This service manual provides procedures for testing and maintaining the Aeonmed model
anesthetic machine. It is not intended to be a complete maintenance document; therefore, it
contains no disassembly or reassembly instructions.
Refer any repairs or adjustments that exceed the scope of this manual to the Service Center of
Beijing Aeonmed Co., Ltd by calling
800-810-8333 or 86-10-88799987-333
This manual contains proprietary information. It is intended for use only by individuals qualified in
the installation and maintenance of the Aeonmed anesthetic machine. Receipt, purchase, or
possession of this document in no way confers or transfers any other rights for the use of this
information. Disclosure or reproduction of the enclosed, without the written permission of Beijing
Aeonmed Co., Ltd is prohibited.
This manual is intended for use only by technicians who have successfully completed Beijing
Aeonmed Co., Ltd training on this product.
Beijing Aeonmed Co., Ltd believes the information herein is accurate but accepts no responsibility
for errors, omissions, or misrepresentation.
1
Aeon7400A Service Manual
1.1 Symbols
Warnings and Cautions indicate all the possible dangers in case of violation of the
stipulations in this manual. Refer to and follow them.
WARNING: indicates potential hazards to operators or patients
CAUTION: indicates potential damage to equipment
Instead of illustrations, other symbols may also be utilized. Not all of them may necessarily appear
in the equipment and manual. The symbols include:
ON(Power)
OFF(Power)
Alternating Current
Protectively earth
Equipotential
Dangerous Voltage
Movement in one direction
SN
Type B equipment
Serial Number
Warning or Caution, ISO
7000-0434
NOTE: refer to the manual,
IEC601-1
This way up
Oxygen flush Button
Lock
Movement in two directions
Inspiration flow
2
Unlock
Expiration flow
Aeon7400A Service Manual
Alarm Silence
Ventilator operate
Directions of Drain Valve
1.2 Warranty considerations
View the reading on the top of
float
Bag operate
Do not make any service repairs on this equipment during the states warranty period. Any
unauthorized work immediately voids the warranty. Aeonmed will not be liable for any repairs
attempted by the owner. Any such attempted repairs other than specified no warranty repairs void
the warranty.
3
Aeon7400A Service Manual
2 Description
This chapter describes the systems, features, controls, and labels on the Aeonmed
anesthetic machine. Also included is a description of the table of technical specifications.
2.1 General description
The Aeon7400A range of continuous flow inhalation anesthetic machines are intended for use
with human patients, in conjunction with anesthetic vaporizers, breathing hoses and patient
connection fittings which comply with the relevant ISO standard or equivalent.
Depending upon the patient circuit selected, the machines can be used in semi closed or closed
circuit configurations.
Aeon7400A is applicable for patient of over 300mL with standard set and for child of over 80mL
with bellows assembly and circles of child.
Aeon7400A equipped with MV200 microprocessor ventilator is pneumatic, electronic-controlled,
pressure-limit and time-cycled ventilator with perfect IPPV ventilating mode and monitoring
system.
2.2 Features of Aeon7400A
Feature Aeon7400A
Model of ventilator MV200
Mode of ventilation IPPV/Manual
I:E ratio 2:1 to 1:4
Respiration frequency 6 to 60bpm
Tidal volume 0 to 1200ml
Power failure alarm Yes
High pressure 20 to 60cmH2O
Low pressure 5 to 20cmH2O
Alarm silence Not more than 120senconds
vaporizer VP200
Circle absorber Yes
Bellows Adult/pediatric
Flowmeter Four tubes, O2/N2O(cascade/continuous)
4
Aeon7400A Service Manual
2.3 Specification
Power
Power line
Fuses
Power usage <=50W
Type Positive pressure volume
shock
Gas input
Gas type O2/N2O
Gas pressure 0.3-0.5Mpa
Environment
Ambient temperature
Relative humidity Not more than 80%
Atmospheric pressure 96kPa to104kPa
Noise of whole set <65dB(A)
Warming up no less than 5min
Miscellaneous
Dimensions 1350(H)×620(D)×650(W)mm
Work surface height 840mm
Writing tablet 540×300mm
Weight 120kg (265lb)
220V±22V, 50Hz±1Hz
φ5×20, 0.5A(F)
φ5×20, 5A(F)
φ5×20, 2A(F)
Type of protection: class I Protection against electrical
Degree of protection: Type BF
+5 °C to +40°C
5
Aeon7400A Service Manual
3 Checking
3.1 Preoperative Checkout procedures
Test interval Preoperative Checkout should be done in the following situation:
Before the first patient’s use everyday.
