Agilent 8510C Spec Sheet

specifications
8510C
Network Analyzer
45 MHz to 110 GHz
Contents
Page
System performance characteristics . . . . . . . . .3
HP 8510E, Option 005, 45 MHz to 20 GHz . . . . . . . . . . .4
HP 8510SX, 45 MHz to 26.5 GHz . . . . . . . . . . . . . . . . . .5
HP 85107B Option 005, 45 MHz to 50 GHz . . . . . . . . . .6
HP 85107B Option 005 & 007, 45 MHz to 50 GHz . . . . .7
HP 85106D millimeter-wave systems,
33 GHz to 110 GHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
HP 8510XF, HP E7340A Option 005,
45 MHz to 85 GHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
HP 8510XF, HP E7350A Option 005,
45 MHz to 110 GHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
HP 85108A pulsed-RF system, 2 GHz to 20 GHz . . . . .12
HP 85108L pulsed-RF system, 45 MHz to 2 GHz . . . . .12
HP 8511A, HP 8511B frequency converters . . . . . . . .13
Accuracy considerations
when using RAMP sweep . . . . . . . . . . . . . . . . . . . . .14
HP 8510C network analyzer capabilities . . . . . . . . . . .15
HP 8510 System software . . . . . . . . . . . . . . . . . . . . . . .21
S-Parameter test sets, with common options . . . . . .22
HP 8510C Accessories
(calibration kits, verification kits, and cables) . . . . . . . .23
Excellence in Network Analysis
The HP 8510C vector network analyzer continues to provide the best performance to meet your new design and test challenges. With unmatched accuracy and convenience, the HP 8510C vector network analyzer makes broadband measurements from 45 MHz to 50 GHz in 2.4 mm coax, from 45 MHz to 110 GHz in 1.0 mm coax and from 33 GHz to 110 GHz in waveguide bands.
The HP 8510C network analyzer measures the magnitude, phase, and group delay of two-port networks to character­ize their linear behavior. Optionally, the HP 8510C network analyzer is also capable of displaying a network’s time domain response to an impulse or a step waveform by computing the inverse Fourier transform of the frequency domain response.
The HP 8510C family of systems is modular. Choose the source(s), S-parameter test set(s), and test accessories to meet your measurement applications. This data sheet presents performance specifications for most standard HP 8510 network analyzer systems. To acquire specifica­tions for system configurations not covered in this data sheet, refer to the HP 8510 Specifications and Performance Verification Software (part number 08510-10033) which is shipped with all HP 8510C network analyzers, test sets, and calibration kits.
2
3
System performance characteristics
This data sheet offers two types of performance numbers to describe the merit of any measurement system: specifications and supplemental characteristics.
Specifications describe the instrument’s warranted performance over the temperature range of 23 ˚C ± 3 ˚C.
Supplemental characteristics are typical but non-war­ranted performance parameters. These are denoted as “typical”, “nominal”, or “approximate.”
To specify the performance of an HP 8510 system, the data sheet lists each system’s dynamic range, measure­ment uncertainty and measurement port characteristics. The glossary below explains the major terms used in the System Performance section of this data sheet.
DYNAMIC RANGE has two descriptions: receiver dynamic range and system dynamic range. In either case, the noise floor (which affects P
min
as defined below) is measured with full two-port error-correction and 1024 averages.
System dynamic range = P
ref–Pmin
, where P
ref
is the nominal or reference power out of port 1 with maximum power delivered from the source and P
min
is the minimum power into port 2 that can be measured above the peaks of the system’s noise floor (10 dB above the average noise floor). System dynamic range is the amount of attenuation that can be measured from a 0 dB reference.
Receiver dynamic range = P
max–Pmin
, where P
max
is
the maximum power that can be input to port 2 before
0.1 dB compression of the test set and P
min
is the mini­mum power into port 2 that can be measured above the peaks of the system’s noise floor (10 dB above the aver­age noise floor). Receiver dynamic range is the system’s full usable dynamic range if the system is considered a receiver. An active device, such as an amplifier, may be required to realize the receiver dynamic range.
Calibration is the process of measuring standards which have fully defined models (and are thus called “known” standards) in order to quantify a network analyzer’s systematic errors based on an error model.
Calibration must be performed within the operating tem­perature specified for the calibration kit. For all calibra­tion kits the operating temperature is 23˚ C ±3˚ C. For a calibration to remain fully verifiable, the temperature of the network analyzer must remain within ±1˚ C around the initial measurement calibration temperature.
Error correction is the process of mathematically removing from the measurement those systematic errors determined by calibration.
MEASUREMENT UNCERTAINTY curves show the worst case uncertainty in reflection and transmission measurements using full two-port error correction with a specified calibration kit. This includes residual systematic errors, as well as system dynamic accuracy, connector repeatability, noise and detector errors. Cable stability and system drift are not included. All measurements assume step sweep mode with 1024 averages unless otherwise specified.
Furthermore, the graphs for reflection measurement uncertainty apply to a one-port device. The graphs for transmission measurement uncertainty assume a well­matched device (S
11
= S22= 0). In the phase uncertainty curves, the phase detector accuracy is better than 0.02 degrees, useful for measurements where only phase changes.
Using the HP 8510 specification and performance verifica­tion software, uncertainty curves can be calculated for non-idealized devices, and specifications can be edited for custom setups.
MEASUREMENT PORT CHARACTERISTICS indi­cate the RF performance of test set port leakages, mis­matches, and frequency response. The specification for the test set’s crosstalk does not include noise.
“Raw” port characteristics refer to the test set’s intrinsic, uncorrected performance. “Residual” port characteris­tics give the test set’s performance after error correction.
4
HP 8510E Option 005
45 MHz to 20 GHz
The following specifications describe the system performance for the HP 8510C network analyzer in the HP 8510E Option 005 (replaces HP 85052D fixed loads with 85052B sliding loads calibration kit) configuration. The system hardware includes the following:
Test set: HP 8514B S-parameter test set RF source: HP 83621B synthesized sweeper Calibration kit: HP 85052B 3.5mm calibration kit Calibration technique: Full two-port calibration with
sliding loads.
Dynamic range (for transmission measurements)
Frequency range (GHz)
0.045–2 2–8 8–20
Maximum power
measured at port 2 +20 dBm +11 dBm +10 dBm
Reference power
at port 1 (nominal) +2 dBm –2 dBm –6 dBm
Minimum power
measured at port 2 –66 dBm –95 dBm –95 dBm
Receiver dynamic range 86 dB 106 dB 105 dB System dynamic range 68 dB 93 dB 89 dB
Measurement uncertainty
Frequency range (GHz)
RESIDUAL 0.045–2 2–8 8–20 Directivity 48 dB 44 dB 44 dB
Source match 40 dB 33 dB 31 dB Load match 48 dB 44 dB 44 dB Reflection tracking ±0.003 dB ±0.003 dB ±0.006 dB Transmission tracking ±0.017 dB ±0.044 dB ±0.084 dB Crosstalk 89 dB 115 dB 110 dB
Frequency range (GHz)
RAW (Typical) 0.045–2 2–8 8–20 Directivity 23 dB 23 dB 14 dB
Source match 17 dB 15 dB 11 dB Load match 17 dB 15 dB 11 dB
Reflection measurements
Transmission measurements
System dynamic range
Magnitude Phase
Magnitude Phase
System performance
5
HP 8510SX
45 MHz to 26.5 GHz
The following specifications describe the system perfor­mance for the HP 8510C network analyzer with the HP 8510SX configuration. The system hardware includes the following:
Test set: HP 8515A S-parameter test set RF source: HP 83631B synthesized sweeper Calibration kit: HP 85052C 3.5mm precision
calibration kit
Calibration technique: TRL two-port calibration
Dynamic range (for transmission measurements)
Frequency range (GHz)
0.045–2 2–8 8–20 20–26.5
Maximum power
measured at port 2 +2 dBm +3 dBm +3 dBm –1 dBm
Reference power
at port 1 (nominal) –5 dBm –9 dBm –14 dBm –25 dBm
Minimum power
measured at port 2 –98 dBm –98 dBm –100 dBm –99 dBm
Receiver dynamic range 100 dB 101 dB 103 dB 98 dB System dynamic range 93 dB 89 dB 86 dB 74 dB
Measurement uncertainty
Measurement port characteristics
Frequency range (GHz)
RESIDUAL 0.045–2 2–8 8–20 20–26.5 Directivity 48 dB 50 dB 50 dB 50 dB
Source match 40 dB 50 dB 50 dB 50 dB Load match 48 dB 50 dB 50 dB 50 dB Reflection tracking ±0.003 dB ±0 dB ±0 dB ±0 dB Transmission tracking ±0.009 dB ±0.004 dB ±0.009 dB ±0.01 dB Crosstalk 114 dB 111 dB 106 dB 95 dB
