1-1. About This Manual .......................................................................................................................... 5
1-2. Features ........................................................................................................................................... 5
3-1. Before Use ..................................................................................................................................... 10
3-2. During Use ...................................................................................................................................... 11
3-3. After Use ........................................................................................................................................ 12
3-4. Power Supply ................................................................................................................................ 12
4. Display Symbols and Key Operation ................................................................................................ 13
7-2. Calibration Using the Internal Mass
7-3. Calibration Test Using the Internal Mass
7-4. Calibration Using an External Weight .......................................................................................... 20
7-5. Calibration Test Using an External Weight .................................................................................. 21
7-6. Correcting the Internal Mass Value 1 (Only for the FZ-CT Series)
7-7. Correcting the Internal Mass Value 2
8. Function Switch and Initialization ...................................................................................................... 24
8-1. Permit or Inhibit .............................................................................................................................. 24
8-2. Initializing the Balance ................................................................................................................... 26
9. Function Table .................................................................................................................................... 27
9-1. Structure and Sequence of the Function Table ........................................................................... 27
9-2. Display and Keys ........................................................................................................................... 27
9-3. Details of the Function Table ......................................................................................................... 28
9-4. Description of the Class "Environment, Display" ......................................................................... 32
9-5. Description of the Item "Data Output Mode" ............................................................................... 34
9-6. Description of the Item "Data Format".......................................................................................... 36
9-7. Data Format Examples ................................................................................................................. 38
9-8. Weighing Units ............................................................................................................................... 40
9-9. Storing Units ................................................................................................................................... 42
13. Password Function ............................................................................................................................ 62
13-1. Balance Software Version1.70 to 1.85 ......................................................................................... 62
13-2. Balance Software Version1.860 or later ...................................................................................... 62
13-3. Using the Password Function ...................................................................................................... 64
13-4. Entering the Password Before Weighing ..................................................................................... 65
13-5. Logging Out ................................................................................................................................... 67
13-6. Storing or Changing the Password .............................................................................................. 67
13-7. How to Change the Password ..................................................................................................... 68
20-3. Other Display ................................................................................................................................. 82
20-4. Asking for Repair ........................................................................................................................... 82
21. Options and Peripherals (Sold Separately) ...................................................................................... 83
24-2. Index ............................................................................................................................................... 90
4
1.Introduction
This manual describes how the FZ-CT / FX-CT series balance works and how to get the most out of it
in terms of performance. Read this manual thoroughly before using the balance and keep it at hand
for future reference.
※ Operation may vary depending on the software version of the balance. To check the software version of
your balance, refer to “18. How to Check Software Version of Balance”.
1-1.About This Manual
This manual consists of the following five parts:
Basic operation
Adapting to the environment
Functions
Interface and communication
Maintenance
1-2. Features
.................................
........
...........................................
.......
......................................
Describes precautions on handling the balance, balance
construction and basic balance operation.
Describes response adjustment and calibration (sensitivity
adjustment).
Describes various functions of the balance.
Describes the interface which transmits data and controls the
balance.
Describes maintenance, error codes, troubleshooting,
specifications and options.
Compact general-purpose balance, can be installed almost anywhere.
Large breeze break, provided as a standard accessory for more accurate weighing.
The breeze break can be removed with one touch, so cleaning is simple.
Casing construction, strong protection against dust and moisture.
Carat pans for weighing jewelry, provided as standard accessories.
Multiple weighing units with most of the common units used around the world.
The default settings of the unit of weight are the metric carat (ct) and grams (g). The unit can be
switched by pressing the
Standard RS-232C serial interface to communicate with a computer and to output the Good
Laboratory Practice (GLP) / Good Manufacturing Practice (GMP) compliant report.
A built-in clock and calendar that can add the time and date to the output data. (Changing the clock
setting can be limited to the administrator. Refer to “15. Password Function”.)
The password function can be used to restrict balance users or changes to the function settings.The key lock function disables the balance key operation and enables the balance to be operated
only by commands from external devices.
Statistical calculation mode to statistically calculate the weight data, and display or output the sum,
maximum, minimum, range (maximum-minimum), average, standard deviation and coefficient of
variation.
MODE key.
Comparator Indicators, displaying the comparison results.
Underhook for below weighing, provided as a standard accessory.
Hold Function, provided for weighing a moving object such as an animal.
As options, the USB interface (FXi-02), the Ethernet interface (FXi-08) and the built-in battery unit
(FXi-09) are available.
