ADA INSTRUMENTS PROF-X2 User Manual [ru]

MEASUREMENT FOUNDATION
Operating manual
Optical theodolite
Model
Manufacturer: ADAINSTRUMENTS Address:WWW.ADAINSTRUMENTS.COM
PROF X2
:
ENG
1.
Review
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2.
Technical parameters
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
3.
Complete set
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
4.
Construction features
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
5.
How to use the instrument
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
6.
Adjusting the instrument
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
7.
Precautions
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
8. Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
9. Exceptions From Responsibility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
10.
Attachement 1- “Certificate of acceptance and sale”
11.
Attachement 2- “Warranty card”
PROF X2
MEASUREMENT FOUNDATION
1. REVIEW
2. TECHNICAL PARAMETERS
Telescope:
- Image formation positive
- Magnification 30х
- Effective aperture of lens 40 mm
- Angle of visual field 1° 30'
- Shortest visual range 2 m
- Stadia multiplication constant 100
- Stadia addition constant 0
- Length 172 mm
Spirit level:
- Tubular level of the alidade 30” / 2 mm
- Circular level 8' / 2 mm
Prof X2 are optical theodolites of medium precision. These instruments are easy to use. Theodolites Prof X2 are used in national defence construction, geodesy and project surveying. They are mainly used in the mapping of vast scale relief-maps and doing survey. It is also suitable for use in general project surveying like municipal construction, communication facilities in mines, etc. Theodolites can be equipped with various kinds of accessories to meet needs of all sorts of surveying work: With tripods:
- ADA Strong, FS 23, M1Y.
- ADA Strongwood, FS 24 (SJJS50). With staff:
- ADA Staff 3, 4, 5. With pole.
PROF X2
3
4
PROF X2
Graduated circle and tape measure:
- Graduation diameter of horizontal circle 94mm
- Graduation interval (H) 1°
- Graduation diameter of vertical circle 76 mm
- Graduation interval (V) 1°
- Graduation interval of microscope scale 1'
- Magnification of the horizontal system 68x
- Magnification of the vertical system 65.4x
Standard error of angle measurement with one taking:
- horizontal Prof X2 2” /10”
- vertical Prof X2 2” /10”
Compensator for vertical circle index:
- working range +/- 2'
Optical plummet:
- magnification 2,5х
- angle of visual field 5°
- range of focusing 0.7 m -
Weight and dimensions of the instrument:
- height 207 mm
- net weight 4.3 kg
- weight of packing case 3.5 kg
3. COMPLETE SET
Theodolite with removable tribrach, case, lens protector from sunlight, plummet, pin, screwdriver, brush, cleaning flannel-cloth, desiccant, rain cover.
MEASUREMENT FOUNDATION
MEASUREMENT FOUNDATION
Fig.1 Optical scheme of theodolite
PROF
X2
5
6
MEASUREMENT FOUNDATION
1.The structure of the instrument is compact and ingenious. It affords easy portability and direct, rapid reading. The newest form of construction at present is adopted in its designing, and it is stable and reliable.
2.This theodolite is equipped with automatic compensator of vertical circle index with V-shape long pendulum, which gives a good performance in resisting high-frequency vibrations. This not only promotes surveying precision, but also, makes the operation easy. It simplifies manipulation procedures, and raises working efficiency.
Fig.2 Photograph of the theodolite
4. CONSTRUCTION FEATURES
1.Optical aimer
2.Focussing sleeve of telescope
3.Reading eyepiece
4.Telescope eyepiece
5.Circle-setting knob
6.Сlamping screw of tribrach
7.Foot screw
8.Tribrach
9.Vertical compensator clamp
10.Micrometer knob
11.Change-over knob
12.Plate level
13.Adjusting screw for plate level
1
2
3
4
5.
6
7
8
10
9
12
11 13
PROF X2
MEASUREMENT FOUNDATION
Fig.3 Photograph of the theodolite
3.The color visual field makes reading soft and smoothing to the eye.
