Specifications describe the instrument’s warranted performance over the temperature range of 0 °C to 40 °C (except as noted). Supplemental characteristics are intended to provide information that is useful in applying the instrument by giving non-warranted performance parameters. These are denoted as SPC (supplemental performance characteristics), typical, or nominal. Warm up time must be greater than or equal to 30 minutes after power on for all specifications.
Source characteristics
Frequency characteristics (Option 4395A-800)
Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10 Hz to 500 MHz
Resolution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1 mHz
Frequency reference
Accuracy
at 23 °C ± 5 °C, referenced to 23 °C . . . . . . . . . . . . . . . . . . . . . . . . . . .< ±5.5 ppm
Aging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .< ±2.5 ppm/year (SPC)
Initial achievable accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . .< ± 1.0 ppm (SPC)
Temperature stability
at 23 °C ± 5 °C, referenced to 23 °C . . . . . . . . . . . . . . . . . . . . . . .< ±2ppm (SPC)
Precision frequency reference (Option 4395A-1D5)
Accuracy
at 0 °C to 40 °C, referenced to 23 °C . . . . . . . . . . . . . . . . . . . . . . . . . .< ±0.13 ppm
Aging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . < ±0.1 ppm/year (SPC)
Initial achievable accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . .< ±0.02 ppm (SPC)
Temperature stability
at 0 °C to 40 °C, referenced to 23 °C . . . . . . . . . . . . . . . . . . . .< ±0.01 ppm (SPC)
Output characteristics
Power range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .–50 dBm to + 15 dBm
Level accuracy
at 0 dBm output, 50 MHz, 23 °C ± 5 °C, . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .±1.0 dB
Level linearity
Output power Linearity
1
-40 dBm ±1.0 dB < -40 dBm ±1.5 dB
Network Measurement
1. At relative to 0 dBm output, 50 MHz, 23 °C ± 5 °C
2
Flatness
at 0 dBm output, relative to 50 MHz, 23 °C ± 5 °C . . . . . . . . . . . . . . . . . . . . . . . .±2 dB
Resolution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0.1 dB
Spectral purity characteristics
Harmonics
at +10 dBm output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .< –30 dBc
Non-harmonics spurious
at +10 dBm output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .< –30 dBc
Noise sidebands
at 10 kHz offset from carrier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .< –95 dBc/Hz
Power sweep range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20 dB max.
Power sweep linearity
deviation from linear power referenced to the stop power level . . . . . . . . . . .±0.5 dB
Impedance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .50 nominal
Return loss
frequency 200 MHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .> 15 dB (SPC)
frequency > 200 MHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .> 7dB (SPC)
Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Type N female
Network Measurement
continued
3
Input characteristics
Frequency range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10 Hz to 500 MHz
Input attenuator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0 to 50 dB, 10 dB step
Full scale input level (R, A, B)
Attenuator setting (dB) Full scale input level
0 –10 dBm 10 0 dBm 20 +10 dBm 30 +20 dBm 40 +30 dBm 50 +30 dBm
IF bandwidth (IFBW) 2, 10, 30, 100, 300, 1 k, 3 k, 10 k, 30 kHz
Note: The IFBW should be set to less than 1/5 of the lowest frequency in
the sweep range.
