This manual applies to all HP 8510C network analyzers, and to all HP
8510A and
85lOB
Performance Upgrade Package
SERIAL NUMBERS
network anlayzers that have been upgraded to HP 8510Cs.
C.07.00
OT
Greater of the
HP Part No. 1157590023
Printed in USA
Jnuary
1994
@Copyright
Hewlett-Packard Co.,
1400 Fountaingrove Pkwy., Santa Rosa, CA 95403 U.S.A.
Contents
1. Introduction
The Upgrade Package
HP 115755 Performance Upgrade
Warranty Information
Upgrade Requirements
Source Compatibility
Installing an Upgrade
Procedure
Step 1. Check the Upgrade Package Contents.
Step 2. Get the Items Not Included in Upgrade Package.
Step 3. Performance Check the HP 8510C System.
Step 4. Follow the Upgrade Instructions.
Checking the HP 85lOC Operation
2.
HP 115755 Performance Upgrade (Rev. C.07.00or Greater)
2-l. HP 115755 Performance Upgrade Package Contents
D-l. Operating System File Compatibility
D-2. HP 85106C mm-Wave Rev. vs System Firmware Rev.
................
.............
................
...........2-l
.................D-l
..........
l-5
A-l
A-2
l-l
l-2
l-3
D-3
Contents-2
HP 8510C Operating System Upgrades
The Upgrade Package
The latest revision number of firmware at the time of shipment is printed on the operating
system disk label. You can determine your instrument’s current firmware revision at power
up when the firmware revision number is displayed. Refer to Table l-l, below, for a list of
requirements to use this upgrade. Refer to Table l-2 for the list of features added to this
upgrade.
HP
11575J
Performance Upgrade
Use this package to upgrade an HP 8510C from revision
C.OG.XX
operating system firmware
to revision C.07.00 or greater.
Warranty Information
Installation of an upgrade does not affect the existing instrument warranty in any way.
Specifically, it does not extend the current instrument warranty. Any parts supplied in an
upgrade package carry a
go-day
replacement-part warranty.
Upgrade Requirements
Table 1-1 lists the instrument and firmware revision requirements of this operating system
upgrade.
Table
14.
Requirements of Each Upgrade Package
Upgrade
Package
HP 115755
Firmware
Revision
Provided
Current
Revision1
Analyzer
Model
Required
HP 851OC*
Firmware
Revision
Required
C.06.00
greater
or
.
1
Current
(Revision 7.00 or greater).
2 Applies to any HP 85lOA or
851OC.
HP 851OC Operating System Upgrades
firmware
is the latest revision available at the time of shipment
8SIOB
that has been upgraded to an HP
Introduction l-l
Source Compatibility
For
complete compatibility with revision
Series source. If an HP 8360 Series source in not used, the Power Domain and Receiver
Calibration functions will not work.
Table
l-2.
HP 8510C Operating System Upgrade History
C.07.00
firmware, your source must be an HP 8360
Upgrade Package
HP 11575J
Added Features .
n
Limit-Line testing capability
w Receiver calibration capibility
n
Dissimilar connector compensation
n
Color printer compatibility
n
Power domain measurements (includes power sweep)
I-2
Introduction
HP 851OC Operating System Upgrades
Installing an Upgrade
Procedure
Step 1. Check the Upgrade Package Contents.
Using the table in Chapter 2, Table 2-l for your upgrade, ensure that the package is complete.
If not, contact your local Hewlett-Packard Sales and Service Office (listed inside the back
cover of this manual) for assistance.
Step 2. Get the Items Not Included in Upgrade Package.
If needed, get items called out in the installation procedures, but are not included in the
upgrade package.
Step 3. Performance Check the HP 8510C System.
Using the instructions in
“Checking the HP 8510C Operation”, check that the system you
plan to upgrade is functioning.
Step 4. Follow the Upgrade Instructions.
Follow the instructions provided in the appendices listed below to install and check your
upgrade installation.
