All safety messages are identified by the following, “
of ANSI Z535.4 (American National Standard Institute: Product Safety Signs
and Labels). The meanings are as follows:
A potentially hazardous situation which, if not avoided,
WARNING
CAUTION
This manual is subject to change without notice at any time to improve the
product.
Product specifications are subject to change without any obligation on the part
of the manufacturer.
When using the SJ-WP, the following safety precautions should always be
followed.
WARNING
CAUTION Avoid installing the scale in direct sunlight, which may cause discoloration
Internal service or adjustment to this product should be performed
by a qualified person.
or malfunctions.
could result in death or serious injury.
A potentially hazardous situation which, if not avoided,
may result in minor or moderate injury.
WARNING
” or “
CAUTION
”,
Do not mix battery types, or new and old batteries. Replace with all new
batteries at the same time.
If the scale is not to be used for a long period of time, remove all batteries
from the battery compartment to avoid leakage.
Avoid overloading the scale.
Avoid using the weighing platform to move the scale, as that could cause
damage to the scale.
Avoid chemical solvents. Clean the scale with water.
IP67
Degrees of protection against water: Protected against temporary submergence.
Degrees of protection against solid foreign objects: Dust-tight.
International Protection of IEC60529.
2
2. PARTS DESCRIPTION
r
y
Sealing diagrams
Weighing pan (SUS304)
Inner pan (ABS)
Screw to fix the pan support
Pan support
The pan support can be removed at times such as
when the scale is cleaned.
Weight sensor (Load cell)
Note: To maintain performance, do not touch
this sensor.
Bubble Spirit level
Adjust the leveling feet so that the bubble is in the
center of circle.
Switch cover
When this cover is opened, the
switches inside are revealed. Keep the cover closed
to maintain waterproof performance.
Battery cove
To maintain waterproof performance, close completel
this cover and secure by screws.
Leveling foot
Screw for
securing
Display and
switch
Screwdriver
Use this screwdriver to open
or close the battery cover.
CAL
Weighing pan
and
RESET
3
3. PREPARATION
y
r
3.1. Installing / Exchanging batteries
The batteries are not included with the product. Prepare 6 x “D” size (R20P or LR20)
dry-cell batteries.
When the
exchange the old batteries with new ones.
1. Loosen the screws for securing the
battery cover using the provided
screwdriver, and remove the battery
cover.
2. Remove all the old batteries from the
battery compartment.
3. Install six new batteries properly
according to the + and - indicators of
polarity in the battery compartment.
4. Close the battery cover, and tighten the
screws for securing the battery cover.
CAUTION -
Do not mix used and new batteries. Do not mix the different types
of batteries. That may cause damage to the batteries or the scale.
mark is displayed,
Batter
cover
Screw fo
securing
Six new
batteries
(R20P/ LR20
/ D size)
Old
batteries
- Take care of the polarity of batteries. The polarity marks are shown
in the battery compartment.
3.2. Setting up the scale
CAUTION Avoid installing the scale in direct sunlight, that may cause
discoloration or malfunctions. Place your SJ-WP on a firm weighing
table so that the scale is level. The scale will not perform accurately
when it is not level.
Place the scale on a firm surface and adjust the feet so that the bubble of the
sprit level is in the center of the circle.
OK NG
4
4.1. Display
Symbols
Symbol Description
4. DISPLAY AND SYMBOLS
STABLE
NET ◄
ZERO ◄ Turns on when zero is displayed.
Comparator LEDs
Weighing units
*
Battery indicator
Turns on when the weight value is stable.
Turns on when the NET weight is displayed.
(The tare operation is in progress.)
Turns on when the comparator results are displayed.
“Ib”, “oz”, “ozt” , “Ib-oz” , “tI-s” , “tI-h” , “tI-t” , “t” , “pcs” , “N” , “g”
and “kg” are available.
A selected unit is displayed.
Turns on while a switch is touched.
The battery indicator changes as the battery capacity
decreases, as shown below:
New Replace the batteries.
5
4.2. Operations and functions of switches
Switch Description
When the scale is turned off:
Touch the switch to turn ON the scale. The scale will be
automatically set to zero (power-on zero).