Noise level (referenced to full scale input level, 23 °C ± 5 °C)
at 10 Hz frequency < 100 Hz, IFBW = 2 Hz . . . . . . . . . . . . . . . . . . . . . .–85 dB (SPC)
at 100 Hz frequency < 100 kHz, IFBW = 10 Hz . . . . . . . . . . . . . . . . . . . . . . . . .–85 dB
at 100 kHz frequency, IFBW = 10 Hz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . -–115 dB
Input crosstalk
for input R + 10 dBm input, input attenuator: . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20 dB
for input A, B input attenuator: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0 dB
at < 100 kHz
R through A, B . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .< –100 dB
others . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .< –100 dB (SPC)
at 100 kHz
R through A, B . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .< –120 dB
others . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .< –120 dB (SPC)
Source crosstalk (for input A, B)(typical for input R)
at + 10 dBm output, < 100 kHz, input attenuator: 0 dB . . . . . . . . . . . . . . . . .< –100 dB
at + 10 dBm output, 100 kHz, input attenuator: 0 dB . . . . . . . . . . . . . . . . .< –120 dB
Multiplexer switching impedance change
at input attenuator 0 dB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .< 0.5% (SPC)
at input attenuator 10 dB and above . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .< 0.1% (SPC)
Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Type-N female
Impedance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .50 nominal
Return loss
Input attenuator
0 dB 10 dB 20 dB to 50 dB
10 Hz frequency < 100 kHz 25 dB
1
25 dB
1
25 dB
1
100 kHz ≤ frequency ≤ 100 MHz 25 dB
1
25 dB 25 dB
1
100 MHz < frequency 15 dB
1
15 dB 15 dB
1
Maximum input level +30 dBm (at input attenuator: 40 dB or 50 dB)
Maximum safe input level +30 dBm or ±7 Vdc (SPC)
Receiver Characteristics
1. SPC
4
Absolute amplitude accuracy (R, A, B)
at –10 dBm input, input attenuator:
10 dB, frequency 100 Hz, IFBW 3 kHz, 23 °C ± 5 °C, . . . . . . . . . . . .< ±1.5 dB
Ratio accuracy (A/R, B/R) (typical for A/B)
at –10 dBm input, input attenuator:
10 dB, IFBW 3 kHz, 23 °C ± 5 °C, . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .< ±2 dB
Dynamic accuracy (A/R, B/R) (typical for A/B)
Input level Dynamic accuracy
1
(relative to full scale input level) frequency 100 Hz
0 dB input level > –10 dB ±0.4 dB
–10 dB input Level –60 dB ±0.05 dB
–60 dB > input level –80 dB ±0.3 dB
–80 dB > input level –100 dB ±3 dB
Figure 1-1. Magnitude dynamic accuracy
Residual responses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .< –80 dB full scale (SPC)
Trace noise (A/R, B/R, A/B)
at 50 MHz, both inputs:
full scale input level –10 dB, IFBW = 300 Hz . . . . . . . . . . . . . .< 0.005 dB rms (SPC)
Stability (A/R, B/R, A/B) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . < ±0.01 dB/°C (SPC)
Phase characteristics
Measurements format . . . . . . . . . . . . . . . . . . .Standard format, expanded phase format
Frequency response (deviation from linear phase) (A/R, B/R) (SPC for A/B)
at –10 dBm input, input attenuator: 10 dB, IFBW 3 kHz, 23 °C ± 5 °C . . . . . .< ±12°
Dynamic accuracy (A/R, B/R) (SPC for A/B)
Input level Dynamic accuracy
1
(relative to full scale input level) frequency 100 Hz
0 dB input level > –10 dB ±3°
–10 dB input Level –60 dB ±0.3°
–60 dB > input level –80 dB ±1.8°
–80 dB > input level –100 dB ±18°
Magnitude Characteristics
1. R input level (B input level for A/B) = full scale input level –10 dB, IFBW = 10 Hz, 23 °C ± 5 °C
Input level (dB)
Magnitude dynamic accuracy
Dynamic accuracy (dB)
Spec Typical
5
Figure 1-2. Phase dynamic accuracy
Trace noise (A/R, B/R, A/B)
at 50 MHz, both inputs:
full scale input level –10 dB, IFBW = 300 Hz . . . . . . . . . . . . . . . . .< 0.04° rms (SPC)
Stability (A/R, B/R, A/B) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .< ±0.1 °/°C (SPC)
Group delay characteristics
Aperture [Hz] . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0.25% to 20% of span
Accuracy
In general, the following formula can be used to determine the accuracy, in seconds, of a specific group delay measurement:
. . . . . . . . . . . .