Frequency range (GHz)
RAW (Typical) 0.045–2 2–8 8–20 20–26.5 Directivity 24 dB 24 dB 28 dB 27 dB
Source match 23 dB 23 dB 16 dB 14 dB Load match 23 dB 23 dB 16 dB 14 dB
Reflection measurements
Transmission measurements
System dynamic range
Magnitude Phase
Magnitude Phase
HP 85107B Option 005
45 MHz to 50 GHz
The following specifications describe the system performance for the HP 8510C network analyzer in the HP 85107B Option 005 (adds step attenuators and bias tees to HP 8517B test set) configuration for 50 GHz measurements. The system hardware includes the following:
Test set: HP 8517B S-parameter test set RF source: HP 83651B synthesized sweeper Calibration kit: HP 85056A 2.4mm calibration kit Calibration technique: Full two-port calibration with
sliding loads
Dynamic range (for transmission measurements)
Frequency range (GHz)
0.045–2 2–20 20–40 40–50
Maximum power
measured at port 2 +17 dBm +8 dBm +4 dBm –3 dBm
Reference power
at port 1 (nominal) +1 dBm –8 dBm –19 dBm –30 dBm
Minimum power
measured at port 2 –76 dBm –97 dBm –91 dBm –90 dBm
Receiver dynamic range 93 dB 105 dB 95 dB 87 dB System dynamic range 77 dB 89 dB 72 dB 60 dB
Measurement uncertainty
Reflection measurements
Reflection measurements
Measurement port characteristics
Frequency range (GHz)
RESIDUAL 0.045–2 2–20 20–40 40–50 Directivity 42 dB 42 dB 38 dB 36 dB
Source match 41 dB 38 dB 33 dB 31 dB Load match 42 dB 42 dB 38 dB 36 dB Reflection tracking ±0.001 dB ±0.008 dB ±0.02 dB ±0.027 dB Transmission tracking ±0.014 dB ±0.043 dB ±0.110 dB ±0.137 dB Crosstalk 99 dB 110 dB 93 dB 81 dB
Frequency range (GHz)
RAW (Typical) 0.045–2 2–20 20–40 40–50 Directivity 22 dB 18 dB 18 dB 18 dB
Source match 20 dB 12 dB 9 dB 9 dB Load match 20 dB 12 dB 9 dB 9 dB
Magnitude Phase
Magnitude Phase
System dynamic range
6
7
HP 85107B Options 005 and 007
45 MHz to 50 GHz
The following specifications describe the system performance for the HP 8510C network analyzer in the HP 85107B Options 005 (adds step attenuators and bias tees to HP 8517B test set) and 007 (adds high power and high dynamic range to HP 8517B test set) configuration for 50 GHz measurements. Specifications not shown are the same as those given for the HP 85107B Option 005. The system hardware includes the following:
Test set: HP 8517B Option 007 S-parameter test set RF source: HP 83651B synthesized sweeper Calibration kit: HP 85056A 2.4mm calibration kit Calibration technique: Full two-port calibration with
sliding loads
Measurement uncertainty
Transmission measurements
HP 85106D
33 GHz to 110 GHz
The following specifications describe the system performance for the HP 8510C network analyzer in the HP 85106D configuration for measurements from 33 to 110 GHz in four waveguide bands. For complete specifica­tions, refer to the HP 85106D system data sheet. The system hardware includes the following:
Test set (two each): HP 85104A series millimeter-wave test set modules
RF sources (two): HP 83621B synthesized sweepers Calibration kit: HP 11644A series waveguide
calibration kit
Calibration technique: TRL two-port calibration
Dynamic range (for transmission measurements)
Frequency range (GHz)
0.045–2 2–20 20–40 40–50
Maximum power
measured at port 2 +17 dBm +13 dBm +6 dBm –4 dBm
Reference power
at port 1 (nominal) +5 dBm +2 dBm –5 dBm –16 dBm
Minimum power
measured at port 2 –71 dBm –92 dBm –89 dBm –91 dBm
Receiver dynamic range 88 dB 105 dB 95 dB 87 dB System dynamic range 76 dB 94 dB 84 dB 75 dB
Dynamic range (for transmission measurements)
Frequency range (GHz) 33–50 40–60 50–75 75–110
Maximum power
measured at port 2 +12 dBm +10 dBm +10 dBm 0 dBm
Reference power
at port 1 (nominal) 0 dBm 0 dBm 0 dBm –3 dBm
Minimum power
measured at port 2 –87 dBm –87 dBm –75 dBm –79 dBm
Receiver dynamic range 99 dB 97 dB 85 dB 79 dB System dynamic range 87 dB 87 dB 75 dB 76 dB
System dynamic range
System dynamic range
8
HP 8510XF system performance HP E7340A Option 005 (45 MHz to 85 GHz)
The following specifications describe the system performance of the HP 8510XF system in the HP E7340A Option 005 configuration, from 0.045 GHz to 85 GHz. The following system configuration was used to generate the specifications:
Test set: HP E7342A Option 005, millimeter controller and two test heads, 45 MHz to 85 GHz RF sources: HP 83621B and 83651B synthesized sweepers (one each) Calibration kit: HP 85059A 1.0 mm precision calibration/verification kit Calibration techniques: SOLT to 50 GHz, and offset-shorts from 50 to 85 GHz
Dynamic range (for transmission measurements)
Frequency range (GHz)
0.045 - 2 2 - 18 18 - 40 40 - 50 50 - 65 65 - 75 75 - 85
Maximum power (in)
measured at port 2 0 dBm 0 dBm +10 dBm +10 dBm –3 dBm –3 dBm –3 dBm
Reference power (out)
at port 1 (nominal) 0 dBm 0 dBm –12 dBm –12 dBm –3 dBm –3 dBm –10 dBm
Minimum power (in)
measured at port 2 –74 dBm –104 dBm –84 dBm –84 dBm –80 dBm –80 dBm –70 dBm
Receiver dynamic range 74 dB 104 dB 94 dB 94 dB 77 dB 77 dB 67 dB System dynamic range 74 dB 104 dB 72 dB 72 dB 77 dB 77 dB 60 dB
Measurement port characteristics
Frequency range (GHz)
RESIDUAL 0.045 - 2 2 - 18 18 - 40 40 - 50 50 - 65 65 - 75 75 - 85 Directivity 30 dB 30 dB 26 dB 24 dB 28 dB 28 dB 28 dB Source match 27 dB 27 dB 23 dB 21 dB 28 dB 28 dB 28 dB Load match 27 dB 27 dB 23 dB 21 dB 28 dB 28 dB 28 dB Reflection tracking ±0.10 dB ±0.10 dB ±0.20 dB ±0.25 dB ±0.30 dB ±0.30 dB ±0.30 dB Transmission tracking ±0.273 dB ±0.273 dB ±0.429 dB ±0.669 dB ±0.322 dB ±0.340 dB ±0.360 dB
Frequency range (GHz)
RAW (Typical) 0.045 - 2 2 - 18 18 - 40 40 - 50 50 - 65 65 - 75 75 - 85 Directivity 20 dB 20 dB 15 dB 15 dB 13 dB 10 dB 10 dB Source match 20 dB 20 dB 15 dB 15 dB 13 dB 12 dB 12 dB Load match 11 dB 11 dB 10 dB 10 dB 10 dB 10 dB 10 dB
9
Reflection measurements
Transmission measurements
Magnitude Phase
Magnitude Phase
Magnitude Phase
Magnitude Phase
Measurement uncertainty
10
HP 8510XF system performance HP E7350A Option 005 (45 MHz to 110 GHz)
The following specifications describe the system performance of the HP 8510XF system in the HP E7350A Option 005 configuration, from 0.045 GHz to 110 GHz. The following system configuration was used to generate the specifications:
Test set: HP E7352A Option 005, millimeter controller and two test heads, 45 MHz to 110 GHz RF sources: HP 83621B and 83651B synthesized sweepers (one each) Calibration kit: HP 85059A 1.0 mm precision calibration/verification kit Calibration techniques: SOLT to 50 GHz, and offset-shorts from 50 to 110 GHz
Dynamic range (for transmission measurements)
Frequency range (GHz)
0.045 - 2 2 - 18 18 - 40 40 - 50 50 - 75 75 - 85 85 - 100 100 - 110
Maximum power (in)
measured at port 2 0 dBm 0 dBm +10 dBm +10 dBm 0 dBm 0 dBm 0 dBm 0 dBm
Reference power (out)
at port 1 (nominal) 0 dBm 0 dBm –12 dBm –12 dBm –7 dBm –12 dBm –12 dBm –12 dBm
Minimum power (in)
measured at port 2 –74 dBm –104 dBm –84 dBm –84 dBm –75 dBm –70 dBm –70 dBm –70 dBm
Receiver dynamic range 74 dB 104 dB 94 dB 94 dB 75 dB 70 dB 70 dB 70 dB System dynamic range 74 dB 104 dB 72 dB 72 dB 68 dB 58 dB 58 dB 58 dB
Measurement port characteristics
Frequency range (GHz)
RESIDUAL 0.045 - 2 2 - 18 18 - 40 40 - 50 50 - 75 75 - 85 85 - 100 100 - 110 Directivity 30 dB 30 dB 26 dB 24 dB 28 dB 28 dB 26 dB 26 dB Source match 27 dB 27 dB 23 dB 21 dB 28 dB 28 dB 26 dB 26 dB Load match 27 dB 27 dB 23 dB 21 dB 28 dB 28 dB 26 dB 26 dB Reflection tracking ±0.10 dB ±0.10 dB ±0.20 dB ±0.25 dB ±0.30 dB ±0.30 dB ±0.30 dB ±0.30 dB Transmission tracking ±0.273 dB ±0.273 dB ±0.429 dB ±0.669 dB ±0.322 dB ±0.360 dB ±0.451 dB ±0.451 dB
Frequency range (GHz)
RAW (Typical) 0.045 - 2 2 - 18 18 - 40 40 - 50 50 - 75 75 - 85 85 - 100 100 - 110 Directivity 20 dB 20 dB 15 dB 15 dB 11 dB 11 dB 11 dB 8 dB Source match 20 dB 20 dB 15 dB 15 dB 11 dB 11 dB 11 dB 10 dB Load match 11 dB 11 dB 10 dB 10 dB 10 dB 10 dB 9 dB 9 dB
11
Reflection measurements
Transmission measurements
Magnitude Phase
Magnitude Phase
Magnitude Phase
Magnitude Phase
Measurement uncertainty
12
HP 85108A Pulsed-RF System
2 to 20 GHz
The following specifications describe the system performance for the HP 8510C network analyzer in HP 85108A configuration. For complete specifications, refer to the HP 85108A system data sheet. The system hardware includes the following:
Test set: HP 85110A S-parameter test set RF sources (one each): HP 83622B and HP 83623L
synthesized sweeper
Calibration kit: HP 85052B 3.5 mm calibration kit Calibration technique: Full two-port calibration with
sliding loads
Other capabilities such as 50 GHz frequency coverage, operating within synchronized pulsed bias, and measure­ments such as spectrum analysis, noise figure and load pull can be added easily without degradation of the raw system performance.
