Agilent 4155C Users Guide V2

User’s Guide
Volume 2
Measurement and Analysis
Agilent 4155C Semiconductor Parameter Analyzer
Agilent 4156C Precision Semiconductor Parameter Analyzer
Agilent Part No. 04156-90020
Printed in Japan January 2001
Edition 1
Legal Notice
The information contained in this document is subject to change without notice.
Copyright © 2001 Agilent Technologies
This document contains information which is protected by copyright. All rights are reserved. Reproduction, adaptation, or translation without prior written permission is prohibited, except as allowed under the copyright laws.
Product Warranty
Agilent Technologies warrants Agilent Technologies hardware, accessories and supplies against defects in materials and workmanship for the period of one year from the warranty start date specified below. If Agilent Technologies receives notice of such defects during the warranty period, Agilent Technologies will, at its option, either repair or replace products which prove to be defective. Replacement products may be either new or like-new.
Warranty service of this product will be performed at Agilent Technologies. Buyer shall prepay shipping charges to Agilent Technologies and Agilent Technologies shall pay shipping charges to return the product to Buyer. However, Buyer shall pay all shipping charges, duties, and taxes for products returned to Agilent Technologies from another country.
Agilent Technologies does not warrant that the operation of Agilent Technologies products will be uninterrupted or error free. If Agilent Technologies is unable, within a reasonable time, to repair or replace any product to a condition as warranted, customer will be entitled to a refund of the purchase price upon prompt return of the product.
The Agilent Technologies products may contain remanufactured parts equivalent to new in performance or may have been subject to incidental use.
The warranty period begins on the date of delivery or on the date of installation if installed by Agilent Technologies. If customer schedules or delays Agilent Technologies installation more than 30 days after delivery, warranty begins on the 31st day from delivery.
Warranty does not apply to defects resulting from (a) improper or inadequate maintenance or calibration, (b) software, interfacing, parts or supplies not supplied by Agilent Technologies, (c) unauthorized modification or misuse, (d) operation outside of the published environmental specifications for the product, or (e) improper site preparation or maintenance.
To the extent allowed by local law, the above warranties are exclusive and no other warranty or condition, whether written or oral, is expressed or implied and Agilent Technologies specifically disclaims any implied warranties or conditions of merchantability, satisfactory quality, and fitness for a particular purpose.
Agilent Technologies will be liable for damage to tangible property per incident up to the greater of $300,000 or the actual amount paid for the product that is the subject of the claim, and for damages for bodily injury or death, to the extent that all such damages are determined by a court of competent jurisdiction to have been directly caused by a defective Agilent Technologies product.
To the extent allowed by local law, the remedies in this warranty statement are customers sole and exclusive remedies. Expect as indicated above, in no event will Agilent Technologies or its suppliers be liable for loss of date or for direct, special, incidental, consequential (including lost profit or date), or other damage, whether based in contract, tort, or otherwise.
For consumer transactions in Australia and New Zealand: the warranty terms contained in this statement, except to the extent lawfully permitted, do not exclude, restrict or modify and are in addition to the mandatory statutory rights applicable to the sale of this product to you.
Assistance
Product maintenance agreements and other customer assistance agreements are available for Agilent Technologies products.
For any assistance, contact your nearest Agilent Technologies Sales Office.
Certification
Agilent Technologies, Inc. certifies that this product met its published specifications at the time of shipment from the factory. Agilent Technologies further certifies that its calibration measurements are traceable to the National Institute of Standards and Technology (NIST), to the extent allowed by the Institutes calibration facility, and to the calibration facilities of other International Standards Organization members.
Agilent 4155C/4156C User’s Guide Vol.2, Edition 1 3
Safety Summary
The following general safety precautions must be observed during all phases of operation, service, and repair of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual may impair the protections provided by the equipment. In addition, it violates safety standards of design, manufacture, and intended use of the instrument. Agilent Technologies, Inc. assumes no liability for customers failure to comply with these requirements.
NOTE Agilent 4155C/4156C/41501B comply with INSTALLATION CATEGORY II for
mains input and INSTALLATION CATEGORY I for measurement input terminals, and POLLUTION DEGREE 2 defined in IEC 1010-1.
Agilent 4155C/4156C/41501B are INDOOR USE products.
NOTE LEDs in Agilent 4155C/4156C/41501B are Class 1 in accordance with IEC 825-1.
CLASS 1 LED PRODUCT.
GROUND THE INSTRUMENT
This is Safety Class I instrument. To minimize shock hazard, the instrument chassis and cabinet must be connected to an electrical ground. The power terminal and the power cable must meet International Electrotechnical Commission (IEC) safety standards.
DO NOT OPERATE IN AN EXPLOSIVE ATMOSPHERE
Do not operate the instrument in the presence of flammable gases or fumes. Operation of any electrical instrument in such an environment constitutes a definite safety hazard.
KEEP AWAY FROM LIVE CIRCUITS
Operation personnel must not remove instrument covers. Component replacement and internal adjustments must be made by qualified maintenance personnel. Do not replace components with power cable connected. Under certain conditions, dangerous voltages may exist even with the power cable removed. To avoid injuries, always disconnect power and discharge circuits before touching them.
DO NOT SERVICE OR ADJUST ALONE
Do not attempt internal service or adjustment unless another person, capable of rendering first aid and resuscitation, is present.
4 Agilent 4155C/4156C Users Guide Vol.2, Edition 1
DO NOT SUBSTITUTE PARTS OR MODIFY INSTRUMENT
Because of the danger of introducing additional hazards, do not install substitute parts or perform any unauthorized modification to the instrument. Return the instrument to a Agilent Technologies Sales and Service Office for services and repair to ensure that safety features are maintained.
DANGEROUS PROCEDURE WARNINGS
Warnings, such as example below, precede potentially dangerous procedures throughout this manual. Instructions contained in the warnings must be followed.
WARNING Dangerous Voltage, capable of causing death, are present in this instrument.
Use extreme caution when handling, testing, and adjusting.
Safety Symbols
The general definitions of safety symbols used on equipment or in manuals are listed below.
Instruction manual symbol: the product will be marked with this symbol when it is necessary for the user to refer to the instruction manual in order to protect against damage to the instrument.
Indicates dangerous voltage and potential for electrical shock. Do not touch terminals that have this symbol when instrument is on.
Protective conductor terminal. For protection against electrical shock in case of a fault. Used with field wiring terminals to indicate the terminal which must be connected to ground before operating equipment.
Frame or chassis terminal. A connection to the frame (chassis) of the equipment which normally includes all exposed metal structures.
Indicates earth (ground) terminal.
Alternating current.
Direct current.
ON (Supply).
Agilent 4155C/4156C User’s Guide Vol.2, Edition 1 5
OFF (Supply).
STANDBY (Supply).
CAT 1
Means INSTALLATION CATEGORY I. Measurement terminals on the rear panel comply with INSTALLATION CATEGORY I.
WARNING The warning sign denotes a hazard. It calls attention to a procedure, practice,
condition or the like, which, if not correctly performed or adhered to, could result in injury or death to personal.
CAUTION The caution sign denotes a hazard. It calls attention to an operating procedure,
practice, condition or the like, which, if not correctly performed or adhered to, could result in damage to or destruction of part or all of the product.
Herstellerbescheinigung
GEÄUSCHEMISSION
Lpa < 70 dB
am Arbeitsplatz
normaler Betrieb
nach DIN 45635 T. 19
Manufacturer’s Declaration
ACOUSTIC NOISE EMISSION
Lpa < 70dB
operator position
normal operation
per ISO 7779
6 Agilent 4155C/4156C Users Guide Vol.2, Edition 1
Printing History
Edition 1: January 2001
Agilent 4155C/4156C User’s Guide Vol.2, Edition 1 7
In This Manual
This manual provides information for measurement and analysis functions of Agilent 4155C/4156C, and consists of the following chapters:
Sweep Measurements
Describes how to perform sweep measurements.
Knob Sweep Measurements
Describes how to perform knob sweep measurements.
Sampling Measurements
Describes how to perform sampling measurements.
Quasi-static C-V Measurements
Describes how to perform quasi-static C-V measurements.
Stress Force
Describes how to perform stress force. Stress function is effective for the reliability testing of semiconductor devices.
Analyzing Measurement Results
Describes how to analyze measurement results manually and automatically.
Measurement Units and Functions
Provides information about measurement units and measurement functions.
Measurement Support Functions
Provides information about several functions which support measurements.
Built-in Functions
Provides information about the built-in mathematical functions and read out functions.
Connecting Measurement Devices
Explains how to connect the measurement devices to the instrument.
8 Agilent 4155C/4156C Users Guide Vol.2, Edition 1
Contents
1. Sweep Measurements
Measurement Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Basic Sweep Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
Subordinate Sweep Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
Synchronous Sweep Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-9
Pulse Sweep Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-11
Defining Measurement Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-14
To Define Measurement Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-16
To Set up Primary Sweep . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-18
To Set up Secondary Sweep . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-19
To Set up Synchronous Sweep . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-20
To Set up Constant Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-21
To Set up SMU Pulsed Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-22
To Set up PGU Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-24
To Set up Stop Condition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-26
To Display Graphics Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-27
To Display List Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-28
To Execute or Stop Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-29
Making a Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-30
Step 1. Prepare for the measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-31
Step 2. Mount your DUT on the test fixture . . . . . . . . . . . . . . . . . . . . . . . . . . 1-33
Step 3. Define the channel assignments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-35
Step 4. Define the user functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-37
Step 5. Set up the measurement parameters . . . . . . . . . . . . . . . . . . . . . . . . . . 1-39
Step 6. Set up the results display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-41
Step 7. Execute the measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-43
2. Knob Sweep Measurements
Measurement Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Available units and functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Agilent 4155C/4156C User’s Guide Vol.2, Edition 1 Contents - 1
Contents
Normal Sweep and Knob Sweep Measurements . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Features of Knob Sweep Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
KNOB SWEEP screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
Analysis of the Knob Sweep Measurement Results . . . . . . . . . . . . . . . . . . . . . 2-9
Executing Measurements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10
To Define Measurement Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-11
To Execute Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-12
To Stop Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-13
To Change Measurement Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-14
3. Sampling Measurements
Measurement Functions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Available Units. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Sampling Interval and Measurement Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Sampling Measurement Data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Sampling Completion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
Source Output Sequence and Time Origin. . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
Linear Sampling Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-13
Thinned-out Sampling Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-16
Logarithmic Sampling Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-19
Defining Measurement Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-24
To Define Measurement Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-26
To Set up Sampling Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-27
To Set up Constant Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-29
To Set up PGU Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-30
To Define Stop Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-32
To Display Graphics Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-33
To Display List Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-34
To Execute or Stop Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-35
Making a Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-36
Contents - 2 Agilent 4155C/4156C Users Guide Vol.2, Edition 1
Contents
Step 1. Prepare for the measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-36
Step 2. Mount your DUT on the test fixture . . . . . . . . . . . . . . . . . . . . . . . . . . 3-36
Step 3. Define the channel assignments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-37
Step 4. Set up the measurement parameters . . . . . . . . . . . . . . . . . . . . . . . . . . 3-38
Step 5. Set up the results display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-40
Step 6. Execute the measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-41
4. Quasi-static C-V Measurements
Measurement Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Available Units. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Capacitance Data Calculation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6
Defining Measurement Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8
To Define Measurement Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-10
To Set up QSCV Sweep Source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-11
To Set up Stop Condition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-12
To Set up Constant Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-13
To Set up PGU Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-13
To Define Measurement Conditions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-14
To Display Graphics Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-15
To Display List Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-16
To Execute or Stop Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-17
Making a Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-18
Step 1. Prepare for measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-18
Step 2. Mount your DUT on the test fixture . . . . . . . . . . . . . . . . . . . . . . . . . . 4-18
Step 3. Define the channel assignments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-19
Step 4. Set up the source parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-20
Step 5. Set up the measurement parameters . . . . . . . . . . . . . . . . . . . . . . . . . . 4-22
Step 6. Set up the results display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-24
Step 7. Execute the measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-25
Agilent 4155C/4156C User’s Guide Vol.2, Edition 1 Contents - 3
Contents
Maximum Measurement Value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-26
Considering Measurement Accuracy. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-28
5. Stress Force
Stress Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Stress Output Channels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Stress Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
Stress Force Sequence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7
Stress Stop Function at Abnormal Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-10
Defining Stress Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-11
To Set up Stress Source Channels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-13
To Set up Stress Condition/Timing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-15
To Set up ac (Pulse) Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-17
To Set up dc Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-19
To Force Stress . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-20
6. Analyzing Measurement Results
Analysis Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
Marker on the GRAPH/LIST: GRAPHICS screen . . . . . . . . . . . . . . . . . . . . . . 6-4
Marker on the GRAPH/LIST: LIST screen . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-6
Cursor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-8
Line Drawing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-9
Scaling Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-11
Overlay Display Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-12
Automatic Analysis Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-12
Manual Analysis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-13
To Specify a Measurement Point on Curve . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-14
To Specify between Measurement Points on Curve . . . . . . . . . . . . . . . . . . . . 6-16
To Display or Move Cursor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-18
To Adjust Display Range to Measurement Curve Automatically . . . . . . . . . . 6-19
Contents - 4 Agilent 4155C/4156C Users Guide Vol.2, Edition 1
Contents
To Zoom the Display Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-19
To Center Display at Cursor Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-20
To Draw Line through Two Specified Points . . . . . . . . . . . . . . . . . . . . . . . . . 6-20
To Draw Line through Specified Point with Specified Gradient . . . . . . . . . . 6-22
To Draw Tangent to Specified Point of Measurement Curve . . . . . . . . . . . . . 6-24
To Draw Regression Line for Specified Region . . . . . . . . . . . . . . . . . . . . . . . 6-26
To Display and Select a Line . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-28
To Display Grid on the Graph . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-28
To Change Data Variable on Graph . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-29
To Change Range of X or Y Axis Scale . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-30
To Change Variable Assigned to X, Y1, or Y2 Axis . . . . . . . . . . . . . . . . . . . . 6-31
To Overlay an Internal Memory Measurement Curve onto Plotting Area . . . 6-32
To Scroll the LIST screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-34
To Display or Move Marker on LIST screen . . . . . . . . . . . . . . . . . . . . . . . . . 6-35
To Change Variables of LIST screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-36
Automatic Analysis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-37
To Draw Line by Specifying Two Points . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-38
To Draw Line by Specifying Gradient and One Point . . . . . . . . . . . . . . . . . . 6-40
To Draw Tangent to Specified Measurement Point . . . . . . . . . . . . . . . . . . . . . 6-42
To Draw Regression Line by Specifying Two Points . . . . . . . . . . . . . . . . . . . 6-44
To Display Marker at Specified Point . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-47
7. Measurement Units and Functions
Measurement Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-3
GNDU - Ground Unit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-3
SMU - Source Monitor Unit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-4
VSU - Voltage Source Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-17
VMU - Voltage Monitor Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-18
PGU - Pulse Generator Unit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-20
Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-23
Voltage and Current Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-23
Agilent 4155C/4156C User’s Guide Vol.2, Edition 1 Contents - 5
Contents
Power Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-27
Measurement Ranging Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-29
Auto Ranging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-30
Limited Auto Ranging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-31
Compliance Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-32
Fixed Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-32
Measurement Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-33
Integration Time. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-33
Overhead Time. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-36
To Specify Source Output Time. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-36
SMU Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-37
Self-calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-38
Zero Offset Cancel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-39
To Measure Offset Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-39
To Perform Offset Cancel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-41
QSCV Zero Offset Cancel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-42
To Measure Offset Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-42
To Perform Offset Cancel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-44
Operation States . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-45
Idle State. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-45
Measurement State . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-45
Stress Force State. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-46
Standby State . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-46
Changing among Operation States . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-46
Output Sequence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-48
Sequential Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-49
Simultaneous Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-51
Contents - 6 Agilent 4155C/4156C Users Guide Vol.2, Edition 1
Contents
8. Support Functions
User Function and User Variable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-3
Output or Measurement Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-3
User Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-5
User Variable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-6
Syntax of Data Variable Name . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-8
Expression . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-9
Standby Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-13
Standby Channels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-13
Standby State . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-13
Available Units and Output Values. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-14
Output Values of non-Standby Channels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-14
Output Sequence of Standby Channels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-15
When Getting Setup File . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-16
To Use Standby Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-17
R-BOX Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-18
Resistance Value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-18
Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-18
Setups . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-20
Circuit Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-21
To Use R-Box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-23
SMU/PG Selector Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-25
Setup and Switching Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-26
Restrictions using Selector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-26
To Use Selector. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-27
Switching Matrix Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-28
Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-28
To Control Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-31
To Control Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-36
To Use Matrix Setup File . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-40
Agilent 4155C/4156C User’s Guide Vol.2, Edition 1 Contents - 7
Contents
Trigger Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-41
Trigger Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-43
Trigger Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-44
9. Built-in Functions
Built-in Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-3
ABS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-4
AT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-4
AVG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-4
COND . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-5
DELTA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-6
DIFF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-7
EXP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-8
INTEG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-8
LGT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-9
LOG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-9
MAVG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-10
MAX . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-11
MIN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-11
SQRT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-11
Read Out Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-12
@CX . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-13
@CY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-13
@CY1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-13
@CY2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-13
@IX . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-14
@IY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-14
@IY1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-15
@IY2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-15
@L1CO . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-16
@L1G . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-16
Contents - 8 Agilent 4155C/4156C Users Guide Vol.2, Edition 1
Contents
@L1G1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-17
@L1G2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-18
@L1X . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-19
@L1Y . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-19
@L1Y1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-19
@L1Y2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-19
@L2CO . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-19
@L2G . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-20
@L2G1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-21
@L2G2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-22
@L2X . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-23
@L2Y . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-23
@L2Y1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-23
@L2Y2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-23
@MI . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-23
@MX . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-24
@MY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-24
@MY1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-24
@MY2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-24
10. Connecting Measurement Devices
Using Test Fixture . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-3
Using Connector Plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-5
To Reduce Leakage Current . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-5
To Measure Low Resistance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-7
Agilent 4155C/4156C User’s Guide Vol.2, Edition 1 Contents - 9
Contents
Contents - 10 Agilent 4155C/4156C Users Guide Vol.2, Edition 1

