Chauvin Arnoux®, Inc. d.b.a. AEMC® Instruments
certifies that this instrument has been calibrated using
standards and instruments traceable to international
standards.
We guarantee that at the time of shipping your
instrument has met its published specifications.
An N.I.S.T. traceable certificate may be requested at
the time of purchase, or obtained by returning the
instrument to our repair and calibration facility, for a
nominal charge.
The recommended calibration interval for this
instrument is 12 months and begins on the date of
receipt by the customer. For recalibration, please use
our calibration services. Refer to our repair and
Thank you for purchasing the Current Probe MR415 or MR416. For best results
dangerous voltages. Type A current sensor as per IEC 61010-2-032.
cycle in accordance with standard ISO 14040.
and with regulations covering EMC.
compliance with Directive WEEE
2002/96/EC.
from your instrument and for your safety, read the following operating instructions
carefully and comply with the precautions for use.
This instrument is compliant with the IEC 61010-2-032 safety standard for
voltages of 300V with respect to earth in measurement category IV, or 600V in
category III.
Symbols
WARNING, risk of DANGER! The operator must refer to these
instructions whenever this danger symbol appears.
Application or withdrawal authorized on conductors carrying
Equipment is protected by double insulation.
Battery.
USB.
Useful information or tip.
Direction of the current.
The product is declared recyclable following an analysis of the life
The CE marking guarantees conformity with European directives
The trash can with a line through it means that in the European
Union, the product must undergo selective disposal for the recycling
of electric and electronic material, in
Definition of Measurement Categories (CAT)
CAT IV Corresponds to measurements taken at the source of low-voltage
installations.
Example: power feeders, counters and protection devices.
CAT III Corresponds to measurements on building installations.
Example: distribution panel, circuit-breakers, machines or fixed industrial
devices.
CAT II Corresponds to measurements taken on circuits directly connected to
low-voltage installations.
Example: power supply to domestic electrical appliances and portable tools.
These instructions are intended to ensure the safety of users and proper
operation of the instrument. Failure to observe these safety instructions may
result in electric shock, fire, explosion, and destruction of the instrument and/or
installations.
The operator and/or the responsible authority must carefully read and clearly
understand the various precautions to be taken in use.
Do not use the instrument on networks on which the voltage or category
exceeds instrument specifications.
Never exceed the protection limits stated in the specifications.
Observe the environmental conditions of use, including relative humidity,
altitude, degree of pollution, and place of use.
Do not use the instrument if it appears to be damaged, incomplete, or not
properly closed.
Before each use, check the condition of the insulation on the leads, housing,
and accessories. Any component on which the insulation is deteriorated
(even partially) must be set aside for repair or scrapping.
When handling the instrument, keep your fingers behind the physical guards.
Use suitable means of protection.
All troubleshooting and metrological checks must be performed by
Upon receiving your shipment, make sure that the contents are consistent with
the packing list. Notify your distributor of any missing items. If the equipment
appears to be damaged, file a claim immediately with the carrier and notify your
distributor at once, giving a detailed description of any damage. Save the
damaged packing container to substantiate your claim.
Ordering Information
AC/DC Current Probe Model MR415 ........................................... Cat. #1200.80
Includes 9V battery, multi-language safety data sheet and user manual.
AC/DC Current Probe Model MR416 ........................................... Cat. #1200.82
Includes 9V battery, multi-language safety data sheet and user manual.
Replacement Parts/Accessories:
Cable – 6’ USB type A to Micro type B ............................................. Cat. #2138.66
Adapter – US Wall plug to USB ........................................................ Cat. #2153.78
The Models MR415 and MR416 are clamp-on current probes that measure DC
currents up to 600A, AC currents up to 400A
AC+DC currents without opening the circuit in which the currents flow. They
indicate the shape and amplitude of the current measured in the form of a
voltage.
These instruments can be used with a multimeter, wattmeter, recorder, and other
instruments. They can be powered by a battery or with 5V
micro‑USB cable.
The MR415 and MR416 include the following features:
overage indicator
power
zero adjustment
Auto Standby feature
one or two ranges, depending on the model (sensitivity 1 and 10mV/A)
micro‑USB connector to connect an external power supply
Before changing batteries: set the switch to OFF and remove the clamp
123
2.1 Battery Installation
from the circuit under measurement.
1.Using a screwdriver, remove the battery compartment cover (1) from the
back of the housing (see Figure 2).
2.Connect the battery to the snap-on connector (2), observing polarity.
3.Place the battery into the battery compartment (3).