Phase accuracy (degree)
Aperture(Hz) x 360 (degree)
Sweep characteristics
Sweep type . . . . . . . . . . . . . . . . .Linear frequency, log frequency, power, list frequency
Sweep direction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Upper direction only
Trigger type . . . . . . . . . . . . . . . . . . . . . . . . . . Hold, single, number of groups, continuous
Trigger source . . . . . . . . . . . . . . . . . . . .Internal (free run), external, manual, GPIB (bus)
Event trigger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .On point, on sweep
Input level (dB)
Phase dynamic accuracy
Dynamic accuracy (degree)
Spec Typical
6
Frequency characteristics
Frequency range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10 Hz to 500 MHz
Frequency readout accuracy
. . . . . . . .±((freq readout[Hz]) x (freq ref accuracy[1]) + RBW[Hz] +
SPAN[Hz]
)) [Hz]
where NOP means number of display points
NOP-1
Frequency reference (Option 4395A-800)
Accuracy
at 23 °C ± 5 °C, referenced to 23 °C . . . . . . . . . . . . . . . . . . . . . . . . . . . . . < ±5.5 ppm
Aging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . < ±2.5 ppm/year (SPC)
Initial achievable accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .< ± 1.0 ppm (SPC)
Temperature stability
at 23 °C ± 5 °C, referenced to 23 °C . . . . . . . . . . . . . . . . . . . . . . . . . < ±2 ppm (SPC)
Precision frequency reference (Option 4395A-1D5)
Accuracy
at 0 °C to 40 °C, referenced to 23 °C . . . . . . . . . . . . . . . . . . . . . . . . . . . < ±0.13 ppm
Aging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .< ±0.l ppm/year (SPC)
Initial achievable accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .< ±0.02 ppm (SPC)
Temperature stability
at 0 °C to 40 °C, referenced to 23 °C . . . . . . . . . . . . . . . . . . . . . . . .< ±0.01 ppm (SPC)
Resolution bandwidth (RBW)
Range
3 dB RBW at span > 0 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1 Hz to 1 MHz, 1-3 step
3 dB RBW at span = 0
. . . . . . . . . . . .3 k, 5 k, 10 k, 20 k, 40 k, 100 k, 200 k, 400 k, 800 k, 1.5 M, 3 M, 5 MHz
Selectivity (60 dB BW/3 dB BW)
at span > 0 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .< 3
Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Auto or manual
Accuracy
at span > 0 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .< ±10%
at span = 0 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .< ±30%
Video bandwidth (VBW)
Range
at span > 0 . . . . . . . . . . . . . . . .3 MHz to 3 MHz, 1-3 step, 0.003 ≤ VBW/RBW ≤ 1
Noise sidebands
Offset from carrier Noise sidebands
1 kHz < 95 dBc/Hz
100 kHz < 108 dBc/Hz
Figure 1-3. Noise sidebands
Spectrum Measurement
Frequency offset [Hz]
Noise sideband [dBc/Hz]
Spec Typical
7
Amplitude range . . . . . . . . . . . . . . . . . . . . . . . . . .displayed average noise level to +30 dBm
Reference value setting range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .–100 dBm to +30 dBm
Level accuracy
at –20 dBm input, 50 MHz, input attenuator: 10 dB, 23 °C ± 5 °C . . . . . . . . . . .< ±0.8 dB
Frequency response
at -20 dBm input, input attenuator: 10 dB, referenced to level at 50 MHz, 23 °C ± 5 °C
frequency 100 Hz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .< ±1.5 dB
frequency < 100 Hz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .< ±1.3 dB
Amplitude fidelity
1
Log scale
2
Range Amplitude fidelity
(dB to reference input lever [dB] [dB]
0 to 30 ±0.0530 to 40 ±0.0740 to 50 ±0.1550 to 60 ±0.3560 to 70 ±0.870 to 80 ±1.8
Linear scale
2
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .< ±3%
Displayed average noise level
at reference value ≤ –40 dBm, input attenuator: auto or 0 dB
at frequency 1 kHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .–120 dBm/Hz
at 100 kHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .–133 dBm/Hz
at 10 MHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . .(–145 + frequency/100 MHz) dBm/Hz
3
Figure 1-4. Typical displayed average noise level
Amplitude Characteristics
1. Fidelity shows an extent of nonlinearity referenced to the reference input level.
2. RBW = 10 Hz, –20 dBm reference value +30 dBm, reference input level = full scale input level –10 dB, 23 ± 5 °C
3. At start frequency 10 MHz
Note: Refer to Input attenuator part for the definition of full scale input level.
Frequency offset [Hz]
Average noise level [dBm/Hz]
Spec Typical
8
Figure 1-5. Typical on-screen dynamic range (center: 100 MHz)
Spurious responses
Second harmonic distortion
at single tone input with full scale input level –10 dB, input signal frequency 100 kHz
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .< –70 dBc, < –75 dBc (SPC)
Third order inter-modulation distortion
at two tones input with full scale input level –16 dB, separation 100 kHz
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .< –75 dBc, < 80 dBc (SPC)
Spurious
at single tone input with full scale input level –10 dB, input signal frequency 500 MHz
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .< –75 dBc
except for the following frequency ranges:
5.6 MHz ± 1 MHz, 30.6 MHz ± 1 MHz, 415.3 MHz ± 1 MHz
Residual response
at reference value setting ≤ –40 dBm, input attenuator: auto or 0 dB . . . . .< –110 dBm
On-screen Dynamic Range
Offset frequency [Hz]
On-screen dynamic range [dBc]
9
Figure 1-6. Typical dynamic range at inputs R, A, and B
Input attenuator
Setting range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0 dB to 50 dB, 10 dB step
Attenuator setting (dB) Full scale input level
0 –20 dBm 10 –10 dBm 20 0 dBm 30 +10 dBm 40 +20 dBm 50 +30 dBm
Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Auto or manual
(In auto mode, the attenuator is set to 20 dB above the reference value; this ensures that the maximum signal level after the attenuator will not be greater than –20 dBm.)
