Agilent 4155C Data Sheet

Agilent 4155C Semiconductor Parameter Analyzer
Agilent 4156C Precision Semiconductor Parameter Analyzer
Data Sheet
S
Introduction
Agilent 4155C and 4156C Basic Functions
• Set measurement and/or stress conditions
• Control measurement and/or stress execution
• Perform arithmetic calculations
• Perform graphical analysis
• Store and recall measurement setups, and measurement and graphical display data
• Dump to printers or plotters for hardcopy output
• Perform measurement and analysis with built-in instrument BASIC
• Self test, Auto calibration
Configuration
The 4155C and 4156C both come standard with I/CV 2.1 Lite automation software. A PC-based instrument controller with I/CV Lite preinstalled and an Agilent 82357A USB/GPIB interface are also included with the standard configuration. You have the option of deleting the controller and cable from your order, but I/CV Lite is always included with
both instruments. If you want the full version of I/CV 2.1, you can request the E5240BU upgrade kit when you order a 4155C or 4156C. For more information about the differences between I/CV 2.1 Lite and I/CV 2.1, please refer to the Agilent I/CV 2.1 Technical Overview, publication number 5988-8474EN.
4155C
4xMPSMU
2xVMU 2xVMU 2xVSU 2xVSU I/CV 2.1 Lite I/CV 2.1 Lite
Standard PC-based controller
and USB/GPIB interface
41501B (Optional)
GNDU
4156C
4xHRSMU
2xPGU (Option)
HPSMU (Option) or 2xMPSMU (Option)
1
SMU: Source Monitor Unit Display resolution: 6 digits at each current range (0.01fA display resolu­ tion at 10pA range) HRSMU: High Resolution SMU (1fA/2µV to 100mA/100V) MPSMU: Medium Power SMU (10fA/2µV to 100mA/100V) HPSMU: High Power SMU (10fA/2µV to 1A/200V) VMU: Voltage Monitor Unit (0.2µV resolution in differential mode) VSU: Voltage Source Unit PGU: Pulse Generator Unit (1 channel) GNDU: Ground Unit
1
Minimum number of installable MPSMU or
PGU is two.
2
Accuracy not guaranteed. Minimum
guaranteedresolution is 1fA at 10pA range.
1
2
S1
Hardware
Specification Condition
The “supplemental” information and “typical” entries in the following specifi­cations are not warranted, but provide useful information about the functions and performance of the instruments.
The measurement and output accuracy are specified at the rear panel connec­tor terminals when referenced to the Zero Check terminal under the follow­ing con-ditions:
1. 23° C ±5° C (double between 5° C to 18° C, and 28° C to 40° C if not noted otherwise)
3. Ambient temperature change less than ±1° C after auto calibration execution.
4. Integration time: medium or long
5. Filter: ON (for SMUs)
6. Kelvin connection (for HRSMU, HPSMU, and GNDU)
7. Calibration period: 1 year
2. After 40 minutes warm-up
Agilent 4156C Precision Semiconductor Parameter Analyzer
HRSMU (High Resolution SMU) Specifications
Voltage Range, Resolution, and Accuracy (HRSMU)
Voltage Set. Set. Meas. Meas. Max. Range Reso. Accuracy Reso. Accuracy Current
±2V 100µV ±(0.02%+400µV) 2µV ±(0.01%+200µV) 100mA
±20V 1mV ±(0.02%+3mV) 20µV ±(0.01%+1mV) 100mA ±40V 2mV ±(0.025%+6mV) 40µV ±(0.015%+2mV) ±100V 5mV ±(0.03%+15mV) 100µV ±(0.02%+5mV)
1
100mA (Vout £20V), 50mA (20V<Vout£40V)
2
100mA (Vout £20V), 50mA (20V<Vout£40V), 20mA (40V<Vout£100V)
Current Range, Resolution, and Accuracy (HRSMU)
Current Set. Set. Meas. Meas. Max. Range Reso. Accuracy Reso. Accuracy V
±10pA 10fA ±(4%+400fA)
±100pA 10fA ±(4%+400fA)
±1nA 100fA ±(0.5%+0.7pA+1fA×Vout) ±10nA 1pA ±(0.5%+4pA+10fA×Vout) 10fA ±(0.5%+2pA+10fA×Vout) 100V ±100nA 10pA ±(0.12%+40pA+100fA×Vout) 100fA ±(0.1%+20pA+100fA×Vout) 100V
±1µA 100pA ±(0.12%+400pA+1pA×Vout) 1pA ±(0.1%+200pA+1pA×Vout) 100V ±10µA 1nA ±(0.07%+4nA+10pA×Vout) 10pA ±(0.05%+2nA+10pA×Vout) 100V ±100µA 10nA ±(0.07%+40nA+100pA×Vout) 100pA ±(0.05%+20nA+100pA×Vout) 100V ±1mA 100nA ±(0.06%+400nA+1nA×Vout) 1nA ±(0.04%+200nA+1nA×Vout) 100V ±10mA 1µA ±(0.06%+4µA+10nA×Vout) 10nA ±(0.04%+2µA+10nA×Vout) 100V ±100mA 10µA ±(0.12%+40µA+100nA×Vout) 100nA ±(0.1%+20µA+100nA×Vout)
1
The accur acy is applicable when offset cancellation has been performed.
2
The offset current specification is multiplied by one of the following
factors depending upon the ambient temperature and humidity (RH = Relative Humidity):
Temperature 5 - 60 60 - 80
5° C to 18° C ×2 ×2 18° C to 28° C ×1 ×2
28° C to 40° C ×2 ×5
3
100V (Iout£20mA) 40V (20mA<Iout£50mA) 20V (50mA<Iout£100mA)
Vout is the output voltage in volts. Iout is the output current in amps.
For example, accuracy specifications are given as ±% of set/measured value (0.04%) plus offset value (200nA+1nA×Vout) for the 1mA range. The offset value consists of a fixed part determined by the set/measure­ ment range and a proportional part that is multiplied by Vout or Vout/100.
