Aemc instruments 603 User Manual

Page 1
CLAMP-ON METER
User Manual
ENGLISH
603
Page 2
2
Page 3
CONTENTS
ORDERING INFORMATION ................................................................................ 7
1 PRESENTATION .......................................................................................... 8
1.1 THE ROTARY SWITCH ................................................................................. 9
1.2 THE FUNCTION BUTTONS ........................................................................ 10
1.3 THE DISPLAY .............................................................................................. 11
1.3.1 Display Symbols ........................................................................... 11
1.3.2 Measurement Capacity Exceeded (OL) ........................................ 12
1.4 THE TERMINALS ........................................................................................ 12
2 THE BUTTONS ........................................................................................... 13
2.1
2.2 (YELLOW) BUTTON (SECOND FUNCTION) ............................................ 14
2.3 BUTTON ............................................................................................. 14
2.4 BUTTON ........................................................................................... 15
2.4.1 Normal Mode ................................................................................ 15
2.4.2 The MAX/MIN Mode + Activation of the HOL D Mode ................... 16
2.4.3 Access to the True Inrush® Mode ( set to ) ................. 16
2.5
2.5.1 Normal Mode ................................................................................ 17
2.5.2 The Hz Function + Activation of the HOLD Mode ......................... 17
2.6 BUTTON ........................................................................................... 18
3 USE ............................................................................................................. 19
3.1 INSTALLING THE BATTERIES ................................................................... 19
3.2 TURNING THE CLAMP-ON METER ON .................................................... 19
3.3 TURNING THE CLAMP-ON METER OFF .................................................. 19
3.4 CONFIGURATION ....................................................................................... 20
3.4.1 Configuring the Maximum Resistance for Continuity .................... 20
3.4.2 Auto Power OFF ........................................................................... 20
3.4.3 Configuring the Threshold for True InRush® Measurement .......... 20
3.4.4 Changing the Default Temperature U nit ....................................... 21
3.4.5 Configuring the Adapter Function Sc ale Factor ............................ 21
3.4.6 Default Configuration .................................................................... 22
3.5 VOLTAGE MEASUREMENT (V) ................................................................. 22
3.6 CONTINUITY TEST ............................................................................... 23
3.6.1 Lead Resistance Compensation ................................................... 23
3.7 RESISTANCE MEASUREMENT Ω ............................................................. 24
3.8 DIODE TEST ......................................................................................... 24
3.9 CURRENT MEASUREMENT (A) ................................................................ 25
3.9.1 AC Measurement .......................................................................... 25
BUTTON ............................................................................................... 13
BUTTON ............................................................................................. 17
3
Page 4
DC Measurement .......................................................................... 26
3.9.2
3.10 STARTING CURRENT OR OVERCURRENT (True InRush®)
MEASUREMENT ......................................................................................... 27
3.11 FREQUENCY MEASUREMENT (HZ) ......................................................... 27
3.11.1 Frequency Measurement (V) ........................................................ 27
3.11.2 Frequency Measurement (A) ........................................................ 28
3.12 TEMPERATURE MEASUREMENT ............................................................ 29
3.12.1 Measurement without External Sensor ......................................... 29
3.12.2 Measurement with External Sensor .............................................. 29
3.13 ADAPTER FUNCTION MEASUREMENT ................................................... 30
4 SPECIFICATIONS ...................................................................................... 31
4.1 REFERENCE CONDITIONS ....................................................................... 31
4.2 SPECIFICATIONS UNDER THE REFERENCE CONDITIONS ................. 31
4.2.1 DC Voltage Measurement ............................................................. 31
4.2.2 AC Voltage Measurement ............................................................. 32
4.2.3 DC Current Measurement ............................................................. 32
4.2.4 AC Current Measurement ............................................................. 33
4.2.5 True Inrush® Measurement ........................................................... 33
4.2.6 Continuity Measurement ............................................................... 33
4.2.7 Resistance Measurement ............................................................. 34
4.2.8 Diode test ...................................................................................... 34
4.2.9 Frequency Measurements ............................................................ 34
4.2.10 Temperature Measurement ........................................................... 35
4.2.11 Adapter Function Measurement .................................................... 36
4.3 ENVIRONMENTAL CONDITIONS .............................................................. 36
4.4 MECHANICAL SPECIFICATIONS .............................................................. 37
4.5 POWER SUPPLY ........................................................................................ 37
4.6 COMPLIANCE WITH INTERNATIONAL STANDARDS ............................. 37
4.7 ENVIRONMENTAL VARIATIONS ............................................................... 38
5 MAINTENANCE .......................................................................................... 39
5.1 WARNING .................................................................................................... 39
5.2 CLEANING ................................................................................................... 39
5.3 REPLACEMENT OF THE BATTERIES ...................................................... 39
6 REP AIR AND CAL IBRATION .................................................................... 40
7 TECHNICAL AND SALES ASSISTANCE .................................................. 40
8 LIMITED WARRANTY ................................................................................ 41
9 WARRANTY REPAIRS .............................................................................. 41
4
Page 5
Thank you for purchasing a Model 603 Clamp-on Meter.
For best results from your instrument and for your safety, read the enclosed operating instructions carefully an d comply with the precautions for use. These products must be only used by qualified and trained users.
Meanings of the symbols used on the device
CAUTION - Risk of Danger! Indicates a WARNING and that the operator
must refer to the user manual for instructions before operating the instrument in all cases where this symbol is marked.
Risk of electric shock. The voltage at the parts marked with this s ymbol may be dangerous.
Refers to a type A current sensor. This symbol signifi es that application around and removal from HAZARDOUS LIVE conductors is permitted.
1.5 V battery
The CE marking indicates compliance with European directives
Double insulation or reinforced insulation In the European Union, this product is subject to a separate collection
system for recycling electrical and electronic components In accordance with directive WEEE 2002/96/EC
AC – Alternating current AC and DC – Alternating and direct current Ground/Earth
5
Page 6
PRECAUTIONS FOR USE
This device complies with safety standards IEC-61010-1 and 61010-2-032 for voltages of 1000V in category IV at an altitude of less than 2000m, indoors, with a degree of pollution not exceeding 2. These safety instructions are intended to ensure the safety of persons and proper operation of the device.
The operator and/or the responsible authority must carefully read and clearly
understand the various precautions to be taken in use.
If this ins trument is used other than as specified, the protection it provides may be
compromised, thereby endangering you.
Do not use the instrument in an explosi ve atmos phere or in the pres ence of fl ammable
gases or fumes.
Do not use the instrument on networks of which the voltage or category exceeds those
mentioned.
Do not exceed the rated ma ximum voltages and currents between terminals or with
respect to earth.
Do not use the i nstrument if it appears to be damaged, incomplete, or not properly
closed.
Before each use, check the condition of the insulation on the leads, housing, and
accessories. Any element of which the insulation is deteriorated (even partiall y) must be set aside for repair or scrapped.
Use leads and ac cessories rated for voltages and categories at least equal to those of
the instrument. If not, an accessory of a lower category lowers the category of the combined Clamp + accessory to that of the accessory.
Observe the environmental conditions of use.Do not modif y the instrument and only use factory replacement parts. Repairs and
adj
ustments must be done by approved qualified personnel.
Replace the batt eries as soon as the symbol app ears on the display of the unit.
Disconnect all leads before opening the battery compartment cover.
Use personal protective equipment when conditions require.Keep your hands away from the unused terminals of the instrument.When handling the tes t probes, alligator clips, and clamp ammeters , keep your fingers
behind the physical guard.
As a safety measure, and to avoid repeated overloads on the inputs of the device,
configuration operations should only be performed when the device is disconnected from all dangerous voltages.
6
Page 7
MEASUREMENT CATEGORIES
Definitions of the measurement cat egories : CAT II: Circuits directly connected to the low-voltage installation.
Example: power supply to household electrical appliances and portabl e tools.
CAT III: Power supply circuits in the installation of the building.
Example: distribution panel, circuit-breakers, fixed industrial machines or devices.
CAT IV: Circuits s upplying the low-voltage installation of the building.
Example: power lines, meters, and protection devices.

