Sensitivity of the load cell is
Accessory connector for power
AD-4430B
DIN Rail Weighing Module with BCD Output
Simplified Instruction Manual
Refer to the instruction manual on the A&D home page
1WMPD4002677E
This Manual
This manual describes how the product works and how to get the most
out of it in terms of performance. Read this manual thoroughly before
using the product and keep it at hand for future reference.
Product s pecifications are subject to change without any obligation on
the part of the manufact urer to notify of changes.
This manual is subject to change without notice at any time to improve
the product. No part of this manual may be photocopied, reproduced,
or translated into another language without the prior written consent of
the A&D Compa ny, limited.
Do not attempt t o repair, modify or disassemb le the product. Doing so
will void the warranty.
2018
3-23-14 Higashi-Ikebukuro, Toshima-ku, Tokyo 170-0013, JAPAN
Telephone: [81] (3) 5391-6132 Fax: [81] (3) 5391-6148.
1. Cautions
Installation and Precautions
1.1.
Before use, confirm the following articles for safe operation.
Avoid vibrati on, shock, extr emely high tempe rature and humi dity, direct sunlight,
dust, splashing water, air containing salt or corrosive gases, places where
inflamma ble gases are p resent.
The operating temperature is -10°C to +50°C (14°F to 122°F).
Ground the module.
Keep cables away from power cables and other sources of electrical noise.
Use a stable DC24 V power source that does not include step down voltage
and noise.
Do not share the earth ground line and power line with other electrical power
equipment.
W hen extending the load cell cable, separate it from the power cable and
electrica l cables with m uch noise.
Do not turn on the module unt il installatio n is complete. Th e module is not
equipped with a switch to turn off.
After the installation is complete, take off the protective cover prior to turning on
the AD-4430B.
Use a shielded load cell cable.
Do not connect more sensors than the allowable number noted in the
specifications.
Cautions During Use
1.2.
The AD-4430B is a precision instrument that measures microvolt output from load
cell. Prevent noise sources such as power lines, radios, el ectric welders or motors
from affecting the instrument.
Do not disassemble the AD-4430B.
-
URL: https://www.aandd.co.jp/
All rights reserved.
- -2- -3- -4-
Voltage requirement DC 24 V +10%, -15%
Power requirement
Load cell excitati on voltage
Operating conditions
External dimensions
Mass
Main display
Accessory
3. External Dimensions
6 W Max.
5V 350 : sensor. Up to four sensors can be connected.
-10 °C to +50 °C, Max 85 %RH (no condensation)
35.3 × 110.0 × 101.3 mm (W×H×D)
Approximately 180 g
The monitor displays measurement data and settings with 7
segments of 5 digits and negative sign. The decimal point is
specifie d at the functio n table.
Power con nector FMC 1.5/2-ST-3.5
35.3
(2.5) 100 1.3
4. Panels
Power Supply Connector
4.1.
...........
+24 V
0 V
4.2.
SIG SIG+
EXC-
SEN-
SEN+
EXC+
SHLD ........... Connect shield of load cell cable.
DC +24 V terminal.
................
DC 0 V t erminal.
Sensor Terminals Of Load Cell
............
The (-) input terminal of signal that is outputted from load cell.
............
The (+) input terminal of signal that is outputted from load cell.
...........
The (-) output terminal to load cell excitation voltage (-).
...........
The (-) input terminal for sensing input (-). (When performing the
4-wire connection, connect between EXC- and SEN-.)
............
The (+) input terminal for sensing input (+). (When performing the
4-wire connection, connect between EXC+ and SEN+.)
...........
The (+) output terminal for load cell excitation voltage (+).
Connector
Connections
W hen connecting and removing the cables, push t he buttons with a driver etc.
We recommend use of pole crimp term inals for the ti ps of cables.
BCD Connector
4.3.
Unit : mm
20
28
39.4
34.5
5.5
36.1
The circuit of BCD is isolated from
the D.C. power supply and load
cell circuit.
Supply D.C. +24 V bet ween BCD
PWR+24V te rminal and COM
terminal.
4.4.
Status Indicat or
Gross : LED lights when indicating gross value.
Net : LED lights when indicating net value.
