Agilent 4280A Data Sheet

This literature was published years prior to the establishment of Agilent Technologies as a company independent from Hewlett-Packard and describes products or services now available through Agilent. It may also refer to products/services no longer supported by Agilent. We regret any inconvenience caused by obsolete information. For the latest information on Agilent's test and measurement products go to:
www.agilent.com find products Or in the US, call Agilent Technologies at 1-800-452-4844 (8am-8pm EST)

H428OA Applications

l
capacitance and Conductance With or Without
Constant DC Bias
HP’s 4280A 1MHz C Meter/C-V Plotter can measure and display capacitance and conductance. A 1 MHz test signal with level of 1OmVrms or 30mVrms is used.
Internal DC bias ( f 1OOV) can be applied when needed.
4280 Front Panel Displays C,G and DC bias
n Test Floating and Grounded Devices
Both Floating and Grounded devices can be tested. This is important because wafer probers will have either floating or grounded chucks. Both configurations are in common use.
n
capacitance and Conductance Using Staircase
Bias Sweep
4280’s built-in + IOOV DC bias supply can be controlled from the front panel to sweep in staircase fashion. Capacitance and conductance can be measured at each step.
Deep Level Transcient Spectroscopy
(DL TS) - Example of Pulse Bias Application
DLTS is used to analyze semiconductor imperfections which effect IC performance. Small imperfection concentrations which are too small for analysis by C-V techniques can often be analyz­ed by DLTS.
m Hard Copy Output
C-V and G-V measurements require no external equipment and are internally synchronized. Results can be output to analog X-Y recorders or to computers via HP-IB (IEEE 488).
n Capacitance and Conductance vs. Time
(Pulse Bias)
Minority carrier lifetime and surface generation velocity in
MIS structures can be obtained using C-t results. 4280A uses two different C-t methods depending on measurement resolu­tion required on the time axis.
Delay time is the time between measurements. When delay time is set to 10 milliseconds or greater, the 4280A will apply a single pulse to the device under test. Then a BURST of measurements are made.
START
When delay time is less than 10 milliseconds, an external pulse generator like HP’s 8112A must be used. The external pulse generator applies repetitive pulses to the device under test. The 4280A makes a single measurement after each pulse.
In the SAMPLING MODE, the 4280A provides synchronization
signals to the external pulse generator.
Time (5)
4
STOP
High speed Gt testing and variable supercooled temperature are major requirements of DLTS test systems. Shown here is system hardware including HP’s 4280A for Gt testing with
pulses of IOms to 32 set in duration. HP’s 8112A pulse gen­erator is added when pulse duration shorter than 10m set is
required.
DLTS is a high-frequency capacitance transient thermal
scanning method useful for observing a wide variety of traps
in semiconductors. This new technique, aimed at studying these
centers, uses the capacitance of a p-n junction or Schottky bar-
rier as a probe to monitor the changes in the charge state of
the centers. Complete Gt characteristics are obtained at multi­ple supercooled temperatures.
Real devices have multiple trap levels and different trap concentrations. Resulting DLTS curve has multiple peaks as shown below.
Temperature (deg C)
Important parameters which can be derived from DLTS in-
clude: 1) surface state density, 2) trap concentration, 3) energy
level of traps, and 4) trap capture cross section.
Advantages offered by HP’s 4280A are to expand the range
of analysis to shallower energy levels by offering resolution to
1 O,us when synchronized to an external pulse generator like HP’s
8112A.
HP’S 4280A Offers New Measurement C;
R6
MC,
Tc
M
VC
S
n
FC
(2
S.,
Wntroduction
Hewlett-Packard’s Model 4280A 1 MHz C Meter/C-V
Plotter offers new measurement capability and flexibility
for the design and production of IC’s. Benefits are im-
proved IC quality and improved engineering productivity.
HP’s 4280A has capability previously requiring the following complicated test set up: 1) capacitancelconduc­tance meter, 2) function generator, and 3) computer for test synchronization.
New measurement capability is featured in transient C-t measurements with 10 microsecond resolution. Such testing is used to analyze deep level impurity concentra­tions which effect IC performance. Transient C-t resolu­tion of 10 microseconds is up to 1000 times better than ever before available. 4280A features 10ms C-t resolu­tion using internal pulse generator. Add HP 8112A pulse generator or equivalent to achieve 10,~s resolution.