HP 85108L Pulsed-RF system
45 MHz to 2 GHz
The following specifications describe the system performance for the HP 8510C network analyzer in HP 85108L configuration. For complete specifications, refer to the HP 85108L system data sheet. The system hardware includes the following:
Test set: HP 85110L S-parameter test set RF sources (two each): HP 83620B Option H80
synthesized sweeper
Calibration kit: HP 85050D 7 mm calibration kit Calibration technique: Full two-port calibration with
broadband loads
Dynamic range (for transmission measurements)
Frequency range (GHz) 2–8 8–18 18–20
Maximum power
(measured) at port 2
1
+11 dBm +11 dBm +11 dBm
Reference power
at port 1 (nominal) +0 dBm –1 dBm –2 dBm
Minimum power
(measured) at port 2 (pulsed) –64 dBm –63 dBm –62 dBm
(cw) –78 dBm –78 dBm –77 dBm
Receiver dynamic range (pulsed) 75 dB 74 dB 73 dB
(cw) 89 dB 89 dB 88 dB
System dynamic range (pulsed) 64 dB 62 dB 60 dB
(cw) 78 dB 77 dB 75 dB
Dynamic range (for transmission measurements)
Frequency range (GHz)
0.045–0.1 0.1–0.5 0.5–2
Maximum power
(measured) at port 2
2
+20 dBm +6 dBm +5 dBm
Reference power
at port 1 (nominal) +0 dBm +1 dBm 0 dBm
Minimum power
(measured) at port 2 (pulsed) –53 dBm –66 dBm –66 dBm
(cw) –81 dBm –95 dBm –96 dBm
Receiver dynamic range (pulsed) 72 dB 72 dB 72 dB
(cw) 101 dB 101 dB 100 dB
System dynamic range (pulsed) 53 dB 67 dB 66 dB
(cw) 81 dB 95 dB 95 dB
System dynamic range (pulsed)
1. This maximum power measurement assumes that the HP 85110A test set has its internal step attenuators set to 0 dB. The test set can handle up to 20 W (+43 dBm) of power if the step attenuators are activated and an isolator is installed (in the port 2 rear panel link).
2. This maximum power measurement assumes that the HP 85110L test set has its internal step attenuators set to 0 dB. The test set can handle up to 50 W (+47 dBm) of power if the step attenuators are activated and proper precautions are taken within the open architecture loops.
13
HP 8511A/8511B frequency converters
45 MHz to 26.5 GHz/45 MHz to 50 GHz
Description
Combining the HP 8511A or 8511B frequency converter with the HP 8510C network analyzer results in a four channel receiver/signal processor operating over a 45 MHz to 26.5 or 50 GHz frequency range. This system offers flexibility in the configuration of a user-supplied signal separation network to meet the needs of custom measurements. The HP 8511A/B contains four separate RF to IF converters all of which can operate over the entire dynamic range of the system. Either the a1 or a2 input must be defined as the reference channel to maintain phase lock and to track the RF source.
Dynamic accuracy
The following plots show the worst case magnitude and phase uncertainty due to IF residuals and detector inaccuracies. These plots exclude uncertainty due to noise, frequency response, directivity, port matches, crosstalk, and connector repeatability. Reference power is –20 dBm.
Measurement uncertainty
Magnitude
Input port characteristics
The following specifications show the uncorrected system characteristics at the four measurement ports.
Frequency range (GHz)
0.045–8 8–20 20–26.5 26.5–40 40–50
Impedance match
(all four ports)
HP 8511A 17 dB 15 dB 9 dB —— —— HP 8511B 17 dB 15 dB 9 dB 9 dB 7 dB
Frequency response tracking
1
HP 8511A ±1 dB ±1 dB ±1 dB —— —— HP 8511B ±1 dB ±1 dB ±3 dB ±3 dB ±3 dB
Crosstalk
2
HP 8511A 115 dB 116 dB 114 dB —— —— HP 8511B 116 dB 114 dB 107 dB 107 dB 105 dB
The following specifications describe the system performance for the HP 8510C network analyzer with the HP 8511A/B frequency converter.
Dynamic range (on all inputs)
Frequency range (GHz)
0.045–8 8–20 20–26.5 26.5–40 40–50
Maximum power
measured at port 2
HP 8511A –10 dBm –10 dBm –15 dBm —— —— HP 8511B –9 dBm –8 dBm –12 dBm –12 dBm –17 dBm
Reference power
at port 1 (nominal)
HP 8511A –20 dBm –20 dBm –20 dBm —— —— HP 8511B –20 dBm –20 dBm –20 dBm –20 dBm –20 dBm
Minimum power
measured at port 2
HP 8511A –112 dBm –115 dBm –113 dBm —— —— HP 8511B –113 dBm –113 dBm –106 dBm –106 dBm –104 dBm
Receiver dynamic range
HP 8511A 102 dB 105 dB 98 dB —— —— HP 8511B 104 dB 105 dB 94 dB 94 dB 87 dB
System dynamic range
HP 8511A 92 dB 95 dB 93 dB —— —— HP 8511B 93 dB 93 dB 86 dB 86 dB 84 dB
Phase
HP 8511A general information
Input ports
Connector type (all inputs): 3.5 mm (f)
Impedance: 50 nominal
HP 8511B general information
Input ports
Connector type (all inputs): 2.4 mm (f)
Impedance: 50 nominal
1. Ratio measurement of any two ports, excludes slope.
2. After error-correction, response and isolation calibration assumes no noise.
14
Accuracy considerations when using ramp sweep
The uncertainty values for the preceding systems assume that the microwave source operates in the synthesized step sweep mode. Selecting ramp sweep saves time; the HP 8510C allows switching between step and ramp sweep without the need to recalibrate
1
.
The frequency accuracy of the HP 8510 in ramp sweep mode is determined by the swept frequency accuracy of the source.
For HP 8360 series synthesizers, the swept frequency accuracy is summarized in the following table:
Sweep width
Accuracy
f 26.5 GHz f >26.5 GHz (% of span, or MHz)
n x 10 MHz n x 10 MHz 0.1% ± time base accuracy >n x 10 MHz and >n x 10 MHz and
400 MHz 800 MHz 1% >400 MHz and >800 MHz and
4 GHz 8 GHz 4 MHz/8 MHz >4 GHz >8 GHz 0.1%
Frequency range n (Multiplier)
10 MHz to <2 GHz 1 2 to <7 GHz 1 7 to <13.5 GHz 2
13.5 to <20 GHz 3 20 to <26.5 GHz 4
26.5 to <38 GHz
2
6
38 to 50 GHz 8
The plots below indicate the typical swept mode accuracy.
Phase errors due to device electrical length
Ramp sweep adds phase uncertainty to the measurement of
electrically long devices. The phase uncertainty ø is
given by the following equation:
ø = (–360/c) x F x L
where c is the propagation velocity in a vacuum (3 x 10
10
cm/sec), F is the frequency accuracy specifica-
tion of the synthesizer in ramp mode, and L is the electri­cal length of the device under test. The following graph shows this uncertainty for the HP 8360 series synthesiz­ers in ramp sweep.
Hewlett-Packard recommends the HP 8360 series synthe­sized sweepers in step sweep mode whenever measuring phase or group delay of electrically long devices, or when­ever the highest system accuracy is required. Ramp sweep is recommended for measurements of electrically short devices, or for applications where maximum trace update rate is desired (for example, tuning).
1. Ramp sweep mode is not available in the HP 8510XF systems.
2. This band is 26.5 to 40 GHz on the HP 83640A.
15
Measurement
Number of display channels: Two display channels are
available. Number of display parameters: The four basic param­eters, S11, S21, S12, S22, can be displayed for either selected channel in either a “four quadrant” or an “over­lay” format. Measurement parameters: S11, S21, S12, S22. Parameters may be redefined by the user for special applications. Conversion to Z1 (input impedance), Z2 (output impedance), Y1 (input admittance), Y2 (output admittance), and 1/S is also provided. Domains available: Frequency, time (Option 010), pulse profile
2
(Option 008), auxiliary voltage (rear panel output
acting as device stimulus, range is ±10 VDC), and power
3
(sweep power level at a CW frequency).
Formats
Cartesian: log/linear magnitude, phase, group delay, SWR, real part of complex parameter, imaginary part of complex parameter. Smith chart: Marker format can be selected as log magni­tude, linear magnitude, R + jX, or G + jB. Polar: Marker format can be selected as log magnitude, linear magnitude, phase, or real and imaginary. Data markers: Five independent data markers read out and display the value of the formatted parameter and stimulus (frequency, time, or auxiliary voltage).
Marker functions
Marker search: Specific trace values can be located, such as MAX, MIN, and target (for example–3.00 dB point) Discrete/continuous: Markers can indicate data at actual data points or they can interpolate between data points to allow the setting of a marker at an exact stimulus value. Delta marker: Marker readout shows difference between active marker and the reference marker (any marker can be used as the reference).