Hold : LED lights when the hold function operates.
Stable : LED lights when the current weighing value is stable.
Zero : LED lights when the weighing value is center zero.
This LED works by selected function at .
Key switches
4.5.
The display switch between gross and net in factory setting.
"The zero key" to perform the zero adjustment in weighing mode.
The key to change a sel ected item or move a flashed figure.
"The tare key" that displays zero for net weighing.
The key to select parameter or increase number in setting mode..
Press the key to turn on the display in standby (OFF mode).
Press and hold the key to turn off the display in weighing mode.
In the s etting mode, this key to store ne w settings. "the enter key".
Function key can be selected the function and use in weighing mode.
The return key or "the escape key" in setting mode.
+
Proceeds to function mode in weighing mode.
+
Proceeds to check mode in the function mode.
+
The condition of the AD-4430B can be updated and be stored.
The mode to check th e AD-4430B.
The mode to calibrate zero point and span of the AD-4430B using span mass.
The AD-4430B measures the voltage of the load cell and displays it.
Specify d ecimal point (
"input voltage at zero calibration (
ſ Perf orm stable meas urement in the ca libration to p revent measurement error.
ſ During a stable measurement, the LED lights.
ſ The decimal point flashes in calibration mode.
Proceeds to calibration in standby (OFF mode).
Operation Mode
4.6.
F unction mode (In weighing mode, + )
Check mode (In function mode, + )
Calibration mode (When display is turned off, + )
5. Calibration
-
function use d to adjust the AD
) in function mode.
(
and "mass value against inp ut voltage at span c alibration (
the "span calibration using mass (
inputted using "digital span" in function mode.
4430B so that t he signal from th e load cell is prop erly converted to mass.
), minimum division ( ) and weighing capacity
supply terminal.
removing the cables.
Calibration is the
)", " inpu t voltage at span c alibration ( )"
)" in calibration mode. These items can be also
)" can be changed by
Span Calibration using Mass
5.1.
Preset a unit, decimal point, minimum division and weighing capacity in funct ion
mode. The span calibration is performed using mass of the weighing capacity.
ſ When and a numb er are displayed, an error has oc curred. Refer t o
"Calibration Errors" for details.
ſ The flashing decimal point means "no weighing value".
ſ Before the calibration, turn on the AD4430B more than 10 minutes so as to
avoid temperature drift (change).
Step 1 When turning off the display, press and hold the
key and press the
Then
Step 2 Press the key to enter calibration mode.
Step 3 Press the key to display .
Step 4 Confirm that the
Step 5 When is displayed, press the
Step 6 Place the m ass on the weighing pan.
Step 7 When
Step 8 Pr ess th e
Step 9 Pr ess th e
5.2.
Display
is displayed. When returning to
weighing m ode, press the key.
Zero Calibration
When skipping zero calibration, press the
key and proceed to step 5.
key. Then is display ed for 2
seconds. W hen canceling s pan calibration and
returnin g to weighing m ode, press the
key twice.
Span Calibration
key. The current span mass is displayed. A
figure for the mass flashes. Specify a new value
for the mass using the
When canceling span calibration and returning
to weighing mode, press the
Confirm that th e
key. Then is displayed for 2
seconds.
from the wei ghing pan. When repea ting span
calibration, press the
displaye d and calibration data is st ored in the
nonvolatile memory of the AD-4430B.
Calibration Error s
Resolution (Weighing
capacity / minimum division)
exceeds the limitati on.
The voltage in zero
calibration is too large.
The voltage in zero
calibration is too small.
Mass value exceeds the
weighing capacity.
Mass valu e is too light f or
calibration.
too small.
The polarity of the load ce ll
output is reversed.
The output of the weighing
capacity is out of range.
key ( + key).
of calibration mode is displayed.
LED is lit and p ress the
and keys.
key twice.
LED is lit and press the
is displayed, remove the mass
key.
key. Then is
key to return to weighing mode.
Cause Treatment
Make the minimum division greater or make the
weighing capacity sm aller. The specified value
depends on specifications of the weighing system.
Check the load cell rating and connection. When
nothing is wrong with the rating and connection,
adjust the load cell output. When the load cell or
A/D converter may be the cause of error, confirm
this by using the check mode.