Input attenuator switching uncertainty
at attenuator: 30 dB, referenced to 10 dB . . . . . . . . . . . . . . . . . . . . . . . . . . . . .< ±1.0 dB
at attenuator: 40 dB, referenced to 10 dB . . . . . . . . . . . . . . . . . . . . . . . . . . . . < ±1.5 dB
Temperature drift . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .< ±0.05 dB/°C (SPC)
Scale
Log 0.1 dB/div to 20 dB/div Linear
at watt . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1.0 x 10
-12
W/div
at volt . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1.0 x 10
-9
V/div
Measurement format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Spectrum or noise (/Hz)
Display unit . . . . . . . . . . . . . . . . . . . . . . . . . . . .dBm (unit of marker: dBm, dBV, dBµV, V, W)
Sweep characteristics
Sweep type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Linear, list
Trigger type . . . . . . . . . . . . . . . . . . . . . . . . . . .Hold, single, number of groups, continuous
Trigger source . .Internal (free run), external, manual, level gate, edge gate, GPIB (bus) Sweep time (excluding each sweep setup time)
RBW SPAN Typical sweep time
1 MHz 500 MHz 190 ms
100 kHz 100 MHz 300 ms
10 kHz 10 MHz 240 ms
1 kHz 1 MHz 190 ms
100 Hz 100 kHz 270 ms
10 Hz 10 kHz 2.0 s
1 Hz 1 kHz 11 s
Zero Span
1
Typical Dynamic Range
1. See the next item for sweep time at zero span
Input level (dB)
(Relative to full scale input level)
Dynamic range (dB)
Sensitivity (1 Hz RBW) Sensitivity ( 100 Hz RBW) 2nd harmonic distortion 3rd order inter-modulation distortion
Second
Third
10
Zero span
RBW Minimum resolution Maximum sweep time
5 MHz 40 ns 1.28 ms
100 kHz 1.28 µs 81.92 ms
3 kHz 40.96 µs 2.62 s
Number of display points
at span > 0 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 to 801 points (automatically set)
at span = 0 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 to 801 points (selectable)
Input characteristics
Input port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .R, A, B
Crosstalk
from any input to other inputs, at the same input attenuator settings< -100 dB (SPC)
Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Type N female
Impedance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .50 nominal
Return loss
Input attenuator
0 dB 10 dB 20 dB to 50 dB
10 Hz frequency < 100 kHz 25 dB
1
25 dB
1
25 dB
1
100 kHz ≤ frequency ≤ 100 MHz 25 dB
1
25 dB 25 dB
1
100 MHz < frequency 15 dB
1
15 dB 15 dB
1
Input level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .+30 dBm max. at input attenuator: 50 dB
Maximum safe input level . . . . . . . . . . . . . . . . . . . . . . . . . . . .+30 dBm or ±7 Vdc (SPC)
11
Gate length
Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6 µs to 3.2 s
Resolution
Range of gate length (T
I
) Resolution
6 µs ≤ T
I
25 ms 0.4 µs
25 ms < T
I
64 ms 1 µs
64 ms < T
I
130 ms 2 µs
130 ms < T
I
320 ms 5 µs
320 ms < T
I
1.28 s 20 µs
1.28s < T
I
3.2 s 100 µs
Gate length
Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 µs to 3.2 s
Resolution
Range of gate delay (T
d
) Resolution
2 µs ≤ T
d
25 ms 0.4 µs
25 ms < T
d
64 ms 1 µs
64 ms <T
d
130 ms 2 µs
130 ms < T
d
320 ms 5 µs
320 ms < T
d
1.28 s 20 µs
1.28 s < T
d
3.2 s 100 µs
Additional amplitude error
Log scale . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .< 0.3 dB (SPC)
Linear scale . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . < 3% (SPC)
Gate control modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Edge (positive/negative) or level
Gate trigger input (external trigger input is used)
Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .BNC female
Level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .TTL
Gate output
Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .BNC female
Level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .TTL
Specifications when Option 4395A-1D6 Time-Gated Spectrum Analysis is Installed
All specifications are identical to the standard Agilent 4395A except the following items.
