Advantest R6246, R6245 Datasheet

Page 1
DC Voltage/Current Generators/Calibrators
Synchronized 2-Channel Measurement Function
R6245/6246
Synchronized 2-Channel Generation, Measurement
Measurement Resolution of 1 µV, 10 fA (SMU220-2), 100 pA (SMU62-20)
Source Range ±220 V, ±2 A (SMU220-2, 2.4 W)/±62 V, ±20 A (SMU62-20, 140 W)
Minimum Pulse Width 100 µs, Minimum Step 700 µs
R6245/6246
DC Voltage/Current Sources/Monitors
The R6245/6246 Series are DC voltage/current sources and monitors having source measurement units (SMUs) with 2 isolated channels. The series covers wide source and measure­ment ranges. It is ideal for measurement of DC characteristics of items ranging from separate semiconductors such as bipolar transistors, MOSFETs and GaAsFETs, to ICs and power de­vices. Further, due to the increased measuring speed and synchronized 2-channel measurement function, device I/O characteristics can be measured with precise timing at high speed which previously difficult to accomplish. Due to features such as the trigger link function and the sequence programming function which automatically per­forms a series of evaluation tests automatically, the new R6245/ 6246 enable much more efficient evaluation tests.
(Photo is R6245)
0.02% High Measurement Resolution of 1 mV, 10 fA
High source Accuracy of ±62000 Columns, 0.03% Wide source Range of ±220 V, ±20 A
High Throughput, 0.8 V/µs (At Current
Measurement Range of 6 mA or more)
Wide Variety of Measurement Functions
• 4 Phenomenon Output (Source/Sink)
• Up to 5 V Remote Sense
• Minimum Step 700 µs Sweep Measurement
(Generation, Measurement, Data Storage)
• Range Switching without Discontinuous Output
• Pulse Measurement with Minimum Pulse Width of
100 µs
• Synchronized 2-Channel Measurement Function
• Search measurement function
Unit Configuration
Model name Channel A Channel B
R6245 SMU220-2 SMU220-2
R6246 SMU62-20 SMU220-2
SMU220-2: 24.2 W max. ±220 V/±2A SMU62-20: 200 W max. ±62 V/±20A
246
Page 2
DC Voltage/Current Generators/Calibrators
Specifications
All accuracy values are for 23˚C ±5˚C, RH 75% max., guaranteed for 6 months. One-day stability is guaranteed for 23˚C ±5˚C, RH 75% max. Measurement accuracy, one-day stability and temperature coefficient are guaranteed with auto zero on, reset time from 1 to 20 PLC.
Maximum output current:
SMU220-2 SMU62-20
±2 A up to 12 V ±20 A up to 7 V
±0.62 A up to 40 V ±10 A up to 20 V
±0.11 A up to 220 V ±2.2 A up to 62 V
Maximum output voltage, maximum follow-up voltage:
SMU220-2 SMU62-20
±220 V up to 0.11 A ±62 V up to 2.2 A
±40 V up to 0.62 A ±20 V up to 10 A
±12 V up to 2 A ±7 V up to 20 A
Voltage source/measurement range: Current source/measurement range:
Unit name
SMU62-20
SMU220-2
SMU220-2
SMU62-20
SMU62-20
Range Source range
600 mV 0 to ±620.00 mV 10 µV 0 to ±620.000 mV 1 µV
6 V 0 to ±6.2000 V 100 µV 0 to ±6.20000 V 10 µV
60 V 0 to ±62.000 V 1 mV 0 to ±62.0000 V 100 µV