Humidity % RH
1, 2
1, 2
1fA ±(4%+20fA+1fA×Vout/100) 1fA ±(4%+40fA+10fA×Vout/100)
2
10fA ±(0.5%+0.4pA+1fA×Vout)
1
2
1, 2
100V
1, 2
100V
2
100V
100
-
Output terminal/connection:
Dual triaxial connectors, Kelvin (remote sensing)
Voltage/Current Compliance (Limiting):
The SMU can limit output voltage or current to prevent damaging the device under test. Voltage: 0 V to ±100 V Current: ±100 fA to ±100 mA Compliance Accuracy: Same as the current (voltage) settling accuracy.
HRSMU Supplemental Information:
Maximum allowable cable resistance when using Kelvin connection (Force,
Sense): 10
Typical voltage source output resistance (Force line/non-Kelvin connection): 0.2 Voltage measurement input resist ance/ current source output resistance:
15
³10
W (10 pA range)
Current compliance setting accuracy for opposite polarity:
3
10 pA to 10 nA range: V/I setting accuracy ±12% of range 100 nA to 100 mA range: V/I setting accuracy ±2.5% of range
Current (mA)
100
50
20
20 40 100
20
-40-
20
-
W
W
HRSMU Measurement and Output Range
Voltage (V)
50
-
100
-
2
Agilent 4155C Semiconductor Parameter Analyzer
MPSMU (Medium Power SMU) Specifications
Voltage Range, Resolution, and Accuracy (MPSMU)
Voltage Set. Set. Meas. Meas. Max. Range Reso. Accuracy Reso. Accuracy Current
±2V 100µV ±(0.03%+900µV+0.3×Iout) 2µV ±(0.02%+700µV+0.3×Iout) 100mA ±20V 1mV ±(0.03%+4mV+0.3×Iout) 20µV ±(0.02%+2mV+0.3×Iout) 100mA
±40V 2mV ±(0.03%+7mV+0.3×Iout) 40µV ±(0.02%+3mV+0.3×Iout)
±100V 5mV ±(0.04%+15mV+0.3×Iout) 100µV ±(0.03%+5mV+0.3×Iout)
1
100mA (Vout £20V), 50mA (20V<Vout £40V)
2
100mA (Vout £20V), 50mA (20V<Vout £40V), 20mA (40V<Vout£100V)
Current Range, Resolution, and Accuracy (MPSMU)
Current Set. Set. Meas. Meas. Max. Range Reso. Accuracy Reso. Accuracy V
±1nA 100fA ±(0.5%+3pA+2fA×Vout) 10fA ±(0.5%+3pA+2fA×Vout) 100V ±10nA 1pA ±(0.5%+7pA+20fA×Vout) 10fA ±(0.5%+5pA+20fA×Vout) 100V ±100nA 10pA ±(0.12%+50pA+200fA×Vout) 100fA ±(0.1%+30pA+200fA×Vout) 100V ±1µA 100pA ±(0.12%+400pA+2pA×Vout) 1pA ±(0.1%+200pA+2pA×Vout) 100V ±10µA 1nA ±(0.12%+5nA+20pA×Vout) 10pA ±(0.1%+3nA+20pA×Vout) 100V ±100µA 10nA ±(0.12%+40nA+200pA×Vout) 100pA ±(0.1%+20nA+200pA×Vout 100V
±1mA 100nA ±(0.12%+500nA+2nA×Vout) 1nA ±(0.1%+300nA+2nA×Vout) 100V ±10mA 1µA ±(0.12%+4µA+20nA×Vout) 10nA ±(0.1%+2µA+20nA×Vout) 100V
±100mA 10µA ±(0.12%+50µA+200nA×Vout) 100nA ±(0.1%+30µA+200nA×Vout)
1
100V (Iout £20V), 40V (20mA<Iout£50mA), 20V (50mA<Iout£100mA)
Vout is the output voltage in volts. Iout is the output current in amps. For example, accuracy specifications are given as ±% of set/measured value (0.1%) plus offset value (30pA+200fA×Vout) for the 100nA range. The offset value consists of a fixed part determined by the set/ measurement range and a proportional part that is multiplied by Vout.
1
2
1
Output terminal/connection:
Single triaxial connector, non-Kelvin (no remote sensing)
Voltage/Current Compliance (Limiting):
The SMU can limit output voltage or current to prevent damaging the device under test. Voltage: 0 V to ±100 V Current: ±1 pA to ±100 mA Compliance Accuracy: Same as the current (voltage) settling accuracy.
MPSMU Supplemental Information:
Typical voltage source output resistance: 0.3
W
Voltage measurement input resistance/ current source output resistance:
13
³10
W (1 nA range)
Current compliance setting accuracy foropposite polarity: 1nA to 10 nA range: V/I setting accuracy ±12% of range 100 nA to 100 mA range: V/I setting accuracy ±2.5% of range
Current (mA)
100
MPSMU Measurement and Output Range
50
20
VSU and VMU specifications are common to both the 4155C and 4156C
VSU (Voltage Source Unit) Specifications
VSU Output Range:
Voltage Meas. Meas. Range Reso. Accuracy
±20V 1mV ± (0.05% of setting +10mV)
1
Specification is applicable under no load
current. Max. Output Current: 100mA
VSU Supplemental Information:
Output resistance: 0.2 W (typical) Maximum load capacitance: 10 µF Maximum slew rate: 0.2 V/µs Current limit: 120 mA (typical) Output Noise: 1 mV rms (typical)
100
-
VMU (Voltage Monitor Unit) Specifications
VMU Differential Mode Range, Resolution, and Accuracy:
1
Diff V Meas. Meas. Range Reso. Accuracy
±0.2V 0.2µV ±(0.03%+10µV+0.3µV×Vi)
±2V 2µV ±(0.02%+100µV+3µV×Vi)
Max. Common Mode Voltage: ± 20V
Note: Vi is the input voltage of VMU2 in volts. For example, accuracy specifications are given as ±% of set/measured value (0.02%) plus offset value (100µV+3µV×Vi) for the 2V range. The differential mode offset value consists of a fixed part determined by the measurement range and a proportional part that is multiplied by Vi.
