A&D FZ-300GD, FZ-600CT, FZ-1200GD, FZ-2000GD, FZ-3000GD Instruction Manual

...
FZ-CT / FZ-GD SERIES FX-CT / FX-GD SERIES
FZ-300CT / FZ-600CT
F
Z-120GD / FZ-200GD / FZ-300GD
F
Z-1200GD / FZ-2000GD / FZ-3000GD
F
X-300CT / FX-600CT
FX-120GD / FX-200GD / FX-300GD
F
X-1200GD / FX-2000GD / FX-3000GD
Jewelry Balance
INSTRUCTION MANUAL
1WMPD4001554A
© 2007 A&D Company Ltd. All rights reserved.
No part of this publication may be reproduced, transmitted, transcribed, or translated
into any language in any form by any means without the written permission of A&D
Company Ltd.
The contents of this manual and the specifications of the instrument covered by this
manual are subject to change for improvement without notice.
Windows, Word and Excel are the registered trademarks of the Microsoft Corporation.
1
Contents
Basic Operation
1. INTRODUCTION .............................................................................................................................3
1-1 About This Manual .....................................................................................................................3
1-2 Features.....................................................................................................................................3
1-3 Compliance................................................................................................................................4
2. UNPACKING THE BALANCE..........................................................................................................7
2-1 Unpacking..................................................................................................................................7
2-2 Installing the Balance.................................................................................................................9
3. PRECAUTIONS.............................................................................................................................10
3-1 Before Use...............................................................................................................................10
3-2 During Use...............................................................................................................................10
3-3 After Use .................................................................................................................................. 11
3-4 Power Supply...........................................................................................................................11
4. DISPLAY SYMBOLS AND KEY OPERATION ...............................................................................12
5. WEIGHING ....................................................................................................................................13
5-1 FZ-CT / FX-CT Series Balance................................................................................................13
5-2 Basic Operation (FZ-CT / FX-CT Series Balances) .................................................................13
5-3 FZ-GD / FX-GD Series Balances.............................................................................................14
5-4 Basic Operation (FZ-GD / FX-GD Series Balances) ................................................................14
ADAPTING TO THE ENVIRONMENT
6. RESPONSE ADJUSTMENT..........................................................................................................15
7. CALIBRATION ...............................................................................................................................16
7-1 Calibration Mode......................................................................................................................16
7-2 Calibration Using the Internal mass (One-Touch Calibration, only for FZ-CT/GD series) ........17
7-3 Calibration Using an External Weight ......................................................................................18
7-4 Calibration Test Using an External Weight...............................................................................20
8. FUNCTION SWITCH AND INITIALIZATION..................................................................................22
8-1 Permit or Inhibit........................................................................................................................22
8-2 Initializing the Balance .............................................................................................................23
FUNCTIONS
9. FUNCTION TABLE ........................................................................................................................24
9-1 Structure and Sequence of the Function Table ........................................................................24
9-2 Display and Keys .....................................................................................................................24
9-3 Details of the Function Table....................................................................................................25
9-4 Description of the Class “Environment, Display”......................................................................28
9-5 Description of the Item “Data Output Mode” ............................................................................30
9-6 Description of the Item “Data Format”......................................................................................31
2
9-7 Data Format Examples ............................................................................................................33
9-8 Units ........................................................................................................................................35
9-9 Storing Units ............................................................................................................................38
9-10 Counting Mode (PCS)............................................................................................................39
9-11 Percent Mode (%) ..................................................................................................................41
9-12 Description of the Item “Application Function” ....................................................................... 42
9-13 Statistical Calculation Mode...................................................................................................43
9-14 Comparator Function .............................................................................................................47