200 V 0 to ±220.00 V 10 mV 0 to ±220.000 V 1 mV
6 nA 0 to ±6.2000 nA 100 fA 0 to ±6.20000 nA 10 fA
60 nA 0 to ±62.000 nA 1 pA 0 to ±62.0000 nA 100 fA
600 nA 0 to ±620.00 nA 10 pA 0 to ±620.000 nA 1 pA
6 µA 0 to ±6.2000 µA 100 pA 0 to ±6.20000 µA 10 pA
60 µA 0 to ±62.000 µA 1 nA 0 to ±62.000 µA 100 pA
600 µA 0 to ±620.00 µA 10 nA 0 to ±620.000 µA1 nA
6 mA 0 to ±6.2000 mA 100 nA 0 to ±6.20000 mA 10 nA
60 mA 0 to ±62.000 mA 1 µA 0 to ±62.0000 mA 100 nA
600 mA 0 to ±620.00 mA 10 µA 0 to ±620.000 mA 1 µA
2 A 0 to ±2.0000 A 100 µA 0 to ±2.00000 A 10 µA 6 A 0 to ±6.2000 A 100 µA 0 to ±6.20000 A 10 µA
20 A 0 to ±20.000 A 1 mA 0 to ±20.0000 A 100 µA
Setting Measurement
resolution range resolution
However, the measurement resolution at a reset time of 100 µs to 500 µs is as follows:
Reset time Measurement resolution (digits)
100 ms 10
200 ms 5
500 ms 2
Voltage/current compliance range:
Unit name
SMU62-20
SMU220-2
SMU220-2
SMU62-20
SMU62-20
*1 100 nA for SMU62-20
Range Maximum setting Minimum setting Setting resolution
600 mV 620.00 mV 1 mV 10 µV
6 V 6.2000 V 0.6201 V 100 µV
60 V 62.000 V 6.201 V 1 mV
200 V 220.00 V 62.01 V 10 mV
6 nA 6.2000 nA 10 pA 100 fA
60 nA 62.000 nA 6.201 nA 1 pA
600 nA 620.00 nA 62.01 nA 10 pA
6 µA 6.2000 µA 0.6201 µA 100 pA
60 µA 62.000 µA 6.201 µA*1 1 nA
600 µA 620.00 µA 62.01 µA 10 nA
6 mA 6.2000 mA 0.6201 mA 100 nA
60 mA 62.000 mA 6.201 mA 1 µA
600 mA 620.00 mA 62.01 mA 10 µA
2 A 2.0000 A 0.6201 A 100 µA 6 A 6.2000 A 0.6201 A 100 µA
20 A 20.000 A 6.201 A 1 mA
Measurement
Synchronized 2-Channel Measurement Function
R6245/6246
Accuracy: Includes calibration accuracy, one-day stability,
temperature coefficient, linearity
One-day stability: At constant power, load Temperature coefficient: At 0˚C to 40 ˚C, RH 75% max.
Voltage source/
compliance range
600 mV 0.03 + 400 µV 0.01 +200 µV 12 + 22 µV
6 V 0.03 + 1.3 mV 0.01 + 500 µV 12 + 55 µV
60 V 0.03 +13 mV 0.01 + 4 mV 12 + 600 µV
200 V 0.03 + 60 mV 0.01 +20 mV 12 + 2.5 mV
Current source/
compliance range
6 nA 0.5 + 6.5 pA + 60 fA 0.1 + 4 pA + 23 fA 150 + 400 fA + 1.8 fA
60 nA 0.5 + 18 pA +600 fA 0.1 + 8 pA +230 fA 120 + 1 pA + 18 fA
600 nA 0.2 + 150 pA + 6 pA 0.035 + 50 pA + 2.3 pA 45 + 6 pA + 180 fA
6 µA 0.05 + 1.5 nA + 60 pA 0.015 + 500 pA + 23 pA 18 + 60 pA + 1.8 pA
60 µA 0.04 + 15 nA + 600 pA 0.015 + 5 nA + 230 pA 18 + 600 pA + 18 pA
600 µA 0.03 + 150 nA + 6 nA 0.01 + 50 nA + 2.3 nA 12 + 6 nA + 180 pA
6 mA 0.03 + 1.5 µA + 60 nA 0.01 + 500 nA + 23 nA 12 + 60 nA + 1.8 nA
60 mA 0.03 + 15 µA + 600 nA 0.01 + 5 µA + 230 nA 12 + 600 nA + 18 nA
600 mA 0.045 + 150 µA + 6 µA 0.015 + 50 µA + 2.3 µA 15 + 6 µA + 180 nA
2 A 0.065 + 1.5 mA + 60 µA 0.03 + 500 µA + 23 µA 22 + 60 µA + 1.8 µA 6 A 0.15 + 1.5 mA + 60 µA 0.12 + 500 µA + 23 µA 40 + 65 µA + 1.8 µA
20 A 0.18 + 15 mA + 600 µA 0.15 + 5 mA + 230 µA 40 + 650 µA + 18 µA
Voltage measurement
range
600 mV 0.02 + 150 µV 0.008 + 80 µ␣ V 7 + 6 µV
