Agilent 4288A Data Sheet

Agilent 4288A Capacitance Meter 1 kHz/1 MHz
Data Sheet
This document contains specifications and supplemental information about Agilent’s 4288A capacitance meter.
2
All specifications apply over a 0°C to 45°C range and 10 minutes after the instrument has been turned on, unless otherwise stated.
Specification (spec.): Warranted performance.
Specifications include guardbands to account for the expected statistical performance distribution, measurement uncertainties, and changes in performance due to environomental conditions.
Supplemental information is intended to provide useful information about the instrument, but is not covered by the product warranty. The information is denoted as typical, or nominal.
Typical (typ.): Expected performance of an
average unit which does not incude guardbands.
Nominal (nom.): A general, descriptive term
that does not imply a level of performance.
Basic specifications
Measurement parameters
• Cp-D, Cp-Q, Cp-Rp, Cp-G
• Cs-D, Cs-Q, Cs-Rs
Where,
Cp: Capacitance value measured using the parallel
equivalent circuit model
Cs: Capacitance value measured using the series
equivalent circuit model
D: Dissipation factor Q: Quality factor (inverse of D) G: Equivalent parallel conductance measured
using the parallel equivalent circuit model
Rp: Equivalent parallel resistance measured using
the parallel equivalent circuit model
Rs: Equivalent series resistance measured using the
series equivalent circuit model
Measurement signals
Table 1. Specifications of measurement signals
1 kHz
1 MHz
Frequency Allowable frequencies 0.99 MHz (1 MHz -1%)
1.01 MHz (1 MHz +1%)
1.02 MHz (1 MHz +2%)
Accuracy ± 0.02%
Range 0.1 V - 1.0 V
Level Resolution 0.1 V
Accuracy ± 5%
Measurement range:
20Ω (nominal)
Output Frequency: 100 pF - 100 nF
impedance 1 kHz Measurement range:
1(nominal)
220 nF - 10µ F
Frequency: 1 MHz 20Ω (nominal)
Measurement cable length
0m, 1m, 2m
Measurement time modes
Short mode, long mode
For information on the measurement time in each mode, refer to “Measurement time” on page 7.
Measurement range selection
Auto (auto range mode), manual (hold range mode)
Measurement range
Table 2. 1 kHz and 1 MHz measurement range
Measurement Measurement
signal frequency: 1 kHz signal frequency: 1 MHz
100 pF 220 pF 470 pF 1 pF 2.2 pF 4.7 pF
1 nF 2.2 nF 4.7 nF 10 pF 22 pF 47 pF
10 nF 22 nF 47 nF 100 pF 220 pF 470 pF
100 nF 220 nF 470 nF 1 nF
1 µF 2.2 µF 4.7 µF
10 µF
For information on measurable range in each measurement mode, refer to “Available measurement ranges” on this page and Tables 6 and 7 on page 4.
Averaging
Table 3. Averaging range and resolution
Range 1 - 256 measurements
Resolution 1
Trigger mode
Internal trigger (Int), manual trigger (Man), external trigger (Ext), GPIB trigger (Bus)
Trigger delay time
Table 4. Range and resolution of trigger delay time
Range 0 - 1.000 s
Resolution 1 ms
3
Measurement display ranges
Table 5 shows the range of the measured value that can be displayed on the screen.
Table 5. Allowable measured value display range
Parameter Measurement display range
Cs, Cp
- 99.9999 µF to - 0.00001 pF, 0,
0.00001 pF to 99.9999 µF
D
- 9.99999 to - 0.00001, 0,
0.00001 to 9.99999
Q
- 99999.9 to - 0.1, 0,
0.1 to 99999.9
Rs, Rp
- 999.999 Mto - 0.001 m, 0,
0.001 mto 999.999 M
G
- 9.99999 kS to - 0.00001 µS, 0,
0.00001 µS to 9.99999 kS
%
- 999.999% to - 0.001%, 0,
0.001% to 999.999%
Available measurement ranges
Table 6 and 7 on page 4 show recommended measurement ranges (recommendation for accurate measurement) and significant measurement ranges (range that does not cause overload) for each measurement range under the condition D (dissipation factor) 0.5.
