Aemc instruments 605 User Manual

User Manual
ENGLISH
POWER CLAMP-ON METER
605
2
CONTENTS
RECEIVING YOUR SHIPMENT ........................................................................... 8
ORDERING INFORMATION ................................................................................ 8
1 PRESENTATION .......................................................................................... 9
1.1 THE ROTARY SWITCH ............................................................................... 10
1.2 THE FUNCTION BUTTONS ........................................................................ 11
1.3 THE DISPLAY .............................................................................................. 12
1.3.1 Display Symbols ........................................................................... 12
1.3.2 Measurement Capacity Exceeded (OL) ........................................ 13
1.4 THE TERMINALS ........................................................................................ 14
2 THE BUTTONS ........................................................................................... 15
2.1
2.2 (YELLOW) BUTTON (SECOND FUNCTION) ............................................ 16
2.3 BUTTON ............................................................................................. 16
2.4 BUTTON ............................................................................................. 17
2.4.1 Normal Mode ................................................................................ 17
2.4.2 The MAX/MIN Mode + Activation of the HOL D Mode ................... 18
2.4.3 Access to the True Inrush® Mode ( set to ) ................... 18
2.5
2.5.1 Normal Mode ................................................................................ 19
2.5.2 The Hz Function + Activation of the HOLD Mode ......................... 19
2.6 BUTTON ........................................................................................... 20
3 USE ............................................................................................................. 21
3.1 INSTALLING THE BATTERIES ................................................................... 21
3.2 TURNING THE CLAMP-ON METER ON .................................................... 21
3.3 TURNING THE CLAMP-ON METER OFF .................................................. 21
3.4 CONFIGURATION ....................................................................................... 22
3.4.1 Configuring the Maximum Resistance for Continuity .................... 22
3.4.2 Auto Power OFF ........................................................................... 22
3.4.3 Configuring the Threshold for True InRush® Measurement .......... 22
3.4.4 Default Configuration .................................................................... 23
3.5 VOLTAGE MEASUREMENT (V) ................................................................. 23
3.6 CONTINUITY TEST ............................................................................... 24
3.6.1 Lead Resistance Compensation ................................................... 25
3.7 RESISTANCE MEASUREMENT Ω ............................................................. 25
3.8 DIODE TEST ......................................................................................... 26
3.9 CURRENT MEASUREMENT (A) ................................................................ 26
3.9.1 AC Measurement .......................................................................... 26
3.9.2 DC or AC+DC Measurement ....................................................... 27
BUTTON ............................................................................................... 15
BUTTON ............................................................................................. 19
3
STARTING CURRENT OR OVERCURRENT (True InRush
3.10
MEASUREMENT ......................................................................................... 28
3.11 POWER MEASUREMENTS W, VA, VAR AND PF ....................................... 29
3.11.1 Single-Phase Power Measurement ............................................... 29
3.11.2 Balanced 3-Phase Power Measurement ....................................... 30
PHASE ROTATION MODE ............................................................... 31
3.12
3.13 FREQUENCY MEASUREMENT (HZ) ......................................................... 32
3.13.1 Frequency Measurement (V) ........................................................ 32
3.13.2 Frequency Measurement (A) ........................................................ 32
3.13.3 Frequency Measurement (W) ....................................................... 33
3.14 MEASUREMENT OF THE LEVEL OF HARMONICS (THD) AND
THE
3.14.1 Measurement of THD and Frequency of the Fundamental (V) ..... 34
3.14.2 Measurement of THD and Frequency of the Fundamental (A) ..... 34
4 SPECIFICATIONS ...................................................................................... 36
4.1 REFERENCE CONDITIONS ....................................................................... 36
4.2 SPECIFICATIONS UNDER THE REFERENCE CONDITIONS ................. 36
4.2.1 DC Voltage Measurement ............................................................. 36
4.2.2 AC Voltage Measurement ............................................................. 37
4.2.3 AC+DC Voltage Measurement ...................................................... 37
4.2.4 DC Current Measurement ............................................................. 38
4.2.5 AC Current Measurement ............................................................. 38
4.2.6 AC+DC Intensity Measurement .................................................... 39
4.2.7 True Inrush® Measurement ........................................................... 39
4.2.8 Continuity Measurement ............................................................... 40
4.2.9 Resistance Measurement ............................................................. 40
4.2.10 Diode test ...................................................................................... 40
4.2.11 Active DC Power Measurements .................................................. 41
