You have just acquired an MTX 1054 (MTX 1052) – congratulations on
your choice and thank you for your confidence in the quality of our
products. It consists of:
• 150 MHz, 4 channel (MTX 1054) or 2 channel (MTX 1052), oscilloscope without display device
• SCOPEin@BOXsoftware
This instrument comes with a data acquisition and pre-processing card and
its own mains supply. It is managed by resident, embedded flash software
that can be updated from a PC via the
SCOPEin@BOX software.
This software communicates with the "host PC" via an ETHERNET
interface.
This instrument has the following operating modes:
"Oscilloscope" Instrument
"Harmonics Analyser" Instrument
"Recorder" Instrument
"SPO" Analogue Persistence display
"FFT" representation
This instrument complies with safety standard EN 61010-1 (2004), single
insulation, relative to electronic measurement instruments and complies with
the EMC standards corresponding to residential and industrial environments.
For optimum service, read this manual carefully and comply with the
operating precautions.
Failure to comply with these warnings and/or user instructions is liable to
cause damage to the equipment. This could be dangerous to the user.
• It is designed for use:
- indoors
- in an environment with pollution level 2
- at an altitude of less than 2000m
- at a temperature between 0°C and 40°C
- with relative humidity of less than 80% up to 31°C
• It can be used for measurements on 300 V CAT II circuits in relation to
earth and can be powered by a 240 V, CAT II network
definition of
measurement
categories
CAT I: Measurement category I corresponds to measurements taken on circuits
not directly connected to the network.
Example: protected electronic circuits
CAT II: Measurement category II corresponds to measurements taken on circuits
Example: power supply for domestic appliances and portable tools
CAT III: Measurement category III corresponds to measurements on building
CAT IV
directly connected to low-voltage installations.
installations.
Example: measurements on distribution panels, cabling, etc.
: Measurement category IV corresponds to measurements taken at the
source of low-voltage installations.
Example
: meters and measurement on overvoltage protection devices.
• Comply with environmental and storage conditions.
• Ensure the three-wire power lead - phase/neutral/earth – that comes with
the instrument is in good condition.
It complies with standard EN 61010-1 (2004): it should be connected to
the instrument as well as the network (variation from 90 to
264 VAC).
Read carefully all the notes preceded by the symbol .
Connect the instrument to an outlet with a ground pin.
The instrument power supply is equipped with an electronic protection
system which is reset automatically when the fault is eliminated.
Be sure not to obstruct the ventilation holes.
As a safety measure, use only suitable leads and accessories supplied
with the instrument or approved by the manufacturer.
When the instrument is connected to the measurement circuits, never
touch an unused terminal.
Warning: Risk of danger.
Refer to the user manual to find out the nature of the potential hazards and
the action necessary to avoid such hazards.
Selective sorting of waste for the recycling of electrical and electronic
equipment. In accordance with the WEEE 2002/96/EC directive: should not
be processed as household waste.
Earth
Warranty
This equipment is warranted to be free of defects in materials or
workmanship, in accordance with the general terms and conditions of sale.
During this period, the manufacturer only can repair the equipment.
The manufacturer reserves the right to carry out repair or replacement of
all or part of the equipment.
In the event that the equipment is returned to the manufacturer, initial
transport costs shall be borne by the customer.
The warranty does not apply in the event of:
• improper use of the equipment or use in connection with
incompatible equipment
• modification of the equipment without explicit authorization from the
manufacturer’s technical services
• repair carried out by a person not certified by the manufacturer
• adaptation for a specific application, not included in the definition of
Before the equipment is opened, it must be disconnected from the mains
supply and the measurement circuits, and the operator must not become
charged with any static electricity. This could cause the destruction of
internal parts.
Any adjustment, servicing or repair of the unit under power must be
undertaken only by qualified personnel, after reading the instructions in
this manual.
A qualified person is a person who is familiar with the installation, its
construction, its use and the hazards that exist. They are authorized to
activate and deactivate the installation and equipment, in compliance with
the safety instructions.
Information and contact details: Get in touch with your nearest distributor.
All the equipment was verified mechanically and electrically before
shipping.
When you receive it, carry out a quick check to detect any damage that
may have occurred during transport.
If necessary, contact our sales department immediately and register any
legal reservations with the carrier.
In the event of reshipping, it is preferable to use the original packaging.
Indicate the reasons for the return as clearly as possible in a note attached
to the equipment.
Error messages See §. Technical specifications P. 143.
Mains power
supply
This is a user manual for the MTX 1052 and MTX 1054.
Most screen copies are made from an MTX 1054.
Check the good condition of the power supply cable to be connected, first
to the back of the instrument and then to a 50-60Hz power socket with an
earth link.
When lit, the LED at the back indicates that the mains voltage is correct
for the oscilloscope.
Connect the oscilloscope and the "Host PC" to the "Ethernet Network" or
directly to one another via the Ethernet twisted cable.
The oscilloscope power supply is designed for:
•
a power supply that can vary from 90 to 264 VAC (nominal range
of use 100 to 240 VAC)
•
a frequency between 47 Hz and 63 Hz.
Fuse
consumption
Type: Time delay
2.5 A
250 V
5 x 20 mm
This protection fuse must only be replaced with an identical model.
Replacement must only be performed by qualified personnel.
Contact your nearest distributor.
Activation Connect the oscilloscope to the 50-60 Hz network.
