A&D AD-4329 User Manual

' a
.-'" '1
DlMinMy(STOi
Weighing Indicator
i-'ii
® [:|f ;I^Ig
WM+PD4000173D
A&D Company, Limited
A This is a hazard aiert mark.
This mark informs you about tha operation of the product
of this manual may be photocopied, reproduced, or translated into another lianguage without the prior writteri consent of A&D Company, Ltd,
Product specifications are subject to change without any obligation on the part of the manufacture.
© 2005 A&D Company Ltd. Ail rights reserved;
Contents
Compliance
1.1. к Compliance with FCC Rules
1.1.2, Compliance with European Directive
Introduction
Installation and Precautions
3.1.1. Installation and Precautions..........................................................................5
3.1.2. The Load Ceil Connections............................................................................5
3.1.3. Adjustment of the Load Cell Output............................................................ 6
3.1.4. Verifying Load Cell Output and Input Sensitivity...
3.1.5. Installing the Option Board.......................................................................... 7
Description of Panels and symbols
4.1.1. Front Panel Description............................................................................... 8
4.1.2. Rear Panel Description.................................................................................... 9
4.1.3. Other Displays and Symbols....
4.1.4. Accessories and Option OP"02.....................................................................10
Calibration
5.1.1. Items of Calibration Mode............................................................................. 11
5.2. Calibration Procedure.................................................................................... 12
5.2.1. The Whole Procedure................................................................................. 12
5.2.2. Preparation
5.2.3. Entering the Calibration Mode.....................................................................13
5.2.4. Minimum Division (Weighing Interval)........................................................14
5.2.5. Maximum Capacity
5.2.6. Zero Calibration
5.2.7. Span Calibration............................................................................................16
5.2.8. Exiting the Calibration Mode........................................................................17
5.3. Weighing Range Function
5.3.1. Setting Division and Range
5.4. Digital Linearization Function......................................................................... 22
5.5. Gravity Compensation Function
5.5.1. The Gravity Acceleration Table....................................................................24
5.6. Calibration Error Code List
.............................................................................................................
...........................................................................3
.............................................................
........................................................................................................ 4
...............................................................................
.......................................
........................................................................8
...................................................................
....................................................................................................... 11
....................................................................................................13
...................................................................................
..................................................................................... 15
...............................................................................
......................................
.................................................................... 23
...............................................................................25
.............
......................19
14
18
3
3
5
6
9
6.
6.1. Changing the Function Settings........................................................................26
6.2. F“Functions.................................................................................................... 27
6.3. CF“Functions................................................................................................... 32
AD4329 Weighing Indicator, instruction Manual, A&D Co„ itd. Page 1
Functions
............................................................................................................
26
ш
I
I
7. Tare.,...,.,
; 7.1,1. Weighing Tare ......;^
7.1.2. Digital Input
7.1.3. ClearingTare............................................................................................... 33
8. Accumulation
8.1.1. Preparation and Specification
8.1.2. Display and Operation................................................................................. 35
9. Comparison
9.1. Comparator Output Relay (OP-02)
9.2. Upper/Lower Limit Comparison....................................................................... 36
9.2.1. Selecting Upper/Lower Limit Comparison................................................. 37
9.2.2. Condition formula for Comparison
9.2.3. Setting Upper/Lower Limit Values...
9.2.4. Example..................................................................................................... .38
9.3. Setpoint Comparison
9.3.1. Selecting Setpoint Comparison...................................................................
9.3.2. Description of Input parameters and Outputs
9.3.3. Setting the Parameters of Setpoint Comparison.
..........
........*....».......:............;.:.....;;....,.л...Л...:,...» 33
......
....................................
............................................................................................... 33
..................................................................................................... 34
.................................................................. 34
......................................................................................................... 36
..........
.......................................................
........................................................
..........................................................
...................................................................................
............................................
.......................................
33
36
37
38
39
39 39 40
10. External Control Input................................... .................................................. 41
11. Standard Serial Data Output (SER.OUT)
11.1.1. Data Format................................................................................................ 43
12. RS-232C Interface............................................................................................ 44
12.1.1. Specification............................................................................................. 44
12.1.2. Data Format................................................................................................ 45
12.1.3. Command Format
12.1.4. Command to Request Data
12.1.5. Commands to Control the Indicator..
12.1.6. Commands to Set Parameters
12.1.7. Related Functions.................................................................................... 49
13. Specifications
13.1. Dimensions.................................................................................................. 51
....................................................................................................
.................................................................................... 46
.....................................................................
.............................................................
...
..........................................
........................................................
........48
42
46
47
50
Page 2
AD-4329 Weighing indicator