Accuracy of Cp, Cs, D, G, and Rs
Tables 10 and 11 on page 5, show the measurement accuracy of Cp, Cs, D, G, and Rs when D 0.1. Note the following when you calculate accuracy with these tables.
• The equation to calculate the accuracy varies depending on the measurement time mode. In tables 10 and 11 on page 5, the upper equation is for the short mode; the lower equation corresponds to the long mode.
• Interpret your calculated accuracy as ± (percentage of the measured value of an error) for Cp and Cs, ± (absolute value of an error) for D, G, and Rs.
When 0.1 < D 0.5, multiply the accuracy obtained from Table 10 or Table 11 by the coefficient in Table 8 below.
Table 8. Coefficient caused by D value @ > 0.1, ≤ 0.5
Parameter Coefficient
Cp, Cs, G, Rs*
1
1 + D
2
D 1 + D
Accuracy of Q
The following equation is used to calculate the accuracy of Q from the accuracy of D.
Equation 1. Equation to calculate Q
Qe =
±Qx
2
x
De
1 Qx xDe
Where, Qe is the accuracy of Q, De is the accuracy of D, and Qx is the measured Q value.
Accuracy of Rp
The following equation is used to calculate the accuracy of Rp from the accuracy of G.
Equation 2. Equation to calculate Rp
Rpe =
±Rpx
2
x
Ge
1 Rpx xGe
Where, Rpe is the accuracy of Rp, Ge is the accuracy of G, and Rpx is the measured Rp value.
Accuracy when ambient temperature exceeds the range between 18°C and 28°C (typical)
When the ambient temperature exceeds the range between 18°C and 28°C, multiply the accuracy obtained above by the coefficient shown in Table 9 below.
Table 9. Coefficient caused by ambient temparature
Coefficient
0°C ambient temperature < 8°C 3 8°C ambient temperature < 18°C 2 18°C ambient temperature 28°C 1 28°C < ambient temperature ≤ 38°C 2 38°C < ambient temperature ≤ 45°C 3
Table 6. Measurable capacitance range when measurement frequency is 1 kHz
Measurement Recommended Significant
range measurement range measurement range
100 pF 68 pF - 150 pF 0 F - 150 pF
220 pF 150 pF - 330 pF 0 F - 330 pF
470 pF 330 pF - 680 pF 0 F - 680 pF
1 nF 680 pF - 1.5 nF 0 F - 1.5 nF
2.2 nF 1.5 nF - 3.3 nF 0 F - 3.3 nF
4.7 nF 3.3 nF - 6.8 nF 0 F - 6.8 nF
10 nF 6.8 nF - 15 nF 0 F - 15 nF
22 nF 15 nF - 33 nF 0 F - 33 nF
47 nF 33 nF - 68 nF 0 F - 68 nF
100 nF 68 nF - 150 nF 0 F - 150 nF
220 nF 150 nF - 330 nF 0 F - 330 nF
470 nF 330 nF - 680 nF 0 F - 680 nF
1 µF 680 nF - 1.5 µF 0 F - 1.5 µF
2.2 µF 1.5 µF - 3.3 µF 0 F - 3.3 µF
4.7 µF 3.3 µF - 6.8 µF 0 F - 6.8 µF
10 µF 6.8 µF - 20 µF 0 F - 20 µF
Table 7. Measurable capacitance range when measurement frequency is 1 MHz
Measurement Recommended Significant
range measurement range measurement range
1 pF 680 fF - 1.5 pF 0 F - 1.5 pF
2.2 pF 1.5 pF - 3.3 pF 0 F - 3.3 pF
4.7 pF 3.3 pF - 6.8 pF 0 F - 6.8 pF
10 pF 6.8 pF - 15 pF 0 F - 15 pF
22 pF 15 pF - 33 pF 0 F - 33 pF
47 pF 33 pF - 68 pF 0 F - 68 pF
100 pF 68 pF - 150 pF 0 F - 150 pF
220 pF 150 pF - 330 pF 0 F - 330 pF
470 pF 330 pF - 680 pF 0 F - 680 pF
1 nF 680 pF - 1.5 nF 0 F - 1.5 nF
Measurement accuracy
Measurement accuracy is defined when all the following conditions are met.