Wait for about one minute before starting the
"SCOPEin@BOX" application software. Refer to the "First Installation"
Reducing
instructions that come with the instrument.
When you exit the "SCOPE in@BOX" software, the remote virtual
oscilloscope switches to reduced consumption mode (except in
"Recorder" Instrument mode). Channels are put on standby but the
microprocessor remains active.
When a new work session is opened, the oscilloscope is automatically
switched to normal consumption.
To save working parameters correctly, exit the "SCOPEin@BOX"
software before disconnecting the instrument from the 50-60 Hz network
or Ethernet network.
"LOCAL" The instrument is directly connected to the control PC via an "Ethernet
"NETWORK" The instrument and control PC can be connected to the ETHERNET
Minimum PC
configuration
required
The instrument can operate in two modes:
twisted cable".
The PC controls one instrument at a time.
network with an untwisted Ethernet cable.
A different IP address must first be programmed on each instrument.
The SCOPEin@BOX software can be activated several times from the PC
to control several instruments at a time. By keeping one instrument
displayed on the PC screen and putting the other instruments as icons, all
the instruments can then be controlled in turn.
With the SCOPEin@BOX software, it is not possible to open an
instrument already open.
• Processor Pentium II or equivalent
• Memory 64 Mb
• Disk space 100 Mb
Installation of
SCOPEin@BOX
• Ports USB 1.1
• Ethernet Network Adapter 10BaseT
• Operating systems - Windows 98 - Millennium - 2000 - XP - Vista
The SCOPEin@BOX software operates with the NI-VISA V3.01
version: this version is included in the installation programme
supplied.
If a more recent version of NI-VISA is already installed on the PC, it
must first be uninstalled.
Please refer to the "First Installation" instructions that come with the
instrument.
ETHERNET and TCP/IP (Transmission Control Protocol/Internet Protocol)
are used to communicate on a company network.
network
Addressing Each piece of equipment under TCP/IP has a physical address (MAC
ADDRESS) and an Internet address (IP).
ETHERNET
physical address
A physical address or MAC ADDRESS, stored in the ROM, identifies each
piece of equipment on the network. The physical address enables the
equipment to determine the source of data "packet" transmission.
The physical address is a number coded over 6 bytes represented in
hexadecimal form.
Equipment manufacturers obtain physical addresses from the IEEE
organisation and assign them to the products manufactured in incremental
order. Each instrument has a unique MAC ADDRESS that cannot be
modified by the user.
IP address
An IP address is coded over 4 bytes, displayed in decimal format.
( Example: 132.147.250.10). Each field may be coded between 0 and
255 and is separated by a decimal point.
Unlike the physical address, the IP address can be modified by the user.
You must ensure that the IP address assigned to the instrument is
unique on your network. If an address is duplicated, network
operation becomes random.
The IP address is made up of two parts:
the network identifier (Network ID) for a given physical network
the host identifier (Host ID) identifying a specific item of equipment on the
same network.
There are 5 addressing classes. Only classes A, B and C are used to
identify the equipment. See below:
Class A
0XXXXXXX XXXXXXXX XXXXXXXX XXXXXXXX
Network ID Host ID
To communicate on the network, the equipment (oscilloscope, PC, printer)
must use a compatible IP address (identical Network ID field).
FTP (File Transfer Protocol) is used in the oscilloscope for fast file transfers to
or from a PC.
To use it, open the browser on the PC and, in the URL field, type the IP
address of the instrument, preceded by "ftp:"
Example: ftp://192.168.3.1
The oscilloscope is an FTP server.
With this protocol, the instrument can function as a Web server. You can
access the usual adjustments:
Display of traces on your PC via a browser (EXPLORER, NETSCAPE, …)
To use it, open the browser on the PC and, in the URL field, type the IP
address of the instrument, preceded by "http:"
Example: http://192.168.3.1
See §. Applications p. 133.
To be able to display the traces, you must install Java Virtual Machine JVM
SUN 1.4.2 (or higher) on your PC. This JVM can be downloaded from the
site:
CHx MATHx MEMx Channel selection
Probe Probe coefficient setting
Volt/div Selection of vertical sensitivity
Coupling Selection of input coupling
Position Setting for the trace vertical position
BWL Bandwidth limitation selection
Autoset CHx Vertical CHx autoset activation buttons
The offset multiplying coefficient of the probe assigns a multiplying
coefficient to the sensitivity of the channel in question.
The variation range is: 0 to 100 000.
The Volt/div channel vertical scale will be modified by the Probe value.
Ensure you reset the Probe coefficient value to 1 by disconnecting the
probe from the input.
Volt/div.
Input coupling AC blocks the DC component input signal and attenuates signals below
Vertical position
BWL
Vertical sensitivity: 15 calibres ranging from 2.5 mV / div. to 100 V / div.
10 Hz.
DC transmits the input signal to the DC and AC components.
GND internally, the instrument links the input of the channel selected at the
0 V reference level (with this coupling, the input impedance 1 MΩ //
13 pF is retained).
Variation range: ±10 div.
There are 4 possible bandwidth limitations for the vertical channel:
none, 15 MHz, 1.5 MHz and 5kHz.
BWL limits the bandwidth of the channel and its trigger circuit, attenuates
display noise and optimises triggering.
Autoset CHx
automatically adjusts vertical sensitivity to the signal present on CHx
channel input.
