Agilent 4291B Data Sheet

New Technologies for Wide Impedance Range Measurements to 1.8 GHz
Product Note 4291-1
HP 4291B RF Impedance/ Material Analyzer
Introduction
With the current trends in the communications and data processing industries requiring higher performance, smaller physical size, lower cost, and higher reliability, accurate and efficient electronic component characterization is an increasingly important part of product design. Many of these advanced products are operating at frequency in the RF range.
Older low-frequency or indirect methods of component evaluation are often not able to provide the high quality impedance parameter information required for
HP 4291B RF Impedance/Material Analyzer with SMD fixtures
understanding component performance under actual operating conditions in the RF range. This product note describes new impedance measurement techniques and innovations contained in the HP 4291B RF Impedance/ Material Analyzer that allow accurate and efficient direct impedance measurement and analysis from 1 MHz to 1.8 GHz. Five general topics will be discussed:
 Limitations of traditional
methods of impedance analysis.
 Extending the frequency range
and impedance magnitude range, while maintaining high accuracy.
1
 Making high accuracy Quality
Factor (Q) and Dissipation Factor (D) measurements at RF.
 Increasing test flexibility by
extending the device-under-test (DUT) location 1.8 meters from the instrument while maintaining accuracy.
 Eliminating fixture errors
critical issues for accurate RF measurements.
Limitations of traditional impedance measurement solutions
Most conventional measuring instruments such as traditional LCR meters and impedance analyzers are limited to lower frequency analysis where the 4-terminal pair method can provide very high accuracy. For higher frequencies a vector network analysis approach is often used by first measuring the reflection coefficient, then calculating impedance values. This indirect method may be useful for impedance values near 50 , but can be inaccurate for impedance values significantly higher or lower than 50 . In addition, calibration and removal of test fixturing errors can be a tedious procedure and in some cases, not possible. The issue of fixturing at RF can be a major
source of error and expense. In general, impedance measurements in the RF range in the past have been very difficult and often yield widely varying results with questionable accuracy. For today's RF component or circuit designer, a new solution is needed to provide the accuracy and test efficiency required for complete RF impedance characterization of circuit elements.
Hewlett-Packard now offers a new analyzer dedicated to the direct measurement of impedance and material parameters from 1 MHz to 1.8 GHz that overcomes many of the limitations of previous approaches. The HP 4291B provides highly accurate RF impedance measurements and offers a family of surface mount device (SMD) test fixtures. With a color display and powerful firmware, up to 15 impedance parameters as well as equivalent circuit models and more are easily measured and displayed. The following topics describe this new analyzer's capabilities and measurement technology.
Figure 1. Ranges of impedance measurement (Accuracy of measurement: 10%)
The HP 4291B achieves its accuracy over these wide impedance ranges using tow new techniques:
 Direct impedance using
the RF I-V method The analyzer uses a new method
to measurement impedance; the RF I-V method. Figure 2 shows the basic principles of the RF I-V method and reflection coefficient method (conventional method using a vector network analyzer).
and reflection coefficient methods, the impedance is given by ratios of the readings of the two voltmeters. Therefore one would expect that the accuracy of both methods would be similar. However, the reflection coefficient method magnifies the measurement error when converting reflection coefficient to impedance. As impedance goes away from 50 , Figure 3 shows that a small changes in reflection coefficient value produces a large change in impedance. In other words, a small error in the reflection coefficient leads to a large error in impedance. (For example, for an impedance of 2 k, a 1 % error of reflection coefficient results in a 24 % error in impedance.)
Figure 3. Relationship between impedance and reflection coefficient
Expanding the range of impedance magnitudes
As shown in Figure 1, the HP 4291B provides an exceptionally wide range of impedance measurements. It is the ideal instrument for measuring very small capacitance (1 pF) and inductance (1 nH) values in the RF range.
