Agilent 4155C Technical Data

Page 1
Agilent 4155C Semiconductor Parameter Analyzer Agilent 4156C Precision Semiconductor Parameter Analyzer
Technical Data

Introduction

Basic Functions
Agilent 4155C and 4156C functions:
Set measurement and/or stress
conditions Control measurement and/or stress
execution Perform arithmetic calculations
 
Display measured and calculated results on the LCD display
Perform graphical analysis
Store and recall measurement setups,
and measurement and graphical display data
Dump to printers or plotters for hard-
copy output
Perform measurement and analysis
with the built-in instrument BASIC
Self test, Auto calibration
Configuration
4155C
4xMPSMU 2xVMU 2x
VSU
41501B
GNDU
(Optional)
4156C
4xHRSMU 2xVMU 2x
VSU
2xPGU(Option)
HPSMU(Option) or 2xMPSMU (Option)
SMU: Source Monitor Unit Display resolution: 6 digits at each current range (0.01fA display resolution at 10pA range) HRSMU: High Resolution SMU (1fA/2µV to 100mA/100V) MPSMU: Medium Power SMU
(
10fA/2µV to 100mA/100V)
HPSMU:High Power SMU
(
10fA/2µV to 100mA/200V)
VMU: Voltage Monitor Unit
(
0.2µV resolution in differential mode)
VSU: Voltage Source Unit PGU: Pulse Generator Unit (1 channel)
GNDU: Ground Unit
*1
*1: Minimum number of installable MPSMU *2: Accuracy not guaranteed. Minimum guaranteed
*1
resolution is 1fA at 10pA range
Page 2

Hardware

Specification Condition
The “supplemental” information and “typical” entries in the following specifi­cations are not warranted, but provide useful information about the functions and performance of the instruments.
The measurement and output accuracy
are specified at the rear panel connector terminals when referenced to the Zero Check terminal under the following conditions:
1. 23°C ± 5°C (double between 5°C to 18°C, and 28°C to 40°C if not noted otherwise)
2. After 40 minutes warm-up
3. Ambient temperature change less than ± 1°C after auto calibration execution.
4. Integration time: medium or long
5. Filter: ON (for SMUs)
6. Kelvin connection (for HRSMU, HPSMU, and GNDU)
7. Calibration period: 1 year