Connecting the USB interface, by cable to a Windows computer, allows transmission of the weight
data to Excel or Word. The Ethernet interface allows connection of the balance to a personal computer
through a LAN. The battery unit allows use of the balance where a power source is not available.
5
The AD-8920A remote display, that is sold separately, allows reading the weighing result in a place
away from the balance. It receives power supply directly from the balance instead of the AC adapter
when connected via RS-232C, so the cable arrangement is simplified.
1-3. Compliance
Compliancewith FCC rules
Please note that this device generates, uses and can radiate radio frequency energy. This device has
been tested and has been found to comply with the limits of a Class A computing device pursuant to
Subpart J of Part 15 of FCC rules. These rules are designed to provide reasonable protection against
interference when this device is operated in a commercial environment. If this unit is operated in a
residential area, it may cause some interference and under these circumstances the user would be
required to take, at his own expense, whatever measures are necessary to eliminate the interference.
(FCC = Federal Communications Commission in the U.S.A.)
6
2. Unpacking the Balance
2-1. Unpacking
The balance is a precision instrument. Unpack the balance carefully. Keep the packing material to be
used for transporting the balance in the future.
See the illustrations to confirm that everything is included.
Install the large breeze break when drafts are present
Top door
Large breeze break
Antistatic coating
Bubble spirit level
Keys
Display
AC adapter
AC adapter ID label
Leveling foot
Weighing pan
Pan support
Breeze break ring
Dust plate
Silver and black colored, one each
How to install the large breeze break
Follow the numbered sequence.
Main unit
Carat pan
Pull out the locking handles, and install the
large breeze break on the main unit.
Push in the locking handles.
Confirm that both of the locking handle
hooks are under the dust plate.
How to open the top door
If removing either of the securing screws
(3) on the top door, the top door may be
pivoted on the remaining screw.
RS-232C serial interface
Main unit rear side
The AC adapter plug is
Grounding terminal
Main unit cover
AX-FXi-31
protected against dust and
may be difficult to insert.
When inserting the plug,
turn the plug while pushing
on it.
AC adapter plug
AC adapter jack
7
Notes
Please confirm that the AC adapter type is correct for your local voltage and receptacle type.
Please use the dedicated AC adapter specified for the balance.
Do not use the AC adapter provided with the balance for other models or equipment with which the AC
adapter may not be compatible.
If you use the wrong AC adapter, the balance and other equipment may not operate properly.
8
2-2.Installing the Balance
Install the balance as follows:
1. Place the balance on a solid weighing table. Refer to "3. Precautions" for installing the balance.
2. Assemble the pan support, weighing pan, breeze break ring and breeze break, on the balance as
shown in the illustration on page 7.
3. Adjust the leveling feet to level the balance. Confirm it using the bubble spirit level.
4. Confirm that the AC adapter type is correct for the local voltage and power receptacle type.
5. Connect the AC adapter to the balance. Warm up the balance for at least one hour with nothing on the
weighing pan.
Note
The AC adapter plug is protected against dust and may be difficult to insert. When inserting
the plug, turn the plug while pushing on it.
9
3.Precautions
To get the optimum performance from the balance and acquire accurate weight data, note the
following:
3-1.Before Use
Install the balance in an environment where the temperature and humidity are not excessive. The best
operating temperature is about 20°C / 68°F at about 50% relative humidity.
Install the balance where it is not exposed to direct sunlight and it is not affected by heaters or air
conditioners.
Install the balance where it is free of dust.
Install the balance away from equipment which produces magnetic fields.
Install the balance in a stable place avoiding vibration and shock. Corners of rooms on the floor at the
ground level are best, as they are less prone to vibration.
The weighing table should be solid, free from vibration and drafts and as level as possible. (An
anti-vibration table or stone table is ideal)
Level the balance by adjusting the leveling feet and check it using the bubble spirit level.
Red circle
Bubble
OK NG
Bubble spirit level
Leveling foot
Make sure that the bubble is in the center of the red circle.
UP
DOWN
DOWN
Leveling foot
UP
When the bubble is off to the left :
Turn the leveling foot on the front right in the
clockwise direction.
When the bubble is off to the backward position :
Turn both leveling feet on the front in the
clockwise direction at the same time.
Ensure a stable power source when using the AC adapter.
Plug in the AC adapter and warm up the balance for at least one hour.
Calibrate the balance before use or after having moved it to another location. Refer to “7. Calibration
(Sensitivity adjustment/Calibration)”.
Caution: Do not install the balance where flammable or corrosive gas is present.