Figure 2. I-V method and reflection coefficient impedance method
As can be seen from the equations in Figure 2, the RF I-V method measures impedance directly, while the reflection coefficient method measures reflection coefficient and concerts it to impedance. In both the RF I-V
2
The RF I-V method measures impedance directly from a ratio of voltage and current, without converting the measured data. Therefore, the RF I-V technique maintains consistent accuracy even if the impedance is significantly larger or smaller than 50 . Thus for measuring non-50 Ω components, the HP 4291B using the RF I-V technique is recommended.
 High/Low impedance circuit
The HP 4291B employs high­impedance and low-impedance circuits, as shown in Figure 4, to expand the range of impedance measurements. When measuring a high-impedance device, accurate measurement of the DUT current is most critical. The high-impedance circuit solves this problem by connecting the current detection circuit directly in series with the DUT, ensuring accurate DUT current measurement and not measuring current flowing in the voltage sensing circuit.
On the other hand, when measuring a low-impedance device, the voltage across the DUT is most critical. In this case, the low-impedance circuit connects the voltage detection circuit directly to the DUT, ensuring accurate voltage measurement and not measuring the voltage drop from the current sensing impedance. By using the right measuring circuit for the impedance being measured, it is possible to extend the range of impedance magnitudes measured for a given accuracy.
Figure 4. High-impedance and low­impedance measuring circuits
High Accuracy Q and D measurements
As shown in Figure 5, the HP 4291B is capable of evaluating a sample with Q = 100 within ±15 % accuracy at 1 GHz. This capability is targeted for evaluating the loss of low-loss components at RF. The accuracy of Q and D measurements depend on the accuracy of phase measurement. See Figure 6.
The HP 4291B improves the accuracy of phase measurements by requiring an additional phase calibration step. (Conventional one-port calibration uses open, short, and 50 load standards.) This type of one-port calibration does not provide satisfactory accuracy for phase measurements because of the phase uncertainty of the 50 standard. By using a low-loss air capacitor as a phase standard, the HP 4291B lowers the phase uncertainty to 1 mrad or less (corresponding to D = 0.001), ensuring improved accuracy for Q and D measurements.
Error-free 1.8m cable extension
The cable connecting the main body of the HP 4291B to the test station has been extended to
1.8 meters without adding additional errors. See Figure 7. The long cable allows easy access to remote DUT locations. This temperature chamber, scanner/handler, or custom test setup, for example.
These two measuring circuits are implemented as two different test heads in the HP 4291B, so users can select and switch the circuits easily to optimize the measurement range and accuracy to best match the DUT impedance.
Figure 5. Q measurement accuracy
Figure 6. Q and D measurement and phase measurement
3
Figure 7. Extended test cable
Normally, extending the cable increases measurement error due to increased noise, temperature differentials, cable resistance, etc. However, as illustrated in Figure 8, the HP 4291B measures the current and voltage signals using the same circuit and alternates the measurement with fast time­division multiplexing.
Since the measurements are made at an alternating interval of several milliseconds, the same cable-induced errors occur in both the current and voltage measurement data. These errors are then canceled out when the impedance is obtain from the ratio of voltage to current.
Furthermore, by measuring current and voltage with the same circuit, errors in the measuring instrument caused by temperature changes are offset in the same manner, resulting in significant temperature characteristic improvement.
Figure 8. Time-division multiplexing of cuurent/voltage measurement
Test fixtures have a large impact on measurement accuracy, especially at higher test frequencies. An important factor in getting accurate measurements is eliminating errors introduced by the DUT fixturing. Calibration insures high accuracy at the plane of calibration (measurement point where the standards are applied), but in actual practice, the test fixture can add additional error terms beyond the calibration plane. This is why fixture error compensation is so important.
Electrical properties of the test fixture (which occur after the calibration point), consist of a phase rotation due to the physical length of the electrodes and other unwanted stray
parasitics between electrodes. Both of these can cause significant measurement errors in the RF band. Conventional measuring instruments often have no convenient method to eliminate them effectively. The HP 4291B uses electrical length compensation to remove the errors caused by phase rotation and OPEN/SHORT compensation (at the DUT location in the fixture) for removing fixture parasitic impedance.