6 V 0.02 + 180 µV 0.008 + 100 µV 7 + 7 µV
60 V 0.025 + 1 mV 0.01 + 500 µV 13 + 40 µV
200 V 0.025 + 5 mV 0.01 + 3 mV 13 + 150 µV
Current measurement
range
6 nA 0.5 + 5.5 pA + 60 fA 0.1 + 3.5 pA + 23 fA 150 + 350 fA + 1.8 fA
60 nA 0.5 + 10 pA + 600 fA 0.1 + 5.5 pA + 230 fA 120 + 600 fA + 18 fA
600 nA 0.2 + 50 pA + 6 pA 0.035 + 25 pA + 2.3 pA 40 + 3 pA +180 fA
6 µA 0.045 + 500 pA + 60 pA 0.015 + 220 pA + 23 pA 18 + 25 pA + 1.8 pA
60 µA 0.035 + 5 nA + 600 pA 0.015 +2.2 nA + 230 pA 15 + 250 pA +18 pA
600 µA 0.025 + 50 nA + 6 nA 0.01 + 22 nA + 2.3 nA 10 + 2.5 nA + 180 pA
6 mA 0.025 + 500 nA + 60 nA 0.01 + 220 nA + 23 nA 10 + 25 nA + 1.8 nA
60 mA 0.025 + 5 µA + 600 nA 0.01 + 2.2 µA + 230 nA 10 + 250 nA + 18 nA
600 mA 0.04 + 60 µA + 6 µA 0.01 + 32 µA + 2.3 µA 12 + 3.8 µA + 180 nA
2 A 0.06 + 600 µA + 60 µA 0.03 + 320 µA + 23 µA 20 + 38 µA +1.8 µA 6 A 0.15 + 600 µA + 60 µA 0.12 + 320 µA + 23 µA 40 + 38 µA + 1.8 µA
20 A 0.18 + 6 mA + 600 µA 0.15 + 3.2 mA + 230 µA 40 + 380 µA + 1.8 µA
Vo: Follow-up voltage (0 V to ±220 V) However, for current ranges, ±0.65 pA/˚C is added to the above values for temperature coefficients of 28˚C to 40˚C. Integration time: Measurement accuracy of 10 ms to 100 µs, the
following full-scale error is added to one day stability,
Measurement range
600 mV ␣ 8 11 18 25 36 57 80
6 V to 200 V 5 7 11 16 22 35 50
6 nA 10 14 22 32 45 71 100
60 nA 7 9 15 21 29 47 66
600 nA 6 8 12 17 25 39 55
6 µA to 2 A 5 7 11 16 22 35 50
Generation linearity:
Overall accuracy One-day stability Temperature coefficient
±(% of setting + range error) ±(ppm of setting + range error)/˚C
Overall accuracy One-day stability Temperature coefficient
±(% of setting + A + A x Vo/1V)
Overall accuracy One-day stability Temperature coefficient
±(% of reading + range error) ±(ppm of reading + range error)/˚C
Overall accuracy One-day stability Temperature coefficient
±(% of reading + A + A x Vo/1 V)
Reset time
10 ms 5 ms 2 ms 1 ms 500 µs 200 µs 100 µs
±(ppm of setting + A + A
±(ppm of reading + A + A
x Vo/1V)˚C
x Vo/1 V)/˚C
Units (digits)
±0.01% of range at 23˚C ± 5˚C, RH 75% or less
247
Page 3
DC Voltage/Current Generators/Calibrators
Synchronized 2-Channel Measurement Function
R6245/6246 (Continued From Previous Page)
Output noise: At no load and maximum load for voltage generation
(Vp-p), at the load resistances given for current generation (Ap-p)
Voltage/current
source range
600 mV 60 µV 300 µV 5 mV 4 mV
6 V 150 µV 500 µV 5 mV 4 mV
60 V 1 mV 2 mV 6 mV 5 mV
200 V 3 mV 4 mV 10 mV 6 mV
6 nA 100 M 500 fA 10 pA 100 pA 100 pA
60 nA 10 M 5 pA 100 pA 1 nA 1 nA
600 nA 1 M 50 pA 1 nA 10 nA 10 nA
6 µA 100 k 500 pA 10 nA 100 nA 100 nA
60 µA 10 k 5 nA 100 nA 1 µA1 µA
600 µA 1 k 50 nA 100 nA 10 µA 10 µA
6 mA 1 k 100 nA 1 µA 10 µA 10 µA
60 mA 1 k 1 µA 10 µA 10 µA 10 µA 600 mA 1 k 10 µA 20 µA 100 µA 100 µA 2 A, 6 A 100 100 µA 200 µA 200 µA 200 µA
20 A 10 1 mA 2 mA 2 mA 2 mA
resistance
Ripple noiseLoad