Use an appropriate calibration weight and
calibrate again.
Use a load cell with higher sensitivity or make
the minimum division greater.
Check the load cell connection.
Use a load cell with a greater rating or make the
weighing capacity smaller.
key
key
mode
key
calibration
LED
key
calibration
key
Example
key
Example
LED
key
mass
Key to store data
key
Digital Linearization ( )
5.3.
Digital lin earization is t he non-linearit y compensation f unction that can rectify or
reduce linearity deviation between zero point and weighing capacity.
Up to four points can be specified except zero. (Refer to )
Relationship of points : Zero =
The high-order correction curve is used so that zero point and individual points
are arranged in a straight line.
Digital linearization includes span calibration.
ſ When and a number are displayed, an error has occurred. Refer to
"Calibration Error" for details .
ſ A flashi ng decimal point means "no wei ghing value".
ſ Before calibration, turn on the AD4430B more than 10 minutes so as to avoid
temperature drift (change).
Step 1 When turni ng off the displ ay, press and hold
the
key and press the key ( +
key). Then of calib ration mode
is displa yed. Press the key to display
.
Step 2 Pr es s th e key to select and
press th e key to enter digital
linearization.
Step 3
Step 4 P lace nothing on the pan.
Step 5 W hen displaying ,
Step 6 P lace the mass on the pan.
Step 7
Step 8
Step 9
Step 10
Step 11 Pre ss t he key to return to weighing mode.
of the zero point is displayed.
LED is displayed, pres s the
While
key to store the weighing value.
Then
press the key to select
a mass value. Specify it using
the
While
press the
the weigh ing value.
Then
for 2 seconds.
operatio n as Step 5 and Step 6 at the
second p oint.
operatio n as Step 5 and Step 6 at the third
point.
operatio n as Step 5 and Step 6 at the fourth
point.
to store new parameters into FRAM and
display
is displayed for 2 sec onds.
and key.
LED is displayed,
key to store
is disp layed
is displayed. Repeat the same
is displayed. Repeat the same
is displayed. Repeat the same
is displayed. Press the key
.
Remove all of mass from t he pan.
< < < <
linearization
LED Stable
perform the
same
operation and
measureme nt.
Place mass
LED Stable
Key to store data
mass
6. Function Mode (Function List)
The function mode stores parameters to c ontrol the weighing modul e.
The parameters are stored even without power supplied.
Outline
6.1.
Types of Functions
...........
Calibration function
...........
Linearity adjustment function
............
Basics function
...........
Hold function
...........
6.2.
-
BCD output functi on
Operation
Select Modes Under Function Mode
..............
Selects a type of select mode (3 upper figures)
..........
Enters a selected mode.
..............
Selects an item under th e selected mode (2 lower figures )
..........
Enters the item.
..........
Stores param eters and returns to weighing mo de.
- -6- -7- -8-
Parameters Selection And Digital Parameters
P Type of parameter selection ( all figures flashing)
..........
Selects a parameter.
......
Activates (store) parameter and returns to select mode.
......
Deactivates parameter and returns to select mode.
D The type to change value ( figure flashing )
.........
Moves the flashing figure.
..........
Changes the value of the flashing figure.
......
Activates (store) value and returns to select mode.
......
Deactivates value and returns to select mode.
6.3.
Function Table
ſ Decimal point depends on .
ſ The input voltage of the zero point ( ), the s pan input voltage ( ), mass for
span input voltage (
ſ If pressing the key while displaying the digital filter ( ), the c urrent weighing
value can be monitored. Press the
Calib ration Function ( )
Item
Range
0 to 0.0000
Minimum division
1 to 50
Weighing capacit y
1 to 99999
0 to 100
Zero tracking time Used with for zero tracking [sec.]
0.0 to 5.0
Zero tracking width Used with for zero tracking [digit]
0.0 to 9.9
Stability detection
0.0 to 9.9
Stability detection
0 to 9
adjustment when
0 to 1
0 to 1
0 to 1
1 to 3
1 to 2
0 to 1
Zero setting when
0 to 1
power is turned on
7.0000
9.9999
9.8500
9.8500
0 to 1
Input voltage at
Input voltage at
Calibration weight
corresponding to
input voltage at span
Gravity a
-7.0000 to
0.0001 to
1 to 99999
9.7500 to
9.7500 to
) are revised in calibrati on mode.
key again to return function mode.