Agilent 4156C Precision Semiconductor Parameter Analyzer

HRSMU (High Resolution SMU) Specifications
Voltage Range, Resolution, and Accuracy (HRSMU)
Voltage Range
± 20V ± 40V ± 100V
*1: 100mA (Vout 20V), 50mA (20V<Vout≤40V) *2: 100mA (Vout 20V), 50mA (20V<Vout≤40V), 20mA (40V<Vout≤100V)
Set. Reso.
1mV
5mV
Set.
Accuracy
±(0.025%+6mV)2mV
Meas. Reso.
Current Range, Resolution, and Accuracy (HRSMU)
Current Range
±A ±10µA ±100µA ±1mA
*1: The accuracy is applicable when offset cancellation has been performed. *2: The offset current specification is multiplied by one of the following factors depending upon the ambient temperature and humidity (RH = Relative Humidity):
*3: 100V (Iout 20mA) 40V (20mA<Iout50mA) 20V (50mA<Iout100mA)
Vout is the output voltage in volts. Iout is the output current in amps. For example, accuracy specifications are given as ±% of set/measured value (0.04%)
plus offset value (200nA+1nA×Vout) for the 1mA range. The offset value consists of a fixed part determined by the set/measuremet range and a proportional part that is multiplied by Vout or Vout/100.
Set. Reso.
10fA±10pA 10fA±100pA 100fA±1nA
±(0.5%+0.7pA+1fA×Vout)
1pA±10nA 10pA±100nA 100pA 1nA 10nA 100nA
±(0.5%+4pA+10fA×Vout)
±(0.12%+40pA+100fA×Vout)
±(0.12%+400pA+1pA×Vout)
±(0.07%+4nA+10pA×Vout)
±(0.07%+40nA+100pA×Vout)
±(0.06%+400nA+1nA×Vout)
±(0.06%+4µA+10nA×Vout)A±10mA
±(0.12%+40µA+100nA×Vout)10µA±100mΑ
5 - 60Temperature
Set.
Accuracy
RHHumidity % 60 - 80
×2×25°C to 18°C ×2×118°C to 28°C ×5×228°C to 40°C
*1*2
*1*2
Meas. Reso.
1fA±(4%+400fA)
±(4%+20fA+1fA×Vout/100)
1fA±(4%+400fA)
*2
±(4%+40fA+10fA×Vout/100)
10fA 10fA 100fA 1pA 10pA 100pA 1nA 10nA 100nA
±(0.5%+0.4pA+1fA×Vout)
±(0.1%+20pA+100fA×Vout)
±(0.1%+200pA+1pA×Vout) ±(0.05%+2nA+10pA×Vout)
±(0.05%+20nA+100pA×Vout)
±(0.04%+200nA+1nA×Vout)
±(0.04%+2µA+10nA×Vout)
±(0.1%+20µA+100nA×Vout)
Output terminal/connection:
Dual triaxial connectors, Kelvin (remote sensing)
HRSMU Measurement and Output Range
Meas.
Accuracy
±(0.01%+200µV)V±(0.02%+400µV)100µV± 2V
±(0.01%+1mV)20µV±(0.02%+3mV)
±(0.015%+2mV)40µV
±(0.02%+5mV)100µV±(0.03%+15mV)
Meas.
Accuracy
±(0.5%+2pA+10fA×Vout)
-100
Max. Current
100mA 100mA
*1 *2
*1*2
*1*2
*2
Voltage/Current Compliance (Limiting):
The SMU can limit output voltage or current to prevent damaging the device under test.
Voltage: 0V to ±100V Current: ±100fA to ±100mA Compliance Accuracy: Same as the
current (voltage) settling accuracy.
HRSMU Supplemental Informa-
tion :
Maximum allowable cable resistance
Max.
when using Kelvin connection (Force,
V
Sense): 10
100V
Typical voltage source output
100V
resistance (Force line/non-Kelvin
100V
connection): 0.2
100V
Voltage measurement input resistance/
100V
current source output resistance:
100V 100V 100V 100V 100V
*3
100
-20-40 20 4
-20
-50
-100
15
10
(10pA range)
Current compliance setting accuracy for opposite polarity: 10pA to 10nA range: V/I setting accuracy ± 12% of range
100nA to 100mA range: V/I setting accuracy ± 2.5% of range
Current (mA)
50
20
0
100
Voltage (V)
2
Page 3
VSU (Voltage Source Unit)
Specifications
VSU Output Range:
Meas.
Voltage Range
Reso.
1mV
±20V *1: Specification is applicable under no load current. Max. Output Current: 100mA
±(0.05% of setting+10mV)
VSU Supplemental Information:
Output resistance: 0.2 Maximum load capacitance: 10µF Maximum slew rate: 0.2V/µs Current limit: 120mA (typical)
Output Noise: 1mV rms (typical)
Meas.
Accuracy
(Voltage Monitor
VMU
Specifications
VMU Measurement Range, Resolution, and Accuracy:
Voltage
Range
VMU Differential Mode Range Resolution, and Accuracy:
Diff V Range
Max. Common Mode Voltage: ± 20V Note: Vi is the input voltage of VMU2 in volts.
Meas. Reso.
Meas. Reso.
±(0.02%+200µV)2µV±2V
±(0.02%+1mV)20µV±20V
Meas.
Accuracy
±(0.03% + 10µV+0.3µV×Vi)µV±0.2V 0.2
±(0.02%+100µV+3 µ V × Vi)2µV±2V
Unit)
Meas.
Accuracy
For example, accuracy specifications are given as ±% of set/measured value (0.02%) plus offset value (1mV+13µV×Vi) for the 2V range. The differential mode offset value consists of a fixed part determined by the measurement range and a proportional part that is multiplied by Vi.
VMU Supplemental Information:
Input Impedance: 1G Input leakage current (@0V): 500pA (Typical)
Measurement noise: 0.01% of range (p-p) (Typical) Differential mode measurement noise:
0.005% of range (p-p) (Typical)

Agilent 4155C Semiconductor Parameter Analyzer

MPSMU
(Medium Power SMU) Specifications
Voltage Range, Resolution, and Accuracy (MPSMU)
Voltage Range
± 20V ± 40V ± 100V
*1: 100mA (Vout 20V), 50mA (20V<Vout≤40V) *2: 100mA (Vout 20V), 50mA (20V<Vout≤40V), 20mA (40V<Vout≤100V)
Set. Reso.
1mV 2mV 5mV
Set.
Accuracy
Meas. Reso.
±(0.02%+700µV+0.3×Iout) V±(0.03%+900µV+0.3×Iout)100µV± 2V
±(0.02%+2mV+0.3×Iout)20µV±(0.03%+4mV+0.3×Iout) ±(0.02%+3mV+0.3×Iout)40µV±(0.03%+7mV)+0.3×Iout) ±(0.03%+5mV+0.3×Iout)100µV±(0.04%+15mV)+0.3×Iout)
Meas.
Accuracy
Current Range, Resolution, and Accuracy (MPSMU)
Current Range
±A ±10µA ±100µA ±1mA
*1: 100V (Iout 20mA), 40V (20mA<Iout≤50mA), 20V (50mA<Iout≤100mA)
Set. Reso.
100fA±1nA 1pA±10nA 10pA±100nA 100pA 1nA 10nA 100nA
±(0.5%+3pA+2fA×Vout)
±(0.5%+7pA+20fA×Vout)
±(0.12%+50pA+200fA×Vout)
±(0.12%+400pA+2pA×Vout)
±(0.12%+5nA+20pA×Vout)
±(0.12%+40nA+200pA×Vout)
±(0.12%+500nA+2nA×Vout)
±(0.12%+4µA+20nA×Vout)A±10mA
±(0.12%+50µA+200nA×Vout)10µA±100
Set.
Accuracy
Meas. Reso.
10fA 10fA 100fA 1pA 10pA 100pA 1nA 10nA 100nA
±(0.5%+5pA+20fA×Vout)
±(0.1%+30pA+200fA×Vout)
±(0.1%+200pA+2pA×Vout)
±(0.1%+3nA+20pA×Vout)
±(0.1%+20nA+200pA×Vout)
±(0.1%+300nA+2nA×Vout)
±(0.1%+2µA+20nA×Vout)
±(0.1%+30µA+200nA×Vout)
Meas.
Accuracy
±(0.5%+3pA+2fA×Vout)
Max. Current
100mA 100mA
Max. V
100V 100V 100V 100V 100V 100V 100V 100V
Output terminal/connection:
Single triaxial connector, non-Kelvin (no remote sensing)
Voltage/Current Compliance (Limiting):
The SMU can limit output voltage or current to prevent damaging the device under test. Voltage: 0V to ±100V
Current: ±1pA to ±100mA Compliance Accuracy: Same as the current (voltage) settling accuracy.
MPSMU Supplemental Information:
Typical voltage source output resistance: 0.3 Voltage measurement input resistance/ current source output resistance: 10 Current compliance setting accuracy for opposite polarity: 1nA to 10nA range: V/I setting accuracy ± 12% of range
100nA to 100mA range: V/I setting accuracy ± 2.5% of range
13
(1nA range)
Vout is the output voltage in volts. Iout is the output current in amps. For example, accuracy specifications are given as ±% of set/measured value (0.1%)
plus offset value (30pA+200fA×Vout) for the 100nA range. The offset value consists of a fixed part determined by the set/measuremet range and a proportional part that is multiplied by Vout.
VSU Specifications
Same as 4156C VSU.
VMU
Specifications
Same as 4156C VMU.
MPSMU Measurem ent and Output Range
-100
Current (mA)
100
50
20
-20-40 20 40
-20
-50
-100
100
Voltage (V)
3
Page 4