Convenience features include the ability to test both floating and grounded devices. Also 428OA’s has capabili­ty to compensate a wide range of stray impedances. This helps eliminate the effect of test fixture residuals.
OFFn nON
cl
:42V = MAXA UNKNOWN
&OATlNG
26
meni Func
Th millis parer
MC
2
W
W428OA Front Panel
Features
1 HP-IB - Construct Your Own System
Standard on 4280A, HP-IB can help you construct an automatic
system. Such systems are used in applications ranging from materials research, device R&D, process engineering, wafer pro­duction and quality assurance.
2 Residual Compensation and Capacitance Offset Compensate test fixture residuals including up to 5 meters of stan­dard cable (HP P/N 81204195). Also use with Option 001 when off-
setting large values of capacitance to obtain extra digit of resolu­tion on 100pFllnF ranges.
3 Capacitance Digital Display Standard capacitance display resolution will be 3% digits or 4% digits depending on test conditions. Option 001 features 5% digit capacitance resolution using capacitance offset function. This display also can show deviation.
4 Wide Capacitance Measurement Range Capacitance measurement range is from 0.001 pF to 1 .QOOOnF.
5 Conductance Digital Display Standard conductance display resolution will be 3% digits or 4% digits depending on test conditions. This display also shows deviation.
6 Wide Conductance Measurement Range Conductance measurement range is from 0.01~5 to 12.000mS.
7 DC Bias and Time Display
DC bias parameters in volts or time in seconds is displayed depen-
ding on operating mode.
8
Wide Range of DC Bias and Time
DC bias can be varied in + IOOV range. Bias can be 1) constant,
2) staircase sweep (C/G-V) or 3) pulsed (C/G-t). V-t display shows values of voltage staircase sweep parameters
in C/G-V modes. These parameters include hold time, and step delay
time. V-t display also shows C/G-t parameters like delay time.
9
Enter Numedc Values of Voltage and Time After selecting bias and time parameters and keying in numeric values, this key enters value into 4280A memory.
10 Numeric Data Ready to Enter This LED must be lit before numeric sweep parameter can be entered.
11 Sweep StartLStop
Start/Stop key controls DC bias and time sweep output. V output
LED lights when bias voltage is applied.
12 Se/act Sweep Parameter or Constant DC Bias
Use up/down keys to select constant DC bias voltage or time sweep
parameter. Enter numeric values using numeric keys.
13
Select Bias Mod&Limit Output Voltage%ld Range
Multi function key selects DC bias mode. Also allows entry of DC bias voltage limit from numeric keys and acts as Range Hold key.
14 Dispfay Deviation, Percent or Deviation Percent
Perform math operations and display results on C/G displays.
15 Change Number of Display D&its
Reduce number of display counts to less than 1000. Use before X-Y recording
16 Setup X-Y Recorder Analog output voltage represents C/G display counts and DC bias/time sweep. Establish origin and size of X-Y plotting area.
MC
i
Note
ME.
A
tc
Prir
3
Dability for Design and Production of /C’S
EUtjCTlON
C-G c
G
- hJ/JTH -
18
‘2.G RANGE , , ,r~ e
&UT0 MANUAL inn hls,
ZERO
25
“is:,.
“...
gPEN
24
SELF
TEST
17TdggarandSwasptUodeContml
In sweep mode, this key selects single or repetative sweep trig-
gered from front panel, or select single sweep tnggerea externally.
, .*
Also can trigger single measurement internally or externally.
18 Be&t Displav Function In C only function, G displays blanks and vice versa. Advantage is test speed approximately doubles.
79 Measulement Range control Manual ranging may be needed during X-Y recording. Also needed using C offset capability to obtain one extra digit of C resolution.
20
Measumment speed Control
Fast mode requires approximately 70ms per measurement in C or
G only modes and 150ms per measurement in C and G mode.
Medium mode requires approximately 100ms per measurement in
C or G only modes and 1 QOms per measurement in C and G mode. Slow mode requires approximately 330ms per measurement in C
or G only mode and 520ms in C and G mode. (Advantage is noise
rejection and in some cases 1 more digit display resolution).