20
-40-
20 40 100
20
-
50
-
100
-
VMU Measurement Range, Resolution, and Accuracy:
Voltage Meas. Meas. Range Reso. Accuracy
±2V 2µV ±(0.02%+200µV) ±20V 20µV ± (0.02%+1mV)
VMU Supplemental Information:
Input Impedance: ³1G W Input leakage current (@0 V): £500 pA Measurement noise: 0.01% of range (p-p) (typical) when integration time is 10 PLC Differential mode measurement noise:
0.005% of range (p-p) (typical) when integration time is short.
Voltage (V)
3
Agilent 41501B SMU and Pulse Generator Expander
HPSMU (High Power SMU) Specifications
Voltage Range, Resolution, and Accuracy (HPSMU)
Voltage Set. Set. Meas. Meas. Max. Range Reso. Accuracy Reso. Accuracy Current
± 2V 100µV ±(0.03%+900µV) 2µV ±(0.02%+700µV) 1A ± 20V 1mV ±(0.03%+4mV) 20µV ±(0.02%+2mV) 1A ± 40V 2mV ±(0.03%+7mV) 40µV ±(0.02%+3mV) 500mA ±100V 5mV ±(0.04%+15mV) 100µV ±(0.03%+5mV) 125mA ±200V 10mV ±(0.045%+30mV) 200µV ±(0.035%+10mV) 50mA
Current Range, Resolution, and Accuracy (HPSMU)
Current Set. Set. Meas. Meas. Max. Range Reso. Accuracy Reso. Accuracy V
±1nA 100fA ±(0.5%+3pA+2fA×Vout) 10fA ±(0.5%+3pA+2fA×Vout) 200V ±10nA 1pA ±(0.5%+7pA+20fA ±100nA 10pA ±(0.12%+50pA+200fA×Vout) 100fA ±(0.1%+30pA+200fA×Vout 200V ±1µA 100pA ±(0.12%+400pA+2pA ±10µA 1nA ±(0.12%+5nA+20pA×Vout) 10pA ±(0.1%+3nA+20pA×Vout) 200V ±100µA 10nA ±(0.12%+40nA+200pA ±1mA 100nA ±(0.12%+500nA+2nA×Vout) 1nA ±(0.1%+300nA+2nA×Vout) 200V ±10mA 1µA ±(0.12%+4µA+20nA ±100mA 10µA ±(0.12%+50µA+200nA×Vout) 100nA ±(0.1%+30µA+200nA×Vout) 1 ±1A 100µA ±(0.5%+500µA+2µA×Vout) 1µA ±(0.5%+300µA+2µA×Vout) 2
1
200V (Iout £50mA), 100V (50mA<Iout£100mA)
2
200V (Iout £50mA), 100V (50mA<Iout£125mA), 40V (125mA<Iout£500mA), 20V (500mA<Iout£1mA)
Vout is the output voltage in volts. Iout is the output current in amps. For example, accuracy
specifications are given as ±% of set/measured value (0.1%) plus offset value (30pA+200fA for the 100nA range. The offset value consists of a fixed part determined by the set/measurement range and a proportional part that is multiplied by Vout.
×Vout) 10fA ±(0.5%+5pA+20fA×Vout) 200V
×Vout) 1pA ±(0.1%+200pA+2pA×Vout) 200V
×Vout) 100pA ±(0.1%+20nA+200pA×Vout 200V
×Vout) 10nA ±(0.1%+2µA+20nA×Vout) 200V
×Vout)
Output terminal/connection:
Dual triaxial connectors, Kelvin (remote sensing)
Voltage/Current Compliance (Limiting):
Voltage: 0V to ±200V Current: ±1pA to ±1A Compliance Accuracy: Same as the current (voltage) settling accuracy.
HPSMU Supplemental Information:
Maximum allowable cable resistance when using Kelvin connection: Force: 0.7W (100mA to 1A) Force: 10W (£100mA) Sense: 10W Typical voltage source output resistance (Force line/non-Kelvin connection): 0.2W Voltage measurement input resistance/current source output resistance:
13
³10
W (1nA range)
Current compliance setting accuracy foropposite polarity: 1nA to 10nA range: V/I setting accuracy ±12% of range 100nA to 1A range: V/I setting accuracy ±2.5% of range
HPSMU Measurement and Output Range
-200
-100 -40 -20
1000
500
125
50
-50
-125
-500
-1000
Current (mA)
Voltage (V)
100 20 40
200
4
PGU (Pulse Generator Unit) Specifications
Modes: Pulse or constant Amplitude: 0Vp-p to 40Vp-p Window: -40.0V to +40.0V Maximum current: ±100mA ±200mA (pulse width: £1ms, average current 100mA) Pulse width: 1.0µs to 9.99s Minimum resolution: 100ns Pulse period: 2.0µs to 10.0s Minimum resolution: 100ns Delay: 0s to 10s Minimum resolution: 100ns Transition time: 100ns to 10ms Minimum resolution: 1ns Output impedance: 50W or low impedance (£1W) Burst count range: 1 – 65535 Pulse parameter accuracy: Period: ±(2% +2ns) Width: ±(3% +2ns)
Delay: ±(2% +40ns) Transition time: ±(5% +10ns) Trigger output: Level: TTL Timing: Same timing and width as PGU1 pulse output
PGU Supplemental Information:
Overshoot: £±5% of amplitude ±10mV (50W output impedance to 50W load) Pulse width jitter: 0.2% + 100ps Pulse period jitter: 0.2% + 100ps Maximum slew rate: 100V/µs (50W output impedance to 50W load) Noise: 0.2% of range (@ DC output)
MPSMU Specifications
Same as 4155C MPSMU.