Item Name Descriptions
Decimal point
position
(A scale / digit)
Zero range
width
Tare and zero
unstable
Tare when the
gross weig ht is
negative
Output when
overflow and
unstable
gross weig ht
net weight
Clear the zero
value
span
of place of
calibration
of place of use
Decimal point position of value.
0 0.0 0.00 0.000 0.0000
Minimum division of value.
1 2 5 10 20 50
Measurement can be displ ayed up to +8 digits
(8 scales) from capacity. Decimal point
depends on
The range that the
Deviation from the calibrated zero point [%].
Used with
time
Used with
The adjustment when value is unstable.
0: Disables both functions.
1: Enables both functions.
Tare when the gross weight is negative.
0: Disabl es tare.
1: Enables tare.
Serial output when overflow and unstable.
0: Disables output.
1: Enables output.
Judgment when the negative gross is exceeded.
1: Gross < -99999
2: Gross < -capacity
3: Gross < -19 digit
Judgment when the negative net is exceeded.
1: Net < -99999 2: Net < -Capacity
Select to clear the zero value.
0: Disable 1: Enable
When turning power on,
0: Disable 1: Enable
Input voltage of the zero point [mV/V]
zero
Input volt age of the full scale [mV/V]
Mass value to equivalent to span input
voltage. (
Gravity acceleration of the place where
calibrating the module. [m /s
Gravity accelerati on of the place where using
the module. [m/s2]
) Decimal point depends on
.
.
key (zero key) works.
for stability detection [sec.]
for stability detection [digit]
2
]
Disable hold 0: Enable 1: Disable
Default
Type
70000
0.0000
3.2000
32000
9.8000
9.8000
key status
)
(depends on
Default
Type
0
P
0.0000
D
0
D
0.0000
D
0
D
0.0000
D
0
D
0.0000
D
0
D
0.0000
D
Default
Type
0000
Binary
number
D
5
P
.
.
.
.
)
20
P
0
P
15
P
1
P
10
-10
D
1
P
10
D
D
1
P
1
P
.
.
.
Linearity Adjustment Function (
Item
Item Name Descriptions
Range
Number of input
0 to 5
points
7.0000 to
Linear-zero I nput voltage for linear-zero input. [mV/V]
7.0000
Linear 1
0 to 99999
Mass value
Linear 1
0.0000 to
Span voltage
9.9999
Linear 2
0 to 99999
Mass value
Linear 2
0.0000 to
Span voltage
9.9999
Linear 3
0 to 99999
Mass value
Linear 3
0.0000 to
Span voltage
9.9999
Linear 4
0 to 99999
0
P
1
P
D
2
D
0.0
0.0
1.0
D
2
P
1
P
1
P
1
P
1
P
1
P
1
P
0
P
D
D
D
D
D
0
Mass value
Linear 4
0.0000 to
Span voltage
9.9999
Basic Function ( )
Item
Item Name Descriptions
Range
Key switch
0000
to 1111
key function
0 to 7
Display refresh
5 to 20
0 to 9
Digital filter
0 to 16
Hold function
1 to 3
-99999 to
99999
-99999 to
99999
-99999 to
99999
1 to 2
1 to 2
1 to 2
Near-zero
Comparison mas s at
Upper limit value
Lower limit value
Comparison mas s of
upper and lower limit
Output logic of upper
and lower limit
Number of adjustment point. If 0, 1 or 2 is
selected, digital linearization is not used.
The mass value for linear 1 input.
Decimal point depends on
The span voltage between linear1 input. [mV/V]
The mass value for linear 2 input.
Decimal point depends on
The span voltage between linear2 input. [mV/V]
The mass value for linear 3 input.
Decimal point depends on
The span voltage between linear3 input. [mV/V]
The mass value for linear 4 input (Full scale).
Decimal point depends on
The span voltage between linear4 input (Full scale). [mV/V]
Figure corresponds to each key (switch).