When the scale is turned on:
Touch and hold the switch to trun OFF the scale.
Tap the switch to ZERO the scale and display zero.
During measurement:
Touch and hold the switch to proceed to the calibration mode.
In the function setting mode to set the parameters:
Tap the switch to change the parameter of the selected item.
During measurement:
Tap the switch to choose a unit specified in the setting mode.
Touch and hold the switch to enter the comparator value
setting mode.
In the function setting mode to set the parameters:
Tap the switch to select the function item.
During measurement:
Tap the switch to tare the scale and display zero (net weight
display).
In the function setting mode to set the parameters:
Other than at the item "unit":
Tap the switch to store new parameters and return to the
weighing mode.
At the item "unit":
Tap the switch to select active / inactive for the displayed unit.
When the scale is turned off:
Touch the
TARE switch to enter the function setting mode.
Further to the above, continue to touch the TARE switch to
restore the function settings to the factory set values.
ON/OFF switch while touching and holding the
When the scale is turned on:
Pressing the switch forcibly turns off the scale.
When the scale is turned on:
By pressing the switch, the scale proceeds to the calibration mode.
6
5. OPERATION
5.1. Basic weighing operation
1. Touch the switch to turn the power ON.
All the symbols except the comparator LEDs are
displayed. When the weight values becomes
stable, the display turns off for a moment and
displays zero (power-on zero) with the weighing
unit used last before turning off.
2. Select a weighing unit using the UNITS switch.
See “6. SELECTING A WEIGHING UNIT” in
detail.
3. When the display doesn’t show zero, tap the
switch to set the display to zero.
4. When using a tare (container), place the
container on the weighing pan, and tap the
TARE switch to set the display to zero.
5. Place the object to be weighed on the weighing
pan or in the container, and wait for the STABLE
indicator to turn on and read the value.
6. Remove the object from the weighing pan.
Container
Weighing pan
Tap
Object to
be weighed
7. Touch and hold the switch. Then, the
scale displays “
” and turns the power off.
0ff
Touch & hold
5.2. Notes about operations
Power-on zero
If the weight is unstable at power ON, the scale displays “
touches the weighing pan, or check if there is strong wind or vibration.
The range for power-on zero is within ±50% (±10% for Legal for Trade models) of
the weighing capacity (kg) at the calibrated zero point. If the scale is powered on
with a load beyond this range, the scale displays “
the weighing pan.
ZERO and TARE
The
is stable.
The switch will zero the scale if the weight value is within ±2% of the
(as ZERO switch) and TARE switches work when the weight value
------
------
”. Remove the load from
”. Check anything
weighing capacity (kg) at the power-on zero point. The ZERO
(ZERO operation)
7
◄
indicator turns on.
The TARE switch will tare the scale and subtract the weight to zero as a tare
weight when the weight is a plus value. In this case the ZERO◄ and NET◄
indicators turn on. (TARE operation)
At the zero point, the net weight display shows the tare weight in negative, and the
ZERO◄ and NET◄ indicators turn on.
(Note: In some countries or areas, the ZERO
while the scale is tared.)
When the scale is tared, weighing range for net loads is reduced by the amount of
the tare weight.
When the ZERO operation is performed in the net weight display, the tare operation
previously done is cleared and the NET◄ indicator turns off.
(Note: In some countries or areas, the ZERO operation will not clear the TARE
operation. Tap the TARE switch after zeroing the scale with nothing
on the weighing pan.)
Auto power-off function
If no switch is touched or tapped and the stable indicator is displayed for a certain
period of time, the scale will automatically turn off. See the function setting “
to set the elapsed time to turn off.
When “ e” or “-e” is displayed (refer to “12.2. Error codes”), the auto power-off
function is enabled.
LCD backlight
The LCD backlight is controlled by the functions “
If no switch is touched or tapped and the weight display continues to be stable for a
certain period of time, the LCD backlight will automatically turn off. The elapsed time
to turn off is set by the function setting “
also selectable.
The function setting “
” adjusts the brightness of the backlight.
l-i
l-it
◄
indicator will not turn on
l-it
”. The backlight always on or off is
” and “
l-i
”.
poff
”
5.3. Weight display resolution
The weight display resolution is a ratio of the minimum display to the weighing
capacity.
