General directions ............................................................................................................................................ 3
Introduction, Precautions and safety measures ............................................................................................................. 3
Special functions ........................................................................................................................................................... 4
Symbols on the instrument ............................................................................................................................................ 5
Warranty, Maintenance, metrological verification, Repair under warranty ..................................................................... 6
Replacing the fuse ......................................................................................................................................................... 7
Rechargeable and primary batteries ............................................................................................................................. 7
Communication interface ............................................................................................................................................... 7
Description of the instruments .........................................................................................................................
Display unit .................................................................................................................................................................. 10
Getting started ................................................................................................................................................ 18
Preparation for use ...................................................................................................................................................... 18
1. MAX MIN AVG mode .............................................................................................................................................. 19
4. "CLAMP" function .................................................................................................................................................... 25
Functions of the switch and keys ................................................................................................................................. 26
How are the various quantities measured? ................................................................................................. 28
1. Voltage measurement ............................................................................................................................................. 28
2. Current measurement ............................................................................................................................................. 29
3. Frequency measurement......................................................................................................................................... 31
6. Diode test ................................................................................................................................................................ 32
8. Temperature measurement ..................................................................................................................................... 34
9. Measurement on an MLI type speed variator .......................................................................................................... 35
10. Resistive power (MTX 3291) ................................................................................................................................. 37
11. dBm decibels in power (MTX 3291) ...................................................................................................................... 38
Technical characteristics of MTX 3290 ........................................................................................................ 40
Technical characteristics of MTX 3291 ........................................................................................................ 50
General characteristics ................................................................................................................................. 61
Power supply ............................................................................................................................................................... 61
Safety, CEM ................................................................................................................................................................ 61
Congratulations! You have just acquired a portable multimeter with a
display digital.
We thank you for this sign of confidence in the quality of our products.
The line of instruments to which it belongs comprises the following models:
Display digital, monochrome,
Power supply 4 R6 primary batteries (AA format) or 4 rechargeable batteries
Counts 6000 60000
Communication - IR/USB
It complies with safety standard NF EN 61010-1 + NF EN 61010-2-030
concerning electronic measuring instruments.
For best results, read this manual closely and observe the precautions of use.
Failure to observe these warnings and/or directions may damage the instrument
and/or its components and may endanger the user.
MTX 3290 MTX 3291
digital, monochrome,
(70x52)
backlit (70x52)
Precautions and
safety measures
before use
during use
•This instrument is been designed to be used as follows:
- indoors
- in an environment of pollution degree 2
- at an altitude of less than 2000m
- at a temperature between -10°C and 55°C
- at a relative humidity below 80% up to 31°C.
•The safety of any system incorporating the instrument is the responsibility
of the system integrator.
•Can be used for measurements on circuits:
- MTX 3290: 600V, CAT III and 300V, CAT IV.
- MTX 3291: 1000V CAT III and 600 CAT IV.
However, some accessories may lead to the use of this instrument on
circuits of a lower voltage and category.
• Comply with the environmental and storage conditions.
• Check the integrity of the guards and insulation of the accessories. Any item
of which the insulation is deteriorated (even partially) must be removed from
service and scrapped. A change of colour of the insulation is a sign of
deterioration.
• Read closely all notes preceded by the
• As a safety measure, use only the appropriate leads and accessories
supplied with the instrument or approved by the manufacturer.
• It is impossible to open the battery or fuse compartment without first
disconnecting the measurement leads.
• During a measurement exceeding 60V
the display unit
• Automatic detection of a connection to the "Ampere" terminal (for both
voltage and current measurements)
• When the maximum permanent voltage or current that can be measured is
exceeded, an intermittent audible signal warns of the risk of an electric
shock.
These multimeters have several features to protect them:
•varistor protection that clips transient voltage surges on the
measurement terminals.
•PTC (Positive Temperature Coefficient) protection against permanent
overvoltages less than or equal to 1000V during resistance,
capacitance, and diode test measurements.
This protection is reset automatically after the overload.
•a fuse that provides protection during current measurements.
- MTX 3290: 10A
- MTX 3291: 11A
DC or 25VAC thesymbol blinks on
Special functions
Automatic
detection
Automatic
switching off
Warning
signals
The number of input terminals is limited to 3: V, COM, A. Connecting the
lead to the "Ampere" terminal automatically selects the corresponding
function.
*When a change of function by the command keypad is
incompatible with the connection of the lead, it triggers an
audible or visual (LEADS) alert.
The current measurement is made with automatic peak range full-scale.
During a current measurement, an audible alert is triggered in the event of a
prolonged absence of current.
If the function is validated ( ), the device is automatically switched off after
30 mn of operation if there has been no action on the front panel during this
time.
*Automatic switching off is disabled:
- in the MAX, MIN, AVG, PEAK Surveillance mode
- in the Communication mode
- if there is a voltage >60V
multimeter.
An intermittent audible signal is emitted in all "Voltage" and "Current" settings if
the max. permanent value the device can measure is exceeded. It is
accompanied by display of the "O. L" acronym and of the
display unit.
This symbol is activated when the voltage on the "V" input exceeds 60V
AC in the "Voltage" setting or when the current injected between the A and
Warning! Using a measuring instrument, a lead, or an accessory belonging to a lower
Symbols on the
instrument
CAT II: Test and measurement circuits directly connected to the points of use
of the low-voltage network (power outlets and other similar points).
E.g.: Measurements on the network circuits of household appliances, portable
tools, and similar devices.
CAT III: Test and measurement circuits connected to parts of the low-voltage
network of the building.
E.g.: Measurements on distribution panels (including secondary meters), circuitbreakers, wiring including cables, bus bars, branch boxes, disconnecting
switches, power outlets in the fixed installation, and industrial appliances and
other equipment, such as motors permanently connected to the fixed
installation.
CAT IV: Test and measurement circuits connected to the source of the lowvoltage network of the building.
E.g.: Measurements on devices installed before the main fuse or the circuitbreaker of the building installation.
measurement or voltage category derates the resulting system (instrument
+ leads + accessories) to the lowest measurement category and/or service
voltage of any of the components.
Risk of electric shock: directions for connection and disconnection of the inputs.
Always connect the probes or adapters to the instrument before connecting
them to the measurement points. Always disconnect the probes or cords from
the measurement points before disconnecting them from the instrument. These
directions apply before the instrument is cleaned.