Auto ModeAutomatic acquisition and refreshment, even in the absence of a trigger event
TriggeredAcquisition and refreshment of the screen for each trigger event
Single strokeAcquisition of the signal and refreshment of the screen on the first trigger after resetting of the trigger
by clicking on
Principal trigger trigger on edge
Pulse trigger on pulse width
Delay delayed trigger
Point trigger after point
TV trigger on video signal
Line trigger on mains supply
Sources selection of the trigger source.
CH1, CH2, CH3 or CH4 (MTX 1054)
CH1, CH2 or EXT (MTX 1052)
Edge selection of the + trigger edge
selection of the - trigger edge
Level trigger level in mV
AUTO LEVEL 50 % automatically adjusts the trigger level to 50% of the
peak-to-peak amplitude of the signal.
Horizontal box
Settings
d. the control
buttons
T/div Sweep coefficient or acquisition time base
H-pos Trigger Horizontal position of the trigger
activates a general AUTOSET
for each active trace) and displays them in an
adjoining window
signals
captures the current traces (transfer of 50,000 points
launches / stops RUN/STOP acquisition
activates the Fast Fourier Transform (FFT) of the
1. Vertical sensitivity of the graphic representation (10 dB/div if log
representation, depends on the sensitivity of the channel on a linear
scale)
2. Position of the origin of the traces in relation to the graphic
representation origin
Horizontal sensitivity of traces: directly linked
order to limit time signal discontinuity effects
to the time base of the time representation
Choice of window type for the FFT calculation in
Choice of vertical representation scale for the curve
If an autoset is carried out with the FFT window active, the
frequency scale will be automatically set so as to position the
fundamental on the first environment division.
Description The Fast Fourier Transform is calculated according to the equation:
Reminder: Activation by clicking on the button in the
Horizontal box or on the icon on the tool bar.
The Fast FOURIER Transform (FFT) is used to calculate the
discrete representation of a signal in the frequency domain, based on its
discrete representation in the time domain.
FFT can be used in the following applications:
• measurement of the different harmonics and the distortion of a signal,
• analysis of a pulse response,
• search for noise source in logic circuits.
The FFT is calculated over 2500 points.
The instrument simultaneously displays the FFT and the trace f(t).
N
1
−
2
X (k) =
12
N
x nj
*( )*exp −
∑
N
n
=−
2
nk
π
N
for k∈ [0 (N – 1) ]
with: x (n): a sample in the time domain
X (k): a sample in the frequency domain
N: resolution of the FFT
n: time index
k: frequency index
The displayed trace represents the amplitude in V or dB of the various
signal frequency components depending on the selected scale.
The DC component of the signal is removed by software.
The finite duration of the study interval results in a convolution in the signal
frequency domain with a function sinx/x.
This convolution modifies the graphic representation of the FFT because of
the lateral lobes characteristic of the sinx/x function (unless the study
interval contains an whole number of periods).
Oscilloscope Instrument - Display
Five types of weighting windows are offered:
•
Rectangular
•
Hamming
•
Hanning
•
Blackmann
•
Flattop
The following table enables the user to choose the type of window
according to the type of signal, the desired spectral resolution and the
amplitude measurement accuracy:
Frequency
resolution
the best
good
good
poor
poor
Window
Rectangular
Hamming
Hanning
Blackman
Flat Top
Type of signal
transient
random
random
random or
mixed
sine wave
Spectral
resolution
poor
reasonable reasonable - 42 dB
good
the best
good
Amplitude
accuracy
poor
reasonable - 32 dB
good
the best
Highest
lateral lobe
- 13 dB
- 74 db
- 93 dB
The following table gives the theoretical maximum amplitude error for each
type of window:
Window
Rectangular
Hamming
Hanning
Blackman
Flat Top
Theoretical max. error in dB
3.92
1.75
1.42
1.13
< 0.01
This error is linked to the calculation of FFT when there is not a whole
number of periods for the signal in the observation window.
Shannon's theorem must be observed, that is to say the sampling
frequency "Fe" must be more than twice the maximum frequency contained
in the signal.
If this condition is not met, spectrum folding phenomena are observed.
For example, if the sampling frequency "Fe" is too low, the following will
occur:
- Truncating of the spectrum beyond "Fe/2"
- Modification of the spectrum below "Fe/2" (due to the overlap of several
The frequency of the fundamental is 10.1 kHz and that of the harmonic
3 to 30.3 kHz and the difference of level between the fundamental and the
first harmonic is 9.56 dB (which corresponds to an amplitude of the 3rd
harmonic, equal to around 33% of that of the fundamental).
Horizontal unit: this is calculated from the sweep coefficient:
Unit (in Hz/div.) =
Vertical unit: 2 possibilities are offered:
12.5
sweep coefficient
Ex: = 6.25 kHz
12.5
2 ms
a) Linear scale: by checking the linear scale in the FFT box
in V/div. = unit of the signal in its time representation V/div.
b) Logarithmic scale: by checking the logarithmic scale
Amplitude of the fundamental 1.40 V frequency 50.6 kHz
www.GlobalTestSupply.com
Oscilloscope Instrument - Display
"Oscilloscope" Instrument (contd.)
Graphic
representation
The FFT representation indicates symmetry in relation to the frequency
origin; only positive frequencies are displayed.
•
The "• " symbol, appearing before one of the options indicates the scale
selected.
•
The (window) MAX can be automatically located by clicking on the
button opposite. Cursor 1 is therefore positioned on the MAX of the
representation on the screen when pressed.