• Warm-up time: 10 minutes or longer
• Ambient temperature: 18°C - 28°C
• Execution of the OPEN compensation
• Measurement cable length: 0 m, 1m, or 2m (16048A/B/D)
• D (dissipation factor) ≤ 0.5
4
±
±
*1. If you select a secondary measurement parameter other than D, calculate D.
5
Table 10. Measurement accuracy of Cp, Cs, D, G, and Rs (measurement frequency: 1 kHz)
Measurement
Measurement parameter
range
Cp, Cs [%] D G [nS] Rs [Ω]
100 pF
0.055 + 0.070 xK 0.00035 + 0.00070 xK (3.5 + 4.5 xK) xCx (90 + 120 xK) / Cx
0.055 + 0.030 xK 0.00035 + 0.00030 xK (3.5 + 2.0 xK) xCx (90 + 50 xK) / Cx
220 pF
0.055 + 0.045 xK 0.00035 + 0.00045 xK (3.5 + 3.0 xK) xCx (90 + 75 xK) / Cx
0.055 + 0.020 xK 0.00035 + 0.00020 xK (3.5 + 1.5 xK) xCx (90 + 35 xK) / Cx
470 pF
1 nF
2.2 nF
4.7 nF
10 nF
22 nF
47 nF 0.055 + 0.030 xK 0.00035 + 0.00030 xK (3.5 + 2.0 xK) xCx (90 + 50 xK) / Cx
100 nF 0.055 + 0.015 xK 0.00035 + 0.00015 xK (3.5 + 1.0 xK) xCx (90 + 25 xK) / Cx
220 nF
470 nF
1 µF
2.2 µF
4.7 µF
10 µF
Table 11. Measurement accuracy of Cp, Cs, D, G, and Rs (measurement frequency: 1 MHz)
Measurement
Measurement parameter
range
Cp, Cs [%] D G [nS] Rs [Ω]
1 pF
0.055 + 0.070 xK 0.00035 + 0.00070 xK (3.5 + 4.5 x K) x Cx (90 + 120 x K) / Cx
0.055 + 0.030 xK 0.00035 + 0.00030 x K (3.5 + 2.0 x K) x Cx (90 + 50 x K) / Cx
2.2 pF
0.055 + 0.045 x K 0.00035 + 0.00045 x K (3.5 + 3.0 x K) x Cx (90 + 75 x K) / Cx
0.055 + 0.020 x K 0.00035 + 0.00020 x K (3.5 + 1.5 x K) x Cx (90 + 35 x K) / Cx
4.7 pF
10 pF
22 pF 0.055 + 0.030 x K 0.00035 + 0.00030 x K (3.5 + 2.0 x K) x Cx (90 + 50 x K) / Cx
47 pF 0.055 + 0.015 x K 0.00035 + 0.00015 x K (3.5 + 1.0 x K) x Cx (90 + 25 x K) / Cx
100 pF
220 pF
470 pF
1 nF
In Tables 10 and 11, Cx is a measured value of the capacitance (Cp or Cs) [nF (for 1 kHz)/pF (for 1 MHz)] and K is defined as follows:
When Cx Cr: K = (1/Vs) x (Cr/Cx) When Cx > Cr: K = 1/Vs Where, Cr is a measurement range
[nF (for 1 kHz)/pF (for 1 MHz)] and Vs is a measurement signal level [V].
Figure 3. Accuracy of Cp and Cs when measurement frequency is 1 MHz (measurement signal level: 1V)
Figure 4 Accuracy of D when measurement frequency is 1 MHz (measurement signal level: 1V)
6
Figure 1, Figure 2, Figure 3, and Figure 4 show the accuracy of Cp, Cs, and D measured within the recom­mended measurement range for each measurement range (refer to Table 6 and Table 7) when D 0.1 and the ambient temperature is between 18°C and 28°C.
Figure 1. Accuracy of Cp and Cs when measurement frequency is 1 kHz (measurement signal level: 1V)
Figure 2. Accuracy of D when measurement frequency is 1 kHz (measurement signal level: 1V)
7
Measurement time
Table 12 shows the measurement time for each measurement time mode.