Group delay characteristics
Group delay is computed by measuring the phase change within a specified step (determined by the frequency span and the number of points per sweep). Aperture: Determined by the frequency span, the num­ber of steps per sweep, and the amount of smoothing applied. Minimum aperture = (frequency span)/(# points –1) Maximum aperture = 20% of the frequency span Range: The maximum delay is limited to measuring no more than ±180 degrees of phase change within the minimum aperture.
Range = 1/(2 x minimum aperture)
For example, with a minimum aperture of 200 kHz, the maximum delay that can be measured is 2.5 µsec. Accuracy: The following graph shows group delay accu­racy at 20 GHz with an HP 8514B test set and an HP 83621A operating in stepped sweep mode. Insertion loss is assumed to be zero.
In general, the following formula can be used to deter­mine the accuracy, in seconds, of a specific group delay measurement.
0.003 x Phase accuracy (deg) + delay (sec) x linearity (Hz) Aperture (Hz)
Depending on the aperture and the device length, the phase accuracy used is either incremental phase accuracy or worst case phase accuracy. The above graph shows this transition.
Source control
All source control is provided from the HP 8510C front panel.
Compatible sources
HP 8360 series synthesized sweeper HP 8340A/B synthesized sweeper
4
HP 8341A/B synthesized sweeper
4
HP 8350B sweep oscillator with HP 835xx RF plug-in
4
(ramp sweep mode only) Sweep limits: Set start/stop or center/span of the stimulus parameter (frequency, time, or auxiliary voltage). Measured # points per sweep: Selectable as 51, 101, 201, 401, or 801 points. In frequency list mode, the number of points can range from 1 to 792.
Sweep modes
Ramp sweep
5
(analog) Stepped sweep (available with all sources except HP 8350B): A faster version of step sweep, called “quick step”, is select­able when using HP 8360 series synthesized sweeper. Frequency list sweep: Define up to 30 different arbitrary sub-sweep frequency ranges by specifying start/stop, center/span, or CW sweeps. Define the number of points or step size for each range. Display all segments or a single segment on-screen. All frequencies are synthesized if using the HP 8340/41 or HP 8360 series synthesized sweepers. Frequency domain only.
Single point (single frequency) Fast CW mode (HP-IB only): Raw data (real and imagi-
nary) is sent immediately to HP-IB as soon as it is taken. Display is blanked in this mode. The source is phase­locked once when entering this mode, but is not re-phase­locked at each point. Must be triggered externally (TTL). Data is available approximately 500 µsec after the trigger pulse is received.
1. The symbol denotes a new feature or capability due to the HP 8510C firmware revision 7.0.
2. Pulse profile domain is not available in the HP 8510XF systems.
3. Power domain requires HP 8360 series sources. Sources with firmware revisions prior to 01 Oct 93 require updating.
4. HP 8340, 8341 and 8350 series sources are not compatible with the HP 8510XF systems.
5. Ramp sweep mode is not available in the HP 8510XF systems.
System capabilities
1
16
Power sweep: From power domain, sweep power at a CW frequency. When combined with the receiver calibration feature, power sweep allows quicker and more accurate absolute-power measurements of an amplifier’s 1-dB gain compression point. Alternate sweep: The two channels, including markers, may be coupled
1
(same source parameters) or uncoupled (different source parameters). Sweep time: Minimum sweep time is automatically selected, depending on the number of points per sweep (and the averaging factor if in stepped mode). Longer sweep times may be entered by the user from 0.1 to 100 seconds. Source power: Set source power (dBm) or power slope (dBm/GHz). For the S-parameter test sets with built-in attenuators, the port 1 or port 2 signal level can be con­trolled by setting the internal attenuator of the test set. Flat power is achievable at the test port when using the user-flatness feature of the HP 8360 series synthesizers. An HP E4418A power meter with the HP 8480 series of power sensors are required to measure power for flatness correction. Multiple frequency control mode: In this mode, the HP 8510C controls up to three frequency ranges indepen­dently: the frequency of the primary source (HP 83621A synthesized sweeper), the frequency of a secondary source (may be phase-locked HP 8350B sweep oscillator), and the frequency of the network analyzer receiver. Frequency control: All frequency ranges can be separately defined as functions of the device under test frequency, by specifying a multiplier (a ratio of integers) and an offset for each frequency. Definition storage: The active multiple frequency mode parameters are stored in non-volatile memory and may be saved on disc with the hardware configuration. Sweep modes: All sweep modes can be used in multiple frequency mode. External LO phase-lock control: In applications where an external LO is used in place of a test set, LO phase­lock control is provided to phase-lock a sweep oscillator with DC FM capability (for example, HP 8350B) to a synthesized sweeper (for example, HP 83621A).
Power leveling and power control
2
Power leveling capability comes as part of the standard HP 8510XF systems. With power leveling, power levels at the test ports are controlled with a typical accuracy of ±1.0 dB with a control range greater than 20 dB. Power leveling modes: These modes are available in both RF and LO power control. System leveling: Power leveled at the test ports and is entirely controlled by the HP 8510XF system. This is also the normal operating mode. Internal leveling: Power leveled at the output port of the source.
External leveling: Leveling is achieved using an external detector. Leveling off: Source set to the unleveled mode. Power leveling detections modes: These modes are available under system leveling. Each mode will deter­mine how the "unleveled" condition is detected. Always: The HP 8510C polls for errors during every sweep. Smart: The HP 8510C polls for errors during the first sweep following a change in frequency, and thereafter only if an error was detected during the first sweep. this is the default mode. Once: The HP 8510C polls for errors only during the first sweep following a change in frequency. Never: The HP 8510C does not poll for errors during any sweep. Power control: Set port power (dBm) or power slope (dBm/GHz). With system leveling selected, the HP 8510XF system will control the RF source, the millimeter-wave controller and the test heads to deliver the user requested power to the test ports.
Vector error correction techniques
Calibration types available: Various calibration types
are available. Once calibrated, the frequency limits may be narrowed using the FREQUENCY SUBSET feature of the HP 8510C. Response/isolation calibration: Compensates for frequency response and directivity (reflection) or frequency response and crosstalk (transmission) of test sets. Requires a short or open circuit and load termination (reflection) or through connection and load termination (transmission). One port calibration: Correction of test set port 1 or port 2 directivity response and source match errors. Requires three known standards, for example, open, short, and load (fixed, sliding, or offset) terminations. Two port calibration: Compensates for port 1 and port 2 directivity, source match, reflection frequency response, load match, transmission, frequency response and crosstalk. Crosstalk (isolation) calibration can be eliminated. Full two-port (traditional): For use with an S-parameter test set, requires three known standards at each port, for example, short, open (or offset short), and load (fixed, sliding, or offset) terminations. A through connection is also required. One path two-port calibration: A two-port calibration for one port Reflection/Transmission test sets, such as with the millimeter-wave systems. Provides a full two-port error-corrected measurement when the test device is turned around and measured in both directions.
1. In the HP 8510XF systems, the two channels are coupled (uncoupled channels is not available).
2. The features or capabilities listed under Power Leveling and Power Control are only available in the HP 8510XF systems.
17
Thru-reflect-line (TRL): A two-port calibration which requires a through connection, a reflection standard, and a reference transmission line. S-parameter test sets only. Line-reflect-match (LRM): A two-port calibration similar to TRL except it uses fixed loads and not transmission line(s) as its impedance reference. LRM is a particularly convenient broadband calibration for non-coaxial environ­ments with accuracy as good as TRL. S-parameter test sets only. Adapter removal calibration: A two-port calibration for noninsertable devices, such as those with identical sexed connectors on both ports. S-parameter test sets only. Two-port to one-port calibration: Create and store a separate one-port calibration set for either port 1 or port 2 from an existing two-port calibration. Reference plane extension: Redefine the plane of mea­surement reference (zero phase) to other than that estab­lished at calibration. A new reference plane is defined in seconds of delay from the test set port and ranges between ±1 second. Connector compensation: Compensates for errors between dissimilar but matable connectors such as
3.5 mm and SMA. Receiver cal: Adjusts non-ratio'ed receiver inputs to absolute power levels; displays absolute power in dBm; requires reference sweep of known source power (using flatness correction). Set Zo: Can redefine the characteristic impedance of a
measurement to other than 50 .
Data averaging: Similar to a variable bandwidth IF filter, this function computes the running average of either a number of data traces in ramp mode or data points in stepped mode. Averaging factors range from 1 to 4096 in powers of 2. In stepped sweep mode, each data point is averaged before being displayed. Trace smoothing: Similar to variable bandwidth video filter, this function computes the moving average of adja­cent data points in a single trace. Smoothing aperture ranges from 0.125% to 20% of the trace width.
Data hardcopy
Data plotting: Hard copy plots are automatically
produced by the HP 8510C when used with an HP-GL compatible plotter using either the system bus (HP-IB compatible) or serial output interface (RS-232-C). Plotter functions: Plot trace(s), graticule(s), marker(s), or text with any plotter pen. Operating and system parameters can also be plotted. Formats: Full or quarter page plots of any parameter. Plot all 4 S-parameters: Plots all four S-parameters in 4-quadrant format. Plotter/printer buffer: Retain trace update of HP 8510 while plotting/printing previous data via either of the two serial output interface (RS-232-C) ports. Total buffer memory available is 400 Kbytes for serial output port 1 and 100 Kbytes for serial output port 2.
Data listings: For printing and plotting, the HP 8510C can send data, via either its HP-IB
1
or its RS-232 inter­face, to the HP ThinkJet, QuietJet, DeskJet, LaserJet, or PaintJet graphics printers. Measurement data, operating parameters and system parameters can all be printed or plotted.
