A&D SK-30K User Manual

Page 1
1WMPD4002371
- 0 -
Page 2
SAFETY PRECAUTIONS
All safety messages are identified by the following the words "WARNING" and "CAUTION". These words mean the following.
Important information to alert you to a situation
WARNING
that might cause serious injury and damage to your property if instructions are not followed.
CAUTION
When using the SK, the following safety precautions should always be followed.
Use only the AC adapter which comes with the scale. Other adapter may cause damage. (AC adapter is optional.)
Internal service or adjustment to this product should be performed by a qualified person.
Avoid installing the scale in direct sunlight, that may cause discoloration or malfunctions.
Do not mix the type of batteries. Replace all other batteries at the same time.
If the scale is not to be used for a long period of time, remove all batteries from the battery compartment to avoid leakage which may cause damage to the instrument.
Avoid overloading that could cause damage to the scale.
Avoid using the weighing pan to move the scale, as that could cause damage to the scale.
Important information that tells how to prevent damage to equipment.
WARNING
CAUTION
- 1 -
Page 3
PARTS DESCRIPTION
- 2 -
Page 4
PREPARATION
1. Installing batteries / Connecting the AC adapters
Batteries
When replacing batteries, be careful not to apply too much force to the weighing pan. An excessive force could damage the scale.
AC adapter (optional)
CAUTION
Remove the battery compartment cover and insert six batteries (R20P/LR20/D size) into the battery compartment, taking extreme care that the polarities plus and minus are observed.
Plug the AC adapter to the AC adapter socket inside the battery compartment. The AC input requirement could be 100, 120, 220, 230, or 240Volts (50/60Hz) depending on the area where used, so please verify that the adapter is correct.
2. Setting up the SK
CAUTION Avoid installing the scale in direct sunlight, that may cause discoloration or malfunctions. Place your SK on a firm weighing table so that the scale is level. (The scale will not perform accurately when it is not level.)
Place the SK on the firm surface so that the sprit level shows that the scale is leve for an accurate weighing.
- 3 -
Page 5
WEIGHING
Before weighing
Automatic Power-Off Function.
If the scale is left ON and the stable indicator is displayed, the automatic power-off function turns the power off after approximately five minutes (factory setting). To disable or enable the function, press the [ON/OFF] key while pressing the [RE-ZERO] key. The scale stores this setting. To verify this, press both the [RE-ZERO] and [UNITS] keys while in the weighing mode. (See "CALIBRATION" about the [UNITS] key.)
poff0 Function disabled poff1 Function enabled
Error Messages
e Overload
----- The scale zero is out of range lb Low battery
Calibration May be Required Before Weighing.
Read "CALIBRATION" first and if necessary, calibrate your scale for accurate weighing.
Weighing procedures
1. Turn on the scale.
Press the [ON/OFF] key to turn on the scale. Then the scale displays all segments for a few seconds and then 0 will be displayed.
2. Start weighing If you do not use a container for weighing,
Press the [RE-ZERO] key to display 0 . Verify the reading is 0 . Place objects on the weighing platform to weigh. When the
reading becomes stable, the stable indicator is displayed.
If you use a container for weighing,
Place an empty container on the platform. Wait for the stable indicator [RE-ZERO] key. Place the objects to be weighed in the container. When the reading becomes stable, the stable indicator
Select the weighing unit, --- Only for the “(k)g-lb” model---
If you use the (k)g-lb model, press the [UNITS] key to select weighing with either (k)g or lb as the unit. Normally scales power up in lb when shipped from the factory. To power up in (k)g, turn the scale off. Press and hold the [UNITS] key, then press the [ON/OFF] key. The scale will start with (k)g. To return to the original setting, repeat this instruction.
to be displayed and press the
is displayed.
- 4 -
Page 6
CALIBRATION
When to Calibrate
Calibration may be required when the scale is initially installed, if it is moved a substantial distance, or in accordance with local regulations. This is necessary because the weight of a mass in one location is not necessarily the same in another location. Also, with time and use, mechanical deviations can occur.
