
Agilent
4349B 4-Channel
High
Service Manual
Resistance
Meter
This
manual
\Instruments
F
or
additional
applies
directly
important
Covered
SERIAL
to
instruments
information
by
This
NUMBERS
Manual"
with
serial number
about
serial
in
Chapter
numbers,
1of
this manual.
prex JP1KD
read
.
Agilent Part No. 04349-90033
Printed in JAPAN March 2001

Notice
The
information
This
document
reserved.
language
contains
No
part
without
contained
proprietary
of
this
the
prior
in
this
document
written
document
information
may
be
consent
is
subject
that
is
photocopied,
of
the
Agilent
to
change
protected
reproduced,
T
echnologies.
without
by
copyright.
or
notice
.
All
translated
rights
to
another
are
Agilent
T
echnologies
Component
1-3-2,
Hyogo
Murotani,
,
651-2241
T
est
Japan,
PGU-Kobe
Nishi-ku,
Japan
Ltd.
Kobe-shi,
c
Copyright Agilent Technologies Japan, Ltd. 1996, 2000, 2001

Manual Printing
A
ugust
1996
:
:
:
:
:
:
:
:
June
2000
:
:
:
:
:
:
:
::
::
History
::
::
::
::
::
::
::
::
::
::
::
:
:
:
:
:
:
:
:
:
:
:
::
::
::
::
:
First
Edition (part
:
:
:
:
:
:
:
:
:
:
:
:
:
::
::
::
::
:
Second
Edition (part
number: 04349-90033)
number:
04349-90033)
March
2001
:
:
:
:
:
:
:
::
::
::
::
::
::
::
:
:
:
:
:
:
:
:
:
:
:
::
::
::
::
::
Third Edition
(part number:
04349-90033)
iii

Safety Summary
The
following
service
W
ARNINGS
In
addition
,
and
elsewhere
it
general
repair
violates
of
safety
this
safety
instrument.
The
A
gilent
requirements
T
echnologies
.
precautions must
instrument. F
in
this
manual
standards
assumes
ailure to
may impair
of design,
no liability
be observed
comply with
the protection
manufacture,
for the
customer's failure
during all
phases of
these precautions
provided by
and intended
operation,
or with
the equipment.
use of
to
comply
the
with
specic
these
Note
Ground The
T
o
avoid
electric
safety
DO
Do
electrical
K
Operating
adjustments
with
with
circuits
DO
earth
NOT
not operate
instrument in
eep
Away
the
power
the
power
before
NOT
ground
Operate
personnel
must
Service
4349B
in
complies
IEC1010-1.
Instrument
shock
hazard,
by
the
In
An
the instrument
such an
From
cable
cable
touching
Live
must
not remove
be
made
connected. Under
removed.
them.
Or
A
with
4349B
the
instrument
supplied
power
Explosive
in
the
environment
Circuits
by qualied
T
o
avoid
djust
Alone
INST
ALLA
is
INDOOR
cable
Atmosphere
presence
constitutes
instrument
maintenance
certain
injuries
TION
CA
USE
product.
chassis
with
of
ammable
covers
personnel.
conditions
,
always
TEGORY
and
cabinet
earth
blade
gasses
a
denite
.
Component
,
dangerous
disconnect
II
.
Do
and
must
or
safety
not
power
POLLUTION
be
connected
fumes
.
Operation
hazard.
replacement
replace
voltages
components
may
and
discharge
DEGREE
to
and
internal
exist
a
of
even
2
any
Do
not
attempt
aid
and resuscitation,
DO
NOT
Because of the danger
perform unauthorized modications to
Technologies Sales and Service
internal
is
Substitute
service
or
adjustment
unless
another
present.
P
arts
Or
Modify
Instrument
of introducing additional hazards
the instrument. Return the instrument to a Agilent
Oce for service and repair to ensure that safety features are
person,
, do not install substitute parts or
maintained.
Dangerous Procedure W
Warnings
, such as the example below
arnings
, precede potentially dangerous procedures throughout
this manual. Instructions contained in the warnings must be followed.
Warning
Dangerous voltages, capable of causing death, are present in this
instrument. Use extreme caution when handling, testing, and adjusting
this instrument.
iv
capable
of
rendering
rst