4.The large aperture and the high magnifying power of the telescope lens ensure the possibility of getting a sharp image quality even under relatively poor lighting conditions.
5.Loosen the screw that fastens tribrach and the upper part of the instrument can be dismounted from it to allow the mounting of accessories. Theodolite is equipped with optical plummet (5), which enables users to center the instrument over the point.
13.Telescope lens
14.Cover plate for index error
15.Eyepiece of optical plummet
16.Circular level
17.Vertical drive
18.Horizontal drive
19.Illuminating mirror for horizontal circle
20.Illuminating mirror for vertical circle
21.Knobs fix the position of axis
20
14
16
21
19
18
17
15
13
21
PROF X2
7
8
PROF X2
5. HOW TO USE THE INSTRUMENT
1. Positioning the tripod.
Place the tripod over the measuring point and set the three legs almost equidistant to it. The place of mounting should be as level as possible. Turn the wing clamps on the legs clockwise. It will enable them to be properly extended. The pointed shoes on the tips of the legs should pierce the surface of the ground somewhat so that the tripod is surely erected and reliable as surveying work is being performed on it.
2. Mounting the instrument on the tripod.
Place the instrument carefully on the tripod and tighten the central screw properly.
3. Adjusting the level of the instrument.
Use the circular level and foot screws of tribrach to roughly adjust the level of the instrument and then use the tubular one to do precision levelling. Levelling with circular level: Turn the screws 1 and 2, as shown in figure 4A, so that the bubble moves to the central line of the circular level, and then turn the screw 3, as shown in figure 4B, so that the bubble is exactly in the center of the level.
3
1
Fig.4A
2
Fig.4B
MEASUREMENT FOUNDATION
MEASUREMENT FOUNDATION
Levelling with the tubular level: Turn the alidade of theodolite first in order to make the tubular level parallel to the line
connecting the centers of any two foot screws (Fig. 5A), and at the same time, turn the two foot screws to enable the bubble to position itself in the middle of the tubular level. Then turn the alidade 90° (Fig. 5B) and also turn the third foot screw to make the bubble move to the middle of the level. Repeat the above described procedure several times so that the offset of the bubble doesn't exceed the allowance for whatever position the alidade is turned.
4. Mounting the instrument properly over the measuring point.
Use an optical plummet to enable the center of the vertical axis of the theodolite to be directly over the plummet line of measuring point. When the optical plummet is used, pull out its visual lens and turn it so that the measuring point is distinctly imaged on the graduated plate. If the center of the measuring point is not in the center of the graduated plate, loosen the central screw of the tripod. Then tighten the central screw of the tripod, and check the levelling of the instrument. Turn the alidade 180° and repeat the above described procedure.
5. Illumination
Turning the position of the light reflection mirror (11) properly will result in a bright illumination of the graduated circle.
Fig.5A
Fig.5B
PROF X2
9
10
6. Angle measurement
1. Turn the telescope to the given point. Use the rough optical sighting device to aim at the object. Then rotate the eyepiece in the counter-clockwise direction until the clear image of reticule. Turn the focussing sleeve of the telescope so that the object will be clearly imaged on the graduated plate.
2. Fix the position of telescope with the help of horizontal and vertical (21) clamps. Turn both horizontal and vertical micrometer screws (17 and 18) so that the telescope is precisely aimed at the target to be measured. At this moment, move the eye left and right, and then up and down. In doing so, no relative displacement should be detected between the target image and the graduation marks on the graduated plate. This means that no parallax exist.
PROF
Рис.6А Рис.6B
X2
7. Graduated circle reading.
After illuminating the graduated circle, adjust the reading eyepiece (4) so that the reading visual field, as shown in Fig.6, can be seen distinctly. The symbols H and V indicate the readings of the horizontal and the vertical circles respectively. There are 60 graduations to every division (1') on the tape and so estimation readings up to 0,1. And the reading error 0,05-0,1 of the graduated circle corresponds to 3"-6" angular sec. The system of readings of vertical angles starts from the horizon 90°. Before making the reading on the vertical circle, it is necessary to turn the locking knob (10) of the automatic compensator to enable the compensator index mechanism to be in a free state. At this moment very distinct sound is audible. And only then you can take the readings of the vertical circle.