When the bubble is off to the right :
Turn the leveling foot on the front left in the
clockwise direction.
When the bubble is off to the forward position:
Turn both leveling feet on the front in the
counterclockwise direction at the same time.
10
3-2.During Use
Discharge static electricity from the material to be weighed
(hereinafter referred to as the sample). When a sample could
have a static charge, the weight data is influenced. Ground
the balance and try the following:
Eliminate static electricity using the optional AD-1683 DC
static eliminator.
Try to keep the ambient humidity above 45%RH.
Use a metal shield case for a charged sample.
Wipe charged plastic samples with a moistened cloth.
This balance uses a strong magnet as part of the balance
assembly, so please use caution when weighing magnetic
materials such as iron. If there is a problem, use the underhook
on the bottom of the balance to suspend the material away
from the influence of the magnet.
Eliminate the temperature difference between a sample and
the environment. When a sample is warmer (cooler) than the
ambient temperature, the sample will be lighter (heavier) than
the true weight. This error is due to a rising (falling) draft
around the sample.
Charged material
Material case
Grounding
Magnetic material
Make each weighing gently and quickly to avoid errors caused
by the evaporation of moisture from the sample or the
absorption of moisture by the sample. Use a pair of long
tweezers to place a sample on the pan to avoid placing a hand
inside the breeze break.
Do not drop things upon the weighing pan, or place a sample
on the pan that is beyond the balance weighing capacity.
Place a sample in the center of the weighing pan.
Do not use a sharp instrument such as a pencil to press the
keys. Use your finger only.
Press the RE-ZERO key before each weighing to eliminate possible errors.
Calibrate the balance periodically so as to eliminate possible errors.
Take into consideration the affect of air buoyancy on a sample when more accuracy is required.
Keep the balance interior free of dust and foreign materials.
The breeze break (The FZ-CT / FX-CT series) is provided as a standard accessory. An anti-static
treatment has been applied to the breeze break components, but they may be charged with static
electricity when they are unpacked or when the humidity is low. If the weighing value is unstable even
when there are no drafts or the balance has a problem with repeatability, remove the breeze break. Or
wipe the clear plates with a moistened cloth, use the optional AD-1683 DC static eliminator or apply an
anti-static spray.
11
3-3.After Use
Avoid mechanical shock to the balance.
Do not disassemble the balance. Contact the local A&D dealer if the balance needs service or repair.
Avoid dust and water so that the balance weighs correctly. Protect the internal parts from liquid spills
and excessive dust.
Do not use organic solvents or chemical cloth to clean the balance.
Clean the balance as described below:
Balance main unit,
Weighing pan
Breeze break
(standard accessory)
Use a soft, lint free cloth that is moistened with a mild detergent to clean.
The edge of the weighing pan is sharp. Use caution when cleaning the pan.
An antistatic treatment has been applied to the breeze break components.
Use a soft, dry, lint free cloth to clean. Cleaning with a cloth that is moistened
with water or a mild detergent repetitively, or washing with water, may
degrade the antistatic effect.
3-4.Power Supply
When the AC adapter is connected, the balance is in the standby mode if the standby indicator is on
(refer to "4. Display Symbols and Key Operation"). This is a normal state and does not harm the
balance. For accurate weighing, allow the balance to reach equilibrium with room temperature, and
then warm up the balance for at least one hour before use.
12
4. Display Symbols and Key Operation
Display symbols
Displays the weight data relative to the weighing capacity, in percentage, in
Key operation
Key operation affects how the balance functions. The basic key operations are:
"Press and release the key immediately" or "Press the key" ....................................................
= normal key operation during measurement
"Press and hold the key" .............................................................................................................
Each key, when pressed or when pressed and held, functions as follows:
the weighing mode (Capacity indicator)
Number of statistical data (Statistical calculation mode)
Displays the setting value in the function mode
Processing indicator
Stabilization indicator
Standby indicator
Processing indicator
Blinking
Response indicators
Comparator indicators
Hold mode indicator
Units
Interval output mode
standby indicator
Weight data
Internal output mode
active indicator
Key When pressed When pressed and held (for 2 seconds)
Turns the display on or off. The standby indicator is displayed when the display is turned off.
The weighing mode is enabled when the display is turned on. When the display is turned on,
the password input (
mode is enabled if the password function is not used. Refer to "13-4.Entering the Password Before Weighing".
This key is available anytime. Pressing the key during operation will interrupt the operation and
turn the display off.
In the weighing mode, turns the minimum
weighing value on or off.