Table 12. Measurement time
Measurement Measurement time time mode (T3 + T4 + T5 of Figure 5)
Short 6.5 ± 0.5 ms
Long 16.5 ± 1.0 ms
Figure 5. Timing chart and measurement time
Table 13 shows the values of T1 - T7 when the following conditions are met.
Display: OFF Measurement range mode: Hold range mode (Hold) Trigger delay time: 0 ms Averaging factor: 1
(The ON/OFF of the compensation and comparator function does not affect the values of T1 - T7)
Table 13. Values of T1 - T7 (Typical)
Measurement Minimum Typical Maximum time mode value value value
T1 Trigger pulse width Independent 1 µs
*2
--
T2 Trigger response time of /READY_FOR_TRIG Independent - 200 µs 350 µs
Trigger response
T3 time of /INDEX Independent - 250 µs 400 µs
and /EOM
T4
Measurement time Analog measurement Short - 4.25 ms ­(T3 + T4 + T5) time (/INDEX) Long - 14.25 ms -
T5
Measurement
Independent - 2.0 ms -
computation time
T6 READY_FOR_TRIG setup time Independent - 15 µs 30 µs
T7 Trigger wait time Independent 0 µs - -
*2. When trigger signal is input through the handler interface.
Display renewal timing
The display is renewed immediately after the measure­ment completes (the /EOM signal goes low level). However, the display is renewed after transferring the measurement data if the external controller transmits a request to read measurement data before end of measurement. (The measurement data is transferred after the display renewal completes if the external controller transmits a request to read the measurement data after the display renewal starts.)
Measurement time (T4) when averaging function is used
When the averaging factor is set to more than 1, T4 is calculated as follows. (See Table 15.)
Measurement data transfer time through GPIB
Table 16 shows the measurement data transfer time when the following conditions are met.
Host computer: HP9000 Series,
Model C200 Display: ON Measurement range mode: Hold range mode (Hold) OPEN/SHORT/LOAD compensation: ON Measurement signal monitor: OFF BIN count function: OFF
Display time
When the display is ON, the instrument needs the time (display time) to renew the displayed items. The display time shown in Table 14 depends on the display page of the instrument setup display area (use the [Show Setting] key in the front panel to select the item you want to display). The ON/OFF of the deviation measurement mode does not affect the display time.
Table 14. Display time
Display page of the instrument Typical setup display area value Page number Display item
1
Frequency and level
of measurement signal
2.5 ms
2
Averaging factor and
measurement cable length
3
Measurement range and
trigger delay time
4
Measurement signal
level monitor values 4.0 ms
5 Multi compensation settings
6
Handler output
(comparator sorting result)
7 to 18 Limit range settings of comparator
19 to 24 BIN count results
8
Table 16. Measurement Data Transfer Time (Typical)
Data transfer format
Comparator
On Off
Required time for data readout of a measurement using :FETC? command
ASCII 2.3 ms 2.1 ms
(Total time of command transmission and data transfer)
Binary 2.5 ms 2.3 ms
Command transmission
ASCII
1.9 ms 1.9 ms
Required time for data readout of Data transfer 1.7 ms 1.5 ms
a measurement using :READ? command Command transmission
Binary
1.9 ms 1.9 ms
Data transfer 1.4 ms 1.2 ms
Required time for data readout of 1000 measurements using
ASCII 1.40 s 1.15 s
:DATA? BUF3 command (Total time of command transmission and data transfer)
Binary 0.65 s 0.49 s
Table 15. Measurement time @ averaging ≥ 2(typical)
Measurement frequency Measurement time mode Equation to calculate T4*3[ms] (typical)
1 kHz
Short T4 = 4.85 x Ave - 0.6
Long T4 = 14.85 x Ave - 0.6
1 MHz
Short T4 = (4
x
100 +Fshift
+ 0.85) xAve -0.6
Long T4 = (14
x
100 +Fshift
+ 0.85) xAve -0.6
100
100
*3. Ave: Averaging factor, F Shift: Frequency Shift setting (-1, 0, 1 or 2)
9
Measurement assistance functions
Compensation function
OPEN compensation, SHORT compensation, LOAD compensation, and OFFSET compensation are all available.
