The information contained in this document is subject to change without notice.
This document contains proprietary information that is protected by copyright. All rights
are reserved. No part of this document may be photocopied, reproduced, or translated to
another language without the prior written consent of Agilent Technologies.
Microsoft®,MS-DOS®,Windows®,Visual C++®,Visual Basic®,VBA® and Excel® are
registered
UNIX is a registered trademark in U.S. and other countries, licensed
exclusively through X/Open Company Limited.
The manual’s printing date and part number indicate its current edition. The printing date
changes when a new edition is printed. (Minor corrections and updates that are
incorporated at reprint do not cause the date to change.) The manual part number changes
when extensive technical changes are incorporated.
October 1999 First Edition
June 2000 Second Edition
January 2007 Third Edition
2
Safety Summary
The following general safety precautions must be observed during all phases of operation,
service, and repair of this instrument. Failure to comply with these precautions or with
specific WARNINGS elsewhere in this manual may impair the protection provided by the
equipment. In addition it violates safety standards of design, manufacture, and intended use
of the instrument.
The Agilent Technologies Company assumes no liability for the customer’s failure to
comply with these requirements.
NOTE4294A comply with INSTALLATION CATEGORY II and POLLUTION DEGREE 2 in
IEC61010-1. 4294A are INDOOR USE product.
NOTELEDs in 4294A are Class 1 in accordance with IEC60825-1.
CLASS 1 LED PRODUCT
•Ground The Instrument
To avoid electric shock hazard, the instrument chassis and cabinet must be connected to
a safety earth ground by the supplied power cable with earth blade.
•DO NOT Operate In An Explosive Atmosphere
Do not operate the instrument in the presence of flammable gasses or fumes. Operation
of any electrical instrument in such an environment constitutes a definite safety hazard.
•Keep Away From Live Circuits
Operating personnel must not remove instrument covers. Component replacement and
internal adjustments must be made by qualified maintenance personnel. Do not replace
components with the power cable connected. Under certain conditions, dangerous
voltages may exist even with the power cable removed. To avoid injuries, always
disconnect power and discharge circuits before touching them.
•DO NOT Service Or Adjust Alone
Do not attempt internal service or adjustment unless another person, capable of
rendering first aid and resuscitation, is present.
•DO NOT Substitute Parts Or Modify Instrument
Because of the danger of introducing additional hazards, do not install substitute parts
or perform unauthorized modifications to the instrument. Return the instrument to a
Agilent Technologies Sales and Service Office for service and repair to ensure that
safety features are maintained.
•Dangerous Procedure Warnings
Warnings, such as the example below, precede potentially dangerous procedures
throughout this manual. Instructions contained in the warnings must be followed.
WARNINGDangerous voltages, capable of causing death, are presenting this instrument. Use
extreme caution when handling, testing, and adjusting this instrument.
3
Safety Symbol
General definitions of safety symbols used on the instrument or in manuals are listed
below.
Instruction Manual symbol: the product is marked with this symbol when it is necessary for
the user to refer to the instrument manual.
Alternating current.
Direct current.
On (Supply).
Off (Supply).
In position of push-button switch.
Out position of push-button switch.
Frame (or chassis) terminal. A connection to the frame (chassis) of the equipment which
normally include all exposed metal structure.
WARNINGThis warning sign denotes a hazard. It calls attention to a procedure, practice,
condition or the like, which, if not correctly performed or adhered to, could result in
injury or death to personnel.
CAUTIONThis Caution sign denotes a hazard. It calls attention to a procedure, practice, condition or
the like, which, if not correctly performed or adhered to, could result in damage to or
destruction of part or all of the product.
NOTENote denotes important information. It calls attention to a procedure, practice, condition or
the like, which is essential to highlight.
Certification
Agilent Technologies Company certifies that this product met its published specifications
at the time of shipment from the factory. Agilent Technologies further certifies that its
calibration measurements are traceable to the United States National Institute of Standards
and Technology, to the extent allowed by the Institution’s calibration facility, or to the
calibration facilities of other International Standards Organization members.
4
Warranty
This Agilent Technologies instrument product is warranted against defects in material and
workmanship for a period corresponding to the individual warranty periods of its
component products. Instruments are warranted for a period of one year. Fixtures and
adapters are warranted for a period of 90 days. During the warranty period, Agilent
Technologies Company will, at its option, either repair or replace products that prove to be
defective.
For warranty service or repair, this product must be returned to a service facility designated
by Agilent Technologies. Buyer shall prepay shipping charges to Agilent Technologies and
Agilent Technologies shall pay shipping charges to return the product to Buyer. However,
Buyer shall pay all shipping charges, duties, and taxes for products returned to Agilent
Technologies from another country.