RECEIVING YOUR SHIPMENT

Upon receiving your shipm ent, make sure that the contents are consistent with the packing list. Notify your d istributor of any missing it ems. If the equipm ent appears to be damaged, file a cla im immediat ely with t he carrier a nd notif y your distrib utor at once, giving a detailed descript ion of any damage. Save the damaged packing container to substantiate your claim.

ORDERING INFORMATION

Clamp-on Meter Model 603 .......................................................... Cat. #2139.31
Includes set of 2 color-coded silicone insulated te st leads, test probes and alligator clips, K-thermocouple with 4mm integrated adapter, soft carrying case, 4x1.5V AA batteries and user manual.
Replacement Parts:
K-thermocouple with 4mm Integrated Adapter ................................................. Cat. #2139.71
Soft Carrying Case ........................................................................................... Cat. #2139.72
Set of 2 Color-coded Silicone Test Leads, Test Probes & Alligator Clips ........ Cat. #2152.05
7
Page 8
1
3
2
7

1 PRESENTATION

ring marks
The Clamp-on Meter Model 603 is a professional electrical measuring i nstrument that combines the following funct i ons :
Current measurementMeasurement of InRush current / overcurrent (True InRushVoltage measurementFrequency measurementContinuity test with buzzerResistance measurementDiode testTemperature measurementAdapter functi on (for use with external transducer s/measurement
devices)
®
)
Item
Jaws with cente
1
(see connection principles)
2 Physical Guard ­3 Rotary Function Switch 1.1 4 Function Buttons 2 5 Backlit Display 1.3
4
5
6
6 Input Terminals 1.4 7 Trigger -
Designation See §
3.5 to
3.12
Figure 1: Clamp-on Meter Model 603
8
Page 9

1.1 THE ROTARY SWITCH

1 2 3 4 5
The rotary switch has six positions. To access the , , , , , functions, set the switch to th e desired function. The functi ons are described in the table below.
6
Figure 2: The Function Rotary Switch
Item Function See §
1 OFF mode – Turns the clamp-on meter off 3.3 2 AC, DC voltage measurement (V) 3.5 3 Continuity test
Resistance measurement
Diode test 4 AC, DC current measurement (A) 3.9 5 Temperature measurement (°C/°F) 3.12 6 Adapter function 3.13
3.6
3.7
3.8
9
Page 10

1.2 THE FUNCTION BUTTONS

1 2 3
4 5 6
Figure 3: The Function Buttons
Item Function See §
1 Holds the last value on the display
Zero correction A
Lead resistance compensation in the continuity and ohmmeter functions
2 Selects the type of measurement and configuration
functions (AC, DC) 3 Enables/disables display backlighting 2.3 4 Enables/disables the MAX/MIN mode
Enables/disables the True InRush 5 Performs Frequency measurements (Hz) 2.5 6 Activation of ΔREL mode – Displays differential and
relative values
DC
®
mode
2.1
3.9.2
3.6.1
2.2
2.4
2.6
10
Page 11

1.3 THE DISPLAY

4
6
1
2
3
7
5
Figure 4: The Display
Item Function See §
1 Mode selection display 2 2 Active measurement value and unit display 3.5 to 3.12 3 Display of the MAX/MIN modes 2.4 4 Type of measurement (AC or DC) 2.2 5 Selected resistance measurement display 1.1 6 Low battery indication 5.3 7 Temperature unit display 3.4.4

1.3.1 Display Symbols

Symbol Designation
AC Alternating current or voltage DC Direct current or voltage
REL
Ref
Max Maximum DC or RMS value
Min Minimum DC or RMS value
V Volt
Relative value, with respect to a reference Reference value Storage of the values and display hold
11
Page 12
Hz Hertz
A Ampere
°C/°F Temperature unit Celsius or Fahrenheit
% Percentage
Ω Ohm
m Milli- prefix
k Kilo- prefix
Lead resistance compensation Continuity test Diode test Auto Power Off disabled Low battery indicator

1.3.2 Measurement Capacity Exceeded (OL)

The OL (Over Load) symbol is displayed when the display capacity is exceeded.