Conclusion
The HP 4291B RF impedance and material analyzer provides highly accurate impedance and material measurements by incorporating new technologies and offering an integrated package including a family of SMD and material fixtures. The analyzer overcomes many limitations of conventional impedance analysis and, for the first time, provides an efficient and accurate measurement solution for passive component analysis over the RF range.
For detailed technical information of the HP 4291B RF impedance and material analyzer, refer the HP 4291B Technical Information in page 5.
For more information, request following literatures from your local HP representative:
HP4291B 1.8 GHz impedance/Material Analyzer Product Overview P/N 5966­1501E
HP 4291B Technical Specifications P/N 5966-1543E
Highly Accurate Evaluation of Chip Capacitors using the HP 4291B Application Note 1300-1 P/N 5966­1850E
Evaluating Chip Inductors using the HP 4291B Application Note 1300-2 P/N 5966-1848E
Permittivity Measurements of PC Board and Substrate Materials using the HP 4291B and HP 16453A Application Note 1300-3 P/N 5966­1847E
Permeability Measurements using the HP 4291B and HP 16454A Application Note 1300-4 P/N 5966-1844E
Electronic Characterization of IC Package Application Note 1300-5 P/N 5966-1849E
Impedance Characterization of Magneto-Resistive Disk Heads Using the HP 4291B Impedance/Material Analyzer Application Note 1300-6 P/N 5966-1096E
On-Chip Semiconductor Device Impedance Mesurements Using the HP 4291B Application Note 1300-7 P/N 5966-1845E
Evaluating Temperature Characteristics using a Temperature Chamber and the HP 4291B Product Note 4291-2 P/N 5966-1927E
Impedance Measurements Using the HP 4291B and the Cascade Microtech Prober Product Note 4291-3 P/N 5966-1928E
Dielectric Comnstant Evaluation of Rough Surfaced Materials Product Note 4291-5 P/N 5966-1926E
4
Appendix.
New Technologies used in the High Frequency Impedance Analyzer
HP 4291B RF Impedance/Material Analyzer Technical information
Abstract
A new one-port impedance analyzer has been developed for analysis of high frequency devices and materials up to 1.8 GHz.
Traditionally, impedances near 50 have been measured accurately by the null method using a directional bridge. However, this new analyzer uses a voltmeter/ammeter method and offers precise measurement capability over a wide impedance range. Furthermore, a special calibration method using a low-loss capacitor realizes an accurate high-Q device measurement.
This paper describes the advantages of these techniques. Impedance traceability of the instrument will also be discussed.
Finally, many types of test fixtures are introduced, because they are a key element in any test system.
1. Introduction
A general impedance measurement schematic using two vector voltmeters is shown in Fig. 1. In this case, the true impedance (Zx) of a device under test (DUT) is determined by measuring the voltages of any two different sets of points in a linear circuit.
Zx = K1 × (1)
where
K1, K2, K3 : complex constant
In the process of deriving the equation above, linearity is assumed but reciprocity is not assumed. Therefore, the existence of active devices in the circuitry is not prohibited.
There are at most three unknown parameters related to the circuit in the equation (1). Once we know these parameters, we can calculate any impedance of the DUT from the measured voltage
K2+Vr
1+K3×Vr
Vr : voltage ratio
E: signal source V1: vector voltmeter1 V2: vector voltmeter2 Zx: DUT
Figure 1. General schematic for impedance measurement using two vector voltmeters
ratio (Vr). The procedure that estimates these circuit parameters is called "calibration" and one method is "Open-Short-Load (OSL) calibration." Calculation of Zx from the measured voltage ratio (Vr) according to equation (1) is called "correction."
2. Transducer
We call a linear circuit such as the one in Fig. 1 (one that relates a signal source, two vector voltmeters and a DUT) a "transducer." Transducers are the key element in impedance measurement. For example, two types of transducers, the "directional bridge" and the transducer in a "voltmeter/ammeter (V-I) method," are compared in terms of sensitivity to the gain variance.
2-1. Directional bridge type
Directional bridges (see Fig. 2-1) are used in many network analyzers.