•
The exact location of the MAX around the active cursor (± 25 div) is
obtained by clicking on the 2nd button opposite. The MAX search zone is
evidenced when the button is pressed by a black rectangle around the
cursor.
•
Manual measurement can be carried out on the frequency
representation with the "unattached cursors"
( §. Measurement Menu Unattached manual cursors.
To avoid distorting the spectral content of the signal and obtain the
most accurate calculation of the FFT, it is advisable to work with a
signal peak-to-peak amplitude of 3 to 7 div.
If the amplitude is too low, accuracy will be reduced, and if it is too high,
over 8 divisions, the signal will be distorted, leading to the appearance of
undesirable harmonics.
Simultaneous time and frequency representation of the signal facilitates
monitoring of changes in the signal amplitude.
Effects of under-sampling on frequency representation:
If the sampling frequency is not correctly adjusted (less than or twice the
maximum frequency of the signal to be measured), the high-frequency
components will be under-sampled and appear in the graphic
representation of the FFT by symmetry (aliasing).
•
The Autoset function enables the above phenomenon to be avoided and
the horizontal scale adapted to make the representation more readable.
In the calculation of the FFT, the type of window applied is selected with the
up/down scroll bars or by clicking on the Window field in the FFT box.
Before calculating the FFT, the oscilloscope weights the signal to be
analyzed by means of a window acting as a band-pass filter. The choice of
window type is essential to distinguish between the various lines of a signal
and to make accurate measurements.
inputs the measuring unit of the channel concerned. This unit can be encoded
using a maximum of 3 characters (e.g.: VAC …)
gives access to the window for definition of the mathematical functions that
can also be directly accessed from the "Vertical" box with a right click on the
CHx channel labels.
A mathematical function can be input by:
1.
automatic input, with the assistance of the predefined functions editor
2.
callup of a ".fct" function file from the FCT file management menu
3.
direct input of the function via the keyboard in the edit window
In all cases, the user can use the edit function manually (maximum of 100
characters).
erases the content of the input box.
Don't forget to check this box if you wish to display the result of this function
before confirming your choice with the OK button.
Whether or not the function is activated, its definition is memorised, even
when the instrument has been turned off, until replaced by a new expression.
cancels the window without modifying the initial definition of the function or its
possible activation.
makes a syntaxical, semantic analysis of the function input and closes the
window, activating or not activating the function if the box
The multiple-choice dialogue boxes help the user to define the elementary
functions on channels (channel inversion, addition, subtraction,
multiplication and division).
2. ".FCT" file
management
Once the elements have been selected, input is validated by pressing
and the elementary function desired is generated (with
automatic scaling management) in the input window.
Mathematical functions stored in ".FCT" extension files can be saved or
recalled.
To call up a function: click on and select the desired file from
the management window.
The function is selected with the mouse and it is loaded with
The mathematical function is then copied into the edit window.
These functions, stored in the project FCT directory are:
•
C1MULC2.FCT
•
SQUARE.FCT
•
DAMPSINE.FCT
The C1MULC2.FCT = CH1*CH2/divv(4) calculates the product of 2 traces,
scaling the result so that it is framed in the screen.
The divv(4) factor is used to optimize the display as long as the source
signals have sufficient dynamics and no overshooting.
We have injected a square signal onto channel CH1 and a triangular signal
on channel CH2, centred on 0 Volts. We represent the result of the MATH3
= C1MULC2.FCT function on channel 3.
SQUARE.FCT
function
This is the definition of a square signal using the first 4 harmonics of a
Fourier series development.
(∗) t = abscissa of the sample (point) in the 50,000-sample (points) depth
acquisition memory.
divh(1) is equivalent to 5,000 samples (points) = 1 horizontal division.
The result of the calculation of a function is always an LSB. To obtain a
vertical division deviation, 32,000 LSBs are needed (amplitudes are
calculated using a virtual 19 dynamic 8 div. virtual ADC).
divv(1) = 1 vertical division = 32,000 LSBs.
With certain mathematical formulae, the calculation time may be long
The function divv (vertical division) is necessary to optimise the display.
math3 = (ch1 * ch2) / divv(1)
Divv(1) can be used to divide by 32,000 (1 vertical division = 32,000 LSBs):
the result of the multiplication is translated into divisions on the screen.
If Vpp of ch1 and ch2 had been 8 vertical divisions, the multiplication
would have had to be divided by divv(4).
When mathematical functions associated with traces are used, the
dynamics of the result obtained must be verified.
Correction of the result of the operations by mathematical functions (divv(),
divvh(), / …) is recommended to optimize the screen display.
Math3 = divv(3) * sin (2 * pi * t / 10 000) blue colour trace.
The trace obtained is a sine wave produced using the sin (sine) function,
according to its mathematical definition (2 x π x Frequency).
The amplitude is 6 divisions (divv(3) x 2 = 3 x 32,000 LSBs x 2).
The period equal to 10,000 samples (2 horizontal divisions) depends on the
time base.
The same trace can be obtained using the divh() function:
Math3 = divv(3) * sin (2 * pi * t / divh(2))
In this example, divh(2) is equivalent to 10,000 samples.
Note: 1 horizontal division = 5,000 samples
The value in seconds of the period T = divh(2) equal to 10,000 samples
(2 horizontal divisions) depends on the time base calibre
(in s/div.)