1 Sweep Measurements

Agilent 4155C/4156C User’s Guide Vol.2, Edition 1
Sweep Measurements
This chapter consists of the following sections which describes how to execute a sweep measurement:
•“Measurement Functions”
•“Defining Measurement Conditions”
•“Making a Measurement”
For details about the measurement setup screens, see Setup Screen Reference manual.
1-2 Agilent 4155C/4156C Users Guide Vol.2, Edition 1
Sweep Measurements

Measurement Functions

Measurement Functions
For sweep measurements, the sweep source channels perform staircase sweep output of voltage or current, while the monitor channels measure voltage or current for each sweep step.
The 4155C/4156C provides three types of sweep measurement:
•“Basic Sweep Measurement
One sweep source (VAR1) is used.
•“Subordinate Sweep Measurement
A primary (VAR1) and secondary sweep source (VAR2) are used.
•“Synchronous Sweep Measurement
A primary (VAR1) and synchronous sweep source (VAR1') are used.
Also, you can set up a combined subordinate and synchronous sweep measurement.
Available Units SMUs and VSUs can be sweep sources (VAR1, VAR2, and VAR1'). A sweep or
constant source SMU can be set to the pulsed source to prevent thermal drift of the DUT.
Output Range Sweep source units use the following output range:
minimum range covers all sweep output values in Linear sweep.
minimum range covers all sweep output values in Log voltage sweep.
optimum range for output value every sweep step in Log current sweep.
Agilent 4155C/4156C User’s Guide Vol.2, Edition 1 1-3
Sweep Measurements Measurement Functions

Basic Sweep Measurement

Basic sweep measurement uses one sweep source (VAR1).
The following sweep types are available:
LIN/LOG
Linear staircase
Logarithmic staircase
SWEEP MODE
Single Source channel sweeps the output from user specified start value to
stop value.
Double Source channel sweeps the output from user specified start value to stop value, then from stop value to start value.
You can select any combination of LIN/LOG and SWEEP MODE as shown in the following table:
1-4 Agilent 4155C/4156C Users Guide Vol.2, Edition 1
Figure 1-1 Basic Sweep Measurement
Sweep Measurements
Measurement Functions
To set up basic sweep measurement, select VAR1 function for desired SMU or VSU on CHANNELS: CHANNEL DEFINITION screen.
Parameters
Also, specify the following parameters for VAR1 on MEASURE: SWEEP SETUP screen.
Parameter Description
sweep mode Single or double sweep.
linear/log Linear or logarithmic sweep. For logarithmic sweep, select the
number steps in one decade as follows:
LOG10 10 steps in one decade.
LOG25 25 steps in one decade.
LOG50 50 steps in one decade.
Agilent 4155C/4156C User’s Guide Vol.2, Edition 1 1-5
Sweep Measurements Measurement Functions
start Start value of sweep. For logarithmic sweep, start must not be
zero. Allowable range of start depends on output range of sweep source. For output range of each measurement channel, refer to Chapter 7.
stop Stop value of single sweep or turning back value of double
sweep. For logarithmic sweep, stop must have same polarity as start, and must not be zero. Allowable range of stop depends on output range of sweep source. For output range of each measurement channel, refer to Chapter 7.
step For linear sweep, step is step increment of sweep. Number
of sweep steps is calculated from start, stop, and step. Calculated number of steps must be in range: 2 to 1001.
For logarithmic sweep, step is invalid. Number of sweep steps is calculated from start, stop, and number of steps in one decade, which is specified by log parameter. Calculated number of steps must be in range: 2 to 1001.
compliance Compliance value of sweep source. This parameter applies to
SMU only. Allowable range of compliance depends on the compliance range of sweep source. For the compliance range of each measurement channel, refer to Chapter 7.
power compliance (Optional) Power compliance value of sweep source. This
parameter applies to SMU only. Allowable range depends on power compliance range of sweep source. For details, refer to Chapter 7.
hold time Time required for DUT to settle after forcing start value.
Allowable range is 0 to 655.35s. Resolution: 10 ms.
delay time Time required for DUT to settle after stepping the output.
Allowable range: 0 to 65.535 s. Resolution: 100 ms
Refer to Setup Screen Reference for setting up these parameters.
1-6 Agilent 4155C/4156C Users Guide Vol.2, Edition 1

Subordinate Sweep Measurement

For subordinate sweep measurement, you set up a secondary sweep source (VAR2) in addition to a primary sweep source (VAR1). After primary sweep is completed, the output of secondary sweep source is incremented or decremented by the specified step value, then the primary sweep source is swept again.
Figure 1-2 Subordinate Sweep Measurement
Sweep Measurements
Measurement Functions
To set up the subordinate sweep measurement, select the following on CHANNELS: CHANNEL DEFINITION screen:
VAR1 function for desired primary sweep source (SMU or VSU).
VAR2 function for desired secondary sweep source (SMU or VSU).
Subordinate sweep measurement has the following restriction:
For the secondary sweep source, only single sweep mode and linear staircase mode are available.
Agilent 4155C/4156C User’s Guide Vol.2, Edition 1 1-7
Sweep Measurements Measurement Functions
Parameters
The parameters for primary sweep source (VAR1) are same as the parameters for sweep source of basic sweep measurement. For secondary sweep source (VAR2), specify the following parameters on MEASURE: SWEEP SETUP screen.
Parameter Description
start Start value of secondary sweep. Allowable range of start
step Step increment of secondary sweep.
number of steps Number of secondary sweep steps.Allowable range: 1 to 128.
NOTE Stop value
Stop value of secondary sweep is calculated from start, step, and number of steps. Allowable range of stop depends on the output range of secondary sweep source. For the output range of each measurement channel, refer to Chapter 7.
compliance Compliance value of secondary sweep source. This parameter
depends on the output range of secondary sweep source. For the output range of each measurement channel, refer to Chapter 7.
applies to SMU only. Allowable range of compliance depends on the compliance range of secondary sweep source. For the compliance range of each measurement channel, refer to Chapter 7.
power compliance (Optional) Power compliance value of secondary sweep source.
This parameter applies to SMU only. Allowable range of power compliance depends on the power compliance range of sweep source. For details, refer to Chapter 7.
1-8 Agilent 4155C/4156C Users Guide Vol.2, Edition 1

Synchronous Sweep Measurement

For synchronous sweep measurement, you set up a synchronous sweep source (VAR1') in addition to a primary sweep source (VAR1). The output of the synchronous sweep source is swept synchronously with the output of the primary sweep source at a constant offset value and ratio.
Figure 1-3 Synchronous Sweep Measurement
Sweep Measurements
Measurement Functions
To set up synchronous sweep measurement, select the following on CHANNELS: CHANNEL DEFINITION screen:
VAR1 function for desired primary sweep source (SMU or VSU).
VAR1' function for desired synchronous sweep source (SMU or VSU).
Synchronous sweep mode has the following restrictions:
For the following, VAR1' is always set to the same mode as VAR1:
linear/log staircase
single/double sweep mode
VAR1 and VAR1' must be same V/I output mode. For example, if VAR1 is set to
V mode, then VAR1' must be set to V or VPULSE mode.
Agilent 4155C/4156C User’s Guide Vol.2, Edition 1 1-9
Sweep Measurements Measurement Functions
Parameters
The parameters for primary sweep source (VAR1) are same as the parameters for sweep source of basic sweep measurement. For synchronous sweep source (VAR1'), specify the following parameters on MEASURE: SWEEP SETUP screen.
Parameter Description
offset Offset between outputs of primary and synchronous sweep
sources.
ratio Ratio between outputs of primary and synchronous sweep
sources.
compliance Compliance value of synchronous sweep source. This
parameter applies to SMU only. Allowable range of compliance depends on the compliance range of synchronous sweep source. For the compliance range of each measurement channel, refer to Chapter 7.
power compliance (Optional) Power compliance value of synchronous sweep
source. This parameter applies to SMU only. Allowable range of power compliance depends on the power compliance range of synchronous sweep source. For details, refer to Chapter 7.
The relationship between the output of primary and synchronous sweep sources is determined by the following equation:
synchronous output = primary output ´ ratio + offset
The synchronous output determined by above equation must not exceed the output range of synchronous sweep source.
1-10 Agilent 4155C/4156C Users Guide Vol.2, Edition 1