4.Replace the battery compartment cover and screw it onto the housing.
Figure 2
2.2 External Power (Optional)
For long‑term measurements, you can connect the clamp to external power via
any micro‑USB adapter that delivers 100mA or more. If external power is
disconnected, the clamp automatically switches to battery operation.
The isolation between the type B micro‑USB connector and the measurement
output is 600V CAT III. This enables you to safely connect the clamp to
measuring instruments with uninsulated inputs. The type B micro‑USB connector
must not be in contact with conductors or uninsulated parts at dangerous voltage.
When operating on external power, the Auto Standby feature is disabled. The
color of the ON indicator shows whether automatic standby is enabled (green) or
disabled (yellow).
Turn on the clamp by pushing the slide switch to the 1mV/A or 10mV/A setting:
1mV/A corresponds to the 600A range (MR415 and MR416)
10mV/A corresponds to the 60A range (MR416)
The green ON indicator should light up:
If indicator blinks, less than 4 hours of battery life remains
If indicator fails to light, replace the battery (see § 4.2)
2.4 Auto Standby
After 10 minutes of operation without user action (such as pressing the DC Zero
button), the clamp automatically enters Standby mode. In this mode, the ON
indicator goes OFF.
To reactivate the clamp, press DC Zero button or change the switch to any
setting other than OFF.
To disable automatic Standby, press and hold down DC Zero when turning the
instrument ON. The ON indicator blinks to indicate that the request has been
applied; and then glows steady yellow when you release the DCZero button.
2.5 DC Zero Adjustment
1.Remove the clamp from the circuit under measurement.
2.Turn the clamp ON.
3.Connect the clamp to the measuring instrument. The phase is on the
red lead.
4.Press the DC Zero button.
5.The OL indicator lights for approximately three seconds to indicate that
the zero adjustment is in progress.
6.If the zero has been correctly adjusted, the OL indicator goes OFF.
If it remains ON, the zero cannot be adjusted. In this case, ensure the
clamp is not on a conductor and that its jaws are correctly closed. Then
press DC Zero again.
Alternately, turn the clamp OFF and then back ON. The previous zero
adjustment will remain in effect.
2.Clamp the jaws around the conductor to be measured. Use the
centering marks on the jaws to position the clamp around the conductor.
If the measurement is to be used in a power calculation, ensure the
arrow on the clamp jaws (see Figure 1) points in the direction of the
current flow: source
3.Release the trigger, ensuring the jaws are completely and correctly
closed.
4.Observe the measurement displayed on the measuring instrument.
5.If the OL indicator lights, the current is too high to be measured. If you
are using the MR416 and the sliding switch is set to the 10mV/A range,
change the setting to 1mV/A.
load.
Figure 3 (MR416 shown)
2.6.2 Converting to Current
The Models MR415 and MR416 can both measure current up to 600A, with 1mV
corresponding to 1A. In addition, the MR416 provides a second measurement
range up to 60A, with 10mV of output corresponding to 1A.
To convert the clamp output to current, divide the voltage reading on the
connected measuring by the V/A coefficient. For example, in the 60A range a
reading of 45mV corresponds to a current of 4.5A.
Position of the conductor Centered on the marks on the jaws
Measurement frequency DC to 65Hz sine wave
External electrical field zero
External DC magnetic field (earth) <40V/m
External AC magnetic field zero
Input impedance ≥ 1MΩ and ≤100pF
The intrinsic uncertainty is the error defined under the reference conditions. It is
expressed as a percentage of the output signal (R) plus an offset in mV:
±(a% R + b)
Except for the battery, the instrument contains no parts that can be
replaced by personnel who have not been specially trained and
accredited. Any unauthorized repair or replacement of a part by an
"equivalent" may gravely impair safety.
4.1 Cleaning
Disconnect the instrument completely.
Use a soft cloth, dampened with soapy water.
Rinse with a damp cloth and dry rapidly with a dry cloth or forced air.
Do not use alcohol, solvents, or hydrocarbons.
Keep the clamp jaws as clean as possible.
4.2 Battery Replacement
The battery must be replaced if the ON indicator remains unlit when the
instrument is turned ON.
1. Disconnect the instrument completely and set the switch to OFF.
2. Remove the battery compartment cover from the instrument casing (see §
2.1).
3. Remove the old battery.
4. Insert the replacement battery into the snap-in battery connector, and place
it into the battery compartment.
5. Replace the battery compartment cover.
Spent batteries must not be treated as ordinary household waste. Take
them to the appropriate collection point for recycling.