2
1 Add One D&it to Capacitanca Resolution Option 001 and C-offset enables one extra digit of resolution on 1OOpF and InF ranges.
22
select Fmm Two Test Levels
Choose 30mVrms or 1OmVrms AC test level.
STORE DIGIT SHIFT
A % A %
X-Y RECORDER
TRIGGER
Self test does not check calibration.
With fixture open, store residual capacitance and conductance.
ENTER
23
Vettfy Normal Operation
24 Compensate Fixture Residuals
Residuals are compensated when CORRECTION ENABLE is turn­ed on.
25 Pmtaction Fmm High Voltaga Bias. When switch is in + 42V position, shorting cap has no effect. Max output is &42V.
When switch is in + 1OOV position - center pin and shield of Remote on/off connector must be shorted to turn bias on. This should be accomplished by removing shorting cap and shorting center pin to shield through remote switch.
26
Test Floating or Gtvundad Devices
Floating devices can be tested over 4280A’s entire measurement
range. Grounded devices can be tested in top two ranges only
(1 OOpF/l mS and 1 nF/l OmS ranges).
27 Two Ten&al Pair
Hi and Low terminals have guard on coaxial shields.
28
Mounting Guides for l&WA and 16081A Fixwas
1608OA, 16081 A and 16082A fixtures mounts directly on 4280A front
panel. These guide holes accept fixture guide pins.
29 Connect Circuit Ground to Chassis Ground
I
9
I
4
n
4280A Specifications
n
428OA Measurement Functions: II C-V Measurement:
Capacitance (C), conductance (G), capacitance and con­ductance (C&G), capacitance vs. DC bias (C-V), conduc­tance vs. DC bias (G-V), capacitance and conductance vs.
DC bias (C&G-V), capacitance vs. time (C-t), conductance vs. time (G-t) and capacitance and conductance vs. time (C&G-t).
AC Test Signal:
Frequency:
OSC
Display:
Maximum Display Counts: C =
Measurement Terminals:
Configuration:
Max Offset Voltage: Max Resistance: Max Allowable Current: +
Connection Mode:
between DUT and Measurement circuit.
Usage
Connection
1 MHz + 0.01 %
Level:
30mVrms If: 10% or 1 OmVrms + 10 %
Max 4% digits (5% digits for Opt 001)
19000,
G = 12000
Two-terminal pair
High and Guard, Low and Guard
+ 1mV (DC Bias OFF) + 20R
1 OOmA
Set connection configuration
Connection Mode
Floating Grounded
To measure
Floated Device Grounded Device
To meawre
To measure
strays for compensation
Measures C-V, G-V or C&G-V characteristics using inter-
nal staircase bias.
n
C-t Measurement:
Measures C-t, G-t or C&G-t characteristics using internal bias source, external pulse bias source, or combination of internal and external bias sources.
Burst Mode:
repetative measurements are made with specified time interval between measurements. Used when delay time setting >lOmS.
Sampling Mode:
single samples taken between pulses. Delay be­tween application of measure voltage and sample
can be specified. Used when delay time setting is
less than 1Oms. This mode requires addition of an
external pulse generator like HP’s 8112A.
Measurement Speed vs. Oscillator Level and
Display Resolution:
Measurement Speed OSC Level Display Digit
pgiigq
W Measurement Accuracy:
Measurement accuracy in the following tables is valid when these conditions are met: 1) r30-minute warm up, 2) ZERO/OPEN calibration is performed, 3) CORRECTION
I
v1 = Internal DC bias source when an external bias source
(pulse generator) is used. V1 and Va can be set either from internal or external DC bias source.
n
C/G Measurement Range
I I
(Error Compensation Of9
I
Measurement Range
C/G Range
C= IOpF C=OOO1pF to 19000pF G = 1 OOllS G = 0 OIBS to 120.00&
C=lOOpF C=O.OlpF to 19O.OOpF C = 0 01 pF to 50 OOpF’ G=lmS G = 0.0001 mS to 1 2000mS G = 0.0001 mS to 1 2000mS
C=lnF C = 0 0001 nF to 1.9000nF G = 1 OmS G = 0 001 mS to 12 OOOmS G = 0.001 mS to 1200CmS
‘Typical values
Floating DUT Grounded DUT
This C/G range IS not available in Grounded DUT
mode
C = 0 0001 nF to 1 7600nF’
I
ENABLE IS ON, and 4) Temperature 23”C+5”C. Note that correction enable compensates for measurement residuals and reduces measurement range by the amount of residual compensation.