Certication
Agilent
of
T
echnologies
shipment
from
measurements
T
echnology
facilities
W
arranty
This
of
Agilent
,
to
other
workmanship
certain
specied
repair
F
or
Agilent
T
echnologies
pay
another
Agilent
T
echnologies
property
of
components
or
warranty
Technologies
all
shipping
country
T
echnologies
the
instrument,
period.
replace
installed
certies
the
are
traceable
the
extent
International
T
echnologies
for
a
period
listed
During
products
service
. Buyer
shall
pay
charges
.
warrants
for
use
with
on
or
factory
.
to
allowed by
instrument
of
one
in
General
the warranty
that
or
repair
shall prepay
shipping
,
duties
an
instrument
that
instrument.
software
that this
Agilent T
the
product met
echnologies further
United
States National
the Institution's
Standards Organization
year
product
from the
is
warranted against
date of
Information
period, Agilent
prove to
,
this
be defective
product
must
shipping
charges
,
and
that
,
or
to
taxes
its
software
will
Agilent
rmware
return
for
products
execute
T
will
and
echnologies
be
its published
certies that
Institute of
calibration facility
members.
shipment, except
of
this manual,
Technologies
.
be
returned to
charges
the
to
product
Agilent
returned
rmware
its
designated
programming
does
uninterrupted
specications at
its calibration
Standards and
,or
defects in
the warranty
will, at
a
service
T
echnologies
to
Buyer
.
However
to
Agilent
by
instruction
not
warrant
or
error
to the
material and
that in
shall
its
option,
facility
T
echnologies
Agilent
that
free
.
the time
calibration
the
case
be
either
designated
and
Agilent
,
Buyer
when
the
operation
of
for
shall
from
the
by
Limitation
The
foregoing
maintenance
misuse
,
preparation
No
other
implied
by
operation
or
warranty
warranties
Of
W
arranty
warranty
Buyer
shall
,
Buyer-supplied
outside
maintenance
is
expressed
of
merchantability
not
apply
the
environmental
.
or
to
software
implied.
and
defects
resulting
or
interfacing,
specications
A
gilent
tness
from
unauthorized
for
the
T
echnologies
for
a
particular purpose
improper
product,
specically
or
inadequate
modication
or
improper
disclaims
.
or
site
the
v

Exclusive Remedies
The
remedies
shall
not
whether
be
liable
based
provided
for
on
contract,
herein
any
direct, indirect,
Assistance
are buyer's
tort,
or
sole and
special, incidental,
any other
exclusive remedies
legal theory
or consequential
.
.A
gilent T
echnologies
damages,
Product
Agilent
F
or
any
A
ddresses
maintenance
T
echnologies
assistance
are
,
provided
agreements
products
contact
at
the
and
.
your
nearest Agilent
back
other
of this
customer
Technologies
manual.
assistance
Sales and
agreements
Service Oce
are
available
.
for
vi

Safety Symbols
General
denitions
Instruction
necessary
Alternating
Direct
On
O
Frame
This
condition
result
This
condition
result
Note
practice
of
safety
current.
(Supply).
(Supply).
or
W
arning
in injury
Caution
in
damage
denotes
,
condition
symbols used
manual
for
the
user
current.
chassis
terminal
sign
or
the
like
or death
sign
or
the
like
to
important
on equipment
symbol:
to
refer
denotes
,
which,
to
denotes a
,
which,
or
destruction
information.
or
the
like
the
product
to the
a
hazard.
if
not
correctly
personnel.
hazard. It
if
not
correctly
of
,
which
or in
manuals are
is
marked with
instruction manual.
It
calls attention
performed
calls
attention
performed
part
or
all
of
the
It
calls
attention
is
essential
to
highlight.
listed below
this symbol
to a
procedure
or adhered
to
a
procedure
or
adhered
product.
to
a
procedure
.
when it
,
practice
to
,
could
,
practice
to
,
could
,
is
,
,
vii

