
Detector Gain Cal Conversion Loss Cal
Chapter 4 Performance Verification
Chapter 5 System Level Servicing and Troubleshooting
Theory of Operation
Troubleshooting Block Diagrams
8510XF Service Quick Reference Guide
iii

Warranty

Certification

DOCUMENTATION WARRANTY

Agilent Technologies certifies that this product met its published specifications at the time of shipment from the factory. Agilent Technologies further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology (NIST, formerly NBS), to the extent allowed by the Institute’s calibration facility, and to the calibration facilities of other International Standards Organization members.
THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED "AS IS," AND IS SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, AGILENT DISCLAIMS ALL WARRANTIES, EITHER EXPRESS OR IMPLIED WITH REGARD TO THIS MANUAL AND ANY INFORMATIONCONTAINEDHEREIN,INCLUDINGBUTNOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. AGILENT SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR PERFORMANCE OF THIS DOCUMENT OR ANY INFORMATION CONTAINED HEREIN. SHOULD AGILENT AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT WITH WARRANTY TERMS COVERING THE MATERIAL IN THIS DOCUMENT THAT CONFLICT WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT WILL CONTROL.

Assistance Product maintenance agreements and other customer assistance agreements

are available for Agilent Technologies products. For assistance, call your local Agilent Technologies office (refer to
“Contacting Agilent” on page v).
iv
8510XF Service Quick Reference Guide

Contacting Agilent

Online assistance: www.agilent.com/find/assist United States
(tel) 1 800 452 4844
New Zealand
(tel) 0 800 738 378 (fax) (+64) 4 495 8950
Malaysia
(tel) 1 800 828 848 (fax) 1 800 801 664
Taiwan
(tel) 0800-047-866 (fax) (886) 2 25456723
Latin America
(tel) (305) 269 7500 (fax) (305) 269 7599
Japan
(tel) (+81) 426 56 7832 (fax) (+81) 426 56 7840
Philippines
(tel) (632) 8426802 (tel) (
PLDT subscriber only):
1 800 16510170
(fax) (632) 8426809 (fax) (
PLDT subscriber only):
1 800 16510288
People’s Republic of China
(tel) (preferred):
800-810-0189
(tel) (alternate):
10800-650-0021 (fax) 10800-650-0121
Canada
(tel) 1 877 894 4414 (fax) (905) 282-6495
Australia
(tel) 1 800 629 485 (fax) (+61) 3 9210 5947
Thailand
(tel) outside Bangkok:
(088) 226 008
(tel) within Bangkok:
(662) 661 3999 (fax) (66) 1 661 3714
India
(tel) 1-600-11-2929 (fax) 000-800-650-1101
Europe
(tel) (+31) 20 547 2323 (fax) (+31) 20 547 2390
Singapore
(tel) 1 800 375 8100 (fax) (65) 836 0252
Hong Kong
(tel) 800 930 871 (fax) (852) 2506 9233
8510XF Service Quick Reference Guide
v

Safety and Regulatory Information

Review this product and related documentation to familiarize yourself with safety markings and instructions before you operate the instrument. This product has been designed and tested in accordance with international standards.
WARNING The WARNING notice denotes a hazard. It calls attention to a
procedure, practice, or the like, that, if not correctly performed or adhered to, could result in personal injury. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.
CAUTION The CAUTION notice denotes a hazard. It calls attention to an operating
procedure, practice, or the like, which, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
Instrument Markings
When you see this symbol on your instrument, you should refer to the instrument’s instruction manual for important information.
!
This symbol indicates hazardous voltages.
The laser radiation symbol is marked on products that have a laser output.
This symbol indicates that the instrument requires alternating current (ac) input.
The CE mark is a registered trademark of the European Community. If it is accompanied by a year, it indicates the year the design was proven.
The C-Tick mark is a registered trademark of the Australian Spectrum Agency.
The CSA mark is a registered trademark of the Canadian Standards Association.
vi
8510XF Service Quick Reference Guide
1SM1-A This text indicates that the instrument is an Industrial Scientific and
Medical Group 1 Class A product (CISPER 11, Clause 4). This ISM device complies with Canadian ICES-001.
Cet appareil ISM est conforme a la norme NMB du Canada. This symbol indicates that the power line switch is ON.
This symbol indicates that the power line switch is OFF or in STANDBY position.

Safety Earth Ground

This is a Safety Class I product (provided with a protective earthing terminal). An uninterruptible safety earth ground must be provided from the main power source to the product input wiring terminals, power cord, or supplied power cord set. Whenever it is likely that the protection has been impaired, the product must be made inoperative and secured against any unintended operation.

BeforeApplying Power Verify that the product is configured to match the available main power

source as described in the input power configuration instructions in this manual. If this product is to be powered by autotransformer, make sure the common terminal is connected to the neutral (grounded) side of the ac power supply.
8510XF Service Quick Reference Guide
vii

Typeface Conventions

Typeface Conventions
Italics
Used to emphasize important information:
Use this software
only
with the xxxxxX system.
Used for the title of a publication:
Refer to the
xxxxxX System-Level User’s Guide
.
Used to indicate a variable:
Type
LOAD BIN
Instrument Display Used to show on-screen prompts and messages that you will see on the
display of an instrument:
The xxxxxX will display the message
[Keycap] Used for labeled keys on the front panel of an instrument or on a
computer keyboard:
Press
[Return].
{Softkey} Used for simulated keys that appear on an instrument display:
Press
{Prior Menu}.
User Entry Used to indicate text that you will enter using the computer keyboard;
text shown in this typeface must be typed
Type
LOAD PARMFILE
filename
.
CAL1 SAVED.
exactly
as printed:
Path Name
Computer Display
Used for examples of programming code:
#endif // ifndef NO_CLASS
Used for a subdirectory name or file path:
Edit the file
usr/local/bin/sample.txt
Used to show messages, prompts, and window labels that appear on a
computer monitor:
The
Edit Parameterswindow will appear on the screen.
Used for menus, lists, dialog boxes, and button boxes on a computer
monitor from which you make selections using the mouse or keyboard:
Double-click
EXIT to quit the program.
viii
8510XF Service Quick Reference Guide

Contents

Notice . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ii
What You’ll Find in This Manual… . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iii
Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iv
Certification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iv
DOCUMENTATION WARRANTY . . . . . . . . . . . . . . . . . . . . . . . . . . iv
Assistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iv
Contacting Agilent . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . v
Safety and Regulatory Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vi
Safety Earth Ground . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vii
Before Applying Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vii
Typeface Conventions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . viii
1. Installation “Preflight” Checkout
System Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
System Diagrams . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Rack Diagram Configured for Coaxial Measurements . . . . . . . . . . 1-2
Rack Diagram Configured for Wafer-Probe Measurements . . . . . . . 1-3
Front View Cabling Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
Rear View Cabling (rack mounted source) . . . . . . . . . . . . . . . . . . . . 1-5
Rear View Cabling (table mounted source) . . . . . . . . . . . . . . . . . . . 1-6
8510XF Option 006 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
2. Upgrading a Source for 8510XF
3. Making 8510XF Adjustments
Procedure 1. Vertical Alignment Adjustment . . . . . . . . . . . . . . . . . . 3-3
Procedure 2. DeGaussing (Demagnetizing) the Display . . . . . . . . . 3-4
Procedure 3. 8360 Series Sources Full User CAL . . . . . . . . . . . . . . 3-5
Procedure 3A. 8360 Series Sources Modulator Offset and Gain
Adjustment (8510XF RF source only) . . . . . . . . . . . . . . . . . . . . 3-5
Procedure 4. Power Level Calibration . . . . . . . . . . . . . . . . . . . . . . . 3-6
4. Performance Verification
Verification Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
When to Verify . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Materials Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Calibration and Frequency Ranges . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Verification Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
General Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Software Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
Verification Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8
Low Band Verification (< 50 GHz) . . . . . . . . . . . . . . . . . . . . . . . . . 4-8
8510XF Service Quick Reference Guide Contents ix
High Band Verification (> 50 GHz) . . . . . . . . . . . . . . . . . . . . . . . . 4-10
CW Frequency Accuracy Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-12
Materials Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-12
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-12
Performance Test Record . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-14
5. System Level Servicing & Troubleshooting
Theory of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
Millimeter Wave Controller . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Frequency Bands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
8510XF Firmware Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
Hardware States . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7
Instrument States . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7
Power Level Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8
Test Plots . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-10
x 8510XF Service Quick Reference Guide
1 Installation “Preflight” Checkout

System Overview The 8510XF network analyzer system has been designed to measure

broadband devices to 110 GHz, on-wafer or coax, fully calibrated, in a single frequency sweep.
Single RF connection to 110 GHz
The 8510XF system provides frequency coverage from 45 MHz to 110 GHz in a single sweep with a single connection.
Millimeter-wave measurements in 1.0 mm coax
With a single connection, fully error-corrected measurements to 110 GHz are made in 1.0 mm coax. Making millimeter-wave measurements in
1.0 mm coax delivers uncompromised performance and improved productivity compared to waveguide.
Convenient millimeter-wave on-wafer measurements
The 8510XF system is designed for convenient on-wafer measurements using the 1.0 mm coax connector to interface with the 1.0 mm wafer probe.
Broadband calibration with built-in firmware
The built-in firmware of the 8510XF system enables you to perform a single calibration from 45 MHz to 110 GHz, either in coax or on-wafer.
8510XF Service Quick Reference Guide
1-1
Rack Diagram Configured for Coaxial Measurements
Installation “Preflight” Checkout

System Diagrams

System Diagrams
Figure 1-1 Rack diagram configured for coaxial measurements
1-2
8510XF Service Quick Reference Guide
Rack Diagram Configured for Wafer-Probe Measurements
Installation “Preflight” Checkout
System Diagrams
Figure 1-2 Rack diagram configured for wafer-probe measurements
8510XF Service Quick Reference Guide
1-3

Front View Cabling Diagram

Installation “Preflight” Checkout
System Diagrams
3.5 mm
Figure 1-3 Front view wiring diagram
Figure 1-4 Test head connectors
1-4
8510XF Service Quick Reference Guide
2.4 mm
SMA
testhead.cdr

Rear View Cabling (rack mounted source)

Installation “Preflight” Checkout
System Diagrams
Figure 1-5 Rear view wiring diagram (rack mounted source)
8510XF Service Quick Reference Guide
1-5

Rear View Cabling (table mounted source)

Installation “Preflight” Checkout
System Diagrams
Figure 1-6 Rear view wiring diagram (table mounted source)
1-6
8510XF Service Quick Reference Guide
Installation “Preflight” Checkout

8510XF Option 006

8510XF Option 006
8510XF Option 006 provides the capability to add an additional 8510 series test set provided Option 001 is installed in the test set. Option 006 includes:
Coupler and amplifier added to millimeter-wave controller
2.4 mm termination added to rear panel
Figure 1-7 8510XF Option 006
The configuration for the 8510XF Option 006 is as shown in Figure 1-8.
Figure 1-8 Multiple test set example application
8510XF Service Quick Reference Guide
1-7
Installation “Preflight” Checkout
8510XF Option 006
1-8
8510XF Service Quick Reference Guide