DC to 100 Hz FAST SLOW
DC to 10 Hz
High frequency noise
DC to 20 MHz
Switching noise:
Range switching noise
Output on/off noise
Range switching noise
Polarity switching noise
Response switching noise
Power source off noise
Voltage source 250 mV + 20 nA x R1 50 mV Current source 250 mV/R1 + 20 nA 100 µA
Voltage source 150 mV 50 mV Current source 250 mV/R1 + 20 nA 50 µA
Voltage measurement 150 mV 50 mV Current measurement 250 mV/R1 50 µA
Voltage source 150 mV 50 mV Current source 50 mV/R1 + 1 nA 50 µA
Voltage source 200 mV 50 mV Current source 80 mV/R1 + 20 nA 50 µA
Maximum value (p-p) Typical value (p-p)
0.6 V 0.6 V
R1 indicates load resistance () Typical values obtained at voltage range 6 V or less, current range 6
mA to 20 A, R1 = 1 k
Settling time: When the current range is 6 mA to 20 A, the time
required to reach ±0.03% of the final value when the output is changed from zero to full-scale. For simple resistance load, load capacity 2.5 pF max., compliance settings at full-scale
Item FAST SLOW
Voltage generation 300 µs 2.5 ms Current generation 500 µs 5.5 ms
Line regulation: ±0.0017% of range, at AC 100 V ± 10% Load regulation:
Voltage generation: ±0.0017% of range or less, with 4-wire
connection at maximum load
Current generation: ±5 ppm x Vo/1V or less, Vo: follow-up voltage
(0 V to ± 220 V)
Output resistance: With 2-wire connection, excluding output
cable
Maximum capacity load: Maximum non-oscillating capacity load
when voltage generation or voltage compliance is operating
Current range Maximum load capacity
6 nA 1 k 3.22 E + 13 0.01 µF
60 nA 100 3.22 E + 12 Ω 0.01 µF
600 nA 10 3.22 E + 11 Ω 0.01 µF
6 µA1 3.22 E + 10 Ω 0.01 µF
60 µA 200 m 3.22 E + 9 0.1 µF
600 µA 100 m 3.22 E + 8 1 µF
6 mA 100 m 3.22 E + 7 Ω 10 µF
60 mA 100 m 3.22 E + 6 Ω 100 µF 600 mA 100 m 3.22 E + 5 Ω 1000 µF 2 A, 6 A 100 m 1 E + 5 Ω 1000 µF
20 A 100 m 1 E + 4 1000 µF
Output resistance
Voltage source Current source
Standard accessory cable (A01010) resistance: 0.2 max. Maximum inductive load: Maximum non-oscillating inductive load
when voltage generation or voltage compliance is operating
Current source/compliance range Maximum inductive load
6 nA to 60 µA 50 µH
600 µA 300 µH
6 mA 500 µH
60 mA to 20 A 1000 µH
CMRR: With 1 k unbalanced impedance between LO and GUARD
terminals, DC and AC 50/60 Hz ± 0.09%
Item
Voltage measurement 60 dB 120 dB
Current measurement 0.01%/1 V 1 ppm/1 V
100 µs to 10 ms 1 PLC to 20 PLC
Reset time
NMRR: At AC 50/60 Hz ± 0.09%
Item
Voltage measurement 0 dB 60 dB
Current measurement 0 dB 60 dB
100 µs to 10 ms 1 PLC to 20 PLC
Reset time
External measurement input: When the current range is measuring
an applied voltage at the MEASURE INPUT terminal, the measured voltage is displayed in the set current range
Voltage measurement accuracy ±(0.02% of reading + 120 µV)*1 Current display accuracy ±(0.375% of reading + 7.6 µA)*2 Maximum allowable applied voltage ±250 V peak max.