Documentation Map
Operation
Operation
Japanese
conformance
Service
Service
repair
the
How
This
is
the
Manual
Manual
edition)
(Agilent
explains
to
published
Manual
Manual (Agilent
4349B
.
T
o
Use
Service
This
Manual
contains adjustments
Chapter
Chapter
recommended
Chapter
1
1
gives
2
General
general
equipment
A
djustments
P
art Number:
specications
specications
P
art
Number:
Manual
for
the
and
repair
Information
information
for
adjustments
04349-90040 for
,
basic
.
04349-90033)
4349B
4-Channel
information,
for servicing
and
English edition,
measurement explanations
and
the 4349B
explains
High
consists
how
Resistance
of
the
following
.
This
to
Meter
chapter
troubleshooting.
04349-97040 for
,and
adjust,
troubleshoot,
.
This
four
lists
how to
manual
chapters
the
verify
and
.
Chapter
specications
Chapter
Chapter
includes
Chapter
Chapter
theory of
2
describes
after
3
Assembly
3
provides
replaceable
4
Troubleshooting
4
provides
the
adjustments
it
has
the
information
assembly
the
information
,
required
been
repaired,
Replacement
on replacing
lists
,
and
on
troubleshooting
or
it
fails the
removal
to
ensure that
performance
the 4349B
and installation
the
assemblies.
4349B
operation and information on faulty assembly isolation.
the
4349B
procedures.
.
The
is
within
tests
.
The information
information
its
published
includes
Appendix A Manual Changes
Appendix A contains
4349Bs manufactured before
Appendix B Default Jumper
manual changes and provides information for using this manual with
the printing date of the manual.
and Switch Settings
Appendix B describes the settings of the jumpers and switches on the 4349B A2 CPU Board
Assembly. The information is required when the A2 assembly is replaced.
viii

Contents
1.
General
Introduction
Instruments Covered
Required Equipment
2.
A
djustments
Introduction
Safety
Required
A
djustment
Adjustable
F
actory-Selected
EEPROM
Related
A
djustment
A
djustments
Channel
Channel
A
Program
Making
Dening
Program
DC
V
oltage
1
Channel
Equipment
Setup
Setup
Channel
Equipment
Setup
Setup
2,
Information
.
.
.
.
.
.
.
.
Consideration
Equipment
T
ools
Components .
Write
djustments
Program
Installation
a
the
Execution
Procedure
A
djustment
1
A
1/2
2/2
1
A
1/2 .
2/2 .
3,
.
Components
Protect
W
orking
Conguration
Source
djustment
.
.
.
.
.
.
.
.
djustment
..
..
..
and
4
A
.
.
.
.
by
This
Manual
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
..
Switch
.
.
.
.
.
.
.
.
.
.
.
Copy
.
.
.
References
(DC
P
ower
.
.
.
Using
.
.
.
.
..
..
.
.
..
Using
..
..
..
.
.
..
..
djustments
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
Supply
.
.
DC
..
..
..
DCV
.
.
.
.
..
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
P
..
..
..
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
..
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
or
DCV
.
.
.
.
.
ower Supply
.
.
.
.
.
.
.
.
.
.
.
.
Calibrator
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
..
.
.
.
.
..
.
.
.
.
.
.
.
.
..
..
.
.
..
.
.
.
..
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
..
.
.
.
.
.
..
.
..
.
.
.
.
.
Calibrator)
..
..
..
.
.
.
.
.
.
.
.
.
.
.
.
..
..
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
..
.
.
.
..
..
..
..
.
.
.
.
.
.
.
.
..
..
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
..
.
.
.
.
.
..
.
.
..
.
.
..
.
.
..
.
..
.
.
..
..
..
..
..
.
.
.
.
.
.
..
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
..
..
.
.
.
.
.
.
..
.
.
..
.
.
.
.
.
.
..
..
.
.
.
.
..
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
..
..
..
..
.
.
.
..
..
.
.
.
.
.
.
.
.
.
.
..
..
.
.
.
.
.
.
.
.
.
.
.
.
.
.
..
.
.
.
.
.
.
.
.
.
.
.
.
..
..
..
.
.
.
.
..
.
.
.
.
.
.
.
.
.
..
.
.
..
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
..
.
.
.
.
..
..
..
.
.
.
.
..
.
.
.
.
.
.
.
.
.
.
..
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
..
.
..
.
.
..
.
..
.
.
.
1-1
. 1-1
.
1-2
.
2-1
.
2-1
.
2-1
.
2-2
. 2-2
.
2-2
.
2-2
.
2-2
.
2-3
.
2-4
.
2-4
.
2-4
.
2-5
.
2-6
.
2-6
.
2-7
.
2-7
.
2-7
2-8
2-9
.
2-10
2-10
.
2-10
2-11
.
2-11
3. Assembly Replacement
Introduction . . . . . . . . . . . . . . . . . . . . . . .
Ordering Information
Restored
Replacing
Replaceable Assembly List
Replaceable Mechanical P
Disassembly Procedures .. ...... ...... ...... ..... .. 3-9
Tools and Fasteners ............................ 3-9
Cover Removal . ..... ...... ...... ...... ...... 3-9
A1 Main Board Removal .......................... 3-10
A2 CPU Board Assembly Removal ..................... 3-10
Exchange Assemblies
the A2 CPU Board Assembly
Removal Procedure .... ...... ...... ...... ..... 3-10
Removal Procedure .... ...... ...... ...... ..... 3-10
...... ...... ...... .....
...... ..... ...... ......
....................
........
arts List . . . . . . .
..................
...... ...... ....
...... ....
.....
3-1
3-1
. 3-1
3-2
3-2
3-5
Contents-1