In the counting or percent mode, enters the
sample storing mode.
Switches the weighing units stored in the function
table. Refer to "9-8. Weighing Units".
Cancels the operation when performing
settings. With the FZ-CT series, enters the
calibration mode using the internal mass.
(One-touch calibration)
Outputs the weight data to a printer or personal
computer using the RS-232C serial interface,
depending on the function table settings.
Confirms the operation when performing
settings.
) is displayed if the password function is used. The weighing
Enters the function table mode. Refer to "9. Function Table".
Enters the response adjustment mode.
Enters the calibration mode (calibration using
an external weight).
No function at the factory setting
By changing the function table:
Outputs "Title block" and "End block" for
GLP/GMP compliant report. (Refer to "10-2.
GLP Report".)
Sets the display to zero.
13
5.Weighing
The default settings of the unit of weight are the metric carat () and grams ( ).
Pressing the MODE key switches the displayed unit, between and .
To use other units, refer to “9-9. Storing Units” on page 37 to store the units.
5-1.Basic Operation (Gram Mode)
1 Place a container (Carat pan) on the weighing pan, if necessary.
Press the RE-ZERO key to cancel the weight (tare). The balance
displays 0.000 .
2 Place a sample on the pan or in the container.
3 Wait for the stabilization indicator to turn on. Read the weighing
value.
While the stabilization indicator is on, pressing the PRINT key
will output the weighing value, using the RS-232C serial interface.
4 Remove the sample and container from the pan.
Note
Peripheral equipment, that is sold separately , such as a
printer or a personal computer is required.
When other units of weight are stored as described in
“9-9. Storing Units”, pressing the MODE key switches the
displayed unit.
Press the SAMPLE key to turn on or off the minimum
weighing value.
When the ON:OFF key is pressed with a container placed on
the weighing pan, the balance displays 0.000 and
weighing is started,
Weighing pan
Carat pan
Sample
Remove the sample
Display off
Weighing pan
Container
14
Zero operation, tare operation and weighing range
When weighing is started
The balance will determine the reference zero point when the ON:OFF key is pressed to enter the
weighing mode.
Depending on the load condition at that time, the balance will automatically judge whether to perform zero or
tare operation. The condition for determining which is used is "power on zero range", and when power on
zero range is exceeded, the tare operation is performed.
Re-zero operation
By pressing the RE-ZERO key, the display can be set to zero.
The re-zero operation with the RE-ZERO key will automatically judge whether to perform zero or tare
operation.
The condition for determining which is used is "zero range", and when zero range is exceeded, the tare
operation is performed.
Weighing range
The range that the balance can weigh varies depending on the model.
When the total amount (net weight + tare weight) displayed for each model exceeds the maximum
display, E is displayed to indicate that the weighing range is exceeded.
When exceeded in the negative direction, -E is displayed.
Model
FZ-1200CT, FX-1200CT
FZ-700CT, FX-700CT
FZ-500CT, FX-500CT
Power on
zero range
Zero range -E display range
Approx. ±125 ct Approx. ±25 ct Approx. -100 ct
Approx. ±25 g Approx. ±5 g Approx. -20g or less
Approx. ±75 ct Approx. ±15 ct Approx. -100 ct
Approx. ±15 g Approx. ±3 g Approx. -20 g or less
Approx. ±50 ct Approx. ±10 ct Approx. -100 ct
Approx. ±10 g Approx. ±2 g Approx. -20 g or less
15
6.Response Adjustment
This function stabilizes the weighing value, reducing the influence on weighing that is caused by drafts
and vibration at the place where the balance is installed.
The function has three stages as follows and can be changed by simple key operation.
Indicator Parameter Response Stability
FAST Fast response, Sensitive value
MID.
SLOW
Slow response, Stable value
Operation
1 Press and hold the MODE key until RESPONSE is
displayed. And then, release the key.
2 Select a setting for the response adjustment by pressing
the MODE key. Either FAST , MID. or SLOW can
be selected.
3 After waiting for a moment, or when the PRINT key is pressed,
the balance displays eend . Then, it returns to the weighing
mode and displays the updated response indicator. The
response indicator remains displayed for a while.
Note
Setting the response rate automatically changes the
"Condition (Cond)" and "Display refresh rate (5pd)"
parameters of "Environment, Display (ba5fnc)" in the
Since the balance's resolution is high, weighing values may change due to gravity and daily environmental
changes. It is necessary to perform calibration (sensitivity adjustment) with the weight in order to keep the weighing
values from changing even if gravity or the environment changes.
