Additel ADT260Ex User Manual

Page 1
260Ex
Handheld Multichannel Reference Recorder
1
Page 2
ADT260Ex Handheld Multichannel Reference Recorder
[Version:2211V01]
Additel
Corporation
Page 3
STATEMENT
This user manual provides operating and safety instructions for the ADT260Ex Handheld Multichannel Reference Recorder. To ensure correct operation and safety, please follow the instructions in this manual. Additel Corporation reserves the right to change the contents and other information contained in this manual without notice.
Page 4
Content
Safety instructions ............................................................................................................................................................ 1
Special safety requirements ............................................................................................................................................. 2
Intended use ...................................................................................................................................................................... 6
1. Introduction ................................................................................................................................................................... 7
1.1 Product Overview ................................................................................................................................................... 7
1.2 Technical specification ........................................................................................................................................... 8
1.2.1 General specification ........................................................................................................................................... 8
1.2.2 Signal measure specification (environmental temp.: 20±10, 1 year accuracy) ....................................... 10
1.2.3 Simulate RTD measure accuracy (environmental temp.: 20±10℃) ........................................................... 11
1.3 Basic Structure ..................................................................................................................................................... 14
1.4 Standard package include .................................................................................................................................... 16
1.5 Power supply description ..................................................................................................................................... 17
2.1 Main interface....................................................................................................................................................... 19
2.1.1 Status bar................................................................................................................................................... 19
2.1.2 APPs list ..................................................................................................................................................... 19
2.1.3 Main function guide .................................................................................................................................... 20
2.2 Control center....................................................................................................................................................... 21
2.2.1 Date and battery ........................................................................................................................................ 21
2.2.2 Diagnostic center and screenshot .............................................................................................................. 21
2.2.3 Shortcut settings ........................................................................................................................................ 21
I
Page 5
3. Data logger .................................................................................................................................................................. 22
3.1 Pressure Measurement ........................................................................................................................................ 23
3.2 RTD measurement ............................................................................................................................................... 25
3.3 Electrical signal measurement ............................................................................................................................. 26
3.3.1 DC Voltage measurement .......................................................................................................................... 26
3.3.2 Current measurement ................................................................................................................................ 26
3.3.3 Switch measurement ................................................................................................................................. 27
3.3.4 Pulse measurement ................................................................................................................................... 27
3.3.5 Frequency measurement ........................................................................................................................... 27
3.4 Filtering ................................................................................................................................................................ 28
3.5 Scaling ................................................................................................................................................................. 29
4. System settings ........................................................................................................................................................... 31
4.1 Bluetooth communication ..................................................................................................................................... 31
4.2 Power management ............................................................................................................................................. 31
4.3 System calibration ................................................................................................................................................ 31
4.4 Services ............................................................................................................................................................... 32
4.4.1 Maintenance .............................................................................................................................................. 32
4.4.2 Restore to factory ....................................................................................................................................... 32
4.4.3 Running information ................................................................................................................................... 32
4.4.4 System upgrade ......................................................................................................................................... 32
4.5 Personalization..................................................................................................................................................... 33
4.5.1 Sound ........................................................................................................................................................ 33
II
Page 6
4.5.2 Language ................................................................................................................................................... 33
4.5.3 Date & Time ............................................................................................................................................... 33
4.6 Product information .............................................................................................................................................. 34
5. Data management ....................................................................................................................................................... 35
6. Data logger .................................................................................................................................................................. 36
6.1 Display mode ....................................................................................................................................................... 36
6.1.1 Single-channel mode ................................................................................................................................. 36
6.1.2 Multi-channel list mode .............................................................................................................................. 38
6.2 Channel data ........................................................................................................................................................ 39
6.3 Channel curve ...................................................................................................................................................... 39
6.3.1 Following Mode .......................................................................................................................................... 39
6.3.2 Eye pattern mode ....................................................................................................................................... 39
6.4 Save the data ....................................................................................................................................................... 39
7. Applications ................................................................................................................................................................. 40
7.1 Differential pressure module ................................................................................................................................ 40
7.2 Units converter ..................................................................................................................................................... 41
7.3 Wiring help ........................................................................................................................................................... 41
7.4 Leak test .............................................................................................................................................................. 41
7.4.1 Leak Test Performing ................................................................................................................................. 42
7.5 PSV test ............................................................................................................................................................... 43
7.5.1 PSV test performing ................................................................................................................................... 43
7.6 Sensor library ....................................................................................................................................................... 44
III
Page 7
Figure Content
Figure 1 Basic Structure .................................................................................................................................................... 14
Figure 2 Adaptor and the power plug ................................................................................................................................ 17
Figure 3 Main interface...................................................................................................................................................... 18
Figure 4 Control center...................................................................................................................................................... 20
Figure 5 Calibrator main screen ........................................................................................................................................ 23
Figure 6 RTD measurement .............................................................................................................................................. 25
Figure 7 Voltage/ frequency/ pulse/ switch wiring .............................................................................................................. 26
Figure 8 Current measurement ......................................................................................................................................... 27
Figure 9 Filter settings ....................................................................................................................................................... 28
Figure 10 Scaling configurations ....................................................................................................................................... 30
Figure 11 Single-channel mode ......................................................................................................................................... 37
Figure 12 Multi-channel list mode ..................................................................................................................................... 38
IV
Page 8