A5
AC
Removal
Keyboard
Removal
LCD
Assembly Removal
Removal
Triaxial
Removal
4.
Troubleshooting
Introduction
Safety
Required
After
Service
Theory
Overall
Overall
Analog
I-V
Digital
P
ower
Troubleshooting
Check
Check
Check
Check
Check
Check 5,
Check
Check
Check
Check
Check
Check
Check
Inlet Board
Procedure .
Assembly Removal
Procedure .
Procedure .
Connector
Procedure
.
.
.
.
Considerations
Equipment
Product
of
Operation
Measurement
Block
Diagram
Section
Converter
Section
Supply
Procedure
1,
2,
3,
4,
6, CPU
7, Any
8, CPU
9, Fuse
10,
11,
12,
.
.
.
.
Section
.
.
References
DC-DC
LCD
Displays
LCD
Operates
Any
Error
CPU
Related
P
ower
Error
Fuse High
and Line
DC-DC
DC-DC
Transformer
Assembly Removal
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
Removal
.
.
.
.
Safety
.
.
Theory
.
.
.
.
.
.
.
.
Converter
Message
LED
Code
Converter
Converter
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
Checks
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
Output
Anything?
Correctly?
Error
Lights?
on
Lead
Switch
Secondary
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
..
.
..
.
.
.
.
.
.
.
.
.
..
on
LCD?
Message?
LED?
V
oltage
.
Input
Open
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
..
.
.
.
.
..
.
.
.
..
..
.
.
.
.
.
.
.
..
..
..
..
Voltage
..
..
.
.
..
.
.
.
.
.
.
.
.
.
.
V
oltage
Output
V
oltage
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
..
.
..
..
.
.
..
.
.
..
.
.
.
.
.
.
..
..
.
.
.
..
.
.
.
.
.
.
.
.
.
.
.
.
V
.
.
.
.
..
.
.
.
.
.
.
.
.
.
.
.
..
..
..
..
.
.
..
.
..
..
.
.
..
.
.
..
..
.
.
..
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
oltages
.
.
.
.
..
.
..
.
..
.
..
.
..
.
..
..
..
..
..
..
..
..
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
..
..
..
..
..
..
..
.
.
.
.
..
.
.
..
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
..
..
..
..
..
..
..
..
.
.
.
.
.
.
..
.
.
..
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
..
..
..
..
..
..
..
.
.
.
.
.
.
.
..
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
..
..
..
..
..
..
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
..
.
..
.
..
..
..
..
..
..
..
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
..
..
..
. 3-10
.
. 3-10
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
..
.
. 4-13
.
. 4-14
..
.
.
. 4-16
. 4-17
3-10
3-10
3-11
3-11
3-12
3-12
4-1
4-1
4-1
4-2
4-3
4-3
4-4
4-4
4-6
4-7
4-8
4-9
4-11
4-11
4-12
4-12
4-12
4-12
4-12
4-14
4-15
A.
Manual
Introduction
Manual
B. Default Jumper and Switch Settings
Introduction . . . . . . . . . .
Jumper and Switch Settings
Index
Contents-2
Changes
.
Changes
.
.
.
.
.
.
.
.
..
.
.
.
.
.
.
..
...... ...... .....
..
..
..
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
..
..
..
..
..
.
.
.
.
.
.
.
.
.
.
.
.
.
.... ...... ...... ...... .
........
A
-1
A-1
B-1
B-1