Accuracy is valid at 4280A front panel with cable length
switch in “0” position. Additional error must be added
when using the 16082A test cables and with cable length
switch in the “lm” position. Add 0.1 % of reading for C and 0.2% of reading for G when 16082A is used.
C accuracies are specified when D I 0.05 when using
C&G, C&G-V and C&G-t display modes. C accuracies are specified when D 5 0.01 when using C only, G only, C-V G-V, C-t and G-t display modes. See page 10 Reference
Data for more.
One pulse is applied then
Repetative pulses are applied with
ACCUIX~
Tab/e
1 -C&G Display Modes:
Use this table in the following measurement modes: 1) C&G,
2) C&G-V and 3) C&G-t when not applying external fast pulses. See tables 3 and 4 for accuracy when applying external fast pulses.
DUT
Connection
Mode
Available
Bias Modes
Range - ‘”
1 NC = C de&y counts m C & G mode with CORRECTION ENABLE OFF
10
l-
1mS ?[02% rag+@+& cntsl +[03% rdg+(3+% cnts] 1 f[04% rdg+(30+$ cnts] 1 ?[05% rdg+(30+% cnts] lnF2 + 10 1 O/o rda + 3 cntsi k (0 2% rdq + 3 cnts) i-(0 3%
1 OmS ‘[1 2% rdg+(3+% cnts] k[l 2% rdg+(3+$ Cntsl +[I 4% rdg+(lO+$J cnts]
*INT Bias OFF,=== 2 /or A ‘EXT SLOW Bias = or JL *EXT SLOW Bias = 0rJ-L
*lNT+ EXT SLOW Bias
0% = 3OmVrms O!X = IOmVrms OSC = 3OmVrms ES = 1OmVrms
k (0 1 % rdg + 5 cnts)
?[02% rdg+(L+& cnts] c[O3% rdg+(i+& cnts] k (0 1 % rdg + 3 cnts) f (0 2% rdg + 3 + cnts) c[O 3%
OPF
Floating DUT Mode Grounded DUT Mode
‘INT Bias OFF, = .q ./or A ‘INT+ EXT SLOW Bias
t(0 2% rdg + 5 cnts)
1
This C/G range IS not wallable in Grounded DUT mode
rdg + 3+ cnts)
rdg+
10 cnts)
+(O 4% rdg + 30 + cnts)
+(04% rdgc 10 cnts) k[l 4% rdg+(20+$ cnts]
6
b
Accuracy Table 2 -
C only and G only Display Modes:
Use this table in the following measurement modes: 1) C only, 2) G only, 3) C-V, 4) G-V, 5) C-t and 6) G-t. Do not use this table when applying external fast pulse bias. See tables 3 and 4 for fast external pulse bias accuracy.
DUT
Connection
Mode
AvaIlable
Boas Modes
1 OpF
C/G Measuring Range
1 NG = G display counts I” C&G mode with CORRECTION ENABLE OFF 2 C accuracy is not specified for C display rPOOpF
1 oofis k [O 25% rdg + (5 +& cnis]
1 OOpF t[O 15% rdg+@+$ cnts] +[025% rdg+(3+$ cnts] -t [0 3 0 h
1mS
lnF2 i-[015X ,,:,+,3+$& cnts] ?[025O/a rdg+(3+ $ cnts] i- [O 3% rdg + (20 + $) cnls] 1 ?[04% rdg+(ZO+& cnts] 1 OmS3 k[l 25% rdg+(3+$ cnts] k[l 25% rdg+(3+ & cnts] ?[l 4% rdg+(ZO +$ cnts] t[l 4% rdg+(ZO+& cnts]
NC = C display counts I” C&G mode with CORRECTION ENABLE OFF
*INT Bias OFF. -Y, 2, / or A
*EXT SLOW Boas z or q
‘lNT+ EXT SLOW Boas
0% = BOmVrms
~[015% rdg+(5+& cnts]
+[025% rdg+(3++ cnts] +[035% rdg+(3+ s cnls]
Floating DUT Mode
NG’
NC’
NC’
&[025% rdg+(5+$) cnts]
k[O35% rdg+(5+ $$ cnts]
*INT Bias OFF z, 2, / 01 A
*EXT SLOW Boas = 01 q
OSC = IOmVrms OSC = 3OmVrms OSC = 1OmVrms
3 G accuracy IS not speded for G display r2mS
*INT + EXT SLOW Bias
This C/G range IS not awlable I” Grounded DUT mode
rdg + (40 ++ cnls] ?[04% rdg+(40+$ cnts]
2 [0 4% rdg + (40 + $ cnts] ?[05% rdg+(40+$ cnts]
Grounded DUT Mode
NG’
1
b
Accuracy Table 3 -
C&G-t Modes:
Use this table in the C&G-t display mode when applying fast pulses
through external fast connector.