Safety instructions

Warning:
To prevent the user from injury, please follow this user manual for use.
To prevent fire, electric shock, or personal injury, please do as follows:
Normal:
Please read the user manual before using the product, especially the Safety Instructions;
Please charge the battery when a low battery level is displayed in case of the measurement abnormal;
Do not expose the battery to fire or short circuit the battery;
Before using the product, please check the appearance of the product for any damage;
If the product is damaged or malfunctions, do not use it, and contact Additel;
Do not touch the metal part of the probes or test cables during use;
Please remove unnecessary probe, cable, or other accessories before using the product;
Do not use damaged or worn cable.
Attention:
To prevent damage this product or the device under tested, please obey the instruction manual for use
Do not use the instrument in a high vibration environment;
Use only the Additel power adaptor and designated battery models;
In case of working with external pressure module, refer to the Safety instructions of its manual.
1
Page 9

Special safety requirements

1. WARNING – DO NOT OPEN WHEN AN EXPLOSIVE ATMOSPHERE IS PRESENT
2. WARNING – DO NOT CHARGE THE BATTERY IN HAZARDOUS LOCATION
3. WARNING – USE ONLY the approved batteries, see Ex instruction
4. The equipment should be protected from high energy impacts
5. Do not touch the non-metallic enclosure or touch with the insulating materials only
6. The batteries used in the equipment should be the same manufacturer, model and electrochemical system
7. The conformal coating applied to PCB boards was declared meet the requirements of IEC 60664-1 and IEC 60664-
3 by manufacturer
8.Tamb:-20℃ to +50℃
9. USB port should only be used in Non-hazardous areas.
10.PX connections to be made with pressure measuring module CDPX-Ex only, the connected cables and flexible
hose used for the external pressure modules are made of insulation materials, so end user should consider the risk of
electrostatic discharge during material selection and installation process
11. Metallic parts (only fasteners) presented a maximum capacitance of total 110pF, end user should consider this risk
for suitable applications
12. No additional input energy limitation are required for Type-C USB port and DC charging port when Um does not
exceed 250V (internal OVP/OCP are designed in line with IEC 60079-11)
2
Page 10
13. Do not remove the silicone rubber protective sleeve from the enclosure of equipment in Hazardous area
14. IS parameters must be observed for different external measuring ports on the different models in accordance with
Ex instruction strictly and completely
15. Intrinsically safe electrical system between IS apparatus and Associated apparatus or other IS apparatus:
Item I.S Interface External Measuring Instrument System
Equipment group IIC IIC IIC
Level of protection ia ia ia
Temperature class T4 T4 T4
Ambient temperature -20°C~+50°C -20°C~+50°C
Voltage Uo Ui(30V)
Current Io Ii(100mA)
Power Po Pi(0.75W)
Cable parameters
Capacitance Co Ci Cc
Inductance Lo Li Lc
L/R ratio / / /
Grounding Isolated Isolated Isolated
Uo Ui Io Ii
3
Page 11
Po Pi Co Ci+Cc Lo Li+Lc
Standard Compliance
EN IEC 60079-0:2018
EN 60079-11:2012
IEC 60079-0:2017 Edition 7.0
IEC 60079-11:2011 Edition 6.0
Ex information for equipment name
National regulations: ATEX directive 2014/34/EU and IECEx scheme 02
Certificate No.:TÜV 20 ATEX 8509 X and IECEx TUR 20.0009X
Ex Marking for EU-Type Examination: II 1 G Ex ia IIC T4 Ga
Ex Marking for IECEx CoC: Ex ia IIC T4 Ga
Ex Marking for UKCA certificate: CSAE 22UKEX1201
Tamb: -20℃~+50℃
IP Rating:IP20
I.S Parameters:
4
Page 12
Additel 260Ex
Handheld Multichannel Reference Recorder
Input parameters, simple connection
Connection Function Ui/V Ii/mA Pi/mW Ci/nF Li/mH
J1, J3 HART (external power and external resistance) 30 100 750 10 0
J2, J3 Current measurement 30 100 750 1 0
J4, J3 Voltage, frequency, and switch measurement 30 100 750 1 0
J1, J2
J5-J8 J9-J12 Resistance measuring 30 100 750 20 0
Loop Circuit Transmitter Current Measuring /
HART(Internal Power and Resistance)
30 100 750 10 0
5
Page 13
Output parameters, simple connection
Connection Function Uo/V Io/mA Po/mW Co/µF Lo/mH
J1, J3
J2, J3 Current measurement 5 0.85 1.1 87 100
J4, J3
J1, J2
J5-J8 J9-J12 Resistance measuring 5 21 26.3 85 30
HART (external power and external
resistance)
Voltage, frequency, and switch
measurement
Loop Circuit Transmitter Current
Measuring / HART(Internal Power and
Resistance)
25.2 79.16 499 0.107 9
5 0.85 1.1 87 100
25.2 79.16 499 0.107 9