Figures
1-1. Serial
2-1. A
2-2. Channel
2-3. Channel
2-4. Channel
2-5.
3-1.
3-2.
4-1.
4-2. 4349B
4-3.
4-4. I-V
4-5. Digital
4-6. P
4-7. Troubleshooting
4-8. DC-DC
4-9.
4-10.
4-11.
4-12.
4-13.
B-1.
Number
djustment
1
1A
1A
Channel
Cover
Cable
Ohm's
Analog
ower
CPU
Location
T
est
DC-DC
Transformer
Jumper
1
Removal
Assembly
Law
Overall
Section
Converter
Section Block
Supply
Converter
Board
of
Pin
Locations
Converter
and
A
A
Assembly
Label
Program
djustment,
djustment,
djustment,
djustment,
.
.
Connection
.
.
.
.
Block
Block
Block
Diagram
Section
Flow
Output
High
Lead
.
Output
Secondary
Switch
Settings
.
.
.
Flow
Setup
Setup
Setup
Setup
.
.
.
.
.
.
Diagram
Diagram
Diagram
Block
.
.
V
oltage
LED
Light
of
A2F1
.
.
.
Connector
V
oltage
.
.
.
.
.
.
1/2
(using
2/2
(using
1/2
(using
2/2
(using
.
.
.
for
Triaxial
.
.
.
.
.
.
.
.
.
Diagram
.
.
.
Check
Sequence
Fuse
.
.
.
Check
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
DC
P
ower
DC
P
ower
DCV
Calibrator)
DCV
Calibrator)
.
.
.
.
Connector
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
..
.
.
.
.
..
..
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
..
.
.
.
.
..
.
.
.
.
.
.
Supply)
Supply)
..
..
.
.
.
..
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
..
.
.
.
.
.
.
.
.
..
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
..
.
.
.
.
.
.
.
.
..
..
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
..
.
.
.
.
.
.
.
.
.
..
.
.
.
.
.
.
.
.
.
.
..
..
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
..
..
.
.
..
.
.
.
.
.
.
.
.
..
.
..
.
.
.
.
..
.
.
.
.
.
.
.
..
.
.
.
..
.
.
.
..
.
.
..
..
.
..
..
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
..
.
.
..
.
.
.
.
.
.
..
..
.
.
..
.
.
.
.
.
.
.
.
.
.
.
.
.
1-1
.
2-3
. 2-8
. 2-9
.
2-10
2-11
.
3-9
3-12
.
4-3
.
4-4
.
4-5
.
4-6
. 4-7
4-8
.
4-10
. 4-11
.
4-13
.
4-14
.
4-15
.
4-16
.
4-17
.
B-1
Tables
1-1.
Required
1-2.
Required
1-3.
Substitute
2-1. Related A
3-1. Replaceable Major Assembly List
3-2. A2 CPU Board Assembly
3-3. Replaceable Assembly List
3-4. Replaceable Mechanical P
3-5. Replaceable Mechanical P
3-6. Replaceable Mechanical P
3-7. Replaceable Mechanical P
4-1. DC-DC Converter Output Voltages (Connected) . ..... ...... ... 4-11
4-2. DC-DC Converter Open Output Voltages . . . . . . . . . . . . . . . . . . . 4-16
A-1. Manual Changes by Serial Number ... ..... ...... ...... . A-1
Equipment
Language
Equipment
djustments . . . . . . . . .
.
.
.
.
.
.
.
.
Extensions
.
, Replaceable Assembly List . . . . . . . . . . . . .
arts List 1 . . . . . . . . . . . . . . . . . .
arts List 2 . . . . . . . . . . . . . . . . . .
arts List 3 . . .
arts List 4 . . . .
.
.
.
.
.
.
.
.
..
.................
...... ...... ...... ..
..
..
..
..
.
.
.
.
.
.
.
.
.
.
.
.
..
..
..
..
..
..
..
.
.
.
.
.
.
.
.
.
.
.
.
..
.
.
.
.
.
...................
.....
.....
...
...
...... ...... ......
.................
.
.
.
.
.
.
Contents-3
1-2
1-2
1-2
2-2
3-3
3-4
3-4
3-5
3-6
3-7
3-8