External analog input: When an external analog signal is added
from the ANALOG INPUT terminal, the output added to the generation value of the set generation range can be obtained. 2 types of gain can be selected. Input and output ranges are decided by the source range.
Voltage source accuracy ±(0.16% of setting + 3.8 mV)*1 Current source accuracy ±(0.5% of setting + 10 µA)*2 Maximum input frequency FAST: 10 kHz, SLOW: 1 kHz, at ±6 V
output voltage, 6 mA to 6 A current range, full-scale
Maximum allowable applied voltage ±10 V peak max. Input resistance Approx. 10 M
V monitor: Voltage between the HI SENSE and LO SENSE terminals
divided at ±6.2 V full-scale is output at V MONITOR terminal
Monitor accuracy ±(0.01% of reading + 1 mV)*1 Output resistance 500 max.
248
Page 4
DC Voltage/Current Generators/Calibrators
Synchronized 2-Channel Measurement Function
R6245/6246
I monitor: Voltage which has been converted from current to voltage
at output current ±6.2 V full-scale is output at I MONITOR terminal
Monitor accuracy ±(35% of reading + 7 mV + 60 µV × Vo/1 V)*2 Vo; Follow-up voltage (0 V to ±220 V) Output resistance 500 max.
*1: Typical value (at 6 V range) *2: Typical value (at 60 mA range)
Generation/Measurement Functions DC generation/measurement: Generation and measurement of DC
voltage and current
Pulse generation/measurement: Generation and measurement of
pulse voltage and current
Sweep generation/measurement: Fixed, linear, log, random, pulse,
linear pulse, log pulse, random pulse
Sweep modes: Single/reverse, repeat 2-channel simultaneous generation/measurement: Synchronized
channel A and B generation and measurement of DC, pulse and sweep
Search Measurement :Binary search measurement. Linear search
measurement. Programming functions : Sequence programming function. All the above functions can be programmed. Max. programs: 128 characters × 100 programs
High-speed sequence programming function. Up to 20 items of DC measurement, pulse measurement, binary/ linear search measurement, time parameter, 16-bit digital output data, and conditional jumps can be programmed.
Results (Lo, Go, Hi, etc.) are output from the handler interface. * Search measurement and programming set via GPIB only. Generation data memory capacity: 2,048 data items/channel (when
random or random pulse is used)
Measurement data memory capacity: 2,048 data items/channel Trigger methods: Measurement or generation can be done with
autotrigger (free-run) and external trigger
Output methods: Bipolar, floating & guarded Output terminals: Tri-axial/binding post *1 Output types: HI OUTPUT, HI SENSE, LO OUTPUT, LO SENSE,
DRIVING GUARD (DG), GUARD
Maximum applied voltage between terminals:
Between HI and LO; 225 Vp-p max. (SMU220-2)/65 Vp-p max. (SMU62-20) Between LO and GUARD; 50 Vp-p max. Between GUARD and outer casing; 500 Vp-p max.
Maximum remote sensing voltage: ±5 V max. *2 Voltage measurement input resistance: 3.22 × 1013 min.
(SMU220-2)/3.22 × 109 min. (SMU62-20)
Voltage input leakage current: ±5 pA/±1 nA max. *3 Guard offset voltage: ±3 mV max. Maximum allowable card capacity: 1000 PF max. *4 Maximum allowable shield capacity: 5000 PF max. *5
*1: SMU220-2/SMU62-20 *2:
Voltage between the HI OUTPUT and HI SENSE terminals, between the LO OUTPUT and LO SENSE terminals and between the HI SENSE and LO SENSE terminals are within the range of the maximum output voltage. Includes voltage drop due to cable resistance.