Pulse/DC Output Voltage and Accuracy (PGU)
Set Voltage Parameter Range Resolution Accuracy
Base ±20V 4mV ±(1% of Base +50mV +1% of Pulse)
±40V 8mV ±(1% of Base +50mV +1% of Pulse)
Pulse ±20V 4mV ±(3% of Base +50mV)
±40V 8mV ±(3% of Base +50mV)
Note: DC output is performed by the Base Parameter.
1
Accuracy is specified at leading edge - trailing edge = 1µs
1
Pulse Range and Pulse Parameter (PGU)
Range Period Width Delay Set resolution
1 2µs -100µs 1µs - 99.9µs 0 - 100µs 0.1µs 2 100µs - 1000µs 1µs - 999µs 0 - 1000µs 1µs 3 1ms - 10ms 0.01ms - 9.99ms 0 - 10ms 10µs 4 10ms - 100ms 0.1ms - 99.9ms 0 - 100ms 100µs 5 100ms - 1000ms 1ms - 999ms 0 - 1000ms 1ms 6 1s - 10s 0.01s - 9.99s 0 - 10s 10ms
Note: Pulse width is defined when leading time is equal to trailing time. PGU2 must be set in the same range as PGU1.
GNDU (Ground Unit) Specifications:
Output Voltage: 0V ±100µV Maximum sink current: 1.6A Output terminal/connection: Single triaxial connector, Kelvin (remote sensing)
GNDU Supplemental Information
Load Capacitance: £1µF Cable resistance: Force £1W Sense £10W
HRSMU, MPSMU, HPSMU Supplemental Information
Maximum capacitive load: 1000pF Maximum guard capacitance: 900pF Maximum shield capacitance: 5000pF Maximum guard offset voltage: ±1mV Noise characteristics (typical, Filter: ON): Voltage source noise: 0.01% of V range (rms) Current source noise: 0.1% of I range (rms) Voltage monitor noise: 0.02% of V range (p-p) Current monitor noise: 0.2% of I Output overshoot (typical, Filter: ON): Voltage source: 0.03% of V range Current source: 1% of I range Range switching transient noise (typical, Filter: ON): Voltage ranging: 250mV Current ranging: 10mV Maximum slew rate: 0.2V/µs
Leading/Trailing Edge Times (PGU)
Range Set Resolution` Accuracy
100ns - 1000ns 1ns ±(5% + 10ns)
0.5µs - 10µs 10ns ±(5% + 10ns)
5.0µs - 100µs 100ns ±(5% + 10ns) 50µs - 1000µs 1µs ±(5% + 10ns)
0.5ms - 10ms 10µs ±(5% + 10ns)
Restrictions: Pulse width < Pulse Period, Delay time < Pulse period, Leading time < Pulse width ´ 0.8 Trailing time < (Pulse period - Pulse width) ´ 0.8 Period, width, and delay of PGU1 and PGU2 must be in the same range. Leading time and trailing timefor a PGU must be in the same range.
5
Capacitance Calculation Accuracy (Supplemental Data)
Accuracy is derived from the current range, voltage range, capacitance measure­ment and leakage current measurement integration times, and the guard capaci­tance of cabling and step voltage. The information in the chart below is based on the following conditions: Voltage Range ±20V; Voltage Step: 100mV; Guard Capacitance : 100pF; Equivalent parallel resistance of DUT: 1 ´ 1015W. The ratio of integration times for capacitance measurement and leakage current measure­ment is 1:1.
HRSMU
Current Integration Max. Meas. Accuracy Range Time Value Resolution Reading % Offset
10pA/ 100pA
1nA 0.5sec 4.5nF 40fF 0.85 280fF
10nA
0.5sec 100pF/1pF 5fF 4.2 70fF 1sec 2pF/20pF 10fF 4.3 90fF 2sec 76pF/760pF 20fF 4.3 130fF
0.1sec 700pF 10fF 0.84 160fF
2sec 18nF 200fF 0.93 740fF
0.1sec 7nF 10fF 0.84 200fF
0.5sec 45nF 40fF 0.85 440fF 2sec 180nF 200fF 0.93 1.4pF 10sec 940nF 1pF 1.3 6.2pF
MPSMU
Integration Max. Meas. Accuracy
Current Time Value Resolution Reading % Offset
0.1sec 700pF 10fF 0.91 170fF
1nA 0.5sec 4.5nF 40fF 0.94 340fF
2sec 18nF 200fF 1.0 1pF
0.1sec 7nF 10fF 0.91 180fF
10nA 0.5sec 45nF 40fF 0.94 480fF
2sec 180nF 200fF 1.0 1.6pF 10sec 940nF 1pF 1.6 7.6pF
Current complicance must be smaller than the current range. The capacitance of the DUT and measurement path must be smaller than the maximum measurement value.
Functions
Measurement Setup
Setting
Fill-in-the-blanks using front-panel or
full-size external keyboard
Load settings from floppy disk or via
the LAN port
Program using internal Instrument
BASIC or via GPIB
HELP Function
Library: Default measure setup, Vce-
Ic, Vds-Id, Vgs-Id, and Vf-If are pre­defined softkeys
User-defined measurement setup
library
Auto file load function on power-up
Measurement
The 4155C and 4156C can perform dc or pulsed force/measure, and stress force. For dc, voltage/ current sweep and sampling (time domain) measurements are available.
6
Voltage/Current Sweep Measurement Characteristics
Each SMU and VSU can sweep using VAR1 (primary sweep), VAR2 (subordinate sweep), or VAR1 (synchronous sweep).