1
General
Information
Introduction
This
chapter
equipment
provides
for
adjustments
Instruments
Agilent
the
characters
T
serial
echnologies
number
are
the
information on
and troubleshooting.
Covered
uses
a
two-section,
label
(
Figure
serial
prex,
by
1-1
and
instruments covered
This
)
attached
the
Manual
nine
to
last
ve
character
the
digits
by this
serial number
instrument's
are
the
sux.
manual
rear
which
panel.
and
required
is
printed
The
rst
on
ve
Figure
An
instrument
number
the
prex
instrument
instrument
manual
manual
part
to
In addition to
errors (Errata) in
Technologies recommends that
manufactured
that
is
dierent
may
be
number
the
newer
is
not
accompanied
.
This
instrument.
listed
from
sheet
after
on
change information, the supplement may contain information for correcting
the manual. T
o keep this manual as current and accurate as possible
you periodically request the latest Manual Changes supplement.
The supplement for this manual is identied by this manual's printing date and its
both of which appear on the manual's title page
available from Agilent T
echnologies. If the serial prex or number of an instrument
than that on the title page of this manual, see
1-1.
Serial
the
printing
the
title
page
those
described
by
a
yellow
Manual
contains \change
Appendix A, Manual Changes
Number
date
of
.
This
in
this
Changes
information" that
Label
this
manual
unlisted
manual.
may
serial
number prex
The
manual
supplement
explains how
have
or
a serial
for
this
have
a
to adapt
part number
. Complimentary copies of the supplement
.F
or information
concerning, a serial number prex that is not listed on the title page or in the Manual
supplement, contact the nearest Agilent Technologies oce.
indicates
new
dierent
the
, Agilent
,
are
is lower
Change
General Information 1-1

Required Equipment
T
able
1-1
lists
lists
the
the
substitutions
the required
required
language
for
the
equipment for
extensions
Multimeter
T
able
adjusting and
to
be used
and
DC V
1-1.
with the
oltage Source
Required
troubleshooting the
adjustment program.
.
Equipment
4349B.
Table
Table
1-3 lists
1-2
Equipment Requirements Recommended
Multimeter DCV
DC
V
oltage Source
RC
Box
V
oltage
No
A
ccuracy:
Range: 1
Substitute
<
to 10
6
0.1
3458A 1 A1,
%
6634A 1 A
Vdc
16340A 1 A
Computer No Substitute HP 9000
RAM
Operating
A
djustment
System
Program
B
ASIC
No
Cable Dual
A
dapter
1
A
djustment
2
Troubleshooting
3
Excluding
CLOCK
CRT
CRTX
CS80
the
HP
Name
A
9826A
Triaxial
A
dapter
T
V
er
.
5.0
5.1
5.1
1
5.0
able
5.1
or higher
substitute
Banana-BNC(m)
T
ee
(m)(f)(f)
1-2.
Required
Name
EDIT
ERR
FGPIB
GRAPH
V
5.1
5.1
5.0
5.2
er
Cable
.
HP
Agilent
Agilent
Agilent
Language
Name
GRAPHX
GPIB
IO
KBD
98616A
Extensions
series 200
4 Mbytes
PN
04349-65005
PN
11001-60001
PN
1250-1551
V
er
.
5.2
5.0
5.1
5.1
Model
Name
MA
T
MS
PDEV
or 300
Qty.Use
3
1 A
1 A
1 A
1 A
1 A
V
er
.
5.1
5.1
5.0
2
T
1
This
language
1-2 General Information
extension
depends
T
able
on
1-3.
the
mass
storage
Substitute
type
used.
Equipment
Equipment Substitution
Multimeter 3455A
3456A
3478A
DC V
oltage Source
1
3245A
DATRON
4000
DATRON 4000A
DATRON 4700
DA
TRON 4707
DATRON 4708
FLUKE 5700A
1
Substituting the DC Voltage Source involves an adjustment
procedure change, refer to \DC Voltage Source (DC Power
Supply or DCV Calibrator)" in Chapter 2 for more details.