DUT
Connection
Mode
AvaIlable Bl% Modes
1OpF
C/G
Measuring
Range
1 NC = C display tour
1 oops
1 OOpF
1mS
1nF
1 OmS
T IS calculated at Nth measwement paint of a C&G-f cuive T = N*delay tome fld)*106 T is in seconds
‘EXT FAST Bias nL ‘EM FAST + INT Bias q + = ‘EXl FAST + EXT SLOW Bias 2 + =
t[(O4++/0 idg+(40+~+2T1*10-6)cnts]
i;[(O5++ rdg+(40+% +T+2T1.10-6) cnts]
?[(04+$)% rdg+(33+3+2T1*10-6) cnts]
+[(05++ rdg+(33+$$+F+2T’*106) cnts]
5 0
k [(l 4 + $ /o rdg + (43 + “+ 2T’ .10~6) cnls]
+[(25+$)% rdg+(43+&+?+2T’&) cnts]
I” C&G-t mode with CORRECTION ENABLE OFF
Accuracy Table 4 -
Floating DUT Mode (Note: Grounded DUT Mode not available)
OSC = 30mVrms
C-t,
G-t Modes:
Use this table when in the C-t or G-t display mode and applying fast
pulses through external fast bias connector
DUT
Connection
Mode
Available ‘EXT FAST Bias q
Bl% Modes
1 OpF
1 oars
‘EXT FAST + INT Bias q+-= ‘EXT FAST + EXT SLOW Bias 4 +=
?[(06 +$/~rdg(40 +!$+= +2T’*lO”)cnts] i[(O7+~)~/~rdg+(40+~+~+2T~~,o-~)cnts]
T’ T’ T’ T’
?[(07 ++ rdg(40 +F+F +2T1*106)cnts] ?[(08 ++rdg+(40+ !!$+~+2T1.10-s)~,,~~]
Floating DUT Mode (Note: Grounded DUT Mode not available)
OSC = 3OmVrms
OSC = 1OmVrms
?[(05+$)% rdg+(40+y+2T’*10-6) cnts]
5 [(O 6 + +/o rdg + (40 + ,$ + + + 2T’ ~10~~) cnts]
,[(05++ rdg+(33+?+2T’&) cnts]
5 0
+[(06+Y)/0 rdg+(33+&+900+2T1.106
k[(l 5+$)% rdg+(43+=$+2T’*@) cnts]
t[(25++ rdg+(43+% +~+2T’*10~6) cnts]
N 1
OSC = IOmVrms
T’ 1
cnts]
1OOpF + [(O 6 +$o rdg +(33 +s+3= + 2T’ *, O-6) cnts]
1mS
1 nF*
1OmS
1 NC = G display cwnts I” C&G mode with CORRECTION ENABLE OFF
NC = C display CO”“& in C&G mode wllh CORRECTION ENABLE OFF T IS calculafed at Nfh meawrement point of a Gt or Gl c”we T = N*delay time (ld)*106 T IS in seconds
T’ T’
*I(07 +~)n/,rdg+(33+4n+il+2T’.10~6)cnts]
T’
?[(l 6 +$ rdg +(43+$+3$+2T’~@)cnts]
T’
?[(2 7 +$o rdg + (43 +s+3+ + 2T’ *, 06) cnts]
T’
NC’ 300
t[(O8 +~)%rdg+(33+~+g~+2T1-,0”)cnts
T’ T’
+[(08 ++/a rdg+(33+~+903+2T1*10j~CnfS]
5 0 NG’
i[(l 7 +ii) /a rdg+(43 + 30+=+2T’+cnls
5 0
?[(27 +.)/o rdg+(43 + 11+,+2T’*106)cnfs
T’
T’
NC’ 900
7
n
Math Functions:
Display measured C/G values as differential values (A),
% ratio (%) or differential %
A = C-C&
% = k
or G-Gref
x 100% or Gx
Gref
(A%)
of the reference value.