12
Measurement functions
Measurement parameters Z, Y, L, C, Q, R, X, G, B, θ Display parameters IZI, 0
z
, R, X, IYI, θy, G, B, IΓI, θγ, Γx, Γy, Cp, Cs,
Lp, Ls, Rp, Rs, D, Q
Display formats
Vertical lin/log scale
Complex plane
Polar/Smith/admittance chart
Sweep parameters
Linear frequency sweep
Logarithmic frequency sweep
List frequency sweep
Power sweep (in dBm unit)
IF bandwidth
2,10, 30, 100, 300, 1k, 3k, 10k, 30k [Hz]
Calibration
OPEN/SHORT/LOAD 3 term calibration
Fixture compensation
Port extension correction
Measurement port type
7-mm
Output characteristics
Frequency range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .100 kHz to 500 MHz
Frequency resolution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1 MHz
Output impedance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .50 nominal
Output level
when the measurement port is terminated by 50
1
. . . . . . . . . . . . . .–56 to +9 dBm
when the measurement port is open . . . . . . . . . . . . . . . . . . .0.71 mVrms to 1.26 Vrms
Resolution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0.1 dBm
Level accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .± (A + B + 6 x F/(1.8 x 10
9
))dB
Where
A = 2 dB B = 0 dB (at 0 dBm ≤ P ≤ + 15 dBm) or B = 1 dB (at –40 dBm P < 0 dBm) or B = 2 dB (at –50 dBm P < –40 dBm) F is setting frequency [Hz], P is output power setting
Option 4395A-010 Impedance measurement
The following specifications are applied when the 43961A impedance test kit is connected to the 4395A.
1. When the measurement port is terminated with 50 , the signal level at the measure­ment port is 6 dB lower than the signal level at the RF OUT port.
13
Measurement accuracy is specified at the connecting surface of the 7-mm connector of the Agilent 43961A under the following conditions:
Warm up time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .> 30 minutes
Ambient temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23 °C ± 5 °C,
within ±1 °C from the temperature at which calibration is performed
Signal level (setting) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0 to +15 dBm
Correction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .ON
IFBW (for calibration and measurement) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .300 Hz
Averaging factor (for calibration and measurement) . . . . . . . . . . . . . . . . . . . . . . . . .8
Figure 1-7. Impedance measurement accuracy
IZI - θ accuracy
IZI accuracy Z
a
= A + (B/IZmI + C x IZmI) x 100 [%]
θ accuracy θ
a
= sin-1(Za/100)
Where, IZ
m
I is IZI measured. A, B, and C are obtained from Figure 1-7.
IYI - θ accuracy
IYI accuracy Ya= A + (B x IYmI + C/IZmI) x 100 [%] θ accuracy θ
a
= sin-1(Ya/100)
Where, IY
m
I is IYI measured. A, B, and C are obtained from Figure 1-7.
Measurement Basic Accuracy
(Supplemental performance characteristics)
Test frequency [Hz]
14
R - X accuracy (depends on D)
Accuracy D 0.2 0.2 < D 5 5 < D
R
a
±Xmx Xa/100[]Ra/cosθ[%] Ra[%]
X
a
Xa[%] Xa/sinθ[%] ±Rmx Ra/100[Ω]
Where,
D can be calculated as: R/X, or
R/(2πf x L
s
), or
R x 2πf x C
s
θ can be calculated as: tan-1(X/R), or
tan
-1
(2πf x Ls/R), or
tan
-1
(1/(R x 2πf x Cs))
R
a
= A + (B/IRmI + C x IRmI) x 100 [%]
X
a
= A + (B/IXmI + C x IXmI) x 100 [%]
R
m
and Xmare the measured R and X, respectively. A, B, and C are obtained
from Figure 1-7.