9-15 Clock and Calendar Function (Only for the FZ-CT/FZ-GD series) ......................................... 48
10. ID NUMBER AND GLP REPORT ................................................................................................50
10-1 Setting the ID Number ...........................................................................................................50
10-2 GLP Report............................................................................................................................50
11. UNDERHOOK .............................................................................................................................53
12. PROGRAMMABLE-UNIT ............................................................................................................54
RS-232C SERIAL INTERFACE
13. RS-232C SERIAL INTERFACE ...................................................................................................55
14. CONNECTION TO Peripheral EQUIPMENT...............................................................................56
14-1 Connection to the AD-8121B Printer......................................................................................56
14-2 Connection to a Computer.....................................................................................................57
14-3 Using Windows Communication Tools (WinCT).....................................................................57
15. COMMANDS ...............................................................................................................................59
15-1 Command List .......................................................................................................................59
15-2 Acknowledge Code and Error Codes ..................................................................................... 60
15-3 Settings Related to RS-232C.................................................................................................61
MAINTENANCE
16. MAINTENANCE ..........................................................................................................................62
17. TROUBLESHOOTING ................................................................................................................63
17-1 Checking the Balance Performance and Environment ..........................................................63
17-2 Error Codes ...........................................................................................................................64
17-3 Asking For Repair ..................................................................................................................65
18. OPTIONS ....................................................................................................................................66
19. SPECIFICATIONS.......................................................................................................................68
20. EXTERNAL DIMENSIONS..........................................................................................................72
21. TERMS/INDEX............................................................................................................................74
3
1. INTRODUCTION
This manual describes how the FZ-CT/GD and FX-CT/GD series balances work and how to get the most out of them in terms of performance.
Read this manual thoroughly before using the balance and keep it at hand for future reference.
1-1 About This Manual
This manual consists of the following five parts:
Basic operation ........................ Describes precautions on handling the balance, balance
construction and basic balance operation.
Adapting to the environment.... Describes response adjustment and calibration.
Functions ................................. Describes various functions of the balance.
RS-232C serial interface.......... Describes the interface which transmits data and controls the balance.
Maintenance ............................ Describes maintenance, error codes, troubleshooting,
specifications and options.
1-2 Features
Compact general-purpose balance, can be installed almost anywhere. Casing construction, strong protection against dust and moisture. Multiple weighing units with most of the common units used around the world.
The default settings of the unit of weight for the FZ-CT and FX-CT series balance are the metric carat (ct) and grams (g). The unit can be switched by pressing the MODE key.
The default setting of the unit of weight for the FZ-GD and FX-GD series balance is grams (g).
Standard RS-232C serial interface to communicate with a computer and to output the Good
Laboratory Practice (GLP) data
Statistical calculation mode to statistically calculate the weight data, and display or output the
sum, maximum, minimum, range (maximum-minimum), average, standard deviation and coefficient of variation.
Comparator Indicators, displaying the comparison results. Small breeze break, provided for the FZ-GD series (all models) and FX-120GD/200GD/300GD; and
large breeze break, provided for the FZ-300CT/600CT and FX-300CT/600CT for more accurate weighing.
Underhook, provided for suspended weighing. Hold Function, provided for weighing a moving object such as an animal.
As options, the USB interface (FXi-02), the Ethernet interface (Fxi-08) and the built-in battery
unit (Fxi-09) are available.
4
1-3 Compliance
Compliance with FCC Rules
Please note that this device generates, uses and can radiate radio frequency energy. This device has been tested and has been found to comply with the limits of a Class A computing device pursuant to Subpart J of Part 15 of FCC rules. These rules are designed to provide reasonable protection against interference when this device is operated in a commercial environment. If this unit is operated in a residential area, it may cause some interference and under these circumstances the user would be required to take, at his own expense, whatever measures are necessary to eliminate the interference. (FCC = Federal Communications Commission in the U.S.A.)
Compliance with EMC Directives
This device features radio interference suppression and safety regulation in compliance
with the following Council Directives Council directive 89/336/EEC EN61326 EMC directive Council directive 73/23/EEC EN60950 Safety of Information Technology Equipment
The CE mark is an official mandatory European marking.
Please note that any electronic product must comply with local laws and regulations when sold or used anywhere outside Europe.