Display control
Display type: Color (raster scan), 7.5” diagonal; display
graticule, 5.5” diagonal. Title: Add custom titles (49 characters maximum) to the display of the HP 8510C. Titles are plotted when making hardcopies of displayed measurements. Adjust display: Control the intensity and background intensity of the display. Also, customize the color, tint, and brightness of the data traces, memory traces, refer ence lines, graticules, text, and warning messages. Select from 16 colors. Default colors can be recalled along with one set of user-defined display values. Control is in the percent of full range. Limit lines: Define up to eight test limit segments per parameter per channel; segments may be any combination of flat lines, sloping lines, or discrete points; limit testing gives pass/fail decision on each sweep. CRT formats: Single channel, dual channel overlay (both traces on one graticule), dual channel split (each trace on separate graticules), four parameter overlay or split.
Trace functions
Display data: Display current measurement data, mem­ory data, or current measurement and memory data simultaneously. Trace math: Vector math (+, –,
×, ÷) of current linear
measurement values and memory data.
Scale resolution
Magnitude: Log format (dB/div): 0.001 to 500
Linear format (units/div): 10×10
-12
to 500
Phase: Cartesian (degrees/div): 10×10
-12
to 500 Polar (degrees/display graticule): 45 Reference value: Ranges between ±500 units (dB, degrees, seconds, etc.) Reference position: Ranges from the 0 (bottom) to 10 (top) graticule position. Auto: Automatically selects scale resolution and reference value to center the trace on the CRT graticules for easy viewing. Electrical delay: Offset measured phase or group delay data by a defined amount of electrical delay, in seconds. Operates similar to an electronic line stretcher. Amount of electrical delay can range between ±1 second. Electrical delay can also be displayed in electrical length (meters) or physical length by entering the appropriate VELOCITY FACTOR, the propagation velocity relative to the speed of light. Waveguide delay: Non-linear electrical delay used to remove the effects of dispersion in waveguide. Table delay: Allows user to input an array or real and imaginary pairs to correct for dispersion in non-coaxial media such as microstrip or coplanar waveguide.
1. When printing/plotting from the HP 8510 system bus (HP-IB compatible) port to a Centronics printer/plotter, use an HP ITEL-45CHVU or HP ITEL-45CHVE HP-to-Centronics Converter.
18
Storage
Internal memory Instrument state: Eight instrument states can be stored
in non-volatile memory via the SAVE menu. They can then be recalled via the RECALL menu. Instrument states include all control settings, memory trace data, active list frequency tables, active calibration coefficients, and custom display titles. Register 8 is reserved for the power-up state, which can be defined by the user. Hardware configurations: One hardware configuration is stored in active non-volatile memory. This configuration is not changed at instrument preset. The hardware config­uration includes all instrument addresses and the multiple frequency mode parameters. Data traces: Eight traces of data can be stored in the trace memories. Traces 1-4 are stored in non-volatile memory. Calibration sets: Eight separate calibration sets may be stored in non-volatile memory. If any 801-point full two­port calibrations are stored, storage may be limited to as few as four calibration sets. Calibration kits: Two calibration kits, including user­modified kits can be stored in the HP 8510 internally allo­cated memory. An internally stored kit is written over when another calibration kit is loaded in the same data storage location. Calibration kits can also be stored to disk. Internal disk drive: The built-in disk drive can be used to store and retrieve different types of data on a 3.5 inch disk. Data files can be stored in either the HP LIF or MS-DOS
®
formats. Diskettes of double sided format or high density format are recommended. External disk drive: Data can also be stored on disk using an external disk drive with command subset SS/80. Data files are stored in Hewlett-Packard’s standard LIF or MS-DOS format.
Disk storage memory requirements
Type of Data to be Stored Memory Required (Kbytes)
Calibration set (full two-port, 801 pts) 234 Calibration kit 2 Instrument state 7 Hardware state 0.5 Machine dump 400 Data data (201 pts)
1 S-parameter 5.5
4 S-parameters 20 Data formatted, raw or memory (201 pts) 5.5 User display 33
Time domain (Option 010)
Description
With the time domain option, data from transmission or reflection measurements is converted from the frequency domain to the time domain using the inverse Fourier transform and presented on the CRT display. The time domain response shows the measured parameter value versus time. Markers may also be displayed in electrical length (or physical length if the relative propagation velocity is entered).
Time stimulus modes
Two types of time domain stimulus waveforms can be sim­ulated during the transformation — a step and an impulse. Although these waveforms are generated mathe­matically with the inverse FFT, the results for linear cir­cuits are the same as would be obtained if the actual time waveforms had been applied and measured. Low pass step: This stimulus, similar to a traditional Time Domain Reflectometer (TDR) waveform, is used to measure low pass devices. Transforming to time low pass requires a sweep over a harmonic set of frequencies includ­ing an extrapolated DC value. The step response is typi­cally used for reflection measurements only. The low pass step waveform displays a different response for each type of impedance (R, L, C), giving useful information about the discontinuities being measured.
Response resolution
1
: In low pass step mode, response resolution is determined by the step rise time (10% to 90%) of the time stimulus. This depends on both the frequency span and the window used (see Windows):
Low pass impulse: This stimulus is also used to measure low pass devices, and is the mathematical derivative of the low pass step response. Transforming to time low pass requires a sweep over a harmonic set of fre­quencies including an extrapolated DC value. The time domain response shows changes in the parameter value versus time. The impulse response can be used for reflec­tion (fault location) or transmission measurements.
Response resolution
1
: In low pass impulse mode, response resolution is defined by the 50% impulse width of the time stimulus. This depends on both the frequency span and the window used (see Windows):
1. Response resolution is the ability to resolve two closely spaced responses of equal magnitude. For example, in time impulse response, two equal responses that are separated in time by less than one impulse width cannot be resolved as two separate responses.
19
Bandpass impulse: The bandpass impulse simulates a pulsed RF signal (with an impulse envelope) and is used to measure the time domain response of band-limited devices. The start and stop frequencies are select-able by the user to any values within the limits of the source and test set used. The bandpass time domain response shows changes in the parameter values versus time. Bandpass time domain responses are useful for both reflection and transmission measurements.
Response resolution
1
: In bandpass impulse mode, response resolution is defined by the 50% impulse width of the time stimulus. This depends on both the frequency span and the window used (see Windows):
Time domain range
The time domain range, the range over which the display is free of response repetition, depends on the frequency span and the number of points as shown in the following graph
2
:
Range resolution: Range-resolution is the ability to locate a single response in time. It is a function of the time span selected and the number of data points. Range­resolution = time span/(number of points –1). Range­resolution better than 1 mm (3 ps) can typically be achieved. This is determined by source stability.
Windows: The windowing function can be used to modify (filter) the frequency domain data to reduce overshoot and ringing in the Time Domain response. Three types of windows are available-minimum, normal, maximum. Typical effective impulse width and sidelobe response to each type of window are shown in the table below:
Kaiser Window bessel Impulse Sidelobes type parameter width (relative to peak)
Minimum 0 Minimum –15 dB Normal 6 1.5x Minimum –50 dB Maximum 13 2.5x Minimum –90 dB
Gating: The gating function can be used to isolate individual Time Domain responses. In converting back to the frequency domain the effects of the responses outside the gate are removed. The location and span of the gate can be controlled by setting either the gate center position and time span or by setting the gate start and stop times.
Measurement throughput summary
The following table shows typical measurement times for an HP 8510C system with full two-port error correction.
Number of points 51 101 201 401 801
Measurement
3
Ramp sweep
4
270 ms 340 ms 470 ms 740 ms 1.3 s
Stepped sweep
Avg factor = 1 1.25 s 2.5 s 5 s 10 s 20 s Avg factor = 128 6.25 s 12.8 s 24 s 49 s 106 s
Time domain conversion550 ms 100 ms 200 ms 400 ms 800 ms HP-IB data transfer to computer
6
Internal format 20 ms 30 ms 50 ms 100 ms 200 ms ASCII format 240 ms 460 ms 900 ms 1.8 s 3.6 s
IEEE 754 floating point format
32 bit 20 ms 40 ms 80 ms 160 ms 320 ms 64 bit 40 ms 80 ms 150 ms 300 ms 590 ms
Remote programming
Interface: HP-IB interface operates according to IEEE
488-1978 and IEC 625 standards and IEEE 728-1982 recommended practices. System interface: The HP 8510C system bus is an HP-IB port used exclusively by the HP 8510C to control and extract information from the other instruments in the system such as the RF source, test set, and the digital plotter. Addressing: The HP-IB addresses of the HP 8510C and all instruments connected to the HP 8510 system inter­face can be verified or set from the HP 8510C front panel via the LOCAL menu. Addresses can range from 0 to 30 decimal.
1. Response resolution is the ability to resolve two closely spaced responses of equal magnitude. For example, in time impulse response, two equal responses that are separated in time by less than one impulse width cannot be resolved as two separate responses.
2. In low pass mode (step or impulse), range is limited by the minimum spacing between frequency domain data points (45 MHz, or 22.5 MHz with 801 points). This limit is labeled on the graph as “low pass limit”.
3. Includes system retrace time, but does not include source bandswitch times (typically 50 msec each). Time domain gating is assumed off.
4. If averaging is used, multiply the above ramp sweep measurement times by the averaging factor to get the total time.
5. Option 010 only, gating off.
6. Measured with an HP 9000 Series 300 computer. Single point data transfers can be accomplished in approximately 1 msec per point using the fast CW mode.
20
The factory selected addresses are the following:
HP 8510C network analyzer 16 Source #1 19 Source #2 18 Test set 20 Plotter 05 Printer 01 Disc drive 00
Pass-through address 17 Pass-through address: Instruments connected to the HP 8510C system bus may be accessed via the pass­through address.