The SJ-WP series has four types of weight display resolution, as shown below.
Low: 1/3,000
Normal: 1/6,000 or 1/7,500 (depending on the weighing capacity)
High: 1/12,000 or 1/15,000 (depending on the weighing capacity)
Maximum: 1/30,000
The factory setting is the normal resolution. Select the resolution according to your
own application in the function setting “
For details about the minimum display and the weighing capacity, refer to “13.
SPECIFICATIONS”.
The weight display resolution of the Legal for Trade models is fixed. The selection in
the function setting “
In the counting mode, the scale works with the maximum resolution regardless of
the weight display resolution selected in the function setting “
re5o
” is not available.
re5o
”.
re5o
”.
8
6. SELECTING A WEIGHING UNIT
6.1. Storing the weighing unit
1. Touch the ON/OFF switch while touching and holding the TARE switch in
order to display
2. Tap the UNITS switch to display
3. Tap the SAMPLE switch to display a unit. Tap the TARE switch to activate or
deactivate the unit. The indicator is displayed for each active unit.
4. Repeat step 3 for other units.
5. Tap the UNITS switch to move to the next function item when finishing the selection.
6. Tap the TARE switch to store new units. The scale returns to the weighing mode.
Tael (Hong Kong jewelry)
Tael (Taiwan)
Tola
Counting unit
Newton
Newtons is a value calculated as follows:
Newtons = (value in grams) x (9.80665 m/s2) / 1000
1.00000 g
1000.00 g
453.59237 g
28.349523125 g
31.1034768 g
-
37.7994 g
37.4290 g
37.5 g
11.6638038 g
-
See below
6.2. Selecting the weighing unit
In the weighing mode, tap the UNITS switch to select a weighing unit.
Each time the UNITS switch is pressed, the unit changes as shown below.
UNITS
Ib
oz
ozt
9
Ib-oz
tl
pcs
t
N
g
kg
7. COUNTING MODE
r
Determines a unit weight (the weight of one piece) from a known sample quantity, and
calculates how many pieces are on the weighing pan using the unit weight.
The unit weight is maintained even if the power is turned OFF.
1. Tap the UNITS switch to se lect “pcs”.
2. Tap the SAMPLE switch to enter the sample
unit weight storing mode.
The numerical value on the left indicates the
number of samples.
3. Tapping the SAMPLE switch allows you to
change the number of samples in the order 5
10 20 50 100 ESC 5.
• By tapping the UNITS switch when “
displayed, the scale exits the sample unit
weight storing mode to the count display.
4. When “-” appears at the right side of the number of
samples, tap the switch to zero the
scale.
If necessary, place a container on the weighing
pan, and tap the TARE switch. Confirm that
the right side of the number of samples shows
zero.
5. Place the correct number of samples on the
weighing pan or in the container.
6. Confirm that the STABLE indicator is turned on.
Tap the
the unit weight. Remove the samples. The
scale is set to count objects with this unit
weight.
UNITS switch to calculate and store
(“pcs” = pieces)
e5C
” is
Containe
Weighing
Samples
pan
Tap
Tap
Tap
Tap
Tap
Number of
samples
Confirm
zero
The total weight of samples should be more than
shown below, regardless of the number of samples.
SJ-3000WP: 2.5 g
SJ-6000WP: 5 g
SJ-15KWP: 12.5 g
SJ-30KWP: 25 g
If not, the display shows “
the display of step 5. Increase the number of
samples (step 3) and try again.
lo ut
” and returns to
7. Place the objects to be counted on the weighing pan.
10
8. COMPARATOR
The scale has three-, five- and seven-level comparators.
Each comparator mode compares the weight value against the preset limit values and
outputs the results using LEDs (yellow / green / red).
Note: When the unit is “lb-oz” or “tl”, this function can not be used.
• Five-level comparator mode:
Uses four comparator values
to compare the weight value
and outputs results in five
levels of LOLO, LO, OK, HI
and HIHI.
• Three-level comparator mode:
Uses two comparator values
(upper and lower limit values)
to compare the weight value
and outputs results in three
levels of LO, OK and HI.