1.4 THE TERMINALS

The terminals are used as follows:
1 2
Figure 5: The Terminals
Item Function
1
COM (black) Input Terminal Jack
2 + Positive (red) Input Terminal Jack
12
Page 13
This mode operates if the MAX/MIN or

2 THE BUTTONS

The buttons respond differently to short, long, and sustained presses. The , and keys provide additional functions as well as the detection and acquisition of parameters complementary to the basic measurements.
Each of these buttons can be used independently of the others or in conjunction with each other. T his makes navigation simple and intuit ive when reviewing measurement results.
It is possible, for example, to either look up in turn s equence the MAX, MIN, etc. values of the RMS voltag e alone, then display relat ive values in parallel
In this section, the icon represents the possible positions of the s witc h for the button’s functionality.
2.1

BUTTON

This button is used to:
Store and look up the last values acquired specif ic to each function (V, A,
, T°, Adp) according to the spec ific m od es prev ious l y ac ti v ated (M A X/MIN, Hz, ΔREL). The present display is then maintain ed while the detection and acquisition of new values continues.
Perform automatic lead resistance compensation (see § 3.6.1). Perform automatic zero correction in A
Successive presses on
long
(> 2 sec)
long
(> 2 sec)
First press: Holds the display of the last value displayed
Second press: Returns to normal display mode (the value of each new measurement is displayed)
ADC Performs automatic zero correction (see
NOTE:
HOLD modes (short press) are first de activated
Performs automatic lead resistance compensation (see 3.6.1)
DC (see § 3.9.2).
Function
3.9.2)
See § 2.4.2 and § 2.5.2 for the button functionality in combination with the
and buttons.
13
Page 14

2.2 (YELLOW) BUTTON (SECOND FUNCTION)

This button is used to select the t ype of measurement (AC , DC) and the secon d functions marked in yellow next to the r el evant positions of the switch.
It can also be used to modify the default values in the configuration mode (see § 3.4).
NOTE: This button is invalid in the MAX/MIN, HOLD and ΔREL modes.
Successive
presses on
- Selects AC or DC. Depending on your choice, the screen displays AC or DC
- C
ycles through the continuity , and diode test modes and returns to the continuity test
elects °C or °F as the temperature unit
- S
Func
tion

2.3 BUTTON

Successive
presses on
- Enables/disables the backlighting of the display
NOTE: The backlighting is automatically disabled at the end of 2 minutes.
Func
tion
14
Page 15

2.4 BUTTON

he MAX and MIN symbols are
If MIN has been selected, MIN
f the HOLD function is activated,
mode. The HOLD function must first be

2.4.1 Normal Mode

This button activates the detection of the MAX and MIN values of the measurements mad e. Max and Min are the extreme mean values in DC and the extreme RMS values in AC.
NOTE: In this mode, the Auto Po wer Off function of the device is automat ically
disabled. The symbol is displayed on the screen.
Successive presses on
First press: Activates detection of the
MAX/MIN values Second press: Displays the MAX or MIN
value successively Third press: Returns to the display of the
short
long (> 2 sec)
OTE : ΔREL function can be used with the functions of the MAX/MIN mode.
N
present measurement without exiting from the mode (the values already detected are not erased)
NOTE: T both displayed, but only the symbol of the measurement selected blinks.
Example: blinks and MAX is lit steadily.
- Exits the MAX/MIN mode. The values previously recorded are then erased.
NOTE: I
it is not possible to exit from the MAX/MIN disabled first.
Func
tion
15
Page 16

2.4.2 The MAX/MIN Mode + Activation of the HOLD Mode

detected.
Successive presses on
short
- Di
splays the MAX/MIN values detected
before the button was pressed.
Func
tion
- When the button is pressed, the last value is held on the display.
NOTE: The HOLD function does not interrupt the acquisition of new MAX, MIN values

2.4.3 Access to the True Inrush® Mode ( set switch to )

This button allows measurement of t he True Inrush® current (starting current, or overcurrent in steady-state operation) for AC or DC current.
Successive presses on
l
ong (>2 sec)
- F
irst press: Enters the True InRush
mode
- "Inrh" is displayed for 3s (the
backlighting blinks)
- The triggering threshold is displayed for
5s (the backlighting is steady)
- "------" is displayed and the "A" symbol
flashes (backlighting turns off)
- After detection and acquisition, the
InRush current measurement is displayed, after the calculations stage "--
----" (backlighting off)
NOTE: The A symbol flashes to indicate
"surveillance" of the signal.
Func
tion
®
short (<2 sec)
Note: A short press
is functional only if a True
InRush value has been
- Second press: Exits the True InRush®
mode (returns to simple current measurement).
- Displays the PEAK+ value of the current
- Displays the PEAK- value of the current
- Displays the RMS True InRush
®
current
NOTE: The A, AC and PEAK values flash
during this sequence.
16
Page 17
2.5
This button is used to display the f requency measurements of a signal.
NOTE: This button is not functional in the DC mode.

BUTTON

2.5.1 Normal Mode

Successive
presses on
Displays:
- The frequency of the signal measured
- The present voltage (V) or current (A) measurement
Function

2.5.2 The Hz Function + Activation of the HOLD Mode

presses on
Successive
- Holds the last frequency reading
- Successively displays the last held frequency, then the voltage or the current
- NOTE: Pressing the button a second time returns to realtime measurement updates.
Function
17
Page 18

2.6 BUTTON

This button is used to display and store the reference value in the unit of magnitude measured, or to display the differential and relative values, in %.
Successive
presses on
Enters the ΔREL mode, to store, then display the reference value. The ΔRef symbol is displayed.
- Displays the differential value: (current value – reference (∆))
short
long (>2 sec) - Exits from the ΔREL mode
NOTE: The “Relative mode
The ΔREL symbol is displayed.
- Displays the relative value in % (current value – reference (∆)) The ΔREL and % symbols are displayed.
- Displays the reference. The ΔRef symbol is displayed
- Displays the current value. The ΔRef symbol blinks.
ΔREL” function can also be used with the functions of
the MAX/MIN mode.
Function
18
Page 19

3 USE

3.1 INSTALLING THE BATTERIES

Insert the batteries supplied with the device as follows:
1. Using a screwdriver, unscrew the battery compartment cover (item 1) from the back of the housing.
2. Ins ert the 4x1.5V AA batteries supplied (item 2), observing polarities.
3. Close the battery compartment cover and sc rew it onto the housing.
2
1
Figure 6 : The Battery Compartmen t

3.2 TURNING THE CLAMP-ON METER ON

With the rotary switch set in the OFF position, turn the switch to the desired function. The display lights (all symbols) f or a few seconds (see
§1.3), then the screen of the function chosen is displayed.
The clamp-on meter is now ready to make measurements .