5
In this case, the bilinear transformation is
1+
Zx = −R
Γ
× (2)
0
1
Γ
where
(Z
R0)
Γ
Vr = = (− ) ×
R
x
= : reflection coefficient
(Z
+
R0)
x
V
2
V
1
= 50 : characteristic impedance
0
1
Γ
8
Parameters corresponding to those in equation (1) are
K1 = - Ro, K2 = 1, K3 = - 1
E: signal source V1: vector voltmeter1 V2: vector voltmeter2 Zx: DUT R@ Ro = 50 Ro: characteristic impedance
Figure 2-1. Directional bridge circuit
2-2. V-I method
Fig. 2-2 shows the simplest transducer in the V-I method. The bilinear transformation in this case is
Zx = R
× Vr (3)
0
where
R
= 50 : resistor that converts DUT current
0
to a voltage
V
Vr =
2
V
1
Parameters corresponding to those in equation (1) are
E: signal source V1: vector voltmeter1 V2: vector voltmeter2 Zx: DUT R: resistor converting DUT current to a voltage R@ Ro = 50
Figure 2-2. V-I method
2-3. Sensitivity to gain variance
In this section we discuss the relationship between the impedance measurement error and the gain variance of the vector voltmeters.
2-3-1. Directional bridge type
We assume that the vector voltmeters in Fig. 2-1 are not ideal but have some gain variance. In this situation, the measured voltages (V1 and V2) and the calculated impedance (Zx) are:
V
= E ×
1
V2 =
Zx = −R
α
1
( − )
8
×
0
1
× E × Γ ×
1+
Γ
1
Γ
α
2
where
α
: gain of the vector voltmeter1
1
α
: gain of the vector voltmeter2
2
Γ
=
8 × Vr ×
V
2
Vr = : voltage ratio
V
1
α
2
α
= : ratio of voltmeter gains
2
α
α
: measured reflection coefficient
r
1
K1 = R, K2 = 0, K3 = 0
6
We define the calculated impedance sensitivity (S) to the voltmeters' gain variance as follows:
δZ
x
Z
S = (4)
x
δ
α
r
α
r
This sensitivity can be considered as the inverse of the "magnification on gain variance." The smaller S is, the smaller the error of the calculated impedance is. In this case, equation (4) is given as follows:
δ
Z
δΓ
S = × ×
x
δΓ
1
= ( ) ×
2
δα
Z
r
2
− R
x
Zx × R
α
r
Z
x
2
o
o
1
( ) × ( ) for |Z
2
R
o
Z
x
| << Ro
x
= 0 for |Zx| = Ro
1
( ) × ( ) for |Z
2
Z
x
R
o
| >> Ro
x
This implies
1) This type of transducer has little sensitivity to the voltmeter gain variance when the DUT impedance is near Ro (50 Ω).
2) The gain variance of the voltmeters behaves as the offset impedance with a magnitude of (1/2)*Ro*| ∆α r/α r | when the DUT impedance is far smaller than Ro.
3) The gain variance of the voltmeters behaves as the offset admittance with a magnitude of (1/2)*Go*| ∆α r/ α r| when the DUT impedance is far larger than Ro.
where
∆αr: change in gain ratio α
r
Go : 1 / Ro
Fig. 2-3 shows this characteristic.
Figure 2-3. Sensitivity to the voltmeter gain variance
2-3-2. V-I method
We also assume that there is some gain variance in the vector voltmeters in Fig. 2-2. The relationship among measured voltages (V1 and V2) and the calculated impedance (Zx) are:
R
V1 = E ×
V2 = E ×
Zx = Ro × Vr
o
Zx + R
Z
x
Zx + R
× α
r
× α
1
o
× α
2
o
where
α
: gain of the vector voltmeter1
1
α
: gain of the vector voltmeter2
2
V
Vr = : voltage ratio
α
2
V
1
α
1
= : ratio of voltmeter gains
r
α
2
In this case, the sensitivity is given by the following: The error ratio (Zx/Zx) is always constant and
equal to unity. For example, if the voltmeter gain
7
ratio ar changes 1%, an impedance error of 1% is caused for any DUT.