Pulse Sweep Measurement

For a sweep measurement, a sweep or constant source SMU can be a pulse source. But only one SMU can be a pulse source. Figure 1-4 shows the relationship between pulse source and other sources.
Figure 1-4 Pulse Source and Other Sources
Sweep Measurements
Measurement Functions
For the pulse sweep measurement, the delay time of the primary sweep source is ignored, and each step of the primary sweep source is synchronized with output of the SMU pulse source. Measurements are made during the pulse output.
The pulse output of PGU is not synchronized with any other source.
Agilent 4155C/4156C User’s Guide Vol.2, Edition 1 1-11
Sweep Measurements Measurement Functions
Figure 1-5 SMU Pulse
Parameters
Specify SMU pulse parameters (MEASURE: SWEEP SETUP):
Parameter Description
pulse period SMU forces the next pulse after specified pulse period.
Allowable range: 5 ms to 1 s. Resolution: 100 ms.
pulse width Time from when SMU output starts to change from base value
to time when SMU starts to return from peak value. Measurements are made while the peak value is output. Allowable range: 0.5 ms to 100 ms. Resolution: 100 ms.
base value The base output value of the SMU pulse.
Be aware that if any of following are true, pulsed SMU channel may not output the pulse period and pulse width you specified:
Measurement range differs from compliance range (lowest range that includes compliance).
Ranging mode is set to auto range or limited auto range.
Multi-channel measurement is set.
1-12 Agilent 4155C/4156C Users Guide Vol.2, Edition 1
NOTE Pulse width
If the measurement settings do not meet the following conditions, pulse width setting of SMU may be insufficient to make measurement. If so, the pulse width is automatically changed to be appropriate.
Number of Meas. Channels: 1
Integration Time: Short
Ranging Mode: Fixed
Sweep Measurements
Measurement Functions
Agilent 4155C/4156C User’s Guide Vol.2, Edition 1 1-13
Sweep Measurements Defining Measurement Conditions

Defining Measurement Conditions

This section describes the sweep measurement tasks.
The basic procedure to test your DUT is as follows:
1 Connecting your DUT to the 4155C/
4156C. See Chapter 10 for procedures.
2 Defining measurement mode and
CHANNELS: CHANNEL DEFINITION 01JAN29 10:57AM
*MEASUREMENT MODE
SWEEP
*CHANNELS
UNIT VNAME INAME MODE FCTN RESISTANCE
SMU1:HR VS IS COMMON CONST 0 ohm
SMU2:HR VG IG V VAR1'
SMU3:HR VD ID V VAR1
SMU4:HR
SMU5:HP 0 ohm
VSU1 -------
VSU2 -------
VMU1 ------- ----- ---- DISCHARGE
VMU2 ------- ----- ---- ON
PGU1 -------
PGU2 -------
GNDU ------- ----
SWEEP
Select Measurement Mode with softkey or rotary knob.
CHANNEL
DEF
MEASURE STBY SERIES
USER
USER
FCTN
VAR
E5250A
S
PROP
B
SWEEP
SAM-
QSCV
DEFAULT MEASURE SETUP
MEM1 M B-Tr VCE-IC
MEM2 M FET VDS-ID
MORE
1/2
NEXT
PAGE
measurement units that you use to make measurement.
PLING
See To Define Measurement Units on page 1-16. Also see Chapter 8 for the user function, standby function, and R-box
control.
3 Setting the source parameters of the units.
MEASURE: SWEEP SETUP 01JAN29 10:58AM
*VARIABLE VAR1 VAR2 VAR1'
UNIT SMU3:HR UNIT SMU2:HR
NAME VD NAME VG
SWEEP MODE SINGLE OFFSET 0.0000 V
LIN/LOG LINEAR RATIO 1.000
START 0.0000 V COMPLIANCE 100.00mA
STOP 2.0000 V POWER COMP OFF
STEP 20.0mV
NO OF STEP 101
COMPLIANCE 100.00mA
POWER COMP OFF
*TIMING
HOLD TIME 0.0000 s
DELAY TIME 0.0000 s *SWEEP STOP AT ANY ABNORM Status
*CONSTANT
UNIT
NAME
MODE
SOURCE --------- --------- --------- ---------
COMPLIANCE --------- --------- --------- ---------
SINGLE
Select Sweep Mode with softkey or rotary knob.
SWEEP
SETUP
MEASURE
SETUP
OUTPUT
SEQ
S
PREV
PAGE
SINGLE
DOUBLE
B
NEXT
PAGE
See following in this section:
•“To Set up Primary Sweep”
•“To Set up Secondary Sweep”
•“To Set up Synchronous Sweep”
•“To Set up Constant Output”
•“To Set up SMU Pulsed Output”
•“To Set up PGU Output”
•“To Set up Stop Condition”
1-14 Agilent 4155C/4156C Users Guide Vol.2, Edition 1
Sweep Measurements
Defining Measurement Conditions
4 Setting the display mode to show
DISPLAY: DISPLAY SETUP 01JAN29 10:58AM
*DISPLAY MODE
GRAPHICS
*GRAPHICS
Xaxis Y1axis Y2axis
NAME VD SQID DSQID
SCALE LINEAR LINEAR LINEAR
MIN 0.000000000 V 0.0000000000 0.0000000000
MAX 2.0000000 V 100.00000000m 100.00000000m
*GRID *LINE PARAMETER
ON ON
*DATA VARIABLES *DATA DISPLAY RESOLUTION
NORMAL
GRAPH-
LIST
measurement results.
ICS
See following in this section:
•“To Display Graphics Results”
•“To Display List Results”
GRAPHICS
Select Display Mode with softkey or rotary knob.
DISPLAY
SETUP
ANLYSIS
SETUP
S
PREV
PAGE
B
NEXT
PAGE
5 Executing the measurement.
See To Execute or Stop Measurement on page 1-29. Also see Chapter 7 for executing self-calibration and zero offset cancel.
6 Results.
GRAPH/LIST: GRAPHICS SHORT 01JAN29 10:59AM
() ()
100.m 100.m
SQID DSQID
10.0m 10.0m /div /div
0.00 0.00
0.00 VD (V) 200 .m /div 2.00
AXISY1MARKER/
LINE SCALING DISPLAY
CURSOR
SWEEP
SETUP
TIMING
SETUP
S
SETUP
B
MARKER
MARKER MIN/MAX
INTER-
MARKER SKIP
AUTO ANALY-
CURSOR
For example, displayed graphically.
OFF
POLATE
OFF
SIS
OFF
Agilent 4155C/4156C User’s Guide Vol.2, Edition 1 1-15
Sweep Measurements Defining Measurement Conditions

To Define Measurement Units

Press Chan front-panel key to define the measurement units. CHANNELS: CHANNEL DEFINITION screen is displayed.
1. MEASUREMENT MODE:
Select SWEEP secondary softkey for sweep measurement.
2. VNAME:
Enter a unique name for voltage variable. For example, enter Vce for collector-emitter voltage. If channel does neither V force nor V measurement, you can omit VNAME.
3. INAME:
Enter a unique name for current variable. For example, enter Ic for collector current. If channel does neither I force nor I measurement, you can omit INAME.
4. MODE:
Select one of the following softkeys:
V : Voltage output (for SMU, VSU, and PGU). V : Grounded mode voltage measurement (for VMU).
I : Current output (for SMU).
VPULSE : Pulsed voltage output (for SMU and PGU).
IPULSE : Pulsed current output (for SMU).
COMMON : Circuit common (for SMU and GNDU).
DVOLT : Differential mode voltage measurement (for VMU).
5. FCTN:
Select one of the following softkeys to set the output function:
CONST : Constant output function (for SMU, VSU, and PGU).
VAR1 : Primary sweep output function (for SMU and VSU).
VAR2 : Secondary sweep output function (for SMU and VSU).
VAR1' : Synchronous sweep output function (for SMU and VSU).
1-16 Agilent 4155C/4156C Users Guide Vol.2, Edition 1
Sweep Measurements
Defining Measurement Conditions
6. DISCHARGE:
Select ON secondary softkey to connect the discharge resistor to VMU input, or OFF to disconnect the resistor. The discharge resistor is used to prevent the VMU inputs from charge up in the idle state.
When DISCHARGE is ON, the discharge resistor is automatically connected to the VMU input, and disconnected from the input in the measurement state.
VNAME and INAME
You can use VNAME and INAME names in user function definitions or for analysis on the GRAPHICS/LIST screens. These names must be 6 or less alphanumeric characters. First character must be alphabet character.
To disable a unit
Select DELETE ROW softkey. The settings for the unit are deleted.
Agilent 4155C/4156C User’s Guide Vol.2, Edition 1 1-17
Sweep Measurements Defining Measurement Conditions

To Set up Primary Sweep

Primary sweep source is the measurement unit defined as VAR1 in the CHANNELS: CHANNEL DEFINITION screen. To set up the primary sweep source, press displayed.
1. VAR1 : SWEEP MODE
Select one of the following softkeys to set the sweep mode:
SINGLE : single sweep mode.
DOUBLE : double sweep mode.
2. VAR1 : LIN/LOG
Select one of the following softkeys to set the sweep step mode:
LINEAR : linear step mode.
LOG XX : logarithmic step mode. XX specifies the number of steps per
3. VAR1 : START
Meas front-panel key. The MEASURE: SWEEP SETUP screen is
decade. XX is 10, 25, or 50.
Enter the sweep start value.
4. VAR1 : STOP
Enter the sweep stop value.
If you select LOG sweep step mode, the polarity of stop value must be same as the polarity of start value.
5. VAR1 : STEP
If you select LINEAR sweep step mode, enter the sweep step value. This field is not available for the LOG mode.
6. VAR1 : COMPLIANCE, POWER COMPLIANCE
Only for SMU. Enter the compliance value, and power compliance (ON or OFF) for the primary sweep source.
NO OF STEP is automatically calculated from START, STOP, and STEP values.
You cannot change UNIT and NAME in this screen. To change the values, go to the CHANNELS: CHANNEL DEFINITION screen.
1-18 Agilent 4155C/4156C Users Guide Vol.2, Edition 1
Sweep Measurements
Defining Measurement Conditions

To Set up Secondary Sweep

Secondary sweep source is the measurement unit defined as VAR2 in the CHANNELS: CHANNEL DEFINITION screen. On the MEASURE: SWEEP SETUP screen, set up the primary sweep source (VAR1), then do following:
1. VAR2: START
Enter the secondary sweep start value.
2. VAR2: STEP
Enter the secondary sweep step value.
3. VAR2: NO OF STEP
Enter the number of steps for the secondary sweep.
4. VAR2 : COMPLIANCE, POWER COMPLIANCE
Only for SMU. Enter the compliance value, and power compliance (ON or OFF) for the secondary sweep source.
SWEEP MODE and LIN/LOG are automatically set to SINGLE and LINEAR, respectively. You cannot change the settings.
The STOP value is automatically calculated from START, STEP, and NO OF STEP values.
You cannot change UNIT and NAME in this screen. To change the values, go to the CHANNELS: CHANNEL DEFINITION screen.
Agilent 4155C/4156C User’s Guide Vol.2, Edition 1 1-19
Sweep Measurements Defining Measurement Conditions

To Set up Synchronous Sweep

Synchronous sweep source is the measurement unit defined as VAR1 in the CHANNELS: CHANNEL DEFINITION screen. VAR1 is available for the measurement units set to the output mode same as the VAR1 output mode. The output value of VAR1' is calculated by the following equation:
VA R1 ' = VAR 1 ´ RATIO + OFFSET
On the MEASURE: SWEEP SETUP screen, set up the primary sweep source (VAR1), then do following:
1. VAR1 : OFFSET
Enter the offset value of the VAR1’ output against the VAR1 output.
2. VAR1 : RATIO
Enter the ratio value of the VAR1 output against the VAR1 output.
3. VAR1 : COMPLIANCE, POWER COMPLIANCE
Only for SMU. Enter the compliance value, and power compliance (ON or OFF) for the synchronous sweep source.
You cannot change UNIT and NAME in this screen. To change the values, go to the CHANNELS: CHANNEL DEFINITION screen.
1-20 Agilent 4155C/4156C Users Guide Vol.2, Edition 1
Sweep Measurements
Defining Measurement Conditions

To Set up Constant Output

Constant voltage/current source is the measurement unit defined as CONST in the CHANNELS: CHANNEL DEFINITION screen. To set up the constant output source, press displayed.
1. CONSTANT : SOURCE
Enter the desired output value of the constant source.
2. CONSTANT : COMPLIANCE
Only for SMU. Enter the compliance value for the constant source.
You cannot change UNIT, NAME, and MODE in this screen. To change the values, go to the CHANNELS: CHANNEL DEFINITION screen.
Meas front-panel key. The MEASURE: SWEEP SETUP screen is
Agilent 4155C/4156C User’s Guide Vol.2, Edition 1 1-21
Sweep Measurements Defining Measurement Conditions