Compliance

1.1.1. Compliance with FCC Rules

Please note thaï this equipment generates, uses and can radiate radio frequency energy. This equipment has been tested and has been found to comply with the
limits of a Class A computing device pursuant to Subpart J of Part 15 of FCC rules. These rules are designed to provide reasonable protection against interference when this equipment is operated in a commercial environment. If this unit is operated
in a residential area it may cause some interference and under these circumstances the user would be required to take, at his own expense, whatever measures are
necessary to eliminate the interference.
(FCC = Federal Communications Commission in the U.S.A.)

1.1.2. Compliance with European Directive

C€ This appliance complies with the statutory EMC (Electromagnetic Compatibility)
directive 89/336/EEC and the Low Voltage Directive 73/23/EEC for safety of electrical equipment designed for certain voltages.
Note: The displayed value may be adversely affected under extreme electromagnetic
influences.
AD-4329 Weighing indicator
Page 3
1. Compliance

2. Introduction

□ The AD-4329 is a compact weighing indicator that amplifies the signal from a load
cell, converts it to digital data and displays it as a mass value.
a This indicator has the following performance:
Input sensitivity: Maximum display: Refresh rate of the display:
Input voltage range:........................
..........................
..........................
..............
...0.2juV/division.
10000 divisions (or 20000 divisions). 10 times/second approximately.
-1 mV ~+15 mV.
□ There are the following standard functions:
□ The upper/lower limit comparison to check a mass value. □ The setpoint comparison for batching applications. □ The accumulation function to totalize these mass values and to count the number
of accumulations.
a There are the following standard interfaces:
□ The standard serial output for a printer. Q The RS-232C serial interface to communicate with a computer.
This interface can request weighing data, enter parameters and control the state of the indicator.
□ The calibration function includes the following functions:
□ Setting of the minimum division (weighing interval) and the maximum capacity. □ Zero and span calibration. a The weighing range function of the multi-interval weighing instrument (scale). □ Digital linearization function.
□ Gravity compensation function. □ There are seven external input terminals controlled by the F function, o The option OP-02 has three output relays used for the upper/lower limit comparison
and setpoint comparison.
2. Introduction
Page 4
AD-4329 Weighing Indicator

3. Installation and Precautions

ЗЛ .1. Instailation and Precautions

The AD-4329 weighing indicator is a precision electronic instrument Handle the Indicator carefully. The operating temperature is -10t: to *f40t: (14“F to 104"^). Do not install the scale in direct sunlight. Misoperation or other problems may be caused by an unstable power source including momentary power failrue or instantaneous noise. Use a stable power source.
Aa
Ad
Do not connect the power cord before the installation Is finished. Please confirm that the local voltage and receptacle type are correct for your scale. The jOperate/Standbvi key @ key) can not cut the power. The load cell and the internal circuit are supplied with power in the standby state. Use shielded cable for ail connections. Connect the cable shields to the shield terminal or case as an earth terminal. Earth ground the indicator. Do not Join the earth ground line with other electric power equipment. There is an earth ground terminal at the power cord receptacle.
Do not install the scale in a place where it is af3t to be charged with static electricity, or where the relative humidity is lower than 45%RH. Plastic and isolators are apt to be charged with static electricity.

