Aemc instruments SR752, SR751 User Manual

AC Current Probe Models SR751 and SR752
User Manual
DESCRIPTION:
The SR751/SR752 (Catalog #2116.31 and #2116.32) are designed for use in industrial environments. The ergonomic design allows them to easily attach to cables or small bus bars. The “circular” jaws guarantee a very good accuracy and low phase shift. The probes have a measurement range up to 1000Arms. continuous and are compatible with any AC voltmeter, multimeter, or other voltage measurement instrument that has the following features: Range and resolution capable of displaying 1mV of output per amp of measured current; voltmeter accuracy of 0.75% or better to take full advantage of the accuracy of the probe; Input impedance of 1M.
WARNING
The safety warnings are provided to ensure the safety of personnel and proper operation of the instrument. Read the instruction completely.
• Use caution on any circuit: potentially high voltages and currents may be present and may pose a shock hazard.
• Do not use the probe if damaged. Always connect the current probe to the measuring device before it is connected around the conductor
• Do not use on non-insulated conductor with a potential to ground greater than 600V CAT III pollution
2. Use extreme caution when clamping around bare conductors or bus bars.
• Before each use, inspect the probe; look for cracks in housing or output cable insulation.
• Do not use clamp in wet environment or in locations that hazardous gases exist.
• Do not use the probe anywhere beyond the tactile barrier.
INTERNATIONAL ELECTRICAL SYMBOLS
This symbol signifies that the current probe is protected by double or reinforced insulation. Use only factory specified replacement parts when servicing the instrument.
This symbol signifies CAUTION! and requests that the user refer to the user manual before using the instrument.
This is a type A current sensor. This symbol signifies that application around and removal from HAZARDOUS LIVE conductors is permitted.
DEFINITION OF MEASUREMENT CATEGORIES
CAT II: For measurements performed on circuits directly connected to the electrical distribution system.
Examples are measurements on household appliances or portable tools.
CAT III: For measurements performed in the building installation at the distribution level such as on
hardwired equipment in fixed installation and circuit breakers.
CAT IV: For measurements performed at the primary electrical supply (<1000V) such as on primary
overcurrent protection devices, ripple control units, or meters.
RECEIVING YOUR SHIPMENT
Upon receiving your shipment, make sure that the contents are consistent with the packing list. Notify your distributor of any missing items. If the equipment appears to be damaged, file a claim immediately with the carrier and notify your distributor at once, giving a detailed description of any damage.
4.37" (111mm)
3.98" (101mm)
2.25" (57mm)
1.22" (31mm)
2.13"
(54.1mm)
1.71"
(43.5 mm)
3.90" (99 mm)
SR751 SR752
ELECTRICAL SPECIFICATIONS
Current Range:
0.1 to 1000A AC, continuous cycle @ 1kHz
Output Signal:
1mV AC/A AC (1V AC at 1000A)
Accuracy and Phase Shift*:
Accuracy:
0.1 to 1A: 2% Reading ± 3µV 1 to 10A: 1% Reading 10 to 100A: 0.5% Reading 100 to 1200A: 0.3% Reading
8.50" (216mm)
Tactical
Barrier
(keep hands below
this barrier)
1.38"
(35 mm)
1.77"
(45 mm)
Phase Shift:
0.1 to 1A: Not Specified 1 to 10A: 2° 10 to 100A: 1° 100 to 1200A: 0.7°
*Reference conditions: 23°C±3°K, 20 to 75% RH, 48 to 65Hz, external magnetic field <40A/m, no DC component, no external current carrying conductor, test sample centered. Load impedance 1M.
Overload: 1200A for 40mn on, 20mn off Accuracy: Per IEC 185