To Set up SMU Pulsed Output

SMU pulse output source is the measurement unit defined as VPULSE or IPULSE in the CHANNELS: CHANNEL DEFINITION screen. For pulsed sweep source, set the function (FCTN) to VAR1, VAR2, or VAR1. For pulsed constant source, set the function to CONST. To set up the SMU pulse output source, press key. The MEASURE: SWEEP SETUP screen is displayed.
1. Set the source output parameters:
For VAR1, refer to “To Set up Primary Sweep” on page 1-18.
For VAR2, refer to “To Set up Secondary Sweep” on page 1-19.
For VAR1’, refer to To Set up Synchronous Sweep” on page 1-20.
For CONST, refer to “To Set up Constant Output” on page 1-21.
2. SMU PULSE: PERIOD
Enter the pulse period value.
3. SMU PULSE: WIDTH
Enter the pulse width value.
Meas front-panel
4. SMU PULSE: BASE
Enter the pulse base value.
You cannot change UNIT and NAME in this screen. To change the values, go to the CHANNELS: CHANNEL DEFINITION screen.
1-22 Agilent 4155C/4156C Users Guide Vol.2, Edition 1
Sweep Measurements
Defining Measurement Conditions
Pulse Parameters
The relation between the PERIOD, WIDTH, and BASE values are as shown in the following figures.
SMU outputs the pulses as shown in figure (a) or figure (b).
Figure (a)
When the function (FCTN) is set to VAR1, VAR2, or VAR1’.
The pulse peak values are the sweep output values calculated from the sweep start, stop, step values, and so on.
Figure (b)
When the function (FCTN) is set to CONST.
The pulse peak value is the output value of the constant output source.
Agilent 4155C/4156C User’s Guide Vol.2, Edition 1 1-23
Sweep Measurements Defining Measurement Conditions

To Set up PGU Output

1. Define PGU to be VPULSE and CONST as described in “To Define Measurement Units on page 1-16.
2. Press
3. Select PGU SETUP primary softkey.
4. In the PERIOD field of PGU1, enter the pulse period value.
5. In the WIDTH field of desired PGU column, enter the pulse width value.
6. In the DELAY TIME field of desired PGU column, enter delay time value.
7. In the PEAK VALUE field of desired PGU column, enter pulse peak value.
8. In the BASE VALUE field of desired PGU column, enter pulse base value.
9. In the LEADING TIME field of desired PGU column, enter the leading-edge
10. In the TRAILING TIME field of desired PGU column, enter the trailing-edge
11. In the IMPEDANCE field of desired PGU column, select:
12. In the PULSE COUNT field, do one of the following:
Meas key in the PAGE CONTROL key group.
transition time.
transition time.
LOW secondary softkey for approximately zero ohm output impedance.
50 ohm secondary softkey for 50 ohm output impedance.
Select FREE RUN secondary softkey to force the pulse continuously.
Or enter the number of pulses to output (for sampling measurement only).
For the pulse period and pulse count values, the values you set for PGU1 are also used for PGU2. The following figure shows the relation between pulse waveform and setup parameters.
1-24 Agilent 4155C/4156C Users Guide Vol.2, Edition 1
Sweep Measurements
Defining Measurement Conditions
To modify the UNIT and NAME fields
Modify the UNIT and NAME fields on the CHANNELS: CHANNEL DEFINITION screen.
Using PGUs as constant voltage source
To use a PGU as a constant voltage source, set the desired PGU as follows:
V in MODE column on the CHANNEL DEFINITION screen
Desired output voltage value in SOURCE field on MEASURE: PGU SETUP
screen.
Agilent 4155C/4156C User’s Guide Vol.2, Edition 1 1-25
Sweep Measurements Defining Measurement Conditions

To Set up Stop Condition

1. Press Meas key in the PAGE CONTROL key group.
2. Move field pointer to SWEEP Status field.
3. Select one of the following softkeys:
CONT AT ANY Sweep will continue even if an abnormal status occurs. If
power compliance is set for an SMU, this softkey is not displayed.
STOP AT ANY ABNORM Sweep will stop if any abnormal status occurs.
STOP AT COMPLIANCE Sweep will stop only if SMU reaches its compliance setting.
This is automatically set when power compliance is set for SMUs, or when 10k ohm, 100k ohm, or 1M ohm is selected in the SERIES RESISTANCE field.
Abnormal status means the following:
SMU reaches its compliance setting.
Current of VSU exceeds approximately ±100 mA.
SMU or VSU oscillates.
A/D converter overflow occurs.
Average current of PGU exceeds ±100 mA.
1-26 Agilent 4155C/4156C Users Guide Vol.2, Edition 1
Sweep Measurements
Defining Measurement Conditions

To Display Graphics Results

1. Press Display key in the PAGE CONTROL key group.
2. Select DISPLAY SETUP primary softkey.
3. In the DISPLAY MODE field, select GRAPHICS secondary softkey.
4. In the X axis column, enter variable name, select axis scale, and enter minimum and maximum values.
5. In the Y1 axis column, enter variable name, select axis scale, and enter minimum and maximum values.
6. If you use Y2 axis, enter variable name, select axis scale, and enter minimum and maximum values in Y2 axis column.
When the pointer is in the NAME row, the allowable variable names appear in the secondary softkey area. To set a variable name, select the desired secondary softkey. The allowable names are names that you already set up on the CHANNEL DEFINITION, USER FUNCTION, and USER VARIABLE screens.
To display a grid on the plotting area
In the GRID field, select ON to display or OFF to not display.
To display line parameters on the plotting area
In the LINE PARAMETER field, select ON to display or OFF to not display. Line parameters are the X and Y intercepts and gradient of the analysis lines.
To display variables on the data variable area
In the DATA VARIABLES fields, select secondary softkey for desired variable. Up to 2 variables can be displayed on the data variable area which is located between the graph area and the marker area.
To change resolution of the data to be displayed
In the DATA DISPLAY RESOLUTION field, select NORMAL for the resolution shown in the specifications, or EXTEND for the ADC full scale resolution. For example, the data display resolution is 1 fA in NORMAL mode, and 10 aA in EXTEND mode for 10 pA measurement range.
Agilent 4155C/4156C User’s Guide Vol.2, Edition 1 1-27
Sweep Measurements Defining Measurement Conditions

To Display List Results

1. Press Display key in the PAGE CONTROL key group.
2. Select DISPLAY SETUP primary softkey.
3. In the DISPLAY MODE field, select LIST secondary softkey.
4. In the LIST area, select the secondary softkey of the variables for which you want to list the measurement results.
When the pointer is in the NAME row, the allowable variable names appear in the secondary softkey area. To set a variable name, select the desired secondary softkey. The allowable names are names that you already set up on CHANNEL DEFINITION, USER FUNCTION, and USER VARIABLE screens.
To display variables on the data variable area
In the DATA VARIABLES fields, select secondary softkey for desired variable. Up to 2 variables can be displayed on the data variable area which is located below the list area.
To change resolution of the data to be displayed
In the DATA DISPLAY RESOLUTION field, select NORMAL for the resolution shown in the specifications, or EXTEND for the ADC full scale resolution. For example, the data display resolution is 1 fA in NORMAL mode, and 10 aA in EXTEND mode for 10 pA measurement range.
1-28 Agilent 4155C/4156C Users Guide Vol.2, Edition 1
Sweep Measurements
Defining Measurement Conditions

To Execute or Stop Measurement

To execute a measurement, press:
Single key in the MEASUREMENT key group for single measurement.
Repeat key in the MEASUREMENT key group for repeat measurement.
Append key in the MEASUREMENT key group for append measurement.
To stop a measurement, press
Sto p key in the MEASUREMENT key group.
Single, Repeat, and Append Measurement
There are three measurement execution modes as follows:
Single measurement Clears GRAPHICS or LIST screen, then executes measurement
one time. Measurement results are displayed on GRAPHICS or LIST screen.
Repeat measurement Executes measurements continuously. Before each
measurement is executed, the GRAPHICS or LIST screen is cleared. Most recent measurement results are displayed on GRAPHICS or LIST screen.
Append measurement Executes measurement one time. Does not clear GRAPHICS or
LIST screen. That is, measurement results are added to the existing results.
Agilent 4155C/4156C User’s Guide Vol.2, Edition 1 1-29
Sweep Measurements Making a Measurement

Making a Measurement

In this section, you learn how to execute the measurements with an 4155C/4156C and to display the measurement results graphically. Id-Vg measurement of a MOS FET is provided as an example. You learn step-by-step how to perform this measurement.
You measure the device under test (DUT) by using the measurement circuit as shown in the following diagram. SMU2 and SMU3 sweep the same voltage to the gate and drain. SMU3 measures the drain current (Id). The source and substrate are connected to circuit common.
You should get result similar to the following figure. Gate voltage Vg (swept from 0 V to 2 V) is assigned to X axis, ÖId assigned to Y2 axis.
is assigned to Y1 axis, and ¶ÖId / ¶Vg is
NOTE For accurate measurements, allow the 4155C/4156C to warm-up for a minimum of
40 minutes after you turn on the instrument, and then execute calibration. For the self-calibration, see Chapter 7.
1-30 Agilent 4155C/4156C Users Guide Vol.2, Edition 1
Sweep Measurements
Making a Measurement

Step 1. Prepare for the measurement

Before executing measurement, configure the 4155C/4156C and accessories.
1. Make sure that the 4155C/4156C is off.
2. Connect the 16442A test fixture to the 4155C/4156C. See figure below.
3. If you use the keyboard, connect it to the 4155C/4156C.
Connection between 4156C and 16442A:
16442A
Kelvin Triaxial Cables
Coaxial Cables
Interlock/LED Cable
4156
4156C cable 16442A 4156C cable 16442A
Intlk
SMU 1
SMU 2
SMU 3
Interlock/LED
Kelvin triaxial
Kelvin triaxial
Kelvin triaxial
a
Intlk VSU 1
c
SMU 1 VSU 2
c
SMU 2 VMU 1
c
SMU 3 VMU 2
Coaxial
Coaxial
Coaxial
Coaxial
b
VSU 1
b
VSU 2
b
VMU 1
b
VMU 2
a. Interlock/LED cable: Agilent 16493J b. Coaxial cable: Agilent 16493B. You do not need to connect VSUs and
VMUs for this measurement.
c. Kelvin triaxial cable: Agilent 16493K
Agilent 4155C/4156C User’s Guide Vol.2, Edition 1 1-31
Sweep Measurements Making a Measurement
Connection between 4155C and 16442A:
Triaxial Cables
16442A
Coaxial Cables
Interlock/LED Cable
4155
4155C cable 16442A 4155C cable 16442A
a
Intlk
SMU 1
Interlock/LED
Triaxial
b
Intlk
SMU 1
VSU 1
Coaxial
c
VSU 1
(blue label)
SMU 2
Triaxial
b
SMU 2
VSU 2
Coaxial
c
VSU 2
(blue label)
SMU 3
Triaxial
b
SMU 3
VMU 1
Coaxial
c
VMU 1
(blue label)
SMU 4
Triaxial
b
SMU 4
VMU 2
Coaxial
c
VMU 2
(blue label)
a. Interlock/LED cable: Agilent 16493J b. Triaxial cable: Agilent 16493C. You do not need to connect SMU4 for
this measurement.
c. Coaxial cable: Agilent 16493B. You do not need to connect VSUs and
VMUs for this measurement.
1-32 Agilent 4155C/4156C Users Guide Vol.2, Edition 1
Sweep Measurements
Making a Measurement

Step 2. Mount your DUT on the test fixture

1. Select a suitable socket module for your DUT.
2. Mount the socket module on the test fixture.
3. Mount your DUT on the socket module.
4. Make connections with four connection cables (miniature banana to pin plug).
You make the following connections:
Source to SMU1
Gate to SMU2
Drain to SMU3
Substrate to SMU1
Both the source and substrate terminals are connected to SMU1.
5. After finishing connections, shut the lid of the test fixture.
Agilent 4155C/4156C User’s Guide Vol.2, Edition 1 1-33
Sweep Measurements Making a Measurement
Wiring for the 4156C
For this measurement, non-Kelvin connections are used. So, connect only the force terminals as shown in the following figure:
Wiring for the 4155C
1-34 Agilent 4155C/4156C Users Guide Vol.2, Edition 1
Sweep Measurements
Making a Measurement