3.1.2. The Load Cell Connections

Connect the load ceil wires to the connector (receptacle), at the rear panel, using the accessory load cell plug.
It is possible to connect a 4 wire cable that 1 pin"2pin and 3pin-4pin are shorted, if the distance between the indicator and a load cell is shorter than 5m. The output voltage of a load ceil is a very sensitive signal. Space the load cell cable away from any noise source.
It is possible to connect eight 350Q load cells.
The load cel! drive is 5VDC ± 5% between EXC+ and EXC-, the maximum current 120mA.
EXC+
EXC­SIG­S1G+
Load ceii
AD-4329 Weighing Indicator
Cable Cable
i\
Positive Sense Positive Excitation Input
Load Cell Plug
Negative Excitation input
Negative Sense
Negative signal Output
Positive signai Outpur
Standard connection
Page 5
Available connection for a cable shorter than 5m.
3. Installation and Precaution
3.1.3. Adjustment of the Load Ceil Output
Caution a Use a inetaf film resistor in the range of 50kQ to 500kQ with a good
temperature coefficient when adding a resistor to adjust a load cell output. Use as a large resistance value as possible in the range in which zero adjustment is possible. Solder this resistor at a point near the load cell or the indicator.
In Case of Reducing the Output Voltage
When the zero output is too large, add a resistor between EXC+and SIG-.
in Case of Adding an Offset Voltage to the Output
When the zero output is too small, add a resistor between EXC+ and S1G+.

3.1.4. Verifying Load Ceil Output and Input Sensitivity

The input sensitivity of the indicator is 0.2pV/division or more. Adapt to the following inequality, when you design a weighing instrument using the indicator and load cell(s).
Caution □ A change in input voltage sensitivity is equivalent to a one division
change of the display. Select as large an Input voltage sensitivity voltage as possible so that the weighing interval becomes stable,
a Consider the leverage If a lever is used.
Weighing instrument using one load ceil.
, Weighing Instrument
using multi-load cell
Verification Example
Design; Load cell
Rated capacity
Rated output
Excitation voltage
Weighing interval
Weighing capacity
E^tB^D
0.2 £
-----------­A
E^Bh^D
0.2^——~ A^N
N=1
A=750 [kg]
8=3 [mVA/I E=5000 [mV] regard the instrument as a good design. D=0.05 [kg]
300 [kg]
A: Rated capacity of load cell [kg] B: Rated output [mVA/]
D: Weighing interval [kg] E: Excitation voltage [mV] N: Number of load cells
6000*3*0.05 , ^
„ -j ^ 0.2. Therefore,
750
3. Insteilatìon and Precaution
Page 6
AD-4329 Weighing Indicator

3,1,5, Installing the Option Board

Caution □ Do not remove screws without the following steps.
This is a procedure for the relay output board (OP-02). Step 1 Remove the power cord and other cables from the indicator. Step 2 Step 3 Step 4
Remove four screws and four boits from the rear panel. Remove two slide-rails, one on each sides. Pull the rear panel and main board from the case. Then the connector between the
main board and the display board may be disconnected smoothly. Step 5 Step 6 Step 7 Step 8
Remove two screws and the small blank panel from the rear panel.
Attach the option board using two screws. Connect the option cable connector to J5 on the main board.
Insert the main board into the both main board guides and insert it as before. Then the
connector between the main board and the display board Is re-connected, if the rear panel will not close completely, retry step 8 to insert the main board correctly.
Step 9 Attach both slide-rails. Fix the rear panel using four screws and four bolts as before.
AD-4329 Weighing indicator
Page 7
3. Installation and Precaution

4.1.1. Front Panel Description

Panels and symbols

No.
1 Standby indicator
2
Data display
3 Weighing unit
State indicator
4
CAL cover
5
CAL key
6 7 Operate/Standby key
Preset tare key
8 9 M+ (Accumulation) key
10 Mode key
Net/Gross key
11
+/" key Zero key
12
>key Tare key
13
A key
Print key
14
Enter key, key
Name •
Description
The indicator lights at the standby state. This state means
connecting the power cord and turning the display off. Weighing data Is displayed. A unit selected in the CF functions. The ^ marks indicate kind of data and weighing state.
M-f- It lights when there is accumuíation data.
it is blinking when displaying accumulation data.
MD The motion detection mark.
It lights when motion is detected.
It turns off while data is stabilized. GROSS It lights when displaying gross data. NET It lights when displaying net data.
PT It lights while entering a preset tare parameter. ZERO it lights when displaying zero. Calibration protection cover. Calibration key. The key to adjust the scale so that mass value is displayed correctly and to enter the CF functions. The key to turn the display on or off. The key to enter preset tare setting mode. The key for the accumulation function. Refer to "8. Accumulation", CF11, F4 and F5.
The key for the accumuiation function and calibration. The key to select a unit of lb or kg. The key to select net or gross. The key to set the polarity of data while in the setting mode. The key to zero a current display.
The key to select a figure while in the setting mode. The key to perform tare. The key to select a value while in the setting mode. The key to output data. The key to store current data while in the setting mode.
4. Description of Panels and symbols
Page 8
AD-4329 Weighing Indicator