δZ
x
Z
x
S = = 1
δ
α
r
α
r
3.Schematic of new RF impedance
analyzer
From the discussion in section 2, we find that the voltmeter gain variance is neither suppressed nor magnified for all DUT impedances in a "V-I method type transducer." This characteristic is desirable for wide impedance measuring capability. Therefore we adopted this type of transducer for our new one-port RF impedance meter.
3-1. Basic transducer circuit
Fig. 3-1 shows the basic circuit of the transducer in New RF Impedance Meter which is modified from Fig. 2-2. The high impedance configuration realizes perfect OPEN and imperfect SHORT conditions. On the other hand the low impedance configuration realizes imperfect OPEN and perfect SHORT conditions. Moreover, this circuit has the following characteristic:
3-2. Actual transducer circuit
Actual transducer circuits are shown in Fig. 3-2 (a) and Fig. 3-2 (b).
We took the following things into account when designing the actual transducer circuits:
(a) for high impedance measurement
* the output impedance at the DUT port is always Ro in
either switch position
E: signal source V1: vector voltmeter1 V2: vector voltmeter2 Zx: DUT R@ Ro: characteristic impedance switch: ON for high impedance measurement switch: OFF for low impedance measurement
Figure 3-1. Basic transducer circuit in New RF impedance Analyzer
(b) for low impedance measurement
E: signal source V1: vector voltmeter1 V2: vector voltmeter2 Zx: DUT R@Ro Ro: characteristic impedance
Figure 3-2. The actual transducer circuit
1) Because a wideband switch with small non­linearity and with small transients over a wide signal range is not easily realized, we divided the circuit in Fig. 3-1 into two separate circuits.
2) Because the minimum frequency of the new meter is 1 MHz, the floating voltmeter (V1),
8
which corresponds to the current meter, is easily realized by using a balun.
3) In order to prevent the error caused by by-pass current we adopted a circuit in which the exciting impedance of the balun is parallel connected not to the current meter (V1) but to the signal source (E).
3-3. Simplified block diagram of new impedance
analyzer
The followings are other key features of the instrument:
1) time division multiplex Two voltmeters are obtained by time division multiplexing one voltmeter. The multiplexing eriod is 2msec. This ensures that the slow drift of the voltmeter gain does not affect the impedance measurement. The signal path after the multiplexer can be extended. This instrument uses a 1.8m cable between the transducer and the instrument main frame. This allows wide flexibility in constructing a test system using automatic device handlers. Good temperature characteristics are also derived even with an extended cable by the single path configuration.
Figure 3-5. (a) Typical errors for impedance magnitude with the transducer for high impedance
Figure 3-5. (b) Typical errors for impedance magnitude with the transducer for low impedance
4. A special calibration for high Q measurement
Normally the accuracy requirement for the impedance phase is greater than that of the impedance magnitude. Our New Impedance Meter has a special, but easy-to-use, calibration for high Q (quality factor) device measurements. We discuss this calibration technique in this section.
(2) impedance ranging
At frequencies below 200 MHz there exists an "expand range." In the expand range there is a gain difference between the voltage channel and the current channel in the earlier stage of the multiplexer. This impedance ranging offers stable measurements for DUTs with impedances that differ greatly from 50 Ω.
Fig.3-3 (a), Fig. 3-3 (b) shows the typical measurement errors for impedanceof the instrument. The errors mainly consist of the uncertainties for standards (STDs) used in the calibration, non-repeatabilities in connections, temperature coefficients, noise and errors in the interpolations. The impedance phase errors can be reduced by using a special calibration (see Section
4).
4.1 Outline of the special calibration
Even if the stability of the instrument is good enough, accurate Q measurements are not performed without correct markings on the phase scale of the instrument. For instance, if we want to measure the Q factor with 10% uncertainty for a DUT whose Q value is almost 100, the uncertainty for phase scaling must be smaller than 1E-3. The phase accuracy of the instrument is determined almost entirely by the uncertainty of the 50 LOAD STD used in the OSL calibration. One method to improve phase measurement accuracy is to use a phase calibrated LOAD STD. However, it is not ensured that phase uncertainty for a calibrated 50 LOAD is smaller than 1E-3 at high frequencies (such as 1 GHz).