Step 3. Define the channel assignments

You set the connection information on the CHANNELS: CHANNEL DEFINITION screen.
1. Turn on the 4155C/4156C. Self-test starts.
2. After self-test is finished, make sure that CHANNELS: CHANNEL DEFINITION screen appears on the screen of the 4155C/4156C. If not, press
Chan front-panel key.
where, softkeys located bottom of screen are the primary softkeys, and softkeys located right side of screen are the secondary softkeys.
3. Make sure that SWEEP is displayed in the MEASUREMENT MODE field. If not, select SWEEP secondary softkey in the MEASUREMENT MODE field.
4. Set the connection information in the CHANNELS area as shown in the following table. For the actual screen, see Figure 1-6.
on Front Panel on Keyboard
To move the pointer, use arrow keys of
MARKER/CURSOR area.
To move the cursor to edit in display area,
To enter "VS" in VNAME field,
To enter "IS" in INAME field,
To set "V" in
use arrow keys of Edit area.
press
. (period) and + keys,
then
Enter.
* and + keys,
press then
Enter.
select V secondary softkey. press
MODE field,
To set "VAR1" in FCTN field,
To set "VAR1" in FCTN field,
select VAR1' secondary softkey.
select VAR1 secondary softkey.
To disable a unit, select DISABLE UNIT
secondary softkey.
use arrow keys
use
Backspace key.
type VS, then press
type IS, then press
Shift-F1 keys.
press
Shift-F4 keys.
Shift-F2 keys.
press
Shift-F7 keys.
press
Enter.
Enter.
Agilent 4155C/4156C User’s Guide Vol.2, Edition 1 1-35
Sweep Measurements Making a Measurement
Figure 1-6 Channel Definition Screen
CHANNELS: CHANNEL DEFINITION 01JAN29 10:57AM
*MEASUREMENT MODE
SWEEP
*CHANNELS
UNIT VNAME INAME MODE FCTN RESISTANCE
SMU1:HR VS IS COMMON CONST 0 ohm
SMU2:HR VG IG V VAR1'
SMU3:HR VD ID V VAR1
SMU4:HR
SMU5:HP 0 ohm
VSU1 -------
VSU2 -------
VMU1 ------- ----- ---- DISC HARGE
VMU2 ------- ----- ---- ON
PGU1 -------
PGU2 -------
GNDU ------- ----
SWEEP
Select Measurement Mode wi th softkey or rotary kno b.
CHANNEL
DEF
USER
FCTN
MEASURE STBY SERIES
USER
VAR
S
E5250A
PROP
B
SWEEP
SAM-
PLING
QSCV
DEFAULT MEASURE SETUP
MEM1 M B-Tr VCE-IC
MEM2 M FET VDS-ID
MORE
1/2
NEXT
PAGE
1-36 Agilent 4155C/4156C Users Guide Vol.2, Edition 1
Sweep Measurements
Making a Measurement

Step 4. Define the user functions

You define the user functions on the CHANNELS: USER FUNCTION DEFINITION screen.
1. Select USER FCTN primary softkey. The CHANNELS: USER FUNCTION DEFINITION screen appears.
2. Enter the user function information as shown in the following table. For the actual screen, see Figure 1-7.
on Front Panel on Keyboard
To move the pointer, use arrow keys of
use arrow keys.
MARKER/CURSOR area.
To move the cursor to edit in display area,
To enter "SQID" in NAME field,
To enter "SQRT(ID)" in DEFINITION field,
To enter "DSQID" in NAME field,
To enter "DIFF(SQID,VG)"
in DEFINITION field,
use arrow keys of Edit area.
press
+, 2, *, / keys, then
Enter.
a
+, 2, 3, p, blue key, (
press keys, and ID secondary softkey, then
/, +, 2, *, / keys, then
press
Enter.
press
/, *, 7, 7, blue key, (,
b
blue key, +, 2, *, /, blue key, , (comma) keys, and
use
type SQID, then press Enter.
type SQRT(ID), then press
) and Enter.
type DSQID, then press Enter.
type DIFF(SQID,VG),
then press VG secondary softkey, then
) and Enter.
To disable a user function,
select DISABLE FUNCTION
press
secondary softkey.
a. Square root operator (Ö
) is defined by "SQRT" built-in function.
b. Partial difference (¶/¶) is defined by "DIFF" built-in function.
Backspace key.
Enter.
Enter.
Shift-F7 key.
Agilent 4155C/4156C User’s Guide Vol.2, Edition 1 1-37
Sweep Measurements Making a Measurement
Figure 1-7 User Function Definition Screen
CHANNELS: USER F UNCTION DEFINIT ION 01JAN29 10:53AM
*USER FUNCTION
NAME UNIT DEFINITION
SQID SQRT(ID)
DSQID DIFF(SQI D,VG)
SQID
Enter U ser Function Name. (max 6 chars.)
CHANNEL
DEF
USER
FCTN
USER
VAR
S
E5250A
PROP
PREV
PAGE
B
DELETE ROW
NEXT
PAGE
Figure 1-7 defines the following two user functions: SQID = ÖId DSQID = SQID / Vg = ¶ÖId / ¶Vg
Where, Id is drain current and Vg is gate voltage.
1-38 Agilent 4155C/4156C Users Guide Vol.2, Edition 1
Sweep Measurements
Making a Measurement

Step 5. Set up the measurement parameters

You set the output parameters on the MEASURE: SWEEP SETUP screen.
1. Press Meas front-panel key. The MEASURE: SWEEP SETUP screen appears.
In the screen, the upper left area defines the VAR1 information, and the upper right area defines the VAR1’ information. See Figure 1-8.
2. Set the VAR1 information as shown below:
on Front Panel on Keyboard
To move the pointer, use arrow keys of
use arrow keys.
MARKER/CURSOR area.
To set "SINGLE" in SWEEP MODE field,
To set "LINEAR" in LIN/LOG field,
To enter "2.000 V" in STOP field,
To enter "20.00 mV" in STEP field,
select SINGLE secondary softkey.
select LINEAR secondary softkey.
2, then Enter.type 2,
press
2, 0, m, then Enter.type 20m,
press
press
Shift-F1 keys.
press
Shift-F1 keys.
then press
then press
Enter.
Enter.
Drain voltage sweeps from 0 V to 2 V with 20 mV step. The current compliance is set to 100 mA.
3. Set the VAR1’ information as shown below:
on Front Panel on Keyboard
To enter "0.000 V" in
press
0, then Enter.type 0, then press Enter.
OFFSET field,
To enter "1.000" in
1, then Enter.type 1, then press Enter.
press
RATIO field,
To force the same voltage to the drain and gate, set RATIO = 1 and OFFSET = 0. Because VAR1 is defined as follows:
(VAR1 output) = RATIO ´ (VAR1 output) + OFFSET
Agilent 4155C/4156C User’s Guide Vol.2, Edition 1 1-39
Sweep Measurements Making a Measurement
Figure 1-8 Sweep Setup Screen
MEASURE: SWEEP S ETUP 01JAN29 10:58AM
*VARIABLE VAR1 VAR2 VAR1'
UNIT SMU3:HR UNIT SMU2:HR
NAME VD NAME VG
SWEEP M ODE SINGLE OFFSET 0.0000 V
LIN/LOG LINEAR RATIO 1.000
START 0.0000 V COMPLIANCE 100.00mA
STOP 2.0000 V POWER C OMP OFF
STEP 20.0mV
NO OF STEP 101
COMPLIANCE 100.00mA
POWER C OMP O FF
*TIMING
HOLD TI ME 0.0000 s
DELAY T IME 0.000 0 s *SWEEP STO P AT AN Y ABNORM Status
*CONSTANT
UNIT
NAME
MODE
SOURCE --------- --------- --------- ---------
COMPLIANCE --------- --------- --------- ---------
SINGLE
Select Sweep Mod e with softkey or rotary knob.
SWEEP
SETUP
MEASURE
SETUP
OUTPUT
SEQ
SINGLE
DOUBLE
B
S
PREV
PAGE
NEXT
PAGE
1-40 Agilent 4155C/4156C Users Guide Vol.2, Edition 1
Sweep Measurements
Making a Measurement

Step 6. Set up the results display

You set the results display information on the DISPLAY: DISPLAY SETUP screen.
1. Press Display front-panel key. The DISPLAY: DISPLAY SETUP screen appears.
2. Make sure GRAPHICS is displayed in the DISPLAY MODE field. If not, select GRAPHIC secondary softkey in the DISPLAY MODE field.
3. Set the X-, Y1-, and Y2-axes information as shown below. For the actual setup, see Figure 1-9.
on Front Panel on Keyboard
To enter "VG" in NAME field,
To set "LINEAR" in SCALE field,
To enter "0.00000 V" in MIN field,
To enter "2.00000 V" in MAX field,
To enter "SQID" in NAME field,
To enter "0.00000 A" in MIN field,
To enter "100.000mA" in MAX field,
To enter "DSQID" in NAME field,
select VG secondary
press
Shift-F3 keys.
softkey.
select LINEAR
press
Shift-F1 keys.
secondary softkey.
press
0, then Enter.type 0, then press Enter.
2, then Enter.type 2, then press Enter.
press
select MORE 1/2, then SQID secondary
Shift-F7 keys, then
press
Shift-F3 keys.
softkeys.
press
0, then Enter.type 0, then press Enter.
1, 0, 0, m, then Enter.type 100m, then press
press
Enter.
select MORE 1/2, then DSQID secondary
Shift-F7 keys, then
press
Shift-F4 keys.
softkeys.
Agilent 4155C/4156C User’s Guide Vol.2, Edition 1 1-41
Sweep Measurements Making a Measurement
Figure 1-9 Display Setup Screen
DISPLAY: DISPLAY SETUP 01JAN29 10:58AM
*DISPLAY MODE
GRAPHICS
*GRAPHICS
Xaxis Y1axis Y2axis
NAME VD SQID DSQID
SCALE LINEAR LINEA R LINEAR
MIN 0.000000000 V 0.00 00000000 0.0000000000
MAX 2.0000000 V 100.00000000m 100.0000 0000m
*GRID *LINE PARAMETER
ON ON
GRAPH-
ICS
LIST
*DATA VARI ABLES *DA TA DISPLAY RESOLUTION
GRAPHICS
Select Display M ode with softkey or rota ry knob.
DISPLAY
SETUP
ANLYSIS
SETUP
NORMAL
S
PREV
PAGE
B
NEXT
PAGE
1-42 Agilent 4155C/4156C Users Guide Vol.2, Edition 1
Sweep Measurements
Making a Measurement

Step 7. Execute the measurement

Press Single front-panel key to execute the measurement.
You will get measurement results as shown in the following example:
GRAPH/LIST: GRAPHICS S HORT 01JAN29 10:59AM
() ()
100.m 100.m
SQID DSQID
10.0m 10.0m /div /div
0.00 0.00
0.00 VD (V) 200.m /div 2.00
AXIS
Y1
MARKER/
CURSOR
LINE SCA LING DISPLAY
S
SETUP
SWEEP
SETUP
TIMING
SETUP
B
MARKER
OFF
MARKER MIN/MAX
INTER-
POLATE
OFF
MARKER SKIP
AUTO ANALY-
SIS
CURSOR
OFF
Agilent 4155C/4156C User’s Guide Vol.2, Edition 1 1-43
Sweep Measurements Making a Measurement
1-44 Agilent 4155C/4156C Users Guide Vol.2, Edition 1

2 Knob Sweep Measurements

Agilent 4155C/4156C User’s Guide Vol.2, Edition 1
Knob Sweep Measurements
This chapter explains how to execute knob sweep measurements.
The knob sweep function is useful in the following cases:
to determine a parameter value for normal sweep
to quickly make a rough measurement of a DUT characteristics
2-2 Agilent 4155C/4156C Users Guide Vol.2, Edition 1
Knob Sweep Measurements

Measurement Functions

Measurement Functions
The knob sweep function allows you to easily perform real-time sweep measurements by rotating the rotary knob on the front panel. This function is useful when you want to quickly make a rough measurement of a DUT characteristic, or when you want to easily define a measurement setup for normal sweep.
To start the knob sweep measurement, press the green key and then the
Single
front-panel key. The 4155C/4156C starts knob sweep measurement, and repeats measurements continuously until this function is stopped. You can change the measurement setups even while the measurements are being performed.
When knob sweep measurements are started, the VAR1 start value and VAR1 sweep range are 0 V or 0 A. You change the sweep range from 0 to the stop value by rotating the knob.
To stop the knob sweep measurements, press
Stop front-panel key, or a PAGE
CONTROL group key. To restart the measurement, press the following key(s):
If you pressed the
If you pressed a PAGE CONTROL key: green key and
Stop key: Single key
Single key

Available units and functions

Table 2-1 shows available units and functions for knob sweep measurement.
Table 2-1 Available Units and Functions for Knob Sweep Measurement
Unit
VAR1 VAR1' VAR2
Output Function Output Mode
CONST
STAND
BY
VI
COMM
ON
Pulse
SMU · n.a. ······n.a. ·- · VSU · n.a. ····----- - VMU -------- -·· ­GNDU ---·---· --- - PGU ---···-- ·-- -
Meas. Mode
V DVOLT I
· means "This is available for knob sweep measurement".
n.a. means "This is not available for knob sweep measurement".
- means "This is not available for this unit".
Agilent 4155C/4156C User’s Guide Vol.2, Edition 1 2- 3
Knob Sweep Measurements Measurement Functions