Transfer formats:
Binary (internal 48 bit floating point complex format) ASCII 32/64 bit IEEE 754 floating point format Form 5 (for transfer to PC’s)
Interface function codes:
SH1, AH1, T6, TE0, L4, LE0, SR1, RL1, PPO, DC1, DT0, CO, E1
General characteristics
Rear panel connectors Sweep in: Input for 0 to 10 V sweep voltage from
compatible sweep oscillator. Stop sweep: Input and output for stop sweep signal from compatible sweep oscillator. 10 MHz in: Input for external 10 MHz reference.
Input level: –10 dBm to +20 dBm, typical Input frequency accuracy: ±0.005% (50 ppm)
20 MHz out: Output of internal 20 MHz reference oscillator. Frequency accuracy: ±0.01%, typical Analog ±10V: Settable output voltage used for auxiliary voltage domain measurements, or analog output for ana­log recorders.
Range: –9.995 to +10.000 volts Linearity: ±0.1%, typical Resolution: 4.88 mV, nominal
Output impedance: 1k
External trigger: TTL input to trigger acquisition of single data point. Delay is included to equalize the single path delay from the test port. Data is taken on negative­going TTL pulse. Pulse repetition period: 1 msec minimum with no averaging. If averaging, add (200 µs x averaging factor). Pulse width: 1 µs minimum.
Environmental
Operating conditions: Temperature: 5 ˚C to 40 ˚C Non-operating conditions: Temperature: –40 ˚C
to +75 ˚C
Power
47.5 to 66 Hz: 100, 120, 200, 240 Vac, ±10%
HP 8510C: 460 VA, maximum Test sets: 145 VA, maximum HP 8510XF millimeter-subsystem: 500 VA, maximum
Weight
HP 8510C: Net, 42 kg (92 lb); shipping, 52 kg (114 lb) HP 8511A/B: Net, 15 kg (33 lb); shipping, 18 kg (40 lb) HP 8514B: Net, 17 kg (38 lb); shipping, 20 kg (45 lb) HP 8515A: Net, 19 kg (41 lb); shipping, 22 kg (48 lb) HP 8517B: Net, 15.5 kg (34.2 lb); shipping, 18.7 kg (41.2 lb) HP 85105A: Net, 20.4 kg (45 lb); shipping, 24.5 kg (54 lb) HP 85104A series modules (each): Net, 6 kg (23 lb);
shipping 6.8 kg (15 lb) HP 8510XF controller: Net, 20.4 kg (45 lb); shipping,
24.5 kg (54 lb) HP 8510XF test heads (each): Net, 11.34 kg (25 lb); shipping, 13.6 kg (30 lb)
Dimensions
TOP
SIDE
REAR
50.8 cm (20 in)42.4 cm (16.7 in)
13.3 cm (5.25 in)
44.5 cm (17.45 in)
(Allow 10 cm, 4 inches
clearance at rear panel
for interconnect cables)
21
System Software
HP 85161B measurement automation software
Description: This software is designed specifically to
operate on an HP 9000 Series 200 or 300 computer for automation of the HP 8510 system including millimeter­wave systems. The software complements the hardware, providing calibration, measurement, and data output capabilities with a minimum of operator interaction. Measurement data can be stored in binary format or in a data file compatible with many CAE design programs.
Operating requirements
Compatible with HP 9000 Series 200 or 300 computer with the following: BASIC Operating System (5.0 or higher) RAM memory (including BASIC): 2 Mbytes, or a PC running HP Basic for Windows revision 6.3 or higher under Windows (3.1/95/NT)
HP 85071B materials measurement software
Description: The HP 85071B software takes broad-band
S-parameter measurements of dielectric and magnetic materials and determines their electromagnetic properties.
The software calculates both the complex permittivity ε
r
(or dielectric constant) and permeability µr, including their loss factors. Depending on the network analyzer and fixtures used, measurements can extend from 100 MHz to 110 GHz. The software offers the choice of four algorithms, each designed to address specific measurement needs.
Operating requirements Standard: Requires MS-DOS on an HP Vectra (or any
100%-compatible PC-AT computer) compatible with Microsoft
®
Windows 2.11 with mouse. Requires >20 Mbyte hard disk and >640 Kbytes RAM. Option 300: Substitutes HP BASIC software for the standard version for operation with HP 9000 Series 300 controllers. Requires BASIC 5.0 or higher and 2 Mbytes of RAM. Refer to the HP 85071B data sheet for more information.
HP 85070B dielectric probe kit
The HP 85070B dielectric probe kit allows convenient non-destructive testing of materials using the open-ended coaxial probe method. The probe, together with its own dedicated software, determines the complex permittivity of a wide variety of liquids, semi-solids, and solids. Since the probe kit measures only permittivity, only non­magnetic materials should be measured. Measurements are efficient and cost-effective because the testing is non-destructive and there is no need for sample preparation or special fixtures. Refer to the HP 85070B data sheet for more information.
22
S-parameter test sets
Description
Combining the HP 8510C network analyzer with an HP 8514B, 8515A, or 8517B results in a system for making full S-parameter measurements. The dual port architecture of the test sets develops a separate reference channel for each incident port. RF switching is done with a single built-in electronic switch.
Test set general information
Ordering information
The following options are available on the test sets: Option 001: Add IF switching. Allows four test sets with different addresses to be connected to the HP 8510 at the same time. The test set in use is selected from the HP 8510C front panel. The 20 MHz IF signal is daisy­chained from the test sets to the HP 8510. IF switching is performed automatically without reconnections. Option 002: Delete step attenuators and bias tees (HP 8514B, 8515A, 8517B only). If attenuators are not required, but bias is required, bias can be applied exter­nally using the HP 11612A/B bias tees.
Option 003: Forward configuration (HP 8514B only). Forward coupler configuration; optimization for forward dynamic range. Option 004: High power configuration (HP 8517B only). Optimized for testing of high power devices (up to 1 watt) by moving the port 2 step attenuator before the b2 sampler. Option 007: High power and high dynamic range configu­ration (HP 8517B only). Adds broadband amplifiers at the input, and before each sampler. Requires HP 8510C firmware revision 7.0.
HP 8514B HP 8515A HP 8517B HP 85110A HP 85110L HP 85105A/HP 85104A
Frequency range (GHz) 0.045 to 20 0.045 to 26.5 0.045 to 50 2 to 20 0.045 to 2 33 to 50 40 to 60 50 to 75 75 to 110
Test ports (port 1 or 2)
Nominal operating power level (dBm) 2 to –6 –5 to –25 +2 to –29 0 to –3 0 0 0 0 –3
+5 to –16
1
Connector type 3.5 mm (m) 3.5 mm (m) 2.4 mm (m) 3.5 mm (m) 7 mm WR-22 WR-19 WR-15 WR-10
Impedance, DC bias 50 nominal, 500 mA, 40 Vdc maximum Waveguide impedance, no bias
Attenuation range
(incident signal) 0 to 90 dB, in 10 dB steps 0 to 60 dB, in 0 to 90 dB, in 10 dB steps N/A N/A N/A N/A
10 dB steps
RF input connector (rear panel)
Max. input power +16 dBm +14 dBm +16 dBm +14 dBm +14 dBm +13 dBm +13 dBm +13 dBm +13 dBm Connector type 3.5 mm (f) 3.5 mm (f) 2.4 mm (f) 3.5 mm (f) 3.5 mm (f) 3.5 mm (f) 3.5 mm (f) 3.5 mm (f) 3.5 mm (f)
1. HP 8517B Option 007.
23
HP 8510C accessories
A wide range of accessories support the HP 8510C network analyzer including calibration kits, verification kits, cables and adapters in 7 mm, 3.5 mm, 2.92 mm. Type-N, 2.4 mm, 1.85 mm and 1.0 mm coax and in the standard waveguide bands. The standards used in the
3.5 mm, Type-N, and 2.4 mm calibration and verification kits use precision slotless connectors
1
(PSC-3.5, PSC-N,
and PSC-2.4).
Calibration kits
Before a network analyzer can make error-corrected mea­surements, the network analyzer’s systematic errors must be measured and removed. Calibration is the process of quantifying these errors by measuring “known”, or preci­sion standards. The calibration kits listed below contain the precision mechanical standards required to calibrate an HP 8510 system. For calibrating an HP 8510 system in the 7 mm, 3.5 mm, 2.92 mm, Type-N, 2.4 mm, 1.85 mm, or
1.0 mm interface, mechanical calibration kits all contain the following:
• Calibration standards to perform full-two port calibration
• Torque wrenches for properly connecting the standards
• Adapters to change the sex of the test port
• A disk for loading the standard definitions into the network analyzer. Option 002
2
provides calibration standard definitions on magnetic tape for use with the HP 8510A/B.
Calibration kits
Three classes of mechanical calibration kits are available: Standard kits contain open circuits, short circuits, and both fixed and sliding terminations in both sexes for all connector types (except 7 mm, a sexless connector). Connector gauges are included in these kits for maintain­ing each standard’s connector interface.
Precision kits have precision 50 airline(s) for
performing the Thru-Reflect-Line (TRL) calibration
2
, the most accurate error-correction technique for coaxial measurements. These kits also contain the open circuit, short circuit, and fixed terminations used for traditional open-short-load calibration techniques. All precision kits except the HP 85052C and 85059A include gauges. All waveguide calibration kits are precision kits and support TRL calibration in their waveguide band. Most kits contain two straight waveguide sections with precision flanges, a flush short circuit, a precision waveguide line section, and either sliding or fixed terminations. Economy kits include the open circuit, short circuit, and fixed termination standards but not sliding terminations or gauges. Gauges can be ordered separately.