A
djustments
Introduction
This
chapter describes
specications
adjustments
optimum
proceed
Note
after it
should be
operating condition.
to Chapter
T
o
warm-up
the adjustments
has been
performed along
4.
ensure
proper
(stabilization)
required to
repaired, or
with periodic
If proper
performance cannot
results
is
when it
and
correct
required
ensure
fails
the
the
performance
maintenance to
be achieved
instrument
before
performing
4349B
operation,
is
within
tests
keep
after
the
adjustments
its
published
.
The
the
4349B
adjustments
a
60
minute
2
in
,
.
Safety
This
operator
Consideration
chapter
safety
adjustments
aware
W
of
arning
the
Required
T
able 1-1
lists the
equipment.
contains
covered
and to
NO
maintain
in this
hazards.
Any
interruption
the
instrument)
make
ground
the
system
Equipment
equipment required
Use
only
calibrated
TE
s
,
CA
UTION
the
chapter
or
instrument
for
equipment
s
,and
W
instrument
should
of
the
protective
be
disconnection
dangerous
any
reason
to perform
when
ARNING
in
a
safe
s
and
performed
ground
of
the
.
Intentional
is
prohibited.
the adjustments
adjusting
which
must
serviceable
only
by
a
conductor
protective
interruption
.T
the
4349B
be
followed
condition.
qualied
(inside
ground
able
1-3
.
person
or
terminal
of
the
lists
the
to
ensure
The
who
is
outside
can
protective
substitute
Adjustments 2-1

Adjustment
No
hand
tools
A
djustable Components
Tools
are
required for
the adjustments
.
There
4349B
are
.
no
All
adjustable
adjustments
components
are
performed
,
such
EEPROM.
F
actory-Selected Components
No
factory-selected
EEPROM
There
are no
Related
There
Option
are
four
001. T
EEPROM write
A
able
components
W
rite
Protect
protect
djustments
adjustments
2-1
lists
for
the
are used
Switch
switches
the
standard
adjustments
replaced.
as variable
by
updating the
in the
in
4349B
that
4349B
the
,
must
4349B
and
capacitors or
adjustment data
.
two
adjustments
be
performed
variable resistors
if
an
stored in
for
assembly
the
the
4349B
has
,
in
the
internal
with
been
2-2 Adjustments
T
able
2-1.
Assembly
A1
Main Board
A2 CPU
A5 A
Board Assy
C Inlet Board Assembly
Assembly
Related
Replaced
without
A
djustments
ROM
Required
All
adjustments
1
None
None
A
djustment
LCD Assembly None
DC-DC Converter Unit None
Keyboard Assembly None
1
If the EEPROM, A2U1, is replaced, all adjustments are required.

Adjustment
The
adjustment
the
computer
data,
and
The
adjustment
installed
start
the
controls
stores
into
your
program
Program
program
the
the
data
program
computer
after
it
listed
in T
4349B
in
the 4349B
runs
on
according to
has
been installed,
able 1-1
and
the other
the
computer
is required
's internal
the procedure
refer to
for the
pieces of
equipment to
EEPROM, A2U1.
listed
in T
able 1-1
described in
\Program Execution
adjustments.
obtain the
.The
program should
\Program Installation
Using the
".
program,
adjustment
be
".
T
o
Figure
each
data
disk.
2-1
menu
both
shows
item.
to
the
the
At
the
4349B
adjustment
end
of each
internal
program
ow.
adjustment, the
EEPROM and
The following
program writes
to the
unit-serial-number dependent
lists the
brief
the
explanation
obtained
for
adjustment
le on
the
INITIAL
Channel
Channel 2 A
Channel 3 A
Channel 4 A
SETUP
1
A
djustment
djustment performs the same as the \Channel 1
djustment performs the same as the \Channel 1 A
djustment performs the same as the \Channel 1 A
Rewrite EEPROM
Figure
denes
values
obtains
scaling. This item should be performed if the performance test
2-1.
the
standards'
as
necessary
all
the
A
djustment
calibration values
.
channel 1
Program
related
data
Flow
. Update
such
as
the calibration
input
resistance
and
fails on
this channel.
Adjustment" for the channel 2.
djustment"
for the channel 3.
djustment" for the channel 4.
writes the adjustment data, which is stored in a unit-serial-number
dependent le, to the EEPROM. Only the adjustment data of
previously adjusted channel(s) will be written.
Adjustments 2-3