Before every patient’s use.
Perform the programs according to requirements after repair or maintenance.
Test schedule is given in the table below:
Before use of the first patient each day Before use of each patient
System check:
Power failure alarm test:
Gas pipeline and gas cylinder test:
Flow control test:
Vaporizer installation and test
Alarm test:
Breathing system test:
Ventilator test:
WARNING: Do not use this system before the operation and maintenance manual
are read and understood.
• Whole system connection
• All warnings and cautions
• Using guide of each system module
Breathing system test:
Ventilator test:
•Testing method of each system module
Before using this system:
• Complete all tests of this section
• Test all the rest of system modules
If test failure, do not use this system. Please contact service representative.
6
Aeon7400A Service Manual
3.1.1 System Checkout
WARNING: make sure the breathing circuit is connected correctly and in good
condition.
Make sure:
1 Equipment is in good condition.
2 All the components are correctly connected.
3 Breathing circuit is correctly connected and in good condition; there is sufficient
absorbent in the breathing system.
4 Vaporizer is in lock position and is filled with sufficient anesthetic.
5 The connection and pressure of pipeline gas supply system are correct.
6 The connected cylinder valve should be closed if there are backup cylinders.
WARNING: Do not leave the cylinder valves open during pipeline gas supply
period; otherwise, cylinder gas supply will be used up and lead to
insufficient supply in case of pipeline malfunction.
7. The required emergency device is ready and in good condition.
8. The device for airway maintenance, organ cannula are ready and in good condition.
9. The applicable anesthetic and emergency medicine are ready.
10. Make sure the truckles are tight and locked and free of motion.
11. Connect the power cord to the AC power outlet. The power indicator light will light up
when power is connected.
If failure, that means no electric power supplying. Exchange other sockets, close breaker,
or replace power cord.
3.1.2 Mains failure alarm test
1 Turn power switch to “I”, stand-by interface appears after self-test.
2 After operating 5 minutes, pull out power cord.
3 Make sure that power off failure alarm occurs, it has the following characteristics:
• Alarm sound;
4 Connect power cord again.
5 Make sure the alarm eliminate.
7
Aeon7400A Service Manual
3.2 Testing the gas supply pipeline and the gas cylinder
CAUTION: A user must confirm that gas supply is connected correctly; there is
no any disconnection, leakage, faulty connection in gas circuits and
pressure indicates correctly. Stop using and check gas connections if
abnormal.
CAUTION: To prevent from damage:
Open cylinder valve slowly.
Never control the flow with excessive force.
Skip step 2 if the system is not using cylinder gas supply.
1 Disconnect all pipeline gas supply and close all the cylinder valves.
•If the readings of the pipeline pressure gauge and cylinder pressure gauge are not
zero.
•Switch on O
supply.
2
• Adjust flow control to middle range.
• Make sure all the pressure gauges are reset to zero except the O
• Switch off O
• Make sure the O
supply.
2
pressure gauge is reset to zero. The low O2 supply alarm should
2
pressure gauge.
2
be on when pressure drops.
2 Make sure cylinders are fully filled:
• Open each cylinder valve.
• Make sure the pressure of each cylinder is high enough. In case the pressure is
insufficient, close the corresponding cylinder valve and install a fully filled cylinder.
3 Test cylinder high pressure leak one by one.
4 Close flowmeters.
5 Open the cylinders.
6 Record the cylinder pressures.
7 Close the cylinder valves.
8
Aeon7400A Service Manual
8 Begin to record the pressures after one minute. If O
pressure drops to 5000 kPa, it
2
means there is a leakage:
•If leakage exists, according to direction of section 5.5, replace a new sheet gasket,
and then tighten T handle.