Math3 = divv(3) * cos (2 * pi * t / divh(2)) blue colour trace
The trace obtained with the cos() function is dephased by 90° in relation to
the one obtained with the sin() function..
If the sine function is programmed on CH2 and the cosine function on CH3
and the dephasing between the 2 channels is measured, we can check this
result:
Direct access to the Trigger menu with the icon opposite.
www.GlobalTestSupply.com
"Oscilloscope" Instrument (contd.)
Oscilloscope Instrument - The "Trigger" menu
Definition
The instrument has "advanced trigger" capability.
• The "Delay" and "Count" trigger modes require parameterization of a
second "auxiliary" trigger source. The auxiliary source may be the same
as the main source.
The trigger choice is validated by exiting from the menu with OK.
If …
… the user exits from the Main tab
… the user exits from the Pulse tab he is in Pulse trigger mode.
etc.
Then …
he is in Main trigger mode.
etc.
• There is only one Holdoff, although it can be programmed from the Main,
Delay, Count, TV and Line tabs.
When Delay or Count is being used, the Holdoff applies the auxiliary
source.
In the other cases, Holdoff applies to the main trigger source.
• Each trigger source has its own specific attributes: Coupling, Level,
Edge, Noise Reject, Filter
Trigger parameters 6 trigger modes: Main
Pulse Delay CountTVLine
each one can be accessed via a tab in the "Trigger parameters" window.
MTX 1054: Choice of main source: channel 1, 2, 3 or 4
MTX 1052: Choice of main source: channel 1, 2 or Ext
+ rising trigger slope
- falling trigger slope
AC - DC - LF reject - HF reject
The trigger symbol takes on the colour of the active trigger channel. The
active coupling of the trigger channel is indicated beside the Trigger symbol
in the "Oscilloscope Trace" window.
AC
TAC symbol
AC coupling (10 Hz to 200 MHz):
blocks the DC component of the signal.
DC coupling (0 to 200 MHz):
allows the entire signal through
T
LF
symbol
Reject of source signal frequencies < 10 kHz facilitates observation of signals with a DC component or an
unwanted low frequency
THF symbol
Rejection of source signal frequencies > 10 kHz: facilitates observation of slow signals with high-frequency noise
www.GlobalTestSupply.com
"Oscilloscope" Instrument (contd.)
adjusts the trigger level by moving the scrollbar with the mouse or directly
entering the value with the keyboard. The variation range is ± 8 vertical
divs.
No Hysteresis ≈ 0.6 div.
Yes Hysteresis ≈ 1.5 div.
Variation range: from 40.00 ns to 10.5 s
disables the trigger for a predefined period
stabilises the trigger on pulse trains.
When adjustment is finished, clicking on the button:
applies the new trigger parameters by exiting the window
applies the new parameters without exiting the window
exits the window without applying the new parameters
Oscilloscope Instrument - The "Trigger" menu
Example
Signal injected on CH1: trains of 4 sine wave signals with a frequency of
4 kHz and amplitude 2.5 Vcc with no DC component, separated by 1 ms.
Oscilloscope adjustment:
- Vertical sensitivity: 0.5 V/div.
- Time base: 500 µs/div.
- Trigger source: channel 1
- Trigger level: 0.250 V
- Edge: rising
The Holdoff stabilises the signal by inhibiting the trigger for a value of
Selection of pulse-width trigger.
In all cases, the effective trigger occurs on the pulse trailing edge.
< triggers on a pulse if its width is less than the value set
= triggers on a pulse if its width is equal to the value set
> triggers on a pulse if its width is greater than the value set
The pulse width is defined by the crossing of the signal with the
vertical Trigger level
MTX 1054: Choice of main source: channel 1, 2, 3 or 4
MTX 1052: Choice of main source: channel 1, 2 or Ext
Pulse type: + positive or - negative
The choice of the edge + (rising) or- (falling) defines the pulse polarity:
edge + defines a positive pulse between and
edge - defines a negative pulse between and
Filters the trigger channel: AC - DC - LF reject - HF reject
Variation range: ±8 div.
Trigger sensitivity changes from ≈ 0.6 div. to ≈ 1.5 div.
range from 20.00 ns to 10.5 s,
our example: 20.00 ns)
"Oscilloscope" Instrument (contd.)
Oscilloscope Instrument - The "Trigger" menu
Example
Signal injected on CH1: trains of 4 negative pulses with amplitude 2.25 Vcc,
no DC component, and a frequency of 10 kHz, separated by 500 µs.
Oscilloscope adjustment:
- Vertical sensitivity: 0.5 V/div.
- Time base: 200 µs/div.
- Trigger mode: Pulse
- Trigger source. : CH 1
- Trigger level: : 0.5 V
- Trigger on pulse: negative
- Trigger condition : "if the pulse width is < 50.05 µs"
The oscilloscope is triggered when the negative pulse width is less than the
specified pulse width (50.05 µs + tolerance).
Measurement of the negative pulse width is triggered on the falling edge
and the trigger is effective on the rising edge, if the pulse width meets the
comparison criterion chosen.
Trigger sensitivity changes from: ≈ 0.6 div. to ≈ 1.5 div.
Variation range: from 40.00 ns to 10.5 s
Signal injected on CH1: trains of 4 pulses with amplitude 2.25 Vcc and a
frequency of 10 kHz, separated by 600 µs.