Agilent 41501B SMU and Pulse Generator Expander

HPSMU (High Power SMU) Specifications
Voltage Range, Resolution, and Accuracy
Voltage Range
± 20V ± 40V ± 100V ± 200V
Current Range
±A ±1
A
±1
00µA
±1mA
*1: 200V (Iout 50mA), 100V (50mA<Iout100mA) *2: 200V (Iout 50mA), 100V (50mA<Iout125mA), 40V (125mA<Iout500mA), 20V (500mA<Iout1mA)
Vout is the output voltage in volts. Iout is the output current in amps.
For example, accuracy specifications are given as ±% of set/measured value (0.1%) plus offset value (30pA+200fA×Vout) for the 100nA range. The offset value consists of a fixed part determined by the set/measuremet range and a proportional part that is multiplied by Vout.
Set. Reso.
1mV
5mV 10mV
Set. Reso.
100fA±1nA 1pA±10nA 10pA±100nA 100pA 1nA 10nA 100nA
±(0.5%+3pA+2fA×Vout)
±(0.5%+7pA+20fA×Vout)
±(0.12%+50pA+200fA×Vout)
±(0.12%+400pA+2pA×Vout)
±(0.12%+5nA+20pA×Vout)
±(0.12%+40nA+200pA×Vout)
±(0.12%+500nA+2nA×Vout)
±(0.12%+4µA+20nA×Vout)A±10mA
±(0.12%+50µA+200nA×Vout)10µA±100mΑ
Set.
Accuracy
±(0.03%+7mV)2mV
Set.
Accuracy
(HPSMU)
Meas. Reso.
±(0.02%+700µV)V±(0.03%+900µV)100µV± 2V
±(0.035%+10mV)200µV±(0.04%+30mV)
Meas. Reso.
10fA 10fA 100fA 1pA 10pA 100pA 1nA 10nA 100nA
±(0.5%+5pA+20fA×Vout)
±(0.1%+30pA+200fA×Vout)
±(0.1%+200pA+2pA×Vout)
±(0.1%+3nA+20pA×Vout)
±(0.1%+20nA+200pA×Vout)
±(0.1%+300nA+2nA×Vout)
±(0.1%+2µA+20nA×Vout)
±(0.1%+30µA+200nA×Vout)
±(0.5%+300µA+2µA×Vout)1µA±(0.5%+500µA+2µA×Vout)100µA±1Α
HPSMU Measurement and Output Range
Meas.
Accuracy
±(0.02%+2mV)20µV±(0.03%+4mV) ±(0.02%+3mV)40µV ±(0.03%+5mV)100µV±(0.04%+15mV)
Meas.
Accuracy
±(0.5%+3pA+2fA×Vout)
Max. Current
1A 1A 500mA 125mA 50mA
PGU (Pulse Generator Unit) Specifications
Modes: Pulse or constant Amplitude: 0Vpp to 40Vpp Window: -40.0V to +40.0V Maximum current:
±200mA (pulse width: 1ms, average current100mA) ±100mA Pulse width: 1.0µs to 9.99s Minimum resolution: 100ns Pulse period: 2.0µs to 10.0s Minimum resolution: 100ns
Delay: 0s to 10s Minimum resolution: 100ns Transition time: 100ns to 10ms Minimum resolution: 1ns Output impedance: 50or low
impedance (1Ω) Burst count range: 1 - 65535
Pulse/DC Output Voltage and Accuracy (PGU)
Set
Parameter
Base
Pulse
Note: DC output is performed by the Base parameter.
Range
±20V ±40V ±20V ±40V
Max. V
200V 200V 200V 200V 200V 200V 200V 200V
*1 *2
ResolutionVoltage
Output terminal/connection:
Dual triaxial connectors, Kelvin (remote sensing)
Voltage/Current Compliance (Limiting):
Voltage: 0V to ±200V Current: ±1pA to ±1A Compliance Accuracy: Same as the
current (voltage) settling accuracy.
HPSMU Supplemental Information:
Maximum allowable cable resistance when using Kelvin connection: Force: 0.7 (100mA to 1A)
Force: 10 (100mA) Sense: 10 Typical voltage source output
resistance (Force line/non-Kelvin connection): 0.2 Voltage measurement input resistance/ current source output resistance:
1000
500
125
-20-40 20 40
-50
-125
-500
-1000
4mV 8mV 4mV 8mV
13
10
(1nA range)
Current compliance setting accuracy for opposite polarity: 1nA to 10nA range: V/I setting accuracy ± 12% of range 100nA to 1A range: V/I setting accuracy ± 2.5% of range
Current (mA)
50
±(1% of Base + 50mV +1% of Pulse) ±(1% of Base + 50mV +1% of Pulse)
Voltage (V)
100-100
Accuracy
±(3% of Base + 50mV) ±(3% of Base + 50mV)
*1
200-200
*1: Accuracy is specified at leading edge - trailing edge = s
4
Page 5
Pulse parameter accuracy Period: ± Width: ±
(2% +2ns)
(3% +2ns)
Delay: ±(2% +40ns) Transition time: ±(5% +10ns) Trigger output
Level: TTL Timing: Same timing and width as PGU1 pulse output
PGU Supplemental Information:
Overshoot: ≤ ±5% of amplitude ±10mV (50output impedance to 50load) Pulse width jitter: 0.2% + 100ps
Pulse period jitter: 0.2% + 100ps Maximum slew rate: 100V/µs (50output impedance to 50load) Noise: 0.2% of range (@ DC output)
MPSMU Specifications
Same as 4155C MPSMU.
(Ground Unit)
GNDU Specifications:
Output Voltage: 0V±100µV Maximum sink current: 1.6A Output terminal/connection: Single triaxial connector, Kelvin (remote sensing)
GNDU Supplemental Information
Load Capacitance: 1µF Cable resistance: Force: ≤1 Sense: ≤10Ω
HRSMU, MPSMU,HPSMU
Supplemental Information
Maximum capacitive load: 1000pF Maximum guard capacitance: 900pF Maximum shield capacitance: 5000pF Maximum guard offset voltage: ±1mV
Noise characteristics (typical, Filter: ON):
Voltage source noise: 0.01% of V range (rms) Current source noise: 0.1% of I range (rms) Voltage monitor noise: 0.02% of V range (p-p) Current monitor noise: 0.2% of I
Output overshoot (typical, Filter: ON): Voltage source: 0.03% of V range Current source: 1% of I range Range switching transient noise (typical, Filter: ON): Voltage ranging: 250mV Current ranging: 10mV Maximum slew rate: 0.2V/µs
,
Pulse Range and Pulse Parameter (PGU)
Range
1 2
0 - 10ms0.01ms -9.99ms1ms - 10ms3
0 - 100ms0.1ms - 99.9ms10ms -100ms4
Note: Pulse width is defined when leading time is equal to trailing time. PGU2 must be set in the same range as PGU1.
Set resolutionDelayWidthPeriod
0.1µs0 - 100µs1µs - 100µs2µs - 100µs 1µs0 - 1000µs1µs - 999µs100µs - 1000µs
10µs
100µs
1ms0 - 1000ms1ms - 999ms100ms - 1000ms5
10ms0 - 10s0.01s - 9.99s1s - 10s6
Range
0.5µs - 10µs
5.0µs - 100.0µs
0.5ms - 10.0ms
Restrictions:
Pulse width < Pulse Period, Trailing time < (Pulse period - Pulse width) × 0.8 Period, width, and delay of PGU1 and PGU2 must be in the same range. Leading time and trailing time for a PGU must be in the same range.
Capacitance Calculation Accuracy (Supplemental Data)
Capacitance Calculation Accuracy (Supplemental Data)
Accuracy is derived from the current range, voltage range, capacitance measurement and leakage current measurement integration times, and the guard capacitance of cabling and step voltage. The information in the chart below is based on the following conditions: Voltage Range: ± 20 V; Voltage Step: 100 mV; Guard Capacitance: 100 pF; Equivalent parallel resistance of DUT: 1 x 1015 W. The ratio of integration times for capacitance measurement and leakage current measurement is 1:1.
1ns100ns - 1000ns
10ns
100ns
Delay time < Pulse period,
Leading time < Pulse width × 0.8
AccuracySet Restrictions
±(5% + 10ns) ±(5% + 10ns) ±(5% + 10ns) ±(5% + 10ns)1µs50µs - 1000µs ±(5% + 10ns)10µs
HRSMU
Current Integration Measurement Resolution Accuracy Range Time Range Reading % Offset
0.5sec 1pF 5fF 4.2 70fF
100pA 1sec 20pF 10fF 4.3 90fF
1nA 0.1sec 700pF 10fF 0.84 160fF
10nA 0.1sec 7nF 10fF 0.84 200fF
2sec 760pF 20fF 4.3 130fF
0.5sec 4.5nF 40fF 0.85 280fF 2sec 18nF 200fF 0.93 740fF
0.5sec 45nF 40fF 0.85 440fF 2sec 180nF 200fF 0.93 1.4pF
10sec 940nF 1pF 1.3 6.2pF
MPSMU
Current Integration Measurement Resolution Accuracy Time Range Reading % Offset
1nA 0.5sec 4.5nF 40fF 0.94 340fF
10nA 0.5sec 45nF 40fF 0.94 480fF
Current complicance must be larger than the current range. The capacitance of the DUT and measurement system must be smaller than the measurement range. The measurement range might be reduced by the leakage current.
0.1sec 700pF 10fF 0.91 170fF
2sec 18nF 200fF 1.0 1pF
0.1sec 7nF 10fF 0.91 180fF
2sec 180nF 200fF 1.0 1.6pF 10sec 940nF 1pF 1.6 7.6pF
5
Page 6