• Seven-level comparator mode (portion weighing mode):
Uses six comparator
values to compare the
weight value and outputs
results in seven levels of
outside the lowest limit,
level 1 (LOLO), level 2
(LO), Level 3 (OK), level
4 (HI), level 5 (HIHI) and
outside the highest limit.
To use the comparator modes, the function settings “
specified and the comparator values must be set.
Using the function setting “
0: five-level comparator mode (Result LED blinks)
1: five-level comparator mode (Result LED lights)
2: three-level comparator mode (Result LED blinks)
3: three-level comparator mode (Result LED lights)
4: Seven-level comparator mode (Result LED blinks)
5: Seven-level comparator mode (Result LED lights)
Cp-l
”, select a comparator mode.
Cp-l
” and “Cp” must be
Using the function setting “
0: No comparison (comparator disabled).
1: To compare all data.
2: To compare all stable data.
3: To compare all data which are ≥ +5d or ≤ -5d.
4: To compare stable data which are ≥ +5d or ≤ -5d.
5: To compare all data which are ≥ +5d.
”, select comparison conditions.
Cp
11
6: To compare stable data which are ≥ +5d.
d = minimum display in kg (Refer to “13. SPECIFICATIONS”.)
In the counting mode, “d” is equal to the minimum weight display of kg mode.
8.1. The formula to compare
Comparison is performed using the formula listed below and the results are output.
The comparator values are common to the weighing and counting mode.
Ignore the decimal point when setting the comparator values.
Example for SJ-6000WP when the setting value is “001000”:
Normal resolution kg 1.000 kg 6.000 kg / 0.001 kg
High resolution kg 0.1000 kg 6.0000 kg / 0.0005 kg
Maximum resolution g 100.0 g 6000.0 g / 0.2 g
Low resolution oz 100.0 oz 210.0 oz / 0.1 oz
Normal resolution oz 10.00 oz 210.00 oz / 0.05 oz
High resolution oz 10.00 oz 210.00 oz / 0.02 oz
Counting mode 1000 pcs
The comparator values are maintained even if the power is turned OFF.
The scale does not judge magnitude relation among the comparator values.
Even if the wrong values are set, no error will be shown.
Displayed value < Level 1 lower limit, or “-e”
Level 1
lower limit
Level 2
lower limit
Level 3
lower limit
Level 3
upper limit
Level 4
upper limit
Display mode Limit value Capacity / Minimum display
≤ Displayed value <
≤ Displayed value <
≤ Displayed value ≤
< Displayed value ≤
< Displayed value ≤
Level 2
lower limit
Level 3
lower limit
Level 3
upper limit
Level 4
upper limit
Level 5
upper limit
( No LEDs on)
( Red LED on)
( Yellow LED on)
( Green LED on)
( Yellow LED on)
( Red LED on)
( No LEDs on)
13
8.2. Entering the comparator values
1. In the weighing mode, touch and hold the UNITS switch to enter the comparator
value setting mode.
2. Enter the comparator values using the following switches.
UNITS To shift the blinking digit to the right.
SAMPLE To increase the value of the blinking digit by one.
TARE To store the setting value and to go to the next limit setting or “
end
The minus sign can be set at the next digit of the least significant digit. The
SAMPLE
switch alter nates the minus sign on and off. The blinking “-” shows minus
and no sign shows plus.
3. When the setting is complete, the scale memorizes the values and displays “
end
then returns to the weighing mode. (At this time, power-on-zero is not performed.)
The SJ-WP series has an auto-tare function to be used with the comparator mode
enabled. If the weight values are in the OK range of comparator limits and stable for a
preset period of time, the scale will automatic 1 ally tare the weight and show zero.
In some countries or areas, the auto-tare function can not be used on the Legal for Trade
models and the selection in the function settings “at”, “
To use the auto-tare function, set the function settings below.
Cp1: Compare all weighing data (other settings may be used depending on
the application).
at 1: Auto-tare function enabled.
at-t 0 ~ 9:
Normal comparion “
Start with display zero after tare operation. Place or take away objects until the
comparison result will show OK. When the stable indicator is ON for a the preset
period of time specified in the function setting “
tare the weight, show zero and be ready for next weighing to repeat.