Oscilloscope programming:
- Vertical sensitivity: 0.5 V/div.
- Time base: 200 µs/div.
- Trigger mode: Count
- Main trigger source: CH 1
- Auxiliary trigger source: CH1
- Number of events: 3
Trigger occurs on the 4th rising edge of the signal (the 1st rising edge on the
auxiliary channel triggers counting, the oscilloscope counts 3 rising edges
on the main channel and acquisition is then triggered).
The "" symbol indicates that the "Repetitive Signal" option has been
selected.
Activation of this option increases the time definition of a trace (up to
100 Gs/s) for a repetitive signal.
For time bases of less than 50 µs/div. (zoom mode not active), the
repetitive signal displayed is reconstituted by adding together successive
acquisitions.
Measurement on a microprocessor timing clock.
Min/Max Acquisition
If the signal is not repetitive, do not use this option as the aggregate
representation could be incorrect.
If Repetitive Signal mode is not selected, the time resolution will be 10 ns
(or 5 ns, if only one channel is active in single stroke). In this mode, all the
counts displayed are updated with each acquisition.
To indicate that repetitive signal mode is not selected, the "Non repetitive
Signal" message is displayed at the top of the window:
allows the signal to be sampled at high frequency (100 MS/s), even for
slow time base speeds. The display represents extreme value samples,
the Min and Max.
It is possible:
• to detect incorrect representation due to under-sampling
• to display short-term events (Glitch, > 10 ns).
Whatever time base is used, short-term events (Glitch, > 10 ns) are
displayed.
The "" symbol indicates that the "Min/Max Acquisition" mode is active.
Display in the Oscilloscope Trace
FFT Trace and
XY Trace windows
of the vertical unit,
the input coupling and
the BWL selection of each active channel.
Display modes
Envelope
Vector
There are two display modes available:
A vector is traced at the centre of the sample.
The minimum and maximum observed on each horizontal position of the
screen are displayed. This mode is used to display drifting in time or
modulation.
The “” symbol indicates the active display mode.
Selecting Zoom Off returns the user to the non-zoom representation of
traces: clicking on the button has the same effect.
A dynamic zoom is obtained by clicking on the button at the foot of
the Oscilloscope Trace or FFT Trace panel.
A static zoom is obtained by clicking on the button on the
Oscilloscope Control panel. The Zoom Off function only applies to the
dynamic zoom.
Enlarge mode (dynamic horizontal zoom) is indicated by the presence of
the 3 buttons opposite, at the foot of the Oscilloscope Trace or FFT Trace
panel.
When the FFT function is activated, the dynamic zoom on the time
representation is not operative.
Validation of XY mode.
The instrument adds a window containing the XY representation to the
current f(t) and FFT representations. The windows are simultaneously
updated.
F(t) and XY
representation of
these signals
The XY source menu is used for assigning one of the 4 traces available
to the X (horizontal) and Y axes (vertical).
MTX 1054: MTX 1052:
Validation of selections using the button opposite.
• Each axis is graduated into 8 divisions.
• The X and Y axes have the nr. of the channel that is assigned to them.
• The " • " symbols indicate the traces selected for each axis.
In XY mode, there are 2 manual measurement cursors
X1 Y1) and (X2 Y2). The vertical calibres of the traces selected for XY
(
display are indicated on the top left of the window.
The manual measurement cursors of the XY Trace window are separate
fromthoseoftheOscilloscopeTracewindow.
www.GlobalTestSupply.com
: trace window:
XY representation
Oscilloscope Instrument - The "Measurement" menu
"Oscilloscope" Instrument (contd.)
The
"Measurement"
menu
Reference
Trace 1
Trace 2
Trace 3
Trace 4
selects the Reference Trace for:
•
automatic measurement
•
phase measurement (automatic or manual)
•
measurement with a manual cursor
Selecting one of the active traces for which automatic or manual
measurements are to be made.
Only active traces can be selected. Inactive traces are shown greyed out.
Automatic
measurements
The "" symbol indicates the reference trace selected.
The measurement reference "Ref: Trace 1, 2, 3, 4" can also be selected
from the toolbar.
Opens the Automatic measurements window.
The 19 automatic measurements are made on the reference trace
selected. All the measurements that can be performed on this trace are
displayed and refreshed.
(- - - -) is displayed for measurements that cannot be performed.
The window is closed by clicking on the x icon.
Activation of automatic measurement does not display the cursors in the
trace display window. For measurements on periodic signals, choose the
time base coefficient so that at least 2 signal periods are displayed on the
screen.
19 automatic
measurements
Vmin
Vmax
Vpp
Vlow
Vhigh
Vamp
Vrms
Vavg
Over+
Tm
Td
W+
W-
DC
Over-
Sum
minimum peak voltage
maximum peak voltage
peak-to-peak voltage
established low voltage
established high voltage
amplitude
rms voltage
average voltage
positive overshoot
rise time
fall time
positive pulse width (at 50 % of Vamp)
negative pulse width (at 50 % of Vamp)
P
period
F
frequency
cyclic ratio
N
number of pulses
negative overshoot
sum of elementary areas (= integral)
•
Measurement
conditions
The measurements are made on the displayed part of the trace.
•
Any change to the signal will lead to updating of the measurements.
They are refreshed in step with acquisition.
•
For greater accuracy of the measurements displayed:
1.
represent at least two complete periods for the signal
2.
choose the calibre and vertical position so that the peak-to-peak
amplitude of the signal to be measured is represented on 4 to 7
divisions of the screen.