4.2.12 Active AC Power Measurements .................................................. 41
4.2.13 Active AC+DC Power Measurements ........................................... 42
4.2.14 Measurement of Apparent AC Po wer ........................................... 43
4.2.15 Measurement of Apparent AC+DC Power .................................... 43
4.2.16 Measurement of Reactive AC Power ............................................ 43
4.2.17 Measurement of reactive AC+DC power ....................................... 44
4.2.18 Calculation of the Power Factor .................................................... 45
4.2.19 Frequency Measurements ............................................................ 45
4.2.20 Specifications in THDr .................................................................. 46
4.2.21 Specifications in THDf ................................................................... 46
4.2.22 Indication of order of the phases ................................................... 46
4.3 ENVIRONMENTAL CONDITIONS .............................................................. 47
4.4 MECHANICAL SPECIFICATIONS .............................................................. 47
4.5 POWER SUPPLY ........................................................................................ 47
4.6 COMPLIANCE WITH INTERNATIONAL STANDARDS ............................. 47
4.7 ENVIRONMENTAL VARIATIONS ............................................................... 48
FREQUENCY OF THE FUNDAMENTAL (NETWORK) ..................... 34
®
)
4
5 MAINTENANCE .......................................................................................... 49
5.1 WARNING .................................................................................................... 49
5.2 CLEANING ................................................................................................... 49
5.3 REPLACEMENT OF THE BATTERIES ...................................................... 49
6 REP AIR AN D CALIBR ATION .................................................................... 50
7 TECHNICAL AND SALES ASSISTANCE .................................................. 50
8 LIMITED WARRANTY ................................................................................ 51
9 WARRANTY REPAIRS .............................................................................. 51
5
Thank you for purchasing a Model 605 Clamp-on Meter.
For best results from your instrument and for your safety, read the enclosed operating instructions carefully an d comply with the precautions for use. These products must be only used by qualified and trained users.
Meanings of the symbols used on the device
CAUTION - Risk of Danger! Indicates a WARNING and that the operator
must refer to the user manual for instructions before operating the instrument in all cases where this symbol is marked.
Risk of electric shock. The voltage at the parts marked with this symbol may be dangerous.
Refers to a type A current sensor. This symbol signifies that application around and removal from HAZARDOUS LIVE conductors is permitted.
1.5 V battery
The CE marking indicates compliance with European directives
Double insulation or reinforced insulation In the European Union, this product is subject to a separate collection
system for recycling electrical and electronic components In accordance with directive WEEE 2002/96/EC
AC – Alternating current AC and DC – Alternating and direct current Ground/Earth
6
PRECAUTIONS FOR USE
This device complies with safety standards IEC-61010-1 and 61010-2-032 for voltages of 1000V in category IV at an altitude of less than 2000m, indoors, with a degree of pollution not exceeding 2. These safety instructions are intended to ensure the safety of persons and proper operation of the device.
The operator and/or the responsible authority must carefully read and clearly
understand the various precautions to be taken in use.
If this instrument is used other than as specified, the protecti on it provides may be
compromised, thereby endangering you.
Do not use the instrument in an explosive atmosphere or in the presence of flammable
gases or fumes.
Do not use the instrument on networks of which the voltage or category exceeds those
mentioned.
Do not exceed the rated maximum voltages and currents between terminals or with
respect to earth.
Do not use the instrument if it appears to be damaged, inc omplete, or not properly
closed.
Before each use, check the condition of the insulation on the leads, housing, and
accessories. Any element of which the insulation is deteriorated (even partiall y) must be set aside for repair or scrapped.
Use leads and accessories rated for voltages and categories at least equal to those of
the instrument. If not, an accessory of a lower category lowers the category of the combined Clamp + accessory to that of the accessory.
Observe the environmental conditions of use. Do not modify the instrument and onl y use factory replacement parts. Repairs and
adjustments must be done by approved qualified personnel.
Replace the batteries as soon as the symbol app ears on the display of the unit.
Disconnect all leads before opening the battery compartment cover.
Use personal protective equipment when conditions require. Keep your hands away from the unused terminals of the instrument. When handling the test probes, alligator clips, and clamp ammeters, keep your fingers
behind the physical guard.
As a safety measure, and to avoid repeated overloads on the inputs of the device,
configuration operations should only be performed when the device is disconnected from all dangerous voltages.
7
MEASUREMENT CATEGORIES
Definitions of the measurement categories: CAT II: Circuits directly connected to the low-voltage installation.
Example: power supply to household electrical appliances and portable tools.
CAT III: Power supply circuits in the installation of the building.
Example: distribution panel, circuit-breakers, fixed industrial machines or devices.
CAT IV: Circuit s supplying the low-voltage instal lation of the building.
Example: power lines, meters, and protection devices.