Calibrate the balance if it is installed for the first time or relocated, or when the weighing values change significantly
in daily inspection.
Calibration means to adjust the weighing value of the balance using the reference weight or internal mass.
Calibration test means to weigh with the reference weight and compare how much the result deviates from the
reference value. (Calibration test does not perform adjustment.)
Calibration (Sensitivity adjustment)
Calibration using the internal mass ------------------- Using the internal mass, adjust the balance with a single
touch. (FZ-CT series)
Calibration using an external weight ----------------- Using an external weight, adjust the balance.
Calibration test (Sensitivity check)
Calibration test using an external weight ----------- Checks the accuracy of weighing using an external weight
(target weight) and outputs the result.
* No adjustment is made.
Notes
When calibrating, be sure to install the provided large breeze break.
Do not allow vibration, drafts, or temperature change to affect the balance during calibration.
Calibration adjusts the balance for accurate weighing.
Besides periodic calibration and before each use, perform calibration when:
The balance is installed for the first time.
The balance has been moved.
The ambient environment has changed.
To output the GLP/GMP compliant report using the RS-232C serial interface, set "GLP output ( info )"
of "Data output ( dout )". For details, refer to "9. Function Table". To display the GLP output, a printer
or personal computer is required.
Calibration test is available only when "GLP output ( info )" of "Data output ( dout )" is set to "1", "2" or
"3".
Note on using an external weight
The accuracy of an external weight will influence the accuracy of weighing. Select an appropriate
-0.0150 g to +0.0150 g FZ-700CT/ FX-700CT 150 g, 100 g, 50 g, 20 g
FZ-500CT/ FX-500CT 100 g, 50 g, 20 g
The calibration weight in bold type: factory setting.
The weight value can be adjusted within the range above.
Display
This indicator means "the balance is measuring calibration data".
Do not allow vibration or drafts to affect the balance while this indicator is
displayed.
17
7-2.Calibration Using the Internal Mass
This function calibrates the balance using the internal mass. The only operation required is to press
the CAL key. (The FX-CT series does not have this function.)
Notes
Level the balance by adjusting the leveling feet and confirm that the bubble is in the center of
the spirit level.
Poor level adjustment may cause a calibration error while using the internal mass.
(One-Touch Calibration, Only for the FZ-CT Series)
Operation
1 Connect the AC adapter and warm up the balance for at least one hour with nothing on the weighing
pan.
2 Press the CAL key.
3 The balance displays Calin and automatically performs calibration using the internal mass. Do not
allow vibration or drafts to affect the balance.
4 The balance displays end after calibration. If the "GLP output (info)" parameter of the function
table is set, the balance displays glp and outputs the calibration report using the RS-232C
interface. Refer to “GLP output (info)” in "10-2. GLP Report".
5 The balance automatically returns to the weighing mode after calibration.
About the internal mass
The internal mass may change due to corrosion or other damage caused by the operating
environment, or due to aging. The internal mass value can be corrected using an external weight. To
maintain the weighing accuracy, perform the internal mass value correction periodically. Refer to "7-4.
Correcting The Internal Mass Value 1" and "7-7. Correcting The Internal Mass Value 2".
18
7-3.Calibration Test Using the Internal Mass
This function tests the balance accuracy using the internal mass.
Calibration test does not perform adjustment.
This is available only when the "GLP output (info)" parameter is set to "1" or "2".
(Only for the FZ-CT Series)
Operation
1 Connect the AC adapter and warm up the balance at least one hour
with nothing on the pan.
2 Refer to "9. Function Table" to set "GLP output (info)"
to "1" or "2".
3 Press and hold the CAL key until CC in is
displayed, then release the key.
4 The balance measures the zero point.
Prevent vibration and drafts to affect the balance.
5 The measured zero point data is displayed.
6 The balance measures the internal mass.
Prevent vibration and drafts to affect the balance.
7 The value of the internal mass is displayed.
The normal range of the value is as follows:
If the displayed value is within the normal range, it indicates
that the calibration sensitivity is correctly adjusted by the
internal mass.
Press and hold
Release
Model Internal mass Normal range
FZ-1200CT 200.000 g
±0.002 g FZ-700CT
100.000 g
FZ-500CT
8 If GLP output is set, glp is displayed and the
calibration test report is output using the RS-232C
interface. Refer to "GLP output (info)" of the function
table, "10-2. GLP Report".
end is displayed after the calibration.