•Perform this step again. If leakage exists all the same, do not use this system.
9 Step 5 ~ 7 should be repeated for all the cylinders. N
O pressure drop in one minute
2
should not exceed 700 kPa.
10 Close all the cylinder valves.
CAUTION: Do not leave the cylinder valves open during pipeline gas supply
period; otherwise, cylinder gas supply will be used up and lead to
insufficient supply in case of pipeline malfunction.
11 Connect pipeline gas supply.
12 Check pipeline pressure according to the table below:
ANSI (U.S. and International), Australia, Canada,
345 kPa (50 psig)
France and Japan
ISO, Italy, Scandinavia, South Africa, Spain and Switzerland 414 kPa (60 psig)
Austria and Germany 500 kPa (75 psig)
9
Aeon7400A Service Manual
3.3 Monitoring Flow Control
WARNING: Refer to Step 1 to 13 of monitoring without oxygen for monitoring
without oxygen.
Refer to Step 1 to 13 of monitoring with oxygen for monitoring with
oxygen.
3.3.1 Monitoring without oxygen
WARNING: The monitoring system cannot be replaced by link system. The fresh
gas containing enough oxygen may not avoid the existence of low
oxygen mixture in the breathing circuit.
If N
O exists, it will pass through the system during the test, which
2
should be securely collected and removed.
Patients may be injured by improper gas mixture. The link system
should not be used if a proper ratio of O
and N2O is not possible.
2
The following procedures can test whether the link system has
serious malfunction; however, it cannot determine whether the
calibration is correct.
CAUTION: The gas flow switch should be adjusted slowly. Do not turn it hard
when the reading of the flowmeter goes beyond the maximum or
minimum flow rate; otherwise, the control valve can be damaged and
the control will not work.
Follow the steps to test the flow control:
1. Connect the pipeline gas supply or open the cylinder valves slowly.
2. Turn clockwise all the flow control till the end (minimum flow).
3. Turn on mains switch.
4. Make sure:
• The oxygen flow is between 25 mL/min and 75 mL/min.
• No gas flowing in any other flow tube.
• Step 5 and step 6 are only applicable for the N
10
O system test.
2
Aeon7400A Service Manual
WARNING: During Step 5 to Step 6, keep link systems working state.
Only adjust testing of control (N
Adjust flow according to order (N
If adjustable range exceeds, please adjust flow control to the nearest
place and perform this step again.
5. To test the flow increase of the link system:
• Turn clockwise the N
• Turn counterclockwise the N
• Set N
O flow control to the rate described in the following table. The O2 flow must be
2
O and O2 flow control till the end (minimum flow).
2
O flow control slowly.
2
higher than the minimum flow limit.
Set N
O flow to
2
O2 flow must be higher than the minimum flow
(liters per minute):
0.6 0.2
1.5 0.5
3 1.0
7.5 2.5
O in step 5 and O2 in step 6).
2
O firstly O2 secondly).
2
(liters per minute):
6. This step tests the function of the Link System when flow is reduced, you should:
Set N
(liters per minute):
O flow to
2
O2 flow must be higher than the minimum flow
(liters per minute):
6.0 2.0
3.0 1.0
0.6 0.2
7. Adjust full flow of all the gas to ensure that the flowmeter float must move smoothly.
8. Shut off the oxygen supply either by closing the oxygen cylinder valve, or by disconnecting
the oxygen pipeline supply.
9. Make sure:
• As pressure decreases, the oxygen-supply failure alarm must continuously sound.
• Disconnect the flow of nitrous oxide and oxygen to be sure that the oxygen flow will be
the last to stop.
•If the oxygen is the driving gas of the ventilator, the oxygen-supply failure alarm must
continuously sound.