2 Upgrading a Source for 8510XF

All 8360 series sources are compatible, but only the 83621A/B or 83651A/B are recommended in order to insure full functionality of the 8510XF.
Table 2-1 shows the upgrade options required.
NOTE 83651/21A and 83650/20A/B will also work for upgraded 8510XF systems
as long as it has test port flatness correction modification or capability, rear panel output (Option 004), and 1 Hz frequency resolution (Option 008) installed.
Table 2-1 Source Upgrades
Model
83620B/21B 83650A/B 83651A/B
83621A 83631A
83620A 83623A 83624A 83630A 83640A
1. Fully compatible at time of shipment.
2. Can be used for 8510XF LO source when upgraded and includes Options 004 and 008.
3. Includes installation.
4. 83623L, 30L, 40L and 50L, with Option 004 and 008, will also work as LO source, but do not have
2 2
2 24 2 2 2
modulation capability (83623L maximum power at highest frequency is + 15 dBm).
Maximum Power at Highest Frequency
+ 13 dBm + 2.5 dBm + 2.5 dBm
+13 dBm + 10 dBm
+ 13 dBm + 17 dBm + 20 dBm + 10 dBm + 26 dBm
Serial Prefix
All No modification required.
< 3103A 3103A 3104A to 3111A
3112A
3103A
3104A to 3111A
3112A 3145A
Required for Test Port Flatness Correction
83601A upgrade kit. 08360-60167 firmware kit 08360-60201 firmware kit No modification required.
08360-60167 firmware kit 08360-60201 firmware kit No modification required. No modification required.
3
1
1
1 1
8510XF Service Quick Reference Guide
2-1
Upgrading a Source for 8510XF
2-2
8510XF Service Quick Reference Guide

3 Making 8510XF Adjustments

NOTE This chapter covers both the CRT and LCD displays. Some display-related
adjustments apply only to instruments with CRT display.
The following table provides adjustments most frequently needed for 8510XF system service. Most of the 8510XF is self-adjusting except for the following:
Table 3-1 Frequently used 8510XF adjustments
Procedure 1 Vertical Alignment Adjustment
2 Degausing (Demagnetizing) the 85101
3 8360 Series Sources Full User CAL 4 Power Level Calibration Insures level power at the test ports Leveling DAC/ EEPROM, A16,
1. Reference subsequent sections in this document.
2. These adjustments apply to a CRT-based display only.
1
Title Adjustment Function Assembly Adjusted
display
2
2
Aligns softkey labels and mechanical softkey buttons.
Demagnetize the display.
A11
and Leveling Compensation Board, A7
8510XF Service Quick Reference Guide
3-1
Making 8510XF Adjustments
Table 3-2 Equipment required for 8510XF adjustments
Equipment Recommended Model Substitute
CRT demagnetizer or bulk tape
1
eraser
Radio Shack Model 44-233 Techno Tool (USA) or Nietronix (Europe) 692PR022
1 to 2 amp pencil sharpening motor (base held
near CRT with motor on), or electric drill Flat-head screwdriver (small) At least 2 inches long, non-conductive none 10 dB pad 8493C 8360 series source with built-in attenuator 85102A/B and test set adjust software Part Number 08510-10024 Rev. A.01.10 Power Meter 437B power meter 438A or EPM-441A power meter Power Sensors:
N/A
N/A
Only available as a special
N/A
50 GHz V band W band
8481A or 8482A 8487A power sensor V8486A power sensor
W8486A power sensor Adapter, 2.4 to 1.0 mm (f-f) 85059A K01 Adapter, V-band to 1.0 mm (f-f) V281C Adapter, W-band to 1.0 mm (f-f) W281C Laptop, personal computer or
controller
2, 3
8510XF port power level calibration
9000 series 200 or larger
08510-100xx check software
BASIC 6.3 or higher
4
E2060B Agilent VEE 4.0 Installation E2120-13605
1. These adjustments apply to a CRT based display only.
2. Minimum 4 megabytes of available memory after BASIC has been loaded.
3. With BASIC for Windows and PCMCIA-GPIB card.
4. Includes language extensions and drivers.
3-2
8510XF Service Quick Reference Guide
Making 8510XF Adjustments

Procedure 1. Vertical Alignment Adjustment

NOTE This procedure applies to a CRT based display only.
NOTE The vertical alignment can be affected by magnetic interference. Before
adjusting the vertical position, be sure the analyzer is in a non-magnetic environment and the CRT is degaussed.
Perform the following steps to adjust the vertical alignment.
1. Switch on the system and allow it to warm up for 60 minutes.
2. Remove the side panel nearest the display to access the vertical position and focus controls. See Figure 3-1.
3. Insert a narrow, non-conductive, flathead screwdriver (2 or more inches long) into the vertical position hole.
4. Adjust the control until the softkey labels align with the softkeys.
Figure 3-1 Vertical adjustment location
8510XF Service Quick Reference Guide
3-3
Making 8510XF Adjustments

Procedure 2. DeGaussing (Demagnetizing) the Display

NOTE This procedure applies to a CRT based display only.
Use any CRT demagnetizer or bulk tape eraser for this procedure. The color monitor display is very susceptible to external magnetic fields, such as metal frame tables, welded cabinets, the earth, unshielded motors, and other sources. The usual symptom is discoloration or slight dimming of the display (usually near a top corner of the CRT). In extreme cases, a total color shift may be observed. For example, a trace that was red may shift to green. This shift does not suggest a problem with the display. It is characteristic of color displays needing demagnetizing. In countries using 50 Hz, some 10 Hz jitter may be observed. If this problem is observed, remove the device causing the magnetic field.
If the display becomes magnetized or if color purity is a problem, cycle the power several times. Leave the instrument off for at least 15 seconds before switching power on. This will trigger the automatic degaussing circuit in the display. If this is insufficient to restore color purity, use a commercially available demagnetizer (either a CRTdemagnetizer or a bulk tape eraser). In
an emergency, you can use a running electric drill or a pencil sharpener that draws 1/2 to 2 amps through its AC motor winding. Follow the
manufacturer’s instructions keeping in mind that it is imperative when demagnetizing a display that the degausser is kept farther than 4 inches (10 cm) from the face of the CRT.Generally, degaussing is done with a slow rotary motion of the degausser, moving it in a circle of increasing radius while simultaneously moving away from the CRT. The following figure illustrates the motion for degaussing the display.
CAUTION Applying an excessively strong magnetic field to the CRT face can
permanently destroy the CRT.
Figure 3-2 Using a Degausser on the display
3-4
8510XF Service Quick Reference Guide
Making 8510XF Adjustments

Procedure 3. 8360 Series Sources Full User CAL

NOTE If the 8360 series source does not have a front panel, use the 8360 source
Full User CAL initiates a full synthesizer calibration. The calibration performed is instrument state dependent. For example, if the synthesizer is in ramp sweep mode, a sweep span and an auto track calibration are performed. If the synthesizer has amplitude modulation active on an SW signal, then RF peaking and AM bandwidth calibrations are performed. For 8510XF purposes, only ramp-sweep mode is needed.
Perform the following calibration procedure:
1. On the 8360, press
2. Select {Full User CAL}. Wait for the calibration to complete (usually less than a minute).
Attenuation needed
If the 8360 series source does not have a built-in attenuator, you will be prompted to connect a 10 dB attenuator to the 8360 series RF output port. Auto track is done as part of Full User CAL.
front panel emulator in the 8510C network analyzer (refer to the “Service Program” section in the 8510C On-Site Service Manual). For the RF source calibration, the front-panel emulator will default to GPIB address 19. For the LO source calibration, change the GPIB address in the front panel emulator to 18.
[PRESET] [USER CAL].
Procedure 3A. 8360 Series Sources Modulator Offset and Gain Adjustment (8510XF RF source only)
Modulator offset and gain adjustment optimizes the modulator in the 8360 series source for maximum performance. This is necessary since the RF source is part of the 8510XF level control (refer to Figure 3-3 on page 3-6). This should be done before the conversion loss calibration is performed on systems with older sources or sources which may be out of calibration.
Adjust the RF source using the procedure “Modulator Offset and Gain Adjustment” in the 8360 Series Synthesized Sweeper Service Guide’s “Adjustments” chapter.
8510XF Service Quick Reference Guide
3-5
Making 8510XF Adjustments

Procedure 4. Power Level Calibration

Power level calibration insures leveled power at ports 1 and 2. To level the power, we sense the IF signal which correlates to RF level. Since the translation function is frequency dependent, we need to measure the translation function using the conversion loss calibration procedure. The power level calibration is accomplished using IF leveling as shown in the following circuit.
Figure 3-3 8510XF level control
3-6
8510XF Service Quick Reference Guide
Making 8510XF Adjustments
Detector Gain Cal Before we begin the conversion loss calibration, we must perform a detector
gain calibration to compensate for the imprecision of the detector board (E7340-60015 A7). Perform the following procedure for the detector gain calibration:
1. Under the
2. Press
3. Press
4. Press
5. Press
Auxiliary Menus, press [System] hard key.
{More} softkey. {RF Power Config} softkey. {More} softkey. {Reset Det Gain Cal} softkey.
6. Follow the on-screen instructions until cal is completed.
CAUTION If the operating system is reloaded, it is necessary to perform the detector
gain calibration.
NOTE It is recommended that this 5-minute calibration be performed once a month.
Additionally, a CW frequency accuracy test should be performed prior to doing the conversion loss and power level flatness calibrations. Refer to
“CW Frequency Accuracy Test” on page 12.
Conversion Loss
Calibration
After performing a detector gain cal, you will be ready to perform the RF to IF conversion loss calibration. This calibration provides the absolute power level information at 25/50 MHz intervals over the entire frequency range of the E7340/50A to compensate for conversion losses in the E7340/50A so that ports 1 and 2 will be power leveled.
a. The external controller with its software changes the 8510XF CW
frequency and reads the power meter.
b. The entire frequency range is stepped in 25/50 MHz increments at
approximately 2200 points per port (25 MHz steps from 45 MHz to 18 GHz, and 50 MHz steps from 18 GHz to 110 GHz).
c. The result is stored in the millimeter controller EEPROM memory
board (A16) after all bands and heads are measured and should be valid for one year or until a test head is repaired or replaced.
8510XF Service Quick Reference Guide
3-7
Making 8510XF Adjustments
To perform the conversion loss calibration:
Use the following procedure to perform the conversion loss calibration.
1. Load the BASIC, version 6.3 or greater.
2. Calibrate and zero the power sensor, and enter all cal factors versus frequency of the 3 or 4 sensors into the text file of the conversion loss calibration program.
3. Connect the power sensor to the left or right test head and press
[ENTER].
4. Create a power sensor cal factor test file. Using MS Windows Notepad, enter the following values in the respective columns:
Column Entry
First Frequency (GHz) Second Cal Factor Third Sensor #3
Save this file as
XFPWRCAL.PMC, and in the same directory as your
conversion loss cal program.
5. Load
6. Press
xfpwcal.bas, and press [ENTER].
{Run}.
7. Select one of the following frequency ranges:
Selection Frequency Range
1 0.045 to 110 GHz 2 2 to 110 GHz 3 0.045 to 85 GHz 4 2 to 85 GHz
8. Select (1) for the left test-head, or (2) for the right test-head.
9. Enter the serial number of the test head up to eight digits.
10. Approximately 20 frequencies will be used to check the 8510XF power control. These frequencies and the corresponding levels will be displayed if all is okay. If not, the display will have an error message which indicates that the 8510XF leveling loop must be corrected.
NOTE Refer to Performance Verification, page 4-1, for details on the types of
computers and software needed to run the Performance Verification and conversion loss calibration programs.
3-8
8510XF Service Quick Reference Guide
4 Performance Verification
Verification Overview
NOTE The verification procedures in this chapter work for all 8510 and 8510XF
hardware configurations by substituting the proper 8510 source, test set, accessories, and millimeter wave controller.
Performance verification software Rev. A.05.00 or greater will be available only in DOS format, and will work with laptops or PCs that have the required accessories installed. Refer to “Materials Required” on page 4-2.