100%
DC Bias Source (continued)
Number of Readings Which Can Be Obtained:
Burst Mode: 1 to 9999 Sampling Mode: 1 to 5/td (td = delay time)
Hold Time (th):
A% = p x
n
Self Test:
Verifies functional operation but does not calibrate.
100%
or $$@ x
100%
H DC Bias Source:
Output Mode:
/or A : Selected when using C-V, G-V,
q, = or OFF: SeSecztk?o;;;n using C-t, G-t or
: or OFF : Selected when using C only or G
Output Resistance: 10Q Max Output Current: Output Voltage Range/Resolution/Accuracy:
~1
O”G55”C
Ranging:
= mode when V Limit/Range hold is in use.
Staircase Sweep Parameter Settings (CV Function Only):
Start/Stop Voltage:
voltage range.
Step Voltage:
matically set as lesser of Start or Stop voltage setting. Example, if Start is 1 .OOOV and Stop is 3O.OV, resolution of Step is O.OlV.
OV to 3.999V: 1mV step
4.OOV to 39.99V: 1OmV step 4O.OV to 2OO.OV: 1OOmV step
Hold Time, Step Delay Time (th, td ):
3ms - 65ms: 1 ms step
0.07s
100.0s - 650.0: 1 OOms step
Hold Time/Step Delay Time Setting Accuracy:
Time 565ms: +[0.02% setting + (loons + bias settling time)]
Time z65ms: +[0.02% setting + (0.5ms + bias
settling time)]
Pulse Bias Parameter Settings (C-t Function Only):
DC/Pulse/Measurement Voltage:
the output voltage range.
OFF, -= (DC), 9, /or /k
C&G-V modes
only modes.
+6mA
Resolution *(% of setting + volts)
‘Accuracy is specified at 23°C + 5°C. Error doubles at
Auto ranging. Fixed range is provided in
Any voltage within the output
Resolution of step voltage is auto-
- 99.99s: 1 Oms step
+ (% of setting + seconds)
Any voltage within
EXT BIAS SLOW 65mS 5th C 1 S = 500~s
*Maxlmum Hold Time Setting I” sampling mode is llmited as
follows.
Function
N = Nth measuiement point, id = delay time
Measuring Speed
Maximum Hold Time
(whichever is shorter)
Delay Time (t(f):
Range: 1 Ops - 32s Resolution: 10~s Z&-j 565ms: 10~s
65ms 5 td 5 1 S: 500/S
fsstdslos: lms
tOs&d~32s: l&tlS
Delay Time Range in Burst Mode:
~1
Delay Time Range in Sampling Mode:
EXT BIAS SLOW: 200~ to 5s EXT BIAS FAST: lO/zs to 5s
Number of C/G-t Readings which can be obtained:
be obtained.
Burst Mode:
Sample Mode:
This determines the resolution which can
1 to 9999 readings
1 to
delay tir?re in seconds (readings)
Hold Time/Delay Time Setting Accuracy: + (%
setting + seconds)
Time 165ms: +[0.02% + (1 OOms*t + Bias Settling
Time)]**
Time r65ms: + [0.2% + (0.5ms + Bias Settling
mime)]**
*I: 10ms when internal bias is used.
l
2:
1Oms to 100ms
when doing C&G-t measurements.
delay will be added to delay time
W DC Bias Sourn: Reference Data
Settling Time: 0.05
of final value. AV = voltage Swing
Ranging Time: Bias Voltage Output Set/Reset Logic Processing
Time: I
20ms
Bias Voltage Setting Change Logic Processing Time ( z mode):
* AV + 1.7 (ms) to within 99.9%
10ms
120ms
31
of
41
8
n Recorder Output:
Output DC voltage proportional to C/G display counts
and voltage/time sweep range.