5
6
7
2. UNPACKING THE BALANCE
2-1 Unpacking
The balance is a precision instrument. Unpack the balance carefully. Keep the packing
material to be used for transporting the balance in the future.
The packing contents depend on the balance model. See the illustrations to confirm that
everything is contained.
FZ-GD / FX-GD series balances
Note
Please confirm that the AC adapter type is correct for your local voltage and receptacle type.
The AC adapter plug is protected against dust and may be difficult to insert. When inserting the plug, turn the plug while pushing on it.
AC adapter
AC adapter ID label
RS-232C serial interface
Main unit rear side
AC adapter jack
Grounding terminal
AC adapter plug
Keys
Weighing pan
Main unit
Display
Bubble spirit level
Pan support
Main unit cove
r
How to assemble the breeze break (For the FZ-GD series (all models) and FX-120GD/200GD/300GD)
Assemble the breeze break as shown below. Follow the numbered sequence.
4
3
1
2
Leveling foot
8
FZ-CT / FX-CT series balances
Be sure to install the breeze break before use.
Note
The large breeze break is delivered in a separate package.
Note
Please confirm that the AC adapter type is correct for your local voltage and receptacle type.
The AC adapter plug is protected against dust and may be difficult to insert. When inserting the plug, turn the plug while pushing on it.
Large breeze break
Carat pan
(Black and silver colored,
one each)
AC adapter
AC adapter ID label
RS-232C serial interface
Main unit rear side
AC adapter jack
Grounding terminal
AC adapter plug
Keys
Weighing pan
Main unit
Display
Main unit cove
r
Follow the numbered sequence.
Bubble spirit level
Leveling foot
Pan support
Breeze ring
Breeze break lower plate
Confirm that the handle's hooks (each side) are under the breeze break lower plate
2
1
1
3
3
How to install the breeze break
Pull out the handles and place the breeze break on the balance main unit. Push in the handles. To open the top door to place a sample: Remove a screw on either side of the top. The top door rotates to open with the other screw as a pivot point.
1
1
2 3
9
2-2 Installing the Balance
Install the balance as follows:
1. Place the balance on a solid weighing table. Refer to “3. PRECAUTIONS” for installing the balance.
2. FZ-GD series (all models) and
FX-120GD /200GD /300GD
Assemble the pan support, weighing pan and small breeze break, on the balance as shown in
the illustration on page 7.
FX-1200GD /2000GD /3000GD Assemble the pan support and weighing pan on the balance as shown in the illustration on page 7.
FZ-300CT /600CT and FX-300CT /600CT Assemble the breeze ring, pan support, weighing pan and large breeze break on the balance as
shown in the illustration on page 8.
3. Adjust the leveling feet to level the balance. Confirm it using the bubble spirit level.
4. Confirm that the AC adapter type is correct for the local voltage and power receptacle type.
5. Connect the AC adapter to the balance. Warm up the balance for at least 30 minutes with nothing on the weighing pan.
Note The AC adapter plug is protected against dust and may be difficult to insert. When inserting the plug, turn the plug while pushing on it.
10
3. PRECAUTIONS
To get the optimum performance from the balance and acquire accurate weight data, note the
following:
3-1 Before Use
Install the balance in an environment where the
temperature and humidity are not excessive. The best operating temperature is about 20°C / 68°F at about 50% relative humidity.
Install the balance where it is not exposed to direct
sunlight and it is not affected by heaters or air conditioners.
Install the balance where it is free of dust.
Install the balance away from equipment which
produces magnetic fields.
Install the balance in a stable place avoiding vibration
and shock. Corners of rooms on the first floor are best, as they are less prone to vibration.
The weighing table should be solid and free from
vibration, drafts and as level as possible.
Level the balance by adjusting the leveling feet and
confirm it using the bubble spirit level.
Ensure a stable power source when using the AC
adapter.
Plug in the AC adapter and warm up the balance for at
least 30 minutes.