RECEIVING YOUR SHIPMENT

Upon receiving your ship m ent, make sure that the contents are consistent with the packing list. Notify your d istributor of any missing it ems. If the equipm ent appears to be damaged, file a cla im immediat ely with t he carrier a nd notif y your distrib utor at once, giving a detailed desc ription of any d amage. Save t he damaged pack ing container to substantiate your claim.

ORDERING INFORMATION

Clamp-on Meter Model 605 .......................................................... Cat. #2139.60
Includes set of 2 color-coded silicone insulated te st leads, test probes and alligator clips, soft carrying case, 4x1.5V AA batteries and user manual.
Replacement Parts:
Soft Carrying Case ........................................................................................... Cat. #2139.72
Set of 2 Color-coded Silicone Test Leads, Test Probes & Alligator Clips ........ Cat. #2152.05
8
1
3
2
7
ring marks

1 PRESENTATION

The Clamp-on Meter Model 605 is a prof essional electrical measuring ins trument that combines the following funct i ons :
Current measurement Measurement of InRush current / overcurrent (True InRush Voltage measurement Frequency measurement Measures harmonic distortion (THD) Continuity test with buzzer Resistance measurement Diode test Power measurements (W, VA, var and PF) Phase order indication
Item
Jaws with cente
1
(see connection principles)
2 Physical Guard ­3 Rotary Function Switch 1.1 4 Function Buttons 2 5 Backlit Display 1.3
4
5
6 Input Terminals 1.4 7 Trigger -
Designation See §
®
)
3.5 to
3.12
6
Figure 1: Clamp-on Meter Model 605
9
1 2 3 4 5
6

1.1 THE ROTARY SWITCH

The rotary switch has six positions. To access the , , , , functions, set the switch to th e desired function. The functi ons are described in the table below.
Figure 2: The Function Rotary Sw itch
Item Function See §
1 OFF mode – Turns the clamp-on meter off 3.3 2 AC, DC, AC+DC voltage measurement (V) 3.5 3 Continuity test
Resistance measurement
Diode test 4 AC, DC, AC+DC current measurement (A) 3.9 5 Power measurements (W, var, VA) and calculation of the
power factor (PF) AC, DC, AC+DC
6
Phase rotation order indicator -
3.6
3.7
3.8
3.11
3.12
10

1.2 THE FUNCTION BUTTONS

1 2 3
4 5 6
Figure 3: The Function Buttons
Item Function See §
1 Holds the last value on the display
Zero correction A
Lead resistance compensation in the continuity and ohmmeter functions
2 Selects the type of measurement and configuration
functions (AC, DC, AC+DC)
Selection of single-phase or 3-phase measurement 3 Enables/disables display backlighting 2.3 4 Enables/disables the MAX/MIN/PEAK mode
Enables/disables the True InRush
measurement (A)
5 Performs Frequency measurements (Hz) and selects
measurement of harmonic distortion (THD)
Display of the powers W, VA, var and PF 6 Activation of ΔREL mode – Displays differential and
relative values
DC/AAC+DC/WDC/WAC+DC
®
mode in current
2.1
3.9.2
3.6.1
2.2
2.4
2.5
2.6
11
AC+DC
Alternating and direct current or voltage
PEAK+
Maximum peak value
4