*3:
When doing ISVM measurement at current generation range of 6 nA/60 µA
*4: Between HI (OUTPUT or SENSE) and DG terminals *5: Between DG and LO (OUTPUT or SENSE) terminals
GPIB interface: IEEE standard, conforms with 488-1978
SH1, AH1, T6, L4, SR1, RL1, PP0, DC1, DT1, C0, E2
Digital output: Internal status or 16 bit setting output
5 V pull-up with 10 k resistance at O/C
Trigger link: Used for synchronizing 2 units of the R6245/6246
Series for generation or measurement. TTL level I/O.
Scanner control: For use in scanner control with ADVANTEST
R7210 or other models. TTL level output.
Trigger input: Used at measurement start, sweep step up, pulse
output trigger. TTL negative pulse 10 ms min.
Trigger output: Used at measurement end, sweep end. Output TTL
negative pulse approx. 15 ms
Interlock input: Used when outputting by interlocking with a fixture Setting Times Hold time setting range:
Setting range Resolution Setting accuracy
100 ms to 65.5350 s 100 µs ± (2% + 40 µs)
Measure delay time setting range:
Setting range Resolution Setting accuracy
100 µs to 65.53500 s 10 µs ±(2% + 40 µs)
Pulse width, pulse period setting range:
Pulse period (Tp) Pulse width (Tw)* Resolution
700 µs to 655.35 ms 100 µs to 655.35 ms 10 µs
655.4 ms to 6.5535 s 100 µs to 6.5535 s 100 µs
6.554 s to 65.535 s 1 ms to 65.535 s 1 ms
* The pulse width resolution is always the same as the pulse period, so the possible setting
range for the pulse width is decided by the pulse period setting.
Setting accuracy:
Pulse period (Tp) ±(2% + 40 µs) Pulse width (Tw) ±(2% + 10 µs)
General Specifications Display:
R6245/6246 Fluorescent display CRT and LED display R6245A LED display only (POWER, ON, FAILED, CH COMMON
SHORT, OPERATE, 4 WIRE, LO-G SHORT)
Setting method:
R6245/6246 By keys, GPIB remote R6245A By GPIB remote only
Preheating time: 30 minutes or more until specified accuracy is
attained
Operating conditions: 0˚C to 40˚C, 85% RH Storage conditions: -25˚C to +70˚C Power requirements: Specify at time of ordering
Option no. Standard Option 32 Option 42 Option 44
Power voltage 90 V to 110 V 108 V to 132 V 198 V to 242 V 216 V to 250 V
Power frequency: 48 Hz to 66 Hz Power consumption: 350 VA (R6245/6245A)/850 VA max. (R6246) Dimensions: Approx. 424 (W) × 177 (H) × 500 (R6245/
6245A)/550 (R6246) (D) mm
Mass: 25 kg max. (R6245/6245A)/33 kg max. (R6246) Accessories:
Item Model Stock no. R6245/6245A R6246
Power Cable
Input/Output Cable A01010 DCB-FM1645 × 01 2 1
A01402 DCB-DD2428 × 01 1 0 A01437 DCB-DD5641 × 01 0 1
Accessories (Sold Separately) PR624501-FJ Semiconductor Parameter Auto Measurement
Software
R12701 Test Fixture A01009-050 Input/Output Cable (0.5 m) A01009-100 Input/Output Cable (1 m) A01009-150 Input/Output Cable (1.5 m) A01009-200 Input/Output Cable (2 m) A01010 Input Cable (1 m) A01038 Input/Output Cable (for R6246 channel A) A02710 Rack Mount Set A (EIA standard, with front handle) A02711 Rack Mount Set A (JIS standard, with front handle) A02720 Rack Mount Set B (EIA standard, without front
handle)
A02721 Rack Mount Set B (JIS standard, without front
handle)
A02703 Front Handle Set A02730 Rack Flange Set A (EIA standard) A02731 Rack Flange Set A (JIS standard) A02615 Slide Rail Set
249