Agilent Technologies warrants that its software and firmware designated by Agilent
Technologies for use with an instrument will execute its programming instruction when
property installed on that instrument. Agilent Technologies does not warrant that the
operation of the instrument, or software, or firmware will be uninterrupted or error free.
Limitation of Warranty
The foregoing warranty shall not apply to defects resulting from improper or inadequate
maintenance by Buyer, Buyer-supplied software or interfacing, unauthorized modification
or misuse, operation outside the environmental specifications for the product, or improper
site preparation or maintenance.
IMPORTANTNo other warranty is expressed or implied. Agilent Technologies specifically disclaims the
implied warranties of merchantability and fitness for a particular purpose.
Exclusive Remedies
The remedies provided herein are buyer’s sole and exclusive remedies. Agilent
Technologies shall not be liable for any direct, indirect, special, incidental, or
consequential damages, whether based on contract, tort, or any other legal theory.
Assistance
Product maintenance agreements and other customer assistance agreements are available
for Agilent Technologies products.
5
For any assistance, contact your nearest Agilent Technologies Sales and Service Office.
Addresses are provided at the back of this manual.
Typeface Conventions
Bold Boldface type is used when a term is defined. For
example: icons are symbols.
ItalicItalic type is used for emphasis and for titles of
The following manuals are available for the 4294A.
•Operation Manual (Agilent P/N: 04294-900x0)
Most of basic information necessary for using 4294A is described in this manual. It
includes the way of installation, preparation, measurement operation including
calibration, performances (specifications), key definitions, and error messages. For
GPIB programming, see the Programming Manual together with “HP Instrument BASIC User's Handbook”.
•Programming Manual (Agilent P/N: 04294-900x1)
The Programming Manual shows how to write and use BASIC program to control the
4294A and describes how HP Instrument BASIC works with the analyzer.
The HP Instrument BASIC User’s Handbook introduces you to the HP Instrument
BASIC programming language, provide some helpful hints on getting the most use
from it, and provide a general programming reference. It is divided into three books,
HP Instrument BASIC Programming Techniques, HP Instrument BASIC Interface
Techniques, and HP Instrument BASIC Language Reference.
This Service Manual is a guide to servicing the 4294A Precision Impedance Analyzer. The
Service Manual provides information about performance testing, adjusting,
troubleshooting, and repairing the 4294A.
13
General Information
Organization of Service Manual
Organization of Service Manual
This manual consists of the major chapters listed below. This section describes the names
of the chapters and the content of each chapter.
•Performance Test provides procedures for performance testing the 4294A.
•Adjustment provides procedures for adjusting the 4294A after repair or replacement of
an assembly. Most of the adjustments update the correction constants stored into the
EEPROM on the A1 CPU. The correction constants are updated by using the
adjustment program.
•Troubleshooting provides troubleshooting procedures to isolate any faulty functional
group or faulty assembly. Faulty assembly isolation procedures for the faulty functional
group follow this chapter.
•Power Supply Troubleshooting provides troubleshooting procedures to isolate any
faulty assembly within the power supply functional group.
•Digital Control Troubleshooting provides troubleshooting procedures to isolate any
faulty assembly within the digital control functional group.
•Service Related Menus documents the functions of the menus accessed from [System],
DIAGNOSTIC TEST. These menus are used to test, verify, and troubleshoot the 4294A.
•Theory of Operation explains the overall operation of the 4294A, the division into
functional groups, and the operation of each functional group.
•Parts Replacement provides part numbers and illustrations of the replaceable
assemblies and miscellaneous chassis parts. This chapter also provides procedures to
disassemble portions of the analyzer when certain assemblies have to be replaced.
•Post Repair Procedures contains the table of related service procedures. It is a table of
adjustments and verification procedures to be performed after repair or replacement of
each assembly.
•Appendices contains manual change information (required to make this manual
compatible with earlier shipment configurations of the analyzer), power requirements
and error messages of the analyzer.
141
Instruments Covered by This Manual
Agilent Technologies uses a two-part, ten-character serial number label (See Figure 1-1)
attached to the instrument’s rear panel. The first five characters are the serial prefix and the
last five digits are the suffix.
Figure 1-1 Serial Number Label
An instrument manufactured after the printing date of this manual may have serial number
prefix that is not listed on the title page. This unlisted serial number prefix indicates the
instrument is different from those described in this manual. The manual for this new
instrument may be accompanied by a yellow Manual Changes supplement or have a
different manual part number. This sheet contains “change information” that explains how
to adapt the manual to the newer instrument.