Output VOltgage: + 1OV for C, G and V/t data.
C or G Data:
V Data:
f 1OV for + 1000 counts (lOmV/count)
- 1OV for Start/Stop voltage, whichever is more negative. + IOV for Start/Stop voltage, whichever is more positive. Max 1000 data points (20mV resolution).
Time (t) Data:
- 1OV at the application of mea-
surement voltage. + 1OV at the final data point. Max
1000 data points (20mV resolution).
Output Voltage Accuracy:
+ (% of output voltage +
volts)
C or G = *(0.5% + 20mV)
V or t = +(0.5% +
Pen
Lift Output:
40mV)
Normally +5V. OV during C-V/t
measurement.
Smoothing Function:
Output filter time constant can be set 0.02s or 0.2s. Initial and post delay time of 2s and 1.5s can be inserted in C-V/t basic function.
Scaling Output:
L.L., Zero and U.R.
m Data Outpui/Remote Contmk
HP - IB:
documentation and support that delivers the shortest
path to a measurement system.
Block Mode Output:
measurement and store measured data in an internal data buffer. Then, packed data can be output. In this mode, front panel display, math function and recorder output do not function. Data with/without error com­pensation can be output.
Maximum Number of Data Which Can be Stored:
C, G, C-V/t or
C-G, C&G-V/t Function: 400 data points
n
Option lull High Resolution Offset
Not just IEEE-488, but the hardware,
Perform C-V/t characteristics
G-V/t
Function: 680 data points
Capacitance Measurement:
Increase C measurement resolutions by one digit on
1 OOpfll nF ranges using capacitance offset reference. This
option cannot be field installed.
c
Offset
can be set by measured data or numeric keys.
C Offset Resolution: C Offset Setting Accuracy:
value + 05pF). This error can be compensated using CORRECTION ENABLE key.
Measurement Accuracy:
accuracy reading before offset plus 2) accuracy of
reading after offset.
High Resolution C Mode ON/OFF Time:
Range: 0 to 1023 picofarads. C offset value
1 picofarad
+(0.2% of reference
Accuracy is the sum of 1)
350m
Measurement Range/Resolution/Display Digit:
Measuring
Time
FAST lOmV/30mV
SLOW lOmV/30mV
n
Reference data are given for information purposes and
should not be considered specifications.
Osc Leve, Measurement Range From Reference Value
MED -~~
1OmV
30mV
Range of Offset Value
lOOpF/lmS
k 19.OOpF + 19O.OpF * 120.0@ kl 200mS
+19.OOOpF k19OOOpF +120.00@s -cl 2000mS
Reference Data:
lnF/lOmS
n Residual Impedance of the l&.WOA:
Radial Lead Contact Block:
Axial Lead Contact Block:
70nH, 50mh2
90nH, 50mQ
n Additional Emrs Due to Test Lead Length:
When using 16081A, 16082A or P/N 8120-4195 coaxial cable up to 5m long, use below figures to estimate addi­tional error.
a
e 51
m
2
b
zi z 0.1
.g B 2 0
0.01 1OP
1
lOOr lm 10m
loop Ill
Test Lead
Length (m)
C(F)
8 Emr Compensation:
Cable Length Compensation:
Residuals of standard cable (HP P/N 8120-4195) up to 5 meters long can be internally compensated.
Zero
Open:
Internally compensate stray capacitance
and conductance of test fixture with fixture open.
Additional L-R Compensation:
not included above using external computer.
Measurement Conditions: Connection Mode: Test Signal Level:
Compensation Range:
L (in C display) and R (in G display).
L:
19.000/4H
R: 190.009
Compensation Accuracy = -+(% rdg + counts):
L: f [0.5% rdg + (20 +$$) counts]
Grounded for compensation
30mVrms
L-R data will be displayed as
Om, lm or 0-5m.