Frequency Residual Residual source Cal kit type range (GHz) Connector Return loss, Return loss, Return loss (dB), directivity (dB) match (dB) and name f
min–fmax
type fixed load sliding load airline @ f
max
@ f
max
@ f
max
STANDARD HP 85050B 0.045–18 7 mm 52 dB, DC–2 GHz 52 dB, 2–18 GHz –––– 52 41 HP 85052B 0.045–26.5 3.5 mm 44 dB, DC–3 GHz 44 dB, 3–26.5 GHz –––– 44 31 HP 85054B 0.045–18 Type-N 48 dB, DC–2 GHz 42 dB, 2–18 GHz –––– 42 32 HP 85056A 0.045–50 2.4 mm 42 dB, DC–4 GHz 36 dB @ 50 GHz –––– 36 32 PRECISION HP 85050C 0.045–18 7 mm 38 dB, DC–18 GHz –––– >60 60 60 HP 85052C 0.045–26.5 3.5 mm 46 dB, DC–2 GHz –––– 50 50 50 HP 85059A 0.045–110 1.0 mm 24 dB, DC–50 GHz –––– –––– 24 21 HP X11644A
1
8.2–12.4 WR-90 42 dB, 8.2–12.4 GHz –––– 50 50 50
HP P11644A
1
12.4–18 WR-62 42 dB, 12.4–18 GHz –––– 50 50 50
HP K11644A
1
18–26.5 WR-42 42 dB, 18–26.5 GHz –––– 50 50 50
HP R11644A 26.5–40 WR-28 –––– 46 dB 50 50 50 HP Q11644A 33–50 WR-22 –––– 46 dB 50 50 50 HP U11644A 40–60 WR-19 –––– 46 dB 50 50 50 HP V11644A 50–75 WR-15 38 dB –––– 50 50 50 HP W11644A 75–110 WR-10 36 dB –––– 50 46 46 ECONOMY HP 85050D 0.045–18 7 mm 38 dB, DC–18 GHz –––– –––– 40 35 HP 85052D 0.045–26.5 3.5 mm 30 dB @ 26.5 GHz –––– –––– 30 25 HP 85054D 0.045–18 Type-N 34 dB @ 18 GHz –––– –––– 34 28 HP 85056D 0.045–50 2.4 mm 26 dB @ 50 GHz –––– –––– 26 23
1. Precision slotless connectors provide greater accuracy and repeatability than standard connectors because the
impedance does not change when it is connected to a device.
2. Not available with the HP 85059A calibration kit
24
Electronic calibration
Electronic calibration (ECal) is a precision, single connec­tion, one- or two-port calibration technique that uses fully traceable and verifiable electronic impedance standards. ECal replaces the traditional calibration technique that uses mechanical standards. ECal requires fewer connec­tions. It removes the intensive operator interaction, which is prone to errors. A full two-port calibration can be accomplished with a single connection of the ECal module and minimal operator interaction. This results in a faster and more repeatable calibration. Calibrations for non-insertable devices are equally convenient and straight forward.
An HP 85097A PC interface module with control software must be ordered to drive all HP ECal modules. The HP 85097A requires a customer-supplied PC meeting the following minimum requirements:
• Windows 95
®
, Windows NT 4.0®or later operating
system
• 486 or later CPU
• 32 MB of RAM
• 10 MB available hard-disk space
• one of the following HP-IB interface cards: HP 82340A/B, 82341C/D, National AT-GPIB/TNT, National AT-GPIB/TNT (plug&play) or National PCI-GPIB
ECal modules are controlled manually or automatically via the HP 85097A PC interface module with control soft­ware. The HP 85097A consists of a PC interface module, control software and a power supply.
The HP 85090 family of RF ECal modules provides cali­bration across the complete frequency range of the HP 8753C/D/E vector network analyzers. The HP 85060 family of microwave ECal modules provides calibrations through 26.5 GHz for the HP 8510B/C and 8719/20/22C/D vector network analyzers.
ECal modules and available options
A, B
Connector type1Frequency range ECal module Available options
model number
7 mm 30 kHz to 6 GHz2 HP 85091A 1BN, 1BP, 910, UK6 7 mm 1 GHz to 18 GHz HP 85060B 001, 1BN, 1BP, 910, UK6 Type-N 30 kHz to 6 GHz2 HP 85092A 00F, 00M, 00A, 1BN, 1BP, 910, UK6 Type-N 1 GHz to 18 GHz HP 85064B 001, 00F, 00M, 00A, 1BN, 1BP, 910, UK6
3.5 mm 30 kHz to 6 GHz2 HP 85093A 00F, 00M, 00A, 1BN, 1BP, 910, UK6
3.5 mm 1 GHz to 26.5 GHz HP 85062B 001,00F, 00M, 00A, 1BN, 1BP, 910, UK6
Options
Option Description 001 Adds a 30 kHz to 6 GHz RF module2 00F Replace f/m connectors on ECal module(s) with f/f connectors 00M Replace f/m connectors on ECal module(s) with m/m connectors 00A Adds male-to-male and female-to-female adapters (also adds a 5/16” 90 N-cm (8 in-lb)
torque wrench to 3.5 mm modules)
1BN Mil-STD 45662 calibration certificate 1BP Mil-STD 45662 calibration certificate with measured data 910 Add an extra operating and service manual UK6 Commercial calibration certificate with measured data
1. For ECal modules with sexed (m-f) connectors, the standard modules have one female and one male connector.
2. RF ECal modules are specified to operate from 300 kHz to 6 GHz, with typical performance down to 30 kHz. A. HP 85060 series modules cover a frequency range of 1 GHz to either 18 or 26.5 GHz. The upper frequency is limited
by the connector cutoff frequency. Each module is supplied with a torque wrench and foam-padded wood storage box.
B. HP 85090 series modules cover a frequency range of 30 kHz to 6 GHz. Each module is supplied with a torque
wrench and foam-padded wood storage box.
25
Verification kits
Verification kits are used to verify that a network analyzer is operating within its specified performance. Hewlett-Packard offers verification kits that include preci­sion airlines, mismatch airlines, and precision fixed atten­uators. All verification kits include measurement data and uncertainties which are traceable to the U.S. National Institute of Standards and Technology (NIST).
Verification kits
Verification Connector Frequency Description kit type range (GHz) (Contents)
HP 85051B 7 mm 0.045–18 10 cm airline, stepped
impedance airline, 20 dB, and 50 dB attenuators
HP 85053B 3.5 mm 0.045–26.5 7.5 cm airline, stepped
impedance airline 20 dB, and 40 dB attenuators
HP 85055A Type-N 0.045–18 10 cm airline, stepped
impedance airline, 20 dB and 50 dB attenuators
HP 85057B 2.4 mm 0.045–50 50 airline, stepped
impedance airline, 20 dB and 40 dB attenuators
HP R11645A WR-28 26.5–40 All contain: HP Q11645A WR-22 33–50 Standard waveguide section HP U11645A WR-19 40–60 Standard waveguide
mismatch
HP V11645A WR-15 50–75 20 dB attenuator HP W11645A WR-10 75–110 50 dB attenuator
Test port return cables
Test port cables are available in the 7 mm, 3.5 mm, Type-N, 2.4 mm, and 1.0 mm connector types
1
. The configurations and performance for all cables are described in the tables on the opposite page. All cable ends connect directly to the special rugged test port of the network analyzer test set.
Hewlett-Packard offers two cable designs: semi-rigid and flexible. Semi-rigid cables offer excellent performance and are suitable for applications where the connectors of the DUT are “in-line” or parallel. Flexible cables are ideal for manufacturing environments since they are more rugged and have a tighter bending radius than semi-rigid cables. Semi-rigid cables are warranted for 90 days; flexible cables are warranted for one year.
Either a single long cable or a shorter cable set can connect a coaxial device to the test set. A single cable with an appropriate test port adapter is best for applica­tions where the DUT requires a connection next to the test port for mechanical rigidity. A set of cables offers the flexibility required to position the test devices away from the test set.
Test port adapter sets
The HP 85130 series test port adapter sets protect the test set port when connecting devices to the test port. These adapters convert the rugged test set port to a connection that can be mated with the device under test. Each set contains a male and a female adapter.
Adapter sets
Adapter Connector type Frequency Return loss set (test port to device) (DC–f
max
) (dB) @ f
max
HP 85130C 3.5 mm
2
to Type-N DC–18 GHz 28
HP 85130D 3.5 mm
2
to PSC-3.5 mm (f) DC–26.5 GHz 28
or 3.5 mm (m)
HP 85130E 2.4 mm
2
to 7 mm DC–18 GHz 26
HP 85130F 2.4 mm
2
to PSC-3.5 mm (f) DC–26.5 GHz 26
or 3.5 mm (m)
HP 85130G 2.4 mm
2
to PSC-2.4 mm (f) DC–50 GHz 23
or 2.4 mm (m)
HP 11904S 2.4 mm
2
to 2.92 mm (4) DC–40 GHz 28
Waveguide to 1.0 mm adapter kits
Adapter kit Frequency range Description (contents) HP V85104A K10 DC–75 GHz WR-15 to 1.0 mm (f)
adapters (4) and 1.0 mm coax cables (2)
HP W85104A K10 DC–110 GHz WR-10 to 1.0 mm (f)
adapters (4) and 1.0 mm coax cables (2)
HP 85043C racked system kit
HP 85043C racked system kit is a rack standing 132 cm (52 in) high, with a width of 60 cm (23.6 in), and a depth of 80 cm (32 in). Complete with support rails and AC power distribution (suitable for 50 to 60 Hz, 100 to 240 VAC), the kit includes rack mounting hardware for all instru­ments. Thermal design is such that no rack fan is needed.