VAR1
Primary sweep controls the staircase (dc or pulsed) voltage or current sweep. Maximum number of steps: 1001 for one VAR1 sweep. Sweep type: linear or logarithmic Sweep direction: Single or double sweep Hold time: Initial wait time or wait time after VAR2 is set: 0 to 655.35s with 10ms resolution Delay time: Wait time from VAR1 step to the start of the measurement: 0 to
65.535s with 100µs resolution
VAR2
Subordinate linear staircase or linear pulsed sweep. After primary sweep is completed, the VAR2 unit output is incremented. Maximum number of steps: 128
VAR1
Staircase or pulse sweep synchronized with the VAR1 sweep. Sweep is made with a user specified ratio and offset value. VAR1 output is calculated as VAR1 = a ´ VAR1 + b, where “a” is the user specified ratio and “b” is the user specified offset value.
CONSTANT
A source unit can be set as a constant voltage or current source depending on the unit.
PULSE
One of the SMUs can be set as a pulse source. Pulse width: 0.5ms to 100ms, 100µs resolution. Pulse period: 5ms to 1s (³pulse width + 4ms), 100µs resolution. SMU pulse setting accuracy (supple­mental information, at fixed range measurement except multichannel measurement): Width: 0.5% + 50µs Period: 0.5% + 100µs Trigger output delay for pulsed measurement: 0 - 32.7ms with 100µs resolution (< pulse width).
Sampling (Time Domain) Measurement Characteristics
Displays the time sampled voltage/ current data versus time. Max. sampling points: 10,001 (linear) Sampling mode: linear, log, and thinned-out
Note: The thinned-out mode is similar to reverse-log sampling. Sampling measurement continues by thinning out older data until the sampling completion condition is satisfied.
Sampling interval range and resolution: Linear scale (auto mode):
60µs to 480µs range: 20µs resolution 480µs to 1s range: 80µs resolution
1s to 65.535s range: 2ms resolution Linear scale (no limit mode), log scale, and thinned-out modes:
560µs (720µs at thinned-out mode)
to 1s range: 80µs resolution
1s to 65.535s range: 2ms resolution
Note: The following conditions must be set when initial interval is less than 2ms.
Number of measurement channels: 1
Measurement ranging: fixed range
Stop condition: disable
Hold time: Initial wait time: 0.03s to 655.35s, 100µs resolution Sampling measurement stop condition:
A condition to stop the sampling can be defined. Sampling interval setting accuracy (supplemental data):
0.5% + 10µs (sampling interval £480µs)
0.5% + 10µs (480µs £sampling interval <2ms)
0.5% + 100µs (2ms £sampling interval)
C-V Measurement Characteristics
Capacitance is a calculated value derived from the following equation:
DQ
C =
DV
DQ is the change in charge when DV,
the step voltage, is applied by the SMU;
DQ is derived from the measurement
current (amps) and the integration time (seconds).
Maximum Measurable Value
Maximum measurable value depends on thecurrent range, integration time, and step voltage (refer to the chart in supplemental data).
Capacitance Calculation Accuracy
Accuracy is dependent on accuracy of the current measurement and voltage measurement and the stray capacitance and leakage current of measurement path, etc. (Refer to the chart in supple­mental data).
Zero Offset
Cancels stray capacitance of the fixtures and test leads.
Leakage Current Compensation
Cancels the inf luence of the leakage current to the capacitance measurement.
Stress Force Characteristics
SMU, VSU, and PGU output can be forced for the user specified period. Stress time set range: 500µs to 31,536,000s (365 days)
Resolution: 100µs (500µs £stress time £10s) 10ms (10s <stress time £31,536,000s) Burst pulse count: 1 - 65,535 (PGU only) Trigger: The 4155C and 4156C output a gate trigger while stress channels are forcing stress.
Knob Sweep
In knob sweep mode, sweep range is controlled instantaneously with the front-panel rotary knob. Only the Channel Definition page need be defined.
Standby Mode
SMUs in “Standby” remain programmed to their specified output value even as other units are reset for the next measurement.
Other Characteristics
Measurement Control: Single, append, repeat, and stop Stress Control: Stress force and stop SMU Setting Capabilities: Limited auto­ ranging, voltage/current compliance, power compliance, automatic sweep abort functions, self-test, and self­ calibration.
Arithmetic and Analysis Functions
Arithmetic Functions
User Functions
Up to six USER FUNCTIONS can be defined using arithmetic expressions. Measured data and analyzed variables from graphics analysis (marker, cursor, and line data) can be used in the computation. The results can be displayed on the LCD.
Arithmetic Operators
+, -, *, /, ^, LGT (logarithm, base 10), LOG (logarithm, base e), EXP (exponent), DELTA, DIFF (differential), INTEG (integration), MAVG (moving average), SQRT, ABS (absolute value), MAX, MIN, AVG (averaging), COND (conditional evaluation).
Physical Constants
Keyboard constants are stored in memory as follows: q:Electron Charge, 1.602177 E-19 C k:Boltzman’s Constant, 1.380658 E-23 e (e): Dielectric Constant of Vacuum,
8.854188 E-12
Engineering Units
The following unit symbols are also available on the keyboard: f (10
-12
p (10 m (10
), n (10-9 ), u or m (10-6 ),
-3
), K (103 ), M (10
Analysis Capabilities
Overlay Graph Comparison
A graphics plot can be stored and later recalled as an overlay plane. Four over­lay planes can be stored. One plane can be overlaid onto the current data.
Marker
Marker to min/max, interpolation, direct marker, and marker slip
Cursor
Long and short, direct cursor.
Line
Two lines, normal mode, grad mode, tangent mode, and regression mode.
Scaling
Auto scale and zoom.
Data Variable Display
Up to two user defined parameters can be displayed on the graphics screen.
Read Out Function
The read out functions are built-in functions for reading various values related to the marker, cursor, or line.
6
), G (10
-15
),
9
)
Automatic Analysis Function
On a graphics plot, the markers and lines can be automatically located using the auto analysis setup. Parameters can be automatically determined using automatic analysis, user function, and read out functions.
User Variable
Display the data on the LCD via GPIB or instrument BASIC.
Output
Display
Display Modes
Graphics and list.
Graphics Display
X-Y or X-Y1/Y2 plot of source current/ voltage, measured current/voltage, time, or calculated USER FUNCTION data.
