Agilent 4156B Technical Data

HP 4155B
This literature was published years prior to the establishment of Agilent Technologies as a company independent from Hewlett-Packard and describes products or services now available through Agilent. It may also refer to products/services no longer supported by Agilent. We regret any inconvenience caused by obsolete information. For the latest information on Agilent's test and measurement products go to:
www.agilent.com find products Or in the US, call Agilent Technologies at 1-800-452-4844 (8am-8pm EST)
Semiconductor Parameter Analyzer HP 4156B Precision Semiconductor Parameter Analyzer
Technical Data
Specifications - July 1997
Introduction
Basic Functions
The HP 4155B and HP 4156B functions:
ì Set measurement and/or stress
conditions
ì Perform measurement and analysis
with the built-in HP Instrument BASIC
ì Self test, Auto calibration
ì Control measurement and/or stress
execution
ì Perform arithmetic calculations ì Display measured and calculated
results on the LCD display
ì Perform graphical analysis
Configuration
HP 4155B
4xMPSMU 2xVMU 2xVSU
HP 4156B
4xHRSMU 2xVMU 2xVSU
ì Store and recall measurement setups,
and measurement and graphical display data
ì Dump to printers or plotters for hard-
copy output
HP 41501B (Optional)
2xPGU(Option)
GNDU
HPSMU(Option) or 2xMPSMU (Option)
*1
SMU: Source Monitor Unit HRSMU: High Resolution SMU (1fA/2µV to 100mA/100V) MPSMU: Medium Power SMU (10fA/2µV to 100mA/100V) HPSMU: High Power SMU (10fA/2µV to 1A/200V) VMU: Voltage Monitor Unit VSU: Voltage Source Unit PGU: Pulse Generator Unit (1 channel) GNDU: Ground Unit
*1: Minimum number of installable MPSMU or PGU is two.
*1
2
Hardware
Specification Condition
The “supplemental” information and “typical” entries in the following specifi­cations are not warranted, but provide useful information about the functions and performance of the instruments.
The measurement and output accuracy
are specified at the rear panel connector terminals when referenced to the Zero Check terminal under the following conditions:
1. 23°C ± 5°C (double between 5°C to 18°C, and 28°C to 40°C if not noted otherwise)
2. After 40 minutes warm-up
3. Ambient temperature change less than ± 1°C after auto calibration execution.
4. Integration time: medium or long
5. Filter: ON (for SMUs)
6. Kelvin connection (for HRSMU, HPSMU, and GNDU)
7. Calibration period: 1 year
HP 4156B Precision Semiconductor Parameter Analyzer
HRSMU (High Resolution SMU) Specifications
Voltage Range, Resolution, and Accuracy (HRSMU)
Voltage Range
± 20V ± 40V ± 100V
*1: 100mA (Vout[20V), 50mA (20V<Vout≤40V) *2: 100mA (Vout[20V), 50mA (20V<Vout≤40V), 20mA (40V<Vout≤100V)
Set. Reso.
1mV
5mV
Set.
Accuracy
±(0.025%+6mV)2mV
Meas. Reso.
Current Range, Resolution, and Accuracy (HRSMU)
Current Range
±A ±10µA ±100µA ±1mA
*1: The accuracy is applicable when offset cancellation has been performed. *2: The offset current specification is multiplied by one of the following factors depending upon the ambient temperature and humidity (RH = Relative Humidity):
*3: 100V (Iout[20mA) 40V (20mA<Iout50mA) 20V (50mA<Iout100mA)
Vout is the output voltage in volts. Iout is the output current in amps. For example, accuracy specifications are given as ±% of set/measured value (0.04%)
plus offset value (200nA+1nA×Vout) for the 1mA range. The offset value consists of a fixed part determined by the set/measuremet range and a proportional part that is multiplied by Vout or Vout/100.
Set. Reso.