Рис.7А Рис.7B
MEASUREMENT FOUNDATION
MEASUREMENT FOUNDATION
6. ADJUSTING THE THEODOLITE
8. Stadia measurement.
There are two pairs of short hairs on the graduated plate of the telescope, i.e. stadia hairs. They are used to measure the distance between the target to be measured and the measurement point of the instrument. Assume the distance between the staff and the measuring point is D. The length cut by a pairs of the stadia hairs on the scale is L and constants of the stadia are K and C which are 100 and 0 respectively. The distance measurement formula is following:
D=KL+C-100L Formula of horizontal distance, taking into account a slope: D=KLC*OS a, where a- a vertical angle of slope Formula for calculation the exceeding between points: H=0,5 kL siin2a +1-v, where l- height of theodolite, v- height of aiming at the staff.
9. Using the rotary circle mechanism.
Theodolites Prof X2 are equipped with a rotary circle knob (5). It enables the reading value of the horizontal graduated circle to be placed in the direction of the target to be measured. To avoid errors, it is necessary to aim at the target, to press the locking catch of the rotary knob (5) and press the knob (5) simultaneously. Then, rotating the knob (5) in the counter-clockwise direction, set the necessary reading. After this procedure press the locking catch again to unlock the limb.
All instruments are adjusted and checked and then transported to users after careful packing. But nevertheless before usage it is necessary to repeat the adjustment as the transportation and storing conditions may tend to engender the possibility of deviation from correct positions adjusted originally. Such adjustment should be done as follows:
1. Adjusting the tubular level of the alidade
Turn the alidade so as to enable the tubular level to be parallel with the line connecting the centers of any two of foot screws.
PROF X2
11
12
MEASUREMENT FOUNDATION
Turn the two screws in opposite directions equally so that the bubble will be in the middle of the level. Turn the alidade 90° and with the third screw make the bubble in the middle of the level. Turn the alidade 90° again. If it's necessary, make the bubble in the center again. Turn the alidade 180°. If the bubble is not in the middle, you should do further adjustment. The foot screw on the half to which the bubble deviates should be turned so as to eliminate the offset value. The other half can be corrected by turning the screw on the tubular level. For convenience, use the special pin.
2. Adjusting the circular level.
After adjusting the tubular level you should adjust the circular level. If the bubble of the circular level deviates from the center, adjust it with three adjusting screws, which are under the level.
3. Adjusting the reticule of the telescope.
Focus the telescope at a plummet about 50 metres away. With micrometer screw, overlay the image of the top of the vertical hair of telescope with the image of plummet of the hair. If they differ in more than 3 strokes, it's necessary to do the adjustment.
1. Screw the protective cap of the graduated plate of the telescope in the counter-clockwise direction and take it off. You see the four adjusting screws on the frame.
2. Loosen the four adjusting screws with the help of screwdriver. Turn the frame of the graduated plate to enable the hair to be vertical. Fix carefully new position of the frame tightening the screws. Repeat the adjustment.
4. Correcting the collimation axis error 2c
Place the theodolite at distance 100 metres from the target. Level the instrument and then check it.
1. Focus the telescope at the target in the position of the instrument “circle left” and take horizontal circle reading Л1.
2. Repeat the same focusing but in the position of the instrument “circle right” Take reading П1.
3. Loosen clamp finger of the support, turn theodolite 180° and then fix the clamp finger again.
4. Calculate the value of collimation axis error: c=0,25[(Л1-П1±180°)=(Л2-П2±180°)]
PROF X2
Loading...
+ 26 hidden pages