3.3 TURNING THE CLAMP-ON METER OFF

The clamp-on meter can be turned off in two ways:
Manually - Turn the switch to the OFF position.
Automatically - After ten minu tes with no act ivit y, the inst rument will tur n
OFF. Thirty (30) seconds before the dev ice is switched off, an audible signal sounds intermittently. To r e-activate the device, pres s any button or turn the rotary switch.
19
Page 20

3.4 CONFIGURATION

As a safety measure, and to avoid repeated overloads on the inputs of the device, configuration operations should only be performed when the device is disconnected from all dangerous voltages.

3.4.1 Configuring the Maximum Resistance for Continuity

To configure the maximum resistance allowed for a continuity:
1. Wit h the switch in the OFF position, hold the (yello w) button down while turning the switch to until the "full screen" displa y ends and a beep is emitted. The display will indicate the value below which the buzzer is activated and the symbol is displayed. The value stored by default is 40. The possible values range between 1and 999Ω.
2. T o change the thresho ld, press the (yellow) button. T he right-hand digit flashes; each press on the (yellow) button inc rements it. To shift to the next digit, apply a long press (>2s) to the (yellow) button.
When the desired value is displayed, turn the switch to another setting. The detection threshold chosen is s tored and a double beep is emitted.

3.4.2 Auto Power OFF

The Auto Power OFF feature is enabled by default. To disable it, perform the following:
1. In t he OFF position, hold the button down while turning t he switch to until the "full screen" display ends and a b eep is emit ted. T he symbol is displayed.
2. When t he button is r eleased, the dev ice is in the v oltmeter f unction in the normal mode.
3. T o return to Auto Power OFF, turn the clamp-on meter OFF and then back ON again.

3.4.3 Configuring the Current Threshold for True InRush® Measurement

To configure the triggering cur rent threshold of the True InRush
1. In the OFF position, hold the button do wn while turning the s witch to until the "full screen" display ends and a beep is emitted. The display will indicate the perc entage overshoot to app ly to the measured current to determine the measurement triggering threshold.
The value stored by default is 10%, representing 110% of the established current measured. The possible values are 5%, 10%, 20%, 50%, 70%, 100%, 150%, and 200%.
20
®
measurement:
Page 21
2. To change the threshold, press the (yellow) button. The value flashes; each press on the (yellow) button displays the ne xt value. To record the chosen threshold, apply a long press (>2s) on the
(yellow) button. A confirmation beep is emitted.
When the desired value is displayed, turn the switch to another setting. The chosen threshold is stored and a double beep is emitted.
NOTE: The starting (InRush) c urrent me asurement triggering threshol d is f ixe d a t 1% of the least sensitive range. T his value i s 1% of 99. 99A or 1A. This thr eshold is not adjustable.

3.4.4 Changing the Default Temperature Unit

To change the measurement unit, ° C or °F:
1. In t he OFF position, hold the (yello w) button down while turning
the switch to until the "full screen" display ends and a beep is emitted. The display will indicate the currently assi gned unit (°C or °F). The default unit is °C.
2. Pressing the (yellow) button toggles between °C and °F.
When the desired unit is displ ayed, turn the switch to another setting. The unit chosen is stored and a double beep is emitted.
NOTE: Pressing the (yellow) button during an active temperature measurement will toggle between °C and °F.
3

.4.5 Configuring the Adapter Function Scale Factor

To configure the Adapter function sc ale factor :
1. From the OFF position, hold the (yellow) button down while
turning the switch to until the "full screen" display ends and a beep is emitted, to enter configuration mode. The display unit indicates the stored scale factor value. The default stored value is 1. The possible values are, in order: 1, 10k, 100k, 100m, 10m, 1m, 100, 10. (see §3.13)
2. To change the value of the scale factor , pres s the ( yellow) button.
The currently active scale factor is displayed. Each press of the (yellow) button displays the next value in the list above.
Once the scale factor displayed has been chosen, turn the switch to another position. The value chosen is stored and a double beep is emitted.
21
Page 22

3.4.6 Default Configuration

To reset the clamp-on meter to its default parameters (factory configuration):
1. In t he OF F posit ion, h old t he (yellow) butt on down while turning the switch to , until the "full screen" display ends and a beep i s emitted. The "rSt" symbol is displayed.
2. Af ter 2 s, the clamp-on meter emits a double beep, then all of the digital symbols of the screen are displayed until the (yellow) button is released. The default paramet ers are then restored:
Continuity detection threshold = 40
True InRush triggering threshold = 10%
Temperature measurement unit = °C
Adapter function scale factor = 1

3.5 VOLTAGE MEASUREMENT (V)

To measure voltage, proceed as follows:
1. Set the switch to .
2. Con nect the blac k lead to th e COM terminal and the red lea d to the "+" terminal.
3. Connect the test probes or the alligator clips to the circuit to be measured. The device selects AC or DC automatically according to which measured value is larger. The AC or DC symbol displ ays blinking in auto detect mode.
To select AC or DC manually, press the (yellow) button to toggle bet ween them. The symbol corresponding to t he c hoice will then display.
The measured value is displayed on the screen.
22
Page 23

3.6 CONTINUITY TEST

Warning: Before performing the test, make sure that the circuit is off and all capacitors have been discharged.
1. Set the switch to ; the symbol is displayed.
2. Con nect the black lead to the COM terminal and the red lead to the "+" terminal.
3. Con nect the test probes or the allig ator clips t o the circuit or component to be measured.
An audible signal is emitted if there is c ontinuity (resistance value is bel ow the maximum threshold, see § 3.4.1) and the measured value is displayed on the screen.

3.6.1 Lead Resistance Compensation

Warning: Before the compensation is executed, the MA X/MIN and HO LD m odes must be de-activated.
To perform automatic compensation of the test lead resistance, proceed as follows:
1. Short-circuit the leads connected to t he meter.
2. Hold the button down until the display unit indicates the lowest value. The device measures the resist ance of the leads.
3. Release the button. The correction and the symbol are displayed. The value displaye d is stored.
NOTE: The correction value is stored only if it is ≤ 2.
Above 2, the value displayed blink s and is not stored.
23
Page 24
3.7 RESISTANCE MEASUREMENT Ω
Warning: Before making a resistance measurement , make s ure that t he circ uit is off and all capacitors have been disch arged.
1. Set the switch to and press the (yellow) button. The Ω symbol is displayed.
2. Con nect the black lead to the COM terminal and the red lead to the "+" terminal.
3. Con nect the test probes or the allig ator clips t o the circuit or component to be measured.
The measured value is displayed on the screen. NOTE: To measure low resistance values, first perform lead resistance
compensation (see § 3.6.1).

3.8 DIODE TEST

Warning: Before perf orming the diode test, make sure that t he circuit is off and all capacitors have been discharge d.
1. Set the switch to and press the (yellow) button t wice. The symbol is displayed.
2. Con nect the black lead to the COM terminal and the red lead to the "+" terminal.
3. Connect the test probes or the alligator clips to the component to be tested.
24
Page 25
The measured value is displayed on the screen.
4. Rever se the leads on the diode and repeat the test.

3.9 CURRENT MEASUREMENT (A)

The jaws are opened by pressing the tr igger on t he body of the meter. The arro w on the jaws of the clamp-on met er (see the diagram below) should point in the presumed direction of current flow, from the generator to the load. Make sure that the jaws have closed correctly after clamping around the conductor.
NOTE: The measurement results are opt imal when the conductor is cent ered in the jaws (aligned with the centeri ng m arks).
The device selects AC or DC automat ic all y a cc ordi ng to which measured value is larger. The AC or DC symbol displays blinking in auto detect mode.

3.9.1 AC Measurement

For an AC current measurement, proceed as follows:
1. Set the switch to and select AC by pressing the (yellow button). The AC symbol is displayed.
2. Clamp the jaws around the conductor to be measured. The device selects AC or DC automatically.
The measured value is displayed on the screen.
25
Page 26

3.9.2 DC Measurement

Set the switch to and select DC if the display does not indicat e "0"; the DC zero must be corrected first.
Step 1: Correction of DC Zero
Important: The clamp must not be closed on the conductor during the DC zero correction. Hold the clamp in t he same position during the whole procedure s o that the correction value will be exact.
Press the button until t he device emits a double beep and disp lays a value near "0". The correction value is stored until the clamp is powered do wn.
NOTE: The correction is eff ected only if the value disp layed is < ±10A , otherwis e the value displayed blinks an d is not st or ed. The clamp must be recalibrat ed (s e e
§ 5.3)
Step 2 : Make a Measurement
1. T he switch is set to . Select DC by pressing the (yellow) button until the desired choice is reached.
2. Clamp the jaws around the conductor to be measured.
The measured value is displayed on the screen.
26
Page 27

3.10 STARTING CURRENT OR OVERCURRENT (True InRush®) MEASUREMENT

To measure a starting current or overcurrent, proceed as foll ows:
1. Set the switch to , correct the DC zero (see §3.9.2), then clamp the jaws around the conductor to be measured.
2. Perform a long press on the button. The InRh symbol is displayed, along with the triggering threshold. The clamp then awaits detection of the True InRush® current. "------" is displayed and the A symbol flashes.
3. Af ter detection and acquisition for 100 ms, the RMS value of the True
®
InRush
current is dis played. Pressing the button will display the
PEAK+/PEAK- values subsequently.
4. A long press on the button o r a change of function on the rotary switch will exit the True InRush
®
mode.
NOTE: The triggering threshold in A is 20A if the initial current is zero (starting of installation). For an established current (overload in an installation) see §3.4.3.

3.11 FREQUENCY MEASUREMENT (HZ)

The frequency measurement is avail able in V and A for AC measureme nts. The measurement is based on a count of zero c rossings (positive-going edges).

3.11.1 Frequency Measurement (V)

To measure the frequency in voltage, proceed as follows:
1. Set the switch to and press the button. The Hz symbol is displayed.
2. Select AC by pressing the (yellow) button u ntil the desired choice is reached.
3. Con nect the black lead to the COM terminal and the red lead to the "+" terminal.
4. Connect the test probes or the alligator clips to the circuit to be measured.
27
Page 28
The measured value is displayed on the screen.

3.11.2 Frequency Measurement (A)

1. Set the switch to and press the button. The Hz symbol is displayed.
2. Select AC by pressing the ( yellow) button until desired choice is reached.
3. Clamp the jaws around the conductor to be m easured.
The measured value is displayed on the screen.
28
Page 29

3.12 TEMPERATURE MEASUREMENT

3.12.1 Measurement without External Sensor

1. Set the switch to .
The temperature (blinking) dis played is the internal temperatur e of the device. It will be equal to the ambient temperature after a sufficiently long thermal stabilization time (at least one hour).

3.12.2 Measurement with External Sensor

The device measures the temperature using a K-thermocouple.
1. Con nect the K-thermocouple t o the COM and “+” input terminals of t he device observing the red and black banana plug polarity.
2. Set the switch to .
3. Place the K-thermocouple on the element or environment to be measured. It must not be at a dangerous v ol tage.
The temperature will be displayed on the screen. To change the unit to either °F or °C, pr ess the (yellow) button.
NOTE:
- If the external sensor is defective, the tem perature displayed blinks.
- If there are lar ge variations of the initial temperatur e environment of the meter and the measurement environment, the measurement must be preceded by a stabilization time.
29
Page 30

3.13 ADAPTER F UNCTION MEASUREMENT

This function makes it possible t o connect any ada pter or sensor that c onverts a measured electrical or physical quantity into a voltage (eith er D C or AC). A direct, immediate reading is obtained without applying a conversion factor.
The measurement mode, AC or DC (the d ef ault ), must be chos e n man ual ly usin g the yellow button.
The scale factor for the adapter or sensor must be chosen in advance in the instrument’s set-up (§3.4.5). The table below indicates the various adapter/sensor sensitivities th at allow a direct reading once the scale fac tor has been chosen.
Adapter/Sensor
Sensitivity
(examples in Amps)
10 mV/kA (0.01 mV/A) 10 k
100 mV/kA (0.1 mV/A) 100 k
1 mV/A 1
10 mV/A 10
100 mV/A 100
1000 mV/A (1 mV/mA) 1 m
10 mV/mA 10 m
100 mV/mA 100 m
Scale Factor to be
Programmed
NOTE: The examples given in Amps are valid for other quantities: humidity (%RH), illumination (lux), speed (m/s), flow, (GPM) etc.
1. Connect the black lead to the COM terminal and the red lead to the "+" terminal.
2. Set the switch to .
3. Select the AC or DC mode.
4. Con nect the adapter according to its directions for use.
30
Page 31

4 SPECIFICATIONS

Measurement range
0.00 to 99.99V
100.0 to 999.9V
1000V (1)
Specified measurement range
0.00 to 9.99V
±(1% R +3cts)
Resolution
0.01V
0.1V
1V
Input impedance
10MΩ

4.1 REFERENCE CONDITIONS

Quantities of Influence Reference Conditions
Temperature: 23°C ±2°C Relative humidity: 45% to 75% Supply voltage: 6.0V ±0.5V Frequency range of the applied signal: 45 to 65Hz Sine wave: pure Peak factor of the applied alternating signal: √2 Position of the conductor in the clamp: centered Adjacent conductors: none Alternating magnetic field: none Electric field: none

4.2 SPECIFICATIONS UNDER THE REFERENCE CONDITIONS

Accuracy is expressed in ± (x% of the reading (R) + y counts (ct)).

4.2.1 DC Voltage Measurement

Accuracy
Note (1) The display indicates "+OL" above + 2000V and "-OL"
0 to 100% of the measurement range
±(1% R + 10cts)
10.00 to 99.99V
below – 2000V, in REL mode. The "-" and "+" signs are manag ed. Above 1000V, a repetitive beep indicates that the voltage being measured is greater than the safety voltage for which the device is guaranteed.
31
±(1% R +3cts)
Page 32

4.2.2 AC Voltage Measurement

Measurement range
1000V RMS
1400V peak (1)
Specified range (2)
0.15 to 9.99V
± (1% R +3cts)
Resolution
0.01V
0.1V
1V
Input impedance
10M
Measurement Range (2)
0.00A to
99.99A
100.0A to
999.9A
1000A
to 3000A (1)
Specified measurement range
2000A
± (3.5% R + 3cts)
Resolution
0.01A
0.1A
1A
0.15 to 99.99V 100.0 to 999.9V
measurement
Accuracy
Note (1) Above 1000V (RMS), a repetitive beep indicates that the voltage
being measured is greater than the safety voltage for which the device is guaranteed.
- Bandwidth in AC = 3 kHz
Note (2) Any value between zero and the min. t hreshold of the measurement
range (0.15V) is forced to show "----" on the display.
Specific Specifications in MAX/MIN mode (from 10Hz to 1kHz, and from
0.30V in AC):
± (1% R + 10cts)
10.00 to 99.99V
0 to 100% of the measurement range
± (1% R +3cts)
Accuracy: add 1% R to the values of the table above.
Capture of the extreme: approximately 100ms.

4.2.3 DC Current Measurement

0 to 100% of the measurement range
Accuracy (2) (zero corrected)
Note (1) The d isplay indicates “+OL” above 6000A and “-OL” below -6000A
Note (2) The r esidual current at zero depends on the remanence. It can be
± (1.5% R +3cts)
± (1% R +
10cts)
in REL mode. The "-" and "+" signs are displayed.
corrected by the “DC zero” function of the HOLD button.
32
± (1% R
+3cts)
2000 to 2500A
± (2.5% R + 3cts)
2500 to 3000ADC
DC
Page 33

4.2.4 AC Current Measurement

Measurement Range
Specified Range (2)
1000 to 1500A
± (2% R +5cts)
Resolution
0.01A
0.1A
1A
Measurement Range
20 to 2000AAC
20 to 3000ADC
Specified Measurement Range
0 to 100% of the measurement range
Accuracy
± (5% R + 5cts)
Resolution
1A
Measurement Range
0.0 to 999.9
Open-circuit Voltage
3.6V
Measurement Current
550µA
Accuracy
± (1% R +5cts)
Buzzer Triggering Threshold
Adjustable from 1 to 999Ω (40Ω is the default)
0.25 to 99.99A 100.0 t o 999.9A 1000 to 2000A (1)
Measurement
Accuracy ± (1% R + 10cts) ± (1% R +3cts)
Note (1) The display indicates “OL” above 3000A in PEAK mode.
- Bandwidth in AC = 1 kHz
Note (2) In AC, any value between zero and the min. threshold of the
measurement range (0.25A) is forced to show “----“ on the display.
Specific Specifications in MAX/MIN mode (from 10Hz to 1kHz, and from
0.30A in AC):
0 to 100% of the measurement range
± (1.5% R +3cts)
1500 to 2000A
Accuracy (with zero corrected): add 1% R to the values of the table
above.
Capture of the extreme: approximately 100ms.

4.2.5 True Inrush® Measurement

Specific Specifications in PEAK mode in True InRush
Accuracy: add ± (1.5% R +0.5A) to the values in the table above.
PEAK capture time: 1ms min. to 1.5ms max.

4.2.6 Continuity Measurement

33
®
(from 10Hz to 1kHz):
Page 34

4.2.7 Resistance Measurement

0.0 to
999.9
1000 to
9999
10.00 to
99.99k
Specified Measurement Range
1 to 100% of the
measurement range
0 to 100% of the
measurement range
Accuracy
± (1% R +5cts)
Resolution
0.1
1
10
Open-circuit Voltage
3.6V
Measurement Current
550µA
100µA
10µA
Measurement Range
0.000 to 3.199VDC
Specified Measurement Range
1 to 100% of the measurement range
Accuracy
± (1% R + 3cts)
Resolution
0.001V
Measurement Current
0.55mA
Indication: junction reversed or open-circuit
"OL" is displayed when the measured
voltage >3.199V
Measurement Range (1)
5.0 to 999.9Hz
1000 to 9999Hz
10 to 19.99kHz
1 to 100% of the
range
Accuracy
± (0.4% R + 1ct)
Resolution
0.1Hz
1Hz
10Hz
Measurement Range (1)
Note (1) Above the maximum display value, the display unit indicates "OL".
The "-" and "+" signs are not managed.
Specific Specifications in MAX/MIN mode:
Accuracy: add 1% R to the values of the table above.
Capture of the extreme: approximately 100ms.

4.2.8 Diode test

Note: The "-" sign is disabled for the diode test function.

4.2.9 Frequency Measurements

4.2.9.1 Voltage
Specified Measurement Range
measurement
34
0 to 100% of the
measurement range
Page 35
4.2.9.2 Current
Measurement Range (1)
5.0 to 999.9Hz
Specified Measurement Range
1 to 100% of the measurement range
Accuracy
± (0.4% R + 1ct)
Resolution
0.1Hz
Function
External temperature
Type of Sensor
K-thermocouple
-60.0° to +999.9°C
-76.0° to +1831.8°F
+1000° to +1200°C
+1832° to +2192°F
Specified Measurement Range
1 to 100% of the
measurement range
0 to 100% of the
measurement range
1% R ±3°C
1% R ±5.4°F
1% R ±3°C
1% R ±5.4°F
0.1°C
0.1°F
1°C 1°F
Note (1) In MAX/MIN mode, the operating range is limited to 1kHz.
If the level of the signal is too low (< 10% of the range, or V<10V or I<10A) or if the frequency is less than 5Hz, the device cannot determine the frequency and displays "----".
Specific Specifications in MAX/MIN mode (from 10Hz to 1kHz):
Accuracy: add 1% R to the values of the table above.
Capture of the extreme: approximately 100ms.

4.2.10 Temperature Measurement

Measurement Range
Accuracy (1)
Resolution
Note (1) The stated external temperature measure ment accuracy does not t ake
the accuracy of the K-thermocouple into account.
Note 2 Use of the thermal time constant (0.7min/°C):
If there is a sudden variation of the temperatur e of the clamp, by 10°C for example, the clamp will be at 99% (cnst = 5) of the final temperature after 0.7min/°Cx1 0°Cx5 = 35 mi n (the ti me const ant of the external sensor must be added to this value).
Specific Specifications in MAX/MIN mode (10Hz to 1kHz):
Accuracy: add 1% R to the values of the table above.
Capture of the extreme: approximately 100ms.
35
Page 36

4.2.11 Adapter Function Measurement

Measurement Range (1)
0.0 to 999.9mV
1.00 to 9.99V
Specified Measurement Range (2)
0 to 100% of the measurement range
Accuracy
1% R +3cts
Resolution
0.1mV
10mV
Input Impedance
10M
Measurement Range (1)
5.0 to 999.9mV
1.00 to 9.99V
Specified Measurement Range (2)
1 to 100% of the
measurement range
0 to 100% of the
measurement range
5.0 to 99.9mV
± (1% R +3cts)
Resolution
0.1mV
10mV
Input Impedance
10M
4.2.11.1 In DC mode
4.2.11.2 In AC mode
Accuracy
Note (1) The basic display is 10,000 counts. The position of the decimal count
and the display of multiples ( m and k) depend on the programming of the scale factor.
In DC, the disp lay in dic at es "+ O L" ab ove +99 99 c ount s a nd "-
OL" below -9999 counts. The "-" and "+" si gns are managed (polarity).
In AC, the display i ndicates "OL" above 9999 counts.
Note (2) - The max. bandwidth is 1kHz.
Specific specifications in MAX/MIN mode (from 10Hz to 1kHz):
± (1% R + 10cts)
100.0 to 999.9mV
1% R +3cts
Accuracy: add 1% R to the values of the table above.
Capture of the extreme: approximately 100ms.

4.3 ENVIRONMENTAL CONDITIONS

Conditions Operating Storage
Temperature -4° to +131°F
(-20° to +55°C)
Relative Humidity (RH) ≤90% up to 131°F (55°C) ≤90% up to 158°F (70°C)
-40° to +158°F
(-40° to +70°C)
36
Page 37

4.4 MECHANICAL SPECIFICATIONS

Polycarbonate
Clamping diameter: 2.36" (60mm)
LCD display unit
Dimension: 1.6 x 1.9" (41 x 48mm)
Housing
Jaws
Screen
Dimension 11.65 x 4.37 x 1.61" (296 x 111 x 41mm) Weight 1.4 lbs (640g) with batteries
Rigid polycarbonate shell with over-molded elastomer covering;
Opening: 2.36" (60mm)
Blue backlighting
UL94 V1

4.5 POWER SUPPLY

Batteries 4x1.5V AA LR6 Battery life >350 hours (without backlighting) Auto Power Off After 10 minutes with no switch and/or but ton activity

4.6 COMPLIANCE WITH INTERNATIONAL STANDARDS

Compliant with standards IEC-61010-1,
Electric Safety
IEC-61010-2-30, and IEC-61010-2-32: 1000V CAT IV.
Electromagnetic
Compatibility
Mechanical strength
Level of Protection
Compliant with standard EN-61326-1 Classification: residential environment
Free fall: 2m (in accordance with standard
IEC-68-2-32)
Housing: IP54 (per standard IEC-60529)
Jaws: IP40
37
Page 38