Intended use

The equipment to be used in Zone 0 hazardous area with gas group IIC.
The equipment is used in oil and gas platforms, oil refineries, chemical and petrochemical plants, pharmaceutical
industry, energy and gas processing industries.
6
Page 14

1. Introduction

1.1 Product Overview

ADT260EX is an intrinsically safe pressure and temperature logger provided by Additel. It can be used in zone 0
environment and has an 8-channel combination capability, supports measurement and logging of pressure, temperature,
barometric pressure and electrical signals. It can keep 10 million records, display the trend curve, and support Bluetooth
communication. Powerful logging capability and high-level intrinsically safe make it widely used in ex-proof pipeline field
/container pressure test, air leak test, inlet and outlet pressure test of pump station, safety valve and ex-proof membrane
test, wellhead pressure test, pipeline filter test, differential pressure transmitter/flow computing test, as well as used in the
routine calibration of pressure devices.
How to contact us
TEL: 1-714-998-6899
Or visit Additel website: www.additel.com
7
Page 15

1.2 Technical specification

1.2.1 General specification

Table 1 General specification
General Specification
Top: 2 channels RTD measurement, 1 channel electric signal measurement, φ4mm banana jacks
Input Channel
Barometric Accuracy ±55Pa
Measurement Rate
Data Storage Logging interval: from 0.1~9999 seconds, log up to 10 million readings (single channel)
Right side: 2 channels for external digital pressure module, Lemo style connection
Bottom: embedded digital pressure module (model ADT158Ex) field switchable.
Internal: 1 embedded atmospheric pressure sensor
mV, V, mA & frequency and RTD: 3 times/ sec
Pressure module: 1~10 times/sec selectable (3 as default)
Barometric: 1 times/ sec
Power
Environmental
8
4000mAh, 14.4Wh explosion-proof intelligent lithium battery, charging time is 6~8 hours, the battery can be
charged independently
Typical working time 100 hours (measurement mode)
Guaranteed temperature range of technical specifications: (-10 ~ 50)°C
*Temperature coefficient: ±5 ppm FS/°C (-20 to -10)°C
Page 16
Operating temperature: (-20 ~ 50)°C
Storage temperature range: (-30 ~ 70)°C
Humidity: 0% to 95% RH, non-condensing
Altitude: 3000 meters
Warm Up Time 10 minutes to fully meet technical specifications
Port Protection Voltage 30V max
ATEX & IECEX: Ex ia IIC T4 Ga (Ta = -20°C to +50°C)
Explosion-proof Grade
CE Certification TUV IEC61326, IEC61010
Rohs Compliance Rohs II Directive 2011/65/EU,EN50581:2012
Protection Level IP67, 1 meter drop test
Communication Isolate USB-TYPEC (slave), Bluetooth
Display 4.4-inch color display capacitive screen, transflective, with LED backlight
Size
Weight 1.65 lb. (0.75 kg)
Warranty Time 1 year
6.97" x 4.13" x 2.04" (177 mm x 105 mm x 52 mm) which does not include the bottom mount ADT158Ex if
CSA: Class I, Division 1, Group A,B,C and D, T4 Class I, Zone 0, AEx ia IIC T4 Ga
Ta = -20°Cto + 50°C Ex ia IIC T4 Ga
UKCA-EX
installed.
9
Page 17
±30V
0.015%RDG + 0.005%FS
0.1mV
Current Measurement
±30mA
0.015%RDG + 0.005%FS
0.1uA
Pulse Count
Loop Power