Warning: Hazard. The operator must refer to the manual each time this danger
symbol is encountered.
Device entirely protected by double insulation or reinforced insulation.
Earth
In the European Union, this product is subject to selective collection for the
recycling of electrical and electronic equipment waste in accordance with
Directive WEEE 2002/96/EC: this equipment must not be treated as ordinary
waste. The spent batteries must not be treated as ordinary waste. Take them in
to the appropriate collection point for recycling.
The CE marking indicates conformity with the European "Low Voltage", "EMC",
"WEEE" and "RoHS" directives.
This equipment is warranted for 3 years against any defect of materials or
workmanship, in accordance with the general terms of sale. During the
warranty period, the instrument may be repaired only by the manufacturer, who
reserves the right to repair the instrument or to replace it or part of it. If the
equipment is returned to the manufacturer, the cost of transport to the
manufacturer is borne by the customer.
The warranty does not apply following:
•improper use of the equipment or use in association with incompatible
equipment
•modification of the equipment without the explicit permission of the
manufacturer's technical staff
• maintenance done by a person not approved by the manufacturer
• adaptation to a particular application not anticipated in the definition of the
equipment or by the user manual
•a shock, a fall, or flooding.
Before opening the instrument, you must disconnect it from line power and
from the measurement circuits and make sure that you are not charged with
static electricity, which might destroy internal components. An adjustment,
maintenance, or repair of the live instrument must be undertaken only by
personnel who are qualified and have familiarized themselves with the
directions in this manual.
We recommend a verification of this instrument at least once a year. For
checking and calibration, contact one of our accredited metrology laboratories
(information and contact details available on request), at our Chauvin Arnoux
subsidiary or the branch in your country.
Unpacking,
repacking
Repair
under warranty
and post warranty
All of the equipment has undergone mechanical and electrical checks before
being dispatched. When you receive it, carry out a quick check to detect any
deterioration that may have occurred during transport. Should the need arise,
immediately contact our sales department and notify the carrier of the customary
reservations.
Use the original packaging to reship the equipment, if possible. Indicate as
clearly as possible, by a note attached to the equipment, the reasons for the
transfer.
For all repairs before or after expiry of warranty, please return the device to your
distributor.
• Disconnect everything connected to the instrument and press the ON/OFF
key (
• Use a soft cloth, moistened with soapy water.
• Rinse with a damp cloth and dry rapidly with a dry cloth or forced air.
• Make sure that no foreign objects interfere with the operation of the device
by which the leads are snapped into place.
).
y Before replacing the fuse (reached by
opening the bottom compartment), disconnect
the instrument from any source of current.
During the replacement, make sure that only a
fuse of the appropriate rating and specified
type is used. Using another type of fuse and
shorting the fuse holder are strictly forbidden.
y Checking the current fuse:
Fuse: SIBA/5019906
MTX 3291: 11A: 10x38 – 1,000V - F
breaking capacity: >20kA
MTX 3290: 10A: 6x32 - 600V - F
breaking capacity: >50kA
Rechargeable
and primary batteries
Active
communication
interface
(MTX 3291 only)
The multimeter is powered by primary or rechargeable batteries (see above).
To charge the rechargeable
batteries (set of 4 NI-MH LSD
batteries), use an external rapid
charger, available as an accessory
*After replacing the batteries,
wait 10s before switching the instrument back on.
The multimeter can communicate with a PC via the USB link.
The basic version includes a USB link using an isolated optical USB cord (type
HX0056Z) and SX-DMM software, plus Labview and Labwindows drivers to
program the devices.
* MTX 3291: They can also be programmed via the SCPI protocol:
- to program via Labview/LW
- to recover data or program the instrument using the software
Orange LEDs around the highly reliable virtual switch indicate the measurement
function chosen. The keys of the switch have priority over the action of the keys
of the keypad. The change from one function to another resets the configuration
of the measurement mode.
Short press Successive short presses
Current measurement
Temperature measurement
Capacitance measurement
Low-impedance AC voltage
measurement (VLowZ)
Selection of the type of probe:
Pt100, Pt1000
Current measurement using a
current clamp
Resistance measurement,
audible continuity
measurement, diode test
Frequency measurement
Voltage measurement
Selection of the transformation ratios
1, 10, 100, 1,000mV/A
The keypad has the following function keys:
The keys are taken into account and applied when pressed. If the key press is
validated, the instrument beeps.
Two types of action are possible:
• Short press Æ press lasting <2 seconds, validated by a beep as
soon as the key press is detected.
•Long press Æ press lasting >2 seconds, validated by a beep as soon as
the key press is detected.
MTX 3290
Successive short presses Long press
Activation/deactivation of storage of
the measurements and of the
quantities at a given time:
- Hold of the display without stopping
the acquisitions. The bargraph
continues to operate normally.
- Exit from the HOLD mode
In the MAX/MIN/AVG PEAK mode,
when the HOLD is active, the blinking
of the "MAX MIN AVG PEAK" symbol
indicates that acquisition continues as
a background task.
Choice of coupling AC, DC, AC+DC:
- Access to various parameters
ÆIn dBm: change of impedance 50,
75, 90, 600 (MTX 3291 only)
ÆIn temperature: the main display
unit indicates the temperature in °C,
the other in °F
ÆIn the REL mode, the key is used
to change from (present value –
reference value) to
Present value – reference value
Reference value
The value is displayed in %.
(MTX 3291 only)
Activation/deactivation of the low-pass
filter 300Hz:
The low-pass filter (4th order) makes
it possible to measure the RMS
voltage delivered by an MLI type
speed controller (for asynchronous
motor).
See curve
, p. 49 and 60.
x100
MTX 3291
- Hold of the display after
stabilization of the
measurement (Auto
HOLD)
When you use this multimeter, you must observe the usual safety rules,
which:
For your safety, use only the leads and accessories (clamp meter, etc.)
supplied with the instrument. Before each use, make sure that they are in
perfect condition.
The rechargeable batteries cannot be recharged in the multimeter.
Press ON/OFF
* Reminder: after replacing the batteries, wait 10s before powering
If the multimeter malfunctions, a long press (>2s) on this key can
In the Volt and Ampere functions, the instrument restarts in AC+DC.
• protect you from electrical hazards,
• protect the multimeter from operator errors.