Negative comparison for take-away “
Take-away check weighing (negative comparison) is the way to compare the
negative weight while taking away objects from a container.
Set the function “
operation mode, the scale operates as “take-away the objects” “OK and stable”
“auto-tare” “take-away the objects” ······.
In this setting, the polarity of LOLO, LO, HI, and HIHI limit values are ignored and
the scale shows the comparator results as below.
Note: To start the take-away check weighing, be sure to use the TARE switch to
tare the weight of the container filled with objects. The ZERO switch may zero
the display, and the scale goes below the zero point by taking out the objects.
Then, the auto-tare function does not work.
When the function “
To start the auto-tare function, usually the container (filled wit h objects) will be placed on
the weighing pan and its weight must be tared using the TARE switch. When the
function “
When all load on the weighing pan is removed, the scale will return to the zero point
and the tare weight will be automatically cleared. If the scale does not return to the
Select the timing to tare automatically to avoid the wrong tare operation, for
example; too early to tare, to take a longer time to go to the next weighing.
Cp-p 0
”
at-t
” (example with “
at-f 1
Cp-p 1
Cp-p 1
” is selected, the scale will tare the initial (container) weight automatically.
” together with the auto-tare function enabled “
at-f 1
Tares the initial (container) weight.” is selected:
” and “
at-t
”, the scale will automatically
Cp-l 0
” is not available.
at-f
”)
at 1
”. In this
zero point, tap the switch to clear the tare weight.
15
10. CALIBRATION
Adjusts the scale for accurate weighing.
Calibrate the scale in the following cases.
When the scale is first installed.
When the scale has been moved.
When the ambient environment has changed.
For regular calibration.
Note: The Legal for Trade models can not be re-calibrated if they have been
sealed.
10.1. Calibration mode
The calibration mode has the following three functions.
• Gravity acceleration correction
• Calibration using a weight
• Restoring the factory set values
How to enter the calibration mode
Method 1:
1. Make sure that the scale is in the weighing mode.
2. Touch and hold the SAMPLE switch until the gravity acceleration value (in this
Note: The above operation is disabled for the Legal for Trade models.
Method 2:
example, “9.7985”) appears and release the switch.
1. Make sure that the scale is in the
weighing mode.
2. Loosen the two screws on the switch
cover and open the switch cover. The
calibration (CAL) switch is located inside.
3. Press the CAL switch. The scale
displays the gravity acceleration value (in
this example, “9.7985”).
Switch cover
CAL
switch
16
Method 1
Touch & hold
Calibration mode
Weighing mode
Power OFF
Method 2
Press
Gravity acceleration
correction
Calibration using
a weight
Restoring the factory
set values
or
To exit from the calibration mode, touch
the switch or Press the CAL
switch. The scale turns the power OFF.
Tap the UNITS switch
to select an item
and
Tap the TARE switch
to execute it.
10.2. Gravity acceleration correction
When the scale is first used or has been moved to another location, it should be
calibrated using a calibration weight.
But if a calibration weight is not available, the gravity acceleration correction will
compensate the scale. Change the gravity acceleration value stored in the scale to the
value of the area where the scale will be used. Refer to the gravity acceleration map at
the end of this manual.
Note: Gravity acceleration correction is not required when the scale will be
calibrated using a calibration weight at the place where it is to be used.
1. Refer to “10.1. Calibration mode” to enter the calibration
mode. The gravity acceleration value is displayed.
2. Tap the TARE switch to enter the gravity
acceleration value setting mode.
3. Change the displayed value using the following switches.
UNITS To shift the blinking digit to the right.
SAMPLE To increase the value of the blinking
digit by one.
4. Tap the TARE switch. The display shows “end” and
returns to the newly stored gravity acceleration value.
Tap
Tap
5. When calibration using a calibration weight is to be
performed, go to step 3 of “10.3. Calibration using a
weight”.
To finish the setting procedure, touch the switch
or the CAL switch. The scale returns to the weighing mode.
17
Tap
10.3. Calibration using a weight
Prepare a weight, preferably a weight with the same value as the weighing capacity of
the scale to be calibrated. Note that the calibration weight value can be changed.