When to Verify After installation of the system is complete, a performance verification is

necessary to assure proper system operation. This initial verification is included with the installation.
After the initial verification, the verification should be repeated once a year. This recommended interval assumes that Agilent cables are used with the system.
If non-Agilent cables, adaptors, or other fixtures are used, the verification schedule must be determined by the user, as the characteristics of these devices are unknown. In establishing a verification schedule, the following factors should be considered:
Frequency of use
Amount of cable movement
Amount of drift occurring between prior verifications
NOTE Performance verification of a system performed at long intervals is not to be
confused with measurement calibration. Measurement calibration typically is performed on a daily basis, or when measurement setup or conditions have changed.
8510XF Service Quick Reference Guide
4-1
Performance Verification

Materials Required The following materials are required to run the tests:

8510XF system with accessories
85059A 1.0 mm Precision Calibration and Verification Kit
PC-based, 8510 Specification and Performance Verification Software
(Rev. A.05.00 or greater; p/n 08510-10033)
Laptop or PC running BASIC for Windows (3.1/95/NT)
(Rev. 6.3 or greater)
GPIB Card for PCs (National Instruments or HP)
PCMCIA Card for Omnibook or laptop
GPIB cable from PC to 8510XF

Calibration and Frequency Ranges

Table 4-1 Calibration storage locations
The 8510XF requires separate verifications for frequencies below and above 50 GHz (see on “Frequency ranges” on page 4-3).
A special two-band calibration procedure is required for the performance verification of the 8510XF system (a measurement calibration is not sufficient). There are up to three different calibrations and corresponding frequency ranges required, which are stored in suggested locations in the 8510 analyzer (Refer to Table 4-1). For more information refer to the section on “Frequency ranges” on page 4-3. The parameters for these calibrations are set up automatically on the analyzer by the performance verification software.
File Location Name:
Cal Set 1 1-50 GHz Cal Set 2 45 MHz Cal Set 3 50-110 GHz
1. This calibration is for Option 005 systems only.
2. The frequency for standard systems is 2-50 GHz.
3. The frequency range shown is for the E7352A system only. The E7342A
system’s frequency range will be: 50-85 GHz.
Frequency Range of Data Stored:
1
(2-50 GHz)
1
3
2
4-2
8510XF Service Quick Reference Guide
Performance Verification
Frequency ranges
For most of the menu items in the software, there are two or more selections that pertain to the 8510XF. These menu selections are differentiated by:
A. The frequency range of the millimeter wave subsystem
E7352A: 110 GHz subsystem E7342A: 85 GHz subsystem Option 005: 45 MHz to 2 GHz
B. The portion of the subsystem’s frequency range that is being verified (the ranges above and below 50 GHz are verified separately).
The menu selections are broken down in Tables 4-2 and 4-3, below.
Table 4-2 System hardware configuration (E7352A -- 110 GHz System)
Hardware Selection (<50 GHz Range) Selection (>50 GHz Range)
Network Analyzer 8510C -- Vector Network Analyzer 8510C -- Vector Network Analyzer Test Set E7352A -- 1.0 mm Subsystem (2-50 GHz)
E7352A005 -- 1.0 mm Subsystem (45MHz-50GHz) Source 8365xXF -- Synthesizer (45 MHz-50 GHz) 836xxXF -- Multiplied Synth. (above 50 GHz) Calibration Kit
Calibration Technique
Test Port Cables DIRECTXF -- for 8510XF verification (1.0 mm-1.0 mm) DIRECTXF -- for 8510XF verification (1.0 mm-1.0 mm) Verification Kit 85059AOV -- 1.0 mm (45 MHz-50 GHz) 85059AOV110 -- 1.0 mm (above 50 GHz)
1. The 85059A --1.0 mm (45MHz-50GHz) [Use w/ 8517A/B] selection is not used for the 8510XF Performance Verification.
1
85059A -- 1.0 mm (45 MHz-50 GHz) [Use w/
E7342A/52A]
BL -- Broadband Load Cal OS -- Offset Short Cal
E7352A110 -- 1.0 mm (50 GHz-110 GHz)
85059A110 -- 1.0 mm (50 GHz-110 GHz)
Table 4-3 System hardware configuration (E7342A -- 85 GHz System)
Hardware Selection (<50 GHz Range) Selection (>50 GHz Range)
Network Analyzer 8510C -- Vector Network Analyzer 8510C -- Vector Network Analyzer Test Set E7342A -- 1.0 mm Subsystem (2-50 GHz)
E7342A005 -- 1.0 mm Subsystem (45 MHz-50GHz) Source 8365xXF -- Synthesizer (45 MHz-50 GHz) 836xxXF --
Calibration Kit
1
85059A -- 1.0 mm (45 MHz-50 GHz) [Use w/
E7342A/52A]
E7342A85 -- 1.0 mm (50 GHz-85 GHz)
Synth. (above 50 GHz) 85059A85 -- 1.0 mm (50 GHz-85 GHz)
Calibration Technique
Test Port Cables DIRECTXF -- for 8510XF verification (1.0 mm-1.0 mm) DIRECTXF -- for 8510XF verification (1.0 mm-1.0 mm) Verification Kit 85059AOV -- 1.0 mm (45 MHz-50 GHz) 85059AOV110 -- 1.0 mm (above 50 GHz)
1. The 85059A --1.0 mm (45MHz-50GHz) [Use w/ 8517A/B] selection is not used for the 8510XF Performance Verification.
BL -- Broadband Load Cal OS -- Offset Short Cal
8510XF Service Quick Reference Guide
4-3
Performance Verification

Verification Setup

Verification Setup

General Preparation Prepare for performance verification by completing the following steps:

1. Measure the environment temperature and humidity. The temperature must be between +20 cannot vary by more than 1
2. Perform a good installation “preflight” checkout on the 8510XF system.
3. Power on the system components in the following order: a. Sources b. Millimeter wave controller c. 8510XF
°C and +26 °C. Additionally, the temperature
°C after calibration.
d. Controller (PC or laptop)
4. Remember to allow at least one hour for warm up of the components.

Software Installation In order to verify performance, you must first install BASIC for Windows

and the 8510C Specification and Performance Verification software on a PC (or controller).
NOTE This manual documents the use of the verification software on a PC. Use of
the software on a controller is not documented.
The PC must have an INTEL 486 processor or higher,and must have at least eight megabytes of memory available. The PC must also have an GPIB card in order to communicate with the 8510XF.
1. Install BASIC for Windows (Revision 6.3 or later) on the PC following the instructions on the first BASIC disk. The program will guide you through the installation process. (For more detailed information, refer to BASIC for Windows, Installing and Using Guide.) If you are installing Revision 7.1 or later, install the Legacy Version.
2. Install the specification and performance verification software by following the instructions on the disk label. The installation program will lead you through the installation. In the process, a new program group will be created, which includes the “Spec8510” icon.
4-4
8510XF Service Quick Reference Guide
Performance Verification
Verification Setup
NOTE Rev. A.05.00 or greater of the performance verification software is
backwards compatible. So, it is acceptable to replace the older version you have on the PC.
3. Run the program by clicking on the “Spec8510” icon or selecting:
Programs, 8510, Spec8510. The following warning messages will be
displayed:
A validBasicfor Windows GPIBdriverfor your board (if any)has not been loaded. You can run the Specifications and Uncertainty portions of the 8510 software but can not run the Verification portion as this requires communication. Press CONTINUE to load and run the 8510 software without GPIB or:
To load the correct driver EDIT this program and Remove the “!” from in front of the LOAD statement for your GPIB board.
After editing, type RE-STORE “AUTOST” to store the modified program. Then, close this window and re-select the Specifications and Verification icon to re-run the program. The 8510 software should now load and run.
4. Edit the AUTOST program according to the instructions in the warning message. (However, if you do not want to run the verification portion of the software, you can skip this step by selecting
CONTINUE.)
NOTE For additional information on running the system specifications, system
uncertainty, and edit system specifications portions of the program, refer to the “Performance Verification and Specifications” chapter in the 8510C On-Site Service Manual.
Start,
5. After editing and storing the
AUTOST file, exit the program by closing the
window and then restart the 8510 Specification and Performance Verification Program.
Software Configuration
NOTE Once the Specifications and Verification Software is running, there is no
functional difference between operation on a PC and operation on a controller.
1. Verify the GPIB cable is connected from the PC to the GPIB bus (also referred to as the public bus) on the analyzer.
2. After the program is started, the program title (8510 Specifications and Performance Verification Software) and copyright notices are displayed. Select
RESUME.
8510XF Service Quick Reference Guide
4-5
Performance Verification
Verification Setup
3. A screen prompt will give you the opportunity to set the date and time. Press the
[Y] key on the PC to skip this step, or press [N] to set the date
and time.
4. The program loads the System Configuration file, and displays the System Hardware Configuration Menu. Use this menu to specify the equipment you are using, and the frequency range you are verifying (refer to “Frequency ranges” on page 4-3).
To set up for the first performance verification, select the proper hardware setup for Low Band Verification (< 50 GHz) using Table 4-2
on page 4-3 for the E7352A 110 GHz system, or Table 4-3 on page 4-3
for the E7342A 85 GHz system. a. Select
menu item.
b. Use the up and down arrow keys on the PC to advance from one
menu item to another.
c. Select
5. The main menu is displayed (see Table 4-4).
Table 4-4 Main menu
Softkey Description
System Config System Hardware and Software Configuration Menu System Specs System Spec Table Menu System Uncert System Uncertainty & Dynamic Accuracy Plot/Table Menu Edit Specs Edit the System Specifications Verify System System Performance Verification Menu Quit Progam Quit this Program
NEXT or PREVIOUS to step through the list of selections for each
DONE when you have made a selection for each menu item.
4-6
8510XF Service Quick Reference Guide
Printer selection
1. From the Main Menu (choosing printer options):
Performance Verification
Verification Setup
a. Select b. Select c. Select
System config. Software Config. Printer Connected to:
2. Select printer output choice (refer to Table 4-5)
Table 4-5 Printer table of selections
Printer Connected To:
GPIB WIN Printer LPT1 LPT2
2
File
1. Windows default printer
2. Under Printer Redirection (File): Verify directory path exists before entering name of path
1
3. Select
Done and Previous Menu to return to Main Menu.
8510XF Service Quick Reference Guide
4-7
Performance Verification