• 4 1.5V alkaline batteries (LR6-AM3-AA) or
• 4 1.2V NI-MH rechargeable batteries of the same type.
to power up the device.
the device back up.
be applied to power down the instrument and then restore normal
operation.
In the BASIC mode, as default, the
device starts up in its elementary
configuration (default values) and in
Ac+DC function.
the V
MTX 3291 only: in the USER mode,
the device restarts in the
configuration and function selected
when it was powered down.
The connections illustrated in this chapter were made with an MTX 3290
multimeter (6,000 points). They would be the same with an MTX 3291 (60,000 pts).
1. Voltage
measurement
superposed on a DC voltage, or DC voltage measurement at high
impedance.
installations. The input impedance <1M serves to avoid
measuring "phantom" voltages due to couplings between the lines.
In all cases, “O.L” is displayed above 1050V (MTX 3291) or 620V (MTX 3290)
and a beep sounds when the measurement exceeds 1000V (MTX 3291) or
600V (MTX 3290).
AC voltage measurement, or measurement of an AC voltage
:
This position is provided to allow measurements in electrical
:
1. Press:
2. Select AC+DC, AC or DC coupling of the signal by pressing
(the default coupling is AC+DC).
Depending on what you select, the screen displays DC, AC or AC+DC.
3. Connect the black lead to the “COM” terminal and the red lead to “V”.
or .
* If the connection is not correct, an audible beep and a visible signal
(LEADS) are activated.
4. Place the test probes on the terminals of the circuit to be measured:
5. Read the measurement value indicated on the display unit.
6. As default, the 2
nd
display unit indicates the frequency, except in DC.
*It is possible to activate the filter in VLowZ, VAC+DC, VAC. The cutoff
frequency of the filter is ≤300Hz.
When a voltage having a frequency above 150Hz is measured, it is
heavily attenuated, and so a large error may be observed. It is
necessary in this case to deactivate the filter to have the full passband.
How are the various quantities measured? (continued)
2. Current
measurement
as an ammeter
1. Press:
2. Select the type of signal, AC+DC, AC, or DC, by pressing
on what you select, the screen displays AC, DC, or AC+DC.
3. Connect the black lead to the “COM” terminal and the red lead to “A”.
.
. Depending
*If the connection is not correct, an audible beep and a visible signal
(LEADS) are activated.
4. Place the test probes in series between the source and the load:
5. Read the measurement value indicated on the display unit.
“O.L” is displayed, if I >20A.
nd
6. As default, the 2
* It is possible to activate the filter in AAC+DC, AAC. The cutoff
frequency of the filter is ≤300Hz.
When a voltage having a frequency above 150Hz is measured, it is
heavily attenuated, and so a large error may be observed. It is
necessary in this case to deactivate the filter to have the full passband.
display unit indicates the frequency, except in DC.
How are the various quantities measured? (continued)
with a current
clamp
1. Press: .
2. Select the type of signal, AC+DC, AC, or DC, by pressing
Depending on what you select, the screen displays AC, DC, or AC+DC.
3. Connect the black lead of the clamp to the "COM" terminal and the red lead of
the clamp to "V".
4. Select the transformation ratio (the same as that of the clamp) - 1mV/A,
10mV/A, 100mV/A, or 1000mV/A - by pressing on "clamp" (
direct reading of the current.
5. Place the clamp around the conductor:
Clamp, V
.
) to have a
7. Read the measurement value indicated on the display unit. The
measurement accuracy is indicated in "Technical characteristics",
§. "Clamp" p. 46.
nd
8. As default, the 2
display unit indicates the transformation ratio in mV/A.
* It is possible to activate the filter in AAC+DC, AAC. The cutoff
frequency of the filter is ≤300Hz.
When a voltage having a frequency above 150Hz is measured, it is
heavily attenuated, and so a large error may be observed. It is
necessary in this case to deactivate the filter to have the full passband.
How are the various quantities measured? (continued)
9. Measurement on an
MLI type speed
variator
Voltage
measurement
1. Press:
2. Select the type of signal, AC+DC, AC, or DC, by pressing
Depending on what you select, the screen displays AC, DC, or AC+DC.
3. Select the filter by pressing
4. Connect the black lead to the “COM” terminal and the red lead to “V”.
5. Place the test probes between two phases of the circuit to be measured:
.
.
.
6. Read the measurement values indicated on the display unit (voltage and
frequency):
In all cases, “O.L” is displayed above 1050V (MTX 3291) or 620V (MTX 3290)
and a beep sounds when the measurement exceeds 1000V (MTX 3291) or
600V (MTX 3290).
The presence of the symbol indicates that the 300Hz filter is active.
*It is very important to leave the filter activated to measure the voltage
and frequency of the signal without being perturbed by the MLI.
These multimeters can be interfaced directly with a PC or other computer using
"SX-DMM" acquisition software:
The transmission rate is 9600 Bauds.
The transmission parameters are fixed (8 data bits, 1 stop bit, no parity).
1. Connect the isolated optical lead to the isolated optical input of the
multimeter (on the side of the multimeter). Mechanical polarization prevents
connection in reverse.
Connect the USB lead to one of the USB ports of the PC.
Install the USB driver on your PC (see the data sheet on the CD provided).
2.
USB
Isolated optical connector
Interface cable
Installing
the "SX-DMM"
software
1. Install the "SX-DMM" software on the PC using the CD.
2. Start the software for data acquisition and study the various display
possibilities (curves, tables, etc.).
*The symbol appears on the display unit when the instrument is
controlled from the PC (REMOTE mode).
For more information, refer to the "Help" menu of the software.
Only values with tolerances or limits are guaranteed values.
Values without tolerances are given for guidance (standard NFC42670).
The technical specifications are guaranteed only after 30 minutes of warming up.
Except as otherwise indicated, they are valid from 10% to 100% of the measurement
range.
DC voltage In the "DC" mode, you measure a direct voltage or the DC component of an AC
voltage (filter activated).
Range
600mV 0 to 600.0mV 0.1mV 0.6% L+2D 10.9M
6V 0 to 6.000V 0.001V
60V 0 to 60.00V 0.01V 10.082M
600V (∗)
(∗) The display indicates "+OL" above +620V and "-OL" above - 620V. Protection:
850Vpk
Specified
measurement range
0 to 600.0V 0.1V 10.008M
Resolution
Intrinsic error Input impedance
0.3% L+2D
10.9M
Secondary measurements and displays: MAX, MIN, AVG
AC and AC+DC
voltages
VACRMSProtection: 850Vpk
With this function, the user can measure the true RMS (TRMS) value of an AC
voltage with its DC component (no capacitive coupling) or without its DC component.