1. Turn the power ON and warm up the scale for at least half an hour.
Change the function setting “
disable the auto power-off function.
2. Refer to “10.1. Calibration mode” to enter the
calibration mode. The gravity acceleration value is
displayed.
3. Tap the UNITS switch and the scale displays “
4. Tap the TARE switch, then “
Confirm that nothing is placed on the weighing pan and
wait for the STABLE indicator to turn on.
5. Tap the
point and displays the value of the calibration weight
(SPAN calibration).
The calibration weight value is equal to the weighing
capacity. (factory setting)
If SPAN calibration is not to be performed, turn the
power OFF to exit from the calibration procedure.
6. To calibrate with a weight different from the weighing
capacity, change the displayed value using the
following switches.
TARE switch. The scale calibrates the zero
UNITS To shift the digit that is blinking to the
right.
poff
” or place something on the weighing pan to
”.
Cal
Cal 0
”is displayed.
Tap
Tap
Tap
Tap
SAMPLE To increase the value of the blinking
digit by one.
Using a weight with the same value as the weighing
capacity is recommended. If other weights are used,
use one with a value greater than two-thirds of the
capacity.
7. Place the calibration weight with the same value as
displayed on the weighing pan, and wait for the
STABLE indicator to turn on.
8. Tap the TARE switch. T he scale calibrates SPAN and
“
” is displayed. Then, the display returns to “
end
To finish the procedure, touch the switch or
press the CAL switch. The scale turns the power
OFF.
Not e: If the scale will be moved to another location, set the gravity acceleration
value for the present location first and calibrate the scale using a weight.
Then, change the gravity acceleration value for the new location.
Cal
”.
Tap
18
10.4. Restoring the calibration data to the factory set values
If the gravity acceleration value or calibration data is changed unintentionally, restore
those values to the factory set values, as follows.
1. Refer to “10.1. Calibration mode” to enter the
calibration mode. The gravity acceleration value is
displayed.
Tap
2. Tap the UNITS switch twice to display “
ClrC
”.
twice
Tap
3. Tap the TARE switch to display “
“no” blinking.
4. Tap the SAMPLE switch.
“
ClrC no
blinking.
To cancel the restoring procedure, tap the
5. When “
TARE switch. The factory set values are restored
and “
Then, the display returns to “
To finish the setting procedure, touch the
switch or press the CAL switch. The scale turns
the power OFF.
” changes to “
switch. The display returns to step 2.
ClrC go
” is displayed.
end
” is displayed, tap the
ClrC go
ClrC
ClrC no
” with “go”
”
” with
Tap
Tap
19
11. FUNCTION SETTINGS
The scale has function settings to specify the scale performance.
The parameters set in the function settings are maintained even if the power is turned
OFF.
Item
11.1. Setting the parameters
1. Turn the power OFF.
Parameter
While
touching & holding
, Touch
2. Touch and hold the
the switch to turn the power ON. The
software version is displayed.
“∗∗∗” indicates the software version number.
3. After about one second, the item is displayed.
4. Change the item or parameter using the following switches.
UNITS To display the next item.
SAMPLE To increase the value of the blinking
digit by one (to change the parameter).
When the parameter is changed,
5. Tap the TARE switch to store the setting value.
After displaying “
mode.
To cancel the setting procedure without storing the value,
touch and hold the
end
TARE switch and touch
the STABLE indicator turns off.
”, the scale goes to the weighing
switch to turn off the scale.
Tap
When changed
Tap
20
11.2.
1. Turn the power OFF.
2. Touch and hold the TARE switch and touch the
Restoring the function settings to the factory set values
switch to turn the power ON and to display
the software version. Release the switch but
continue to touch the TARE switch until “
with “no” blinking is displayed.
3. Tap the SAMPLE switch.
“
Clrf no
4. When “
The factory set values are restored. After displaying
“
end
To cancel the restoring procedure, touch and hold the
” changes to “
Clrf go
”, the scale goes to the weighing mode.
switch to turn off the scale.
” is displayed, tap the TARE switch.
Clrf go
” with “go” blinking.