\
Upgrade Package
Firmware Revision
HP 115751, C.07.00
Firmware
Table l-3. Upgrade Package Appendix Information
Appendix Contents
Appendix
Appendix B Loading the HP 8510C Operating System
Appendix C Backing Up the Operating System
Appendix D Instrument State Files
A
Installing the EEPROM
.
7
HP
SglOC
Operating System Upgrades
Introduction l-3
Checking the HP 8510C
F%Fy
CONNEC!
HP-IB
f
WETUORK
ANRLYZER
,
SOURCE
. Q
0
a
ab
SOURCf
MODULE
INTERFACE
RF IN
TEST SET
5
SOURCE CONTROL
EST
SET
CA8l.E (HP
MERCONNECT
8516A ONLY)
IN/OUT
0
RF
OUTPUT
HP-B
c
POWER SPUllER HAS 3.5mm
(F)
Figure l-l. Equipment Setup and HP 8511 A-Specific Connections
HP 8510C Operating System Upgrades
Introduction
l-5
:
t
HP 11575J Performance Upgrade
(Rev. C.07.00 or Greater)
Use this package to upgrade an HP 851OC from revision C.06.00(or greater) operating system
firmware to the current firmware revision at the time of shipment. Chapter
the requirements for this upgrade. Features added to this upgrade are listed in Chapter 1,
Table l-2.
Procedure Overview
1,
Table l-l, lists
Note.
Step 1. Install the new EEPROM, Appendix A.
Step 2. Load the new operating system, Appendix B.
Step 3. Backup the new operating system, Appendix C.
Step 4. Update documentation, inserting the new documents in appropriate manuals.
Step 5. Become aware of instrument state limitations, Appendix 0.
You should have checked system performance already. If not, return to
Chapter
Operation.”
1,
read and follow the instructions in “Checking the HP 85lOC
-
Upgrade Package Contents
Table
2-l.
HP 11575J Performance Upgrade Package Contents
HP 11575J
Item
I Performance
1 Installation manual
EEPROM
Uwrade
Data Disk
1 Qty 1
I
HP Part Number
1
I
85101-80116
1
85101-80114
1
I
Performance Upgrade
2-l
Installing the EEPROM
Procedure
1. Ensure a static-safe work station. Check for the following:
a. The static mat sits on a clean, flat, sturdy surface.
b. The static mat has an earth ground connection.
c. The static mat has a connected wrist strap.
2. Install the EEPROM:
a. Remove the HP 85101 display processor from the system.
b. Remove the top cover of the display processor.
c. Refer to Figure A-l. Remove the A8 assembly.
A
.
HP 8510C Operating
System
Figure A-1. A8 Assembly Location
Upgrades
Install EEPROM A-l
d. Install EEPROM
sockets on the A8 assembly (see Figure A-2).
(85101-80114)
rom the upgrade kit in any one of the empty EEPROM
f
Note
e. Replace the A8 assembly; replace the top cover.
f. Return the display processor to the system; reconnect all cables and line cords.
g.
Continue with
If an EEPROM socket needs to be opened, remove the EEPROM containing
the previous revision system firmware. All of the sockets can be filled without
any adverse affects to system performance.
.
“Making Sure the EEPROM is Installed
Properlv.”
0
A0 ASSEMBLY
Figure A-2. EEPROM Installation Location
A-2 Install EEPROM
HP 8510C Operating System Upgrades
i
:
t
i
,
Instrument State Files
HP
85lOC
Network Analyzers
Not all instrument state
Table D-l lists the firmware revisions and their file compatibility.
Be aware that if you save instrument state files (on tape or disk) on an analyzer with a given
firmware revision, you may not be able to use those files in an instrument having a more
recent firmware version.
The first vertical column in Table D-l. lists the firmware revision number under which your
work disk files were created. The horizontal row lists the upgrade operating system revision
numbers.
n To read Table D-l, look for the revision number you used to save data on disks.
H
Then look for the current operating system (horizontal axis) installed in your instrument.