These are enabled at weigh ing mode.
Figure 4 fig. 3 fig. 2 fig. 1 fig.
disable
Parameter 0: Enable 1: Disable
0: None 4: Momentary switch
1: Print com mand 5: Display exchange
2: Hold 6: Tare clear
3: Alternative switch 7: Zero clear
20 times/s 10 times/s 5 times/s
rate
0: None
1: Zero tracking in progress
2: Alarm (Zero range s et ting error, over)
3:
4: Zero band
display
5: HI output (Over the upper limit value)
6: OK output (Between upper and lower limit values)
7: LO output (Below the lower limit value)
8: User input 1
9: User output 1
Selects a cutoff frequency.
0: None 6: 20.0 Hz 12: 2.8 Hz
1: 100.0 Hz 7: 14.0 Hz 13: 2.0 Hz
2: 70.0 Hz 8: 10.0 Hz 14: 1.4 Hz
3: 56.0 Hz 9: 7.0 Hz 15: 1.0 Hz
4: 40.0 Hz 10: 5.6 Hz 16: 0.7 Hz
5: 28.0 Hz 11: 4.0 Hz
1: Hold immediately
2: Peak hold
3: Averaging hold
The refer ence value for near-zero.
Decimal point depends on
1: Gross weight 2: Net weight
near-zero
Reference value for t he upper limit.
Decimal point depends on
Reference value for the lower limit.
Decimal point depends on
1: Gross weight 2: Net weight
The logic of the comparator function output.
1: Positive logic 2: Negative logic
Hold Function ( )
Item
Item Name
Range
Average
0.00
to 9. 99
0.00
to 9. 99
0 to 2
0 to 1
0.00
to 9. 99
0 to 99999
0 to 1
Item
Range
1 to 4
1 to 3
5 to 10 00
1 to 2
1 to 2
1 to 2
1 to 2
0 to 5
0 to 12
Ԭ1: GRS/NET(1 8) for input
Ԭ2: HOLD(34) for i nput
time
Star t wai t
time
Condition of
automatic
start
Release using
control input
Release time
Release using
fluctuation range
Release at
near-zero
BCD Output Function ( )
Item Name
Data
outputting
mode
rate
Output logic of
data transfer
Output logic of
negative sign
Output logic
of status
Output logic
of strobe
Input
selection
Input(17)
Output
selection
Output(33)
Descripti ons
Time to calc ulate the aver age. [sec.]
0.00 : Not used
Time to wait before commencing hold or
averaging. [sec.]
Condition for starting the hold or averaging.
0: Not used 2: Above the near-zero range
1: Above the near-zero range, and stable
Release when control input is falling.
0: Do not release 1: Release Ԭ2
Release after a set amou nt of time has
passed. [ sec.] 0.00 : Continue
Release when fluctuation from the holding
value exceeds a set value. Dec imal point
depends on
Release when the weighing value is in the
near-zero range.
0: Do not release. 1: R elease.
. 0 : Continue
Descripti ons
1: Weighing value 3 : Net weight Ԭ1
2: Gross weight 4: Data specified at BCD in put
1: Stream 3: Manual printing
2: A utomatic printin g
5 tim es/sec.
10 times/sec. (Depends on
20 times/sec.
100 times/sec.
1000 times/sec.
1: Negative logic 2: Positive logic
1: Negative logic 2: Positive logic
1: Negative logic 2: Positive logic
1: Negative logic 2: Positive logic
0: None
1: Clear z ero
2: Clear tare
3: Changing between gross and net
4: Print com mand
5:
key
0: None
1: Stabilization
2: During tare
3: Zero band
4: Hold busy state
5: HI output (Over upper limit)
6: OK output (W ithin upper and lower limits)
7: LO output (Under lower limit)
8: During operating weighing (On)
9: During operating weighing (1 Hz)
10: During operating weighing (50 Hz)
11: Alarm (Zero correction error and tare error)
12: Busy
key
Default
Type
0.00
D
0.00
D
0
P
1
P
0.00
D
0
D
0
P
Default
Type
1
P
1
P
)
20
P
2
P
2
P
2
P
2
P
3
P
1