Verification Procedures

Verification Procedures
Low Band Verification (< 50 GHz)
The following steps assume the hardware configuration has been previously setup for Low Band Verification (< 50 GHz) in step 4 on page 4-6 under Software Configuration.
1. From the Main Menu select Verification Menu.
2. Select
3. Select for frequencies <50 GHz. See Table 4-1 on page 4-2 for the calibrations required and suggested calibration storage locations (also refer to the “Two-Band Calibration Procedure for 8510XF Performance Verification” art foldout at the end of this chapter).
a. Install the 1.0 mm female to female cable (8.8 cm, part number
b. A full two-port calibration is required. A table of calibration
Serial Numbers to enter serial numbers of system and components. System Cal and follow the program prompts to calibrate the system
11500-60001) onto the right test head (port 2) prior to calibration. This cable is part of the 85059A 1.0 mm Precision Calibration and Verification Kit and is considered to be a test port cable.
standards required for each frequency range can be found in the 85059A 1.0 mm precision calibration and verification kit manual, or on the art foldout at the end of this chapter.
VerifySystem to go to the System Performance
CAUTION Do not remove this test port cable once the Calibration/Verification process
has begun. If the test port cable becomes loose or is removed during the calibration/verification process, the calibration is invalid.
CAUTION Do not pull on the connectors on the ends of the test port cable—this will
damage the cable. Avoid damaging the cable once it is connected to port 2 by carefully making the thru connection between the two test heads. This is best done by placing the two test heads on a smooth surface and gently sliding the right test head (port 2) towards the left test head (port 1).
4-8
8510XF Service Quick Reference Guide
Performance Verification
Verification Procedures
c. To make the thru connection, gently slide the right test (port 2)
towards the left test head (port 1) a little bit at a time, while turning the threaded ring on the left head test port connector by hand onto the cable attached to the right test head. Do not use the threaded ring on the test port connector to pull the cable into the connector. Repeat this process until the cable is firmly seated into port 1, then make the threaded ring barely finger tight. Finally use the torque wrench on the port 1 threaded ring and a backup wrench on the cable to tighten the connection.
4. Select
Select Standard to go to the Standard Selection Menu.
5. To verify the system, use the Verification Standards provided with the 85059A 1.0 mm Precision Calibration and Verification Kit (refer to
Table 4-6). Using the menu select the verification standard and enter the
standard’sserial number. The verification standards may be measured in any order.
Table 4-6 Verification standards from 1.0 mm calibration and verification kit
Type of Standard Part Number
Mismatched Thru Adapter 85059-60016 Match Thru Adapter 85059-60017
NOTE It is recommended to connect the Verification Standards to the cable on
port 2 prior to attempting to connect to port 1. Then follow the recommended procedure for making a thru connection outlined under step 3 above to avoid damaging the cable.
6. Use defaults for
Measurement (If Required): 2
7. If your system has Option 005, verify that
Required)
8. Select
9. Select
is set to Yes. Otherwise, set to No.
Done and follow the program prompts. Measure Data to measure the verification standards. Follow the
Cal Set for Broadband Measurement: 1 and Cal Set for 45 MHz
in the Standard Menu.
Cal Set for 45 MHz Measurement (If
program prompts.
10. When verification has been completed view the data, then print or save (if required). If needed the verification standard can be remeasured by selecting
Repeat Measure.
8510XF Service Quick Reference Guide
4-9
Performance Verification
Verification Procedures
11. Select Select Standard to return to the Standard Selection Menu. Repeat
steps 5 through 10 above for the other verification standard (refer to Table 4-6 on page 4-9).
12. Select
Prior Menu until the Main Menu appears.
HighBandVerification (> 50 GHz)
NOTE Before verification of the high band frequencies it is necessary to:
1) select new hardware for the system, and
2) perform a calibration (for the frequency band > 50 GHz).
1. From the Main Menu select
2. Select the proper hardware setup for High Band Verification (> 50 GHz) using Table 4-2 on page 4-3 for the E7352A 110 GHz System or
Table 4-3 on page 4-3 for the E7342A 85 GHz System.
3. Select
4. Select
5. Select
Done and Prior Menu to return to Main Menu. Verify System to go to the System Performance Verification Menu. System Cal and follow the program prompts to calibrate the system
for frequencies >50 GHz. See Table 4-1 on page 4-2 for the calibrations required and suggested calibration storage locations (refer to step 3 on
page 4-8).
System Config then Hardware Config.
6. Select
Select Standard to go to the Standard Selection Menu.
7. To verify the system use the Verification Standards provided with the 85059A 1.0 mm Precision Calibration and Verification Kit (refer to
Table 4-6 on page 4-9). Using the menu select the verification standard
and enter the standard’s serial number. The verification standards may be measured in any order.
NOTE It is recommended to connect the Verification Standards to the cable on
port 2 prior to attempting to connect to port 1. Then follow the recommended procedure for making a thru connection outlined under step 3
on page 4-8 to avoid damaging the cable.
8. Change default for
Cal Set for Broadband Measurement from 1 to 3.
NOTE It is suggested that the High Band Calibration be stored in Cal Set 3 (Refer
to Table 4-1 on page 4-2).
4-10
8510XF Service Quick Reference Guide
Performance Verification
Verification Procedures
9. Verify Cal Set for 45 MHz Measurement (If Required) is set to No.
10. Select
11. Select
Done and follow the program prompts. Measure Data to measure the verification standards. Follow the
program prompts.
12. When verification has been completed view the data, then print or save (if required). If needed the verification standard can be remeasured by selecting:
13. Select
Repeat Measure.
Select Standard to return to the Standard Selection Menu. Repeat
steps 7 through 12 above for the other verification standard (refer to Table 4-6 on page 4-9).
14. Select
15. Select
Prior Menu until the Main Menu appears. Quit Program to exit the program, then close the program window to
finish.
8510XF Service Quick Reference Guide
4-11
Performance Verification

CW Frequency Accuracy Test

CW Frequency Accuracy Test
NOTE The CW frequency accuracy test should be performed prior to doing the
conversion loss calibration adjustment or detector gain calibration.
Source frequency accuracy is tested across the entire sweep range for 8360 sources. The CW frequency accuracy is measured with a frequency counter.

Materials Required The following materials are required to run the tests:

5343 Option 001, 10 MHz to 26.5 GHz frequency counter
10 dB pad
NOTE If the source and test set operate below 500 MHz, connect the test set output
to the 10 Hz to 500 MHz BNC connector on the frequency counter. The input switch on the frequency counter must also be in the 10 Hz to 500 MHz position.

Procedure 1. Connect the equipment as shown in Figure 4-1.

Figure 4-1 CW frequency accuracy equipment setup
2. To preset the instruments, press
{FACTORY PRESET}
4-12
8510XF Service Quick Reference Guide
. Disregard any error messages.
INSTRUMENT STATE [RECALL] {MORE}
Performance Verification
CW Frequency Accuracy Test
3. To set the frequency using the analyzer front panel, press STIMULUS
[CENTER] [MENU] {SINGLE POINT}
.
Enter the start frequency of the source.
4. Measure RF and LO frequencies with the counter, and record these values on the test record at the end of this section.
5. From the analyzer front panel, enter the frequency per the tables at the end of this section.
NOTE Be sure to connect RF and LO output to the 500 MHz to 26.5 GHz input on
the frequency counter. Also set the input switch to the 500 MHz to 26.5 GHz position.
6. Measure the frequency with the counter,then record the value on the test record at the end of this section.
In Case of Difficulty If the measured values do not meet the specifications listed on the test
record, refer to your source manual for adjustment and troubleshooting instructions.
8510XF Service Quick Reference Guide
4-13
Performance Verification
CW Frequency Accuracy Test