Range
600mV
6V
60V
600V 2)
1) See the typical curve of the 300Hz filter.
2) The LCD indicates "+OL" above +620V, "-OL" below -620V or above 620VRMS.
Operating
range
0 to
600.0mV
0 to
6.000V
0 to
60.00V
0 to
600.0V
Specified
easurement
range 3)
60.0 to
600.0mV
0.600 to
6.000V
6.00 to
60.00V
60.0 to
600.0V
Resolution
0.1mV
0.001V
0.01V 10.082 M
0.1V 10.008 M
Uncertainty
(±)
2% L+
0.25%x
[F(kHz)-1]L
±5D
2% L+
0.18%x
[F(kHz)-1]L
±3D
Additional
uncertainty F
(Hz) 1)
45<F<65Hz
0.3% L
typ.
at 100Hz
0 .7% L
typ.
at 150Hz
1.8% L
typ.
at 300Hz
30% L
typ.
Pass
band
10Hz to
20kHz
10Hz to
20kHz
@ 1kHz
Input
impedance
// <50 pF
10.9 M
10.9 M
Peak
factor
3 to
500mV
3 to 5V
3 to 50V
3 to
500V
3) From 1kHz, the measurement must exceed 15% of the range.
Secondary measurements and displays: FREQ (AC coupling), MAX, MIN, AVG,
PEAK
Technical characteristics of the MTX 3290 (continued)
Currents
DC current
Three possible modes: DC, AC, AC+DC
In DC mode, you can measure a direct current or the DC component of an
alternating current.
In the AC and AC+DC modes, you can measure the true RMS (TRMS) value of
an alternating current with/without its direct component (no capacitive coupling in
"DC" mode).
Particular reference conditions
6mA range
: Measuring a strong current for a long time can cause a temperature
:
rise in some components. In this case, it is necessary to wait some time for the
metrological characteristics specified in 6mA to be restored.
Range
6mA 0 to 6.000mA 0.002 to
60mA 0 to 60.00mA
600mA 0 to 600.0mA 0.2 to 600.0mA0.1mA 1.2% L ± 2D 0.58mV/mA
6A 0 to 6.000A 0.200 to 6.000A0.001A 1.2% L ± 3D 0.05V/A
10 A/20A
(∗)
Operating
range
0 to 20.00A 0.20 to 20.00A0.01A 1.2% L ± 2D
Specified
measurement
range
6.000mA
.02 to 60.00mA0.01mA 1.2% L ± 2D 3mV/mA
Resolution
1µA 1.2% L ± 5D
Uncertainty
(±)
Voltage
drop
25mV/mA
0.05V/A
Protection
Fuse
10A/600V
>50kA
AAC RMS current
The display indicates "OL" above 19.99A. The symbol blinks and a beep
sounds above 10A.
(∗) Acceptable overload: 10A to 15A for 30s max. with a pause of 5min between
2 measurements. Ambient temp. 35°C max.
Secondary measurements and displays: MAX, MIN, AVG
Range
6mA
60mA
600mA
6A
10A/20A
(∗)
The display indicates "OL" above 19.99A. The symbol blinks and a beep
Operating
range
0 to
6.000mA
0 to
60.00mA
0 to
600.0mA
0 to
6.000A
0 to
20.00A
Specified
easuremen
range
0.600 to
6.000mA
6.00 to
60.00mA
60.0 to
600.0mA
0.600 to
6.000A
1.00 to
10.00A
Resolution
1µA 1.7% L ± 5D
0.01mA
0.1mA
0.001A 1.7% L ± 5D 2.8 to 5A 0.05V/mA
0.01A 1.5% L ± 3D 3.7 to 8A 0.05V/mA
Uncertainty (±)
40Hz to 20kHz
(**)
1.5% L ± 3D
Peak factor
2.6 to
5mA
2.6 to
50mA
2.6 to
500mA
Voltage
drop
25mV/mA
3mV/mA
0.58mV/mA
Protection
Fuse
10A/600V
>50kA
sounds above 10A.
Secondary measurements and displays: FREQ (AC coupling) MAX, MIN, AVG,
PEAK
(∗) Acceptable overload: 10A to 15A for 30s max. with a pause of 5min between 2
measurements. Ambient temp. 35°C max.
(∗∗) Additional uncertainty with the 300Hz filter.
Technical characteristics of the MTX 3290 (continued)
AAC+DC TRMS current Warning: The sum AC+DC must never exceed the range, 600mA, or
60mA, or 6mA, or 6A, or 10A, as the case may be.
Range
Operating
range
Specified
measure-
ment range
esolution
Uncertainty AC
40Hz at 20kHz
(±) (∗∗)
Additional
uncertainty
DC (±)
Peak
factor
Voltage
drop
Protection
6mA
60mA
600mA
6A
10A
/20A*
The display indicates “OL” above 19.99A. The symbol blinks and a beep
0 to
6.000mA
0 to
60.00mA
0 to
600.0mA
0 to
6.000A
0 to
20.00A
0.060 to
6.000mA
6.00 to
60.00mA
60.0 to
600.0mA
0.600A to
6.000A
0.60A to
20.00A
0.01mA
0.1mA
0.001A
0.01A
1µA
1.7% L +[0.08%x
(FkHz-1)] L
±5D
1.5% L +[0.08%x
(FkHz-1)] L
±3D
1.7% L+[0.08%x
(FkHz-1)] L
±5D
1.5% L+ [0.08%x
(FkHz-1)] L
±3D
±15D
±13D
±10D 2.8 to 5A 0.05V/mA
±10D 3.7 to 8A 0.05V/mA
2.6 to
5mA
2.6 to
50mA
2.6 to
500mA
25mV/mA
3mV/mA
0.58mV/mA
Fuse
10A/600V
>50kA
sounds above 10A.
(*) Acceptable overload: 10A to 15A for 30s max. with a pause of 5min between 2
measurements. Ambient temp. 35°C max.
Secondary measurements and displays: F (AC coupling), MAX, MIN, AVG, PEAK
(**) Additional uncertainty with the 300Hz filter.
Technical characteristics of the MTX 3290 (continued)
Frequency
Main frequency
measurement
In this setting, you can measure the frequency of a voltage.