Note
Make
\UUT:
the
adjustment
sure
4349B"
that
the
correct
window
data
accuracy
line
frequency
preceding
.
the
Main
(50 or
Menu.
60Hz) has
This selection
been selected
will improve
in the
Program
The
adjustment
The
installation
Conguration
Making
Copy
prevent
harddisk
Dening
Dene
types
the
a
the
contents
accidental
and
the
the
and
the
program
W
conguration
conguration
Note
Refer
to
the
1.
Turn
on
the
Installation
program
can
be
divided
".
orking
of
Copy
the
deletion
store
the
original
Conguration
of
GPIB
addresses
depends
has
following
changed.
DC
V
oltage
to \DC
on
Voltage
procedure
computer
must
be
installed
into
adjustment
or
destruction
disk
the
adjustment
of the
equipment
Source
Source
to
and
bring
in
the
two
stages
,
\Making
program
in
a
of
safe
disk
the
place.
equipment.
multimeter and
conguration, it
selection
dene the
up
(DC
the
involves
P
ower
conguration.
B
ASIC
computer
a
to
a working
program
The
dc voltage
must be
the
Supply
system.
using
the
W
orking
disk or
les
.Use
conguration
source
redened if
adjustment
or
DCV
Calibrator)"
following
Copy" and
to the
the working
procedures
\Dening the
harddisk to
disk or
includes
to
be
used.
the equipment
procedure change
for
more
the
Because
. Refer
details
.
the
.
2. Set
3. Load
4.
5.
the
Mass
exists using
the \
LOAD
"TE_A4349B"
Run
the
loaded
As
the
program
Storage
the
MSI
TE_A4349B
4
program
instructs
Unit
command.
"
Return
address.
6. As the program instructs
GPIB address
Note
.
Once the types and GPIB addresses of both the Multimeter and DC V
Source are set, they cannot be changed unless the conguration is redened.
Specier
le
5
(or
by
,
select
from
pressing
4
Enter
the
the
to
the
drive/directory
adjustment
5
).
RUN
.
type
of
program
the
Multimeter
, select the type of the DC V
where
disk
to
the
by
be
adjustment
pressing
used
and enter
program
its GPIB
oltage Source to be used and enter its
oltage
2-4 Adjustments

Program
P
erform
the
Execution
following
steps
to
start
the
adjustment
program.
Note
1.
Turn
required
2.
Set
exists
3.
Load
4.
Execute
Note
Note
on
the
the
language
Mass
using
the
the
Install the
computer
Storage
the
MSI
program,
loaded
In
the
If
the
change
test
adjustments
program prior
and
extensions
Unit
command.
\
ADJ4349B
program
program,
adjustments
switch
errors
at
have
to execution.
bring
up
the
B
ASIC
.
Specier
by
A2SW1-1
"
by
pressing
the
being
to
the
pressing
4349B is
performed
to \CLOSE",
drive/directory
RUN
referred as
power-on. (Return
been
completed.)
system.
LOAD
"ADJ4349B"
.
follow
the switch
Refer
Refer to
Refer
\UUT".
so that
to Figure
\Program Installation
to
the
T
able
where
right
the
to
4
after
4349B
the
the
5
Return
EEPROM
\OPEN"
B-1
adjustment
(or
will
for
1-2
for the
5
).
4
Enter
replacement,
ignore
position
the
switch
".
program
the
self
when
location.
the
Adjustments 2-5