List Display
Measurement data and calculated USER FUNCTION data are listed in conjunction with VAR1 step number or time domain sampling step number. Up to eight data sets can be displayed.
Display
8.4-inch diagonal color active matrix LCD, 640 dot (H) ´ 480 dot (V). More than 99.99% of the pixels on an LCD are active.
Hard Copy Functions
Graphics Hard Copy
Measured data and all data appearing on the LCD can be output via GPIB, parallel printer port, or network interface to supported HP plotters or printers. PCL, HR PCL (high-resolution PCL), and HP-GL formats are sup­ported (selectable).
Text Hard Copy
Print out setup information or mea­sured data list as ASCII text via GPIB, parallel printer port, or network interface to supported HP plotters or printers. PCL, HR PCL, and HP GL formats are supported (selectable).
Hard Copy File
Hard copy output can be stored to an internal or external mass storage device instead of sending it to a printer or plotter. The data can be stored in PCL, HR PCL, TIFF, HR TIFF (high­resolution TIFF), or HP GL formats.
Hard Copy via Network Interface
The network interface has lpr client capability.
High-Resolution (HR) Mode
This file mode is available for cases where an extremely clean print-out or plot is desired.
Note: High-resolution mode takes significantly greater CPU time to generate, so its use is recommended for final reports only.
7
Data Storage
Mass storage device: Built-in 3.5-inch floppy disk drive Media: 3.5-inch 2HD or 2DD diskette Format type: HP LIF and DOS User area: 1.44Mbyte (2HD) or 720Kbyte (2DD) File types:
Auto start program file, initial setup file, measurement setup file, mea surement setup/result file, stress setup file, customize file, hard copy data file, and Instrument BASIC
program and data file. Format of data made by the HP BASIC program: Data made by the HP BASIC program and data made by the Instrument BASIC program are compatible. Network mass storage device: An NFS mountable mass storage device File types: Auto start program file, initial setup file, measurement setup file, measure­ ment setup/result file, stresS setup file, customize file, and hard copy data file. Maximum number of files allowed per directory on network mass storage device: 199 Data storage (supplemental data): 2HD DOS format: Available bytes: 1457K (byte) File size: Measurement setup: 3843 (byte) Stress setup: 601 (byte) Measurement setup/result (Typical data): 15387 (byte) (VAR1: 101, VAR2: 5) Customized system setup: 1661 (byte) Hardcopy data: 30317 (byte) (Monochrome PCL 75DPI file) Hardcopy data: 38702 (byte) (monochrome TIFF file)
Note: For LIF format, the total number of files is limited to 199.
Repeating and Automating Test
Instrument Control
Agilent 4155C and 4156C function control: Internal or external computer controls the 4155C and 4156C functions via the GPIB interface Command sets: SCPI command set Agilent FLEX command set Agilent 4145B command set Program Memory: Using the Agilent FLEX command set, the user can store program code in the 4155C or the 4156C. The maximum number of subprograms is 255 (8 bit). External instrument remote control: Control external equipment via the GPIB interface.
8
Instrument BASIC
Instrument BASIC is a subset of HP BASIC. Functions: Arithmetic operation, binary opera tion, string manipulation, logical operation, array operation, program flow control, event-initiated branch ing, program editing and debugging support, mass storage operation, instrument control, real-time clock, softkey operation, and graphics. Agilent 4145B automatic sequence program (ASP) typing aid:
4145B ASP-like syntax softkeys are available in instrument BASIC. A 4145B ASP file cannot be read by the 4155C or 4156C.
Remote control:
Instrument BASIC is remote controllable from an external
computer via the GPIB interface. Instrument BASIC memory area (supplemental data):
Program (text) area: 16K (byte)
Variable/stack area: 500K (byte)
Common variable area: 600K (byte)
Note: The memory size for common variable is decreased when hard copy or disk operation is performed.
Trigger
Input:
External trigger input starts a sweep or sampling measurement or can be used as a trigger input for continuing
an Instrument BASIC program. Input Level: TTL level, negative or positive edge trigger Output: External edge trigger outputs can be generated by the start of a sweep measurement, the start of each sweep step in a staircase sweep, the start of each pulse leading edge for an SMU in pulse mode, and the issuance of an an IBASIC trigger out command execution. In addition, you can set the trigger signal to be active during the Stress Force State. If you have a 41501A/B with PGU option, you can output a synchronized trigger output through the 41501A/B trigger output. Output Level: TTL level, negative or positive logic
4145B Data Compatibility and Syntax Commands
Setup and data file
Measurement setup and data from the 4145B can be loaded.
GPIB program
GPIB programs for the 4145B can be used when the 4145B command set is selected.
Note: There is a possibility that GPIB programs for the 4145B will need to be modified.
Interfaces
GPIB interface: SH1, AH1, T6, L4, SR1, RL1, PP0,
DC1, DT1, C1, C2, C3, C4, C11, E2 Parallel interface: Centronics RJ45: Ethernet IEEE 802.3 10BASE-T for a 10Mbps CSMA/CD local area network External keyboard: Compatible PC-style 101-key keyboard (mini DIN connector) Interlock and LED connector R-BOX control connector Trigger in/out SMU/PGU selector control connector (41501B)
Sample Application Programs
Flash EEPROM test TDDB Constant I (Electromigration test) V-Ramp test J-Ramp test SWEAT GO/NO-GO test HCI degradation test Charging pump test
Sample VEE Program
Vth measurement using the 4155C or 4156C, the E5250A, and a wafer prober.
VXI plug&play Drivers
VXI plug&play drivers for the 4155C and 4156C
Supported VXI plug&play operating systems:
Microsoft Windows 95, 98, NT, 2000 Professional, and XP Professional
Format
Tree-structured function panel. Panel mode for hardware configura­tion and manual parameter setting. Parameter mode for variable definition and I/O configuration.