Particular reference conditions
: 150mV <U <600V
When the switch is set to Hz, the 300Hz filter is not in service.
Protection: 850Vpk
Range Operating range
60Hz 10.00 to 60.00Hz 10.00 to 60.00Hz 0.01Hz
600Hz 10.0 to 600.0Hz 10.00 to 600.0Hz 0.1Hz
6kHz 0 to 6.000kHz 0.010 to 6.000kHz 0.001kHz
60kHz 0 to 60.00kHz 0.01 to 60.00kHz 0.01kHz
600kHz 0 to 200.0kHz 0.1 to 200.00kHz 0.1kHz
Specified
measurement
range
Resolution Intrinsic error
0.1% L ±1D
Below 10Hz, or if the signal detection level is inadequate, the reading is forced to
zero.
* The measured period in ms is available on the second display unit.
Secondary
frequency
measurement
You can measure the frequency and magnitude of a voltage or of a current
simultaneously.
Same accuracy as in the "Hz" setting
Particular reference conditions
: 150mV <U <600V
0.15A <I <10A
Max. frequency measurable in volts: 20kHz
Max. frequency measurable in amperes: 20kHz
When the switch is set to VLowZ, Volts or Ampere, if the 300Hz filter is activated,
the measurable frequency remains within the limits of the PB of the filter.
Below 10Hz, or if the signal detection level is inadequate, the reading is forced to
“-----“.
Technical characteristics of the MTX 3290 (continued)
Resistance
Ohmmeter In this setting, the user can measure a resistance.
Capacitance
Particular reference conditions
:
The (+COM) input must not have been overloaded following the accidental
application of a voltage to the input terminals with the switch set to or T°.
If this happens, the return to normal may take about ten minutes.
Protection: 850Vpk
Range
600
6k0 to 6.000k 0.001k
60k0 to 60.00k 0.01k 12.6A
600k0 to 600.0k 0.1k 1.26A
6M0 to 6.000M 0.001M 1.5% L ±3D 240nA
60MΩ
Specified
measurement
range
0 to 600.0 ∗
0 to 60.00M
Resolution Uncertainty
0.1 0.5% L ±2D 850µA
0.5% L ±2D
0.01M3% L ±3D 29nA
Measurement
current
126.6A
(∗) REL measurements
Open-circuit
voltage
<5V
Capacitance
meter
In this setting, the user can measure the capacitance of a capacitor.
Range
6nF
60nF 0 to 60.00nF 0 to 60.00nF0.01nF 1.5% L ±8D 1.26µA 400ms
600nF 0 to 600.0nF 0 to 600.0nF0.1nF 1.5% L ±5D 1.26µA 400ms
6µF 0 to 6.000µF 0 to 6.000µF0.001µF 1.5% L ±5D 12.6µA 0.125 s/µF
60µF 0 to 60.00µF 0 to 60.00µF0.01µF 1.5% L ±5D 126.6µA 0.125 s/µF
600µF 0 to 600.0µF 0 to 600.0µF0.1µF 3.5% L ±5D 850µa 0.125 s/µF
6mF 0 to 6.000mF 0 to 6.000mF1µF 4.5% L ±5D 850µa 17 s/mF
60mF 0 to 60.00mF 0 to 60.00mF10µF 6.5% L ±5D 850µa 17 s/mF
Operating
range
0.100 to
6.000nF
Specified
measurement
range
0.100 to
6.000nF
Resolution
0.001nF 2.5% L ±30D 1.26µA 400ms
Intrinsic
error
Measurement
current
Measurement
The use of wires that are very short and shielded is strongly recommended.
Technical characteristics of the MTX 3290 (continued)
Diode Test
Audible continuity
Clamp
DC current
Range Resolution Accuracy Open-circuit voltage
3V 1mV 2% L ±3D <5V <1.1mA
Measurement
current
Audible signal triggered if <40mV ±10mV
Protection: 850Vpk
Range Resolution Accuracy
600 0.10.5% L ±3D <5V <1.1mA 850Vpk
Open-circuit
voltage
Measurement
current
Protection
Response time <100ms
Triggering threshold: <30 ±5
Protection: 850Vpk
You can measure a current using various current clamps and obtain a direct
reading of the current by selecting the correct transformation ratio, which must be
the same as that of the clamp.
If the signal detection level is insufficient, the value is forced to "-----"
The input impedance is approximately 10M.
*Add the error of the clamp to the intrinsic error of the multimeter, specified
in the tables below.
Range
Ratio
1mV/A
10mV/A
100mV/A
1000mV/A
600mA 6A 60A 600A 6000A
Resolution
Accuracy 0.6%L ±2D 0.6%L ±2D 0.3%L ±2D
Resolution
Accuracy 0.6%L ±2D 0.6%L ±2D 0.3%L ±2D
Resolution 0.1mA 0.001A 0.01A
Accuracy 0.6%L ±2D 0.6%L ±2D 0.3%L ±2D
Resolution 0.1mA 0.001A
Accuracy 0.6%L ±2D 0.3%L ±2D
0.001A 0.01A 0.1A
0.01A 0.1A 1A
Secondary measurements and displays: MAX, MIN, AVG and transformation ratio of
the sensor
AAC RMS current
Range
Ratio
1mV/A
10mV/A
100mV/A
1000mV/A
600mA 6A 60A 600A 6000A
Resolution
Accuracy
Resolution
Accuracy
Resolution 0.1mA 0.001A 0.01A
Accuracy
Resolution 0.1mA 0.001A
Accuracy 2% L ±5D 2% L ±3D
Peak factor 3 @ 500mA @ 5A @ 50A @ 500A @ 5000A
2% L ±5D (∗)
0.001A 0.01A 0.1A
2% L ±5D (∗)
2% L ±5D 2% L ±3D
0.01A 0.1A 1A
2% L ±5D (∗)
2% L ±5D 2% L ±3D
2% L ±5D 2% L ±3D
Secondary measurements and displays: MAX, MIN, AVG and transformation ratio
of the sensor
300Hz filter: if the filter is active, see "300Hz filter" curve for the additional
uncertainty. (∗) : see "Frequency response" curve, p. 47.
Only values with tolerances or limits are guaranteed values.
Values without tolerances are given for guidance (standard NFC42670).