MULTI compensation function
OPEN/SHORT/LOAD compensation for 64 channels
• The LOAD compensation standard value can be
defined for each channel.
Display
Meter has a 40-column x2-row LCD display.
Deviation measurement function
Deviation from reference value and percentage of deviation from the reference value can be output as the result.
Comparator function
BIN sort The primary parameter can be sorted into 9 BINS,
OUT_OF_BINS, AUX_BIN, and LOW_C_REJECT. The secondary parameter can be sorted into High, In, and Low.
Limit setup An absolute value, deviation value, and %
deviation value are available for setup.
Bin count Countable from 0 to 999999.
Low C reject function
Extremely low measured capacitance values can be automatically detected as a measurement error.
Measurement signal level monitor function
• Measurement voltage and measurement current can be monitored.
• Level monitor accuracy (typical): ±(3% + 1 mV)
Data buffer function
Up to 1000 measurement results can be read out in a batch.
Save/recall function
Up to 10 setup conditions can be written to and read from the built-in non-volatile memory.
Resume function
• Recovers the instrument setup automatically saved at power-down at the next power-on.
• Memory retention time (typical): 72 hours (23°C ±5°C)
Key lock function
The front panel keys can be locked.
GPIB interface
Complies with IEEE488.1, 2 and SCPI.
Handler interface
The input and output signals are negative logic and optically-isolated open collector signals.
Output signal Bin1 - Bin9, Out of Bins, Aux Bin, P-Hi,
P-Lo, S-Reject, INDEX, EOM, Alarm, OVLD, Low C Reject
Input signal Keylock, Ext-Trigger
Scanner interface
The input and output signals are negative logic and optically-isolated open collector signals.
Output signal INDEX, EOM
Input signal Ch0 - Ch5, Ch valid, Ext-Trigger
Measurement circuit protection
See Table 17 for the maximum discharge withstand voltage that the internal circuit can be protected if a charged capacitor is connected to the UNKNOWN terminal is as follows.
NOTE: Discharge capacitors before connecting them to the
UNKNOWN terminal (or a test fixture).
Table 17. Maximum discharge withstand voltage (typical)
Maximum discharge Range of the capacitance
withstand voltage value C of DUT
1000 V C < 2 µF
2/C V C 2 µF
Figure 6. Maximum discharge withstand voltage (typical)
EMC
European Council Directive 89/336/EEC IEC 61326-1:1997+A1
CISPR 11:1990 / EN 55011:1991 Group 1, Class A IEC 61000-4-2:1995 / EN 61000-4-2:1995
4 kV CD / 4 kV AD
IEC 61000-4-3:1995 / EN 61000-4-3:1996
3 V/m, 80-1000 MHz, 80% AM
IEC 61000-4-4:1995 / EN 61000-4-4:1995
1 kV power / 0.5 kV Signal
IEC 61000-4-5:1995 / EN 61000-4-5:1995
0.5 kV Normal / 1 kV Common
IEC 61000-4-6:1996 / EN 61000-4-6:1996
3 V, 0.15-80 MHz, 80% AM
IEC 61000-4-11:1994 / EN 61000-4-11:1994
100% 1 cycle
Note: When tested at 3 V/m according to IEC 61000-4-3:1995 / EN 61000-4-3:1996, the measurement accuracy is double the accuracy of basic specification when the test frequency is 1 kHz and the instrument measurement range is 100 pF.