10fA±10pA 10fA±100pA 100fA±1nA
±(0.5%+0.7pA+1fA×Vout)
1pA±10nA 10pA±100nA 100pA 1nA 10nA 100nA
±(0.5%+4pA+10fA×Vout)
±(0.12%+40pA+100fA×Vout)
±(0.12%+400pA+1pA×Vout)
±(0.07%+4nA+10pA×Vout)
±(0.07%+40nA+100pA×Vout)
±(0.06%+400nA+1nA×Vout)
±(0.06%+4µA+10nA×Vout)A±10mA
±(0.12%+40µA+100nA×Vout)10µA±100mΑ
5 - 60Temperature
Set.
Accuracy
RHHumidity % 60 - 80
×2×25°C to 18°C ×2×118°C to 28°C ×5×228°C to 40°C
*1*2
*1*2
Meas. Reso.
1fA±(4%+400fA)
±(4%+20fA+1fA×Vout/100)
1fA±(4%+400fA)
*2
±(4%+40fA+10fA×Vout/100)
10fA 10fA 100fA 1pA 10pA 100pA 1nA 10nA 100nA
±(0.5%+0.4pA+1fA×Vout)
±(0.1%+20pA+100fA×Vout)
±(0.1%+200pA+1pA×Vout) ±(0.05%+2nA+10pA×Vout)
±(0.05%+20nA+100pA×Vout)
±(0.04%+200nA+1nA×Vout)
±(0.04%+2µA+10nA×Vout)
±(0.1%+20µA+100nA×Vout)
Output terminal/connection:
Dual triaxial connectors, Kelvin (remote sensing)
HRSMU Measurement and Output Range
Meas.
Accuracy
±(0.01%+200µV)V±(0.02%+400µV)100µV± 2V
±(0.01%+1mV)20µV±(0.02%+3mV)
±(0.015%+2mV)40µV
±(0.02%+5mV)100µV±(0.03%+15mV)
Meas.
Accuracy
±(0.5%+2pA+10fA×Vout)
-100
Max. Current
100mA 100mA
*1 *2
*1*2
*1*2
*2
Voltage/Current Compliance (Limiting):
The SMU can limit output voltage or current to prevent damaging the device under test. Voltage: 0V to ±100V
Current: ±100fA to ±100mA Compliance Accuracy: Same as the current (voltage) settling accuracy.
HRSMU Supplemental Informa­tion:
Maximum allowable cable resistance
Max.
when using Kelvin connection (Force,
V
Sense): 10
100V
Typical voltage source output
100V
resistance (Force line/non-Kelvin
100V
connection): 0.2
100V
Voltage measurement input resistance/
100V
current source output resistance:
100V 100V 100V 100V 100V
*3
100
-20-40 20 40
-20
-50
-100
15
10
(10pA range)
Current compliance setting accuracy for opposite polarity: 10pA to 10nA range: V/I setting accuracy ± 12% of range
100nA to 100mA range: V/I setting accuracy ± 2.5% of range
Current (mA)
50
20
100
Voltage (V)
3
VSU (Voltage Source Unit) Specifications
VSU Output Range:
Meas.
Voltage
Reso.
Range
1mV
±20V *1: Specification is applicable under no load current. Max. Output Current: 100mA
±(0.05% of setting+10mV)*1
VSU Supplemental Information:
Output resistance: 0.2 Maximum load capacitance: 10µF Maximum slew rate: 0.2V/µs Current limit: 120mA (typical)
Output Noise: 1mV rms (typical)
Meas.
Accuracy
VMU (Voltage Monitor Unit) Specifications
VMU Measurement Range, Resolution, and Accuracy:
Voltage Range
Meas. Reso.
VMU Differential Mode Range Resolution, and Accuracy:
Meas.
Diff V
Reso.
Range
±(0.03%+100µV+1.3µV×Vi)1µV±0.2V
±(0.02%+1mV+13µV×Vi)2µV±2V
Max. Common Mode Voltage: ± 20V Note: Vi is the input voltage of VMU2 in volts.
Meas.
Accuracy
±(0.02%+200µV)2µV±2V
±(0.02%+1mV)20µV±20V
Meas.
Accuracy
For example, accuracy specifications are given as ±% of set/measured value (0.02%) plus offset value (1mV+13µV×Vi) for the 2V range. The differential mode offset value consists of a fixed part determined by the measurement range and a proportional part that is multiplied by Vi.