General Specifications
Temperature range
Operating: +10°C to +40°C (if using floppy disk drive) +5°C to +40°C (if not using floppy disk drive) Storage: -22°C to +60°C
Humidity range
Operating:
20% to 80% RH, non-condensing and wet bulb temperature £29°C (if using floppy disk drive) 15% to 80% RH, non-condensing and wet bulb temperature £29°C (if not using floppy disk drive) Storage: 5% to 90% RH , non-condensing and wet bulb temperature £39°C
Altitude
Operating: 0 to 2,000 m (6,561 ft) Storage: 0 to 4,600 m (15,091 ft)
Power requirement
90V to 264V, 47 to 63 Hz
Maximum VA
4155C and 4156C: 450VA 41501B: 350 VA
Regulatory Compliance
EMC: EN 61326-1:+A1, AS/NZS 2064.1 Safety:
CSA C22.2 NO.1010.1 (1992),
IEC 61010-1:+A2/EN 61010-1:+A2
UL3111-1:1994 Certification: CE, CSA, NRTL/C, C-Tick
Dimensions
4155C and 4156C: 235mm H ´ 426mm W ´ 600mm D 41501B: 190mm H ´ 426mm W ´ 600mm D
Weight (approx.)
4155C and 4156C: 21kg 41501B: 16kg (option 412, HPSMU + 2 ´ PGU)
4155C and 4156C Furnished Accessories
Triaxial cable, 4 ea. (4155C) Kelvin triaxial cable, 4 ea. (4156C) Coaxial cable, 4 ea. Interlock cable, 1 ea. Keyboard, 1 ea. User manual, 1 set Sample application program disk, 1 ea. Sample VEE program disk, 1 ea. VXIplug&play drivers disk for the 4155C and 4156C, 1 ea. VXIplug&play drivers disk for the E5250A, 1 ea. LAN Interface Test Adapter, 1 ea.
Accessory Specifications
Specification Condition
The “supplemental information” and “typical”entries in the following specifi­cations are not warranted, but provide useful information about the functions and performance of the instruments (23°C ±5°, 50% RH).
16440A SMU/Pulse Generator Selector
The 16440A switches either an SMU or PGU to the associated output port. You can expand to 4 channels by adding an additional 16440A. The channel 1 PGU port provides a “PGU OPEN” function, which can disconnect the PGU by opening a semiconductor relay. The 16440A cannot work without two pulse generator units of the
41501A/B (SMU and Pulse Generator Expander). Channel configurations: Two channels (CH1, CH2) CH1: INPUT ports: 2 (SMU and PGU, PGU port has additional series semiconductor relay) OUTPUT port: 1 CH2: INPUT ports: 2 (SMU and PGU) OUTPUT port: 1
Voltage and Current Range
Input port Max. V Max. I
SMU 200V 1.0A PGU 40V 0.2A (AC Peak)
Supplemental Information (at 23°C ± 5°C, 50%RH)
SMU port leakage current: < 100fA @100V SMU port residual resistance (typical):
0.2W SMU port stray capacitance (typical @1MHz): Force « Common: 0.3pF Force « Guard: 15pF Guard « Common: 130pF PGU port residual resistance: 3.4W PGU port OFF capacitance (typical): 5pF PGU port OPEN capacitance (typical): 700pF (@ 1MHz, Vin - Vout = 0V)
PGU port signal transfer characteristics
Overshoot: < 5% of pulse amplitude (@20ns leading and trailing time, 50W pulse generator source impedance, 50pF and 1MW in parallel load).
General Specifications
Dimensions: 50mm H ´ 250mm W ´ 275mm D Approximate weight: 1.1kg
16441A R-BOX
The 16441A R-BOX adds a selectable series resistor to the SMU output. You can select the resistor from the setup page, and the voltage drop due to the series resistor is automaticallycompen­sated for in the measurement result. Measurement limitations with the 4155C and 4156C and R-BOX:
If you measure device characteristics including negative resistance over 1MW with the 4155C/4156C and R-BOX, there is a possibility that they cannot be measured. There is a possibility that the 4155C and 4156C cannot perform measurements because of DUT oscillations even with the R-BOX. Whether oscillation occurs or not depends upon the DUT and
measurement conditions. Number of SMU channels that can add a resistor: 2 Resistor values: 1MW, 100kW, 10kW, 0W (each channel) Resistance accuracy:
0.3% (at 23°C ±5°C, between input­ output terminal) Maximum voltage: 200V Maximum current: 1A (0W selected) Kelvin connection: Kelvin connection is effective only when 0W is selected.
Supplemental Information (at 23°C ± 5°C, 50%RH)
Leakage current: <100fA @ 100V
General Specifications
Dimensions: 72mm H ´ 250mm W ´ 270mm D Approximate weight: 1.6kg
16442A Test Fixture
Channel Information
SMU: 6 channels (1 triaxial connector per channel) 3 channels (1 Kelvin triaxial connector per channel) VSU: 2 channels (1 BNC connector per channel) VMU: 2 channels (1 BNC connector per channel) PGU: 2 channels (1 BNC connector per channel) GNDU: 1 channel (1 triaxial connector) INTLK: 6-pin connector
Supplemental Information (at 23°C ± 5°C, 50% RH)
SMU channel: Leakage current: 10pA max @200V (Force or Sense « Common) Stray capacitance: 15pF max (Force or Sense « Common) Stray capacitance: 3pF typical (Force or Sense « Other SMU) Residual resistance: 60mW typical (Force, Sense) Guard capacitance: 70pF max (Force or Sense « Guard) VSU channel residual resistance: 60mW typical VMU channel residual resistance: 60mW typical PGU channel characteristic impedance: 50mW typical GNDU channel residual resistance: 40mW typical (Force, Sense)
General Specifications
Temperature range: Operating: +5°C to +40°C Storage: -40°C to +70°C Humidity range; Operating: 5% to 80% RH (no condensation) Storage: 5% to 90% RH at 65°C (no condensation) Dimensions: 140 mm H ´ 260 mm W ´ 260 mm D Weight (approx.): 2.5kg
9
Automation Software
I/CV 2.1 Lite
Overview
Agilent I/CV 2.1 Lite provides automated test solutions for semi­conductor characterization. It sup­ports the Agilent 4155C and 4156C, the Agilent E5270 Series, the Agilent E5250A Low Leakage Switch, the Keithley 707 Switch­ing Matrix, the Agilent 4284A and 4294A LCR meters, and many pop­ular semiautomatic wafer probers. I/CV 2.1 Lite also provides wizard­based test development, test ex­ecution, and sequencing along with data logging and post- analysis tools on Microsoft® Windows.®
10
Software Functions
Interactive Measurements
I/CV 2.1 Lite includes Agilent ICS as the default measurement tool. ICS provides point-and-click measurements, intuitive matrix control, and graphical analysis capabilities for semiconductor parametric measurements. Created setups can be used as measurement algorithms in the script editor.