G - B accuracy (depends on D)
Accuracy D 0.2 0.2 < D 5 5 < D
G
a
±Bmx Ba/100[S] Ga/cosθ[%] Ga[%]
B
a
Ba[%] Ba/sinθ[%] ±Gmx Ga/100[Ω]
Where,
D can be calculated as: G/B, or
G/(2πf x C
p
), or
G X 2πf x L
p
θ can be calculated as: tan-1(B/G), or
tan
-1
(2πf x Cp/G), or
tan
-1
(1/(G x 2πf x Lp))
G
a
= A + (B/IGmI + C x IGmI) x 100[%]
B
a
= A + (B/IBmI + C x IBmI) x 100[%]
G
m
and Bmare the measured G and B, respectively. A, B, and C are obtained from
Figure 1-7.
D accuracy
Accuracy D 0.2 0.2 < D
D
a
Za/100 (Za/100) x (1 + D2)
Where, Z
a
is IZI accuracy.
L accuracy (depends on D)
Accuracy D 0.2 0.2 < D
L
a
La/100 La(1 + D)
Where, L
a
= A + (B/IZlI + C x IZl) x 100[%]
IZ
l
I = 2πf x Lm, f is frequency in Hz, and Lmis measured L. A, B, and C are obtained
from Figure 1-7.
C accuracy (depends on D)
Accuracy D 0.2 0.2 < D
C
a
C
a
Ca(1 + D)
Where, C
a
= A + (B/IZcI + C x IZcI) x 100[%]
IZ
c
I = 2πf x Cm, f is frequency in Hz, and Cmis measured C. A, B, and C are obtained
from Figure 1-7.
Display
LCD
Size/type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8.4 inch color LCD
Number of pixels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .640 x 480
Effective display area . . . . . . . . . . . . . . . . . . . . . . .160 mm x 115 mm(600 x 430 dots)
Number of display channels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2
Format single, dual (split or overwrite)
Number of traces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
For measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 traces
For memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 traces
Data math . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .gain x data – offset,
gain x (data - memory) – offset, gain x (data + memory) – offset, gain x (data/memory ) – offset
Data hold . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Maximum hold, minimum hold
Marker
Number of markers
Main marker . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .l for each channel
Sub-marker . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7 for each channel
marker . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1 for each channel
Hard copy
Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . .Dump mode only (including color dump mode)
Storage
Built-in flexible disk drive
Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3.5 inch, 1.44 MByte, or 720 KByte,
1.44 MByte format is used for disk initialization
Memory . . . . . . . . . . . . . . . . . . . . . . . . . . .512 KByte, can be backed up by flash memory
GPIB
Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .IEEE 488.1-1987, IEEE 488.2-1987,
IEC 625, and JIS C 1901-1987 standards compatible.
Interface function . . . . . . . . . . . . . . . . . . . . . . . . .SH1, AH1, T6, TEO, L4, LEO, SR1, RL1,
PP0, DC1, DT1, C1, C2, C3, C4, C11, E2
Data transfer formats . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .ASCII,
32 and 64 bit IEEE 754 floating point format,
DOS PC format (32 bit IEEE with byte order reversed)
Printer parallel port
Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .IEEE 1284 Centronics standard compliant
Printer control language . . . . . . . . . . . . . . . . . . . . . . . . . . .PCL3 printer control language
Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .D-SUB (25-pin)
15
Common to Network/Spectrum/
Impedance Measurement
16
Option 4395A-001 DC voltage/current source
The setting of Option 4395A-001 DC voltage/current source is independent of channel 1 and channel 2 settings.
Voltage
Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .–40 V to +40 V
Resolution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1 mV
Current limitation
at voltage setting = –25 V to +25 V . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .±100 mA
at voltage setting = –40 V to –25 V, 25 V to 40 V . . . . . . . . . . . . . . . . . . . ±20 mA
Current
Range –20 µA to -100 mA, 20 µA to 100 mA
Resolution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20 µA
Voltage limitation
at current setting = –20 mA to +20 mA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .±40 V
at current setting = –100 mA to –20 mA, 20 mA to 100 mA . . . . . . . . . . . . . . ±25 V
Accuracy
Voltage
at 23 °C ± 5 °C . . . . . . . . . . . . . . . . . . . . . . .±(0.1% + 4 mV + I
dc
1
[mA] x 5 [] mV)
Current
at 23 °C ± 5 °C . . . . . . . . . . . . . . . . . . . . . . .±(0.5% + 30 µA + V
dc
2
[V]/10 [k] mA)
Probe power
Output voltage . . . . . . . . . . . . . . . . .+15 V (300 mA), –12.6 V (160 mA), GND nominal
Specifications when instrument BASIC is operated
Keyboard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . PS/2 style 101 English keyboard
Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .mini-DIN
8 bit I/0 port
Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .D-SUB (15-pin)
Level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .TTL
Number of input/output bit . . . . . . . . . . . . . . . . . . . . . .4 bit for input, 8 bit for output
Figure 1-8. 8 bit I/O port pin assignments
24-bit I/O interface
Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .D-SUB (36-pin)
Level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .TTL
I/O . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-bit for input or output, 16-bit for output