Calibrate the balance before use or after having
moved it to another location.
Caution
Do not install the balance where flammable or corrosive gas is present.
3-2 During Use
Discharge static electricity from the material to be
weighed (hereinafter referred to as the sample). When
a sample could have a static charge, the weight data is
influenced. Ground the balance and try the following:
Eliminate static electricity using the optional
AD-1683 DC static eliminator. Try to keep the ambient humidity above 45%RH. Use a metal shield case for a charged sample. Wipe charged plastic samples with a moistened
cloth.
11
This balance uses a strong magnet as part of the
balance assembly, so please use caution when weighing magnetic materials such as iron. If there is a problem, use the underhook on the bottom of the balance to suspend the material away from the influence of the magnet.
Eliminate the temperature difference between a sample and the environment. When a sample
is warmer (cooler) than the ambient temperature, the sample will be lighter (heavier) than the true weight. This error is due to a rising (falling) draft around the sample.
Make each weighing gently and quickly to avoid errors caused by the evaporation of moisture
from the sample or the absorption of moisture by the sample. For balances with a minimum weighing value of 0.001g or less, use a pair of long tweezers to place a sample on the pan to avoid placing a hand inside the breeze break.
Do not drop things upon the weighing pan, or place a sample on the pan that is beyond the
balance weighing capacity. Place a sample in the center of the weighing pan.
Do not use a sharp instrument such as a pencil to press the keys. Use your finger only.
Press the RE-ZERO key before each weighing to eliminate possible errors.
Calibrate the balance periodically so as to eliminate possible errors.
Take into consideration the affect of air buoyancy on a sample when more accuracy is required.
Keep the balance interior free of dust and foreign materials.
The breeze break (FZ-GD series (all models) and FZ-300CT/600CT and
FX-300CT/600CT/120GD/200GD/300GD) is provided as an accessory. An anti-static treatment has been applied to the breeze break components, but they may be charged with static electricity when they are unpacked or when the humidity is low. If the weight value is unstable even when there are no drafts or the balance has a problem with repeatability, remove the breeze break. Or wipe the clear plates with a moistened cloth, use the optional AD-1683 DC static eliminator or apply an anti-static spray.
3-3 After Use
Avoid mechanical shock to the balance.
Do not disassemble the balance. Contact the local A&D dealer if the balance needs service or repair.
Do not use organic solvents to clean the balance. Clean the balance with a lint free cloth that
is moistened with warm water and a mild detergent.
The edge of the weighing pan is sharp. Use much care when cleaning the pan.
Avoid dust and water so that the balance weighs correctly. Protect the internal parts from liquid
spills and excessive dust.
3-4 Power Supply
When the AC adapter is connected, the balance is in the standby mode if the standby indicator is on
(refer to “4. DISPLAY SYMBOLS AND KEY OPERATION”). This is a normal state and does not harm the balance. For accurate weighing, warm up the balance for at least 30 minutes before use.
12
4. DISPLAY SYMBOLS AND KEY OPERATION
Key operation
Key operation affects how the balance functions. The basic key operations are:
“Press and release the key immediately” or “Press the key” = normal key operation during measurement
“Press and hold the key”
Display symbols
Each key, when pressed or when pressed and held, functions as follows:
Key When pressed When pressed and held
Turns the display on or off. The standby indicator is displayed when the display is turned off. The weighing mode is enabled when the display is turned on.
This key is available anytime. Pressing the key during operation will interrupt the operation and turn the display off.
In the weighing mode, turns the minimum weighing value on or off.
In the counting or percent mode, enters the sample storing mode.
Enters the function table mode. Refer to “9. FUNCTION TABLE”.
Switches the weighing units stored in the function table. Refer to “9-9 Storing Units”.
Enters the response adjustment mode.
Cancels the operation when performing settings.
Enters the calibration mode.
Outputs the weight data to a printer or personal computer using the RS-232C serial interface, depending on the function table settings.
Confirms the operation when performing settings.