Clrf no
”
Tap
Tap
21
11.3. Function list
Item
Auto power-off
function
poff
Parameter
0
1
2
3
Description
Auto power-off function disabled
Turns off after 5 minutes
Turns off after 10 minutes
Turns off after 15 minutes
Turns the power OFF
automatically
Weight display
resolution
re5o
Weighing unit
Unit
Zero tracking
trc
Weighing
stability /
response
speed
Cond
Backlight
control
l-it
Brightness of
backlight
l-i
Decimal point
dp
Factory setting
SAMPLE Proceeds to the next unit
TARE
UNIT Proceeds to the next setting
4
5
0
1
2
3
0
1
0
1
2
3
4
0
1
2
3
4
5
0
1
2
3
4
0
1
Turns off after 30 minutes
Turns off after 60 minutes
1/3,000
1/6,000 or 1/7,500
1/15,000 or 1/12,000
1/30,000
Selects whether a unit is active or
inactive
Zero tracking function disabled
Zero tracking function enabled
Weak stability & quick response
Strong stability & slow response
Backlight always off
Backlight always on
Turns off 5 seconds after stabilizing
Turns off 10 seconds after stabilizing
Turns off 15 seconds after stabilizing
Turns off 30 seconds after stabilizing
Dark
Bright
Dot
Comma
Changes the
minimum display
See “6. SELECTING
A WEIGHING UNIT”
Tracks the zero drift
Response =
time from placing an
object on the pan to
turning on the stable
indicator
Sets the timing to
trun off the backlight.
Backlight turns on by
weight change or
switch operation.
Adjusts the brightness
of the backlight
22
Item
Comparator mode
Parameter
0
Description
Five-level (Result LED blinks.)
Cp-l
Comparison
conditions
Cp
Comparator
LED brightness
Cp-i
Normal/Negative
comparison
Cp-p
Auto-tare
function
at
Auto-tare timing
at-t
Auto-tare of the
initial weight
at-f
Factory setting
1
2
3
4
5
0
1
2
3
4
5
6
0
1
2
3
4
0
1
0
1
0
1
2
3
4
5
6
7
8
9
0
1
Five-level (Result LED lights.)
Three-level (Result LED blinks.)
Three-level (Result LED lights.)
Seven-level (Result LED blinks.)
Seven-level (Result LED lights.)
Comparator disabled
Compares all data
Compares all stable data
Compares all data of ≥ +5d or ≤ -5d
Compares stable data of ≥ +5d or ≤ -5d
Compares data of ≥ +5d
Compares stable data of ≥ +5d
Dark
Bright
Normal comparison
Negative comparison for take-away
check weighing
Auto-tare function disabled
Auto-tare function enabled
Immediately after OK and stable
0.5 second after OK and stable
1.0 second after OK and stable
1.5 seconds after OK and stable
2.0 seconds after OK and stable
2.5 seconds after OK and stable
3.0 seconds after OK and stable
4.0 seconds after OK and stable
5.0 seconds after OK and stable
10 seconds after OK and stable
Function disabled
Tares the initial (container) weight.
Sets comparator
mode
Sets comparison
conditions
d = minimum display
in kg
Adjusts LED
brightness of
comparison result
Refer to
“9. AUTO-TARE”
Refer to
“9. AUTO-TARE”
Timing to tare
automatically after
the comparison OK
and stable weight.
Automatic operation
23
12. MAINTENANCE
12.1. Notes on maintenance
Do not disassemble the scale. Contact your local A&D dealer if the scale needs
service or repair.
Use the original packaging for transportation.
Do not use organic solvents to clean the scale. Use a warm lint free cloth
dampened with a mild detergent.
Calibrate the scale periodically to maintain the weighing accuracy.
12.2. Error codes
Overload error
Indicates that an object beyond the weighing capacity
has been placed on the weighing pan.
Underload error
Power-on zero error
Unit weight error
CAL error
Low battery
Other
Remove the object from the weighing pan.
Indicates that the weight sensor receives a strong
upward force. Check if there is anything sandwiched
around the weighing pan. There is a possibility that the
weight sensor or internal circuit may have a problem.
Indicates that the power is turned on with a load beyond
the power-on zero range, or the weight value too
unstable to perform power-on zero. Remove the load, or
check if there is wind, vibration or anything touching the
weighing pan.