The technical specifications are guaranteed only after 30 minutes of warming up. Except
as otherwise indicated, they are valid from 10% to 100% of the measurement range.
In the “DC” mode, you measure a direct voltage or the DC component of an AC voltage
(filter activated).
60mV range: Measuring a strong current or measuring a current for a long time may
cause a temperature rise of some components.
Protection: 1414 Vpk
Range
60mV 1) 0 to 60.000mV 0.001mV 0.5% L+35D
600mV 0 to 600.00mV 0.01mV 0.5% L+25D
6V 0 to 6.0000V 0.0001V
60V 0 to 60.000V 0.001V
600V 0 to 600.00V 0.01V
Specified
measurement range
Resolution Intrinsic error
0.05% L+25D
Input impedance
10.612M
10.9M
10.9M
10.082M
10.008M
AC and AC+DC
voltages
VACRMS
1000V 2) 0 to 1000.0V 0.1V 0.07% L+25D
10.008M
1) This range is accessible only with the Range key.
Input impedance: approx. 10.6M // 50 pF
2) The display indicates "+OL" above +1050V and "-OL" below -1050V.
Secondary measurements and displays: MAX, MIN, AVG
With this function, the user can measure the true RMS (TRMS) value of an AC voltage
with its DC component (no capacitive coupling) or without its DC component.
60mV range: Measuring a strong current or measuring a current for a long time may
cause a temperature rise of some components.
Technical characteristics of the MTX 3291 (continued)
AC RMS
V
(continued)
VLowZ AC RMS
Protection: 1414 Vpk
1) See the typical curve of the 300Hz.
2) This range is accessible only with the Range key
Input impedance: approx. 10.6M // 50 pF
3) The LCD indicates “+OL” above +1050V, “-OL” below -1050V or above 1050VRMS.
4) From 1kHz, the measurement must exceed 15% of the range.
Secondary measurements and displays: FREQ (AC coupling), MAX, MIN, AVG, PEAK
The pass band (between 3dB down points) is reduced to 300Hz, if the filter is activated. In
VLowZ, there is no 60mV range.
The frequency measurement is made like the measurement in a 300Hz pass band.
Range
600mV
6V
60V
600V 0 to 600.00V
Operating
range
0 to
600.00mV
0 to
6.0000V
0 to
60.000V
Specified
measurement
range 3)
60.00 to
600.00mV
0.6 to
6.0000V
6.000 to
60.000V
60.00 to
600.00V
Resolution
0.01mV
0.0001V
0.001V 3 to 50.0V
0.01V 3 to 500.0V
Uncertainty (±)
1% L+ 0.25%x
[F(kHz)-1] L
± 30D
0.5% L+
0.18%x
[F(kHz) -1] L
± 25D
Additional
uncertainty
F(Hz) 1)
45<F<65Hz
0.3% L typ.
at 100Hz
0.7% L typ.
at 150Hz
1.8% L typ.
at 300Hz
30% L typ.
Pass band
≅ 300 k
Input
impedance
//<50 pF
3 to
500.0mV
3 to 5.0V
VAC+DC TRMS
Protection: 1414 Vpk
2)
1000V
1) See the typical curve of the 300Hz.
2) The LCD indicates “+OL” above +1050V, “-OL” below -1050V or above 1050VRMS.
0 to
1000.0V
60 to
1000.0V
0.1V
1.42 to
1000.0V
3) From 1kHz, the measurement must exceed 15% of the range.
Secondary measurements and displays: FREQ (AC coupling), MAX, MIN, AVG, PEAK
60mV range: Measuring a strong current or measuring a current for a long time may cause a
temperature rise of some components.
Range
60mV 2)
600mV
6V
60V
600V
Operating
range
0 to
60.000mV
0 to
00.00mV
0 to
6.0000V
0 to
60.000V
0 to
600.00V
Specified
measurem.
range 4)
6.000 to
60.000mV
60.00 to
600.00mV
0.6 to
6.0000V
6.000 to
60.000V
60.00 to
600.00V
Resolution
0.001mV
0.01mV
0.0001V
0.001V
0.01V
Additional
uncertainty
DC (±)
± 15D
Uncertainty
AC (±)
1.5% L
±35D
0.8% L
+ 0.18%x
[F(kHz)-1] L
±30D
0.5% L
+ 0.18%x
[F(kHz)-1] L
±25D
Additional
uncertainty
F(Hz) 1)
45<F<65Hz
0.3% L typ.
at 100Hz
0.7% L typ.
at 150Hz
1.8% L typ.
at 300Hz
30% L typ.
Pass band
≈ 400Hz
10Hz to
50kHz
10Hz to
100kHz
Input
impedance
//<50 pF
10.612 M
10.9 M
10.9 M
10.082 M
10.008M
Peak factor
500mV
3 to 5V
3 to
50mV
3 to
3 to
50V
3 to
500V
0 to
3)
1000V
1) See the typical curve of the 300Hz.
2) This range is accessible only with the Range key - Input impedance: approx. 10.6M//50 pF
1000.0V
60 to
1000.0V
0.1V
10.008 M
1.42 to
1000V
3) The LCD indicates “+OL” above +1050V, “-OL” below -1050V or above 1050VRMS.
4) From 1kHz, the measurement must exceed 15% of the range.
Secondary measurements and displays: FREQ (AC coupling), MAX, MIN, AVG, PEAK
Technical characteristics of the MTX 3291 (continued)
Currents
DC current
Three possible modes: DC, AC, AC+DC
In DC mode, you can measure a direct current or the DC component of an
alternating current.
In the AC and AC+DC modes, you can measure the true RMS (TRMS) value of
an alternating current with/without its direct component (no capacitive coupling in
"DC" mode).
Particular reference conditions:
600µA and 6mA ranges
: Measuring a strong current for a long time may cause a
temperature rise of some components. In this case, it is necessary to wait some
time for the metrological characteristics specified in these ranges.
Range
600µA
6mA
60mA
600mA
Operating
range
0 to
600.00µA
0 to
6000.0mA
0 to
60.000mA
0 to
600.00mA
Specified
measurement
range
0.02 to
600.00µA
0.002 to
6.0000mA
0.020 to
60.000mA
0.20 to
600.00mA
Resolution
0.01µA 1% L ±25D 10mV/mA
0.1µA 0.8% L ±25D 25mV/mA
0.001mA 0.8% L ±20D 3mV/mA
0.01mA 0.8% L ±20D 0.58mV/mA
Uncertainty
(±)
Voltage drop Protection
Fuse
11A/1000V
>20kA
6A
10A/20A
(∗)
The display indicates “OL” above 19.99A. The
0 to
6.0000A
0 to
20.000A
0.2000 to
6.0000A
0.200 to
20.000A
0.0001A 0.8% L ±20D 0.05V/A
0.001A 0.8% L ±20D 0.05V/A
symbol blinks and a beep
sounds above 10A.