AS / NZS 2064.1/2 Group 1, Class A
Safety
European Council Directive 73/23/EEC IEC 61010-1:1990+A1+A2 / EN 61010-1:1993+A2
INSTALLATION CATEGORY II, POLLUTION DEGREE 2 INDOOR USE
IEC60825-1:1994 CLASS 1 LED PRODUCT
CAN / CSA C22.2 No. 1010.1-92
General specifications
Power source
Voltage 90 VAC - 132 VAC, 198 VAC - 264 VAC
Frequency 47 Hz - 66 Hz
Power consumption Max. 35 W /Max. 100 VA
Operating environment
Temperature 0°C - 45°C Humidity ( 40°C, no condensation) 15% - 95% RH
Altitude 0 m - 2000 m
Storage environment
Temperature -40°C - 70°C Humidity ( 65°C, no condensation) 0% - 90% RH
Altitude 0 m - 4572 m
Weight
3 kg (nominal), 6.6 lbs
Outer dimensions
320 (width) x 100 (height) x 300 (depth) mm (nominal)
12.6(W) x 3.9(H) x 11.8(D) inches (nominal)
Figure 7. 4288A Front view
Figure 8. 4288A Rear view
Figure 9. 4288A Side view
10
11
When measurement parameter is Cp-G
The following is an example of calculating accuracy of Cp and G, assuming that measured result of Cp is
8.00000 nF and measured result of G is 1.00000 µ S.
The accuracy of Cp is the same as that in the example of Cp-D.
From Table 10, the equation to calculate the accuracy of G is (3.5 + 2.0 xK) xCx. Substitute K = 2.5 (same as Cp-D) and 8.00000 nF of the measured Cp result into this equation. The accuracy of G is (3.5 + 2.0 x2.5)
x
8.00000 = 68 nS(0.068 µS).
Therefore, the true G value exists within 1.00000 ± 0.068 µS, that is, 0.932 µS to 1.068 µS.
When measurement parameter is Cp-Rp
The following is an example of calculating accuracy of Cp and Rp, assuming that measured result of Cp is
8.00000 nF and measured result of Rp is 2.00000 M.
The accuracy of Cp is the same as that in the example of Cp-D.
From Table 10, the equation to calculate the accuracy of G is (3.5 + 2.0 xK) xCx. Substitute K = 2.5 (same as Cp-D) and 8.00000 nF of the measured Cp result into this equation. The accuracy of G is (3.5 + 2.0 x2.5)
x
8.00000 = 68 nS. Then, substitute the obtained G accuracy into Equation 2. The accuracy of Rp is ± (2
x 10
6)2
x
68 x10
-9
/(1 2 x10
6
x
68 x10
-9
) = ± 0.272
x
10
6
/(1 0.136) , that is, -0.23944 Mto 0.31481 M.
Therefore, the true Rp value exists within 1.76056 Mto
2.31481 M.
When measurement parameter is Cs-Rs
The following is an example of calculating accuracy of Cp and Rs, assuming that measured result of Cs is
8.00000 nF and measured result of Rs is 4.00000 k.
Because the Cs accuracy is D = 2 xπxFreq xCs xRp = 2 xπx10
3
x8 x
10
-9
x4x
10
3
= 0.2> 0.1, multiply 0.13%
(the result obtained for Cs-D) by 1 + D
2
. The result is
0.13 x (1 + 0.2
2
) = 0.1352%.
From Table 10, the equation to calculate the accuracy of Rs is (90 + 50 xK)/Cx. Substitute K = 2.5 (same as Cs-D) and 8.00000 nF of the measured Cs result into this equation. The accuracy of G is (90 + 50 x2.5)/8.00000 =
26.875 . Because D>0.1, multiply the result by 1 + D
2
as in the case of Cs. The final result is 27.95 Ω.
Therefore, the true Cs value exists within
8.00000 ± (8.00000 x0.1352/100) = 8.00000 ± 0.01082 nF, that is, 7.98918 nF to 8.01082 nF, and the true Rs value exists within 4.00000 ± 0.02795 k, that is, 3.97205 to
4.02795 k.
Sample calculations of measurement accuracy
This section describes an example of calculating the measurement accuracy for each measurement parame­ter, assuming the following measurement conditions:
Measurement signal frequency 1 kHz Measurement signal level 0.5 V Measurement range 10 nF Measurement time mode Short mode Ambient temperature 28°C
When measurement parameter
is Cp-D (or Cs-D)
The following is an example of calculating accuracy of Cp (or Cs) and D, assuming that measured result of Cp (or Cs) is 8.00000 nF and measured result of D is 0.01000.