No function at the factory setting
By changing the function table: Outputs “Title block” and “End block” for GLP report.
Sets the display to zero.
13
5. WEIGHING
5-1 FZ-CT / FX-CT Series Balances
The default settings of the unit of weight for the FZ-CT / FX-CT series balances are the metric carat
(ct) and grams (g).
Pressing the MODE key switches the displayed unit, between ct and g.
To use other units, refer to “9-9 Storing Units” on page 38 to store the units.
5-2 Basic Operation (FZ-CT / FX-CT Series Balances)
1 Place a container (carat pan) on the weighing pan, if
necessary.
Press the RE-ZERO key to cancel the weight (tare). The
balance displays 0.000 ct. (The decimal point position depends on the balance model.)
2 Place a sample on the pan or in the container.
3 Wait for the stabilization indicator to be displayed. Read
the value.
While the stabilization indicator is on, pressing the PRINT key will output the weight value, using the RS-232C serial interface.
Note Peripheral equipment, that is sold separately , such as a printer or a personal computer is required.
4 Remove the sample and container from the pan.
Notes When other units of weight are stored as described in “9
-9
Storing Units”,
pressing the MODE key switches the
displayed unit.
Press the SAMPLE key to turn on or off the minimum weighing value.
When the ON:OFF key is pressed with a container placed on the weighing pan, the balance displays 0.000 ct and weighing is started,
Sample
Remove the sample
Weighing pan
Carat pan
ct
ct
Weighing pan
Carat pan
(Display off)
ct
14
5-3 FZ-GD / FX-GD Series Balances
The default setting of the unit of weight for the FZ-GD / FX-GD series balances is grams (g).
To use other units, refer to “9-9 Storing Units” on page 38 to store the units.
5-4 Basic Operation (FZ-GD / FX-GD Series Balances)
1 Place a container on the weighing pan, if necessary. Press the RE-ZERO key to cancel the weight (tare). The
balance displays 0.000 g (The decimal point position depends on the balance model.)
2 Place a sample on the pan or in the container.
3 Wait for the stabilization indicator to be displayed. Read
the value.
While the stabilization indicator is on, pressing the PRINT key will output the weight value, using the RS-232C serial interface.
Note Peripheral equipment, that is sold separately , such as a printer or a personal computer is required.
4 Remove the sample and container from the pan.
Notes When other units of weight are stored as described in “9
-9
Storing Units”,
pressing the MODE key switches the
displayed unit.
Press the SAMPLE key to turn on or off the minimum weighing value.
When the ON:OFF key is pressed with a container placed on the weighing pan, the balance displays 0.000 g and weighing is started,
Weighing pan
Container
(Display off)
Sample
Remove the sample
Weighing pan
Container
15
6. RESPONSE ADJUSTMENT
This function stabilizes the weight value, reducing the influence on weighing that is caused by drafts and vibration at the place where the balance is installed.
The function has three stages as follows and can be changed by simple key operation.
Indicator Description
FAST
Fast response, but prone to drafts and vibration. Good for target weighing.
MID.
SLOW
Slow response, but strong against drafts and vibration Good for weighing which requires a stable display.
Operation
1 Press and hold the MODE key until RESPONSE is
displayed. And then, release the key.
2 Press the MODE key to select a rate of the response
adjustment. Either FAST, MID. or SLOW can be selected.
3 After a few seconds of inactivity, or when the PRINT
key is pressed, the balance displays eend . Then, it returns to the weighing mode and displays the updated response indicator. The
response indicator
remains displayed for a while.
Note Setting the response rate automatically changes the “Condition (Cond)” and “Display refresh rate (5pd)” parameters of “Environment, Display (ba5fnc)” in the function table, as shown below:
FZ-CT / FX-CT series
Indicator Cond (Condition)
5pd (Display refresh rate)
FAST 0 1 MID. 1 0 SLOW 2 0
FZ-GD / FX-GD series
Indicator Cond (Condition)
5pd (Display refresh rate)
FAST 0 2 MID. 1 0 SLOW 2 0
When the balance is to be used with other setting combinations, set each parameter in the function table.