(∗) Acceptable overload: 10A to 20A for 30s max. with a pause of 5min between 2
measurements. Ambient temp. 35°C max.
Secondary measurements and displays: MAX, MIN, AVG
Technical characteristics of the MTX 3291 (continued)
AAC RMS current
Range Operating range
600µA
6.000mA
60mA
600mA
6A
10A/20A
(∗)
The display indicates “OL” above 19.99A. The
0 to
600.00µA
0 to
6.0000mA
0 to
60.000mA
0 to
600.00mA
0 to
6.0000A
0 to
20.000A
Specified
measurement
range
60 to
600.00µA
0.6000 to
6.0000mA
6.000 to
60.000mA
60.00 to
600.00mA
0.6000 to
6.000A
1.000 to
20.000A
Resolution
0.01µA
0.1µA
0.001mA
0.01mA
0.0001A
0.001A
Uncertainty (±)
40Hz to 20kHz
(**)
1.5% L
±30D
1.2% L+[0.08%
x (FkHz-1)] L
±25D
1% L+[0.08%
x (FkHz-1)] L
±25D
1% L +[0.1%
x (FkHz-1)] L
±25D
1.2% L+ [0.1%
x (FkHz-1)] L
±25D
symbol blinks and a beep
Peak factor
2.6 to
500µA
2.6 to
5mA
2.6 to
50mA
2.6 to
500mA
Voltage
drop
10mV/µA
25mV/mA
3mV/mA
0.58mV/mA
2.8 to 5A 0.05V/mA
3.7 to 8A 0.05V/mA
Protection
Fuse
/1000V
>20kA
sounds above 10A.
Secondary measurements and displays: FREQ (AC coupling), MAX, MIN, AVG,
PEAK
(∗) Acceptable overload: 10A to 20A for 30s max. with a pause of 5min between 2
measurements. Ambient temp. 35°C max.
(∗∗) Additional uncertainty with the 300Hz filter.
11A
AAC+DC TRMS currentWarning: the sum AC+DC must never exceed the range, 600mA, or
60mA, or 6mA, or 600µA or 6A, or 10A, as the case may be.
The AC component must represent at least 5% of the amplitude of the AC+DC total
for it to be possible to measure it.
Range
600µA
6mA
60mA
600mA
6A
10A/20A
The display indicates “OL” above 19.99A. The
(∗)
Operating
range
0 to
600.00µA
0 to
6.0000µA
0 to
60.00mA
0 to
600.00mA
0 to
6.0000A
0 to
20.00A
Specified
measurement
range
60 to
600.00µA
0.6000 to
6.0000mA
6.000 to
60.000mA
60.00 to
600.00mA
0.6000 to
6.000A
0.600 to
20.000A
Resolution
0.01µA
0.1µA
0.001mA
0.01mA
0.0001A
0.001A
Uncertainty (±)
40Hz to 20kHz
(**)
1.5% L
±20D
1% L +[0.08% x
(FkHz - 1)]L
±25D
1% L +[0.08%x
(FkHz - 1)]L
±25D
1% L+[0.1%x
(FkHz-1)]L
±25D
1.2% L+ [0.1%x
(FkHz-1)]L
±25D
Peak
factor
Voltage
drop
±20D
±15D
2.6 to
500µA
2.6 to 5mA 25mV /mA
2.6 to
50mA
2.6 to
500mA
2.8 to 5A 0.05V /mA
3.7 to 8A 0.05V /mA
symbol blinks and a beep sounds
above 10A.
(∗) Acceptable overload: 10A to 20A for 30s max. with a pause of 5min between 2
measurements. Ambient temp. 35°C max.
Secondary measurements and displays: FREQ (AC coupling), MAX, MIN, AVG, PEA
(∗∗) Additional uncertainty with the 300Hz filter.
Technical characteristics of the MTX 3291 (continued)
Frequency
Main frequency
measurement
Secondary
frequency
measurement
In this setting, you can measure the frequency of a voltage.
Particular reference conditions
: 150mV <U <600V
When the switch is set to Hz, the 300Hz filter is not in service.
Protection: 1414 Vpk
Range Operating range
60Hz 10.00 to 60.00Hz 10.00 to 60.00Hz 0.01Hz
600Hz 10.0 to 600.0Hz 10.0 to 600.0Hz 0.1Hz
6kHz 0 to 6.000kHz 0.010 to 6.000kHz 0.001kHz
60kHz 0 to 60.00kHz 0.01 to 60.00kHz 0.01kHz
600kHz 0 to 200.0kHz 0.1 to 200.0kHz 0.1kHz
Specified
measurement
range
Resolution Intrinsic error
0.1% L ±1D
Below 10Hz, or if the signal detection level is inadequate, the reading is forced to
zero.
* The measured period in ms is available on the second display unit.
You can measure the frequency and magnitude of a voltage or of a current
simultaneously.
Same accuracy as in the "Hz" setting
Particular reference conditions
0.15A <I <10A
Max. frequency measurable in volts: 100kHz
(except 60mV range Æ 400Hz and
600mV range Æ 50kHz)
Max. frequency measurable in amperes: 20kHz
When the switch is set to VLowZ, Volts or Ampere, if the 300Hz filter is activated,
the measurable frequency remains within the limits of the PB of the filter.
Below 10Hz, or if the signal detection level is inadequate, the reading is forced to
“-----“.
Technical characteristics of the MTX 3291 (continued)
Resistance
Ohmmeter In this setting, the user can measure a resistance.
Capacity
Particular reference conditions
:
The (+COM) input must not have been overloaded following the accidental
application of a voltage to the input terminals with the switch set to or T°.
If this happens, the return to normal may take about ten minutes.