1.2.2 Signal measure specification (environmental temp.: 20±10, 1 year accuracy)

Table 2 Signal measure specification
Specification Range Accuracy Resolution Note
RTD Measurement
Voltage Measurement
Frequency Measurement
Switch On-Off
Measurement
Note 1: When the environmental temperature is (-20~-10) ℃, the temperature coefficient is:
(1)Measure of voltage, current and RTD±5ppmFS/℃.
Note 2: Input characteristics:
(1) Voltage range:-300~300mV input impedance >100 MΩ; 10
Accuracy
0~400Ω
±300mV 0.015%RDG + 0.005%FS 1uV
0.01~50000Hz (auto range) 0.005% RDG + 2 last digit
0 ~ 9999999, optional rising edge and falling edge Min threshold voltage: 2.5V
20V ± 10%, max output impedance: 320Ω, max load current: 25mA
0.01%RDG+0.005%FS
Units: Hz, kHz, MHz, CPM, CPH, s, ms, us
Inspection voltage: (3 ~ 30)V
Response speed: < 10ms, supports wet and dry switch
[1]
1mΩ Excitation current: 1 mA
Impedance: >100MΩ
Impedance: >1MΩ
Impedance: < 40Ω
6-digit auto-
resolution
Min threshold voltage: 2.5V
Page 18
(2) Voltage range:-30~30V input impedance >1MΩ;
(3) Current measure input impedance<40Ω;
Note 3: The excitation power supply for RTD measure is 0.2mA, and each position has 4-wire, 3-wire, and 2-wire
methods. The accuracy is as follows: (given accuracy data in the table in based on 4-wire)
(1) The accuracy of the 3-wire method is 10mΩ more than the accuracy of the 4-wire method;
(2) The accuracy of the 2-wire method is 50mΩ more than the accuracy of the 4-wire method.
Note 4Minimum threshold voltage for frequency and pulse measure2.5V;
Note 5The frequency measure supports following units: Hz, kHz, MHz, CPM, CPH, s, ms, μs
Note 6There are rising edge and falling edge trigger mode available for pulse measure.

1.2.3 Simulate RTD measure accuracy (environmental temp.: 20±10℃)

Table 3 Simulate RTD measure accuracy
Sensor type
PT10(385) -200850
PT25(385) -200850
Temperature(℃) Annual accuracy(℃)
-200200 0.59
200600 0.72
600850 0.82
-200200 0.27
200600 0.35
11
Page 19
PT50(385)
PT100(385)
PT100(391) PT100(3916) PT100(3926)
PT200(385)
PT400(385)
PT500(385)
PT1000(385)
-200850
-200850
-200850
-200850
-200850
-200850 -200200
600850
-200200
200600
600850
-200200
200600
600850
-200200
200300
300600
600850
-2000
0200
200600
600850
-200200
200600
600850
0.41
0.16
0.22
0.27
0.10
0.16
0.20
0.32
0.34
0.41
0.48
0.15
0.18
0.25
0.30
0.16
0.22
0.27
0.10
12
Page 20
200600 0.16
600850 0.20
Cu10(427) -200260 -200260 0.56
Cu50(428) -50150 -50150 0.13
Cu100(428) -50150 -50150 0.08
Ni100(617) Ni100(618)
Ni120 (672) -80260 -80260 0.05
Ni1000 -50150 -50150 0.07
-60180
-600 0.06
0180 0.05
Note: It complies with the international temperature scale ITS90, depends on the maximum tolerance of the RTD
measurement and simulate signal output.
13
Page 21

1.3 Basic Structure

Figure 1 Basic Structure
14
Page 22
Table 4 Basic Structure
Subject Content Description
1 Display+ capacitive touch screen Display area, touchable
2 Keys For operating the device
3 Lemo port A Connect to external pressure module.
4 Lemo port B Connect to external pressure module.
5 Adaptor port Power supply from adaptor.
6 USB slave port For USB communication.
7 Electric test plug Connection for electric test, including cables and HART communication.
8 Pressure module For pressure measure
15
Page 23

1.4 Standard package include

Table 5 Accessories
Accessories
Test leads 5 sets(10pcs)
Ex-proof USB Cable type A to type C (For Ex models only) 1 pc
Included
Optional
110V/220V external power adaptor 1 pc
Chargeable Li-ion battery 1 pc
Hanging strap with magnet 1 pc
ISO 17025 accredited calibration certificate 1 pc
ADT158Ex Built-in digital pressure module
ADT161Ex External digital pressure module
9530 Additel/ACal calibration software
9060 Pressure module connection cable
9905 Hard carrying case for handheld calibrators and readouts with space for two RTDs
9918-SC Soft carrying case, with space for handheld instrument, test leads, and accessaries
Note: in case of changes on the accessories list, please refer to the packing list in shipping.
16
Page 24

1.5 Power supply description

(1) By Ex-proof Li-ion battery
Supply the power by 1 pc BP3640Ex Ex-Proof Intelligent Li-ion battery.
Ex-Proof Li-ion battery is chargeable by external power independently.
The power adaptor is adaptable for standard in various countries.
Do not expose the battery to fire or short circuit the battery.
Use only the Additel power adaptor and designated battery models.
Figure 2 Adaptor and the power plug
17
Page 25
2. Overview for display and basic functions
The ADT260Ex will automatically go to Data Logger function after powered on. It can also be returned to main
interface for all the functions. (Refer to Figure.3).
Figure 3 Main interface
18
Page 26

2.1 Main interface

In the main interface, there are three sections from top to bottom: status bar, APPs list and main function guide, as
shown in Figure 3.