The Position of the [UNITS] key for the Calibration
The (k)g model has no [UNITS] key, but the key for the calibration is located under the overlay.
Calibration Using a Calibration Weight
Location of the [UNITS]
Calibrate the scale with the following calibration weight.
SK-1000: 1000g ± 0.1g SK-2000: 2000g ± 0.2g SK-5000: 5000g ± 0.5g SK-10K: 10kg ± 1g SK-20K: 20kg ± 2g SK-30K: 30kg ± 2g
1. Enter calibration mode.
Press the [ON/OFF] key to turn the power off. While pressing both the [RE-ZERO] and [UNITS] keys, press the [ON/OFF] key. Then, Cal will be displayed.
2. Zero calibration
Press the [RE-ZERO] key to calibrate zero. Cal0 will be displayed. When you have proceeded to the display other than Cal0 , press
the [ON/OFF] key to turn the power off and start over again.
Wait for the stable indicator to be displayed and press the [RE-ZERO] key. Calf will be displayed in a few seconds. If the zero calibration only is to be done, press the [UNITS] key. The scale will show end and automatically return to the weighing mode.
3. Span calibration
Place the weight at the center of the platform for accurate calibration. Wait for the stable indicator press the [RE-ZERO] key. The scale will show end and the scale will automatically return to the weighing mode.
- 5 -
to be displayed and
Page 7
Calibration by gravity compensation
If you have calibration weights at the capacity of the scale, you can calibrate the scale with the weights instead of compensating for the gravity acceleration.
The scale is calibrated at 9.798 m/sec acceleration at your location is not the same, calibrate the scale by compensating for the gravity acceleration. (See “The Value of Gravity at Various Locations” and “World Map” and find your location value.)
1. Enter the calibration mode.
Press the [ON/OFF] key to turn the power off. While pressing both the [RE-ZERO] and [UNITS] keys, press the [ON/OFF] key. Then,
Cal will be displayed.
2. Select a digit to be changed.
Press the [UNITS] key. The previously set value, for example 979,8 , will be displayed. A digit after the comma “,” will be changed by pressing the [RE-ZERO] key. Press the [UNITS] key to move the comma “,” and select a digit to be changed.
3. Set the new value.
2
at the factory. If the gravity
Change the value of the gravity acceleration using the [RE-ZERO] and [UNITS] keys.
4. Save the value in memory.
Press and hold the [UNITS] key until end is displayed. The scale will automatically return to the weighing mode.
- 6 -
Page 8
SPECIFICATIONS
W
Model
Capacity
Resolution
Non-linearity ±1g ±2g ±4g ±0.01 kg ±0.02 kg ±0.04 kg Repeatability 0.5g 1g 2g 0.005 kg 0.01 kg 0.02 kg Span drift ±0.015% / °C TYP (5°C~35°C / 41°F~95°F)
Operating temperature
Display 25 mm / 0.98 inches, 7 segment liquid crystal display Display update Approximately 10 times per second Power 6 x R20P / LR20 / “D” size batteries or AC adapter
Battery life
Platform size 230 mm (W) x 190 mm (D) / 9.05 in(W) x 7.48 in(D) Dimensions 244 (W) x 232 (D) x 137 (H) mm / 9.61 (W) x 9.13 (D) x 5.39 (H) inches.