Figure 1-9. 24-bit I/O interface pin assignment
1. Current at DC source connector.
2. Voltage at DC source connector.
17
Table 1-1. Signal source assignment
Pin No. Signal name Signal standard
1 GND 0 V 2 INPUT1 TTL level, pulse input (pulse width: 1 µs or above) 3 OUTPUT1 TTL level, latch output 4 OUTPUT2 TTL level, latch output 5 OUTPUT PORT A0 TTL level, latch output 6 OUTPUT PORT A1 TTL level, latch output 7 OUTPUT PORT A2 TTL level, latch output 8 OUTPUT PORT A3 TTL level, latch output
9 OUTPUT PORT A4 TTL level, latch output 10 OUTPUT PORT A5 TTL level, latch output 11 OUTPUT PORT A6 TTL level, latch output 12 OUTPUT PORT A7 TTL level, latch output 13 OUTPUT PORT B0 TTL level, latch output 14 OUTPUT PORT B1 TTL level, latch output 15 OUTPUT PORT B2 TTL level, latch output 16 OUTPUT PORT B3 TTL level, latch output 17 OUTPUT PORT B4 TTL level, latch output 18 OUTPUT PORT B5 TTL level, latch output 19 OUTPUT PORT B6 TTL level, latch output 20 OUTPUT PORT B7 TTL level, latch output 21 I/O PORT C0 TTL level, latch output 22 I/O PORT C1 TTL level, latch output 23 I/O PORT C2 TTL level, latch output 24 I/O PORT C3 TTL level, latch output 25 I/O PORT D0 TTL level, latch output 26 I/O PORT D1 TTL level, latch output 27 I/O PORT D2 TTL level, latch output 28 I/O PORT D3 TTL level, latch output 29 PORT C STATUS TTL level, input mode: LOW, output mode: HIGH 30 PORT D STATUS TTL level, input mode: LOW, output mode: HIGH 31 WRITE STROBE SIGNAL TTL level, active low, pulse output
(width: 10 µs; typical) 32 +5 V PULLUP 33 SWEEP END SIGNAL TTL level, active low, pulse output
(width: 20 µs; typical) 34 +5 V +5 V, 100 mA MAX 35 PASS/FAIL SIGNAL TTL level, PASS: HIGH, FAIL: LOW, latch output 36 PASS/FAIL WRITE STROBE SIGNAL
TTL level, active low, pulse output
(width: 10 µs; typical)
18
Input and output characteristics
External reference input
Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10 MHz ± 100 Hz (SPC)
Level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .-5 dBm to +5 dBm (SPC)
Input impedance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .50 nominal
Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .BNC female
Internal reference output
Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10 MHz nominal
Level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0 dBm (SPC)
Output impedance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .50 nominal
Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .BNC female
Reference oven output (Option 4395A-1D5)
Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10 MHz nominal
Level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0 dBm (SPC)
Output impedance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .50 nominal
Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .BNC female
External trigger input
Level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .TTL
Pulse width (Tp) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 µs typically
Polarity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .positive/negative selective
Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .BNC female
External program Run/Cont input
Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .BNC female
Level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .TTL
Gate output (Option 4395A-1D6)
Level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .TTL
Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .BNC female
Figure 1-10. Trigger signal (external trigger input)
General Characteristics
Positive trigger signal Negative trigger signal
19
S-parameter test set interface
Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .D-SUB (25-pin)
Caution
Do not connect a printer to this connector. If you connect a printer with the S-parameter test set interface connector (TEST SET-I/0 INTERCONNECT), it may cause damage to the printer.