16
7. CALIBRATION
7-1 Calibration Mode
Calibration / Calibration test
Calibration Using the Internal mass (only for FZ-CT/GD series) Calibration using an external weight Calibration test using an external weight (Calibration test does not perform calibration.)
Terms
The following terms are defined as follows: External weight = A weight that you have. Referred to as a calibration weight when used for
calibration. Calibration weight = A weight used for calibration Target weight = An external weight used for calibration test
Caution
Calibration adjusts the balance for accurate weighing. Besides periodic calibration and before each use, perform calibration when:
• The balance is installed for the first time.
• The balance has been moved.
• The ambient environment has changed.
Do not allow vibration or drafts to affect the balance during calibration.
To output the data for GLP using the RS-232C serial interface, set “GLP output ( info )” of
“Data output ( dout )”. For details, refer to “9. FUNCTION TABLE”.
Calibration test is available only when “GLP output ( info )” of “Data output ( dout )” is set to “1” or “2”.
Caution on using an external weight
The accuracy of an external weight will influence the accuracy of weighing. Select an
appropriate weight as listed below:
Model Usable calibration weight Adjustable range
FZ-300CT / FX-300CT 50 g, 20 g FZ-600CT/ FZ-120GD FX-600CT/ FX-120GD
100 g, 50 g
FZ-200GD / FX-200GD 200 g, 100 g FZ-300GD / FX-300GD 300 g, 200 g, 100 g
-0.050 g to +0.050 g
FZ-1200GD / FX-1200GD 1000 g, 500 g, FZ-2000GD / FX-2000GD 2000 g, 1000 g FZ-3000GD / FX-3000GD 3000 g, 2000 g, 1000 g
-0.50 g to +0.50 g
The calibration weight in bold type: factory setting The calibration weight value can be adjusted within the range above.
Display
This indicator means “the balance is measuring calibration data”.
Do not allow vibration or drafts to affect the balance while this indicator is displayed.
17
7-2 Calibration Using the Internal mass
(One-Touch
Calibration, only for
FZ-
CT/GD
series
)
This function calibrates the balance using the internal mass. The only operation required is to press
the CAL key. (The FX-CT/GD series does not have the one-touch calibration function.)
Operation
1 Connect the AC adapter and warm up the balance for at least 30 minutes with nothing on the
weighing pan.
2 After carefully installing the small breeze break, press the CAL key.
3 The balance displays Calin and performs calibration using the internal mass. Do not allow
vibration or drafts to affect the balance.
4 The balance displays end after calibration. If the “GLP output (info)” parameter of the
function table is set to “1” or “2”, the balance displays glp and outputs “Calibration Report” using the RS-232C interface or stores the data in memory. For details on the calibration report format, refer to “10-2 GLP Report”.
5 The balance will automatically return to the weighing mode after calibration.
About the internal mass
The mass of the internal mass may change due to corrosion or other damage caused by the operating environment, or due to aging. Check the internal mass periodically. Correct the internal mass value as necessary.
To maintain the weighing accuracy, perform the calibration using an external weight periodically, as described below.
18
7-3 Calibration Using an External Weight
This function calibrates the balance using an external weight.
Operation
1 Connect the AC adapter and warm up the balance
for at least 30 minutes with nothing on the pan.
2 Press and hold the CAL key until Calout is
displayed, and then release the key.
3 The balance displays Cal 0 .
If you want to change the calibration weight (a
list of usable weights is shown on page 16), press the SAMPLE key and proceed to step 4.
If you use the calibration weight value stored in
the balance, proceed to step 5.
4 Specify the calibration weight value as follows:
SAMPLE key To switch the display condition
to: “All of the segments blinking” (calibration weight selection mode) or “The last two digits blinking” (value adjustment mode).
RE-ZERO key (To increase the value) MODE key (To decrease the value)
To select the calibration
weight or adjust the value.