Protection: 1414 Vpk
Range
600
6k0 to 6.0000k 0.0001k
60k0 to 60.000k 0.001k 12.6A
600k0 to 600.00k 0.01k 1.26A
6M0 to 6.0000M 0.0001M 1.5% L ±30D 240nA
60MΩ0 to 60.000MΩ
Specified
measurement
range
0 to 600.00 (∗)
Resolution Uncertainty
0.01 0.2% L ±20D 1mA
0.2% L ±20D
0.001M3% L ±30D 29nA
Measurement
current
126.6A
REL measurements
Open-circuit
voltage
<5V
Capacitance meter In this setting, the user can measure the capacitance of a capacitor.
Range
6nF 0.100 to 6.000nF 0.100 to 6.000nF0.001nF 2% L ±30D
60nF 0 to 60.00nF 0 to 60.00nF 0.01nF 1% L ±8D
600nF 0 to 600.0nF 0 to 600.0nF 0.1nF 1% L ±5D 1.26µA 400ms
6µF 0 to 6.000µF 0 to 6.000µF 0.001µF 1% L ±5D 12.6µA 0.125 s/µF
60µF 0 to 60.00µF 0 to 60.00µF 0.01µF 1% L ±5D 126.6µA 0.125 s/µF
600µF 0 to 600.0µF 0 to 600.0µF 0.1µF 3% L ±5D 1mA 0.125 s/µF
6mF 0 to 6.000mF 0 to 6.000mF1µF 4% L ±5D 1mA 17 s/mF
60mF 0 to 60.00mF 0 to 60.00mF10µF 6% L ±5D 1mA 17 s/mF
The use of wires that are very short and shielded is strongly recommended.
You can measure a current using various current clamps and obtain a direct
reading of the current by selecting the correct transformation ratio, which must be
the same as that of the clamp.
If the signal detection level is insufficient, the value is forced to "-----"
The input impedance is approximately 10M.
Open-circuit
voltage
Measurement
current
Measurement
current
Protection
* Add the error of the clamp to the intrinsic error of the multimeter, specified
in the tables below.
DC current
Range
Ratio
1mV/A
10mV/A
100mV/A
1000mV/A
600mA 6A 60A 600A 6000A
Resolution
Accuracy 0.5% L ±2D 0.5% L ±2D 0.05% L ±2D
Resolution 0.001A 0.01A 0.1A
Accuracy
Resolution 0.1mA 0.001A 0.01A
Accuracy 0.5% L ±2D 0.5% L ±2D 0.05% L ±2D
Resolution
Accuracy 0.5% L ±2D0.05% L ±2D
Secondary measurements and displays: MAX, MIN, AVG and transformation ratio
of the sensor
Technical characteristics of the MTX 3291 (continued)
Temperature
Pt100/Pt1000
The user can measure the temperature by means of a Pt100/Pt1000 sensor.
Range Measurement current Resolution Accuracy Protection
"Active" protection by PTC thermistor
Display in °C/°F possible
Peak
SURV
MIN, MAX, AVG
Resistive power
Duty cycle
-200°C to
+800°C
<1mA (Pt100)
<0.1mA (Pt1000)
0.1°C
0.1% L ±1°C
1414 Vpk
Add 1% L + 30 D to obtain the accuracy corresponding to the function and the
range.
Fmax 1kHz (1ms)
Protection 1414 Vpk
Add 0.2% L+2D to obtain the accuracy corresponding to the function and the
range.
Acquisition time of the extrema approximately 100ms
Protection 1414 Vpk
Display of the resistive power with respect to a reference resistance measured on
the installation and saved in memory using the HOLD key (600 is the default)
The function determined is: (measured AC+DC voltage)
2
/ VRref
Range DC, AC and AC+DC
Resolution 1mW
Accuracy: 2 x accuracy V
Max. measurement voltage 1000V
AC+DC
AC (in %)
Protection 1414 Vpk
Unit of display W
Display of the measurement in % of a logical signal (TTL, CMOS, etc.) in
"AC+DC" mode
DC+ duty cycle = θ
DC- Duty cycle = T - θ
Resolution 0 .01%
θ
T
Minimum duration for θ10 μs
Maximum duration for T 0 .8 s
Minimum duration for T 200 μs [5kHz]
Nominal range 5 to 95% typical
Sensitivity (10V range) >10% of the range, Freq <1kHz
>20% of the range, Freq >1kHz
Absolute error on the duty
cycle, expressed in % absolute ± [0.1%+0.045%*(RC-50)], Freq <1kHz
± [0.5%+0 .06% *(RC-50)], Freq >1kHz
Protection 1414 Vpk
Technical characteristics of the MTX 3291 (continued)
Pulse width
(
) ()
Depending on frequency counter triggering conditions.
Resolution 10µs
Minimum pulse width 100µs
Accuracy 0.1% ±10µs
Maximum duration of a period 1.25s (0.8Hz)
Triggering threshold 20% of the range except 1000VAC range
This threshold is: positive in
, negative in .
Additional error on the measurement due to the slope at the zero crossing:
See §. Measurement of duty cycle.
Protection 1414 Vpk
dBm Display of the measurement in dBm with respect to a resistance reference
chosen by the user from among 50Ω, 75Ω, 90Ω, and 600Ω, (default value 600Ω)
Resolution 0.1dBm
Absolute error in dBm 0.09 x relative error V
AC expressed in %
Additional calculation error 0.1dBm
Measurement range 10mV to 1000V
Protection 1414 Vpk
Operation
of the audible beep
Beep reporting a valid key
Beep reporting an invalid key
High-pitched
sound
Low-pitched
sound
Successive beeps reporting an overshoot of the danger
threshold (alarm)
Successive beeps reporting recording of the MAX, MIN, PEAK
Altitude <2000m
Reference range 23°C ±5°C
Specified range of use -10°C to 55°C
Influence of temperature see §. Influences
Relative humidity 0% to 80% from 0°C to 31°C
0% to 70% from 40°C to 55°C
limited to 70% for the 6 and 60 ranges
Dust- and water-tightness IP67 (in the event of immersion, under 1m of
water for 30 mn, it is necessary to let the
water flow off or to let the unit dry before
putting it back into service).
Storage range - 20°C to 70°C
The multimeter is powered by primary or rechargeable batteries:
Measurement category of the "measurements" MTX3290: 600V CAT III and 300V CAT IV with respect to earth
MTX3291: 1000V CAT III and 600V CAT IV with respect to earth
This instrument is designed in conformity with the EMC standards in force and
its compatibility has been tested in accordance with the following standards: