AEMC Instruments 6000-36-2-0.1 User manual

AmpFlex® Flexible AC Current Probes
USER MANUAL
www.GlobalTestSupply.com
www.GlobalTestSupply.com
Table of Contents
Warning ..................................................................................................... 2
International Electrical Symbols ................................................................ 2
Definition of Measurement Categories ...................................................... 3
Receiving Your Shipment .......................................................................... 3
Packaging .................................................................................................. 3
Description ................................................................................................ 4
Features .................................................................................................... 6
Accessories ............................................................................................... 6
Standard Models ....................................................................................... 7
Specifications ............................................................................................ 8
Instrument Compatibility .......................................................................... 12
Operation ................................................................................................. 12
Making Measurements with the AmpFlex® .......................................... 12
Tips for Making Precise Measurements: ............................................. 12
Typical Response Curves:................................................................... 16
Maintenance ............................................................................................ 18
Warning ............................................................................................... 18
Battery Replacement ........................................................................... 18
Cleaning............................................................................................... 18
Repair and Calibration ............................................................................ 19
Technical and Sales Assistance ............................................................. 19
www.GlobalTestSupply.com
AmpFlex® - Flexible AC Current Probes
Warning
These safety warnings are provided to ensure the safety of personnel and proper operation of the instrument.
Read the instruction manual completely and follow all the safety information before attempting to use or service this instrument.
Wear protective clothing and gloves as required.
Use caution on any circuit: potentially high voltages and
currents may be present and may pose a shock hazard.
Read the safety specifications section prior to using the current probe. Never exceed the max voltage ratings given.
Safety is the responsibility of the operator. The AmpFlex must be used only by qualified personnel using applicable safety precautions.
®
ALWAYS de-energize the circuit before wrapping the AmpFlex
®
around bare conductors, bus bars, or near live
parts. Do not wrap on live conductors.
ALWAYS connect the electronic module to the display device before wrapping the AmpFlex
®
around the sample
being tested.
ALWAYS inspect the module, sensor, sensor cable, and output terminals prior to use. Replace any defective parts immediately. Use only factory parts.
®
NEVER use the AmpFlex
on electrical conductors rated
above 600V CAT IV; 1000V CAT III.

International Electrical Symbols

This symbol signifies that the current probe is protected by double or reinforced insulation. Use only factory specified replacement parts when servicing the instrument.
This symbol signifies CAUTION! and requests that the user refer to the user manual before using the instrument.
This symbol refers to a type B current sensor. Do not apply around or remove from HAZARDOUS LIVE conductors without additional protective means (de-energizing the circuit or wearing protective clothing suitable for high voltage work).
- 2 -
www.GlobalTestSupply.com
AmpFlex® - Flexible AC Current Probes

Definition of Measurement Categories

CAT IV: For measurements performed at the primary electrical supply
(<1000V) such as on primary overcurrent protection devices, ripple control units, or meters.
CAT III: For measurements performed in the building installation at the
distribution level such as on hardwired equipment in fixed installation and circuit breakers.
CAT II: For measurements performed on circuits directly connected to
the electrical distribution system. Examples are measurements on household appliances or portable tools.

Receiving Your Shipment

Upon receiving your shipment, make sure that the contents are consistent with the packing list. Notify your distributor of any missing items. If the equipment appears to be damaged, file a claim immediately with the carrier and notify your distributor at once, giving a detailed description of any damage.

Packaging

Your AmpFlex® consists of the following items:
Flexible probe with electronic module
User manual
9V battery
- 3 -
www.GlobalTestSupply.com
AmpFlex® - Flexible AC Current Probes

Description

The AmpFlex® is a flexible AC current transformer composed of a flexible sensor and an electronic module. The flexible sensor permits measurements on conductors where standard clamp-on probes could not be used. In particular, it can be installed in tight spaces, around breaker panels, around cable bundles, around wide or large bus bars, or even wrapped around irregular shapes. The Shape Memory feature enables the user to “pre-shape” the sensor before inserting it between or around conductors. This feature facilitates closing, enhances user safety, and alleviates the drooping effect typically associated with flexible sensors.
®
The AmpFlex standard transformers. The transformation principle is based on an air core. It presents virtually no load to the system under test, has a low phase shift and excellent frequency response, and cannot be damaged by overloads.
The sensor assembly is waterproof and insulated for 600V CAT IV; 1000V CAT III. The AmpFlex designed with materials and components to meet international agency approvals.
The AmpFlex for direct readings on DMMs, data loggers, oscilloscopes, and power or harmonic meters. TRMS measurements are taken when connected to a TRMS meter. The AmpFlex AC component of the measured signal is measured.
The length of the flexible sensor can be selected in lengths of 24", 36", and 48" in length. Consult the factory for custom lengths, ranges and/or features.
is lightweight and does not use magnetic cores like
®
meets EN 61010-1, is CE marked, and is
®
has an mV output proportional to the current measured
®
is insensitive to DC currents and only the
- 4 -
www.GlobalTestSupply.com
1. Positive: Red Banana Plug (+)
Figure 1
AmpFlex®
AmpFlex® - Flexible AC Current Probes
2. Common: Black Banana Plug (-)
3. Power ON Indicator (Green LED)
4. Overload Indicator (Red LED)
5. Range Selection Switch
6. AmpFlex
®
Connector/Latch
7. Flexible Sensor (Diameter 0.5", 12.5mm)
8. Lead from Sensor to Module (6.5 ft, 2m)
9. Electronic Module - Descriptive Label on Back
Case (range, model, etc.)
- 5 -
www.GlobalTestSupply.com
AmpFlex® - Flexible AC Current Probes
Features
Models to measure from 0.5Arms to 30,000Arms
Accuracy 1% of Reading
TRMS measurements when connected to a TRMS instrument
No core saturation or damage if overloaded
Overrange LED for measurement circuitry
600V CAT IV; 1000V CAT III; EN 61010; CE Mark
Waterproof sensor
9V battery for typical 150 hour continuous operation
Shape Memory for custom pre-shaping of sensor before use
(no drooping)
Very high frequency response
Low phase shift for power
Insensitive to DC, measures only AC component on DC + AC signals
Excellent linearity
Lightweight

Accessories

Banana (Female) / BNC (Male) Adapter ..................... Cat. # 2118.46
®
For connection of the AmpFlex BNC terminals on scopes and other displaying instruments.
to SLII Models L101, L102, L562,
- 6 -
www.GlobalTestSupply.com
AmpFlex® - Flexible AC Current Probes
300-24-2-10
30/300A
100/10mV/A
24” (610mm)
8” (190mm)
2112.88
1000-24-1-1
1000A
1mV/A
24” (610mm)
8” (190mm)
2112.39
1000-24-2-1
100/1000A
10/1mV/A
24” (610mm)
8” (190mm)
2112.98
1000-36-2-1
100/1000A
10/1mV/A
36” (910mm)
11” (290mm)
2113.00
3000-24-1-1
3000A
1mV/A
24” (610mm)
8” (190mm)
2112.46
3000-36-1-1
3000A
1mV/A
36” (910mm)
11” (290mm)
2112.48
3000-24-2-1
300/3000A
10/1mV/A
24” (610mm)
8” (190mm)
2113.05
3000-36-2-1
300/3000A
10/1mV/A
36” (910mm)
11” (290mm)
2112.00
3000-48-2-1
300/3000A
10/1mV/A
48” (1220mm)
15” (390mm)
2112.01
6000-36-2-0.1
600/6000A
1/0.1mV/A
36” (910mm)
11” (290mm)
2113.21
30000-24-2-0.1
3000/30,000A
1/0.1mV/A
24” (610mm)
8” (190mm)
2113.33

Standard Models

Model
Measurement
Range
Output
Sensor Length
All of our AmpFlex® models are designed to be used with recorders or power analyzers with a voltage AC input. A scale factor may need to be entered into the recorder to display the exact value.
Custom lengths and ranges are available. Contact us:
Max
Conductor
Size
Cat. #
- 7 -
www.GlobalTestSupply.com
AmpFlex® - Flexible AC Current Probes
Model
300A
30/300A
1000A
100/1000A
Measurement
Range
Accuracy
±1% +200mA
±1% +200mA
±1% +200mA
±1% +200mA
100mV/A or
10mV/A
10mV/A or
1mV/A
Crest Factor
(mid-range)
Phase Shift
@ 50/60Hz
Residual Noise
200mA
200mA
200mA
200mA
Model #
3000A
300/3000A
3000/30000A
600/6000A
Measurement
Range
Accuracy
±1% +200mA
±1% +200mA
±1% +1A
±1% +500mA
10mV/A or
1mV/A
1mV/A or
0.1mV/A
1mV/A or
0.1mV/A
Crest Factor
(mid-range)
Phase Shift
@ 50/60Hz
Residual Noise
200mA
200mA
1A
500mA

Specifications

5 to 300A 5 to 300A 5 to 1000A 5 to 1000A
Output Signal
10mV/A
3.0 3.0 9.0 9.0
1.5° 1.3° 0.7° 0.7°
1mV/A
5 to 3000A 5 to 3000A 5-30000A 5 to 6000A
Output Signal
1mV/A
3.0 3.0 3.0 > 10
0.7° 0.7° 0.5° 0.5°
*Reference Conditions: 25°C ± 5°K, 20 to 75% RH, 1 minute warm-up, battery at 9V ± 0.5V, conductor center, external DC magnetic field < 40A/m, no external AC magnetic field, no external electrical field, 10 to 100Hz, sine wave. See accuracy curves for low currents.
- 8 -
www.GlobalTestSupply.com
AmpFlex® - Flexible AC Current Probes
Common Specifications
ELECTRICAL
Accuracy: 1% of reading ± residual noise
Frequency Range: 10 - 20,000Hz with current derating
Signal Output: 4.5V max
Working Voltage: 600V CAT IV; 1000V CAT III
Frequency Influence:
See Accuracy vs. Frequency curves on pages 16 and 17
Frequency Limitation:
See current derating curves (note no limitation on 300A Range) page 17
Influence of adjacent conductor in contact with sensor and with AC signal: 0.2% typical, 2% maximum
Influence of conductor position in sensor: 0.5% typical, 4% max
Influence of shape of sensor: Oblong shape: 0.2% typical, 1% max
Common Mode Rejection: 100dB typical, 80dB min
ENVIRONMENTAL SPECIFICATIONS
Operating Temperature Range: -14° to 131°F (-10° to +55°C)
Storage Temperature Range: -40° to 158°F (-40° to +70°C)
Influence of Temperature:
Sensor: -14° to 194°F (-10° to 90°C); 0.15% per 18°F (10°C) typical,
0.5% per 18°F 10°C max Module: -14° to 131°F (-10° to 55°C); 0.15% per 18°F (10°C) typical,
0.5% per 18°F (10°C) max
Influence of Relative Humidity:
10 to 90% RH: 0.2% typical, 0.5% maximum
Operating Relative Humidity:
50° to 86°F (10° to 30°C); 85° ± 5% RH (without condensation) 104° to 122°F (40° to 50°C); 45° ± 5% RH (without condensation)
Altitude:
Operating: 0 to 2000m, working voltage derating above Non-operating: 0 to 12,000m
- 9 -
www.GlobalTestSupply.com
AmpFlex® - Flexible AC Current Probes
MECHANICAL SPECIFICATIONS
Module Output: Two 4 mm safety banana jacks
Standard 3/4" (19mm) spacing
Battery: 9V Alkaline (NEDA 1604A, IEC 6LR61) recommended
Battery Life: Useable from 9V to 7V, 150 hours typical (continuous use)
Low Battery: Power ON indicator (green LED) when battery voltage
7V, LED blinks when battery voltage is low
Overload Indication: Red LED ON indicates the selected range is overloaded. Module output may not reflect the actual measurement
Dimensions (sensor):
24", 36", 48" & 60" nominal (± 1"), other lengths optional
Dimensions (Electronic Module): 4.9 x 2.5 x 1.1" (124 x 64 x 28mm)
Weight:
AmpFlex AmpFlex AmpFlex AmpFlex
®
48" with battery: 0.95 lb
®
60" with battery: 1.13 lb
®
24" with battery: 0.74 lb
®
36" with battery: 0.89 lb
Connection Cable Length (sensor to module): 6.5 ft (2m)
Colors: Red sensor with dark gray connector and module, black
connection cable (sensor to module)
Drop Test: Per IEC 68-2-32
Vibration: Per IEC 68-2-6
Mechanical Shock: Per IEC 68-2-27
Weatherproofing: Module: IP40 (EN 60529)
SENSOR SPECIFICATIONS:
Weight: 10.8oz (302.4g)
Bend Radius: 0.75" (19mm) minimum
Bending Life: >10000 without performance deterioration
Waterproofing: IP 65
Resistance to Chemicals: Resistant to oils and aliphatic hydrocarbons
Diameter: 12mm ± 0.5mm
- 10 -
www.GlobalTestSupply.com
AmpFlex® - Flexible AC Current Probes
Outer Sheath Material: Polyurethane, UL94V0
Dielectric Strength: 7500V
Latch Spring Life: >10,000 maneuvers
MATERIAL SPECIFICATIONS
Module:
UL 94V2, Color dark gray, Polycarbonate
Sensor Latch:
Material: Lexan 500R, UL94V0
Cable Assembly to Sensor:
UL 94V0, 1000V rating
SAFETY SPECIFICATIONS
Electrical:
Double insulation or reinforced insulation between primary or secondary and outer case per EN 61010
600V CAT IV; 1000V CAT III Pollution Degree 2
7.50kV, 50/60Hz, dielectric between secondary and the outer case
Electromagnetic Compatibility
Immunity: meets BF EN 61326-1 Industrial environment category
- Electrostatic discharge (meets EN 61000-4-2) 8kV in air – level 3 – class B 4kV on contact – level 2 – class B
- 10V/m radiated electromagnetic field (in line with EN 61000-4-3) – class B
- Rapid transients (in line with EN 61000-4-4) 1kV – level 2 – class B
- Electric shocks (in line with EN 61000-4-5) 6kV – class B
- 11 -
www.GlobalTestSupply.com
AmpFlex® - Flexible AC Current Probes

Instrument Compatibility

The AmpFlex® is compatible with any multimeter, AC voltmeter, or other voltage measuring instrument with an input impedance greater than 1M. To achieve the best overall accuracy, use the AmpFlex
®
with a AC voltmeter having an accuracy of 0.75% or better.

Operation

Please make sure that you have already read and fully understand the WARNING section on page 2.

Making Measurements with the AmpFlex®:

Connect the electronic module to the AC Volt range of your DMM or measuring instrument. Select the appropriate module output voltage range. If the current magnitude is unknown, and if AmpFlex ranges, select the lowest mV/A output setting.
Wrap the flexible core around the conductor to be tested. If possible within range, select the higher mV/A AmpFlex
®
output range to obtain the best resolution. Do not exceed specified current range for the output. Do not use on selected range if overload LED goes on.
Read the displayed value on the DMM and divide it by the range selected (i.e. if reading = 2.59V with the 10mV/A output range, the current flowing through the probe is 2590mV ÷ 10 = 259A).
For best accuracy, carefully center the conductor inside the flexible core, and avoid if possible, the proximity of other conductors which may create noise and interference (particularly near the latch).
True RMS measurements are obtained when the AmpFlex connected to a True RMS meter. Note that the DC component is not measured.

Tips for Making Precise Measurements:

When using the AmpFlex® with a meter, it is important to select the range that provides the best resolution. Failure to do this may result in measurement errors.
®
For best results, select the highest AmpFlex and the most sensitive meter range for this output.
Make sure the DMM or measuring instrument can accurately measure
AC. Certain inexpensive DMM have poor resolution and accuracy
mV when measuring low mV AC.
For best accuracy, center the AmpFlex
®
measured. (See figure 2).
output signal possible
around the conductor to be
®
has two
®
is
- 12 -
www.GlobalTestSupply.com
AmpFlex® - Flexible AC Current Probes
For best accuracy center the AmpFlex®
Tips for Making Precise Measurements (continued):
To increase the sensitivity or measure on low currents, the AmpFlex® may be wrapped several times around the conductor. Remember to
divide your reading by the number of turns for the actual measurement (see Figure 4).
The overall measurement accuracy is the sum of the AmpFlex accuracy and the displaying instrument accuracy.
®
Figure 2
- 13 -
www.GlobalTestSupply.com
AmpFlex® - Flexible AC Current Probes
One turn around a conductor carrying 250AAC
Tips for Making Precise Measurements (continued):
The AmpFlex® may be doubled around the conductor to be measured to double the output. (See Figures 3 and 4 to show the different values on a DMM while measuring 250A
AC.)
Figure 3
- 14 -
www.GlobalTestSupply.com
AmpFlex® - Flexible AC Current Probes
Double the turns to double the output in low­current applications or for higher sensitivity
Tips for Making Precise Measurements (continued):
www.GlobalTestSupply.com
Figure 4
- 15 -
AmpFlex® - Flexible AC Current Probes