PRINT key To store the new weight
value. Even if the AC adapter is removed, the data is maintained in non-volatile memory.
CAL key To cancel the operation and
return to Cal 0 .
Select a weight
To next page
Press and hold the key
Release the key
e.g. Calibration weight value
50.012 g
ct
19
5 Confirm that there is nothing on the pan and press
the PRINT key. The balance measures the zero point. Do not allow vibration or drafts to affect the balance. The balance displays the calibration weight value.
6 Place the displayed calibration weight on the pan
and press the PRINT key. The balance measures the calibration weight. Do not allow vibration or drafts to affect the balance.
7 The balance displays end . Remove the weight
from the pan.
8 When the GLP output is set, the balance displays
1 glp and outputs “Calibration Report”.
9 The balance will automatically return to the
weighing mode.
10 In the gram mode, place the calibration weight on
the pan and confirm that the value displayed is within ±2 digits of the specified value. If it is not within the range, check the ambient conditions such as drafts and vibration, also check the weighing pan. Then, repeat steps 1 to 9.
20
7-4 Calibration Test Using an External Weight
This function tests the balance weighing accuracy using an external weight and outputs the result. This is available only when the “GLP output (info)” parameter is set to “1” or “2”. (Calibration test does not perform calibration.)
Operation
1 Connect the AC adapter and warm up the balance
for at least 30 minutes with nothing on the pan.
2 Press and hold the CAL key until CCout is
displayed, and then release the key.
3 The balance displays CC 0 .
If you want to change the target weight (a list of
usable weights is shown on page 16), press the SAMPLE key and proceed to step 4.
If you use the target weight value stored in the
balance, proceed to step 5.
4 Specify the target weight value as follows:
SAMPLE key To switch the display
condition to: “All of the segments blinking” (target weight selection mode) or “The last two digits blinking” (value adjustment mode).
RE-ZERO key (To increase the value) MODE key (To decrease the value)
To select the target weight or
adjust the value.
PRINT key To store the new weight
value. Even if the AC adapter is removed, the data is maintained in non-volatile memory.
CAL key To cancel the operation and
return to CC 0 .
Select a weight
To n ex t pa g e
Press and hold the key
Release the key
e.g. Target weight value
50.012 g
ct
21
5 Confirm that there is nothing on the pan and press the
PRINT key. The balance measures the zero point and displays the measured value. Do not allow vibration or drafts to affect the balance.
The balance displays the target weight value.
6 Place the displayed target weight on the pan and press
the PRINT key. The balance measures the target weight and displays the measured value. Do not allow vibration or drafts to affect the balance.
7 The balance displays end . Remove the weight from the
pan.
8 When the GLP output is set, the balance displays glp1
and outputs “Calibration Test Report”.
9 The balance will automatically return to the weighing
mode.
From previous page
Targ et we igh t
GLP output
ct
22
8. FUNCTION SWITCH AND INITIALIZATION
8-1 Permit or Inhibit
The balance stores parameters that must not be changed unintentionally. There are two switches for the purpose of protecting these parameters. Each switch can select either “permit” or “inhibit”. The “inhibit” protects parameters against unintentional operations.
FZ-CT/ FZ-GD series FX-CT/ FX-GD series
Function switch
- Function table
- Calibration using the internal mass
- Calibration using an external weight
- Internal mass correction
- Function table
- Calibration
Switches (FZ-CT/ FZ-GD series)
Switches (FX-CT/ FX-GD series)
Function table
 0 
To inhibit changes to the function table
 1 
To permit changes to the function table
Calibration using the internal mass (One-touch calibration)
 0 
To inhibit calibration using the internal mass
 1 
To permit calibration using the internal mass
Calibration using an external weight
 0 
To inhibit calibration using an external weight
 1 
To permit calibration using an external weight
Not used
Internal mass correction
 0 
To inhibit correction
 1 
To permit correction
(The display shown left indicates the factory settings.)