From Table 10, the equation to calculate the accuracy of Cp (or Cs) is 0.055+0.030 xK and the equation to calculate the accuracy of D is 0.00035 + 0.00030 xK. The measurement signal level is 0.5, the measurement range is 10 nF, and the measured result of Cp (or Cs) is
8.00000 nF. Therefore, K = (1/0.5) x(10/8.00000) = 2.5. Substitute this result into the equation. The result is: the accuracy of Cp (or Cs) is 0.055 + 0.030 x2.5 = 0.13% and the accuracy of D is 0.00035 + 0.00030 x2.5 = 0.0011.
Therefore, the true Cp (or Cs) value exists within
8.00000 ± (8.00000 x0.13/100) = 8.00000 ± 0.0104 nF, that is, 7.9896 nF to 8.0104 nF, and the true D value exists within 0.01000 ± 0.0011, that is, 0.0089 to 0.0111.
When measurement parameter is Cp-Q (or Cs-Q)
The following is an example of calculating accuracy of Cp (or Cs) and Q, assuming that measured result of Cp (or Cs) is 8.00000 nF and measured result of Q is 20.0.
The accuracy of Cp (or Cs) is the same as that in the example of Cp-D.
From Table 10, the equation to calculate the accuracy of D is 0.00035 + 0.00030 xK. Substitute K = 2.5 (same as Cp-D) into this equation. The accuracy of D is 0.00035 + 0.00030 x2.5 = 0.0011. Then, substitute the obtained D accuracy into Equation 1. The accuracy of Q is ± (20.0)
2
x
0.0011/(1 20.0 x0.0011) = ±0.44/(1 0.022), that is,
-0.43 to 0.45.
Therefore, the true Q value exists within 19.57 to 20.45.
±
±
±
±
Agilent Technologies’ Test and Measurement Support, Services, and Assistance
Agilent Technologies aims to maximize the value you receive, while minimizing your risk and problems. We strive to ensure that you get the test and measurement capabilities you paid for and obtain the support you need. Our extensive support resources and services can help you choose the right Agilent products for your applications and apply them successfully. Every instrument and system we sell has a global warranty. Support is available for at least five years beyond the production life of the product. Two concepts underlie Agilent’s overall support policy: “Our Promise” and “Your Advantage.”
Our Promise
Our Promise means your Agilent test and measurement equipment will meet its advertised performance and functionality. When you are choosing new equipment, we will help you with product information, including realistic performance specifications and practical recommendations from experienced test engineers. When you use Agilent equipment, we can verify that it works properly, help with product operation, and provide basic measurement assistance for the use of specified capabilities, at no extra cost upon request. Many self-help tools are available.
Your Advantage
Your Advantage means that Agilent offers a wide range of additional expert test and measurement services, which you can purchase according to your unique technical and business needs. Solve problems efficiently and gain a competitive edge by contracting with us for calibration, extra-cost upgrades, out-of­warranty repairs, and on-site education and training, as well as design, system integration, project management, and other professional engineering services. Experienced Agilent engineers and technicians worldwide can help you maximize your productivity, optimize the return on investment of your Agilent instruments and systems, and obtain dependable measurement accuracy for the life of those products.
Key literature
Agilent 4288A Capacitance Meter product overview, publication number 5980-2861EN.
Web resources
Please visit our component manufacturer industry area at:
www.agilent.com/find/component_test
By internet, phone, or fax, get assistance with all your test and measurement needs
Online assistance:
www.agilent.com/find/assist
Phone or Fax United States: (tel) 1 800 452 4844
Canada: (tel) 1 877 894 4414 (fax) (905) 282-6495
Europe: (tel) (31 20) 547 2323 (fax) (31 20) 547 2390
Japan: (tel) (81) 426 56 7832 (fax) (81) 426 56 7840
Latin America: (tel) (305) 269 7500 (fax) (305) 269 7599
Australia: (tel) 1 800 629 485 (fax) (61 3) 9210 5947
New Zealand: (tel) 0 800 738 378 (fax) 64 4 495 8950
Asia Pacific: (tel) (852) 3197 7777 (fax) (852) 2506 9284
Product specifications and descriptions in this document subject to change without notice.
Copyright ©2000 Agilent Technologies Printed in USA December 28, 2000 5988-0362E
Loading...