Figure 1-11. S-parameter test set interface pin assignments
External monitor output
Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .D-SUB (15-pin HD)
Display resolution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .640 x 480 VGA
Operation conditions
Temperature
Disk drive non-operating condition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0 °C to 40 °C
Disk drive operating condition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10 °C to 40 °C
Humidity
at wet bulb temperature ≤ 29 °C, without condensation
Disk drive non-operating condition . . . . . . . . . . . . . . . . . . . . . . . . . .15% to 95% RH
Disk drive operating condition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15% to 80% RH
Altitude . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0 to 2,000 m
Warm up time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .30 minutes
Non-operation conditions
Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .-20 °C to 60 °C
Humidity
at wet bulb temperature ≤ 45 °C, without condensation . . . . . . . . . . .15% to 95% RH
Altitude . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0 to 4,572 m
Others
EMC . . . . . . . . . . . .Complies with CISPR 11 (1990) / EN 55011(1991) : Group 1, Class A
Complies with EN 50082-1 (1992) / IEC 1000-4-2 (1995) : 4 kV CD, 8 kV AD
Complies with EN 50082-1 (1992) / IEC 801-3 (1984) : 3 V/m
Complies with EN 50082-1 (1992) / IEC 1000-4-4 (1995) :1 kV / Main, 0.5kV /Signal line
Complies with IEC 1000-3-2 (1995) / EN 61000-3-2 (1995) Complies with IEC 1000-3-3 (1994) / EN 61000-3-3 (1995)
Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Complies with IEC 1010-1 (1990),
Amendment 1(1992), Amendment 2 (1995)
Certified by CSA-C22.2 No.1010.1-92
Power requirements . . . . . . .90 V to 132 V, or 198 V to 264 V (automatically switched),
47 to 63 Hz, 300 VA max.
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21 kg (SPC)
Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .425 (W) x 235 (H) x 553 (D) mm
Agilent Email Updates
Agilent Technologies Test and Measurement Support, Services, and Assistance
Agilent Technologies aims to maximize the value you receive, while minimizing your risk and problems. We strive to ensure that you get the test and measurement capabilities you paid for and obtain the support you need. Our extensive support resources and services can help you choose the right Agilent products for your applications and apply them successfully. Every instrument and system we sell has a global warranty. Support is available for at least five years beyond the production life of the product. Two concepts underlie Agilents overall support policy: Our Promise and Your Advantage.
Our Promise
Our Promise means your Agilent test and measurement equipment will meet its advertised performance and functionality. When you are choosing new equipment, we will help you with product information, including realistic performance specifications and practical recommendations from experienced test engineers. When you use Agilent equipment, we can verify that it works properly, help with product operation, and provide basic measurement assistance for the use of specified capabilities, at no extra cost upon request. Many self-help tools are available.
Your Advantage
Your Advantage means that Agilent offers a wide range of additional expert test and measurement services, which you can purchase according to your unique technical and business needs. Solve problems efficiently and gain a competitive edge by contracting with us for calibration, extra-cost upgrades, out-of-warranty repairs, and onsite education and training, as well as design, system integration, project management, and other professional engineering services. Experienced Agilent engineers and technicians worldwide can help you maximize your productivity, optimize the return on investment of your Agilent instruments and systems, and obtain dependable measurement accuracy for the life of those products.
Agilent T&M Software and Connectivity
Agilents Test and Measurement software and connectivity products, solutions and developer network allows you to take time out of connecting your instruments to your computer with tools based on PC standards, so you can focus on your tasks, not on your connections. Visit
www.agilent.com/find/connectivity
for more information.
By internet, phone, or fax, get assistance with all your test & measurement needs
Online Assistance:
www.agilent.com/find/assist
Product specifications and descriptions in this document subject to change without notice.
© Agilent Technologies, Inc. 2001, 2003 Printed in USA, May 13, 2003 5965-9340E
Phone or Fax United States:
(tel) 800 452 4844
Canada:
(tel) 877 894 4414 (fax) 905 282 6495
China:
(tel) 800 810 0189 (fax) 800 820 2816
Europe:
(tel) (31 20) 547 2323 (fax) (31 20) 547 2390
Japan:
(tel) (81) 426 56 7832 (fax) (81) 426 56 7840
Korea:
(tel) (82 2) 2004 5004 (fax) (82 2) 2004 5115
Latin America:
(tel) (305) 269 7500 (fax) (305) 269 7599 Taiwan: (tel) 0800 047 866 (fax) 0800 286 331
Other Asia Pacific Countries:
(tel) (65) 6375 8100 (fax) (65) 6836 0252 Email: tm_asia@agilent.com
www.agilent.com/find/emailupdates
Get the latest information on the products and applications you select.
Loading...