eight (
approximately.) Calibration weight
Accessories This manual Options AC adapter
SK-1000 SK-2000 SK-5000 SK-10K SK-20K SK-30K
1000 g
2.2 lb
0.5 g
0.001 lb
2000 g
4.4 lb 1 g
0.002 lb
5000 g
11 lb
2 g
0.005 lb
10 kg
22 lb
0.005 kg
0.01 lb
20 kg
44 lb
0.01 kg
0.02 lb
30 kg
66 lb
0.02 kg
0.05 lb
-10°C~40°C / 14°F~104°F, Less than 85%RH
Approximately 600 hours with manganese type
1200 hours with alkaline cells at 20°C / 68°F
1.6 kg / 3.53 lb 1.9 kg / 4.19 lb
1000g±0.1g 2000g±0.2g 5000g±0.5g 10kg ±1g 20kg ±2g 30kg ±2g
THE VALUE OF GRAVITY AT VARIOUS LOCATIONS
Amsterdam 9.813 m/s2 Manila 9.784 m/s2 Athens 9.800 m/s2 Melbourne 9.800 m/s2 Auckland NZ 9.799 m/s Bangkok 9.783 m/s Birmingham 9.813 m/s Brussels 9.811 m/s Buenos Aires 9.797 m/s2 Ottawa 9.806 m/s2 Calcutta 9.788 m/s CapeTown 9.796 m/s Chicago 9.803 m/s Copenhagen 9.815 m/s Cyprus 9.797 m/s2 Singapore 9.781 m/s2 Djakarta 9.781 m/s2 Stockholm 9.818 m/s2 Frankfurt 9.810 m/s2 Sydney 9.797 m/s2 Glasgow 9.816 m/s2 Taichung 9.789 m/s2 Havana 9.788 m/s2 Taiwan 9.788 m/s2 Helsinki 9.819 m/s2 Taipei 9.790 m/s2 Kuwait 9.793 m/s2 Tokyo 9.798 m/s2 Lisbon 9.801 m/s2 Vancouver, BC 9.809 m/s2 London (Greenwich) 9.812 m/s Los Angeles 9.796 m/s Madrid 9.800 m/s2 Zurich 9.807 m/s2
2
Mexico City 9.779 m/s2
2
Milan 9.806 m/s2
2
New York 9.802 m/s2
2
Oslo 9.819 m/s2
2
Paris 9.809 m/s2
2
Rio de Janeiro 9.788 m/s2
2
Rome 9.803 m/s2
2
San Francisco 9.800 m/s2
2
Washington DC 9.801 m/s2
2
Wellington NZ 9.803 m/s2
- 7 -
Page 9
WORLD MAP
- 8 -
Page 10
R
3-23-14 Higashi-Ikebukuro, Toshima-ku, Tokyo 170-0013 JAPAN Telephone: [81] (3) 5391-6132 Fax: [81] (3) 5391-6148
A&D ENGINEERING, INC.
1756 Automation Parkway, San Jose, California 95131 U.S.A. Telephone: [1] (408) 263-5333 Fax: [1] (408)263-0119
A&D INSTRUMENTS LIMITED <UK Office>
Unit 24/26 Blacklands Way, Abingdon Business Park, Abingdon, Oxfordshire Telephone: [44] (1235) 550420 Fax: [44] (1235) 550485
A&D INSTRUMENTS LIMITED <German Sales Office>
Große Straße 13 b 22926 Ahrensburg Deutschland Telefon: [49] (0) 4102 459230 Telefax: [49] (0) 4102 459231
A&D Australasia Pty Ltd.
32 Dew Street, Thebarton, South Australia 5031 AUSTRALIA Telephone: [61] (8) 8301-8100 Fax: [61] (8) 8352-7409
A&D KOREA Limited 한국에이.엔.디(주)
대한민국 서울시 영등포구 여의도동 36-2 맨하탄 빌딩 8 ( 8th Floor, Manhattan Bldg. 36-2 Yoido-dong, Youngdeungpo-ku, Seoul, KO 전화: [82] (2) 780-4101 팩스: [82] (2) 782-4280
A&D RUS CO., LTD. Компания ЭЙ энд ДИ РУС
121357, Российская Федерация, г.Москва, ул. Верейская, дом 17 ( Bldg. 17, Vereyskaya st., Moscow, 121357 RUSSIAN FEDERATION ) тел.: [7] (495) 937-33-44 факс: [7] (495) 937-55-66
A&D Instruments India Private Limited
( 509, Udyog Vihar, Phase-
: 91-124-4715555
, Gurgaon - 122 016, Haryana, India )
: 91-124-4715599
- 9 -