10. Turn all of the flow control valve knobs completely clockwise to the minimum flow.
11. Reconnect oxygen pipeline supplies or slowly open the oxygen cylinder valve.
12. Turn off mains supply.
11
Aeon7400A Service Manual
3.3.2 Monitoring with Oxygen
WARNING: The monitoring system cannot be replaced by link system. The fresh
gas containing enough oxygen may not avoid the existence of low
oxygen mixture in the breathing circuit.
O exists, it will pass through the system during the test, which
If N
2
should be securely collected and removed according to safe and
eligible methods.
Patients may be injured by improper gas mixture. The link system
should not be used if a proper ratio of O
CAUTION: Before continuous testing, perform test of the O2 monitoring device
according to step 8 in section 3.6.
and N2O is not possible.
2
Follow the steps to test the flow control:
1. Connect the pipeline gas supplies, or slowly open the cylinder valve.
2. Turn all of the flow control valve knobs completely clockwise to the minimum flow.
3. Turn on mains switch.
4. Make sure:
• The oxygen flow is between 25 mL/min and 75 mL/min.
• No gas flowing in any other flow tube.
• Step 5 and step 6 are only applicable for the N
O system test.
2
WARNING: During Step 5 to Step 6, keep link systems working state.
Only adjust testing of control (N
Adjust flow according to order (N
O in step 5 and O2 in step 6).
2
O firstly O2 secondly).
2
The oxygen sensor being used must be calibrated correctly.
12
Aeon7400A Service Manual
5 To test the flow increase of the link system:
• Turn clockwise the N
• Turn counterclockwise the N
O and O2 flow control till the end (minimum flow).
2
O flow control slowly.
2
•Make sure that the oxygen flow is increasing. The concentration of the oxygen tested
must ≥ 21% during the complete process.
6 To test the flow increase of the link system:
• Set the nitrous oxide flow to 9.0 L/min.
• Set the oxygen flow to 3/min or higher.
• Turn the flow control valve knob of the oxygen clockwise slowly.
• Be sure that the oxygen flow is getting reduced. The concentration of the oxygen tested
must ≥ 21% during the complete process.
7 Adjust all of the gas full flow to ensure that the flowmeter floats must move smoothly.
8 Shut off the oxygen supply either by closing the oxygen cylinder valve, or by disconnecting
the oxygen pipeline supply.
9 Make sure:
• As pressure decreases, the oxygen-supply failure alarm must continuously sound.
• Disconnect the flow of nitrous oxide and oxygen to be sure that the oxygen flow will be
the last to stop.
•If oxygen is the driving gas of the ventilator, the oxygen-supply failure alarm must
continuously sound.
10 Turn all of the flow control valve knobs completely clockwise to the minimum flow.
11 Reconnect oxygen pipeline supplies or open the oxygen cylinder valve slowly.
12 Turn off mains supply.
13
Aeon7400A Service Manual
3.4 Installing and testing of vaporizer
3.4.1 Installation
WARNING: Do not take the vaporizer away from the bypass valve with its locking
lever locked.
Do not use more than one vaporizer at the same time in this system.
Install vaporizers in accordance with the following steps:
1. The vaporizer must be disassembled and reinstalled if its top is not horizontal.
2. Set the locking lever of the vaporizer so that it is locked.
3. Try to lift the vaporizer directly upwards so as to separate itself from the bypass valve, but do
not pull the vaporizer forwards. Be careful not to rotate it on the bypass valve.
4. As the vaporizer is taken away from the bypass valve, reinstall the vaporizer and then follow
step 1 to step 3. Do not use this system if you cannot put return the vaporizer to a horizontal
position on the bypass valve.
5. Try on opening two vaporizers at the same time.
• Testing any possible instance of each combination.
• If more than one vaporizer can be opened at the same time, disassemble and reinstall
them, then perform step 1 to step 5.
3.4.2 Testing Vaporizer Back Pressure
CAUTION: About performance testing of vaporizer refer to relevant instruction
for use.
14
Aeon7400A Service Manual
3.5 Testing alarm
1 Connect reservoir bag to patient end.
2 Set bag/ventilator switch to ventilator control.