9 The balance automatically returns to the weighing mode.
GLP output
19
7-4.Calibration Using an External Weight
This function calibrates the balance using an external weight.
Operation
1 Connect the AC adapter and warm up the balance
for at least one hour with nothing on the pan.
2 Press and hold the CAL key until Cal out is
displayed, and then release the key.
3 The balance displays Cal 0 .
If you want to change the calibration weight (a list of usable
weights is shown on page 17), press the SAMPLE key
and proceed to step 4.
If you use the calibration weight value stored in the
balance, proceed to step 5.
4 Specify the calibration weight value as follows:
SAMPLE key To switch between the calibration weight
selection mode (all of the segments
blinking) and the weight value adjustment
mode (the last three digits blinking).
RE-ZERO (+) / MODE (-) keys
To select the calibration weight to use or
adjust the value. (Refer to page 17.)
PRINT key To store the new calibration weight value.
Even if the AC adapter is removed, the data
is maintained in non-volatile memory.
CAL key To cancel the operation and return to Cal 0 .
5 Confirm that there is nothing on the pan and press the PRINT key.
The balance measures the zero point. Do not allow vibration or
drafts to affect the balance. The balance displays the calibration
weight value.
Place the weight
Press and hold
Release
Release
Select a weight
Example:
New value
50.0120g
6 Place the displayed calibration weight on the pan and press
the PRINT key. The balance measures the calibration weight.
Do not allow vibration or drafts to affect the balance.
7 The balance displays end . Remove the weight from the pan.
8 When the GLP output is set, the balance displays glp and
outputs the calibration report.
9 The balance automatically returns to the weighing mode.
Remove
10 Press the MODE key to select the gram mode. Place the
calibration weight on the pan and confirm that the value
GLP output
displayed is within ±0.002 g of the calibration weight value. If it
is not within the range, check the ambient conditions such as
drafts and vibration, also check the weighing pan. Then, repeat
steps 1 to 9.
20
S
R
7-5.Calibration Test Using an External Weight
This function tests the balance weighing accuracy using an external weight and outputs the result.
This is available only when the “GLP output (info)” parameter is set to "1" , "2" or "3". (Calibration test
does not perform calibration.) .
Operation
1 Connect the AC adapter and warm up the balance
for at least one hour with nothing on the pan.
Press and hold
2 Press and hold the CAL key until CC out is
displayed, and then release the key.
3 The balance displays CC 0 .
If you want to change the calibration weight (a list of usable
weights is shown on page 17), press the SAMPLE key
and proceed to step 4.
If you use the calibration weight value stored in the
balance, proceed to step 5.
4 Specify the calibration weight value as follows:
SAMPLE key To switch between the calibration weight
selection mode (all of the segments
blinking) and the weight value adjustment
mode (the last three digits blinking).
RE-ZERO (+) / MODE (-) keys
To select the calibration weight to use or
adjust the value. (Refer to page 17.)
PRINT key To store the new calibration weight value.
Even if the AC adapter is removed, the
data is maintained in non-volatile memory.
CAL key To cancel the operation and return to CC 0 .
Release
Release
elect a weight
Example:
New value
50.0120g
5 Confirm that there is nothing on the pan and press
the PRINT key. The balance measures the zero
point and displays the measured value. Do not allow
vibration or drafts to affect the balance. The balance
displays the target weight value.
6 Place the displayed target weight on the pan and press
the
PRINT key. The balance measures the target weight
and displays the measured value. Do not allow vibration
or drafts to affect the balance.
7 The balance displays end . Remove the weight from
the pan.
8 When the GLP output is set, the balance displays glp
and outputs “Calibration Test Report”.
9 The balance automatically returns to the weighing mode.
Target weight
emove
GLP output
21
P
r
7-6.Correcting the Internal Mass Value 1
The FZ-CT series balance can correct the internal mass value using an external weight.
Method 1: Calibrate the balance using the internal mass. Measure the external weight.
Calculate the correction value. And store it in the balance.
The adjustable range of the internal mass value is as follows:
Model Correction Reference Value Adjustable Range
FZ-1200CT 200 g
FZ-500CT
100 g
(Only for the FZ-CT Series)
-0.0150 g to +0.0150 g FZ-700CT
Operation
Example : Correcting the weight value by +0.0030 g in 200 g using the FZ-1200CT.
If correcting the weight value by +0.0030 g in 100 g, use the correction value of +0.0060 g as
the correction reference value is 200 g.
200.0000g 200.0030g
1 Perform calibration using the internal mass (one-touch calibration).
Place the external weight on the weighing pan to obtain the correction value.