H
If your previously saved disk information is saved with a compatible operating system
revision, there is a “yes” printed in the vertical column where the two columns meet.
H
If there is a “no” printed there, the data from your work disk must be recreated and saved
under the new operating system.
Caution
The analyzer successfully loads incompatible instrument state files. When
the incompatible file is recalled, however, the display blanks, and the front
panel does not respond. Do not press (PRESET). See “Recovering from an Error
State” for instructions about recovering properly from a crashed system.
fles
are interchangeable between the various firmware revisions.
Table D-l. Operating System File Compatibility
Revisions
C.06.00-system stored files
I
C.06.50~system stored files
I
C.07.00-system
I
HP 851OC Operating
stored
System
Firmware
I
I
files
I
Upgrades
RevisionFirmware Revision
C.06.00
Yes
Yes
No
C.06.50 to C.06.56C.07.00 or greater
No
must be recreated)
I (
I
I
Yes
No
Firmware Revision
I
I
~ I
Instrument State Files D-l
No
No
Yes
I
I
I
Recovering from an Error State
To recover an instrument from the error state described above:
Using a narrow tool (such as a small screwdriver), press the TESTbutton on the analyzer
front panel.
:
t
To clear the incompatible instrument state:
Press (SAVE), then select each instrument state register.
.
.
Do not press
(=)I
m
w
Pressing
data may remain corrupted and cause the analyzer to fail at a later time.
Do not cycle power. This does’not work if the incompatible instrument state
was loaded into instrument state 8.
(PRESEi;)
may allow the instrument to partially recover. However the
Recreating the Instrument State
In the case of file incompatibility just described, you must manually recreate the desired
instrument state, then reload it on disk.
If you encounter an incompatibility after upgrading your instrument to firmware revision
C.06.50 or greater, and you do not have time to recreate earlier instrument states, reload the
old firmware. This gives you full use of instrument state files created on an instrument with
the earlier firmware revision.
Remember
Machine dump files contain instrument states, and have the same
compatibility rules.
D-2 Instrument State
Files
HP 851OC Operating System Upgrades
HP
85lOC
Millimeter-Wave Systems
Caution
The analyzer successfully loads incompatible revision-version files. When
incompatible files are recalled, however, the display goes blank and there is no
front panel response. Do not press (PRESET). Refer to “Recovering from an
Error State” for instructions about recovering properly from a crashed system.
HP
85106C
Table D-2 defines the compatibility of HP 85106C configuration disk revisions versus analyzer
firmware revisions.
Table D-2. HP 85106C mm-Wave Rev. vs System Firmware
If the network analyzer in your HP 85106C millimeter-wave system fails after you load the
system configuration disk, order the millimeter-wave system configuration disk revision that is
correct for your firmware.
HP 851OC Operating
System
Upgrades
Instrument State Files D-3
Hewlett-Packard Sales and Service Off ices
IN THE UNITED STATES IN AUSTRALIA
California
Hewlett-Packard Co.
1421
South Manhattan Ave.
P.O. Box
Fullerton, CA
4230
92631
(714)999-6700
Hewlett-Packard Co.
301
E. Evelyn
Mountain View, CA 94039
(415)6942000
Colorado
Hewlett-Packard Co.
24
Inverness Place, East
Englewood,. CO
80112
(303)649-5000
Georgia
Hewlett-Packard Co.
2000
South Park Place
P.O. Box
Atlanta, GA 30339
105005
(404)955-1500
Illinois
Hewlett-Packard Co.
5201
Tollview
Rolling Meadows, IL
(312) 255-9800
Drive
60008
New Jersey
Hewlett-Packard Co.
120 W. Century Road
Paramus,NJ
07653
(201)265-5000
Texas
Hewlett-Packard Co.
930
E. Campbell Rd.
Richardson, TX
75081
Hewlett-Packard Australia Ltd.
31-41
Joseph Street
Blackburn, Victoria
895-2895
IN CANADA
Hewlett-Packard (Canada) Ltd. IN PEOPLE’S REPUBLIC