Bias networks
The HP 11612 bias networks apply DC as close to the device as possible, bypassing the test set's internal shunt resistor. These bias networks are especially useful when measuring DC parameters of semiconductor devices.
Maximum Maximum
Model number Frequency range current voltage HP 11612A K10/K20345 MHz–26.5 GHz 0.5 amps 40 volts HP 11612A K12/K223400 MHz–26.5 GHz 2.0 amps 40 volts HP 11612B K11/K21345 MHz–50 GHz 0.5 amps 40 volts
Network analyzer
test set
Network analyzer
test set
Test port
adapter
Device
under
test
Single cable
Device
under
test
Cable set
1. To measure Type-N devices, use a pair of 7 mm cables and the 7 mm-to-Type-N adapters provided in the HP 85054B, D calibration kits.
2. Special rugged female connector specifically for connecting to the network analyzer test port, but does not mate with a standard male connector.
3. Special option number K1x refers to port 1, K2x refers to port 2.
26
Test port return cable specifications
Single cables for HP 8514B and 8515A test sets (3.5 mm)
Connector type Frequency Length
2
Return Insertion loss Stability
1, 2
(test port to device) (GHz) cm (inch) loss (dB) (dB) (f in GHz) ±magnitude (dB) ±Phase (degrees)
HP 85131C semi-rigid cable 3.5 mm
3
to DC–26.5 81 (32) 17 0.43 f +0.3 <0.06 0.16 (f) +0.5
PSC-3.5 mm (f) (2.5 dB @ f
max
)
HP 85131E flexible cable 3.5 mm
3
to DC–26.5 96.5 (38) 16 0.35 f +0.3 <0.22 0.16 (f) +0.8
PSC-3.5 mm (f) (2.1 dB @ f
max
)
HP 85132C semi-rigid cable 3.5 mm
3
to 7 mm DC–18 81 (32) 17 0.35 f +0.3 <0.06 0.16 (f) +0.5
(1.8 dB @ f
max
)
HP 85132E flexible cable 3.5 mm
3
to 7 mm DC–18 97.2 (38.25) 17 0.35 f +0.3 <0.22 0.16 (f) +0.8
(1.8 dB @ f
max
)
Cable set for HP 8514B and 8515A test sets (3.5 mm)
Connector type Frequency Length
2
Return Insertion loss Stability
1, 2
(test port to device) (GHz) cm (inch) loss (dB) (dB) (f in GHz) ±magnitude (dB) ±Phase (degrees)
HP 85131D semi-rigid cable set 3.5 mm
3
to DC–26.5 53 (21) 16 0.30 f +0.2 <0.06 0.16 (f) +0.5
PSC-3.5 mm (f) or (1.8 dB @ f
max
)
3.5 mm (m)
HP 85131F flexible cable set 3.5 mm
3
to DC–26.5 62.2 (24.5) 16 0.25 f +0.2 <0.12 0.13 (f) +0.5
PSC-3.5 mm (f) or (1.5 dB @ f
max
)
3.5 mm (m)
HP 85132D semi-rigid cable set 3.5 mm
3
to 7 mm DC–18 53 (21) 17 0.25 f +0.2 <0.06 0.16 (f) +0.5
(1.3 dB @ f
max
)
HP 85132F flexible cable set 3.5 mm
3
to 7 mm DC–18 62.9 (24.75) 17 0.25 f +0.2 <0.12 0.13 (f) +0.5
(1.3 dB @ f
max
)
Single cables for HP 8517B test set (2.4 mm)
Connector type Frequency Length
2
Return Insertion loss Stability
1, 2
(test port to device) (GHz) cm (inch) loss (dB) (dB) (f in GHz) ±magnitude (dB) ±Phase (degrees)
HP 85133C semi-rigid cable 2.4 mm
3
to DC–50 81 (32) 15 0.84 f +0.3 <0.06 0.18 (f)
PSC-2.4 mm (f) (5.6 dB @ f
max
)
HP 85133E flexible cable 2.4 mm
3
to DC–50 113 (44) 12.5 0.58 f +0.35 <0.25 0.8 +0.16 (f)
PSC-2.4 mm (f) (4.45 dB @ f
max
)
HP 85134C semi-rigid cable 2.4 mm
3
to DC–26.5 81 (32) 16 0.46 f +0.3 <0.06 0.18 (f)
PSC-3.5 mm (f) (2.7 dB @ fmax)
HP 85134E flexible cable 2.4 mm
3
to DC–26.5 97.2 (38.25) 16 0.46 f +0.3 <0.22 0.16 (f) +0.8
PSC-3.5 mm (f) (2.7 dB @ fmax)
HP 85135C semi-rigid cable 2.4 mm
3
to 7 mm DC–18 81 (32) 17 0.46 f +0.3 <0.06 0.18 (f)
(2.25 dB @ fmax)
HP 85135E flexible cable 2.4 mm
3
to 7 mm DC–18 97.2 (38.25) 17 0.46 f +0.3 <0.22 0.16 (f) +0.8
(2.25 dB @ fmax)
Cable set for HP 8517B test set (2.4 mm)
Connector type Frequency Length
2
Return Insertion loss Stability
1, 2
(test port to device) (GHz) cm (inch) loss (dB) (dB) (f in GHz) ±magnitude (dB) ±Phase (degrees)
HP 85133D semi-rigid cable set 2.4 mm
3
to DC–50 53 (21) 15 0.55 f +0.2 <0.06 0.16 (f)
PSC-2.4 mm (f) or (3.7 dB @ fmax)
2.4 mm (m)
HP 85133F flexible cable set 2.4 mm
3
to DC–50 72 (28) 12.5 0.48 f +0.25 <0.17 0.8 +0.16 (f)
PSC-2.4 mm (f) or (3.64 dB @ fmax)
2.4 mm (m)
HP 85134D semi-rigid cable set 2.4 mm
3
to DC–26.5 53 (21) 16 0.31 f+0.2 <0.06 0.18 (f)
PSC-3.5 mm (f) or (1.8 dB @ fmax)
3.5 mm (m)
HP 85134F flexible cable set 2.4 mm
3
to DC–26.5 62.9 (24.75) 16 0.31 f+0.2 <0.12 0.13 (f) +0.5
PSC-3.5 mm (f) or (1.8 dB @ fmax)
3.5 mm (m)
HP 85135D semi-rigid cable set 2.4 mm
3
to 7 mm DC–18 53 (21) 17 0.31 f+0.2 <0.06 0.18 (f)
(1.5 dB @ fmax)
HP 85135F flexible cable set 2.4 mm
3
to 7 mm DC–18 62.9 (24.75) 17 0.31 f+0.2 <0.12 0.13 (f) +0.5
(1.5 dB @ fmax)
1. Phase stability of semi-rigid/flexible cables is specified with a 90 degree bend and a 4" radius.
2. Cable length and stability are supplemental characteristics.
3. Special rugged female connector specifically for connecting to the network analyzer test port, but does not mate with a standard male connector.
27
Related literature
5091-8970E HP 8510 Network Analyzer Color Brochure 5965-8837E HP 8510 System Solutions 5091-8967E HP 8510 Family Network Analyzer Configuration Guide 5091-8969E HP 85103 Upgrade Package to the HP 8510 5964-4229E HP 85106D Millimeter-wave Network Analyzer System
Product Overview
5091-8965E HP 85108A/L CW/Pulsed Network Analyzer Systems
Product Overview
5965-9888E HP 8510XF 110 GHz Single-Sweep Systems
Product Overview
5963-3743E HP 85060 Series and HP 85090 Series Electronic Calibration
Modules and PC Interface
MS-DOS®is U.S. registered trademark of Microsoft Corporation. Microsoft®is a U.S. registered trademark of Microsoft Corporation. Windows®and MS Windows®are U.S. registered trademarks of Microsoft Corporation.
28
For more information about Hewlett-Packard test and measure­ment products, applications, services, and for a current sales office listing, visit our web site, http://www.hp.com/go/tmdir. You can also contact one of the following centers and ask for a test and measurement sales representative.
United States:
Hewlett-Packard Company Test and Measurement Call Center P.O. Box 4026 Englewood, CO 80155-4026 1 800 452 4844
Canada:
Hewlett-Packard Canada Ltd. 5150 Spectrum Way Mississauga, Ontario L4W 5G1 (905) 206 4725
Europe:
Hewlett-Packard European Marketing Centre P.O. Box 999 1180 AZ Amstelveen The Netherlands (31 20) 547 9900
Japan:
Hewlett-Packard Japan Ltd. Measurement Assistance Center 9-1, Takakura-Cho, Hachioji-Shi, Tokyo 192, Japan Tel: (81) 426-56-7832 Fax: (81) 426-56-7840
Latin America:
Hewlett-Packard Latin American Region Headquarters 5200 Blue Lagoon Drive, 9th Floor Miami, Florida 33126, U.S.A. (305) 267 4245/4220
Australia/New Zealand:
Hewlett-Packard Australia Ltd. 31-41 Joseph Street Blackburn, Victoria 3130, Australia 1 800 629 485
Asia Pacific:
Hewlett-Packard Asia Pacific Ltd. 17-21/F Shell Tower, Times Square, 1 Matheson Street, Causeway Bay, Hong Kong Tel: (852) 2599 7777 Fax: (852) 2506 9285
Data Subject to Change Copyright © 1999 Hewlett-Packard Company Printed in U.S.A. 3/99 5091-8484E
Loading...