In addition to the normal OPEN-SHORT-LOAD STDs, using a low-loss air-capacitor as the second LOAD (LOAD2), whose dissipation factor (D) is kept below 1E-3 at around 1 GHz, offers the following feature:
9
* the uncertainty for the measured phase is
decreased from the phase uncertainty of the 50 LOAD (LOAD1) to the uncertainty of D of the low-loss capacitor (LOAD2) for almost all DUT impedances.
4.2 Details of the modified OSL calibration using an additional load
We want to have the calibration method which reduces the error in phase measurement in spite of the existence of phase error for the 50 LOAD. Consider the case that we have the 50 LOAD STD whose impedance phase is not known but impedance magnitude is known. In this situation how about adding another LOAD (LOAD2) whose impedance magnitude is not known but impedance phase is known? We use a low-loss capacitor as the second LOAD. The number of unknown circuit parameters are still three at most. However, two more unknowns related to STDs are added. Let us define the problem. There are eight real unknown parameters:
1) circuit parameter K1 (two real parameters)
2) circuit parameter K2 (two real parameters)
3) circuit parameter K3 (two real parameters)
Zlmi : measured impedance for LOAD STD (i = 1 : LOAD1, i = 2 : LOAD2) Zlsi : true impedance for LOAD STD (i = 1 : LOAD1, i = 2 : LOAD2) Zls
= Zabs_ls1*EXP(j*θls1)
1
Zls
= Zabs_ls2*EXP(j*θls2)
2
θls1 = θ2 - θ1 + θls2 Zabs_ls2 = A1/A2*Zabs_ls1 Zabs_ls1 : impedance magnitude for LOAD1 (50) : known θls2 : impedance phase for LOAD2 (low-loss capacitor) : known
θ1 = arg((1 - Zlm θ2 = arg((1 - Zlm
*Yom)/(Zlm1 - Zsm))
1
*Yom)/(Zlm2 - Zsm))
2
A1 = |(1 - Zlm1 *Yom)/(Zlm1 - Zsm)| A2 = |(1 - Zlm2 *Yom)/(Zlm2 - Zsm)|
For the actual cases these circuit parameters are expressed by far more complicated equations. Therefore
we adopted a simpler procedure consisting of two steps:
4) impedance phase qls1 of the 50 LOAD
(one real parameter)
5) impedance magnitude Zabs_ls2 of the low-loss capacitor LOAD2 (one real parameter)
We solved this problem analytically. For the simplest case where both the OPEN and the SHORT STD are ideal, the three circuit parameters are found as follows :
K1 = A × Zls × R
Z
K2 =
K3 = −Ysm x R
sm
R
o
o
o
(5)
where
Ro : characteristic impedance A = (1 - Zlmi *Yom)/(Zlmi - Zsm) Yom : measured admittance for OPEN STD Zsm : measured impedance for SHORT STD
1) step 1
*Regard the impedance of the 50 LOAD as
Zls1 = 50+j*0 (that is, the phase of LOAD1 is zero).
*Find the circuit parameters K1, K2, and K3 by
normal OSL calibration using the LOAD value (Zls1).
*Execute correction for LOAD2 and get the
corrected impedance (Zcorr2).
*Calculate the phase difference (∆θ) between the
phase of Zcorr2 and the true phase of LOAD2
2) step 2 *Modify the impedance of LOAD1 to Zls1'
whose phase is -∆θ and whose impedance magnitude is still 50
*Calculate the circuit parameters again by
normal OSL calibration using modified LOAD impedance Zls1'.
10
Although this is an approximate method, actually just performing the two steps is accurate enough for our purpose. We call this method the "modified OSL calibration."