If the measurement cannot be performed, " - - - -" is displayed . For
example, if the time base chosen does not enable 2 complete signal
periods to be represented:
Manual
measurements
(dt, dv)
Cursor measurements
The blue and yellow measurement cursors are displayed as soon as the
menu is activated.
The two measurements made are:
dX = dt
dY = dv
The measurements performed and the cursors are linked to the selected
reference trace (see §. Reference Measurement).
•
The "" symbol indicates that the manual measurements
(time deviation between the two cursors)
(voltage deviation between the two cursors).
(dt, dv) are active.
• The measurement cursors can be moved directly with the mouse.
• If the unattached cursors option is not activated (see §. Measurement).
the cursors will remain linked to the reference trace during movement.
If the option is active, the cursors can be moved anywhere in the trace
display window.
• The dt and dv measurements in relation to the selected reference
are indicated in the measurement display area.
Example: (1)dt = dX = 1.05 ms, dv = dY = 1.21V
If manual phase measurement is selected:
The three cursors are unattached and can be placed anywhere in the
trace display window:
The blue and yellow cursors determine the reference period for
calculation of the phase and the dephasing value displayed depends on
the position of the black cursor in relation to these 2 cursors.
For manual measurement of the phase, a signal on the screen is all that
saves and calls up the following files: settings (.CFG)
trace (.TRC)
samples (.TXT)
function (.FCT)
Saving of the trace selected to its volatile reference memory:
Example: Trace 1 in Ref. 1. 1
The 4 traces have their reference memory.
The time base coefficient and vertical sensitivity of the reference are
indicated in the colour of the reference:
ExampleREF1: 1ms, 200mV and REF2: 1ms,50mV
• Only the 500 counts displayed on the screen are stored (not the 50
kcounts acquired); as a result, a Ref. x cannot be zoomed.
• All active channels can be simultaneously referenced with the icon
on the toolbar.
Saving a Trace (in the non-volatile memory, the PC hard disk): 50,000
counts are saved.
It can be saved in two formats: ".TRC" or ".TXT".
Saving files to recall them in the trace window
The files saved will take the extension .TRC; they can be recalled via the
"Memory Trace Recall TRC" menu.
The files saved have the extension .TXT; they cannot be recalled by the "Trace
Recall TRC" menu for screen display.
standard format for use in another programme (spreadsheet, etc.).
( Example:
However, they can be exported in a
Microsoft EXCEL) using the menu "Util Export to EXCEL".
www.GlobalTestSupply.com
Oscilloscope Instrument - The "Memory" menu
"Oscilloscope" Instrument (contd.)
Example of saving
a .TRC trace file
When the menu "Memory Save.'TRC’ Trace1" menu is selected, the
following window is displayed:
• Choose the save directory.
Enter the name of the file to be saved using the keyboard ( :
• Click on to save the file.
The name of the file saved takes the extension .TRC .
• Trace1 is saved as a .TXT file via the menu "Memory Save ‘*.TXT
Trace1". The name of the file saved then takes the extension .TXT (text
format) .
On the traces zoomed, the value of dX between the X1 and X2 cursors is:
dX = 73.9µs for the CH2, 3, 4 traces and dX = 148µs for MEM1.
When a trace is recalled, "MEMx" appears in the destination trace channel
zone. The sensitivity, coupling and band limitation become those of the
trace restored (they cannot be modified).
This is unique and cannot be modified by the user. It identifies the
instrument on the network.
IP addressThe user may keep the default IP address or enter a new one via the keyboard.
Subnet mask Input of the network mask
Gateway Programming of the gateway IP address (if a gateway is
used)
Validation of the new configuration settings.
Exit without validation
www.GlobalTestSupply.com
Oscilloscope Instrument - The "Util" menu
"Oscilloscope" Instrument (contd.)
Print …
Export to EXCEL…
This window allows the selection of the panel(s) that you wish to print. The
paper orientation (Portrait/Landscape) is selected with the switch opposite.
Start printing
Exit without printing
•
either by clicking on the icon on the toolbar
•
or via the menu "Util Export to EXCEL".
The following menu appears: ”Loading samples …”
It indicates the transfer of 50,000 samples corresponding to each trace
active at the time of the click.
Once the transfer is finished, the Capture Trace and Export to Excel
The memory zone to be exported corresponds to the one displayed in the
black frame of the first trace, itself represented on the lower graph. It can be
delimited using the Horizontal Zoom and by moving the frame with
the mouse or the buttons opposite.
The time necessary for an export to EXCEL depends on the number of
samples to be exported.
Display of data concerning the operation of the instrument since it was first
used:
•
the number of times switched on
•
the number of hours of use
•
the date last calibrated
•
the recommended date for the next calibration.
firmware …
The instrument time is automatically set to that of the PC when a working
session is set up.
When a working session is closed, the instrument switches to low
consumption mode, if not in recorder mode.
It automatically switches to normal consumption when a new working
session is set up.
•
Select the new version of the embedded software to be loaded.
By logging on to the www.chauvin-arnoux.com web site, the user can
download updates after inputting identification details.
A product support technician will answer any questions via the email
address.
www.GlobalTestSupply.com
Oscilloscope with SPO Persistence Instrument
Parallèle
"Oscilloscope with SPO Persistence" instrument
Selection
Presentation
Smart Persistence Oscilloscope (SPO) mode is activated from the
Instrument menu.
