Agilent 4100, 4200, 4210 User Manual

Agilent Microwave Plasma Atomic Emission
Models 4100, 4200, and 4210
NOTE
NOTE
ON
Spectrometers
Long Term Shut Down and Start up Instructions

Shut down the system

To shut down the system:
1 Rinse the spray chamber by aspirating water (or the appropriate solvent for your application)
for a few minutes.
When running organic samples, it is recommended that the spray chamber be cleaned and dried thoroughly between analyses.
2 Remove solution tubing from the solution. 3 Pump all solution from the sample line, nebulizer and spray chamber with the peristaltic pump
and then continue to run for another 30 seconds to ensure that the sample introduction system is solution free.
4 Choose Plasma Off from the Plasma button drop-down arrow to extinguish the plasma.
The peristaltic pump stops automatically when the plasma is extinguished.
The green Plasma enable button, located on the front of the MP-AES instrument, is intended to be used only in an emergency. It is not intended to be used every time you want to extinguish the plasma.
If it is used to extinguish the plasma, you will need to reset the Plasma enable button to the position before the plasma can be re-ignited.
5 Loosen the pressure on the peristaltic pump tubing by releasing the pressure bars and lifting
the tubing out of the grooves in the tube retainer to increase the pump tubing lifetime. To do this:
a Push up the pressure bar tensioners (1 in Figure 1).
This releases the tensioner from the pressure bar.
b Swing the pressure bar backwards (2 in Figure 1). c Lift the tubing out of the grooves in the tube retainer (3 in Figure 1).
6 Dispose of all pump tubes. 7 Remove all hazardous materials and samples from in and around the equipment and store
appropriately. Special precautions are required for HF applications systems.
Agilent Microwave Plasma Atomic Emission Spectrometers Long Term Shut Down and Start up Instructions
WARNING
NOTE
search for “Agilent
2
3
1
Warning – Hazardous Materials
Perform the appropriate decontamination procedure if hazardous material has been spilled on or inside the instrument.
8 Clean any sample spills. 9 Remove sample introduction components (plasma torch, spray chamber and nebulizer) and
store in original packaging.
10 Turn off power to the instrument at the instrument ON/OFF switch. 11 Close all open applications and shut down the computer. 12 Shut off all gas supplies at the regulators/isolation valves and gas cylinders where applicable. 13 Disconnect the instrument from the mains power. 14 Turn off the laboratory exhaust system. 15 Shutdown and disconnect any accessories according to the instructions provided in their
Figure 1 Peristaltic pump components
manual.
1 Pressure bar tensioner 2. Pressure bar 3. Tube retainer
The torch should be cleaned of any deposits, dirt or residue.
For further information about cleaning the torch see the MP-AES User’s Guide or MP-AES Easy Fit Torch Maintenance” on www.agilent.com

Start up the system

To start up the system:
1 Check that the exhaust and intake lines are secured to the MP-AES instrument. 2 Ensure the gas lines are connected to the MP-AES instrument and that the gas supplies are
turned on and set to the correct pressures.
3 Check that the USB and power cables are plugged in. 4 Check that the pre-optics window is clean and correctly installed and that the interlock is
engaged.
5 Check that the torch is clean and in good condition. 6 Insert the torch and completely close the torch handle. 7 Insert the nebulizer into the spray chamber. 8 Fit the spray chamber socket to the ball joint on the base of the torch and secure using the
torch clamp.
2
Loading...
+ 2 hidden pages