4.3 Phase measurement error using modified OSL calibration
We considered the following error factors during phase measurement in the modified OSL calibration:
1) uncertainty for impedance magnitude of LOAD1
For example, the relationship between the phase measurement error and the uncertainty for the impedance phase of LOAD2 (∆θls2) is shown in Fig. 4-1. Fig. 4-2 shows the relationship between the phase measurement error and the uncertainty for admittance magnitude of OPEN (∆Yopen).
From the above discussion the phase measurement error when using the modified OSL calibration is mainly determined by
* uncertainty for the impedance phase of LOAD2 * uncertainty for the admittance magnitude of
OPEN
2) impedance phase of LOAD1
3) impedance magnitude of LOAD2
4) uncertainty for impedance phase of LOAD2
5) uncertainty for admittance magnitude of OPEN
Notice that factor 2 and factor 3 do not cause any error if we use the analytical solution. By computer simulations the phase error is shown to behave as follows:
1) Sensitivity of phase measurement error due to the uncertainty of impedance magnitude for LOAD1 is small.
2) Sensitivity of phase measurement error due to the impedance phase of LOAD1 is small.
3) Sensitivity of phase measurement error due to the impedance magnitude of LOAD2 is small.
4) Uncertainty for the impedance phase of LOAD2 directly affects the phase measurement error.
5) Sensitivity of the phase measurement error due to the uncertainty for admittance magnitude of OPEN (| Yopen|) is reduced to |Ro*Yopen|*(Copen/Cls2) in the case of reactive DUT. In the case of resistive DUTs, the sensitivity is the same as in the normal OSL calibration.
where
Ro : characteristic impedance = 50 Copen : capacitance of OPEN STD Cls2 : capacitance of LOAD2 (low-loss
capacitor)
Now we evaluate these two items. The D factor for the capacitor (3pF) used in the calibration can be small because its dimensions are small and the space between the inner and outer conductors is filled almost completely with air. The D value is verified as 500E-6 at 1GHz from the residual resistance measurement at the series resonant frequency. The D factor is in proportion to Freq due to the skin effect. By using zero as the D value for the capacitor during calibration a phase measurement error of 500E-6 is realized at 1 GHz. On the other hand, the uncertainty for the OPEN capacitance is +/-5fF at most.
This leads to a phase measurement error less than +/-100E-6 at 1 GHz. In all, a phase measurement uncertainty of 500E-6 is realized by using the modified OSL calibration.
5. Impedance tracability
For an impedance performance check using the top-down method [1] we set up a kit traceable to the National Standards. This kit is calibrated annually at our standard laboratory. Two major items in the kit are the 50 LOAD and 10 cm long, 50 beadless air line. The 50 LOAD is desirable because its frequency characteristic for impedance is very flat. The structure of the air line is very simple. Therefore, it is easy to predict its frequency characteristic and it is convenient to realize various impedances by changing frequencies with the OPEN or SHORT terminated.
The traceability path for the kit is shown in Fig. 5. The impedance characteristic of OPEN ended and SHORT ended air lines can be calculated theoretically from their dimensions and resistivity.
11
1.5
[2] However, it is not easy to design the syst offring calibrated dimensions with the individual kit.
Freq. = 1GHz |Zopen| : impedance of OPEN STD @ 2 k Dqls2: uncertainty for impedance phase of LOAD2 (low-loss
capacitor) = 500E-6
Figure 4-1. Relationship between the phase measurement error and the uncertainty for the impedance phase of LOAD2
Therefore, the periodical calibration of dimensions is performed only to the reference air line of the standard laboratory. Calibration for the individual air line is executed by the network analyzer calibrated from the reference air line. The 50 LOAD calibration is done mainly by the quarter wave impedance method and DCR measurement. The OPEN termination is calibrated by a capacitance bridge at low frequencies and by the network analyzer at high frequencies. The SHORT termination is treated as an ideal one. Uncertainties for it consists of the skin effect and non-repeatabilities.