Script Editor
The script editor provides a wizard-based interface for build­ing test scripts used in the execu­tion of automated tests. It allows access to libraries of built- in software components that sup­port functions for creating test plans. Components include:
• Automated sub-die prober movement
• Switch connection execution
• Test algorithm execution
• Pass/Fail determination and
processing
• Conditional branching: IF, ELSE
• Looping: FOR, WHILE
• User variable creation
• User prompts
• Message displays
• Test script commenting
Wafer Prober Navigation
I/CV 2.1 Lite provides support for popular semiautomatic prob­ers as well as several automatic probers. Probe plans can be defined that include sub-die movement for performing auto­mated test of multiple modules or individual devices across a wafer. Interactive prober control can also be implemented for analytical applications.
Test Execution
Test scripts can be executed for either manual or automated tests. Manual tests are used for single devices or single modules (which can include several devices) on a manual prober. Automated tests are used for wafer tests combing semiautomatic prober control with die or module test scripts. Test wizards provide step-by-step instructions for entering runtime information, selection of wafer navigation plans, selection of test plans, and starting a test.
Auto-Analysis and Test Reporting
Parametric quantities from test data can be extracted and standard reports and graphs can be generated. Supported graphs and reports include:
• Color wafer maps
• Histograms
• Parameter statistics
• Parametric values vs. die location
• Tables of I-V or C-V curve data
Software Measurement Tool Support
Test algorithms can be created using the following tools:
• Agilent ICS
• Microsoft VBScript (resident in the script editor)
Computer System Requirements
Operating System
Microsoft Windows 2000 Professional or XP Professional with Service Pack 1
CPU
300 MHz Pentium II-class (500 MHz Pentium III-class or faster recommended)
Hard Disk
5 GB available space
(20 GB recommended)
Memory
128 MB for Windows 2000 Professional (256 MB recommended)
256 MB for Windows XP Professional
Disk Drive
CD-ROM
Software Security
Parallel or USB port required to attach security key
Control I/F
Supported GPIB card (see requirements below)
GPIB Card Support
Agilent
Card
82341C (ISA) 82357A
(USB/GPIB)
Agilent I/O Library L.02.01 required
National Instruments
Card
PCI-GPIB
GPIB-USB-A
Windows 2000 Windows XP Pro.
Professional (Service Pack 1)
X
X X
Windows 2000 Windows XP Pro.
Professional (Service Pack 1)
X X
X X
Prober Support
Cascade Microtech
S 300 with Nucleus version 2.1 or 2.5 Summit 12k with Nucleus version
2.1 or 2.5
Electroglas
2001 and 408X
SUSS MicroTec
All SUSS MicroTec probe stations using Prober Bench NT v4.2
Vector Semiconductor
AX-2000 / VX-3000, Version 3.2 or later
Supported Measurement Instruments
• E5270 Series of Parametric Measurement Solutions
• 4155A/B/C Semiconductor Parameter Analyzer
• 4156A/B/C Precision Semicon­ ductor Parameter Analyzer
• 4284A Precision LCR Meter
• 4294A Impedance Analyzer*
• E5250A Low Leakage Switch Mainframe
• Keithley 707 Switch
* VBScript libraries are supplied.
11
For more information about Agilent and its products, go to www.agilent.com.
For more information about Agilent Technologies semiconductor test products, applications, and services, visit our website: www.agilent.com/ go/semiconductor or you can call one of the centers listed and ask to speak with a semiconductor test sales representative.
Americas
Brazil (11) 4197-3600 Canada (French) 1 877 894-4414 Canada (English) 1 800 447-8378 Mexico 33 134-5841 United States 1 800 447-8378
Asia/Asia Pacific Australia 1 800 629-485 China 1 800 276-3059 Hong Kong 852 2599 7889 India 91/11 690-6156 Japan 0120 421-345 Malaysia 1 800 880-780 New Zealand 0 800 738 378 Philippines 1 800 1651-0135 Singapore 1 800 276-3059 South Korea 080 778-0011 Taiwan 0 800 047-662 Thailand 1 800 2758-5822
Europe Austria (01) 25 125-7183 Belgium (0) 2 404-9380 Denmark 080301040 Finland 20 547-9999 France (0) 825 010710 Germany (0) 18 05 24-63 34 Greece 20 547-9999 Ireland 016158393 Italy 02 92 60 8333 Luxembourg (0) 2 404-9340 Netherlands (0) 20 547-9999 Poland 20 547-9999 Russia 20 547-9999 Spain 91 631 3383 Sweden 020 120-9975 Switzerland (Italian) (0) 2 92 60 8484 Switzerland (German) (0) 1 735-9300 Switzerland (French) (0) 825 010 700 United Kingdom (0) 7004 222-222
Middle East Israel 20 547-9999
Technical data subject to change without notice. Microsoft and Windows are U.S. registered trademarks of Microsoft Corporation. © Copyright 2002 Agilent Technologies Printed in USA April 21, 2003
5988-9238EN
S1
Loading...