Correct the internal mass by +0.0030g in 200g.
Calibrate with the corrected internal mass.
The same mass
Corrected mass value
2 With the factory setting, the balance cannot correct the internal mass
value. Refer to "8. Function Switch And Initialization" and set the
internal mass value correction switch to 1.
3 In the weighing mode, press and hold the SAMPLE key to
display ba5fnc .
4 Press the SAMPLE key several times until C5 in 1 is displayed.
If C5 in 1 is not displayed, perform step 2.
5 Press the PRINT key.
Correct the internal mass value using the following keys:
RE-ZERO key
MODE key
PRINT key
CAL key
6 Press the CAL key to return to the weighing mode.
7 Press the CAL key to perform calibration using the internal mass.
8 Confirm that the internal mass value has been corrected properly by
placing the external weight on the weighing pan.
If the internal mass value has not been corrected properly, repeat the
procedure to adjust the correction value.
...
........
........
...........
To increase the value by one.
To decrease the value by one.
To store the correction value and
display the next menu item of the
function table.
To cancel the operation and return to the
weighing mode.
The balance displays
-0.0150 g after +0.0150 g.
Press and hold
ess several
times
22
A
7-7.Correcting the Internal Mass Value 2
The FZ-CT series balance can correct the internal mass value using an external weight.
Method 2: Calibrate the balance using the external weight. The balance performs automatic
calibration using the internal mass, corrects the internal mass value and stores it in the
balance. The corrected mass value is maintained in non-volatile memory even if the AC
adapter is removed.
The usable calibration weights and the adjustable range are as follows:
Model Correction Reference Value Adjustable Range
FZ-1200CT 250 g, 200 g, 100 g, 50 g, 20 g
FZ-500CT 100 g, 50 g, 20 g
The calibration weight in bold type: factory setting.
The calibration weight value can be adjusted within the range above.
(Only for the FZ-CT Series)
-0.0150 g to +0.0150 g FZ-700CT 150 g, 100 g, 50 g, 20 g
Operation
1 With the factory setting, the balance cannot correct the internal mass
value. Refer to "8. Function Switch And Initialization" and set the
internal mass value correction switch to 1.
Press and hold
2 In the weighing mode, press and hold the SAMPLE key to
display ba5fnc .
3 Press the SAMPLE key several times until C5 in 21 is
displayed. If C5 in 2 is not displayed, perform step 1.
4 Press the PRINT key to display Cal 0 .
Refer to "7-4. Calibration Using an External Weight", to
perform the calibration.
5 After calibration, remove the weight.
The balance displays . Press the PRINT key.
6 Cal. is displayed and the balance corrects the internal
mass value automatically
7 After correcting the internal mass value, the balance
displays Cal.
automatically using the corrected internal mass value.
8 The balance displays end and returns to the weighing
mode.
9 Confirm that the internal mass value has been
corrected properly by placing the external weight
used for the correction procedure on the
weighing pan.
If the internal mass value has not been corrected
properly, repeat the procedure to adjust the
correction value.
in
and performs the calibration
djusting the internal mass value
Calibrating with internal mass
Press several times
Calibration using an
external weight
Place the weight
Remove the weight
23
8. Function Switch and Initialization
8-1. Permit or Inhibit
The balance stores parameters that must not be changed unintentionally. There are two switches for
the purpose of protecting these parameters. Each switch can select either “permit” or “inhibit”.
“Inhibit” protects parameters against unintentional operations.
Function selection can be restricted with the password function. When using balances with software
version 1.860 or later, the password function can be used to restrict calibration (sensitivity adjustment)
functions.
The switches for function selection" are as follows:
FZ-CT series FX-CT series
Function switch
- Function table
- Calibration using the internal mass
- Calibration using an external weight
- Internal mass correction
- Function table
- Calibration using an external weight
Operation
1 Press the ON:OFF key to turn off the display.
2 While pressing and holding the PRINT key and the SAMPLE key, press the ON:OFF key.
The balance displays p5 .
3 Press the PRINT key. Then the balance displays the function switches.
4 Set the switches using the following keys.
SAMPLE key
RE-ZERO key
PRINT key
CAL key
....
...
........
...........
To select the switch to change the parameter. The selected switch blinks.
To change the parameter of the switch selected.
0 : To inhibit changes. 1 : To permit changes
To store the new parameter and return to the weighing mode.
To cancel the operation ( Clr is displayed.)
To return to the weighing mode, press the CAL key once again.