4.1.2. Rear Panel Description

No.
15 Power connector
16 Grounding terminal 17 RS-232C interface 18 Blank panel
19 Fuse holder
20 External input
21 Serial output 22
Load cell connector
Name Description
Use accessory power cord. Please Confirm that the iocal voltage and receptacle type are correct for your scale.
Refer to "12. RS-232C Interface" for the detail. Option space (for relay output). Use 0.5A time lag fuse for AC1OOV to AC120V. Use 0.316A time lag fuse for AC200V to AC240V. Seven control inputs selected function by the F functions. Refer to FI 0 to 16.
Printer port. Refer to "11. Standard Serial Output" for the details. Connect accessory load cell plug.

4.1.3. Other Displays and Symbols

Blank Decimai point
r r
AD-4329 Weighing Indicator
standby display.
Zero error when turning display on. If the I mode I key is
pressed, a current weighing value may be displayed. Over load display. Remove any load from the load cel!
immediately. It may cause damage to the indicator. Example of an error display.
Page 9
4. Description of Panels and symbols
4.1.4. Accessories and Option OP“02
1 1
(1) 1
1
JM-NJC-207-PF
JA-TCP0576 JA-TCP0586 FS-EAWK-500MA
Accessories
Instruction manual Load cell piuq Power cord 7pin Din connector
Spin Din connector
0.5A or 0.315A time lag fuse 1 FS-EAWK-315MA
Rubber foot 4
10-SJ-5023
Caution Please Confirm that the receptacle type and focal voltage is correct for
your indicator (scale).
Comparator relay output
Accessory
OP-02
Cable
AD4329-02
1
4. Description of Panels and symbols
Page 10
AD-4329 Weighing Indicator

5. Calibration

This weighing Indicator, converts an input voltage from a load cel! to the "mass*' value,
and displays it. Calibration is the adjustment function so that the scale (indicator) can
weigh correctly.

5:1,1, items of Calibration Mode

There are seven items In the calibration function.
Basic Items
The minimum division.
The maximum capacity
Zero calibration
Span calibration
...
...................
..................
.........................
Comment This function Is the fundamental starting point to
..................... .
.....
_______________________
Setting the weighing interval.
Setting the maximum display.
When unloading any weight on the weighing unit
( there is nothing on the weighing pan), this function performs adjustment so as to display the zero point mark. Select an adjustment method from the. "weighing input" or "digital input".
weigh anything, and influences the performance of scale (indicator).
The function to measure an input voltage variation
by loading to the weighing unit correctly. Select an adjustment method from the "weighing arbitrary mass", "weighing capacity mass" or "digital input".
_____
Optional Items (Sub-functions)
Range function..........................
pigital linearization function
Gravity compensation function... The function to revise weighing error between the
Caution □ The maximum display is less than or equal to 10000 divisions. This number
is calculated from the maximum capacity divided by the minimum division. □ Check the accuracy of weighing instrument periodically. □ Recommended mass, use a mass heavier than 2/3 maximum capacity.
AD-4329 Weighing Indicator
...........
The function for a multi-interval scale. Select "dual
range scale" or "triple range scale". Each range has
parameters of "weighing range" and "division"
The function to revise a linearity deviation using
weighing points, (up to three weighing points)
calibration location and another weighing location using gravity acceleration.
Page 11
5. Calibration
□ Calibrate the scale, if It is moved to another location or the environment
'has changed.
□ It Is not necessary to set the gravity acceleration correction, when calibrating
the scale with a calibration mass at the place where the scale is used.
□ Enter the stable weighing data while the MD mark is turned off. If
unstable data Is used, it may cause a weighing error. Arrange the condition using the FOO filter function.
□ The span calibration needs the zero calibration data. We recommend that
you perform the span calibration Immediately after the zero calibration.
□ Perform the digital linearization function immediately after the zero
calibration. And perform the span calibration continuously.
□ Select a decimal point and weighing unit at the CF function CFOO, CF 01. □ If you use, the dual or triple range function of the multi-interval scale,
perform the “Range Function", “Zero Calibration" and "Span Calibration".