Performance Test Record

Record the measured frequencies for RF source readings on Table 4-7, and for LO source on Table 4-8.
Table 4-7 Performance test record for CW frequency accuracy test (RF source)
Instrument Model: _________________ Report Number: __________________________ Date: _______
Frequency Minimum Specification Recorded Results Maximum Specification Uncertainty
45 MHz 44.999955 MHz 45.000045 MHz 10 Hz 2 MHz 1.999998 GHz 2.000002 GHz 10 Hz
2
20 GHz 26 GHz 26.4999735 GHz 26.5000256 GHz 5 kHz
1. The measurement uncertainty is quoted for these performance tests using only the recommended models specified in Table 8-1 of the “Equipment Required” chapter of the On-Site Service Manual. The quoted uncertainty represents limits of 3 times the equivalent standard deviation (3s) and is intended to represent 90% confidence level.
2. For 83620A/B or 83621A/B only.
19.99998 GHz 20.00002 GHz 4 kHz
1
Table 4-8 Performance test record for CW frequency accuracy test (LO source)
Instrument Model: _________________ Report Number: __________________________ Date: _______
Frequency Minimum Specification Recorded Results Maximum Specification Uncertainty
1
45 MHz 44.999955 MHz 45.000045 MHz 10 Hz 2 MHz 1.999998 GHz 2.000002 GHz 10 Hz
2
20 GHz
1. The measurement uncertainty is quoted for these performance tests using only the recommended models specified in Table 8-1 of the “Equipment Required” chapter of the On-Site Service Manual. The quoted uncertainty represents limits of 3 times the equivalent standard deviation (3s) and is intended to represent 90% confidence level.
2. For 83620A/B or 83621A/B only.
19.99998 GHz 20.00002 GHz 4 kHz
4-14
8510XF Service Quick Reference Guide
Two-Band Calibration Procedure for
8510XF Performance Verification
The 8510XF requires separate verifications for frequencies below and above 50 GHz (see "Frequency Ranges" in this same chapter).
A special two-band calibration procedure is required for the performance verification of the 8510XF System (a single full-band measurement calibration is equivalent).not
There are up to three different full 2-port calibrations and corresponding frequency ranges required, which are stored in suggested locations in the 8510 analyzer.
(E7352L or
Calibration Storage Locations
File Location Name:
Cal Set 1
Cal Set 2 45 MHz
Cal Set 3 50–110 GHz
1. This calibration is for Option 005 systems only.
2. The frequency for standard systems is 2–50 GHz.
3. The frequency range shown is for the E7352A system only. The E7342A system's frequency range will be: 50–85 GHz
F
or more information refer to the section on "Frequency Ranges" in this same chapter. The parameters for these calibrations are set up automatically on the analyzer by the performance verification software.
Frequency Range of Data Stored:
1–50 GHz (2–50 GHz)
12
1
3
Calibration Techniques and Standards
Calibration Kit Label:
Frequency
Range
DC–50 GHz
3
Calibration
Technique
1
SOLT
"Open" Standard
(Reflection)
Open
Disk File Name:
"Short" Standard
CK_MMKA1A
(Reflection)
Short 3
50–110 GHz Offset Shorts Short 3 Short 1
1. SOLT = Short, Open, Load and Thru standards
2. The broadband load is a combination of a lossy delay line plus a 50 GHz load.
3. Also used for 45 MHz frequency point.
Calibration Techniques:
A variety of calibration techniques can be used with the 8510XF, including the following:
(Short-Open-Load-Thru) Used for performance verification 45 MHz to 50 GHz.
SOLT
Offset Shorts
TRL
(Thru-Reflect-Line)**
TRM
(Thru-Reflect-Match)**
** Note: Not used for performance verification
The calibration technique used depends in part on the type of calibration kit used. The 1.0 mm calibration kit (85059A) was designed specifically for the 8510XF, and uses a combination of two calibration techniques: SOLT calibration for frequencies up to 50 GHz, and offset shorts calibration for frequencies above 50 GHz. This combination of techniques provides the best possible calibration results on the 8510XF.
(Shorts with different offset lengths) Used for performance verification 50 GHz
to 110 GHz.
E7342L)
1.00mmA.1A
s1
8510C
MM Controller
Left
Test
Head
Full 2-port calibration system configuration
for performance verification in 1.0 mm coax.
"Load" Standard
(Reflection)
50 GHz Load Thru
Short 4 Short 2
83651B
83621B
Cable F-F (8.8 cm) 11500-60001
From 85059A calibration verification kit
"Thru" Standard
(not used for
1-port cal)
Thru Load BB
E7341A (Std. or Opt. 005)
Right Test Head (E7352R or E7342R)
Cable F-F (8.8 cm)
"Isolation"
Standard
50 GHz Load
Clarifying Connector Sex:
The calibration standard labels that appear in the 8510XF softkey calibration menus specify connector sex as "(M)" or "(F)". The sex designator refers to the sex of the test port connector to which the calibration standard is connected ( the sex of the calibration standard connector).not
Offset Shorts:
Male Test Port
The "OPEN" category is therefore renamed "OPEN/SHORT," and the "LOAD" category is renamed "LOAD/SHORT" on the calibration menu softkey labels.
Female Calibration Standard
Offset shorts are used in the place of opens and loads
Full 2-Port Reflection
Cal Menu
(S11):
SHORTS
(S22):
SHORTS
DONE
THRU
THRU
THRU
THRU
TRANS.
DONE
OMIT
LOAD
LOAD
DONE
CLASS11A
CLASS11B
CLASS11C
CLASS22A
CLASS22B
CLASS22C
REFD
FWDT
FWDM
REVT
REVM
TRAD
OMII
FWDI
REVI
ISOD
(F)
(M)
Lossy Delay Line
(M) (F)
(F) (M)
50 GHz Load
"FWD/REV ISOL'N LOAD"
OPEN/SHORT
LOAD/SHORT
OPEN/SHORT
LOAD/SHORT
Full 2-Port
Cal Menu
REFLECT'N
ISOLATION
2-PORT CAL
2
TRANS-
MISSION
SAVE
Cal Cal Set Select Menu
REFL
TRAN
ISOL
SAV2
REFLECT'N
Full 2-Port
Transmission
Cal Menu
FWD. TRANS
FWD. MATCH
REV.TRANS.
REV. MATCH
Full 2-Port
Isolation
Cal Menu
ISOLATION
FWD ISOL'N
REV ISOL'N
ISOLATION
For: < 50 GHz use a 50 GHz load only. For: > 50 GHz a lossy delay line is required in addition
to a 50 GHz load.*
*
Note: If you use a 50 GHz load only, a bad calibration will
result due to reflection from the load above 50 GHz.
at frequencies above 50 GHz.
Open/Short
(F) OPEN
(F) SHORT3
(M) OPEN
(M) SHORT3
DONE
OPEN/SHORT
Shorts Menu
(F) SHORT3
(F) SHORT1
(M) SHORT3
(M) SHORT1
DONE
SHORTS
Load/Short
Menu
50 GHz LOAD
(F) SHORT4
(F) SHORT2
(M) SHORT4
(M) SHORT2
DONE
LOAD/SHORT
(M)
(M)
(F)
(F)
DONE
(M)
(M)
(F)
(F)
DONE
(F)
(M)
(M)
(M)
(F)
(F)
DONE
rr51c

8510XF Full 2-Port Calibration Menus

5 System Level Servicing & Troubleshooting

This chapter describes the basic servicing and troubleshooting techniques for the 8510XF system.
8510XF Service Quick Reference Guide
5-1
System Level Servicing & Troubleshooting

Theory of Operation

Theory of Operation
The 8510XF applies RF test signals to the DUT, and compares these incident signals to whatever is reflected from (or transmitted through) the DUT. The incident, reflected, and transmitted signals are separated out by means of directional couplers, and mixed down to create a set of four IF inputs to the analyzer, as illustrated below.
Figure 5-1 Basic operation of the system
5-2
8510XF Service Quick Reference Guide
System Level Servicing & Troubleshooting
Theory of Operation
At this simplified level of description, system operation does not differ from that of other 8510-based systems. What makes the 8510XF unique is the means by which it covers its extremely wide frequency range.
Figure 5-2 shows how the 8510XF test set is divided into frequency bands.
For the sake of simplicity, this figure only shows half of the test set. An actual test set, with two base units and two test heads, would include twice as many mixers. Note that all bands make use of the same output/input paths, the same connection to the DUT, and the same directional couplers.
Figure 5-2 Modules of the 7352A test set
8510XF Service Quick Reference Guide
5-3
System Level Servicing & Troubleshooting
Theory of Operation

Millimeter Wave Controller

The controller has more functions than are illustrated in Figure 5-2 on
page 5-3. It controls the switching of RF paths in the microwave subheads
and test heads at different points in the system’s frequency range. It also collects together the IFs generated by all mixers in the microwave subhead, the millimeter test heads, and the controller itself, and uses a multiplexer to furnish the appropriate IFs to the network analyzer at each moment in time.

Frequency Bands Each frequency band uses its own set of mixers. The bands are as follows.

Band Included In
.045 to 2 GHz 8510XF systems with Option 005 2 to 50 GHz All 8510XF systems 50 to 85 GHz
50 to 75 GHz 75 to 110 GHZ E7350A systems/E7352A subsystems
Frequency Multiplication The frequency range of the system greatly exceeds that of the system’s RF
source (the 83651B has a maximum frequency of 50 GHz). Frequencies above 50 GHz are generated through harmonic multiplication of the RF source. The signals from the LO source are also multiplied, in the sense that the mixers in certain ranges are designed to use a harmonic of the LO input, rather than the fundamental input. The details for each band are discussed below.
E7340A systems/E7342A subsystems E7350A systems/E7352A subsystems
75 to 110 GHz Band In the 75 to 110 GHz range, the signal from the RF source is applied to a
frequency tripler in the test head. The mixers for this range are located in the test head, and designed to use the 18th harmonic of the LO input frequency. This applies to E7350A systems and E7352A subsystems only.
50 to 85 GHz or
50 to 75 GHz Band
In the 50 to 85 GHz range, the signal from the RF source is applied to a frequency doubler in the test head. The mixers for this range are located in the test head, and designed to use the 14th harmonic of the LO input frequency. For E7350A systems and E7352A subsystems, this band covers the 50 to 75 GHz range.
2 to 50 GHz Band In the 2 to 50 GHz range, the signal from the RF is not multiplied; it is
applied directly to the mixers. The mixers for this range are located in the microwave subhead. In a portion of this range (from 18 to 50 GHz) the mixers use the third harmonic of the LO input frequency. In the remainder of this range (from 2 to 18 GHz), the mixers use the fundamental of the LO input frequency.
5-4
8510XF Service Quick Reference Guide
System Level Servicing & Troubleshooting
Theory of Operation
0.045 to 2 GHz Band In the .045 to 2 GHz range, the signal from the RF source is not multiplied; it
is applied directly to the mixers. The mixers for this range are located in the millimeter wave controller, and are designed to use the fundamental of the LO input frequency. This applies to 8510XF systems equipped with Option 005 only.
8510XF Service Quick Reference Guide
5-5
System Level Servicing & Troubleshooting

8510XF Firmware Specifications

8510XF Firmware Specifications
Operating Parameters Frequency Limits: 45 MHz to 85 GHz (E7340A/E7342A with Option 005)
or 45 MHz to 110 GHz (E7350A/E7352A with Option 005) Band Crossings:
Table 5-1 Band crossings (45 MHz- 85 GHz)
RF Frequency RF Harm LO Harm Band
45 MHz - 2 GHz 1 1 0 2 - 18 GHz 1 1 1 18 - 50 GHz 1 3 1 50 - 64.2 GHz 2 12 2
64.2 - 85 GHz 2 14 2
Table 5-2 Band crossings (45 MHz- 110 GHz)
RF Frequency RF Harm LO Harm Band
45 MHz - 2 GHz 1 1 0 2 - 18 GHz 1 1 1 18 - 50 GHz 1 3 1 50 - 64.2 GHz 2 12 2
64.2 - 75 GHz 2 14 2 75 - 85.5 GHz 3 14 3
85.5 - 97.7 GHz 3 16 3
97.7 - 110 GHz 3 18 3
Step Mode Internally implemented as frequency list mode.
51 points
101 points
201 points
401 points
801 points
List Mode Up to 801 points.
5-6
8510XF Service Quick Reference Guide
System Level Servicing & Troubleshooting
8510XF Firmware Specifications
Sweep Time Similar to 8510C.

Hardware States The hardware states specific to the 8510XF are as shown in Table 5-3.

Table 5-3 Hardware States Specific to 8510XF
System Parameter
System Phaselock Multi Source Leveling Source #1 (RF) Leveling Source #2
1. These menus are integrated into the firmware and are not available as softkeys.
2. Source leveling is set to system for normal (default) 8510XF operation. The source will indicate external leveling for the RF source and internal leveling for the LO source.
1
1
2
2
Frequency Range (45 MHz to 50 GHz)
None None On On System (External) System (External) System (Internal) System (Internal)
Frequency Range (50 to 85 GHz, or 50 to 110 GHz)
Receiver: Constant frequency = 0.020000000 GHz

Instrument States Table 5-4 shows the instrument states specific to the 8510XF system.