24
Switches (FZ-CT series)
(The display shown left indicates the factory settings)
Function table
To inhibit changes on the function table
To permit changes on the function table
Calibration using the internal mass (One-touch calibration)
Software version 1.860 or later
Software version 1.43 to 1.85
To inhibit calibration using the internal mass
To inhibit calibration using the internal mass
1
*
To inhibit calibration using the internal mass when a user
() or a guest () is logged in.
To permit calibration using the internal mass
Calibration using an external weight
Software version 1.43 to 1.85
Software version 1.860 or later
To inhibit calibration using an external weight
To inhibit calibration using an external weight
1
*
To inhibit calibration using the external mass when a user
() or a guest () is logged in.
To permit calibration using an external weight
Not used
Internal mass value correction
To inhibit correction
To permit correction
* 1 The operation is permitted when logged in as an administrator ().
Switches (FX-CT series)
(The display shown left indicates the factory settings)
Function table
To inhibit changes on the function table
To permit changes on the function table
Not used
Calibration using an external weight
To inhibit calibration using an external weight
To permit calibration using an external weight
Not Used
Not Used
Software version 1.43 to 1.85
Software version 1.860 or later
1
*
To inhibit calibration using the external mass when a user
To inhibit calibration using an external weight
() or a guest () is logged in.
25
8-2.Initializing the Balance
This function restores the following parameters to factory settings.
Calibration data
Function table
Sample unit mass value (counting mode), 100% reference mass value (percent mode)
External weight value
Function switch settings
Statistical data
Internal mass value (Only for the FZ-CT series)
Note
Be sure to calibrate the balance after initialization.
Operation
1 Press the ON:OFF key to turn off the display.
While pressing and holding the PRINT key and the SAMPLE key,
press the ON:OFF key. The balance displays p5 .
2 Press the SAMPLE key to display Clr .
3 Press the PRINT key to displayClr no .
(To cancel the operation, press the CAL key.)
4 Press the RE-ZERO key to displayClr go .
5 Press the PRINT key to initialize the balance.
The balance will automatically return to the weighing mode.
Press and hold Press
26
9.Function Table
The function table reads or rewrites the parameters that are stored in the balance. These parameters
stored, even if the AC adapter is removed, are maintained in non-volatile memory.
9-1.Structure and Sequence of the Function Table
The function table menu consists of two layers. The first layer is the "Class" and the second layer is the
"Item". Each item stores a parameter.
Example
This example sets "Auto print mode A ()" for "Data output mode ()" and "10 digits ()"
for "Auto print difference ()".
Start, From weighing mode
Press several times
Class
Press
End, To weighing mode
Press and hold
Item
Press one time
Press one time
Press two times
Press two times
Item : "Data output mode"
Parameter : "Auto print mode A"
Item : "Auto print difference"
Parameter : "10 digits"
Press one time
9-2. Display and Keys
Display/Key Description
The symbol indicates that the parameter displayed is in effect.
When pressed and held (for approx. 2 seconds) in the weighing mode, enters the
function table mode.
Selects the class or item in the function table mode.
Changes the parameter.
When a class is displayed, moves to an item in the class.
When an item is displayed, stores the new parameter and displays the next class.
When an item is displayed, cancels the new parameter and displays the next class.
When a class is displayed, exits the function table mode and returns to the weighing
mode.
27
9-3.Details of the Function Table
Class Item and Parameter Description
0 Fast response, sensitive value
1
2 Slow response, stable value
0 Stable range is ±1 digit
1
2 Stable range is ±3 digits
0 OFF
1 ON
The stabilization indicator illuminates when
the display fluctuation is within the range.
With "
Holds the display when stable in animal
weighing mode. With "
turns on.
0 OFF
1 Normal
2 Strong
Keeps zero display by tracking zero drift.
3 Very strong
0 Approx. 5 times/second
1
Approx.
10 times/second
0 Point (.)
1 Comma (,)
0 OFF
1 ON
0 OFF
1 ON (10 minutes)
0 Displays
Does not display
1
0 Does not sound
1 Sounds
Refer to "9-10. Clock and
calendar function".
Period to refresh the display
Decimal point format
Turns on the weighing mode display when
the AC adapter is connected.
Turns off the display after 10 minutes of
inactivity.
Select whether or not to sound the beep
when operating on keys.
Confirms and sets the time and date. The
time and date are added to output data.
0 No comparison
1
Comparison, excluding "near zero" when stable value or overloaded
2
Comparison, including "near zero" when stable value or overloaded