Normal Sweep and Knob Sweep Measurements

Table 2-2 compares the normal sweep measurement performed by measurement front-panel keys and knob sweep measurement by the front-panel knob.
Table 2-2 Comparison of Sweep Measurement and Knob Sweep Measurement
Item Sweep Measurement Knob Sweep Measurement
Spacing of VAR1 linear or log linear
Sweep Mode of VAR1 single or double single or double
Number of Steps for VAR1 1 to 1001 1 to 1001
Hold Time 0 to 655.35 s 0 to 655.35 s
Power Compliance available not available
Measurement Ranging Mode auto, limited auto, or fixed
compliance range
Standby Function available available
Measurement Channel 1 to 8 ch 1 ch only
Output Sequence can set
can set
b
Trigger Function available not available
Integration Time short, medium, or long 80 ms
a
Measurement range is automatically set according to specified compliance value.
b
Settings on the MEASURE: OUTPUT SEQUENCE screen also apply to knob
sweep measurement.
a
2-4 Agilent 4155C/4156C Users Guide Vol.2, Edition 1
Knob Sweep Measurements
Measurement Functions

Features of Knob Sweep Function

The following are parameters that are for knob sweep measurement only or that have a different meaning or range from normal sweep measurement.
LIN/LOG mode
Only linear mode is available. Even if you set LOG on the MEASURE: SWEEP SETUP screen, the knob sweep is a linear sweep measurement.
VAR1 Range
Sweep range of VAR1 source output. If you do not set the VAR1 RANGE, the default is the stop value specified for the VAR1 channel on the MEASURE: SWEEP SETUP screen.
Polarity
The following sweep polarities are available for the knob sweep function of the VAR1 source. To set the polarity, select VAR1 SETUP and POLARITY softkeys.
+ (positive) Used to set the sweep output in the positive X direction.
- (negative) Used to set the sweep output in the negative X direction. +/- (bipolar) Used to set the sweep output in both the positive and negative X
directions.
To increase the source absolute value, rotate the rotary knob in clockwise.
To decrease the source value toward 0, rotate the rotary knob in counterclockwise. When measurement curves reach 0, the curves remain at 0 even if you continue to rotate the rotary knob counterclockwise.
Agilent 4155C/4156C User’s Guide Vol.2, Edition 1 2- 5
Knob Sweep Measurements Measurement Functions
Step Time
Step time is the time width of a sweep step as shown in the following figure. For knob sweep measurements, you cannot set the delay time. Instead, you set the step time, which you can only set on the KNOB SWEEP screen.
Setup range is 0.5 ms to 100 ms, with 100 ms resolution.
For normal sweep measurement, the step time depends on the measurement time. For knob sweep measurement, step time is always this specified value.
Measurement Channel
You select the measurement channel by selecting the Y-AXIS ASSIGN softkey on the KNOB SWEEP screen, then selecting the desired secondary softkey. You can select one measurement channel only, so the Y2 axis is not available on the KNOB SWEEP screen.
default measurement channel
When an SMU is set to VAR1
Measurement channel is the VAR1 channel.
When a VSU is set to VAR1
Measurement channel is the first found channel that can measure. Searching order is:
SMU1 ® .....® SMU6 ® VMU1 ® VMU2.
restrictions
If you use series resistance for VAR1 channel and VAR1 channel is V force mode, only VAR1 measurement channel can be assigned to Y axis.
NOTE Measurement Resolution
When performing knob sweep measurement, measurement resolution of each measurement unit is worse than the measurement resolution of normal sweep measurements.
2-6 Agilent 4155C/4156C Users Guide Vol.2, Edition 1
Knob Sweep Measurements
Measurement Functions
Sweep Step Value
For the VAR1 channel, you do not set the step value. You can consider the step value to be the amount you rotate the knob. Then, the sweep is performed for the specified number of steps. The STEP field on the MEASURE: SWEEP SETUP screen has no meaning.
Initial value: 0, Step value automatically set: 0 to VAR1 range/number of steps.
Number of Steps
For the VAR1 channel, you set the number of steps on the KNOB SWEEP screen. So, for the knob sweep function, the number of steps for VAR1 has no relation to the NO OF STEP setting on the MEASURE: SWEEP SETUP screen.
Start Value
The start value is always 0, and does not depend on the polarity. You cannot set the start value. So, the START setting on the MEASURE: SWEEP SETUP screen has no meaning for the knob sweep function.
Stop Value
Stop value is always step value ´ number of steps. You cannot set the stop value. The measurement is continuously repeated from 0 to the stop value until the
Stop
front-panel key is pressed or the KNOB SWEEP screen is changed to another screen.
Measurement Range
If SMU is the measurement channel:
Compliance range is used. That is, the measurement range is set to the lowest range that includes the compliance value that is set on the MEASURE: SWEEP SETUP screen.
If VMU is the measurement channel:
20 V range is used for the grounded measurement mode, and the 2 V range is used for the differential measurement mode.
Agilent 4155C/4156C User’s Guide Vol.2, Edition 1 2- 7
Knob Sweep Measurements Measurement Functions

KNOB SWEEP screen

To start the knob sweep measurement, press the green key and then the Single front-panel key. The 4155C/4156C displays KNOB SWEEP screen, and starts measurements. To stop the knob sweep measurement, press the or a PAGE CONTROL group key.
Sto p front-panel key
Cursor
On the KNOB SWEEP screen, the long cursor is always displayed, and you cannot turn it off. In the CURSOR field, coordinate values of the cursor are displayed in X, Y order.
X axis setting
X axis always plots the VAR1 source value. Maximum value of X axis is the setting value of the VAR1 RANGE secondary softkey of VAR1 SETUP softkey group.
Y axis setting
Y axis always plots the measurement data of the measurement channel. You can select the measurement channel by using the secondary softkeys of the Y-AXIS ASSIGN softkey group. Maximum value of Y axis scale is the compliance value of the measurement channel.
2-8 Agilent 4155C/4156C Users Guide Vol.2, Edition 1
Knob Sweep Measurements
Measurement Functions

Analysis of the Knob Sweep Measurement Results

On the KNOB SWEEP screen, you cannot use analysis functions and user functions. But you can analyze the knob sweep measurement result by quitting knob sweep mode as shown below:
1. Select the SETUP COPY primary softkey on the KNOB SWEEP screen.
2. Press the
The knob sweep results are displayed on the GRAPH/LIST screen. Then you can use analysis functions.
Graph/List front-panel key.
To Use User Function
User function is not available for the knob sweep mode. If you want to use user functions in GRAPH/LIST screen group after the knob sweep measurements, do following before entering to the knob sweep mode:
1. Define user functions on CHANNELS: USER FUNCTION DEFINITION screen.
2. Enter user function names in DATA VARIABLES field of DISPLAY: DISPLAY SETUP screen.
3. On GRAPH/LIST: GRAPHICS screen, select DISPLAY SETUP primary softkey, then set DATA VAR softkey to ON.
After getting the measurement data on the knob sweep mode, and quitting the knob sweep mode as described above, the results of user function will be displayed on the GRAPH/LIST screen.
Agilent 4155C/4156C User’s Guide Vol.2, Edition 1 2- 9
Knob Sweep Measurements Executing Measurements

Executing Measurements

You can easily execute the knob sweep measurements, as shown below:
1. Defines measurement units on the CHANNELS: CHANNEL DEFINITION screen.
2. Presses the green key, and then Single key.
NOTE In the knob sweep mode, user functions and user variables are not available. See “To
Use User Function” on page 2-9.
To use PGU in the pulse output mode (MODE=VPULSE), set up the MEASURE: PGU SETUP screen as same as the normal sweep mode. For setting up the PGU SETUP screen, refer to Chapter 1.
2-10 Agilent 4155C/4156C Users Guide Vol.2, Edition 1
Knob Sweep Measurements
Executing Measurements

To Define Measurement Units

Press Chan front-panel key to define the measurement units. CHANNELS: CHANNEL DEFINITION screen is displayed.
1. MEASUREMENT MODE:
Select SWEEP secondary softkey.
2. VNAME and INAME:
Enter a unique name for voltage or current variable. These names must be 6 or less alphanumeric characters. First character must be alphabet character. For example, enter Vce for collector-emitter voltage. If channel does neither output nor measurement, you can omit VNAME and INAME.
3. MODE:
Select one of the following softkeys to set the output mode:
V : Voltage output (for SMU, VSU, and PGU). V : Grounded mode voltage measurement (for VMU).
I : Current output (for SMU).
VPULSE : Pulsed voltage output (for PGU).
IPULSE : Not available.
COMMON : Circuit common (for SMU and GNDU).
DVOLT : Differential mode voltage measurement (for VMU).
4. FCTN:
Select one of the following softkeys to set the output function:
CONST : Constant output function (for SMU, VSU, and PGU).
VAR1 : Primary sweep output function (for SMU and VSU).
VAR2 : Secondary sweep output function (for SMU and VSU).
VAR1' : Not available.
To disable a unit
Select DELETE ROW softkey. The settings for the unit are deleted.
Agilent 4155C/4156C User’s Guide Vol.2, Edition 1 2-11
Knob Sweep Measurements Executing Measurements

To Execute Measurements

1. Press the green key, then Single front-panel key. The KNOB SWEEP screen is displayed, and knob sweep measurement starts.
During measurements, self-test, or forcing stress, this operation is ignored.
If you want to change the Y-axis parameter, press Y-AXIS ASSIGN primary softkey. Then select a secondary softkey for the Y-axis parameter you want.
To start knob sweep measurement again, press
2. Rotate the rotary knob to stretch or shrink the sweep range. Knob sweep measurement is executed, and measurement curve appears on the graph.
Stop front-panel key and
Single front-panel key only.
Warning messages
If the CHANNELS or MEASURE screen group have incorrect settings for knob sweep measurements when starting the knob sweep, a warning message is displayed, then the STOP and CONT primary softkeys are available.
Select STOP softkey to know the incorrect setting. The setting is highlighted on the screen.
Select CONT softkey to perform knob sweep measurement with the following settings:
If VAR1 function (FCTN) has been set:
The unit works as CONST channel. The output value is VAR1 start value.
If VPULSE or IPULSE mode (MODE) has been set for the VAR1 or VAR2 unit:
The unit works as V or I channel.
If the power compliance function (POWER COMP) has been set:
The function does not work. Same as POWER COMP = OFF.
2-12 Agilent 4155C/4156C Users Guide Vol.2, Edition 1
Knob Sweep Measurements
Executing Measurements
Example
The following figure shows an example to set both X axis and Y axis display regions to positive.

To Stop Measurement

To stop the knob sweep measurement, press Stop front-panel key.
This returns the 4155C/4156C operation state to the previous state. For example, if the knob sweep measurement starts from the idle state, the operation state returns to the idle state.
Starting knob sweep again
To start the knob sweep measurement again, press the following front-panel key:
Single The knob sweep measurement will start from the point where it
was stopped.
Green, Single The knob sweep measurement will start from 0 V or 0 A.
Agilent 4155C/4156C User’s Guide Vol.2, Edition 1 2-13
Knob Sweep Measurements Executing Measurements

To Change Measurement Conditions

To change the measurement conditions, use the following primary softkeys:
Softkey Description
DISPLAY SETUP Used to change graph display. The following
secondary softkeys are available:
X-AXIS REGION Selects the X-axis display range
from +, -, or +/-.
Y-AXIS REGION Selects the Y-axis display range
from +, -, or +/-.
X-AXIS DISPLAY Selects the X-axis direction
from NORMAL or REVERSE.
Y-AXIS DISPLAY Selects the Y-axis direction
from NORMAL or REVERSE.
GRID Sets grid on or off.
VAR1 SETUP Used to change VAR1 sweep source setup. The
following secondary softkeys are available:
SWEEP MODE Sets sweep mode to SINGLE or
DOUBLE.
POLARITY Sets polarity of VAR1 output to
POS, NEG, or BIPOLAR.
VAR1 RANGE Sets VAR1 sweep range. This
sets X-axis scale.
NO OF STEPS Sets number of sweep steps.
COMPLIANCE Sets VAR1 compliance value.
This sets Y-axis scale.
HOLD TIME Sets hold time.
STEP TIME Sets step time.
2-14 Agilent 4155C/4156C Users Guide Vol.2, Edition 1
Knob Sweep Measurements
Executing Measurements
Softkey Description
VAR2 SETUP Used to change VAR2 sweep source setup. The
following secondary softkeys are available:
VAR2 START Sets VAR2 sweep start value.
VAR2 STEP Sets VAR2 sweep steps.
VAR2 POINTS Sets number of sweep steps.
COMPLIANCE Sets VAR2 compliance value.
CONST SETUP Used to change CONST source setup. Secondary
softkeys are available for selecting CONST source. Select a secondary softkey to change the CONST source output value.
To copy knob sweep setups
Select SETUP COPY primary softkey. This copies knob sweep setups to the MEASURE: SWEEP SETUP and DISPLAY: DISPLAY SETUP screens. Then you can see the setups on the screens, and use the setups for the normal sweep measurement.
Note that the power compliance function is not available for the knob sweep measurement. So the POWER COMP column is set to OFF.
Agilent 4155C/4156C User’s Guide Vol.2, Edition 1 2-15
Knob Sweep Measurements Executing Measurements
2-16 Agilent 4155C/4156C Users Guide Vol.2, Edition 1