Freq. = 1GHz Ro: characteristic impedance = 50 |Zopen| : impedance of OPEN STD @ 2k Dq1 = |Ro*DYopen|*Copen/Cls2 |DYopen| : uncertainty for admittance magintude of OPEN = 30mS Copen = 82 fF, Cls2 = 3pF
Figure 4-2. Relationship between the phase measurement error
6. Test fixtures
In the actual measurement test fixtures corresponding to different shaped DUTs are needed. As the frequency range goes up higher, people need fixtures that are able to handle smaller devices. We have developed four types of fixtures:
* a fixture for surface mount devices (SMDs) with
bottom electrodes * a fixture for SMDs with side electrodes * a fixture for very small SMDs * a fixture for lead components
To reduce the error at the fixture terminal it is necessary to minimize the length from the reference APC-7 plane to the fixture terminal and to minimize the connection non-repeatability.
The new fixtures improved repeatability by almost five times compared with our old ones.The typical non-repeatability of the SMD fixtures is
12
* +/-50 pH, +/- 30 m for SHORT measurement * +/- 5 fF, +/- 2 µS for OPEN measurement
Furthermore, we installed "fixture compensations" in the firmware which corresponds to the corrections at the reference plane. This reduces the errors generated in the circuit between the reference plane and the fixture terminal. The best choice for the compensation is the OSL. However, it is not easy to prepare a LOAD having excellent frequency characteristics. As the more realistic way we prepare the following compensation functions:
* fixture port extension * OPEN-SHORT correction at the fixture plane
Fig. 6 shows the typical additional error when using the above corrections at the same time. These values are almost three times better than the errors for our former type of fixtures.
Figure 6. Typical additional errors after fixture compensations
7. Conclusion
Selection of "transducers" is important for accurate impedance measurement . A new type of transducers, which are based on the voltmeter/ ammeter method and have wide impedance measuring capability, are proposed. By adopting these transducers the New RF Impedance Analyzer has been developed. We also proposed a new phase calibration technique modified from the OSL calibration. It utilizes a low-loss capacitor as the second LOAD. This calibration enables accurate Q measurements.
References
[1] Hewlett Packard, "Traceability and the HP8510
Network Analyzer," Nov. 1 1985
[2] Robert E. Nelson, Marlene R. Coryell,
"Electrical Parameters of Precision, Coaxial, Air-Dielectric Transmission Lines," NBS Monograph96, June 30 1966
13
For more information about Hewlett­Packard test & measurement products, applications, services, and for a current sales office listing, visit our web site, http://www.hp.com/go/tmdir. You can also contact one of the following centers and ask for a test and measure­ment sales representative.
United States:
Hewlett-Packard Company Test and Measurement Call Center P.O. Box 4026 Englewood, CO 80155-4026 1 800 452 4844
Canada:
Hewlett-Packard Canada Ltd. 5150 Spectrum Way Mississauga, Ontario L4W 5G1 (905) 206 4725
Europe:
Hewlett-Packard European Marketing Centre P.O. Box 999 1180 AZ Amstelveen The Netherlands (31 20) 547 9900
Japan:
Hewlett-Packard Japan Ltd. Measurement Assistance Center 9-1, Takakura-Cho, Hachioji-Shi, Tokyo 192, Japan Tel: (81) 426 56 7832 Fax:(81) 426 56 7840
APC-7 ® is a U.S. registered trademark of the Bunker Ramo Corporation.
14
Latin America:
Hewlett-Packard Latin American Region Headquarters 5200 Blue Lagoon Drive 9th Floor Miami, Florida 33126 U.S.A. Tel: (305) 267-4245
(305) 267-4220
Fax:(305) 267-4288
Australia/New Zealand:
Hewlett-Packard Australia Ltd. 31-41 Joseph Street Blackburn, Victoria 3130 Australia Tel: 1 800 629 485 (Australia)
0800 738 378 (New Zealand)
Fax:(61 3) 9210 5489
Asia Pacific:
Hewlett-Packard Asia Pacific Ltd. 17-21/F Shell Tower, Times Square, 1 Matheson Street, Causeway Bay, Hong Kong Tel: (852) 2599 7777 Fax:(852) 2506 9285
Copyright © 1998 Hewlett-Packard Company Printed in U.S.A. 10/98 5966-2046E
Loading...