General Information
Instruments Covered by This Manual
1. General Information
In addition to change information, the supplement may contain information for correcting
errors (Errata) in the manual. To keep this manual as current and accurate as possible,
Agilent Technologies recommends that you periodically request the latest Manual Changes
supplement. The supplement for this manual is identified by this manual’s printing data
and is available from Agilent Technologies. If the serial prefix or number of an instrument
is lower than that on the title page of this manual, see Appendix A, Manual Changes. For
information concerning, a serial number prefix that is not listed on the title page or in the
Manual Changes supplement, contact the nearest Agilent Technologies office.
1 15
General Information
Require Equipment
Require Equipment
Table 1-1 lists the recommended equipment for performing maintenance on the 4294A.
*1.P:Performance Test A:Adjustment
*2.16048A/D cannot be used for a substitute.
*3.furnished with the 4294A
*1
1 17
General Information
Require Equipment
181
2. Performance Test
2 Performance Test
This chapter contains the performance test for the 4294A Precision Impedance Analyzer.
These performance tests are used to verify that the analyzer’s performance meets its
specifications.
19
Performance Test
Performance Test
Performance Test
Introduction
This section provides the test procedures used to verify that the 4294A’s specifications are
met. The performance tests can also be used for incoming inspection, and for verification
after troubleshooting or adjustment. If the performance tests indicate that the 4294A is
NOT operating within the specified limits, check your test setup, then proceed with
troubleshooting if necessary.
War m Up Time
Allow the 4294A to warm up for at least 30 minutes before you execute any of the
performance tests
Ambient Conditions
Perform all performance tests in ambient conditions of 23 °C ± 5 °C, ≤ 70% RH.
Performance Test Interval
The performance test should be performed periodically. The recommended test interval is
12 months.
NOTEThe test interval depends on maintenance of use and the environmental conditions under
which the instrument is used. You may find that the test interval could be shortened or
lengthened; however, such a decision should be based on substantial quantitative data.
Performance Test Record and Calculation Sheet
Performance test record lists all test points, acceptable test limits, test result entry columns,
and measurement uncertainties. The listed measurement uncertainties are valid only when
the recommended test equipment is used.
The calculation sheet is used as an aid for recording raw measurement data, and for
calculating the performance test results.
The procedure for using the calculation sheet and performance test record is;
1. Photo copy the calculation sheet.
2. Follow the performance test procedure and record the measurement values, the
4294A’s reading, etc., into the specified column on the calculation sheet.
3. Calculate the test result using the appropriate equation given on the calculation sheet,
and record the test result into the Test Result column of the performance test record.
Test Equipment
Table 1-1 lists the recommended equipment for performing maintenance on the 4294A.
202
Performance Test
Performance Test
Frequency Accuracy Test
This test checks the frequency accuracy of the internal frequency reference with a
frequency counter and a frequency standard.
Specification
Frequency Accuracy: ± 20 ppm at (23±5) ºC (without Option 1D5)
± 0.13 ppm at 0 ºC to 55 ºC (with Option 1D5)
Test Equipment
DescriptionRecommended Model
Frequency Counter5334B or 53181A
Frequency Standard5071A
4TP to BNC Interface Boxp/n 04294-61002
BNC(m) -BNC(m) Cable, 61 cmp/n 8120-1839
2. Performance Test
BNC(m)-BNC(m) Cable, 122 cmp/n 8120-1840
Procedure
1. Connect the test equipment as shown in Figure 2-1. For testing the 4294A with Option
1D5, connect a BNC(m)-BNC(m) cable between the EXT REF Input connector and the
REF OVEN connector on the analyzer rear panel.
NOTEIf testing the 4294A without Option 1D5, the frequency standard is not required.
2 21
Performance Test
Performance Test
Figure 2-1 Frequency Accuracy Test Setup
2. Initialize the frequency counter. Then set the controls as follows.
Gate Time: 1 sec.
INT/EXT Switch (rear panel, 5334B only)
*1.If the frequency standard is not connected, set the switch to INT.