Typical Response Curves:

- 16 -
www.GlobalTestSupply.com
AmpFlex® - Flexible AC Current Probes

Typical Response Curves (cont.):

- 17 -
www.GlobalTestSupply.com
AmpFlex® - Flexible AC Current Probes

Maintenance

Warning

For maintenance use only specified replacement parts.
To avoid electrical shock, do not attempt to perform any servicing
unless you are qualified to do so.
®
Do not perform any service while the AmpFlex
To avoid electrical shock and/or damage to the instrument, do not get
water or other foreign agents into the electronic module.
Also see warning on page 2.

Battery Replacement

If the power ON indicator (green LED) blinks or does not light up, replace the battery.
®
Remove the AmpFlex
To replace the battery, open rear case, replace battery and
reassemble. The green LED should go on when the module is turned on.
from any circuit before replacing the battery.

Cleaning

It is important to keep the probe sensor latch mating surfaces clean and prevent foreign bodies from hampering the closing. The sensor may be gently cleaned with a soft cloth, soap and water. Dry immediately after cleaning. Avoid water penetration into the electronic module.
Make sure the sensor, electronic module, and all leads are dry before any further use.
is on any circuit.
- 18 -
www.GlobalTestSupply.com
AmpFlex® - Flexible AC Current Probes

Repair and Calibration

To ensure that your instrument meets factory specifications, we recommend that it be submitted to our factory Service Center at one-year intervals for recalibration, or as required by other standards or internal procedures.
For instrument repair and calibration:
You must contact our Service Center for a Customer Service Authorization number (CSA#). This will ensure that when your instrument arrives, it will be tracked and processed promptly. Please write the CSA# on the outside of the shipping container. If the instrument is returned for calibration, we need to know if you want a standard calibration, or a calibration traceable to N.I.S.T. (includes calibration certificate plus recorded calibration data).
(O
r contact your authorized distributor)
Costs for repair, standard calibration, and calibration traceable to N.I.S.T. are available.
NOTE: You must obtain a CSA# before returning any instrument.

Technical and Sales Assistance

If you are experiencing any technical problems, or require any assistance with the proper operation or application of your instrument, please call, fax or e-mail our technical support hotline:
www.GlobalTestSupply.com

Limited Warranty

The instrument is warranted to the owner for a period of 2 years from the date of original purchase against defects in manufacture. This limited warranty is given by AEMC was purchased. This warranty is void if the unit has been tampered with, abused or if the defect is related to service not performed by AEMC Instruments.
Full warranty coverage and product registration is available on our website at:
Please print the online Warranty Coverage Information for your records.
What AEMC
If a malfunction occurs within the warranty period, you may return the instrument to us for repair, provided we have your warranty registration information on file or a proof of purchase. AEMC option, repair or replace the faulty material.
®
Instruments will do:
®
Instruments, not by the distributor from whom it
.
®
Instruments will, at its
®

Warranty Repairs

What you must do to return an Instrument for Warranty Repair:
First, request a Customer Service Authorization Number (CSA#) by phone or by fax from our Service Department (see address below), then return the instrument along with the signed CSA Form. Please write the CSA# on the outside of the shipping container. Return the instrument, postage or shipment pre-paid to:
Ca
ution: To protect yourself against in-transit loss, we recommend you
insure your returned material.
NOTE: You must obtain a CSA# before returning any instrument.
www.GlobalTestSupply.com
NOTES
AmpFlex® - Flexible AC Current Probes
www.GlobalTestSupply.com
99-MAN 100122 v24 08/19