Functions

Measurement Set-up

Setting
Fill-in-the-blanks using front-panel or
full-size external keyboard
Load settings from floppy disk or via the LAN port
Program using internal Instrument
BASIC or via GPIB
HELP Function Library: Default measure setup, Vce-Ic,
Vds-Id, Vgs-Id, and Vf-If are pre­defined softkeys
User-defined measurement setup
library Auto file load function on power-up
Measurement
The 4155C and 4156C can perform dc or pulsed force/measure, and stress force. For dc, voltage/current sweep and sampling (time domain) measurements are available.

Voltage/Current Sweep Measurement Characteristics

Each SMU and VSU can sweep using VAR1 (primary sweep), VAR2 (subordinate sweep), or VAR1| (synchronous sweep).
VAR1
Primary sweep controls the staircase (dc or pulsed) voltage or current sweep. Maximum number of steps: 1001 for one VAR1 sweep. Sweep type: linear or logarithmic Sweep direction: Single or double sweep Hold time: Initial wait time or wait time after VAR2 is set: 0 to 655.35s with 10ms resolution Delay time: Wait time from VAR1 step to the start of the measurement: 0 to 65.535s with 100 µs resolution
VAR2
Subordinate linear staircase or linear pulsed sweep. After primary sweep is completed, the VAR2 unit output is incremented. Maximum number of steps: 128
6
|
VAR1
Staircase or pulse sweep synchronized with the VAR1 sweep. Sweep is made with a user specified ratio and offset value. VAR1 VAR1 user specified ratio and b is the user specified offset value.
|
output is calculated as
|
= a × VAR1 + b, where a is the
CONSTANT
A source unit can be set as a constant voltage or current source depending on the unit.
PULSE
One of the SMUs can be set as a pulse source. Pulse width: 0.5ms to 100ms, 100µs resolution. Pulse period: (5ms to 1s ( pulse width + 4ms),
100µs resolution. SMU pulse setting accuracy (supplemental information, at fixed range measurement except multi- channel measurement): Width: 0.5% + 50µs Period: 0.5% + 100µs Trigger output delay for pulsed measurement: 0 - 32.7ms with 100µs resolution (< pulse width).

Sampling (Time Domain) Measurement Characteristics

Displays the time sampled voltage/ current data versus time. Maximum sampling points: 10,001 (linear) Sampling mode: linear, log, and thinned-out
Note: The thinned-out mode is similar to reverse-log sampling. Sampling measurement continues by thinning out older data until the sampling completion condition is satisfied.
Sampling interval range and resolution: Linear scale (auto mode): 60µs to 480µs range: 20µs resolution
480µs to 1s range: 80µs resolution 1s to 65.535s range: 2ms resolution
Linear scale (no limit mode), log scale, and thinned-out modes: 560µs (720µs at thinned-out mode)
to 1s range: 1s to 65.535s range:
Note: The following conditions must be set when initial interval is less than
umber of measurement channels: 1
N
 
Measurement ranging: fixed range Stop conditi
80µs resolution
: disable
on
2ms resolution
2ms.
Hold time: Initial wait rime: 0.03s to 655.35s, 100µs resolution Sampling measurement stop condition: A condition to stop the sampling can be defined. Sampling interval setting accuracy (supplemental data):
0.5% + 10µs (sampling interval 480µs)
0.5% + 10µs (480µs sampling interval <2ms)
0.5% + 100µs (2ms sampling interval)

C-V Measurement Characteristics

Capacitance is calculated value derived from the following equation:
dQ C =
dQ is the charge when dV, the step voltage, is applied by the SMU; dQ is derived from the measure­ment current (amps) and the integration time (seconds).
Measurement Range:
Measurement range depends on the current range, integration time, and step voltage (refer to the chart in supplemental data).
Capacitance Calculation Accuracy:
Accuracy is dependent on accuracy of the current measurement and voltage measure­ment, and the stray capacitance and leakage current of measurement system etc (Refer to the chart in supplemental data).
Zero Offset:
Cancels stray capacitance of the fixtures and test leads.
Leakage Current Compensation:
Cancels the influence of the leakage cur­rent to the capacitance measurement.
dV

Stress Force Characteristics

SMU, VSU, and PGU output can be forced for the user specified period. Stress time set range: 5000µs to 31,536,000s (365 days)
Resolution: 100µs (500µsstress time10s)
10ms (10s<stress time≤31,536,000s) Burst pulse count: 1 - 65,535 (PGU only) Trigger:
The 4155C and 4156C output a gate trigger while stress channels are forcing stress.
Page 7
Knob Sweep
In the knob sweep mode, sweep range is controlled instantaneously with the front-panel rotary knob. Only the Channel Definition page needs to be defined.
Standby Mode
SMUs in Standby remain programmed to their specified output value even as other units are reset for the next measurement.
Other Characteristics
Limited auto-ranging, voltage/current compliance, power compliance, automatic sweep abort functions, self-test, and self-calibration.

Arithmetic and Analysis Functions

Ar
ithmetic Functions
User Functions
Up to six USER FUNCTIONS can be defined using arithmetic expressions. Measured data and analyzed variables from graphics analysis (marker, cursor, and line data) can be used in the computation. The results can be displayed on the LCD.
Marker
Marker to min/max, interpolation, direct marker, and marker slip
Cursor
Long and short, direct cursor.
Line
Two lines, normal mode, grad mode, tangent mode, and regression mode.
Scaling
Auto scale and zoom.
Data Variable Display
Up to two user defined parameters can be displayed on the graphics screen.
Read Out Function
The read out functions are built-in functions for reading various values related to the marker, cursor, or line.
Automatic Analysis Function
On a graphics plot, the markers and lines can be automatically located using the auto analysis setup. Parameters can be automatically determined using automatic analysis, user function, and read out functions.
User Variable
Display the data on the LCD via GPIB or instrument BASIC.
parallel printer port, or network interface to supported HP plotters or printers. PCL, HR PCL (high-resolution PCL), and HP GL formats are supported
(selectable).
Text Hard Copy
Print out setup information or measured data list as ASCII text via GPIB, parallel printer port, or network interface to supported HP plotters or printers. PCL, HR PCL, and HP GL formats are supported (selectable).
Hard Copy File
Hard copy output can be stored to an internal or external mass storage device instead of sending it to a printer or plotter. The data can be stored in PCL, HR PCL, TIFF, HR TIFF (high­resolution TIFF), or HP GL formats.
Hard Copy via Network Interface
The network interface has lpr client capability.
High-Resolution (HR) Mode
This file mode is available for cases where an extremely clean print-out or plot is desired.
Note: High resolution mode takes signifi­cantly greater CPU time to generate, so its use is recommended for final reports only.

Data Storage

Arithmetic Operators
+, -, *, /, ^, LGT (logarithm, base 10), LOG (logarithm, base e), EXP (exponent), DELTA, DIFF (differential), INTEG (integration), MAVG (moving average), SQRT, ABS (absolute value), MAX, MIN, AVG (averaging), COND (conditional evaluation).
Physical Constants
Keyboard constants are stored in memory as follows: q: Electron Charge, 1.602177 E-19 C k: Boltzmans Constant, 1.380658 E-23 ε:
Dielectric Constant of Vacuum,
8.854188 E-12
Engineering Units
The following unit symbols are also available on the keyboard: f (10
-12
p (10
3
K (10
Analysis Capabili
Overlay Graph Comparison
A graphics plot can be store recalled as an overlay plane. Four overlay planes can be stored. One plane can be overlaid onto the current data
), n (10
), M (10
-9
), u or µ (10
6
), G (10
ties
9
-6
)
d and later
-15
), m (
),
-3
10
),
.

Output

Display
Display Modes
Graphics and list.
Graphics Display
X-Y or X-Y1/Y2 plot of source current/voltage, measured current/voltage, time, or calculated USER FUNCTION data.
List Display
Measurement data and calculated USER FUNCTION data are listed in conjunction with VAR1 step number or time domain sampling step number. Up to eight data sets can be displayed.
Display
8.4 inch diagonal color active matrix LCD, 640 dot (H) × 480 dot (V)
Hard Copy Functions
Graphics Hard Copy
Measured data and all data appearing on the LCD can be output via GPIB,
Mass storage device: Built-in 3.5 inch flexible disk drive Media: 3.5 inch 2HD or 2DD diskette Format type: HP LIF and DOS User area:
1.44Mbyte (2HD) or 720Kbyte (2DD) File types: Auto start program file, initial setup file, measurement setup file, measurement setup/result file, stress setup file, customize file, hard copy data file, and instrument BASIC program and data file. Format of data made by the HP BASIC program: Data made by the HP BASIC program and data made by the instrument BASIC program are compatible. Network mass storage device: An NFS mountable mass storage device File types: Auto start program file, initial setup file, measurement setup file, measurement setup/result file, stress setup file, customize file, and hard copy data file.
7
Page 8
Maximum number of files allowed per directory on network mass storage device: 199 Data storage (supplemental data): 2HD DOS format: Available bytes: 1457K (byte) File size: Measurement setup: 3843 (byte) Stress setup: 601 (byte) Measurement setup/result (Typical data): 15387 (byte) (VAR1: 101, VAR2: 5) Customized system setup: 1661 (byte) Hardcopy data: 30317 (byte) (Monochrome PCL 75DPI file) Hardcopy data: 38702 (byte) (monochrome TIFF file)
Note: For LIF format, the total number of files is limited to 199.
Repeating and Automating
Test
Instrument Control
Agilent 4155C and 4156C function control: Internal or external computer controls the 4155C and 4156C functions via the GPIB interface Command sets: SCPI command set
Agilent FLEX command set Agilent 4145B command set Program Memory: Using the command set, the user can store program code in the 4155C or the 4156C. The maximum number of subprograms is 256 (8 bit). External instrument remote control: Control external equipment via the GPIB interface.
Instrument BASIC
Instrument BASIC is a subset of HP BASIC. Functions: Arithmetic operation, binary operation, string manipulation, logical operation, array operation, program flow control, event-initiated branching, program editing and debugging support, mass storage operation, instrument control, real-time clock, softkey operation, and graphics.
Agilent 4145B automatic sequence program (ASP) typing aid: 4145B ASP-like syntax softkeys are available in instrument BASIC. A 4145B ASP file cannot be read by the 4155C or 4156C.
4155C/4156C Agilent FLEX
Remote control: Instrument BASIC is remote controllable from an external computer via the GPIB interface. Instrument BASIC memory area (supplemental data): Program (text) area: 16K (byte) Variable/stack area: 500K (byte) Common variable area: 600K (byte)
Note: The memory size for common variabl is decreased when hard copy or disk operation is performed.
Trigger
Input: External trigger input starts a sweep or sampling measurement or can be used as a trigger input for continuing an Instrument BASIC program. Input Level:
TTL level, negative or positive edge trigger Output: External trigger can be generated by the following events: start of each sweep measurement step, start of each pulse (SMU) output, while the stress source is forcing, and Instrument BASIC trigger out command execution. Output Level: TTL level, negative or positive logic

4145B Data Compatibility and Syntax Commands

Setup and data file
Measurement setup and data from the 4145B can be loaded.
GPIB program
GPIB programs for the 4145B can be used when the 4145B command set is selected.
Note: There is a possibility that GPIB programs for the 4145B will need to be modified.

Interfaces

GPIB interface: SH1, AH1, T6, L4, SR1, RL1, PP0, DC1, DT1, C1, C2, C3, C4, C11, E2 Parallel interface: Centronics RJ45:
Ethernet IEEE 802.3 10BASE-T for a 10Mbps CSMA/CD local area network External keyboard: Compatible PC-style 101-key keyboard (mini DIN connector)
Interlock and LED connector R-BOX control connector Trigger in/out SMU/PGU selector control connector (41501B)

Sample Application Programs

Flash EEPROM test TDDB Constant I (Electromigration test) V-Ramp Test J-Ramp Test SWEAT GO/NO-GO Test HCI degradation test
Sample VEE
Vth measurement using the 4155C or 4156C, the E5250A, and a wafer prober.
plug&play
VXI
VXIplug&play drivers for the 4155C and 4156C
Supported VXIplug&play operating systems:
Windows NT Windows 95
Format
Tree-structured function panel.
Panel mode for hardware configura- tion and manual parameter setting. Parameter mode for variable definition and I/O configuration.
Program
Drivers

General Specifications

Temperature range
Operating: +10°C to +40°C (if using floppy disk drive) +5°C to +40°C (if not using floppy disk drive) Storage: -22°C to +60°C
Humidity range
Operating: 20% to 80% RH, non-condensing and wet bulb temperature 29°C (if using floppy disk drive) 15% to 80% RH, non-condensing and
wet bulb temperature 29°C (if not using floppy disk drive) Storage: 5% to 90% RH , non­ condensing and wet bulb temperature
39°C
8
Page 9
Altitude
Operating: 0 to 2,000 m (6,561 ft) Storage: 0 to 4,600 m (15,091 ft)
Power requirement
90V to 264V, 47 to 63 Hz
Maximum VA
4155C and 4156C: 450VA 41501B: 350 VA
Regulatory Compliance
EMC:
EN 61326-1:+A1, AS/NZS 2064.1
Safety:
CSA C22.2 NO.1010.1-1992 IEC 61010-1:+A2/EN 61010-1:+A2 UL3111-1:1994
Certification: CE, CSA, NRTL/C, C-Tick
Dimensions:
4155C and 4156C: 235mm H × 426mm W × 600mm D 41501B: 190mm H × 426mm W × 600mm D
Weight (approx.):
4155C and 4156C: 21kg 41501B: 16kg (option 412, HPSMU + 2×PGU)
4155C and 4156C Furnished Accessories
Triaxial cable, 4 ea. (4155C) Kelvin triaxial cable, 4 ea. (4156C) Coaxial cable, 4 ea.
Interlock cable, 1 ea. Keyboard, 1 ea. User manual, 1 set Sample application program disk, 1 ea. Sample VEE program disk, 1 ea VXIplug&play drivers disk for the 4155C and 4156C, 1 ea. VXIplug&play drivers disk for the E5250A, 1 ea.