EXT
*1
3. Press [Preset] to initialize the 4294A. Then set the controls as follows.
SettingOperation
Frequency Span: 0 Hz[Span] - [0] - [×1]
Center Frequency: 10 MHz[Center] - [1] - [0] - [M/m]
2. Press Reset key (blue,⇒) to initialize the multimeter, then Set it as follows.
a. Press [ACV] to set the measurement mode to AC voltage.
b. Press S(blue - N Rdgs/Trig),⇓,⇓,⇓ to display SETACV.
c. Press ⇒,⇓,⇓,⇓ to display SYNC, then press [Enter]
d. Press [NPLC] - [1] - [0] - [0] - [Enter]
3. Press [Preset] to initialize the 4294A. Then set the controls as follows.
SettingOperation
Frequency Span: 0 Hz[Span] - [0] - [×1]
Center Frequency: 1 kHz[Center] - [1] - [k/m]
OSC Level: 1 V[Source] - LEVEL - [1] - [×1]
OSC Level Monitor: ON[Display] - OSC MON
Number of Points[Sweep] - NUMBER OF POINTS - [2] - [×1]
4. Press [Trigger] - Single for a single sweep measurement:
5. Record the multimeter reading and the OSC Level Monitor Reading to the calculation
sheet.
6. Calculate the monitor accuracy according to the calculation sheet, and record the
monitor accuracy into the calculation sheet.
NOTEOPEN Reading is recorded in the Step22.
242
7. Confirm that the monitor accuracy is within the monitor accuracy limit.
NOTEIf the monitor accuracy is beyond the limit, OSC Level Test fails.
8. Repeat from Step 4 to 7 for all OSC level and frequency setting in Table 2-1
Table 2-1 OSC Level Accuracy Test (Interface box) Settings 1
Performance Test
Performance Test
OSC LevelFrequency
1.0 V1 kHz
100 kHz
0.5 V40 Hz
1 kHz
10 kHz
100 kHz
0.25 V1 kHz
100 kHz
0.125 V1 kHz
100 kHz
0.064 V1 kHz
100 kHz
0.016 V1 kHz
100 kHz
2. Performance Test
Accessory: Nothing, 1 MHz to 110 MHz
9. Connect the power sensor to the power meter. Calibrate the power meter for the power
sensor.
10. Set the power meter measurement mode to “Watt”
11. Disconnect the BNC(m)-BNC(m) Cable from the interface box, and connect the
equipment as shown in
NOTEThe OSC level must be set to 0.75V because impedance mismatching occurs between the
4294A and the power sensor. The calculation sheet and the test record say the OSC level is
0.5 V.
13. Press [Trigger] - Single for a single sweep measurement:
14. Record the following values according to the calculation sheet.
a. OSC level monitor reading
b. Power meter reading
c. OSC level calculated from the power meter reading.
d. OSC level monitor accuracy.
15. Confirm that the OSC level monitor accuracy is within the monitor accuracy limit.
NOTEIf the monitor accuracy is beyond the limit, OSC Level Test fails.
262
Performance Test
Performance Test
16. Repeat from Step 13 to 15 for all frequency settings in Table 2-2
Table 2-2 OSC Level Accuracy Test (Interface box) Settings 2
Frequency
1 MHz
10 MHz
15 MHz
16 MHz
50 MHz
110 MHz
17. Disconnect the all test equipment from the 4294A, then connect the OPEN termination
directly to the 4294A UNKNOWN terminals.
18. Press [Source] - LEVEL - [.] - [5] - [×1] to set the 4294A OSC level to 0.5 V.
19. Press [Trigger] - Single for a single sweep measurement:
20. Read the level monitor reading for all OSC level and frequency setting in Table 2-2,
then record them to the OPEN Reading column in the calculation sheet.
21. Press [Trigger] - Single for a single sweep measurement:
22. Read the level monitor reading for all OSC level and frequency setting in Table 2-1,
then record them to the OPEN Reading column in the calculation sheet.
23. Calculate the test results according to the calculation sheet, then record them to the test
record.
With 42942A Terminal Adapter
24. Disconnect the all test equipment from the 4294A, then connect the equipment as
shown in
Figure 2-4
2. Performance Test
2 27
Performance Test
Performance Test
Figure 2-4 42942A Setup
25. Setup the 42942A as follows.
a. Press [Cal],Adapter,7mm 42942A,SETUP.
b. Connect the OPEN termination furnished with the 42942A to the 7mm connector.
c. Press PHASE COMP to measure the phase compensation data.
d. Press OPEN to measure the open compensation data.
e. Connect the SHORT termination furnished with the 42942A instead of the OPEN
termination.
f. Press SHORT to measure the short compensation data.
g. Connect the 50 Ω termination furnished with the 42942A instead of the SHORT
termination.
h. Press LOAD to measure the load compensation data.
i. Press done to activate the measured compensation data.
26. Connect the instruments as shown in Figure 2-5.