Indicates that total weight of samples is too light to set
the unit weight in the counting mode. Increase the
number of samples and try again
Indicates that the calibration procedure is canceled
because the calibration weight is too light.
Check that the weighing pan is installed properly and the
mass of the calibration weight is correct.
Indicates that the batteries have run out.
Replace them with new batteries.
There may be an internal malfunction.
(∗indicates an error number.)
Note: If the error persists or other errors occur, contact your local A&D dealer.
24
13. SPECIFICATIONS
Specifications
Model
Capacity 3kg 6kg 15kg 30kg
Minimum display “d”
Counting
Repeatability (SD) 0.5 g 1 g 2 g 5 g
Linearity ±1 g ±2 g ±5 g ±10 g
Sensitivity drift ±50ppm / °C (5°C~35°C / 41°F~95°F)
Display
Display update 20 times per second
Operating
temperature
Power 6 x R20P / LR20 / “D” size batteries
Battery life
(Approximately)
Pan size 232 (W) x 192 (D) mm / 9.13 (W) x 7.56 (D) in.
Dimensions 236(W) x 260 (D) x 148 (H) mm / 9.3 (W) x 10.2 (D) x 5.8 (H) in.
Mass Approximately 4 kg / 9 lb
Accessories This manual,Screwdriver
* Factory setting
260
148
SJ-3000WP SJ-6000WP SJ-15KWP SJ-30KWP
0.001 kg 0.002 kg 0.005 kg 0.01 kg
0.0005 kg* 0.001 kg* 0.002 kg* 0.005 kg*
0.0002 kg 0.0005 kg 0.001 kg 0.002 kg
0.0001 kg 0.0002 kg 0.0005 kg 0.001 kg
Number of samples: 5, 10, 20, 50 or 100 pieces
Maximum count: 120,000
Weight display: 7 segment LCD with Backlight,
Character height: 26 mm
Comparator LEDs: red / yellow / green / yellow / red
-10°C~40°C / 14°F~104°F, Less than 85%RH
5000 hours with alkaline cells at 20°C (LED & Backlight off)
2000 hours with manganese cells at 20°C (LED & Backlight off)
236
192
232
Unit: mm
155
180
25
Other weighing units
Model
Capacity
g
Minimum display
Capacity
lb
Minimum display
Capacity
oz
Minimum display
Capacity
ozt
lb-oz
Catty-tl
(HG)**
Catty-tl
(HJ)**
Catty-tl
(T)**
Tola
* Factory setting
** Catty-tael, HG: Hong Kong General / Singapore, HJ: Hong Kong Jewelry, T: Taiwan
Minimum display
Capacity
Minimum d i sp l a y0.1 oz 0.1 oz 0.1 oz 0.1 oz
Capacity
Minimu m d i s pl a y0.01 tl 0.1 tl 0.1 tl 0.1 tl
Capacity
Minimu m d i s pl a y0.01 tl 0.1 tl 0.1 tl 0.1 tl
Capacity
Minimu m d i s pl a y0.01 tl 0.1 tl 0.1 tl 0.1 tl
Capacity
Minimum display
SJ-3000WP SJ-6000WP SJ-15KWP SJ-30KWP
3000 g 6000 g 15000 g 30000 g
1 g 2 g 5 g 10 g
0.5 g* 1 g* 2 g* 5 g*
0.2 g 0.5 g 1 g 2 g
0.1 g 0.2 g 0.5 g 1 g
6.6 lb 13 lb 33 lb 66 lb
0.002 lb 0.005 lb 0.01 lb 0.02 lb
0.001 lb* 0.002 lb* 0.005 lb* 0.01 lb*
0.0005 lb 0.001 lb 0.002 lb 0.005 lb
0.0002 lb 0.0005 lb 0.001 lb 0.002 lb
105 oz 210 oz 520 oz 1050 oz
0.05 oz 0.1 oz 0.2 oz 0.5 oz
0.02 oz* 0.05 oz* 0.1 oz* 0.2 oz*
0.01 oz 0.02 oz 0.05 oz 0.1 oz
0.005 oz 0.01 oz 0.02 oz 0.05 oz
96 ozt 193 ozt 480 ozt 960 ozt