2.1.1 Status bar

On the top of the main interface shows the status information, including:
1. System time: real-time of the system;
2. Battery: real-time display the battery level and charging status. Icon indicates the adaptor is connected.
3. Bluetooth: icon indicates Bluetooth function is on, while icon indicates off;
4. Module A: icon indicates the external module A is connected, while indicates not connected;
5. Module B: icon indicates the external module B is connected, while indicates not connected;
6. Message center: icon flashes means there is message, warning or abnormal, icon means no message;
7. 24V status: icon indicates 24V power supply is on, icon disappears indicates the 24V power supply is off;
8. Screen lock: icon means the screen is locked.

2.1.2 APPs list

The APPs list shows all the applications provided in the device, including differential pressure module, unit converter,
wiring help, leak test, PSV test and sensor library. Click the button , or use the left and right key to switch between
interfaces of the APPs list.
19
Page 27

2.1.3 Main function guide

The main function guide at the bottom of the interface provides access to three main functions of the device: data
logger, data manager and setup.
Figure 4 Control center
20
Page 28

2.2 Control center

In any interface of the device, the control center (Figure 4 Control Center) can be called up by physical key
provides detailed information of date, battery, diagnostic center, screenshot functions, and some shortcuts items.

2.2.1 Date and battery

In the control center show the current date and precise battery percentage.

2.2.2 Diagnostic center and screenshot

Click icon in the control center to open the diagnostic center, which shows the real-time diagnostic hardware status
of the device. If some of the function fails, check the cause here, as in Figure 5. Click icon in the diagnostic center to
see the historic abnormal log. Click icon to take the snapshot, which can be viewed or deleted in data manager APP.

2.2.3 Shortcut settings

The control center provides series of shortcut settings:
1. Click icon to lock the screen, any touch operation and physical keys except the control center and power are
invalid during lock.
2. Click icon to open and close 24V loop power supply.
3. Click icon to open and close the backlight.
4. Click icon to open and close Bluetooth.
5. Click icon to enable and disable system sound.
. It
21
Page 29

3. Data logger

The ADT260Ex can be used for measuring and recording many signal types, including the signal of the internal
pressure module, external pressure module A and B, differential pressure module, barometric pressure, 2-wire RTD and
electrical signal measurement, totally 8 measurement features available. The electrical signal measurement supports
voltage, current, switch, pulse, and frequency. On the main page, there can be up to 6 measuring items display at the
same time.
On the middle of the channel there is the current measured signal of this channel. The icon of the signal type and the
corresponding range information is displayed on the upper left; Click the icon in the upper right corner to set the
current channel in the pop-up menu; Click the icon in the lower right corner to perform the zero on the current
channel. If the signal type selected in the channel has multiple values for display (for example, after enabling the scaling,
it shows concurrently the scaled value and initial value), the channel will also display these signal values below the main
signal value. In case of too many channels displayed on the main page, the device will automatically simplify the content
to fit the screen.
22
Page 30
Figure 5 Calibrator main screen

3.1 Pressure Measurement

The device can measure the pressure by internal pressure module, external pressure module A/B, differential
pressure, and barometer. The external pressure A and B are only available when the external pressure module is
connected. The differential pressure function is available when it is enabled in the “Pressure Delta ” APP and both the
23
Page 31
two modules are online. Select the proper module in the Channel settings to measure the pressure. Click the pressure
unit above the pressure measurement value to switch the pressure unit in the pop-up unit selection interface.
Click the menu icon in the pressure measurement channel and click the "Settings" menu in the pop-up menu to set
the pressure type, resolution, stability, tare, and other functions of the current pressure measurement channel.
Table 6 Pressure measurement settings
Subject Valid Val ue Description
Resolution 4/5/6 Resolution of current measurement channel
Pressure type GP/ AP Current measurement channel pressure type
Measurement frequency 1~10Hz
Stability enable Enable/disable Whether to turn on the stability judgment function
Stability value
Stabilization time Number Time to judge the pressure stability
Tare enable Enable/disable Enable or disable tare function
Tare value Pressure value Correction value of the tare function
Pressure value or percentage of
24
range
Data acquisition frequency of the current measurement
channel
A value for judging whether the pressure is stable. If the
measured value fluctuates less than this value within the
stable time, the calibrator will judge as stable
Page 32