Table 5-4 Instrument states specific to the 8510XF system
System Parameter
RF Power source #1 LO Power source #2 Z
o
Delay Coaxial Coaxial
1,2
1,2
Frequency Range (45 MHz to 50 GHz)
10 dBm 10 dBm +10 dBm +10 dBm 50 ohm 50 ohm
Frequency Range (50 to 85 GHz or 50 to 110 GHz)
3
1. The 8510XF normally operates in step mode (ramp mode is not available).
2. Source power is set/read with internal source leveling selected. Normal (default) operation of the 8510XF is with system source leveling selected.
3. Zo is normally set to one Ohm for calibrations/measurements using waveguide test port connectors.
8510XF Service Quick Reference Guide
5-7
System Level Servicing & Troubleshooting

Power Level Check

Power Level Check
A quick operational check may be performed by measuring the power levels of the 8510XF user parameters a levels listed in Table 5-5 on page 5-9 gives the user a high level of confidence that the system is operating properly. For complete system verification, see Performance Verification, page 4-1.
, b1, a2, and b2. Observing the appropriate
1
1. If needed, save the current instrument state by pressing
[SAVE]
and selecting 1-8. Press INSTRUMENT STATE [RECALL], {MORE}, {FACTORY
PRESET}
to restore the default factory instrument state. Change the
INSTRUMENTSTATE
instrument settings as indicated under “Power level check (low RF
power)” on page 5-10.
2. Connect a short to test port 1. Press
PARAMETER [MENU], then {USER 1 a1}.
The forward reference signal path power level is displayed. The power level should be approximately as indicated in Table 5-5 and Figure 5-3
on page 5-11.
3. Select
{USER 4 b1}. The forward reflection signal path power level is
displayed. The power level should be approximately as indicated in
Table 5-5 and Figure 5-6 on page 5-12.
4. Connect a short to test port 1. In order to measure the reverse reference signal path, the parameter must be redefined. Press:
{USER 3 a2} {REDEFINE PARAMETERS} {DRIVE} {DRIVE:PORT2} {REDEFINE DONE}
The reverse reference signal path signal level is displayed. The power level should be approximately as indicated in Table 5-5 and Figure 5-5
on page 5-12.
5. In order to measure the reverse reflection signal path, the parameter must be redefined. Press:
{USER 2 b2} {REDEFINE PARAMETERS} {DRIVE} {DRIVE:PORT2} {REDEFINE DONE}
The reverse reflection signal path power level is displayed. The power level should be approximately as indicated in Table 5-5 and Figure 5-4
on page 5-11.
5-8
8510XF Service Quick Reference Guide
System Level Servicing & Troubleshooting
Power Level Check
6. Connect Port 1 and Port 2 together. Select {USER 2 b2} {REDEFINE
PARAMETERS} {DRIVE} {DRIVE:PORT1}
. The forward transmission signal path power level is displayed. The power level should be approximately 2 dB less than what was measured with a short on each test port.
NOTE The following measurements show the approximate RF signal levels
incident at the first frequency conversion stage and are given in dBm even though the marker value is read out in dB.
NOTE Figures 5-3 through 5-6 are typical graphs from which the data in Table 5-5
was derived. These graphs are typical 8510XF user channel traces with offset shorts connected to Port 1 and Port 2. The RF source power level must be set so that it does not activate the IF OVERLOAD running error message.
Table 5-5 Typical power levels for power level check (low RF power) of the mm-wave
sub-system
Raw Channel Power
Frequency Range (GHz) 75-110 (E7350A/E7352A only)
a
1 (USER 1)
b
1 (USER 4)
a
2 (USER 3)
b
2 (USER 2)
Frequency Range (GHz)
0.045 to 50
5 to 35 30 to 50 35 to50
5 to 35 30 to 50 35 to50
5 to 35 30 to 50 35 to50
5 to 35 30 to 50 35 to50
Frequency Range (GHz) 50 to 85 (E7340A/E7342A) 50 to 75 (E7350A/E7352A)
NOTE 0.1 dB compression, 8 dBm; typical, 10 dBm maximum; damage,
+15 dBm.
If any of the power levels shown in Table 5-5 and Figures 5-3 through 5-6 are not observed (within approximately
±5 dB), consult the “Service and
Troubleshooting” paragraphs in this Quick Reference Guide. An Agilent Customer Engineer may be contacted for service assistance (see “Contacting
Agilent” on page v).
Additional graphs—for help with troubleshooting—are provided in the following sections:
“Power level check (high RF power)” on page 5-13
“Operator’s check (normal operating condition)” on page 5-16
8510XF Service Quick Reference Guide
5-9
System Level Servicing & Troubleshooting
Power Level Check

Test Plots Power level check (low RF power)