3 Sampling Measurements

Agilent 4155C/4156C User’s Guide Vol.2, Edition 1
Sampling Measurements
This chapter explains how to execute sampling measurements, and consists of the following sections:
•“Measurement Functions”
•“Defining Measurement Conditions”
•“Making a Measurement”
For details about how to enter or input setup data, refer to Setup Screen Reference manual.
3-2 Agilent 4155C/4156C Users Guide Vol.2, Edition 1
Sampling Measurements

Measurement Functions

Measurement Functions
For a sampling measurement, you can monitor current or voltage changes at a DUT while forcing constant current, constant voltage, or pulsed constant bias.
The 4155C/4156C provides the following three types of sampling measurement according to the sampling interval:
•“Linear Sampling Measurement”
•“Thinned-out Sampling Measurement”
•“Linear Sampling Measurement”

Available Units

Available units and functions for sampling measurement are shown below:
Output Function Output Mode
Unit
VAR1 VAR1' VAR2 CONST STANDBY V I COM V I
SMUn.a.n.a.n.a.
VSU n.a. n.a. n.a.
VMU
GNDU
PGU
---- - --- -·-
---· - --· ---
---· · ·--·--
· · ···
·· ·-----
Pulse
n.a.
n.a. means "This is not available for sampling measurement".
· means "This is available for sampling measurement".
- means "This is not available for this unit".
For sampling measurements, only the PGU output can be pulsed.
The pulse output timing from PGU is not synchronized with the timing of sampling measurement.
Meas. Mode
··
Agilent 4155C/4156C User’s Guide Vol.2, Edition 1 3- 3
Sampling Measurements Measurement Functions

Sampling Interval and Measurement Time

When the sampling interval enough longer than the actual measurement time, measurement unit repeats measurement every specified sampling interval. However, if the sampling interval is less than the measurement time, measurement unit cannot repeat measurements every specified interval. For example, if the measurement time is one and a half the specified sampling interval, the interval of measurement is two times the sampling interval. See Figure 3-1 which explains the operation of the sampling measurement.
Measurement time depends on the measurement condition: integration time, measurement range, and so on. So if you want to execute sampling measurement with the specified sampling interval, you need to know the actual measurement time upon your measurement setup, and set the sampling interval value enough longer than the actual measurement time. You can see typical measurement time by repeating the sampling measurements with several sampling interval settings. See Sampling Measurement Data on page 3-7.
Measurement time is given by the following fomula:
Tmeas = Tinteg + Toh
where,
Tmeas : Measurement time.
Tinteg : Integration time.
Toh : Overhead time caused by the following elements:
range changing time during measurement (when measurement ranging mode is set to auto or limited auto)
range changing time at measurement start (when using measurement range less than the compliance value)
3-4 Agilent 4155C/4156C Users Guide Vol.2, Edition 1
Figure 3-1 Sampling Measurement Operation Summary
e
e
Case 1. Sampling Interval > Meas. Ti me
SMU Output
PGU Output
Meas. Time
Hold Time
2
4
Sampling Interval
3
Trigger TriggerTrigger Trigger Trigger
6
Sampling Interval
5
Sampling Measurements
Measurement Functions
Tim
Sampling Interval
1
Starts Sampling Stops Sampling
Case 2. Sampling Interval < Meas. Time
SMU Output
PGU Output
Meas. Time
Hold Time
2
1
Starts Sampling Stops Sampling
Sampling Interval
3
Trigger TriggerTrigger Trigger Trigger
Total Sampling Time (TOTAL SAMP. TIME)
(TOTAL SAMP. TIME is one of the stop co ndition.)
4
6
Sampling Interval
5
Total Sampling Time (TOTAL SAMP. TIME)
(TOTAL SAMP. TIME is one of the stop conditi on.)
Sampling Interval
7
Tim
7
Agilent 4155C/4156C User’s Guide Vol.2, Edition 1 3- 5
Sampling Measurements Measurement Functions
Sampling measurement is executed as explained below:
1. Forces constant current, constant voltage, or pulsed constant bias.
2. Waits hold time.
3. Triggers one point measurement.
4. Measurement unit executes measurement. Measurement result data is stored in memory.
5. Triggers one point measurement. Interval of trigger is same as Sampling Interval.
6. (Case 1) Measurement unit executes measurement if it is ready to measure. Measurement result data is stored in memory.
(Case 2) Measurement unit waits next trigger if it is busy or in measurement.
7. Repeats steps 5 and 6 until that a sampling completion condition is satisfied.
In Figure 3-1, sampling measurement stops when the completion condition total sampling time is satisfied.
Number of measurement data stored in memory depends on the sampling completion condition. Maximum number is specified by the NO. OF SAMPLES field of the MEASURE: SAMPLING SETUP screen. However the measurement will be immediately stopped if a sampling completion condition is satisfied before reaching the maximum number. For the sampling completion condition, see Sampling Completion on page 3-9.
3-6 Agilent 4155C/4156C Users Guide Vol.2, Edition 1
Sampling Measurements
Measurement Functions

Sampling Measurement Data

Measurement parameters of sampling measurement are set to the NAME column of the DISPLAY: DISPLAY SETUP screen. Available parameters and example parameters for the NAME field are listed in the table below:
Parameter
Name
@TIME Measurement start time. This is the time the measurement unit
starts one point measurement. This is different from timing of the measurement trigger sent every sampling interval.
@INDEX Data index. Integer. This is the index numbered to measurement
data stored in memory.
V1 for example, SMU1 voltage output value or measured value.
I1 for example, SMU1 current output value or measured value.
Measurement start time can be expressed by the following fomula. This formula is available for the measurement points before starting the discarding operation for the linear sampling or thinned-out sampling. For logarithmic sampling, this is available for the measurement points in the first decade.
@TIME = Thold + Tinterval ´ [(@INDEX - 1) + N]
where,
@TIME : Measurement start time.
Thold : Hold time.
Tinterval : Sampling interval.
Meanings of Parameter
@INDEX : Data index.
N : Number of triggers between two nearest measurement points. This
value is 0 if the sampling interval is enough longer than the measurement time.
Agilent 4155C/4156C User’s Guide Vol.2, Edition 1 3- 7
Sampling Measurements Measurement Functions
For example, if Thold=10 ms, Tinterval=5 ms, and @TIME values are as shown below, estimated measurement time is 15 ms to 20 ms, and there are 3 triggers between @INDEX=1 and @INDEX=2.
@TIME (for @INDEX=1) = 10 ms = 10 + 5 ´ [(1 - 1) + 0] ms
@TIME (for @INDEX=2) = 30 ms = 10 + 5 ´ [(2 - 1) + 3] ms
Sampling Interval
Hold Time
TIME
Meas.time
@TIME(at @INDEX=1) @TIME(at @INDEX=2)
To Use Multiple Measurement Units
If you define multiple measurement parameters in the NAME column of the DISPLAY: DISPLAY SETUP screen, sampling measurement is executed by using multiple measurement units. Differences between this measurement and the measurement using only one unit are shown below:
Measurement Sequence
Measurement units start measurement in the order below: Parameters for GRAPH: X ® Y1 ® Y2
Parameters for LIST: Order of No. assigned for the parameters
Value of @TIME
@TIME stores the time the first measurement unit starts measurement. @TIME does not store the time another unit starts measurement.
Measurement Time
Measurement time is sum of the measurement time by all units. To execute sampling measurement with the specified sampling interval, the sampling interval must be enough longer than the measurement time.
3-8 Agilent 4155C/4156C Users Guide Vol.2, Edition 1
Sampling Measurements
Measurement Functions

Sampling Completion

The sampling measurement completes when one of the following conditions is satisfied:
Stop condition
The stop condition is satisfied. See below.
Total sampling time
The specified total sampling time has elapsed.
Available for linear and thinned-out sampling. Setting TOTAL SAMP.TIME to auto or no limit disables this sampling completion condition.
Number of sampling points
The specified number of samples has elapsed.
Available for logarithmic sampling. For linear sampling, setting TOTAL SAMP.TIME to auto enables this sampling completion condition.
Stop front-panel key
The
Stop front-panel key is pressed.
GPIB Command
The 4155C/4156C receives GPIB command to stop sampling.
Emergency Condition
An emergency condition occurs on the 4155C/4156C.
Interlock Open
Interlock terminal opens due to high voltage.
Stop Condition
The stop condition is defined by using the STOP CONDITION table of MEASURE: SAMPLING SETUP screen. This function stops the measurement as shown below.
1. Compares the value of the parameter set to NAME field and the value defined in THRESHOLD field.
2. Counts how many times the selected EVENT occurs.
3. When the count reaches the value defined in EVENT NO. field, sampling is stopped immediately.
Agilent 4155C/4156C User’s Guide Vol.2, Edition 1 3- 9
Sampling Measurements Measurement Functions
To use this function, the INITIAL INTERVAL value must be set to 2 ms or more. The INITIAL INTERVAL is the minimum resolution of the sampling interval. For details about the INITIAL INTERVAL, see “Linear Sampling Measurement” on page 3-13, “Thinned-out Sampling Measurement” on page 3-16, or “Logarithmic Sampling Measurement” on page 3-19.
To set up the stop condition, specify the following parameters on the MEASURE: SAMPLING SETUP screen.
Parameter Description
ENABLE/ DISABLE Enables or disables the stop condition.
ENABLE DELAY Delay time. in second. This is the time from starting sampling
measurement to enabling this function. Allowable range: 0 to INITIAL INTERVAL ´ 32767 s. Resolution: INITIAL INTERVAL.
NAME Name of measurement data or user function to monitor for stop
condition.
Val of EVENT.
THRESHOLD Threshold value at which to stop sampling measurement.
Th of EVENT.
EVENT Event for stop condition.
Val > Th True if NAME parameter value is greater than
THRESHOLD value.
Val < Th True if NAME parameter value is less than
THRESHOLD value.
|Val| > |Th| True if absolute NAME parameter value is
greater than absolute THRESHOLD value.
|Val| < |Th| True if absolute NAME parameter value is less
than absolute THRESHOLD value.
EVENT NO. Target value of the count the event occurs (true). When the
count of true is this value, sampling is immediately stopped. Allowable range: 1 to 200.
3-10 Agilent 4155C/4156C Users Guide Vol.2, Edition 1
Sampling Measurements
Measurement Functions

Source Output Sequence and Time Origin

Source unit output sequence and the time origin depends on the setup value of the OUTPUT SEQUENCE MODE OF SAMPLING field in the MEASURE: OUTPUT SEQUENCE screen. The following two modes are available for the field.
SIMULTANEOUS mode
All source unit starts output at same timing. This timing is defined as the Time Origin. See figure below.
Agilent 4155C/4156C User’s Guide Vol.2, Edition 1 3-11
Sampling Measurements Measurement Functions
SEQUENTIAL mode
Source units starts output in the order defined in the OUTPUT SEQUENCE table of the MEASURE: OUTPUT SEQUENCE screen. Time Origin is when the last source reaches the specified output value. See figure below.
If there is pulse bias sources (PGUs), they start to force pulse base value in the order shown above, and start to force pulse bias at the Time Origin.
3-12 Agilent 4155C/4156C Users Guide Vol.2, Edition 1
Sampling Measurements
Measurement Functions