3.2 RTD measurement

The connection of RTD measurement is shown in figure 6. Click the switching menu and select the RTD measurement
.
Figure 6 RTD measurement
Click the "Settings" of the RTD measurement channel to set various parameters for the functions.
Table 7 RTD settings
Subject Val id Value Description
Sensor type RTD Select the type of thermal resistance sensor
Wires 2/3/4 Select the RTD wire
Temperature unit
Resolution 0~0.001 Temperature display resolution
K / /
Temperature unit selection
25
Page 33

3.3 Electrical signal measurement

3.3.1 DC Voltage measurement

Please connect correctly as shown in the figure 7. Then switch the measurement signal to voltage measurement. To
ensure measurement accuracy and adapt to more usage scenarios, two different ranges can be selected for voltage
measurement in this device: (-30~30)V and (-300~300)mV , the user should select the appropriate
measurement item according to the use situation.
Figure 7 Voltage/ frequency/ pulse/ switch wiring

3.3.2 Current measurement

Please connect correctly as shown in the figure 8. Then switch the measurement signal to current measurement (-
30~30)mA .
26
Page 34

3.3.3 Switch measurement

Figure 8 Current measurement
The wiring method of the switch measurement function is the same as the voltage measurement. After wiring, please
switch the measurement signal channel to switch measurement .

3.3.4 Pulse measurement

The wiring method of the pulse measurement function is the same as the voltage measurement. After wiring, please
switch the measurement signal channel to (0 ~9999999) . Pulse measurement allows the user to set the pulse
counting method. In the pulse measurement mode, enter the setting menu to select whether the pulse counting uses
rising or falling edges for counting.

3.3.5 Frequency measurement

The wiring of the frequency measurement function is the same as the voltage measurement. After wiring, please
27
Page 35
switch the measurement signal channel to frequency measurement (0.01~50k)Hz .

3.4 Filtering

The ADT260Ex provides two filtering methods: first-order linear filtering and moving average filtering to process data to
meet the needs of different scenarios. Click the menu button of the measurement channel, select the filter menu item
in the pop-up menu, and the filter setting interface will be displayed (see Figure 9).
Figure 9 Filter settings
28
Page 36
Table 8 Filter configurations
Subject Val id Value Description
Enable/disable Enable/disable Enable or disable the filter function
Method first-order linear filter or moving average filter Filtering method
Filter coefficient 0 ~ 1 Only available when the first-order linear filter is selected
Number of samples 1~50 Only available when the moving average filter is selected

3.5 Scaling

The scaling function can convert electrical signals into other signals. Click the menu button of the measurement
channel, select the scaling menu item in the pop-up menu, and the parameter configurations related to scaling will be
displayed (see Figure 10).
Table 9 Scaling configurations
Subject Vali d Va lu e Description
Enable/disable Enable/disable Enable or disable the scaling function
Input range Number Range before scaling
Output range Number Range after scaling
Resolution 0~0.000001 The number of decimal places displayed after scaling
Transfer function Linear, square, or square root Scaling transfer function
29
Page 37
Figure 10 Scaling configurations
30
Page 38

4. System settings

On the main page of the device, click the system setting button in the middle of the bottom to enter the system
setting interface. System
settings.

4.1 Bluetooth communication

The device has the function of Bluetooth communication. Click the "Bluetooth Communication" menu item in the
system setting interface to enter the Bluetooth communication setting interface. The Bluetooth communication setting
interface provides the enabling and disabling of the Bluetooth communication function and the query function of the
Bluetooth device name and physical address.

4.2 Power management

Click the "Power Management" menu item in the system setting interface to enter the power management interface. In
the power management interface, the user can enable and disable the screen backlight, set the time for automatically
turning off the backlight,

4.3 System calibration

Click "System Calibration" in the system setting interface and enter the calibration password "123456" in the pop-up
password input box to enter the system calibration interface. In the system calibration interface, you can calibrate all the
measurement and output signal in the device. The calibration process is as follows:
1. Select the item to be calibrated in the signal list.
maintenance, system calibration, and various basic system settings are provided in the system
automatic standby, and automatic shutdown.
31
Page 39
2. Use a high-precision standard device, after fully warming up, follow the calibration guide in the interface and click the
"Start" button to start the calibration.
3. According to the reference calibration point provided on the interface, select the appropriate standard value, and enter
it.
4. Click the "Finish" button to send the calibration data to the module to complete the calibration.
Click the "Restore" button at the bottom of the calibration interface to restore to the factory calibration data of the
device. Click the “Cancel zero" button to clear the zeroed data of the current signal.