1.3 THE DISPLAY

7
1
2
3
5
Figure 4: The Display
Item Function See §
1 Mode selection display 2 2 Display of the measurement value and unit 3.5 to 3.12 3 Display of the MAX/MIN/PEAK modes 2.4 4 Type of measurement (AC or DC) 2.2 5 Total 3-phase power measurements 3.11.2 6 Selected resistance mode display 1.1 7 Low battery indication 5.3

1.3.1 Display Symbols

Symbol Designation
6
AC Alternating current or voltage DC Direct current or voltage
REL
Ref
Relative value, with respect to a reference Reference value Storage of the values and display hold
Max Maximum DC or RMS value
Min Minimum DC or RMS value
12
PEAK-
Minimum peak value
Balanced total 3-phase power measurement
W
Watt
var
Reactive power
VA
Apparent power
PF
Power factor
THDf
Total harmonic distortion with respect to the fundamental
Total harmonic distortion with respect to the true RMS
V Volt
Hz Hertz
A Ampere
% Percentage
Ω Ohm
m Milli- prefix
k Kilo- prefix
THDr
value of the signal.
Phase order indication
Lead resistance compensation Continuity test Diode test Auto Power Off disabled Low battery indicator
The display of “rdy”, for “ready”, indicates that the device is ready (Phase Order Indication function).

1.3.2 Measurement Capacity Exceeded (OL)

The OL (Over Load) symbol is displayed when the display capacity is exceeded.
13

1.4 THE TERMINALS

The terminals are used as follows:
1 2
Figure 5: The Terminals
Item Function
1
COM (black) Input Terminal Jack
2 + Positive (red) Input Terminal Jack
14
ADC
W
AC+DC
This mode operates if the MAX/MIN or

2 THE BUTTONS

The buttons respond differently to short, long, and sustained presses. The , and keys provide additional functions as well as the
detection and acquisition of parameters complementary to the basic measurements.
Each of these buttons can be used independently of the others or in conjunction with each other. T his makes navigation simple and intuit ive when reviewing measurement result s.
It is possible, for ex ample, to either look up i n sequence the MAX, MIN, etc. values of the RMS voltage alone, or l ook up in sequence all of the MAX (or MIN, or PEAK) values of all power measurements (W, VA, var, etc.).
In this section, the icon represents the possible posi tions of the switch for the button’s functionality.
2.1

BUTTON

This button is used to:
Store and look up the last values acquired specific to each funct ion (V, A,
, W) according to the specific modes previously activated (MAX/MIN).
The present display is then maintai ned while the detection and acquisition of new values continues.
Perform automatic lead resistance compensat ion (see § 3.6.1). Perform automatic zero correction in
ADC/
NOTE: This button is invalid for the Phas e Order Indication function.
Successive
presses on
long
(> 2 sec)
long
(> 2 sec)
A
AC+DC
W
DC
First press: Holds the last value displayed
Second press: Returns to normal display mode
(the value of each new measurement is displayed)
Performs automatic zero correction (see 3.9.2) NOTE:
HOLD modes (short press) are first de activated
Performs automatic lead resistance compensation (see 3.6.1)
See § 2.4.2 and § 2.5.2 for the button functionality in com bination with the
and buttons.
AC+DC
Function
and WDC/
(see § 3.9.2).
AC+DC
15

2.2 (YELLOW) BUTTON (SECOND FUNCTION)

This button is used to select the t ype of m easureme nt (AC, DC, AC+DC) and the second functions marked in yellow ne xt to the relevant positions of the switch.
It can also be used to modify the default values in the configuration mode (see § 3.4).
NOTE: This button is invalid in the MAX/MIN/PEAK, HOLD and ΔREL modes.
Successive
presses on
short
long (>2 sec)
- Selects AC, DC or AC+DC. Depending on your choice, the screen displays AC, DC or AC+DC
- Cycles through the continuity , and diode test modes and returns to the continuity test
- Resets the measurement process for the phase order indication function
- Displays the total 3-phase power of a balanced system ( is displayed). Press again to return to the display of the single­phase power ( is off)
Function

2.3 BUTTON

Successive
presses on
- Enables/disables display backlighting
NOTE: The backlighting is automatically disabled at the end of 2 minutes.
16
Function