4.7 ENVIRONMENTAL VARIATIONS

Influence
VAC
Hz
V A
10Hz to 1kHz
400Hz to 2kHz
1% R
4% R
1% R + 1ct
5% R + 1ct
Position of
the conductor
(f≤400Hz)
Adjacent
RMS
Conductor
the clamp
Application of
the clamp
1.4 to 3.5,
1400V peak
Condition of
Influence
Temperature
Humidity 10 to 90% RH
Frequency
in the
jaws
conductor carrying a
current of
150A
DC or
enclosed by
Range of Influence
-4° to +131°F
(-20° to +55°C)
1kHz to 3kHz
10Hz to 400Hz
Any position
on the internal
perimeter of
the jaws
Conductor
touching the
external
perimeter of
the jaws
0 to 500ARMS V
Measurement
Influenced
VDC
A*
T°C
V A
A 2% R 4% R + 1ct
A 40 dB 45 dB
Typical MAX
-
0.1% R/10°C 1% R/10°C
(0.2% R +1°C)/10°C
0.1% R/10°C + 2ct
0.1% R 0.1% R + 1ct
8% R 1% R
< 1ct
0.5% R/10°C + 2ct
1.5% R/10°C + 2ct
(0.3% R +2°C)/10°C
0.1% R/10°C + 3ct
0.1% R/10°C
9% R + 1ct 1% R + 1ct
1ct
a voltage on
Peak factor
*Note in Temperature: Influenc e s pecified until 1000 A DC
0 to 1000VDC
or RMS
limited
to 1500A peak
A
A (AC) V (AC)
38
< 1ct
1% R 1% R
3% R + 1ct
3% R + 1ct
Page 39

5 MAINTENANCE

5.1 WARNING

Remove the test leads on any input before opening the case.
Do not operate the clamp-on meter without a battery case cov er.
To avoid electrical shock, do not attempt to perform any servicing
unless you are qualified to do so.
To avoid electrical shock and/or damage to the instrument , do not get water or other foreign agents into the probe.

5.2 CLEANING

Disconnect everything connected to the device and set the switch to
OFF. Use a soft cloth moistened with soapy water. Rinse with a damp cloth
and dry quickly using a dry cloth or for ced air. Dry completely before putting bac k into use.

.3 REPLACEMENT OF THE BATTERIES

5
The symbol indicates th at the batt eries are low. Whe n this s ymbol appea rs on the display unit, the batteries must be replaced. The measurements and specifications are no longer guaranteed.
To replace the batteries, proceed as f ollows:
1. Disc onnect the measurement leads from t he i nput terminals.
2. Set the switch to OFF.
3. Using a screwdriver, unscrew the battery compartment cover from the back of the housing (see §3.1).
4. Remov e the used batteries and rep lace them with 4x1.5V AA batteries, observing the polarities.
5. Close the battery compartment cover and screw it onto the housing.
39
Page 40

6 REP AIR AND CALIBRA TION

To ensure that your instrument meets factory specifications, we recommend that it be submitted to our factory Service Center at one-year intervals for recalibration, or as required by other standards or internal procedures.
For instrument repair and calibration:
You must contact our Service Center for a Customer Service Authorization number (CSA#). This will ensure that when your instrument arrives, it will be tracked and processed promptly. Please write the CSA# on the outside of the shipping container. If the instrument is returned for calibration, we need to know if you want a standard calibration, or a calibration traceable to N.I.S.T. (includes calibration certificate plus recorded calibration data).
Chauvin Arnoux 15
Faraday Drive
Dover, NH 03820 USA Tel: (800) 945-2362 (Ext. 360)
(603) 749-6434 (Ext. 360)
Fax: (603) 742-2346 or (603) 749-6309 repair@aemc.com
(Or contact your authorized distributor) Costs for repair, standard calibration, and calibration traceable to N.I.S.T. are
available.
NOTE: All customers must obtain a CSA# before returning any instrument.
®
, Inc. d.b.a. AEMC® Instruments

7 TECHNICAL AND SALES ASSISTANCE

If you are experiencing any technical problems, or require any assistance with the proper operation or application of your instrument, please call, mail, fax or e-mail our technical support hotline:
®
C
hauvin Arnoux
Foxborough Boulevard
200 Foxborough, MA 02035, USA
Phone: (800) 343-1391
(508) 698-2115
Fax: (508) 698-2118 techsupport@aemc.com
ww.aemc.com
w
NOTE: Do not ship instruments to our Foxborough, MA address.
, Inc. d.b.a. AEMC® Instruments
40
Page 41

8 LIMITED WARRANTY

The Model 603 is warranted to the owner for a period of three years from the date of original purchase against defects in manufacture. This limited warranty is given by AEMC
®
Instruments, not by the distributor from whom it was purchased. This warranty is void if the unit has been tampered with, abused or if the defect is related to service not performed by AEMC
®
Instruments.
Full warranty coverage and product registration is available on our website at www.aemc.com/warranty.html
.
Please print the online Warranty Coverage Information for your records.
If a malfunction occurs within the three-year period, you may return the instrument to us for repair, provided we have your warranty registration information on file or a proof of purchase. AEMC option, repair or replace the faulty material.
®
Instruments w ill, at its
REGISTER ONLINE AT: www.aemc.com

9 WARRANTY REPAIRS

What you must do to return an Instrument for Warranty Repair: First, request a Customer Service Authorization Number (CSA#) by
phone or by fax from our Service Department (see address below), then return the instrument along with the signed CSA Form. Please write the CSA# on the outside of the shipping container. Return the instrument, postage or shipment pre-paid to:
®
hauvin Arnoux
C 15 Faraday Drive • Dover, NH 03820 USA
Tel: (800) 945-2362 (Ext. 360)
(603) 749-6434 (Ext. 360)
Fax: (603) 742-2346 or (603) 749-6309 repair@aemc.com
, Inc. d.b.a. AEMC® Instruments
Caution: To protect yourself against in-transit loss, we recommend you insure your returned material.
NOTE: All customers must obtain a CSA# before returning any instrument.
41
Page 42
NOTES:
42
Page 43
Page 44
Chauvin Arnoux®, Inc. d.b.a AEMC® Instruments
99-MAN 100369 v6 06/17
15 Faraday Drive • Dover, NH 03820 USA
www.aemc.com
Loading...