5.2. Calibration Procedure

5.2.1. The Whole Procedure

Essential Items
Step A Preparation for calibration. Step B Minimum division. Step C Maximum capacity. Step B Zero calibration. Step G Span calibration
Optional Items
Step D Weighing range function,
i Perform this function just
before zero calibration and
span calibration.
Step F Digital linearization function.
Perform this function after the zero calibration immediately. And perform the span calibration continuously.
Step H Gravity compensation function.
Perform this function after span calibration.
5. Calibration Page 12
AD-4329 Weighing Indicator

5.2.2. Preparation

step 1 Keep the following conditions to calibrate the scale (indicator) correctly.
□ Maintain a constant temperature, stable power and stable input voltage from the
load ceil. □ Avoid direct sunshine and in the neighborhood of an air conditioner. □ Do not install the scale (indicator) where there is a strong magnetic field.
Step 2 Turn the display on during 10 minutes.

5.2.3. Entering the Calibration Mode

Step 3 Press the ICAlj key In the norma! weighing mode. The
two seconds when entering the calibration mode.
MODE! key.... Parameters are not revised and the next item is displayed.
CALI key........Parameters are stored in the indicator and Snd I is displayed.
Press the jOperate/Standbvl key to turn the display off.
Press and hold the ¡NET/GROSSl key and press the |CAU key
................
....All parameters are not changed
and the calibration mode is. finished. Press the jOperate/Standby key to turn the display off after displaying CBnCEL
CHL (O
is displayed for
Turn the display on
for 10 miniuies
CAL
m ,n
AD-4329 Weighing Indicator
Page 13
Zero
Calibration^
Span Calibration.
^
5. Calibration

5.2.4. Minimum Division (Weighing Interval)

Step 4
Step 5
Check the | d-xx I display. The xx is minimum division (01,02, 05,10, 20,50).
Select a minimum division using the following keys,
A
key... The key to select a parameter, key... The key to store a parameter and
proceed to step 6.
mode! key..The key not to change a parameter
and proceed to step 6.
M-H key ..... The key to proceed to "5.6. Gravity
Compensation Function".
To Gravity
Compesation Function

5.2.5. Maximum Capacity

Step 6 Check the [ ixxxx | display. The xxxx is current Maximum capacity
capacity. Select a new maximum capacity using the foliowing keys.
>
key
.......... The key to select a figure.
A
key..
.........
The key to select a number.
key.......... The key to store a parameter
Minimum division
To Maximum Capacity
and proceed to step 7 (of the zero calibration).
Imod^ key
PHESETTABg key,,
..........
The key not to change a
parameter and proceed to step 7 Punction
(of the zero calibration).
Weighing Range Function".
.....
The key to porceed to "5.3.
To Weighing Range
To Zero Calibration
5. Calibration
Page 14
AD-4329 Weighing Indicator

5.2.6. Zero Calibration

Step 7 Check the
Select a zero calibration method to adjust the zero point
Weighing input
(Normal way)
Digital input
Weighing Input
Step 8 Place nothing on the weighing unit. Press the key
after the MD mark has turned off. Proceed to steplO. MOPg key
Caution Do not press the [ *-■ key while the MD mark is lit
(detecting motion). Arrange the condition using
the FOO filter function.
CfiL 0
.......
display.
The adjustment method with nothing on the weighing unit. The numerical input to
enter a load cell output
voltage.
The key not to change the zero point
data and proceed to step 10.
To step 8
To step 9
Zero calibration
Digital Input
Step 9 Pressing the @ key, a stored input voltage parameter
of the zero point is displayed in the unit of mV/V. Adjust the input voltage using the following keys.
key.....The key to select a figure,
key.....The key to select a number.
key.....The key to store the zero point parameter
and proceed to step 10.
MODE] key..The key not to change parameter and
proceed to step 10.
To Span Calibration
AD-4329 Weighing Indicator
Page 15
5. Calibration
Loading...
+ 39 hidden pages