Accessory Specifications

Specification Condition
The supplemental information and typical entries in the following specifications are not warranted, but provide useful information about the functions and performance of the instruments. 23°C ± 5°C, 50% RH.
16440A
SMU/Pulse Generator Selector
The 16440A switches either an SMU or PGU to the associated output port.
You can expand to 4 channels by adding
an additional 16440A. The channel 1 PGU port provides PGU OPEN function, which can disconnect the PGU by opening a semiconductor relay. The
16440A cannot work without two pulse generator units of the 41501A/B (SMU and Pulse Generator Expander). Channel configurations: Two channels (CH1, CH2) CH1: INPUT ports: 2 (SMU and PGU, PGU port has additional series semiconductor relay) OUTPUT port: 1
CH2: INPUT ports: 2 (SMU and PGU) OUTPUT port: 1
Voltage & Current Range
Input port
Supplemental Information (at
23°C ± 5°C, 50% RH)
SMU port leakage current: < 100fA @100V SMU port residual resistance (typical):
0.2 SMU port stray capacitance (typical @1MHz): Force Common: 0.3pF Force Guard: 15pF Guard Common: 130pF
PGU port residual resistance: 3.4 PGU port OFF capacitance (typical): 5pF PGU port OPEN capacitance (typical): 700pF (@ 1MHz, Vin - Vout = 0V)
PGU port signal transfer charac
Overshoot: < 5% of pulse amplitude (@20ns leading and trailing time, 50 pulse generator source impedance, 50pF and 1M in parallel load).
General Specifications
Dimensions: 50 mm H × 250 mm W × 275 mm D Weight (approx.): 1.1kg
Max IMax. V
1.0 A200 VSMU
0.2A (AC peak)40VPGU
teristics
16441A R-BOX
The 16441A R-BOX adds a selectable series resistor to the SMU output. You can select the resistor from the setup page, and the voltage drop due to the series resistor is automatically compensated for in the measurement result. Measurement limitations with the 4155C and 4156C and R-BOX
If you measure device characteristics including negative resistance over
1M with the 4155C/4156C and R-BOX, there is a possibility that they cannot be measured.
There is a possibility that the 4155C and 4156C cannot perform measurements because of DUT oscillations even with the R-BOX. Whether oscillation occurs or not depends upon the DUT and
measurement conditions. Number of SMU channels that can add resistor: 2 Resistor values: 1M, 100k, 10k, 0 (each channel) Resistance accuracy:
0.3% (at 23°C±5°C, between input­ output terminal) Maximum voltage: 200V Maximum current: 1A (0 selected) Kelvin connection: Kelvin connection is effective only when 0is selected.
Supplemental Information (at
23°C ± 5°C, 50% RH)
Leakage current: <100fA @ 100V
General Specifications
Dimensions: 72 mm H × 250 mm W × 270 mm D Weight (approx.): 1.6kg
16442A Test Fixture
Channel Information
SMU: 6 channels (1 triaxial connector/ channel) 3 channels (1 Kelvin triaxial connector/ channel) VSU: 2 channels (1 BNC connector/channel) VMU: 2 channels (1 BNC connector/channel) PGU: 2 channels (1 BNC connector/channel) GNDU: 1 channel (1 triaxial connector) INTLK: 6 pin connector
Supplemental Information (at
23°C ± 5°C, 50% RH)
SMU channel: Leakage current: 10pA max @200V (Force or Sense Common) Stray capacitance: 15pF max (Force or Sense Common) Stray capacitance: 3pF typical (Force or Sense Other SMU)
Residual resistance: 60m typical (Force, Sense) Guard capacitance: 70pF max (Force or Sense Guard)
9
Page 10
VSU channel residual resistance:
60m typical
VMU channel residual resistance:
60m typical
PGU channel characteristic impedance: 50 typical GNDU channel residual resistance:
40m typical (Force, Sense)
General Specifications
Temperature range
Operating: +5°C to +40°C
Storage: -40°C to +70°C Humidity range Operating: 5% to 80% RH (no condensation) Storage: 5% to 90% RH at 65°C (no condensation) Dimensions: 140 mm H × 260 mm W × 260 mm D Weight (approx.): 2.5kg
For more information about Agilent Technologies semiconductor test products, applications,and services,visit our website:
www.agilent.com/go/semiconductor
or you can call one of the centers listed and ask to speak with a semiconductor test sales representative. For more information about other Agilent test and measurement products,go to
www.agilent.com United States:
1 800 452 4844
Canada:
1 877 894 4414
Europe:
(31 20) 547 2000
Japan:
(81) 426 56 7832 Fax: (81) 426 56 7840
Latin America:
(305) 269 7500 Fax: (305) 269 7599
Australia/New Zealand:
1 800 629 485 (Australia) Fax: (61 3) 9272 0749 0 800 738 378 (New Zealand) Fax: (64 4) 495 8950
Asia Pacific:
(852) 2599 7889 Fax: (852) 2506 9233
Taiwan:
(886 2) 717 9524 Fax: (886) 2 718 9860
Korea:
(822) 769 0800 Fax: (822) 786 1083
Singapore:
(65) 1 800 292 8100 Fax: (65) 275 0387
Product specifications and descriptions in this document subject to change without notice. Copyright ©2000 Agilent Technologies Printed in U.S.A. 11/00
5988-0301EN
Loading...