The RF source power level must be set so that the IF OVERLOAD running error message is not activated when obtaining these graphs. The graphs
Figures 5-3 through 5-6 were obtained with the following settings:
1. Factory preset
2. Shorts connected to Ports 1 and 2
3. RF Source leveling: internal (no conversion loss calibration)
4. Disconnect the BNC cable from the millimeter-wave controller ALC output to the RF Source ALC input.
5. LO Source leveling: system (default)
6. RF Source power: 10 dBm (default)
NOTE In order for no IF OVERLOAD running error message to occur, it may be
necessary to set the RF Source power level lower.
7. LO Source power: +10 dBm (default)
8. Response reference value: 0 dB (default)
9. Response reference position: 8 Div
5-10
8510XF Service Quick Reference Guide
System Level Servicing & Troubleshooting
Power Level Check
Figure 5-3 User 1 (a1)
Figure 5-4 User 2 (b2)
8510XF Service Quick Reference Guide
5-11
System Level Servicing & Troubleshooting
Power Level Check
Figure 5-5 User 3 (a2)
Figure 5-6 User 4 (b1)
5-12
8510XF Service Quick Reference Guide
System Level Servicing & Troubleshooting
Power Level Check
Power level check (high RF power)
The IF OVERLOAD running error message is normal when obtaining these graphs. A power level of +10 dBm will over drive the 8510XF so you can see the intermodulation leveling differences. The graphs Figures 5-7 through
5-10 were obtained with the following settings:
1. Factory preset
2. Shorts connected to Ports 1 and 2
3. RF Source leveling: internal (no conversion loss calibration)
4. Disconnect the BNC cable from the millimeter-wave controller ALC output to the RF Source ALC input.
5. LO Source leveling: system (default)
6. RF Source power: +10 dBm
NOTE Disregard IF OVERLOAD running error message.
7. LO Source power: +10 dBm (default)
8. Response reference value: 0 dB (default)
9. Response reference position: 8 Div
8510XF Service Quick Reference Guide
5-13
System Level Servicing & Troubleshooting
Power Level Check
Figure 5-7 User 1 (a1)
Figure 5-8 User 2 (b2)
5-14
8510XF Service Quick Reference Guide
System Level Servicing & Troubleshooting
Power Level Check
Figure 5-9 User 3 (a2)
Figure 5-10 User 4 (b1)
8510XF Service Quick Reference Guide
5-15
System Level Servicing & Troubleshooting
Power Level Check
Operator’s check (normal operating condition)
The graphs Figures 5-11 through 5-14 were obtained with the following settings (normal default settings) after performing a factory preset:
1. Shorts connected to Ports 1 and 2
2. RF Source leveling: system (default—with conversion loss calibration)
3. BNC cable connected from millimeter-wave controller ALC output to the RF source ALC input (default).
4. LO Source leveling: system (default)
5. Port 1 and Port 2 power level: 15 dBm (default)
6. Response reference value: 20 dB
7. Response reference position: 5 Div (default)
NOTE For additional information on normal operating conditions, refer to the
“System Operation Test” section in the “Installation” chapter of the 8510XF Operating and Service Manual.
5-16
8510XF Service Quick Reference Guide
System Level Servicing & Troubleshooting
Power Level Check
Figure 5-11 User 1 (a1)
Figure 5-12 User 2 (b2)
8510XF Service Quick Reference Guide
5-17
System Level Servicing & Troubleshooting
Power Level Check
Figure 5-13 User 3 (a2)
Figure 5-14 User 4 (b1)
5-18
8510XF Service Quick Reference Guide
What's Wrong?
1. Do you have any self test failures, running error messages (beeping) or raw channel power (unratioed power) problems? = MAGIC QUESTION.
2. Failed performance tests?
3. Display or Front Panel abnormalities, measurement errors, or display hang-ups?
4. Low or no power symptoms?
Do a Pre-Operational Check of the
8510XF System
Check Front and Rear Panel control settings, cables and GPIB addresses. Check the 85101 free run switch on the A5 Processor Board (all zeros), and proper firm­ware revisions for source and 8510XF (you may need to reload the Operating System in an extreme case). Check that the 8510XF firmware is XF.01.00 or greater.
First, turn on the Sources, then the Test Set, then the 8510XF, or the system cabinet power switch. Then check all troubleshooting LED's. See the 8510XF system level troubleshooting block diagram for these LED's. See installation for additional information in performing the Preflight Checkout.
If NO Obvious Failure Type
Prove 85101 is OK
Obvious Self-Test, running error, or unratioed power test failure. Refer to the 8510XF Service Manual, part number E7350-90001, Service and Troubleshooting.
Obvious Failure Type
No Obvious Failure Type
The best approach to repair the 8510XF at this point is to verify that the 85101C is good. Once it is known to be good, it will act as your diagnostic controller.
Remove the external controller, printer, etc. and the cables from the 8510XF system.
Run the 8510XF Service Program to verify that the 8510XF is OK to a 95% confidence level. To do this, run all the service program tests in the "85101C Display/Processor Service Program menu (#1)." These tests take about ten minutes. Refer to the 8510XF Service Program Flowchart. See partial flowchart on next page.
NOTE: the above tests can be skipped if the 85101C passes the power up 85101 self test. The self test has only a 70% confidence level, so use it accordingly.
To Sheet 2
Obvious Failure Type
Obvious Calibration or Verification Problem
Read the 8510XF Calibration and Verification and Cable manuals on how to inspect, gage, clean, and use the calibration and verification devices, waveguide, and test port return cables. Make sure, if your 85059A Verification Kit has been certified to 110 GHz within the past year.
Obvious Software Problem
Try the backup standard software by reloading the Operating System firmware.
It might be a known software bug. See STARS bug report. Call the factory with details for checkout and duplication of unknown and new bugs!
Obvious Power Supply Problem
CHECK POWER SUPPLIES
Check all main (110V/220V) fuses and power switches. Remove 85101, 85102 and E7342A/52A instrument covers and check LED power and service indicator lights. Recheck by measuring each power supply voltage with a DVM.
The 85102 power switch LED is supplied by the +5V supply in the 85101.
83621A/B
LO Source
45 MHz to 20 GHz
83651A/B
RF Source
45 MHz to 50 GHz
IF, 75-110 GHz
Band
IF, 50-75 GHz
Band
IF, 2-50 GHz
Band
Reference IFsTest IFs
IF, 45 MHz-2 GHz
Band
Option 005
IF, 75-110 GHz
Band
IF, 50-75 GHz
Band
IF, 2-50 GHz
Band
IF, 45 MHz-2 GHz
Band
Option 005
Simplified Signal Path of Unratioed Power Level Tests
Test Head ( 1 of 2 )
Millimeter
Wave
Controller
Option 005
EPROM ID on the
GPIB Board A12
LO
1r
L
I
R
LO
L
I
R
45 MHz to
2 GHz
RF
1t
RF
IF
IF
50 GHz-
110 GHz
45 MHz-
50 GHz
SLOPE AMP
8-18 dB
LO
L
IF
I
R
LO
L
IF
I
R
2r
RF
2-50 GHz
<2 GHz
Splitter
r
2t
RF
2-50 GHz
Microwave Subassembly Millimeter Wave Subassembly
<2 GHz
Splitter
t
PREAMP
PREAMP
LO
IF
L
I
LO
L
IF
I
LO
IF
L
I
LO
L
IF
I
R
R
R
R
4r
RF
3r
RF
4t
RF
3t
RF
X3
X2
SLOPE AMP
75-110 GHz
50-75 GHz
<50 GHz
75-110 GHz
50-75 GHz
<50 GHz
75-110 GHz
50-75 GHz
<50 GHz
Combiner
Splitter
Combiner
Splitter
Combiner
Splitter
r
RF
(Incident)
a1, a2
(Reflected or Transmitted)
test b1, b2
t
r
45 MHz - 110 GHz
RF
E7340
flowcht1.cdr
Rev. 2 4/22/98
Refer to Qualified
Personnel Only!
Test Port
( 1 of 2)
8510XF System Service Flow Diagram (Sheet 1of 2)
5-17/5-18
From Sheet 1
85101 is OK
NOTE: From this point, the problem is probably one of the remaining instruments; the 85102, sources, test heads, the mm-wave controller or their interfaces.
You can use the 8510XF, power-up self tests, running error messages, a1, a2, b1, and b2 raw channel displays, troubleshooting LED's, and passed installation data and displays to help pinpoint the problem to an instrument. But, with the diagnostic aids that are now available for the 8510XF, the quickest way to isolate a problem to an instrument is using the technique of instrument sub­stitution using the diagnostic aids as follows:
To prove that the 85102 is OK, run the built-in 85102 IF/Detector Service Program Number 2. This verifies that the 85102 is good to an 80% confidence level. It takes less than a minute to run and pinpoint failures to a board.
Prove 85102 is OK
85102 is OK
Partial Flowchart
Quick Reference
GPIB
GPIB
Main Service Functions Menu
Looping Self Tests
1 A5 Processor Eprom 2 A5 Processor Ram 3 A7 Data Bus 4 A4 Display Processor 5 A4 Display Ram 6 A7 Timer/Clock/RS-232 7 A7 Public HP-IB 8 A7 System Bus 9 Interrupt System 10 A5 Multiplier 11 A7 Disk Controller 12 A6 Non-Volatile Memory 13 IF Detector Data 14 Keyboard
ENTER selection, then press =MARKER
System Commands
15 Run Main Program 16 Memory Operations 17 Rerun Self Test 18 Repeat Test Loop
Disk Commands
19 Load Program Disk 20 Record Program Disk 21 Initialize Disk
Service Commands
22 Run Service Program 23 Diagnose a Failure
If All Four Channels Fail the
Unratioed Power Level Tests
If all four channels fail the Unratioed Power Level Tests, the most likely causes are a failure in the RF or LO Sources signal going to the Controller test head and associated cables that effect all four channels (see LO path and RF path). A Power Supply can also cause these paths to fail.
Remove the GPIB and RF power cables from the 8510XF system to isolate the sources. Refer to the appropriate source manual to troubleshoot the 83651B and 83621B. For 8360 series sources use the front panel emulator 08510-60022.
NOTE: Do not try to measure the RF source power in source #1 in the 8510XF system by disconnecting the RF cable. Measuring the RF from the source breaks the leveling loop and causes the source to supply maximum power output. Instead, try placing a coupler in the RF cable path and measure the power that way. Consider substituting known good sources and cables for the suspected sources, if they are available.
(20 minute test) To prove that the E7342A/52A Millimeter Sub-Systems are at fault use the part number 85102-60210 5-way power splitter. This substitutes for the mm-wave controller test heads and sources and proves whether they are at fault. Also run the 8510C Service Program Menu number 3.
Finally, perform the RF Source built-in diagnostics and the E7342A/52A un-ratioed power troubleshooting. Refer to the Operator's Check , and perform this check to troubleshoot the un-ratioed power levels. This tests the RF Sources mm-wave controller and test heads. This verifies that the RF Source, mm-wave controller and test heads are operating to a 90% confidence level.
If All Four Channels are OK, the 8510XF is OK and You are Done!
If One, Two, or Three Channels Fail the Unratioed Power Tests. If the power levels on at least one but not all of the four channels fail, the most likely cause of failure is the test head or mm-wave Controller. See E7342A/52A manual for service procedures for they are manual procedures and can­not be abbreviated here.
Below describes how test set manual works for the 8510XF. To help narrow the problem to the faulty instrument, the service adapter and service tools are used to emulate the operation of the source and test set and check unratioed power levels into the 85102 IF/Detector.
Run the Service Program using the Test Set GPIB commands tests. These tests can find approximately 15% of all E7342A/52A mm-wave Controller failures and requires about one minute to complete. The digital functionality of the mm­wave Controller is checked with this test, not the RF paths. Finally, use the failed unratioed power test to find the bad path/paths and assembly.
NOTE: Do not perform Test number 5 "Increment Active Attenuator" in the "Test Set GPIB Commands" menu. The mm-wave Controller does not contain any attenuators.
If One, Two, or Three Channels Fail the Unratioed Power Tests
8510 Service Program Menu
85101 Display/Processor Service Program 1
85102 IF/Dectector Service Program 2
Test Set GPIB Service Program 3
Return to Main Service Functions Menu F[,]