4.4 Services

4.4.1 Maintenance

This function is only open for manufacturer.

4.4.2 Restore to factory

Enter the password for this function, default password is 123456.
Restore to the factory default settings, will not restore the system calibration data.

4.4.3 Running information

To view the device running information. It includes the barometric reading, current battery voltage and electric module
information.

4.4.4 System upgrade

Users can upgrade the device firmware via the USB-C.
32
Page 40

4.5 Personalization

Click the "Personalization" menu item in the system setting interface to enter the personalized setting interface. In the
personalized setting interface, users can set the device's sound, language type, current time, and date, and time date
format according to their own preferences.

4.5.1 Sound

Change the sound level by adjusting the volume bar. It also provides sound configuration, as in Table 10.
Table 10 Sound settings
Subject Vali d Va lu e Comment
Sound enable On / Off Sound enable setting
Prompt sound On / Off Prompt sound setting
Touch sound On / Off Touch sound setting
Over range sound On / Off Over range sound setting
Alarm sound On / Off Alarm sound setting

4.5.2 Language

The device provides a multi-language user interface. Use this menu to change from the provided languages.
After the language is selected, the device needs to be restarted for the changes to take effect.

4.5.3 Date & Time

The device provides customizable settings for time and date, as shown in Table 11.
33
Page 41
Table 11 Date & time
Subject Vali d Va lu e Comment
Date
Time
Date format Y-M-D / M-D-Y / D-M-Y Date format setting
Separator
24 hours enable 24-hour or 12-hour format
2000-1-1 2099-12-31
00:00 23:59
-/.
Date setting
Time setting
Date separator setting

4.6 Product information

Product information is read-only information, includes basic information and the module information.
Basic information: including model, serial number and system version information.
Module information: including EM board version, ES board version and TMS board version.
External module information: including external module A and external module B.
34
Page 42

5. Data management

On the main page of the device, click the file management button in the middle bottom to enter the data
management interface. The file management interface is classified and managed by functional modules, and the data
saved by each function is managed in the corresponding folder, which is convenient for users to browse.
The functional modules that can save data files are: snapshot, data logger, PSV test and leak test. In the corresponding
folder, the users can open the operation menu through the menu button in the upper right corner and delete and
batch delete the files in it.
35
Page 43

6. Data logger

The most important function of the device is the data logger. Click the menu icon at the top right of the main
interface and click "Start logging" to enter the logging configuration interface. Here it is available to set the interval for
each logging point and the total number & time of loggings. After finishing the configuration, click OK to start.
During the process, the measured values of all currently selected channels and the time are recorded and saved in a
file. After the logging starts, the progress bar at the bottom and the percentage will show the progress, and the time
beside the progress bar will show the remaining time.
Click the stop icon on the upper right to terminate the current logging, and you can choose whether to store the data
that has been recorded during the process.

6.1 Display mode

The ADT260Ex supports single-channel mode and multi-channel list mode, click the icon at the left of the progress
bar to switch the display mode.

6.1.1 Single-channel mode

Under this mode, both the curve and data will be displayed at the same time, as shown in figure 11. In case of several
channels need to be recorded, users can switch to the curve and data of other channels manually or automatically.
36
Page 44
Figure 11 Single-channel mode
Click the data area to switch the channel mode. When the top right area shows , it means manual switch at
present. Click the left and right icon or the physical direction key to switch the channels. When the top right area shows
or , it means automatic switch mode is activated, and the device will automatically switch to next channel every 5s
or 30s. In this mode, it can also be switched to manual mode at any time.
37
Page 45

6.1.2 Multi-channel list mode

Under this mode, there can be the curve or data of at most 3 channels to be displayed at one page, as shown in figure
12. The channel can fit with the sizes and when logging more than 3 channels, there will be the icon for page switching,
click the icon or the physical direction key to switch between pages.
Figure 12 Multi-channel list mode
Click the icon at top right to switch to curve display. And click the icon to switch back to data display. When
38
Page 46
using curve, click the curve area to switch the curve mode.

6.2 Channel data

It includes the real time measured value, maximum, minimum, average, peak to peak value and standard deviation.

6.3 Channel curve

The Y axis of the channel curve can adapt to the scale based on current value. The curve supports following mode (1x,
2x and 4x) and eye pattern mode. Click the curve area to switch the mode of curve.

6.3.1 Following Mode

The curve will always show the latest data. When the display is fully filled with the curve, the X axis will move forward
by half, and new data will be continually recorded on the curve. For 1x following, every recorded point will be added for
one point on the curve; For 2x following, every two recorded points will be added for one point on the curve; and for 4x
following, every four recorded points will be added for one point on the curve.