VMU Supplemental Information:
Input Impedance: 1G Input leakage current (@0V): 500pA (Typical)
Measurement noise: 0.01% of range (p-p) (Typical) Differential mode measurement noise:
0.005% of range (p-p) (Typical)
HP 4155B Semiconductor Parameter Analyzer
MPSMU (Medium Power SMU) Specifications
Voltage Range, Resolution, and Accuracy (MPSMU)
Voltage Range
± 20V ± 40V ± 100V
*1: 100mA (Vout[20V), 50mA (20V<Vout≤40V) *2: 100mA (Vout[20V), 50mA (20V<Vout≤40V), 20mA (40V<Vout≤100V)
Set. Reso.
1mV 2mV
5mV
Set.
Accuracy
Meas. Reso.
±(0.02%+700µV+0.3×Iout) V±(0.03%+900µV+0.3×Iout)100µV± 2V
±(0.02%+2mV+0.3×Iout)20µV±(0.03%+4mV+0.3×Iout) ±(0.02%+3mV+0.3×Iout)40µV±(0.03%+7mV)+0.3×Iout) ±(0.03%+5mV+0.3×Iout)100µV±(0.04%+15mV)+0.3×Iout)
Meas.
Accuracy
Current Range, Resolution, and Accuracy (MPSMU)
Current Range
±A ±10µA ±100µA ±1mA
*1: 100V (Iout[20mA), 40V (20mA<Iout≤50mA), 20V (50mA<Iout≤100mA)
Vout is the output voltage in volts. Iout is the output current in amps. For example, accuracy specifications are given as ±% of set/measured value (0.1%)
plus offset value (30pA+200fA×Vout) for the 100nA range. The offset value consists of a fixed part determined by the set/measuremet range and a proportional part that is multiplied by Vout.
Set. Reso.
100fA±1nA 1pA±10nA 10pA±100nA 100pA 1nA 10nA 100nA
±(0.5%+3pA+2fA×Vout)
±(0.5%+7pA+20fA×Vout)
±(0.12%+50pA+200fA×Vout)
±(0.12%+400pA+2pA×Vout)
±(0.12%+5nA+20pA×Vout)
±(0.12%+40nA+200pA×Vout)
±(0.12%+500nA+2nA×Vout)
±(0.12%+4µA+20nA×Vout)A±10mA
±(0.12%+50µA+200nA×Vout)10µA±100mΑ
Set.
Accuracy
Meas. Reso.
10fA 10fA 100fA 1pA 10pA 100pA 1nA 10nA 100nA
±(0.5%+5pA+20fA×Vout)
±(0.1%+30pA+200fA×Vout)
±(0.1%+200pA+2pA×Vout)
±(0.1%+3nA+20pA×Vout)
±(0.1%+20nA+200pA×Vout)
±(0.1%+300nA+2nA×Vout)
±(0.1%+2µA+20nA×Vout)
±(0.1%+30µA+200nA×Vout)
MPSMU Measurem ent and Output Range
VSU Specifications
Same as HP 4156B VSU.
Meas.
Accuracy
±(0.5%+3pA+2fA×Vout)
-100
VMU Specifications
Same as HP 4156B VMU.
Max. Current
100mA 100mA
*1 *2
Max. V
100V 100V 100V 100V 100V 100V 100V 100V
*1
Output terminal/connection:
Single triaxial connector, non-Kelvin (no remote sensing)
Voltage/Current Compliance (Limiting):
The SMU can limit output voltage or current to prevent damaging the device under test. Voltage: 0V to ±100V
Current: ±1pA to ±100mA Compliance Accuracy: Same as the current (voltage) settling accuracy.
MPSMU Supplemental Informa­tion:
Typical voltage source output resistance: 0.3 Voltage measurement input resistance/ current source output resistance:
13
10
(1nA range)
Current compliance setting accuracy for opposite polarity: 1nA to 10nA range: V/I setting accuracy ± 12% of range
100nA to 100mA range: V/I setting accuracy ± 2.5% of range
Current (mA)
100
50
20
-20-40 20 40
-20
-50
100
Voltage (V)
-100
Loading...
+ 7 hidden pages