NOTES:
1. Hex digits A thru F are assigned to [G/n], [M/u], [k/m], [X1], [+/-], and [.] respectively.
2. To make a selection, type the number indicated, then press =MARKER.
3. The default value on data entry is zero.
copyright © Agilent Technologies 1987, 1988, 1990, 1998, 2000
85101 Display/Processor
Service Program Menu
CPU Board Tests (A5) 1
I/O Board and Front Panel Tests (A1, A2, A7) 2
Display Board and CRT Tests (A4, A11) 3
Non-Volatile Memory Board Tests (A6) 4
Return to 8510 Service Program Menu F [,]
E7340
flowcht2.cdr
Rev. 2 4/22/98
8510XF System Service Flow Diagram (Sheet 2of2)
5-19/5-20
HOWTOREADTHISBLOCKDIAGRAM:
THISDIAGRAMSHOWSATYPICAL851OXFSYSTEM CONFIGURATION.NOTETHATTHECIRCUITBOARDSAND ASSEMBLIESHAVENUMBERSPRECEDEDBYTHELETTER 'A',THATNUMBERIDENTIFIESTHEASSEMBLY.THEOTHER NUMBERSREFERTOTHELISTONSHEET2.
USETHISBLOCKDIAGRAMASAQUICKREFERENCEGUIDE TOSYSTEMFAILURESASWELLASAGUIDETOLED LOCATIONANDMEANING.TESTSETADDRESSAND CONFIGURATIONSWITCHSETTINGSAREALSOSHOWN.
NUMBERS1–14
THESENUMBERSIDENTIFYTHESELFTESTTHAT EXERCISESTHEPARTICULARASSEMBLY,MOSTLYONTHE 85101DISPLAY/PROCESSOR.FORMOREINFORMATION, REFERTO'SELF'TESTFAILURESINTHEON–SITESERVICE MANUAL.
NUMBERS15–36
THESENUMBERSIDENTIFYTHEBUILT–INRUNNING ERRORMESSAGETHATWILLAPPEARONTHE8510ICRTIF ASYSTEMFAILUREISDETECTED.MOSTLYONTHE85102 IF/DETECTOR,TESTSET,ANDSOURCE.THENUMBERS IDENTIFYTHEASSEMBLIESWHERETHEERRORIS DETECTEDORGENERATED.FORMOREINFORMATION, REFERTO'RUNNINGERRORMESSAGES'INTHEON–SITE SERVICEMANUAL.
NUMBERS37–41
THESENUMBERSIDENTIFYRUNNINGERRORMESSAGES ASSOCIATEDWITH8360–SERIESSOURCESONLYAND APPEARONTHE85101CRT.
NUMBERS42–45
THESEAREEXTRATROUBLESHOOTING
TESTSETDEFAULTSETTINGS:Option006.
THEDEFAULTSTATEOFTHE8517A/BTESTSETS(TURN POWEROFF,PULLA4GPIBBOARDANDTURNPOWER BACKON)ISTOHAVEALLSAMPLERSENABLED,A14'SVTO BIASEDON,PORT1AND2ATTENUATORSAT0dB,ANDTHE SWITCHSPLITTERAPPLYINGRFPOWERTOPORT1.
UNIQUETOOLS
TORXDRIVERT-108710-1623(ForChassis)
TORXDRIVERT-158710-1622(ForChassis)
ALLENWRENCH#5020-0289(ForRPGKnob)6=1/16IN.
85101C
(Address16)
IF/Display
Interconnect
Source LanguageSW setto111 toallow programs toselect language.
MMWave
Controller
(Address23)
RFSource
(Address19)
Language111
LOSource
{
(Address18)
Language111
85102B
83651B
83621B
Option006
CoupledRFOutput
RFInput
LOInput
ALCOutput
RFOutput
EXTALCIn
RFOutput
SelfTestIndicatorsforMillimeterWaveControllerGPIBBoard
IftheACTIVELEDontheFrontPaneldoesnotlightwithintwoseconds ofturnon,orlightsimmediately,thetesthasnotpassedit’sSelfTest. YoucandeterminewhatpartoftheSelfTestfailedbydeterminingwhich GPIBLEDsontheA12Assemblyarelit.TheTablebelowshowsthe possibleSelfTestfailures.
REM
ON
ON
OFF
ElapsedTime
AfterTurn-On
(inseconds)
0to0.5
ONBriefly
0.52to2.0
After2sec.
SelfTest
Indication
PWON
FailROMTest
PassROMTest
FailRAMTest
PassRAMTest
Twosecondsafterturn-onallfourGPIBLEDsshouldgoOFF,andthe FrontPanelACTIVELEDshouldlightsimultaneously.Aftergoingout briefly,thetwoGPIBLEDswilllightaccordingtothestateoftheTest Set.IfyouareusingmultipleTestSets,thefirstTestSetinlinewillbe theactiveTestSetuntilanotherischosen.
IfSelfTestFailstoRun
IftheportionofmemorywhichcontainstheSelfTestprogrammingisfaulty, theSelfTestwillnotrunproperly.Thefollowingconditionsindicatethatthe SelfTestROMsarefaulty:
AllLEDsflashbrieflyandgoOFF AllLEDsflashbrieflyandstayON ACTIVELEDgoesONbeforeGPIBLEDsgoOUT ACTIVELEDdoesnotgoONatall(checkLEDoperationbeforeproceeding)
8in-lbs20mm8710-1764
8in-lbs5/16in8710-1765
4in-lbs6mm8710-2079
A12GPIBLEDs
LSN
TLK
SRQ
ON ONONONONON
OFF
ON ON
OFF
OFF
OFF
OFF
OFF
OFF OFF
OFF
TORQUEWRENCHES
8510XF
(RearView)
40103ARack
FillerPanel
40104ARack
FillerPanel
8510CInterconnect
TestSetInterconnect
TestSetIFInterconnect
SystemBus
GPIB
GPIB
SystemBus
GPIB
SystemBus
40102ARack
FillerPanel
46298M
DrawerUnit
GPIBSystemBus
GPIBSystemBus
10MHzIn
TriggerIn
10MHzRefInput
10MHzRefOutput
TriggerOutput
10MHzRefOutput
10MHzREF
TriggerIN
10MHzREF
83621A/B
LOSource
45MHzto20GHz
83651A/B
RFSource
45MHzto50GHz
IF,75-110GHz
Band
IF,50-75GHz
Band
IF,2-50GHz
Band
IF,45MHz-2GHz
Band
Option005
IF,75-110GHz
Band
IF,50-75GHz
Band
IF,2-50GHz
Band
IF,45MHz-2GHz
Band
Option005
Whenchanging A16: SaveEEPROM fromoldA16 levelingDAC boardtoretain calibration constants.
Millimeter
Wave
Controller
Option005
RegulatorBoard
9PowerSupply
LEDsforALL
PowerSupplies
+5VLEDFront
Panel
PowerSwitch
LO
1r
L
I
R
4LEDs
LO
L
I
R
2GHz
RF
1t
RF
IF
GPIBBoard
IF
45MHzto
GPIBSystemBus
TestHead(1of2)
SLOPEAMP
50MHz­110MHz
45MHz-
50GHz
8-18dB
LO
2r
L
IF
I
R
LO
L
IF
I
R
SupplyLEDsforallPowerSupplies
RF
2-50GHz
<2GHz
2t
RF
2-50GHz
<2GHz
MicrowaveSubassembly
BiasBoardsandPower
Splitter
Splitter
75-110GHz
X3
PREAMP
PREAMP
LO
50-75GHz
X2
SLOPEAMP
<50GHz
Combiner
Splitter
r
4r
IF
IF
r
IF
IF
RF
L
R
I
LO
L
I
R
LO
L
R
I
LO
L
I
R
75-110GHz
3r
RF
50-75GHz
45MHzto50GHzBothONDS7andDS9
50GHzto75GHzDS7OFFDS9ON
75GHzto110GHzDS9OFFDS7ON
Combiner
Splitter
r
(Incident)
<50GHz
DS7andDS9LEDs(Yellow)
referencea1,a2
RF
4t
RF
75-110GHz
(Reflectedor
3t
RF
50-75GHz
<50GHz
Combiner
Splitter
Transmitted)
testb1,b2
t
RF
t
MillimeterWaveSubassembly +/-7VoltONLEDFrontPanel
Connector-TestPort
*
1mmM/F (partnumber 85104-60061)
TestPort
(1of2)
*
9
ToSheet2
rr52c
8510XFSystemLevelTroubleshootingBlockDiagram(CRT/LCD)(Sheet1of2)
GPIB System Bus
From Sheet 1
9
10 MHz REF
Trigger IN
Self Test (1-14), Running Error Messages (15-41), Unratioed Power Test (42), LEDs/GPIB Addresses/Configuration Switch Settings
PC/CONTROLLER
GPIB
(VGA Type
Connector)
85101-60263
A11 DISPLAY SECTION
(In 8510XF for blanking only)
69261
85102-69272
B1 FAN
08756-20073
FAN REPLACEMENT KIT
for 85102A
08756-60074
GPIB
MOTHERBOARD
CAPACITORS
69272
GPIB
7167
0950-3488
SELF TESTS
1. A5 PROCESSOR EPROM (self test and service program PROM using checksum)
2. A5 PROCESSOR RAM (test processor memory using data complement technique)
3. A7 DATA BUS TEST (test I/O data bus)
4. A4 DISPLAY PROCESSOR
5. A4 DISPLAY RAM
6. A7 TIMER/CLOCK/RS – 232
7. A7 PUBLIC GPIB (Read/Write test of GPIB register)
8. A7 SYSTEM BUS (Read/Write) test of system bus register)
9. INTERRUPT SYSTEM (test interrupts)
10. A5 MULTIPLIER (PERFORMS a complex multiply)
11. A7 DISC CONTROLLER
12. A6 NON-VOLATILE MEMORY TEST (read/write test)
13. IF DETECTOR DATA (read/write test data and address lines)
14. KEYBOARD (read 85101C keyboard for a stuck key)
RUNNING ERROR MESSAGES
15. FAILED PRETUNE/LOST PRETUNE
16. NO IF FOUND
17. PHASE LOCK FAILURE
18. PHASE LOCK LOST
19. VTO OVER-RANGE
20. ADC CAL FAILED
21. IF CAL FAILED – REFERENCE; IF CAL FAILED – TEST; IF CAL FAILED – BOTH (channels)
22. ADC NOT RESPONDING
23. IF OVERLOAD (External level of 8360 can cause this.)
24. AUTORANGE CAL FAILED
25. SOURCE SLEEP SYNC ERROR
26. SWEEP TIME TOO FAST
27. SOURCE (GPIB) SYNTAX ERROR
28. TEST SET (GPIB) SYNTAX ERROR
29. SYSTEM BUS ADDRESS ERROR
30. SYSTEM BUS SRQ ERROR
31. UNABLE TO LOCK TO EXT 10 MHz REF
32. ERROR: SYSTEM KEY NOT INSTALLED; CAUSED BY SECURITY KEY MISSING, BAD I/O, BAD A8 CONNECTOR, BAD MOTHER-BOARD. CAUTION: OPTIONAL FUNCTION: NOT INSTALLED CAUTION: TIME DOMAIN KEY NOT INSTALLED
33. DISC HARDWARE PROBLEM
34. DISC COMMUNICATION ERROR
35. INITIALIZATION FAILED
36. DISC MEDIA WEARING OUT - REPLACE SOON
8360 – SERIES SOURCES RUNNING ERROR MESSAGES
37. FAILURE – FAULT INDICATOR ON
38. FAILURE – RF UNLOCKED
39. FAILURE – OVERMODULATION
40. FAILURE – SELF TEST FAILURE
41. INVALID SETUP – CHECK SYSTEM BUS CONFIGURATION
OTHER TROUBLESHOOTING HINTS
42. Perform the unratioed power test for a1, a2, b1, and b2. Refer to 'System Maintenance, Troubleshooting' in the 8510XF Operating and Service Manual, E7350-90001.
43. The A4 GSP and the A11 display must be working in order for a normal display to be on the CRT.
44. If self test 11 passes, try to initialize a blank disc to further test the A2 disc drive.
45. A18 (85102) ANALOG TO DIGITAL CONVERTER TEST When you first turn the instrument ON, if the noise on the trace is greater than 1 dB, then the A18 board may be faulty.
If you cycle the AC power (Off, then On), the noise may disappear. If it does, there is probably a malfunction with the A18 A/D board although no apparent symptoms are present. There also may be no Running Error Messages, Self Test Failures or Service Program test failures.
If this trace noise disappears when the POWER is cycled, the test has FAILED and the A18 board should be REPLACED.
NOTE: A16, A23, A22 are not used for 8510XF operation; that is, phase lock none, except for some signal transfers from LPTHLD, etc.
rr518c
8510XF System Level Troubleshooting Block Diagram (CRT) (Sheet 2 of 2)
From Sheet 1
GPIB System Bus
9
10 MHz REF
Trigger IN
Self Test (1-14), Running Error Messages (15-41), Unratioed Power Test (42), LEDs/GPIB Addresses/Configuration Switch Settings
PC/CONTROLLER
GPIBGPIB
(VGA Type Connector)
85101-60263
MOTHERBOARD
CAPACITORS
85101-60298
85101-60300
1
A16 INVERTER BOARD
0950-3379 (A16 Only)
(In 8510XF for blanking only)
69261
69272
B1 FAN
08756-20073
FAN REPLACEMENT KIT
for 85102A
08756-60074
69272
85101-60296
GPIB
A15 LCD ASSEMBLY
85101-60297 (Includes A16)
7167
0950-3488
SELF TESTS
1. A5 PROCESSOR EPROM (self test and service program PROM using checksum)
2. A5 PROCESSOR RAM (test processor memory using data complement technique)
3. A7 DATA BUS TEST (test I/O data bus)
4. A14 DISPLAY PROCESSOR
5. A14 DISPLAY RAM
6. A7 TIMER/CLOCK/RS – 232
7. A7 PUBLIC GPIB (Read/Write test of GPIB register)
8. A7 SYSTEM BUS (Read/Write test of system bus register)
9. INTERRUPT SYSTEM (test interrupts)
10. A5 MULTIPLIER (PERFORMS a complex multiply)
11. A7 DISC CONTROLLER
12. A6 NON-VOLATILE MEMORY TEST (read/write test)
13. IF DETECTOR DATA (read/write test data and address lines)
14. KEYBOARD (read 85101C keyboard for a stuck key)
RUNNING ERROR MESSAGES
15. FAILED PRETUNE/LOST PRETUNE
16. NO IF FOUND
17. PHASE LOCK FAILURE
18. PHASE LOCK LOST
19. VTO OVER-RANGE
20. ADC CAL FAILED
21. IF CAL FAILED – REFERENCE;
IF CAL FAILED – TEST; IF CAL FAILED – BOTH (channels)
22. ADC NOT RESPONDING
23. IF OVERLOAD (External level of 8360 can cause this.)
24. AUTORANGE CAL FAILED
25. SOURCE SLEEP SYNC ERROR
26. SWEEP TIME TOO FAST
27. SOURCE (GPIB) SYNTAX ERROR
28. TEST SET (GPIB) SYNTAX ERROR
29. SYSTEM BUS ADDRESS ERROR
30. SYSTEM BUS SRQ ERROR
31. UNABLE TO LOCK TO EXT 10 MHz REF
32. ERROR: SYSTEM KEY NOT INSTALLED; CAUSED BY SECURITY KEY MISSING, BAD I/O, BAD A8 CONNECTOR, BAD MOTHER-BOARD. CAUTION: OPTIONAL FUNCTION: NOT INSTALLED CAUTION: TIME DOMAIN KEY NOT INSTALLED
33. DISC HARDWARE PROBLEM
34. DISC COMMUNICATION ERROR
35. INITIALIZATION FAILED
36. DISC MEDIA WEARING OUT - REPLACE SOON
8360 – SERIES SOURCES RUNNING ERROR MESSAGES
37. FAILURE – FAULT INDICATOR ON
38. FAILURE – RF UNLOCKED
39. FAILURE – OVERMODULATION
40. FAILURE – SELF TEST FAILURE
41. INVALID SETUP – CHECK SYSTEM BUS CONFIGURATION
OTHER TROUBLESHOOTING HINTS
42. Perform the unratioed power test for a1, a2, b1, and b2. Refer to
'System Maintenance, Troubleshooting' in the 8510XF Operating and Service Manual, E7350-90001.
43. The A14 GSP and the A15 display must be working in order for a normal
display to be on the LCD.
44. If self test 11 passes, try to initialize a blank disc to further test the A2
disc drive.
45. A18 (85102) ANALOG TO DIGITAL CONVERTER TEST
When you first turn the instrument ON, if the noise on the trace is greater than 1 dB, then the A18 board may be faulty.
If you cycle the AC power (Off, then On), the noise may disappear. If it does, there is probably a malfunction with the A18 A/D board although no apparent symptoms are present. There also may be no Running Error Messages, Self Test Failures or Service Program test failures.
If this trace noise disappears when the POWER is cycled, the test has FAILED and the A18 board should be REPLACED.
NOTE: A16, A23, A22 are not used for 8510XF operation; that is, phase lock none, except for some signal transfers from LPTHLD, etc.
rr53c.cdr
8510XF System Level Troubleshooting Block Diagram (LCD) (Sheet 2 of 2)
Loading...