Compensate residuals
R: *[1.2% rdg + (lo+ &$-) counts]
1NR = resistance display counts
NL = inductance display counts
Reference Data: (Continued)
l C/G Measurement Time:
Total Measurement Time is the sum of 1) Net Measure-
ment Time, 2) Internal Error Compensation Time, 3) Math
Function Time, and 4) Display Time.
The table below shows Total Measurement Time in milliseconds. Net Measurement Time is shown in parentheses.
Ranging Time: OSC Level Changing Time: Connection Mode Changing Time:
Initial Setting Time:
10Oms
40ms
11 Oms
650ms
Time Constant of C/G Measurement Circuit:
50~s when measurement time interval is <lOms in
FAST q mode and < 1OOms in MED mode, in the
C-t, G-t, and C&G-t functions.
EXT FAST: 1~s EXT SLOW: 6fl
WW Characteristic Measun?ment Time:
2 * A + (B+C) * INTEGER (F+ 1) where A = Hold Time, B = Step Delay Time, C = C/G
Measurement Time, D = Step Voltage, E = Start
Voltage, and F = Step Voltage
n C-t Characteristic Measurement Time
Burst Mode Measurement Time:
(delay time l no. of readings).
Sampling Mode Measurement Time:
measurement time for n data points.
Measurement Time of nth Data:
C/G-t Measurement: See below figure C&G-t Measurement: Measurement time in below figure times two.
Hold time plus
Sum of
W2 Measurement Envm Caused by High D
Devices:
.05sDrl:
floating DUT rnoi: using EXT. BIAS FAST add Ton
1OpFIlOOpF ranges and D% on 1nF range.
Drl:
Add =
Multiply C error by (1 + D2).
rdg to C .error. Except for
n G Measurement Ems Caused by High Q
Devices:
0.55Grl: Q>l:
n
Genera/ Specifications
Add Q/10% to G measurement error
Multiply G measurement error*(l + Q2)
W Operating Temperature:
0°C - 55°C and Relative Humidity 190% at 40°C
n Power Requirement:
1 OOV, 12OV, 220V rlr 10 O/o, 240V + 5% - 10% , 48 - 66H2,
max 140VA
n Dimensions:
426mm (W) x 177mm (H) x 499mm (D) or 16.7” (W)
x 7.0”
(H) x 19.7” (D)
n Weight;
Approximately 15.6kg or 34.3 Ibs.
n
Ordering lnforma tion
n Standard Instrument:
4280A IMHz C Meter/C-V Plotter. . . . . . . . .
moptions:
Opt.
WI:
High Resolution Offset Capacitance
Measurement increases C Resolution by 1 digit on lOOpFl1 nF Ranges. Cannot be field installed.
Opt.
907: Front Handle Kit
(HP P/N 5061-0090)
Opt.
908: Rack Flange Kit
Opt Opt 910:
(HP P/N 5061-0078)
909: Rack Flange & Front Handle Kit
(HP P/N 5061-0084)
Extra Manual
(HP P/N 04280-90000)
n Accessories:
16D58A:
1609OA: 1609lA: 16092A1 16083A:
HP P/N 12!50-0929:
n External DC Bias:
Response Time of EXT SLOW Bias Circuit:
to within 99.9% of final value.
For more information, call your local HP Sales Office or nearest Regional Office: l Eastern (201) 265-5000; l Midwestern (312) 2559600; l Southern (404) 9551500; l Western (213) 9707500; l Canadian (416) 676.9430. Ask the operator for instrument sales. Or write Hewlett-Packard, 1601 Page Mill Road, Palo Alto, CA 94304. In Europe: Hewlett-Packard .%A., 7, rue du Bois-du-&in, P.O. Box, CH 1217 Meyrin 2, Geneva, Switzerland. In Japan: Yokogawa-Hewlett-Packard Ltd., 29-21, Takaido-Higashi Scheme. Suginami-ku, Tokyo 168.
Printed in USA
loops
Data Subject To Change
Test Fixture for Packaged Semicon-
ductor Testing Test Fixture (furnished with 4280A) Test Leads, 2 Meter, Double Shielded. Test Leads, 1 Meter, Single Shielded.
Pulse Bias Noise Clipper (for use with
external Pulse Generator)
Remote ON/OFF Shorting Cap
(furnished with 4280A)
5953-6949
Loading...