6.3.2 Eye pattern mode

The curve will always show the samples of all recorded data. When the display is fully filled with the curve, the X axis
will be compressed to half.

6.4 Save the data

When the data logging is complete, the device will automatically display the record save interface, and the record
name, operator, time & date information need to be entered. Then the logging will be saved as a file. For the previous
saved data, it can be found in the “Data logging management” menu or go to the “Data management” function.
39
Page 47
functions channel list.

7. Applications

7.1 Differential pressure module

The ADT260Ex provides the function of combining two pressure modules (any 2 of the internal pressure module,
external pressure module A and B) into a differential pressure module. This provides a convenient differential
pressure(DP) test solution in a high static pressure environment. The set parameters of the DP module are shown in the
table below.
Table 12 Pressure delta setting
Project Effective value Description
Enable or disable the differential pressure module. When the
Enable Enable/ disable
Resolution 4/5/6 The resolution of differential pressure
Range Numbers Differential module measuring range
Calculate type The difference between two modules The composition method of differential pressure module
Real time data ------ or Real time differential pressure value
differential pressure module is enabled, the differential
pressure module channels will be displayed in the related
The real-time differential pressure value is displayed only when
the differential pressure module is enabled, and the pressure
modules are all online. Otherwise, the real-time differential
pressure value is displayed as ---------
40
Page 48
Zero offset
zeroing
zeroing
offset will be saved and used for calculation
the zero offset to 0
Pressure module A S/N of the pressure module
Pressure module B S/N of the pressure module
Zero
Zero offset generated by performing
Zero offset when performing differential pressure module
Perform zero to the differential pressure module, the zero
Cancel zero
Cancel the zero-offset created when performing the zero, reset

7.2 Units converter

The device provides unit converter for pressure units and temperature units, users can select multiple pressure units or
temperature units for conversion. Click the unit converter icon on the main interface of the device to enter the unit
converter. Click icon on the right top to select and switch between different unit types. Click the unit under the value
display to select and switch between different units.

7.3 Wiring help

The wiring help function in the device can help users to connect the wires correctly under various working conditions of
the calibrator. It can be entered by clicking the "Wiring help" menu in each channel, or by clicking the "wiring help" icon in
the main interface of the device.

7.4 Leak test

The device provides a pressure leak test function, which can perform self-checking on the seal of the device. Click on
41
Page 49
The pressure type being recorded, view detail according to
device’s module
the pressure module column to set the test parameters, see the table below:
Table 13 Leak test setting
Subject Valid value Description
Pressure type GP/ AP
Pressure unit Pressure unit Select corresponding pressure unit
Waiting time Numbers The pressurizing time of external module
Test time Numbers Total record time
Unit of the rate Second/ minute Displayed unit after calculation

7.4.1 Leak Test Performing

1. Connect to air circuit
a. Connect the device to the air circuit;
b. Click the start icon at the bottom of the screen, apply pressure to the leak test point;
c. Start counting down the waiting time, and record the real-time pressure at the end of the waiting time as the initial
pressure;
d. Start the test, count down the test time, and start to calculate the real-time leakage, real-time leakage = initial
pressure-real time pressure;
e. Record the pressure at the end of the test time as the end pressure;
f. The entire leak detection process ends, and the final leakage = initial pressure - end pressure;
42
Page 50
2. The entire process will be displayed in stages in the leak curve at the bottom of the screen.

7.5 PSV test

The device provides function of PSV test, to test the safety pressure of the safety valves. Click the pressure module on
the top, to set relevant parameters, see table below:
Table 14 PSV test setting
Subject Valid value Description
Pressure module Internal pressure module, external module A or B Select the pressure module
Test time 00:00:00 ~ 00:10:00 Set test time
Pressure unit Different modules support different pressure units Select corresponding pressure unit

7.5.1 PSV test performing

1. Connect air circuit
a. Connect the device to the air circuit;
b. Press the Start icon on the right of the screen to start performing, apply the pressure gradually to the safety valve;
c. Count down the test time and record the real-time pressure value and create the pressure curve, then record the
maximum pressure during the process;
d. Count down complete, the entire leak test process finishes, save the data;
2. The entire process will be shown in the leak test curve at the bottom of the screen
43
Page 51

7.6 Sensor library

To meet the needs of custom sensor types, the ADT260Ex provides a sensor library function. In the sensor library
function, users can define new sensor types according to their needs and set relevant parameters in the sensor. The
sensor library supports a total of four types of custom sensors: ITS-90, CVD, RTD and TC.
44
Page 52
45
Loading...