SPO Persistence:
•
displays unstable, transient phenomena and glitch
•
displays the evolution of the signal over a period of time, jitter and
modulation in the same way as when an analogue oscilloscope is used
•
causes acquisition to persist for a set period of time in order to observe
trace aggregation.
The light intensity or colour assigned to the point on the screen diminishes
if not renewed when a new acquisition process is implemented.
Acquisition is made according to 3 dimensions:
- time
- amplitude
- occurrence, which is a new dimension.
Acquisition SPO processing optimises the detection of transitory phenomena:
without SPO with SPO
Acquisition tasks and processing are
serial.
1 acquisition = 1 display
Acquisition
Representation on the screen of 500
points out of the 50,000 points
acquired.
Display of a segment to link the two
points.
Traitement
Affichage
Acquisition tasks and processing are in
parallel.
The number of acquisitions per second
can be multiplied by 100. The idle time
between two acquisitions is thus
considerably reduced.
N acquisitions = one display
Acquisition
Representation on the screen of 50,000
points acquired using an appropriate
compression system.
Display of a cloud of points not
interconnected. No interpolation.
Traitement
rapide
Affichage
Occurrence SPO brings a statistical dimension to the breakdown of samples.
The colour or light intensity highlight signal irregularities. They also enable a
distinction to be made between rare points and frequent points.
These settings can be modified by adjusting the persistence period.
The Vertical menu limits the user to the choice of the vertical unit.
Mathematical functions cannot be defined.
Ditto Oscilloscope mode.
The Horizontal menu limits the user to the selection/deselection of Min/Max
acquisition mode.
The Display menu limits the user to activation/deactivation of display of the
grid or units, coupling and limitation of the band of each channel active on
the trace.
The Measurement menu is limited to manual measurement with unattached
cursors and manual phase measurement.
This menu enables traces to be saved/recalled to/from .PER files and
instrument settings in .CFG files.
Tools
"?"
This menu is identical to oscilloscope mode but no EXCEL export is
possible.
This menu is identical to the one in Oscilloscope mode.
Adjustment of the main trigger threshold level using the mouse or
keyboard.
Adjustment of the auxiliary trigger level using the mouse or keyboard. This
adjustment is only active if the Exterior trigger Type is selected (otherwise
the Level2 box is greyed out).
This window indicates the trigger type of the channel. Recorder mode
enables a condition to be simultaneously monitored for each active
channel.
Pretrig is
monitored for
each type of
trigger.
No trigger: if all the channels are in this mode, the instrument observes
the trace indefinitely (continuously). When stopped, only 50,000 points are
saved.
••••
Lower than:
triggering takes place when the
signal drops below the Level1
threshold
lower trigger
threshold.
••••
Lower/higher than
triggering takes place when the
signal drops below or rises above
the threshold.
••••
Higher than:
lower trigger
threshold
upper trigger
triggering takes place when the
signal rises above the threshold.
••••
Outside:
triggering takes place when the
signal goes outside the window
threshold
thresholds
upper trigger
the window
defined by the two thresholds,
Level1 and Level2.
Trigger:
A half-division hysteresis is applied to prevent ill-timed triggers.
- Channel 1 is set with a 1.00V “Greater than” trigger for Level1.
- Channel 2 is set with a “Exterior” type trigger defined by a Level1 = 5.00V
and a Level2 = 4.00V.
- Channels 3 and 4 do not have any trigger.
(∗) MATH4 for the
MTX 1052
In this case, the trigger takes place on CH1 when the signal exceeds a level
of 1.00V.
(∗)
There is no trigger on CH2 because the signal amplitude is within the
window defined by Level1 = 5.00V and Level2 = -4.00V and the
programmed trigger condition is: "Outside" the window specified.
- Channel 1 is set with a 2.5V “Greater than” trigger for Level1.
- Channel 2 is set with an “Exterior" type trigger.
In this case, triggering takes place on channel CH2 since the condition on
channel CH1 is not met.
(∗)
Triggering takes place on the rising edge of CH2 when the
signal on channel CH2 exceeds 1.00V and goes out of the window specified
by "Level1 = 1.00V and Level2 = -4.00V".
Display of the sensitivity, coupling and bandwidth of the channels activated
2.
Types of trigger selected on the channels
3.
Levels of trigger associated with the channels
4.
Current status of acquisition
5.
Transition to the Next/Previous 10 faults
6.
Position "0 V" of the channels
7.
Fault separator
8.
Number of the 10 faults displayed
9.
Selection of the fault to be displayed
3.
4.
Trace display box
File capture
Number of files
created
Trigger types
Trigger higher than the last channel activated
Trigger lower than the last channel activated
Trigger higher/lower than the last channel activated
Trigger outside the window of the last channel activated
The colour of the level indicator is that of the channel activated.
This button transfers the 50,000 points corresponding to a recording to the
PC and analyses them.
When this button is pressed, two additional windows are opened after
downloading:
•
Capture: Recorder Control
•
Capture: Recorder Trace
(∗)
(∗) MATH4 for
the MTX 1052
This panel indicates the values of the various parameters used to capture
this recording:
•
vertical,
•
horizontal
•
and trigger
at the moment the user clicks on the capture button.
It is associated with the Capture: Recorder Trace panel (next page)
When one of the 2 windows is closed, they disappear at the same time.