Agilent 4291B Quick Start Guide

HP 4291B RF Impedance/Material Analyzer
Quick Start Guide
SERIAL NUMBERS
This manual applies directly to instruments with serial number prex
JP1KE and above, or whose rmware is version 1.xx. For additional
important information about serial numbers, read \Serial Number" in
Appendix A of this manual.
HP Part No. 04291-90031
Notice
The information contained in this document is subject to change
without notice.
This document contains proprietary information that is protected by
copyright. All rights are reserved. No part of this document may be
photocopied, reproduced, or translated to another language without
the prior written consent of the Hewlett-Packard Company.
Hewlett-Packard Japan, LTD.
Kobe Instrument Division
1-3-2, Murotani, Nishi-ku, Kobe-shi,
Hyogo, 651-22 Japan
R
MS-DOS
APC-7
is a U.S. registered trademark of Microsoft Corporation.
R
is a U.S. registered trademark of Bunker Ramo Corporation.
c
Copyright 1997, 1998 Hewlett-Packard Japan, LTD.
Manual Printing
History
The manual printing date and part number indicate its current
edition. The printing date changes when a new edition is printed.
(Minor corrections and updates that are incorporated at reprint do not
cause the date to change.) The manual part number changes when
extensive technical changes are incorporated.
Safety Summary
Note
Note
Ground The
Instrument
December 1997
December 1998
:::::: ::::::: ::::::: :::::: ::::::: ::::::: ::
:::::: ::::::: :::::: ::::::: ::::::: ::::::
Second Edition
First Edition
The following general safety precautions must be observed during all
phases of operation, service, and repair of this instrument. Failure to
comply with these precautions or with specic
WARNINGS
elsewhere
in this manual may impair the protection provided by the equipment.
In addition it violates safety standards of design, manufacture, and
intended use of the instrument.
The Hewlett-Packard Company assumes no liability for the customer's
failure to comply with these requirements.
HP 4291B comply with INSTALLATION CATEGORY II and POLLUTION
DEGREE 2 in IEC1010-1. HP 4291B are INDOOR USE product.
LEDs in HP 4291B are Class 1 in accordance with IEC825-1.
CLASS 1 LED PRODUCT
To avoid electric shock hazard, the instrument chassis and cabinet
must be connected to a safety earth ground by the supplied power
cable with earth blade.
DO NOT Operate In An
Explosive Atmosphere
Keep Away From Live
Circuits
DO NOT Service Or
Adjust Alone
Do not operate the instrument in the presence of ammable gasses or
fumes. Operation of any electrical instrument in such an environment
constitutes a denite safety hazard.
Operating personnel must not remove instrument covers. Component
replacement and internal adjustments must be made by qualied
maintenance personnel. Do not replace components with the power
cable connected. Under certain conditions, dangerous voltages may
exist even with the power cable removed. To avoid injuries, always
disconnect power and discharge circuits before touching them.
Do not attempt internal service or adjustment unless another person,
capable of rendering rst aid and resuscitation, is present.
iii
DO NOT Substitute
Parts Or Modify
Instrument
Because of the danger of introducing additional hazards, do not
install substitute parts or perform unauthorized modications to the
instrument. Return the instrument to a Hewlett-Packard Sales and
Service Oce for service and repair to ensure that safety features are
maintained.
Dangerous Procedure
Warnings
Warning
Safety Symbols
Warnings
, such as the example below, precede potentially dangerous
procedures throughout this manual. Instructions contained in the
warnings must be followed.
Dangerous voltages, capable of causing death, are present in this
instrument. Use extreme caution when handling, testing, and
adjusting this instrument.
General denitions of safety symbols used on equipment or in manuals
are listed below.
Instruction manual symbol: the product is marked
with this symbol when it is necessary for the user to
refer to the instruction manual.
Alternating current.
Direct current.
On (Supply).
O (Supply).
In position of push-button switch.
Out position of push-button switch.
Frame (or chassis) terminal. A connection to the
frame (chassis) of the equipment which normally
include all exposed metal structures.
This
Warning
sign denotes a hazard. It calls attention
to a procedure, practice, condition or the like, which,
if not correctly performed or adhered to, could result
in injury or death to personnel.
This
Caution
sign denotes a hazard. It calls attention
to a procedure, practice, condition or the like, which,
if not correctly performed or adhered to, could result
in damage to or destruction of part or all of the
product.
This
Note
sigh denotes important information. It
calls attention to a procedure, practice, condition or
the like, which is essential to highlight.
Axed to product containing static sensitive devices
use anti-static handling procedures to prevent
electrostatic discharge damage to component.
iv
Typeface
Conventions
Bold
Italics
Boldface type is used when a term is dened.
For example:
icons
are symbols.
Italic type is used for emphasis and for titles
of manuals and other publications.
Italic type is also used for keyboard entries
when a name or a variable must be typed in
place of the words in italics.For example:
copy
lename
means to type the word
copy
to type a space, and then to type the name of
a le such as
file1
.
,
Computer
4
HARDKEYS
NNNNNNNNNNNNNNNNNNNNNNNNNN
SOFTKEYS
Computer font is used for on-screen prompts
and messages.
5
Labeled keys on the instrument front panel
are enclosed in45.
Softkeys located to the right of the LCD are
enclosed in
NNNN
N
.
v
Certication
Hewlett-Packard Company certies that this product met its
published specications at the time of shipment from the factory.
Hewlett-Packard further certies that its calibration measurements
are traceable to the United States National Institute of Standards and
Technology, to the extent allowed by the Institution's calibration
facility, or to the calibration facilities of other International Standards
Organization members.
Warranty
Limitation Of
Warranty
This Hewlett-Packard instrument product is warranted against defects
in material and workmanship for a period of one year from the date
of shipment, except that in the case of certain components listed
in
General Information
of this manual, the warranty shall be for
the specied period. During the warranty period, Hewlett-Packard
Company will, at its option, either repair or replace products that
prove to be defective.
For warranty service or repair, this product must be returned to a
service facility designated by HP. Buyer shall prepay shipping charges
to HP and HP shall pay shipping charges to return the product to
Buyer. However, Buyer shall pay all shipping charges, duties, and
taxes for products returned to HP from another country
.
HP warrants that its software and rmware designated by HP for use
with an instrument will execute its programming instruction when
property installed on that instrument. HP does not warrant that
the operation of the instrument, or software, or rmware will be
uninterrupted or error free.
The foregoing warranty shall not apply to defects resulting from
improper or inadequate maintenance by Buyer, Buyer-supplied
software or interfacing, unauthorized modication or misuse,
operation outside the environmental specications for the product, or
improper site preparation or maintenance.
No other warranty is expressed or implied. HP specically disclaims
the implied warranties of merchantability and tness for a
particular purpose.
vi
Exclusive
Remedies
The remedies provided herein are buyer's sole and exclusive
remedies. HP shall not be liable for any direct, indirect, special,
incidental, or consequential damages, whether based on contract, tort,
or any other legal theory.
Assistance
Product maintenance agreements and other customer assistance
agreements are available for Hewlett-Packard products.
For any assistance, contact your nearest Hewlett-Packard Sales and
Service Oce.Addresses are provided at the back of this manual.
vii

Contents

1. Introduction
Using This Manual .................. 1-1
Brief Description of the HP 4291B . . . . . . . . . . . 1-3
Additional Features . . . . . . . . . . . . . . . . . 1-3
2. Installation and Set Up Guide
Incoming Inspection ...... ...... ..... 2-1
Power Requirements ...... ...... ..... 2-5
Power Cable .. ...... ...... ...... 2-5
Replacing the Fuse . . . . . . . . . . . . . . . . 2-7
Fuse Selection .......... ...... ...
Procedure .....................
Operation Conditions . . . . . . . . . . . . . . . . .
Ventilation Requirements ...............
Instruction for Cleaning . . . . . . . . . . . . . . . .
Rack Mounting . . . . . . . . . . . . . . . . . . . .
Rack/Handle Installation ...... ...... ...
Option 1CN Handle Kit ...............
Installing the Handle .. ...... ...... .
Option 1CM Rack Mount Kit . . . . . . . . . . . . .
Mounting the Rack ........ ...... ..
Option 1CP Rack Mount & Handle Kit ........ 2-10
Mounting the Handle and Rack ..........
Connecting the Test Station . . . . . . . . . . . . . .
Connecting the Test Head . . . . . . . . . . . . . . .
Selecting the Test Head . . . . . . . . . . . . . . . 2-12
Connecting the Test Head to the Test Station . . . . . 2-12
Connecting a BNC Adapter (Option 1D5 Only) ..... 2-13
Connecting a Keyboard . . . . . . . . . . . . . . . . 2-13
2-7
2-7
2-8
2-8
2-8
2-8
2-9
2-9
2-9
2-9
2-9
2-10
2-11
2-12
3. Impedance Measurement Quick Start
Overview .. ...... ...... ..... ...
Basic Measurement Flow ..............
Required Equipment ................
1. Power ON . . . . . . . . . . . . . . . . . . . . .
2. Calibration ....................
Calibration Procedure . . . . . . . . . . . . . . . .
3. Connecting the Test Fixture ............ 3-8
Selecting a Test Fixture . . . . . . . . . . . . . . . 3-8
Connecting the Test Fixture to the Test Head . . . . . 3-9
4. Setting the Electrical Length of the T
est Fixture . . . 3-10
5. Fixture Compensation ...............
Performing SHORT Compensation ..........
Selecting the Shorting Device ........ ...
SHORT Compensation Key Sequence . . . . . . . .
3-1
3-2
3-3
3-4
3-4
3-5
3-11
3-12
3-12
3-14
Contents-1
Performing OPEN Compensation . . . . . . . . . . . 3-14
OPEN Compensation Key Sequence .. ...... 3-14
6. Connecting the DUT to the Test Fixture . . . . . . . 3-15
7. Setting up the Analyzer .......... .... 3-15
7-1. Setting up forjZj-vs. Frequency . . . . . . . . 3-16
7-2. Setting up for C - D vs. Frequency ....... 3-17
7-3. Setting up for C - D vs. OSC Level ....... 3-18
7-4. Setting up for C - D vs. dc-V (Option 001 only) . . 3-19
Display Annotations ................ 3-20
8. Measuring the DUT ................ 3-21
Performing Automatic Scaling . . . . . . . . . . . . 3-21
9. Analyzing a Measurement Result .. ...... .. 3-21
Using the Marker . . . . . . . . . . . . . . . . . . 3-21
Reading a Measured Value . . . . . . . . . . . . . 3-21
Moving the Marker to the Minimum Point .... . 3-22
Clearing the Marker .. ...... ...... . 3-23
Equivalent Circuit Analysis ............. 3-23
Approximation . . . . . . . . . . . . . . . . . . . 3-23
Simulation . . . . . . . . . . . . . . . . . . . . . 3-24
Clearing the Approximated Constant Data ..... 3-24
Clearing the Simulated Data . . . . . . . . . . . .
10. Measuring Other DUT . . . . . . . . . . . . . . .
For More Information . . . . . . . . . . . . . . . . .
4. Dielectric Material Measurement Quick Start
Overview .. ...... ...... ...... ..
Brief Measurement Theory .............
Basic Measurement Flow ..............
Required Equipment ................
1. Power ON . . . . . . . . . . . . . . . . . . . . .
2. Calibration ....................
Calibration Procedure . . . . . . . . . . . . . . . .
3. Connecting the Test Fixture ...... ..... .
Connecting the Test Fixture to the Test Head . . . . . 4-9
Selecting the Test Fixture . . . . . . . . . . . . . .
4. Input Load Thickness .. ...... ...... . 4-11
5. Fixture Compensation ............... 4-12
Performing SHORT Compensation .......... 4-12
SHORT Compensation Key Sequence . . . . . . . . 4-12
Performing OPEN Compensation . . . . . . . . . . . 4-13
OPEN Compensation Key Sequence .. ......
Performing LOAD Compensation . . . . . . . . . . .
LOAD Compensation Key Sequence ........
6. Input the MUT Thickness . . . . . . . . . . . . . .
7. Placing the MUT in the Test Fixture . . . . . . . . . 4-15
8. Setting up the Analyzer .. ...... ...... 4-16
8-1. Setting up for
0
"
- tanvs. Frequency ..... 4-17
r
Display Annotations ................ 4-18
8-2. Setting up for a Cole-Cole Plot ...... ... 4-19
Display Annotations ................ 4-20
9. Measuring the MUT .. ...... ...... ..
Performing Automatic Scaling . . . . . . . . . . . .
Using the Marker . . . . . . . . . . . . . . . . . .
Marker for Frequency/OSC Level/dc bias
Characteristics ................
3-25
3-25
3-25
4-1
4-2
4-3
4-4
4-5
4-5
4-6
4-9
4-10
4-13
4-14
4-14
4-15
4-21
4-21
4-21
4-21
Contents-2
Marker for a Cole-Cole Plot .... ...... .. 4-22
10. Measuring Other MUTs .............. 4-22
For More Information . . . . . . . . . . . . . . . . . 4-22
5. Magnetic Material Measurement Quick Start
Overview .. ...... ...... ..... ... 5-1
Brief Measurement Theory ............. 5-2
Basic Measurement Flow .............. 5-3
Required Equipment ................ 5-4
1. Power ON . . . . . . . . . . . . . . . . . . . . . 5-5
2. Calibration .................... 5-5
Calibration Procedure . . . . . . . . . . . . . . . . 5-6
3. Connecting the Test Fixture ............ 5-9
Selecting Fixture and Holder ............ 5-9
Connecting the Test Fixture to the Test Head . . . . . 5-9
Selecting the Test Fixture . . . . . . . . . . . . . . 5-11
4. Fixture Compensation ............... 5-11
Performing SHORT Compensation .......... 5-11
SHORT Compensation Key Sequence . . . . . . . . 5-12
5. Input the MUT Size .... ...... ...... 5-12
6. Placing the MUT on the Test Fixture ........ 5-13
7. Setting up the Analyzer .. ...... ......
0
7-1. Setting up for
7-2. Setting up for
7-3. Setting up for
00
-
r
0
r
0
r
vs. Frequency . . . . . . 5-15
r
00
-
vs. OSC Level ...... 5-16
r
00
-
vs. Dc-I (Option 001 only) 5-17
r
Display Annotations .. ...... ...... ..
8. Measuring the MUT .. ...... ...... ..
Performing Automatic Scaling . . . . . . . . . . . .
9. Analyzing a Measurement Result .. ...... ..
Using the Marker . . . . . . . . . . . . . . . . . .
Reading a Measured Value . . . . . . . . . . . . .
Moving the Marker to the Maximum Point ..... 5-20
Clearing the Marker .. ...... ...... .
10. Measuring Other MUTs ..............
For More Information . . . . . . . . . . . . . . . . .
5-14
5-18
5-19
5-19
5-19
5-19
5-19
5-21
5-21
5-21
6. Features Common to All Modes
Performing User Dened Calibration . . . . . . . . . . 6-2
User Dened Calibration Description . . . . . . . . . 6-2
Performing User Dened Fixture Compensation .... 6-3
User Dened Fixture Compensation Description ... 6-3
Performing Averaging . . . . . . . . . . . . . . . . .
Performing Point Averaging . . . . . . . . . . . . .
Performing Sweep Averaging ............ 6-4
Averaging Description ............... 6-4
Using Point Delay and Sweep Delay ...... .... 6-5
Making a Point Delay Measurement ......... 6-5
Making a Sweep Delay Measurement . . . . . . . . . 6-5
Delay Description . . . . . . . . . . . . . . . . . . 6-5
Changing the Number of Measurement Points ..... 6-6
Number of Measurement Points Description ..... 6-6
Applying Dc Bias (Option 001 Only) ..........
Monitoring the OSC Level or the Dc Bias Level . . . . .
Performing a GO/NO-GO Test ...... ...... .
Editing a Limit Line Table . . . . . . . . . . . . . .
6-3
6-3
6-6
6-7
6-8
6-8
Contents-3
Executing a Limit Line Test............. 6-10
Performing a List Sweep Measurement . . . . . . . . . 6-10
Editing a Sweep List ...... ...... .... 6-10
Activating a List Sweep . . . . . . . . . . . . . . . 6-11
Constant Current/Voltage Level Measurement ..... 6-11
Auto Level Control Measurement for OSC Level Sweep . 6-12
Placing Characters on the Screen ........... 6-13
Placing Letters on the Screen ............ 6-14
Editing the Label and Changing the Position or Color . 6-14
Splitting the Measurement Display . . . . . . . . . . . 6-15
Performing Independent Sweeps for Each Channel . . 6-15
Using the Marker . . . . . . . . . . . . . . . . . . . 6-16
Reading a Value Using the Marker . . . . . . . . . . 6-16
Reading Values by Using the Sub-markers ...... 6-16
Searching for a Specic Point ............ 6-17
Searching for Maximum/Minimum Points . . . . . . 6-17
Searching for Points that Have the Target Value .. 6-17
Storing Your Trace to the Memory Traces ....... 6-17
Storing the Data Trace to the Memory Trace . . . . . 6-17
Storing More Traces to Memory Traces . . . . . . . 6-17
Displaying Memory Traces ...... ...... .
Displaying a Memory Trace Only When It Is Selected.
Clearing the Memory Traces . . . . . . . . . . . . .
Using the Trace Math Function ............
6-18
6-18
6-18
6-19
Calculating Using the Data Trace and Memory Trace . 6-19
Turning O the Data Math Function . . . . . . . . .
Saving and Recalling .. ...... ...... ...
6-19
6-19
Saving an Analyzer Setting and Measurement Trace . 6-20
Recalling a Saved Analyzer Setting and Measurement
Trace . . . . . . . . . . . . . . . . . . . . . .
6-20
Saving a Display Image to a TIFF File ........ 6-20
Saving Measured Data for a Spreadsheet . . . . . . .
Purging a File ...... ...... ...... .
6-21
6-21
Initializing a Disk/Memory Disk for Use .. ..... 6-21
Printing .......................
6-21
Printing a Display Image ........ ..... . 6-21
Logging the Key Sequence into a Program ....... 6-22
Measurement Techniques ........ ...... . 6-23
Smoothing the Trace ................ 6-23
Reducing Trace Noise . . . . . . . . . . . . . . . . 6-23
Increasing the Sweep Speed . . . . . . . . . . . . .
6-23
When the Measurement Result Does Not Seem Correct 6-23
Contents-4
A. Manual Changes
Introduction .....................
Manual Changes ...................
Serial Number ....................
A-1
A-1
A-2
B. Maintenance
Performance Verication ............... B-1
Repair . . . . . . . . . . . . . . . . . . . . . . . . B-1
Possible Problems and Their Solution ........ B-1
Replacement of Center Conductor Collet . . . . . . . . B-2
C. Fixture Compensation Procedures for the
HP 16191A and HP 16193A
HP 16191A . . . . . . . . . . . . . . . . . . . . . . C-1
SHORT Compensation . . . . . . . . . . . . . . . . C-1
SHORT Compensation Key Sequence . . . . . . . . C-2
OPEN Compensation .. ...... ...... .. C-3
OPEN Compensation Key Sequence .. ...... C-3
Connecting DUT .................. C-4
HP 16193A . . . . . . . . . . . . . . . . . . . . . . C-5
SHORT Compensation . . . . . . . . . . . . . . . . C-5
SHORT Compensation Key Sequence . . . . . . . . C-6
OPEN Compensation .. ...... ...... .. C-7
OPEN Compensation Key Sequence .. ...... C-7
Connecting DUT .................. C-7
D. Fixture Compensation Procedures for the HP 16194A
Exchanging the device holder . . . . . . . . . . . . .
Connecting the Test Fixture . . . . . . . . . . . . . .
Setting . . . . . . . . . . . . . . . . . . . . . . . .
SHORT compensation . . . . . . . . . . . . . . . . .
Short compensantion key syquence .........
Open compensation . . . . . . . . . . . . . . . . . .
Open compensation Key syquence . . . . . . . . . .
Load compensation . . . . . . . . . . . . . . . . . .
Load compensation Key sequence . . . . . . . . . .
DUT.........................
D-1
D-2
D-3
D-4
D-4
D-5
D-5
D-6
D-6
D-7
E. Calibration and Compensation
Calibration . . . . . . . . . . . . . . . . . . . . . .
Compensation .................... E-1
OPEN/SHORT Fixture Compensation and Electrical
Length Compensation ...... ...... .. E-1
OPEN/SHORT/LOAD Fixture Compensation . . . . . . E-2
Index
E-1
Contents-5
Figures
1-1. Manual Reference Flow . . . . . . . . . . . . . . . 1-1
2-1. Contents . . . . . . . . . . . . . . . . . . . . . . 2-4
2-2. HP Part Numbers of Power Cables . . . . . . . . . . 2-6
2-3. Rack Mount Kits Installation ............ 2-9
2-4. Connecting the Test Station to the Mainframe .... 2-11
2-5. Connecting the Test Head to the Test Station . . . . . 2-12
2-6. Connecting a BNC Adapter ............. 2-13
2-7. Connecting a Keyboard . . . . . . . . . . . . . . . 2-14
3-1. Basic Flow for Impedance Measurements . . . . . . . 3-2
3-2. Required Equipment ................ 3-3
3-3. Calibration . . . . . . . . . . . . . . . . . . . . . 3-4
3-4. ...... ...... ..... ...... ...
3-5. ...... ...... ..... ...... ...
3-6. ...... ...... ..... ...... ...
3-7. ...... ...... ..... ...... ...
3-8. ...... ...... ..... ...... ...
3-9. ...... ...... ..... ...... ...
3-10. ...... ...... ..... ...... ...
3-11. ...... ...... ..... ...... ...
3-12. ...... ...... ..... ...... ...
3-13. ...... ...... ..... ...... ...
3-5
3-5
3-5
3-6
3-6
3-6
3-7
3-7
3-9
3-9
3-14. Connecting the Test Fixtures (HP 16192A) .. .... 3-10
3-15. Fixture Compensation ...............
4-1. Dielectric Material Measurement . . . . . . . . . . .
4-2. Basic Flow for Dielectric Material Measurements . . .
4-3. Required Equipment ................
3-11
4-2
4-3
4-4
4-4. Calibration . . . . . . . . . . . . . . . . . . . . . 4-5
4-5. ...... ...... ..... ...... ... 4-6
4-6. ...... ...... ..... ...... ... 4-6
4-7. ...... ...... ..... ...... ... 4-6
4-8. ...... ...... ..... ...... ... 4-7
4-9. ...... ...... ..... ...... ...
4-10. ...... ...... ..... ...... ...
4-11. ...... ...... ..... ...... ...
4-12. ...... ...... ..... ...... ...
4-13. ...... ...... ..... ...... ...
4-7
4-7
4-8
4-9
4-9
4-14. Connecting the Test Fixtures ............ 4-10
4-15. Complex Plane Annotation ............. 4-20
5-1. Magnetic Material Measurement . . . . . . . . . . . 5-2
5-2. Basic Flow for Magnetic Material Measurements ... 5-3
5-3. Required Equipment ................ 5-4
5-4. Calibration . . . . . . . . . . . . . . . . . . . . . 5-5
5-5. ...... ...... ..... ...... ...
5-6. ...... ...... ..... ...... ...
5-7. ...... ...... ..... ...... ...
5-8. ...... ...... ..... ...... ...
5-6
5-6
5-6
5-7
Contents-6
5-9. .......................... 5-7
5-10. .......................... 5-7
5-11. .......................... 5-8
5-12. Dimensions of the MUT Holder ........ ... 5-9
5-13. .......................... 5-10
5-14. .......................... 5-10
5-15. Connecting the Test Fixtures (HP 16454A Small) ... 5-10
5-16. Dimensions of the MUT . . . . . . . . . . . . . . . 5-13
6-1. Point Averaging and Sweep Averaging . . . . . . . . 6-4
6-2. Point Delay and Sweep Delay ............ 6-5
6-3. Number of Points . . . . . . . . . . . . . . . . . . 6-6
6-4. Level Monitor Function . . . . . . . . . . . . . . . 6-8
6-5. Limit Line Example . . . . . . . . . . . . . . . . . 6-8
6-6. List Sweep Example ...... ...... .... 6-10
6-7. Label Function . . . . . . . . . . . . . . . . . . . 6-13
6-8. Split Display .................... 6-15
6-9. Uncoupled Channels ................ 6-15
6-10. Marker List .................... 6-16
A-1. Serial Number Plate .... ...... ...... A-2
D-1. Exchanging the Device Holder . . . . . . . . . . . . D-1
D-2. Connecting the HP 16194A to the Measurement
Terminal .. ...... ...... ......
D-3. ..........................
D-4. ..........................
D-5. Open compensation . . . . . . . . . . . . . . . . .
D-6. Load compensation . . . . . . . . . . . . . . . . .
D-7. DUT . . . . . . . . . . . . . . . . . . . . . . . .
E-1. OPEN/SHORT Fixture Compensation . . . . . . . . .
D-2
D-3
D-4
D-5
D-6
D-7
E-2
Contents-7
Tables
2-1. Contents . . . . . . . . . . . . . . . . . . . . . . 2-2
2-2. Fuse Selection .... ...... ...... ... 2-7
2-3. Rack Mount Kits ...... ...... ...... 2-8
3-1. Test Fixtures Specication (1/2) ........... 3-8
3-2. Dimension of Shorting Devices .... ...... . 3-12
3-3. MEASUREMENT Block Parameter Setting ...... 3-16
3-4. STIMULUS Block Parameter Setting ......... 3-16
3-5. MEASUREMENT Block Parameter Setting ...... 3-17
3-6. STIMULUS Block Parameter Setting ......... 3-17
3-7. MEASUREMENT Block Parameter Setting ...... 3-18
3-8. STIMULUS Block Parameter Setting ......... 3-18
3-9. MEASUREMENT Block Parameter Setting ...... 3-19
3-10. STIMULUS Block Parameter Setting ......... 3-19
4-1. MEASUREMENT Block Parameter Setting ...... 4-17
4-2. STIMULUS Block Parameter Setting .........
4-17
4-3. MEASUREMENT Block Parameter Setting ...... 4-19
4-4. STIMULUS Block Parameter Setting .........
5-1. MUT Size For Test Fixtures .............
4-19
5-9
5-2. MEASUREMENT Block Parameter Setting ...... 5-15
5-3. STIMULUS Block Parameter Setting .........
5-15
5-4. MEASUREMENT Block Parameter Setting ...... 5-16
5-5. STIMULUS Block Parameter Setting ......... 5-16
5-6. MEASUREMENT Block Parameter Setting ...... 5-17
5-7. STIMULUS Block Parameter Setting .........
A-1. Manual Changes by Serial Number . . . . . . . . . .
5-17
A-1
A-2. Manual Changes by Firmware Version ...... .. A-1
Contents-8
Introduction
1
This Quick start Guide walks you through system setup and
initial power-on, shows you how to make basic measurements,
and explains commonly used features. Normally, you will use
this guide when you are rst learning to use the HP 4291B
RF Impedance/Material Analyzer.
Using This Manual
The organization of this Quick start Guide is described below.
Figure 1-1 shows how to use the manual. Note that chapters 3, 4, and
5 are independent of each other.You need only read the chapter that
pertains to your measurement (or you can read all three).
Figure 1-1. Manual Reference Flow
Introduction 1-1
Chapter 1. Introduction
This chapter provides a brief description of the HP 4291B and
how to use this manual.
Chapter 2. Installation and Set Up Guide
This chapter provides the information necessary for performing
an incoming inspection and setting up your analyzer.You should
refer to this chapter if the analyzer has not been inspected and
set up for operation.
Chapter 3. Impedance Measurement Quick Start
This chapter provides a quick start guide for impedance
measurement.
Chapter 4. Dielectric Material Measurement Quick Start
This chapter provides a quick start guide for dielectric material
measurement. Dielectric material measurement is available when
Option 002 is installed.
Chapter 5. Magnetic Material Measurement Quick Start
This chapter provides a quick start guide for magnetic material
measurement. Magnetic material measurement is available when
Option 002 is installed.
Chapter 6. Features Common to All Modes
This chapter provides the procedures for the analyzer key
features and measuremnent technique that can be used in all
modes of operation (impedance, dielectric, and magnetic).
Appendix A. Maintenance
This appendix provides information on performance verication,
repair, line voltage setting changes, and fuse replacement.
Appendix B. Compensation Procedure for HP 16191A and HP
16193A
This appendix provides the compensation procedure for the
HP 16191A and HP 16193A. This appendix is referred in \5.
Fixture Compensation" in Chapter 3.
Appendix C. Calibration and Compensation
This appendix provides the explanation of Calibration and
Compensation.
1-2 Introduction
Brief Description
of the HP 4291B
For surface-mount component evaluation and material testing, the
HP 4291B RF Impedance/Material Analyzer is an integrated package
designed to provide accurate testing at higher frequencies.
For component manufacturers, RF and digital equipment designers,
and material researchers, the HP 4291B provides these capabilities:
Broad frequency coverage from 1 MHz to 1.8 GHz for testing RF
components and materials.
Improved measurement accuracy and repeatability over an
impedance range of 0.1 to 50 k
Surface-mount-device (SMD) test xtures for dierent sizes of chip
capacitors and inductors.
Dielectric test xture and built-in function for measuring
permittivity, including Cole-Cole plot relaxation time (with Option
002 and HP 16453A).
Magnetic test xture and built-in function for measuring
permeability of ferrite material (with Option 002 and HP 16454A).
Direct impedance and material parameter measurement versus time,
humidity, or temperature (with an external temperature chamber).
Additional Features
Impedance Measurement
Evaluate components with dc bias up to6100 mA and640 V
(with Option 001). (See \Applying Dc Bias (Option 001 Only)" in
Chapter 6.)
Monitor test signals applied to your DUTs. (See \Monitoring the
OSC Level or the Dc Bias Level" in Chapter 6.)
Simulate a component with equivalent circuit analysis. (See
\Equivalent Circuit Analysis" in Chapter 3.)
Others
Store the measurement data and analyzer's settings on the DOS
and LIF compatible 1.4 MB oppy disk drive and memory disk.
(See \Saving and Recalling" in Chapter 6.)
Markers and marker utilities (See \Using the Marker" in
Chapter 6.)
Limit lines for go/no-go testing (See \Performing a GO/NO-GO
Test" in Chapter 6.)
Frequency linear/log/list sweep (See \Performing a List Sweep
Measurement" in Chapter 6 for list sweep.)
Introduction 1-3
Installation and Set Up Guide
This chapter provides installation and setup instructions. It contains
the following information:
Incoming Inspection
Power Requirements
Replacing the Fuse
Operation Conditions
Ventilation Requirements
Instruction for Cleaning
Rack/Handle Installation
Connecting the Test Station
Connecting the Test Head
Connecting a BNC Adapter
Connecting a Keyboard
2
Incoming
Inspection
Warning
To avoid hazardous electrical shock, do not turn on the analyzer
when there are signs of shipping damage to any portion of the
outer enclosure (for example, covers, panel, or display)
Inspect the shipping container for damage. If the shipping container
or cushioning material is damaged, it should be kept until the
contents of the shipment have been checked for completeness and
the analyzer has been checked mechanically and electrically
. The
contents of the shipment should be as listed in Table 2-1. If the
contents are incomplete, if there is mechanical damage or defect, or if
the analyzer does not pass the power-on selftests, notify the nearest
Hewlett-Packard oce. If the shipping container is damaged, or the
cushioning material shows signs of unusual stress, notify the carrier
as well as the Hewlett-Packard oce. Keep the shipping materials for
the carrier's inspection.
Installation and Set Up Guide 2-1
Table 2-1. Contents
Description HP Part Number
RF Impedance/Material Analyzer HP 4291B
Mainframe
Test Station
High Impedance Test Head
Documents
Operating Manual
Quick Start Guide
Programming Manual
2
3
3
1
04291-90020
04291-90021
3
04291-90027
HP Instrument BASIC Users
Handbook
3
E2083-90005
Calibration Kit
0STermination 04191-85302
0Termination 04191-85300
50 Termination 04291-65006
LOW-LOSS Capacitor 04291-60042
Carrying Case
3
04291-60041
Sample Program Disk 04291-18000
mini-DIN Keyboard C3757-60401
Power Cable See \Power Cable"
Option 012 only
Low Impedance Test Head
Option 013 only
High Temperature High-impedance
Test Head
Temperature Coecient
Measurement Program Disk
Fixture Stand
4
Pad
4
4
04291-18001
04291-60121
04291-09001
2-2 Installation and Set Up Guide
1
Delete when Option 011 is ordered.
2
Delete when Option 0B0 is ordered
3
This item is not shown in Figure 2-1 .
4
See the temperature coecient measurement chapter of
Operation Manual
.
Option 014 only
High Temperature Low-impedance
Test Head
Table 2-1. Contents (continued)
Description HP Part Number
Temperature Coecient
Measurement Program Disk
Fixture Stand
1
Pad
Option 0BW only
Service Manual
Option 1D5 only
BNC Adapter 1250-1859
1
See the temperature coecient measurement chapter of
2
This item is not shown in Figure 2-1 .
1
2
1
04291-18001
04291-60121
04291-09001
04291-90111
Operation Manual
.
Installation and Set Up Guide 2-3
2-4 Installation and Set Up Guide
Figure 2-1. Contents
Power
Requirements
The HP 4291B requires the following power source:
Voltage : 90 to 132 Vac, 198 to 264 Vac
Frequency : 47 to 63 Hz
Power : 300 VA maximum
Power Cable
Warning
In accordance with international safety standards, this instrument
is equipped with a three-wire power cable. When connected to an
appropriate ac power outlet, this cable grounds the instrument
frame.
The type of power cable shipped with each instrument depends on
the country of destination. Refer to Figure 2-2 for the part numbers
of the power cables available.
For protection from electrical shock, the power cable ground
must not be defeated.
The power plug must be plugged into an outlet that provides a
protective earth ground connection.
Installation and Set Up Guide 2-5
2-6 Installation and Set Up Guide
Figure 2-2. HP Part Numbers of Power Cables
Replacing the
Fuse
Fuse Selection
Procedure
Select proper fuse according to the Table 2-2.
Table 2-2. Fuse Selection
Fuse Rating/Type Fuse Part Number
5A 250Vac
UL/CSA type 2110-0030
Time Delay
For ordering the fuse,contact your nearest Hewlett-Packard Sales and
Service Oce.
Perform the following steps to exchange the fuse:
Caution
Lever a small minus screwdriver to
dismount the fuse holder above the AC
line receptacle on the rear panel.
Use the proper fuse for the line voltage selected. Use only fuses with
the required current rating and of the specied type as replacements.
DO NOT use a mended fuse or short-circuit the fuse-holder in order to
by-pass a blown fuse. Find out what caused the fuse to blow!
To check or replace the fuse, pull the fuse
holder and remove the fuse.To reinstall
the fuse, insert a fuse with the proper
rating into the fuse holder.
Installation and Set Up Guide 2-7
Operation
Conditions
The HP 4291B must be operated under within the following
environment conditions, and sucient space must be kept behind the
HP 4291B to avoid obstructing the air ow of the cooling fans.
Temperature
Disk drive non-operating condition
Disk drive operating condition
:::::: ::::::: :::::: :
::::::: :::::: ::::::: :::
0Cto40C
10Cto40C
Humidity
@wet bulb temperature<29C, without condensation
Disk drive non-operating condition
Disk drive operating condition
Warm up time
::::::: :::::: ::::::: ::::::: :::::: ::::::: ::::
:::::: ::::::
::::::: :::::: :::
15%to95%RH
15%to80%RH
30 minutes
Note
Ventilation
Requirements
Instruction for
Cleaning
Rack Mounting
The HP 4291B must be protected from temperature extremes which
could cause condensation within the instrument.
To ensure adequate ventilation, make sure that there is adequate
clearance of at least 180mm behind and 60mm sides
.
To prevent electrical shock, disconnect the HP 4291B power cable
from the receptacle before cleaning. Wipe with a dry cloth or a soft
cloth that is soaked with water and wrung tightly without undue
pressure to clean the casing. Do not attempt to clean the HP 4291B
internally.
Rack mounting information is provided with the rack mount kit.
If the kit was not ordered with the analyzer as an option, it may
be ordered through the nearest Hewlett-Packard oce. The part
numbers of the rack mount kit are shown in Table 2-3.
2-8 Installation and Set Up Guide
Table 2-3. Rack Mount Kits
Option Description HP Part Number
1CN Handle Kit 5063-9229
1CM Rack Mount Kit 5063-9216
1CP Rack Mount & Handle Kit 5063-9223
Rack/Handle
Installation
The HP 4291B can be rack mounted and used as a component in a
measurement system. Figure 2-3 shows how to rack mount the
HP 4291B.
Option 1CN Handle Kit
Option 1CM Rack
Mount Kit
Figure 2-3. Rack Mount Kits Installation
Option 1CN is a handle kit containing a pair of handles and the
necessary hardware to attach them to the instrument.
Installing the Handle
1. Remove the adhesive-backed trim strips1
front sides of the HP 4291B.
2. Attach the front handles3
3. Attach the trim strips4
Option 1CM is a rack mount kit containing a pair of anges and the
necessary hardware to mount them to the instrument in an equipment
rack with 482.6 mm (19 inches) horizontal spacing.
Mounting the Rack
1. Remove the adhesive-backed trim strips1
front sides of the HP 4291B.
2. Attach the rack mount ange2
the HP 4291B using the screws provided.
to the sides using the screws provided.
to the handles.
to the left and right front sides of
from the left and right
from the left and right
Installation and Set Up Guide 2-9
3. Remove all four feet (lift bar on the inner side of the foot, and
slide the foot toward the bar).
Option 1CP Rack
Mount & Handle Kit
Option 1CP is a rack mount kit containing a pair of anges and the
necessary hardware to mount them to an instrument which has
handles attached, in an equipment rack with 482.6 mm (19 inches)
spacing.
Mounting the Handle and Rack
1. Remove the adhesive-backed trim strips1
front sides of the HP 4291B.
2. Attach the front handle3
on the left and right front sides of the HP 4291B using the screws
provided.
3. Remove all four feet (lift bar on the inner side of the foot, and
slide the foot toward the bar).
and the rack mount ange5
from the left and right
together
2-10 Installation and Set Up Guide
Loading...
+ 128 hidden pages