Linear Sampling Measurement

Linear sampling mode keeps a constant sampling interval that is the interval of measurement trigger. And if the measurement units are ready to measure, the units start measurement, and the result data is stored in memory. This is repeated until one of the sampling completion conditions is satisfied.
However, if both the following two conditions occur, linear sampling mode changes the sampling interval to two times the previous sampling interval, and continues sampling measurement.
number of sampling points reaches specified NO.OF SAMPLES
sampling completion condition is not satisfied
Example Operation
This example assumes the following sampling setup:
INITIAL INTERVAL value is longer than the measurement time
NO.OF SAMPLES value is set to 10
TOTAL SAMP.TIME is long (for example, 50 ´ INITIAL INTERVAL). Do
not set to AUTO which enables the number of sampling points sampling completion condition.
1. Executes one point measurement, and stores data in memory. Repeats this 10 times every sampling interval (INITIAL INTERVAL setting value) because of the sampling interval enough longer than the measurement time.
2. If the sampling completion condition is not satisfied after 10 points measurement, linear sampling mode changes the sampling interval to two times the INITIAL INTERVAL, and continues sampling measurement.
To store new measurement data, linear sampling mode discards a data every couple of nearest data as shown above. Data is updated every measurement.
Agilent 4155C/4156C User’s Guide Vol.2, Edition 1 3-13
Sampling Measurements Measurement Functions
3. If the sampling completion condition is not satisfied after additional 5 points measurement, linear sampling mode changes the sampling interval to two times the previous interval, and continues sampling measurement. Measurement data is updated as described in step 2.
4. This discarding and doubling of the sampling interval is repeated until the sampling completion condition is satisfied. By the end of the measurement, 10 measurement result data is stored in memory.
Parameters
To set up the linear sampling measurement, specify the following parameters on MEASURE: SAMPLING SETUP screen. See Table 3-1.
Parameter Description
MODE Sampling mode. LINEAR.
INITIAL INTERVAL The sampling interval for first NO.OF SAMPLES samples.
Allowable range: 60 ms to 65.535 s.
NO.OF SAMPLES Number of data to be stored by end of measurement.
Maximum: 10001. If there are multiple measurement units, this value must be 10001/(number of units) or less.
TOTAL SAMP.TIME Total sampling time. Does not include HOLD TIME. This is
the time from the 1st trigger to the sampling completion. One of the sampling completion conditions, so sampling stops after this time elapses.
Allowable range when
INITIAL INTERVAL £ 480 ms:
AUTO
Allowable range when
INITIAL INTERVAL ´ (NO.OF SAMPLES - 1) sec to
11
1 ´ 10
sec, AUTO, or NO LIMIT.
INITIAL INTERVAL > 480 ms:
AUTO Enables the number of sampling points sampling
completion condition.
NO LIMIT Disables the total sampling time sampling completion
condition
3-14 Agilent 4155C/4156C Users Guide Vol.2, Edition 1
Sampling Measurements
Measurement Functions
HOLD TIME Hold time. This is the time from starting source output
to first trigger. If this value is 0, first @TIME value is 0.
Allowable range when
INITIAL INTERVAL ³2 ms:
0 to 655.35 s with 100 ms resolution.
Allowable range when
INITIAL INTERVAL <2 ms:
- 30 ms to 655.35 s with 100 ms resolution.
Table 3-1 Effective Parameter Values
INITIAL INTERVAL
NO.OF SAMPLES
TOTAL SAMP.TIME
HOLD TIME
60 ms to 480 ms560 ms to 1.92 ms 2 ms to 65.535 s
Max. 10001/(number of measurement units)
AUTO AUTO/NO LIMIT/
INITIAL INTERVAL ´ (NO.OF SAMPLES - 1) s
11
to 1 ´ 10
s
- 30 ms to 655.35 s, 100 ms resolution 0 to 655.35 s, 100 ms resolution
Stop Condition DISABLE DISABLE/ENABLE
Measurement Units
Measurement Range
Integration Time
e
a
b
1
d
FIX FIX/AUTO/LIMITED
Max. 8
c
Short Short/Medium/Long
a
Number of units (SMUs or VMUs) used for measurements.
b
If voltage source SMU is connected to R-BOX, only this SMU can be used for the
measurement.
c
Maximum 8 units if the 41501 is used (6 SMUs and 2 VMUs). If multiple
measurement units are used, total measurement time will be more than the sum of the measurement time by all units.
d
If compliance value is more than the measurement range, range change when
starting measurement makes long measurement time. Also if ranging mode is set to AUTO or LIMITED, range changing makes long measurement time.
e
Integration time is an element of the measurement time. If automatic measurement
data compensation is executed, the measurement time will be more than two times the integration time.
Agilent 4155C/4156C User’s Guide Vol.2, Edition 1 3-15
Sampling Measurements Measurement Functions

Thinned-out Sampling Measurement

Thinned-out sampling mode operates like the linear sampling mode. Difference is that the sampling interval is not changed in the thinned-out sampling measurement. So even if both the following two conditions occur, thinned-out sampling mode does not change the sampling interval, and continues sampling measurement.
number of sampling points reaches specified NO.OF SAMPLES
sampling completion condition is not satisfied
Example Operation
This example assumes the following sampling setup:
INITIAL INTERVAL value is longer than the measurement time
NO.OF SAMPLES value is set to 10
1. Executes one point measurement, and stores data in memory. Repeats this 10 times every sampling interval (INITIAL INTERVAL setting value) because of the sampling interval enough longer than the measurement time.
2. If the sampling completion condition is not satisfied after 10 points measurement, thinned-out sampling mode keeps the sampling interval, and continues sampling measurement.
To store new measurement data, thinned-out sampling mode discards a data every couple of nearest data as shown above. Data is updated every measurement.
3-16 Agilent 4155C/4156C Users Guide Vol.2, Edition 1
Sampling Measurements
Measurement Functions
3. If the sampling completion condition is not satisfied after additional 5 points measurement, thinned-out sampling mode keeps the sampling interval, and continues sampling measurement. Data is updated as described in step 2.
4. This discarding is repeated until the sampling completion condition is satisfied. By the end of the measurement, 10 measurement result data is stored in memory.
Parameters
To set up the thinned-out sampling measurement, specify the following parameters on MEASURE: SAMPLING SETUP screen. See Table 3-2.
Parameter Description
MODE Sampling mode. THINNED OUT.
INITIAL INTERVAL The sampling interval during thinned-out sampling.
Allowable range: 720 ms to 65.535 s.
NO.OF SAMPLES Number of data to be stored by end of measurement.
Maximum: 10001. If there are multiple measurement units, this value must be 10001/(number of units) or less.
TOTAL SAMP.TIME Total sampling time. Does not include HOLD TIME. This is
the time from the 1st trigger to the sampling completion. One of the sampling completion conditions, so sampling stops after this time elapses. Allowablerange:NOLIMIT,or
INITIAL INTERVAL ´ (NO.OF SAMPLES - 1) sec to
11
1 ´ 10
sec
NO LIMIT Disables the total sampling time sampling completion
condition
HOLD TIME Hold time. This is the time from starting source output to
first trigger. If this value is 0, first @TIME value is 0.
Allowable range when
INITIAL INTERVAL ³2 ms:
0 to 655.35 s with 100 ms resolution.
Allowable range when
INITIAL INTERVAL <2 ms:
- 30 ms to 655.35 s with 100 ms resolution.
Agilent 4155C/4156C User’s Guide Vol.2, Edition 1 3-17
Sampling Measurements Measurement Functions
Table 3-2 Effective Parameter Values
INITIAL INTERVAL
NO.OF SAMPLES
TOTAL SAMP.TIME
HOLD TIME
720 ms to 1.92 ms 2 ms to 65.535 s
Max. 10001/(number of measurement units)
NO LIMIT/
INITIAL INTERVAL ´ (NO.OF SAMPLES - 1) s
11
to 1 ´ 10
- 30 ms to 655.35 s,
100 ms resolution
s
0to655.35 s, 100 ms resolution
Stop Condition DISABLE DISABLE/ENABLE
Measurement Units
a
Measurement Range
Integration Time
a
Number of units (SMUs or VMUs) used for measurements.
b
If voltage source SMU is connected to R-BOX, only this SMU can be used for the
e
b
1
d
FIX FIX/AUTO/LIMITED
Max. 8
c
Short Short/Medium/Long
measurement.
c
Maximum 8 units if the 41501 is used (6 SMUs and 2 VMUs). If multiple
measurement units are used, total measurement time will be more than the sum of the measurement time by all units.
d
If compliance value is more than the measurement range, range change when
starting measurement makes long measurement time. Also if ranging mode is set to AUTO or LIMITED, range changing makes long measurement time.
e
Integration time is an element of the measurement time. If automatic measurement
data compensation is executed, the measurement time will be more than two times the integration time.
3-18 Agilent 4155C/4156C Users Guide Vol.2, Edition 1
Sampling Measurements
M
Measurement Functions

Logarithmic Sampling Measurement

Logarithmic sampling mode plots the measurement data on the X-axis (@TIME) set to the logarithmic scale by doing the following operation. See Figure 3-2.
1. Forces constant current, constant voltage, or pulsed constant bias.
2. Waits hold time.
3. Triggers one point measurement.
4. Measurement unit executes measurement. Measurement result data is stored in memory.
5. Triggers one point measurement. Interval of trigger is constant (setting value of INITIAL INTERVAL).
6. Measurement unit executes measurement if it is ready to measure. Measurement result data is stored in memory.
Measurement unit waits next trigger if it is busy or in measurement.
7. Repeats steps 5 and 6 until that a sampling completion condition is satisfied. Logarithmic sampling mode retains only the measurement data that can plot the data on the logarithmic X-axis in almost the same interval.
Figure 3-2 Example Operation of Logarithmic Sampling
Starts Sampling
Hold Time
3
2
1
4
easurement
Sampling Interval
5
Data of first decade
6
7
Discard
Second decadeFirst decade
Data of second decade
1/10 decade
TIME
This is the example to plot LOG10 data.
@TIME
Agilent 4155C/4156C User’s Guide Vol.2, Edition 1 3-19
Sampling Measurements Measurement Functions
@TIME Value
@TIME value of measurement data is determined by MODE, INITIAL INTERVAL, NO. OF SAMPLES, and HOLD TIME parameters. Where MODE decides number of measurement points in 1 decades. For example, LOG10 mode obtains 10 data per 1 decade.
An example to get measurement data in logarithmic sampling measurement is explained below. This example assumes the following settings. See also Figure 3-2.
MODE = LOG10 (10 data / 1 decade)
INITIAL INTERVAL = 10 ms
NO.OF SAMPLES = 20
HOLD TIME = 10 ms
STOP CONDITION = DISABLE
If sampling interval is enough longer than measurement time:
INITIAL INTERVAL value decides the range of a decade.
10 ms to 100 ms (1st decade) Sampling is executed at the following @TIME
value: 10 ms, 20 ms, 30 ms, 40 ms, 50 ms, 60 ms, 70 ms, 80 ms, 90 ms, 100 ms. LOG10 mode allows to have 10 data in 1 decade. Therefore all measurement data is stored in memory.
100 ms to 1 s There are 90 sampling points in this range. Number of data can
remain is only 10. They can plot the data on the X-axis in almost the same interval.
@TIME values are as follows: 140 ms, 170 ms, 210 ms, 270 ms, 330 ms, 410 ms, 520 ms, 650 ms, 810 ms, 1.02 s.
If sampling interval is less than measurement time:
Measurement time decides the range of a decade. If the measurement time is 18 ms, interval of measurement is 20 ms, and the following data are stored in memory:
20 ms to 200 ms (1st decade) Sampling is executed at the following @TIME
value: 20 ms, 40 ms, 60 ms, 80 ms, 100 ms, 120 ms, 140 ms, 160 ms, 180 ms, 200 ms. LOG10 mode allows to have 10 data in 1 decade. Therefore all data is stored in memory.
200 ms to 2 s There are 90 sampling points in this range. Number of data can
remain is only 10. They can plot the data on the X-axis in almost the same interval.
3-20 Agilent 4155C/4156C Users Guide Vol.2, Edition 1
Sampling Measurements
Measurement Functions
Rule to determine @TIME:
@TIME value is determined by the following rule. Data measured at @TIME=Tlog are stored in memory.
Tlog ³ Ttarget | Tlog - Ttarget | < | Ttarget - Tprev |
where,
Tlog Data stored in @TIME. Actual measurement point.
Ttarget Target value of @TIME. The values can plot data on the logarithmic
X-axis in the same interval completely.
Tprev Tlog - sampling interval. Actual measurement point.
Discard
Tprev Tlog
Sampling Interval
Ttarget
BA
Measurement
Stored in @TIME
Ttarget < Tlog
A > B
TIME
A= | Ttarget - Tprev | B=|Tlog - Ttarget |
Agilent 4155C/4156C User’s Guide Vol.2, Edition 1 3-21
Sampling Measurements Measurement Functions
Parameters
To set up the logarithmic sampling measurement, specify the following parameters on MEASURE: SAMPLING SETUP screen. See Table 3-3.
Parameter Description
MODE Sampling mode. LOG10, LOG25, or LOG50.
MODE Number of data in 1 decade
LOG10 10 LOG25 25 LOG50 50
INITIAL INTERVAL The sampling interval during logarithmic sampling.
Allowable range: 560 ms to 65.535 s. If this value is 560 ms to 10 ms, number of measurement
data may be less than the specified number of data for top 2 decades. Then sampling continues to get all samples.
NO.OF SAMPLES Number of data to be stored by end of measurement. One
of the sampling completion conditions, so sampling stops after this point elapses. Maximum 11 decades.
MODE Maximum value
LOG10 111 LOG25 276 LOG50 551
HOLD TIME Hold time. This is the time from starting source output to
first trigger. If this value is 0, first @TIME value is 0.
Allowable range when
INITIAL INTERVAL ³2 ms:
0 to 655.35 s with 100 ms resolution.
Allowable range when
INITIAL INTERVAL <2 ms:
- 30 ms to 655.35 s with 100 ms resolution.
Example:
HOLD TIME=1.003 s, and measurement interval is 3 ms,
decade and its range are as shown below:
decade Range ( in sec )
1st decade 1.003 to 1.030 ( 3 m +1 to 30 m +1 ) 2nd decade 1.030 to 1.300 ( 30 m +1 to 300 m +1 ) 3rd decade 1.300 to 4.00 ( 300 m +1 to 3+1 ) 4th decade 4 to 31 ( 3+1 to 30+1 ) 5th decade 31 to 301 ( 30+1 to 300+1 )
3-22 Agilent 4155C/4156C Users Guide Vol.2, Edition 1
Loading...